ELECTRONIC Counter TYPE AG5xEZD, AG5xEZ and AG4xEZD


Add to my manuals
27 Pages

advertisement

ELECTRONIC Counter TYPE AG5xEZD, AG5xEZ and AG4xEZD | Manualzz
ELECTRONIC Counter
EZ /
TYPE AG5xEZD, AG5xEZ and AG4xEZD
EZD
with Reed Sensors or Wiegand Sensors
Operating Instructions
Please read and keep the operating instructions.
Bopp & Reuther Messtechnik GmbH
Am Neuen Rheinhafen 4 - 67346 Speyer
Postfach 1709 - 67327 Speyer
Phone: +49(6232) 657-0
Fax: +49(6232) 657-505
[email protected]
www.bopp-reuther-mt.de
We reserve the right to change the.
measurements, weights and other
technical data.
Printed in the
Federal Republic of Germany
A-EN-01213-00Rev B
Latest Modific.: 0909-01
1 of 23
Electronic Counter
Table of Contents
1 General Description ................................................................................................................................... 3
1.1 Proper Use .......................................................................................................................................... 3
1.2 Transport, Delivery and Storage ......................................................................................................... 3
1.3 Warranty.............................................................................................................................................. 3
1.4 Repairs ................................................................................................................................................ 3
2 System Description.................................................................................................................................... 4
3 Model AG4xEZD (Wiegand Sensor with NAMUR Power Supply) ............................................................ 5
3.1 Technical Data for AG4xEZD.............................................................................................................. 5
3.2 Electrical Connection .......................................................................................................................... 6
3.3 Terminal Diagram for the Variation with Original Pulses (AG4xEZD) ............................................... 7
3.4 Terminal Diagram for the Variation with Scaled Pulses (AG4xEZD) ................................................ 8
4 Model AG5xEZD (Reed Sensor with NAMUR Power Supply) .................................................................. 9
4.1 Technical Data for AG5xEZD.............................................................................................................. 9
4.2 Electrical Connection .......................................................................................................................... 9
4.3 Terminal Diagram for the Variant with Original Pulses (AG5xEZD) ................................................. 10
4.4 Terminal Diagram for the Variant with Scaled Pulses (AG5xEZD)................................................... 11
5 Model AG5xEZ (Reed Sensor with Battery Power Supply)..................................................................... 12
5.1 Technical Data for AG5xEZ .............................................................................................................. 12
5.2 Terminal Diagram AG5xEZ ............................................................................................................... 12
6 Assembly and Installation ........................................................................................................................ 13
6.1 Safety Instructions............................................................................................................................. 13
6.2 Installation Instructions ..................................................................................................................... 13
6.2.1 Temperature Ranges, Casing Distance to Sensor..................................................................... 13
7 Operating Overview ................................................................................................................................. 14
7.1 Taking in Operation........................................................................................................................... 14
7.2 Display and Control Elements........................................................................................................... 14
7.2.1 Operating When Set To Operation Mode 1................................................................................ 15
7.2.2 Operating When Set To Operation Mode 2................................................................................ 16
7.2.4 Changing the Operation Mode ................................................................................................... 18
7.3 Configuration of the Electronic Meter (Option) ................................................................................. 18
7.3.1 SensorPort.................................................................................................................................. 18
7.3.1.1 Record Card "General"............................................................................................................ 19
7.3.1.2 Record Card "Constants" ........................................................................................................ 19
7.3.1.3 Record Card "Operation Type"................................................................................................ 20
7.3.2 Operating Module (BEZW) ......................................................................................................... 21
8. Alarm and Warning Messages ............................................................................................................... 23
Bopp & Reuther Messtechnik GmbH
Page 2 of 23
Electronic Counter
1 General Description
1.1 Proper Use
The electronic counter, EZ and EZD, may only be mounted on oval wheel meters from Bopp &
Reuther Messtechnik GmbH. They are for the display of the volume, flow and pulse transmission.
The technical data, the information in the configuration data sheet, the information on the data plate
and from the meter data plate are authoritative for the employment of the meters.
The electronic meters of type AG5xEZD and AG5xEZ can be used for all oval wheel meters with
Reed sensors, AG5x, and the electronic meters of type AG4xEZD can be used for all oval wheel
meters with Wiegand sensors.
The manufacturer is not liable for any damages, which occur due to improper use or use, which is
not accordance with these instructions.
If employed in an explosion-endangered area, the valid regulations and standards are to be
systematically observed.
1.2 Transport, Delivery and Storage
Inspection of the delivery:
The shipment is to be checked for completeness upon receipt. Compare the instrument data with
the information on the packing slip and the order documents.
Report any transport damage immediately after receiving the delivery. Damages, which are
reported later, will not be recognized.
The instruments are to be protected against wetness, moisture, dirt, shocks and damage.
1.3 Warranty
The scope and the period of warranty are found in the contractual terms of delivery.
