Invacare IRC 1730 Network Card User Manual

Add to my manuals
36 Pages

advertisement

Invacare IRC 1730 Network Card User Manual | Manualzz
Operator’s Manual
Stratos™Portable Plus
Compressor Nebulizer
System
Model Number: IRC 1730
Dealer: This manual MUST be given to the end user.
User: BEFORE using this product, read this manual and save
for future reference.
For more information regarding
Invacare products, parts, and services,
please visit www.invacare.com
WARNING
DO NOT USE THIS PRODUCT OR ANY AVAILABLE
OPTIONAL EQUIPMENT WITHOUT FIRST
COMPLETELY READING AND UNDERSTANDING
THESE INSTRUCTIONS AND ANY ADDITIONAL
INSTRUCTIONAL MATERIAL SUCH AS OWNER’S
MANUALS, SERVICE MANUALS OR INSTRUCTION
SHEETS SUPPLIED WITH THIS PRODUCT OR
OPTIONAL EQUIPMENT. IF YOU ARE UNABLE TO
UNDERSTAND THE WARNINGS, CAUTIONS OR
INSTRUCTIONS, CONTACT A HEALTHCARE
PROFESSIONAL, DEALER OR TECHNICAL
PERSONNEL BEFORE ATTEMPTING TO USE THIS
EQUIPMENT - OTHERWISE, INJURY OR DAMAGE
MAY OCCUR.
ACCESSORIES WARNING
Invacare products are specifically designed and
manufactured for use in conjunction with Invacare
accessories. Accessories designed by other manufacturers
have not been tested by Invacare and are not
recommended for use with Invacare products.
Stratos™Portable Plus
2
Part No 1150755
TABLE OF CONTENTS
TABLE OF CONTENTS
SPECIAL NOTES ............................................... 5
INTRODUCTION .............................................. 6
REGISTER YOUR PRODUCT .............................. 7
FEATURES ....................................................... 8
TYPICAL PRODUCT PARAMETERS ................... 9
SUPPLEMENTAL INFORMATION ..................... 10
Compressor........................................................................................ 10
Extension Cords................................................................................ 10
HANDLING ................................................... 11
Shipping Instructions......................................................................... 11
Unpacking............................................................................................ 11
Inspection ............................................................................................ 11
SECTION 1—IMPORTANT SAFEGUARDS ......... 12
To Reduce the Risk of Electrocution: .......................................... 13
SECTION 2—OPERATING INSTRUCTIONS ....... 17
Setup..................................................................................................... 17
Treatment Instructions .................................................................... 19
SECTION 3—BATTERIES ................................. 21
Part No 1150755
3
Stratos™Portable Plus
TABLE OF CONTENTS
TABLE OF CONTENTS
Battery Specifications ....................................................................... 21
Conditioning the Battery................................................................. 23
Using the Battery............................................................................... 24
Charging the Battery ........................................................................ 25
SECTION 4—CLEANING AND DISINFECTING ... 26
Cleaning the Nebulizer .................................................................... 26
Cleaning the Mouthpiece................................................................. 28
Changing the Nebulizer ................................................................... 28
Cleaning the Compressor ............................................................... 28
SECTION 5—MAINTENANCE .......................... 30
Routine Care ...................................................................................... 30
Storage................................................................................................. 31
Filter Care........................................................................................... 31
SECTION 6—TROUBLESHOOTING ................... 32
LIMITED WARRANTY .................................... 34
Stratos™Portable Plus
4
Part No 1150755
SPECIAL NOTES
SPECIAL NOTES
Signal words are used in this manual and apply to hazards or unsafe practices which could result in personal injury or property damage. Refer to the table below for definitions of the signal words.
SIGNAL
WORD
MEANING
DANGER
Danger indicates an imminently hazardous situation which, if
not avoided, will result in death or serious injury.
WARNING
Warning indicates a potentially hazardous situation which, if
not avoided, could result in death or serious injury.
CAUTION
Caution indicates a potentially hazardous situation which, if
not avoided, may result in property damage.
