Agilent IO Libraries M.01.xx or higher

Agilent IO Libraries
M.01.xx or higher
Installation Instructions for
use with the Agilent
ChemStation Rev. B.03.01 or
higher
Agilent Technologies
Agilent Technologies
Notices
© Agilent Technologies, Inc. 2004,
2005-2007
No part of this manual may be reproduced in any form or by any means
(including electronic storage and
retrieval or translation into a foreign
language) without prior agreement
and written consent from Agilent
Technologies, Inc. as governed by
United States and international copyright laws.
Manual Part Number
G2170-90129
Edition
02/2007
Printed in Germany
Agilent Technologies
Hewlett-Packard-Strasse 8
76337 Waldbronn
Warranty
The material contained in this document is provided “as is,” and is subject to being changed, without notice,
in future editions. Further, to the maximum extent permitted by applicable
law, Agilent disclaims all warranties,
either express or implied, with regard
to this manual and any information
contained herein, including but not
limited to the implied warranties of
merchantability and fitness for a particular purpose. Agilent shall not be
liable for errors or for incidental or
consequential damages in connection
with the furnishing, use, or performance of this document or of any
information contained herein. Should
Agilent and the user have a separate
written agreement with warranty
terms covering the material in this
document that conflict with these
terms, the warranty terms in the separate agreement shall control.
Technology Licenses
The hardware and/or software described in
this document are furnished under a license
and may be used or copied only in accordance with the terms of such license.
Restricted Rights Legend
If software is for use in the performance of a U.S. Government prime
contract or subcontract, Software is
delivered and licensed as “Commercial computer software” as defined in
DFAR 252.227-7014 (June 1995), or
as a “commercial item” as defined in
FAR 2.101(a) or as “Restricted computer software” as defined in FAR
52.227-19 (June 1987) or any equivalent agency regulation or contract
clause. Use, duplication or disclosure
of Software is subject to Agilent Technologies’ standard commercial
license terms, and non-DOD Departments and Agencies of the U.S. Government will receive no greater than
Restricted Rights as defined in FAR
52.227-19(c)(1-2) (June 1987). U.S.
Government users will receive no
greater than Limited Rights as
defined in FAR 52.227-14 (June 1987)
or DFAR 252.227-7015 (b)(2) (November 1995), as applicable in any technical data.
Safety Notices
CAUTION
A CAUTION notice denotes a
hazard. It calls attention to an
operating procedure, practice, or
the like that, if not correctly performed or adhered to, could
result in damage to the product
or loss of important data. Do not
proceed beyond a CAUTION
notice until the indicated conditions are fully understood and
met.
WA R N I N G
A WARNING notice denotes a
hazard. It calls attention to an
operating procedure, practice,
or the like that, if not correctly
performed or adhered to, could
result in personal injury or
death. Do not proceed beyond a
WARNING notice until the indicated conditions are fully understood and met.
Contents
Contents
1
General Information
5
2
Configuring the Agilent 82350 GPIB Interface Drivers under Windows XP
Professional
7
Installing New IO Libraries
8
3
4
5
6
Configuring the Agilent 82357 USB - GPIB Interface Driver under Windows XP
Professional
19
Installing New IO Libraries for USB - GPIB Interface on Windows XP
20
Upgrading Existing IO Libraries
35
Upgrading IO Libs under Windows XP
36
Working with your Agilent IO Libraries
43
What has been changed by the Installation?
44
Troubleshooting
45
Missing files during installation
Problems with Plug and Play option GPIB cards
Problems with Plug and Play option using 82357A USB-GPIB Interface
46
51
52
3
Contents
4
Installation Instructions for use with the Agilent ChemStation Rev.
B.03.01 or higher
1 General Information
To use the GPIB 82350A or 82350B board, or as well for the 82357 USB-GPIB
Interface, it is necessary to install or upgrade the corresponding driver and
configuration software, located in the “\IO_Libs” directory on the Agilent
ChemStation CD-ROM (CD1 Installation) or the Agilent GP-IB Libraries for
Instrument control CD-ROM. To install the I/O Libraries proceed according to
the steps described in this manual.
The Agilent ChemStation B.03.01 and greater requires version M.01.01.04 of
the I/O Libraries. These libraries are supported under Windows XP
Professional.
Be sure to follow the correct installation description.
