Thermo Scientific Orion pH Electrodes

Thermo Scientific Orion pH Electrodes
ROSS and the COIL trade dress are trademarks of
Thermo Fisher Scientific Inc.
PerpHecT ROSS electrodes are protected by U.S. patent
6,168,707.
AQUAfast, Cahn, ionplus, KNIpHE, No Cal, ORION,
ORION Series A meters and 900A printer are protected
perpHect, PerpHecT, PerpHecTion, pHISA, pHuture, Pure
by U.S. patents 5,198,093, D334,208 and D346,753.
ROSS Ultra, Sure-Flow, Titrator PLUS and TURBO2 are
registered trademarks of Thermo Fisher.
1-888-pHAX-ION, A+, All-in-One, Aplus, AQUAsnap,
AssuredAccuracy, AUTO-BAR, AUTO-CAL, AUTO
DISPENSER, Auto-ID, AUTO-LOG, AUTO-READ, AUTOSTIR, Auto-Test, BOD AutoEZ, Cable-Free, CERTI-CAL,
CISA, DataCOLLECT, DataPLUS, digital LogR, DirectCal,
DuraProbe, Environmental Product Authority, Extra Easy/
Extra Value, FAST QC, GAP, GLPcal, GLPcheck, GLPdoc,
ISEasy, KAP, LabConnect, LogR, Low Maintenance
Triode, Minimum Stir Requirement, MSR, NISS, OneTouch, One-Touch Calibration, One-Touch Measurement,
Optimum Results, Orion Star, Pentrode, pHuture MMS,
pHuture Pentrode, pHuture Quatrode, pHuture Triode,
Quatrode, QuiKcheK, rf link, ROSS Resolution, SAOB,
SMART AVERAGING, Smart CheK, SMART STABILITY,
Stacked, Star Navigator 21, Stat Face, The Enhanced
Lab, ThermaSense, Triode, TRIUMpH, Unbreakable pH,
Universal Access are trademarks of Thermo Fisher.
Guaranteed Success and The Technical Edge are service
marks of Thermo Fisher.
PerpHecT meters are protected by U.S. patent
6,168,707.
ionplus electrodes and Optimum Results solutions are
protected by U.S. patent 5,830,338.
ROSS Ultra electrodes are protected by U.S. patent
6,793,787.
ORP standard is protected by U.S. patent 6,350,367.
No Cal electrodes with stabilized potential are protected
by U.S. patent 7,276,142.
© 2009 Thermo Fisher Scientific Inc. RDO is a
registered trademark of In-Situ Inc., Fort Collins, CO
USA. Microsoft and Windows are registered trademarks
of Microsoft Corporation in the United States and/or
other countries. All other trademarks are the property of
Thermo Fisher Scientific Inc. and its subsidiaries.
The specifications, descriptions, drawings, ordering
information and part numbers within this document are
subject to change without notice.
This publication supersedes all previous publications on
this subject.
2009 Thermo Scientific Orion Laboratory Products Catalog
Water, Sage, Sensing the Future, SensorLink, ROSS,
C-09ORION-E 0209 Rev. A
Thermo Scientific Orion
Laboratory Products Catalog
pH l ISE l Dissolved Oxygen l Conductivity l Colorimetry
www.thermo.com/water
Environmental Instruments
Water Analysis Instruments
9001:2008
North America
166 Cummings Center
Beverly, MA 01915 USA
Toll Free: 1-800-225-1480
Tel: 1-978-232-6000
Dom. Fax: 1-978-232-6015
Int’l Fax: 978-232-6031
Europe
P.O. Box 254, 3860 AG Nijkerk
Wallerstraat 125K, 3862 BN
Nijkerk, Netherlands
Tel: (31) 033-2463887
Fax: (31) 033-2460832
Asia Pacific
Blk 55, Ayer Rajah Crescent
#04-16/24, Singapore 139949
Tel: 65-6778-6876
Fax: 65-6773-0836
www.thermo.com/water
Thermo Scientific Orion Service Program Options
Our leadership shows up in everything we do including:
A letter of introduction from
Pekka Ronka, Vice President and General Manager
to our valued customers.
