Spot Vital Signs® LXi Service Manual

Spot Vital Signs® LXi Service Manual

Welch Allyn Spot Vital Signs LXi

BP

SYS mm g

DIA mm g

TEMP

1 2 7

73

9

8 5

4 WEST

F

AM

Reading

01

SpO2

%

/min PULSE

62

Patient ID 0 2 4

Weight

Height

Respiration

Pain

Clear

Send/Next Reading

Enter

Service Manual

BP

DIA mm g

TEMP

1 2 7

73

9

8 5

SpO2

Patient ID 0 2 4

Weight

Height

Respiration

%

F

4 WEST

AM

Reading

01

/min PULSE

62

Pain

Clear

Send/Next Reading

Enter

A

Un

Welch Allyn Spot Vital Signs LXi

Service Manual

iv

Welch Allyn Spot Vital Signs LXi

Copyright 2012 Welch Allyn. All rights are reserved. To support the intended use of the product described in this publication, the purchaser of the product is permitted to copy this publication, for internal distribution only, from the media provided by Welch Allyn. No other use, reproduction, or distribution of this publication, or any part of it, is permitted without written permission from Welch Allyn.Welch Allyn assumes no responsibility for any injury to anyone, or for any illegal or improper use of the product, that may result from failure to use this product in accordance with the instructions, cautions, warnings, or statement of intended use published in this manual.

Welch Allyn, Spot Vital Signs, SureBP Technology, and SureTemp are registered trademarks of Welch

Allyn.

Braun ThermoScan is a registered trademark of the Braun Company.

LNCS is a trademark of, and SET, LNOP, and Masimo are registered trademarks of, Masimo Corporation.

Possession or purchase of a Masimo SpO

2

-equipped device does not convey any express or implied license to use the device with unauthorized sensors or cables which would, alone or in combination with this device, fall within the scope of one or more of the patents relating to this device.

Nellcor and Oxi-Max are registered trademarks of Nellcor Puritan Bennett Inc.

Health o meter is a registered trademark of Sunbeam Products, Inc. used under license.

Software in this product is Copyright Welch Allyn or its vendors. All rights are reserved. The software is protected by United States of America copyright laws and international treaty provisions applicable worldwide. Under such laws, the licensee is entitled to use the copy of the software incorporated with this instrument as intended in the operation of the product in which it is embedded. The software may not be copied, decompiled, reverse-engineered, disassembled, or otherwise reduced to human-perceivable form. This is not a sale of the software or any copy of the software; all right, title, and ownership of the software remain with Welch Allyn or its vendors.

For information about any Welch Allyn product, call Welch Allyn Technical Support:

USA +1 800 289 2501

Canada +1 800 561 8797

European Call Center +353 46 90 67790

Germany +49 695 098 5132

Latin America +1 305 669 9003

Singapore +65 6419 8100

United Kingdom +44 207 365 6780

Reorder #4500-89

Manual Material No. 718440 Ver. A

Australia +61 2 9638 3000

China +86 21 6327 9631

France +33 15 5695 849

Japan +81 3 6383 0852

Netherlands +31 202 061 360

South Africa +27 11 777 7555

Sweden +46 85 853 6551

Welch Allyn

4341 State Street Road

Skaneateles Falls, NY

13153 USA

Regulatory Affairs Representative

Welch Allyn Limited

Navan Business Park

Dublin Road

Navan, County Meath, Republic of Ireland

Printed in USA

Service Manual

1 - Safety Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1

Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1

Warnings and Cautions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1

General Warnings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1

Blood Pressure Warnings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3

Temperature Warnings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

SpO

2

Warnings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

General Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

Blood Pressure Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

Temperature Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

SpO

2

Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

Electrostatic Discharge (ESD) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

Safety Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

Button Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

Connection Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

Agency Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

2 - Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11

Purpose and Scope. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11

Other Applicable Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11

Miscellaneous Mounting Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12

Wall Mount Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12

Mobile Stand Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14

Radio Mounting Accessory. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16

Barcode Reader Mounting Accessory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17

Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18

Spot Vital Signs LXi . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18

Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18

Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

User Serviceable Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

Technical Assistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

Field Replacement Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

Service Loaners . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

Service Intervals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

Spot Vital Signs LXi Configurations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20

Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20

Display Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22

Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24

Braun ThermoScan PRO 4000 Lock. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25

Blood Pressure Hose and Cuff . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25

Thermometer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26

SpO

2

Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27

Radio . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27

Quick Reference Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27

AC Power Transformer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28

Printer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28

Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29

v

vi

Welch Allyn Spot Vital Signs LXi

3 - Functional Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31

Self Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31

Internal Configuration Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32

4 - Spot Vital Signs LXi Technical Overview . . . . . . . . . . . . . . . . . . . . 37

System Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37

Battery System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37

Battery Charger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37

“CPU I/O” Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37

Standby Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38

CPU Core Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38

+5V Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38

Clock/Calendar Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38

Mod F NIBP Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38

SpO

2 and Thermometer Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38

LCD Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38

External Load (RS-232) Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39

SureTemp Plus Thermometer Interface. . . . . . . . . . . . . . . . . . . . . . . . . . . . 39

Braun ThermoScan PRO 4000 Thermometer Interface . . . . . . . . . . . . . . . . 39

SpO

2

Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39

Radio Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40

Bar Code Scanner Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40

Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40

Printer Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41

Interconnect Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42

5 - Maintenance and Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43

Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43

Spot Vital Signs LXi . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43

Blood Pressure Cuff . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43

Blood Pressure Hose and Cable. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43

SureTemp Plus Thermometer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44

Braun ThermoScan PRO 4000 Thermometer . . . . . . . . . . . . . . . . . . . . . . . 44

SpO

2

Sensors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45

Radio . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45

Printer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45

Battery Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46

Spot Vital Signs LXi . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46

Braun ThermoScan PRO 4000 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46

Printer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47

Spot Vital Signs LXi Firmware Upgrade CD . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48

Spot Vital Signs LXi Radio Configuration CD . . . . . . . . . . . . . . . . . . . . . . . . . . . 48

Product Disposal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48

6 - Calibration and Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49

Terms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49

Required Test Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50

Pneumatic test assembly setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50

Power Supply Test Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52

Service Manual

vii

Battery Removal / Power Shutdown . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53

Voltage Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55

Pressure verification test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57

Blood pressure calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62

NIBP board Initialization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62

Functional tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67

Temperature Functional Check. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77

Temperature test setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77

Changing the Calibration Set Point Temperature on the 9600 Plus Calibration

Tester . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78

Welch Allyn SureTemp Plus Thermometers. . . . . . . . . . . . . . . . . . . . . . . . . . . . 78

Braun ThermoScan Pro 4000 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82

Changing the Calibration Set Point Temperature . . . . . . . . . . . . . . . . . . . . . 85

7 - Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89

Performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89

Blood Pressure Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89

Temperature Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90

SpO

2

Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91

Masimo Sensor Accuracy Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91

Masimo Patents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91

Nellcor Sensor Accuracy Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92

Nellcor Patents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92

Mechanical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93

Electrical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93

Patents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93

Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93

Environmental. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93

Radio. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94

Mechanical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94

Configuration Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94

Guidance and Manufacturer’s Declaration For Spot Vital Signs LXi Radio . . . . . 95

8 - Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99

Error Codes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99

Radio Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 110

Service Work Checklist. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111

Spot LXi Radio Configuration Tool . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 112

9 - Supplies and Accessories. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113

Blood Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113

Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114

Pulse Oximetry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115

Masimo Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115

Nellcor Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117

Miscellaneous. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 118

Service Contracts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119

viii

Welch Allyn Spot Vital Signs LXi

10 - Field Replaceable Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121

11 - Disassembly and Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 127

Tool list . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 127

Key Pad Disassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132

SpO

2

Circuit Board Disassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 133

Masimo . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 133

Nellcor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 134

Blood Pressure Circuit Board Disassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . 136

LCD Inverter Ballast Board Disassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 137

LCD Disassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 137

Pump Disassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 138

RS232-Communication Circuit Board Disassembly . . . . . . . . . . . . . . . . . . . . . 138

Fan and Power Circuit Board Disassembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . 139

Thermometry Circuit Board Disassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 139

SureTemp Plus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 139

Braun ThermoScan PRO 4000 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 141

12 - Repair Test Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143

General Unit Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143

A-D Noise Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143

Leak Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143

Inflation Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143

Dump Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143

Pneumatic Accuracy Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143

Valve Control Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 144

Voltage Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 144

Blank Mode Current Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 144

Back Light (Idle) Current Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 144

Valve/Pump Mode Current Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 144

Interface Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 144

Temperature Option Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 144

Accuracy Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 144

Temperature Probe Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145

SpO

2

Option Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145

SpO

2

Functional Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145

SpO

2

Mode Current Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145

Fail Safe Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145

Over Pressure Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145

Over 15 mmHg . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145

Interconnect Diagram 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 147

Interconnect Diagram 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 148

1

Safety Summary

Introduction

All users of the Spot Vital Signs LXi must read and understand the safety summary, and all additional specific warnings and cautions located throughout the documentation.

Warnings and Cautions

Familiarize all operating personnel with the general safety information in this summary.

Specific warnings and cautions are also found throughout this manual.

General Warnings

A warning statement in this manual identifies a condition or practice, which if not corrected or discontinued immediately, could lead to patient injury, illness, or death.

These warnings pertain entirely to Spot Vital Signs LXi.

WARNING The information in this manual is a comprehensive guide to the operation of Spot Vital Signs LXi. For best results, read this manual thoroughly before using the device.

WARNING Spot Vital Signs LXi is designed for medical clinician use. Although this manual may illustrate medical spot-check techniques, only a trained clinician who knows how to take and interpret a patient’s vital signs should use this device.

WARNING Spot Vital Signs LXi is not intended for use in environments that are without health care practitioner supervision.

WARNING Spot Vital Signs LXi is not intended for continuous monitoring. Do

not leave the device unattended while taking measurements on a patient.

WARNING To ensure data integrity, save readings and clear the Spot Vital Signs

LXi display between patients.

WARNING The Spot Vital Signs LXi is not defibrillator proof.

WARNING Spot Vital Signs LXi is not intended for use during patient transport.

WARNING This device is not suitable for use in the presence of a flammable anesthetic mixture with air or oxygen or nitrous oxide. An explosion may result.

1

2

Safety Summary Welch Allyn Spot Vital Signs LXi

WARNING To ensure patient safety, use only accessories and supplies (i.e., cuffs, hoses, temperature probes, SpO

2 sensors, etc.) recommended for or supplied with Spot Vital Signs LXi. Using unapproved accessories with Spot Vital

Signs LXi can affect patient and/or operator safety.

WARNING Prevent fluid from entering any connectors on the device. Should this occur, dry the connectors with warm air. Check the accuracy of all operating functions.

WARNING Every three months, inspect the blood pressure cuff, SpO

2 and other accessories for fraying or other damage. Replace as necessary.

cable,

WARNING Do not use Spot Vital Signs LXi on patients who are on heart/lung machines.

WARNING Electric shock hazard. There are no user-serviceable parts inside Spot

Vital Signs LXi other than battery replacement (see “Battery Replacement” on page 46). Only an individual specifically trained and approved for the repair and/or

verification of this specific Welch Allyn product may perform maintenance procedures specifically described in this manual. For service, refer the device to an Authorized Service Center.

WARNING This device is not intended for hand-held use during operation.

WARNING Do not autoclave.

WARNING This device complies with current required standards for electromagnetic interference and should not present problems to other equipment or be affected by other devices. As a precaution, avoid using this device in close proximity to other equipment.

WARNING Welch Allyn is not responsible for the integrity of any mounting installation. Welch Allyn recommends that the customer contact their Biomedical

Engineering Department or maintenance service to ensure professional installation for safety and reliability of any mounting accessory.

WARNING The Spot Vital Signs LXi consists of high-quality precision parts.

Protect it from severe impact and shock. A qualified service technician must check any Spot Vital Signs LXi that is dropped or damaged for proper operation prior to further use. Do not use the Spot Vital Signs LXi if you notice any signs of damage. Contact the Welch Allyn Technical Support Department for assistance.

WARNING Do not use an SpO

2 finger clip sensor and a blood pressure cuff simultaneously on the same limb. Doing so may result in inaccurate pulse rate and perfusion readings.

WARNING All signal input and output (I/O) connectors are intended for connection of only devices complying with IEC 60601-1, or other IEC standards

(for example, IEC 60950) as appropriate to the device. Connecting additional devices to the Spot LXi may increase leakage currents. To maintain operator and patient safety, it is necessary to consider the requirements of IEC 60601-1-1.

WARNING For proper patient electrical isolation, use only a Welch Allyn power supply (4500-101A) to charge Spot Vital Signs LXi.

WARNING Do not use the manufacturer supplied charger for the weight scale while it is attached to Spot LXi.

Service Manual Safety Summary

3

Blood Pressure Warnings

These warnings pertain to the Spot Vital Signs LXi blood pressure feature.

WARNING Spot Vital Signs LXi is not intended to measure BLOOD PRESSURE on neonatal patients. The AAMI SP10:2002 standard defines neonates as children

28 days or less of age if born at term (37 weeks gestation or more); otherwise up to 44 gestational weeks.

WARNING To ensure pediatric blood pressure accuracy and safety, the Child

Reusable Two-Piece Blood Pressure Cuff (4500-01), Infant Durable One-Piece

Cuff (5082-82-4MQ), and the Infant Disposable One-Piece Cuff (5082-92-4MQ) are the smallest cuffs approved for use with young children and infants. The child’s arm must fit within the range markings on the cuff.

WARNING Avoid compression of the blood pressure hose or cuff tubing of Spot

Vital Signs LXi. This may cause system errors to occur in the device.

WARNING Patients who are experiencing moderate to severe arrhythmias may give inaccurate blood pressure measurements.

WARNING Spot Vital Signs LXi does not operate effectively on patients who are experiencing convulsions or tremors.

WARNING Use only Welch Allyn blood pressure cuffs and/or hoses. Using other manufacturers’ blood pressure cuffs and/or hoses may produce inaccurate blood pressure readings.

WARNING When several blood pressure measurements are taken on the same patient, regularly check the cuff site and extremity for possible ischemia, purpura, and/or neuropathy.

WARNING Do not place the cuff on any extremity that is used for intravenous infusions or any area where circulation is compromised.

WARNING Excessive cuff tightness may cause venous congestion and discoloration of the limb.

WARNING Wrapping the cuff too loosely (preventing proper inflation) may result in errors.

WARNING Do not change the connector(s) on the blood pressure cuff tubing of this device to luer type. Luer type connectors are commonly used in intravenous infusion systems. Using the luer connectors on blood pressure cuff tubing creates the risk that the blood pressure tubing could be mistakenly connected to a patient's intravenous line, resulting in the introduction of air into the patient's circulatory system.

4

Safety Summary Welch Allyn Spot Vital Signs LXi

Temperature Warnings

These warnings pertain to the Spot Vital Signs LXi temperature feature.

SureTemp

®

Plus

These warnings are specific to the SureTemp Plus thermometer option.

WARNING Use only Welch Allyn probe covers. Using other manufacturers’ probe covers or no probe cover may produce temperature measurement errors and/or inaccuracy.

WARNING Always use a probe cover whenever coming into contact with a patient.

WARNING Long-term continuous monitoring beyond three to five minutes is not recommended in any mode.

WARNING Oral/axillary probes (blue ejection button at top of probe) and blue removable probe wells are used for taking oral and axillary temperatures only.

Rectal probes (red ejection button) and red removable probe wells are used for taking rectal temperatures only. Use of the probe at the wrong site will result in temperature errors. Use of the incorrect removable probe well could result in patient cross-contamination.

WARNING The thermometer connectors and probe are not waterproof. Do not immerse or drip fluids on these items. Should this occur, dry the connectors and probe with warm air. Check all functions for proper operation and accuracy.

WARNING Do not take an axillary temperature through patient’s clothing. Direct probe cover to skin contact is required.

WARNING Do not autoclave.

WARNING Use Welch Allyn single-use disposable probe covers to limit patient cross-contamination.

WARNING Incorrect insertion of probe can cause bowel perforation.

WARNING Washing hands greatly reduces the risk of cross-contamination and nosocomial infection.

WARNING To ensure optimal accuracy, always confirm that the correct mode is selected.

Service Manual Safety Summary

5

Braun Thermoscan PRO 4000

These warnings are specific to the Braun ThermoScan PRO 4000 thermometer option.

WARNING Keep the probe window clean, dry, and undamaged at all times to ensure accurate measurements. To protect the probe window, always keep the thermometer in the storage cover while transporting or when not in use.

WARNING Only use Braun ThermoScan probe covers with this thermometer.

Using other manufacturer’s probe covers or no probe cover may produce temperature measurement errors and/or inaccuracies. If the thermometer is used

without a probe cover attached, clean the lens (see “Braun ThermoScan PRO

4000 Thermometer” on page 44).

WARNING Do not autoclave.

WARNING The thermometer is not waterproof. Do not immerse or drip fluids on it. Should this occur, dry the thermometer with warm air. Check for proper operation and accuracy.

SpO

2

Warnings

These warnings pertain to the Spot Vital Signs LXi SpO

2 feature.

WARNING Only use Spot Vital Signs LXi with Masimo or Nellcor SpO

2 option with Masimo or Nellcor brand sensors and accessories, respectively. Using the wrong or unapproved sensors or cables may cause improper performance.

WARNING The SpO

2 sensor and extension cables are intended for use only for pulse oximetry measurements. Do not attempt to connect these cables to a PC or any similar device.

WARNING Before using, carefully read the sensor Directions for Use, including all warnings, cautions, and instructions.

WARNING Do not use a damaged sensor or pulse oximetry cable or a sensor with exposed optical components.

WARNING Incorrect application or a long duration of use of an SpO

2 sensor may cause tissue damage. Inspect the sensor site periodically as directed in the sensor’s Directions for Use.

WARNING Certain ambient environmental conditions, sensor application errors, and certain patient conditions may affect SpO

2 readings and pulse signal.

WARNING Do not immerse the sensor or patient cables in water, solvents, or cleaning solutions (the sensors and connections are not waterproof). Do not use irradiation, steam, or ethylene oxide for sterilization.

WARNING The SpO

2 as an apnea monitor.

in the Spot Vital Signs LXi device is not intended for use

WARNING Consider the SpO

2 an early warning device. As a trend toward patient deoxygenation is indicated, use laboratory instruments to analyze blood samples to completely understand the patient’s condition.

6

Safety Summary Welch Allyn Spot Vital Signs LXi

WARNING Tissue damage can be caused by incorrect application or duration of use of a Nellcor OxiMax sensor. Inspect the sensor site as directed in the sensor

Directions for Use.

WARNING Do not use the sensors during magnetic resonance imaging (MRI) scanning. Induced current could potentially cause burns. The MS board pulse oximeter may affect the MRI image, and the MRI unit may affect the accuracy of the oximetry measurements.

WARNING Carefully route patient cabling to reduce the possibility of patient entanglement or strangulation.

WARNING Failure to cover the Nellcor OxiMax sensor site with opaque material in high ambient light conditions may result in inaccurate measurements.

General Cautions

A caution statement in this manual identifies a condition or practice, which if not corrected or discontinued immediately, could lead to equipment failure, equipment damage, or data loss.

These cautions pertain to the entire Spot Vital Signs LXi device.

Caution If the accuracy of any measurement is in question, check the patient’s vital sign(s) with an alternate method and then check to verify the device is functioning properly.

Caution Place the device on a secure surface or use one of the optional mounting accessories.

Caution Do not place fluids on or near the device.

Caution It is recommended that the device is used within stated operating

temperature ranges (see “Environmental” on page 93). The device will not meet

its performance specifications if used outside these temperatures ranges.

Caution Always unplug the AC power transformer from the outlet before moving the mobile stand to a new location.

Caution The basket has a three-pound weight limit. Take care not to exceed this limit.

Caution When connecting a Health o meter

® weight scale, only operate the scale with battery power. Remove and discard the scale's AC power cord.

Caution When using the Health o meter scale, remove the two hex nuts on the

RS-232 cable, as supplied, before screwing the cable into the scale base.

Service Manual Safety Summary

7

Blood Pressure Cautions

These cautions pertain to the Spot Vital Signs LXi blood pressure feature.

Caution Minimize extremity and cuff motion during blood pressure readings.

Caution If the blood pressure cuff is not at heart level, note the difference in reading due to the hydrostatic effect. Add the value of 1.80 mmHg (.2 kPa) to the displayed reading for every inch (2.5 cm) above heart level. Subtract the value of

1.80 mmHg (.2 kPa) from the displayed reading for every inch (2.5 cm) below heart level.

Caution Proper blood pressure cuff size and placement is essential to the accuracy of the blood pressure determination. See Reusable Two-Piece Cuff

Measurements or Durable One-Piece Cuff Measurements of the Directions for

Use for sizing information.

Caution The position and physiologic condition of the subject can affect a blood pressure reading.

Temperature Cautions

These cautions pertain to the Spot Vital Signs LXi temperature feature.

Caution The SureTemp Plus feature only operates with the probe well in place.

Caution Do not use alkaline batteries in the Braun ThermoScan PRO 4000.

Caution Biting the probe tip may result in damage to the probe.

SpO

2

Cautions

These cautions pertain to the Spot Vital Signs LXi SpO

2 feature.

Caution The pulse oximeter is calibrated to determine the percentage of arterial oxygen saturation of functional hemoglobin. Significant levels of dysfunctional hemoglobin such as carboxyhemoglobin or methemoglobin may affect the accuracy of the measurement.

Caution Some sensors may not be appropriate for a particular patient. If at least 10 seconds of perfusion pulses cannot be observed for a given sensor, change sensor location or sensor type for perfusion to resume.

Caution Physiological conditions, medical procedures, or external agents that may interfere with the pulse oximeter’s ability to detect and display measurements include dysfunctional hemoglobin, arterial dyes, low perfusion, dark pigment, and externally applied coloring agents such as nail polish, dye, or pigmented cream.

Caution When selecting a sensor, consider the patient’s weight and activity level, the adequacy of perfusion, the available sensor sites, the need for sterility, and the anticipated duration of monitoring.

8

Safety Summary

Electrostatic Discharge (ESD)

Welch Allyn Spot Vital Signs LXi

Electrostatic discharge is a sudden current flowing from a charged object to another object or to ground. Electrostatic charges can accumulate on common items such as foam drinking cups, cellophane tape, synthetic clothing, untreated foam packaging material, and untreated plastic bags and work folders, to name only a few.

Electronic components and assemblies, if not properly protected against ESD, can be permanently damaged or destroyed when near or in contact with electrostatically charged objects. When you handle components or assemblies that are not in protective bags and you are not sure whether they are static-sensitive, assume that they are static-sensitive and handle them accordingly.

