Combo Thermometer Operating
814-065
Combo Thermometer
Operating Instructions
The Combo Thermometer is not only a non-contact infrared thermometer but also a conventional penetration
thermometer. Choose the mode required, either the IR mode (infrared) or the TC On (probe). To use the noncontact measurement, simply press the IR button and point the thermometer at the target to display the surface
temperature (in the upper display). To use the probe, press the probe button, unfold and insert into the object
being measured, wait until the temperature stabilizes and the reading will be displayed in the lower display.
Please note: the Combo Thermometer will only display either the IR or probe reading.
Minimum or Maximum Mode (Infrared only)
To utilize the Combo’s minimum or maximum mode, first turn the instrument on by pressing the Infrared
measurement button, then press the mode button once for minimum or twice for maximum function. The MIN or
MAX icon will flash, then press the measurement button to confirm the minimum or maximum mode. The Combo
will display the minimum or maximum reading only.
Lock Mode (Infrared only)
The lock mode is particularly useful for continuous monitoring of temperatures. To utilize the Combo’s lock
mode, first turn the instrument on by pressing the Infrared measurement button, then press the mode button
three times for the lock mode function. The LOCK icon will flash, then press the Infrared button to confirm the
lock measurement mode. The Combo will continuously display the temperature for up to 60 minutes or until the
Infrared button is pressed.
°C or °F Mode
To change the Combo from °C to °F or from °F to °C, first turn the instrument on by pressing the Infrared
measurement button, then press the mode button four times, the °C or °F icon will flash, press the Infrared
button to change, then press again to confirm the mode.
Emissivity
The Combo is supplied with a default emissivity of 0.95. The emissivity of the Combo can be changed from
0.1 (10E) to 1 (100E). Changes should only be carried out by experienced personnel. To change the Combo’s
emissivity, first press the Infrared button, then press the mode button five times for emissivity function. The 95E
will flash on the LCD screen, then press the Infrared button to adjust the emissivity value, press the mode button
again to exit the set up screen. For information relating to the emissivity of specific materials, please contact
ThermoWorks or visit www.thermoworks.com. Note: non-contact infrared thermometers are not recommended
for use in measuring the temperature of shiny or polished metals.
Specifications
Measuring Range (IR) -27 to 428˚F / -33 to 220˚C
* IR accuracy can be affected by the emissivity
setting, target spot size, and rapid changes in
Accuracy (IR)* ±2%, or 2˚C, whichever is greater
ambient temperature.
Repeatability (IR) 0.1˚C at 25˚C, 0.5˚C at 80˚C
Measuring Range (probe) -67 to 626˚F / -55 to 330˚C
Accuracy (probe) ±1%, or 1.8˚F / 1°C, whichever is greater
Response Time (probe) 10 to 15 seconds typical
Resolution
0.1˚ from -9.9 to 199.9°; 1° beyond
Units Switchable C or F
Emissivity Default 0.95, adjustable 0.1 to 1.0 in 0.01 increments
Distance/Target Ratio 5:1 (1” target diameter at 5” distance)
Battery CR2032, included
814-065 Combo Thermapen
Utah, U.S.A.
Ph: 801-756-7705 Fax: 801-756-8948
www.thermoworks.com
Made in China
Combo Thermometer Operating Instructions
The Combo Thermometer is not only a non-contact infrared thermometer but also a conventional
penetration thermometer. Choose the mode required, either the IR mode (infrared) or the TC On (probe).
To use the non-contact measurement, simply press the IR button and point the thermometer at the target
to display the surface temperature (in the upper display). To use the probe, press the probe button,
unfold and insert into the object being measured, wait until the temperature stabilizes and the reading
will be displayed (in the lower display). Please note: the Combo Thermometer will only display either
the IR or probe reading.
MINIMUM OR MAXIMUM MODE (Infrared only)
To utilize the Combo’s minimum or maximum mode, first turn the instrument on by pressing the Infrared
measurement button, then press the mode button once for minimum or twice for maximum function.
The MIN or MAX icon will flash, then press the Infrared measurement button to confirm the minimum or
maximum mode. The Combo will display the minimum or maximum reading only.
