Cistermiser easyflush wave, easyflush walkaway electronic flush valve Installation guide
Below you will find brief information for electronic flush valve easyflush wave, electronic flush valve easyflush walkaway. This document provides instructions for installing the easyflush wave and easyflush walkaway electronic flush valves. It covers topics such as sensor positioning, power connections, and testing and commissioning. It also includes a troubleshooting section to help you resolve common problems.
Advertisement
Advertisement
easyflush
installation guide
1
Introduction the easyflush valve provides electronic flushing and filling of the WC.
There are two versions to suit different applications: easyflush wave is operated by the user bringing their hand towards the infrared sensor. The dual flush is standard.
A part flush is triggered by a quick hand pass (under 1 second) and a full flush is triggered by a slightly longer hand pass (over 1 second).
easyflush walkaway is activated when the user stands up or leaves the cubicle, but it can also be activated by the user bringing their hand toward the sensor; this function can be disabled by the installer. The dual flush is standard. A part flush is triggered by short user occupancy (under 45 seconds) and a full flush by long user occupancy (over 45 seconds).
0118 969 1611 | [email protected] | www.cistermiser.co.uk
supplied parts
1
Side inlet
2
Bottom inlet
3
Outlet
4
Inlet washer and nut
5
Stem washer with boss
6
Stem washer
7
Stem nut
8
Cap seal
9
Cap washer
10 Cap nut
11 Valve unit
12 Float
13 Inlet hose
14 Sensor unit
15 Sensor gasket
16 Sensor nut
17 Hole adaptor
18 Clamp plate
19 Screw packer
20 Usage label
21 Mains adaptor*
22 Screws*
23 Battery case*
24 Sticky pad
25 Flush pipe**
1
2
3
4
5
optional extras
Infrared Configuration Unit (ICU)
Multi Product Power Supply Unit (PSUC)
Side Front Adaptor (SFA)
6
7
8
9
10
11
12
13 21
14
22
15
16
17
18
19
20
WC Flushing
1 sec.
2 sec.
23
24
25 x 4 x 4
* Where supplied
** When pre-assembled in a cistern
0118 969 1611 | [email protected] | www.cistermiser.co.uk
2
Valve operation
Water supply feed
Outlet drop valve
Inlet float valve
Float
Plunger
Discharge
3
Sensor positioning
cistern mounted sensor
(WAVE only)
Top Front panel mounted sensor
(WAVE only)
300 – 400mm
WC
Side
NOTE:
Do not position sensor near toilet roll holder or disabled bar.
WC panel mounted sensor
(WALKAWAY only)
WC
Exposed installation Concealed installation
NOTE:
Reflective surfaces and materials such as hi-vis jackets may cause the sensor to activate when not expected.
0118 969 1611 | [email protected] | www.cistermiser.co.uk
Concealed installation
4
Installation schematic
C x 4 or x 4
B
A
X
Drill hole to the required diameter.
See
B for guidance
X
– Walkaway version only.
If this distance is less than 12cm it is recommended that the
Hand Activation is disabled. Refer to section 10.
Pipework diagram suggestion only
NOTE:
For chemical water treatment.
If the water system has been treated with chemical dosing, ensure the system is thoroughly flushed before fitting any
Cistermiser products. Concentrated chemicals in dead legs can damage the product and result in failure. If the water is treated with Chlorine Dioxide (ClO2), ensure concentration levels are maintained below 5ppm.
0118 969 1611 | [email protected] | www.cistermiser.co.uk
A
Valve installation
Side inlet Bottom inlet Connect hose
Fit float
Inlet hose
Fit outlet
Stem nut
NOTE:
For extra narrow cisterns fit outlet & valve pre-fitted together.
Latch valve
0118 969 1611 | [email protected] | www.cistermiser.co.uk
B
Panel mounted sensor
Thin wall
3 – 15mm
Medium thickness wall
15 – 30mm
Thick wall
30mm+
Ø38mm
Sensor nut (
16)
Ø44mm
Sensor nut (
16)
6mm packer
(See note*) (
19)
M4 screw
(
22)
Ø38mm
Sensor gasket (
15)
Apply sanitary silicone sealant
Sensor gasket (
15)
Clamp plate
(reversible) (
18)
Apply sanitary silicone sealant
NOTE:
*If wall thickness is under 35mm use packer to prevent screw causing damage to sensor
Apply sanitary silicone sealant
C
Cistern mounted sensor – wave version only
Sensor fixing Mounting options
Sensor fastening nut
Sensor gasket
NOTE:
Use hole adaptor for holes between 40 and 50mm. A hole diameter larger than 50mm is a non standard installation.
