ATX/ATYSeries

ATX/ATYSeries
C054-E066C
UniBloc Analytical Balances
ATX /ATY Series
Durable, high-performance aluminum alloy mass sensor
UniBloc
3 Benefits of UniBloc.
The ATX/ATY series incorporates a one-piece aluminum alloy mass sensor
technology (UniBloc), first introduced by Shimadzu for precision balances
in 1989. It excels in performance, and resists deterioration and damage
by ordinary impacts. The UniBloc’s compact, uniform structure replaces 70
parts found in a conventional electromagnetic balance sensor assembly
and ensures stable temperature characteristics, excellent response time
and stable corner-load performance.
The UniBloc design permits a consistency of production that assures
reliability and a long operational life.
Touch-key Calibration
Automated calibration can be started by pressing keys. (ATX series)
Also, your external calibration weights can be used for span calibration.
(All models)
2
Reliability
Stability
Response
Easy Setting Best fit to weighing application
Patented
Quickly adjust the desired ratio of stability and response for every application,
Indicator
even during measurement, with one-touch operation.
Responsiveness priority
Stability priority
kly
Quic
・When you want to do a quick
measurement
・When you want to raise the
operation efficiency
・When you want to measure or blend
the determination of liquid or
powder samples
Precisely!
・When the numerical value is unstable
・When using in the vibration
environment
・When the airflow exists constantly and
the numerical value fluctuates
frequently
Check the overview of AT on our website.
Check the easy setting video on YouTube.
http://www.shimadzu.com/an/balance/at.html
https://www.youtube.com/watch?v=9zVWxjx-J_g&list=
PLCPY11zjvhZPmTi_xW1oGJeSuI-YR6qaV&index=1
3
Calibration record
You can leave a record of execution of calibration. With serial
number and ID of balance.
Large pan size
This model has the largest weighing pan in the class
(91mm diameter).
Multiple weighing units
In addition to grams (g), weigh in ct, mg, oz, etc. or a custom
conversion units.
Expanded Piece Counting Function
Unit weights of up to 5 different samples can be easily entered,
stored and recalled for use.
Comparator Function
Compare samples to target values or pass/fail criteria and clearly
indicate the results.
% measurement
The weight of the sample is converted to a percentage of the
reference weight.
Power saving function
When weighing operation ends. Power automatically turns off
after a fixed or pre-set time.
WindowsDirect Communication Function
Send balance data to Excel or other Windows applications
without any data communication software installation required.
By combining standard AutoPrint functions with typical
spreadsheet functions, even difficult applications can be easily
automated.
*RS232C interface is needed.
Password lock
In order to ensure that the menu settings are not changed by
mistake, the person managing the Balance controls the
password and can prohibitmenu operation.
Formulation Mode
Convenient for making many measurements of minute samples
and seeking the total mass.
ATX/ATY Series
UniBloc Analytical Balances
4
Weighing results and analysis results can be stored in a database. Reports can be generated according to results in a prescribed format.
Analytical Balance
UHPLC
AT series
Nexera X2
Weighing results by Balance
Analysis results by LC
Database
Analysis results by LC
LabSolutions software can store
analysis results and weighing results
in a database at the same time.
Weighing results by Balance
* The optional software (LabSolutions
Balance) is required in order to connect
with LabSolutions.
Report example
Analysis results by LC
Weighing results
The report can be generated
automatically according to results.
ATX/ATY Series
UniBloc Analytical Balances
5
ATX/ATY Series
Dimensions
432
244
356
426
240
222
136
8
102
170
138
185
338
ø91
163
213
26
162
307
Specifications
Options
ATX/ATY series
Model
Capacity
ATX84
ATX124
ATX224
ATX324
ATY64
ATY124
ATY224
ATY324
82g
120g
220g
320g
62g
120g
220g
320g
Description
EP-100 Printer
Minimum Display
0.1mg
EP-110 Printer
Repeatability (Standard Deviation)
≤0.1mg
I/O-RS cable
Linearity
± 0.2mg
Protection Cover
Stabilisation Time*1
Operating Temperature and
Humidity Limits
5-40˚C 20-85%*2
Temperature Coefficient for
Sensitivity (10-30˚C)
± 2 ppm/ ˚C
Pan Size (mm) approx.
Power Requirement
Internal Calibration
USB conversion kit
ø91
213(W) × 356(D) × 338(H)
Main Body Dimensions (mm) approx.
Main Body Weight (kg) approx.
STABLO-AP Ionizer
Approx. 3.0 seconds
6.2
6.0
12V, 1A
-
*1 The Stabilisation time is a representative value
*2 No condensation
For Research Use Only. Not for use in diagnostic procedures.
This publication may contain references to products that are not available in your country. Please contact us to check the availability of these
products in your country.
Company names, products/service names and logos used in this publication are trademarks and trade names of Shimadzu Corporation, its
subsidiaries or its affiliates, whether or not they are used with trademark symbol “TM” or “®”.
Third-party trademarks and trade names may be used in this publication to refer to either the entities or their products/services, whether or not
they are used with trademark symbol “TM” or “®”.
Shimadzu disclaims any proprietary interest in trademarks and trade names other than its own.
www.shimadzu.com/an/
The contents of this publication are provided to you “as is” without warranty of any kind, and are subject to change without notice. Shimadzu
does not assume any responsibility or liability for any damage, whether direct or indirect, relating to the use of this publication.
© Shimadzu Corporation, 2017
First Edition: March 2010, Printed in Japan 3655-02715-30AIK