IPI Indicators
pressure
Industrial Pressure Calibrator
IPI Indicators
» Wide pressure range
IPI100A
IPI030C
IPI100C
IPI300C
IPI500C
IPI015G
IPI01KG
IPI02KG
IPI03KG
IPI05KG
IPI10KG
7 bar absolute / 100 psi
2 bar / 30 psi
7 bar / 100 psi
21 bar / 300 psi
35 bar / 500 psi
1 bar / 15 psi
70 bar / 1,000 psi
140 bar / 2,000 psi
200 bar / 3,000 psi
350 bar / 5,000 psi
700 bar / 10,000 psi
Intrins
ically
safe
» High accuracy
±0.05% of F.S. for positive pressure
Vacuum ranges to 500 psi (35 bar)
» High flexibility
Easily disassemble the test system and use
the indicator for other tasks
» True field indicator
Lightweight and portable with full temperature-compensation, long battery life and
large display for easy visibility
» Record min and max readings
Capture the min and max pressure readings
for safety valve applications
» Complete marine programme
Part of a complete programme of marine
approved temperature, pressure and signal
calibrators; including temperature sensors
» ATEX and CSA certification
The IPI Indicator is ATEX and CSA certified
and designed for use in potentially explosive
environments
JOFRA IPI indicators bring together the ease of an analog gauge with the
accuracy and easy-to-read display of a digital calibrator.
Use the JOFRA IPI in applications from safety valve checks to system
pressure verification.
Furthermore the IPI indicator is ATEX and CSA certified for use in potentially explosive environments such as oil refineries, chemical plants and
offshore platform, where there is a risk of inflammable gases.
This series is designed to meet your pressure measurement applications
and make the tasks easier. The IPI offers 18 different pressure units,
long battery life, high accuracy, and even serial communications. The
accuracy of the IPI rivals that of a pressure calibrator and is temperature
compensated for shop or in process applications. This versatile unit is
available as a stand alone indicator or in a complete test system.
You can perform a calibration locally without returning the IPI unit to the
manufacturer. All you need is an accurate pressure reference. If you do
require factory calibration, the pump and indicator are independent and
only the indicator needs to be returned.
The JOFRA IPI is available as an indicator or in one of 6 test ready systems that are complete and equipped to meet your pressure measuring
or testing needs.
ISO 9001 Manufacturer
Specification Sheet
SS-IPI
The JOFRA IPI digital pressure indicator takes the concept
of an analog test gauge, and brings it to a new level. The
IPI combines the accuracy of digital technology with the
simplicity of an analog gauge, and achieves performance,
ease-of-use, and a feature set unmatched in the pressure
measurement world. Setup of the IPI is fast and straightforward, through a menu-driven display, with minimal text,
and intuitive functions, that is simple enough to allow the
indicator to be used anywhere in the world, without the
need for multilingual displays.
Protective boot
Standard configuration IPI units and
those supplied with systems are
delivered with a protective boot.
Units with back mounting and the
24 VDC power supply are not delivered with the protective boot due
to design limitations.
Clear dual line display
ON / OFF
The IPI display does more than
just show a pressure reading. The
user can check temperature and
exact battery voltage through
the use of the keypad. The
large, 5 1/2 digit, 0.65 in
(1.65 cm) character display
is large and easy-to-read;
even from a distance.
Icons are also included
to indicate battery life,
engineering units, and
there is a bar graph that
shows the percentage of
scale reading. All of this
information can be read
in low light with the use of
the backlight.
The IPI is delivered with the auto-shutoff active and set
to 30 minutes: like many of our other instruments.
This feature can be reduced from 30 minutes
down to 1 minute in 1 minute increments. In
cases where a constant reading is necessary, this can also be turned completely
off to allow for continuous operation.
ZERO
Keep the high accuracy:
ZERO the instrument before
every test.
MAX / MIN
A MIN/MAX function reads
the maximum and minimum
pressure that have been automatically stored. Monitor a
safety valve or look for pressure
spikes in the process. The IPI
also have the ability to check the
minimum and maximum readings
during the ransient.
Pressure connections
1/4” NPT male lower connection is standard. A 1/4” BSP male
adapter is included. The IPI may be
ordered with a back mounting configuration.
This design has the 1/4 in. NPT male connection on the back of the case vice the lower manifold
connection. This allows for use in panel mounted applications or in applications where flat mounting is necessary
such as in an overhead. This configuration is not delivered
with a protective boot.
External power
Sometimes it is more convenient to take advantage of an
existing power source rather than using batteries. The IPI
may be ordered with a 24 VDC power supply option. The unit
has terminals to accept the positive and negative leads to
allow for easy connection to the power supply. These units
will not be supplied with batteries and batteries should not
be installed when the unit is connected to 24 VDC power.
2
www.jofra.com
ENTER
Save selections and turn
the backlight on and off.
CONFIG
This is the key to access all user-settable functions. The
CONFIG key is used in conjunction with the ▼ and ▲
functions on the ZERO and MAX/MIN keys to select and
change different functions.
You can set one of 20 engineering units, change the autoshutoff function setting, display the actual battery voltage,
display the actual temperature (in °C or °F), turn the dampening on or off, change the sample rate, and set the Tare
value. In addition to the 20 available engineering units on the
IPI, you can create your own unit to meet your measurement
need. For example, if you need to take a level measurement,
simply determine the pressure that equates to one unit of
that measurement and set that as your engineering unit. The
application may require gallons, feet, inches, meters, liters, or
similar; with the IPI you can take the measurement.
Serial communications
The IPI has a serial port that is accessible though the back
of the case. This feature uses standard ASCII commands to
allow for extraction of data from the IPI while taking readings.
This requires the optional serial cable.
Industrial Pressure Calibrator • IPI
Hazard location information / approvals
An explosive atmosphere is defined as a ”mixture with air,
under atmospheric conditions, of flammable substances in
the form of gases, vapours, mists or dusts in which, after
ignition has occurred, combustion spreads to the entire
unburned mixture”.
The standard in the European Union has been set
with the 9/94/EC Directive, commonly called ATEX
(“Atmosphères Explosibles,” French for explosive
atmospheres).