NOTICE
The information contained in this document is subject to
change without notice.
Part No 1150755
5
Stratos™Portable Plus
INTRODUCTION
INTRODUCTION
Your Invacare Stratos Portable Plus Compressor Nebulizer System is the latest in portable compressor nebulizer technology. The purpose of the nebulizer system is to convert your liquid medication into a fine mist that can be drawn into your lungs to relieve your respiratory condition. The nebulizer system is very portable and lightweight. For even greater convenience, ask your dealer about the optional battery pack that is sold separately.
To maximize your benefit from the nebulizer system, we recommend you read and understand the information in the operating manual. By following your doctor’s advice and these simple instructions, the nebulizer system will become a valuable part of your daily treatment routine.
Stratos™Portable Plus
6
Part No 1150755
REGISTER YOUR PRODUCT
REGISTER YOUR PRODUCT
The benefits of registering include:
1. Safeguarding your investment.
2. Ensuring long-term maintenance and servicing of your product.
3. Receiving updates with product information, maintenance tips
and industry news.
Register ONLINE at
warranty.invacare.com
Please have your model number and purchase date available to
complete your registration.
Any registration information you submit will only be used by
Invacare Corporation and protected as required by applicable laws
and regulations.
Part No 1150755
7
Stratos™Portable Plus
FEATURES
FEATURES
Compressor
Outlet
On/Off Button
Inlet Air Filter
AC/DC Power
Outlet
Stratos™Portable Plus
8
Part No 1150755
TYPICAL PRODUCT PARAMETERS
TYPICAL PRODUCT PARAMETERS
SOUND LEVEL:
58 dBA
OPERATING
TEMPERATURE RANGE:
50°F - 104°F (10°C to 40°C); 10 to 95% RH
FOAM INTAKE FILTER:
Expanded Polyurethane
SAFETY SYSTEMS:
High temperature compressor shutdown.
REGULATORY LISTING:
In compliance with EN 60601-1
DIMENSIONS
WITHOUT BATTERY:
WITH BATTERY:
5.1 x 3.6 x 1.6 inches (13cm x 9cm x 4cm)
5.1 x 3.6 x 2.6 inches (13cm x 9cm x 7cm)
*WEIGHT
WITHOUT BATTERY:
WITH BATTERY:
1 lbs (450 g)
1.4 lbs (650 g)
POWER SUPPLY:
100-120 VAC 50-60 Hz, 12 VDC, 1.4 A
MAXIMUM PRESSURE:
29 psi (200 kPa)
MAXIMUM FLOW RATE:
10 L/min
STORAGE CONDITIONS:
-13°F to 158°F (-25° to 70°C); 10 to 95%
RH
OPERATING FLOW
RATE:
3.5 L/min @ 14.5 PSI operating pressure
MODE OF OPERATION:
Intermittent Use 30 min ON/30 min OFF
NOTE: *Weight is without nebulizer and tubing.
Part No 1150755
9
Stratos™Portable Plus
SUPPLEMENTAL INFORMATION
SUPPLEMENTAL
INFORMATION
Compressor
If the unit shuts down frequently or fails to start, you may have a low voltage situation. Low voltage can also be suspected when:
• The compressor fails to start or motor does not get up to full power or speed.
• Other motor operated appliances fail to operate properly or too many motor operated appliances are on same circuit.
Extension Cords
Make certain extension cord is in good condition. Make certain extension cord wire is NOT LESS than 18 gauge for 25 feet.
Stratos™Portable Plus
10
Part No 1150755
HANDLING
HANDLING
Shipping Instructions
The shipping container has been designed to assure protection of this device.
If this device is to be reshipped by common carrier, it should be packed in the same carton.
Unpacking
1. Check for any obvious damage to the carton or its contents. If damage is evident, DO NOT use. Notify carrier/Invacare or your local Invacare dealer.
2. Remove all the loose packing from the carton.
3. Carefully remove all the components from the carton.
NOTE: Retain carton and packing materials for use in shipping or storing the unit until use is required.
