pH Electrode Troubleshooting Guide

pH Electrode Troubleshooting Guide
pH Electrode Troubleshooting Guide
Drifting/Erratic Readings
Frozen pH Readings
Contamination on sensing glass – Clean your electrodes regularly.
Visit www.hannainst.com/cleaning for a selection of application-specific
cleaning solutions.
Broken electrode – The pH meter will display the same value when placed
in different buffers or samples. This indicates a crack or break in the
sensing glass.
Clogged junction – Depending on the sample clogging the electrode,
use application specific cleaning solutions.
Visit hannainst.com/cleaning for a list of application specific
cleaning solutions.
How to calculate offset and slope
If the junction is constantly clogging due to measurement in semi solid or
viscous samples, use a pH electrode that has an open junction design like the
FC200. Visit hannainst.com/electrodes for a complete list of electrodes.
Low conductivity sample – Use an electrode that has a high junction
flow rate or add high purity Potassium Chloride (KCl) to increase
conductivity. The triple ceramic junction of the HI1053 allows for
stable readings in low conductivity samples.
Electrode is not properly hydrated – Use HI70300L storage solution
and soak for at least 2-3 hours.
Electrical noise interference —Noise from rectifiers, motors, pumps or
ballasts can interfere with the high impedance measuring circuit. Direct
measurement can be taken with an amplified electrode or electrodes with
a matching pin, otherwise take a grab sample and measure away from the
electronics. The Hanna H99XXX series of meters are application specific
and have amplified electrodes built in. Visit hannainst.com/amplified for a
full selection.
Inaccurate Readings
Wiping a pH electrode with tissue— Wiping or rubbing an electrode
can cause static electricity buildup, removal of the hydration layer on the
electrode bulb, or scratching of the electrode surface. If removal of liquid
from electrode bulb is required be sure to blot with lint-free paper towel
(e.g. Kimwipes®).
Improper calibration – Make sure that pH electrode is rinsed with
distilled water (DI) between calibration buffers to prevent
cross-contamination, and the electrode is at thermal equilibrium
with the buffer. Be sure to use fresh buffer for each calibration.
(For meters that have mV setting) The procedure below is based on
calibration buffers at 25°C. At this temperature the 100% theoretical
slope is 59.16 mV/pH change from pH 7.01. Buffer values differ based on
temperature, but temperature compensation corrects for these changes.
To perform a slope check put the meter in mV mode.
Step 1 Measure mV of pH 7.01 buffer and record value
Step 2 Measure mV value of pH 4.01 buffer and record value
Step 3 Calculate the absolute mV difference (pH 4.01 value – pH 7.01 value)
Step 4 Calculate the slope (mV difference/3)/59.16) = Slope)
Examples
Electrode 1
pH 7.01 = -15 mV
pH 4.01 = +160 mV
Absolute mV difference is +160 mV – (-15 mV) = +175 mV
Slope = (175/3)/59.16 = 98%
Electrode 2
pH 7.01 = +15 mV
pH 4.01 = +160 mV
Absolute mV difference is +160 mV – (+15 mV) = +145 mV
Slope = (145/3)/59.16 = 82%
Conclusion: Electrode 1 is working properly while electrode 2 has an
unacceptable slope. Change the fill solution, clean, condition, and calibrate
the electrode, if this does not improve the slope, replace the electrode.
pH Electrodes Life Span (< 6 months)
Testing high temperature samples
Elevated temperatures reduce the life span of pH electrodes. At room
temperature (25°C), a pH electrode will typically last 1 to 2 years. A general
rule is that for every 25°C increase, the electrode life will decrease by ½.
Operating Temperature
25°C
50°C
75°C
100°C
Hanna recommends using distilled water (DI) for rising electrodes;
however, deionized , RO, pure, or demi water, all work.
(For meters that have mV setting) Check offset and slope of electrode. First
put the meter in mV mode then place the electrode in pH 7.01 buffer, the
mV should read ±30 mV; if outside of this range, try cleaning the electrode.
Slope (difference in mV from pH 7.01 and pH 4.01) must be 150-186mV
(85%-105%). If the slope is less than 85% or more than 105% then use
fresh buffers, change fill solution (refillable electrodes only), and clean
electrode. If the slope is not between 85% to 105%, replace electrode.
Buffers will change their value after they have been opened. pH 10.01 buffer
is especially susceptible to contamination from atmospheric CO2 diffusion.
pH 10.01 buffer should be used within 1-3 months after opening the bottle.
pH 4.01 and pH7.01 should be used within 3-6 months of opening.
If measuring samples at temperatures greater than 50°C, use a pH
electrode with high temperature (HT) glass such as the HI1043.
Storing a pH electrode in purified water – Purified water (DI) will
cause an osmotic effect on the internal reference solution and will decrease
electrode life. If using a refillable pH electrode, replace fill solution; if using a
gel-filled electrode, the electrode will have to be replaced. When not in use,
be sure to store electrodes in storage solution HI70300L is recommended.
Calibrating and measuring at different temperatures—Either use
a meter that has automatic temperature compensation, or calibrate and
measure at the same temperature. Note that the buffer pH at various
temperatures will change. These differences are noted on buffers bottles.
Clean Regularly | Calibrate Often | Condition Always
hannainst.com | Tel. 800-426-6287 | sales@hannainst.com
Average Lifespan
1 to 2 years
6 to 12 months
3 to 6 months
<1 month
Solutions with hydrofluoric acid will dissolve the glass. Use electrodes
with HF resistant glass. The HI1143 will resist HF up to 2 g/L @ pH 2
at temperatures less than 60°C.
Hanna has put together this guide to serve as a quick reference tool.
Always remember to consult the instruction manual or contact us
directly for detailed instructions for your specific needs.
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project

Download PDF

advertising