Raytek Raytek RAYMR1SBSF Ratio Infrared Temperature Sensor

Raytek Raytek RAYMR1SBSF Ratio Infrared Temperature Sensor
Raytek® MARATHON™ Series
MR1S
High Performance Infrared Ratio Thermometers
MR1S model shown with
optional ThermoJacket
1.800.561.8187
www.
.com
information@itm.com
MR1S Highlights
■
Specifications
Provides accurate measurement of
targets that are: partially occluded,
obstructed by smoke, steam, or
particulates, moving, smaller than
the instrument field of view.
Temperature Ranges:
MR1SA
MR1SB
MR1SC
600ºC to 1400°C (1112ºF to 2552°F)
700ºC to 1800°C (1292ºF to 3632°F)
1000ºC to 3000°C (1832ºF to 5432°F)
Detector
Si/Si layered detector, nominal 1µm
Measures from 600º to 3000ºC
(932º to 4532ºF) with three models
Accuracy
±(0.5% Tmeas +2º C); Tmeas in ºC*
■
Fast response time down to 10 mSec
Temperature Resolution
1°C or °F
■
One-color or two-color operation
Response Time
10 mSec
Simultaneous analog and digital outputs
Emissivity one-color 0.10 to 1.0 in 0.01 increments
Slope two-color
0.85 to 1.150 in 0.001 increments
■
■
■
■
■
■
■
■
Repeatability±0.3% full scale
Programmable relay output (dualtemperature setpoints or “fail-safe”)
Signal ProcessingPeak Hold and Averaging
* No attenuation
Unique “dirty window” alarm
(attenuation measurement
US Patent No. 5,815,410)
Bi-directional RS485 communications
Supports up to 32 Marathon Series
sensors on a multipoint network
Optical
Typical Spot Size at Minimum Focus Distance
Windows DataTemp Multidrop Software
(operates under WIN 3.1/95/98/NT4
Windows 2000, XP)
Models
D:S*
Minimum Spot Size Diameter*
MR1SASF
44:1
14 mm (0.55 inch) @ 600mm (24 inch)
MR1SBSF
82:1
7.3 mm (0.29 inch) @ 600mm (24 inch)
Field Calibration Software
MR1SCSF
130:1
4.6 mm (0.19 inch) @ 600mm (24 inch)
MR1SACF
44:1
7.0 mm (0.28 inch) @ 300mm (12 inch)
MR1SBCF
82:1
3.7 mm (0.15 inch) @ 300mm (12 inch)
MR1SCSF
130:1
2.3 mm (0.09 inch) @ 300mm (12 inch)
®
®
Marathon Ratio Thermometers (MR1S) use
two-color measurement to provide superior
accuracy in demanding applications. The
MR1S thermometers features an advanced
electro-optical design, digital electronics
and built-in user interface—all combined
in a rugged, compact housing. MR1S
thermometers are ideal in situations where
the target may be obstructed due to
atmospheric smoke or other particulates,
where the target is moving, or where the
target is smaller than the field-of-view.
MR1S thermometers allow users to
switch between one-color and two-color
operation, which simplifies setup and
provides continued measurement flexibility.
All Marathon Series thermometers include
bidirectional, RS-485 serial communications
between the sensor (or a local area network
of sensors) on the factory floor and a
personal computer in the control room.
This enables remote setup, monitoring,
calibration, and maintenance, especially
valuable for installation in hard-to-reach
locations.
1.800.561.8187
www.
*Typical spot size diameter=Minimum Focus Distance / Typical D:S
100% Best performance realized when % signal reduction <95%
Consult MR1S Operator’s Manual for further information.