Side/front adaptor
NOTE:
Side/front adaptor required.
(Not supplied with standard product but available as an extra. Contact Cistermiser for further details or order on-line at www.cistermiser.co.uk).
0118 969 1611 | [email protected] | www.cistermiser.co.uk
5
Power connections
NOTE:
Only one power source should be connected at any time, either mains or batteries.
It is not possible to connect both simultaneously.
If using the battery case, open it and fit 4 Alkaline
AA (LR-6) batteries as indicated. Reseal and fasten into place on panel using screws (22) or with the sticky pads (24). If used in an exposed installation the battery case can be hung inside the cistern by hooking it over the edge.
If using the mains adaptor, fasten into place on panel in a dry location using screws (22) or with the sticky pads (24) and connect the unterminated mains cable to a 50Hz 230V AC single phase supply via a 1A fused spur (not supplied).
electrical connection
As shown below, connect the spade connectors from the sensor unit to the solenoid prongs taking care to connect the wires according to the colour coding on the label. If these are not long enough they can be extended up to 1 metre.
Connect the power connector from the sensor to the mains adaptor, or in the case of a battery powered installation, the battery case.
NOTE:
Remove the lens protection label from the sensor BEFORE connecting to the power.
When the power is first connected the LED in the sensor flashes amber. This is normal and lasts only a few seconds.
Colour conventions are brown for positive and grey for negative.
Connections
Sensor valve unit Power connector
Power options
Battery case requires 4 x AA alkaline batteries
Mains adaptor requires 1A fused spur
Multi product power supply unit ((Not supplied, contact Cistermiser for further details).
Suitable for powering up to 20 units
0118 969 1611 | [email protected] | www.cistermiser.co.uk
6
Testing and commissioning
Restore the water supply and check all joints and connections for leaks. The cistern will begin filling until the float rises sufficiently to close the water inlet valve.
Check the water level to ensure the cistern fills to the required full flush volume – typically six litres is sufficient – adjust the position of the float if necessary to allow more or less water in for a full flush.
testing operation refill duration setting
It may be necessary for it to be increased or decreased depending on the supply water pressure. The refill duration should be matched to the time it takes the cistern to complete its refill.
Follow the Advanced Configuration instructions in section 9.
NOTE:
During installation it is not uncommon for an unexpected flush to occur as a result of a sensor being unintentionally activated during the refill time delay. To check that operation is correct it is important to take the refill time into account.
the system
When installing the easyflush for the first time air may be trapped in the valve and supply line.
This will cause the valve to fluctuate repeatedly between the open and closed positions. To expel air from the system activate the valve by placing a hand in front of the sensor whilst the valve is in this repeating cycle. It may be necessary to do this several times before the action ceases and the valve operates normally.
wave version:
Test the installation by placing a hand within 10cm of the sensor for 1 second.
The easyflush should enter a part flush, lasting
2 seconds (configurable, see section 9 Advanced setting guide). Place a hand within 10cm of the sensor for more than 2 seconds to activate a full flush, lasting 8 seconds. After operating, the easyflush will not flush again until the cistern is full, but the easyflush does ‘remember’ attempts to flush. This is called a memory flush.
walkaway version:
Same as above to operate wave or sit on WC for 6 seconds to activate a part flush, lasting 2 seconds and for more than 45 seconds to activate a full flush, lasting 8 seconds.
explanation of the memory flush
(rapid green flashing)
If a second flush is requested whilst the cistern is refilling, the easyflush will not flush immediately but will ‘remember’ the request. The sensor will show a rapid green flashing during this time. After the refill duration (see below) the easyflush will flush, completing either a part or full flush as requested.
additional configuration
Additional configuration is possible with an Infrared
Configuration Unit (ICU) remote control.
(Not supplied; contact Cistermiser). See section 10 for further details.
NOTE:
For grey water/rain water harvesting.
Ensure adequate filtering is fitted, a 10µm filter is recommended.
NOTE:
For chemical water treatment.