• Perform all service procedures in a static-protected environment. Always use techniques and equipment designed to protect personnel and equipment from electrostatic discharge.

• Remove static-sensitive components and assemblies from their static-shielding bags only at static-safe workstations - a properly grounded table and grounded floor mat and only when you are wearing a grounded wrist strap (with a resistor of at least 1 megohm in series) or other grounding device.

• Use only grounded tools when inserting, adjusting, or removing static-sensitive components and assemblies.

• Remove or insert static-sensitive components and assemblies only with monitor power turned off.

• Insert and seal static-sensitive components and assemblies into their original staticshielding bags before removing them from static-protected areas.

Always test your ground strap, bench mat, conductive work surface, and ground cord before removing components and assemblies from their protective bags and before beginning any disassembly or assembly procedures.

Service Manual Safety Summary

9

Symbols

The following symbols are associated with the Spot Vital Signs LXi.

Safety Symbols

Button Symbols

Identifies information within the manual to avoid injury.

Caution: consult accompanying documents

Handle with Care

MAX

Storage Humidity

Class II Equipment

Type BF Equipment

Recycle the product separate from

other disposables, see “Product

Disposal” on page 48.

Mode of Operation: Continuous

IPXØ

Pb

Identifies information within the manual to avoid equipment failure.

Internally Powered, Lead Acid

Battery

Transport Temperature

Recycle

Equipment is not protected against the ingress of liquid.

On/Off

Non-ionizing radiation (RF transmitter)

DC Power In

Navigation Buttons Select

MEM

Memory Blood Pressure

Power On/Off

Connection Symbols

USB Connection Serial Port Connection

10

Safety Summary

Agency Symbols

Welch Allyn Spot Vital Signs LXi

0297

CONFORMS TO:

UL STD 60601-1

IEC 60601-1

The CE mark on this product indicates that it has been tested to and conforms with the provisions noted within the 93/42/EEC Medical Device Directive.

Regulatory Affairs Representative

Welch Allyn Limited

Navan Business Park

Dublin Road

Navan, County Meath, Republic of Ireland

2

Overview

Purpose and Scope

The Spot Vital Signs LXi Service Manual is intended as a reference for maintenance and repair to the field replaceable unit (FRU) level and are listed on

page 121 .

Note

Opening the Spot Vital Signs device case requires recalibration of specific functions. This includes any internal components that were replaced. Calibration requires the use of the Welch Allyn Spot LXi Tool.

This manual provides the technical qualified service person with troubleshooting information, repair procedures, and calibration and performance verification instructions. A technical overview of the Spot Vital Signs LXi subsystems is provided as an introduction to the device’s circuitry and pneumatics.

This manual is intended for the technical qualified service personnel. Service training classes on Welch Allyn’s products are available. Contact Welch Allyn Technical Support for information.

Other Applicable Documents

The Spot Vital Signs LXi Directions for Use manual is also available. Refer to this document for information other than maintenance and repair.

Welch Allyn 9600 Plus Calibration Tester Directions for Use - for all models.

Braun ThermoScan PRO 4000 User’s Guide - for models 450E0, 45NE0, 45ME0.

Masimo Directions for Use - for models 45MT0, 45ME0.

Nellcor Directions for Use - for models 45NT0, 45NE0.

11

12

Overview

Miscellaneous Mounting Accessories

Wall Mount Kit

WALL MOUNT KIT

EF: 4701-62

aterial #: 711890 Rev. B

1

3

Welch Allyn Spot Vital Signs LXi

2

Attach bracket to wall before assembling basket.

}

4

5

Welch Allyn is NOT responsible for the integrity of any wall mounting interface.

Ensure you are using the appropriate hardware to mount the device to a surface.

Welch Allyn recommends that the customer contact their Biomedical

Engineering Department or maintenance service to ensure professional installation, safety, and reliability of any mounting accessory.

Service Manual Overview

13

14

Overview

Mobile Stand Kit

Welch Allyn Spot Vital Signs LXi

Service Manual Overview

15

16

Overview

Radio Mounting Accessory

A

1

2

5

4

3

B

Welch Allyn Spot Vital Signs LXi

Service Manual

Barcode Reader Mounting Accessory

Overview

17

18

Overview Welch Allyn Spot Vital Signs LXi

Warranty

Spot Vital Signs LXi

Welch Allyn warrants that the Spot Vital Signs LXi Radio products meet the labeled specifications of the products and will be free from defects in materials and workmanship that occur within one year after the date of purchase.

The date of purchase is: 1) the date specified in our records if you purchased the Product directly from us, 2) the date specified in the warranty registration card that we ask you to send to us, or 3) the date of purchase of product from the authorized Welch Allyn distributor as documented from a receipt from said distributor.

This warranty does not cover damage caused by: 1) handling during shipping, 2) use or maintenance contrary to labeled instructions, 3) alteration or repair by anyone not authorized by Welch Allyn, and 4) accidents.

If a Product or accessory covered by this warranty is determined to be defective because of defective materials, components, or workmanship, and the warranty claim is made within the warranty period described above, Welch Allyn will, at its discretion, repair or replace the defective Product or accessory free of charge.

You must obtain a service notification number from Welch Allyn to return your Product before you send it to Welch Allyn’s designated service center for repair. Contact Welch

Allyn Technical Support.

THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED,

INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY

AND FITNESS FOR A PARTICULAR PURPOSE. WELCH ALLYN’S OBLIGATION UNDER

THIS WARRANTY IS LIMITED TO REPAIR OR REPLACEMENT OF PRODUCTS

CONTAINING A DEFECT. WELCH ALLYN IS NOT RESPONSIBLE FOR ANY INDIRECT OR

CONSEQUENTIAL DAMAGES RESULTING FROM A PRODUCT DEFECT COVERED BY

THE WARRANTY.

Accessories

Refer to the manufacturer’s Directions for Use for the Masimo and Nellcor finger sensor and cable warranty.

Refer to the Directions for Use provided with the Welch Allyn Blood Pressure Cuff for warranty information.

The SureTemp Plus probe is covered by a one-year warranty and the SureTemp Plus probe well is covered by a 90-day warranty against original defects in material and workmanship.

Probe covers are intended for single-use only.

The Braun ThermoScan PRO 4000 is covered by a three-year warranty against original defects in material or workmanship.

The printer is covered by a one-year warranty against original defects in material or workmanship.

The barcode scanner is covered by a five-year warranty against original defects in material or workmanship.

Service Manual

Service

Overview

19

Caution Unauthorized repairs will void the warranty.

All repairs on products under warranty must be completed by Welch Allyn or Welch Allyn certified service personnel in accordance with warranty guidelines. See the Welch Allyn

Partners In Care

™ web site ( http://www.welchallyn.com/promotions/services/default.htm

) for a description of the different levels of services offered.

User Serviceable Parts

For user serviceable parts available see “Supplies and Accessories” on page 113.

Technical Assistance

If you have an equipment problem that you cannot resolve, call the Welch Allyn Service

Center nearest you for assistance. Technical service telephone support is available on normal business days.

If you are advised to return a product to Welch Allyn for repair or routine maintenance, schedule the repair with the service center nearest you.

Before returning a product for repair, you must obtain authorization from Welch Allyn.

Service personnel will give you a Service Notification number. Please note this number on the outside of your shipping box. Returns without a Service Notification number will not be accepted for delivery.

Field Replacement Units

See “Field Replaceable Units” on page 121.

Service Loaners

Service loaners or exchanges may be provided depending on the specific Partners In

Care

™ agreement in place for the device. Loaners for products repaired while under the original warranty, may be provided free of charge and are shipped within 48 hours of notification of need.

For service repairs outside of warranty or outside of service agreement are available for a nominal charge and shipment is subject to availability. Loaners are shipped pre-paid; however, this charge is added to the service charges.

Service Intervals

Calibrate the Spot Vital Signs LXi on an annual basis for blood pressure, temperature, and

SpO2 accuracy.

20

Overview Welch Allyn Spot Vital Signs LXi

Spot Vital Signs LXi Configurations

REF

450T0

450E0

45MT0

45ME0

45NT0

45NE0

Table 1. Available Versions of Spot Vital Signs LXi

Description

SureBP with SureTemp Plus Thermometer

SureBP with Braun ThermoScan PRO 4000 Thermometer

SureBP with Masimo SpO

2 and SureTemp Plus Thermometer

SureBP with Masimo SpO

2 and Braun ThermoScan PRO 4000 Thermometer

SureBP with Nellcor SpO

2 and SureTemp Plus Thermometer

SureBP with Nellcor SpO

2 and Braun ThermoScan PRO 4000 Thermometer

Controls

Figure 1. Spot Vital Signs LXi Front Panel with SureTemp Plus Thermometer

BP

SYS mm g

DIA mm g

TEMP

1 2 7

73

9 8 5

SpO2

Patient ID 0 2 4

Weight

Height

Respiration

%

4 WEST

F

AM

Reading

01

PULSE /min

62

Pain

Clear

Send/Next Reading

Enter

SureTemp Plus thermometer:

remove the probe from the probe well to take a temperature.

Removable probe

well: store the temperature probe here when not in use; remove to clean or replace to eliminate cross-contamination.

Navigation button: move throughout the options in the Navigation Window or to increase/decrease parameter units.

Probe cover storage area:

storage space for one box of probe covers.

Select button:

accepts the current option.

Power button: controls power to the device.

Charging LED: indicates when device is plugged in and charging the battery.

Blood Pressure Start/Stop button:

initiates a new blood pressure cycle.

Pressing again aborts an active blood pressure measurement.

Memory button:

recalls up to 50 most recent readings.

Service Manual Overview

21

Figure 2. Spot Vital Signs LXi Front Panel with Braun ThermoScan PRO 4000 Thermometer

Braun ThermoScan PRO 4000

thermometer: store the probe covers toward the back of the thermometer housing; remove the thermometer to access the covers.

DIA mm g

BP

SYS mm g

1

TEMP

2 7

73

9 8

5

SpO2

%

Patient ID 0 2 4

Weight

Height

Respiration

F

4 WEST

AM

Reading

01

PULSE /min

62

Pain

Clear

Send/Next Reading

Enter

A

Navigation button: move throughout the options in the Navigation Window or to increase/decrease parameter units.

Memory button: recalls up to 50 most recent measurements.

Charging LED: indicates the thermometer battery is charging.

Charging LED: indicates when device is plugged in and charging the battery.

Power button: controls power to the device.

Select button:

accepts the current option.

Blood Pressure Start/Stop button:

initiates a new blood pressure cycle.

Pressing again aborts an active blood pressure measurement.

22

Overview Welch Allyn Spot Vital Signs LXi

Display Window

Note

Before using Spot Vital Signs LXi for the first time, you must program it using the initial configuration screen.

1.

Press the Power button. The display window shows the initial configuration screen.

Figure 3. Initial Configuration Screen

Change Local Defaults Menu

Model No.: 4 NTO

Serial No. 200 040004

Language:

BP Units:

Temperature Units: eight Units:

Weight Units:

Date Format:

Date:

Time Format:

Time:

Select

2.

Use the Select button to access the options and accept the entries, and use the

Navigation buttons to move through the menu.

3.

The word “Exit” appears at the bottom of the list after you have programmed all items in the menu. You must program all items before you can start to use the device.

4.

Scroll to Exit and press the Select button to save the entries.

Service Manual Overview

23

The liquid crystal display (LCD) may indicate any of the following: systolic blood pressure

(mmHg or kPa), diastolic blood pressure (mmHg or kPa), MAP (mmHg or kPa), temperature (

F or C), temperature mode, pulse rate, pulse signal level, SpO

2 percent, department location, date, time, record number, height (in or cm), weight (lb or kg), respiration rate, pain level, connectivity signal strength, and battery charge level.

Figure 4. Display Window

X

Connectivity indicator: shows the status of Spot LXi Spot LXi if the device is sending a wireless reading.

Reading sent: indicates a successful wireless send reading attempt.

Reading not sent: indicates an unsuccessful wireless send reading attempt.

Location identifier: shows user programmed location information. If no location is entered, this area displays the date.

Systolic and Diastolic

display: if MAP is turned on, the screen toggles between the systolic and diastolic values, and the word “MAP” and the MAP value.

Temperature display and

indicator: shows the temperature in Fahrenheit or

Celsius.

SpO

2

display: shows the percent saturation of arterial hemoglobin. For devices with

SpO

2 only.

Navigation Window: shows menu options, error messages, or temperature conversion.

BP

SYS mm g

DIA mm g

TEMP

1 2 7

73

9 8 5

SpO2

Patient ID 0 2 4

Weight

Height

Respiration

%

F

4 WEST

AM

Reading

01

PULSE /min

62

Pain

Clear

Send/Next Reading

Enter

Pulse Signal Bar Graph: shows the strength of the pulses detected.

Pulse display: shows the pulse rate.

Clock: shows the current time.

Reading number indicator:

shows how many readings are stored in memory. Spot Vital

Signs LXi retains up to 50 readings in memory.

Battery level indicator: shows the battery charge level.

Thermometer probe setting

indicators: shows temperature mode. For devices with SureTemp

Plus thermometer only.

Thermometer probe setting

indicators: shows Monitor

Mode. For devices with SureTemp

Plus thermometer only.

Out-of-range indicator: shows the patient’s temperature reading above or below the measurement range limits. For devices with

SureTemp Plus thermometer only.

Clinician ID icons:

The figure icon signifies that Spot

Vital Signs LXi requires the clinician to enter the Clinician ID.

The check mark signifies Spot Vital

Signs LXi has recorded the Clinician

ID.

For devices with a bar code scanner only.

24

Overview Welch Allyn Spot Vital Signs LXi

Connections

Use the following instructions to connect the blood pressure hose, thermometer probe, and optional attachments to the Spot Vital Signs LXi.

Figure 5. Spot Vital Signs LXi Side and Rear Panel Connections

Icon ID label

CTIONS ort

MAT: 703956

SureTemp Plus thermometer connection port (for units with SureTemp Plus only)

Port I for external device connection

USB connection for PC

(behind plug)

Port II for external device connection

Battery door

I

II

Mobile stand or wall mount screw connection

SpO

2 cable connection port

(for units with SpO

2 only)

Blood pressure hose connection port

DC power connection port

Service Manual Overview

25

Braun ThermoScan PRO 4000 Lock

Press the lock tab toward the Braun ThermoScan PRO 4000 thermometer until it clicks. To release the housing, insert the lock release pin into the lock release hole until the lock tab snaps back.

Figure 6. Spot Vital Signs LXi with Braun ThermoScan PRO 4000 Lock

Lock release pin

Lock tab

Lock release hole

Blood Pressure Hose and Cuff

Identify and have available the Spot Vital Signs LXi, blood pressure cuff, and the blood pressure hose.

1.

Inspect the blood pressure hose; notice that one end has a single, gray connector fitting and the other end has two white fittings.

2.

Squeeze the side tabs on the gray connector and completely push the blood pressure hose connector into the blood pressure hose connection port until it clicks into place

(see Figure 5 on page 24).

3.

Twist the white connectors on the blood pressure hose and cuff connectors together.

26

Overview Welch Allyn Spot Vital Signs LXi

Thermometer

Spot Vital Signs LXi is available with either the SureTemp Plus thermometer or the Braun

ThermoScan PRO 4000 thermometer.

SureTemp Plus

SureTemp Plus is available with two probes and matching wells; one for oral/axillary temperatures (blue ejection button and probe well) and one for rectal temperatures (red ejection button and probe well).

WARNING Always use a probe cover whenever coming into contact with a patient.

Caution The SureTemp Plus feature only operates with the probe well in place.

1.

Align the probe well with the tabs facing up and down into the round opening of the

SureTemp Plus housing on the right side of Spot Vital Signs LXi. Push it into place.

2.

Align the temperature probe connector with the SureTemp Plus thermometer connection port on the back of the Spot Vital Signs LXi (see Figure 5 on page 24). You can only insert the connector into the port one way.

3.

Press the tab on the connector and push it until it clicks into place.

4.

Insert the temperature probe into the probe well.

Braun ThermoScan PRO 4000

Caution Do not use alkaline batteries in the Braun ThermoScan PRO 4000 thermometer. Use the rechargeable battery pack supplied by Welch Allyn.

1.

Open the package of rechargeable batteries and follow the installation guide provided.

2.

Open the box of probe covers as directed on the box and slide the box into the metal guides toward the back of the thermometer housing with the opening at the top and perforation facing forward.

3.

Hold the Braun thermometer at a 45° angle then insert the probe and the top of the thermometer into the housing.

4.

Lower the bottom portion of the thermometer into the housing until it snaps into place. If you do not properly seat the thermometer, it could fall out of the holder and become damaged.

5.

Slide the thermometer housing into the thermometer slot on the right side of the

Spot Vital Signs LXi device.

6.

Push the lock tab forward to prevent the thermometer housing from falling out of

Spot Vital Signs LXi (see

Figure 6 on page 25).

To release the lock, insert the lock release pin into the lock release hole.

Service Manual Overview

27

SpO

2

Sensor

Spot Vital Signs LXi is available with a wide variety of SpO

2 reusable finger clip sensor.

sensors and ships with a

1.

Align the shape and pin configuration of the extension cable connector to the SpO

2 cable connection port on the left side of the Spot Vital Signs LXi device.

2.

Push the connector firmly into the SpO

2 into place (see Figure 5 on page 24).

cable connection port until you hear it click

3.

Align the opposite end of the extension cable to the sensor cable connector and firmly push them together.

WARNING Use only Masimo or Nellcor SpO2 sensors and accessories with the

Spot Vital Signs LXi with Masimo or Nellcor configurations, respectively. Using the wrong or unapproved sensors or cables may cause improper performance.

Radio

CTIO NS ort

MAT: 703956

I

II

The Spot Vital Signs LXi identifies which serial port to use after activating the Radio in the internal configuration mode.

Welch Allyn offers two different radios:

• 802.11b radio and

• 802.11a/b/g radio

Only use Welch Allyn

Radios

COMM

LINK

POWER

Quick Reference Card

Attach the Quick Reference Card to the Spot Vital Signs LXi handle, mobile stand, or wall mount using the supplied plastic cable tie.

28

Overview Welch Allyn Spot Vital Signs LXi

AC Power Transformer

Note

To assure proper electrical isolation use only the Welch Allyn SPOT Vital Signs AC power transformer/charger.

The operator can use the Spot Vital Signs LXi with AC or battery power (after charging the battery).

1.

Insert the round transformer connector into the AC power connection port on the back of the Spot Vital Signs LXi (see Figure 5 on page 24).

2.

Insert the line cord into the line connector on the transformer then plug the power cord on the transformer into the AC main power source to charge the battery.

Printer

Once the printer is enabled in the Internal Configuration Mode, you can use plain or label paper. If you use label paper, you must clean the printer heads after every 10 rolls. A head cleaner kit is provided with each package of label paper available from Welch Allyn.

To connect the printer to the Spot LXi:

1.

Connect the the large end of the cable (405210) to the Spot LXi.

2.

Connect the small end of the cable (405210) to the Spot LXi printer.

Connect to Spot

LXi

Connect to Spot

LXi printer

To charge the Spot LXi printer:

1.

Remove the guard on the power supply body (713703 or BC1300M) and attach the appropriate power supply plug/adapter.

Service Manual Overview

29

2.

Connect the Spot LXi printer power supply to the Spot LXi printer.

3.

Charge the Spot LXi printer for at least 9 hours before initial use.

To enable the Spot LXi printer:

1.

Connect the Spot LXi printer to the Spot LXi.

2.

Enable the printer in the Internal Configuration Mode of the Spot LXi ( see “Internal

Configuration Mode” on page 32

).

Battery

Charge the Spot LXi battery and printer battery (if using) for 9 hours before initial use.

Charge the device an additional hour if it includes a Braun ThermoScan PRO 4000 thermometer.

Note

While charging the Spot LXi, it is acceptable to charge the Spot LXi printer with the power supply (713703 or BC1300M).

While Spot Vital Signs LXi is charging, the charging LED (~) flashes and the battery level indicator segments on the display continuously sequence. When the battery is fully charged, the charging LED stops flashing and the battery level indicator will stop sequencing.

If the device includes a Braun ThermoScan PRO 4000 thermometer, the charging LED below the thermometer will illuminate orange as it is charging. When the battery is fully charged, the LED will power off.

If the Spot LXi has a mounted, external printer that is connected via the cable, the charging LED on the printer flashes orange and green when the Spot LXi is plugged into the AC main power source. If the Spot LXi has a mounted, external printer that is not connected via the cable, the charging LED on the printer will not flash.

Note

There is no hazard associated with leaving the battery in the device, even if the device is not used for long periods of time.

30

Overview Welch Allyn Spot Vital Signs LXi

3

Functional Overview

This functional verification procedure helps to confirm the proper operation of the Spot

Vital Signs LXi and options. This procedure supports the requirements of routine preventative maintenance. It is not necessary to disassemble the Spot Vital Signs LXi to perform this procedure.

For the calibration procedures, see

“Calibration and Verification” on page 49. If the Spot

Vital Signs LXi fails certain functional tests or a circuit board is replaced, the device may require calibration. Always perform this functional verification procedure after performing any calibration.

Self Test

The Spot Vital Signs LXi performs a self-test each time the device is powered on. Press the Power button to turn the device on or off. Upon each power up, the display turns on, two audible beeps occur, and the Spot Vital Signs LXi displays the model and serial numbers. If the internal self-check is successful, the display shows its normal functions

(see

Figure 4 on page 23) with all values blank, and the device is ready for operation. If the

self-check fails, an error code is shown in the Navigation Window (see “Error Codes” on page 99).

Spot Vital Signs LXi automatically powers off when not used for 30 minutes.

31

32

Functional Overview Welch Allyn Spot Vital Signs LXi

Internal Configuration Mode

You can change several device operating parameters in the Internal Configuration Mode.

When changed, these settings become the default power-up settings. You will also see non-changeable device configurations for technical service purposes.

To Enter the Internal Configuration Mode:

1.

Turn the Spot Vital Signs LXi off.

2.

Press and hold the Select and Power buttons for 5 seconds. The device enters the

Internal Configuration Mode and the Configuration Menu screen appears on the display.

Figure 7. Internal Configuration Mode Menu

Configuration Menu

Version

Numbers

Battery

Location

Identifier

Date/Time

Blood Pressure

Temperature

Contrast

Manual

Parameters

External

Devices

Save Readings

Buttons

Lockout

Defaults

Event Log

Select

3.

Use the Navigation buttons to move through the menu options and then press the

Select button to access the options or accept a change. See the following tables for descriptions of the menu options.

4.

Press the Power button to exit the Internal Configuration Mode.

