AL200 Automatic Lensometer User’s Guide ®

AL200 Automatic Lensometer User’s Guide ®
AL200
Automatic Lensometer®
User’s Guide
©2012 AMETEK, Inc.
AMETEK is a registered trademark of AMETEK, Inc.
Reichert, Reichert Technologies and Lensometer are registered trademarks of Reichert, Inc.
All other trademarks are property of their respective owners.
The information contained in this document was accurate at time of publication. Speciications are subject to
change without notice. Reichert Technologies reserves the right to make changes in the product described in this
manual without notice and without incorporating those changes in any products already sold.
ISO 9001/13485 Certiied – Reichert products are designed and manufactured under quality processes meeting
ISO 9001/13485 requirements.
No part of this publication may be reproduced, stored in a retrieval system, or transmitted in any form or by any
means, electronic, mechanical, recording, or otherwise, without the prior written permission of Reichert, Inc.
Caution: Federal law restricts this device to sale by or on the order of a licensed physician. Rx Only.
2
13600-101 Rev. F
Table of Contents
Preliminary Information
Warnings and Cautions
Symbol Information
4
6
Introduction
Congratulations on your Purchase
Indications for use
Contraindications
7
7
7
Installation Features and Functions
Contents
Setup
Application of Input Power
Disconnection of Input Power
Icon Guide
8
8
9
9
11
Instrument Initialization
Setup Mode
12
Operation
Single Vision Auto Mode
Progressive Auto Mode
Bifocal Auto Mode
Single Vision Manual Mode
Progressive Manual Mode
Bifocal Manual Mode
Prism Measurement
Review Mode
13
14
16
18
19
21
22
23
Troubleshooting
Troubleshooting Table
24
Maintenance / Speciications
Cleaning and Maintenance Procedures
AL200 Speciications
25
28
Warranty
AL200 Warranty Terms
29
Notes
13600-101 Rev. F
30
3
Warnings and Cautions
Reichert Technologies is not responsible for the safety and reliability of this instrument when:
• Assembly, disassembly, repair, or modiication is made by unauthorized dealers or persons.
• Instrument is not used in accordance with this User’s Guide.
WARNING: AN INSTRUCTION THAT DRAWS ATTENTION TO RISK OF INJURY OR DEATH.
WARNING: UNITED STATES FEDERAL LAW AND EUROPEAN REGULATIONS REqUIRE ThAT
ThIS DEVICE BE PURChASED ONLy By A PhySICIAN OR A PERSON ACTING ON BEhALF OF A
PhySICIAN.
WARNING: ThIS INSTRUMENT ShOULD BE USED IN STRICT ACCORDANCE WITh ThE
INSTRUCTIONS OUTLINED IN ThIS USER’S GUIDE. ThE SAFETy OF ThE OPERATOR AND ThE
PERFORMANCE OF ThE INSTRUMENT CANNOT BE GUARANTEED IF USED IN A MANNER NOT
SPECIFIED By REIChERT TEChNOLOGIES.
WARNING: DO NOT REPAIR OR SERVICE ThIS INSTRUMENT WIThOUT AUThORIzATION FROM ThE
MANUFACTURER. ANy REPAIR OR SERVICE TO ThIS INSTRUMENT MUST BE PERFORMED By
EXPERIENCED PERSONNEL OR DEALERS WhO ARE TRAINED By REIChERT SO ThAT CORRECT
OPERATION OF ThIS INSTRUMENT IS MAINTAINED OR SERIOUS INJURy TO ThE OPERATOR OR
PATIENT MAy OCCUR.
WARNING: MODIFICATIONS TO ThIS INSTRUMENT IS NOT ALLOWED. ANy MODIFICATION TO ThIS
UNIT MUST BE AUThORIzED By REIChERT SO ThAT CORRECT OPERATION IS MAINTAINED OR
SERIOUS INJURy TO ThE OPERATOR OR PATIENT MAy OCCUR.
WARNING: IF ThIS INSTRUMENT IS MODIFIED, APPROPRIATE INSPECTION AND TESTING MUST
BE CONDUCTED TO ENSURE CONTINUED SAFE USE OF ThIS INSTRUMENT.
WARNING: TO AVOID RISK OF ELECTRIC ShOCK, ThIS EqUIPMENT MUST ONLy BE CONNECTED
TO A SUPPLy MAINS WITh PROTECTIVE EARTh OR DAMAGE TO ThE OCULAR RESPONSE
ANALyzER AND/OR INJURy TO ThE OPERATOR OR PATIENT MAy OCCUR.
WARNING: ENSURE ThAT ThE VOLTAGE APPLIED TO ThE UNIT IS ThE SAME AS ThE VOLTAGE
ThAT IS INDICATED ON ThE DATA PLATE OR DAMAGE TO ThE INSTRUMENT AND/OR INJURy TO
ThE OPERATOR OR PATIENT MAy OCCUR.
WARNING: ThIS INSTRUMENT MUST BE PLUGGED INTO AN OUTLET WITh AN EARTh GROUND.
