Teledyne 3000TA Oxygen Equipment Operating instructions


Add to my manuals
89 Pages

advertisement

Teledyne 3000TA Oxygen Equipment Operating instructions | Manualzz

OPERATING INSTRUCTIONS FOR

MODEL 3000TA

Trace Oxygen Analyzer

P/N M66316

3/09/11

DANGER

Toxic gases and or flammable liquids may be present in this monitoring system.

Personal protective equipment may be required when servicing this instrument.

Hazardous voltages exist on certain components internally which may persist for a time even after the power is turned off and disconnected.

Only authorized personnel should conduct maintenance and/or servicing.

Before conducting any maintenance or servicing, consult with authorized supervisor/manager.

3000TA-

Copyright © 2011 Teledyne Analytical Instruments

All Rights Reserved. No part of this manual may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into any other language or computer language in whole or in part, in any form or by any means, whether it be electronic, mechanical, magnetic, optical, manual, or otherwise, without the prior written consent of Teledyne

Analytical Instruments, 16830 Chestnut Street, City of Industry, CA 91749-1580.

Warranty

This equipment is sold subject to the mutual agreement that it is warranted by us free from defects of material and of construction, and that our liability shall be limited to replacing or repairing at our factory (without charge, except for transportation), or at customer plant at our option, any material or construction in which defects become apparent within one year from the date of shipment, except in cases where quotations or acknowledgements provide for a shorter period. Components manufactured by others bear the warranty of their manufacturer. This warranty does not cover defects caused by wear, accident, misuse, neglect or repairs other than those performed by Teledyne or an authorized service center. We assume no liability for direct or indirect damages of any kind and the purchaser by the acceptance of the equipment will assume all liability for any damage which may result from its use or misuse.

We reserve the right to employ any suitable material in the manufacture of our apparatus, and to make any alterations in the dimensions, shape or weight of any parts, in so far as such alterations do not adversely affect our warranty.

Important Notice

This instrument provides measurement readings to its user, and serves as a tool by which valuable data can be gathered. The information provided by the instrument may assist the user in eliminating potential hazards caused by his process; however, it is essential that all personnel involved in the use of the instrument or its interface, with the process being measured, be properly trained in the process itself, as well as all instrumentation related to it.

The safety of personnel is ultimately the responsibility of those who control process conditions. While this instrument may be able to provide early warning of imminent danger, it has no control over process conditions, and it can be misused. In particular, any alarm or control systems installed must be tested and understood, both as to how they operate and as to how they can be defeated. Any safeguards required such as locks, labels, or redundancy, must be provided by the user or specifically requested of Teledyne at the time the order is placed.

Therefore, the purchaser must be aware of the hazardous process conditions. The purchaser is responsible for the training of personnel, for providing hazard warning methods and instrumentation per the appropriate standards, and for ensuring that hazard warning devices and instrumentation are maintained and operated properly.

Teledyne Analytical Instruments, the manufacturer of this instrument, cannot accept responsibility for conditions beyond its knowledge and control. No statement expressed or implied by this document or any information disseminated by the manufacturer or its agents, is to be construed as a warranty of adequate safety control under the user’s process conditions.

Trace Oxygen Analyzer

Specific Model Information

The instrument for which this manual was supplied may incorporate one or more options not supplied in the standard instrument.

Commonly available options are listed below, with check boxes. Any that are incorporated in the instrument for which this manual is supplied are indicated by a check mark in the box.

Instrument Serial Number: _______________________

Options Included in the Instrument with the Above Serial Number:

3000TA-C: In addition to all standard features, this model also has separate ports for zero and span gases, and built-in control valves. The internal valves are entirely under the control of the 3000TA electronics to automatically switch between gases in synchronization with the analyzer’s operations

19" Rack Mnt: The 19" Relay Rack Mount units are available with either one or two 3000 series analyzers installed in a standard 19" panel and ready to mount in a standard instrument rack.

Sensor Options Available for the Instrument with the Above Serial

Number:

 Insta-Trace B2C (Default)

 A2C

 L2C

 L2CL

 Insta-Trace A2C

3000TA-

Important Notice

Model 3000TA-EU complies with all of the requirements of the

Commonwealth of Europe (CE) for Radio Frequency Interference,

Electromagnetic Interference (RFI/EMI), and Low Voltage Directive

(LVD).

The following International Symbols are used throughout the

Instruction Manual. These symbols are visual indicators of important and immediate warnings and when you must exercise CAUTION while operating the instrument. See also the Safety Information on the next page.

STAND-BY: Instrument is on Stand-by, but circuit is active

GROUND:

Protective Earth

CAUTION: The operator needs to refer to the manual for further information. Failure to do so may compromise the safe operation of the equipment.

CAUTION: Risk of Electrical Shock

Trace Oxygen Analyzer

Safety Messages

Your safety and the safety of others is very important. We have provided many important safety messages in this manual. Please read these messages carefully.

A safety message alerts you to potential hazards that could hurt you or others. Each safety message is associated with a safety alert symbol.

These symbols are found in the manual and inside the instrument. The definition of these symbols is described below:

GENERAL WARNING/CAUTION : Refer to the instructions for details on the specific danger. These cautions warn of specific procedures which if not followed could cause bodily Injury and/or damage the instrument.

CAUTION:

H

OT

S

URFACE

W

ARNING

: This warning is specific to heated components within the instrument. Failure to heed the warning could result in serious burns to skin and underlying tissue.

WARNING:

E

LECTRICAL

S

HOCK

H

AZARD

: Dangerous voltages appear within this instrument. This warning is specific to an electrical hazard existing at or nearby the component or procedure under discussion. Failure to heed this warning could result in injury and/or death from electrocution.

No

Symbol

Technician Symbol: All operations marked with this symbol are to be performed by qualified maintenance personnel only.

NOTE: Additional information and comments regarding a specific component or procedure are highlighted in the form of a note.

3000TA-

CAUTION: THE ANALYZER SHOULD ONLY BE USED FOR THE

PURPOSE AND IN THE MANNER DESCRIBED IN

THIS MANUAL.

IF YOU USE THE ANALYZER IN A MANNER OTHER

THAN THAT FOR WHICH IT WAS INTENDED,

UNPREDICTABLE BEHAVIOR COULD RESULT

POSSIBLY ACCOMPANIED WITH HAZARDOUS

CONSEQUENCES.

This manual provides information designed to guide you through the installation, calibration and operation of your new analyzer. Please read this manual and keep it available.

Occasionally, some instruments are customized for a particular application or features and/or options added per customer requests.

Please check the front of this manual for any additional information in the form of an Addendum which discusses specific information, procedures, cautions and warnings that may be peculiar to your instrument.

Manuals do get lost. Additional manuals can be obtained from

Teledyne at the address given in the Appendix. Some of our manuals are available in electronic form via the internet. Please visit our website at: www.teledyne-ai.com.

Trace Oxygen Analyzer

This is a general purpose instrument designed for use in a nonhazardous area. It is the customer's responsibility to ensure safety especially when combustible gases are being analyzed since the potential of gas leaks always exist.

The customer should ensure that the principles of operation of this equipment are well understood by the user. Misuse of this product in any manner, tampering with its components, or unauthorized substitution of any component may adversely affect the safety of this instrument.

Since the use of this instrument is beyond the control of Teledyne, no responsibility by Teledyne, its affiliates, and agents for damage or injury from misuse or neglect of this equipment is implied or assumed.

3000TA-

Table of Contents

Safety Messages ........................................................................... v  

Introduction ................................................................................... 1  

1.1 Overview 1  

1.2 Typical Applications

1.3 Main Features of the Analyzer

1

1

 

 

1.4 Model Designations

1.5 Front Panel (Operator Interface)

2

 

3

 

1.6 Recognizing Difference Between LCD & VFD

1.7 Rear Panel (Equipment Interface)

4  

5

 

Operational Theory ....................................................................... 7  

2.1 Introduction 7

 

2.2 Micro-Fuel Cell Sensor

2.2.1 Principles of Operation

7

7

 

 

2.2.2 Anatomy of a Micro-Fuel Cell

2.2.3 Electrochemical Reactions

8

 

9

 

2.2.4 The Effect of Pressure

2.2.5 Calibration Characteristics

2.3 Sample System

2.4 Electronics and Signal Processing

10

 

10

 

11

 

13

 

Installation ................................................................................... 17  

3.1 Unpacking the Analyzer 17

 

3.2 Mounting the Analyzer 17

 

3.3 Rear Panel Connections

3.3.1 Gas Connections

3.3.2 Electrical Connections

3.3.2.1 Primary Input Power

3.3.2.2 50-Pin Equipment Interface Connector

3.3.2.3

  RS-232 Port 27  

19  

19

 

21

 

21

 

22

 

Trace Oxygen Analyzer

3.4 Installing the Micro-Fuel Cell

3.5 Testing the System

29

 

29  

Operation ..................................................................................... 31  

4.1 Introduction 31

 

4.2 Using the Data Entry and Function Buttons

4.3 The System Function

4.3.1 Tracking Oxygen Readings During Calibration and

Alarm Delay

4.3.2 Setting up an Auto-Cal

4.3.3 Password Protection

4.3.3.1 Entering the Password

31

 

33

 

34

 

36

 

37

 

37

 

4.3.3.2 Installing or Changing the Password

4.3.4 Logout

4.3.5 System Self-Diagnostic Test

4.3.6 Version Screen

4.3.7 Showing Negative Oxygen Readings

4.4 The Zero and Span Functions

4.4.1 Zero Cal

4.4.1.1

 

Auto Mode Zeroing

4.4.1.2 Manual Mode Zeroing

4.4.1.3

 

Cell Failure

43  

44

 

44

 

45

 

45

 

38  

40

 

41

 

42

 

42

 

4.4.2 Span Cal

4.4.2.1 Auto Mode Spanning

4.4.2.2 Manual Mode Spanning

4.4.3 Span Failure

4.5 The Alarms Function

4.6 The Range Function

4.6.1 Setting the Analog Output Ranges

4.6.2 Fixed Range Analysis

4.7 The Analyze Function

4.8 Signal Output

46  

46

 

47

 

49

 

49

 

51

 

52

 

53

 

54

 

54

 

Maintenance ................................................................................. 57  

5.1 Routine Maintenance 57

 

5.2.1 Storing and Handling Replacement Cells 57

 

3000TA-

5.2.2 When to Replace a Cell

5.2.3 Removing the Micro-Fuel Cell

5.2.4 Installing a New Micro-Fuel Cell

5.2.5 Cell Warranty

5.3 Fuse Replacement

5.4 System Self Diagnostic Test

5.5 Major Internal Components

5.6 Cleaning

5.7 Troubleshooting

Appendix ...................................................................................... 67  

A-1 Model 3000TA Specifications 67

 

A-2 Recommended 2-Year Spare Parts List

A-3 Drawing List

A-4 19-inch Relay Rack Panel Mount

68

70

70

 

 

 

A.5 Application notes

A-5 Material Safety Data Sheet

71

 

75

 

58

 

58  

59

 

61

 

61

 

62

 

63  

64

 

64

 

Trace Oxygen Analyzer

List of Figures

Figure 1-1: Model 3000TA Front Panel ........................................... 3

 

Figure 1-2: Model 3000 TA Rear Panel ........................................... 5

 

Figure 2-1: Micro-Fuel Cell .............................................................. 8

 

Figure 2-2. Cross Section of a Micro-Fuel Cell (not to scale) .......... 8

 

Figure 2-3. Characteristic Input/Output Curve for a Micro-Fuel

Cell ............................................................................. 11

 

Figure 2-4: Piping Layout and Flow Diagram for Standard Model . 12

 

Figure 2-5: Flow Diagram .............................................................. 13

 

Figure 2-6: 3000TA Internal Electronic Component Location ........ 14

 

Figure 2-7: Block Diagram of the Model 3000TA-EU Electronics . 15

 

Figure 3-1: Front Panel of the Model 3000TA ............................... 18

 

Figure 3-2: Required Front Door Clearance .................................. 18

 

Figure 3-3: Rear Panel of the Model 3000TA ................................ 19  

Figure 3-4: Equipment Interface Connector Pin Arrangement ....... 22

 

Figure 3-5: Remote Probe Connections ........................................ 27

 

Figure 3-6: FET Series Resistance ............................................... 27

 

Figure 5-1: Removing the Micro-Fuel ............................................ 59

 

Figure 5-2: Removing Fuse Block from Housing ........................... 61

 

Figure 5-3: Installing Fuses ........................................................... 62

 

Figure 5-4: Rear-Panel Screws ..................................................... 64

 

Figure A-1: Single and Dual 19" Rack Mounts .............................. 70

 

3000TA-

List of Tables

Table 3-1: Analog Output Connections Pin Function .................... 23

 

Table 3-2: Alarm Relay Contact Pins ............................................ 24

 

Table 3-3: Remote Calibration Connections .................................. 25

 

Table 3-4: Range ID Relay Connections ....................................... 26

 

Table 3-5: Commands via RS-232 Input ....................................... 28

 

Table 5-1: Self Test Failure Codes ................................................ 62

 

Trace Oxygen Analyzer Introduction

Introduction

1.1 Overview

The Teledyne Analytical Instruments Model 3000TA Trace

Oxygen Analyzer is a versatile microprocessor-based instrument for detecting oxygen at the parts-per-million (ppm) level in a variety of gases. This manual covers the Model 3000TA General Purpose flushpanel and/or rack-mount units only. These units are for indoor use in a nonhazardous environment.

