DustTrak DRX Aerosol Monitor Models 8533 and 8534

DustTrak DRX Aerosol Monitor Models 8533 and 8534
H E A L T H
A N D
S A F E T Y
Exposure Monitoring
DUSTTRAK™ DRX Aerosol Monitor
Models 8533 and 8534
Features and Benefits
Real-Time Dust Monitoring Takes a Giant Leap Forward
Only the new DUSTTRAK™ DRX Aerosol Monitor can
simultaneously measure both mass and size fraction—no
other monitor can do both. DUSTTRAK DRX monitors are
battery-operated, data-logging, light-scattering laser
photometers that give you real-time aerosol mass readings.
They use a sheath air system that isolates the aerosol in the
optics chamber to keep the optics clean for improved
reliability and low maintenance. Suitable for clean office
settings as well as harsh industrial workplaces, construction
and environmental sites and other outdoor applications.
DUSTTRAK DRX monitors measure aerosol contaminants such
as dust, smoke, fumes and mists.
Simultaneously measure size-segregated mass fraction
concentrations corresponding to PM1, PM2.5, Respirable,
PM10, and Total PM size fractions
Easy to program, easy to operate
Industrial/occupational hygiene surveys
New graphical user interface with color touch-screen
Indoor air quality investigations
Perform in-line gravimetric analysis for custom reference
calibrations
Outdoor environmental monitoring
Baseline trending and screening
Automatic zeroing (with optional zero module) minimizes the
effect of zero drift
Point source monitoring
Engineering control evaluations
Engineering studies
Remote and process monitoring
Epidemiology studies
Aerosol research studies
Applications
TRUST. SCIENCE. INNOVATION.
DUSTTRAK DRX Aerosol Monitors Are Light Years
Ahead of Any Other
Desktop Models: Ideal for Long-Term Surveys
and Remote Monitoring Applications
These new laser photometers simultaneously measure five sizesegregated mass fraction concentrations at once—something no
other monitor can do. Both the desktop and handheld monitors are
continuous real-time 90° light-scattering laser photometers that
simultaneously measure size-segregated mass fraction
concentrations corresponding to PM1, PM2.5, Respirable, PM10, and
Total PM fractions. They combine both particle cloud (total area of
scattered light) and single particle detection to achieve mass fraction
measurements.
Manual and programmable data logging functions also make
DUSTTRAK DRX desktop monitors ideal for unattended applications.
They come with USB (device and host), Ethernet, and analog and
alarm outputs allowing remote access to data. User adjustable alarm
setpoints for instantaneous or 15-minute short-term excursion limit
(STEL) are available on desktop models. The alarm output with userdefined setpoint alerts you when upset or changing conditions occur.
This size-segregated mass fraction measurement technique is
superior to either a basic photometer or optical particle counter
(OPC). It delivers the mass concentration of a photometer and the
size resolution of an OPC.
Photometers can be used at high mass concentration, but they
do not give any size information (unless used with size selective
inlet conditioners) and significantly underestimate large particle
mass concentrations.
All DUSTTRAK DRX desktop monitors have four unique features:
Gravimetric sampling capability using a 37-mm filter cassette
which can be inserted in-line with the aerosol stream allowing
you to perform an integral gravimetric analysis for custom
reference calibrations.
They can be zeroed automatically using the external zeroing
module. This optional accessory is used when sampling over
extended periods of time. By zeroing the monitor during
sampling, the effect of zero drift is minimized.
STEL alarm feature for tracking 15-minute average mass
concentrations when alarm setpoint has been reached for
applications like monitoring fugitive emissions at hazardous
waste sites.
Standard and advanced calibration capabilities. The DUSTTRAK
DRX Aerosol Monitor has two calibration factors: a photometric
calibration factor (PCF) and a size calibration factor (SCF). The
PCF accounts for the photometric response difference between
A1 Test Dust and the aerosol under measurement, while the SCF
accounts for the aerodynamic size difference.
OPC’s provide size and count information; however, they do not
provide any mass concentration information and cannot be used
in high mass concentration environments.
Easy to Program and Operate
The new graphical user interface with color touch-screen puts
everything at your fingertips. The easy-to-read display shows realtime mass concentration and graphical data as well as other
statistical information along with instrument pump, laser and flow
status, and much more. Perform quick walk-through surveys or
program the instrument’s advanced logging modes for long-term
sampling investigations. Program start times, total sampling times,
logging intervals, alarm setpoints and many other parameters. You
can even set up the instrument for continuous unattended operation.
– The primary goal of the standard calibration is to obtain the
SCF for the aerosol of interest. The standard calibration
process is very easy and does not require comparison to
gravimetric samples. Measure with and without a PM2.5
impactor, and the instrument takes the ratio of these two size
distributions and compares this reading to the PM2.5 impactor
transmission efficiency curve to calculate the SCF. However,
the absolute mass concentration may not be as accurate as
the advanced calibration.
– The advanced calibration method yields high size segregated
mass concentration accuracy. It involves two separate
gravimetric measurements to obtain PCF and SCF in
sequence. The advanced calibration will accurately measure
size segregated mass concentrations.
Handheld Models: Perfect for Walk-Through Surveys All Desktop Models
Hot swappable batteries
and Single-Point Data Collection Applications
DUSTTRAK DRX handheld models are lightweight and portable. They’re
perfect for industrial hygiene surveys, point source location monitoring,
indoor air quality investigations, engineering control evaluations /
validation, and for baseline trending and screening. Like desktop
models, they have manual and programmable data logging functions. In
addition, they have single-point data logging capability. Single-point
data collection is used for walk-through industrial hygiene surveys and
indoor air quality investigations.
