Peak Scientific Genius 3031 Gas Generator Installation Guide

Below you will find brief information for Gas Generator Genius 3031. This guide provides detailed instructions for installing the Genius 3031 gas generator. Its purpose is to supply on-demand gas to your analytical instrument. The guide covers unpacking, removing transit brackets, connecting drain lines, and setting up power supply. It also provides instructions for purging the system and connecting the generator to your instrument.

PDF Download
Document
Genius 3031 Installation Guide | Manualzz

Advertisement

Advertisement

Advertisement

Key features

  • On demand gas supply
  • Purge run
  • Transit bracket removal
  • Drain line connection
  • Power supply
  • Teflon tube connection
  • Pressure gauge
  • Voltage requirement

Frequently asked questions

It operates on voltages of 230VAC +/- 10%. This means it will operate between 207V and 253V. If the generator is operated on a supply voltage of less than 220V, the internal compressors will work harder and the service life of the compressors will be reduced.

The purge run should be left on for 30 minutes to purge the system.

The maximum distance away from the instrument the generator can be placed still using 6mm (4mm internal diameter) or ¼” (3/16” internal diameter) tubing is 10m.