1.2 Typical Applications

A few typical applications of the Model 3000TA are:

 Monitoring inert gas blanketing

 Air separation and liquefaction

Chemical reaction monitoring

Semiconductor manufacturing

Petrochemical process control

Quality assurance

Gas analysis certification.

1.3 Main Features of the Analyzer

The Model 3000TA Trace Oxygen Analyzer is sophisticated yet simple to use. The main features of the analyzer include:

 A 2-line alphanumeric display screen, driven by microprocessor electronics, that continuously prompts and informs the operator.

High resolution, accurate readings of oxygen content from low ppm levels through 25%. Large, bright, meter readout.

Nylon cell block. (Stainless steel optional)

Introduction 3000TA-

Advanced Micro-Fuel Cell, designed for trace analysis, has a one year warranty and an expected lifetime of two years.

Versatile analysis over a wide range of applications.

Microprocessor based electronics: 8-bit CMOS microprocessor with 32 kB RAM and 128 kB ROM.

Three user definable output ranges (from 0-10 ppm through 0- 250,000 ppm) allow best match to users process and equipment.

Air-calibration range for convenient spanning at 20.9 %.

Auto Ranging allows analyzer to automatically select the proper preset range for a given measurement. Manual override allows the user to lock onto a specific range of interest.

Two adjustable concentration alarms and a system failure alarm.

Extensive self-diagnostic testing, at startup and on demand, with continuous power-supply monitoring.

CE Compliance.

RS-232 serial digital port for use with a computer or other digital communication device.

Four analog outputs: two for measurement (0–1 VDC and

Isolated 4–20 mA DC) and two for range identification.

Convenient and versatile, steel, flush-panel or rackmountable case with slide-out electronics drawer.

1.4 Model Designations

3000TA: Standard model.

3000TA-C: In addition to all standard features, this model also has separate ports for zero and span gases, and built-in control valves. The internal valves are entirely under the control of the 3000TA electronics, to automatically switch between gases in synchronization with the analyzer’s operations.

Trace Oxygen Analyzer Introduction

1.5 Front Panel (Operator Interface)

The standard 3000TA is housed in a rugged metal case with all controls and displays accessible from the front panel. See Figure 1-1.

The front panel has thirteen buttons for operating the analyzer, a digital meter, an alphanumeric display, and a window for viewing the sample flowmeter.

Figure 1-1: Model 3000TA Front Panel

Function Keys: Six touch-sensitive membrane switches are used to change the specific function performed by the analyzer:

Analyze Perform analysis for oxygen content of a sample gas.

System Perform system-related tasks (described in detail in chapter 4, Operation.).

Span Span calibrate the analyzer.

Zero Zero calibrate the analyzer.

Alarms Set the alarm setpoints and attributes.

Range Set up the 3 user definable ranges for the instrument.

Data Entry Keys: Six touch-sensitive membrane switches are used to input data to the instrument via the alphanumeric VFD display:

Introduction 3000TA-

Left & Right Arrows Select between functions currently displayed on the VFD screen.

Up & Down Arrows Increment or decrement values of functions currently displayed.

Enter Moves VFD display on to the next screen in a series.

If none remains, returns to the Analyze screen.

Escape Moves VFD display back to the previous screen in a series. If none remains, returns to the Analyze screen.

Digital Meter Display: The meter display is a LED device that produces large, bright, 7-segment numbers that are legible in any lighting. It produces a continuous readout from 0-10,000 ppm and then switches to a continuous percent readout from 1-25%. It is accurate across all analysis ranges without the discontinuity inherent in analog range switching.

Alphanumeric Interface Screen: The VFD screen is an easy-to-use interface from operator to analyzer. It displays values, options, and messages that give the operator immediate feedback.

Flowmeter: Monitors the flow of gas past the sensor. Readout is 0.2 to

2.4 standard liters per minute (SLPM).

Standby Button: The Standby button turns off the display and outputs, but circuitry is still operating.

CAUTION: THE POWER CABLE MUST BE UNPLUGGED TO

FULLY DISCONNECT POWER FROM THE

INSTRUMENT. WHEN CHASSIS IS EXPOSED OR

WHEN ACCESS DOOR IS OPEN AND POWER

CABLE IS CONNECTED, USE EXTRA CARE TO

AVOID CONTACT WITH LIVE ELECTRICAL

CIRCUITS.

Access Door: For access to the Micro-Fuel Cell, the front panel swings open when the latch in the upper right corner of the panel is pressed all the way in with a narrow gauge tool. Accessing the main circuit board requires unfastening rear panel screws and sliding the unit out of the case.

1.6 Recognizing Difference Between LCD & VFD

LCD has GREEN background with BLACK characters. VFD has

DARK background with GREEN characters. In the case of VFD - NO

CONTRAST ADJUSTMENT IS NEEDED.

Trace Oxygen Analyzer Introduction

1.7 Rear Panel (Equipment Interface)

The rear panel, shown in Figure 1-2, contains the gas and electrical connectors for external inlets and outlets. Those that are optional are shown shaded in the figure. The connectors are described briefly here and in detail in the Installation chapter of this manual.

Figure 1-2: Model 3000 TA Rear Panel

Power Connection Universal AC power source.

Gas Inlet and Outlet One inlet (must be externally valved) and one exhaust out. Three inlet when “C” option ordered.

RS-232 Port Serial digital concentration signal output and control input.

Remote Valves Used in the 3000TA for controlling external solenoid valves only.

50-Pin Equipment Interface Port:

Analog Outputs 0–1 VDC concentration plus 0-1 VDC range ID, and isolated 4–20 mA DC plus 4-20 mA DC range ID.

Alarm Connections 2 concentration alarms and 1 system alarm.

Introduction 3000TA-

Remote Span/Zero Digital inputs allow external control of analyzer calibration.

Calibration Contact To notify external equipment that instrument is being calibrated and readings are not monitoring sample.

Range ID Contacts Four separate, dedicated, range relay contacts. Low, Medium, High, Cal.

Network I/O Serial digital communications for local network access. For future expansion. Not implemented at this printing.

Optional:

Calibration Gas Ports (Auto Cal Option) Separate fittings for zero, span and sample gas input, and internal valves for automatically switching the gases.

Note: If you require highly accurate Auto-Cal timing, use external

Auto-Cal control where possible. The internal clock in the

Model 3000TA is accurate to 2-3 %. Accordingly, internally scheduled calibrations can vary 2-3 % per day.

Trace Oxygen Analyzer Operational Theory

Operational Theory

2.1 Introduction

The analyzer is composed of three subsystems:

1. Micro-Fuel Cell Sensor

2. Sample System

3. Electronic Signal Processing, Display and Control

The sample system is designed to accept the sample gas and transport it through the analyzer without contaminating or altering the sample prior to analysis. The Micro-Fuel Cell is an electrochemical galvanic device that translates the amount of oxygen present in the sample into an electrical current. The electronic signal processing, display and control subsystem simplifies operation of the analyzer and accurately processes the sampled data. The microprocessor controls all signal processing, input/output and display functions for the analyzer.

2.2 Micro-Fuel Cell Sensor

2.2.1 Principles of Operation

The oxygen sensor used in the Model 3000T series is a Micro-Fuel

Cell designed and manufactured by Analytical Instruments. It is a sealed plastic disposable electrochemical transducer.

The active components of the Micro-Fuel Cell are a cathode, an anode, and the 15% aqueous KOH electrolyte in which they are immersed. The cell converts the energy from a chemical reaction into an electrical current in an external electrical circuit. Its action is similar to that of a battery.

There is, however, an important difference in the operation of a battery as compared to the Micro-Fuel Cell: In the battery, all reactants are stored within the cell, whereas in the Micro-Fuel Cell, one of the reactants (oxygen) comes from outside the device as a constituent of the sample gas being analyzed. The Micro-Fuel Cell is therefore a hybrid between a battery and a true fuel cell. (All of the reactants are stored externally in a true fuel cell.)

Operational Theory 3000TA- EU

2.2.2 Anatomy of a Micro-Fuel Cell

The Micro-Fuel Cell is a cylinder only 11/4 inches in diameter and

11/4 inches thick. It is made of an extremely inert plastic, which can be placed confidently in practically any environment or sample stream. It is effectively sealed, although one end is permeable to oxygen in the sample gas. The other end of the cell is a contact plate consisting of two concentric foil rings. The rings mate with spring-loaded contacts in the sensor block assembly and provide the electrical connection to the rest of the analyzer. Figure 2-1 illustrates the external features.

Figure 2-1: Micro-Fuel Cell

Refer to Figure 2-2, Cross Section of a Micro-Fuel Cell, which illustrates the following internal description.

Figure 2-2. Cross Section of a Micro-Fuel Cell (not to scale)

At the top end of the cell is a diffusion membrane of Teflon®, whose thickness is very accurately controlled. Beneath the diffusion membrane lies the oxygen sensing element—the cathode—with a surface area almost 4 cm 2 . The cathode has many perforations to ensure

Trace Oxygen Analyzer Operational Theory sufficient wetting of the upper surface with electrolyte, and it is plated with an inert metal.

The anode structure is below the cathode. It is made of lead and has a proprietary design which is meant to maximize the amount of metal available for chemical reaction.

At the rear of the cell, just below the anode structure, is a flexible membrane designed to accommodate the internal volume changes that occur throughout the life of the cell. This flexibility assures that the sensing membrane remains in its proper position, keeping the electrical output constant.

The entire space between the diffusion membrane, above the cathode, and the flexible rear membrane, beneath the anode, is filled with electrolyte. Cathode and anode are submerged in this common pool. They each have a conductor connecting them to one of the external contact rings on the contact plate, which is on the bottom of the cell.

2.2.3 Electrochemical Reactions

The sample gas diffuses through the Teflon membrane. Any oxygen in the sample gas is reduced on the surface of the cathode by the following HALF REACTION:

O

2

+ 2H

2

O + 4e – → 4OH – (cathode)

(Four electrons combine with one oxygen molecule—in the presence of water from the electrolyte—to produce four hydroxyl ions.)

When the oxygen is reduced at the cathode, lead is simultaneously oxidized at the anode by the following HALF REACTION:

Pb + 2OH – → Pb +2 + H

2

O + 2e – (anode)

(Two electrons are transferred for each atom of lead that is oxidized. Therefore it takes two of the above anode reactions to balance one cathode reaction and transfer four electrons.)

The electrons released at the surface of the anode flow to the cathode surface when an external electrical path is provided. The current is proportional to the amount of oxygen reaching the cathode. It is

Operational Theory 3000TA- EU measured and used to determine the oxygen concentration in the gas mixture.

The overall reaction for the fuel cell is the SUM of the half reactions above, or:

2Pb + O

2

→2PbO

(These reactions will hold as long as no gaseous components capable of oxidizing lead—such as iodine, bromine, chlorine and fluorine—are present in the sample.)

The output of the fuel cell is limited by (1) the amount of oxygen in the cell at the time and (2) the amount of stored anode material.

In the absence of oxygen, no current is generated.

2.2.4 The Effect of Pressure

In order to state the amount of oxygen present in the sample in parts-per-million or a percentage of the gas mixture, it is necessary that the sample diffuse into the cell under constant pressure.

