Welch Allyn 300 Xenon Light Source User manual
Below you will find brief information for Xenon Light Source 300. The Xenon Light Source 300 is intended to be used as a 300 watt light source to supply illumination for surgical fiber optic headlights. It features a Cermax Xenon lamp with a life expectancy of more than 500 hours and accepts fiber optic bundles from various manufacturers. The unit includes a turret for selecting the correct fiber optic bundle type and a brightness control knob for adjusting the light intensity.
Advertisement
Advertisement
XENON 300 LIGHT SOURCE
Thank you for purchasing the Welch Allyn Xenon 300 Light Source. By following the use and care guidelines given in this booklet, you will be rewarded with years of dependable, trouble-free service from your new light source.
WARNING: All users of this product should be thoroughly trained in the medical procedures appropriate to the equipment. Furthermore, they should read and understand the instructions for all other equipment used in conjunction with this product.
Contents
1 - Conventions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Warnings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Indications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Contraindications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2 - Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
System Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Set-Up and Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3 - Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Cleaning Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Lamp Care and Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Fuse Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Basic Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
4 - Technical Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
5 - Warranty and Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Limited Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
i
ii
Contents Welch Allyn Xenon 300 Light Source
WARNING:
Extremely high energy light. Some fiber optic bundles will not withstand the light intensity of this light source. USE ONLY WELCH ALLYN FIBER OPTIC BUNDLE or
SUNOPTICS TECHNOLOGIES (formerly CUDA FiberOptics - 904-737-7611) CLAD
ROD (CR) MODEL FIBER OPTIC BUNDLES.
2
Chapter 1 Welch Allyn Xenon 300 Light Source
1
Conventions
Symbols
Power “ON”
Protective Earth Ground
Type BF
CAUTION; consult Directions
For Use
Power “STANDBY”
~
Alternating Current
Warning
3
4
Chapter 1 Conventions Welch Allyn Xenon 300 Light Source
Warnings
A warning statement in this manual identifies a condition or practice which, if not corrected or discontinued immediately, could lead to patient injury, illness, or death.
WARNING It is the user’s responsibility to check the specifications of any device attached to this light source to verify that it can transmit the intense light output without generating high temperatures and heat build-up in the device that can cause serious burns.
WARNING High intensity light can cause burns to tissue even if the tip of the fiber optic cable or attached terminal device is cool. Burns are caused by the absorption of light by tissue and the subsequent conversion of that light energy into thermal energy. There is serious risk of igniting fires if the energized light cable fibers are placed on flammable materials, and full-thickness burns can be caused by prolonged exposure to concentrated light energy. Anesthetized patients are particularly susceptible to burn injuries via this mechanism. IT IS THE
USER’S RESPONSIBILITY TO DETERMINE AND MAINTAIN THE MINIMUM
SAFE DISTANCE BETWEEN THE END OF THE ENERGIZED FIBER-OPTIC
LIGHT PATH AND ANY LIVING TISSUE OR FLAMMABLE MATERIAL.
WARNING When used to power the Welch Allyn MFI Headlight system, we recommend a standard 16” working distance separating the headlight and the patient. Under no circumstances should the energized headlight be placed closer than 10” to any object.
WARNING There is a risk of injury to tissue from exposure to the intense illumination. The risk of injury is reduced by using the minimum level of illumination necessary, by minimizing exposure time, and by avoiding close stationary viewing.
WARNING The high intensity light at the front of the light source and at the tip of a fiber optic bundle may give rise to high temperatures. To minimize the risk of injury, avoid contact with tissues.
WARNING As with any bright light, do not look into the beam. Permanent eye injury may result.
WARNING Total system risk current - excluding the patient applied instrument - should not exceed 100 microamps. an isolation transformer is required if total system risk current exceeds 100 microamps when accessories are interconnected.
WARNING POSSIBLE EXPLOSION HAZARD. Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide.
