Single-Detector “Quantum/Canon 50G”: Floor Mounted System

Single-Detector “Quantum/Canon 50G”: Floor Mounted System
DRX-Ascend System
Specifications
JULY 21, 2011
DRX-Ascend System Specifications
TABLE OF CONTENTS
Integrated Operator Console ..................................................................................................................................................2
Software Options ....................................................................................................................................................................3
Systems (Generator and X-ray Tube Combinations) .............................................................................................................3
X-ray Generators ....................................................................................................................................................................3
X-ray Tubes ............................................................................................................................................................................4
High Voltage Cables ...............................................................................................................................................................4
Tube Stands............................................................................................................................................................................4
Collimators ..............................................................................................................................................................................5
Dose Area Product Measuring System ..................................................................................................................................5
Tables .....................................................................................................................................................................................5
Wall Stand...............................................................................................................................................................................6
DR Detectors ..........................................................................................................................................................................6
Detector / Cassette Holders....................................................................................................................................................7
INTEGRATED OPERATOR CONSOLE
1019850
1019876
1019926
22” Wide-Screen Touch-Panel LCD Monitor Operator Console [RC-OPC-W]
• Operator’s graphical user interface for Modality Worklist query, patient registration & exam configuration, image
acquisition, image preview & QC, integrated generator, DICOM Print & Store
• Wide-screen all-in-one display presents the DR operator interface combined with the full-featured generator console
• Touch-panel screen appropriately displays alpha or numeric software-keyboard whenever operator entry is required
Generator “DRX-Series” Integration to DRX Console [Q-CONNECT-CS]
• Integrated Software & License for DRX Series
• Generator mini-control with on/off switch and rotor prepare/expose hand switch
• All generator service configuration, calibration, and diagnostics operated from DR console
DRX-Ascend Console
• PC hardware HP5700; 500GB HDD, 1.8 GHz Duo Core PC, 4 GB RAM, Keyboard, Mouse (Cat # 1010834)
• Includes QG-DIG-DRX: DRX integration for x-ray generator for pre-configured technique for each view
• Includes EVP-Plus Image Processing (Cat # 1010719)
• High-order, multi-frequency image processing ; high-contrast and low-contrast visualization of bone and soft tissue
• Consistent fixed-look body part processing and Improved Image processing of un-known body parts
• Choice of adaptive image processing “Looks”
• Simple orthogonal reprocessing controls
• Includes DICOM Modality Worklist; automated Patient registration; configurable query filter and polling interval (Cat #
1010750)
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
Includes IHE Scheduled Workflow and Patient Information Reconciliation (Cat# 1010826)
Includes CD/DVD image burning in DICOM format with DICOM Viewer (Cat # 1010768)
Barcode scanner for patient registration and wireless detector enable/disable
Procedure Code Mapping software to automatically map RIS codes to specific views for each examination
Grid detection and suppression software automatically detects and suppresses gridline artifacts
Automated and manually adjusted Black Masking eliminates flare for better viewing
Low Exposure Optimization reduces noise in areas of low exposure to preserve detail
Annotations including Text, Distance Measurements, Angles, Cobb Angle, Comments, Electronic Markers,
Image manipulations for Flip. Rotate, Pan, Zoom, Invert Grayscale, Brightness, Latitude, and Detail Contrast
DICOM-Store and DICOM-Print (multiple connections allowed) and DICOM Storage Commit
IHE Consistent Presentation of Images for consistent presentation of images at PACs destinations
MPPS color-scheme status for scheduled, started, completed, arrived.
Security Audit Log supports privacy regulation requirements
Cisco Security Agent monitors for evidence of malicious software; provides threat protection and mitigation
Repeat/Reject Analysis and exposure data monitors staff performance, productivity, and patient care
Configurable anonymous Emergency/Trauma Defaults
Pre-configured SNOMED codes for multi-accession procedure mapping, hanging protocols, and patient orientation
Exam Tutor displays positioning icons as visual coach for each unexposed view in an exam
Preview image displayed for each acquisition; configurable for “auto-accept” and ready for next exposure.