Service Manual Functional Overview

33

Table 2. Configuration Menu Options

Setting

Version Numbers*

Battery*

Location Identifier

Date/Time

Description

Displays the software and hardware version numbers in the Spot Vital Signs LXi device.

Displays the battery level.

Allows the entry of the device’s location (e.g., the department name). Follow the display prompts and use the Navigation and Select buttons to enter up to 10 characters.

Changes the date and time formats or updates the actual date and time. See Table 3,

“Date/Time Menu Options” for available settings.

Blood Pressure

Temperature

Contrast

Manual Parameters

Changes the blood pressure options. See Table 4, “Blood Pressure Menu Options” for available settings.

Changes the temperature options. See Table 5, “Temperature Menu Options” for available settings.

Changes the Display Contrast options. Use the left/right navigation buttons to adjust.

Changes the manual parameters defaults. See Table 6, “Manual Parameters Menu

Options” for available settings.

External Devices

Save Readings

Buttons Lockout

Defaults

Enables or disables available external devices. See Table 7, “External Devices Menu

Options” for available settings.

Saves the current patient reading at a preselected time interval or upon request. See

Table 8, “Save Readings Menu Options” for available settings.

Secures Spot Vital Signs LXi so unauthorized people cannot use the device or access data without enacting the proper key sequence. To override the buttons lockout feature, simultaneously press the Left Navigation button and the Select button.

Allows the user to select the default settings for the device and reset the unit to the default settings. See Table 9, “Change Local Defaults Options” for available settings.

Event Log* Displays the recent button presses, errors, measurements, measurement sites, battery state changes, and patient reading send events.

* Displayed information only; operator cannot change.

Table 3. Date/Time Menu Options

Setting

Date Format

Date

Time Format

Time

Description

Displays the date in one of the following styles:

• mm/dd/yyyy example: July 16, 2005 = 07/16/2005

• dd/mm/yyyy example: 16 July 2005 = 16/07/2005

Changes the date on the Display Window and in patient readings. If a location is entered (see “

Location Identifier

“ in

Table 2, “Configuration Menu Options”

), the date will not appear on the Display Window; the location will.

Displays the time in one of the following styles:

• 12-hour example: 5:00 PM

• 24-hour example: 17:00

Changes the time on the Display Window.

34

Functional Overview Welch Allyn Spot Vital Signs LXi

Table 4. Blood Pressure Menu Options

Setting

BP Calibration Check

Blood Pressure Units

Description

Prepares the Spot Vital Signs LXi for calibration. Only qualified personnel should verify

the Spot Vital Signs LXi blood pressure calibration. For more details, see “Blood pressure calibration” on page 62.

mmHg or kPa.

Mean Arterial Pressure (MAP) On or off.

Table 5. Temperature Menu Options

Setting

Temperature Units

Temperature Mode

Description

Fahrenheit (°F) or Celsius (°C).

SureTemp Plus models only: Oral, Pediatric Axillary, Adult Axillary, and Last Mode. In

Last Mode the device takes the next temperature in the mode in which the previous temperature was measured. Rectal Mode is available only when the rectal probe (red ejection button) and probe well are attached.

Table 6. Manual Parameters Menu Options

Setting

Height

Height Units

Height Default

Weight

Weight Units

Weight Default

Respiration

Pain Level

Description

On or off.

Inches (in) or centimeters (cm).

Changes the default patient height displayed in the Navigation Window.

On or off. Even if weight is enabled here, if weight scale is enabled in the External

Devices Menu, you cannot manually enter the weight.

Pounds (lb) or kilograms (kg).

Changes the default patient weight displayed in the Navigation Window.

On or off.

On or off.

Service Manual Functional Overview

35

Table 7. External Devices Menu Options

Setting

Information System

Barcode Patient ID

Barcode Clinician ID

Mandatory Sign In

Clear on Send/Save:

Clear on Power Cycle:

Weight Scale

Wireless Module

Printer

Printer Paper

Description

On or off. You must enable this option to send patient readings wired or wirelessly.

On or off. You must enable this option to send patient readings wirelessly.

On or off. You must enable this option to require mandatory sign in of a clinician.

Yes or no. You must enable this option to require the sign-in of the clinician.

Yes or no. You must enable this option to clear the Clinician ID after sending or saving the measurements.

Yes or No. You must enable this option to clear the Clinician ID after powering the Spot

Vital Signs LXi off.

On or off. Spot Vital Signs LXi connects to a scale and the weight shows on the display

window (see “Spot Vital Signs LXi Technical Overview” on page 37 for scale details).

None or DPAC. You must enable 802.11 B to send patient readings wirelessly. The radio is available as an accessory only with Welch Allyn Connex.

On or off.

Plain or labels. Only available if the Printer is enabled.

Table 8. Save Readings Menu Options

Setting

Save Mode

Auto Save Interval

Reading Full Action

Description

Manual or automatic. If automatic, Spot Vital Signs LXi saves readings at a preselected time interval. For either option, Spot Vital Signs LXi automatically saves the measured parameters into memory before automatically powering off when not used for 30 minutes.

Changes the amount of time before automatically saving the current patient reading.

Only available if Automatic Save Mode is enabled.

Auto Overwrite, Ask Overwrite, Do not Overwrite. Spot Vital Signs LXi can save 50 patient readings in memory. Upon reaching reading 51, the device may automatically overwrite reading 1, ask the user if he/she wants to overwrite reading 1, or disable the ability to take another reading until at least one reading is erased.

Table 9. Change Local Defaults Options

Setting

Language

BP Units

Temperature Units

Height Units

Weight Units

Date Format

Time Format

Description

English, Dansk, Nederlands, Suomi, Français, Deutsch, Italiano, Norsk, Español,

Português, Svenska, or Chinese.

mmHg or kPa.

Fahrenheit (°F) or Celsius (°C).

Inches (in) or centimeters (cm).

Pounds (lb) or kilograms (kg).

Displays the date in one of the following styles:

• mm/dd/yyyy example: July 16, 2005 = 07/16/2005

• dd/mm/yyyy example: 16 July 2005 = 16/07/2005

Displays the time in one of the following styles:

• 12-hour example: 5:00 PM

• 24-hour example: 17:00

36

Functional Overview Welch Allyn Spot Vital Signs LXi

4

Spot Vital Signs LXi Technical

Overview

System Description

The Spot Vital Signs LXi automatically measures systolic and diastolic pressure (excluding neonates), pulse rate, temperature (oral, adult axillary, pediatric axillary, rectal, and ear), and pulse oximetry (SpO

2

) as well as calculates Mean Arterial Pressure (MAP).

Furthermore, Spot Vital Signs LXi allows the entry of height, weight, respiration rate, and pain level. Spot Vital Signs LXi also calculates Body Mass Index (BMI) following height and weight entry.

The device is intended to be used by clinicians and medically qualified personnel. It is available for sale only upon the order of a physician or licensed health care provider.

Refer to the Spot Vital Signs LXi Directions for Use manual for complete information.

Battery System

Spot Vital Signs LXi uses a fuse and a shunt diode on the main board as well as a fuse in the battery assembly for protection against reverse battery connections. The fuse is located on the main board in the negative battery line. In the event of a reverse battery connection, a blown fuse will require replacement. The negative line location is more likely to short the battery (+) wire to ground than short the battery (-) wire to the VBUS node.

Battery Charger

Provides two levels of power to fast-charge or to float the battery at fixed voltages using software control. The fast-charge level rapidly brings the battery up to near-full capacity (7

Volts) using software to not overcharge the battery. The float level, set per the manufacturer’s recommendation, slowly tops-off the battery and can stay on indefinitely.

“CPU I/O” Power Supply

Provides power to the I/O pins of the DragonBall MPU and the attached peripheral IC’s.

The nominal voltage is set to a value of 3.15 Volts so that both the 3.0 Volt MPU and 3.3

Volts peripherals can operate within specifications. Spot Vital Signs LXi uses buck switching topology for efficiency and an on-off power latch directly controls the power supply.

37

38

Spot Vital Signs LXi Technical Overview Welch Allyn Spot Vital Signs LXi

Standby Mode

The Standby Mode conserves battery power. The device goes into Standby Mode if it is not used for two minutes. Press any button to bring the Spot Vital Signs LXi out of

Standby Mode.

CPU Core Power Supply

Powers the MPU core only and operates at the recommended Dragonball voltage range of 1.8 +/- 0.1 Volts; 2.0 Volts absolute max.

The +3.15 Volt supply powers the +1.8 Volt supply for efficiency and, more importantly, for power sequencing. During power-on the +1.8 Volt supply cannot exceed the +3.15 Volt supply. The R207/C182 further delays the +1.8 Volt supply. During power-off, D106 prevents the +1.8 Volt supply from exceeding the +3.15 Volt supply by more than a partial diode drop.

+5V Supply

Provides supply to miscellaneous loads that require more than 3.15 Volts. The Power

Latch controls the supply and the average load current is small, so efficiency is not important.

Clock/Calendar Power

The Clock/Calendar U19 receives power from two sources. During power on the clock runs from the +5 Volt supply and during power down the clock receives power directly from the battery potential using a voltage divider. Capacitor C131 provides power to the

U19 during the battery exchanges. Zener D102 prevents overvoltage to U19 in the case of operation directly from the charger.

Mod F NIBP Power

This module is powered directly from VBUS through an EMI filter. Software via BP-PWR-

ON-BAR controls the power. Level shifter Q115 / Q104 assures that the module remains off when the MPU is unpowered.

SpO

2

and Thermometer Power

A common structure supplies SpO

2 and thermometry. A single push-pull driver delivers square wave power to a pair of isolation transformers, one for SpO

2 and the other for thermometry. THERM-PWR-BAR activates the driver under software control. Level shifter

Q118, Q119 assures that the module remains off when the MPU is unpowered.

In the push-pull (forward) configuration, VBUS is multiplied by the transformer turns ratio and supplied to the input of the rectifier diodes. This is filtered and passed through lownoise 5-Volt linear regulators to supply power to the parameter modules.

LCD Power

The LCD data interface connects directly to the +3.15 Volt MPU I/O power supply. This powers the driver ICs inside the LCD pane and assures the data interface to the LCD is always powered when the unit is powered.

Service Manual Spot Vital Signs LXi Technical Overview

39

The contrast (bias) supply to the LCD panel and the LCD CCFL backlight both derive power indirectly from a pre-regulator. A software control can shut down the pre-regulator.

For efficiency, it is implemented as a buck switcher operating directly from the battery.

The primary purpose of the pre-regulator is to protect the backlight inverter / ballast from battery voltages that it cannot tolerate and to control current out of the unregulated ballast. The voltage is set to the low side of 5 Volts to reduce battery consumption, while still driving the CCFL with adequate voltage.

External Load (RS-232) Power

There are two power outlet means on the Spot Vital Signs LXi – RS-232 port power and the Braun PRO 4000 charger.

Two isolated power supplies provide power to each of the external RS-232 ports. This power is available to external devices on pin 9 of each serial connector whenever the software enables the corresponding RS-232 port.

SureTemp Plus Thermometer Interface

The SureTemp Plus OEM module is located in an optional pod and is connected to the main PCB via a flex cable.The Thermometer Probe connection is located on a separate PC board. It is connected to the SureTemp Plus OEM module by a ribbon cable. The electrical interface to the thermometer pod is a 6-pin ribbon connector, Molex 52271-0690 or equivalent.

Braun ThermoScan PRO 4000 Thermometer Interface

The interface to the Braun ThermoScan PRO 4000 thermometer is located in the optional pod and connects to the main PCB via a flex cable. The communications link is a single duplex current loop that also charges the Braun battery pack.

The presence or absence of current passing through the Braun ThermoScan PRO 4000 charge contacts controls the interface charge LED, located within the pod, which mimics the Braun ThermoScan Pro 4000 Base Station (type 6021).

SpO

2

Description

A shielded flat-flex cable connects the user interface sensor connector J8 to the SpO

2 electronics.

The Masimo SpO

2 module mates to the Masimo Adaptor Board that, in turn, mates to the

Main board. This provides power and communications to the module from the Main board.

The Nellcor SpO

2 module mounts directly to the Main board. This provides power and communications to the module, and brings the sensor signals to the main board where

EMI and ESD can be controlled. The sensor signals pass through the main board to the mating connector for the shielded flat-flex cable.

40

Spot Vital Signs LXi Technical Overview Welch Allyn Spot Vital Signs LXi

Radio Description

For the wireless radio description, refer to the manufacturer’s Directions for Use.

The radio interfaces to the Spot Vital signs LXi through an RS-232 female connector with a

9-pin male/female null modem connection.

Figure 8. 9-pin Null Modem Pinout

MA 0051 (DB9M/F) Adapter Pin out:

8

9

G

09 Male

2

3

4

5

1 + 6

7

7

9

G

09 Female

3

2

4

8

1 + 6

5

Bar Code Scanner Description

The Spot Vital Signs LXi has a 2D scanner (IMAGETEAM™ 4600). This scanner is programmed specifically for Welch Allyn. Scanners not purchased through Welch Allyn will not work with the Spot Vital Signs LXi.

Interface

The bar code scanner interfaces to Spot Vital Signs LXi through an isolated RS-232 DB-9 female connector. Use Welch Allyn part number 4500-910 for the barcode option.

Service Manual Spot Vital Signs LXi Technical Overview

41

Printer Description

The SPOT Vital Signs LXi interfaces with printer kit 4500-500.

Table 10. LED Pattern Table

Pattern

Constant Green

Fast Flashing Green

Long Green - Short Red

Short Green - Short Red

Short Green - Long Orange

Long Orange Flash

Fast Flashing Orange

Battery

Running

Running

Battery Low

Battery Low

Trickle Charging

Trickle Charging

Fast charging

Fast Flashing Red

No Light

Error Condition

Flat or in sleep mode

* Spool may have resulted from Paper Out or Head Up conditions)

Buffer Mode

Normal

Spool *

Normal

Spool *

Normal

Spool *

Spool (Printing prohibited)

Spool (Printing prohibited)

No hardware power switch is fitted, as power control is either automatic or by command from the host computer. To save power, the printer enters sleep mode after a period of inactivity, factory-set to 30 seconds.

The printer external power supply is used to charge the printer batteries.

42

Spot Vital Signs LXi Technical Overview Welch Allyn Spot Vital Signs LXi

Interconnect Diagram

(For drawing details

see “Interconnect Diagram 1” on page 147

and

see “Interconnect

Diagram 2” on page 148.

A9

A10

A11

A6

A7

A8

Item

A1

A1-F1

A2

A3

A4

A5

A5-F1

Description

PCA, Main, Spot Vital Signs LXi

Item

W1

Description

Assy, Power Harness, Spot Vital Signs LXi

Fuse, FlastBlo, 5A W1-A1 PCA, Power Interface, Spot Vital Signs LXi

PCA, Serial Interface, Spot Vital Signs LXi W2 Flex, Main, Spot Vital Signs LXi

Display, LCD, 320 x 240, Mono

Inverter, LCD

Assy, Battery, Spot Vital Signs LXi

Wire Protection Fuse

Assy, Fan Unit Cooling

PCA, Masimo Adapter, Spot Vital Signs LXi

Masimo MS-11 SpO

2

Shield, Masimo EMI

W3

W4

W5

W6

W7

W8

W9

Cbl Flx, 1/2mm, 15 POS

Assy, Ballast Harness, Spot Vital Signs LXi

Assy, LCD Gnd Harness

Assy, Cable, Flat Flex, SpO

2

Wire Clip, Side Entry,.375 Wide

Foam Strip, 2 Sided Adhesive

Cbl, Flx, 1mm, 6POS

Cbl, Flx, 1mm, 8POS

Cbl, Flx, Braun, Spot Vital Signs LXi Tape, Copper, Masimo Shield

PCA, SureTemp Plus, OEM w/o EDAC

PCBA, BP OEM, (MOD F)

Assy, Pump, and Tubing, Spot Vital Signs

LXi

A11-F1 Filter, Air, 1/8 Barbs

A12 Assy, Transducer Tubing, Spot Vital Signs

LXi

A13

A14

A15

PCA, EDAC, Spot Vital Signs LXi

Nellcor MP506 SpO

2

PCB

PCA, Braun Interface, Spot Vital Signs LXi

BLK

BRN

RED

ORN

GRN

BLU

PNK

Black

Brown

Red

Orange

Green

Blue

Pink

Wire Color

5

Maintenance and Service

Cleaning

WARNING Before performing any maintenance or service to the Spot Vital

Signs LXi, disconnect the AC power line from the electrical outlet.

Spot Vital Signs LXi

Caution Prevent water or other fluids from entering any connectors. If the connectors get wet, dry them with warm air. Check all measurement functions.

Caution Do not sterilize or autoclave the Spot Vital Signs LXi device.

As necessary, clean the device with a cloth slightly dampened with either 70% isopropyl alcohol, 10% chlorine bleach solution, or mild detergent in water, or use PDI Sani-System

Cloths. Never immerse Spot Vital Signs LXi in any type of fluid.

Blood Pressure Cuff

Before washing the Reusable Two-Piece Cuff, remove the two connectors and bladder, close off tubes with plugs (available as accessory 5082-163), and close the hook and loop fasteners. After washing, allow the cuff to air dry. Re-assemble the tube fittings.

Never press with an iron.

Blood Pressure Hose and Cable

Wipe the pressure hose with a damp cloth moistened in a mild detergent solution. Do not immerse hose.

43

44

Maintenance and Service Welch Allyn Spot Vital Signs LXi

SureTemp Plus Thermometer

Temperature Probe

Caution DO NOT immerse or soak the probe in any type of fluid.

Caution DO NOT use steam, heat, or gas sterilization on the probe.

Caution DO NOT autoclave the probe.

Press down on the connector tab and slide the connector out of the port to remove the temperature probe.

Regularly wipe the probe with a cloth dampened with warm water and a mild detergent solution, a 70% isopropyl alcohol solution, or a 10% chlorine bleach solution.

Removable Probe Well

Caution DO NOT use hard or sharp objects to clean the probe well. This could damage the probe well and cause the device to not function properly.

Caution DO NOT use steam, heat, or gas sterilization on the probe well.

Caution DO NOT autoclave the probe well.

1.

Remove the temperature probe from Spot Vital Signs LXi (see “Temperature Probe” on page 44).

2.

Grasp the well under the probe opening and pull up gently to remove it from the device.

3.

Swab the inner and outer surface of the probe well with a cloth dampened with a mild detergent solution, 70% isopropyl alcohol, or 10% chlorine bleach solution. Immerse the probe well in mild detergent solution as necessary for cleaning.

4.

Dry all surfaces thoroughly before re-assembling the device (see “SureTemp Plus” on page 26) for reassembly instructions.

Braun ThermoScan PRO 4000 Thermometer

Use a soft cloth slightly moistened with alcohol to clean the thermometer display and exterior. Do not use abrasive cleaners.

Damage to the probe tip or the presence of dirt or cerumen on the probe tip can affect the accuracy of your temperature measurement. To clean the tip, gently wipe it with a cotton swab slightly moistened with alcohol and immediately wipe dry with a clean cotton swab.

Allow to dry at least five minutes before taking a temperature.

Every month, clean the Braun ThermoScan PRO 4000 charging contacts within the Spot

Vital Signs LXi housing with a swab slightly dampened with alcohol.

Service Manual Maintenance and Service

45

SpO

2

Sensors

WARNING Do not immerse the sensor in water, solvents, or cleaning solutions

(the sensors and connections are not waterproof). Do not use irradiation, steam, or ethylene oxide for sterilization.

Clean the reusable SpO

2 the sensor.

sensor with a 70% isopropyl alcohol solution. Do not immerse

Radio

Refer to Manufacturer’s Directions for Use.

Printer

The printer label kit comes with a two-step head cleaner. Follow the directions provided with the cleaner.

46

Maintenance and Service Welch Allyn Spot Vital Signs LXi

Battery Replacement

Spot Vital Signs LXi

Caution Only use the Welch Allyn 4500-84 lead acid battery. Using an incorrect battery will cause damage to the Spot Vital Signs LXi and void the warranty.

Caution Do not break the shrink-wrap around the battery.

If necessary, replace the internal battery after heavy use or the battery no longer charges.

1.

Power off the Spot Vital Signs LXi and disconnect the AC power transformer cord.

2.

Remove the four screws holding the battery door using a phillips-head screwdriver.

Remove the battery door to expose the battery.

3.

Tip the Spot Vital Signs LXi backward and slide the battery out. Disconnect the oneway connector and then attach a new battery to the connector as shown. The oneway connector ends only connect one way. Do this as quickly as possibly to prevent loss of clock time.

Figure 9. Battery Removal and One-way Connector

One-way connector

4.

Slide the battery into the compartment as far as it will go. Push the connector down into the case next to the battery.

5.

Replace the battery door and tighten each of the four screws.

6.

Connect the AC power transformer to the Spot Vital Signs LXi and allow the new battery to charge for approximately 9 hours. The rechargeable batteries in the Braun

ThermoScan PRO 4000 thermometer require an additional 1 hour to charge. You can use the Spot Vital Signs LXi during this charging period via the AC power cord.

If Spot Vital Signs LXi displays the E38 error code after power up, set the date (see

“Date/Time Menu Options” on page 33).

The battery is a non-spillable lead-acid battery and must be recycled or disposed of properly according to national or local regulations.

Braun ThermoScan PRO 4000

Caution Do not use alkaline batteries in the Braun ThermoScan PRO 4000 thermometer.

Service Manual Maintenance and Service

47

Welch Allyn supplies a rechargeable battery pack with the Braun ThermoScan PRO 4000 thermometer.

1.

Open the battery compartment.

2.

Remove the battery pack and replace with a new battery pack, verify the poles are in the right direction.

3.

Slide the battery door back in until it snaps into place.

If the battery is completely discharged, the LEDs will not illuminate. Allow the unit to charge at least 15 minutes before proceeding.

The battery is a rechargeable battery and must be recycled or disposed of properly according to national or local regulations.

Printer

Caution Do not use alkaline batteries in the printer.

1.

Open the battery compartment.

2.

Remove the battery pack and replace with a new battery pack, verify the poles are in the right direction.

3.

Slide the battery door back in until it snaps into place.

Allow the printer to charge at least 9 hours before proceeding.

The battery is a rechargeable battery and must be recycled or disposed of properly according to national or local regulations.

48

Maintenance and Service Welch Allyn Spot Vital Signs LXi

Spot Vital Signs LXi Firmware Upgrade CD

This CD upgrades the Spot Vital Signs LXi unit software version. A detailed procedure is available on the upgrade CD.

Spot Vital Signs LXi Radio Configuration CD

The software on this CD configures the Spot Vital Signs LXi Radio for use on a local wireless network.