Inspection
1. Examine exterior for nicks, dents, scratches or other damage. Inspect all components.
Part No 1150755
11
Stratos™Portable Plus
SECTION 1—IMPORTANT SAFEGUARDS
SECTION 1—IMPORTANT
SAFEGUARDS
WARNING
SECTION 1 - IMPORTANT SAFEGUARDS contains
important information for the safe operation and use of
this product.
CAUTION
“Caution: Federal law restricts this device to sale or rental by or on order of a physician, or any other practitioner licensed by the law of the State in which he/she practices to use or order the use of this device.”
When using electrical products, especially when children are present, basic safety precautions should ALWAYS be followed, including the following.
READ ALL INSTRUCTIONS BEFORE USING
Stratos™Portable Plus
12
Part No 1150755
SECTION 1—IMPORTANT SAFEGUARDS
DANGER
Risk of electric shock. DO NOT disassemble. Refer servicing to qualified service personnel.
FIRE RISK ‐ DO NOT use this product in an oxygen enriched environment (i.e – oxygen tent).
To Reduce the Risk of Electrocution:
ALWAYS unplug this product IMMEDIATELY after using.
DO NOT use while bathing.
DO NOT place or store product where it can fall or be pulled into a tub or sink.
DO NOT reach for product that has fallen into water. Unplug IMMEDIATELY.
DO NOT place in or drop into water or other liquid.
DO NOT reach for a product that has fallen into water. Unplug IMMEDIATELY.
Part No 1150755
13
Stratos™Portable Plus
SECTION 1—IMPORTANT SAFEGUARDS
WARNING
To Reduce the Risk of Burns,
Electrocution, Fire and Injury to Persons:
This product should NEVER be left unattended when plugged in.
Close supervision is necessary when this product is used by, on, or near children and physically challenged individuals.
Use this product only for its intended use as described in this manual and only with the prescribed medication. Use this product only under physician’s direction.
NEVER operate this product with a damaged cord or plug, if it is not working properly, if it was dropped or damaged, or if it was dropped into water. Call qualified service personnel for examination and repair.
Keep the nebulizer tubing, system power cord, and unit away from heated or hot surfaces, including space heaters, electric blankets, stoves and similar electrical appliances.
NEVER drop or insert objects into any opening or hose.
NEVER use while sleeping or drowsy.
Stratos™Portable Plus
14
Part No 1150755
SECTION 1—IMPORTANT SAFEGUARDS
WARNING
NEVER block the air openings of the product or place it on a soft surface, such as a bed or couch, where the air openings may be blocked. Keep the air openings free from lint, hair and the like. Blocked openings may cause the unit to shut down. Contact dealer IMMEDIATELY.
The Stratos Portable Plus Compressor Nebulizer System uses the surrounding air to effectively administer the prescribed medication through the nebulizer. The user and administrator should be aware of surroundings so as not to have contaminated air filtered through the compressor nebulizer system. (Examples can be air contaminated with paint or cleaning fumes, weed killer, car exhaust, smoking, etc.) DO NOT use the Stratos Portable Plus Compressor Nebulizer System in these environments. Otherwise, serious bodily injury may occur.
This unit is oil‐less. DO NOT lubricate.
This equipment is NOT suitable for use in the presence of a flammable anesthetic mixture with air or with nitrous oxide.
DO NOT operate the compressor while it is still in the carrying case.
Part No 1150755
15
Stratos™Portable Plus
SECTION 1—IMPORTANT SAFEGUARDS
WARNING
When not administering treatment, keep out of the reach of small children. Some of the parts of the nebulizer system are small enough for children to swallow. The tubing may present a strangulation hazard
NEVER touch the unit with wet hands.
DO NOT pull on the power cord to unplug the unit from an electrical outlet.
DO NOT overload wall outlets.
Pump maintenance MUST be performed by a qualified service technician.
SAVE THESE INSTRUCTIONS
Stratos™Portable Plus
16
Part No 1150755
SECTION 2—OPERATING INSTRUCTIONS
SECTION 2—OPERATING
INSTRUCTIONS
WARNING
Remove the compressor from the carrying case before
use.
