INSTRUCTIONS FOR USE
Derived Result for Direct Bilirubin
DBIL
Direct Bilirubin
Rx ONLY
Intended Use
For in vitro diagnostic use only.
The derived test DBIL uses VITROS Chemistry Products BuBc Slides and VITROS Chemistry Products TBIL Slides on
VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System to estimate the concentration of direct bilirubin in serum or plasma.
Summary and Explanation of the Test
Direct bilirubin (DBIL) is used for evaluating liver and biliary diseases. An increased direct bilirubin occurs in both intrahepatic and extrahepatic biliary tract obstructions. Hepatocellular causes of elevation include hepatitis, cirrhosis, and advanced neoplastic states. Direct bilirubin is also increased in Dubin-Johnson syndrome and in Rotor syndrome.
Principles of the Procedure
The value for direct bilirubin, as reported by VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS
5600 Integrated System, includes all bilirubin fractions that react in the Jendrassik-Grof reaction without the addition of promoters. These include conjugated (mono- and diconjugates) and delta bilirubins.
VITROS Chemistry and Integrated Systems use two types of VITROS Slides (TBIL and BuBc) to determine the concentration of bilirubin components in human serum.
Total bilirubin (TBIL) is measured by the TBIL Slide as the diazo derivatives of unconjugated Bu, conjugated, and covalently protein-bound “delta” bilirubins. Conjugated and unconjugated bilirubins are measured simultaneously on the BuBc Slide using a dual-wavelength spectral measurement. Direct bilirubin values are calculated on the systems using these measured results and the following equation: direct bilirubin = TBIL – Bu
Test Type and Conditions
Derived
Reagents
VITROS TBIL and BuBc Slides are used to determine the concentration of direct bilirubin in human specimens.
IMPORTANT: For information on slide ingredients, labeling, handling, storage, stability, and preparation, refer to the Instructions for Use for the VITROS BuBc Slides and the
VITROS TBIL Slides.
Specimen Requirements
Refer to the Instructions for Use for the VITROS BuBc Slides and the VITROS TBIL Slides.
Testing Procedure
Refer to the Instructions for Use for the VITROS BuBc Slides and the VITROS TBIL Slides.
Calibration
Required Calibrators
VITROS Chemistry Products Calibrator Kit 4
Version 8.0
Pub. No. MP2-77_EN 1 of 4
DBIL
Direct Bilirubin
INSTRUCTIONS FOR USE
Quality Control
Calibrator Preparation, Handling, and Storage
Refer to the Instructions for Use for VITROS Calibrator Kit 4.
Calibration Procedure
Refer to the operating instructions for your system.
When to Calibrate
Refer to the Instructions for Use for the VITROS BuBc Slides and the VITROS TBIL Slides.
Calculations
Refer to the Instructions for Use for the VITROS BuBc Slides and the VITROS TBIL Slides.
Traceability of Calibration
Refer to the Instructions for Use for the VITROS BuBc Slides and the VITROS TBIL Slides.
Quality Control
Do not monitor the derived test.
Results
Reporting Units and Unit Conversion
The VITROS Chemistry and Integrated Systems may be programmed to report DBIL results in conventional, SI, and alternate units.
Conventional Units mg/dL
SI Units
µmol/L (mg/dL x 17.1)
Alternate Units mg/L (mg/dL x 10.0)
Limitations of the Procedure
Substances that interfere with the measurement of total bilirubin or unconjugated bilirubin, as indicated in the Instructions for Use for the VITROS BuBc Slides and the VITROS TBIL Slides, will affect the results for direct bilirubin.
Refer to the Instructions for Use for the VITROS BuBc Slides and the VITROS TBIL Slides.
Expected Values
Reference Interval
This reference interval is the central 95% of results from an internal study of 110 apparently healthy adults from a working population (85 females and 25 males).
Note: Do not use this derived test for specimens from neonatal patients less than 14 days old. Refer to the TBIL Test Methodology.
Conventional Units
(mg/dL)
0.0–0.4
SI Units
(μmol/L)
0–7
Alternate Units
(mg/L)
0–4
Each laboratory should confirm the validity of these intervals for the population it serves.
Performance Characteristics
Refer to the Instructions for Use for the VITROS BuBc Slides and the VITROS TBIL Slides.
2 of 4 Pub. No. MP2-77_EN Version 8.0
INSTRUCTIONS FOR USE
Glossary of Symbols
Glossary of Symbols
DBIL
Direct Bilirubin
Revision History
Date of Revision
2015-10-12
2014-09-05
2012-02-28
2010-11-01
2008-09-18
2004-09-13
2003-06-30
Version
8.0
7.0
6.0
5.0
4.0
3.0
2.0
Description of Technical Changes*
• Prescription Use statement added.
• Updated EC Representative address.
• Added USA to legal manufacture address
Glossary of Symbols: added Date of Manufacture
Glossary of Symbols: updated
Added information for the VITROS 4600 Chemistry System
• Added information for the VITROS 5600 Integrated System
• Glossary of Symbols – Updated
• Minor wording and formatting changes
• Calibration Procedure – updated operator’s manual to operating instructions
• Glossary of Symbols – updated data
• New organization and sections consistent with IVD Directive
• Summary and Explanation of the Test – corrected wording
• Principles of the Procedure – corrected wording
• Calibration – updated
• Quality Control – updated
• Performance Characteristics – changed to “Refer to the Instructions for Use for the VITROS BuBc Slides and the VITROS TBIL Slides.”
Version 8.0
Pub. No. MP2-77_EN 3 of 4
DBIL
Direct Bilirubin
INSTRUCTIONS FOR USE
Revision History
Date of Revision
2002APR19
Version
1.0 – English
Description of Technical Changes*
New format, technically equivalent to 11/96.
only
* The change bars indicate the position of a technical amendment to the text with respect to the previous version of the document.
When this Instructions For Use is replaced, sign and date below and retain as specified by local regulations or laboratory policies, as appropriate.
Signature
Obsolete Date
Ortho-Clinical Diagnostics
Felindre Meadows
Pencoed
Bridgend
CF35 5PZ
United Kingdom
Ortho-Clinical Diagnostics, Inc.
100 Indigo Creek Drive
Rochester, NY 14626
USA
VITROS is a registered trademark of Ortho-Clinical Diagnostics,
Inc.
© Ortho-Clinical Diagnostics, Inc., 2002-2015
4 of 4 Pub. No. MP2-77_EN Version 8.0