The JOFRA IPI Industrial Pressure Indicator is ATEX
approved by KEMA as complying with the Essential Health
and Safety Requirements related to the design and construction of equipment and protective systems intended
for use in potentially explosive atmospheres given in
Annex II in the directive, and with the following rating: II 3
G EEx nA IIB T6 (Ta=–10°C... +55°C).
The JOFRA IPI Industrial Pressure Indicator is
also certified by CSA as conforming to relevant
Canadian and USA standards with the following
rating: Class 1, Div. 2, Groups A-D.
See the definitions regarding hazardous locations in NFPA
70, Article 500 or CSA C22.1 Section 18. NFPA 70, Article
500 and CSA C22.1 Section 18.
Field recalibration
The IPI does not need to be returned to the factory for
calibration. If you have a reliable and accurate pressure
reference or a local laboratory, you may re-calibrate the
IPI locally. This feature is password protected.
Sampling rate
The IPI sampling rate is user selectable. If you want to
capture fast system transients, the IPI can take a reading
10 times per second. Conversely, you may want to conserve battery life and only need periodic samples. Then,
you can select 1 sample per every 2 seconds. This works
well for in process and panel mounted applications. You
can also choose the accepted instrument sampling rate of
3 samples per second for normal use or to allow for filtering, you can select 1 sample per second.
JOFRACAL CALIBRATION SOFTWARE
JOFRACAL calibration software ensures easy calibration
of RTD´s, thermocouples, transmitters, thermoswithes,
pressure gauges and pressure switches. JOFRACAL can
be used with all JOFRA calibration instruments. When
used with ASM-800 signal multi scanner, JOFRACAL can
perform a simultaneous semi automatic calibration on up
to 24 pressure and/or temperature devices under test in
any combination.
JOFRACAL software controls the complete calibration
procedure, stores the results and provides a calibration
audit trail through hard-copy certificates. All calibration
data are stored for each sensor to monitor drift and optimise recalibration intervals. A scheduler feature allows
planning of future calibrations.
Damping adjustment
The damping function can be turned on or off on the IPI.
This allows for readings to be integrated, which accounts for
momentary changes such as those from pulsing sources.
Tare
Beyond zeroing the IPI, you may have to account for
residual pressure. The Tare feature allows you to take care
of that error and prevent the manual calculation of the difference. This can be used in combination with the custom
engineering units to make the level measurements easier.
Temperature display and compensation
Because the IPI is designed for in-process tasks, temperature
compensation is included to make the job easier. This allows
the IPI to maintain accuracy over the measurement range. The
measured temperature may be checked from the keypad.
IPI • Industrial Pressure Calibrator
www.jofra.com
3
FUNCTIONAL SPECIFICATIONS
Display
Pressure; compound ranges
Display........................................................... Backlight, blue
Display resolution ........................... 5 digit floating decimal
Bar graph........................................ 20 segment, 0 to 100%
Display indicators...............................Engineering units icon
...................................................Low battery indication icon
.............................. Battery life indication utilizing bar graph
............................................. Measured temperature display
Display update.......................................... twice per second
bar......................................................-0.82 to 2, 7, 21 or 35
psi ...............................................-12 to 30, 100, 300 or 500
Pressure; gauge ranges
bar...................................... 0 to 1, 70, 140, 200, 350 or 700
psi............................. 0 to 12, 1,000, 2,000, 5,000 or 10,000
Pressure; absolute ranges
bara.........................................................................0.07 to 7
psia...........................................................................1 to 100
Engineering units
User defined.................................... One user-definable unit
User selectable......................................................... 20 units
(PSI, Bar, kg/cm2, inH2O (4 °C, 20 °C or 60 °F), ftH2O (4 °C, 20
°C or 60 °F), mmH2O (4 °C and 20 °C), cmH2O (4 °C and 20 °C),
mH2O (4 °C and 20 °C), KPa, mBAR, inHg, mmHg, Torr)
Not all units are availibale in all ranges.
Pressure accuracy
Pressure............................................................ ±0.05% F.S.
Full temperature compensation.......0 to 50°C / 32 to 122°F
Vacuum (100, 300, 500 psi indicator/
7, 21, 35 bar indicator)....................... ±0.25% F.S.
Vacuum (30 psi indicator/2 bar indicator)........... ±0.1% F.S.
F.S. (full scale) is the numerical value of the positive pressure
range. Accurracy includes hysteresis, nonlinearity, repeatability
and reference standard uncertainty, 1 year typical long-term stability, operated inside the rated temperature span and pressure
range.
Requiring frequently zeroing (Gauge/diff.) or entering of reference
pressure (Absolute).
Pressure accuracy ambient temp. (0 to 50°C / 32 to 122°F).
Power supply
Battery........................................................... (3) AA Alkaline
Battery life .......... 1500 operational hours without backlight
Battery life .......... 2000 operational hours without backlight
..............................................................at low sampling rate
Battery life ..................150 operational hours with backlight
Low battery indicator..............................................at 3 VDC
Optional input port...................... 24 VDC connection 1) 2) 3)
1) Units delivered with 24 VDC power supply connection are not supplied with batteries
2) Batteries should not be installed when using 24 VDC power supply as damage to the IPI may occur
3) Units with 24 VDC power supply connection cannot be ordered with the protective boot option
Instrument dimensions
Indicator LxWxH...... 125 x 111 x 38 mm / 4.9 x 4.4 x 1.5 in
Indicator weight (including battery).................1.6 lb / 0.7 kg
Input port...........................1/4” NPT Male lower connection
Optional input port......... 1/4” NPT Male back connection 1)
3) Units with back connection cannot be ordered with the protective boot option
Shipping dimensions
Serial communication interface
Indicator LxWxH.... 250 x 160 x 100 mm / 9.8 x 6.3 x 3.9 in
Indicator weight...............................................1.8 lb / 0.8 kg
Instrument case
Connector .......................................................... Stereo jack
Serial .......... 0-3 VDC, 9600 baud, 8 data, no parity, 1 stop
Protocol ...................................... ASCll command language
1) Does not apply if the serial connection is in use or if the
external 24 VDC power option is used.
Rating...........................................................NEMA 4/IP65 1)
Media compatibility
Approvals - IPI system
Liquids and gasses compatible with 316 stainless steel.