If the total pressure increases, the rate that oxygen reaches the cathode through the diffusing membrane will also increase. The electron transfer, and therefore the external current, will increase, even though the oxygen concentration of the sample has not changed. It is therefore important that the sample pressure at the fuel cell (usually vent pressure) remain relatively constant between calibrations.

2.2.5 Calibration Characteristics

Given that the total pressure of the sample gas on the surface of the

Micro-Fuel Cell input is constant, a convenient characteristic of the cell is that the current produced in an external circuit is directly proportional to the rate at which oxygen molecules reach the cathode, and this rate is directly proportional to the concentration of oxygen in the gaseous mixture. In other words it has a linear characteristic curve, as shown in

Figure 2-3. Measuring circuits do not have to compensate for nonlinearities.

In addition, since there is zero output in the absence oxygen, the characteristic curve has close to an absolute zero (within ± 1 ppm oxygen). In practical application, zeroing may still used to compensate for the combined zero offsets of the cell and the electronics. (The

Trace Oxygen Analyzer Operational Theory electronics is zeroed automatically when the instrument power is turned on.)

Figure 2-3. Characteristic Input/Output Curve for a Micro-Fuel

Cell

2.3 Sample System

The sample system delivers gases to the Micro-Fuel Cell sensor from the analyzer rear panel inlet. Depending on the mode of operation either sample or calibration gas is delivered.

The Model 3000TA sample system is designed and fabricated to ensure that the oxygen concentration of the gas is not altered as it travels through the sample system. The sample encounters almost no dead space. This minimizes residual gas pockets that can interfere with trace analysis.

The sample system for the standard instrument incorporates 1/4 inch tube fittings for sample inlet and outlet connections at the rear panel. For metric system installations, 6 mm adapters are supplied with each instrument to be used if needed. The sample or calibration gas flows through the system is monitored by a flowmeter downstream from

Operational Theory 3000TA- EU the cell. Figure 2-4 shows the piping layout and flow diagram for the standard model.

Figure 2-4: Piping Layout and Flow Diagram for Standard Model

Figure 2-5 is the flow diagram for the sampling system. In the standard instrument, calibration gases (zero and span) can be connected directly to the Sample In port by teeing to the port with appropriate valves.

The shaded portion of the diagram shows the components added when the –C option is ordered. The valving is installed inside the 3000TA-C enclosure and is regulated by the instruments internal electronics.

Trace Oxygen Analyzer Operational Theory

Figure 2-5: Flow Diagram

2.4 Electronics and Signal Processing

The Model 3000TA Trace Oxygen Analyzer uses an 8031 microcontroller with 32 kB of RAM and 128 kB of ROM to control all signal processing, input/output, and display functions for the analyzer.

System power is supplied from a universal power supply module designed to be compatible with any international power source. Figure

2-6 shows the location of the power supply and the main electronic PC boards.

The signal processing electronics including the microprocessor, analog to digital, and digital to analog converters are located on the motherboard at the bottom of the case. The preamplifier board is mounted on top of the motherboard as shown in the figure. These boards are accessible after removing the back panel. Figure 2-7 is a block diagram of the Analyzer electronics.

Operational Theory 3000TA- EU

Figure 2-6: 3000TA Internal Electronic Component Location

In the presence of oxygen the cell generates a current. A current to voltage amplifier converts this current to a voltage, which is amplified in the second stage amplifier.

The second stage amplifier also supplies temperature compensation for the oxygen sensor output. This amplifier circuit incorporates a thermistor, which is physically located in the cell block. The thermistor is a temperature dependent resistance that changes the gain of the amplifier in proportion to the temperature changes in the block. This change is inversely proportional to the change in the cell output due to the same temperature changes. The result is a signal that is temperature independent. The output from the second stage amplifier is sent to an 18 bit analog to digital converter controlled by the microprocessor.

Trace Oxygen Analyzer Operational Theory

Figure 2-7: Block Diagram of the Model 3000TA-EU Electronics

Operational Theory 3000TA- EU

The digital concentration signal along with input from the control panel is processed by the microprocessor, and appropriate control signals are directed to the display, alarms and communications port. The same digital information is also sent to a 12 bit digital to analog converter that produces the 4-20 mA DC and the 0-1 VDC analog concentration signal outputs, and the analog range ID outputs.

Signals from the power supply are also monitored, and through the microprocessor, the system failure alarm is activated if a malfunction is detected.

Trace Oxygen Analyzer Installation

Installation

Installation of the Model 3000TA Analyzer includes:

1. Unpacking

2. Mounting

3. Gas connections

4. Electrical connections

5. Installing the Micro-Fuel Cell

6. Testing the system.

3.1 Unpacking the Analyzer

Although the analyzer is shipped complete, certain parts, such as fuses and sensors, are wrapped separately to be installed on site as part of the installation. Carefully unpack the analyzer and inspect it for damage. Immediately report any damage or shortages to the shipping agent.

3.2 Mounting the Analyzer

The Model 3000TA is for indoor use in a general purpose area. It is

NOT for hazardous environments of any type.

The standard model is designed for flush panel mounting. Figure 3-

1 is an illustration of the 3000TA standard front panel and mounting bezel. There are four mounting holes—one in each corner of the rigid frame. The Drawings section in the rear of this manual contains outline dimensions and mounting hole spacing diagrams.

On special order, a 19" rack-mounting panel can be provided. For rack mounting, one or two 3000 series analyzers are flush-panel mounted on the rack panel. See Appendix for dimensions of the mounting panel.

Installation 3000TA-

Figure 3-1: Front Panel of the Model 3000TA

All operator controls are mounted on the control panel, which is hinged on the left edge and doubles as the door that provides access to the sensor and cell block inside the instrument. The door is spring loaded and will swing open when the button in the center of the latch

(upper right corner) is pressed all the way in with a narrow gauge tool

(less than 0.18 inch wide), such as a small hex wrench or screwdriver

Allow clearance for the door to open in a 90-degree arc of radius 7.125 inches. See Figure 3-2.

Figure 3-2: Required Front Door Clearance

Trace Oxygen Analyzer Installation

3.3 Rear Panel Connections

Figure 3-3 shows the Model 3000TA rear panel. There are ports for gas, power, and equipment interface. The Zero In and Span In ports are not included on the standard model, but are available as options.

Figure 3-3: Rear Panel of the Model 3000TA

3.3.1 Gas Connections

Before using this instrument, it should be determined if the unit will be used for pressurized service or vacuum service and low pressure applications. Inspect the restrictor kit that came with the unit. The kit consist of two restrictors and a union for 1/4” diameter tubing. Notice that the two 1-3/4” long, 1/4” diameter tubing are restrictors. It has an open end and a closed end with a small circular orifice. The restrictor without the blue sticker is for low pressure and vacuum service. For high pressure applications (5-50psig), use the restrictor that has a blue sticker on the body.

For pressurized service (> 5psig) , use the restrictor with the blue dot and union from the restrictor kit and attach it to the Sample In port.

Installation 3000TA-

The small circular orifice should face away from the back of the unit

(against the direction of gas flow).

For positive pressures less than 5 psig use the low-pressure restrictor without the blue dot in the Sample-in line.

For vacuum service (5-10 in Hg), use the restrictor without the blue dot sticker and union but attach it to the Exhaust Out port. The small circular orifice should face toward the back of the unit (against the direction of gas flow).

Remove the blue sticker from the restrictor before using.

Warning: Operating the unit without restrictors can cause damage to the micro-fuel cell.

The unit is manufactured with 1/4 inch tube fittings, and 6 mm adapters are supplied for metric system installations. For a safe connection:

1. Insert the tube into the tube fitting, and finger-tighten the nut until the tubing cannot be rotated freely, by hand, in the fitting. (This may require an additional 1/8 turn beyond finger-tight.)

2. Hold the fitting body steady with a backup wrench, and with another wrench rotate the nut another 1-1/4 turns.

SAMPLE IN:

In the standard model, gas connections are made at the SAMPLE IN and EXHAUST OUT connections. Calibration gases must be tee'd into the Sample inlet with appropriate valves.

The gas pressure in should be reasonably regulated. Pressures between 2 and 50 psig are acceptable as long as the pressure, once established, will keep the front panel flowmeter reading in an acceptable range (0.1 to 2.4 SLPM). For non-pressurized sample or very low pressure, (2 psig or less) vacuum service plumbing is recommended.

Exact figures will depend on your process.

If greater flow is required for improved response time, install a bypass in the sampling system upstream of the analyzer input.

Trace Oxygen Analyzer Installation

EXHAUST OUT:

Exhaust connections must be consistent with the hazard level of the constituent gases. Check Local, State, and Federal laws, and ensure that the exhaust stream vents to an appropriately controlled area if required.

ZERO IN and SPAN IN (Optional):

These are additional ports for inputting span gas and zero gas.

There are electrically operated valves inside for automatic switching between sample and calibration gases. These valves are completely under control of the 3000T Electronics. They can be externally controlled only indirectly through the Remote Cal Inputs, described below.

Pressure, flow, and safety considerations are the same as prescribed for the SAMPLE IN inlet, above.

3.3.2 Electrical Connections

For safe connections, no uninsulated wiring should be able to come in contact with fingers, tools or clothing during normal operation.

CAUTION: USE SHIELDED CABLES. ALSO, USE PLUGS THAT

PROVIDE EXCELLENT EMI/RFI PROTECTION. THE

PLUG CASE MUST BE CONNECTED TO THE CABLE

SHIELD, AND IT MUST BE TIGHTLY FASTENED TO

THE ANALYZER WITH ITS FASTENING SCREWS.

ULTIMATELY, IT IS THE INSTALLER WHO ENSURES

THAT THE CONNECTIONS PROVIDE ADEQUATE

EMI/RFI SHIELDING.

3.3.2.1

P RIMARY I NPUT P OWER

The power cord receptacle and fuse block are located in the same assembly. Insert the power cord into the power cord receptacle.

CAUTION: POWER IS APPLIED TO THE INSTRUMENT'S

CIRCUITRY AS LONG AS THE INSTRUMENT IS

CONNECTED TO THE POWER SOURCE. THE RED

SWITCH ON THE FRONT PANEL IS FOR SWITCHING

POWER ON OR OFF TO THE DISPLAYS AND OUT-

PUTS ONLY.

The universal power supply requires a 85–250 V ac, 47-63 Hz power source.

Installation 3000TA-

Fuse Installation: The fuse block, at the right of the power cord receptacle, accepts US or European size fuses. A jumper replaces the fuse in whichever fuse receptacle is not used. Fuses are not installed at the factory. Be sure to install the proper fuse as part of installation. (See

Fuse Replacement in chapter 5, maintenance.)

3.3.2.2

50-P IN E QUIPMENT I NTERFACE C ONNECTOR

Figure 3-4 shows the pin layout of the Equipment Interface connector. The arrangement is shown as seen when the viewer faces the rear panel of the analyzer. The pin numbers for each input/output function are given where each function is described in the paragraphs below.

Figure 3-4: Equipment Interface Connector Pin Arrangement

Analog Outputs: There are four DC output signal pins—two pins per output. For polarity, see Table 3-1. The outputs are:

0–1 VDC % of Range: Voltage rises linearly with increasing oxygen, from 0 V at 0 ppm to 1 V at full scale ppm. (Full scale = 100% of programmable range.)

0–1 VDC Range ID: 0.25 V = Low Range, 0.5 V = Medium

Range, 0.75 V = High Range, 1 V = Air

Cal Range.

4–20 mA DC % Range: Current increases linearly with increasing oxygen, from 4 mA at 0 ppm to 20 mA at full scale ppm. (Full scale =

100% of programmable range.)

4–20 mA DC Range ID: 8 mA = Low Range, 12 mA = Medium

Range, 16 mA = High Range, 20 mA =

Air Cal Range.