Any warranty claims require that the instrument is properly installed and placed into operation
according to the operating instructions, which apply to the instrument. The required installation
placing in operation and servicing work may only be performed by qualified and authorized
personnel.
1.4 Repairs
The instrument is to be sent to Bopp & Reuther Messtechnik GmbH for repairs.
For all work on the opened instrument, the direct short-circuiting of the installed lithium cell is not
allowed. An explosive atmosphere could be ignited.
The valid regulations concerning the opening and repairing of electrical equipment are to be
observed.
For the battery fed model, AG5xEZ, the replacement of the lithium cell, which is specially equipped
with a cable connector and current-limiting resistor, is allowable in an Ex-area.
The smaller battery (battery buffered RAM), which is firmly soldered, for the NAMUR model
(AG5xEZD and AG4xEZD) may only be replaced by Bopp & Reuther Messtechnik GmbH or an
authorized repair facility.
Bopp & Reuther Messtechnik GmbH
Page 3 of 23
Electronic Counter
2 System Description
These operating instructions apply to the NAMUR powered instruments, AG5xEZD and AG4xEZD,
and the battery-powered instrument AG5xEZ.
You can recognize the instrument type, which your are using, from the information printed in the
terminal area of the instrument and on the data plate of the meter. The model code includes, in
addition to the items shown below, the name of the meter.
ATEX
Terminal
strip
Terminal name
14 13 12 11 10 9
8
7
6
5
4
3
2
1
ge
-
+
-
+
-
+
-
+
gn
ge
2
gn
4
1
Sensors
3
2
1
NAMUR signal
Year Manuf: 2001
SN 2-40-xxxxx-xxx
Model code
Type: Ag4xEZD
Electronic number
Electr.-No. 4N-xxxx
Serial No.
Bopp & Reuther Messtechnik GmbH
Carl-Reuther-Str. 1 68305 Mannheim
Manufacturer
Connection for
interface
type IF232
Slide switch for selection of pulse
output for Namur signal 1
Original (O)
(S) Scaled
Overview of the instruments described:
Pulse output with supply electric
circuits (Namur)
Battery instrument without supply
electric circuits
Reed Sensor
AG5xEZD
AG5xEZ
Wiegand Sensor
AG4xEZD
Not available
The following shows the technical data for each instrument type with the associated terminal
diagram.
The data, which apply to the corresponding model for your instrument, are the sole references for
connecting your instrument.
Bopp & Reuther Messtechnik GmbH
Page 4 of 23
Electronic Counter
3 Model AG4xEZD (Wiegand Sensor with NAMUR Power Supply)
3.1 Technical Data for AG4xEZD
Ignition protection type:
"Intrinsically safe“ EEx ib II C T4 Ta=-20...+70°C
according to
EN 50014, EN 50020 and DMT 01 ATEX E 072 X.
Connection to approved intrinsically safe power supply
device according to EN 60947-5-6 (NAMUR)
Control circuit:
1 to 4 channels
Terminals 1+2 ; 3+4 ; 5+6 ; 7+8 :
Four intrinsically safe connections for pulse output
according to NAMUR, power supply approx. 8 VDC
Security related data for intrinsically safe
circuit max. 30 V, 100 mA, 166 mW, Ci=15 nF, Li=4 µH
Terminals 11+12 ; 13+14:
Two sensor electrical circuits for Wiegand sensors,
together max. 3.6 V, 1 mW, 3.7 mA, Ca=20µF,
La=1000mH
Terminals 9+10:
Potential equalization/ground, connected
with the meter casing
Service plug:
Safe for connection of passive contact points
Casing protection type:
IP 65 according to EN 60529
Ambient temperature for function:
-10°C to +55°C
Ambient temperature for storage:
-10°C to +70°C
Cable connection:
EMC adapted cable fitting
M20 x 1.5 for max. cable outside Ø 12 mm
(previously: PG 13.5)
Cable specification:
2 to 8 wire, shielded, pairs intertwisted
Max. cable length:
Ca − Ci
in meters, but line resistance of
Ck
max. 25 Ω per wire max. 110nF
l=
Ca:
Admissible external capacity of the
Ci:
Ck:
power supply device in nF
Internal capacity of the EZD = 15nF
Specific cable capacity in nF/m.
Display:
8-position LCD display with 9 mm digit height, decimal
point and an identification field below the LCD display
Data backup:
battery buffered RAM
Bopp & Reuther Messtechnik GmbH
Page 5 of 23
Electronic Counter
EMC:
Harmonized standards and recommendations applied:
•
EN50082-2
Interference immunity industrial area
•
EN50081-1
Spurious radiation living area (emission)
•
EN55011: 1998+A1:1999 Group 1, Class B (critical
values and measurement method)
•
NAMUR recommendation NE21
EX:
EC Type approval certificate DMT 01 ATEX E 072 X
CE:
The instrument corresponds to the EC guidelines
94/9/EC (explosion protection) and 89/336/EEC
(electromagnetic compatibility) including all published
changes and amendments as of today, May 28, 2001.