WARNING
The cover of the light source should never be removed.
Electrical shock hazard due to high internal voltage. there are no user serviceable parts inside this light source except for the lamp and fuse, which are accessible through the bottom and side door panels. refer all service to Welch Allyn.
WARNING
T8.0L 250V
Replace fuses as marked. See Fuse Replacement Section.
Directions for Use Chapter 1 Conventions
5
Cautions
A caution statement in this manual identifies a condition or practice which, if not corrected or discontinued immediately, could lead to equipment failure, equipment damage, or data loss.
Caution Consult users manual for important information.
Caution This system should only be used by qualified medical personnel.
Caution
To reduce the chance of direct visual exposure to the light, set the light intensity knob to the minimum level and plug the fiber optic bundle into the light source before turning the power ON.
Caution When the unit is on, never disconnect the fiber optic bundle from the light source or the accessory from the end of the fiber bundle. High intensity light from the end of the fiber optic bundle and light source can ignite flammable materials (i.e. drapes).
Caution Grounding reliability is achieved only when connected to hospital-use or hospital-grade receptacles. Inspect electrical plug and cord routinely. Do not use if damage is discovered.
Caution
PROVIDE VENTILATION TO PREVENT OVERHEATING. Do not cover or drape the light source.
Caution
Do not use the turret as a light attenuator or with the turret misaligned with the light port, overheating will result.
Caution
The ends of the fiber bundle get hot during use. Allow to cool before removing.
Caution Do not operate light source without lamp module in place.
Caution If this light source is used during invasive surgery, the failure of a component within the system, causing loss of illumination, may directly cause risk to the patient. Therefore, it is recommended to have alternative means of illumination available.
Caution Since the light source uses a custom lamp, always have a spare lamp available for replacement. Use only Welch Allyn replacement lamp number
04980. Use caution because lamp may be hot. Allow the lamp to cool before replacing.
Caution
IPXØ - Equipment not protected against the ingress of water. Do not use or store liquids above or on the light source.
6
Chapter 1 Conventions
Indications
For use with the Welch Allyn Headlight System
Contraindications
• Neonate transillumination
• Ophthalmic procedures
• For use with Rhinolaryngoscopes
• For use with Endoscopes
Welch Allyn Xenon 300 Light Source
2
Components
System Description
The Welch Allyn 49801 light source is intended to be used as a 300 watt light source to supply illumination for surgical fiber optic headlights.
Power Switch - “l” indicates “ON” position
Fiber Optic Bundle Turret - accepts fiber optic bundles for
ACMI, Wolf, Olympus and Storz.
Brightness Control - increases or decreases light output intensity level.
Cooling Fans - provides cooling to light source
1
SERIAL
NUMBER
C22.2
NO. 601.1
W W
US US MP MP
OLY OLY
DO NOT REMOVE LAMP IMMEDIATELY AFTER
OPERATION.
ALLOW LAMP TO COOL 5 MINUTES.
TO
THE
PREVENT OVERHEATING, REPLACE ONLY WITH
SAME TYPE AND RATING OF LAMP OR LAMP
MODULE.
LAMP
SAME TYPE AND RATING OF LAMP OR LAMP
READ INSTRUCTIONS BEFORE REPLACING
OR LAMP MODULE.
DO
IN
NOT OPERATE THE LIGHTSOURCE WITHOUT LAMP MODULE
PLACE.
FOR LAMP REPLACEMENT,
WITH PART NUMBER 04980,
REPLACE
14 VDC 300 W
3
US
MP
OLY
W
2
Active light port
Power Module - accepts
4
AC power cord.
Fuse Holders - contains the main fuses.
R
N344
74227
REF 49801
200-240V ~
L IS T E D R
/
IEC601-1
50 60Hz, 6.5A UL2601
C
L IS T E D R
CAN/CSA
C22.2
NO. 601.1
9
5
6
DO NOT REMOVE LAMP IMMEDIATELY AFTER
OPERATION.