• Minimized ramp up time
• Ongoing support
• Flexible data reporting
and analysis
• Low cost of ownership
• Comprehensive support solutions for
all phases of the product life cycle
• Immediate access to experts worldwide
Dear Valued Customer,
I would like to take this opportunity to welcome you to the NEW 2009 Thermo Scientific
Orion Laboratory Products Catalog, and to introduce myself as the new Vice-President &
General Manager of Water Analysis Instruments.
For over 40 years Thermo Scientific Orion products have set the standard for innovation
and quality in electrochemistry. Today, Water Analysis Instruments continue to meet all
your electrochemistry needs with the highest quality products, service, and support.
Service Programs
Flexible service programs provide
fixed-price solutions, offering benefits
over the traditional time and materials,
per-quote based model. So you get
improved ROI, minimized downtime,
unmatched flexibility, better staff
utilization and increased productivity.
Titration Product Service Options
In this catalog you will find a comprehensive selection of meters, electrodes, solutions
and accessories for pH, ion selective electrodes (ISE), conductivity and dissolved
oxygen. We also feature innovative new products for water analysis, including:
•
Orion Star™ RDO® Meters, pages 50-51
Introducing rugged optical dissolved oxygen technology for maintenance-free,
better accuracy measurements.
•
Orion DUAL STAR™ Meters, pages 48-49
The next generation of Orion 720A/920A meters make it easier than ever to
simultaneously measure pH and ISE.
•
Colorimetry and Turbidity Meters, pages 74-85
Ideal for EPA monitoring and reporting of critical water analysis parameters.
•
Green Electrodes, page 18
Go green with environmentally-friendly electrodes that meet RoHS compliance
and contain no lead, mercury or hazardous materials.
•
ROSS® pH/ATC Triode™ Electrodes, page 8
Providing the convenience of built-in temperature with the performance and
accuracy of ROSS technology.
We are excited about the new products and welcome you to contact our technical
experts at (800) 225-1480 or fax (978) 232-6015 with your product questions.
Thank you for your continued support of Thermo Scientific Orion products. I am excited
to be a part of this world–class organization, and look forward to leading the company
into a future of innovation and excellence.
Sincerely,
Pekka Ronka
Vice President & General Manager
Water Analysis Instruments
A variety of service options are now
available for all titration products to
get you up and running quickly and
to make sure you are operating at
the highest performance possible.
Set-up services and training options
are available to minimize ramp up
time so results are expedited rapidly
and accurately. Avoid downtime by
choosing a preventative maintenance
program that meets your high
demands for quality and accurate
results, or select a two or three year
extended warranty to ensure you are
minimizing the possibility of downtime
due to an unforeseen failure.
Orion Star™ Series Service Options,
Calibration and Validation
All Thermo Scientific Orion Products In
Warranty Support
You’ll feel confident your results
are accurate and exact when taking
advantage of our calibration and
validation options. We have a range
of testing, validation and calibration
programs to satisfy even the most
exacting standard. Our dedicated
validation and calibration consultants
offer in-depth expertise and a
full understanding of compliance
requirements. It is recommended that
these options be performed annually,
based on your application demands.
Orion Star series and PerpHecT® meters
carry a three-year warranty that includes
return-to-factory, repair or replace
on a reasonable efforts basis, phone
support, web support, and any software
maintenance releases. Russell meters
include a 2 year warranty.
Stay in specification compliance by
simply sending your Orion Star series
meter to us, where calibration and
validation will be performed to your
choice of standards. We will return
your meter along with a certificate
of validation that includes calibration
details and test results.
IQ/OQ Package
Save time and money when you
purchase the IQ/OQ package to
ensure that your meter performs
to the manufacturer’s published
specifications. Full IQ/OQ
documentation is available for the
Orion Star series meters to perform
a comprehensive installation and
operation on your meter and satisfy
the requirements of regulated
industries such as pharmaceutical.
For product failures in the first thirty
days after shipment, we issue a product
replacement, no questions asked.
Go to thermo.com/water for
comprehensive warranty and service
option information.
Spare Parts
To help you keep your instruments
operating at the highest levels of
accuracy and efficiency, we offer factory
authorized parts at prices comparable to
those of generics. All parts, accessories,
and consumables are performancecertified and fully warranted to perform
to our products’ rigorous design
specifications.