CE Conformity............... EN61326: 1998, EN60079-0: 2006, ................................................................. EN60079-15:2005
Environmetal
Storage temperature.....................-20 to 70°C / -4 to 158°F
Operating temperature................-10 to 55°C / -14 to 131°F
Pressure connection
All ranges ......................................................1/4” NPT male
The JOFRA IPI systems are type approved by Det Norske
Veritas. Find the certificate at www.jofra.com
Approval, Certificate no...........................................A-10549
Ex approvals - IPI indicator only
Adapters to 1/4” BSP male are included as standard
CSA.......................................Class 1, Div. 2, Groups A-D 1)
ATEX................. II 3 G EEx nA IIB T6 (Ta=–10°C... +55°C) 1)
Pressure overload
1) The 24 volt version is not CSA or ATEX approved
1 to 35 bar / 15 to 500 psi……...............................3X range
70 to 350 bar / 1,000 to 5,000 psi .........................2X range
700 bar / 10,000 psi ……………..........................1.5X range
Overload alarm ………………………………….OL on display
Overload alarm range ……………………….…… 1.2X range
111 mm
38 mm
36 mm
99,3 mm
125 mm
IPI
25,4 mm
4
www.jofra.com
Industrial Pressure Calibrator • IPI
JOFRA IPI PRESSURE RANGES
This table shows the resolutions that can be obtained by the IPI indicators throughout all engineering units.
IPI30C
Vacuum to 30 psi
Vacuum to 2 bar
IPI100A / IPI100C
Vacuum to 100 psi
Vacuum to 7 bar
IPI300C
Vacuum to 300 psi
Vacuum to 21 bar
IPI500C
Vacuum to 500 psi
Vacuum to 35 bar
psi
-12.000
30.000
-12.000
100.00
-12.000
300.00
-12.00
500.00
inH2O@4°C
-332.17
830.42
-332.2
2768.1
-332.2
8304.2
-332
13840
inH2O@20°C
-332.76
831.89
-332.7
2773.0
-332.7
8318.9
-333
13865
inH2O@60°F
-332.50
831.24
-332.5
2770.8
-332.5
8312.4
-332
13854
ftH2O@4°C
-27.681
69.202
-27.68
230.67
-27.68
692.02
-27.7
1153.4
ftH2O@20°C
-27.730
69.324
-27.73
231.08
-27.73
693.24
-27.7
1155.4
ftH2O@60°C
-27.708
69.270
-27.71
230.90
-27.71
692.70
-27.7
1154.5
inHg@0°C
-24.432
61.081
-24.43
203.60
-24.43
610.81
-24.4
1018.0
Torr
-620.6
1551.5
-620.6
5171.5
-620
15514
-621
25858
bar
-0.8300
2.0000
-0.8300
7.0000
-0.8300
21.000
-0.8300
35.000
mbar
-830.0
2000.0
-830.0
7000.0
-830.00
21000
-830
35000
kPa
-83.00
200.00
-83.00
700.00
-83.0
2100.0
-83.0
3500.0
kg/cm2
-0.8464
2.0394
-0.8464
7.1380
-0.846
21.414
-0.846
35.690
cmH2O@4°C
-846.4
2039.5
-846.4
7138.2
-846
21415
-846
35691
cmH2O@20°C
-847.9
2043.1
-847.9
7150.8
-847
21452
-848
35754
mH2O@4°C
-8.464
20.395
-8.464
71.382
-8.46
214.15
-8.46
356.91
mH2O@20°C
-8.479
20.431
-8.479
71.508
-8.48
214.52
-8.48
357.54
mmHg@0°C
-622.6
1500.1
-622.6
5250.4
-622
15751
-623
26252
Resolution obtained
by the IPI indicator
IPI015G
0 to 15 psi
0 to 1 bar
IPI01KG
0 to 1,000 psi
0 to 70 bar
IPI02KG
0 to 2,000 psi
0 to 140 bar
IPI03KG
0 to 3,000 psi
0 to 200 bar
IPI05KG
0 to 5,000 psi
0 to 350 bar
IPI10KG
0 to 10,000 psi
0 to 700 bar
psi
15.000
1000.0
2000.0
3000.0
5000.0
10000
inH2O@4°C
415.21
27681
55361
83042
N/A
N/A
inH2O@20°C
415.95
27730
55459
83189
N/A
N/A
inH2O@60°F
415.62
27708
55416
83124
N/A
N/A
ftH2O@4°C
34.601
2306.7
4613.5
6920.2
11534
23067
ftH2O@20°C
34.662
2310.8
4621.6
6932.4
11554
23108
ftH2O@60°C
34.635
2309.0
4618.0
6927.0
11545
23090
inHg@0°C
30.540
2036.0
4072.1
6108.1
10180
20360
Torr
775.73
51715
N/A
N/A
N/A
N/A
bar
1.0000
70.000
140.00
200.00
350.00
700.00
mbar
1000.0
70000
N/A
N/A
N/A
N/A
kPa
100.00
7000.0
14000
20000
35000
70000
kg/cm2
1.0197
71.380
142.76
203.94
356.90
713.80
cmH2O@4°C
1019.7
71382
N/A
N/A
N/A
N/A
cmH2O@20°C
1021.5
71508
N/A
N/A
N/A
N/A
mH2O@4°C
10.197
713.82
1427.6
2039.5
3569.1
7138.2
mH2O@20°C
10.215
715.08
1430.2
2043.1
3575.4
7150.8
mmHg@0°C
750.06
52504
N/A
N/A
N/A
N/A
Resolution obtained
by the IPI indicator
Imperial ranges
Metric ranges
Imperial ranges
Metric ranges
IPI • Industrial Pressure Calibrator
www.jofra.com
5
JOFRA IPI System A
•
•
•
•
0 to 1 bar (15 psi)
Vacuum to 2 bar (30 psi )
Vacuum to 7 bar (100 psi)
Vacuum to 21 bar (300 psi)
This system includes the
JOFRA IPI calibrator to­ge­ther
with one of the following pneumatic hand pumps: T-960 and
T-970, depending on pressure
range. System A is an easyto-use single hand operated
calibration system. No need for a flat surface to operate the
system - just hold it in your hand.
The System A comes in an aluminium carrying case with
cutouts for fittings, hose, and the complete assembled calibration unit - no time is required to assemble the unit every
time you need it. The calibrator (indicator) can at any time
be used separately for other pressure test tasks.
A special quick connector between the pump and the indicator makes it possible to seperate the system in seconds
and to swivel the indicator for easy viewing.