Gravimetric reference sample capability
Long life internal pump
TRAKPRO Data Analysis Software
Auto zeroing module (optional accessory)
STEL alarm setpoint
New Software Makes Monitoring Easier than Ever
TRAKPRO™ Data Analysis Software allows you to set up and program
directly from a PC. A new feature is the ability for remote programming
and data acquisition from your PC via wireless (922 MHz or 2.4 GHz)
communications or over an Ethernet network. As always, you can print
graphs, raw data tables and statistical and comprehensive reports for
recordkeeping purposes.
DUSTTRAK DRX Aerosol Monitor Features
All Models
Li-Ion rechargeable batteries
Internal and external battery charging capabilities
Outlet port for isokinetic sampling applications
User serviceable sheath flow and pump filters
Logged test pause and restart feature
Logged test programming
All Handheld Models
Long life internal pump
Single-point data collection for walk through surveys
TRAKPRO Data Analysis Software
– Color touch screen—either manual mode or program mode
– TRAKPRO™ Data Analysis Software via a PC
Battery Performance
User adjustable standard or advanced custom calibration settings
Instantaneous alarm settings with visual and audible warnings
Model 8533 (typical)
6600 mAH Li-Ion Battery Pack (P/N 801680)
Real-time graph display
Battery Runtime (hours)
View statistical information during and after sampling
On-screen instrument status indicators: FLOW, LASER and FILTER
Filter service indicator for user preventative maintenance
1 Battery
2 Batteries
up to 6
up to 12
Charge Time * (hours) in DUSTTRAK
4
8
Charge Time* (hours) in external battery charger
(P/N 801685)
4
8
Model 8534 (typical)
3600 mAH Li-Ion Battery Packs (P/N 801681)
Battery
Battery Runtime (hours)
up to 6
Charge Time * (hours) in DUSTTRAK
4
Charge Time* (hours) in external battery charger
(P/N 801686)
4
*of a fully depleted battery
H E A L T H
A N D
S A F E T Y
Exposure Monitoring
Specifications
Models 8533 and 8534
DUSTTRAK™ DRX Aerosol Monitor
Sensor Type
90° light scattering
Particle Size Range
0.1 to ≅15 µm
Aerosol Concentration Range
8533 Desktop
8534 Handheld
Weight
Handheld
Desktop
0.001 to 150 mg/m3
0.001 to 150 mg/m3
2.9 lb (1.3 kg), 3.3 lb (1.5 kg) with battery
3.5 lb (1.6 kg), 4.5 lb (2.0 kg)–1 battery,
5.5 lb (2.5 kg)–2 batteries
Display
Size Segregated Mass Fractions for PM1, PM2.5, Respirable, PM10 and
Total. All displayed
Communications
8533
Resolution
±0.1% of reading or 0.001 mg/m3, whichever is greater
8534
Zero Stability
±0.002 mg/m3 per 24 hours at 10 sec time constant
Power–AC
Switching AC power adapter with universal line cord included, 115–
240 VAC
Flow Rate
3.0 L/min
Analog Out
8533
Flow Accuracy
±5% of factory set point, internal flow controlled
Alarm Out
8533
Temperature Coefficient
+0.001 mg/m3 per °C
Operational Humidity
0 to 95% RH, non-condensing
Time Constant
User adjustable, 1 to 60 seconds
Data Logging
5 MB of on-board memory (>60,000 data points)
45 days at 1 minute logging interval
Log Interval
User adjustable, 1 second to 1 hour
Physical Size (HWD)
Handheld
8534
Screen
8533
8534
5.7 in. VGA color touchscreen
3.5 in. VGA color touchscreen
Gravimetric Sampling
8533
Removable 37 mm cartridge (user supplied)
CE Rating
Immunity
Emissions
EN61236-1:2006
EN61236-1:2006
TSI Incorporated
USA
UK
France
Germany
India
China
Singapore
Specifications are subject to change without notice. TSI, the TSI logo, DUSTTRAK, and TRAKPRO are
trademarks of TSI Incorporated. Microsoft and Windows are trademarks of Microsoft Corporation.
4.9 x 4.8 x 12.5 in.
(12.5 x 12.1 x 31.6 cm)
5.3 x 8.5 x 8.8 in.
(13.5 x 21.6 x 22.4 cm)
Desktop
- 500 Cardigan Road, Shoreview, MN 55126-3996 USA
Tel: +1 800 874 2811
Tel: +44 149 4 459200
Tel: +33 491 11 87 64
Tel: +49 241 523030
Tel: +91 80 41132470
Tel: +86 10 8251 6588
Tel: +65 6595 6388
E-mail: info@tsi.com
Website: www.tsi.com
E-mail: tsiuk@tsi.com
Website: www.tsiinc.co.uk
E-mail: tsifrance@tsi.com Website: www.tsiinc.fr
E-mail: tsigmbh@tsi.com Website: www.tsiinc.de
E-mail: tsi-india@tsi.com
E-mail: tsibeijing@tsi.com
E-mail: tsi-singapore@tsi.com
Contact your local TSI Distributor or visit our website www.tsi.com for more detailed
specifications.
P/N 6001982 Rev D (A4)
User selectable output, 0 to 5 V or
4 to 20 mA
User selectable scaling range
Relay or audible buzzer
Relay
Non-latching MOSFET switch
User selectable set point
–5% deadband
Connector 4-pin, Mini-DIN connectors
Audible buzzer
Operational Temp
32 to 120°F (0 to 50°C)
Storage Temp
-4 to 140°F (-20 to 60°C)
USB (Host and Device) and Ethernet. Stored
data accessible using flash memory drive
USB (Host and Device). Stored data
accessible using flash memory drive
Copyright © 2011 by TSI Incorporated
Printed in U.S.A.
TRUST. SCIENCE. INNOVATION.