Fluke 60 Series Handheld Infrared Thermometers

Fluke 60 Series Handheld Infrared Thermometers
Thermometers
Fluke 60 Series Handheld Infrared Thermometers
Specifications
Backlit display for use in
poorly lit areas
Fluke 61
Fluke 65
Fluke 62
Fluke 63
Fluke 66
Fluke 68
Yes
Yes
Yes
Yes
Yes
Yes
Holds temperature
readings
Fluke 61
Fluke 65
Fluke 62
Fluke 63
Fluke 66
Fluke 68
Yes
Yes
Yes
Yes
Yes
Yes
Range
Fluke 61
Fluke 65
Fluke 62
Fluke 63
Fluke 66
Fluke 68
-18 to 275
-40 to 500
-30 to 500
-32 to 545
-32 to 600
-32 to 760
MIN/MAX/AVG
Fluke 61
Fluke 65
Fluke 62
Fluke 63
Fluke 66
Fluke 68
N/A
MIN/MAX
MAX
MAX
All
All
Datalogging
Fluke 61
Fluke 65
Fluke 62
Fluke 63
Fluke 66
Fluke 68
N/A
N/A
N/A
N/A
Yes
Yes
Distance to spot ratio
Fluke 61
Fluke 65
Fluke 62
Fluke 63
Fluke 66
Fluke 68
8:1
8:1
10:1 at 80% energy
12:1
30:1
50:1
Hi-Lo Alarm
Fluke 61 N/A
ºC (-0 to 525 ºF)
ºC (-40 to 932 ºF)
ºC (-20º to 932 ºF)
ºC (-25 to 999 ºF)
ºC (-25 to 1100 ºF)
ºC (-25 to 1400 ºF)
Fluke 65
Fluke 62
Fluke 63
Fluke 66
Fluke 68
N/A
N/A
Yes
Yes
Yes
Optional RTD probe
(contact temperature
measurement)
Fluke 61
Fluke 65
Fluke 62
Fluke 63
Fluke 66
Fluke 68
N/A
N/A
N/A
N/A
Yes
Yes
Emissivity
Fluke 61
Fluke 65
Fluke 62
Fluke 63
Fluke 66
Fluke 68
Fixed at 0.95
Fixed at 0.95
Fixed at 0.95
Fixed at 0.95
Digitally adjustable (from 0.1 to 1.0 by 0.01)
Digitally adjustable (from 0.1 to 1.0 by 0.01)
Measurement Accuracy
Fluke 61
± 3 °C (5 °F)
-18 °C (0 °F) to -1 °C (30 °F)
above -1 °C (30 °F) to 275 °C (525 ± 2% of reading or 2 °C (3.5 °F)
°F)
(With the instrument stable between 23-25
°C)
Fluke 65
Above 0 ºC (32 ºF) : ± 2 ºC (± 4 ºF)
Below 0 ºC (32 ºF) : ± 5 ºC (± 9 ºF)
(With instrument stable at 23ºC ± 3 ºC)
Fluke 62
10 ºC to 30 ºC (50 ºF to 86 ºF): ± 1 ºC (2 ºF)
Outside 10 ºC to 30 ºC (50 ºF to ± 1.5 ºC (3 ºF) or ± 1.5 % of reading,
whichever is greater
86 ºF):
Fluke 63, 66, 68
-32 to -26 ºC (-25
to -15 ºF):
-26 to -18 ºC (-15
to 0 ºF):
-18 to 23 ºC (-0 to
73 ºF):
±3 ºC (±5 ºF)
±2.5 ºC (±4 ºF)
±2 ºC (±3 ºF)
For targets above 23 ºC (73 ºF): ±1% of reading or ±1 ºC
(±2 ºF), whichever is greater
© 1995 - 2005 Fluke Corporation