DO NOT REMOVE OR DEFEAT ThE EARTh GROUND CONNECTION ON POWER INPUT CONNECTOR OR ThE UNIT’S POWER CORD OF ThIS INSTRUMENT OR DAMAGE TO IT AND/OR INJURy TO
ThE OPERATOR OR PATIENT MAy OCCUR.
WARNING: ThE EqUIPMENT OR SySTEM ShOULD NOT BE USED ADJACENT TO OR STACKED
WITh OThER EqUIPMENT AND ThAT IF ADJACENT OR STACKED USE IS NECESSARy, ThE
EqUIPMENT OR SySTEM ShOULD BE OBSERVED TO VERIFy NORMAL OPERATION IN ThE
CONFIGURATION IN WhICh IT WILL BE USED.
WARNING: ThIS INSTRUMENT IS NOT SUITABLE FOR USE IN ThE PRESENCE OF FLAMMABLE
ANESThETIC MIXTURES, SUCh AS OXyGEN OR NITROUS OXIDE.WARNING: ThE USE OF
ACCESSORIES OR CABLES OThER ThAN ThOSE SPECIFIED, WITh ThE EXCEPTION OF ThOSE
SOLD By ThE MANUFACTURER AS REPLACEMENT PARTS FOR INTERNAL COMPONENTS, MAy
RESULT IN INCREASED EMISSIONS OR DECREASED IMMUNITy OF ThE EqUIPMENT OR SySTEM.
4
13600-101 Rev. F
Warnings and Cautions
(Continued)
CAUTION: AN INSTRUCTION THAT DRAWS ATTENTION TO THE RISK OF DAMAGE TO THE
PRODUCT.
CAUTION: ThE INTERNAL CIRCUITRy OF ThE INSTRUMENT CONTAINS ELECTROSTATIC DISChARGE SENSITIVE DEVICES (ESDS) ThAT MAy BE SENSITIVE TO STATIC ChARGES PRODUCED
By ThE hUMAN BODy. DO NOT REMOVE ThE COVERS WIThOUT TAKING PROPER PRECAUTIONS.
CAUTION: ThIS INSTRUMENT IS NOT INTENDED TO BE CONNECTED TO EqUIPMENT OUTSIDE
ThE CONTROL OF REIChERT INC. OR MUST BE TESTED TO AN APPLICABLE IEC OR ISO STANDARDS.
CAUTION: DO NOT USE SOLVENTS OR STRONG CLEANING SOLUTIONS ON ANy PART OF ThIS INSTRUMENT AS DAMAGE TO ThE UNIT MAy OCCUR. SEE MAINTENANCE SECTION FOR DETAILED
CLEANING INSTRUCTION.
CAUTION: USE OF AMMONIA BASED CLEANERS ON ThE LIqUID CRySTAL DISPLAy (LCD) MAy
CAUSE DAMAGE TO DISPLAy. SEE MAINTENANCE SECTION FOR DETAILED CLEANING INSTRUCTION.
CAUTION: MEDICAL ELECTRONIC EqUIPMENT NEEDS SPECIAL PRECAUTIONS REGARDING EMC
AND NEEDS TO BE INSTALLED AND PUT INTO SERVICE ACCORDING TO ThE EMC INFORMATION
PROVIDED IN ThE ACCOMPANyING DOCUMENTS.
CAUTION: PORTABLE AND MOBILE RF COMMUNICATIONS EqUIPMENT CAN AFFECT MEDICAL
ELECTRICAL EqUIPMENT.
CAUTION: ThIS INSTRUMENT IS NOT TO BE USED NEAR hIGh-FREqUENCy EMITTING SURGICAL
EqUIPMENT.
13600-101 Rev. F
5
Symbol Information
The following symbols appear on the instrument:
Caution symbol indicating important operating and maintenance
instructions that are included in this User’s Guide.
Alternating Current Power.
Protective Earth Connection.
ON / OFF.
2012
Date of Manufacture.
REF
Catalog Number.
S/N
Serial Number.
Waste of Electrical and Electronic Equipment.
Compliance to Medical Device Directive 93/42/EEC.
Authorized to mark given by Intertek ETL Semko for conformance
with electrical standards.
Accompanying Documents must be consulted.
Authorized Representative in European Community
Fragile Contents in Shipping Container - handle with care
Keep Dry - Package shall be kept away from rain.
This Way Up - Indicates correct upright position of package
6
13600-101 Rev. F
Introduction
Congratulations on the purchase of your new AL200 Auto Lensometer®.
The AL200 will provide you with fast, accurate and reliable measurements of eyeglass lenses for many
years. The instrument has an innovative microprocessor controlled lens measurement system that reduces
operator error and provides precise, repetitive measurements for single, multifocal, and progressive lenses.
This User’s Guide is designed as a training and reference manual for operation, maintenance, and troubleshooting. We recommend that you read it carefully prior to use and follow the instructions in the guide to
ensure optimum performance of your new instrument. Properly trained eyecare professionals such as ophthalmologists, optometrists, opticians and eye care technicians should operate this instrument. All parts of this ME
system are suitable for use within the patient environment.