The conformity with the tested type is confirmed by the
application of the CE symbol
3.2 Electrical Connection
Loosen four screws on the casing cover. Carefully remove the cover and attach it to the casing with
one of the cover screws so that the cable connections are relieved of strain.
Insert the two-wire connection cable for the NAMUR power supply through the cable fitting and
connect to the terminals 1 and 2.
When operating with 2-wire technology connection must always be made to 1 and 2. This
provides power to the electronics.
Using the slide switch in the terminal area, you can select if the original or scaled pulse output is
placed on terminals 1 and 2. See diagram in chapter 2 "System Description."
Power supply acc. NAMUR approx. 8 VDC
Bopp & Reuther Messtechnik GmbH
Page 6 of 23
Electronic Counter
3.3 Terminal Diagram for the Variation with Original Pulses (AG4xEZD)
This terminal diagram must be used, if the slide switch in the terminal area is in the original pulse
(O) position. See diagram in chapter 2 "System Description."
+ 1
-
Original pulses
Connection Namur power supply device
2
+ 3
Connection optional
-
4
+ 5
-
Scaled pulses
+ 7
-
Connection optional
6
Connection optional
8
9
Shielding in metal cable fitting
10
gn 11
Wiegand sensor 1
ge 12
gn 13
Wiegand sensor 2
ge 14
1+2
Volume pulse output (original impulses), fmax 1 kHz, pulse duration 0.5 ms
3+4
Volume pulse output original impulses, inverted to 1+2, fmax 1 kHz, pulse duration 0.5
ms
5+6
Volume pulse output scaled, fmax 100 Hz,
with original frequencies to approx. 8 Hz the Pulse duration depends on the relation
between original and output frequency,
above this the pulse-pause relation is 1:1,
or if the "warning message" in SensorPort was activated:
warning output, but then no scaled pulse is possible.
7+8
Alarm output
9+10
Potential ground
11+12
Input 1 connected with Wiegand sensor 1
13+14
Input 2, connected with Wiegand sensor 2
All outputs correspond to EN 60947-5-6 (EN 50227)
Bopp & Reuther Messtechnik GmbH
Page 7 of 23
Electronic Counter
3.4 Terminal Diagram for the Variation with Scaled Pulses (AG4xEZD)
This terminal diagram must be used, if the slide switch in the terminal area is in the scaled pulse
(S) position. See diagram in chapter 2 "System Description."
+ 1
-
Scaled pulses
+ 3
-
Connection optional
4
+ 5
-
Original pulses
Connection optional
6
+ 7
-
Connection Namur power supply device
2
Connection optional
8
9
Shielding in metal cable fitting
10
gn 11
Wiegand sensor 1
ge 12
gn 13
Wiegand sensor 2
ge 14
1+2
Volume pulse output, scaled, fmax 100 Hz,
with original frequencies to approx. 8 Hz the Pulse duration depends on the relation
between original and output frequency,
Above this the impulse-pause relation is 1:1
or if the "warning message" in SensorPort was activated:
warning output, but then no scaled pulse is possible.
3+4
Volume pulse output original impulses, inverted to 5+6, fmax 1 kHz, pulse duration 0.5
ms
5+6
Volume pulse output original pulse, fmax 1 kHz, pulse duration 0.5 ms
7+8
Alarm output
9+10
Potential ground
11+12
Input 1, connected with Wiegand sensor 1
13+14
Input 2, connected with Wiegand sensor 2
All outputs correspond to EN 60947-5-6 (EN 50227)
Bopp & Reuther Messtechnik GmbH
Page 8 of 23
Electronic Counter
4 Model AG5xEZD (Reed Sensor with NAMUR Power Supply)
4.1 Technical Data for AG5xEZD
Ignition protection type:
"Intrinsically safe “ EEx ib II C T4 Ta=-20...+70°C
according to
EN 50014, EN 50020 and DMT 01 ATEX E 072 X.
Connection to approved intrinsically safe power supply
device according to EN 60947-5-6 (NAMUR)
Control circuit:
1 to 4 channels
Terminals 1+2 ; 3+4 ; 5+6 ; 7+8 :
Four intrinsically safe connections for pulse output
according to NAMUR; power supply approx. 8 VDC
Security related data for intrinsically safe
circuit max. 30 V, 100 mA, 166 mW, Ci=15 nF, Li=4 µH
Terminals 11+12 ; 13+14:
Two sensor electric circuits for Reed sensors,
together max. 10.6 V, 697 mW, 437 mA
Terminals 9+10:
Potential equalization/ground, connected
with the meter casing
Service plug:
Safe for connection of passive contact points
Casing protection type:
IP 65 according to EN 60529
Ambient temperature for function:
-10°C to +55°C
Ambient temperature for storage:
-10°C to +70°C
Cable connection:
EMC adapted cable fitting
M20 x 1.5 for max. cable outside Ø 12 mm
(previously: PG 13.5)
Cable specification:
Max. cable length:
2 to 8 wire, shielded, pairs intertwisted
Ca − Ci
in meters, but line resistance of
Ck
max. 25 Ω per wire max. 110nF
l=
Ca:
Admissible external capacity of the
supply device in nF
Ci:
Internal capacity of the EZD = 15nF
Ck:
Specific cable capacity in nF/m.