Trace Oxygen Analyzer Installation

Table 3-1: Analog Output Connections Pin Function

Pin Function

3 + Range ID, 4-20 mA, floating

4

5

– Range ID, 4-20 mA, floating

+ % Range, 4-20 mA, floating

6

8

23

24

7

– % Range, 4-20 mA, floating

+ Range ID, 0-1 VDC

– Range ID, 0-1 VDC, negative ground

+ % Range, 0-1 VDC

– % Range, 0-1 VDC, negative ground

Alarm Relays: The nine alarm-circuit connector pins connect to the internal alarm relay contacts. Each set of three pins provides one set of Form C relay contacts. Each relay has both normally open and normally closed contact connections. The contact connections are shown in Table 3-2. They are capable of switching up to 3 amperes at 250 V ac into a resistive load. The connectors are:

 Threshold Alarm 1:

 Can be configured as high (actuates when concentration is above threshold), or low (actuates when concentration is below threshold).

Can be configured as failsafe or non-failsafe.

Can be configured as latching or non-latching.

 Can be configured out (defeated).

Threshold Alarm 2:

Can be configured as high (actuates when concentration is above threshold), or low (actuates when concentration is below threshold).

Can be configured as failsafe or non-failsafe.

Can be configured as latching or non-latching.

Can be configured out (defeated).

System Alarm:

 Actuates when DC power supplied to circuits is unacceptable in one or more parameters. Permanently

Installation 3000TA-

 configured as failsafe and latching. Cannot be defeated.

Actuates if self test fails.

Reset by pressing button to remove power. Then press

again and any other button EXCEPT System to resume.

Further detail can be found in Chapter 4, Section 4-5.

Table 3-2: Alarm Relay Contact Pins

45 Threshold Alarm 1, normally closed contact

28 Threshold Alarm 1, moving contact

46 Threshold Alarm 1, normally open contact

42 Threshold Alarm 2, normally closed contact

44 Threshold Alarm 2, moving contact

43 Threshold Alarm 2, normally open contact

36 System Alarm, normally closed contact

20 System Alarm, moving contact

37 System Alarm, normally open contact

Digital Remote Cal Inputs: Accept 0 V (off) or 24 VDC (on) inputs for remote control of calibration. (See Remote Calibration

Protocol below.) See Table 3-3 for pin connections.

Zero: Floating input. 5 to 24 V input across the + and – pins puts the analyzer into the Zero mode. Either side may be grounded at the source of the signal. 0 to 1 volt across the terminals allows

Zero mode to terminate when done. A synchronous signal must open and close the external zero valve appropriately. See

Remote Probe Connector. (The –C option internal valves operate automatically.)

Span: Floating input. 5 to 24 V input across the + and – pins puts the analyzer into the Span mode. Either side may be grounded at the source of the signal. 0 to 1 volt across the terminals allows Span mode to terminate when done. A synchronous signal must open and close external span valve appropriately. See Figure 3-5

Trace Oxygen Analyzer Installation

Remote Probe Connector. (The –C option internal valves operate automatically.)

Cal Contact: This relay contact is closed while analyzer is spanning and/or zeroing. (See Remote Calibration Protocol below.)

Table 3-3: Remote Calibration Connections

Pin Function

9

11

10

12

+ Remote Zero

– Remote Zero

+ Remote Span

– Remote Span

Remote Calibration Protocol: To properly time the Digital

Remote Cal Inputs to the Model 3000TA Analyzer, the customer's controller must monitor the Cal Relay Contact.

When the contact is OPEN, the analyzer is analyzing, the Remote

Cal Inputs are being polled, and a zero or span command can be sent.

When the contact is CLOSED, the analyzer is already calibrating. It will ignore your request to calibrate, and it will not remember that request.

Once a zero or span command is sent, and acknowledged (contact closes), release it. If the command is continued until after the zero or span is complete, the calibration will repeat and the Cal Relay Contact

(CRC) will close again.

For example:

1. Test the CRC. When the CRC is open, Send a zero command until the CRC closes (The CRC will quickly close.)

2. When the CRC closes, remove the zero command.

3. When CRC opens again, send a span command until the

CRC closes. (The CRC will quickly close.)

4. When the CRC closes, remove the span command.

When CRC opens again, zero and span are done, and the sample is being analyzed.

Installation 3000TA-

Note: The Remote Valve connections (described below) provides signals to ensure that the zero and span gas valves will be controlled synchronously. If you have the –C Internal valve option—which includes additional zero and span gas inputs— the 3000T automatically regulates the zero, span and sample gas flow.

Range ID Relays: Four dedicated Range ID relay contacts. The first three ranges are assigned to relays in ascending order—Low range is assigned to Range 1 ID, Medium range is assigned to Range 2 ID, and

High range is assigned to Range 3 ID. The fourth range is reserved for the Air Cal Range (25%). Table 3-4 lists the pin connections.

Table 3-4: Range ID Relay Connections

Pin Function

21 Range 1 ID Contact

38

22

Range 1 ID Contact

Range 2 ID Contact

39

19

18

34

35

Range 2 ID Contact

Range 3 ID Contact

Range 3 ID Contact

Range 4 ID Contact (Air Cal)

Range 4 ID Contact (Air Cal)

Network I/O: A serial digital input/output for local network protocol. At this printing, this port is not yet functional. It is to be used for future options to the instrument. Pins 13 (+) and 29 (–).

Remote Valve Connections: The 3000TA is a single-chassis instrument, which has no Remote Valve Unit. Instead, the Remote Valve connections are used as a method for directly controlling external sample/zero/span gas valves. See Figure 3-5.

Trace Oxygen Analyzer Installation

Figure 3-5: Remote Probe Connections

The voltage from these outputs is nominally 0 V for the OFF and

15 VDC for the ON conditions. The maximum combined current that can be pulled from these output lines is 100 mA. (If two lines are ON at the same time, each must be limited to 50 mA, etc.) If more current and/or a different voltage is required, use a relay, power amplifier, or other matching circuitry to provide the actual driving current.

In addition, each individual line has a series FET with a nominal

ON resistance of 5 ohms (9 ohms worst case). This can limit the obtainable voltage, depending on the load impedance applied. See

Figure 3-6.

Figure 3-6: FET Series Resistance

3.3.2.3 RS-232 P ORT

The digital signal output is a standard, full duplex RS-232 serial communications port used to connect the analyzer to a computer,

Installation 3000TA- terminal, or other digital device. It requires a standard 9-pin D connector.

The output data is status information, in digital form, updated every two seconds. Status is reported in the following order:

 The concentration in ppm or percent

The range in use (HI, MED, LO)

The span of the range (0-100 ppm, etc)

Which alarms—if any—are disabled (AL–x DISABLED)

Which alarms—if any—are tripped (AL–x ON).

Each status output is followed by a carriage return and line feed.

Three input functions using RS-232 have been implemented to date. They are described in Table 3-5.

Table 3-5: Commands via RS-232 Input as <enter> Immediately starts an autospan. az <enter> Immediately starts an autozero. st <enter> Toggling input. Stops/Starts any status message output from the RS-232, until st <enter> is sent again.

The RS-232 protocol allows some flexibility in its implementation.

Table 3-6 lists certain RS-232 values that are required by the 3000TA implementation.

Table 3-6: Required RS-232 Options

Parameter Setting

Baud 2400

Byte 8

Parity none

Stop 1

Message Interval 2 seconds. When CRC opens again, zero and span are done,

Trace Oxygen Analyzer Installation

3.4 Installing the Micro-Fuel Cell

The Micro-Fuel Cell is not installed in the cell block when the instrument is shipped. Install it before the analyzer is placed in service.

Once it is expended, or if the cell is exposed to air for too long, the

Micro-Fuel Cell will need to be replaced. The cell could also require replacement if the instrument has been idle for too long.

When the micro-Fuel Cell needs to be installed or replaced, follow the procedures in Chapter 5, Maintenance, for removing and installing cells.

3.5 Testing the System

Before plugging the instrument into the power source:

 Check the integrity and accuracy of the gas connections.

Make sure there are no leaks.

Check the integrity and accuracy of the electrical connections. Make sure there are no exposed conductors

Verify that the restriction device has been properly installed

(see section 3.3.1).

Check that inlet sample pressure is within the accepted range

(se section 3.3.1).

Power up the system, and test it by repeating the Self-

Diagnostic Test as described in Chapter 4, Section 4.3.5.

Trace Oxygen Analyzer Operation

Operation

4.1 Introduction

Once the analyzer has been installed, it can be configured for your application. To do this you will:

 Set system parameters:

 Establish a security password, if desired, requiring

Operator to log in.

 Establish and start an automatic calibration cycle, if desired.

Calibrate the instrument.

Define the three user selectable analysis ranges. Then choose autoranging or select a fixed range of analysis, as required.

Set alarm setpoints, and modes of alarm operation (latching, failsafe, etc).

Before you configure your 3000TA these default values are in effect:

Ranges: LO = 100 ppm, MED = 1000 ppm, HI = 10,000 ppm.

Auto Ranging: ON

Alarm Relays: Defeated, 1000 ppm, HI, Not failsafe, Not latching.

Zero: Auto, every 0 days at 0 hours.

Span: Auto, at 000008.00 ppm, every 0 days at 0 hours.

If you choose not to use password protection, the default password is automatically displayed on the password screen when you start up, and you simply press Enter for access to all functions of the analyzer.

4.2 Using the Data Entry and Function Buttons

Data Entry Buttons: The ◄►arrow buttons select options from the menu currently being displayed on the VFD screen. The selected option blinks.

Operation 3000TA-

When the selected option includes a modifiable item, the ▲/▼ arrow buttons can be used to increment or decrement that modifiable item.

The Enter button is used to accept any new entries on the VFD screen. The Escape button is used to abort any new entries on the VFD screen that are not yet accepted by use of the Enter button.

Figure 4-1 shows the hierarchy of functions available to the operator via the function buttons. The six function buttons on the analyzer are:

 Analyze.

This is the normal operating mode. The analyzer monitors the oxygen content of the sample, displays the percent of oxygen, and warns of any alarm conditions.

 System.

The system function consists of six subfunctions that regulate the internal operations of the analyzer:

 Auto-Cal setup

Password assignment

Self-Test initiation

Checking software version

 Logging out.

Zero . Used to set up a zero calibration.

Span.

Used to set up a span calibration.

Alarms.

Used to set the alarm setpoints and determine whether each alarm will be active or defeated, HI or LO acting, latching, and/or failsafe.

Range.

Used to set up three analysis ranges that can be switched automatically with auto-ranging or used as individual fixed ranges.

Any function can be selected at any time by pressing the appropriate button (unless password restrictions apply). The order as presented in this manual is appropriate for an initial setup.

Each of these functions is described in greater detail in the following procedures. The VFD screen text that accompanies each operation is reproduced, at the appropriate point in the procedure, in a

Monospaced type style. Pushbutton names are printed in Oblique type.

Trace Oxygen Analyzer Operation

Figure 4-1: Hierarchy of Functions and Sub functions

4.3 The System Function

The subfuctions of the System function are described below.

Specific procedures for their use follow the descriptions:

 Auto-Cal

: Used to define an automatic calibration sequence and/or start an Auto-Cal.

 PSWD:

Security can be established by choosing a 5 digit password (PSWD) from the standard ASCII character set.

Operation 3000TA-

(See Installing or Changing a Password, below, for a table of

ASCII characters available.) Once a unique password is assigned and activated, the operator MUST enter the

UNIQUE password to gain access to set-up functions which alter the instrument's operation, such as setting the instrument span or zero setting, adjusting the alarm setpoints, or defining analysis ranges.

After a password is assigned, the operator must log out to activate it. Until then, anyone can continue to operate the instrument without entering the new password.

Only one password can be defined. Before a unique password is assigned, the system assigns TETAI by default.

This allows access to anyone. After a unique password is assigned, to defeat the security, the password must be changed back to TETAI.

Logout: Logging out prevents an unauthorized tampering with analyzer settings.

More : Select and enter More to get a new screen with additional subfunctions listed.

Self–Test : The instrument performs a self-diagnostic test to check the integrity of the power supply, output boards and amplifiers.

Version : Displays Manufacturer, Model, and Software

Version of the instrument.

Show Negative : The operator selects whether display can show negative oxygen readings or not.