Product Service Options
Calibration Options
Description
Cat. No.
2 Year Extended
Warranty
3 Year Extended
Warranty
Start-up
Preventative
Maintenance
Cat. No.
Star Series Calibration
and Validation
960 Titrator
096000
096000EXT2
096000EXT3
—
—
111200CAL
3-Star pH Benchtop
—
121200CAL
3-Star pH Portable
CBODEZ00PM
111300CAL
3-Star DO Benchtop
121300CAL
3-Star DO Portable
111400CAL
3-Star Cond Benchtop
121400CAL
3-Star Cond Portable
111500CAL
4-Star pH/ISE Benchtop
121500CAL
4-Star pH/ISE Portable
111600CAL
4-Star pH/DO Benchtop
121600CAL
4-Star pH/DO Portable
111700CAL
4-Star pH/Cond Benchtop
121700CAL
4-Star pH/Cond Portable
121800CAL
5-Star pH/DO/Cond Portable
111900CAL
5-Star pH/DO/Cond/ISE Benchtop
121900CAL
5-Star pH/DO/Cond/ISE Portable
STAR-IQOQ
Star Series IQ/OQ Package
960 Titrator
BOD AutoEZ
096002
10060020
096002EXT2
—
096002EXT3
—
—
CBODEZ00SU
Preventative Maintenance (PM) – contracts and Start-up/Training (SU) are available in N. America only and are represented by
Catalog Numbers ending in PM and SU.
Extended Warranties (EXT) – are available world wide and are represented by Catalog Numbers ending in EXT2 (2 year) and EXT3 (3 year).
table of contents
Thermo Scientific Orion Laboratory Products
Thermo Scientific Orion Colorimetry
Thermo Scientific Orion pH Electrodes
pH Electrode Families
2
Colorimetry Theory Overview
pH Electrode Features
3
AQUAfast® IV Advanced Turbidity Meter
76
74-75
pH Electrode Selection Guide
4-5
AQUAfast® AQ3010 Turbidity Meter
77
pH Electrode Maintenance, Storage and Cleaning
6-7
AQUAfast® AQ3070 Chlorine Meter
78
AQUAfast® AQ3700 Colorimeter
79
22
Digestion Test Reagents
80
23
COD Test Systems
81
Silver Billet and Karl Fischer Electrodes
23
AQUAfast®
AQ4000 Colorimeter
82
ATC Probes
24
AQUAfast®
Chemistries Overview
83
Stirrers
25
AQUAfast®
Colorimeter Reagent Chemistries
pH Electrodes
Redox/ORP and
Sure-Flow®
8-21
ORP Triode™
Electrodes
Reference Electrodes
pH Buffers, Solutions and Test Kits
Electrode Connectors and Cables
26-27
28
Thermo Scientific Orion Ion Selective Electrodes (ISE)
Ion Selective Electrode Theory
Customer Services
Cross Reference
86-91
Index
92-96
29
Ion Selective Electrode Selection Guide
30-31
Fluoride Ion Selective Electrodes
32
Ammonia Ion Selective Electrodes
33
Nitrate Ion Selective Electrodes
34
ISE Solutions and Accessories
35
Thermo Scientific Orion Conductivity Cells
Conductivity Cell Families
36-37
Conductivity Cell Selection Guide
38-39
Water Quality Specialist Tip
The following icons are designed to help you locate the
best product for your application.