ALLOW LAMP TO COOL 5 MINUTES.
DO
IN
PLACE.
TO
THE
PREVENT OVERHEATING, REPLACE ONLY WITH
SAME TYPE AND RATING OF LAMP OR LAMP
MODULE.
LAMP
READ INSTRUCTIONS BEFORE REPLACING
OR LAMP MODULE.
NOT OPERATE THE LIGHTSOURCE WITHOUT LAMP MODULE
FOR LAMP REPLACEMENT,
WITH PART NUMBER 04980,
REPLACE
14 VDC 300 W
Lamp Module Access Door - access for lamp replacement.
Handle - means to carry light source.
8
7
8
Chapter 2 Components Welch Allyn Xenon 300 Light Source
Set-Up and Operation
1.
Unpack unit and visually inspect it for damage that may have occurred during shipment.
2.
Plug power cord into light source power module and the remaining end into a hospital grade receptacle.
3.
Select type of fiber optic bundle port (ACMI, Wolf, Olympus, Storz) and rotate turret to align with light port.
4.
Insure that the ends of the fiber optic bundle are clean before each use. Any residue on the ends of the fiber optic will absorb the high intensity light and will overheat or damage the fiber optic bundle.
5.
Insert fiber optic bundle into light port and connect the headlight luminare. Set the light intensity to minimum level. This prevents inadvertent shining into the patient’s or user’s eyes when turned on.
Caution When the unit is on, NEVER disconnect the fiber optic bundle from the light source or the headlight from the fiber optic bundle. High intensity light from the end of the fiber optic bundle and light source can ignite flammable materials
(i.e. drapes).
6.
Turn the unit on.
7.
Set light intensity to minimum level necessary for adequate visualization.
8.
Turn unit off when not in use.
9.
Allow the light source to cool before removing fiber optic bundle. The ends of the fiber optic bundle get hot during use..
Caution With the output of high intensity (Xenon) light sources, it is possible to convey a large amount of light energy and to concentrate this energy in a very small area. This may result in thermal injury to the tissue.
Caution The risk of injury is increased under the following conditions:
• Prolonged close contact or close stationary viewing of the mucosa or tissue.
• Using a high intensity light source at high output levels.
• Using multiple instruments to illuminate the same area.
Caution The following recommendations will reduce the risk of thermal injury:
• Use the minimum level of illumination necessary for adequate visualization.
• When possible, avoid close stationary viewing.
• Minimize exposure time.
3
Maintenance
Cleaning Instructions
Note
Do not sterilize or immerse the Welch Allyn Xenon 300 Light Source.
1.
Turn power off and disconnect the power cord from both the wall outlet and the
Welch Allyn Xenon 300 Light Source.
2.
Wipe the external surfaces clean with a cloth dampened with mild soap and water or use one of the following cleaning/disinfecting solutions. DO NOT IMMERSE.
• 70% isopropyl alcohol
• 10% mild bleach solution
3.
Wipe the power cord clean using the same cleaning/disinfecting solutions listed above. Wipe dry with a clean cloth. Take care not to get the plug prongs wet. DO NOT
IMMERSE.
4.
Do not plug the unit into a wall outlet until it is thoroughly dry.
Caution Clean the ends of the fiber optic bundle before each use. Any residue
(including cleaning and disinfecting solutions) will absorb high intensity light on the surface of the fiber optic. Absorbed light will overheat the residue causing damage to the fiber optic bundle.
9
10
Chapter 3 Maintenance Welch Allyn Xenon 300 Light Source
Lamp Care and Replacement
The lamp in the Welch Allyn Xenon 300 Light Source will typically provide more than 500 hours of service. The symptoms indicating the need for replacement are:
175 Hours
HOURMETER ASSY
0 HR
FRONT
TO REMOVE LAMP
PULL OUT.