Product Disposal

Do not dispose of this product as unsorted municipal waste. Prepare this product for reuse or separate collection as specified by Directive 2002/96/EC of the European

Parliament and the Council of the European Union on Waste Electronic and Electrical

Equipment (WEEE). If this product is contaminated, this directive does not apply.

For more specific disposal information, see www.welchallyn.com/weee,or contact Welch

Allyn Customer Service at +44 207 365 6780.

6

Terms

Calibration and Verification

BP test volume repair fixture. Equipment with four separate 100cc, 250cc, and two

500cc linearity volume cylinders. Referred to in this procedure as “Test Volume”.

Calibration. Process to change the outputs of the UUT that are out of range. After saving, the changes become permanent until the next calibration.

Pneumatic test assembly. Equipment that connects the UUT, the pressure meter, the squeeze bulb, and the BP test volume repair fixture.

Spot LXi Service Tool. Software package to verify and calibrate the Spot LXi device.

Service computer. Computer with the repair software loaded and ready to operate.

Functional Tests. Process to check the outputs of the UUT.

UUT. Unit under test.

49

50

Calibration and Verification Welch Allyn Spot Vital Signs LXi

Required Test Equipment

Description

Power supply, 0-10 Volts, 2 amps capacity or better, fine voltage adjustment required

Digital multi meters (DMM)

Battery simulator connector, 24 inch or shorter

Patch cords- banana plug to banana plug, Red, 18 AWG, 24 inch or shorter

Patch cords- banana plug to banana plug, Black, 18 AWG, 24 inch or shorter

USB Cable, 2.0/5-Pin Mini-B Cable

Calibrated pressure meter

PC running Spot Lxi Service Tool (Ver 3.0 or greater)

9600 Plus Temperature Calibration Tester

Nellcor-approved SpO2 simulator, for devices with Nellcor option

Masimo SpO2 Tester, for devices with Masimo option

Thermometer Calibration Key, Optional

Blood pressure two-tube hose with Y fitting

Blood pressure two-tube hose (standard cuff)

BP test volume repair fixture (includes items below)

Note This is the new Test Volume needed for passing the NIBP Linearity Test

BP test volume repair fixture. Must have at least the following volume cylinders: 100cc,

250cc, and 500cc

Squeeze bulb with bleeding screw

Square (4-way) tube fitting connector

12-inch (30.5 cm) tubing with.25-inch (6.35 mm) ID

Part Number

4500-905

01802-110

SRC-MAX

1795

06138-000

5082-183

4500-30

407672

3

2

Qty

2

4

Pneumatic test assembly setup

1.

Connect the squeeze bulb with bleed screw to a square (4-way) tube fitting connector, a calibrated pressure meter, a test volume, and to a blood pressure two-tube, Y-fitting hose as shown in

Figure 1 on page 51.

Service Manual

Figure 1. Pneumatic test assembly setup

Test volume

500 CC

LINEARITY

100 CC

500 CC

250 CC

Calibration and Verification

51

52

Calibration and Verification Welch Allyn Spot Vital Signs LXi

Power Supply Test Setup

1.

Connect the power supply to the Battery Simulator connector, banana plug patch cords, digital multi meter (DMM) for voltage monitoring, DMM for current monitoring.

Figure 2. Power Supply Test Setup

Power supply

DMM

DMM

DMM connections

Test the power supply output as shown in

Figure 3

. Observe the correct polarity.

Figure 3. Power Supply

Service Manual Calibration and Verification

53

2.

Turn on the power supply and adjust to 6.5 +/- 0.1V DC before connecting the UUT.

Turn off power supply.

3.

Disconnect the UUT battery and connect the external power supply to the UUT, using the Battery Simulator. Observe the correct polarity.

Caution DO NOT connect the Spot LXi AC power adapter to the UUT power connection port. This will result in damage to the UUT, AC power adapter, or the

Battery Simulator power supply.

4.

Turn on the Power Supply.

Note

All tests and verification in this procedure require battery voltage input adjusted to

6.5V DC; Except as called out in the Voltage Calibration section.

Battery Removal / Power Shutdown

Removal of the UUT battery, or power shutdown of the Battery Simulator external power supply, will require reset of the date and time prior to any further testing, calibration, or verification activities.

Time & Date Reset:

1.

Power on the UUT in configuration mode by pressing the Check and Power On buttons on the UUT. Hold the Check button until the UUT beeps twice. The UUT screen displays any POST errors detected.

Figure 4. POST Errors Detected

2.

Click the Check button on the UUT. The Configuration Menu opens.

3.

Scroll to Date/Time on the UUT Configuration Menu and select.

54

Calibration and Verification

Figure 5. Date/Time Menu

Welch Allyn Spot Vital Signs LXi

4.

Scroll to Date on the UUT and select. Using the navigation keys, adjust the correct date and select. Scroll to Time and select. Using navigation keys, adjust the correct time and select.

Figure 6. Set Date, Set Time

5.

Scroll to Exit and select.

Service Manual

Figure 7. Exit Menu

Calibration and Verification

55

6.

Reset the UUT using the power button. Do NOT turn off the power supply. The next power up of the UUT screen displays the correct date and time without errors.

Voltage Calibration

1.

Remove the battery from the UUT. Connect the UTT to a Battery Simulator with the external power supply set to 6.5 +/- 0.1V DC. (Follow the steps in

“Power Supply Test

Setup” on page 52 .)

2.

Connect the service PC running the Spot Lxi Service Tool to the UUT with the USB

Cable. Open the Spot LXi Service tool from the PC.

3.

Power on the UUT in configuration mode by pressing the Check and Power On buttons on the UUT. Hold the Check button until the UUT beeps twice.

56

Calibration and Verification

Figure 8. Configuration Menu

Welch Allyn Spot Vital Signs LXi

4.

Verify the power supply is at 6.5 +/- 0.1V DC at the battery connector simulator leads.

5.

Verify that the UUT is connected and communicating with Spot Lxi Service Tool.

Select Unit > Information. The Unit information window opens. Click OK to close.

6.

Select Calibrate > Voltage on the service computer. The UUT screen goes blank.

Figure 9. Calibrate Voltage

7.

Carefully adjust the power supply to 5.5 + 0.1V DC at the battery connector simulator leads. A fine voltage adjustment will be needed to make this adjustment.

Service Manual Calibration and Verification

57

Note

Do not adjust power supply below 5.5V DC or the UUT will go into shutdown mode.

8.

Read the voltage on the DMM that is connected to the battery simulator leads.

9.

Enter this voltage reading into the Actual Voltage field and click the Calibration button.

10. Click the OK button to accept the inputs. Click Yes to make the changes permanent.

11. Enter your initials in the Calibration Signature field and click OK. The UUT resets and powers up in normal mode.

Figure 10. Calibration Signature Field

Note

After “Reset” and power up, the UUT detects the low 5.5V DC supply setting and goes into auto shutdown. Adjust power back to 6.5V DC.

Pressure verification test

Spot LXi devices require annual calibration.

Note

If this is a new NIBP board, go to

“NIBP board Initialization” on page 62 and

carefully follow the instructions to initialize this board.

1.

Connect the UUT to the pneumatic test assembly using the two-tube blood pressure hose with the Y-fitting.

Figure 11. Blood Pressure Hose with Y-fitting

2.

Connect the test volume hose to the 500cc cylinder.

58

Calibration and Verification Welch Allyn Spot Vital Signs LXi

3.

Power on the UUT in configuration mode by pressing the Check and Power On buttons on the UUT. Hold the Check button until the UUT beeps twice. The configuration screen displays.

Figure 12. Configuration Menu

Scroll to Blood Pressure and select it

Figure 13. Blood Pressure Menu

Select BP Calibration Check.

Service Manual

Figure 14. BP Calibration Check

Calibration and Verification

59

4.

Open the Spot LXi Service Tool on the service computer and select Test >

Calibration. A Test Calibration box displays on the screen.

Figure 15. Test Calibration with Open Valve

5.

Verify the pressure is at or near zero and the voltage is within 0.2V DC of the power supply (6.5V DC). Verify the valve status is open and the pump status is off.

6.

Select Close Valve on the UUT. The valve status changes to closed.

60

Calibration and Verification

Figure 16. BP Calibration Check Close Valve

Welch Allyn Spot Vital Signs LXi

Figure 17. Test Calibration with Closed Valve

7.

Select Start Cuff Inflation. The pump starts inflating the cuff. Verify the pump status is on and the pressure is rising.

Service Manual

Figure 18. Start Cuff Inflation

Calibration and Verification

61

8.

Inflate the pressure up to 250 mmHg. Select the Stop Pump on the UUT to stop the inflation. Fine tune the pressure reading to 250 mmHg with the squeeze bulb.

Figure 19. Fine Tune the Pressure Reading to 250 mmHg

9.

Check the reading on the pressure meter against reading on the UUT.

a.

Verify the pressure on the UUT is within the pressure tolerance in the table. Use the squeeze bulb to fine tune the pressure at each inflation pressure limit. If the pressure is within tolerance, go to the next inflation pressure.

62

Calibration and Verification Welch Allyn Spot Vital Signs LXi

Inflation pressure Pressure tolerance

250 mmHg +/- 2.0 mmHg

150 mmHg +/- 1.5 mmHg

Inflation pressure Pressure tolerance

50 mmHg +/- 1.0 mmHg

0 mmHg +/- 1.0 mmHg b.

If all readings are within specification, go to

“Functional tests” on page 67 .

c.

If the UUT is not within the pressure tolerance, open the valve to bleed pressure to zero. Select Calibrate > Pressure on the repair software. Proceed to

“Blood pressure calibration” on page 63. UUT needs calibration.

d.

If replacement of the NIBP board is required, first proceed to

“NIBP board

Initialization” on page 62 and then proceed to

“Blood pressure calibration” on page 63.

Blood pressure calibration

Note

Complete the voltage calibration before beginning a blood pressure calibration.

If all readings were within specifications during

“Pressure verification test” on page 57,

skip this section and go to

“Functional tests” on page 67 .

NIBP board Initialization

Note

All New NIBP Boards must be initialized before calibration.

1.

Power on the UUT in configuration mode by pressing the Check and Power On buttons on the UUT. Hold the Check button until the UUT beeps twice. The configuration screen displays.

Figure 20. Configuration Menu

Service Manual Calibration and Verification

63

2.

Open the Spot LXi Service Tool on the service computer and select Calibrate > Reset

Defaults. The Reset Defaults window displays.

Figure 21. Reset Defaults Window

3.

Click Reset > Initialize > OK. The NIBP board is initialized.

4.

Perform the

“Voltage Calibration” on page 55 before proceeding to

“Blood pressure calibration” .

Blood pressure calibration

1.

Power on the UUT in configuration mode by pressing the Check and Power On buttons on the UUT. Hold the Check button until the UUT beeps twice.

Figure 22. Blood Pressure Menu

64

Calibration and Verification Welch Allyn Spot Vital Signs LXi

2.

The configuration screen displays, scroll to Blood Pressure and select it. Select BP

Calibration Check.

Figure 23. BP Calibration Check

3.

Open the Spot LXi Service Tool on the service computer and select Calibrate >

Pressure. The Calibrate Pressure window opens.

Figure 24. Calibrate Pressure Window

Service Manual Calibration and Verification

65

4.

Connect the UUT to the pneumatic test assembly using the two-tube blood pressure hose with the Y-fitting. (See

“Blood Pressure Hose with Y-fitting” on page 57.)

5.

Connect the pneumatic test assembly hose to the 500cc cylinder of the test volume and verify the following:

• the pressure is at zero mmHg,

• the bleed screw on squeeze bulb is closed,

• valve on UUT is in the open state.

6.

On the Spot LXi Service Tool select Calibrate Zero button.

7.

Close Valve on the UUT and use the squeeze bulb to manually inflate the UUT to 200 mmHg. Let stabilize.

Figure 25. Fine Tune the Pressure Reading to 200 mmHg

8.

Enter the pressure meter reading in the Calibration Gain field on the service computer. Be as precise as possible. If pressure meter has resolution of 1.00 mmHg make sure it reads 200.00 without switching to 199.00 or 201.00. Enter 200.00 in

Calibration Gain field.

66

Calibration and Verification

Figure 26. Calibration Gain Field

Welch Allyn Spot Vital Signs LXi

9.

Click Calibrate Gain on the service computer.

10. Click OK to accept the inputs. Click Yes to make the changes permanent.

11. Enter your initials in the Calibration Signature field and click OK. The UUT resets and powers up in normal mode.

Figure 27. Calibration Signature Field

12. Blood Pressure calibration is complete.

Service Manual Calibration and Verification

67

Functional tests

Note

Perform the Functional Tests sequentially. Fill out the

“Service Work Checklist” on page 111 with the Calibration and Functional Test data obtained as you perform

these steps.

Each test is software driven. Selecting the Test button begins the test.

Connect the correct Test Volume cylinder called out in the procedure (in parenthesis below) or the test will fail.

Test specification details:

Leak test

Dump test

Inflation test

Valve control test

Inflation linearity

6mm maximum in 15 seconds with a volume of (100cc) at 250 mmHg

From 260 to less than 15 mmHg in 10 seconds or less with a volume of (500 cc)

From 5 to 210 mmHg in 7 seconds or less with a volume of (250 cc)

From 160 mmHg with a volume of (100 cc)

Open the valve for the following time intervals and pressure drops:

•10 msec = 4 to 12 mmHg

•15 msec = 4 to 15 mmHg

•25 msec = 4 to 25 mmHg

Connect a standard cuff hose between the UUT and the 500cc Linearity input on the Test Volume and inflate.

•Measure the pressure at 6, 9, and 12 seconds and compare to the expected readings at these intervals.

•Also check the noise level at 7, 10, and 13 seconds into the inflation step.

Functional test steps:

1.

If the UUT is connected to the Battery Simulator, set the external power supply to 6.5

+/- 0.1V DC.

Note

The Spot LXi AC power adapter should NOT be connected to the UUT.

If the UUT is NOT connected to the battery simulator, connect the Spot LXi AC power adapter to the UUT.

2.

Power on UUT in normal mode.

3.

Connect the UUT to the pneumatic test assembly using the two-tube blood pressure hose with the Y-fitting. (See

“Blood Pressure Hose with Y-fitting” on page 57.) Verify

that the bleed valve on the squeeze ball is closed.

4.

Open the Spot LXi Service Tool on the service computer and select Test >

Pneumatics. The Test Pneumatics window opens.

68

Calibration and Verification

Figure 28. Test Pneumatics Window

Welch Allyn Spot Vital Signs LXi a.

Connect the test specific cylinder in parenthesis; (100cc), (500cc), (250cc) b.

Sequentially select each pneumatic test, starting with “Leak (100cc)”. ClickTest.

A successful test will display “Passed”.

Figure 29. Passed Leak Test

c.

Perform each of the tests listed after connecting the test specific cylinder. The

“Inflation Linearity (Cuff)” test may use the 500cc cylinder in place of the adult cuff.

Service Manual

Figure 30. Inflation Linearity Cuff Test

Calibration and Verification

69

Figure 31. Inflation Linearity 500cc Cylinder

500 CC

LINEARITY

100 CC

500 CC

250 CC

d.

Click OK after completing the last test to close the Pneumatics test window.

5.

On the Spot LXi Service Tool select Test > Current Levels. Perform the tests listed.

The Test Current Levels window opens.

70

Calibration and Verification

Figure 32. Test Current Levels Window

Welch Allyn Spot Vital Signs LXi a.

Sleep State > Test, screen is off and the UUT is idle. Verify that the current is less than 500 mA. Click OK after completing the measurement.

Figure 33. Sleep State

b.

Valve/pump Mode > Test, the UUT is fully active with the valve closed and the pump running. Verify that the current is less than 2.0 amperes in this state. Click

OK after completing the measurement.

c.

Normal Mode; screen is lit, but the UUT is inactive. The UUT current draw is less than 1.25 amperes in this state.

d.

Click OK after completing the last test to close the Test Current Levels window.

6.

On the Spot LXi Service Tool select Test > Noise Levels > Test. The Test Noise

Levels window opens.

Service Manual

Figure 34. Test Noise Levels Window

Calibration and Verification

71

a.

The Test Noise Levels test measures the amount of noise on the UUT A-D pressure channel. The sample is 1 second in length with 0 mmHg applied. The maximum allowed noise is .050 mmHg.

b.

Click OK after completing the test to close the Test Noise Levels window.

7.

On the Spot LXi Service Tool select Test > Buttons. The Test Buttons window opens.

Figure 35. Test Buttons

a.

Press each button on the UUT and verify the appropriate button lights on the Spot

LXi Service Tool window.

72

Calibration and Verification

Figure 36. UUT Buttons

Welch Allyn Spot Vital Signs LXi b.

Click OK after completing the test to close the Buttons Test window.

8.

On the Spot LXi Service Tool select Test > Interface. The Test Interface window opens.

Figure 37. Test Interface Window

a.

Perform all commands in this test, verifying the screen on the UUT responds correctly with each command:

Normal

All On

All Off

Test Pattern

Backlight On/Off, = blank dark UUT display screen

Service Manual

Power Indicator On/Off

Charge Indicator On/Off

Enunciator On/Off, Toggles enunciator tone on and off.

Figure 38. Normal

Calibration and Verification

73

Figure 39. All On

74

Calibration and Verification

Figure 40. All Off

Welch Allyn Spot Vital Signs LXi

Figure 41. Test Pattern

Service Manual

Figure 42. Backlight On/Off

Calibration and Verification

75

Figure 43. Power Indicator On/Off

Figure 44. Charge Indicator On/Off

b.

Click OK after completing the tests to close the Test Interface window.

To perform the over pressure test:

1.

Verify that the UUT is in normal mode.

2.

Re-install the pneumatic test assembly with the 250cc cylinder connected.

3.

On the Spot LXi Service Tool select Unit >Status. A Unit Status window displays.

76

Calibration and Verification

Figure 45. Unit Status Window

Welch Allyn Spot Vital Signs LXi

4.

Check the Log Status Information box in the lower right hand corner of the Spot LXi

Service Tool. The Event logs display in rapid succession.

5.

Immediately initiate a blood pressure reading. As the pressure increases, start applying additional pressure with the squeeze bulb.

6.

Increase the pressure to the UUT until an E10 error displays.

7.

Uncheck the Log Status Information box and click OK after completing the test to close the Unit Status box.

Scroll through the event log to locate the last ‘no’ answer, directly followed by the first ‘yes’ answer in the column labeled over pressure.

Service Manual

Figure 46. Event Log

Calibration and Verification

77

8.

Scroll to the right to see the mmHg pressure that these events occurred.

9.

Select the higher pressure reading when the E10 error occurred. Verify the pressure is between 296 and 329 mmHg.

Note

The following tests are for verification only. There is no updating of calibration parameters.

Temperature Functional Check

Temperature test setup

The 9600 Plus Calibration Tester takes approximately 20 minutes to heat to the lowest setting. When testing several thermometers at all three temperatures, it is recommended to test all probes at one Calibration Set Point Temperature before proceeding to the next

Calibration Set Point Temperature.

78

Calibration and Verification Welch Allyn Spot Vital Signs LXi

To further expedite testing start at the lowest Calibration Set Point Temperature. The 9600

Plus Calibration Tester does not have an internal fan, this causes a longer cool down time than warm up time.

Refer to the 9600 Plus Calibration Tester Directions for Use manual for specific information regarding the LCD window or the control buttons.

1.

Choose the proper mains plug insert and slide it over the two prongs in the power converter.

Figure 47. Power Adapter and Mains Plug Inserts

Two prongs in the power adapter

11.8593./US

(US)

US

11.8593./UK

(UK)

UK

11.8593./AUS

(AUS)

AUS

11.8593./EU

(EU)

EU

2.

Plug the power adapter into the 9600 Plus Calibration Tester and the opposite end into a wall outlet.

3.

Place the 9600 Plus Calibration Tester on a level surface away from sunlight, drafts, and other sources of heat or cold.

4.

Observe the Set Point Mode in the upper left hand corner of the LCD display. If the unit displays a “D”, it is in Default Mode and will heat to the lowest Set Point

Temperature. If you do not want to conduct testing at this Set Point Temperature, press and hold the Temperature Selection button to select the desired setting. The temperature display will flash before staying on continuously to indicate the 9600 Plus

Calibration Tester has stabilized and is ready for use.

Changing the Calibration Set Point Temperature on the 9600 Plus

Calibration Tester

To scroll from one set point to the next, press and hold the Temperature Selection button (up/down button) until a beep is heard. The newly selected set point appears in the upper left corner of the LCD display. The device's current temperature is displayed, starts to flash, and continues flashing until the cavity reaches the equilibrium at the new set point.

Welch Allyn SureTemp Plus Thermometers

Caution Store thermometers for testing in the same room as the 9600 Plus

Calibration Tester for approximately 30 minutes prior to testing to allow for thermal accommodation.

Service Manual Calibration and Verification

79

To begin functional verification of the SureTemp Plus thermometer:

1.

Verify that the UUT is ON and is operating in normal mode.

2.

Remove the thermometer probe from the probe well and clean it with either a 70% isopropyl alcohol solution, or a 10% chlorine bleach solution. Let the thermometer air dry. Do not apply a probe cover.

3.

Verify that the blue (oral) probe well is in place.

Insert the thermometer into the 9600 Plus Calibration Tester device thermistor device port (small port).

Figure 48. 9600 Plus Calibration Tester with the Welch Allyn SureTemp Plus Thermometer

4.

Verify that the thermometer enters Monitor Mode, (approximately one minute), the snail symbol and “Monitor” are displayed on the UUT display.

Figure 49. Temperature Monitor Mode

5.

After the UUT enters “Monitor” mode wait for approximately one minute or until the temperature on the thermometer is stable for ten seconds.

a.

Verify that the UUT thermometer's temperature reading is within the temperature tolerance in the table.

Calibration Set Point Temperature

96.8°F (36.0°C)

105.8°F (41.0°C)

Temperature tolerance

+/- 0.2°F (+/- 0.1°C)

+/- 0.2°F (+/- 0.1°C)

80

Calibration and Verification Welch Allyn Spot Vital Signs LXi b.

When verifying several UUT thermometers at both temperatures, it is recommended to verify all UUT thermometers at one Calibration Set Point

Temperature before proceeding to the next Calibration Set Point Temperature.

c.

Changing the Calibration Set Point Temperature (9600 Plus Calibration Tester):

To scroll from one set point to the next, press and hold the Temperature

Selection button (up/down button) until a beep is heard. The newly selected set point appears in the upper left corner of the LCD display. The device's current temperature is displayed, starts to flash, and continues flashing until the cavity reaches the equilibrium at the new set point.