DO NOT operate unit without filter.
Avoid operating this product in a dusty environment.
Otherwise, premature compressor wear could occur.
Setup
NOTE: For this procedure, refer to FIGURE 2.1 on page 18.
1. Wash your hands before you begin the treatment procedure.
2. Place the device on a table so that the controls can be reached when seated.
3. Open the nebulizer by rotating the top clockwise while securely holding the bottom. 4. Make sure the cone is correctly positioned in the cup. Part No 1150755
17
Stratos™Portable Plus
SECTION 2—OPERATING INSTRUCTIONS
WARNING
DO NOT exceed the maximum allowed quantity of six ml
of medication. The maximum capacity of the nebulizer
medication cup is 6 ml.
CAUTION
NEVER pour medication into the compressor outlet.
Damage to the compressor will occur.
5. Add the prescribed amount of medication to the cup of the nebulizer using an eyedropper, pre‐measured dose or ampoule (FIGURE 2.1).
Eyedropper
Nebulizer
Cup
FIGURE 2.1 Adding
Medication
7. Attach one end of the tubing onto the compressor outlet and connect the other end to the nebulizer.
8. In high humidity conditions, moisture or condensation can occur in the tubing. If this happens, simply turn on the compressor with the tubing attached and run for 2 ‐ 3 minutes or hang the tubing so the moisture can drain out.
9. Connect the mouthpiece to the top section of the nebulizer (if required).
6. Close the nebulizer.
Stratos™Portable Plus
18
Part No 1150755
SECTION 2—OPERATING INSTRUCTIONS
10. To use with battery, charge for 6 hours. Refer to Conditioning the Battery on page 23.
11. To use the nebulizer system with the AC power cord, connect the power cord plug into the power outlet on the side of the nebulizer compressor system. Connect the power cord to the AC adapter and insert the plug from the AC adapter into the electrical outlet.
12. To use the nebulizer system with the DC auto adapter, connect the DC auto adapter to the power outlet on the side of the nebulizer compressor system. Connect the auto adapter plug of the DC cord to the auto adapter receptacle.
DANGER
To avoid the risk of possible electrical shock, the nebulizer
system MUST be operated on the specified power source.
Treatment Instructions
NOTE: For this procedure, refer to FIGURE 2.2 on page 20.
NOTE: The Invacare Stratos Portable Plus Nebulizer System has been designed for intermittent use. Normal treatment time should be five to ten minutes. Turn off the nebulizer system after 30 minutes of use and allow the device to cool down 30 minutes before turning it on again.
Part No 1150755
19
Stratos™Portable Plus
SECTION 2—OPERATING INSTRUCTIONS
1. Sit in a relaxed and upright position. 2. Turn on the device by pushing the On/Off button. 3. Begin treatment by placing the nebulizer mouthpiece between your teeth and close your mouth. Inhale deeply and slowly as the aerosol begins to flow and then exhale slowly.
FIGURE 2.2 Using the
Nebulizer
4. When treatment is completed, turn off the nebulizer system and unplug the power adapter.
5. Remove the mouthpiece from the nebulizer.
6. Remove the nebulizer and follow the cleaning instructions. Refer to Cleaning and Disinfecting on page 26.
Stratos™Portable Plus
20
Part No 1150755
SECTION 3—BATTERIES
SECTION 3—BATTERIES
Battery Specifications
Battery type
Li-ion
Supply voltage
12 VDC (use only original adapter)
Complete charge
6 hours
Capacity
1800 mAh
Autonomy
Up to 90 minutes at full charge
The optional battery used with the Invacare Stratos Portable Plus Nebulizer System requires a few simple precautions to ensure proper operation and to avoid damage to the nebulizer system or to the battery. When used correctly, the battery can operate the nebulizer system for approximately 90 minutes. Battery operation may vary due to charge conditions and care.
WARNING
Use only the original batteries supplied by Invacare.
Carefully follow the instructions for operating the unit.