Please retain this guide for future reference and to share with other users. Additional copies can be obtained
from your authorized Reichert Technologies dealer or contact our Customer Service department directly at:
Tel:716-686-4500
Fax:716-686-4555
www.reichert.com
Indications for use
A lensometer is indicated for measuring prescription lenses.
Contraindications
None
13600-101 Rev. F
7
Installation Features and Functions
Contents
The items listed below should be included in the AL200 packaging container. If any of these items are missing, please contact Reichert Customer Service.
•
•
•
•
•
•
AL200 Auto Lensometer
Dust Cover (P/N 12621-044)
Spare Nosepiece Cover (P/N 12621-047-001
Flat, open-ended wrench (for nosepiece tip removal) (P/N 13600-050)
Power Cord (WCBL10018 (110 VAC) or WCBL10027 (230 VAC)
Information Packet that contains the following:
Users Guide (13600-101)
quick Reference Guide (13600-104)
Setup
Place the AL200 on a table in an environment that is clean, dry, temperature controlled, and away from direct
sunlight and sources of bright light. Refer to the Speciications section of this manual for operating conditions.
If applicable, connect the printer interface cable from the printer to the RS-232C connector on the rear connection panel of the AL200. Tighten the screws, inger tight, on the RS-232C cable.
Insert the female end of the power cord into the power input receptacle on the rear of the AL200. Attach the
other end of the power cord to a power source of the correct voltage as indicated on the data plate. Turn on
the power switch.
Screen
Contrast
Control
RS232
Power Input
On / Off
Switch
AL200 Rear Panel
8
13600-101 Rev. F
Installation Features and Functions (Continued)
Application of Input Power
WARNING: CARE MUST BE TAKEN TO ARRANGE ThE CABLES FOR ThE ACCESSORIES
SUCh ThAT ThEy DO NOT PRESENT A TRIPPING hAzARD TO ThE EXAMINER OR A DANGER TO ThE PATIENT.
WARNING: POSITION ThIS INSTRUMENT SO ThAT IT IS NOT DIFFICULT TO OPERATE ThE
DISCONNECTION DEVICE (PLUG).
WARNING: DO NOT REMOVE ThE OUTSIDE COVERS OF ThE UNIT OR ATTEMPT TO REPAIR ANy INTERNAL PARTS. REPAIR AND SERVICE OF ThE UNIT MUST BE PERFORMED
By EXPERIENCED PERSONNEL OR DEALERS ThAT ARE TRAINED By REIChERT.
CAUTION: ENSURE ThAT ThE VOLTAGE APPLIED TO ThE UNIT IS ThE SAME AS ThE VOLTAGE ThAT IS INDICATED ON ThE DATA PLATE NEXT TO ThE INPUT CORD RECEPTACLE OR
DAMAGE TO ThE UNIT MAy OCCUR.
CAUTION: FOR CONTINUED PROTECTION AGAINST ThE RISK OF FIRE, ANy REPLACEMENT OF DAMAGED FUSES MUST BE IN ACCORDANCE WITh ThE RATING AS INDICATED
IN ThE SPECIFICATIONS SECTION OF ThIS MANUAL.
1. After the unit is in its secure location, apply the correct input voltage to the instrument using the Power
Cord supplied with the unit.
2. Press down on the “|” located on the ON/OFF Switch. The power inlet is located on the back of the unit
(Refer to page 10, item 8, for its location).
3. If necessary, adjust the screen contrast by turning the knob located on the rear of the AL200.
4. Read the User’s Guide and the quick Reference Card before operating this instrument.
Disconnection of Input Power
1. At any time, the power switch can be set to OFF. The unit does not have a power down sequence. To
terminate operation of this instrument, press the ON / OFF switch to the OFF position (O).
2. If this instrument is intended to be OFF for an extended period of time, it can be disconnected from power
by detaching the power cord from its receptacle.
13600-101 Rev. F
9
Installation Features and Functions (Continued)
LCD Screen (most common screen features shown)
LCD SCREEN
hOLD BUTTON
NOSEPIECE
LENS hOLDER LEVER
LENS MARKER LEVER
TABLE hEIGhT LEVER
LENS TABLE
AL200 Features
10
13600-101 Rev. F
Installation Features and Functions (Continued)
Icon Guide
The AL200 utilizes an Icon-based operating system. Icons located along the bottom
of the screen represent the functions that the buttons beneath them perform. Some
buttons perform different functions, depending on which mode the instrument is in.
Users should become familiar with the AL200 icons in order to understand the functions of each mode. Icons utilized on the AL200 are shown and described below.
Operation Icons:
Setup - Press to enter the Setup mode
Review - Press to enter the Review mode - used for viewing measurement data
Right / Left - Press to switch from the right lens to the left lens
Left / Right - Press to switch from the left lens to the right lens
Print - Press to print measurement results
Clear - Press to clear any data on the screen
Auto / Manual - Press to toggle between automatic and manual modes
Progressive - Press to change to progressive lens measurement mode
Single Vision - Press to change to single vision lens measurement mode
Bifocal - Press to change to bifocal measurement mode
Setup Menu Icons:
Up - Press to move up through menu choices
Down - Press to move down through menu choices
Left - Press to move left through menu choices
Right - Press to move right through menu choices
Enter - Press to conirm menu choices and exit setup mode
13600-101 Rev. F
11
Setup Mode
The AL200 comes conigured with the most commonly used instrument settings.