DO NOT REMOVE
IF LAMP IS HOT.
8AA 1104
SERIAL NUMBER
• Low light output
• The unit requires more than four ignition strikes to start the lamp going into operation.
• The lamp life indicator reads more than 500 hours.
The light source is designed for simple and quick changing of the lamp module, which houses the lamp.
Removing the Lamp
1.
Turn unit off and unplug power cord from its power source.
2.
Allow the lamp to cool.
3.
Unlock and open access door located on the side of the light source.
4.
Grasp the two handles and pull the entire lamp module out of the light source.
Inserting a Lamp
:
S
MPU
OLY
W
1.
Orient the new lamp module so that the window of the lamp is facing the front of the light source.
2.
Carefully align the connection pins with the brass sockets on the lamp module and push firmly to insure proper connection.
3.
Close and lock the access door.
Caution DO not incinerate used lamps to avoid risk of explosion.
Directions for Use
Fuse Replacement
Chapter 3 Maintenance
11
DO NOT REMOVE LAMP IMMEDIATELY AFTER
OPERATION.
ALLOW LAMP TO COOL 5 MINUTES.
TO
THE
PREVENT OVERHEATING, REPLACE ONLY WITH
SAME TYPE AND RATING OF LAMP OR LAMP
MODULE.
LAMP
READ INSTRUCTIONS BEFORE REPLACING
OR LAMP MODULE.
DO
IN
NOT OPERATE THE LIGHTSOURCE WITHOUT LAMP MODULE
PLACE.
FOR LAMP REPLACEMENT, REPLACE
WITH PART NUMBER 04980, 14 VDC 300 W
1.
Turn unit off prior to checking the fuses.
2.
Unplug the light source.
3.
Use a flat blade screwdriver to remove fuse covers and check for blown fuse.
4.
Replace the blown fuses with the same size and rating.
5.
Replace fuse covers.
Ordering Information
Part Information
49800
Welch Allyn Xenon 300 Light Source
100-120/200-240 VAC, 50-60 Hz, 5.5A
50-60 Hz
USA Plug type
Replacement Parts
04980-U
76400
Replacement Lamp Module
USA Power Cord
12
Chapter 3 Maintenance
Basic Troubleshooting
Welch Allyn Xenon 300 Light Source
Symptom Possible Problem Remedy
Power does not come on.
Power plug is not plugged into the receptacle.
Power switch is off.
Fuse is blown.
Lamp door is not closed tightly.
Lamp module is not installed incorrectly.
No light output.
Power plug is not plugged into receptacle.
Power switch is off.
Fuse is blown.
Lamp door is not closed tightly.
Lamp module is installed incorrectly.
Turret misaligned with light port.
Lamp burned out.
Fiber bundle is not connected.
Light attenuator is set to minimum.
Plug in power cord.
Turn on power switch.
Replace fuse.
Close lamp door completely.
Reinstall lamp module correctly.
Plug in power cord.
Turn on power switch.
Replace fuse.
Close lamp door completely.
Reinstall lamp module.
Set turret to detent position (light port aligned).
Replace lamp.
Connect fiber bundle correctly.
Set the light attenuator to minimum level needed for adequate visualization.
Replace lamp.
Lamp flickers or dims.
Light source turns off after a few minutes of operation.
Low light output
Lamp is getting old.
Obstructed air intake; overheating causes thermal switch to trip.
Fan not running; overheating causes thermal switch to trip.
Accessory damaged.
Lamp is old.
Light transmission rod is damaged or contaminated.
Several ignition attempts needed to start lamp.
Three or fewer ignition attempts are normal.
Allow unit to cool (5-10 minutes). Remove obstruction(s).
Contact Welch Allyn.
Replace accessory.
Replace lamp.
Clean with isopropyl alcohol and allow to dry.
Use care not to chip edges.
Contact Welch Allyn for repair.
Replace lamp if four or more ignition attempts are required.