Any other kind of use constitutes an improper use and is
therefore dangerous. Invacare cannot be held liable for
any damage caused by improper, incorrect or
unreasonable use.
Part No 1150755
21
Stratos™Portable Plus
SECTION 3—BATTERIES
CAUTION
In case of failure and/or malfunction, turn the unit off and
read the operating manual. DO NOT attempt to open the
plastic case or to replace any components.
DO NOT short circuit battery contacts. Make sure
battery contacts DO NOT come in contact with keys,
coins or other metal objects.
The battery has protection from overheating and will not
operate if temperature is 104ºF (40ºC) or higher. Allow
the battery to cool before restarting.
WARNING
To charge battery, use only the original power cable (AC)
or auto adapter cable (DC). Using non-original equipment
may result in damage to the battery.
Charging of the battery requires at least 6 hours. Failure
to condition the battery prior to its first use may result in
a reduced battery life expectancy.
DO NOT dispose of the battery pack in the household
waste. Proper disposal at an appropriate collection site
provided in your local area is strongly recommended.
DO NOT dispose of batteries in fire, as they could explode.
Stratos™Portable Plus
22
Part No 1150755
SECTION 3—BATTERIES
Conditioning the Battery
NOTE: The definition of conditioning the battery is simply charging the battery for a full six hour cycle before the initial use of the product.
1. Disconnect the battery from the compressor.
2. Connect the power cable (AC) or auto adapter cable (DC) to the battery.
3. Charge the battery for a full six hour cycle.
The orange light will indicate if the battery is properly charging. Battery charging to full capacity is dependent upon many factors. Charging of the battery requires at least 6 hours. Failure to condition the battery prior to its first use may result in a reduced battery life expectancy. To assure maximum life performance of the battery after conditioning, leave the battery connected to a power source when not in use. The battery operation time depends upon the charge condition of the battery, the temperature, the storage time of the battery prior to use, the amount of medication used for treatment, and the number of charge/discharge cycles. The manufacturer of the battery recommends the battery be conditioned every one or two months.
Part No 1150755
23
Stratos™Portable Plus
SECTION 3—BATTERIES
Using the Battery
NOTE: For this procedure, refer to FIGURE 3.1 on page 24.
1. Fit the front hook on the battery top into the recess on the bottom of the compressor.
2. Press the battery and compressor together until the release button on the front of the battery clicks.
3. Gently pull the compressor and battery pack in opposite directions to ensure the tabs have seated properly.
4. To remove the battery Recess
Compressor
pack, press the release (hidden
button on the front of the from
view)
battery pack while pulling the battery and compressor apart.
NOTE: The GREEN light on the battery will not illuminate Hook
while the nebulizer is being Battery
powered. The ORANGE light flashes when the battery power Release
level is low. When the Connect Cable HERE
Button
ORANGE light flashes, FIGURE 3.1 Using the
immediately find an alternative Battery
power source (i.e. a wall socket).
Stratos™Portable Plus
24
Part No 1150755
SECTION 3—BATTERIES
Charging the Battery
NOTE: For this procedure, refer to FIGURE 3.1 on page 24.
1. Press in the release button on the front of the battery pack and separate the battery from the compressor. 2. Connect the power cable (AC) to the AC adapter and cable provided with your STRATOS Portable Plus Compressor.
3. Connect the AC adapter and cable to the battery.
4. Plug the power cable into the appropriate power source.
The ORANGE light will turn on indicating that the charging cycle has begun. The battery will reach full charge after at least six hours. Completion of the charge cycle is indicated by a GREEN battery light. Full charge will provide up to ninety minutes of operation.
NOTE: In order to ensure maximum performance of the battery, keep the battery plugged in and charging when not in use. This will not overcharge or damage the battery.
NOTE: The battery must be plugged into a power source for the battery to charge. The battery will not charge when it is attached to the compressor nebulizer system.
NOTE: Never leave the battery connected to either power cable when the cable is disconnected from the power source. This will cause the battery to discharge.