TRAK/HLD : The operator sets whether the instrument analog outputs track the concentration change during calibration and sets a time delay for the concentration alarms after calibration

4.3.1 Tracking Oxygen Readings During Calibration and

Alarm Delay

The user has the option of setting the preference as to whether the analog outputs track the display readings during calibration or not. To set the preference, press the System key once and the first System menu will appear in the VFD display:

Trace Oxygen Analyzer Operation

TRAK/HLD Auto-Cal

PSWD Logout More

TRAK/HLD should be blinking. To enter this system menu press the Enter key once:

Output Sttng: TRACK

Alarm Dly: 10 min or

Output Sttng: HOLD

Alarm Dly: 10 min

In the first line, TRACK or HOLD should be blinking. The operator can toggle between TRACK and HOLD with the Up or Down keys. When TRACK is selected, the analog outputs (0-1 VDC and 4-20 mA) and the range ID contacts will track the instrument readings during calibration (either zero or span). TRACK is the factory default.

When HOLD is selected, the analog outputs (0-1 VDC and 4-20 mA) and the range ID contacts will freeze on their last state before entering one of the calibration modes. When the instrument returns to the Analyze mode, either by a successful or an aborted calibration, there will be a three-minute delay before the analog outputs and the range ID contacts start tracking again.

The concentration alarms freeze on their last state before entering calibration regardless of selecting HOLD or TRACK. But, when HOLD is selected the concentration alarms will remain frozen for the time displayed in the second line of the TRAK/HLD menu after the analyzer returns to the Analyze mode.

The factory default is three minutes, but the delay time is programmable. To adjust to delay time use the Left or Right arrow keys. When the time displayed on the second line blinks, it can be adjusted by

Pressing the Up or Down keys to increase or decrease its value. The minimum delay is 1 minute, the maximum is 30.

This preference is stored in non-volatile memory so that it is recovered if power is removed from the instrument.

Operation 3000TA-

4.3.2 Setting up an Auto-Cal

When proper automatic valving is connected (see Chapter 3,

Installation), the analyzer can cycle itself through a sequence of steps that automatically zero and span the instrument.

Note: If you require highly accurate Auto-Cal timing, use external

Auto-Cal control where possible. The internal clock in the

Model 3000TA is accurate to 2-3 %. Accordingly, internally scheduled calibrations can vary 2-3 % per day.

To setup an Auto–Cal cycle:

Choose System from the Function buttons. The LCD will display five subfunctions.

TRAK/HLD Auto—Cal

PSWD Logout More

Use ◄►arrows to blink Auto—Cal , and press Enter. A new screen for Span/Zero set appears.

Span OFF Nxt: 0d 0h

Zero OFF Nxt: 0d 0h

Press ◄►arrows to blink Span (or Zero ), then press Enter again.

(You won’t be able to set OFF to ON if a zero interval is entered.) A

Span Every ... (or Zero Every ...) screen appears.

Span Every 0 d

Start 0 h from now

Use ▲/▼ arrows to set an interval value, then use ◄ arrows to move to the start-time value. Use ▲/▼ arrows to set a start-time value.

To turn ON the Span and/or Zero cycles (to activate Auto-Cal):

Press System again, choose Auto—Cal, and press Enter again. When the Span/ Zero values screen appears, use the ◄► arrows to blink the

Span (or Zero) OFF/ON field. Use ▲/▼ arrows to set the OFF/ON field to ON. You can now turn these fields ON because there is a nonzero span interval defined.

Trace Oxygen Analyzer Operation

4.3.3 Password Protection

If a password is assigned, then setting the following system parameters can be done only after the password is entered: span and

zero settings, alarm setpoints, analysis range definitions, switching between autoranging and manual override, setting up an auto-cal, and assigning a new password. However, the instrument can still be used for analysis or for initiating a self- test without entering the password.

If you have decided not to employ password security, use the default password TETAI. This password will be displayed automatically by the microprocessor. The operator just presses the Enter key to be allowed total access to the instrument’s features.

Note: If you use password security, it is advisable to keep a copy of the password in a separate, safe location.

4.3.3.1

E NTERING THE P ASSWORD

To install a new password or change a previously installed password, you must key in and ENTER the old password first. If the default password is in effect, pressing the ENTER button will enter the default TETAI password for you.

Press System to enter the System mode.

TRAK/HLD Auto—Cal

PSWD Logout More

Use the ◄►arrow keys to scroll the blinking over to PSWD , and press Enter to select the password function. Either the default TETAI password or AAAAA place holders for an existing password will appear on screen depending on whether or not a password has been previously installed.

T E T A I

Enter PWD or

Operation 3000TA-

A A A A A

Enter PWD

The screen prompts you to enter the current password. If you are not using password protection, press Enter to accept TETAI as the default password.

If a password has been previously installed, enter the password using the ◄►arrow keys to scroll back and forth between letters, and the ▲/▼ arrow keys to change the letters to the proper password. Press Enter to enter the password.

If the password is accepted, the screen will indicate that the password restrictions have been removed and you have clearance to proceed.

PSWD Restrictions

Removed

In a few seconds, you will be given the opportunity to change this password or keep it and go on.

Change Password?

<ENT>=Yes <ESC>=No

Press Escape to move on, or proceed as in Changing the

Password, below.

4.3.3.2

I NSTALLING OR C HANGING THE P ASSWORD

If you want to install a password, or change an existing password, proceed as above in Entering the Password. When you are given the opportunity to change the password:

Change Password?

<ENT>=Yes <ESC>=No

Press Enter to change the password (either the default TETAI or the previously assigned password), or press Escape to keep the existing password and move on.

If you chose Enter to change the password, the password assignment screen appears.

Trace Oxygen Analyzer Operation

T E T A I

<ENT> To Proceed or

A A A A A

<ENT> To Proceed

Enter the password using the ◄►arrow keys to move back and forth between the existing password letters, and the ▲/▼ arrow keys to change the letters to the new password. The full set of 94 characters available for password use are shown in the table below.

Characters Available for Password Definition:

A B C D E F G H I J

K L M N O P Q R S T

U V W X Y Z [ ¥ ] ^

` a b c d e f g h i j k l m n o p q r s t u v w x y z { |

}

!

" # $ % & ' (

3 4 5 6 7

= > ? @

8 9 : ; <

When you have finished typing the new password, press Enter. A verification screen appears. The screen will prompt you to retype your password for verification.

A A A A A

Retype PWD To Verify

Wait a moment for the entry screen. You will be given clearance to proceed.

A A A A A

<ENT> TO Proceed

Operation 3000TA-

Use the arrow keys to retype your password and press Enter when finished. Your password will be stored in the microprocessor and the system will immediately switch to the Analyze screen, and you now have access to all instrument functions.

If all alarms are defeated, the Analyze screen appears as:

0.0 ppm Anlz

Range: 0 — 100

If an alarm is tripped, the second line will change to show which alarm it is:

0.0 ppm Anlz

AL—1

Note: If you log off the system using the logout function in the system menu, you will now be required to re-enter the password to gain access to Span, Zero, Alarm, and Range functions.

4.3.4 Logout

The Logout function provides a convenient means of leaving the analyzer in a password protected mode without having to shut the instrument off. By entering Logout, you effectively log off the instrument leaving the system protected against use until the password is reentered. To log out, press the System button to enter the System function.

TRAK/HLD Auto—Cal

PSWD Logout More

Use the ◄►arrow keys to position the blinking over the Logo ut function, and press Enter to Log out. The screen will display the message:

Trace Oxygen Analyzer Operation

Protected Until

Password Reentered

4.3.5 System Self-Diagnostic Test

The Model 3000TA has a built-in self-diagnostic testing routine.

Pre-programmed signals are sent through the power supply, output board and sensor circuit. The return signal is analyzed, and at the end of the test the status of each function is displayed on the screen, either as OK or as a number between 1 and 3. (See System Self Diagnostic Test in

Chapter 5 for number code).

The self diagnostics are run automatically by the analyzer whenever the instrument is turned on, but the test can also be run by the operator at will. To initiate a self diagnostic test during operation:

Press the System button to start the System function.

TRAK/HLD Auto—Cal

PSWD Logout More

Use the ◄►arrow keys to blink More, then press Enter.

Version Self—Test

Use the ◄►arrow keys again to move the blinking to the Self–

Test function. The screen will follow the running of the diagnostic.

RUNNING DIAGNOSTIC

Testing Preamp — 83

During preamp testing there is a countdown in the lower right corner of the screen. When the testing is complete, the results are displayed.

Power: OK Analog: OK

Preamp: 3

Operation 3000TA-

The module is functioning properly if it is followed by OK. A number indicates a problem in a specific area of the instrument. Refer to

Chapter 5 Maintenance and Troubleshooting for number-code information. The results screen alternates for a time with:

Press Any Key

To Continue...

Then the analyzer returns to the initial System screen.

4.3.6 Version Screen

Move the ◄►arrow key to More and press Enter. With Version blinking, press Enter. The screen displays the manufacturer, model, and software version information.

4.3.7 Showing Negative Oxygen Readings

For software version 1.4.4 or later, the instrument only displays oxygen readings that are positive or zero. The instrument can be reconfigured to show negative readings if sensor output drifts below zero. This situation may arise after the instrument has been zeroed, as time progresses the sensor may drift below the zero calibration setpoint.

To show negative oxygen readings on the display:

Press the System key.

TRAK/HLD Auto-Cal

PSWD Logout More

Use the Right or Left arrow keys and select More. Press Enter.

Version Self-Test

Show_Negative=NO

Use the Right or Left Arrow keys and select “Show_Negative=NO”.

Use the Up or Down key to toggle from NO to YES.

Press the Escape key twice to return to the analyze mode.

Trace Oxygen Analyzer Operation

This preference is stored in non-volatile memory, so this configuration is remembered after a power shutdown. If the instrument is cold started, it will go back to default (not showing negative oxygen readings).

4.4 The Zero and Span Functions

Note: Zeroing is not required in order to achieve the published accuracy specification of this unit.

Zeroing will eliminate offset error contributed by sensor, electronics, and internal and external sampling system and improve performance beyond published specification limits.

The analyzer is calibrated using zero and span gases.

Any suitable oxygen-free gas can be used for zero gas as long as it is known to be oxygen free and does not react adversely with the sample system.

Although the instrument can be spanned using air, a span gas with a known oxygen concentration in the range of 70–90% of full scale of the range of interest is recommended. Since the oxygen concentration in air is 20.9% (209,000 ppm), the cell can take a long time to recover if the instrument is used for trace oxygen analysis immediately following calibration in air.

Connect the calibration gases according to the instructions given in

Section 3.4.1, Gas Connections, observing all the prescribed precautions.

Shut off the gas pressure before connecting it to the analyzer, and be sure to limit the pressure to 40 psig or less when turning it back on.

Readjust the gas pressure into the analyzer until the flowrate (as read on the analyzer’s SLPM flowmeter) settles between 0.15 and 2.4

SLPM (approximately 0.2 - 5 SCFH).

If you are using password protection, you will need to enter your password to gain access to either of these functions. Follow the instructions in Sections 4.3.3 to enter your password. Once you have gained clearance to proceed, you can enter the Zero or Span function.

Operation 3000TA-

4.4.1 Zero Cal

The Zero button on the front panel is used to enter the zero calibration function. Zero calibration can be performed in either the automatic or manual mode. In the automatic mode, an internal algorithm compares consecutive readings from the sensor to determine when the output is within the acceptable range for zero. In the manual mode, the operator determines when the reading is within the acceptable range for zero. Make sure the zero gas is connected to the instrument. If you get a CELL FAILURE message skip to Section 4.4.1.3.

4.4.1.1

A UTO M ODE Z EROING

Press Zero to enter the zero function mode. The screen allows you to select whether the zero calibration is to be performed automatically or manually. Use the ▲/▼ arrow keys to toggle between AUTO and MAN zero settling. Stop when AUTO appears, blinking, on the display.

Zero: Settling: AUTO

<ENT> To Begin

Press Enter to begin zeroing.

####PPM Zero

Slope=#### ppm/s

The beginning zero level is shown in the upper left corner of the display. As the zero reading settles, the screen displays and updates information on Slope (unless the Slope starts within the acceptable zero range and does not need to settle further).

Then, and whenever Slope is less than 0.08 for at least 3 minutes, instead of Slope you will see a countdown: 5 Left, 4 Left, and so forth.

These are five steps in the zeroing process that the system must complete, AFTER settling, before it can go back to Analyze.

####PPM Zero

4 Left=### ppm/s

Trace Oxygen Analyzer Operation

The zeroing process will automatically conclude when the output is within the acceptable range for a good zero. Then the analyzer automatically returns to the Analyze mode.