The JOFRA IPI delivery always includes a protective boot
for the indicator. All IPI indicators used in System A are
delivered calibrated in both pressure and vacuum.
Only the IPI indicator is approved for use in potentially explosive
atmospheres.
T-960 and T-970 pumps for system A
Model T-960 pneumatic pressure pump features an
extended range vernier adjustment for precisely controlling
pump pressure and comfortable pistol grip handle.
Model T-970 pneumatic pressure pump is similar to the T960, but generate pressure up to 40 bar (580 psi).
Pressure range, T-960...................... 0 to 2 bar / 0 to 30 psi
Pressure range, T-970...................0 to 40 bar / 0 to 580 psi
Type...................................................................... Pheumatic
Test medium..................................................................... Air
Operation................................................................... Scissor
“O”-rings...................................................................Buna-N
Wetted parts....................... Aluminium, brass, stainless steel,
...................................................................... nylon, Nylatron GS
Connection to test object....................... Hose 0.6 m / 24 in
........................with 1/4” BSP and NPT female terminations
Size..........................21.6 x 12.1 x 6.2 cm / 8.5 x 4.8 x2.4 in
Weight.............................................................1.4 kg / 3.0 lb
Ordering
Part No. Description
T-960 Pump: 0 to 2 bar (30 psi)
T-970 Pump: 0 to 40 bar (580 psi)
Included with System A delivery
• Standard delivery (see page 12)
• Hand pump T-960 or T-970
• 3/8” BSP female adapter with 1/4” BSP and NPT
female terminations
6
www.jofra.com
• 0.6 m / 24 in hose with 1/4” BSP and NPT female
terminations
• 1 roll of Teflon tape
• Protective carrying case
• Quick Connector between pump and calibrator
Accessories
Part No. Description
75P014 Service kit for T-960 pump
75P016 Service kit for T-970 pump
T-733-2 Pressure hose for T-960/970, 0.6 meter std.
T-733-3 Pressure hose for T-960/970, 0.5 meter
T-733-4 Pressure hose for T-960/970, 1.0 meter
T-733-5 Pressure hose for T-960/970, 2.0 meter
T-733-6 Pressure hose for T-960/970, 5.0 meter
65R191 Set of BSP female fittings, teflon tape, packings
* Fitting size : 1x1/8”, 1x3/8”, 1x1/2”
65R192 Set of NPT female fittings, teflon tape, packings
* Fitting size : 1x1/8”, 1x1/4”, 1x3/8” 1x1/2”
60R191 1 set of packing and seals for 65R191 and 65R192
123958 RS232 cable with stereo Jack connector, 2m/6ft
124716 4x 1,5 Volt rechargeable batteries
124718 Charger for rechargeable batteries - 115/230 VAC
SPK-HPC-005 Quick connector set for IPI system A / B
Shipping Information
Size....................570 x 180 x 350 mm / 22 x 7,1 x 13,8 inch
Weight.........................................................5,5 kg / 12.13 lb
Ordering information System A
Order no. Description
Pressure indicator
IPIIndustrial Pressure Indicator
Indicator pressure range
015G 0 to 1 bar (15 psi)
100A 0.070 to 7 bar absolut (100 psi)
030C -0.82 to 2 bar (30 psi)
100C -0.82 to 7 bar (100 psi)
300C -0.82 to 21 bar (300 psi)
500C -0.82 to 35 bar (500 psi)
NONE Pump system only (no calibrator)
Boot
B With boot (standard for indicator with battery power and lower mounting –
and on all system kits)
Power supply
XBatteries (standard)
Mounting
XLower mount (standard)
Base model number
AXX IPI System A (AXX pump T-960)
AHX IPI System A (AHX pump T-970)
Certification
G NIST traceable certificate (standard)
H Accredited certificate
(optional quotation basis)
IPI030CBXXAXXG Sample order number
JOFRA IPI 2 bar (30 psi) indicator with boot and in pressure System AXX with pneumatic pressure handpump and
NIST traceable calibration certificate.
Industrial Pressure Calibrator • IPI
JOFRA IPI System B
• Vacuum to 2 bar (30 psi )
• Vacuum to 40 bar (580 psi)
This system includes the dual
function pneumatic hand pump.
With this pump you can calibrate both vacuum and pressure applications. It takes just a
few seconds to switch between
vacuum and pressure, just push
one button (valve) on the handpump.
The System B comes in an aluminium carrying case with
cutouts for fittings, hose and the complete assembled calibration unit - no time required to assemble the unit every time.
The indicator and the pump are fitted with a special quick
connector that makes it possible to separate the system in
seconds without any tool. This quick connector also makes
it possible to use the assembled system with the indicator
twisted to any angle.
Accessories
Part No. Description
75P017 Service kit for T-975 pump
T-733-2 Pressure hose for T-975, 0.6 meter std.
T-733-3 Pressure hose for T-975, 0.5 meter
T-733-4 Pressure hose for T-975, 1.0 meter
T-733-5 Pressure hose for T-975, 2.0 meter
T-733-6 Pressure hose for T-975, 5.0 meter
65R191 Set of BSP female fittings, teflon tape, packings
* Fitting size : 1x1/8”, 1x3/8”, 1x1/2”
65R192 Set of NPT female fittings, teflon tape, packings
* Fitting size : 1x1/8”, 1x1/4”, 1x3/8” 1x1/2”
60R191 1 set of packing and seals for 65R191 and 65R192
123958 RS232 cable with stereo Jack connector, 2m/6ft
124716 4x 1,5 Volt rechargeable batteries
124718 Charger for rechargeable batteries - 115/230 VAC
SPK-HPC-005 Quick connector set for IPI system A / B
Shipping Information
Size....................570 x 180 x 350 mm / 22 x 7,1 x 13,8 inch
Weight.........................................................5,5 kg / 12.13 lb
Only the IPI indicator is approved for use in potentially explosive
atmospheres.
T-965 and T-975 hand pumps for system B
Model T-965 and T-975 pneumatic pressure pumps features both vacuum and pressure generation. Quick switch
(valve) makes it easy to switch from vacuum to pressure
measure­ments. A built-in release valve, vernier valve for
fine adjustment and dual pressure output enable safe and
simple operation. The unit offers both metric and imperial
threads on the reference connection and the hose.