To begin functional verification of SureTemp Plus without thermometer (Optional):

If the UUT fails the temperature verification using the SureTemp Plus thermometer; then perform this separate functional test of the UUT temperature circuitry.

1.

Remove probe and probe well completely.

2.

Insert calibration key (part number 06138-000) and replace probe well.

Figure 50. Calibration Key and Probe Well

3.

Insert and then remove probe from probe storage channel to reset thermometer.

Service Manual

Figure 51. Reset Thermometer

Calibration and Verification

81

4.

Wait for display test, and then observe display.

5.

Display should read 36.3°C +/- 0.1°C or 97.3°F +/- 0.2°F.

Figure 52. Temperature Display in F°

6.

Remove calibration key and install probe well.

7.

If the UUT temperature circuitry verified successfully and the temperature verification failed using thermometer probe; then replace SureTemp Plus thermometer probe.

82

Calibration and Verification Welch Allyn Spot Vital Signs LXi

Braun ThermoScan Pro 4000

Caution Store thermometers for testing in the same room as the 9600 Plus

Calibration Tester for approximately 30 minutes prior to testing to allow for thermal accommodation.

To begin functional verification of the Braun ThermoScan PRO 4000 thermometer:

1.

Clean the probe tip with a cotton swab slightly moistened with isopropyl alcohol, remove excess alcohol with a clean cotton swab, and let air dry for 5 minutes.Do not use any chemical other than alcohol to clean the probe window.

2.

Place the Braun ThermoScan Pro 4000 in Calibration Mode using the following steps.

a.

Push and release the I/O mem button to turn the product on. The display shows symbols and functions. The Braun ThermoScan Pro 4000 performs an automatic self check. After a few seconds you hear a beep and see two dashed lines, a sound symbol, and ° C or ° F on the display.

Figure 53. Braun Self Check

b.

Push the I/O mem button again and keep it pressed for the following steps:

• After approx. 3 seconds the “OFF” symbol flashes on the display (keep pressing the button).

• When you hear a beep, release the button immediately.

• The Braun TheromScan Pro 4000 is now in calibration check mode and the display is flashing and showing the “CAL” symbol.

• Apply a new probe cover. Insert the Braun ThermoScan Pro 4000 into the

9600 Plus Calibration Tester ear device port (large port). Place the probe firmly into the ear device port.

Service Manual

Figure 54. Braun Ear Device Port

Calibration and Verification

83

3.

Wait approximately three seconds, press the thermometer start button, and watch for the “Exac Temp” light to flash.

Figure 55. Press Start Button

84

Calibration and Verification Welch Allyn Spot Vital Signs LXi

4.

Leave the thermometer in the 9600 Plus Calibration Tester until you hear a beep.

Remove the Pro 4000 Thermometer from the 9600 Plus Calibration Tester and read the temperature in the thermometer’s display.

a.

Verify that the UUT thermometer's temperature reading is within the temperature tolerance in the table.

Calibration Set Point Temperature

96.8°F (36.0°C)

105.8°F (41.0°C)

Temperature tolerance

+/- 0.2°F (+/- 0.1°C)

+/- 0.2°F (+/- 0.1°C) b.

Wait one full minute before taking another reading with the same thermometer.

Repeated measurements in short sequence may cause higher readings.

c.

When verifying several UUT thermometers at both temperatures, it is recommended to verify all UUT thermometers at one Calibration Set Point

Temperature before proceeding to the next Calibration Set Point Temperature.

d.

Changing the Calibration Set Point Temperature (9600 Plus Calibration Tester):

To scroll from one set point to the next, press and hold the Temperature

Selection button (up/down button) until a beep is heard. The newly selected set point appears in the upper left corner of the LCD display. The device's current temperature is displayed, starts to flash, and continues flashing until the cavity reaches the equilibrium at the new set point.

5.

The Braun ThermoScan PRO 4000 thermometer will exit CAL mode after four minutes.

To begin functional verification of the Braun ThermoScan PRO 4000 thermometer

Down Load:

1.

Clean the probe tip with a cotton swab slightly moistened with isopropyl alcohol, remove excess alcohol with a clean cotton swab, and let air dry for 5 minutes. Do not use any chemical other than alcohol to clean the probe window.

2.

Insert the Braun ThermoScan PRO 4000 thermometer into the UUT and wait for the

“Exac Temp” light to go out.

Service Manual

Figure 56. Braun Exac Temp Light

Calibration and Verification

85

3.

Remove the Braun ThermoScan PRO 4000 thermometer from the UUT.

4.

Apply a new probe cover. Insert the Braun ThermoScan Pro 4000 into the 9600 Plus

Calibration Tester ear device port (large port). Place the probe firmly into the large port. See

(Figure 54)“Braun Ear Device Port” on page 83. The temperature selection

is non-specific for this particular test.

5.

Wait approximately three seconds, press the thermometer Start button, and watch for the “Exac Temp” light to flash. See

“Braun Exac Temp Light”

.

6.

Leave the thermometer in the 9600 Plus Calibration Tester until you hear a beep.

Remove the Pro 4000 Thermometer from the 9600 Plus Calibration Tester and read the temperature in the thermometer's display.

7.

Insert the Braun ThermoScan PRO 4000 thermometer into the UUT (should be on and in normal mode). The reading from the Braun ThermoScan PRO 4000 thermometer should transfer automatically to the UUT display. Verify this action.

Changing the Calibration Set Point Temperature

To scroll from one set point to the next, press and hold the Temperature Selection button until a beep is heard. The newly selected set point appears in the upper left corner of the

LCD display. The device’s current temperature is displayed, starts to flash, and continues flashing until the cavity reaches the equilibrium at the new set point.

To begin functional verification of the Masimo SpO2

Use the “Masimo Tester” to verify the Masimo SpO2 functionality.

86

Calibration and Verification

Figure 57. Masimo Tester

Welch Allyn Spot Vital Signs LXi

1.

With the UUT OFF, orient the Masimo Tester such that the mini-D connector mates with the SpO2 connector on the UUT; or use appropriate Masimo extension cable to connect the Masimo Tester and UUT.

2.

Power ON the Spot LXi UUT and confirm the SpO2 reading in the Display Window is

81%± 3% and the pulse reading is 61 bpm ± 1 bpm.

Figure 58. SpO2 Reading

3.

Place the thumb and index finger on the gray buttons on either side of the Masimo

Tester connector or extension cable connector; press the buttons firmly, and gently pull to remove the tester.

Service Manual Calibration and Verification

87

Note

The functionality of the SpO2 module is not adjustable. If the SpO2 is not functioning properly, or the SpO2 and pulse performance cannot be verified against the calibrated simulator, contact Welch Allyn Technical Support

page iv

.

To begin functional verification of the Nellcor SpO2

Use a Nellcor-approved SpO2 simulator (SRC-MAX) to verify the Nellcor SpO2 functionality.

Figure 59. Nellcor SpO2 Simulator

1.

With the UUT ON and completely booted, orient the simulator such that the connector mates with the SpO2 connector on the UUT; or use appropriate Nellcor extension cable to connect the simulator and UUT.

Note

The UUT MUST be powered on prior to connecting the Nellcor SpO2 simulator.

2.

On the Nellcor simulator change the SpO2 percentage from default 75% to 90%.

3.

Confirm the SpO2 reading in the Display Window is 90%± 1% and the pulse reading is 60 bpm ± 1 bpm.

88

Calibration and Verification

Figure 60. SpO2 Reading and Pulse Reading

Welch Allyn Spot Vital Signs LXi

4.

Remove the simulator by placing the thumb and index finger on either side of the simulator connector or extension cable connector; press the buttons firmly and gently pull to remove.

Note

The functionality of the SpO2 module is not adjustable. If the SpO2 is not functioning properly, or the SpO2 and pulse performance cannot be verified against the calibrated simulator, contact Welch Allyn Technical Support

page iv

7

Specifications

Performance

This section describes normal ranges for Spot Vital Signs LXi.

Blood Pressure Accuracy

Blood pressure accuracy meets or exceeds AAMI SP10:2002 standards for non-invasive blood pressure accuracy (± 5 mmHg mean error, 8 mmHg standard deviation). Blood pressure accuracy is validated for pressure measurement using the upper arm only.

Cuff Pressure Range

Systolic Range

Diastolic Range

Blood Pressure Determination Time

0 to 300 mmHg

60 to 250 mmHg

30 to 160 mmHg

Typical: 15 seconds

Mean Arterial Pressure Range

Pulse Rate Range

(using SpO

2 determination)

Pulse Rate Range

(using Blood Pressure determination)

Pulse Rate Accuracy (using SpO

2 determination)

40 to 190 mmHg

25 to 240 bpm

35 to 199 bpm

Without Motion: 25 to 245 bpm ± 3 digits

1

With Motion: normal physiologic range (55 to 125 bpm) ± 5 digits

Low Perfusion: 25 to 245 bpm ± 3 digits

1

±5.0%

Pulse Rate Accuracy

(using Blood Pressure determination)

Overpressure Cutoff

1

Specification applies to device performance and was validated with Biotek and Nellcor simulators.

315 mmHg ±15 mmHg

89

90

Specifications

Temperature Specifications

Temperature Range

SureTemp Plus

Braun ThermoScan PRO 4000

Calibration Accuracy

SureTemp Plus

Braun ThermoScan PRO 4000 for displayed temperature ranges

Display Resolution

SureTemp Plus Predict Time

Oral

Adult Axillary

Pediatric Axillary

Rectal

Welch Allyn Spot Vital Signs LXi

80° to 110° F (26.7° to 43.3° C)

68° to 108° F (20° to 42.2° C)

±0.2° F (0.1° C) (Monitor Mode)

± 0.4° F (±0.2° C) 95.9° to 107.6° F (35.5° to 42° C)

±0.5° F (±0.25° C) (outside this temperature range)

0.1° F or ° C

Approx. 4 to 6 seconds

Approx. 12 to 15 seconds (age 18 years and older)

Approx. 10 to 13 seconds (age 17 years and younger)

Approx. 10 to 13 seconds

Service Manual Specifications

91

SpO

2

Specifications

Masimo Sensor Accuracy Guide

Accuracy specified when used with Masimo SET pulse oximetry monitors or with licensed Masimo SET pulse oximetry modules using PC series patient cables, during no motion. Numbers present ± 1 standard deviation. Plus or minus one standard deviation represents 68% of the population. SpO accuracy from 25 to 240 bpm.

2 accuracy from 70% to 100%. Pulse rate

Performance Measurement Range

Perfusion

Accuracy

Pulse Rate Accuracy

SpO

2

: 1 to 100%

Pulse Rate: 25 - 240 beats per minute (BPM)

0.02% to 20%

Saturation: 70% to 100%

No Motion: Adults, Pediatrics ± 2 digits

Neonates ± 3 digits

Motion: Adults, Pediatrics ± 3 digits

Neonates ± 3 digits

Low Perfusion: Adults, Pediatrics ± 2 digits

Neonates ± 3 digits

Pulse Rate: 25 to 250 bpm

No Motion: Adults, Pediatrics, Neonates ± 3 digits

Motion: Adults, Pediatrics, Neonates ± 5 digits

Low Perfusion: Adults, Pediatrics, Neonates ± 5 digits

Table 1. Masimo Sensor Accuracy Guide

Sensor

LNOP-ADT

LNOP-PDT

LNOP-NEO

LNOP-NEOPT

LNOP-DCI

LNOP-DCIP

Weight Range

> 30 kg

10 to 50 kg

< 10 kg

< 1 kg

> 30 kg

10 to 50 kg

Saturation Accuracy

No Motion Motion

± 2%

± 2%

± 3%

± 3%

± 3%

± 3%

± 2%

± 2%

± 3%

± 3%

± 3%

± 3%

Pulse Rate Accuracy

No Motion Motion

± 3 bpm

± 3 bpm

± 5 bpm

± 5 bpm

± 3 bpm

± 3 bpm

± 3 bpm

± 3 bpm

± 5 bpm

± 5 bpm

± 5 bpm

± 5 bpm

Masimo Patents

The Masimo sensors and cables are covered under one or more of the following U.S.A.

patents: 5,758,644; 5,823,950; 6,011,986; 6,157,850; 6,263,222; 6,501,975; and other applicable patents listed at www.masimo.com/patents.htm

.

92

Specifications Welch Allyn Spot Vital Signs LXi

Nellcor Sensor Accuracy Guide

Accuracy specifications are based on controlled hypoxia studies with healthy, non-smoking adult volunteers over the specified saturation SpO

2

SpO

2 readings were compared to SaO

2 range. Pulse oximeter values of drawn blood samples measured by hemoximetry. All accuracies are expressed as + “X” digits. This variation equals + one standard deviation (+ 1 SD), which encompasses 68% of the population.

Table 2. OxiMax Sensor Models, Single Patient Use

Sensor Models

MAX-AI

1

SpO

2

Range

70% to 100%

± 2

MAX-PI* ± 2

MAX-II

MAX-RI

2

1

The accuracy specification under motion conditions is ± 3. For a definition of motion, contact Nellcor

Technical Services or your local Nellcor representative.

2

The accuracy specification has been determined between saturations of 80% to 100%.

± 2

± 3.5

Table 3. OxiCliq Sensor Models, Single Patient Use

Sensor Models

OXICLIQ-PI

SpO

2

Range

70% to 100%

± 2.5

Table 4. Reusable Sensor Models

Sensor Models

D-YS (Infant to Adult)

D-YS and D-YSE

D-YS and D-YSPD

DS-100A

OXI-A/N (Adult/neonate)

OXI-P/I (Pediatric/infant)

SpO

2

Range

70% to 100%

± 3

± 3.5

± 3.5

± 3

Adult: ± 3

Neonate: ± 4

± 3

Nellcor Patents

Covered by one or more of the following U.S. patents and foreign equivalents:

5,485,847; 5,676,141; 5,743,263; 6,035,223; 6,226,539; 6,411,833; 6,463,310; 6,591,123;

6,708,049; 7,016,715; 7,039,538; 7,120,479; 7,120,480; 7,142,142; 7,162,288; 7,190,985;

7,194,293; 7,209,774; 7,212,847; 7,400,919.

Service Manual Specifications

93

Mechanical

Dimensions

Weight

Mounting

Portability

Height: 10.63“ (27 cm)

Length/Braun: 8” (20.32 cm)

Length/SureTemp Plus: 7.5” (19 cm)

Depth: 5.25” (13.34 cm)

7.5 lbs (3.4 kg)

Self-supporting on rubber feet

Custom mobile stand

Custom wall mount

May be hand-carried when held by the rear handle

Electrical

Power Requirements

Patient-rated isolation transformer is connected to AC mains:

100-240V, 50-60Hz, 0.4A or internal power: 6Vdc, 6Ah

Patents

For patent information, please visit www.welchallyn.com/patents

Batteries

Spot Vital Signs LXi

Braun ThermoScan PRO 4000 thermometer

External Printer

Sealed Lead Acid Battery (4500-84) with external charger (4500-101A)

Rechargeable Battery Pack (53020-0000)

Ni-Mh Rechargable Battery Pack (4500-505)

Includes a separate charger (713703 or BC1300M)

• The Spot Vital Signs LXi battery is 90 to 100% charged after 6 hours of charging. The

Spot LXi battery automatically charges when Spot LXi is powered through the AC power transformer. An operator can use the device while the battery is charging; however, the battery charges faster when the instrument is not in operation.

• The rechargeable batteries in the Braun ThermoScan PRO 4000 thermometer require an additional 1 hour to charge.

• Charge the external printer battery 9 hours before initial use.

Environmental

WARNING This device is not suitable for use in the presence of a flammable anesthetic mixture with air or oxygen or nitrous oxide. An explosion may result.

94

Specifications Welch Allyn Spot Vital Signs LXi

Operating Temperature

Storage/Transport Temperature

Relative Humidity

Operating Altitude

Radio

Mechanical

Dimensions

Mounting

Operating Temperature

Storage Temperature

50° to 104° F (10° to 40° C)

Device with SureTemp Plus: -13

 to 131F (-25 to 55C)

Device with Braun ThermoScan PRO 4000: -4

 to 122F (-20 to 50C)

15 to 95% (non-condensing)

-557 to 16,000 ft. (-170 to 4877 m)

3.93” L x 2.38” W x 1.42” T

99.78mm L x 60.5 mm W x 36.1mm T

Nylon Bracket

-40° C to +85° C

-40° C to +125° C

Configuration Options

DHCP Enabled or

Not

Encryption

If not DHCP:

• IP Address

• Subnet Address

• Default

Gateway

Address

• Primary DNS

Wireless Security

Mode

Disabled

WPA2 Security

(WPA2-PSK and

WPA2-AES only)

CCMP

EAP Settings

(WPA2-AES only)

EAP Type

Channel

Default WEP

Key 1-4

SSID

WPA Pass

Phrase

WPA2 Pass

Phrase

WEP Authentication

PSK Format Hexadecimal

WEP 64-bit

CCMP and TKIP

TLS

ASCII

WEP 128-bit

TTLS

WPA2-PSK

PEAP

WPA2-AES

Auto

Service Manual Specifications

95

Guidance and Manufacturer’s Declaration For Spot Vital

Signs LXi Radio

Emissions and Immunity Information

Electromagnetic Emissions

The Spot Vital Signs LXi is intended for use in the electromagnetic environment specified below. The customer or user of the Spot Vital Signs

LXi should assure that it is used in such an environment.

Emissions Test

RF emissions

CISPR 11

RF emissions

CISPR 11

Harmonic emissions

IEC 61000-3-2

Voltage fluctuations/ flicker emissions

IEC 61000-3-3

Compliance

Group 1

Class B

Class A

Complies

Electromagnetic Environment - Guidance

The Spot Vital Signs LXi uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.

The Spot Vital Signs LXi is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.

Electromagnetic Immunity

The Spot Vital Signs LXi is intended for use in the electromagnetic environment specified below. The customer or user of the Spot Vital Signs

LXi should assure that it is used in such an environment.

Immunity Test IEC 60601

Test Level

± 6 kV contact Electrostatic discharge

(ESD)

± 8 kV air

IEC 61000-4-2

Electrical fast transient/ burst

±2 kV for power supply lines

IEC 61000-4-4

Surge

IEC 61000-4-5

Compliance Level

± 6 kV contact

± 8 kV air

±2 kV for power supply lines

Electromagnetic Environment - Guidance

Floors should be wood, concrete, or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.

Mains power quality should be that of a typical commercial or hospital environment.

±1 kV for input/output lines

±1 kV differential mode

±1 kV for input/output lines

±1 kV differential mode Mains power quality should be that of a typical commercial or hospital environment.

±2 kV common mode ±2 kV common mode

96

Specifications Welch Allyn Spot Vital Signs LXi

Voltage dips, short interruptions, and voltage variations on power supply input lines.

IEC 61000-4-11

Power frequency

(50/60Hz) magnetic field

IEC 61000-4-8

>95% dip in 0.5 cycle

60% dip in 5 cycles

30% dip for 25 cycles

>95% dip in 5 seconds

3 A/m

>95% dip in 0.5 cycle

60% dip in 5 cycles

30% dip for 25 cycles

>95% dip in 5 seconds

3 A/m

Mains power quality should be that of a typical commercial or hospital environment. If the user of the Spot Vital Signs LXi requires continued operation during power mains interruptions, it is recommended that the Spot Vital Signs LXi be powered from an uninterruptible power supply or battery.

Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.

Electromagnetic Immunity

The Spot Vital Signs LXi is intended for use in the electromagnetic environment specified below. The customer or user of the Spot Vital Signs

LXi should assure that it is used in such an environment.

Immunity Test IEC 60601

Test Level

Conducted RF

IEC 61000-4-6

Radiated RF

IEC 61000-4-3

3 Vrms

150 kHz to 80 MHz

3 V/m

80 MHz to 2.5 GHz

Compliance

Level

Electromagnetic Environment - Guidance

Portable and mobile RF communications equipment should be used no closer to any part of the Spot Vital Signs LXi, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.

3 Vrms

Recommended separation distance

d = (1.17)

P

3 V/m d = (1.17)

P

80 MHz to 800 MHz

d = (2.33)

P

800 MHz to 2.5 GHz where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in metres (m).

Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, a should be less than the compliance level in each frequency range.

b

Interference may occur in the vicinity of equipment marked with the following symbol:

Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.

Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.

a

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio,

AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Spot

Vital Signs LXi is used exceeds the applicable RF compliance level above, the electrocardiograph should be observed to verify normal operation.

If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the electrocardiograph.

b

Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

Service Manual Specifications

97

Recommended Separation Distances Between Portable and Mobile RF Communications Equipment and the Spot Vital Signs LXi

The Spot Vital Signs LXi is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or user of the Spot Vital Signs LXi can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Spot Vital Signs LXi as recommended below, according to the maximum output power of the communications equipment.

Rated Max. Output

Power of Transmitter

(W)

0.01

0.1

1

10

100

Separation Distance According to Frequency of Transmitter (m)

150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz

d = (1.17)

P

0.117

0.37

1.17

3.70

11.70

d = (1.17)

0.117

0.37

1.17

3.70

11.70

P

d = (2.33)

0.233

0.74

2.33

7.37

23.30

P

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.

Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.

98

Specifications Welch Allyn Spot Vital Signs LXi

8

Troubleshooting

This chapter provides helpful information in troubleshooting the Spot Vital Signs LXi.

Error Codes

The following tables of error codes provide a quick reference of the descriptions and probable causes of error codes.

Table 1. Informational Error Codes

Internal

Event Log

Code

Number

_n/a

Displayed

Device Code

Internal Event Log Description

_n/a

_n/a

_n/a

_n/a

_n/a

_n/a

_n/a

_n/a

_n/a

_n/a

_n/a

Corrective Action

Copyright notice displayed on powers up splash screen.

Message displays when the user attempts to start another print while a print job is currently printing.

Printing is already in progress

Clear current reading.

Erase all readings.

Information screen telling the user how to operate the location identifier edit menu in Configuration Mode.

Message displays when Mod F cannot perform a STEP NIBP measurement and switches to normal method.

Use the navigation keys to edit characters and the SELECT key to stop editing.

Restrict patient movement.

Message displays when in MEM mode and there are no readings to display.

External devices configuration menu message when the user tries to exit the menu having enabling the information system, but not that Patient ID.

Message displays below txtPIDMsgB.

User must enable the Patient ID if using the information system.

Information displays when the user enables device in external devices configuration menu.

Message displays below txtPIDMsgA

User must press SELECT and correct the problem.

User must connect the Spot Vital Signs

LXi to an external device.

Line displays below error/information message if the user can clear the error/ condition.

User must press SELECT when completed.

User must press ACCEPT to clear.