NOTE
Before installing the GPIB Interface Driver, verify that this communication component is
supported for your instrument in use, see manual “Installing your ChemStation” of your
appropriate instrument or in the table below.
Table 1
Supported devices with the 82357A USB-GPIB Interface
Instrument
• Capillary
Supports 82357A USB-GPIB Interface
Electrophoresis
Yes
• HP
1090 LC system
Yes
• HP
1046 FLD Detector
Yes
Agilent Technologies
5
1
6
General Information
Instrument
Supports 82357A USB-GPIB Interface
• HP
Yes
1049 ECD Detector
• Agilent
1200/1100 modules
No
• Agilent
GC Systems
No
• Agilent
A/D Systems
No
• Agilent
LC/MS Systems
No
• Agilent
GC/MS Systems
No
Installation Instructions for use with the Agilent ChemStation Rev.
B.03.01 or higher
2 Configuring the Agilent 82350 GPIB
Interface Drivers under Windows XP
Professional
Installing New IO Libraries
Agilent Technologies
8
7
2
Configuring the Agilent 82350 GPIB Interface Drivers under Windows XP Professional
Installing New IO Libraries
Installing New IO Libraries
1 Turn the Computer ON. As Windows XP Professional has started, the user
is notified about new detected hardware via tool tip in the right corner of
the task bar. The Found New Hardware Wizard will start some while later. At
this moment, select Cancel (see Figure 1 on page 8).
Figure 1
Found New Hardware Wizard
2 Using the Windows Explorer, locate and start ‘m0101.exe’ in the \IO_Libs
directory of your ChemStation CD-ROM (CD 1 Installation). Follow the
instructions in the installation Wizard. This will install the Standard
Interface Control Library (SICL) on your computer, update the registry and
add a new Program Group called “Agilent I/O Libraries”.
8
Configuring the Agilent 82350 GPIB Interface Drivers under Windows XP Professional
Installing New IO Libraries
2
The InstallShield Wizard opens and starts the IO Libraries installation.
Figure 2
Install Shield Wizard
3 When the title screen appears (see Figure 3 on page 10), click Next to go the
License Agreement screen and then click Yes to accept the licenses terms
and to display the Readme Information.
9
2
Configuring the Agilent 82350 GPIB Interface Drivers under Windows XP Professional
Installing New IO Libraries
Figure 3
Installation screen IO Libraries
4 Click Next to display the Select Type of Installation screen.
5 Choose installation option Full Installation, for an installation of both the
SICL libraries and VISA components. The System will go to the next screen
and display the current settings (see Figure 4 on page 11).
10
Configuring the Agilent 82350 GPIB Interface Drivers under Windows XP Professional
Installing New IO Libraries
Figure 4
2
Installation options
6 By default SICL is installed in C:\Program Files\Agilent\IO libraries. VISA
is installed by default in C:\Program Files\VISA\winnt. Click Next if you
accept the default setting, or go back to set the directory you want via
Custom Installation option. By clicking Next the IO Libraries will be
installed.
7 The IO Libraries have been successfully installed now. Finish the
installation without viewing the Readme file and without running the IO
Config. Leave the screen with Finish and reboot the system.
8 As Windows XP Professional starts after power cycling, a tool tip will notify
about new hardware and the Found New Hardware Wizard will start again.
Select Option 2 and click Next (see Figure 5 on page 12).
11
2
Configuring the Agilent 82350 GPIB Interface Drivers under Windows XP Professional
Installing New IO Libraries
Figure 5
Found New Hardware Wizard
9 Select Option 2 and click Next (See Figure 6 on page 13).
12
2
Configuring the Agilent 82350 GPIB Interface Drivers under Windows XP Professional
Installing New IO Libraries
Figure 6
Install Hardware Device Drivers
10 Select 82350 PCI and click Next (see Figure 7 on page 14). The installation
of the device driver will start.
13
2
Configuring the Agilent 82350 GPIB Interface Drivers under Windows XP Professional
Installing New IO Libraries
Figure 7
Select a Device Driver
11 Select Finish (see Figure 8 on page 15). The user will get a notification by the
task bar that the hardware is installed and ready to use. Also the IO Config
Utility will be placed now in the task bar.
