CA-210 Color Analyzer
CA-210 Color Analyzer
PRODUCT SPECS:
What are the specs for the product?
Receptor
Silicon photocell
Measurement area
Φ27mm
Acceptance angle
±2.5°; Pointing function: By LED
Measurement distance
30±10mm
Display range
Luminance
0.01 to 1000cd/m2
Chromaticity
Displayed in 4- or 3-digit value (selectable)
Measurement
0.10 to 1000cd/m2
Luminance
range
Accuracy (for
white) *1
Repeatability
±2%, ±1 digit of reading (Temperature: 23°C ±2°C; Relative
humidity:(40±10)%)
0.10 to 0.99cd/m2: 0.2%+1digit; 1.0 to 1000cd/m2: 0.1%+1digit
(2σ) *1
Chromaticity
Measurement
0.10 to 1000cd/m2
range
Accuracy *1
0.10 to 4.99cd/m2: ±0.008 for white5.00 to 39.99cd/m2: ±0.005 for
white40.00 to 1000 cd/m2: ±0.003 for white160 cd/m2: ±0.002 for white;
± 0.004 for monochrome *2
Repeatability
0.10 to 0.19cd/m2: 0.0150.20 to 0.49cd/m2: 0.0080.50 to 1.99cd/m2:
0.0032.00 to 1000 cd/m2: 0.001
(2σ) *1
Measurement
xyLv*3
0.10 to 1.99 cd/m2: 5 (4.5) 4 cd/m2 or greater: 20 (17)
speed
(meas./sec.)
Page
1
of
3
CA-210 Color Analyzer
(meas./sec.)
Display
Digital
xyLv, XYZ, TΔuvLv, u'v'Lv, RGB analyze
Analog
ΔxΔyΔLv, R/G B/G ΔG, ΔR B/R G/R
LCD
16 characters × 2 lines (with backlight)
SYNC mode
NTSC, PAL, EXT, UNIV, INT
Measurement subject
Vertical synchronous frequency: 40 to 200Hz
Memory channels
100
Analyzer function
Standard function
Interface
USB (Rev. 1.1) or RS-232c (38,400 bps maximum)
Multi-point measurement Max.
5 points (using optional 4-Probe Expansion Board CA-B14)
Software Solutions
SDK software (supplied as standard accessory)
Operating temperature/humidity
Temperature: 10 to 28°C; Relative humidity 70% or less with no
condensation. Luminance change: ±2% ±1 digit of reading for white;
Chromaticity change: ±0.002 for white, ±0.006 for monochrome from
reading of Konica Minolta's standard LCD *1, 160.0 cd/m2 at 23°C
40%RH
range
Storage temperature/humidity
range
0 to 28°C: Relative humidity 70% or less with no condensation; 28 to
40°C: Relative humidity 40% or less with no condensation
Input voltage range
100 to 240V~50/60Hz, 50VA
Dimensions (W × H × D)
Main body: 340 × 127 × 216mmProbe: Φ49 × 204mm
Page
2
of
3
CA-210 Color Analyzer
Weight
Main body: 3.58kg; Probe: 520g
For more details, or for details regarding Small Universal Measuring Probe and LCD Flicker Measuring Probes,
please request a catalog from your sales represenatative.
*1:
Chromaticity and luminance measured under Konica Minolta's conditions (Standard LCD at 6500K and 9300K)
*2:
The luminance for monochrome is measured when the reading of luminance for white is 160 cd/m2.
*3:
For single probe and USB; figures in parentheses for RS-232C with baud rate of 38,400 bps.
*
Specifications are subject to change without notice.
Page
3
of
3