DICOM Image transmission configurable as DX or CR modality
Image re-assignment for patient/study/view reconciliation/correction
Configurable high/low watermarks for image retention, with continuing retention of patient demographics
Page 2 of 7
DRX-Ascend System Specifications
SOFTWARE OPTIONS
1022995
1022979
Administrative Analysis Software
• Detailed statistics and analysis of detector use, radiographer use, repeat rate and exposure index
Tube and Line Visualization Software
• with a single x-ray exposure, create a high contrast / high sharpness companion image to visualize PICC lines and
catheters
• Customer defined studies to automatically create a companion image for improved technologist productivity
• Send only original image to PACS or send original plus companion image to PACS, customer configurable
SYSTEMS (GENERATOR AND X-RAY TUBE COMBINATIONS)
• Generators and X-ray Tubes are not ordered separately with individual CAT numbers
• Each combination of {Generator + X-ray Tube} has a unique CAT number that is used for ordering, according to the
following table:
X-RAY GENERATORS
QG-50
1021336
1021328
QG-50-SE
1011097
QG-65
1019827
50 kW / 125 kVp “ULTRA HF” Deluxe Radiographic Generator
• ULTRA High Frequency Power, 120 kHz PLUS
• X-Ray generator designed for complete integration to Digital Workstation, for easy operator use
• 50 kW maximum output (according to IEC 601)
• kVp Range: 40 - 125 kVp, in 1 kVp increments
• mA Range: 25 - 650
• mAs Range: 0.025 - 800
• Timer Range: 0.001 - 6.300 seconds (or as limited in the QG-DIG-DRX interface)
• Self-Diagnostics, Anode Heat Unit monitor, Error Messaging, Auto shut-off timer, History reporting log
• Input power 380 - 480 VAC (± 5%) Three Phase
• Input power 208 VAC Three Phase Input
50 kW / 125 kVp “ULTRA HF” Deluxe “STORED ENERGY” Generator
• Digital Imaging Ready, ULTRA High Frequency Power, 120 kHz PLUS
• 50 kW maximum output; (according to IEC 60601)
• mA Range: 25 to 600
• kVp Range: 40 to 125 kVp, in 1 kVp increments (150kVp optional)
• mAs Range: 0.025 – 500
• Timer Range: 0.001 – 6.3 seconds
• “APR” Anatomical Programmed Radiography (100 APR Views / 5000 Techniques) for std. and custom views
• Large Graphic Color LCD display for APR and technique information, includes date/time feature
• Self-Diagnostics, Anode Heat Monitor, Error Messaging, Auto Shut-off Timer, History Reporting Log, RS-232 Port
• STORED ENERGY technology permits full operation using a standard wall outlet or alternate power source
• * Input power (50/60 Hz): 105 - 130 VAC, 10 Amp (+/- 10%) Single Phase; (or 220 – 240 VAC for Int’l)
65 kW / 150 kVp “ULTRA HF” Deluxe Radiographic Generator
• ULTRA High-Frequency Power, 120 kHz
• X-Ray generator designed for complete integration to Digital Workstation, for easy operator use
• 65 kW maximum output (according to IEC 601)
• kVp Range: 40 - 150 kVp, in 1 kVp increments
• mA Range: 25 - 800
• mAs Range: 0.025 - 800
• Timer Range: 0.001 - 6.300 seconds (or as limited in the QG-DIG-DRX interface)
• Self-Diagnostics, Anode Heat Unit monitor, Error Messaging, Auto shut-off timer, History reporting log
• Nominal Input power 380 - 480 VAC (± 5%) Three Phase
Page 3 of 7
DRX-Ascend System Specifications
QG-80
1011063
80 kW / 150 kVp “ULTRA HF” Deluxe Radiographic Generator
• ULTRA High Frequency Power, 120 kHz PLUS