14
2
Configuring the Agilent 82350 GPIB Interface Drivers under Windows XP Professional
Installing New IO Libraries
Figure 8
Finish Installation GPIB card
12 Start the IO Config Utility by clicking in the Icon
Run IO Config.
and choose
13 From the list of available interface types choose the appropriate one, i.e.
82350 PCI GPIB in case you have the Agilent 82350 board installed (see
Figure 9 on page 16).
15
2
Configuring the Agilent 82350 GPIB Interface Drivers under Windows XP Professional
Installing New IO Libraries
Figure 9
IO Libraries Configuration
14 Select Configure. The configuration utility now attempts to detect the
selected GPIB interface in the PC in order to assign the correct resources to
the card.
15 Set the SICL Interface Name to hp82341, the Bus Address to 30 and the
System Controller flag as shown in Figure 10 on page 17. Please note that the
appearance of the dialog slightly differs depending on the type of GPIB
board installed.
16
Configuring the Agilent 82350 GPIB Interface Drivers under Windows XP Professional
Installing New IO Libraries
Figure 10
NOTE
2
Configuring GBIB Cards
You need to type in the SICL interface name hp82341 with lower case letters for both cards.
16 Select OK. The configured GPIB board should now be listed under Configured
Interfaces in the Interface Configuration Screen. Close the window by
clicking OK. The installation is complete. Note that you must restart to
initialize the IO Libraries.
17
2
18
Configuring the Agilent 82350 GPIB Interface Drivers under Windows XP Professional
Installing New IO Libraries
Installation Instructions for use with the Agilent ChemStation Rev.
B.03.01 or higher
3 Configuring the Agilent 82357 USB GPIB Interface Driver under
Windows XP Professional
Installing New IO Libraries for USB - GPIB Interface on Windows XP
20
Agilent Technologies
19
3
Configuring the Agilent 82357 USB - GPIB Interface Driver under Windows XP Professional
Installing New IO Libraries for USB - GPIB Interface on Windows XP
Installing New IO Libraries for USB - GPIB Interface on
Windows XP
The USB-GPIB Interface driver (IO libraries) needs to be installed prior to the
connection of the Interface device.
1 Using the Windows Explorer, locate and start ‘m0101.exe’ in the \IO_Libs
directory of your ChemStation CD-ROM (CD1 Installation). Follow the
instructions in the installation Wizard. This will install the Standard
Interface Control Library (SICL) on your computer, update the registry and
add a new Program Group called “Agilent I/O Libraries”.
The InstallShield Wizard opens and starts the IO Libraries installation.
Figure 11
Install Shield Wizard
2 When the title screen appears (see Figure 3 on page 10), click Next to go the
License Agreement screen and then click Yes to accept the licenses terms
and to display the Readme Information.
20
Configuring the Agilent 82357 USB - GPIB Interface Driver under Windows XP Professional
Installing New IO Libraries for USB - GPIB Interface on Windows XP
Figure 12
3
Installation screen IO Libraries
3 Click Next to display the Select Type of Installation screen.
4 Choose installation option Full Installation, for an installation of both the
SICL libraries and VISA components. The System will go to the next screen
and display the current settings (see Figure 4 on page 11).
21
3
Configuring the Agilent 82357 USB - GPIB Interface Driver under Windows XP Professional
Installing New IO Libraries for USB - GPIB Interface on Windows XP
Figure 13
Installation options
5 By default SICL is installed in C:\Program Files\Agilent\IO libraries. VISA
is installed by default in C:\Program Files\VISA\winnt. Click Next if you
accept the default setting, or go back to set the directory you want via
Custom Installation option. By clicking Next the IO Libraries will be
installed.
6 The IO Libraries have been successfully installed now. Finish the
installation without viewing the Readme file and without running the IO
Config. Leave the screen with Finish and reboot the system.
7 Plug the 82357A USB-GPIB Interface into the GPIB connector of the GPIB
card of the instrument. If a GPIB cable is used to connect multiple modules
(e.g. HP 1090 mainframe and DAD), the USB Interface device needs to be
plugged into the GPIB connector (see Figure 14 on page 23).
22
Configuring the Agilent 82357 USB - GPIB Interface Driver under Windows XP Professional
Installing New IO Libraries for USB - GPIB Interface on Windows XP
Figure 14
3
82357A USB-GPIB Interface on e.g. HP 1090/DAD
8 Connect the USB connector of the 82357A USB-GPIB interface into the USB
port of your Computer. The hardware wizard detects a USB device and the
Found New Hardware Wizard will start (see Figure 15 on page 24).
