NETWORK ANALYZERS
NETWORK ANALYZERS
RF Network Analyzer, 30 kHz to 6 GHz
HP 8753D
30 kHz to 6 GHz frequency range
Integrated S-parameter test set
Integrated 1-Hz resolution synthesized source
Optional time-domain and swept-harmonic measurements
Up to 110 dB of dynamic range
Group delay and deviation from linear phase
Save/recall to built-in disk drive
Built-in accuracy enhancement
HP 8753D
HP 8753D RF Network Analyzer
The HP 8753D RF network analyzer provides affordable excellence in RF
network measurement for laboratory and production testing. It has an
integrated S-parameter test set for longer-lasting calibrations, exception-
al reliability, and improved resistance to electrostatic discharge. The
HP 8753D gives you a complete solution for characterizing active or pas-
sive networks, devices, or components from 30 kHz to 6 GHz—with a
cost savings over the previous model with a test set.
The network analyzer has two independent display channels for
simultaneous display of reflection and transmission, magnitude and
phase, or time-domain and swept-frequency measurements. Easy-to-use
softkeys let you access measurement functions quickly, and you can view
results in overlay or split-screen format on the crisp color display.
Maximum Versatility and Performance
An integrated, synthesized source provides > 100 mW of output power,
1 Hz frequency resolution, and linear, log, list, power, and CW sweep
types. Three tuned receivers allow independent power measurements or
simultaneous ratio measurements over a wide dynamic range of 105 dB
at 6 GHz (with Option 006 frequency extension) or 110 dB at 3 GHz (stan-
dard). The integrated test set allows you to measure transmission and
reflection characteristics of a device to 6 GHz without a frequency doubler.
TRL*/LRM*! calibration has been added for convenient, accurate
measurements in noncoaxial environments. A high-stability frequency
reference, Option 1D5, improves the frequency accuracy of measure-
ments on high-Q devices such as SAW resonators or dielectric resonance
filters. For configuration flexibility, Option 011 deletes the built-in test set
so that you can select your own. The HP 8753D Option 011 works with
the HP 85046A/B, 85047A, and other test sets.
Productivity Enhancements
A test-sequence function allows rapid, repeated execution of complex
tasks with a single keystroke. In sequencing mode, you make a measure-
ment once from the front panel, and the analyzer stores the keystrokes
so that no additional programming is required. You can also use a test
sequence to control external devices through the parallel port.
Other productivity enhancements include a built-in disk drive sup-
porting LIF and DOS formats, a faster CPU clock rate, non-volatile mem-
ory of 512 KB, serial and parallel interfaces, a DIN keyboard interface,
and a real-time clock for time-stamping of printouts and files. Limit test-
ing, arbitrary frequency testing, and marker tracking functions are
included. Segmented calibration and interpolative error correction allow
you to apply vector accuracy enhancement over a subset of the analyzer’s
calibrated frequency range.
Nonlinear Device Testing
For nonlinear device characterization, Option 002 adds harmonic mea-
surement capability. Swept second- and third-harmonic levels of an
amplifier can be displayed directly or in dBc. With the press of a button,
you can measure harmonics down to -40 dBc. Power meter calibration
provides leveled absolute power to devices that are sensitive to absolute
input or output levels. The HP 8753D automatically controls an HP 436A,
437B, or 438A power meter to set the power anywhere in the test config-
uration with power meter accuracy. The analyzer also performs mixer
tracking and conversion loss measurements. Both fixed and swept IF
measurements can be made.
Time-Domain Analysis
With Option 010, you can view responses in the time domain. The analyz-
er computes the inverse FFT of the frequency-domain data to display the
reflection or transmission coefficient versus time. Two time-domain
analysis modes enable you to view the step or impulse response of your
device, or to remove unwanted responses such as connector mismatch
using gating.
"TRL* and LRM* are three-sampler implementations of tha through-reflect-line and line-
reflect-match calibration techniques.
Hewlett-Packard’s fax service provides technical information 24 hours a day +800-800-5281
Specifications Summary
Test Set
Integrated S-parameter with complete forward and reverse mea-
surements in 50 Q (standard) or 75 Q (Option 075). External test sets
supported with Option 011.