Default settings can be changed in the setup mode. Once changes are made, all
settings are stored until further changes are made.
Upon initialization the AL200 will display the
Reichert logo and then proceed to the default
measurement screen (single vision, auto
mode). To enter the SETUP mode from any
measurement screen, simply press the button
under the setup icon button on the lower left
side of the screen.
Entering
SETUP Mode
Once in SETUP mode, operators can navigate
and activate menu options using the buttons
under the arrow icons. Current instrument
settings appear highlighted. To change any
setting, simply use the up and down arrow buttons to select the desired menu, then use the
left and right arrow buttons to move between
the choices. Once the desired settings have
been highlighted, press the enter button to
conirm the settings and exit setup mode.
SETUP Screen
SETUP Options
12
SETUP options are as follows:
STEPS
Operators can select from 0.25, 0.12, and 0.01 diopter increments.
CYL
Operators can select from minus, plus, or mixed cylinder mode.
PRISM
Operators can choose to display no prism (off), or prism in
Base (U, D, I, O), or Axis (diopters or millimeters).
ABBE
The Abbe setting should be selected based on the type of optical
material to be measured. Choose from Normal (60), Medium (45)
or Low (30).
TONE
Operators can choose to turn the AL200 tone on or off.
SLEEP
Operators can select from 5 minutes, 10 minutes, or no sleep mode.
LANG
Operators can select from the following display languages:
Eng (English), Fra (French), Esp (Spanish), Deu (German),
Por (Portuguese), and Ita (Italian).
EDLINE
Operators can choose from D Line or E Line measurement scales.
13600-101 Rev. F
Operation
Single Vision Auto Measurement Mode
Start Screen
Upon initialization the AL200 will display the
Reichert logo screen and perform a brief
calibration check. After the calibration is complete, press any button to enter Single Vision
Auto Measurement mode.
NOTE: Do not place any lenses or obstructions in front of the aperture during calibration.
Default
Measurement
Screen
Single Vision
Auto Mode
The AL200’s default measurement mode is
Single Vision, Automatic. This mode is used
for fast, easy measurement of a pair of single
vision spectacle lenses. The instrument is
now ready to measure the right lens, indicated
by the highlighted R on the upper left side of
the screen.
Centering the lens
Place the spectacles on the lens table and
secure the right lens with the lens holder. Using the table base levers and your hand, move
the lens until the large cross is positioned
directly over the small cross in the middle of
the centering target. If the operator inserts
a progressive lens in single vision mode, the
progressive icon may lash to indicate that the
mode should be changed.
Optical Center
When optical center has been located, the
AL200 tone will sound and the measurements
will be locked on the screen. Upon removal
of the lens the AL200 will automatically switch
to the left lens, indicated by the highlighted L
on the upper right side of the screen. Repeat
the measurement process for the left lens. All
measurements will remain on the screen until
the operator chooses to print or clear the data.
13600-101 Rev. F
13
Operation
Progressive Auto Measurement Mode
Default
Measurement
Screen
Single Vision
Auto Mode
Upon initialization the AL200 will display the
Reichert logo screen and perform a brief
calibration check. After the calibration is complete, press any button to continue.
NOTE: Do not place any lenses or obstructions in front of the aperture during calibration.
To switch to Auto Progressive mode, simply
press the lens type button once.
Auto
Progressive
mode
Centering the lens
Distance
Measurement
Once in Auto Progressive mode the AL200
is ready to measure the right lens of a pair of
progressive spectacles, indicated by the highlighted R in the upper left corner of the screen.
To measure the distance segment of the lens,
place the spectacles on the lens table and secure the right lens with the lens holder. Using
the table base levers and your hand, move the
lens until the large cross is positioned directly
over the small cross in the middle of the centering target.
When the lens has been properly positioned,
the AL200 tone will sound and the distance
measurement will be locked on the screen.
An arrow will appear on the screen instructing
the operator to move the lens upward to begin
measuring the progressive zone (add).
NOTE: The design of certain progressive
lenses may prohibit auto distance measurement. In these cases operators must press
the hold button to lock the distance value on
the screen.
14
13600-101 Rev. F
Operation
Progressive Auto Measurement Mode (Continued)
Measuring Add
Power
Arrows Indicate
Path to Maximum
Add Value
Maximum Add
Measured
As the lens is moved through the add zone,
black lines will ill the progressive area on
the screen indicating the amount of add being detected. In addition, the add power (in
diopters) is displayed at the bottom corner of
the screen.
Arrows to the right and left of the add zone
will indicate if the operator is moving outside
the progressive zone. Continue following the
direction of the arrows until the maximum add
power is reached.
NOTE: The black lines may not completely ill
the progressive area.