4.4.1.2

M ANUAL M ODE Z EROING

Press Zero to enter the Zero function. The screen that appears allows you to select between automatic or manual zero calibration. Use the ▲/▼ keys to toggle between AUTO and MAN zero settling. Stop when MAN appears, blinking, on the display.

Zero: Settling: Man

<ENT> To Begin

Press Enter to begin the zero calibration. After a few seconds the first of five zeroing screens appears. The number in the upper left hand corner is the first-stage zero offset. The microprocessor samples the output at a predetermined rate. It calculates the differences between successive samplings and displays the rate of change as Slope= a value in parts per million per second (ppm/s).

####ppm Zero

Slope=#### ppm/s

Note: It takes several seconds for the true Slope value to display.

Wait about 10 seconds. Then, wait until Slope is sufficiently close to zero before pressing Enter to finish zeroing.

Generally, you have a good zero when Slope is less than 0.05 ppm/s for about 30 seconds. When Slope is close enough to zero, press

Enter . In a few seconds, the screen will update.

Once span settling completes, the information is stored in the microprocessor, and the instrument automatically returns to the Analyze mode.

4.4.1.3

C ELL F AILURE

Cell failure in the 3000TA is usually associated with inability to zero the instrument down to a satisfactorily low ppm reading. When this occurs, the 3000TA system alarm trips, and the LCD displays a failure message.

Operation 3000TA-

#.# ppm Anlz

CELL FAIL/ ZERO HIGH

Before replacing the cell:

 Check your span gas to make sure it is within specifications.

 Check for leaks downstream from the cell, where oxygen may be leaking into the system.

If there are no leaks and the span gas is OK, replace the cell as described in Chapter 5, Maintenance.

4.4.2 Span Cal

The Span button on the front panel is used to span calibrate the analyzer. Span calibration can be performed using the automatic mode, where an internal algorithm compares consecutive readings from the sensor to determine when the output matches the span gas concentration.

Span calibration can also be performed in manual mode, where the operator determines when the span concentration reading is acceptable and manually exits the function.

4.4.2.1

A UTO M ODE S PANNING

Press Span to enter the span function. The screen that appears allows you to select whether the span calibration is to be performed automatically or manually. Use the ▲/▼ arrow keys to toggle between

AUTO and MAN span settling. Stop when AUTO appears, blinking, on the display.

Span: Settling: AUTO

<ENT> For Next

Press Enter to move to the next screen.

Calib. Holding time

Cal hold: 5 min

Trace Oxygen Analyzer Operation

This menu allows the operator to set the time the analyzer should be held in the span mode, after the readings of the analyzer settle. Five minutes is the default, but it could be adjusted anywhere from 1 to 60 minutes by using the UP or DOWN keys.

Press Enter to move to the next screen.

Span Val: 000008.00

<ENT>Span <UP>Mod #

Use the ▲/▼ arrow keys to enter the oxygen-concentration mode.

Use the ◄►arrow keys to blink the digit you are going to modify. Use the ▲/▼ arrow keys again to change the value of the selected digit.

When you have finished typing in the concentration of the span gas you are using (209000.00 if you are using air), press Enter to begin the Span calibration.

#### ppm Span

Slope=#### ppm/s

The beginning span value is shown in the upper left corner of the display. As the span reading settles, the screen displays and updates information on Slope. Spanning automatically ends when the span output corresponds, within tolerance, to the value of the span gas concentration. Then the instrument automatically returns to the analyze mode.

4.4.2.2

M ANUAL M ODE S PANNING

Press Span to start the Span function. The screen that appears allows you to select whether the span calibration is to be performed automatically or manually.

Span: Settling:MAN

<ENT> For Next

Operation 3000TA-

Use the ▲/▼ keys to toggle between AUTO and MAN span settling. Stop when MAN appears, blinking, on the display. Press Enter to move to the next screen.

Press Enter to move to the next screen.

Calib. Holding time

Cal hold: 5 min

This menu allows the operator to set the time the analyzer should be held in the auto span mode. It does not affect anything in Manual

Mode. Just press Enter to continue.

Span Val: 000008.00

<ENT>Span <UP>Mod #

Press ▲ (<UP>) to permit modification (Mod #) of span value.

Use the arrow keys to enter the oxygen concentration of the span gas you are using (209000.00 if you are using air). The ◄►arrows choose the digit, and the ▲/▼ arrows choose the value of the digit.

Press Enter to enter the span value into the system and begin the span calibration.

Once the span has begun, the microprocessor samples the output at a predetermined rate. It calculates the difference between successive samplings and displays this difference as Slope on the screen. It takes several seconds for the first Slope value to display. Slope indicates rate of change of the Span reading. It is a sensitive indicator of stability.

#### % Span

Slope=#### ppm/s

When the Span value displayed on the screen is sufficiently stable, press Enter. (Generally, when the Span reading changes by 1% or less of the full scale of the range being calibrated for a period of ten minutes it is sufficiently stable.) Once Enter is pressed, the Span reading changes to the correct value. The instrument then automatically enters the Analyze function.

Trace Oxygen Analyzer Operation

4.4.3 Span Failure

The analyzer checks the output of the cell at the end of the span. If the raw output of the cell is less than 0.5 uA/ppm O

2

, the span will not be accepted. The analyzer will return to the previous calibration values, trigger the System Alarm, and display in the VFD:

Span Failed!!

This message will be shown for five seconds and the instrument will return to the Analyze mode. In the upper right hand corner of the

VFD display “FCAL” will be shown. This message flag will help the operator troubleshoot in case calibration was initiated remotely. To reset the alarm and the flag message, the unit must be turned off by cycling the standby key . It will reset if the next span cycle is correct.

A trace cell is unlikely to fail span. As explained before, when the sensor reaches the end of its useful life, the zero offset begins to rise until the analyzer finds the zero unsatisfactory. Nevertheless, feeding the wrong span gas or electronics failure could set this feature off at the end of the span. Consider this before replacing the cell.

4.5 The Alarms Function

The Model 3000TA is equipped with 2 fully adjustable concentration alarms and a system failure alarm. Each alarm has a relay with a set of form “C" contacts rated for 3 amperes resistive load at 250 V ac.

See Figure in Chapter 3, Installation and/or the Interconnection Diagram included at the back of this manual for relay connections.

The system failure alarm has a fixed configuration described in

Chapter 3 Installation.

The concentration alarms can be configured from the front panel as either high or low alarms by the operator. The alarm modes can be set as latching or non-latching, and either failsafe or non-failsafe, or, they can be defeated altogether. The setpoints for the alarms are also established using this function.

Decide how your alarms should be configured. The choice will depend upon your process. Consider the following four points:

1. Which if any of the alarms are to be high alarms and which if any are to be low alarms?

Setting an alarm as HIGH triggers the alarm when the oxygen concentration rises above the setpoint. Setting an

Operation 3000TA- alarm as LOW triggers the alarm when the oxygen concentration falls below the setpoint.

Decide whether you want the alarms to be set as:

 Both high (high and high-high) alarms, or

 One high and one low alarm, or

 Both low (low and low-low) alarms.

2. Are either or both of the alarms to be configured as failsafe?

In failsafe mode, the alarm relay de-energizes in an alarm condition. For non-failsafe operation, the relay is energized in an alarm condition. You can set either or both of the concentration alarms to operate in failsafe or non-failsafe mode.

3. Are either of the alarms to be latching?

In latching mode, once the alarm or alarms trigger, they will remain in the alarm mode even if process conditions revert back to non-alarm conditions. This mode requires an alarm to be recognized before it can be reset. In the non-latching mode, the alarm status will terminate when process conditions revert to non- alarm conditions.

4. Are either of the alarms to be defeated?

The defeat alarm mode is incorporated into the alarm circuit so that maintenance can be performed under conditions which would normally activate the alarms.

The defeat function can also be used to reset a latched alarm.

(See procedures, below.)

If you are using password protection, you will need to enter your password to access the alarm functions. Follow the instructions in section 4.3.3 to enter your password. Once you have clearance to proceed, enter the Alarm function.

Press the Alarm button on the front panel to enter the Alarm function. Make sure that AL–1 is blinking.

AL—1 AL—2

Choose Alarm

Trace Oxygen Analyzer Operation

Set up alarm 1 by moving the blinking over to AL–1 using the

◄►arrow keys. Then press Enter to move to the next screen.

AL—1 1000 ppm HI

Dft—N Fs—N Ltch—N

Five parameters can be changed on this screen:

 Value of the alarm setpoint, AL–1 #### ppm (oxygen)

Out-of-range direction, HI or LO

Defeated? Dft–Y/N (Yes/No)

Failsafe? Fs–Y/N (Yes/No)

Latching? Ltch–Y/N (Yes/No).

To define the setpoint, use the ◄►arrow keys to move the blinking over to AL–1 ####. Then use the ▲/▼ arrow keys to change the number. Holding down the key speeds up the incrementing or decrementing. (Remember, the setpoint units are ppm O

2

.)

To set the other parameters use the ◄►arrow keys to move the blinking over to the desired parameter. Then use the ▲/▼ arrow keys to change the parameter.

Once the parameters for alarm 1 have been set, press Alarms again, and repeat this procedure for alarm 2 (AL–2).

To reset a latched alarm, go to Dft– and then press either

A times or

V

two times. (Toggle it to Y and then back to N.)

two

–OR –

Go to Ltch– and then press either ▲ two times or ▼ two times.

(Toggle it to N and back to Y.)

4.6 The Range Function

The Range function allows the operator to program up to three concentration ranges to correlate with the DC analog outputs. If no ranges are defined by the user, the instrument defaults to:

Low = 0–100 ppm

Med = 0–1,000 ppm

High = 0–1 0,000 ppm

Operation 3000TA-

The Model 3000TA is set at the factory to default to autoranging.

In this mode, the microprocessor automatically responds to concentration changes by switching ranges for optimum readout sensitivity. If the current range limits are exceeded, the instrument will automatically shift to the next higher range. If the concentration falls to below 85% of full scale of the next lower range, the instrument will switch to that range. A corresponding shift in the DC percent-of-range output, and in the range ID outputs, will be noticed.

The autoranging feature can be overridden so that analog output stays on a fixed range regardless of the oxygen concentration detected. If the concentration exceeds the upper limit of the range, the DC output will saturate at 1 VDC (20 mA at the current output).

However, the digital readout and the RS-232 output of the concentration are unaffected by the fixed range. They continue to read accurately with full precision. See Front Panel description in Chapter 1.

The automatic air calibration range is always 0-25 % and is not programmable.

4.6.1 Setting the Analog Output Ranges

To set the ranges, enter the range function mode by pressing the

Range button on the front panel.

L—100 M—1000

H—1 0000 Mode—AUTO

Use the ◄►arrow keys to blink the range to be set: low (L), medium (M), or high (H).

Use the ▲/▼ arrow keys to enter the upper value of the range (all ranges begin at 0 ppm). Repeat for each range you want to set. Press

Enter to accept the values and return to Analyze mode. (See note below.)

Note: The ranges must be increasing from low to high, for example, if range 1 is set as 0–100 ppm and range 2 is set as 0–1,000 ppm, range 3 cannot be set as 0– 500 ppm since it is lower than range 2.

Ranges, alarms, and spans are always set in ppm units (over the entire 0-250,000 ppm range), even though all concentration-data outputs

Trace Oxygen Analyzer Operation change from ppm units to percent when the concentration is above

10,000 ppm.

4.6.2 Fixed Range Analysis

The autoranging mode of the instrument can be overridden, forcing the analyzer DC outputs to stay in a single predetermined range.

To switch from autoranging to fixed range analysis, enter the range function by pressing the Range button on the front panel.

Use the ◄►arrow keys to move the blinking over AUTO.

Use the ▲/▼ arrow keys to switch from AUTO to FX/LO, FX/M ED, or FX/H I to set the instrument on the desired fixed range (low, medium, or high).

L—100 M—1000

H—1 0000 Mode—FX/ LO

or

L—100 M—1000

H—1 0000 Mode—FX/MED

or

L—100 M—1000

H—1 0000 Mode—FX/ HI

Press Escape to re-enter the Analyze mode using the fixed range.

Note: When performing analysis on a fixed range, if the oxygen concentration rises above the upper limit (or default value) as established by the operator for that particular range, the output saturates at 1 VDC (or 20 mA). However, the digital readout and the RS-232 output continue to read the true value of the oxygen concentration regardless of the analog output range.

