Affinity HIT440 Module


Add to my manuals
2 Pages

advertisement

Affinity HIT440 Module | Manualzz

Affi nity HIT440 Module

- Effi cient Hearing Aid Testing

Hearing instrument testing

With HIT440 Hearing Instrument Testing module the

Affi nity takes hearing aid evaluation to a whole new level in its ease of use and fl exibility.

Directional microphones

Directional hearing aid microphones can be tested in the integral test box and all tests can be directly compared to previous measurements made on the same hearing aid.

All client data can be saved directly to NOAH or

Interacoustics’ own IABaseII database during the hearing aid fi tting. You can also choose not to use a database for quick and easy hearing aid evaluations. While preloaded standard ANSI and IEC protocols including coil measurements can get you up and running quickly you can also create your own protocols or import specialized tests such as those that are developed specifi cally by hearing aid manufacturers.

Easy reporting

You can print a customized report at the end of the session or gain immediate access to the data from previous test sessions. You no longer have to print the data and paste it in a fi le – now you can save it and verify hearing aid functionality over time, even compare tests right on your PC screen.

Other benefi ts of the HIT440 module:

• Easy upgrade with new tests

• Tests digital hearing aids

• Report page for session notes

Fit for the future

The HIT440 module is designed to test traditional hearing aids as well as digital hearing aids. Special stimuli and test procedures are available and include

ICRA noise and real speech. Most importantly is the open architecture of the Affi nity software platform. As new technology is developed future hearing aids will require special stimuli that are not known today. While you may have had to buy a new hearing aid testing device to keep up with hearing aid technology in the past, the Affi nity has the ability to simply load new test stimuli and procedures on its platform.

leading diagnostic solutions

Affi nity HIT440 Module

- Effi cient Hearing Aid Testing

General Technical Specifi cations

Standards:

Safety: EN 60601-1, Class I, Type B,

UL 2601-1, CAN/CSA-C22.2 No. 601.1-

M90, EN 60601-1-1. EMC: EN 60601-

1-2.

Computer Communication:

Built-in USB1.1 computer interface.

Optical isolation available.

Medical CE-mark:

The CE-mark indicates that Interacoustics

A/S meets the requirements of the Annex

II of the Medical Device Directive 93/42/

EEC. Approval of the quality system is made by TÜV – identifi cation no. 0123.

PC Minimum Requirements:

Processor/clock speed: 800 MHz Pentium 3 or better. System RAM: Minimum 128 MB

Screen Resolution: Minimum 800 x 600

Harddisk space: Minimum 500 MB free disk space. Operating System: Windows 98

SE/2000/ XP/ME.

Construction: Painted metal cabinet bottom and moulded plastic top.

Weight: 5.5 kg / 12.1 lbs.

Dimensions: (LxWxH): 42x38x15 cm/

16.5x15x5.5 inches.

Technical Specifi cations - Affi nity HIT440 Module

Hearing Aid Analyzer: EN 60118-0,

EN 60118-7, ANSI S3.22.

Frequency Range: 100-10000Hz.

Intensity Accuracy: ± 1.5 dB

Stimulus Distortion: Less than 1 % THD.

Coupler Microphone Ranges: 40-145dB.

PreProgrammed Protocols:

HIT module comes with a set of Test

Protocols loaded. Additional Test

Protocols can be designed by user, or easily imported into the system.

Frequency Resolution:

1/3, 1/6, 1/12 and 1/24 octave or 1024 point FFT.

FF Loudspeaker Output: Max 6 W into

8ohms, max 10 W into 4 ohms.

Frequency Accuracy: ± 1%.

Telecoil drive in Test box: 10 – 100 mA/ meter.

Stimulus Signal: Warble Tone, Pure Tone,

Random noise, Pseudo random noise, Band limited white noise, Chirp, ICRA, Real

Speech, any other sound fi le (automatic calibration available).

Battery Simulator: Standard types are selectable, Custom types within 1.1 – 1.6 V,

0 – 25 Ohm range.

Sweep Speed: 1.5 - 12 sec.

Test box:

Built in test box holds telecoil drive as well as special dual speaker set for checking directional microphone function.

Stimulation Intensity Range: 40-100 dB

SPL in 1 dB steps.

Noah: 3.0.

Included Parts:

Affi nity HIT440 CD

2CC coupler with microphone and adaptors for

ITE, BTE and Body Style HA

Coupler seal wax

Reference Microphone

Standard USB Cable

Power Cable 120 or 230V

Operation and Multilingual CE Manual

Optional Parts:

Battery Adapters BAA675, BAA13, BAA312,

BAA10, BAA5

Couplers 1.2CC and 0.6CC: ITE, BTE, Body

Ear Simulator

TBS25M External Test Chamber incl. cables.

ACC60 Affi nity carrying Case

Calibration Adaptor

Sales and Service in your area: Interacoustics A/S

Phone: +45 6371 3555

Fax: +45 6371 3522

E-mail: [email protected]

Web: www.interacoustics.com

Mail: Interacoustics A/S

DK-5610 Assens, Denmark

Available test box tests:

IEC118, ANSI 3.22, Custom:

OSPL90

Full On Gain

Input/Output

Attack/Recovery Time

Reference Test Gain

Frequency Response

Equivalent Input Noise

Harmonic Distortion

Intermodulation Distortion

Battery Current Drain/Battery Life Time

Microphone Directionality

Coil Frequency Response

Coil Harmonic Distortion

Coil Full-On Gain Response

advertisement