To lock the add power on the screen the operator simply presses the hold button on the top
of the AL200. The tone will sound and all right
lens measurement values will be locked on the
screen. Upon removal of the lens the AL200
will automatically switch to the left lens, indicated by the highlighted L on the upper right
side of the screen. Repeat the measurement
process for the left lens. All measurements
will remain on the screen until the operator
chooses to print or clear the data.
hold Button
13600-101 Rev. F
15
Operation
Bifocal Auto Measurement Mode
Default
Measurement
Screen
Single Vision
Auto Mode
Upon initialization the AL200 will display the
Reichert logo screen and perform a brief
calibration check. After the calibration is complete, press any button to continue.
NOTE: Do not place any lenses or obstructions in front of the aperture during calibration.
To switch to Auto Bifocal mode, simply press
the lens type button twice.
Auto
Bifocal
mode
Centering the lens
Distance
Measurement
16
Once in Auto Bifocal mode the AL200 is ready
to measure the right lens of a pair of bifocal
spectacles, indicated by the highlighted R in
the upper left corner of the screen.
To measure the distance segment of the lens,
place the spectacles on the lens table and secure the right lens with the lens holder. Using
the table base levers and your hand, move the
lens until the large cross is positioned directly
over the small cross in the middle of the centering target.
When the lens is properly positioned, the
AL200 tone will sound and the distance measurements will be locked on the screen. An
arrow will appear on the screen instructing the
operator to move the lens to measure the add
power.
13600-101 Rev. F
Operation
Bifocal Auto Measurement Mode (Continued)
Measuring Add
Power
13600-101 Rev. F
Move the lens to position the add power
segment over the aperture. The AL200 will
automatically detect the add power and lock
the value on the screen. Upon removal of
the lens the AL200 will automatically switch
to the left lens, indicated by the highlighted L
on the upper right side of the screen. Repeat
the measurement process for the left lens. All
measurements will remain on the screen until
the operator chooses to print or clear the data.
17
Operation
Single Vision Manual Measurement Mode
Default
Measurement
Screen
Single Vision
Auto Mode
Upon initialization the AL200 will display the
Reichert logo screen and perform a brief
calibration check. After the calibration is complete, press any button to continue.
NOTE: Do not place any lenses or obstructions in front of the aperture during calibration.
To switch to Single Vision Manual mode, simply press the Auto / Manual button.
Single Vision
Manual
mode
18
Once in Single Vision Manual mode the AL200
is ready to measure the right lens of a pair of
single vision spectacles, indicated by the highlighted R in the upper left corner of the screen.
If preferred, the left lens can be measured irst
by pressing the Right / Left lens button.
Centering the lens
Place the spectacles on the lens table and
secure the right lens with the lens holder. Using the table base levers and your hand, move
the lens until the large cross is positioned
directly over the small cross in the middle of
the centering target. If the operator inserts
a progressive lens in single vision mode, the
progressive icon may lash to indicate that the
mode should be changed.
Optical Center
Press the hold button when the desired positioning has been reached. The AL200 tone
will sound and measurements will be locked
on the screen. Remove the lens and press
the Right / Left lens button to measure the
next lens. Repeat the measurement process
for the left lens. All measurements will remain
on the screen until the operator chooses to
print or clear the data.
13600-101 Rev. F
Operation
Progressive Manual Measurement Mode
Default
Measurement
Screen
Single Vision
Auto Mode
Upon initialization the AL200 will display the
Reichert logo screen and perform a brief
calibration check. After the calibration is complete, press any button to continue.
NOTE: Do not place any lenses or obstructions in front of the aperture during calibration.
To switch to Progressive mode, simply press
the lens type button once, then press the A/M
button to enter manual mode.
Manual
Progressive
mode
Centering the lens
Distance
Measurement
Once in Progressive Manual mode the AL200
is ready to measure the right lens of a pair of
progressive spectacles, indicated by the highlighted R on the upper left side of the screen.
If preferred, the left lens can be measured irst
by pressing the right / left lens button.
To measure the distance segment of the lens,
place the spectacles on the lens table and
secure the right lens with the lens holder. Using the table base levers and your hand, move
the lens until the large cross is positioned over
the small cross in the middle of the centering
target.
Press the hold button when the desired positioning has been reached. The AL200 tone
will sound and measurements will be locked
on the screen. An arrow will appear on the
screen instructing the operator to move the
lens upward to begin measuring the progressive zone (add).
Continued on Page 17
13600-101 Rev. F
19
Operation
Progressive Manual Measurement Mode (Continued)
Measuring Add
Power
Arrows Indicate
Path to Maximum
Add Value
Maximum Add
Measured
As the lens is moved through the add zone,
black lines will ill the progressive area on
the screen indicating the amount of add being detected. In addition, the add power (in
diopters) is displayed at the bottom corner of
the screen.
Arrows to the right and left of the add zone
will indicate if the operator is moving outside
the progressive zone. Continue following the
direction of the arrows until the maximum add
power is reached.
NOTE: The black lines may not completely ill
the progressive area.