power
Display:
8-position LCD display with 9 mm digit height, decimal
point and an identification field below the LCD display
Data backup:
battery buffered RAM
4.2 Electrical Connection
Loosen four screws on the casing cover. Carefully remove the cover and attach it to the casing with
one of the cover screws so that the cable connections are relieved of strain.
Insert the two-wire connection cable for the NAMUR power supply through the cable fitting and
connect to the terminals 1 and 2.
When operating with 2-wire technology connection must always be made to 1 and 2. This
provides power to the electronics.
Using the slide switch in the terminal area, you can select if the original or scaled pulse output is
placed on terminals 1 and 2. See diagram in chapter 2 "System Description."
Bopp & Reuther Messtechnik GmbH
Page 9 of 23
Electronic Counter
4.3 Terminal Diagram for the Variant with Original Pulses (AG5xEZD)
This terminal diagram must be used, if the slide switch in the terminal area is in the original pulse
(O) position. See diagram in chapter 2 "System Description."
+ 1
Original pulses
-
2
+ 3
-
Connection Namur power
supply device
Connection
optional
4
+ 5
Scaled pulses
-
6
+ 7
-
8
Connection
optional
Connection
optional
9
Shielding in metal cable fitting
10
ge 11
Reed sensor 1
ge 12
gn 13
Reed sensor 2
gn 14
1+2
Volume pulse output original pulse, fmax 1 kHz, pulse duration 0.5 ms,
Pulse-pause relation about 1:1.7
3+4
Volume pulse output phase shifted to 1+2, fmax 1 kHz
The phase shift depends on the impulse pause relation at output 1+2 (approx. 66° with
impulse pause relation 1:1,7)
5+6
Volume pulse output scaled, fmax 100 Hz,
with original frequencies to approx. 8 Hz the Pulse duration depends on the relation
between original and output frequency,
above this the impulse-pause relation is 1: 1, or
if the "warning message" in SensorPort was activated:
warning output, but then no scaled pulse is possible.
7+8
Alarm output
9+10
Potential ground
11+12
Input 1, connected with Reed sensor 1
13+14
Input 2, connected with Reed sensor 2
All outputs according to EN 60947-5-6 (EN 50227)
Bopp & Reuther Messtechnik GmbH
Page 10 of 23
Electronic Counter
4.4 Terminal Diagram for the Variant with Scaled Pulses (AG5xEZD)
This terminal diagram must be used, if the slide switch in the terminal area is in the scaled pulse
(S) position. See diagram in chapter 2 "System Description."
+ 1
Scaled pulses
-
2
+ 3
-
Connection Namur power
supply device
Connection
optional
4
+ 5
Original pulses
-
6
+ 7
-
8
Connection
optional
Connection
optional
9
Shielding in metal cable fitting
10
ge 11
Reed sensor 1
ge 12
gn 13
Reed sensor 2
gn 14
1+2
Volume pulse output, scaled, fmax 100 Hz,
with original frequencies to approx. 8 Hz the Pulse duration depends on the relation
between original and output frequency,
above this the impulse-pause relation is 1:1
or if the "warning message" in SensorPort was activated:
warning output, but then no scaled pulse is possible.
3+4
Volume pulse output original impulses, phase shifted to 5+6, fmax 1 kHz.
The phase shift depends on the impulse pause relation at the output 5+6
(approx. 66° with impulse pause relation 1:1,7)
5+6
Volume pulse output original impulse, fmax 1 kHz,
impulse pause relation approx. 1:1,7
7+8
Alarm output
9+10
Potential ground
11+12
Input 1, connected with Reed sensor 1
13+14
Input 2, connected with Reed sensor 2
All outputs according to EN 60947-5-6 (EN 50227)
Bopp & Reuther Messtechnik GmbH
Page 11 of 23
Electronic Counter
5 Model AG5xEZ (Reed Sensor with Battery Power Supply)
5.1 Technical Data for AG5xEZ
Ignition protection type:
"Intrinsically safe “ EEx ib II C T4 Ta=20°C...+40°C according to EN 50014, EN 50020
and
DMT 01 ATEX E 072 X
Power supply:
3.6 V battery, service life about 5 years
Casing protection type:
IP 65 according to EN 60529
Ambient temperature:
-10°C to +55°C
Ambient temperature for function
in non-Ex-area:
-10°C to +55°C
Ambient temperature for storage:
-10°C to +70°C
Service plug:
Safe for connection of contact points
Display:
8-position LCD display with 9 mm digit height,
decimal point and an identification field below the
LCD display
5.2 Terminal Diagram AG5xEZ
Plug-in for power supply battery
1
Terminal
strip
J3
1
2
3
4
+ - + Reed1 Reed2
2
J1
+ Battery
Terminal name
Model code
Type: Ag5xEZ
Year Manuf: 2001
Electronic number
Electr. No. xx-xxxx
SN 2-40-xxxxx-xxx
Serial No.