Test Port Output
Frequency Characteristics
Range: 30 kHz to 3 GHz (std.); 30 kHz to 6 GHz (Option 006)
300 kHz to 3 GHz (Option 011) |
30 kHz to 6 GHz (Option 011, 006)
Resolution: 1 Hz
Accuracy: +10 ppm at 25 C+ 5" C
Output Characteristics
Power range: —85 to 10 dBm
Resolution: 0.05 dB
Sweep Range: 25 dB
Level Accuracy: +1.0 dB relative to 0 dBm output level
Level Linearity: (15 to +5 dBm} +0.2 dB typical 30 kHz to 300 kHz
(+5 to +10 dBm) +0.5 dB typical 30 kHz to 300 kHz
Impedance: 50 Q
2nd Harmonic: <-25 dBc at +10 dBm {16 MHz to 3 GHz)
3rd Harmonic: <-25 dBc at +10 dBm (16 MHz to 2 GHz)
Nonharmonic Spurious (typical)
Mixer-related: <-30 dBc at +10 dBm
Test Port Input Characteristics
Frequency Range: 30 kHz to 3 GHz (std.); 30 kHz to 6 GHz
(Option 006)
Average Noise Level
3 kHz BW: -82 dBm (< -3 GHz), -77 dBm (3 to 6 GHz)
10 Hz BW: -102 dBm (< —3 GHz), -97 dBm (3 to 6 GHz)
Maximum Input Level: +10 dBm
Damage Level: +26 dBm or 35 Vdc
Impedance: 50 Q (75 € with Option 075)
Harmonics: (Option 002)
2nd Harmonic: <-15 dBc at+8 dBm
3rd Harmonic: <-30 dBc at +8 dBm
Harmonic Measurement Accuracy (25 +5°C)
16 MHz to 3 GHz +1 dB
3 GHz to 6 GHz + 3 dB (with Option 006)
Harmonic Measurement Dynamic Range
—40 dBc (output=-10 dBm, input <-15 dBm)
Group Delay Characteristics
Range: 1/(2 x minimum aperture)
Aperture (selectable)
Maximum: 20% of frequency span
Minimum: (frequency span)/(no. of pts. —1)
Group Delay Accuracy (in seconds): + (phase accuracy in degrees)/
(360 x aperture in Hz)
Physical Characteristics
Size: + mm W x 222 mm H x 508 mm D (16.75 in x 8.75 in x
20.0 in
Weight: 34 kg (75 [b) net; 37 kg (82 Ib) shipping
Upgrade Kits
Upgrade kits retrofit the latest operating systems or add optional mea-
surement capability to existing HP 8753A/B/C/D network analyzers.
HP 11883A Harmonic Measurements Upgrade
This upgrade kit adds harmonic measurement capability (Option 002) to
an HP 8753C/D network analyzer. This kit includes installation at an HP
service center,
NETWORK ANALYZERS
HP 11884A 6 GHz Receiver Upgrade
This kit extends the operating frequency range of the HP 8753B/C
receiver from 3 GHz to 6 GHz. To make transmission/reflection mea-
surements above 3 GHz, the HP 85047A S-parameter test set is required.
This kit includes installation at an HP service center.
HP 11884B 6 GHz Upgrade for HP 8753D Standard
This kit extends the operating frequency range of the HP 8753D from
3 GHz to 6 GHz. No additional test set is needed. Includes installation at
an HP service center. Not compatible with Option 075 or Option 011.
HP 11884C 6 GHz Upgrade for HP 8753D Option 011
This kit extends the operating frequency range of the HP 8753D Option
011 from 3 GHz to 6 GHz. Includes installation at an HP service center.
Not compatible with Option 075.
HP 85019B Time-Domain Upgrade Kit
This upgrade kit adds time-domain analysis capability (Option 010) to an
existing HP 8753C/D network analyzer. This kit is user-installable.
Transmission/Reflection Test Sets
Transmission/reflection test sets provide the capability to simultaneo-
usly measure the reflection and transmission characteristics of two port
devices. The device must be physically turned around to measure its
reverse direction characteristics. These test sets are used with the
HP 8753A/B/C or the HP 8753D Option 011 only.