Press the hold button when the desired add
value has been reached. The AL200 tone will
sound and all right lens measurement values
will be locked on the screen. Remove the lens
and press the Right / Left lens button to measure the next lens. Repeat the measurement
process for the left lens. All measurements
will remain on the screen until the operator
chooses to print or clear the data.
hold Button
20
13600-101 Rev. F
Operation
Bifocal Manual Measurement Mode
Default
Measurement
Screen
Single Vision
Auto Mode
Upon initialization the AL200 will display the
Reichert logo screen and perform a brief
calibration check. After the calibration is complete, press any button to continue.
NOTE: Do not place any lenses or obstructions in front of the aperture during calibration.
To switch to Bifocal mode, simply press the
lens type button twice, then press the A/M
button to enter manual mode.
Manual
Bifocal
mode
Centering the lens
Distance
Measurement
13600-101 Rev. F
Once in Manual Bifocal mode the AL200 is
ready to measure the right lens of a pair of
bifocal spectacles, indicated by the highlighted
R in the upper left corner of the screen. If
preferred, the left lens can be measured irst
by pressing the right / left lens button.
To measure the distance segment of the lens,
place the spectacles on the lens table and secure the right lens with the lens holder. Using
the table base levers and your hand, move the
lens until the large cross is positioned directly
over the small cross in the middle of the centering target.
Press the hold button when the desired positioning has been reached. The AL200 tone
will sound and measurements will be locked
on the screen. An arrow will appear on the
screen instructing the operator to move the
lens to measure the add power.
21
Operation
Bifocal Manual Measurement Mode (Continued)
Measuring Add
Power
Move the lens to position the add power segment over the aperture. Press the hold button
when the add value has been measured. The
AL200 tone will sound and all right lens measurement values will be locked on the screen.
Remove the lens and press the Right / Left
lens button to measure the next lens. Repeat
the measurement process for the left lens. All
measurements will remain on the screen until
the operator chooses to print or clear the data.
Prism Measurement
Marked Lens
Prism measurements for a marked lens are obtained by measuring at the indicated
measurement point on the lens.
Unmarked Lens
Prism measurements in an unmarked lens are obtained by:
• First applying a non-permanent ink dot on the lens where the patient’s distance
vision is focused (aligned with the center of the pupil). If an add segment is present,
a second dot may be applied to facilitate the add reading
• Prism (and Sphere, Cylinder, and Axis) measurements are obtained by measuring
at the ink dot on the lens.
NOTE: Measurements for prism are displayed in either the x-y (x, y displacement in
mm), or P-b (prism in diopters) as indicated in the SETUP mode (Prism option).
NOTE: If a lens is not marked with the optical center, an inaccurate prism
measurement may be displayed.
22
13600-101 Rev. F
Operation
Review Mode
The Review mode is used to view the current lens measurement data in all available formats (+/- Cylinder, Spherical Equivalent, Contact Lens conversion, Prism in
Diopters, Prism in Millimeters, Prism in Base/Axis). Complete information for both
lenses is displayed. Measurement formats displayed are described below:
Review Mode
Orig
+/CL
S eq
Shows the lens data in the measured format
Converts the original data to plus or minus cylinder
Converts the original measurement data to contact lens data
Shows the spherical equivalent of the original measurement data
Base Shows prism value in Base format
mm/ax Shows prism value in Millimeters / Axis
d/ax
Shows prism value in Diopters / Axis
Obstruction Viewer
The AL200’s obstruction viewer is a real-time view of the camera image inside the
instrument. See picture below.
The obstruction viewer is particularly helpful in identifying obstructions that could affect the measurements of a lens, such as excessive dirt or markings on the lens.
The obstruction viewer also assists with viewing the rim of a lens while measuring
progressive lenses, as well as viewing the line on a bifocal lens.
The obstruction viewer is always on the screen and there is no need to adjust any
settings.
Obstruction
Viewer
13600-101 Rev. F
23
Troubleshooting
Power
Screen blank
- Instrument may be in sleep mode - push any button.
- LCD contrast may be set incorrectly.
Adjust knob on the back of the AL200.
Instrument will not turn on
- Ensure power cord is properly seated in the power input
receptacle on the back of the AL200. Ensure that the instrument is plugged into a wall outlet of the correct voltage.
- Check the fuses located in the fuse holder on the back of the
AL200. Replace fuses if required. Fuse speciications are
shown in the speciication table on page 24.
Measurement
Data on screen with no lens in place
- There may be dirt inside the aperture.
Refer to page 23 for cleaning instructions.
Unstable / erroneous readings
- Instrument may have been turned on with a lens in place,
causing improper calibration. Remove all lenses from AL200
and reboot the instrument.
- There may be dirt inside the aperture.
Refer to page 23 for cleaning instructions.
Obstruction detected message
- Ensure that there are no physical obstructions in front of the
aperture.
NOTE: if the marking pens are not fully returned to their resting position they may obstruct the aperture.
- There may be dirt inside the aperture.
Refer to page 23 for cleaning instructions.