• X-Ray generator designed for complete integration to Digital Workstation, for easy operator use
• 80 kW maximum output (according to IEC 601)
• kVp Range: 40 - 150 kVp, in 1 kVp increments
• mA Range: 25 - 800
• mAs Range: 0.025 - 800
• Timer Range: 0.001 - 6.300 seconds (or as limited in the QG-DIG-DRX interface)
• Self-Diagnostics, Anode Heat Unit monitor, Error Messaging, Auto shut-off timer, History reporting log
• Nominal Input power 380 - 480 VAC (± 5%) Three Phase
1021393
1021435
GENERATORS: STANDARD ITEMS
Automatic Exposure Control: (AEC) electronics for “ULTRA HF” generators [QG-AEC]
Exposure Hand Switch: 2 position with retractable coil cord and holder [R80-HS]
1011105
1021427
GENERATORS: OPTIONAL ITEMS (for 40kW and 50kW only, not for SE)
High Speed Starter (Internal): Anode acceleration to 10,000 RPM, w/ dynamic braking [Q-HSS]
150 kVp Output: Increases generator output to 150 kVp [QG-150]
X-RAY TUBES
1011188
1011162
1011246
1011220
4” Radiographic X-Ray Tube (Dunlee DA 1036) [R10-TD36]
• 0.6 / 1.2 mm Focal spots sizes with 400,000 Heat Unit capacity
• 150 kVp, High/Standard speed rotor, 13.5° anode target angle, 90° horn angle
4” Radiographic X-Ray Tube (Toshiba E7254FX) [R10-T600]
• 0.6/1.2 mm focal spot sizes with 400,000 Heat Unit capacity
• 150 kVp, High/standard speed rotor, 12° anode target angle, 90° arms
4” Radiographic X-Ray Tube (Varian RAD-92) [R10-T920V]
• 0.6 / 1.2 mm focal spot sizes with 600,000 Heat Unit capacity
• 150 kVp, High/standard speed rotor, 12° anode target angle, 90° arms
4” Radiographic X-Ray Tube (Dunlee DA 1092) [R10-TD92]
• 0.6/1.2 mm focal spot sizes with 600,000 Heat Unit capacity
• 150 kVp, High/standard speed rotor, 12° anode target angle, 90° arms
HIGH VOLTAGE CABLES
1021344
1011261
1011279
One Pair, 12 meters (40 feet) long with Federal Terminals [R70-40]
One Pair, 15 meters (50 feet) long with Federal Terminals [R70-50]
One Pair, 18 meters (60 feet) long with Federal Terminals [R70-60]
1021443
Deluxe Floor Mounted Tubestand [QS-550]
• Floor Mounted tubestand with 10 ft. long tracks and 98” (249 cm) of longitudinal travel
• Deluxe Handgrips: multi-function, fingertip controls for horizontal, vertical, transverse, and longitudinal movements
• Includes “All Locks” release switch and auto-stop sensor for horizontal/vertical adjustments
• Vertical Travel of 60.5” (154 cm) with minimum floor to focus distance of 13.75” (35 cm)
• FAIL-SAFE electromagnetic braking system and integral counterbalancing ensure safe, easy use
• Column Rotation (±90°), Transverse Arm (10 inch travel)
• Tube Angulations of ±135° with detents at 0°, ±90°
• Cable concealment and management system
Deluxe Floor Mounted Tubestand – NO Transverse Travel [QS-550-NT]
• Same as QS-550, but without Transverse travel
TUBE STANDS
1021468
1021476
1011139
TUBE STANDS: STANDARD ITEMS for QS-550 (not for QS-550-NT)
Trunnion Rings: For tube head rotation of -20°/+45° with dual angle guides [QS-55T4]
TUBE STANDS: OPTIONAL ITEMS
“TECHVISION” 2-way generator integration and Image Preview for “DRX-Series” [QG-TV-IP-F]
• Multi-Color touch-panel screen mounted at the tubestand handgrip
• Allows Technologist to remain close to the patient during exams improving patient care and productivity
• Full “2-way” generator integration with DR console for in-room control and display of generator parameters
