CardioSoft Diagnostic System Resting ECG

CardioSoft Diagnostic System Resting ECG
GE Healthcare
CardioSoft Diagnostic System
Resting ECG application
Clinically connected. Simply smart.
The CardioSoft* Diagnostic System Resting ECG application transforms your
physician office PC into a resting ECG device. The Cardiac Acquisition Module
(CAM-14) with the CAM USB A/T interface is easily connected from the patient
to your PC’s standard USB port. The CAM-14 provides high-quality ECG
presentation for accurate cardiac assessment of your patients. Its
outstanding digital signal processing reduces artifact for clearer tracings.
• Comprehensive GE Marquette* 12SL* ECG analysis program including
ACI-TIPI, gender-specific criteria, 12SL with Right Ventricular Involvement
measurement and serial presentation of ECGs; 12SL is one of the most
validated ECG analysis programs in the industry and has been cited in
over 150 independent publications
• Multiple configurable report formats are available
• Full disclosure data can be stored in the CardioSoft Diagnostic System
database and easily accessed remotely via CS Web; robust editing
functions make report generation simple and fast
• Seamless connectivity with other GE resting ECG devices such
as MAC* 800, MAC 1200, MAC 1600, MAC 3500, MAC 5500 and
with EMR** systems
• DICOM formatted resting ECG report generated by CardioSoft Diagnostic
System is automatically exported to PACS workstations, making physician
over-read fast and efficient while speeding up patient throughput
• Seamless connectivity to GE MUSE* Cardiology Information
System for even more workflow efficiencies
• Multiple networking options are available (consult CardioSoft
Diagnostic System Networking product sheet)
** Check EMR vendors’ compatibility with your GE representative.
Specifications
Signal processing
ST measurements
ST amplitudes, slope
Signal processing technique
Incremental median updating
Baseline correction
Cubic Spline
ECG output
Real-time ECG/QRS beep/TTL synchronization output
Heart rate
Automatic arrhythmia detection, documentation and annotation
Full disclosure ECG
Beat-to-beat ECG record and event review
Reanalysis
Re-analyze after manual correction of median beats and measurements
(HEART Algorithm)
ECG program
(Optional) 12SL adult and pediatric ECG analysis program
Additional ECG function
Vectorcardiography
Technical specifications
Communications/storage
MUSE systems compatible via diskette; network (optional)
MUSE Web compatible for retrieval view and printing of MUSE
system data
CS Web compatible for report viewing
PDF export of final reports (auto export and custom file name)
Microsoft® Word export of configured reports
XML or Excel® export of specified data
EMR connectivity
EMR Gateway
Data acquisition (via CAM-14)
Technology
Active, “Type BF” floating isolated powered 14-channel
acquisition module with built-in lead-fail detection and
lead prep impedance measurement
Sampling rate
4,000 samples per second, per lead across 15 leads
Dynamic range
320 mV, ±10 mV signal superimposed on ±150 mV DC offset
Resolution
4.88 µV/LSB @ 500 Hz
Noise
<15 µV peak-to-peak noise over 0.01 to 150 Hz
(-3dB) bandwidth
ECG analysis frequency
500 Hz
High pass filter
0.01 (or 0.05 Hz, special use) with DC offset control
Low pass filter
20, 40, 100, 150 Hz (selectable)
Line filter
50.0 or 60.0 Hz notch filter (selectable)
Baseline correction
Cubic Spline algorithm
Common mode rejection
Measured: 100dB, calculated: >140 dB (123 dB with
AC filter disabled)
Impedance
>10 M Ohms @ 10 Hz, defibrillator protected
Patient leakage
<10 µA
Pace detect
Orthogonal LA, LL and V6; 750 µV @ 50 µs
Display type
User provided
Monitored leads
12, 15
Displayed leads
Number on screen 3, 6, 12 or 15
Display format
4 x 2.5, 4 x 2.5 + 1 rhythm, 2 x 6, 6 rhythm, 3 rhythm
Display speeds
25, 50 mm/s
Display sensitivity/gain
2.5, 5, 10, 20, 40 mm/mV
Computer specifications
Microprocessor
Minimum: Pentium 4 Class Processor 1.6 GHz (Windows XP Professional);
Pentium 4 Class Processor 2 GHz (Windows 7 Professional)
RAM
Minimum: 512 MB (Windows XP Professional); 1 GB (Windows 7 Professional)
Hard drive
Minimum: 20 GB (depending on number of tests to save); 4 GB of free memory
SW installation
CD-ROM drive
Pointer
Mouse
Graphics adapter
Minimum: SVGA 1024 x 768
Recommended: SXGA 1280 x 1024
Interfaces
Minimum: 2 USB ports, CD-RW, SD card, network interface card (recommended),
Serial RS232 for each device using this interface type
Operating system
Windows XP/Professional SP3; Windows 7 Professional (32 bit and 64 bit,
with SP1);
Printer
HP P3015dn (printer not available for purchase from GE Healthcare in all regions. Confirm availability with your local GE representative)
Additional software for export functionality Microsoft® Word and Excel® (optional)
Networking LAN
Wireless: 802.11G (optional)
TCP/IP interface
Citrix®
Citrix application software not sold, installed or supported by GE Healthcare
GE Healthcare
P.O. Box 900, FIN-00031 GE, Finland
GE Direct United Kingdom: +44 (0)800 0329201
www.gehealthcare.com
©2012 General Electric Company – All rights reserved.
General Electric Company reserves the right to make
changes in specifications and features shown herein,
or discontinue the product described at any time
without notice or obligation.
*GE, GE Monogram, CardioSoft, CASE, Marquette, 12SL,
MAC and MUSE are trademarks of General Electric
Company.
GE Healthcare, a division of General Electric Company.
Citrix is a trademark of Citrix Systems, Inc.
HP LaserJet is a trademark of Hewlett-Packard
Company.
Microsoft, Excel, Word, Windows, and Windows Server
are trademarks of Microsoft Corporation
Pentium is a trademark of Intel Corporation.
EMEA DOC1047262 5/12
(EMEA English DOC1047262)
Was this manual useful for you? yes no