Lens positioning is abnormally slow
- Dark lenses are more dificult to read and may cause
positioning to be slower than normal.
- There may be dirt inside the aperture.
Refer to page 23 for cleaning instructions.
- The lens being measured is dirty or scratched. Cleaning the
lens may help.
For all other problems or questions please contact Reichert Technical Support.
Contact information is provided on the back page of this manual.
24
13600-101 Rev. F
Cleaning and Maintenance
Perform the following procedures as necessary to maintain proper operation of the AL200.
WARNING: ALWAyS WEAR EyE PROTECTION WhEN PERFORMING CLEANING AND
MAINTENANCE.
CAUTION: DO NOT USE STRONG CLEANING SOLUTIONS OR SOLVENTS ON ANy PART OF ThE
AL200 OR DAMAGE TO ThE INSTRUMENT MAy OCCUR.
LCD Screen Cleaning
Gently clean any dirt or contaminants off of the LCD screen using a lint-free cotton cloth lightly dampened with
a mild cleaning solution that is safe for plastic. Be sure to wipe off any residual solution using a lint-free cotton
cloth.
Housing Cleaning
Gently clean any dirt or contaminants off the housing and the LCD screen using a lint-free cotton cloth lightly
dampened with a mild cleaning solution that is safe for plastic. Be sure to wipe off any residual solution using a
lint-free cotton cloth moistened with clean water.
Pen Cleaning
The AL200 utilizes pre-illed, self contained, felt tip marking pens. Under normal circumstances these pens will
require no maintenance except periodic replacement as required. However, if excess ink gathers on the tip of the
pens, it may be wiped off with a lint-free cotton cloth moistened with clean water.
Nosepiece Cover Cleaning / Removal and Installation
WARNING: ALCOhOL IS FLAMMABLE. KEEP AWAy FROM hEAT AND FLAMES.
Clean the outside of the Nosepiece Cover with a lint-free cotton cloth moistened with alcohol. If the inside surface of the Nosepiece Cover is contaminated, remove the Nosepiece Cover and clean the inside with a lint-free
cotton swab moistened with alcohol. If the Nosepiece Cover is damaged or torn, replace it. Refer to the following
diagrams for nosepiece removal and installation instructions.
Using a small, lat screwdriver, push the
blade under the outer edge of the rubber
nosepiece cover. Grasp the edge with
your ingers and pull the nosepiece cover
out of the groove and off of the metal
nosepiece. Clean or replace the nosepiece
as required.
13600-101 Rev. F
25
Cleaning and Maintenance (Continued)
Nosepiece Cover Cleaning / Removal and Installation (continued)
After the nosepiece is cleaned, or a new nosepiece cover is obtained, invert (lip
inside out) the nosepiece cover for installation onto the AL200 nosepiece.
Position the cleaned or new nosepiece cover over the front of the metal nosepiece
on the AL200.
Push the outer edges of the nosepiece cover over the metal nosepiece. Use your
ingers to seat the rubber nosepiece cover securely into the groove in the metal
nosepiece.
Nosepiece Lens Cleaning
CAUTION: DEBRIS AND CONTAMINATION INSIDE ThE NOSEPIECE MAy CAUSE
MEASUREMENT ERRORS. USE ONLy CLEAN, DRy, OIL-FREE COMPRESSED AIR TO CLEAN
INSIDE ThE NOSEPIECE.
A) Fully retract the lensholder and remove the nosepiece cover
as described on page 22.
B) Using the provided wrench, remove the metal tip of the nosepiece assemly.
C) Using canned air spray a few short bursts of air into the opening.
D) Reinstall the metal nosepiece tip and tighten using the provided wrench. Insure that the tip is tightened so that it can
not be loosened by hand. Install the nosepiece cover when
complete.
* Suggested source of canned air:
EnviroTech Duster 1671 Tech Spray.
26
E) Remove all lenses and obstructions from the AL200. Turn
the instrument off, and then back on to initiate the calibration
cycle. Insure that all sphere, cylinder, and axis data on the
screen reads zero.
13600-101 Rev. F
Cleaning and Maintenance (Continued)
Fuse Replacement
Fuses are located next to the power inlet. Replace fuses with only a rating as indicated in the General Speciications section of this manual.
WARNING: DISCONNECT POWER BEFORE ATTEMPTING TO REMOVE ThE FUSES OR SERIOUS
INJURy OR DEATh MAy OCCUR.
Replace the fuses in the Power Input Module with the fuses indicated in the Speciications section of this manual.
1. Remove input power to the instrument and push up on the tab in the
middle of the Power Input Module to release the Fuse holder. Refer
to Figure .
2. Pull the fuse holder out of the input module.
3. Install new fuses that are indicated in the Speciication section of this
manual into the Fuse holder.
4. Push the Fuse holder into the Power Input Module until it snaps into
place.
NOTE: Replacement of this fuse must be performed by qualiied service
personnel only.