Specifications Summary
HP 85044A HP 850448
Impedance: 50 Q 752
Frequency Range: 300 kHz to 3 GHz 300 kHz to 2 GHz
Directivity:’ 35 dB to 1.3 GHz 35 dB to 1.3 GHz
30 dB to 3.0 GHz 30 dB to 2.0 GHz
Typical Tracking
Transmission Magnitude,
Phase:"*’
0.3 MHz to 2.0 MHz +1.5dB, +10" +1.5d8B,+10*
2.0 MHz to Fm. +1.5 dB, +10 +15 dB, +10°
Reflection Magnitude,
Phase:
0.3 MHz to 2.0 MHz +1.5 dB, +25 +1.5 dB, +25
2.0 MHz to Frax +1.5 dB, +10° +1.5 dB, +10
Effective Source Match *’
(test ports):
0.3 MHz to 2.0 MHz 14dB 14dB
2.0 MHz t0 1.3 GHz 20dB 17 dB
1.3 GHz to F max 16 dB 16 dB
RF Connectors
Test Port: Precision 7 mm 75 Q type-N
(female)
All Others: 50 Q type-N 50 Q type-N
(female) (female)
Physical Characteristics
Size: 191 mm W x 62 mm H x 204 mm D (7.5 in x 2.44 in x 8.0 in)
Weight: Net, 1.7 kg (3.8 Ib); shipping, 2.0 kg (4.4 Ib)
'Degrees, specified as deviation from linear phase.
Fmax is the upper frequency limit of the associated test set.
Can be improved through accuracy enhancement.
To have a Hewlett-Packard representative help you place an order or to get more information call 1-800-452-4844
NETWORK ANALYZERS
S-Parameter Test Sets
HP 8753D Series
S-Parameter Test Sets
The S-parameter test sets provide the capability to measure reflection
and transmission characteristics (including S-parameters) of two port
devices in either direction with a single connection. The test sets are con-
trolled from the analyzer and include programmable step attenuators
These test sets are used with the HP 8753A/B/C or the HP 8753D
Option 011 only
HP 85046A/B S-Parameter Test Sets
The HP 85046A/B test sets provide the capability to simultaneously mea-
sure the transmission and reflection characteristics of 50 and 75 ohm
devices, respectively
Specifications Summary
HP 85046A HP 850468
impedance 50Q 750
Fraquency Range 300 kHz to 3 GHz 300 kHz to 2 GHz
Directivity 35 dB to 1.3 GHz 35 dB to 1.3 GHz
30 dB to 3.0 GHz 30 dB to 2.0 GHz
Typical Tracking
Transmission Magnitude,
Phase:
0.3 MHz to 2.0 MHz
2.0 MHz to Fox
Reflection Magnitude
ase
0.3 MHz to 2.0 MHz
2.0 MHz to Fm.
Effective Source Match
{test ports)
0.3 MHz to 2.0 MHz 14dB 14dB
2.0 MHz to 1.3 GHz 20 dB 17 dB
2.0 MHz to Fm. 16 dB 16 dB
RF Connectors
Test Ports
+1.5 dB, +20
+1.5 dB, +10
+1.5 dB, +20
+1.5 dB, +10
+1.5 dB, +25
+1.5 dB, +10
+1.5 dB, +25
+1.5 dB, +10
Precision 7 mm 75 Q type-N
(female)
50 Q Type-N
(female)
All Others 50 Q2 Type-N
(female)
Includes: Four 190-mm (7.5 in) cables with Type-N (male) connectors for
connection to the HP 8753. One HP 8753 test set interconnect cable
Physical Characteristics
Size: 426 mm W x 90 mm H x 508 mm D (16.75 in x 3.5 in x 20 in)
Weight: Net, 6.8 kg (15 Ib); shipping, 9.1 kg (20 Ib)
HP 85047A S-Parameter Test Set
The HP 85047A test set includes a frequency doubler that can be
switched in to measure 3 MHz to 6 GHz in a single sweep or switched out
to measure 300 kHz to 3 GHz in a single sweep. The HP 8753B/C con-
trols the frequency doubler. (The HP 8753D Option 006 and 011 with
built-in 6 GHz source does not use the frequency doubler.) Option 006
(6 GHz receiver) is required to activate the HP 85047A
Specifications Summary
impedance: 50 Q
Frequency Ranges: 300 kHz to 3 GHz
3 MHz to 6 GHZ
300 kHz to 6 GHz (HP 8753D Option 006 and 011)
Directivity: 300 kHz to 1.3 GHz: 35 dB
1.3 GHz to 3 GHz: 30 dB
3 GHz to 6 GHz: 25 dB
'Degrees, specified as deviation from linear phase.