Part No 1150755
25
Stratos™Portable Plus
SECTION 4—CLEANING AND DISINFECTING
SECTION 4—CLEANING AND
DISINFECTING
Following the cleaning instructions after each use will prevent any remaining medication in the bottle from drying and causing the device to nebulize ineffectively.
Cleaning the Nebulizer
WARNING
To prevent possible risk of infection due to
contamination, follow these instructions:
• Clean the nebulizer before initial operations.
• Clean the nebulizer after each aerosol treatment.
• ALWAYS use fresh cleaning solution.
• DO NOT towel dry the nebulizer components after
cleaning.
NOTE: Your physician and/or an Invacare dealer may specify certain cleaning procedures. If so, follow their recommendations. If not, clean according to the following instructions:
1. Remove the nebulizer and mouthpiece from the tubing.
Stratos™Portable Plus
26
Part No 1150755
SECTION 4—CLEANING AND DISINFECTING
2. Disassemble nebulizer chamber by turning counterclockwise and separating.
3. Fill two plastic containers or bowls:
A. Washing: Hot water/detergent solution.
B. Soaking: Hot water/vinegar solution (one part vinegar to three parts water).
4. Thoroughly clean the nebulizer and mouthpiece in hot water/detergent solution.
5. Remove from the water/detergent solution and rinse with clear hot tap water.
6. Soak in hot water/vinegar solution for 30 minutes.
7. Rinse with hot tap water again and air dry thoroughly.
NOTE: If using medical disinfectant cleaners, follow manufacturer’s instructions carefully.
8. Keep the outer surface of the tubing dust‐free by wiping regularly.
The nebulizer air tubing does not have to be washed internally because only filtered air passes through the tubing.
Part No 1150755
27
Stratos™Portable Plus
SECTION 4—CLEANING AND DISINFECTING
Cleaning the Mouthpiece
1. Boil the mouthpiece in water for a maximum of five minutes or autoclave at a maximum of 121° C for ten minutes.
2. Allow the mouthpiece to air dry.
3. Once dry, the mouthpiece may be stored inside the nebulizer top.
Changing the Nebulizer
The nebulizer is designed for single patient use.
The nebulizer MUST be replaced after a long period of inactivity, or when it is obstructed by dry medication, dust, etc. The reusable nebulizer should last from six months to one year with general use. The disposable nebulizer should last 30 days.
Cleaning the Compressor
DANGER
DO NOT remove outer case of this unit. Removing the
case creates a serious electrical shock hazard. All
disassembly and maintenance of this unit MUST be done
by a qualified technician.
Stratos™Portable Plus
28
Part No 1150755
SECTION 4—CLEANING AND DISINFECTING
WARNING
Unplug unit before cleaning.
DO NOT clean using abrasive cleaners.
DO NOT submerge in water to clean. Clean outer case
with damp cloth.
CAUTION
If fluid gets into the area around the compressor outlet,
wipe the area off with a dry cloth and let the nebulizer sit
on its front end for five minutes. Then run the compressor
for two minutes with no nebulizer or tubing adapter in
place.
1. Make sure the compressor is off and is not plugged into any power source.
2. Clean the compressor daily by using a soft cloth moistened with water or a mild detergent.
Part No 1150755
29
Stratos™Portable Plus
SECTION 5—MAINTENANCE
SECTION 5—MAINTENANCE
DANGER
DO NOT remove the compressor case. Failure to comply
could result in electrical shock. All disassembly MUST be
performed by an authorized Invacare service center.
WARNING
All maintenance on your Invacare Stratos Portable Plus
Nebulizer system MUST be performed by an authorized
Invacare service center. Maintenance manuals and parts
lists are available through Invacare upon request.
Routine Care
To keep your device in the best condition and protect the unit from damage, follow these directions:
• DO NOT subject the device or the parts to any strong shocks such as dropping the device on the floor.
• DO NOT store the device in extreme hot or cold temperatures, high humidity or under direct sunlight.