Best, Better, Good
Electrode ratings based on the electrode
precision
Best Seller
Our most popular customer products
39
Mercury-Free
Environmentally safe, mercury-free
pH electrodes
RDO® Optical Dissolved Oxygen Probes
40
Dissolved Oxygen Probe Families
41
TRIS Compatible
pH electrodes for use with TRIS,
protein and sulfide samples (calomel
replacements)
Conductivity Standards and Accessories
Thermo Scientific Orion Dissolved Oxygen Probes
Dissolved Oxygen Probe Selection Guide
Dissolved Oxygen Solutions and Accessories
42-43
43
Thermo Scientific Orion Meters
Meter Overview
44-45
Meter Selection Guide
46-47
Orion DUAL
STAR™
Meter
48-49
Orion Star™ Series Meters and Kits
Star and Star Plus Navigator
PerpHecT®
LogR™
21™
Meters
Software
Superior Temperature Control
pH electrodes that respond exceptionally
well to sample temperature changes
Lead-Free
Environmentally safe electrodes that
use lead-free glass, which exceeds the
requirements of RoHS
Pb
Compliant with EPA testing methods
50-68
69
Compliant with USP requirements
70
Intrinsically Safe Meters
71
Meter Accessories
72
AutoTration™
73
- 500 Autosampler
84-85
Hazardous Materials
Additional shipping charges may apply
2009 Sourcebook
1
pH electrodes
Thermo Scientific Orion pH Electrode Families
The best selection of pH electrodes
Page
ROSS
Ultra®
Premium pH Electrodes
8
The best of the best! The fastest response, highest
accuracy and best reproducibility that you can get
with a pH electrode – plus an industry best warranty.
ROSS® Premium pH Electrodes
13
Electrodes designed to measure samples
as small as 0.5 μL in containers as small as
384 well plates.
2
Thermo Scientific Orion Laboratory Products Catalog
Double Junction pH Electrodes
20
20
Isolated Ag/AgCl reference system that prevents silver
from coming in contact with the sample. Great for
measuring TRIS buffer, sulfide and protein samples.
Economy/Basic pH Electrodes
21
Value-priced, durable and low maintenance gel-filled
pH electrodes.
Redox/ORP
22
The ideal choice for measuring the oxidationreduction potential of samples and performing
redox titrations.
16
ATC Probes
18
Green pH electrodes contain no lead, mercury or
other hazardous substance, which makes these
electrodes environmentally friendly, allows for easy
and hassle-free electrode disposal and meets all the
requirements of RoHS.
Micro Ag/AgCl pH Electrodes
Unique reference system that provides quick,
accurate measurements. Great for measuring TRIS
buffer, sulfide and protein samples. No calibration
required.
15
A large variety of quality electrodes for a wide range
of applications. Includes specialty pH electrodes for
unique or challenging pH measurements.
Green pH Electrodes
20
The pH electrode is surrounded by a stainless blade
for measuring meat, cheese, and sludge samples.
Low maintenance polymer-filled double junction
electrodes with an extended life, fast response and
clog-resistant junction.
Standard Ag/AgCl pH Electrodes
pH Electrodes
KNIpHE® pH Electrodes
Designed to provide the best pH and temperature
measurements when used with the 320, 350 or
370 PerpHecT digital LogR™ meters. Includes the
PerpHecT ROSS micro pH electrode, which can
measure samples as small as 15 μL in 384 well plates;
is TRIS buffer, protein and sulfide compatible; and can
be used on any pH meter with a BNC connection.
AquaPro Professional pH Electrodes
No
10
Precise pH measurements with unmatched
stability, reliability and rapid response regardless of
temperature or sample composition.
PerpHecT® pH Electrodes
Page
Cal®
24
Automatic temperature compensation
probes measure sample temperatures and
adjust pH measurements by correcting the
electrode slope according to the measured
temperature.
Stirrer Probe
25
Convenient and economical answer to sample stirring.
Powered by Star benchtop meters.