*Fmax is the upper frequency limit of the associated test set.
*Can be improved through accuracy enhancement
Typical Tracking
Transmission Magnitude, Phase
300 kHz to 3 GHz: +1.5 dB, +10
3 GHz to 6 GHz: +0.5, — 2.5 dB, +20
Reflection Magnitude, Phase
300 kHz to 3 GHz: +1.5 dB, +10
3 GHz to 6 GHz: +1.5 dB, +20
Effective Source Match
300 kHz to 1.3 GHz: 20 dB
1.3 GHz to 3 GHz: 16 dB
3 GHz to 6 GHz: 14 dB
RF Connectors
Test Ports: Precision 7 mm
All Others: 50 Q Type-N (female)
Includes: Four 190 mm (7.5 in) cables with Type-N (male) connectors for
connection to the HP 8753, one HP 8753 test set interconnect cable
Physical Characteristics
Size: 426 mm W x 90 mm H x 508 mm D (16.75 in x 3.5 in x 20 in)
Weight: Net, 10 kg (22 Ib); shipping, 15 kg (33 Ib)
Special Test Sets
HP 8753D Option K36 Duplexer Test Adapter enables single-connection
error corrected measurement of return loss at Tx, Rx, and Ant ports, and
insertion loss from Tx-to-Ant, Ant-to-Tx, Rx-to-Ant, and Ant-to-Rx ports
For applications requiring Tx-to-Rx and Rx-to-Tx measurements
Option K39 three-port test set is recommended
Contact HP for technical specifications and information about addi-
tional special options
Solid-State Switching
Solid-state switching allows for simultaneous measurement of forward
and reverse parameters and continuous update of all four S-parameters as
required for two-port error correction (used to achieve best possible mea-
surement accuracy). Option 009 replaces the standard solid-state RF test
port switch with a mechanical RF switch. HP 8753 system specifications
for standard and Option 009 test sets are identical. Nominal insertion loss
of the solid-state switch is less than 2 dB (at 3 GHz) or 3 dB (at 6 GHz),
relative to a mechanical switch
HP 86389A/B Solid-State Switch Upgrade Kits
The HP 86389A/B kits retrofit any existing HP 85046A/B and HP 85047A
S-parameter test set by replacing the mechanical RF test port switch with
a solid-state RF switch. This solid-state switch allows for simultaneous
measurement of forward and reverse parameters and continuous mea-
surement of all four S-parameters (required for two-port error correction)
The HP 86389A retrofits HP 85046A/B test sets, and the HP 86389B
retrofits HP 85047A test sets. HP 8753C network analyzers with firmware
revision 4.0 or higher support solid-state test sets.
Hewlett-Packard's fax service provides technical information 24 hours a day 1-800-800-5281
HP 86205A/86207A
Accessories
HP 86205A/86207A RF Bridges
The HP 86205A/86207A high directivity RF bridges offer unparalleled
performance in a variety of general-purpose applications. They are ideal
for accurate reflection measurements and signal leveling applications.
