Brochure - Rainbow Care

Brochure - Rainbow Care
PR-TRENDvent-eng-0610-03.pdf 1 21.01.2011 09:53:14
Dimensions (w x h x l)
330 mm x 110 mm x 280 mm
Mass
4,2 kg
Power Supply
100 V - 250 V AC, 50 - 60 Hz; 24 V DC / 5 A
Internal Battery
Lithium-Ion-Rechageable, 24 V; 2.3 Ah
Average Power Consumption.
52 W
Noise Level
< 30 dB(A) overall
< 25 dB(A) at 10 hPa
Parameters
PCV/APCV
PSV
CPAP
S
ST
T
IPAP
4 - 40 hPa
4 - 40 hPa
2 - 20 hPa
4 - 40 hPa
4 - 40 hPa
4 - 40 hPa
PEEP
0 - 20 hPa
0 - 20 hPa
2 - 20 hPa
2 - 20 hPa
2 - 20 hPa
Frequency
4 - 50 bpm
4 - 50 bpm; Off
4 - 50 bpm
4 - 50 bpm
Insp.Time
0,3 - 8,0 s
EPAP
TRENDvent
Technical Data
Accessories
The functional bag was developed especially for the ventilator TRENDvent.
Through the bag window, the functions of the device can be controlled.
For the accessories an enlargeable compartment with separat space for
the humidifier is integrated. The wide belts care for a good wearing comfort
as backpack or sling bag. Many attachment features at the bag allow an
easy assembly at wheel chair.
The external rechargable battery is intended as autonomous power supply
for all HOFFRICHTER ventilators including TRENDvent.
It was especially developed for homecare use, use in hospitals and effort in
mobile use.
The rechargable battery enables an additonal using time of 12 hours (in
standard adjustments).
0,3 - 8,0 s
Apnoea Limit
Off; 3 - 60 s;
Off; 3-60 s
Auto
[as alarm]
1-5
1-5
1-5
1-5
1 - 5; Auto
1 - 5; Auto
Off; 1 - 5; Auto
Ramp
1-5
Insp.Trigger
Off; 1 - 5; Auto 1 - 5; Auto
3 - 60 s; Auto
Exp.Trigger
10 - 90%; Auto
10 - 90%; Auto 10 - 90%; Auto
Max. Insp.Time
1 - 10 s
1 - 10 s
1 - 10 s
Vt. min.
0.05 - 2 l; Off
0.05 - 2 l; Off
0.05 - 2 l; Off
0.05 - 2 l; Off
0.05 - 2 l; Off
Add. Pressure
3 - 10 hPa
3 - 10 hPa
3 - 10 hPa
3 - 10 hPa
3 - 10 hPa
The remote alarm box is a wire alarm system signalling alarms of the
TRENDvent for home use. The alarm is generated by the ventilator which is
not within the operator’s immediate hearing range. There is no difference
between alarm type and alarm priority in alarm signalling.
The FiO2-Sensor-Set measures the oxygen at the air outlet of the ventilator which shows the actual value in the display of the device.
Alarm Parameters
Max. Volume
0.2 - 2.5 l; Off
0.2 - 2.5 l; Off
0.2 - 2.5 l; Off
0.2 - 2.5 l; Off
0.2 - 2.5 l; Off
0.2 - 2.5 l; Off
Min.Volume
0.01- 2 l; Off
0.01- 2 l; Off
0.01- 2 l; Off
0.01- 2 l; Off
0.01- 2 l; Off
0.01- 2 l; Off
10 - 120 bpm;
10 - 120 bpm;
10 - 120 bpm;
10 - 120 bpm;
10 - 120 bpm;
Off
Off
Off
Off
Off
Max. Frequency 10 - 120 bpm;
Off
Max. Oxygen
30 - 100 %; Off 30 - 100 %; Off 30 - 100 %; Off 30 - 100 %; Off 30 - 100 %; Off 30 - 100 %; Off
Min. Oxygen
18 - 90 %; Off
Pres. Difference 1 - 10 hPa
Leakage
18 - 90 %; Off
18 - 90 %; Off
18 - 90 %; Off
18 - 90 %; Off
18 - 90 %; Off
1 - 10 hPa
1 - 10 hPa
1 - 10 hPa
1 - 10 hPa
1 - 10 hPa
AquaTREND uni, the optional therapy air humidifier is able to be adjusted
quickly with only one click. Easy cleaning and a large filling volume (300 ml)
increases the user friendliness.