Press. range, T-965.......-0.65 bar (-9.4 psi) to 2 bar (30 psi)
Press. range, T-975... -0.91bar (-13.2psi) to 35 bar (500 psi)
Type...................................................................... Pheumatic
Test medium..................................................................... Air
Operation................................................................... Scissor
“O”-rings...................................................................Buna-N
Wetted parts........Anodized aluminium, nickel-plated brass,
............................................................ stainless steel, Nylon
Connection to test object....................... Hose 0.6 m / 24 in
........................with 1/4” BSP and NPT female terminations
Size.........................21.6 x 12.1 x 6.2 cm / 8.5 x 4.8 x 2.4 in
Weight.............................................................1.4 kg / 3.0 lb
Ordering
Part No. Description
T-965 Pump: -0.65 to 2 bar / -19 inHg to 30 psi
T-975 Pump: -0.91 to 40 bar / -27 inHg to 580 psi
Included with System B delivery
• Standard delivery (see page 12)
• Hand pump T-965 or T-975 with a 0.6 m / 24 in hose
with 1/4” BSP and NPT female terminations
• 3/8” BSP female adapter with 1/4” BSP and NPT
female terminations
• 1 roll of Teflon tape
• Protective carrying case
• Quick Connector between pump and calibrator
IPI • Industrial Pressure Calibrator
Ordering information System B
Order No.Description
Pressure indicator
IPIIndustrial Pressure Indicator
Indicator pressure range
100A 0.070 to 7 bar absolute (100 psi)
030C -0.82 bar to 2 bar (30 psi)
100C -0.82 bar to 7 bar (100 psi)
300C -0.82 bar to 21 bar (300 psi)
500C -0.82 bar to 35 bar (500 psi)
NONE Pump system only (no calibrator)
Boot
B With boot (standard for indicator with battery power and lower mounting –
and on all system kits)
Power supply
X Batteries (standard)
Mounting
XLower mount (standard)
Base model number
BXX IPI System B (BXX pump T-965)
BHX IPI System B (BHX pump T-975)
Certificate
G NIST traceable certificate (standard)
H Accredited certificate
(optional quotation basis)
IPI300CBXXBXXG Sample order number
JOFRA IPI 21 bar (300 psi) indication with boot and in
pressure System BXX kit with pneumatic combination
handpump and NIST traceable calibration certificate.
www.jofra.com
7
JOFRA IPI System C
•
•
•
•
•
0
0
0
0
0
to
to
to
to
to
35 bar (500 psi)
70 bar (1,000 psi)
140 bar (2,000 psi)
200 bar (3,000 psi)
350 bar (5,000 psi)
This system consists of an
IPI indicator together with a
hydraulic, high pressure hand
pump T-620 or T-620H, which
features an oil reservoir to prime the system. System C is
an easy-to-use single-hand operated test system. No need
for a flat surface (table) to operate the system. The system
includes release valve and vernier valve for fine adjustment
to enable safe and simple operation of the pump.
The System C comes in a protective carrying case with
cutouts for fittings, hose, and the complete assembled
test unit - no time adder to assemble the unit every time.
Easy and fast connection between pump and indicator
makes it easy to use the pressure indicator separately for
other pressure test tasks.
Only the IPI indicator is approved for use in potentially explosive
atmospheres.
Included with System C delivery
• Standard delivery (see page 12)
• Hand pump T-620 or T-620H
• 0.6 m (24 in) hose with 1/4” NPT female and 1/4” BSP
terminations
• 1 roll of Teflon tape
• Protective carrying case
Accessories
Part No. Description
T-656 Service kit for T-620 and T-620H pumps
T-649 Pressure hose for T-620/T-620H pumps
65P175 1 m hose with 1/4” BSP female termination Max. pressure 350 bar (5,000 psi)
65P180 2 m hose with 1/4” BSP female termination
Max. pressure 350 bar (5,000 psi)
123958 RS232 cable with stereo Jack connector, 2m/6ft
124716 4x 1,5 Volt rechargeable batteries
124718 Charger for rechargeable batteries - 115/230 VAC
MGAAA/GL Oil for T-620 & T-1 pump AAA OIL in
1-GALLON CAN
MGAAA/QT Oil for T-620 & T-1 pump AAA OIL in 1-
QUART CAN
Shipping Information
Size....................560 x 188 x 350 mm / 22 x 7,4 x 13,8 inch
Weight.........................................................5,5 kg / 12.13 lb
AMETEK hydraulic handpump
This easy-to-use hand pump is specially designed for high
pressure applications. The pump has a built-in reservoir,
vent valve, and vernier valve for fine adjustment. The
pump has a dual pressure output - one for the test object
and one for the IPI indicator.
Pressure range, T-620..................... 0 to 200 bar / 3,000 psi
Pressure range, T-620H.................. 0 to 350 bar / 5,000 psi
Type........................................................................ Hydraulic
Test medium..................................................................... Oil
Reservoir capacity.......................................200 ml / 0.5 pint
Operation...............................................................Pistol-grip
“O”-rings...................................................................Buna-N
Wetted parts...........Aluminum, brass, stainless steel, Lexan
Connection to test object....................... Hose 0.6 m / 24 in
........................with 1/4” BSP and NPT female terminations
Size.........................21.6 x 12.1 x 6.2 cm / 8.5 x 4.8 x 2.4 in
Weight.............................................................1.4 kg / 3.0 lb
Ordering
Part No. Description
T-620 Pump: 0 to 200 bar (3,000 psi)
T-620H Pump: 0 to 350 bar (5,000 psi)
Ordering information IPI System C
Order No.Description
Pressure indicator
IPIIndustrial Pressure Indicator
Indicator pressure range
500C -0.82 to 35 bar (500 psi)
01KG 0 to 70 bar (1,000 psi)
02KG 0 to 140 bar (2,000 psi)
03KG 0 to 200 bar 3,000 psi)
05KG 0 to 350 bar 5,000 psi)
NONE Pump system only (no calibrator)
Boot
B With boot (standard for indicator with battery power and lower mounting –
and on all system kits)
Power supply
X Batteries (standard)
Mounting
XLower mount (standard)
Base model number
CXX IPI System C (CXX pump T-620)
CHX IPI System C (CHX pump T-620H)
Certificate
G NIST traceable certificate (standard)
H Accredited certificate
(optional quotation basis)
IPI01KGBXXCXXG Sample order number
JOFRA IPI 70 bar (1,000 psi) indicator with boot and in
pressure System CXX with hydraulic pressure handpump
and NIST traceable calibration certificate.