Specifications Summary
HP 86205A HP 86207A
impedance: 50 Q 752
Frequency Range: 300 kHz to 6 GHz 300 kHz to 3 GHz
Directivity: 30 dB, 0.3 MHz to 5 MHz 30 dB, 0.3 MHz to 5 MHz
40 dB, 5 MHz to 2 GHz
30 dB, 2 GHz to 3 GHz
(typical) 20 dB, 3 GHz to 5 GHz
(typical) 16 dB, 5 GHz to 6 GHz
(<3 GHz) 16.0 dB, + 0.15 dB/GHz
(> 3 GHz) 16.5 dB, — 0.20 dB/GHz
Insertion Loss: 1.5 dB, + 0.1 dB/GHz
Maximum Input: 25 dBm
RF Connectors: 50 Q Type-N (female)
Physical Characteristics
Size: 93 mm H x 160 mm W x 23 mm D (3.7 in x 6.3 in x 1 in)
Weight: Net, 0.57 kg (1.3 Ib); shipping, 1.8 kg (4 Ib)
40 dB, 5 MHz to 1.3 GHz
35 dB, 1.3 GHz to 2 GHz
30 dB, ? GHz to 3 GHz
Coupling Factor:
75 Q Type-N (female)
HP 11850C/D Three-Way Power Splitters
Specifications Summary
HP 11850C HP 11850D
Impedance: 50 Q 75 ©
Frequency Range: DC to 3 GHz DC to 2 GHz
Tracking: +0.25 dB, +3” +0.2 dB, + 2.5°
Equivalent Source Match 30dB at 1.3 GHz 30 dB at 1.3 GHz
(ratio or leveling): 20 dB at 3 GHz 20 dB at 3 GHz
Nominal Insertion Loss: 9.5 dB + 1 dB/GHz 7.8 dB
Input Port Match:
DC to 1.3 GHz 20 dB 20 dB
1.3 GHz to Fr. 10dB 10dB
RF Connectors
RF input: 50 £2 Type-N 50 Q Type-N
(female)
All Others: 50 © Type-N 75 Q Type-N
(female)
Frox is the upper frequency limit of the associated power splitter.
HP 11851B RF Cable Kit
This kit includes three 610-mm (24-in) 50 Q cables phase matched to 4° at
1.3 GHz and one cable 860 mm (34-in). Connectors are type-N (male).
HP 11852B 50 2 /75 Q Minimum Loss Pad
The HP 11852B is a low SWR minimum loss pad required for measure-
ments on 75 Q devices with the HP 8753C/D receiver.
Frequency Range: DC to 2.0 GHz
Insertion Loss: 5.7 dB
Return Loss: 75 Q typically > 30 dB, 50 Q typically > 26 dB
NETWORK ANALYZERS
Accessories
HP 8753D Series
Maximum Input Power: 250 mW (+24 dBm)
RF Connectors: 50 Q type-N (f) and 75 Q type-N (m) standard,
50 € type-N (m) and 75 Q type-N (f) Option 004
Type-N Accessory Kits
Each kit contains a type-N (female) short, a type-N (male) short, two type-
N (male) barrels, two type-N (female) barrels, and a storage case.
HP 11853A 50 Q Type-N Accessory Kit
The HP 11853A accessory kit furnishes the RF components required for
measurement of devices with 50 Q type-N connectors using the
HP 11850C, 85044A, 85046A, or 85047 A
HP 11855A 75 Q Type-N Accessory Kit
The HP 11855A accessory kit furnishes the RF components required for
measurement of devices with 75 Q type-N connectors using the
HP 11850D, 85044B, or 85046B. This kit also contains a 75 Q type-N
(male) termination.
BNC Accessory Kits
The BNC accessory kit contains two type-N (male) to BNC(female)
adapters, two type-N (male) to BNC (male) adapters, two type-N (female)
to BNC (female) adapters, two type-N (female) to BNC (male) adapters,
a BNC (male) short, and a storage case.
HP 11854A 50 Q BNC Accessory Kit
The HP 11854A accessory kit furnishes the RF components required for
measurement of devices with 50 Q BNC connectors using the
HP 11850C, 85044A, 85046A, or 85047A.
HP 11856A 75 Q BNC Accessory Kit
The HP 11856A furnishes RF components required for measurement of
devices with 75 Q BNC connectors using the HP 11850D, 85044B, or
85046B. This kit also contains a 75 Q BNC (male) termination.
