SK-7601 - KAISE CORPORATION
AC DIGITAL CLAMP METER
INSTRUCTION MANUAL
SK-7601
12. TEMPERATURE COEFFICIENT : 0℃ to 18℃, 28℃ to 40℃ ; add accuracy at
23℃±5℃×0.1 / ℃
13. SAFETY LEVEL : CE marking approved (IEC-61010-1, IEC-61010-2-032 CATⅢ300V
and EMC Test passed)
14. POWER SUPPLY : 3V CR2032 battery x 1
15. POWER CONSUMPTION : 3mA max.
16. CONTINUOUS OPERATING TIME : Approx. 70 hours
17. AUTO POWER OFF : Power turns off automatically after approx. 12 minutes
18. CONDUCTOR DIAMETERS : φ27mm max.
19. DIMENSIONS & WEIGHT : 150(H)×52(W)×24(D)mm, Approx. 110g
20. ACCESSORIES : 1004 Carrying Case, 3V CR2032 Battery (x1, installed),
Instruction manual
21. OPTIONAL ACCESSORIES : 880 Line Separator
2-2. MEASUREMENT SPECIFICATION
(23℃±5℃,
<80%RH in non-condensing)
AC CURRENT (∼A)
40.00A
±1.8%rdg±5dgt
300.0A
FOR SAFETY MEASUREMENTS!!
To prevent an electrical shock hazard to the operator and/or damage to the
instruments, read this instruction manual carefully before using the
instrument. WARNINGS with the symbol
on the instrument and this
instruction manual are highly important.
Important Symbols :
The symbol listed in IEC 61010-1 and ISO 3864 means "Caution (refer
to instruction manual)".
The symbol in this manual advises the user of an
WARNING electrical shock hazard that could result in serious
injury or even death.
CAUTION
The symbol in this manual advises the user of an
electrical shock hazard that could cause injury or
material damages.
WARNING
Measurement on High Power Line (High Energy Circuits) is very dangerous.
It sometimes includes High Surge Voltage that could cause dangerous arcs
of explosive short in the instrument and could result in serious injury to the
operator. When measuring high power line, be careful not touch any part
of the circuit.
Average rectification
Accuracy(50/60Hz)
Range
Resolution
Max. Input Current
10mA
300A
(300V line max.)
100mA
Frequency Response : 0A to 200A ; 40Hz to 500Hz (add 1.5%rdg+3dgt to the above accuracy.)
NOTE : Add 3dgt for 0A to 0.5A.
Correct knowledge of electric measurements is essential to avoid unexpected
danger such as operator's injury or damage to the instrument. Read the following
precautions carefully for safety measurements.
1. UNPACKING AND INSPECTIONS
Confirm if the following items are contained in the package in good condition.
If there is any damage or missing items, ask your local dealer for replacement.
1. Digital Clamp Meter
2. Carrying Case (1004)
3. Battery (3V CR2032)
4. Instruction Manual
1 pce.
1 pce.
1 pce. (installed)
1 pce.
2. SPECIFICATIONS
3-1. WARNINGS
WARNING 1. Checks of the Instrument
Before measurement, check if there are no damage to the instrument. Dust,
grease and moisture must be removed.
WARNING 2. Warning for High Power Line Measurement
High Power Line (High Energy Circuits) such as distribution transformers, bus bars
and large motors are very dangerous. High Power Line sometimes includes High
Surge Voltage that could cause explosive short in the instrument and could result
in shock hazard. Generally, shock hazard could occur when the current between
the circuit, that involves more than 30V AC or 42.4V DC, and ground goes up to
0.5mA or more. When measuring high power line, do not touch the live line and
keep enough distance.
WARNING 3. Maximum Input Observance
Do not measure any current that might exceed the specified maximum input value.
3-2. PRECAUTION FOR USE
CAUTION : Do not measure AC high-frequency current. Clamp head
becomes heated and could damage the instrument.
CAUTION 1.
CAUTION 2.
CAUTION 3.
1. DISPLAY (LCD)
b. Units and Symbols : ∼, A, DH, BAT, AUTO, APO and decimal point.