8
www.jofra.com
Industrial Pressure Calibrator • IPI
JOFRA IPI System D
•
•
•
•
•
0
0
0
0
0
to
to
to
to
to
35 bar (500 psi)
70 bar (1,000 psi)
140 bar (2,000 psi)
200 bar (3,000 psi)
350 bar (5,000 psi)
This system consists of an IPI
indicator together with a rugged, hydraulic, high pressure screwpump. System D is an
easy-to-use test system. Place the pump on a flat surface
and turn the handle to generate pressure. You can adjust
the indicator to fit the best viewing angle. The system
includes a manifold for connection of the test device, a
fine adjustment vernier valve (optional), and a reservoir for
extra oil or water (optional).
The System D comes in a protective carrying case with
cutouts for fittings and the complete assembled test unit
- no time adder to assemble the unit every time. Easy
and fast connection between pump and indicator makes
it easy to use the pressure calibrator separately for other
pressure test tasks.
Only the IPI indicator is approved for use in potentially explosive
atmospheres.
AMETEK screwpump
The 65-P016 and 65-P017 screwpumps are designed
for very easy generation of high pressure. The screwpumps has a 4 connection mani­fold. Fine adjust­ment,
vernier valve, liquid reservoir, and fittings are optional. The
AMETEK screwpump may be delivered for 2 different test
media: hydraulic oil or water.
Pressure range................................ 350 bar / 0 to 5,000 psi
Type........................................................................ Hydraulic
Test media.............................................. Hydraulic oil, water
Operation...........................................................Screw pump
“O”-rings...................................................................Buna-N
Wetted parts...................... Aluminum, brass, stainless steel
Connections to test objects................. 4 x 1/4” BSP female
Size.....................36.6 x 19.5 x 10.5 cm / 14.4 x 7.7 x 4.1 in
Weight.............................................................3.7 kg / 8.2 lb
Ordering
Part No. Description
65-P016 Screw pump oil: 0 to 350 bar (5000 psi) 1)
65-P017 Screw pump water: 0 to 350 bar (5000 psi) 1)
1) 250 ml (0.85 oz) reservoir
Included with System D delivery
• Standard delivery (see page 12)
• Screwpump 65-P016 for hydraulic oil or 65-P017 for
water
• 1 roll of Teflon tape
• 10 bonded seals
• 6 mm Allen key
• Protective carrying case
• Banjo Connector Set
IPI • Industrial Pressure Calibrator
Accessories
Part No. Description
60P013 Vernier valve fine adjustment for 65-P016/017
60P016 Packing set for screw pump oil part no. 65-P016
60P017 Packing set for screw pump water part no. 65-P017
65-R975 Reservoir set with seals, valve and bonded seal
60R135 Valve for oil reservoir
65P175 1 m. Pressure hose, (Max. 350 bar / 5,000 psi)
* 1/4” BSP female to 1/4” BSP male
65P180 2 m. Pressure hose, (Max. 350 bar / 5,000 psi)
* 1/4” BSP female to 1/4” BSP male
60I156 1.5 m. Pressure hose, (Max. 700 bar/10,000 psi)
* 1/4” BSP female to 1/4” BSP male
60I157 5 meter, Pressure hose, (Max. 700 bar / 10,000 psi)
* 1/4” BSP female to 1/4” BSP male
65N000 Extension tube, 120 mm (Max. 350 bar / 5.000 psi)
* 1/4” BSP female to 1/4” BSP male
123958 RS232 cable with stereo Jack connector, 2m/6ft
124716 4x 1,5 Volt rechargeable batteries
124718 Charger for rechargeable batteries - 115/230 VAC
50-REP 615 Quick connector set, female 1/4”BSP
to 1/4” BSP
SPK-HPC-004 Banjo Connector Set for pump
Shipping Information
Size....................730 x 220 x 430 mm / 28,7 x 8,7 x 17 inch
Weight............................................................12 kg / 26.5 lb
Ordering information IPI System D
Order No.Description
Pressure indicator
IPIIndustrial Pressure Indicator
Indicator pressure range
500C -0.82 to 35 bar (500 psi)
01KG 0 to 70 bar (1,000 psi)
02KG 0 to 140 bar (2,000 psi)
03KG 0 to 200 bar (3,000 psi)
05KG 0 to 350 bar (5,000 psi)
NONE Pump system only (no calibrator)
Boot
B With boot (standard for indicator with battery power and lower mounting –
and on all system kits)
Power supply
X Batteries (standard)
Mounting
XLower mount (standard)
Base model number
DOX IPI System D (DOX pump 65-P016)
DWX IPI System D (DWX pump 65-P017)
Certificate
G NIST traceable certificate (standard)
H Accredited certificate
(optional quotation basis)
IPI01KGBXXDOXG Sample order number
JOFRA IPI 70 bar (1,000 psi) indicator with boot and in
pressure System DOX with hydraulic oil screwpump and
NIST traceable calibration certificate.
www.jofra.com
9
JOFRA IPI System E
•
•
•
•
•
0
0
0
0
0
to
to
to
to
to
70 bar (1,000 psi)
140 bar (2,000 psi)
200 bar (3,000 psi)
350 bar (5,000 psi)
700 bar (10,000 psi)
This system consists of an IPI
indicator together with a rugged, hydraulic, high pressure pump.
System E is an easy-to-use test system. The hydraulic
pump makes it very easy to prime the system. You can
adjust the indicator to fit the best viewing angle. The system includes a manifold for connection of two test devices
and the fine adjustment vernier valve is standard.
Accessories
Part No. Description
60P013 Vernier valve for fine adjustment
60P014 Packing set for Jack pump part no. 65-P014
60I156 1.5 m. Pressure hose, (Max. 700 bar/10,000 psi)
* 1/4” BSP female to 1/4” BSP male
60I157 5 meter, Pressure hose, (Max. 700 bar / 10,000 psi)
* 1/4” BSP female to 1/4” BSP male
123958 RS232 cable with stereo Jack connector, 2m/6ft
124716 4x 1,5 Volt rechargeable batteries
124718 Charger for rechargeable batteries - 115/230 VAC
50-REP 615 Quick connector set, female 1/4”BSP
to 1/4” BSP
The System E comes in a protective carrying case with
cutouts for fittings and the complete assembled test unit
- no time adder to assemble the unit every time. Easy
and fast connection between pump and indicator makes
it easy to use the pressure calibrator separately for other
pressure test tasks.