Test Port Cables
HP 11857D 50 Q APC-7 Test Port Cables
The HP 11857D includes two precision 61-cm (24-in) cables, phase
matched to 2° at 1.3 GHz for use with the HP 8753D, 85046A or 85047A
S-parameter test sets. Connectors are 50 Q APC-7.
HP 11857B 75 Q Type-N Test Port Cables
The HP 11857B includes two precision 61-cm (24-in) cables, phase
matched to 2° at 1.3 GHz for use with the HP 8753D Option 075 or
HP 85046B S-parameter test set. One cable has 75 Q type-N (male)
connectors on both ends; the other has one type-N (male) and one type-N
(female) connector.
HP 85043D System Cabinet
The HP 85043D system cabinet has been ergonomically-designed
specifically for the HP 8753 and the HP 85046A/B or 85047A S-parameter
test sets. The 132-cm (52-in) system cabinet includes a bookcase, a draw-
er, and a convenient work surface.
Calibration Kits
The calibration kits in the HP 8753 family contain precision standards
used in accuracy enhancement procedures to characterize the systemat-
ic errors of an HP 8753 measurement system.
HP 85031B 7-mm Calibration Kit
The HP 85031B Calibration Kit contains a set of precision 7-mm fixed
terminations, and a one-piece open/short circuit used to calibrate the
HP 8753 and its 50 Q test sets for measurement of devices with precision
7-mm connectors. This kit is specified 300 kHz to 6 GHz.
To have a Hewlett-Packard representative help you place an order or to get more information call 1-800-452-4844
2
NETWORK ANALYZERS
Accessories (cont'd)
HP 8753D Series
HP 85032B 50 Q2 Type-N Calibration Kit Ordering Information Price
The HP 85032B calibration kit contains precision 50 © type-N standards HP 8753D Network Analyzer, 30 kHz to 6 GHz $34,500
used to calibrate the HP 8753 and its 50 2 test sets for measurement of Integrated network analyzer with built-in color display,
devices with 50 Q type-N connectors. Precision phase-matched 7 mm to S-parameter test set, disk drive, and 30 kHz to 3 GHz
50 Q type-N adapters are included for accurate measurements of non- synthesized source. Standard 50 Q version has two
insertable devices. Standards include fixed terminations, open circuits, 7-mm test ports.
and short circuits in both sexes. This kit is specified from dc to 6 GHz. Opt 002 Harmonic Measurement Capability +$3,500
ont 006 6 GHz Frequency Extension hens
- ‘ : . | pt 010 Time-Domain Capability +$5,300
HP 85032E 50 Q Type-N Economy Calibration Kit Opt 011 Delete Built-in Test Set “$6 500
The HP 85032E calibration kit contains a type-N (m) fixed termination Opt 07575 Q Impedance $0
and a one-piece type-N (mn) open/short circuit. The kit is specified from Opt 1D5 High-Stability Frequency Reference +$1,000
dc to 6 GHz. HP 85047A 50 Q S-Parameter Test Set—6 GHz $11,000
Opt 009 Mechanical tons Port poate” Rite
- i ! ; pt 913 Rackmount Kit {5062-4069 +$41
HP 85033D 3.5-mm Calibration Kit HP 85046A 50 Q S-Parameter Test Set—3 GHz $9,180
The HP 85033D calibration kit contains fixed loads and open and short Opt 009 Mechanical Test Port Switch —$1,020
circuits in both sexes to calibrate the HP 8753 and 50 Q test sets for mea- Opt 913 Rackmount Kit (5062-4069) +$41
surement of devices with precision 3.5-mm and SMA connectors. Phase- HP 85046B 75 Q S-Parameter Test Set—300 kHz-2 GHz $9,180
matched 7-mm to 3.5-mm adapters for male and female connectors are Opt 009 Mechanical Test Port Switch -$1,020
included for use with 7-mm test port cables. This kit is specified from dc Opt 913 Rackmount Kit (5062-4069) +$41
to 6 GHz. HP 85029B Precision 7-mm Verification Kit $1,735
н Opt 001 Data for HP 7028 к 2
- i f i P 85031B Precision 7-mm Calibration Kit ` 1,225
НР 650568 75 2 Туре N Calibration Kit HP 85032B 50 Q Type-N Calibration Kit $1,940