Stratos™Portable Plus
30
Part No 1150755
SECTION 5—MAINTENANCE
Storage
1. Put the nebulizer and mouthpiece in the carrying case.
2. Coil the air tubing.
3. Wrap the AC cord in a coil.
4. Zip the bag closed.
5. Store the device in a safe area.
NOTE: Store the repackaged device in a dry area. DO NOT place other objects on top of the repackaged device.
Filter Care
The accessible filter MUST be replaced after one month of use or when it turns GREY. The filter should be replaced if the nebulizer has been in storage for three months or longer.
CAUTION
Use only the filters provided with the Invacare Stratos
Portable Plus Nebulizer device.
DO NOT use cotton or any other material.
DO NOT wash or clean the filter.
Use only Invacare filters.
DO NOT operate without a air filter.
Part No 1150755
31
Stratos™Portable Plus
SECTION 6—TROUBLESHOOTING
SECTION 6—TROUBLESHOOTING
PROBLEM
SOLUTION
Turn the unit OFF for a minimum of 40
Nebulizer overheats. The
minutes to allow the nebulizer to cool.
nebulizer has an internal
thermal protector to
prevent the device from
overheating. If the internal
temperature reaches the
preset level, the unit will
automatically shut down, and
the power LED will turn
RED.
The nebulizer does not
operate.
Check all power sources to the nebulizer.
Push the ON/OFF button again.
Medication does not nebulize properly.
Check for proper connection of the tubing
adapter.
Make sure the nebulizer is in an upright position
and not tilted beyond 30º.
Make sure the nebulizer is not empty.
Clean the nebulizer as described in the manual.
Make sure the cone has been properly placed in
the nebulizer cup.
Replace the nebulizer.
Check the intake filter and replace it if it is dirty.
Nebulizer delivers no mist.
Make sure the cone is in the cup and properly
seated.
Stratos™Portable Plus
32
Part No 1150755
SECTION 6—TROUBLESHOOTING
PROBLEM
SOLUTION
Nebulizer does not work,
The thermal protector has been activated due
and the light indicator is red. to:
a. The device operating outside the conditions
outlined in this manual.
b. The device has been operating near heat
sources or in hot environments.
Allow the nebulizer to cool down for a
minimum of 40 minutes. Wait until the RED
light turns off before turning the nebulizer on
again.
Optional battery is not
working optimally.
Battery was not properly conditioned (charged
for at least 6 hours before first use).
If it has been in use for approximately four
months, it may be time to recondition the
battery.
Optional battery is not
charging.
Battery will not charge if connected to the
nebulizer.
Check all power cord connections and cord
conditions.
Check power outlet.
Part No 1150755
33
Stratos™Portable Plus
LIMITED WARRANTY
LIMITED WARRANTY
PLEASE NOTE: THE WARRANTY BELOW HAS BEEN
DRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLE
TO PRODUCTS MANUFACTURED AFTER JULY 4, 1975.
This warranty is extended only to the original purchaser who
purchases this product when new and unused from Invacare
Corporation or a dealer. This warranty is not extended to any
other person or entity and is not transferable or assignable to
any subsequent purchaser or owner. Coverage under this
warranty will end upon any such subsequent sale or other
transfer of title to any other person. This warranty gives you
specific legal rights and you may also have other legal rights
which vary from state to state.
Invacare warrants this product when purchased new and unused
to be free from defects in materials and workmanship for a
period of three years from date of purchase from Invacare or a
dealer, except that for batteries, the warranty period is six
months; with a copy of the seller’s invoice required for coverage
under this warranty. If within such warranty periods any such
product shall be proven to Invacare’s satisfaction to be defective,
such product shall be repaired or replaced, at Invacare’s option.
This warranty does not apply to normal wear and tear or failure
to adhere to the product instructions, nor does it include
shipping charges incurred in replacement part installation or
repair of any such product. Invacare’s sole obligation and your
exclusive remedy under this warranty shall be limited to such
repair and/or replacement.
Stratos™Portable Plus
34
Part No 1150755
LIMITED WARRANTY
For warranty service, please contact the Invacare service
department during normal business hours at the toll-free
telephone number identified on the back cover of this manual.