2. OPERATING PRINCIPLE : Σ⊿ conversion
3. RANGE SELECTION : Auto range
4. OVERLOAD INDICATION : "OL" indication more than 400A
5. BATTERY WARNING : "BAT" indication at approx. 2.4V or less
6. SAMPLING RATE : 2.5 times per second
7. DISPLAY HOLD : Hold indicating values by DH Key
8. OVERLOAD PROTECTION : 400A AC (300V Line) for 1 minute
9. DIELECTRIC STRENGTH : 3.7kV AC for 1 minute (between case and metal part)
10. OPERATING TEMPERATURE & HUMIDITY : 0℃ to 40℃, 80%RH or less
in non-condensing
11. STORAGE TEMPERATURE & HUMIDITY : -10℃ to 50℃, 70%RH or less
in non-condensing.
EN - 1
4-2. POWER Key
Press this key to turn the clamp meter on. To turn off, press it once again.
4-3. DH Key : Display Hold
Hold LCD displayed value by pressing this key. ("DH" lights up)
To release it : Press DH key once again.
4-4. Clamp Head
Clamp a single conductor in the center of clamp head.
NOTE : Measurement cannot be done if the several conductors are clamped.
4-5. Safety Line
The line to protect yourself against electrical shock hazard. Always keep your
fingers behind the safety line.
about electric measurements must not use this instrument.
Do not measure the electricity in naked of barefooted to
protect yourself from electrical shock hazard.
Away the instrument from hot and humid conditions like in
the car. Do not apply hard mechanical shock or vibration.
Do not polish the case or attempt to clean it with any cleaning
fluid like gasoline or benzine. If necessary, use silicon oil or
antistatic fluid.
Remove the battery when the instrument is out of use for a
long time. The exhausted battery might leak electrolyte and
corrode the inside.
4. NAME ILLUSTRATION
WARNING
To avoid electrical shock, detach the clamp meter from the circuit when to
replace the battery and turn it off.
Replace the battery when "BAT" lights up on LCD.
1. Turn off the clamp meter.
2. Loosen a screw of battery cover and remove it in the direction of an arrow.
3. Remove the exhausted battery. Insert a tweezers (or similar tool) into the side
hole of the battery case and push the battery up.
4. Insert a new 3V CR2032 battery in correct polarity, + mark up.
5. Fix the battery cover and tighten the screw.
NOTE : Remove the batteries when the instrument is out of use for a long time.
The exhausted batteries might leak electrolyte and corrode the inside.
NOTE : Installed battery is for inspection purpose and might be exhausted earlier
than the new battery.
screw
1. INSTRUCTION MANUAL
3V CR2032
battery
2. BATTERY
Hole
One 3V CR2032 battery is installed in this instrument. When battery is exhausted,
"BAT" lights up on LCD. For the replacement, refer to "6-2. BATTERY REPLACEMENT".
LCD displays "OL" when the input value exceeds 400A AC.
WARNING
Do not measure any current that might exceed 300A AC (300V line maximum)
to avoid electrical shock hazard and/or damage to the instrument .
4. AUTO POWER OFF
Approx. 12 minutes after the last operation of the POWER or DH keys the clamp
meter will automatically turn off.
NOTE : There is small current consumption even in the auto power off or power
off conditions.
NOTE : Press POWER Key twice to turn on the clamp meter after the auto power off.
5. SYMBOL MARK
Caution (refer to instruction manual.)
Do not apply around, or remove from HAZARDOUS LIVE conductors.
Alternating Current (AC)
Earth (Ground)
Double Insulation
5-2. AC CURRENT MEASUREMENT (∼A)
WARNING
●
●
●
※ Insert a tweezers (or similar tool) in this
hole and push the battery up.
fig. 4
6-2. PERIODICAL CHECK AND CALIBRATION
Periodical check and calibration is necessary to make safety measurements and
to maintain the specified accuracy. The recommended check and calibration term
is once a year and after the repair service. This service is available at KAISE
AUTHORIZED SERVICE AGENCY through your local dealer.