4
Technical Information
Specifications
Electrical
Dimensions
Lamp Operating Power:
Lamp Current:
Input Voltage:
Lamp:
Lamp Life:
300 watts DC, nominal
21 Amps DC, nominal
100-120/200-240 VAC, 50/60 Hz, 5.5A
Cermax® Xenon Lamp
500 hours
Fiber Optic Bundle Interface
Turret Fiber Bundle Port accepts ACMI, Wolf, Olympus, and Storz
Environment
Size:
Weight:
12.75” (324mm)W x 4.75” (121mm) H x 12.5” (318mm) D
12.5 lbs (5.7kg)
Operating:
Transport/Storage: -20 o
10 o
C to 40 o
C, 95% R.H., 500hPa - 1060 hPa, Altitude
C to 49 o
C, 95% R.H. Max, 500hPa - 1060hPa, Altitude
Fuses
5mm fuse, T 8.0L 8 amp 250V.
13
14
Chapter 4 Technical Information
Approvals
UL 2601-1
CAN/CSA C22.2
No601.1-M90
IEC 601-1
IEC 601-1-2:1993
Welch Allyn Xenon 300 Light Source
Equipment Classification
Class 1 Equipment
Continuous Operation
Patents
Patents Pending
Equipment Disposal: Do not incinerate this equipment at the end of useful life for risk of lamp explosion. Equipment contains no hazardous materials and can be discarded without environmental risks.
5
Warranty and Service
Limited Warranty
The Welch Allyn Xenon 300 Light source is guaranteed by Welch Allyn against all manufacturing defects. Welch Allyn will repair or replace, free of charge, any parts of its own manufacture that prove to be defective for reasons other than misuse, neglect, damage in shipment, or normal wear. This warranty is in effect for one year from original date of purchase.
Service
Service Policy
A Welch Allyn Service Center must perform or approve all repairs on products under warranty. Unauthorized repairs will void the warranty. Qualified electronics personnel or a
Welch Allyn Service Center should repair products out of warranty.
Technical Assistance
If you have an equipment problem that you cannot resolve, you may call the Welch Allyn
Service Center nearest you for assistance. Call the service center on normal business days at the phone numbers listed on page 16.
If you are advised to return a product to Welch Allyn for repair or routine maintenance, schedule the repair with the service center nearest you.
Before returning a product for repair you must obtain authorization from Welch Allyn. Our service personnel will give you a Service Notification number. Note this number on the outside of your shipping box. Service centers do not accept deliveries without an Service
Notification number.
15
16
Chapter 5 Warranty and Service
Welch Allyn Service Centers
For Service or Repair
USA 800 535 6663
315 685 4560
Canada 800 561 8797
European Call Center 353 46 9067790
Germany 49 7477 9271 70
Latin America 305 669 9003
Singapore 65 6419 8100
United Kingdom 44 1332 368968
Welch Allyn Xenon 300 Light Source
Australia 61 2 9638 3000
800 074 793
China 86 21 6327 9631
France 33 1 55 69 58 49
Japan 81 3 3219 0071
Netherlands 31 157 505 000
South Africa 27 11 777 7555
Welch Allyn, Inc.
4341 State Road
P.O. Box 220
Skaneateles Falls, NY
13153-0220
Tel: (315) 685-4560
Printed in U.S.A.
Part No. 704264 Rev. B
![](http://s3.manualzz.com/store/data/022767929_1-3c50562a7f9426e503b0f4b313500672-210x147.png)
Advertisement
Key features
- 300 watt light source
- Cermax Xenon lamp
- 500+ hour lamp life
- Turret for fiber optic bundle selection
- Brightness control knob
- Accepts ACMI, Wolf, Olympus, and Storz fiber optic bundles
- UL 2601-1, CAN/CSA C22.2, No601.1-M90, IEC 601-1, IEC 601-1-2:1993 approved