Manufacturer
Bopp & Reuther Messtechnik GmbH
Carl-Reuther-Str. 1 68305 Mannheim
Connection for
interface
ATEX
J2
Bopp & Reuther Messtechnik GmbH
Page 12 of 23
Electronic Counter
6 Assembly and Installation
6.1 Safety Instructions
Before assembly and placing in operation, read and observe these maintenance instructions and
the operating instructions for oval wheel meters.
6.2 Installation Instructions
The installation of the complete oval wheel meter with an electronic meter is to be done according
to the installation instructions for oval wheel meters.
The electronic meter, according to the ignition protection type " intrinsically safe", EEx ib IIC T4,
can be employed along with the pulse trigger in an Ex-area.
The installation must be done in accordance with the national regulations and in particular with
respect to explosion protection. In Europe EN 60079-14 (VDE 0165 Part 1) applies.
In addition to the separately listed ambient temperature limits for the proper functioning, the
absolute temperature limits from the type approval certificate must be observed also, if the
instrument is to be employed in an hazardous area. The influence of the material being measured
must be considered in every case.
Place shielding in the metal cable fitting.
Only for AG4xEZD and AG5xEZD:
Connection cable to the evaluation device, 2 to 8 wires depending on the instrument variant, wire
2,
cross section 0.75 mm pairs intertwisted with joint shielding. For maximum length see the formula
in the "technical data."
6.2.1 Temperature Ranges, Casing Distance to Sensor
MM= material being measured, U= ambient
AG41EZD
AG42-44EZD
AG50EZ
1. Casing distance 0 mm
1. Casing distance 50 mm
1. Casing distance 0 mm
Class
TU
TMM
Class
TU
TMM
Class
TU
TMM
T4
+55°C
+55°C
T4
+48°C
+110°C
T4
+40°C
+40°C
minimum -10°C
-10°C
T4
+55°C
+55°C
minimum -10°C
-10°C
-40°C
for all classes
for all classes
minimum -10°C
for all classes
2. Casing distance 57 mm
2. Casing distance 350 mm
Class
TU
TMM
Class
TU
TMM
T3
+52°C
+170°C
T3
+47°C
+170°C
T4
+53°C
+135°C
T4
+49°C
+135°C
-40°C
T4
+55°C
+55°C
minimum -10°C
-40°C
minimum -10°C
for all classes
for all classes
Bopp & Reuther Messtechnik GmbH
Page 13 of 23
Electronic Counter
7 Operating Overview
7.1 Taking in Operation
Place the oval wheel meter in operation according to the installation and maintenance instructions
for oval wheel meters.
The electronic meters are delivered, if nothing to the contrary is prescribed, according to operation
mode 2 with permanent flow display and require no adjustments. For the models AG4xEZD and
AG5xEZD, the NAMUR power supply must be connected.
7.2 Display and Control Elements
You can switch the information shown in the display using a switching magnet. To do this, just
touch the marked button with the enclosed switching magnet. Some switching functions react with
a slight time delay after you touch them with the switching magnet.
Ovalradzähler
Reset
Button
Reset
Identification Window
Liter
Display
Liter/min
Bopp & Reuther
Messtechnik
Decimal Point
Button for
Switching the
Display
Bopp & Reuther Messtechnik GmbH
Page 14 of 23
Electronic Counter
7.2.1 Operating When Set To Operation Mode 1
This operation type is identified with "Flow Display Disappears After 10 Sec." in the configuration
software.
The standard display shows the volume on the main meter. You can switch to other functions using the
switching magnet.
The display changes depending on how long you touch the "Display" button.
Each display can be identified with the symbols in the identification window.
U
•
If you touch the "Display" button for less than 1 second, the display switches temporarily to the
flow display. This is indicated by the symbol "u“ under the lower left digit position. After 10
seconds, the display returns automatically to the main meter.
•
If you touch the "Display" button between 1 to 2 seconds you move to the K factor and then to the
display test. Then you return automatically to the main meter.
•
If the switching magnet touches "Display" for more than 2 seconds, the return flow quantity meter,
the malfunction quantity meter and the sum meter are shown. The display sequence ends with
the display of the main meter.
Touching the "Reset" button resets the main meter immediately and adds the number to the sum
meter. At the same time the return flow and malfunction quantity meters are reset to zero.