• DR “Image Preview”, Accept/Reject, and previous Image recall for the current patient
• Study Management synchronized with DR console for selecting/sequencing anatomical views
Page 4 of 7
DRX-Ascend System Specifications
COLLIMATORS
1021369
1011519
1021377
“Progeny MC150” Manual Collimator [R40-M-P]
• Laser light for patient and cassette tray positioning, plus rectangular light field with cross hair markings
• Lamp/Timer Feature; includes spare projection lamp
• Swivel Mount allowing 360° with 90° detents
• Radiation field size scale for 40” & 72” SID (100cm & 180cm)
• Integrated SID tape measure
“Progeny Pinnacle” Motorized Selectable Collimator [R40-S-P]
• Motorized collimation system with digital display
• Microprocessor controlled shutter system, provides pre-formatted sizing and SID selection
• Swivel Mount allowing 360° with 90° detents
• Laser Light localizer for accurate cassette and patient positioning,
• Integrated SID tape measure
• Automatic Lamp/Timer Feature; includes spare projection lamp
• Radiation field size scale 40”- 72” SID range (100cm & 180cm)
Auto-Shutter APR-Controlled Collimator (Huestis) [R40-A-S]
• Fully integrated with DR Console for pre-programmed automated field size by body part
• Collimation “HOLD” feature leaves collimator blades unchanged for next body part
• Collimator field light controlled from DR console
• Requires QG-APRC Generator interface option for DR APR collimator (CAT 1021385)
DOSE AREA PRODUCT MEASURING SYSTEM
1011659
Dose Monitoring: Integrated Dose Area Product (DAP) Meter [R35-002]
• Provides digital readout of Dose Area Product into the DICOM header for each acquired image
• Provides digital readout of accumulated Dose Area Product for a sequence of exposures
• Requires R35-003 Mounting Rails (CAT 1021575)
• Requires QG-DAP Generator interface option for integrated DAP (CAT 1021583)
TABLES
1021351
1011329
1021567
1021401
1021492
1021526
1011592
Float-Top Radiographic Table [QT-740]
• 650 lb. (295.5 kg.) Patient Weight Capacity
• Tabletop length: 85” (216 cm) with 32” (82 cm) of longitudinal travel
• Tabletop width: 35.5” (90 cm) with 11.5” (29.2 cm) of transverse travel; (EXTRA WIDE DESIGN, FOR LARGE
PATIENT COMFORT)
• “Flat-Top” Table design for easy patient transfer and cleaning, with low absorption material
• FAIL-SAFE electromagnetic braking system ensure safe, easy use
• Recessed Foot switch and table Hand Control switch (for floating-top function)
• Includes adjustable patient handgrips, mounted along concealed accessory rails on the sides of table
• Power input: 115 VAC/60 Hz/single phase/10 Amp (Specify for International use: 220 VAC/single phase/50Hz)
“QUIET-LIFT” Elevating / Float-Top Radiographic Table: for Direct Digital (DR) [QT-750]
• 650 lbs. (295.5 kg.) Patient Weight Capacity
• Elevating Range of 21” - 32.5” (53 - 83 cm) with collision avoidance electronics and safety lock-out control switch
• Tabletop length: 85” (216 cm) with 32” (82 cm) of longitudinal travel
• Tabletop width: 35.5’ (90 cm) with 11.5” (29.2 cm) of transverse travel; (Extra wide design, for large patient comfort)
• Flat top design for easy patient transfer and cleaning, with low absorption material
• FAIL-SAFE electromagnetic braking system ensure safe, easy use
• Recessed Foot Switches for all table movements, with float-top hand control switch