Only the IPI indicator is approved for use in potentially explosive
atmospheres.
Shipping Information
Size....................730 x 220 x 430 mm / 28,7 x 8,7 x 17 inch
Weight............................................................13,5 kg / 30 lb
AMETEK high pressure hydraulic oil pump
The 65-P014 hydraulic pump is designed for high pressure applications. The pump has a 4 connection output
manifold. Fine adjustment vernier valve and fittings are
optional.
Ordering information IPI System E
Pressure range.............................. 0 to 700 bar / 10,000 psi
Type........................................................................ Hydraulic
Test media......................................................................... Oil
Operation............................................................. Jack pump
“O”-rings...................................................................Buna-N
Wetted parts...................... Aluminum, brass, stainless steel
Connections to test objects................. 4 x 1/4” BSP female
Size.....................36.5 x 20.0 x 14.2 cm / 14.4 x 7.9 x 5,6 in
Weight......................................................... 5.3 kg / 11.6 lbs
Indicator pressure range
01KG 0 to 70 bar (1,000 psi)
02KG 0 to 140 bar (2,000 psi)
03KG 0 to 200 bar (3,000 psi)
05KG 0 to 350 bar (5,000 psi)
10KG 0 to 700 bar (10,000 psi)
NONE Pump system only (no calibrator)
Ordering
Part No. Description
65-P014 Hydraulic pump for oil: 0 to 700 bar (10,000 psi) 1)
1) 450 ml (15.22 oz) reservoir
Included with System E delivery
•
•
•
•
•
•
•
10
Standard delivery (see page 12)
Hydraulic pump 65-P014 for hydraulic oil
Vernier valve
1 roll of Teflon tape
10 bonded seals
6 mm Allen key
Protective carrying case
www.jofra.com
Order No.
Description
Pressure indicator
IPIIndustrial Pressure Indicator
Boot
B With boot (standard for indicator with battery power and lower mounting –
and on all system kits)
Power supply
X Batteries (standard)
Mounting
X Lower mount (standard)
Base model number
EXX IPI System E (EXX pump 65-P014)
Certificate
G NIST traceable certificate (standard)
H Accredited certificate
(optional quotation basis)
IPI01KGBXXEXXG Sample order number
JOFRA IPI 70 bar (1,000 psi) indicator with boot an in
pressure system System EXX with high pressure pump for
hydraulic oil and NIST traceable calibration certificate.
Industrial Pressure Calibrator • IPI
JOFRA IPI System F
•
•
•
•
•
•
0
0
0
0
0
0
to
to
to
to
to
to
35 bar (500 psi)
70 bar (1,000 psi)
140 bar (2,000 psi)
200 bar (3,000 psi)
350 bar (5,000 psi)
700 bar (10,000 psi)
The System F hydraulic pump
makes it very easy to prime the system. It contains 75
in3/1.23 l of liquid. The system includes a manifold for connection of two test devices and the fine adjustment vernier
valve is standard. The System F comes in a protective carrying case with cutouts for fittings. Easy and fast connection between pump and calibrator makes it easy to use the
pressure calibrator separately for other pressure test tasks.
Only the IPI indicator is approved for use in potentially explosive
atmospheres.
AMETEK Type T hydraulic oil or water pump
The type T pump is designed for high pressure applications up to 15,000 psi (1,000 bar). The pump and system
may be ordered with either oil or a water/alcohol mixture
as the pressure medium. The pump is constructed of
300 series stainless steel and Monel allowing for the use
of other hydraulic media. There are three available seal
packages for the system: Buna-N, Viton®, and EPT.1) The
pump features a dual pressure output manifold, fine adjustment vernier valve, relief valve, and dual volume control
for rapid pressure increase at lower pressures and easier
pumping at higher pressures.
Pressure range........................... 0 to 1,000 bar / 15,000 psi
Type........................................................................ Hydraulic
Test media.............................................. Hydraulic oil, water
Reservoir capacity............................................1.23 l / 75 in3
Operation............................................................. Jack pump
“O”-rings.......... Buna-N (standard) or EPT / Viton (optional)
Wetted part........................................ Stainless steel, Monel
Connections to test objects.................... 1/4” and 1/2” BSP ............ terminations; 1/4” and 1/2” PT female terminations
Size................................. 79.4x22.9x50.8 cm / 31.3x9x20 in
Weight............................................................ 9.1 kg / 20 lbs
Ordering
Part No.Description Seals
T-1 0 to 1,000 bar (15,000 psi) water - Buna N 2)
T-1/VITON 0 to 1,000 bar (15,000 psi) water - Viton 3)
T-1/EPT 0 to 1,000 bar (15,000 psi) water - EPT 4)
T-1/OIL 0 to 1,000 bar (15,000 psi) oil
- Buna N 2)
T-1/OIL/VITON 0 to 1,000 bar (15,000 psi) oil
- Viton 3)
T-1/OIL/EPT 0 to 1,000 bar (15,000 psi) oil
- EPT 4)
1) Customer must verify compatibility of seals and pump materials with pressure medium.
Viton is a registered trademark. 2) Buna N are the standard ”O” ring material for the
hydraulic testers. They are used with water/alcohol or non corrosive oils. 3) Viton ”O” ring
material is used for all corrosive oils other than Skydrol. 4) EPT ”O” ring material is specifically for Skydrol Oil used for hydraulic systems in airplanes, it is very corrosive.