99

100

Troubleshooting Welch Allyn Spot Vital Signs LXi

Table 1. Informational Error Codes

Internal

Event Log

Code

Number

_n/a

Displayed

Device Code

Internal Event Log Description

_n/a

_n/a

Corrective Action

User decides whether to reset everything back go factory default values.

Error condition occurs causing the unit to automatically shut down. The user cannot clear this message.

Spot Vital Signs LXi will shut down in 1 minute.

User confirms to start printing.

22

23

24

25

19

20

21

15

16

17

18

Table 2. Blood Pressure Error Codes

11

12

13

Internal

Event Log

Code

Number

10

Displayed

Device Code

Internal Event Log Description

C01

C03

E10

C04

User abort.

Inflation too quick.

Overpressure

Major air leak.

14 C05 Excessive arm movement

C07

C07

C02

C04

E20

E20

Auto cycle skipped, system busy.

Auto cycle skipped, NIBP reached too soon.

Auto zero failure.

Could not reach target pressure.

Missed A/D sample.

A/D timeout failure.

NIPB NOT AVAILABLE.

26

E11

C06

C05

C12

C13

Malfunction

Reinflation needed.

Cycle took too long.

Ambient temp out of range.

Battery failure

Corrective Action

Check tubing and connections.

Check patient.

Excessive inflation time; check for air leaks.

Excessive noise, check cuff placement; restrict patient movement.

Device will power down.

Device will power down.

Auto zero failure; check air obstruction; limit patient movement.

Excessive inflation time; check for air leaks.

General internal malfunction. Call

Technical Support.

General internal malfunction. Call

Technical Support.

Correct probe tip conditions. If problem persists, replace probe. If problem still persists, replace module.

User should try another cycle. If problem persists, replace probe.

Measurement was outside of device’s measurement range.

Check cuff placement; restrict patient movement.

Check internal temperature at probe well.

Verify ambient temperature does not exceed operating specification.

Battery failure; use wall transformer.

Service Manual Troubleshooting

101

Table 2. Blood Pressure Error Codes

Internal

Event Log

Code

Number

27

Displayed

Device Code

Internal Event Log Description

E11

28

29

32

33

34

41

42

43

44

45

46

47

48

49

50

51

52

E11

C02

C06

C06

C05

C05

C05

C05

C05

C05

C05

C05

C05

C05

Corrective Action

> 10mm residual pressure > minutes.

>15mm residual pressure >3 mins.

Couldn’t bleed pressure < 9 mmHg

Verify that the supply voltage is within specification.

Verify that the supply voltage is within specification.

Auto zero failure, check tubing for kinks and connection integrity, limit patient movement.

Try another predict cycle. If the problem persists, replace probe.

Replace probe.

Thermistor pulled away from the tip or heater broken.

Probe not responsive. Probe not characterized/calibrated.

Probe not characterized/calibrated.

Low inflation; first step had largest value, first step’s value > systolic value.

Replace probe.

Measurement was outside of Spot Vital

Signs LXi measurement range.

Check connections; restrict patient movement.

Dump too soon; last step had largest value, next to last step’s value > diastolic value; will not normally occur in autodump mode.

Step too big systolic; difference in pressure between step above and step below systolic is too great.

Step too big diastolic; difference in pressure between step above and step below diastolic is too great.

Range error: calculated BP is outside of measurement range.

Excessive noise; check cuff placement and limit movement.

Excessive noise; check cuff placement and limit movement.

Excessive noise; check cuff placement and limit movement.

Pressure reverse systolic; pressure of step above systolic is lower than pressure of step below systolic.

Excessive noise; check cuff placement and limit movement.

Pressure reverse diastolic; pressure of step above systolic is lower than pressure of step below diastolic.

Excessive noise; check cuff placement and limit movement.

Step too big MAP; difference in pressure between adjacent steps near MAP is too great.

Too few steps; less than 5 steps

Excessive noise; check cuff placement and limit movement.

Excessive noise; check cuff placement and limit movement.

Bad value diastolic; value of step below diastolic and step below that step are not consistent.

Excessive noise; check cuff placement and limit movement.

Value reverse systolic; value of step above systolic is greater than value of step below systolic.

Excessive noise; check cuff placement and limit movement.

Value reverse diastolic; value of step above systolic is greater than value of step below diastolic.

Excessive noise; check cuff placement and limit movement.

102

Troubleshooting Welch Allyn Spot Vital Signs LXi

Table 2. Blood Pressure Error Codes

Internal

Event Log

Code

Number

53

Displayed

Device Code

Internal Event Log Description

C05

54

55

56

57

58

59

60

61

62

63

64

65

66

C05

C05

C05

C05

C05

C05

C05

C05

C05

C05

C05

C05

C05

Corrective Action

No info - can’t be produced by Mod F if they don’t exist in the code.

No info - can’t be produced by Mod F if they don’t exist in the code.

No info - can’t be produced by Mod F if they don’t exist in the code.

Excessive noise; check cuff placement and limit movement.

Excessive noise; check cuff placement and limit movement.

Excessive noise; check cuff placement and limit movement.

Excessive noise; check cuff placement and limit movement.

Out of memory; deflation contains too many steps. SW sets allowable limit of steps that can be stored and used for BP determination for a given cycle.

Retarget error; cycle tried to retarget too many times. Only one allowed retarget allowed per cycle.

Repump error: Tried to repump too many times. Limit is set to 3 repumps. This can occur on a cycle with bad signal to noise ratio or if a subject’s pulses are so small that pulses above the diastolic ratio are not detectable with the lowest allowed thread threshold.

Systolic curve fit error; the steps on the systolic side of the envelope could not be curve fit. Causes include: too few or too many steps sent to curve fit routine; math error in curve fit routine.

Excessive noise; check cuff placement and limit movement.

Excessive noise; check cuff placement and limit movement.

Excessive noise; check cuff placement and limit movement.

Diastolic curve fit error; the steps on the systolic side of the envelope could not be curve fit. Causes include: too few or too many steps sent to curve fit routine; math error in curve fit routine.

Systolic slope error; slope of curve fit of systolic points has a slope out of range.

Can occur with bad cuff fit, rapidly changing BP etc.

Excessive noise; check cuff placement and limit movement.

Excessive noise; check cuff placement and limit movement.

Diastolic slope error; slope of curve fit of diastolic points has a slope out of range.

Can occur with bad cuff fit, rapidly changing BP etc.

Systolic extrapolation error; the systolic pressure that is represented by the systolic ratio is outside of the data sent to the curve fit routine.

Excessive noise; check cuff placement and limit movement.

Excessive noise; check cuff placement and limit movement.

Diastolic extrapolation error; the systolic pressure that is represented by the diastolic ratio is outside of the data sent to the curve fit routine.

Excessive noise; check cuff placement and limit movement.

Peak range error; the height of the peak of the envelope is outside normal limits.

Excessive noise; check cuff placement and limit movement.

Reinflation too late.

Excessive noise; check cuff placement and limit movement.

Service Manual Troubleshooting

103

68

69

70

78

79

5600

5601

5602

5603

5604

Table 2. Blood Pressure Error Codes

Internal

Event Log

Code

Number

67

Displayed

Device Code

Internal Event Log Description

C06

Err

C06

C06

C06

C06

Err

Err

Err

Err

Err

Corrective Action

High heart rate.

Unable to determine blood pressure; pulse rate may be out-of-range.

Too many reinflates.

Problem reading the module's EEPROM or saving to the module’s EEPROM.

Restrict patient movement.

Verify that the supply voltage is correct.

Try to make changes. If problem persists, replace module.

Problem reading the probe’s EEPROM.

Verify that the supply voltage is correct.

Try to make changes.

The probe non-volatile error log memory error detection mechanism detected error.

Measurement was outside of device’s measurement range.

The module’s non-volatile error log memory error detecting mechanism detected an error.

Invalid ModF algorithm.

Measurement was outside of device’s measurement range.

Invalid ModF init code.

Invalid ModeF PMode.

Invalid ModF target pressure.

Invalid ModF configuration.

87

88

85

86

83

84

81

82

Table 3. POST Error Codes

Internal

Event Log

Code

Number

80

Displayed

Device Code

Internal Event Log Description

E30 FPROM checksum error.

E35

E36

E37

E38

E31

E32

E33

E34

RAM test failure.

Factory EEprom checksum error.

User EEprom checksum error.

A/D converter failure.

SpO2 module failure.

Diatek module failure

Printer problem.

Real time clock failure.

Corrective Action

General internal malfunction; contact

Technical Support.

General internal malfunction; contact

Technical Support.

General internal malfunction; contact

Technical Support.

General internal malfunction; contact

Technical Support.

General internal malfunction; contact

Technical Support.

General internal malfunction; contact

Technical Support.

General internal malfunction; contact

Technical Support.

General internal malfunction; contact

Technical Support.

Date and time not set; see “Set the Date and Time Configurations” in the

Directions for Use.

104

Troubleshooting Welch Allyn Spot Vital Signs LXi

90

91

92

93

94

95

Table 3. POST Error Codes

Internal

Event Log

Code

Number

89

Displayed

Device Code

Internal Event Log Description

E39 No calibration signature.

E40

E41

E42

E43

Err-E44

E44 E45

Bad NIBP calibration record.

Stepper valve bad.

8bit substitute for 16bit errors.

Error log checksum error.

POST que full.

POST que full.

Corrective Action

General internal malfunction; contact

Technical Support.

General internal malfunction; contact

Technical Support.

General internal malfunction; contact

Technical Support.

Internal communications error.

Disconnect the battery and wait 5 minutes. Reconnect the battery, and then see “Set the Date and Time

Configurations” in the Directions for Use.

General internal malfunction; contact

Technical Support.

General internal malfunction; contact

Technical Support.

General internal malfunction; contact

Technical Support.

128

129

130

131

124

125

126

127

Table 4. SureTemp Plus Temperature Error Codes

Internal

Event Log

Code

Number

121

Displayed

Device Code

Internal Event Log Description

E0.4

122 E0.1

Probe is over temperature (112° F /

44.4° C).

Excessive heater energy.

123 23 Host interface error.

30

31 bv

29

26 b^

A^!

Av!

Device outside operating temperature range.

Device outside operating temperature range.

Invalid SureTemp algorithm.

Battery voltage over the algorithm’s maximum value.

Battery voltage under the algorithm’s minimum value.

Battery voltage not set.

Prediction algorithm not set.

Ambient temperature not set.

Corrective Action

Correct probe tip conditions. If problem persists, replace probe.

Try another predict cycle. If problem persists, replace probe.

Try another predict cycle. If problem persists, replace probe.

Decrease ambient temperature.

Increase ambient temperature.

Try another predict cycle. If problem persists, replace probe.

Contact Technical Support.

Contact Technical Support.

Try another predict cycle. If problem persists, replace probe.

Try another predict cycle. If problem persists, replace probe.

Try another predict cycle. If problem persists, replace probe.

Service Manual Troubleshooting

105

Table 4. SureTemp Plus Temperature Error Codes

Internal

Event Log

Code

Number

132

Displayed

Device Code

Internal Event Log Description

E0.5

133

134

145

146

147

148

149

150

151

161

162

163

165

152

159

160

166

E0.9

E0.9

E0.2

E0.2

E4.1

E4.1

E4.0

E4.0

E0.2

E0.9

bv b^

E4.2

E4.8

E4.9

E4.3

E0.8

Corrective Action

No probe temperature rise.

Try another predict cycle. If problem persists, replace probe.

Calibrate probe or replace probe.

Probe EEPROM is uninitialized. Bad probe gain.

Probe EEPROM is uninitialized. Bad probe response value.

A/D measurement overrun - pulse too long.

Calibrate probe or replace probe.

A/D Measurement underrun - pulse too short.

A/D measurement overrun - RCAL pulse too long.

A/D measurement underrun - RCAL pulse too short.

Patient or environmental temperature conditions may be too low. Verify conditions at 80° F and 50° F respectively. If conditions are valid and problem persists, replace probe. If problem persists, replace the SureTemp

Plus Thermometer module.

Patient or environmental temperature conditions may be too high. Verify conditions at 110° F and 104° F respectively. If conditions are valid and problem persists, replace probe. If problem persists, replace the SureTemp

Plus Thermometer module.

If problem persists, replace the SureTemp

Plus Thermometer module.

If problem persists, replace the SureTemp

Plus Thermometer module.

A/D measurement overrun - PTB pulse too long.

If problem persists, replace the SureTemp

Plus Thermometer module.

A/D measurement underrun - PTB pulse too short.

If problem persists, replace the SureTemp

Plus Thermometer module.

A/D measurement overrun - timeout.

Probe EEPROM is uninitialized.

Displayed when battery voltage is too low.

Displayed when battery voltage is too high.

If problem persists, replace the SureTemp

Plus Thermometer module.

Bad probe A/D calibration.

Reference power supply malfunction.

If problem persists, replace the SureTemp

Plus Thermometer module.

Calibrate device.

Device not calibrated.

Missing probe well - cycle will not start.

Insert probe well.

Device EEPROM save failure.

Verify that the batteries are fresh. Try to make changes again. If problem persists, replace the SureTemp Plus Thermometer module.

EEPROM CRC Failure.

(CONFIG_TRANSFER_FAILURE)

Verify that the batteries are fresh. Try again, if problem persists, replace the component that is indicated.

106

Troubleshooting Welch Allyn Spot Vital Signs LXi

Table 4. SureTemp Plus Temperature Error Codes

Internal

Event Log

Code

Number

167

Displayed

Device Code

Internal Event Log Description

E4.4

168

169

170

171

172

173

174

175

180

181

182

183

184

185

E4.5

E4.6

E0.6

71

72

73

E4.7

E0.7

E5.0

E5.0

E5.0

E5.0

E5.0

E5.2

Corrective Action

Device EEPROM read/save failure.

(CONFIG_START_FAILURE)

Device EEPROM read/save failure.

(CONFIG_SLAVE_ACK_FAILURE)

Device EEPROM read/save failure.

(CONFIG_MASTER_ACK_FAILURE)

Verify that the batteries are fresh. Try to make changes again. If problem persists, replace the SureTemp Plus Thermometer module.

Verify that the batteries are fresh. Try to make changes again. If problem persists, replace the SureTemp Plus Thermometer module.

Verify that the batteries are fresh. Try to make changes again. If problem persists, replace device.

Probe EEPROM read failure.

(CONFIG_NULL_CHAR_FAILURE)

Verify that the batteries are fresh. Try to make changes again. If problem persists, replace the SureTemp Plus Thermometer module.

Try another predict cycle. If problem persists, replace probe.

Device EEPROM access failure.

(CONFIG_ACQUIRE_FAILURE)

Device EEPROM access failure.

(CONFIG_RELEASE_FAILURE)

Device EEPROM access failure.

(CONFIG_INVALID_PTR_FAILURE)

Device EEPROM initialization failure.

(CONFIG_INITIALIZE_FAILURE)

Probe EEPROM is un-initialized.

(CONFIG_UNINITIALIZED_PROBE)

Try another predict cycle. If problem persists, replace probe.

Try another predict cycle. If problem persists, replace probe.

If problem persists, replace the SureTemp

Plus Thermometer module.

Verify that the batteries are fresh. If problem persists, replace probe.

Heater stuck on. (ERROR_HEATER_ON) If problem persists, replace the SureTemp

Plus Thermometer module.

Heater stuck off. (ERROR_HEATER_OFF) If problem persists, replace the SureTemp

Plus Thermometer module.

Heater on when should be off.

(ERROR_HEATER_VOLTAGE_HIGH_Q)

Heater on when should be off.

(ERROR_HEATER_VOLTAGE_Q_HIGH_

IMPEDANCE)

Heater is off when it should be on.

(ERROR_HEATER_VOLTAGE_LOW)

Heater hardware failsafe failure.

(ERROR_HEATER_FAILSAFE)

If problem persists, replace the SureTemp

Plus Thermometer module.

If problem persists, replace the SureTemp

Plus Thermometer module.

If problem persists, replace the SureTemp

Plus Thermometer module.

If problem persists, contact Technical

Support.

Table 5. SpO

2

Error Codes

Internal

Event Log

Code

Number

1

3

Displayed

Device Code

Internal Event Log Description

--

Missing or disconnected sensor.

No pulse from sensor.

Corrective Action

Adjust or replace sensor.

Adjust or replace sensor.

Service Manual Troubleshooting

107

5

6

Table 5. SpO

2

Error Codes

Internal

Event Log

Code

Number

4

Displayed

Device Code

Internal Event Log Description

E7 Communications failure.

C8

C9

8-second timeout interval

Defective or bad SpO

2 sensor.

SpO

2 time limit exceeded.

Corrective Action

Retake reading; contact Technical

Support.

Replace sensor.

Remove sensor from patient.

108

Troubleshooting Welch Allyn Spot Vital Signs LXi

Table 6. Comms Errors

1013

1014

1015

1016

1017

1018

1019

1009

1010

1011

1012

1005

1006

1007

1008

1001

1002

1003

1004

Internal

Event Log

Code

Number

1000

Displayed

Device Code

Internal Event Log Description

Err

Err

Err

Err

Err

Err

Err

Err

Err

Err

Err

Err

Err

Err

Err

Err

Err

Err

Err

Err

Corrective Action

Invalid Operation in this Context.

Auto zero in progress.

Bad checksum.

Unimplemented command.

Buffer Overflow.

Overrun error.

Wrong data count.

Internal error, contact Technical Support.

Internal error, contact Technical Support.

Internal error, contact Technical Support.

Internal error, contact Technical Support.

Internal error, contact Technical Support.

Internal error, contact Technical Support.

Internal error, contact Technical Support.

FPROM erase error.

FPROM programming error

Bad Motorola record.

A/D channel locked.

Internal error, contact Technical Support.

Internal error, contact Technical Support.

Internal error, contact Technical Support.

Internal error, contact Technical Support.

RTC not present.

Internal error, contact Technical Support.

Attempt to reference nonexisting cycle.

Internal error, contact Technical Support.

Data range error.

Alarm pair range error.

Internal error, contact Technical Support.

Internal error, contact Technical Support.

No POST error to clear.

Can’t clear this POST error.

Command not command type.

NAK from ST3800 protocol.

No wireless link.

Internal error, contact Technical Support.

Internal error, contact Technical Support.

Internal error, contact Technical Support.

Internal error, contact Technical Support.

Internal error, contact Technical Support.

Table 7. Comms Host Errors

2002

2003

2004

2005

2006

Internal

Event Log

Code

Number

2001

Displayed

Device Code

Internal Event Log Description

Err

Err

Err

Err

Err

Err

Timeout.

Incorrect response header.

Incorrect response checksum.

Too much received data.

User cancelled.

User paused.

Corrective Action

Internal error, contact Technical Support.

Internal error, contact Technical Support.

Internal error, contact Technical Support.

Internal error, contact Technical Support.

Internal error, contact Technical Support.

Internal error, contact Technical Support.

Service Manual Troubleshooting

109

Table 8. Printer Errors

Internal

Event Log

Code

Number

5400

Displayed

Device Code

Internal Event Log Description

Err

5401

5402

5403

Err

Err

Err

Corrective Action

Problem with the printer

Displayed when printing started and print head is not closed in place.

Check all options; if problem persists, contact Technical Support.

Printer is not responding.

Displayed when print started and printer is out of paper.

Printer head will not shut all the way or too much paper.

Add paper.

Table 9. Spot Vital Signs LXi Errors

Internal

Event Log

Code

Number

5500

5501

5502

Displayed

Device Code

Internal Event Log Description

Err

Err

Corrective Action

Unknown error code.

Displayed when battery is too low for unit to continue operating.

Charge battery; use wall transformer.

Key stuck error causing the unit to be automatically shut down.

110

Troubleshooting Welch Allyn Spot Vital Signs LXi

Table 10. 802.11 b Radio Errors

LED LED Color

Power Off

Red

Amber

Link

Green

Off

Red

Green

Comm Off

Red

Blinking Red

Amber

Blinking Green

Green

Function

Bridge is not receiving power.

Bridge failed its Power On Self Test (POST) and is not configured for wireless communication.

Bridge passed its POST but is not configured for wireless communication.

Bridge passed its POST and is configured for wireless communication.

Bridge is not receiving power.

Bridge is searching for an Access Point.

Wireless network and MAC have associated with an Access Point.

No power or no wireless TCP session is established and no physical serial connection is detected.

No wireless TCP session is established; a physical connection is detected.

A physical serial connection was detected and there is serial traffic present on that connection, but no wireless TCP session is established.

A wireless TCP connection is established but no physical serial connection is detected (i.e. no serial cable is attached to the Bridge).

A wireless TCP session is established, a physical serial connection is detected, and the Bridge is transmitting or receiving data across the wired serial port.

A wireless TCP session is established, a physical serial connection is detected, but there is n o active data movement across the wired serial port.

NOTE:The 802.11a/b/g radio has no functional LED displays.

Radio Test

Refer to Manufacturer’s Directions for Use.

Service Manual

Service Work Checklist

Spot Vital Signs

Model Number Serial Number BP Cycle Count Technician

Troubleshooting

111

Date

Test

Unit SW Version

SPO2 SW Version

Thermometry SW Version

Unit Pressure @ 0 mmHg

Unit Pressure @ 50 mmHg

Unit Pressure @ 150 mmHg

Unit Pressure @ 250 mmHg

Test Voltage @ 5.5 V

Current Test – Sleep State

Current Test – Patient DAQ

Current Test – Valve Pump

Noise Level

Button Test

Interface Test

Leak Test

Dump Test

Inflation Test

Valve Control Test

Over Pressure Test

SPO2 Sensor - Masimo

SPO2 Sensor - Nellcor

Temperature Calibration Suretemp Plus

Temperature Calibration Braun Pro 4000

Temperature Cal Key Suretemp Plus

Radio Ping Test

Test Data

Pass or Fail

Pass/Fail

N/A

N/A

N/A

Test Specification

N/A

N/A

N/A

0 mmHg (+/- 1 mmHg)

50 mmHg (+/- 1 mmHg)

150 mmHg (+/- 1.5 mmHg)

+/- 2.0 mmHg

+0.1 VDC

< 0.5 A

< 1.25 A

< 2.0 A

< 0.05 mmHg

Pass/Fail

Pass/Fail

<= 6mmHg in 15 seconds

< 10 Seconds

< 7 Seconds

Pass/Fail

296 to 329 mmHg

81% +/- 3% 61bpm +/- 1bpm

90% +/- 1% 60bpm +/- 1bpm

96.8 +/- 0.2°F 105.8 +/- 0.2°F

96.8 +/- 0.2°F 105.8 +/- 0.2°F

97.3 +/- 0.2°F

Pass

112

Troubleshooting Welch Allyn Spot Vital Signs LXi

Spot LXi Radio Configuration Tool

This CD configures the Spot Vital Signs LXi Radio to a local network. See the Welch Allyn

Spot LXi Radio Configuration Procedure (80012309) for further information.