• Adjustable patient handgrips along concealed accessory rails
• Power input: 115 VAC / 60 Hz / single phase / 10 Amp (Specify for International use: 220 VAC / single phase / 50Hz)
TABLE: STANDARD ITEMS
“DR-Tray” for Landscape/Portrait Positioning of DRX Wireless Detector [Q-DRT-DRXT-R]
• Digital Tray with “ROTATING Feature”, allows for one-handed manual rotation of the digital receptor within tray
Ionization Chamber for Digital: with three (3) Fields for use with Table [R80-AEC]
10:1, 103L, 100cm focus anti-scatter Grid; for use with Table [R20-1010M]
TABLE: OPTIONAL ITEMS
Patient Support Straps: for use with Radiographic Tables [QT-PLS]
• Designed for wide variation of patient sizes
• Large 6” wide Velcro-type design, for easy and fast patient securing
• Includes table mounting blocks, for attachment along the tables accessory rail
Patient Compression Device [R90-CB]
Page 5 of 7
DRX-Ascend System Specifications
WALL STAND
1019900 (L)
1019918 (R)
“VERTI-Q” Vertical Wall Stand for Direct Digital (DR); indicate Left or Right hand load [QW-420-D]
• Vertical travel: 60.0” (153 cm)
• Minimum center pixel-to-Floor Distance: 13.75” (35 cm)
• Features the exclusive “EZ-GLIDE” Hand control for easy and precise movement, Grip rotates +105º
• FAIL-SAFE electromagnetic vertical braking system and integral counterbalancing ensure safe, easy use
• Heavy Duty Single-column structure; floor-to-wall or floor-to-ceiling mount
• Includes QW-DIG-INT: QW-420 Interface for Direct Digital (DR) Receptor Mounting and Rotating Tray
1021534 (L)
1021559 (R)
1023019 (L)
1023027 (R)
WALL STAND: CONFIGURATION (select one and specify Right or Left to match Wall Stand)
“DR-Tray” for Landscape/Portrait Positioning of DRX Wireless Detector [Q-DRT-DRXx-R]
• Digital Tray with “ROTATING Feature”, allows for one-handed manual rotation of the digital receptor within tray
Housing for fixed 4343 detector [Q-FDR-V-x]
• Grid is removable
1021401
1021484
WALL STAND: STANDARD ITEMS
Ionization Chamber with three (3) Fields; for use with QW-420 Wall Stand [R80-AEC]
10:1, 103L, 150cm focus anti-scatter Grid; for use with QW-420 Wall Stand [R20-1010L]
1011469
1011493
WALL STAND: OPTIONAL ITEMS
Patient “Side Mounted” Handgrips: (one set), Mounts to “VERTI-Q” Wall Stand [QW-HG20]
Patient “Overhead” Handgrip: Mounts to “VERTI-Q” Wall Stand [QW-HG30]
DR DETECTORS
1018670
1018654
1021591
1021625
DRX-1C “Thin-Lightweight Low-Dose” Wireless 35x43cm DR Detector (CsI) [R-DRX-1C]
• Band 1 / 2 Antenna 5.15 – 5.25 GHz
• Cesium-Iodide (Csl) scintillator on Amorphous Silicon TFT
• ~65% DQE (RQA-5 beam quality) typical
• Imaging Area: 42.7cm x 35.6cm (17” x 14”)
• Pixel Array: 3032 x 2520 pixels
• Pixel Size: 139 micron (Nyquist Resolution 3.6 lp/mm)
• 14-bit A/D
• Preview Image <8 seconds
• Dimensions: 38.35 x 45.95 x 1.55 cm (15 x 18 x 0.6 in.)
• Weight: 8.5 lbs. (3.85kg)
DRX-1 “Thin-Lightweight” Wireless 35x43cm Portable DR Detector (Gd2O2S:Tb) [R-DRX-1]
• Band 1 / 2 Antenna 5.15 – 5.25 GHz
• Gadolinium Oxysulfide scintillator, Amorphous Silicon TFT
• ~32% DQE (RQA-5 beam quality) typical
• Imaging Area: 42.7cm x 35.6cm (17” x 14”)
• Pixel Array: 3056 x 2544 pixels
• Pixel Size: 139 micron (Nyquist Resolution 3.6 lp/mm)
• 14-bit A/D
• Preview Image <8 seconds
• Dimensions: 38.35 x 45.95 x 1.55 cm (15 x 18 x 0.6 in.)