An ingenious air-guidance system prevents water from getting into the
therapy device.
20 - 200 l/min; 20 - 200; l/min; 20 - 200; l/min; 20 - 200; l/min; 20 - 200; l/min; 20 - 200; l/min;
Off
Off
Off
Off
Off
Off
Specifications and Performance
Max. Stable Pressure Limit
60 hPa
Max. Working Pressure
40 hPa
Min. Stable Pressure Limit
0 hPa
Min. Working Pressure
0 hPa
Max. Flow
250 l/min
Technical Requirements for Accessories
Oxygen Inlet:
Bacterial Filter:
Type of Connection Port
Quick Coupling
Connections
22 / 15 mm Cone (according to EN 1281-1)
Manufacturer:
Pressure
< 1000 hPa
Resistance
< 2.3 hPa at 60 l/min
Flow
< 15 l/min
CE-Mark required!
HOFFRICHTER GmbH
Mettenheimer Straße 12/14
19061 Schwerin
Germany
Phone: +49 385 39925 0
Fax:
+49 385 39925-25
E-Mail: info@hoffrichter.de
www.hoffrichter.de
Technical specifications are subject to technical modifications.
PR-TRENDvent-eng-0610-03
Sales:
For more noblesse
PR-TRENDvent-eng-0610-03.pdf 2 21.01.2011 09:53:16
TRENDvent
• Usage of Leakage- and Valve Tubing Systems
With an easy switching, the system can be used with a
leakage circuit or a valve circuit.
PCV Pressure Controlled Ventilation
• The ventilation archetype is pressure- and time-controlled.
Airway Pressure
IPAP
controlled
inspiration
controlled
inspiration
controlled
inspiration
PEEP
The wide, big handle guarantees an
easy transport of the device.
PCV - Pressure Controlled Ventilation
• If the trigger is active ((A)PCV), the patient can request
additionnal controlled inspiration breathes.
• Invasive and non invasive ventilation for adults and children
• Leakage- and valve ventilation
• Clear menu structure
• Up to three ventilation presettings
IPAP
Airway Pressure
• Pressure modes with optional volume guarantee
controlled
inspiration
assisted
controlled
inspiration
assisted
controlled
inspiration
spontaneous
breath trigger
spontaneous
breath trigger
All connections for the tubes are
located next to each other on the
frontside of the device.
controlled
inspiration
A PCV - Assisted Pressure Controlled Ventilation
• Extreme low noise level
PSV Pressure Supported Ventilation
• In this mode, normally every breath is triggered by the patient. The exception
is the activation of the background frequency, caused by missing spontaneuos
breathing of the patient.
If the device is operated with a single line patient circuit with an external expiration valve,
the ventilation can be done invasively, e.g. via a tracheostoma and also non invasively via a
mask.
Controlled
Inspiration
Airway Pressure
The well-arranged user interface is very easy to handle. High comfort and maximum safety
are the results in clinical and home use.
The optional humidifier
AquaTREND uni can be clicked
on easyly.
PEEP
• Battery and mains operation
TRENDvent was developed for the pressure controlled respiration therapy and ventilation
of adults and children. Six different ventilation modes are available for an individual therapy
with a minimum breathing volume of 50 ml.
The operating concept impresses
by the first contact. As soon as
the panel is touched, the useable
areas and the display are illuminated. The parameters are adjusted with the Control Pad. The
cleaning of the surface is as easy
as can be. The operating panel is
made of one single piece of glas
without any gaps.
The function of the panel can be
locked for more safety.
IPAP
PEEP
Spontaneous Breath Spontaneous Breath
Trigger
Trigger
PSV - Pressure Supported Ventilation
At the backside of the device, the
connections for the power supply
(external supply), data transmission, a COM-connector, a
connector for the remote alarm
box and an USB-plug for the link
between device and personal
computer, are located.