13600-101 Rev. F
27
Speciications:
Physical Dimensions
Size:
Weight, unpacked: 23.0 lbs. (10.4 Kg)
height: 19.8 in. (50.2 cm), Width: 10.5 in. (26.7 cm), Depth: 14.0 in. (35.6 cm)
Electrical
Voltage: 100-240, VAC 50/60 hz
Power: 50 VA, Fuses: Time-Lag, 6.30 mA 250V, 5x20mm, RohS
Measurement Range:
Types of Lenses Measured:
Single Vision, Bifocal/Trifocal, and Progressive Addition
Performance:
Spherical Power ..................................................................-25D to + 25D
Increments ....................................................... 0.01D, 0.12D or 0.25D
Cylindrical Power ................................................................ -10D to + 10D
Increments ....................................................... 0.01D, 0.12D or 0.25D
Axis: .......................................................................................... 0° to 179°
Increments ........................................................................................ 1°
Add: ....................................................................................... 0D to + 10D
Increments ........................................................ 0.01D, 0.12D or 0.25D
Prism: ..................................................................................... 0 D to 10 D
Increments ........................................................ 0.01D, 0.12D or 0.25D
Lens Diameters Accommodated:
Lens Blanks ......................................................................... 28 to 100 mm
Operational Conditions
Environmental:
The environmental conditions are as follows:
Operating:
Temperature 10° C (50° F) to 35° C (95° F)
Relative humidity: 30% to 75%
Atmospheric Pressure: 80 (23.6 in. hg) to 106 kPa (31.3 in. hg)
Transportation & Storage:
Temperature -40° C (-40° F) to +70° C (158° F).
Relative humidity: 10% to 80% (non-condensing)
Atmospheric Pressure: 50 (14.8 in. hg) to 106 kPa (31.3 in. hg)
35°C
10°C
75%
30%
106 kPa
80 kPa
Exposure to extreme temperature conditions indicated above must not exceed 15 weeks.
Disposal
This product does not generate any environmentally hazardous residues. At the end of its product service life,
follow your local laws and ordinances regarding the proper disposal of this equipment.
Software Revision
The software revision can be obtained by contacting Reichert, Inc. The serial number identiies the manufacture date and will provide access to the software version.
28
13600-101 Rev. F
Warranty
This product is warranted by Reichert, Inc. against defective material and workmanship under normal use for a
period of one year from the date of invoice to the original purchaser. (An authorized dealer shall not be considered an original purchaser.) Under this warranty, Reichert’s sole obligation is to repair or replace the defective part
or product at Reichert’s discretion.
This warranty applies to new products and does not apply to a product that has been tampered with, altered in
any way, misused, damaged by accident or negligence, or which has had the serial number removed, altered or
effaced. Nor shall this warranty be extended to a product installed or operated in a manner not in accordance
with the applicable Reichert instruction manual, nor to a product which has been sold, serviced, installed or repaired other than by a Reichert factory, Technical Service Center, or authorized Reichert Dealer.
Lamps, bulbs, charts, cards and other expendable items are not covered by this warranty.
All claims under this warranty must be in writing and directed to the Reichert factory, Technical Service Center,
or authorized instrument dealer making the original sale and must be accompanied by a copy of the purchaser’s
invoice.
This warranty is in lieu of all other warranties implied or expressed. All implied warranties of merchantability or
itness for a particular use are hereby disclaimed. No representative or other person is authorized to make any
other obligations for Reichert. Reichert shall not be liable for any special, incidental, or consequent damages for
any negligence, breach of warranty, strict liability or any other damages resulting from or relating to design, manufacture, sale, use or handling of the product.
PATENT WARRANTy
If notiied promptly in writing of any action brought against the purchaser based on a claim that the instrument infringes a U.S. Patent, Reichert will defend such action at its expense and will pay costs and damages awarded in
any such action, provided that Reichert shall have sole control of the defense of any such action with information
and assistance (at Reichert’s expense) for such defense, and of all negotiation for the settlement and compromise
thereof.
PRODUCT ChANGES
Reichert reserves the right to make changes in design or to make additions to or improvements in its products
without obligation to add such to products previously manufactured.
CLAIMS FOR ShORTAGES
We use extreme care in selection, checking, rechecking and packing to eliminate the possibility of error. If any
shipping errors are discovered:
1.
Carefully go through the packing materials to be sure nothing was
inadvertently overlooked when the unit was unpacked.
2.
Call the dealer you purchased the product from and report the
shortage. The materials are packed at the factory and none
should be missing if the box has never been opened.
3.
Claims must be iled within 30 days of purchase.
CLAIMS FOR DAMAGES IN TRANSIT
Our shipping responsibility ceases with the safe delivery in good condition to the transportation company. Claims
for loss or damage in transit should be made promptly and directly to the transportation company.
If, upon delivery, the outside of the packing case shows evidence of rough handling or damage, the transportation company’s agent should be requested to make a “Received in Bad Order” notation on the delivery receipt. If
within 48 hours of delivery, concealed damage is noted upon unpacking the shipment and no exterior evidence of
rough handling is apparent, the transportation company should be requested to make out a “Bad Order” report.
This procedure is necessary in order for the dealer to maintain the right of recovery from the carrier.
13600-101 Rev. F
29