Upon receiving notice of an alleged defect in a product, Invacare
will issue a serialized return authorization. It shall be the
responsibility of the purchaser to pack the product(s) or part(s)
in an appropriate manner to avoid shipping damage and return
the product(s) or part(s), at the purchaser’s expense, to either
Invacare’s plant or service center as specified by Invacare in
advance. The defective product(s) or component part(s) must
be returned for warranty inspection using the serial number as
identification within 30 days of the return authorization date.
DO NOT return any product(s) or part(s) to our factory or
service center without prior consent. C.O.D. shipments will be
refused – please pre-pay shipping charges.
LIMITATIONS AND EXCLUSIONS: THE FOREGOING
WARRANTY SHALL NOT APPLY TO SERIAL NUMBERED
PRODUCTS IF THE SERIAL NUMBER HAS BEEN REMOVED
OR DEFACED; PRODUCTS SUBJECTED TO NEGLIGENCE,
ACCIDENT, IMPROPER OPERATION, MAINTENANCE, OR
STORAGE; PRODUCTS MODIFIED WITHOUT INVACARE’S
EXPRESS WRITTEN CONSENT (INCLUDING, BUT NOT
LIMITED TO, MODIFICATION THROUGH THE USE OF
UNAUTHORIZED PARTS OR ATTACHMENTS); PRODUCTS
DAMAGED BY REASON OF REPAIRS MADE TO ANY
COMPONENT WITHOUT THE SPECIFIC CONSENT OF
INVACARE; PRODUCTS DAMAGED BY CIRCUMSTANCES
BEYOND INVACARE’S CONTROL; OR PRODUCTS
REPAIRED BY ANYONE OTHER THAN INVACARE OR AN
INVACARE SERVICE CENTER. SUCH EVALUATION SHALL
BE SOLELY DETERMINED BY INVACARE.
Part No 1150755
35
Stratos™Portable Plus
THE FOREGOING EXPRESS WARRANTY IS EXCLUSIVE AND
IN LIEU OF ANY OTHER WARRANTIES WHATSOEVER,
WHETHER EXPRESSED OR IMPLIED, INCLUDING THE
IMPLIED WARRANTIES OF MERCHANTABILITY AND
FITNESS FOR A PARTICULAR PURPOSE, AND THE SOLE
REMEDY FOR VIOLATIONS OF ANY WARRANTY
WHATSOEVER, SHALL BE LIMITED TO REPAIR OR
REPLACEMENT OF THE DEFECTIVE PRODUCT PURSUANT
TO THE TERMS CONTAINED HEREIN. THE APPLICATION
OF ANY IMPLIED WARRANTY WHATSOEVER SHALL NOT
EXTEND BEYOND THE DURATION OF THE EXPRESS
WARRANTY PROVIDED HEREIN. INVACARE SHALL NOT BE
LIABLE FOR ANY CONSEQUENTIAL OR INCIDENTAL
DAMAGES WHATSOEVER.
SOME STATES DO NOT ALLOW THE EXCLUSION OR
LIMITATION OF INCIDENTAL OR CONSEQUENTIAL
DAMAGE, OR LIMITATION OF HOW LONG AN IMPLIED
WARRANTY LASTS, SO THE ABOVE EXCLUSION AND
LIMITATION MAY NOT BE APPLICABLE.
THIS WARRANTY SHALL BE EXTENDED TO COMPLY
WITH STATE/PROVINCIAL LAWS AND REQUIREMENTS.
Invacare Corporation
www.invacare.com
USA
One Invacare Way
Elyria, Ohio USA
44036-2125
800-333-6900
All rights reserved. Trademarks are
identified by the symbols ™ and ®. All
trademarks are owned by or licensed
to Invacare Corporation unless
otherwise noted.
©2008 Invacare Corporation
In Florida
2101 E. Lake Mary Blvd.
Sanford, Florida
32773
800-832-4707
Part No 1150755
Rev A - 1/08

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Related manuals

Download PDF

advertisement