Electrical
Outputs
0/4–20 mA; RS-485, 2-wire/4-wire, networkable to 32 sensors; Relay (SPST 48V, 300 mA, response time < 2 mSec)
Power Requirements
24 VDC, 500 mA, ±20%
Compliance
CE low voltage directive; EN 61326
General
Environmental Rating
NEMA-4 (IEC 529, IP 65)
Ambient Temperature
Without cooling option
With air cooling
With water cooling
With ThermoJacket (water cooled)
0º
0º
0º
0º
Storage Temperature
-20º to 70ºC (-4º to 158ºF)
Relative Humidity
10% to 95% non-condensing
Shock (Electronics Housing)
MIL-STD-810D (IEC 68-2-27)
Vibration (Electronics Housing)
MIL-STD-810D (IEC 68-2-6)
Weight
Sensing Head
With air/water cooled housing
0.48kg (17oz)
0.8kg (28oz)
.com
to
to
to
to
50ºC (32º to 120ºF)
120ºC (32º to 250ºF)
175ºC (32º to 350ºF)
315ºC (32º to 600ºF)
information@itm.com
User Interface
Accessories
Polarizing filter end cap reduces visual light intensity in high
temperature applications (XXXTPFEC)
Air purge collar (XXXTXXACAP) *
Swivel bracket (XXXTXXACSB) * Right angle sighting accessory (MR1S) (XXXTXXACRS)*
Pipe adapter (Compatible with Sighting Tubes)
(XXXTXXACPA) * *Compatible with standard unit and "W" option (not ThermoJacket)
Flow Regulators
Water flow regulator (water cooling) (XXXTXXWR)
Air purging flow/regulator assembly with air filter
(XXXTXXAR)
Cooling air flow regulator (high capacity) (XXXTXXCAFR)
Physical Dimensions
Thermojacket Housing Rated to 315 °C (600 °F)
MR1S Housing
Mounting Flange for ThermoJacket (XXXTXXMF)** 198
(7.78)
Ø 57
(2.24)
161
(6.3)
ThermoJacket housing; for Marathon MA/MR
(RAYTXXTJ1)
Adjustable mounting base (XXXTXXMB)** 20 16
(.8) (.63)
155
(6.1)
Adjustable Pipe Adapter assembly (XXXTXXAPA) **
Ø 24
(.99)
Ø 24
(0.95)
20
(0.79)
mm (inches)
MR1S Housing with Air and Water Cooling (“W” option)
197
(7.76)
– 27 NPT (2 PLCS)
61
(2.4)
56
(2.2)
23
(0.9)
57
(2.24)
12
(0.47)
12
(0.47)
28
(1.1)
16
(.63)
48
(1.9)
Ø38
(1.5)
1 1/2 – 20 UN–2A
35
(1.38)
97
(3.83)
158
(6.22)
DUIT
20
(0.79)
3.3
(0.13)
** For use with ThermoJacket only
Sighting Tubes
300mm (12in.) SIGHTING TUBE, ceramic
(up to 1500°C) (XXXTSTC12)
300mm (12in.) SIGHTING TUBE stainless steel
(up to 800°C) (XXXTST12)
300mm (12in.) SIGHTING TUBE, carbon steel; 45 degree
end-cut with slotted weep hole at base. (BEESIGHTT)
Ø24
(0.95)
Ø76
(3)
Ø24
(0.95)
Mounting flange for use with sighting tubes (XXXTXXMST)** 1 1/2 – 20 UN–2A
Ø 24
(0.95)
FOR 3/4 EMT CONDUIT
35
(1.38)
Ø 38
(1.5)
mm (inches)
NIST Calibration Certification 2132258
Power Supply (24VDC, 110/220VAC input) and Marathon
Terminal Block mounted in a NEMA 4 (IP65) enclosure
(RAYMAPB)
Power Supply 24VDC 1.1A Switching power supply with
universal input (110/220V) (XXX2CDCPSS)
Spare Marathon Terminal Block Accessory (XXXMATB)
Spare Marathon Terminal Block in a NEMA-4
enclosure (XXXMATBN4)
ThermoJacket Housing
RS485/RS232 CONVERTER
DB25 to Terminal Strip Interface Converter, for direct wiring
between a serial interface and the Marathon terminal block
(XXX485CVT)
1.800.561.8187
www.
.com
information@itm.com
Model
RAYMR1S
Code A
A
B
C
Description
Raytek Infrared Ratio Thermometer
Temperature Range
600ºC to 1400ºC (1112ºF to 2552ºF) 44:1 D:S
700ºC to 1800ºC (1292ºF to 3272ºF) 82:1 D:S
1000ºC to 3000ºC (1832ºF to 5432ºF) 130:1 D:S
Code B
SF
CF
Code C
Focus Option
Standard focus variable optics 600mm (24") to ∞
Close focus variable optics 300mm (12") to 600mm (24")
Cooling Option
W
XXXMRCERT
Typical Model
Number
Coolable Housing for cooling in ambient temperatures up to 175ºC (350ºF), includes Lens Air Purge Collar;
NOTE: For cooling in ambient temperatures in excess of 175ºC (350ºF), see ThermoJacket
NIST Calibration Certification (must be ordered with MR1S as a separate line item)
MR1SBCF
© 2013 Raytek Corporation (3111807 Rev. J) 4/2013
Raytek, the Raytek logo and DataTemp are registered trademarks of Raytek Corporation.
Windows is a registered trademark of Microsoft Corporation.
Specifications subject to change without notice.
1.800.561.8187
www.