9 Select Option 1 and click Next and the installation will start (see Figure 15 on
page 24).
23
3
Configuring the Agilent 82357 USB - GPIB Interface Driver under Windows XP Professional
Installing New IO Libraries for USB - GPIB Interface on Windows XP
Figure 15
Found New Hardware Wizard Windows XP, first detection
10 To initialize the interface select Option 1 and click Next and the initialization
will start (see Figure 16 on page 25).
24
Configuring the Agilent 82357 USB - GPIB Interface Driver under Windows XP Professional
Installing New IO Libraries for USB - GPIB Interface on Windows XP
Figure 16
3
Found New Hardware Wizard Windows XP, first detection
11 The system will initialize the USB device. This might take some seconds (see
Figure 17 on page 26).
25
3
Configuring the Agilent 82357 USB - GPIB Interface Driver under Windows XP Professional
Installing New IO Libraries for USB - GPIB Interface on Windows XP
Figure 17
Initializing the 82357A USB-GPIB Interface in Windows XP, first detection
12 Select Finish to close the wizard after the initialization Hardware driver(see
Figure 18 on page 27). The initialization of the USB device is finished, the
installation still needs to be done.
26
3
Configuring the Agilent 82357 USB - GPIB Interface Driver under Windows XP Professional
Installing New IO Libraries for USB - GPIB Interface on Windows XP
Figure 18
Finish Device Driver Installation in Windows XP
13 The hardware wizard detects now the USB-GPIB device for a second time
and the Found New Hardware Wizard starts again (see Figure 19 on page 28).
14 Select Option 1 and click Next (see Figure 19 on page 28).
27
3
Configuring the Agilent 82357 USB - GPIB Interface Driver under Windows XP Professional
Installing New IO Libraries for USB - GPIB Interface on Windows XP
Figure 19
Found New Hardware Wizard Windows XP, second detection
15 Select Option 1 and click Next to start the installation (see Figure 20 on
page 29).
28
Configuring the Agilent 82357 USB - GPIB Interface Driver under Windows XP Professional
Installing New IO Libraries for USB - GPIB Interface on Windows XP
Figure 20
3
Found New Hardware Wizard Windows XP, second detection
16 The system initializes the 82357A USB-PGIB Interface device. This might
take some seconds (see Figure 21 on page 30). Please wait until the
configuration screen appears, see step 14.
29
3
Configuring the Agilent 82357 USB - GPIB Interface Driver under Windows XP Professional
Installing New IO Libraries for USB - GPIB Interface on Windows XP
Figure 21
Installing the 82357A USB-GPIB Interface in Windows XP, second detection
17 The system automatically detects the installed USB-GPIB Interface. Select
Accept to start the IO Configuration Utility (see Figure 22 on page 31).
30
Configuring the Agilent 82357 USB - GPIB Interface Driver under Windows XP Professional
Installing New IO Libraries for USB - GPIB Interface on Windows XP
Figure 22
3
USB- GPIB Interface detected
18 Set the SICL Interface Name to hp82341 and the Bus Address to 30 as shown
in Figure 23 on page 32. Press OK to finish the configuration.
31
3
Configuring the Agilent 82357 USB - GPIB Interface Driver under Windows XP Professional
Installing New IO Libraries for USB - GPIB Interface on Windows XP
Figure 23
NOTE
Configuring 82357A USB-GPIB Interface
You need to type in the SICL interface name hp82341 with lower case letters.
19 Select Finish to close the wizard after the installation of the Hardware
driver(see Figure 24 on page 33). The installation of the USB device is
finished, the new hardware wizard will show a taskbar message in the right
hand corner of the Windows XP system: “Your new Hardware is found and
ready to use”.
32
Configuring the Agilent 82357 USB - GPIB Interface Driver under Windows XP Professional
Installing New IO Libraries for USB - GPIB Interface on Windows XP
Figure 24
3
Finish Device Driver Installation in Windows XP
33
3
34
Configuring the Agilent 82357 USB - GPIB Interface Driver under Windows XP Professional
Installing New IO Libraries for USB - GPIB Interface on Windows XP
Installation Instructions for use with the Agilent ChemStation Rev.