The following diagram shows the operational sequence for operation type 1
Main Meter (Volume)
Magnet
1 s < Touching < 2 s
Magnet
Touching > 2 s
K Factor
Return flow quantity meter
Further switching
automatically
after 10 s
Further switching
automatically after 10 s
Display test
Flow
Automatic return
after 10 s
Malfunction quantity meter
Automatic return
Magnet
Touching < 1 s
Further switching
automatically after 10 s
Totalizer
Automatic return
after 10 s
Bopp & Reuther Messtechnik GmbH
Page 15 of 23
Electronic Counter
7.2.2 Operating When Set To Operation Mode 2
This operation type is identified with "Flow Display Permanent" in the configuration software.
The user can decide if he wants the display to continuously show the volume of the main meter or the
flow. By using the switching magnet you can switch from the main meter to the other functions.
The display changes depending on how long you touch the "Display" button.
Each display can be identified with the symbols in the identification window.
U
•
If, starting from the main meter, you touch the "Display" button for less than 1 second, the display
switches to the flow display, until you touch the button again. This is indicated by the symbol "u“
under the lower left digit position. There is no automatic return to the main meter.
•
If, starting from the main meter, you touch the "Display" button between 1 to 2 seconds you move
to the K factor and then to the display test. Then there is an automatic switching to the flow
display. To return to the main meter, touch the "Display" button with the switching magnet for less
than 1 second.
•
If, in the main meter display, the switching magnet touches the "Display" button for more than 2
seconds, the return flow quantity meter, the malfunction quantity meter and the sum meter are
shown. The display sequence ends with the display of the flow.
Touching the "Reset" button resets the main meter immediately and adds the number to the sum
meter. At the same time the return flow and malfunction quantity meters are reset to zero.
The following diagram shows the operational sequence for operation mode 2
Main Meter
Magnet Touching < 1sek
Magnet Touching
Magnet
1 s < Touching < 2 s
K Factor
Further switching
automatically
Magnet
Touching > 2 s
Return flow quantity meter
Further switching
automatically after 10 s
Display test
Flow
Further switching
automatically after 10 s
Malfunction quantity meter
Further switching
automatically after 10 s
Totalizer
Further switching
automatically after 10 s
Bopp & Reuther Messtechnik GmbH
Page 16 of 23
Electronic Counter
7.2.3 Operating When Set To Operation Mode 3 (only for model AG5x...)
Requirements:
1. Both Reed sensors must be connected.
2. The flow display is "Permanent."
3. The functions "Forward and Return Identification" and "Reaction to Alarm and
Warning" are activated.
When meter runs backwards, the warning message "E13" appears.
Table: Return Identification
Function
Type of Return Identification
Description
Main meter
decrements
during return
Return meter is
decremented after
return
OFF
OFF
Malfunction:
When the meter runs backwards, the return quantity is
interpreted as a malfunction quantity. The return and the
malfunction quantity meters are incremented.
The warning message can only be deleted with the switching
magnet or through the serial interface with the SensorPort
program.
During the execution of the reset-function the sum counter
increases the amount of the main counter.
The return quantity goes lost.
ON
OFF
Normal Operation:
When the meter runs backwards, the return meter is
incremented and the main meter is decremented. There is
no effect on the malfunction quantity meter. When the meter
runs forwards after this, the warning message is deleted
automatically.
During execution of the reset-function the sum counter is
increased by the total of the main and return counters.
OFF
ON
Malfunction:
When the meter runs backwards, the return amount is
interpreted as a malfunction quantity. The return and
malfunction quantity meters are incremented.
The warning message can only be deleted with the switching
magnet or through the serial interface with the SensorPort
program.
If the warning “E13” is not deleted during following counter
advance the return counter is decremented and the main
counter stops until the return counter achieves “0”
During execution of the reset-function the sum counter is
increased by the difference of the main and return counters.
ON
ON
Normal Operation:
When the meter runs backwards, there is no effect on all
meters.
When afterwards the meter runs forwards, the warning
message is deleted automatically.
During the execution of the reset-function the sum counter is
increased by the amount of the main counter.
The remaining functions correspond to operation type 2.
Bopp & Reuther Messtechnik GmbH
Page 17 of 23
Electronic Counter
7.2.4 Changing the Operation Mode
When the meter is delivered, the operation type has already been set according to instructions from
the customer. If, however, a change of the operation type is desired, this can be done using the
SensorPort program in conjunction with the IF232 interface. See chapter 7.3.
7.3 Configuration of the Electronic Meter (Option)
There are two possibilities for configuring the instrument:
1. using the SensorPort operating program through the serial interface
2. using the operating module (BEZW)
7.3.1 SensorPort
You can operate the instrument with a PC or laptop and the SensorPort 2 operating software in
conjunction with the IF232 interface. The meter data can be read out and the configuration data
can be changed.
The general SensorPort functions are available in separate operating instructions.