Operation 3000TA-

4.7 The Analyze Function

Normally, all of the functions automatically switch back to the

Analyze function when they have completed their assigned operations.

Pressing the Escape button in many cases also switches the analyzer back to the Analyze function. Alternatively, you can press the Analyze button at any time to return to analyzing your sample.

4.8 Signal Output

The standard Model 3000TA Trace Oxygen Analyzer is equipped with two 0–1 VDC analog output terminals accessible on the back panel

(one concentration and one range ID), and two isolated 4–20 mA DC current outputs (one concentration and one range ID).

See Rear Panel in Chapter 3, Installation, for illustration.

The signal output for concentration is linear over the currently selected analysis range. For example, if the analyzer is set on range that was defined as 0–100 ppm O

2

, then the output would be:

Voltage Signal

Output (VDC)

Current Signal

Output (mA dc) ppm O

2

0 0.0 4.0

10 0.1 5.6

20 0.2 7.2

30 0.3 8.8

40 0.4 10.4

50 0.5 12.0

60 0.6 13.6

70 0.7 15.2

80 0.8 16.8

90 0.9 18.4

100 1.0 20.0

The analog output signal has a voltage which depends on the oxygen concentration AND the currently activated analysis range. To relate the signal output to the actual concentration, it is necessary to know what range the instrument is currently on, especially when the analyzer is in the autoranging mode.

To provide an indication of the range, a second pair of analog output terminals are used. They generate a steady preset voltage (or

Trace Oxygen Analyzer Operation current when using the current outputs) to represent a particular range.

The following table gives the range ID output for each analysis range:

Range Voltage (V) Current (mA)

CAL (0-25%) 1.00 20

IMPORTANT: In the event of loss of flow through the analyzer, if the vent is vented to a location of high oxygen content, oxygen will back diffuse through the vent line and in most cases quickly saturate the cell with oxygen which can then require a quite long purge down time for the sensor when then exposed to low oxygen concentrations. In the event that flow is to be interrupted into the analyzer, it is suggested that the user do one of the following:

1. Bag the sensor in nitrogen during this time

2. Install a shut off valve on the vent port of the analyzer or somewhere within the users sample system.

Operation 3000TA-

Trace Oxygen Analyzer Maintenance

Maintenance

5.1 Routine Maintenance

Aside from normal cleaning and checking for leaks at the gas connections, routine maintenance is limited to replacing Micro-Fuel cells and fuses, and recalibration. For recalibration, see Section 4.4

Calibration.

Warning: See warnings on the title page of this manual.

5.2 Cell Replacement

The L-2 Micro-Fuel Cell is a sealed electrochemical transducer with no electrolyte to change or electrodes to clean. When the cell reaches the end of its useful life, it is replaced. The spent fuel cell should be discarded according to local regulations. This section describes fuel cell care as well as when and how to replace it.

5.2.1 Storing and Handling Replacement Cells

To have a replacement cell available when it is needed, it is recommended that one spare cell be purchased 9-10 months after commissioning the 3000TA, or shortly before the end of the cell's one year warranty period.

CAUTION: DO NOT STOCKPILE CELLS. THE WARRANTY

PERIOD STARTS ON THE DAY OF SHIPMENT.

The spare cell should be carefully stored in an area that is not subject to large variations in ambient temperature (75°F nominal, 24°C) or to rough handling.

WARNING: The sensor used in the model 3000TA Trace Oxygen

Analyzer uses electrolytes which contain toxic substances, mainly Lead and potassium hydroxide, that can be harmful if touched, swallowed, or

Inhaled. Avoid contact with any fluid or powder in or

Maintenance 3000TA- around the unit. What may appear to be plain water could contain one of these toxic substances. In case of eye contact, immediately flush eyes with water for at least 15 minutes. Call physician. (See appendix,

Material Safety Data Sheet.)

CAUTION: DO NOT DISTURB THE INTEGRITY OF THE CELL

PACKAGE UNTIL THE CELL IS TO ACTUALLY BE

USED. IF THE CELL PACKAGE IS PUNCTURED AND

AIR IS PERMITTED TO ENTER, THE CELL WILL

REQUIRE AN EXCESSIVELY LONG TIME TO REACH

ZERO AFTER INSTALLATION (1-2 WEEKS!).

5.2.2 When to Replace a Cell

The characteristics of the Micro-Fuel Cell show an almost constant output throughout its useful life and then fall off sharply towards zero at the end. Cell failure in the 3000TA is usually characterized inability to zero the instrument down to a satisfactorily low ppm reading. When this occurs, the 3000TA system alarm trips, and the LCD displays a failure message.

#.# ppm Anlz

CELL FAIL/ ZERO HIGH

Before replacing the cell: a) b)

Check your span gas to make sure it is within specifications.

Check for leaks downstream from the cell, where oxygen may be leaking into the system.

If there are no leaks and the span gas is OK, replace the cell.

5.2.3 Removing the Micro-Fuel Cell

The Micro-Fuel cell is located inside the stainless steel cell block behind the front panel (see Figure 5-1). To remove an existing cell:

1. Remove power to the instrument by unplugging the power cord at the power source.

2. Open the front panel door by pressing the release button on the top right corner of the door all the way in with a narrow gauge tool, such as a small screwdriver, and releasing it.

Trace Oxygen Analyzer Maintenance

3. With one hand placed underneath the cell block ready to catch the Micro-Fuel cell, lift up on the stainless steel gate in front of the cell block. This releases the cell and cell holder from the block. The cell and holder will fall out in your hand.

WARNING: Risk of electric shock high voltage exposed at the end of enclosure!

Figure 5-1: Removing the Micro-Fuel

5.2.4 Installing a New Micro-Fuel Cell

It is important to minimize the amount of time that a Teledyne

Trace Oxygen Sensor is exposed to air during the installation process.

The quicker the sensor can be installed into the unit, the faster your TAI

O

2

sensor will recover to low O

2

measurement levels.

Maintenance 3000TA-

CAUTION: DO NOT TOUCH THE SENSING SURFACE OF THE

CELL. IT IS COVERED WITH A DELICATE TEFLON

MEMBRANE THAT CAN LEAK WHEN PUNCTURED.

THE SENSOR MUST BE REPLACED IF THE

MEMBRANE IS DAMAGED.

Before installing a new cell, check the O-ring in the base of the cell holder. Replace if worn or damaged.

Place the cell on the holder with the screen side facing down.

Note: There is a small location hole drilled in the holder. This hole mates with a guide pin on the bottom rear of the cell block. The hole in the cell block holder must align with the guide pin on the cell block.

Step 1. Remove power from instrument.

Step 2. Remove the old sensor (if installed) from the analyzer.

Step 3. Purge the analyzer at approximately 1 SCFH flow rate with

N

2

(or applicable sample gas with the sensor holder removed).

Step 4. Remove sensor from double bag storage.

Step 5. Remove sensor shorting button.

Step 6. Place sensor on sensor holder so that the gold contact plate of the sensor is facing up towards the sky.

Step 7. Install sensor and sensor holder into cell block.

Step 8. With O-ring in place, align the guide pin with the hole on the cell holder. Then, with the holder, lift cell into the cell block.

Step 9. Push the gate on the cell block down so that the slots on the side of the gate engage the locating screws on the side of the block. This forces the holder into position and forms a gastight seal.

Step 10. Purge system with sample or zero gas.

Step 11. Power-up.

Note: If steps 4 through 10 are accomplished quickly (elapsed time less than 15 seconds), recovery to less than 1ppm level should occur in less than 8 hours.

Trace Oxygen Analyzer Maintenance

5.2.5 Cell Warranty

The Class L-2 Micro-Fuel cell is typically used in the Model

3000TA. Other cell options are available. See page iii for the cell shipped with your analyzer.

The L-2 cell is a long life cell and is warranted for 1 year from the date of shipment. Note any Addenda attached to the front of this manual for special information applying to your instrument.

With regard to spare cells, warranty period begins on the date of shipment. The customer should purchase only one spare cell (per section

5.2.1). Do not attempt to stockpile spare cells.

The L-2 cell is not designed applications where CO

2

is a major component in the sample, however concentrations of 1,000 ppm or less will not adversely affect the cell performance. Consult the factory for available options for either intermittent or continuous CO

2

exposure.

If a cell was working satisfactorily, but ceases to function before the warranty period expires, the customer will receive credit toward the purchase of a new cell.

If you have a warranty claim, you must return the cell in question to the factory for evaluation. If it is determined that failure is due to faulty workmanship or material, the cell will be replaced at no cost to you.

Note: Evidence of damage due to tampering or mishandling will render the cell warranty null and void.

5.3 Fuse Replacement

1. Place small screwdriver in notch, and pry cover off, as shown in Figure 5-2.

Figure 5-2: Removing Fuse Block from Housing

Maintenance 3000TA-

2. To change between American and European fuses, remove the single retaining screw, flip Fuse Block over 180 degrees, and replace screw.

3. Replace fuse as shown in Figure 5-3.

4. Reassemble Housing as shown in Figure 5-2.

American Fuses European Fuses

Figure 5-3: Installing Fuses

5.4 System Self Diagnostic Test

1. Press the System button to enter the system mode.

2. Use the ◄►arrow keys to move to More, and press Enter.

3. Use the ◄►arrow keys to move to Self-Test, and press

Enter . The following failure codes apply:

Table 5-1: Self Test Failure Codes

Power

0 OK

1 5 V Failure

2 15 V Failure

Analog

0 OK

1

2

DAC A (0–1 V Concentration)

DAC B (0–1 V Range ID)

Trace Oxygen Analyzer Maintenance

Preamp

0 OK

1 Zero too high

2 Amplifier output doesn't match test input

Failed

5.5 Major Internal Components

The Micro-Fuel cell is accessed by unlatching and swinging open the front panel, as described earlier. Other internal components are accessed by removing the rear panel and sliding out the entire chassis.

See Figure 5-4, below. The gas piping is illustrated in Figure 2-4, and the major electronic components locations are shown in Figure 2-5, in chapter 2.

WARNING: See warnings on the title page of this manual.

The 3000TA contains the following major components:

 Analysis Section

 Micro Fuel Cell (L-2 or other)

 Stainless steel cell block

Sample system

Power Supply

Microprocessor

Displays

 5 digit LED meter

 2 line, 20 character, alphanumeric, VFD display

RS-232 Communications Port.

See the drawings in the Drawings section in back of this manual for details.

Maintenance 3000TA-

Figure 5-4: Rear-Panel Screws

To detach the rear panel, remove only the 14 screws marked with an X.

5.6 Cleaning

If instrument is unmounted at time of cleaning, disconnect the instrument from the power source. Close and latch the front-panel access door. Clean outside surfaces with a soft cloth dampened slightly with plain clean water. DO NOT use any harsh solvents such as paint thinner or benzene.

For panel-mounted instruments, clean the front panel as prescribed in the above paragraph. DO NOT wipe front panel while the instrument is controlling your process.

5.7 Troubleshooting

Problem: Erratic readings of the Oxygen concentration as reported by the analyzer.

Possible Cause:

The analyzer may have been calibrated in an inaccurate fashion.

Solution:

Trace Oxygen Analyzer Maintenance

Turn the analyzer off, then back on again. Press the System key when prompted by the analyzer “Press System for default Values”. This will return the analyzer to its default settings in calibration and zero values.

If erratic behavior continues replace the sensor.

Possible Cause:

Atmospheric Oxygen may be diffusing in through the vent and affecting the oxygen level which the sensor sees.

Solution:

Increase flow rate and/or length or vent tubing in order to dilute or minimize the diffusion of oxygen from the vent back to the sensor.

Problem: Inaccurate zero operation (i.e. the user has zeroed the analyzer accidentally on gas much higher than one would normally use for a zero gas).

Solution:

Turn the analyzer off, then back on again. Press the System key when prompted by the analyzer "Press System for default Values". This will return the analyzer to its default settings in calibration and zero values.

Now proceed to carefully calibrate and zero the analyzer.

Maintenance 3000TA-

Trace Oxygen Analyzer Appendix

Appendix

A-1 Model 3000TA Specifications

Packaging: General Purpose

 Flush panel mount (Standard).

 Relay rack mount. Contains either one or two instruments in one 19" relay rack mountable plate

(Optional).