If the water system has been treated with chemical dosing, ensure the system is thoroughly flushed before fitting any
Cistermiser products. Concentrated chemicals in dead legs can damage the product and result in failure. If the water is treated with
Chlorine Dioxide (ClO2), ensure concentration levels are maintained below 5ppm.
0118 969 1611 | [email protected] | www.cistermiser.co.uk
7
Usage advice and specification
Minimum working pressure:
Maximum working pressure:
0.5 bar dynamic
6 bar static
NOTE:
The easyflush is a flush and fill type valve. It uses the incoming water pressure to lift the flush mechanism. At pressures below 0.5 bar dynamic, the valve will not perform as intended.
Back siphonage protection:
Class 4 type double check valve – integral in inlet.
factory settings
Range (EF Wave):
Range (EF Walkaway):
Refill duration after part flush:
Refill duration after full flush:
Part flush duration:
Full flush duration:
Power requirements:
Normal battery life:
~18cm.
~65cm (~9cm for hand activation).
48 seconds.
80 seconds.
2 seconds.
8 seconds.
Either 6V from 4 x alkaline AA (LR6) batteries or 6V DC regulated from mains adaptor. (1A fused spur required).
2 years under typical usage conditions. A single red flashing of the LED indicates that the batteries are low and need to be replaced.
Clean with soap and water only.
Infrared lens can be polished with a soft cloth.
Cleaning:
Lens care:
electronic specification
Control classification:
Maximum load:
Rated temperature range:
Action classification:
Pollution classification:
Ingress protection:
Independent.
7.5W 1.25A (6VDC) EMC emissions tested at load.
0-40 ̊C.
Type 1.Y.
Degree 1.
IP65.
0118 969 1611 | [email protected] | www.cistermiser.co.uk
8
Component dimensions
84mm 212mm 93mm
Ø44mm or Ø50mm please specify flush pipe size
48mm 27mm
0118 969 1611 | [email protected] | www.cistermiser.co.uk
9
Advanced settings guide easyflush walkaway
Carry out only if settings need to be changed
1
Disconnect power, wait for 5 seconds and reconnect.
2
When flashing amber.
Place hand 4-6cm from sensor until constant green
, then remove hand.
3
You are now in configuration mode
If a hand is not placed over the sensor, it will go into normal operation mode.
Sensing range of adjustment
A Wait for single red flashing
LED to enter sensing range mode.
B Briefly place your hand in front of the sensor until the green LED flashes slowly
. Range configuration mode will then be ready.
C Stand at the distance from the sensor you would like to set at maximum range. Wait 8 seconds until constant green
LED shows.
Sensor distance is now set.
A
C
Refill duration adjustment
Wait for double red flashing
LED to enter refill time.
x 2
B During double red flashing
, briefly place your hand in front of the sensor. The valve will flush and the cistern will refill.
A green flashing
LED will be seen.
Once the cistern has refilled completely and the water has ceased to flow, move your hand back into the line of sight of the sensor. The refill time has now been set.
Part flush duration adjustment
A
C
Wait for flashing triple red
LED to enter part flush time mode.
x 3
B During the triple red flashing briefly place your hand in front of the sensor. The valve will start to flush and a green flashing
LED will be seen. HOLD HAND STEADY.
When the valve has flushed for the desired part flush time, move your hand out of the line of sight of the sensor. The water will cease to flow and the part flush time will be set.
4 After configuration the unit will re-enter configuration mode.
flash amber
, giving opportunity to
0118 969 1611 | [email protected] | www.cistermiser.co.uk
easyflush wave
Carry out only if settings need to be changed
1
Disconnect power, wait for 5 seconds and reconnect.
2
When flashing amber.
Place hand 4-6cm from sensor until constant green
, then remove hand.
3
You are now in configuration mode
If a hand is not placed over the sensor, it will go into normal operation mode.
Sensing range of adjustment
A Wait for single red flashing
LED to enter sensing range mode.
B Briefly place your hand in front of the sensor until the green LED flashes slowly
. Range configuration mode will then be ready.
C Move your hand to the distance from the sensor you would like to set as maximum range. Wait 8 seconds until constant green
LED shows.
Sensor distance is now set.
Refill duration adjustment
A
C
Wait for double red flashing
LED to enter refill time.
x 2
B During double red flashing briefly place your hand in front of the sensor. The valve will flush and the cistern will refill.
