user instructions and warranty
camel
freedom
through lightweight solutions
camel
MI0202 issue 5
The CAMEL by Mangar, is an inflatable lifting device
which is designed to sit up and lift a fallen person
from the floor. When inflated, the CAMEL lifts the
patient into a raised seated position ready to stand
with or without assistance.
This equipment may be used with or without a helper
as required.
The CAMEL is powered by a Mangar Airflo Compressor
and controlled using a 4-Way Hand Control. Please
refer to the separate Airflo compressor instructions
before operating.
PLEASE BE SURE THAT YOU HAVE READ AND
UNDERSTOOD BOTH THESE AND THE SEPARATE
AIRFLO COMPRESSOR INSTRUCTIONS BEFORE
OPERATING THIS EQUIPMENT.
Patents Pending
CONTENTS
Pg 3
Standard Package Contents
Pg 3
Connecting the Hand Control and Airflo
Pg 3
Extra Air Flow
Pg 4
Airflo ‘Auto’ Control Button
Pg 4
Positioning the patient on the CAMEL
Pg 4
Using the CAMEL
Pg 6
After Use
Pg 6
Maintenance
Pg 6
Routine Cleaning & Disinfecting
Pg 6 Specifications
Pg 7 Symbols used on packaging
Pg 7
2
Warranty
WARNINGS
• This equipment MUST only be used by persons
familiar in its use.
• The medical state of a person needing to be raised
MUST be assessed before commencing a lift.
• Ensure that the surface on which the CAMEL is to
be placed is free from sharp objects.
• If a slide sheet has been used to transfer the
patient onto the CAMEL it must be removed from
under the patient before commencing a lift.
• Sections 2, 3 and 4 MUST always be fully inflated in
sequence from the bottom up, in order to maintain
stability. NOTE: The backrest may be partially
inflated before any other section is raised. Once
the lowest section is fully inflated, the backrest
can be raised further as required.
camel
MI0202 issue 5
Standard Package Contents - fig. 1
A
C
1 x CAMEL (A) including carrying bag (not shown)
1 x 4-Way Hand Control (B)
1 x Airflo Compressor (C) with user instructions
B
1
A
Connecting the Hand Control and Airflo - fig. 2
B
•
Remove the CAMEL from the carrying bag,
unfasten the straps and unroll.
•
Connect the four hoses from the hand control
to the corresponding numbered and coloured
connectors on the CAMEL (A).
•
Connect the hose (B) from the hand control to
the Airflo compressor.
•
Position the Airflo compressor so that it will not be
in the way while positioning the patient. Ensure
that the hoses are not kinked.
2
Extra Air Flow - fig. 3
•
D
Select the ‘High’ Air Flow using the switch on the
underside of the Airflo compressor (D).
Normal
3
High
WARNING: DO NOT operate the Normal/High air
flow switch while the compressor is running.
freedom through lightweight solutions
3
camel
MI0202 issue 5
Airflo ‘Auto’ Control Button - fig. 4
The ‘Auto’ function must be selected in order to
operate the 4-Way Hand Control. When selected, the
Airflo will work automatically when the hand control
buttons are pressed.
•
To switch the ‘Auto’ function on, press and hold
the button (A) for a few seconds until the LED
(B) lights up.
•
To switch the ‘Auto’ function off, Press any of the
Airflo buttons.
WARNING: Pressing any of the Airflo buttons will
cancel the ‘Auto’ function and the hand control will
not work.
B
A
4
Positioning the patient on the CAMEL - fig. 5
•
Position the CAMEL at the side of the patient so
that the patient’s head is level with the CAMEL
badge (A).
•
Transfer the patient onto the CAMEL. The head
should lie on the CAMEL badge (A) and the body
should lie centrally.
5
Using the CAMEL
The following instructions apply equally when
the CAMEL is being operated by a person raising
themselves without assistance, or by an assistant
raising a patient.
WARNING: If a slide sheet has been used to transfer
the patient onto the CAMEL it must be removed
from under the patient before commencing a lift.