The connection of the interface is done using the interface plug on the connection circuit board.
See the diagram in chapter 2, System Description.
U
The IF 232 interface may only be used in atmospheres, which are not explosion-endangered.
zum PC = To PC
Bopp & Reuther Messtechnik GmbH
Page 18 of 23
Electronic Counter
7.3.1.1 Record Card "General"
Name:
A short text with 16 characters can be entered or read by the user.
Tag number:
The TAG address (measurement point number) is displayed.
Material being measured:
Name of the material being measured
User message:
A short text with 32 characters can be entered or read by the user.
Type designation:
Type designation of the entire instrument
Manufacturer:
Bopp & Reuther Messtechnik GmbH
Year of manufacture:
Year of manufacture of the electronics
Instrument identification:
Serial number of instrument
Representative:
Customer:
Order number:
Number of the order, to which this meter is assigned
7.3.1.2 Record Card "Constants"
Sum meter:
The sum of the volumes measured is displayed, which were
accumulated before Reset function was activated the last time:
"Reset of all meters except for the sum meter" (N).
Main meter:
The volume is displayed, which was counted since the last
activation of the above Reset button. This quantity is transferred to
the sum meter when the reset button is touched.
Return quantity meter:
The volume of the return quantity is displayed, which was counted
since the last activation of the above reset function. The return
identification is only possible for the model AG5x..., when operation
type 3 is set.
Malfunction quantity meter:
The volume of the malfunction quantity is displayed, which was
counted since the last activation of the above reset function. The
malfunction quantity is counted after the occurrence of:
- an alarm message;
- the warning message E.13 during malfunction operation of the
return identification (operation type 3, only for model AG5x...);
- creeping motion amount suppression;
when the function "Reaction to Alarm and Warning" is activated.
Calibration factor in Pulses/L.
K Factor:
Flow attenuation factor:
The attenuation factor in seconds affects the display information
and the output.
Flow decimal points:
Number of digits after the decimal point in the flow display
Bopp & Reuther Messtechnik GmbH
Page 19 of 23
Electronic Counter
Quantity decimal points:
The number of digits after the decimal point in the quantity displays
The maximum number of digits after the decimal points is
dependent on the K factor setting:
"3"
if K Factor > 100
"2"
if 10 < K Factor ≤ 100"
"1"
if 1< K Factor ≤ 10
"0"
if K Factor ≤ 1
The following manufacturer's constants may only be changed by the customer upon instructions
from Bopp & Reuther. They serve for a later extension for the local Bureau of Standards.
IF5V:
Standard value = 5
IF5V-min:
Standard value = 2
K15:
Standard value = 15,000
Actual:
Standard value = 8
Actual min:
Standard value = 2
IR-max:
Standard value = 15
IR-min:
Standard value = 2
f-max:
Standard value = 1000
f-min:
Standard value = 1
ZP:
Standard value = 2000
7.3.1.3 Record Card "Operation Type"
Flow display:
Permanent (operation type 2 or 3):The user can
set the display with the switching magnet to
continuously show the main meter or flow.
Disappears after 10 sec. (operation type 1):The
instrument shows the main meter continuously. A
display of the flow quantity is possible for a limited
time.
Suppression of creeping motion quantities:
Main meter is decremented during return:
Active: If the flow value is in the range of
Q < 0.3 * Qmin, the electronic meter shows this
with the display of the malfunction meter (when
showing volume), or the flow, which blinks every
second. The flow is counted by the malfunction
quantity meter.
Qmin[ℓ/min.] = f-min[Hz] / K Factor[Pulses/ℓ] * 60
See 7.2.3
Return meter is decremented during return:
See 7.2.3
Identification of time errors:
Active: Identification of errors E.08 (battery
replacement required) and E.18 (calibrating cycle
required), if the function "Reaction to Alarm and
Warning" is activated.
Forwards and return identification:
Active: The return identification (operation type 3)
is possible. Only for model AG5x....
Bopp & Reuther Messtechnik GmbH
Page 20 of 23
Electronic Counter
Output behavior:
Only for models with NAMUR power supply.
Warning: Depending on the position of the slide
switch, output 3 (terminals 5+6) or output 1
(terminals 1+2) is set as warning output. The
identification of warning messages is possible, if
the function "Reaction to Alarm and Warning" is
activated.
Scaled pulses: Depending on the position of the
slide switch, output 3 or output 1 is set to scaled
pulses.
Reaction to alarm and warning:
Active: Alarm and warning messages are shown
on the display. The corresponding alarm or
warning output is switched.
Pulse output scaled:
Only for models with NAMUR power supply.
Scaling factor of the scalable pulse output in
Pulses/L, which is dependent of the K factor
setting:
"10" or "100" if K Factor > 100
"1" or "10" if 10 < K Factor ≤ 100
"0.1" or "1" if 1 < K Factor ≤ 10
"0.01" or "0.1" if K Factor ≤ 1
Flow display:
Selection of the flow unit as L/min., L/h or m³/h.