B.03.01 or higher
4 Upgrading Existing IO Libraries
Upgrading IO Libs under Windows XP
Agilent Technologies
36
35
4
Upgrading Existing IO Libraries
Upgrading IO Libs under Windows XP
Upgrading IO Libs under Windows XP
To upgrade the IO Libraries please followi the outlined steps below:
1 Using the Windows Explorer, locate and start m0101.exe in the \IO_Libs
directory of your ChemStation CD-ROM (CD 1 Installation). Follow the
instructions in the installation Wizard. This will upgrade the Standard
Interface Control Library (SICL) on your computer, update the registry and
update the Program Group called “Agilent I/O Libraries”.
The InstallShield Wizard opens and starts the IO Libraries upgrade.
Figure 25
Install Shield Wizard
2 An upgrade screen appears that shows the current version installed and the
new version that will be installed (see Figure 26 on page 37).
36
Upgrading Existing IO Libraries
Upgrading IO Libs under Windows XP
Figure 26
4
Upgrade IO drivers screen
3 When the upgrade screen appears, click Next to continue to the License
Agreement screen and then click Yes to accept the license terms, click Next
to display the Readme Information and click Next again to get to the current
settings.
4 By clicking Next the system installs the SICL libraries and VISA
components.
37
4
Upgrading Existing IO Libraries
Upgrading IO Libs under Windows XP
Figure 27
SICL libraries and VISA components
5 Select Run IO Config to check configuration of the interface card (see
Figure 27 on page 38).
6 The installed interface card or the USB-GPIB Interface appears already in
the Configured Interface List. Please verify VISA Name GBPI0 and SICL
Name hp82341. Otherwise select the configured interface and enter via
Edit (see Figure 28 on page 39).
38
Upgrading Existing IO Libraries
Upgrading IO Libs under Windows XP
Figure 28
4
IO Libraries Configuration
7 If necessary set the SICL Interface Name to hp82341, the Bus Address to 30
and in case of the 82350A/B GPIB cards set the System Controller flag. Please
note that the appearance of the dialog differs slightly depending on the type
of GPIB board installed.
39
4
Upgrading Existing IO Libraries
Upgrading IO Libs under Windows XP
Figure 29
40
Configuration screens for 82350A/B GPIB cards
Upgrading Existing IO Libraries
Upgrading IO Libs under Windows XP
Figure 30
NOTE
4
Configuration screens for 82357A USB-GPIB Interface
You need to type in the SICL interface name hp82341 with lower case letters for all cards as
well for the USB-GPIB Interface.
8 When installation is complete, leave the screen with OK. Note that you must
restart to initialize the IO Libraries for the GPIB cards.
41
4
42
Upgrading Existing IO Libraries
Upgrading IO Libs under Windows XP
Installation Instructions for use with the Agilent ChemStation Rev.
B.03.01 or higher
5 Working with your Agilent IO
Libraries
What has been changed by the Installation?
Agilent Technologies
44
43
5
Working with your Agilent IO Libraries
What has been changed by the Installation?
What has been changed by the Installation?
Depending on the installation option you selected, SICL and/or VISA were
installed and several program icons were created. In addition, an Agilent IO
Libraries Control was created.
During installation, the Agilent IO Libraries program group was created.
Figure 31
44
Agilent IO Libraries program group
Installation Instructions for use with the Agilent ChemStation Rev.
B.03.01 or higher
6 Troubleshooting
Missing files during installation
46
Problems with Plug and Play option GPIB cards
51
Problems with Plug and Play option using 82357A USB-GPIB
Interface
52
Agilent Technologies
45
6
Troubleshooting
Missing files during installation
Missing files during installation
During installation is might be possible that the system is looking for some IO
installation files which are not copied during the installation process. In this
case the system will ask you to manually locate the files during the installation
process.
1 The manual process might take place during the installation of the driver,
after choosing next to start installation (see Figure 32 on page 46).
Figure 32
Driver Installation Screen
2 The wizard asks for the file agtgpibclass.dll as shown in Figure 33 on
page 47.
46
Troubleshooting
Missing files during installation
Figure 33
6
Searching for file
3 Browse to C:\windows\system32\ and click OPEN (See Figure 34 on
page 47).