A green flashing
LED will be seen.
Once the cistern has refilled completely and the water has ceased to flow, move your hand back into the line of sight of the sensor. The refill time has now been set.
Part flush duration adjustment
A
C
Wait for flashing triple red
LED to enter part flush time mode.
x 3
B During the triple red flashing briefly place your hand in front of the sensor. The valve will start to flush and a green flashing
LED will be seen. HOLD HAND STEADY.
When the valve has flushed for the desired part flush time, move your hand out of the line of sight of the sensor. The water will cease to flow and the part flush time will be set.
4 After configuration the unit will flash amber giving opportunity to re-enter configuration mode.
0118 969 1611 | [email protected] | www.cistermiser.co.uk
10
Infrared Configuration Unit (ICU) guide
NOTE:
Not supplied but available from Cistermiser or any major plumbing merchant.
1
2
3
4
5
6
7
8
9
button descriptions
Activates cleaning mode
Activates ICU configuration mode
Decreases setting
Increases setting
Checks the setting being altered
Saves changes and exits ICU configuration mode
Quits ICU configuration mode without saving changes
Configures sensor range
Configures part flush time
Configures cistern refill time
Toggles Wave function ON/OFF (Walkaway ONLY)
12-hour hygiene cycle activation
Dual flush activation
Autorange setting of sensor range
Siphonic trap refill activation
Resets to default factory settings
entering configuration mode
Point the ICU towards the easyflush sensor and push the configuration button. Activation is most effective when the configuration button is held down as the ICU is brought close to the sensor.
It can take up to 3 seconds for the product to sense the ICU. The easyflush will return to normal operation if there are no button presses for 30 seconds.
configuring sensor range
Point the ICU at the easyflush sensor and press the
1 sensor range button (the sensor blinks green).
Decrease or increase the sensor range by pressing the and buttons respectively.
The sensor blinks red when the min or max value is reached.
Press the button to check the sensor range setting - the sensor displays the current setting by flashing green; see table.
For Wave version
Number of flashes 1
Range (cm approx) 6
2
9
3
11
4
15
5
17
For Walkaway version
Number of flashes 1 2 3 4 5
Range (cm approx) 45 50 53 56 58
Save setting and exit ICU configuration mode by pressing the button.
Exit without saving by pressing the button.
0118 969 1611 | [email protected] | www.cistermiser.co.uk
configuring part flush time
Point the ICU at the easyflush sensor and press the
2 flush time button (the sensor blinks green).
Decrease or increase the part flush time by pressing the and buttons respectively. The sensor blinks red when the min or max value is reached.
Press the button to check the part flush time setting - the sensor displays the current setting by flashing green; see table.
Number of flashes 1
Part flush time (sec) 1
2
2
3
3
4
4
Save setting and exit ICU configuration mode by pressing the button.
Exit without saving by pressing the button.
5
5
NOTE:
The full flush volume is set by setting the float height in the cistern.
configuring refill time
Point the ICU at the easyflush sensor and press the
3 refill time button (the sensor blinks green).
Decrease or increase the refill time by pressing the and buttons respectively. The sensor blinks red when the min or max value is reached.
Press the button to check the refill time setting - the sensor displays the current setting by flashing green; see table.
Number of flashes 1 2 3 4 5 6 7
Refill time (sec) 0 20 40 60 80 100 120
Save setting and exit ICU configuration mode by pressing the button.
Exit without saving by pressing the button.
activating and de-activating the hand wave flushing (walkaway version only)
Point the ICU at the easyflush sensor and press the 4 (the sensor blinks green). By default the hand wave function is on.
Pressing the and button switches the hand wave function on or off respectively.
Press the button to check the setting - the sensor flashes green once if function is off or twice if it is on.
Save setting and exit ICU configuration mode by pressing the button.
Exit without saving by pressing the button.
activating the 12-hour hygiene flush cycle
Point the ICU at the easyflush sensor and press the hygiene cycle button (the sensor blinks green).
Pressing the and buttons switches the hygiene flush function on or off respectively.
Press the button to check the setting - the sensor flashes green once if function is off or twice if it is on.
Save setting and exit ICU configuration mode by pressing the button.
Exit without saving by pressing the button.
activating the dual flush function
Point the ICU at the easyflush sensor and press the 6 (the sensor blinks green).