The flow value is calculated automatically.
7.3.2 Operating Module (BEZW)
The electronic meter can also be configured within the Ex-area with four passive function keys.
The following data can be changed:
K Factor
The number of digits after the decimal point for the volume display (main meter, return meter and
malfunction quantity meter).
Resetting of the main meter with automatic transfer to the sum meter.
Programming
Mode
Programmiermodus
P
(+)
BEZW
EZd
Bopp & Reuther Messtechnik GmbH
Return
(-)
P
BEZW
EZ
Page 21 of 23
Electronic Counter
The connection of the operating module is done through the interface plug on the connection circuit
board. See diagram in chapter 2, System Description.
Setting K factor:
P-Key
Programming key. The meter switches to the programming mode. The third
identification field from the left is highlighted. The K factor and the decimal point
are shown.
+ (>) Key
The value of the digit, which is marked by the identification field, is increased by 1
when you push the key.
- (<) Key
The value of the digit, which is marked by the identification field, is reduced by 1
when you push the key.
Return key
When you push the return key, the next digit to the right of the digit, which is
presently marked, is marked by the identification field and can now be set.
When all digits have been set, the K factor is accepted by pushing the return key again and the
display then moves to the main meter. The last position of the identification field on the right is
highlighted. You can now set the decimal point for the volume display.
Setting the decimal point:
You can change the decimal point in the programming mode in the main meter using the + / - (> /
<) keys.
When you push the return key, the decimal point is accepted and the highlighting of the
identification field disappears and the display is shown in its basic status. At the same time, the
decimal points of the malfunction quantity meter and the return meter are changed along with the
main meter.
Setting Meter to Zero:
The simultaneous pushing of the + / - (> / <) keys in the basic status sets the display of the main
meter, the return meter and the malfunction quantity meter to zero. At the same time, the sum
meter, which cannot be erased, is increased internally by the amount in the main meter. This
insures that no volume amount is lost.
Bopp & Reuther Messtechnik GmbH
Page 22 of 23
Electronic Counter
8. Alarm and Warning Messages
Alarm and warning messages are shown, when the function "Reaction to Alarm and Warning" is
activated.
For meters with NAMUR power supply (AG4XEZD, AG5XEZD), if an alarm or warning message is shown
on the LCD display, the corresponding alarm or warning output is switched to "low." The warning output
(output 1 or output 3) is activated, if the output behavior is set to "Warning."
The alarm and warning messages can be tested using the SensorPort program. Use the menu
"Instrument/Self-test."
The alarm and warning messages can be reset with the switching magnet by touching the "Display"
button or through the serial interface with the SensorPort program. Use the menu "Instrument/Reset
Meter," function "Delete Alarm and Warning Messages."
After this, the display shows the return meter and the malfunction quantity meter for four times for two
seconds each.
The alarm and warning messages can also be deleted with the reset function "Reset all Meters" (R).
Caution! This reset process erases all meter data.
Alarm
No.
Warning No. Meaning
E.01
1 of 3 errors. An alarm error, which serves as a later extension for the local Bureau
of Standards.
E.02
Outage of the NAMUR power supply for models AG5xEZD, AG4xEZD or battery
power supply for the model AG5xEZ.
E.03
Number of return pulses > critical alarm value (IRMax)
E.04
Number of malfunction pulses > critical alarm value (IF5V)
E.05
Number of malfunction pulses > critical alarm value (ACTUAL). An alarm error,
which serves as a later extension for the local Bureau of Standards.
E.06
Q > 1.1 * Qmax
Qmax is calculated with the following formula:
Qmax[L/min.] = f-max[Hz] / K-Factor[Pulses/L] * 60
E.07
Memory error EEPROM (error when reading K factor)
E.08
Battery replacement is required. A time error, which is dependent on the service life
of the battery.
E.11
2 of 3 errors. A warning message, which serves as a later extension for the local
Bureau of Standards.
E.12
Run reset sequence (hardware reset). Only for models AG5xEZD, AG4xEZD.
E.13
Number or return pulses > critical warning value (IRMin)
E.14
Number or malfunction pulses > critical warning value (IF5VMin)
E.15
Number of malfunction pulses > critical warning value. The warning value, which
serves as a later extension for the local Bureau of Standards. (ACTUALMin)
E.16
Qmax < Q < 1.1 * Qmax
E.17
Error when running reset.
A warning message, which appears after the reset sequence if there is an outage of
the battery power supply, only for the model AG5xEZ, or a power supply
interruption. Display reset, display test and resetting of all meters except for the
sum meter, without transferring the main meter data to the sum meter.
E.18
Calibration cycle required. Time error warning, which serves as a later extension for
the local Bureau of Standards.
Bopp & Reuther Messtechnik GmbH
Page 23 of 23

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Related manuals

advertisement