Sensor: Teledyne L-2 trace analysis

Micro-Fuel Cell. Cell Block: 316 stainless steel.

Sample System: All wetted parts of 316 stailess steel. 90 %

Response Time: 65 seconds at 25 °C (77

°F).

Ranges: Three user definable ranges from 0–10 ppm to 0–250,000 ppm, plus air calibration range 0- 250,000 ppm (25 %).

Autoranging with range ID output.

Alarms: One system-failure alarm contact to detect power failure or sensor-zero failure.

Two adjustable concentration threshold alarm contacts with fully programmable setpoints.

Displays: Two-line by 20-character, VFD screen, and one 5 digit LED display.

Power: Universal power supply 85-250 VAC, at

47-63 Hz, 0.9 A MAX.

Operating Temperature: 0-50 °C (32-122 °F)

Humidity: 99% Altitude: 1,609 m

Accuracy: ±2% of full scale at constant temperature.

±5% of full scale over operating

Appendix 3000TA- temperature range (except 0-10 ppm analysis range) once thermal equilibrium is reached.

±1 ppm on 0-10 ppm analysis range, once thermal equilibrium is reached.

0-1 VDC range ID

4-20 mA DC—isolated—percent-ofrange,

4-20 mA DC—isolated—range ID.

19 cm high, 24.9 cm wide, 31 cm deep

(6.96 in high, 8.7 in wide, 12.2 in deep).

A-2 Recommended 2-Year Spare Parts List

1

1

2

C62368-A

Back Board

Front Board

Trace Preamplifier Board

C73870-A

F 1296

Main Computer Board

Fuse, 2A, 250 V 5x20 mm Slow

Blow

1 O165

1* Micro-Fuel Cell

* See page iii for sensor used in this instrument

A minimum charge is applicable to spare parts orders.

Note: Orders for replacement parts should include the part number (if available) and the model and serial number of the instrument for which the parts are intended.

Trace Oxygen Analyzer

Orders should be sent to:

TELEDYNE Analytical Instruments

16830 Chestnut Street

City of Industry, CA 91749-1580

Phone (626) 934-1500, Fax (626) 961-2538

Web: www.teledyne-ai.com or your local representative.

Appendix

Appendix 3000TA-

A-3 Drawing List

D66316 Final Assembly/Outline Drawing

D77800 Final Assembly Trace Oxygen Analyzer 3000TA-EU

Series

A-4 19-inch Relay Rack Panel Mount

Figure A-1: Single and Dual 19" Rack Mounts

(dimensions in mm)

Trace Oxygen Analyzer Appendix

A.5 Application notes

3000 SERIES ANALYZERS

APPLICATION NOTES ON PRESSURES AND FLOW

RECOMMENDATIONS

The 3000 series analyzers require reasonably regulated sample pressures. While the 3000 analyzers are not sensitive to variations of incoming pressure provided they are properly vented to atmospheric pressure. The pressure must be maintained as to provide a useable flow rate trough the analyzer. Any line attached to sample vent should be 1/4 or larger in diameter.

FLOW RATE RECOMMENDATIONS:

A usable flow rate for a 3000 series analyzer is one which can be measured on the flowmeter. This is basically .2-2.4 SLPM. The optimum flow rate is 1 SLPM (mid scale).

Note: Response time is dependent on flow rate, a low flow rate will result in slow response to O

2 changes in the sample stream. The span flow rate should be the approximately same as the sample flow rate.

CELL PRESSURE CONCERNS:

The sensors used in 3000 series analyzers are optimized to function at atmospheric pressure. At pressures other than atmospheric the diffusion rate of O

2 will be different than optimum value. Higher pressures will produce faster O

2 diffusion rates resulting in higher O reading and shorter cell life. To use a 3000 series analyzer at a cell

2 pressure other than atmospheric, the analyzer must be calibrated with a known calibration gas at the new cell pressure to adjust for the different diffusion rate. Cell pressures below 2/3 atmospheric are not recommended because they tend to cause excessive internal expansion which may result in seal failure.

For operation at cell pressures other than atmospheric, care must be taken not to change the sample pressure rapidly or cell damage may occur. For cell pressures above atmospheric, caution must be exercised to avoid over pressuring the cell holder. (Percent analyzers will require some type of cell retainer to prevent the cell from being pushed out by

Appendix 3000TA- the pressure). For operation at pressures below atmospheric pressure a suffix C (clamped) cell is required.

RESTRICTION DEVICES:

For proper operation, all 3000 series analyzers require a flow restriction device. This device is typically a restrictor or a valve. This restriction device serves two functions in the sample path. The first function is to limit the flow rate of the sample through the analyzer. A restrictor is chosen to operate over a range of pressures and provide a useable flow rate over that range.

The second function that the restriction device provides is a pressure drop. This device is selected to provide the only significant pressure drop in the sample path.

RESTRICTOR KIT

The current revision of the 3000 series analyzers are supplied with a kit containing two restrictors and a union which are user installed.

These parts supplied to give the end user more flexibility when installing the analyzer. The restrictor kit is suitable for high and low positive pressure applications as well as vacuum service (atmospheric pressure sample) applications (see manual for installation instructions).

The standard restrictor (BLUE DOT) is recommended for pressures between 5 psig and 50 psig. For positive low pressure application (5 psig or less) the un-marked restrictor is better suited . For non-pressurized sample applications the marked restrictor should be used and configured for vacuum service.

For extremely low positive pressure applications (less than 2 psig) the vacuum service configuration should provide higher performance

(higher flow rates). For vacuum service the end user must supply a vacuum pump and a bypass valve for the pump. A vacuum level of 5 -

10 inches of mercury should provide the optimum flow rate.

CAUTION: FLOW RESTRICTORS HAVE VERY SMALL ORIFICES

AND MAY BE PLUGGED BY SMALL PARTICLES

(.005” DIA OR LARGER) A SAMPLE FILTER MUST

BE INCLUDED IN THE SAMPLE LINE PRIOR TO THE

RESTRICTOR! A 60 MICRON FILTER IS

RECOMMENDED.

3000TA EXAMPLES:

Trace Oxygen Analyzer Appendix

Example 1: With an incoming pressure of 10 psig the standard restrictor

(blue dot) will provide a flow rate of .76 SLPM. Up-stream of the restrictor the sample line pressure will be 10 psig, while down stream

(including the cell) the pressure will be at atmospheric pressure.

(analyzer vented to atmospheric pressure) Note, all other pressure drops in the sample path are insignificant at these flowrates. This insures that the cell operates at atmospheric pressure. At very high flow rates (off scale of flow-meter), pressure drops other than the restriction device could become significant, and result in pressurizing the cell.

Example 2: A 3000TA is configured for vacuum service as follows. The un-marked restrictor is placed in the sample vent port. The downstream end of the restrictor is then connected to a vacuum pump and bypass valve. The bypass valve is adjusted to provide a flow rate of 1 SLPM.

The sample pressure between the pump and the restrictor will be approximately -7 inches of mercury, while the pressure in the balance of the sample system including the cell will be approximately at atmospheric pressure. (Provided the sample flow into the analyzer is not blocked.)

BYPASS:

To improve the system response, a bypass can be added to increase the sample flow rate to the analyzer by a factor of ten. A bypass provides a sample flow path around the analyzer of 2-18 scfh, typically.

CALIBRATION GAS:

For 3000 series analyzers r with the Auto-Cal option, the customer must supply control valves (or restrictors) for any SPAN or ZERO gas source which is attached to the Auto-Cal ports. The valve should be adjusted to the same flow rate as the sample gas. When restrictors are used, the gas pressure must be adjusted to achieve the proper flow rate.

OPERATION WITHOUT A RESTRICTORDEVICE:

Operation without a restrictor device is not recommend as mentioned above. A 3000TA without any flow restrictor device was tested on 11-19-97. This results in a flow rate of 2.4 slpm at 1 psig. This is a cv of 0.023 for the standard sample system.

Appendix 3000TA-

REFERENCE: FLOW_1 .XLS & FLOW _2.XLS for information on flow rates at various pressures.

TAI PART NUMBERS

Restrictor Kit:

Union (ss)

LP. Restrictor

Std. Restrictor

A68729

U11

R2323

R2324

(low pressure /vac. service )

Blue Dot

Nut N73

Ferrule F73

Ferrule F74 Both ferrules are required

CONVERSIONS:

1 PSI

1 SCFH

=

=

2.04 INCHES OF MERCURY (in. Hg.)

0.476 SLPM

Trace Oxygen Analyzer Appendix

A-5 Material Safety Data Sheet

Section I - Product Identification

Product Name: Micro-fuel Cells

Mini-Micro-fuel Cells

Super Cell, all classes except T-5F

Electrochemical Oxygen Sensors, all classes

Manufacturer: Teledyne Electronic Technologies

Analytical Instruments

Address: 16380 Chestnut Street,

City of Industry, CA 91749

Phone: (626) 961-9221

Technical Support: (626) 934-1673

Environment, Health and

Safety:

(626) 934-1592

Date Prepared: 11/23/98

Section II - Physical and Chemical Data

Chemical and Common

Names:

CAS Number:

Potassium Hydroxide (KOH), 15% (w/v)

Lead (Pb), pure

KOH 1310-58-3

Pb 7439-92-1

Melting Point/Range:

KOH (15% w/v)

-10 to 0 °C

Pb (pure)

328 °C

Boiling Point/Range:

Specific Gravity: pH:

Solubility in Water:

Percent Volatiles by Vol.:

Appearance and Odor:

100 to 115 °C

1.09 @ 20 °C

1744 °C

11.34

>14 N/A

Completely soluble Insoluble

None N/A

Colorless, odorless solution

Grey metal, odorless

Appendix 3000TA-

Section III -Physical Hazards

Potential for fire and explosion: The electrolyte in the Micro-fuel Cells is not flammable. There are no fire or explosion hazards associated with Micro-fuel

Cells.

Potential for reactivity: The sensors are stable under normal conditions of use. Avoid contact between the sensor electrolyte and strong acids.

Section IV - Health Hazard Data

Primary route of entry: Ingestion, eye/skin contact

Exposure limits: OSHA PEL:

ACGIH TLV:

0.05 mg./cu.m. (Pb)

2 mg/ cu.m. (KOH)

Effects of overexposure

Ingestion: The electrolyte could be harmful or fatal if swallowed.

Oral LD50 (RAT) = 3650 mg/kg

Eye: The electrolyte is corrosive; eye contact could result in permanent loss of vision.

Dermal: The electrolyte is corrosive; skin contact could result in a chemical burn.

Inhalation: Liquid inhalation is unlikely.

Signs/symptoms of exposure: Contact with skin or eyes will cause a burning sensation and/or feel soapy or slippery to touch.

Medical conditions aggravated by exposure: None

Carcinogenicity: NTP Annual Report on Carcinogens: Not listed

LARC Monographs: Not listed

OSHA: Not listed

Other health hazards: Lead is listed as a chemical known to the State of California to cause birth defects or other reproductive harm.

Trace Oxygen Analyzer Appendix

Section V - Emergency and First Aid Procedures

Eye Contact: Flush eyes with water for at least 15 minutes and get immediate medical attention.

Skin Contact: Wash affected area with plenty of water and remove contaminated clothing. If burning persists, seek medical attention.

Ingestion: Give plenty of cold water. Do not induce vomiting. Seek medical attention. Do not administer liquids to an unconscious person.

Inhalation: Liquid inhalation is unlikely.

Section VI - Handling Information

NOTE: The oxygen sensors are sealed, and under normal circumstances, the contents of the sensors do not present a health hazard. The following information is given as a guide in the event that a cell leaks.

Protective clothing: Rubber gloves, chemical splash goggles.

Clean-up procedures: Wipe down the area several times with a wet paper towel. Use a fresh towel each time.

Protective measures during cell replacement:

Before opening the bag containing the sensor cell, check the sensor cell for leakage. If the sensor cell leaks, do not open the bag. If there is liquid around the cell while in the instrument, put on gloves and eye protection before removing the cell.

Disposal: Should be in accordance with all applicable state, local and federal regulations.

NOTE: The above information is derived from the MSDS provided by the manufacturer. The information is believed to be correct but does not purport to be all inclusive and shall be used only as a guide.

Teledyne Analytical Instruments shall not be held liable for any damage resulting from handling or from contact with the above product.