9

Supplies and Accessories

Blood Pressure

Table 1. Reusable Blood Pressure Cuffs (1 per pack)

REF

4500-01

4500-02

Size

Child

Adult

REF

4500-03

4500-04

Size

Large Adult

Thigh

Table 2. Reusable Blood Pressure Cuffs (1 per pack)

REF Size

REUSE-07-2MQ Infant

REUSE-08-2MQ Small Child

REUSE-09-2MQ Child

REUSE-10-2MQ Small Adult

REF Size

REUSE-11-2MQ Adult

REUSE-12-2MQ Large Adult

REUSE-13-2MQ Thigh

Table 3. Disposable Blood Pressure Cuffs (5 per box)

REF Size

SOFT-07-2MQ Infant

SOFT-08-2MQ Small Child

SOFT-09-2MQ Child

SOFT-10-2MQ Small Adult

REF Size

SOFT-11-2MQ Adult

SOFT-12-2MQ Large Adult

SOFT-13-2MQ Thigh

Table 4. Miscellaneous Supplies and Accessories

REF

4500-30

4500-32

Description

Blood Pressure Hose (5ft/1.5M)

Blood Pressure Hose (8ft/2.4M)

REF

4500-31

Description

Blood Pressure Hose (10ft/3M)

113

114

Supplies and Accessories Welch Allyn Spot Vital Signs LXi

Temperature

Table 5. SureTemp Plus

REF

02895-000

02895-100

02894-0000

02894-1000

06138-000

01802-110

05031-101

20500-025M

400194

400195

400196

Description

SureTemp Plus Oral Probe and Well (9 feet/2.7M)

SureTemp Plus Rectal Probe and Well (9 feet/2.7M)

SureTemp Plus Oral Well

SureTemp Plus Rectal Well

SureTemp Plus Temperature Calibration Key

9600 Plus Calibration Tester

Disposable SureTemp Plus Probe Covers (1,000 covers, packaged 25/box)

Probe Cover-Box 25

Assy, Suretemp Plus Components, Spot Vital Signs LXi

PCA, Suretemp Plus OEM without EDAC Calibrated

PCA, EDAC, Spot Vital signs LXi

Table 6. Braun ThermoScan PRO 4000

REF

04000-200

04000-600

05075-800

04000-800

01802-110

53020-0000

4500-53

4500-54

Description

Braun ThermoScan PRO 4000 Thermometer (for North America, South America, and

Asia Pacific)

Braun ThermoScan PRO 4000 Thermometer (for Europe, Middle East, and Africa)

Braun ThermoScan PRO 4000 Disposable Probe Covers (Case of 800 covers for North

America, South America, and Asia Pacific)

Braun ThermoScan PRO 4000 Disposable Probe Covers (Case of 800 covers for Europe,

Middle East, and Africa)

9600 Plus Calibration Tester

Braun ThermoScan PRO 4000 Rechargeable Battery Pack

Braun Unlocking Key

Repair-Braun with rechargeable batteries

Service Manual Supplies and Accessories

115

Pulse Oximetry

Masimo Accessories

Table 7. Adhesive Sensors: Single-Patient Use

Catalog # Description

NEO-WRAP-RP Replacement Wrap Neo Adhesive

LNCS-INF-3 Disposable Adhesive Finger Sensor-Infant

INF-WRAP-RP Replacement Wrap Infant Adhesive

YI-AD

YI-FM

Multisite Adhesive Wrap AD/PED/NEO, YI

Sensor

Multisite Foam Wrap AD/PED/NEO, YI Sensor

LNCS-YI

LNCS-TC-I

Multisite Reusable Sensor

Reusable Ear Sensor

LNCS-NEO-L-3 Disposable Adhesive Finger Sensor- NEO/Adult

LNCS-ADTX Adhesive Finger Sensor - Adult (20 per case)

LNCS-PDTX Adhesive Finger Sensor - Pediatric (20 per case)

LNCS INF-L Adhesive Finger Sensor - Infant (20 per case)

LNOP-ADT

LNOP-PDT

Adult sensor

Pediatric sensor

LNOP-NEO Neonatal sensor

LNOP-NEOPT SofTouch neonatal preterm sensor

Weight Range

>30 kg

10 to 50 kg

3 to 20 kg

>66 lbs (30 kg)

22 to 110 lbs

(10 to 50 kg)

<22 lbs (10 kg)

<2.2 lbs (1 kg)

Quantity

LNCS-ADTX

LNCS-PDTX

LNCS INF-L

20

20

20

20

Table 8. Reusable Sensor

Catalog #

LNCS-TC-I

LNCS-YI

LNCS-DCI

Description

Reusable Ear Sensor

Multisite Reusable Sensor

Finger sensor - adult

LNCS-DCIP Finger sensor - pediatric

LNOP-DCI Finger clip probe - adult

LNOP-DCIP Finger clip probe - pediatric

Weight Range Quantity

>66 lbs (30 kg)

10 to 50 kg

>66 lbs (30 kg)

10 to 50 kg

Table 9. Sensor Cables

Catalog # Description

LNC-4-WA 4-foot cable with DB-9 connector for LNCS

LNC-10-WA 10-foot cable with DB-9 connector for LNCS

PC-04 4-foot cable with sensor connector

Weight Range

NA

NA

NA

Quantity

1

1

1

1

1

1

1

116

Supplies and Accessories

Table 9. Sensor Cables

Catalog #

PC-08

PC-12

Description

8-foot cable with sensor connector

12-foot cable with sensor connector

Welch Allyn Spot Vital Signs LXi

Weight Range

NA

NA

Quantity

1

1

Service Manual Supplies and Accessories

117

Nellcor Accessories

Table 10. OxiMax Adhesive Sensors: Single-Patient Use

Catalog #

MAX-AI

MAX-PI

Description

MAX-A Adhesive Sensor, adult

MAX-P Adhesive Sensor, pediatric

MAX-II MAX-I Adhesive Sensor, infant

MAX-RI MAX-R Adhesive Sensor, adult nasal

Weight Range

>66 lbs (30 kg)

22 to 110 lbs

(10 to 50 kg)

6.5 to 44 lbs

(3 to 20 kg)

>110 lbs (50 kg)

Quantity

Case of 24

Case of 24

Case of 24

Case of 24

Table 11. OxiMax OxiCliq

®

Sensors: Reusable Cable with Adhesive Sensor Bandage

Catalog #

OC-3

OXICLIQ-PI

Description

OxiCliq Sensor Cable (3 ft / 91cm)

OxiCliq P, pediatric

Weight Range

N/A

22 to 110 lbs

(10 to 50 kg)

Quantity

1

Case of 24

Table 12. OxiMax Reusable Sensors

Catalog #

DS-100A

OXI-A/N

Description

Durasensor

®

DS-100A finger-clip sensor, adult

Oxiband

®

OXI-A/N, adult/neonatal

OXI-P/I

D-YS

D-YSE

D-YSPD

Oxiband OXI-P/I, pediatric/infant

Dura-Y

®

D-YS, multisite sensor

D-YSE ear clip for Dura-Y sensor

PediCheck

D-YSPD pediatric spot-check sensor

Weight Range

>88 lbs (40 kg)

< 6.5 lbs or > 88 lbs

(<3 kg or >40 kg)

6.5 lbs to 88 lbs

(3 to 40 kg)

>2.2 lbs (1 kg)

>66 lbs (30 kg)

6.5lbs to 88 lbs

(3 to 40 kg)

Quantity

1

1

1

1

1

1

Table 13. OxiMax Sensor Cables

Catalog #

DOC-10

Description

DOC-10 (10 ft/3M)

Weight Range

N/A

Quantity

1

118

Supplies and Accessories

Miscellaneous

4500-505

4500-510

4500-520

4500-900

4500-905

4500-906

4500-907

4500-915

4500-202

4500-400

4500-402

4500-404

4500-406

4500-408

4500-450

4500-500

4500-920

4500-921

4500-922

4500-923

4500-13

4500-14

4500-75

4500-89

4500-150E

4500-100

4500-101A

4500-200

REF

4500-60

4500-62

4500-84

705310

705298

4500-11

4500-12

Description

Mobile Stand, Spot Vital Signs LXi

Wall Mount, Spot Vital Signs LXi

Lead Acid Battery

Directions for Use, Spot Vital Signs LXi English

Quick Reference Card

Child Inf. Bag, 2 Tube Monitor

Adult Bag, 2 Tube Monitor

Large Adult Inf. Bag, 2 Tube Monitor

Thigh Inf. Bag, 2 Tube Monitor

Kit, Desk Plate, Spot Vital Signs LXi

Service Manual, Spot Vital Signs LXi

In-Service CD, Spot Vital Signs LXi

Carrying Case, Spot Vital Signs LXi

AC Power Transformer (desktop transformer, line cord not included)

Cuffkit, Child, Adult, Thigh, 2T Cuff & Bag

Cuffkit, Small Child, Child, Small Adult 2T

Line Cord (United States/Canadian/Japanese version)

Line Cord (European version)

Line Cord (United Kingdom version)

Line Cord (Australian version)

Line Cord (South African version)

Transformer Mount, Spot Vital Signs LXi

Kit, Printer, Spot Vital Signs LXi

Printer Rechargeable Replacement Battery

Printer Paper (10 rolls)

Label Paper (10 rolls) and Cleaning Kit

Kit, Dev Conn SDK Developers

Repair Tool Software CD Kit

Spot Vital Signs LXi Configuration CD

Spot Vital Signs LXi Firmware Loader CD

Image Team 4600 2D Bar Code Scanner Kit

802.11b (DEPAC) Radio Kit

Radio Directions for Use

802.11 a/b/g Spot Vital Signs LXi Radio Kit

Spot Vital Signs LXi a/b/g Radio Directions for Use

Welch Allyn Spot Vital Signs LXi

Service Manual Supplies and Accessories

119

REF

4500-925

4500-926

4500-927

5200-08

Description

USB 2.0 Cable for Wired Connectivity

Cable for Wired Connectivity, Keyspan (USB to serial adapter)

Spot Vital Signs LXi USB/serial Cable Kit

Calibration T Fitting

Service Contracts

See the Welch Allyn Partners In Care™ web site ( http://www.welchallyn.com/promotions/ services/default.htm

) for a description of the different service contracts offered.

120

Supplies and Accessories Welch Allyn Spot Vital Signs LXi

10

Field Replaceable Units

The following list identifies the available FRUs for Spot Vital Signs LXi. To order an FRU, contact Welch Allyn Technical Support.

Table 1. Spot Vital Signs LXi Base Components

401250

421051-12

421051-9

59P586

620028-E

620217

700101

700102

700103

700105

700110

707237

700119

Component Number

400163

400170

403912

400200

400203

400207

400387

4500-901

4500-902

4500-84

4500-101A

400720

400732

401082

Object Description

ASSY, POWER HARNESS, Spot Vital Signs LXi

ASSY, PUMP HARNESS

PCA, MAIN, Spot Vital Signs LXi

BASE UNIT, SURETEMP, NELLCOR - Spot Vital Signs LXi

BASE UNIT COMPONENTS, Spot Vital Signs LXi

LANGUAGE COMPONENTS, ENG, Spot Vital Signs LXi

PBCA BP OEM, (MOD F)

Spot Vital Signs LXi Pump and Pump Tubing Kit

Spot Vital Signs LXi Pump Tubing Kit

Lead Acid Battery

AC Power Transformer

ASSY, BALLAST HARNESS, Spot Vital Signs LXi

ASSY, BATTERY, Spot Vital Signs LXi

Assy, Bezel and Window, Spot Vital Signs LXi

Assy, Fan Unit Cooling

TUBING,1/8 X 1/4 X 2.50

TUBING,1/8 X 1/4 X 0.95

FOOT, RUBBER

LABEL, CAUTION

FITTING, “T”, 1/8 X 1/16X 1/8

HSG, HANDLE, LXi

HSG, HANDLE INSERT, LXi Ultra

HSG, REAR, LXi

HSG, BATT DOOR, Spot Vital Signs LXi

BUMPER, LCD, Spot Vital Signs LXi

SWITCH ARRAY, Spot Vital Signs LXi-New

INVERTER, LCD

121

122

Field Replaceable Units Welch Allyn Spot Vital Signs LXi

Table 1. Spot Vital Signs LXi Base Components

703077

703118

703354

703355

705284

705298

705310

703842

705653

704423

4500-89

401427

700863

704375

701040

701502

700847

421023-501

4500-500

4500-501

713703

HOUSING, FRONT, Spot Vital Signs LXi

HSG, PANEL, SPO2-BLANK, Spot Vital Signs LXi

SERVICE MANUAL, Spot Vital Signs LXi

LABEL, ICON ID, Spot Vital Signs LXi

CBL,FLX, 1/2mm,15POS,1.0”

FLEX, MAIN, Spot Vital Signs LXi

FITTING, PLASTIC L, 1/16 X 1/16

Valve,Pneumatic,6PSI,5V,TH

STANDOFF, PCB, BROACHING, #4 X .500 LG

BATTERY CLIP, AA, SINGLE

DISPLAY, LCD, 320X240, BLITE, MONO, 3.3V, 5.7"

FILTER, AIR, 1/8 BARBS

Spot Vital Signs LXi Warranty Card English

QRC, Spot Vital Signs LXi English

DFU, Spot Vital Signs LXi English

TUBING, SILICONE, .063ID X .125OD X 5.50

TUBING, SILICONE, .125IDx.25ODx1.75LG

ASSEMBLY, SPOT PUMP REPAIR

KIT, PRINTER, SPOT ULTRA (LXi)

Spot LXi PRINTER UPGRADE KIT

SPOT LXi PRINTER POWER SUPPLY

Table 2. SureTemp Plus Components

Component Number

02692-100

02894-0000

400194

400195

400196

700111

700112

700858

700859

70999-0000

02692-100

Object Description

PROBE ASSY,ORAL 9FT

ASSY,PRB WLL,BLUE,OEM,M690_692

ASSY, SURETEMP+, COMPONENTS, Spot Vital Signs LXi

PCA, SURETEMP+, OEM W/O EDAC, CALIBRATED

PCA, EDAC, Spot Vital Signs LXi

HOUSING, THERM OEM POD, Spot Vital Signs LXi

PLATE, POD MOUNTING, Spot Vital Signs LXi

CBL,FLX,1MM,6POS

CBL,FLX,1MM,8POS

LABEL,SURETEMP TECHNOLOGY

PROBE ASSY,ORAL 9FT

Service Manual

Table 3. Braun ThermoScan PRO 4000 Components

Component Number

400652

400831

700112

700820

700822

704365

700859

703161

703162

703164

703165

703192

Object Description

PCA, BRAUN INTERFACE, Spot Vital Signs LXi

BRAUN COMPONENTS, Spot Vital Signs LXi

PLATE, POD MOUNTING, Spot Vital Signs LXi

HSG, BOTTOM, BRAUN POD, Spot Vital Signs LXi

HSG, BRAUN POD, Spot Vital Signs LXi

WINDOW, BRAUN POD, Spot Vital Signs LXi

CBL,FLX,1MM,8POS

SECURITY LOCK, BRAUN POD, Spot Vital Signs LXi

LATCH, BRAUN POD, Spot Vital Signs LXi

CONTACT, BRAUN POGO, Spot Vital Signs LXi

SPRING, COMPRESSION

COVER, BRAUN SECURITY, Spot Vital Signs LXi

Table 4. Masimo SpO

2

Components

LNCS-DCI

LNCS-DCIP

LNCS-INF-L

LNCS-PDTX

LNOP-ADT

LNOP-DCI

LNOP-DCIP

LNOP-NEO

Component Number

008-0648-00

008-0692-01

008-0960-00

008-1013-00

400192

400205

400555

400610

704422

705651

711350

LNC-10-WA

LNC-4-WA

LNCS-ADTX

Object Description

MASIMO LNOP-DCI ADULT REUSABLE SENSOR

MASIMO - 8' INTERFACE CABLE

MASIMO-LNCS-DCI ADULT REUSABLE SENSOR

SP02 CABLE, LNCS, MASIMO, MINI-D, 4’, EXP

PCA, MASIMO OEM, Spot Vital Signs LXi

MASIMO SPO2 COMPONENTS, Spot Vital Signs LXi

PCA, MASIMO ADAPTOR, Spot Vital Signs LXi

ASSY, CABLE, FLAT FLEX, SPO2

HSG, PANEL, SPO2-MASIMO, LXi

HOUSING, FRONT MASIMO, Spot LXi

MASIMO MS-11 SP02, STD, SW V4.6.0.2

MASIMO, 10’ EXT CABLE, LNCS/DB-9

MASIMO, 4’ EXT CABLE, LNC/DB-9

MASIMO, LNCS-ADTX ADULT ADHSV, 20 CASE

MASIMO, LNCS-DCI ADULT REUSABLE SENSOR

MASIMO, LNCS-DCIP PED REUSABLE SENSOR

MASIMO, LNCS-INF-L INFANT ADHSV, 20 CASE

MASIMO, LNCS-PDTX PED ADHSV, 20 CASE

DISPOSABLE ADHESIVE FINGER CLIP PROBE-AD

REUSABLE FINGER CLIP PROBE-ADULT

REUSABLE FINGER CLIP PROBE-PEDIATRIC

DISP ADHESIVE FINGER CLIP PROBE-NEONATE

Field Replaceable Units

123

124

Field Replaceable Units

Table 4. Masimo SpO

2

Components

Component Number

LNOP-NEOPT

LNOP-PDT

PC-04

PC-08

PC-12

Object Description

DISPOSABLE FINGER SENSOR-NEONATE

DISP ADHESIVE FINGER CLIP PROBE-PEDIATRIC

4-FOOT CABLE W/ SENSOR CONN

8-FOOT CABLE W/ SENSOR CONN

12-FOOT CABLE W/ SENSOR CONN

Table 5. Nellcor SpO

2

Components

DEC-4

DS-100A

D-YS

D-YSE

D-YSPD

MAX-AI

MAX-II

MAX-PI

Component Number

400183

400201

400610

620377-1

704421

704870

DOC-10

MAX-RI

OC-3

OXI-A/N

OXICLIQ-PI

OXI-P/I

Object Description

PCA, NELCOR OEM, Spot Vital Signs LXi

NELLCOR SPO2 COMPONENTS, Spot Vital Signs LXi

ASSY, CABLE, FLAT FLEX, SPO2

“NELLCOR WORKS HERE” LABEL

HSG, PANEL, SPO2-NELLCOR, LXi

EVSPNELLCOR NELL3 PULSE OXIMETRY

CABLE, SPO2 EXTENSION, NELLCOR

EXTENSION CABLE 4 FT.

DURASENSOR,ADULT,DS-100A

DURA-Y OXYGEN TRANSDER (1 sensor-40 wraps)

EAR CLIP (USE WITH DURA-Y SENSOR)

PEDICHECK PEDIATRIC SPOT CHECK

NELLCOR OXIMAX ADULT SENSOR (SU CS/24)

NELLCOR OXIMAX INFANT SENSOR (SU CS/24)

NELLCOR OXIMAX PED SENSOR (SU CS/24)

NELLCOR OXIMAX ADULT NASAL SEN (CS/24)

NELLCOR OXICLIQ SENSOR CABLE

NELLCOR OXIBAND AD/NEO TRANSDUCER 1/50

PEDIATRIC OXYGEN TRANSDUCER

NELLCOR OXIBAND PED/INF TRANSDUCER 1/50

Welch Allyn Spot Vital Signs LXi

Service Manual Field Replaceable Units

125

Table 6. Radio Components

4500-922

4500-923

4500-905

4500-906

4500-907

4500-925

4500-926

4500-927

404684

404685

Component Number

708920

703380

704622

4500-915

703878

4500-920

4500-921

Object Description

Spot Vital Signs LXi Wireless Radio (A/B/G)

Spot Vital Signs LXi Wireless Radio Mounting Bracket

Spot Vital Signs LXi Wireless Radio Mounting Instructions

Image Team 4600 2D Barcode Scanner Kit

Null Modem 9-Pin M/F 4-40

Kit, Spot Vital Signs LXi 802.11b Radio

802.11 b radio Directions for Use

Kit, Wireless Spot Vital Signs LXi 802.11a/b/g Radio

802.11a/b/g Radio Directions for Use

Repair Tool CD, Spot Vital Signs LXi

Spot Vital Signs LXi Configuration CD

Spot Vital Signs LXi Firmware Loader CD

Cable for wired connectivity

Cable for wired connectivity, Keyspan (USB to serial adapter)

USB 2.0/serial cable kit

Spot Vital Signs LXi Firmware Upgrade CD in case

Spot Vital Signs LXi Radio Configuration CD in case

126

Field Replaceable Units Welch Allyn Spot Vital Signs LXi

11

Disassembly and Repair

This chapter provides the instructions for removing and replacing serviceable modules in the Spot Vital Signs LXi.

In general, re-assembly procedures are the reverse order of the disassembly procedures.

WARNING Electric shock hazard. There are no user-serviceable parts inside Spot

Vital Signs LXi other than battery replacement (see “Battery Replacement” on page 46). Only an individual specifically trained and approved for the repair and/or

verification of this specific Welch Allyn product may perform maintenance procedures specifically described in this manual. For service, refer the device to an Authorized Service Center.

Note

Always disconnect the sealed lead-acid battery in the Spot Vital Signs LXi before performing any repair function.

Tool list

Description

Blood Pressure Test Volume Repair Fixture

Spot LXi Service Tool Software

Service Manual, Spot Vital Signs LXi

Spot Vital Signs LXi Directions for Use

802.11b Radio

802.11b Radio Directions for Use

802.11a/b/g Radio

802.11a/b/g Radio Directions for Use

Tester, Calibration, 9600 Plus

SureTemp Oral Probe

Blood Pressure Y-Tube, No Fittings 1/8 Tube

Welch Allyn Cuff

Welch Allyn Blood Pressure Hose, 5ft.