Pressing the and buttons switches the dual flush function on or off respectively. Press the button to check the setting - the sensor flashes green once if function is off or twice if it is on.
Save setting and exit ICU configuration mode by pressing the button.
Exit without saving by pressing the button.
configuring sensor range using the autorange function
If the cubicle door is opposite the sensor, ensure that the cubicle door is closed or ajar, but not fully open.
Point the ICU at the easyflush sensor and press the 7 button.
Immediately stand clear of the sensor. Sensor blinks green for 5 seconds, then a steady green when setting complete. The sensor measures the background reflections and sets the sensor range to an appropriate setting.
Save setting and exit ICU configuration mode by pressing the button.
Exit without saving by pressing the button.
0118 969 1611 | [email protected] | www.cistermiser.co.uk
11
Frequently asked questions indicators for normal sensor function after user activation
Wave:
• single green flash once a second (if part flush)
• double green flash once a second (if full flush)
• constant rapid green flash (if re-flush has been requested)
Walkaway:
• 3 seconds between green flashes
(if a presence is detected)
• single green flash once a second (if part flush)
• double green flash once a second (if full flush)
• constant rapid green flash (if re-flush has been requested)
no water is entering the cistern
No obvious indicator
The sensor is covered
Ensure that the water supply is reaching the valve. Minimum working pressure 0.5 bar dynamic, maximum 6.0 bar static.
Air in inlet hose (normally found with low pressure). Ensure loop is facing downwards.
Blockage. Ensure the filters on the inlet and outlet side of the valve are clear.
Anti vandal mode has been triggered. The object/debris needs to be removed from the sensor and the valve will resume normal functionality.
the valve is not working at all
The sensor light does not flash
Ensure the power supply is connected. If mains power is being when a hand is placed in front used through the mains adaptor check that the mains adaptor is of it.
working by reverting to the battery pack. Remove the mains adaptor when using batteries.
Sensor is flashing red slowly or
Low or no battery power; change batteries. If operated by mains not at all when hand is placed power, check wiring then contact Cistermiser.
in front of it.
No obvious indicator
Ensure there is a good water supply and pressure of 0.5 to 6 bar.
continuous flow into the pan from the cistern
No obvious indicator
No obvious indicator
Check the seal of the flush valve and clean away any debris or scale.
Water flowing through the overflow; ensure that the water entering the cistern is being shut off by adjusting the float accordingly. If the float is set too high, water will continuously flow into the cistern and out of the overflow.
Sensor is flashing red slowly or
Low or no battery power; change batteries. If operated by mains not at all when hand is placed power, check wiring then contact Cistermiser.
in front of it.
0118 969 1611 | [email protected] | www.cistermiser.co.uk
11
Frequently asked questions
WC flushes when in use
Working as normal otherwise
Ensure that the sensor is mounted in the correct position. Refer to installation schematic (section 3).
Ensure the sensor range is correctly configured. Refer to the advanced setting guide (section 9).
other issues
Shuddering noise
Double red flash
Red flash once every second
Air in the system. Flush the valve repeatedly to purge the system.
Faulty wiring. Check wiring carefully.
Faulty solenoid. Call Cistermiser for advice.
Low or no battery power. Change batteries. If operated by mains power, check wiring then contact Cistermiser.
Red flash twice every 3 seconds
Sensor covered or heavily scratched. Uncover or polish out scratches.
cistermiser product warranty and extended warranty
Cistermiser products are guaranteed for
12 months from the date of manufacture.
The guarantee is for faulty products and parts only: there is no labour warranty. If you believe your product is faulty, please either contact
Cistermiser directly on 0118 969 1611 or at [email protected], with a photograph and the serial number, to diagnose the cause of the problem.
The warranty on Cistermiser products can be extended, within one year of date of manufacture, at no cost to three years from the date of installation by completing the enclosed warranty card or at www.cistermiser.co.uk/warranty.
Please make a note of the serial number and take a photograph of the installation before you leave site.
0118 969 1611 | [email protected] | www.cistermiser.co.uk
EF2/17_C-03
DDA

Public link updated
The public link to your chat has been updated.
Advertisement
Key features
- Electronic flushing and filling
- Dual flush functionality
- Infrared sensor activation
- Walkaway activation
- Adjustable sensor range and flush duration
- Battery or mains power operation
- Easy installation and commissioning
- Hygiene flush cycle