Figure 34
File Location
47
6
Troubleshooting
Missing files during installation
4 Select OK as shown in Figure 35 on page 48.
Figure 35
Choosing file
5 The wizard asks for the file agt82350.sys when an Agilent 82350 card is
installed as shown in Figure 36 on page 48.
Figure 36
Searching for file agt82350.sys
6 Browse to C:\windows\system32\drivers\ and click Open (See Figure 37 on
page 49).
48
Troubleshooting
Missing files during installation
Figure 37
6
File location
7 Select OK as shown in Figure 38 on page 49.
Figure 38
Select file
49
6
Troubleshooting
Missing files during installation
8 After the driver installation is completed press Finish. The configuration of
the driver can be done using the IO Config Utility described in the
installation procedures.
Figure 39
50
Finishing installation
6
Troubleshooting
Problems with Plug and Play option GPIB cards
Problems with Plug and Play option GPIB cards
Operation systems using Plug and Play may show problems during the
installation of the appropriate drivers for the GPIB (PCI) cards.
In order to overcome plug and play related problems please follow the steps to
install the card successfully.
1 If already installed, uninstall the Agilent IO Libraries from Control Panel,
Add/Remove Programs.
2 Remove the GBIB Interfaces (if installed) listed under the Device Manager
and shut down the computer. For Windows XP: Control Panel, double-click
System, Hardware Tab, Device Manager, GPIB Interfaces
3 Physically remove the GPIB card from the computers PCI Interface.
4 Restart the computer and log on to the system. Install the Agilent IO
Libraries revision M.01.01.04. Do not configure the card this time and shut
down the computer.
5 Install the GPBI card again and turn on the computer - the new hardware
found wizard will indicate that a PCI card is found. Select the second choice
to search from a list of known devices and there select the option for
Agilent Technologies, since the IO Libraries are already installed.
6 If the new hardware wizard does not run, check the device manager to
verify that the GPIB interfaces is listed and installed correctly.
7 To configure the GPIB card please follow the instructions stated in the first
sections of this booklet. Continue at step 13.
8 Start the IO Config Utility by clicking in the Icon
Run IO Config.
and choose
51
6
Troubleshooting
Problems with Plug and Play option using 82357A USB-GPIB Interface
Problems with Plug and Play option using 82357A USB-GPIB
Interface
Operating systems using Plug and Play may show problems during the
installation of the appropriate drivers for the USB-GPIB Interface.
In order to overcome problems related to Plug and Play, please follow the steps
to install the card successfully.
1 Verify the status of the USB-GPIB Interface in the Device Manager: Start,
Settings, Control Panel, double-click System, Hardware Tab, Device Manager, GPIB
Interfaces:
52
Troubleshooting
Problems with Plug and Play option using 82357A USB-GPIB Interface
Figure 40
6
Device Manager Windows XP USB-GPIB Interface, correctly installed
2 If a yellow exclamation mark is attached to the USB-GPIB Interface (see
Figure 41 on page 54) the driver needs to be installed a second time as in
“Installing New IO Libraries for USB - GPIB Interface on Windows XP” on
page 20.
53
6
Troubleshooting
Problems with Plug and Play option using 82357A USB-GPIB Interface
Figure 41
Device Manager Windows XP after first installation of USB-GPIB Interface
3 After the second installation, the system detects the USB device
automatically and allows you to configure the device properly. Select Accept
to start the IO Configuration Utility (see Figure 42 on page 55).
54
6
Troubleshooting
Problems with Plug and Play option using 82357A USB-GPIB Interface
Figure 42
USB-GPIB Interface detected
4 It is also possible to start the IO Config Utility by clicking in the Icon
and choose Run IO Config.
5 From the list of available interface types, choose the 82357 USB to GPIB (see
Figure 43 on page 56).
55
6
Troubleshooting
Problems with Plug and Play option using 82357A USB-GPIB Interface
Figure 43
IO Libraries Configuration
6 Select Configure. The configuration utility attempts to detect the selected
USB-GPIB interface in the PC in order to assign the correct resources to the
card.
7 Verify the correct settings mentioned in step 13 of the USB-GPIB Interface
Installation for Windows XP:
• SICL Interface Name to hp82341
• Bus Address to 30
8 Leave the 82357A USB/GPIB configuration screen with OK.
9 Leave the IO Library configuration screen with OK.
56
Troubleshooting
Problems with Plug and Play option using 82357A USB-GPIB Interface