Included with System F delivery
•
•
•
•
Standard delivery (see page 12)
Type T Hydraulic pump
1 roll of Teflon tape
Protective carrying case
IPI • Industrial Pressure Calibrator
Accessories
Part No. Description
T-250 Service kit for T pump with Buna N O-rings
T-250 Service kit for T pump with VITON N O-rings
T-250 Service kit for T pump with EPT N O-rings
60I156 1.5 m. Pressure hose, (Max. 700 bar/10,000 psi)
* 1/4” BSP female to 1/4” BSP male
60I157 5 meter, Pressure hose, (Max. 700 bar / 10,000 psi)
* 1/4” BSP female to 1/4” BSP male
123958 RS232 cable with stereo Jack connector, 2m/6ft
124716 4x 1,5 Volt rechargeable batteries
124718 Charger for rechargeable batteries - 115/230 VAC
MGAAA/GL Oil for T-620 & T-1 pump AAA OIL in
1-GALLON CAN
MGAAA/QT Oil for T-620 & T-1 pump AAA OIL in 1-
QUART CAN
Shipping Information
Size....................730 x 220 x 430 mm / 28,7 x 8,7 x 17 inch
Weight...............................................................18 kg / 40 lb
Ordering information IPI System F
Order No.Description
Pressure indicator
IPIIndustrial Pressure Indicator
Indicator pressure range
500C -0.82 to 35 bar (500 psi)
01KG 0 to 70 bar (1,000 psi)
02KG 0 to 140 bar (2,000 psi)
03KG 0 to 200 bar (3,000 psi)
05KG 0 to 350 bar (5,000 psi)
10KG 0 to 700 bar (10,000 psi)
NONE Pump system only (no calibrator)
Boot
B With boot (standard for indicator with battery power and lower mounting –
and on all system kits)
Power supply
X Batteries (standard)
Mounting
X Lower mount (standard)
Base model number
FWB IPI System F (pump T-1)
FWV IPI System F (pump T-1/VITON)
FWE IPI System F (pump T-1/EPT)
FOB IPI System F (pump T-1/OIL)
FOV IPI System F (pump T-1/OIL/VITON)
FOE IPI System F (pump T-1/OIL/EPT)
Certificate
G NIST traceable certificate (standard)
H Accredited certificate
(optional quotation basis)
IPI10KGBXXFOEG Sample order number JOFRA IPI 700 bar (10,000 psi) indicator with boot and in
pressure System F with Type T pump for oil, EPT seals
and NIST traceable calibration certificate.
www.jofra.com
11
ORDERING INFORMATION - JOFRA IPI INDICATOR
Order No.Description
Pressure indicator
IPIIndustrial Pressure Indicator
Indicator pressure range
100A 0.07 to 7 bar absolute (100 psi)
030C -0.82 to 2 bar (30 psi)
100C -0.82 to 7 bar (100 psi)
300C -0.82 to 21 bar (300 psi)
500C -0.82 to 35 bar (500 psi)
015G 0 to 1 bar (15 psi)
01KG 0 to 70 bar (1,000 psi)
02KG 0 to 140 bar (2,000 psi)
03KG 0 to 200 bar (3,000 psi)
05KG 0 to 350 bar (5,000 psi)
10KG 0 to 700 bar 10,000 psi ()
Boot
B With boot (standard for indicator with battery power and lower mounting - and on all system kits)
X No boot (standard for indicators with back mounting and/or 24 VDC powers supply connection)
Power supply
X Batteries (standard)
E Connection for external 24 VDC power supply 1) Mounting
XLower mount (standard)
BBack mount
Base model number
IND IPI indicator with standard accessories for configuration
Certification
G NIST traceable certificate (standard)
H Accredited certificate (optional – on quotation basis)
IPI030CBXXINDG Sample order number
JOFRA IPI 2 bar (30 psi) indicator with boot, standard accessories, and
NIST traceable calibration certificate.
1) The 24 volt version is not CSA or ATEX approved
AMETEK Test & Calibration Instruments
A business unit of AMETEK Measurement & Calibration
Technologies Division offering the following industry
leading brands for test and calibration instrumentation.
JOFRA Calibration Instruments
Temperature Calibrators
Portable dry-block calibrators, precision thermometers
and liquid baths. Temperature ranges from
-90°C(-130°F) to 1205°C(2200°F). Temperature sensors
for industrial and marine use.
Pressure Calibrators
Convenient electronic systems ranging from -25 mbar to
1000 bar - fully temperature-compensated for problemfree and accurate field use.
Signal Instruments
Process signal measurement and simulation for easy
control loop calibration and measurement tasks.
M&G Pressure Testers & Pumps
Pneumatic floating-ball or hydraulic piston dead weight
testers with accuracies to 0.015% of reading. Pressure
generators delivering up to 1,000 bar.
Lloyd Instruments
Materials testing machines and software from Lloyd
Instruments guarantees expert materials testing
solutions. The comprehensive program also covers
Texture Analysers to perform rapid, general food testing
and detailed texture analysis on a diverse range of foods
and cosmetics.
Davenport Polymer Test Equipment
Allows measurement and characterization of moisturesensitive PET polymers and polymer density.
STANDARD DELIVERY
Chatillon Force Measurement
The hand held force gauges and motorized testers
have earned their reputation for quality, reliability and
accuracy and they represent the de facto standard for
force measurement.
• IPI indicator
• Calibration certification
performance traceable to NIST
• (3) AA batteries
Newage Testing Instruments
Hardness testers, durometers, optical systems and
software for data acquisition and analysis.
• Adapter to 1/4” BSP male
• User manual
• Protective boot
www.jofra.com
UK
AMETEK Calibration Instruments
Tel +44 (0)1243 833 302
jofra@ametek.co.uk
France
Denmark
Singapore
USA
China
AMETEK Denmark
Tel +45 4816 8000
jofra@ametek.com
AMETEK S.A.S.
Tel +33 (0)1 30 68 89 40
general.lloyd-instruments@ametek.fr
AMETEK Mansfield & Green
Tel +1 (800) 527 9999
cal.info@ametek.com
Germany
India
AMETEK GmbH
Tel +49 (0)2159 9136 510
info.mct-de@ametek.de
AMETEK Instruments India Pvt Ltd.
Tel +91 22 2836 4750
jofra@ametek.com
AMETEK Singapore Pte Ltd
Tel +65 6484 2388
jofra@ametek.com
AMETEK Inc. - Shanghai
Tel +86 21 5868 5111
AMETEK Inc. - Beijing
Tel +86 10 8526 2111
AMETEK Inc. - Guangzhou
Tel +86 20 8363 4768
jofra.sales@ametek.com.cn
Information in this document is subject to change without notice. ©2011, by AMETEK, Inc., www.ametek.com. All rights reserved.
12
www.jofra.com
Pub code SS-IPI Issue 1111
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project

Download PDF