• Weight: 8.5 lbs. (3.85kg)
Stationary 43x43cm DR Detector (Gd2O2S:Tb) for QW-420-D (Wall Stand) [DRX-V4343-G]
• Gadolinium Oxysulfide scintillator, Amorphous Silicon TFT
• ~32% DQE (RQA-5 beam quality) typical
• Imaging Area: 43cm x 43cm (17” x 17”)
• Pixel Array: 3072x3072 (9 million pixels)
• Pixel Size: 139 micron (Nyquist Resolution 3.6 lp/mm)
• 14-bit A/D
• Preview Image <8 seconds
• * Includes Power Box and Gigabit Ethernet interface
Stationary 43x43cm DR Detector (CsI) for QW-420-D (Wall Stand) [DRX-V4343C]
• Cesium-Iodide (CsI) scintillator on Amorphous Silicon TFT (Csi)
• ~65% DQE (RQA-5 beam quality) typical
• Imaging Area: 43cm x 43cm (17” x 17”)
• Pixel Array: 3072x3072 (9 million pixels)
• Pixel Size: 139 micron (Nyquist Resolution 3.6 lp/mm)
• 14-bit A/D
• Preview Image <8 seconds
• * Includes Power Box and Gigabit Ethernet interface
Page 6 of 7
DRX-Ascend System Specifications
1022433
1022458
1021054
1021856
1023035
1021864
1021872
1021898
1021906
1021914
1021955
1021922
1021930
DRX Detector Battery
• Lithium-polymer Technology
• “Smart” battery technology prevents overcharge
• Size 21 x 15 x 0.67 cm
• Weight 0.4 kg (12.4 oz.)
• Voltage / Energy 14.8V DC, 2.1Ah (nominal) capacity
• Charge Capability 190 maximum images per charge
• 14 days per charge in undisturbed “Sleep Mode”
• Expected Life 500 charge / discharge cycles results in ~80% full charge energy
DRX Detector Battery Charger
• Three slot charger with “Smart” charge capability and visible charge state indication
• Size 38 x 14 x 18 cm
• Weight: 2.26 kg (5 lb.)
• Power Output: 12 V to 16.8 V, Constant Current/Constant Voltage, Lithium
• Ion charge method, 1A max charge current.
• Electrical Ratings 100-240VAC, 50/60 Hz, 1.0A
Wireless Access Point
• Model: D-Link DAP-2553 (for EU)
• Model: D-Link DAP-2553 (for USA)
• Model: D-Link DAP-2553 (for CANADA)
• Model: D-Link DAP-2553 (for KOREA)
• Model: D-Link DAP-2553 (for AUSTRALIA)
• Model: D-Link DAP-2553 (for EGYPT)
• Model: D-Link DAP-2553 (for SINGAPORE)
• Model: D-Link DAP-2553 (for TAIWAN)
• Size 198mm x 120mm x 32mm (7.8” x 4.8” x 1.3”)
• Input 100-240V AC, 50–60 Hz
• Model: Cisco AIR-LAP1252AG-E-K9 (for EU)
• Model: Cisco AIR-LAP1252AG-A-K9 (for USA)
• Model: Cisco AIR-LAP1252AG-N-K9 (for Australia)
DETECTOR / CASSETTE HOLDERS
1014034
1014042
1008705
1008713
Lateral Cassette Holder (for Portable DR panels, CR or Film Cassettes) [QT-LCH]
• Articulating swivel arm (±90°) allows variety of angled positions, along table-top
• Accommodates portable DR, CR, or film cassettes up to 14” X 17”
• Mounts to either side of Table Rails
Lateral Cassette Holder PLUS (for Portable DR panels, CR or Film Cassettes) [QT-LCH-P]
• Assists in acquiring cross-table lateral exams, as well as many other special views