LOCK MODE (Infrared only)
The lock mode is particularly useful for continuous monitoring of temperatures. To utilize the Combo’s
lock mode, first turn the instrument on by pressing the Infrared measurement button, then press the
mode button three times for the lock mode function. The LOCK icon will flash, then press the Infrared
measurement button to confirm the lock measurement mode. The Combo will continuously display the
temperature for up to 60 minutes or until the Infrared measurement button is pressed.
°C OR °F MODE
To change the Combo from °C to °F or from °F to °C, first turn the instrument on by pressing the Infrared
measurement button, then press the mode button four times, the °C or °F icon will flash, press the
Infrared button to change, then press again to confirm the mode.
EMISSIVITY
The Combo infrared thermometer is supplied with a default emissivity of 0.95. The emissivity of the
Combo can be changed from 0.1 (10E) to 1 (100E). Changes should only be carried out by experienced
personnel. To change the Combo’s emissivity, first press the Infrared button, then press the mode button
five times for emissivity function. The 95E will flash on the LCD screen, then press the Infrared button to
adjust the emissivity value, press the mode button again to exit the set up screen. For information relating
to the emissivity of specific materials, please contact ThermoWorks or visit www.thermoworks.com.
Note: non-contact infrared thermometers are not recommended for use in measuring the temperature
of shiny or polished metals.
LCD ERROR MESSAGES
The Combo incorporates visual diagnostic messages as
follows:
Hi or Lo is displayed when the temperature being measured
is outside of the range of the instrument, Hi when higher
than +428°F and Lo when lower than -27.4°F.
Er2 is displayed when the Combo is exposed to rapid
changes in the ambient temperature. Er3 is displayed when
the ambient temperature of the Combo exceeds 32°F or
+122°F. In both cases you should allow plenty of time
(minimum 30 minutes) for the Combo to stabilize to the
working/room temperature (32 to +104°F).
For all other error messages it is necessary to reset the Combo. To reset, turn the instrument off, remove
the battery and wait for a minimum of one minute, insert the battery and turn on. If the error message
remains please contact the ThermoWorks Service Department for further assistance.
BATTERY
The Combo incorporates visual low battery indication as follows:
‘Battery OK’
measurements are
possible
‘Battery Low’ battery needs to
be replaced, measurements are
possible
‘Battery Exhausted’
measurements are
not possible
BATTERY REPLACEMENT
When the Low Battery icon indicates the battery is low, it should be replaced immediately with a CR2032
lithium cell. The battery is located under the twist cover at the rear of the Combo. When inserting the new
battery, be sure to slide the coin cell battery under metal spring-loaded clip first, then snap into place.
Please note: It is important to turn the instrument off before replacing the battery otherwise the Combo
may malfunction.
STORAGE & CLEANING
The sensor lens is the most delicate part of the Combo. The lens should be kept clean at all times, care
should be taken when cleaning the lens using only a soft cloth or cotton swab with water or medical
alcohol, allowing the lens to fully dry before using the Combo, do not submerge any part of the unit. The
Combo should be stored at room temperature between +50 to +104°F.
EMC/RFI
Readings may be affected if the unit is operated within a radio frequency electromagnetic field strength of
approximately 3 volts per metre, but the performance of the instrument will not be permanently affected.
PRODUCT WARRANTY & GUARANTEE
This instrument carries a one-year warranty and guarantee against defects in either components or
workmanship. During this period, products that prove to be defective will, at the discretion of ThermoWorks,
be either repaired or replaced without charge. The product warranty and guarantee does not cover
damage caused by fair wear and tear, abnormal storage conditions, incorrect use, accidental misuse,
abuse, neglect, misapplication or modification, or use with non-ThermoWorks hardware/ software. No
warranty of fitness for a particular purpose is offered and the user assumes the entire risk of using
the product. Any liability by ThermoWorks is limited only to the replacement of defective materials or
workmanship, and ThermoWorks accepts no responsibility for consequential loss. In line with our policy
of continuous development, we reserve the right to amend our product specification without prior notice.
* IR accuracy can be affected by the emissivity setting, target spot size, and rapid changes in ambient
temperature.
Utah, U.S.A.
Ph: 801-756-7705 Fax: 801-756-8948
www.thermoworks.com