The HP 85036B calibration kit contains precision 75 © type-N standards HP 85032E 50 © Type-N Economy Calibration Kit $665
used to calibrate the HP 8753 and its 75 © test sets for measurement of HP 85033D 3.5-mm Calibration Kit $2,950
devices with 75 Q type-N connectors. Standards include fixed terminations, НР 85036B 75 Q Type-N Calibration Kit $2 040
open circuits, and short circuits in both sexes. Precision phase-matched HP 85039A Type-F Calibration Kit $2 350
adapters are included for accurate measurements of non-insertable devices. HP 85060C Electronic Calibration Contro! Unit $6,100
This kit is specified from dc to 3 GHz. HP 85060 Series Electronic Calibration Modules Price varies
HP 85043D System Rack $3,265
HP 85039A Type-F Calibration Kit HP 11883A Harmonic Measurements (Option 002) $3,570
The HP 85039A calibration kit contains type-F male open, short, and load Tr A 6 GHz Receiver (Option 006) Upgrade $3 570
standards to calibrate the HP 8753 and 75 Q test sets for measurement of HP 11884B 6 GHz (Option 006) Upgrade Kit $6.000
devices with 75 © type-F connectors. The kit also includes type-N to type- for HP 8753D '
F adapters. This kit is specified from dc to 3 GHz. HP 11884C 6 GHz (Option 006) Upgrade Kit $6,000
| | | | for HP 8753D Option 011
HP 85060 Series Electronic Calibration Modules and HP 85019B Time Domain (Option 010) Upgrade $5,410
Control Unit (HP8753B/C/D)
HP 86389A Solid-State Switch Upgrade Kit $1,530
This series provides electronic calibration (ECal) capability for the (for HP 85046A/B Test Sets)
HP 8753D. The usual calibration kit standards are replaced by two solid- HP 86389B Solid-State Switch Upgrade Kit $1,530
state calibration modules that can be programmed by a control unit to (for HP 85047A Test Sets)
present many different impedances to the test ports. A full two-port cali- HP 86205A 50 Q Bridge $1,325
bration can be done with a single connection in just a few minutes, with HP 86207A 75 Q Bridge $1,325
less chance for error and less wear on connectors. ECal requires an HP 8753D
HP 85060C electronic calibration control unit and HP 85060 series cali- Opt K36 Duplexer Test Adapter +$4,000
bration modules of the appropriate connector type. — OptK39Three-Port Test Set Adapter +$7,000
HP 11850C 50 Q Power enter $970
: . ; HP 11850D 75 Q Power Splitter $1,735
Verification Kits HP 11851B 50 Q/Type-N RF Cable Kit $1,020
Measuring known devices, other than the calibration standards, is a con- HP 118528 50 Q/75 Q Minimum Loss Pad $510
venient way of verifying that the HP 8753 measurement system is operat- HP 11853A 50 Q Type-N Accessory Kit $510
ing properly. HP 11854A 50 Q BNC Accessory Kit $510
НЕ 11855A 75 Q Te Accessory Kit pel
- i I i P 11856A 75 Q BNC Accessory Kit 510
HP 850298 7 mm Verification Kit HP 11857B 75 Q Type-N Test Port Extension Cables $1,485
The HP 85029B verification kit contains a set of precision 7-mm devices, НР 118570 50 Q APC-7 Test Port Extension Cables $1,275
with data traceable to NIST, used to verify the calibrated performance of
an HP 8753A/B/C/D measurement system. The devices have precision S For off-the-shelf shipment, call 800-452-4844,
7-mm connectors and include a 20 dB pad, a 50 dB pad, and a mismatch
attenuator. The verification process requires only an HP 85031B calibra-
tion kit, and an HP 85029B verification kit.
Option 001 is intended solely for use with the HP 8702B lightwave
component analyzer. Option 001 adds verification data that is compatible
with the HP 8702B.
Hewlett-Packard’s fax service provides technical information 24 hours a day 1-800-800-5281
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project

Download PDF

advertising