• Provides positioning of receptor/cassette either on or below the table to eliminate image cut-off
• Angulation ability: complete receptor can be angled by 35° for best patient positioning
• Variable Height Adjustable
• Multi-Articulating swivel arm (±90°) allows variety of angled positions
• Accommodates portable DR, CR, or film cassettes up to 14” X 17”
• Mounts to either side of Table Rails
Rolling Lateral Cassette Holder (for Portable DR Panels or CR / Film Cassettes) [QR-LCH]
• Quick and Secure maneuverability around the x-ray exam room
• Articulating swivel arm (±90°) allows variety of angled positions
• Accommodates portable DR, CR, or film cassettes up to 14” X 17”
• Assists in acquiring cross-table lateral exams, as well as many other specialty views
• Provides positioning of image receptor either at or below
• Table top to eliminate image cut-off during cross table examinations
• Height Adjustable positioning with Vertical travel of up to 45.9"
• (116.6 cm); Single-Handed Vertical Adjustment Locking Mechanism
• Receptor Rotation: +/- 90° for more angle flexibility
• Receptor Tilt: 0° through 90°
• Quick Locking 5" Caster Wheels
Table Top Lateral Cassette Holder (for Portable DR Panels or CR / Film Cassettes) [QT-TCH]
• Quick and Easy positioning on the x-ray table
• Accommodates portable DR, CR, or film cassettes up to 14” X 17”
• Quick locking side adjustments hug the imaging plate in place allowing for the imaging plates to be positioned either
horizontally or vertically
• Assists in acquiring cross-table lateral exams, as well as many other specialty views
• Effortless maneuverability with table top scratch protection
Page 7 of 7
Chapter 7 Diagrams
4" [102]
(5) .750 [19] DIAMETER MOUNTING
HOLES 28" [711] APART
28" [711]
61/2" [166]
181/2" [470]
5'-9"
[1753]
35.55 [903]
1'-61/2" [467]
C.G.
52.34 [1329]
±5" [127]
SEE NOTE 1
OF
TUBESTAND
ROTATION
BASE PLATE FOOTPRINT
(SEE DETAIL UPPER
LEFT)
72.00" SID
[1829]
45.50 [1156]
DELUXE
FREESTANDNG
TUBESTAND
QS-550
QUIET -LIFT
ELEVATING,
FLOAT TOP
TABLE
QT-750
92.00
[2337]
SEE NOTE 9
5" [127]
35.55
[903]
5" [127]
3 " [88.9]
TABLE TOP
TO FILM
PLANE
1/2
72.75 MAX.
[1848]
13.00 MIN.
[330]
20.00
[508]
84.00
[2134]
REF
" [19] REF
72.75 MAX
[1848]
2" [49.9]
3/4"
[19]
REF
3/4
40"
[1016]
16.00
[406]
4"
[102]
12.75 [324]
FOCAL
SPOT
85.00
[2159]
83.00
[2108]
16.00
[406]
21.00
[533]
74.25 MAX
[1886]
13.00 MIN
[330]
34.00
[864]
6.00 [152]
27.00
[686]
1.75 [44]
BUCKY TOP
TO FILM
PLANE
2"
[51]
9'-4" [2842]
32.50
[826]
5.00
[127]
.312 [8] DIA
MOUNTING HOLES
40" SID
[1016]
48.00
[1219]
SEE NOTE 4
SEE NOTE 3
13.75 MIN
[349]
SEE NOTE 2
10' [3048]
Figure 7-2. Q-Rad "DS Series" Floor-Mounted
System Dimensions
Q-Rad Radiographic System
Quantum Medical Imaging, LLC
Revision B
7-4
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project

Download PDF

advertising