Cable, USB A to 5 pin mini

DPAC Wireless Module

Needle Nose Pliers

Wire Cutter

Part Number

407672

4500-905

704432

705310

4500-920

4500-921

4500-922

4500-923

01802-110

02895-000

5082-183

Soft-11-2MQ

4500-30

704889

706514

Qty Source

1 Welch Allyn

1 Welch Allyn

1 Welch Allyn

1 Welch Allyn

1 Welch Allyn

1 Welch Allyn

1 Welch Allyn

1 Welch Allyn

1 Welch Allyn

1 Welch Allyn

1 Welch Allyn

1 Welch Allyn

1 Welch Allyn

1 Welch Allyn

1 Welch Allyn

1 Tool/Supply Store

1 Tool/Supply Store

127

128

Disassembly and Repair Welch Allyn Spot Vital Signs LXi

Description

Tweezers

#2 Phillips Screwdriver

T10 Torx Driver

4” PVC Pipe

Soldering Station

Cable Tie Tool - T9921

Part Number

GS2B

*Setra Pressure Meter

*Netech Pressure Meter

Nellcor Sensor Ext Cable (4’ Cable)

Nellcor Sensor Ext Cable (8’ Cable)

Nellcor Sensor

Masimo SpO

2

Tester

Nellcor SRC-MAX SpO

2

Tester

Masimo Patient Cable

Masimo Adult SpO2 Reusable Sensor

Digital Volt Meter with 4 1/2 Digit Display

Power Supply: 0-20 Vdc adjustable with 0-3A output

Wireless Router

Ethernet Cable

IBM compatible computer Windows 2000, XP,

NT4

* Only requires one of the pressure meters

2270-01

2000 in

Dec-4

Dec-8

DS-100A

1795

SRC-MAX

1005

1269

Qty Source

1 Tool/Supply Store

1 Tool/Supply Store

1 Tool/Supply Store

1 Tool/Supply Store

1 Tool/Supply Store

1 Marsh Electronics

+1 800 877-8919

1 Setra +1 800 257 3872

1 Netech +1 800 547 6557

1 Nellcor +1 800 635 5267

1 Nellcor +1 800 635 5267

1 Nellcor +1 800 635 5267

1 Masimo +1 800 326-4890

1 Nellcor +1 800 635 5267

1 Masimo +1 800 326-4890

1 Masimo +1 800 326-4890

2 Electronics Supply Store

1 Electronics Supply Store

1 Electronics Supply Store

1 Electronics Supply Store

1

To disassemble Spot Vital Signs LXi:

1.

Disconnect the power and all accessories from the Spot Vital Signs LXi.

2.

Remove the four screws holding the battery door using a phillips-head screwdriver.

Remove the battery door to expose the battery.

Service Manual Disassembly and Repair

129

MAT: 70

3956

Top, back housing screws (2 places)

I

II

Battery Door

Screws (4 places)

3.

Tip the Spot Vital Signs LXi backward and slide the battery out. Disconnect the oneway connector.

4.

Remove the two screws inside the battery housing that are identified with arrows molded into the housing and the two screws at the top of the back upper housing.

5.

Hold the device together, lay the back housing on the bench, and carefully lift the top housing off.

6.

Remove the screw and slightly lift the circuit board from the right facing side and disconnect the flex cable (J103).

130

Disassembly and Repair Welch Allyn Spot Vital Signs LXi

7.

Observe the locations of the two tubes before disconnecting. Lift and disconnect the wire harnesses and the pneumatic tubing. The two wire harnesses are a 6-wire harness that connects to J104 and a 2-wire harness that connects to J80 on the small, mounted blood pressure board.

8.

Disconnect the red electrical connector and the gold ribbon connector. This is located on the main board. Lift up the tab on the white cable retainer so cable can slide out after it is unplugged.

Gold Connector

Red Connector

9.

Lift the flap on the beige connector (large ribbon cable) to remove the main board.

This releases the serial board cable from connector J108.

J108

Service Manual Disassembly and Repair

131

Cut jumper

JP2

Note

This step only applies to devices configured with Masimo SpO

2 sensors. Cut jumper JP2 with diagonal cutting pliers twice so that it is open. Cut it on each end on the dotted lines as illustrated, so that it is an open connection.

132

Disassembly and Repair

Key Pad Disassembly

1.

Gently pull the button switch array out of the cover.

2.

Align and push the new button switch array into place.

Welch Allyn Spot Vital Signs LXi

Service Manual Disassembly and Repair

133

SpO

2

Circuit Board Disassembly

Note

To assure proper SpO

2

Allyn specified part.

operation, replace the SpO

2 board using only the Welch

To assure patient electrical isolation, after the main board is nearly back in position, verity that the SpO

2 flex cable is freely floating in space and is not pressed up against the main board.

Masimo

Note

Masimo configurations ship with an EMI foil shield, copper tape, and Nomex insulator. These components are removed from the illustrations to show clarity of detail. These components must be reinstalled to comply with EMC requirements.

1.

Lay the LCD flat on an ESD mat.

2.

Find the double-stack of circuit boards and remove the four corners screws. Carefully lift the circuit board straight up.

134

Disassembly and Repair Welch Allyn Spot Vital Signs LXi

3.

Remove the three screws on the circuit board located behind the LCD.

4.

Slightly rock the circuit board back and forth while lifting straight up.

5.

Replace with a new circuit board.

Nellcor

1.

Lay the LCD flat on an ESD mat.

2.

Remove the three screws on the circuit board located behind the LCD.

Service Manual Disassembly and Repair

135

3.

Slightly rock the circuit board back and forth while lifting straight up.

4.

Replace with a new circuit board and set the dip switches to match the original board

(dip 1 and 2: up / dip 3 and 4: down).

136

Disassembly and Repair Welch Allyn Spot Vital Signs LXi

Blood Pressure Circuit Board Disassembly

Note

When replacing the NIBP board see “NIBP board Initialization” on page 62 for

instructions.

Remove the four screws and carefully lift the circuit board straight up. There is one stationary connector.

Service Manual Disassembly and Repair

137

LCD Inverter Ballast Board Disassembly

This is a small board held in place with two-sided tape. It is located towards the bottom of the main board. Remove the connector at each end and pull straight up while taking care not to damage the board because the adhesive tape has considerable sticking strength.

LCD Disassembly

1.

Remove the main board. At the bottom of this board is a circuit board with a pink connector at the bottom.

2.

Pull the brown tabs of the top connector forward and remove the ribbon cable.

Inspect connector tabs for wear and replace if worn.

3.

Turn the board over and pull back the soft rubber holders. Pull the cable from the lug to remove the grounding strap and remove the LCD.

4.

Replace with a new LCD.

138

Disassembly and Repair Welch Allyn Spot Vital Signs LXi

Pump Disassembly

Note

For proper blood pressure operation, replace the pump using only the Welch Allyn specified part.

To assure patient electrical isolation, route the pump wires through the rear housing clip feature (near the pump terminals) and held in place with a tiewrap to the holes in the rear housing at the back of the battery compartment.

To assure patient electrical isolation, do not modify the length of the pump wires.

1.

Snip the wire ties (2 places).

2.

Disconnect the pneumatic tubing and unhook the wiring.

3.

Replace with a new pump.

RS232-Communication Circuit Board Disassembly

1.

Remove the pneumatic connector and associated hoses from the pump and the two screws in the rear housing. Remove the associated hoses.

2.

Release the catch and remove the cover and inner handle assembly.

3.

Remove the connection port panel.

4.

Remove the two screws inside the back housing and the ribbon cable that hold the

RS232 board in place. Take care since the connector is easily broken.

5.

Slide the board straight out and remove the wire harness.

6.

Replace with a new circuit board.

Service Manual Disassembly and Repair

139

Fan and Power Circuit Board Disassembly

Note

To prevent buildup of hydrogen gas, replace the fan using only the Welch Allyn specified part.

This is a small board held in place with two-sided tape and one connector. Remove the screw before disconnecting the connector.

Thermometry Circuit Board Disassembly

Caution To assure patient electrical isolation, trap the flex cable to the thermometer pod behind the clip on the main housing adjacent to the main board connection point.

Caution To assure patient electrical isolation, verify that the correct flex cable is used to connect the thermometer pod to the main board. The Braun and

SureTemp Plus flex cables are slightly different in length and are not interchangeable.

Caution To protect the user from high voltage, properly install the thermometer pod insulating paper separating the LCD ballast wires from the housing crack.

SureTemp Plus

Caution Do not attempt to repair or clean the solder joints on either SureTemp

Plus board. Incorrect flux or technique can degrade thermometer accuracy.

1.

Remove the two screws located at the thermometer housing with a T-10 torx wrench.

Save the Nomex.

2.

Locate the path of the notched tabs on the housing. Slide the housing up and out following this path and set the case aside.

140

Disassembly and Repair Welch Allyn Spot Vital Signs LXi

3.

Remove the four screws from the mounting plate and save the Nomex. This exposes the circuit board.

4.

Remove the three Torx screws (T10) on the board (one on bottom and one on either side of the probe well cover) and lift the board out of the housing and orient the part as shown.

5.

Disconnect all connectors and replace with a new board.

To replace the temperature connector, remove the one screw holding the temperature board and connector. Verify the Nomex is properly mounted upon reassembly.

When replacing the SpO

2/ blood pressure housing, verify the orange and white twisted wire is pushed as far into the handle housing as possible. Failure to do so can cause the wires to stretch when the SpO

2

/blood pressure housing is inserted.

Service Manual Disassembly and Repair

141

Braun ThermoScan PRO 4000

1.

Remove the two screws located at the thermometer housing with a T-10 torx wrench.

Save the Nomex.

2.

Locate the path of the notched tabs on the housing. Slide the housing up and out following this path and set the case aside.

3.

Remove the four screws that hold the metal shield in place.

4.

Remove the springs and posts. Carefully rock the board back and forth while lifting straight up.

5.

Disconnect all connectors and replace with a new board.

6.

Reassembly is the reverse of disassembly.

142

Disassembly and Repair Welch Allyn Spot Vital Signs LXi

12

Repair Test Specifications

This appendix refers to the Spot Vital Signs LXi without pneumatics (tubing and cuff), temperature probe, SpO

2 probe, or main battery attached, unless otherwise noted.

Use the Repair Software for the performing the tests on the Spot Vital Signs LXi. The standard test voltage, unless otherwise stated is 6.5 Vdc (+0/-0.25 Vdc). Unless otherwise stated, all calibrated volumes are ±10cc of the stated volume.

General Unit Test

A-D Noise Test

The A-D Noise Test is defined as the amount of noise on the Spot Vital Signs LXi A-D pressure channel; over a 1 second sample time while 0.0 mmHg is applied to the Device pressure port. The maximum limit is 0.050 mmHg.

Leak Test

The Leak Test is defined as the amount of pressure drop that is recorded over a 15-second interval with a 100cc cylinder attached to the Spot Vital Signs LXi pressure port and that volume having a stabilized pressure of 250 mmHg. The maximum limit is 6 mmHg.

Inflation Test

The Inflation Test is defined as the amount of time the Spot Vital Signs LXi pump can inflate a 250 cc cylinder from 5 mmHg to 210 mmHg. The maximum limit is 7 seconds.

Dump Test

The Dump Test is defined as the amount of time it takes Spot Vital Signs LXi to deflate a

500 cc cylinder from 260 mmHg to less than 15 mmHg. The maximum limit is 10 seconds.

Pneumatic Accuracy Test

Enter the Internal Configuration Mode to perform this test (see “Temperature Functional

Check” on page 77).

Perform an Auto-zero before starting this test (see “Temperature Functional Check” on page 77). The Pneumatic Accuracy test is defined as the comparison of Spot Vital Signs

LXi’s pressure measurement and applied pressure at 0, 50, 150, and 250 mmHg. The

143

144

Repair Test Specifications Welch Allyn Spot Vital Signs LXi specification for each reading is within ±3 mmHg of the target pressure except for the 0 mmHg reading which is within ±1.0 mmHg.

Valve Control Test

Perform the valve control test using the Welch Allyn Spot LXi Service Tool, see “Valve control test” on page 67.

Voltage Calibration

Calibrate the Spot Vital Signs LXi battery voltage measurement circuit at 5.5 + 0.1V DC.

After performing a successful battery measurement circuit calibration, Spot Vital Signs

LXi’s memory will store a calibration signature of up to four printable characters.

Blank Mode Current Test

The Blank Mode Current Test is defined as the amount of current drawn through the battery terminals. All LCD segments, the back light, and the SpO

2

The limit is 200 mA maximum.

mode are all turned off.

Back Light (Idle) Current Test

The Back Light Current Test is defined as the amount of current drawn through the battery. All LCD segments and the back light are turned on while the SpO

2 off. The limit is 400 mA maximum.

mode is turned

Valve/Pump Mode Current Test

The Valve/Pump Mode Current Test is defined as the amount of current that is drawn through the battery terminals; Spot Vital Signs LXi is in the Blank Mode while the valve and pump are actuated to on. The limit is 2.0 A maximum.

Interface Test

When given the proper commands, the Display Window will turn on or off all segments and turn on or off the back light. When given the proper command, the buzzer sounds to verify operation. The user determines the pass/fail criteria.

Temperature Option Requirements

Accuracy Test

Verify the accuracy of the temperature module is within ±0.2° F for readings with a normal temperature of 97.3° F (36.3° C) using a Cal Key (5200-25). Verify Spot Vital Signs

LXi can read a temperatures of 96.4° F (35.8° C) and 106° F (41.1° C) within ±0.3° F/±0.2°

C using a Welch Allyn 9600 Plus Calibrator.

Service Manual Repair Test Specifications

145

Temperature Probe Test

Spot Vital Signs LXi displays “ORL” after you remove the blue thermometer probe from the blue probe well.

SpO

2

Option Requirements

These tests are only valid on Spot Vital Signs LXi models with the Masimo or Nellcor

SpO

2 option.

SpO

2

Functional Test

See

“To begin functional verification of the Masimo SpO2”

or

“To begin functional verification of the Nellcor SpO2” on page 87 for the functional test.

SpO

2

Mode Current Test

The SpO

2

Mode Current Test is defined as the amount of current, less the Blank Mode

Current, that is drawn through the battery terminals, placing Spot Vital Signs LXi in the

Blank Mode, actuating the SpO

2 mode and applying any SpO maximum limit is 120 mA maximum.

2 signal to the device. The

Fail Safe Test

Over Pressure Test

Verify that Spot Vital Signs LXi can detect over pressure on the pneumatic system between 296 mmHg and 329 mmHg.

Over 15 mmHg

Verify that Spot Vital Signs LXi can detect static pressure over 15 mmHg for 180 seconds.

146

Repair Test Specifications Welch Allyn Spot Vital Signs LXi

Service Manual

147

Interconnect Diagram 1

3

D

C

B

A

BATTERY

-

+

A5

BATTERY

BLK

RED

1

2

1

A5-F1

J4

CHARGER center = +

W1-A1-JP2

1

2

+

1

2

W1-A1-J3

W1-A1

POWER

INTERFACE

BOARD

+ RED

BLK

5

BLK

RED

W1

POWER HARNESS

BLU

GRN

ORN

WHT

+

+

+

1

BAT-IN

4

5

6

1

2

3

SER-GND

BAT-GND

CHG-IN

CHG-GND

J104

POWER

Ferrite

10

11

12

7

8

9

13

14

15

1

2

3

4

5

6

J106

LCD

LCD-D0

LCD-D1

LCD-D2

LCD-D3

LCD-VEE

LCD-VDD

LCD-LACD

LCD-LSCLK

LCD-LP

LCD-FLM

DGND

A1-F1

FUSE 5A

A6

FAN

FAN

W3

FFC

MNT8

MNT7

A10-J80

PUMP

2 1

1

VBUS

DGND

D107

J105

NIBP CONTROL

DGND

A10

NIBP BOARD

A10-J81

VALVE

9

10

11

7

8

12

13

14

15

1

2

3

4

5

6

FRM

LOAD

CP

DISP-ON

VDD

VSS

VEE

D7/NC

D6/NC

D5/NC

D4/NC

D3

D2

D1

D0

Pinout on LCD is reversed from standard Molex assignment.

MNT9

MNT10

A3

LCD

PNK

WHT

1

CN2

1

3

A4

INVERTER /

BALLAST

CN1

1 1

2

5

RED

BLK

W4

DGND

X1

W5

1

DGND

X3

+3.15V

A1-J5

JTAG

12

14

16

18

20

2

4

6

8

10

11

13

15

17

19

5

7

9

1

3

NOPOP

DGND

W7

FFC

+3.15V

A1

J110

MAIN

BOARD

DGND

2

S3

RESET

1

GND-ZC

J103

THERMOMETER

GND-ZB

MNT4

J107

BACKLIGHT

J102

SPO2 CONTROL

DGND

J108

SERIAL

INTERFACE

A1-JP2

0

1 2

Cut out for MS-11

J101

SENSOR

SERA-PF-BAR

SERA-EN

SERA-CTS-BAR

SERA-RXE

SERA-TX-BAR

SERA-RTS-BAR

+HST-IF

+5V

SER-SYNC

SERB-PF-BAR

SERB-EN

SERB-CTS-BAR

SERB-RXE

SERB-TX-BAR

SERB-RTS-BAR

USB-ID-BAR

USB-SUSPND

USB-OE-BAR

USB-VP

USB-VM

USB-RCV

USB-VPO

USB-VMO

DGND

J6 P6

SPO2 SENSOR

GND-ZBI

ANODE

CATH

INNER

INNER

DIGI-CAL

OUTER

LED(1)

OUTER

LED(8)

DIGI-GND

1

2

3

7

8

9

4

5

6

10

1

MNT5 GND-ZBX

20

21

22

23

24

25

26

27

28

29

30

17

18

19

14

15

16

7

8

9

4

5

6

10

11

12

13

1

2

3

1

MNT6

1

P102 P101

A7

MASIMO

ADAPTOR BOARD

CONTROL SENSOR

A10-P6

SENSOR

A11-F1

FILTER

A11

PUMP &

TUBING

M

A12

TRANSDUCER

TUBING

Ferrite

MNT23

1

P3

A8-J3

MNT24

MNT21

A8

MASIMO SPO2

BOARD

Shield

1

P1

A8-J1

6

1

A9-J1

CONTROL

A9

SURE TEMP

PLUS OEM

BOARD

A9-J3

PROBE

8

7

6

5

4

3

2

1

W8

FFC

MNT22

PRODUCT ENCLOSURE

W2

FFC

BLOOD

PRESSURE

CUFF

4 3 2

30

1

J10

J90

1

8

GND-IO

1

J11

VBUS

POWER

A2

DCD 1

DSR

RXD

6

2

RTS 7

TXD

CTS

3

8

DTR

ZC+5V

4

9

5

GND-ZE DTE

SERIAL

INTERFACE

BOARD

J2

SERIAL II

J3

USB

DEVICE

W6

SPO2 FPC ASSY

DCD

DSR

1

6

RXD 2

RTS 7

TXD

CTS

3

8

DTR

ZC+5V

4

9

5

GND-ZD DTE

J1

SERIAL I

A13

THERMOMETER

INTERFACE

(EDAC) BOARD

J8

SPO2

Mini-D

J5

SURETEMP

PLUS

PROBE

Note: This page depicts interconnect for basic unit plus Masimo

SPO2 and SureTemp Plus Thermometer options.

THIS INFORMATION IS THE

PROPERTY OF WELCH

ALLYN. IT IS SUBMITTED IN

CONFIDENCE AND IS NOT

TO BE DISCLOSED OR

UTILIZED WITHOUT

WRITTEN PERMISSION.

CHECK

APPROVALS

-

PROJ. MGR.

-

Source document:

40700-HDD

DATE

PCB Mat'l # (ref)

PCA Mat'l # (ref)

TITLE

WELCH ALLYN

Spot Lxi Interconnect Diagram

SIZE

B

NUMBER

DATE 12/6/05

DRAWN BY

SHEET

K Burdick

1 of 2

REV

A

1

D

C

B

A

148

Interconnect Diagram 2

5

D

C

B

A

Welch Allyn Spot Vital Signs LXi

4 3 2 1

5

5

6

7

8

9

3

4

1

2

10

11

12

13

14

15

LCD-D0

LCD-D1

LCD-D2

LCD-D3

LCD-VEE

LCD-VDD

LCD-LACD

LCD-LSCLK

LCD-LP

LCD-FLM

DGND

J106

LCD

BAT-IN

4

5

6

1

2

3

SER-GND

BAT-GND

CHG-IN

CHG-GND

J104

POWER

A1-F1

FUSE 5A

DGND

VBUS

D107

MNT8

MNT7

J105

NIBP CONTROL

DGND

MNT9

MNT10

DGND

X1

DGND

X3

+3.15V

J5

JTAG

12

14

16

18

20

2

4

6

8

10

11

13

15

17

19

1

3

5

7

9

NOPOP

DGND

+3.15V

A1

J110

MAIN

BOARD

DGND

2

S3

RESET

1

J107

BACKLIGHT

DGND

J108

SERIAL

INTERFACE

SERA-PF-BAR

SERA-EN

SERA-CTS-BAR

SERA-RXE

SERA-TX-BAR

SERA-RTS-BAR

+HST-IF

+5V

SER-SYNC

SERB-PF-BAR

SERB-EN

SERB-CTS-BAR

SERB-RXE

SERB-TX-BAR

SERB-RTS-BAR

USB-ID-BAR

USB-SUSPND

USB-OE-BAR

USB-VP

USB-VM

USB-RCV

USB-VPO

USB-VMO

W9

FFC

GND-ZC

J103

THERMOMETER

GND-ZB

MNT4

J102

SPO2 CONTROL

1

A1-JP2

0

2

J101

SENSOR

Cut out for MS-11

DGND

J6

SPO2 SENSOR

GND-ZBI

ANODE

CATH

INNER

INNER

DIGI-CAL

OUTER

LED(1)

OUTER

LED(8)

DIGI-GND

7

8

9

10

1

2

3

4

5

6

MNT5 GND-ZBX

1

A14-J2

CONTROL

MNT6

1

A14-J1

SENSOR

A14

NELLCOR

SPO2

BOARD

12

13

14

15

16

7

8

9

10

11

5

6

3

4

1

2

22

23

24

25

26

27

28

29

30

17

18

19

20

21

4 3

W6

FPC

J8

SPO2

Mini-D

PRODUCT ENCLOSURE

6

1

A15-J1

CONTROL

A15-J2

WELL

CONTACTS

A15

BRAUN

INTERFACE

BOARD

CHG/DATA

GND

2

1

Note: This page depicts interconnect for Nellcor SPO2 and Braun

Thermometer options only.

THIS INFORMATION IS THE

PROPERTY OF WELCH

ALLYN. IT IS SUBMITTED IN

CONFIDENCE AND IS NOT

TO BE DISCLOSED OR

UTILIZED WITHOUT

WRITTEN PERMISSION.

CHECK

APPROVALS

-

PROJ. MGR.

-

Source document:

40700-HDD

DATE

PCB Mat'l # (ref)

PCA Mat'l # (ref)

TITLE

WELCH ALLYN

Spot Lxi Interconnect Diagram

SIZE

B

NUMBER

DATE 12/6/05

DRAWN BY

SHEET

K Burdick