•
Press the number 1 button on the hand control
to partially inflate the backrest section until the
patient is in a comfortable position - fig. 6
•
Press the number 2 button to inflate the bottom
section until it becomes rigid - fig. 7
6
7
4
A
camel
MI0202 issue 5
•
Continue to press buttons 3 and 4 in sequence
until the sections become rigid, in exactly the
same manner - fig. 8
•
Press button 1 to fully raise the backrest and arrive
at a sitting position - fig. 9
•
Whether assisted or used alone, the patient should
now be in a position to stand or transfer - fig. 10
8
9
WARNING: In order to maintain stability, sections
must always be fully inflated from the bottom up, in
sequence.
10
Note: only the backrest (section 1) may be partially
inflated before the lower sections.
freedom through lightweight solutions
5
camel
MI0202 issue 5
After Use
• To deflate, disconnect each of the air hoses from
the CAMEL by pushing the release collar on the
connector away from the hose and then pulling
the hose clear of the connector. NOTE: The initial
‘rush’ of air as the hoses are disconnected from
the CAMEL, can be reduced by first disconnecting
the air hose from the Airflo and then pressing
each of the buttons on the hand control.
•
A
Roll up the CAMEL with the hand control in the
hollow (A) - fig. 11 secure the straps and place
in the carrying bag - fig. 12
Maintenance
Other than routine cleaning, the CAMEL is
maintenance free. The backrest section or the
main lifting section can be replaced separately if
required.
11
We recommend that a thorough examination of the
CAMEL is carried out regularly; the frequency of
examination may need to be increased depending
on usage and operating environment.
If required, Mangar International can provide a
thorough examination and full maintenance within an
annual service contract. Please contact our Customer
Service department for further details (See page 7
for further information).
Routine Cleaning
Inflate the CAMEL, sponge clean with a detergent
solution and dry thoroughly before storing in a dry
place. Do not dry using hot air.
12
Disinfecting
Inflate the CAMEL, pre-clean with a detergent
solution and then sponge with a disinfecting solution
or alcohol impregnated wipe. Allow to air dry before
storage. Do not dry using hot air.
Specifications
Maximum user weight:
Seat height (approx):
Overall length (approx):
Overall width (approx):
Deflated & rolled (in bag):
Length :
Diameter:
Weight (Inc. Hand Control):
Maximum air pressure:
Max / Min temperature:
6
320kg (50 st)
56cm
112cm
70cm
76cm
20cm
6.5kg
0.35bar (5psi)
+50°C / -15°C
56 cm
70 cm
112 cm
camel
MI0202 issue 5
symbols used on packaging
Keep packaging dry
Recycle packaging
Handle with care
Do not dispose of in household waste
Max/Min storage temperature
Max/Min humidity when transporting in
pressurised aircraft hold
Net & gross weights
Max/Min pressure when transporting in
pressurised aircraft hold
Stack no more than five high
Max/Min temperature when transporting
in pressurised aircraft hold
Warranty
Mangar International guarantees this product for a period of two years.
The guarantee is given against defects in material and workmanship and comprises piece,
parts and labour. The guarantee is not valid in the case of fair wear and tear, wilful damage,
common neglect, misuse or unauthorised alteration or repair by unqualified persons.
The following statements do not affect the consumer’s statutory rights. This product is sold on the understanding
that in the event of any defect of manufacture or material appearing within two years of the date of receipt of this
product, the defect will be rectified free of charge providing that:
(a) Reasonable evidence is provided (e.g. purchase invoice, delivery note) that the product was not purchased
more than two years prior to the date of the claim.
(b) The defect is not attributable to accidental damage (either in transit or otherwise), misuse or unauthorised repair.
CAMEL serial number:
.............................
Airflo serial number:
.............................
Date of purchase:
.............................
Invoice/Delivery note number:.............................
We strive to provide our customers with reliable and durable products of the highest quality.
In addition to a two year guarantee, and for added peace of mind, our servicing division,
Primeserve, also offers a variety of service options. Choose from an Extended Parts Warranty,
Call-Out Protection and an Annual Service and Clean.
To find out how we can provide complete cover for your Mangar International product call our
Customer Service Department on Freephone 0800 280 0485.
Mangar International reserves the right to alter product specifications and/or any of the information contained within this document without notice.
freedom through lightweight solutions
7
freedom through lightweight solutions
Mangar International
Presteigne
Powys
Wales, UK
LD8 2UF
tel: +44 (0)1544 267674
fax:+44 (0)1544 260287
customerservice@mangar.co.uk
www.mangarinternational.co.uk
Patents Pending
MI0202 issue 5
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project

Download PDF

advertising