3. OVERLOAD INDICATION
●
CLAMP HEAD
6-1. BATTERY REPLACEMENT
5-1. PREPARATION FOR USE
WARNING 1. Children and the persons who do not have enough knowledge
WARNING 2.
Alternating current (AC)
Low battery warning
Auto range
Display Hold
Unit of Current
Auto Power Off
The following symbol marks shown on the instrument and instruction manual are
listed in IEC 61010-1 and ISO 3864.
3-3. GENERAL WARNINGS AND CAUTIONS
2-1. GENERAL SPECIFICATIONS
a. Numerical Display : 3 3/4 digit, 4000 count, Maximum reading 4000, 12mm high.
fig. 2
:
:
:
:
:
:
Read INSTRUCTION MANUAL carefully to understand the specification and functions
properly. "3. SAFETY PRECAUTIONS" is very important for safety measurement.
Do not put your fingers over the safety line while current measurement.
Thank you for purchasing KAISE "SK-7601 AC DIGITAL CLAMP METER". To
obtain the maximum performance of this instrument, read this Instruction
Manual carefully, and take safe measurement.
6. MAINTENANCE
∼
BAT
AUTO
DH
A
APO
5. MEASUREMENT PROCEDURES
3. SAFETY PRECAUTIONS
WARNING 4. Safety Line
INTRODUCTION
4-1. LCD
Do not measure any current that might exceed the maximum input value,
300A AC (300V line maximum).
Read "3. SAFETY PRECAUTIONS" carefully to avoid electrical shock hazard
and serious damage to the instrument.
Always keep your fingers behind the safety line.
Do not touch any part of the power line or the circuit to be measured.
6-3. REPAIR
Repair service is available at KAISE AUTHORIZED SERVICE AGENCY through
your local dealer. Pack the instrument securely with your name, address,
telephone number and problem details, and ship prepaid to your local dealer.
Check the following items before asking repair service.
1.
2.
3.
4.
5.
Check the battery connection, polarity, and capacity.
Confirm if the over input, exceeding the maximum input value, is not applied.
Confirm that measured accuracy is adopted in the operating environment.
Confirm that the body of this instrument has no cracks or any other damages.
Check if the instrument is not affected by the strong noise generated from the
equipment to be measured or measuring surroundings.
WARRANTY
SK-7601 is warranted in its entirety against any defects of material or workmanship under normal
use and service within a period of one year from the date of purchase of the original purchaser.
Warranty service is available at KAISE AUTHORIZED SERVICE AGENCY through your local
dealer. Their obligation under this warranty is limited to repairing or replacing SK-7601 returned
intact or in warrantable defect with proof of purchase and transport charges prepaid. KAISE
AUTHORIZED DEALER and the manufacturer, KAISE CORPORATION, shall not be liable for any
consequential damages, loss or otherwise. The foregoing warranty is exclusive and in lieu of all
other warranties including any warranty of merchantability, whether expressed or implied.
This warranty shall not apply to any instrument or other article of equipment which shall have
been repaired or altered outside of KAISE AUTHORIZED SERVICE AGENCY, nor which have
been subject to misuse, negligence, accident, incorrect repair by users, or any installation or use
not in accordance with instructions provided by the manufacturer.
KAISE AUTHORIZED DEALER
POWER Key
1. Press POWER Key to turn on the clamp meter.
2. Open Clamp Head and clamp a single conductor.
NOTE : Measurement cannot be done if the
several conductors are clamped.
3. Read the measurement value on LCD.
4. After finishing the measurement, remove the
clamp head from the conductor. Press POWER
Key to turn off.
DH Key
LCD
SAFETY LINE
EN - 2
422 Hayashinogo, Ueda City, Nagano Pref., 386-0156 Japan
TEL : +81-268-35-1601 / FAX : +81-268-35-1603
E-mail : sales@kaise.com http://www.kaise.com
Available Function : Display Hold (refer to 4-3)
fig. 3
fig. 1
EN - 3
Product specifications and appearance are subject to change without notice due to continual improvements.
EN - 4
70-1301-7601-1 0906