19
pH electrodes
Thermo Scientific Orion pH Electrode Features
Select features to meet the needs of your applications
Electrode Styles
Standard Size
Semi-micro
Micro
Rugged Bulb
Spear Tip
Flat Surface Tip
12 mm electrode
diameter for use in
a wide variety of
sample sizes
6 to 8 mm electrode
diameter for sample
sizes down to 200 μL
1 to 5 mm electrode
diameter for samples
as small as 0.5 μL
& containers as small
as 384 well plates
When glass is
required, the extra
durable pH bulb
prevents breakage
For piercing solid or
semi solid samples
and measuring small
volume samples
For measuring
surfaces of solid
or gel samples and
measuring small
volume samples
Electrode Body Materials
Glass Body
• Easy to clean
• Compatible with virtually any sample,
including solvents
Epoxy Body
• Extremely durable and rugged
to prevent breakage
• Value-priced
Electrode Fill Types
Refillable
Polymer Filled (Sealed)
Gel Filled (Sealed)
• Easy maintenance, filling
solution required
• Low maintenance, no filling solution
needed
• Low maintenance, no filling
solution needed
• Periodic filling and draining needed
• Sealed reference
• Sealed reference
• Long expected life span
• Easy to use
• Easy to use
• Suited for a wide range of applications
• Suited for a wide range of applications
• General purpose for everyday use
• 0.01 to 0.02 pH precision
• 0.02 pH precision
• 0.05 to 0.1 pH precision
• Best response time
• Better response time
• Good response time
• 1 to 2 year warranty
(6 months for Ag/AgCl micro electrodes)
• 1 year warranty
• 3 to 6 month warranty
Electrode References
ROSS® Reference
Double Junction Ag/AgCl Reference
Single Junction Ag/AgCl Reference
• 0.01 pH precision
• 0.02 pH precision
• 0.02 to 0.1 pH precision
• Best measurement response time
• Better measurement response time
• Good measurement response time
• Ideal for TRIS, protein and sulfide samples
• Ideal for TRIS, protein and sulfide samples
• General purpose for everyday use
• Variety of body styles and types
• Variety of body styles and types
• Variety of body styles and types
• Refillable or gel design
• Refillable, polymer or gel design
• Refillable or gel design
• Best temperature response
• Good temperature response
• Good temperature response
Electrode Junctions
Sure-Flow®, Sleeve and Open
Ceramic and Glass Capillary
Wick and Glass Fiber
• Best junction for dirty, difficult samples
and demanding lab use
• Better junction for routine lab use in
a variety of samples
• Good junction for routine lab or field use
in buffers, water and aqueous samples
• Junction is clog-free and easy to clean
• Junction is high quality and durable
• Junction with rugged epoxy electrodes
• Ideal for thick or viscous samples,
compatible with all sample types
• Ideal for most lab applications and
sample types
• Ideal for basic lab applications and
aqueous samples
2009 Sourcebook
3
pH electrodes
Thermo Scientific Orion pH Electrode Selection Guide
Sample Type
Electrode
Recommendations
ROSS Ultra®
ROSS®
0.01
0.01
Electrode Rating
pH Precision
BEST
BETTER
Biological/Pharmaceutical –
TRIS buffer, proteins, enzymes
Electrodes should have a ROSS or double junction
Ag/AgCl reference (no sample contact with silver)
8102BNUWP
8107BNUMD
8156BNUWP
8157BNUMD
8102BN, 8104BN
815600, 8165BNWP
8172BNWP
Education/Student Use
Electrodes should have an epoxy body for
added durability
8107BNUMD
8156BNUWP
8157BNUMD
815600
8165BNWP
Emulsions – Foods, cosmetics, oils
Electrodes should have a Sure-Flow or open junction to
prevent the electrode from clogging
8165BNWP
8172BNWP
Emulsions – Petroleum products, paint
Electrodes should have a glass body that resists damage
from the sample and a Sure-Flow or open junction to
prevent the electrode from clogging
8172BNWP
Extreme pH – Acid with fluoride
Electrodes must be resistant to hydrofluoric acid
*930101 module requires 9300BNWP handle and 900200
reference electrode
Flat Surfaces – Cheese, meat, agar
Electrodes should have a flat-surface tip and ROSS or
double junction Ag/AgCl reference (no sample contact
with silver)
8135BNUWP
8135BN
Flat Surfaces – Paper
Electrodes should have a flat-surface tip
8135BNUWP
8135BN
General Purpose – Most sample types
All electrodes are suitable for general
purpose measurements
8102BNUWP
8107BNUMD
8156BNUWP
8157BNUMD
8102BN, 8104BN
815600, 8165BNWP
8172BNWP
Harsh Environments –
Field or plant use, rugged use
Electrodes should have an epoxy body for
added durability and be polymer or gel filled for
easy maintenance
8107BNUMD
High Ionic Strength –
Acids, bases, brines, pH > 12 or pH < 2
Electrodes should have a Sure-Flow or open junction
for better contact with the sample and more stable
measurements
Large Sample Sizes – Tall flasks
Electrodes should have a long body that fits the container
Low Ionic Strength – Treated effluent,
deionized water, distilled water
Electrodes should be refillable for better contact with the
sample and more stable measurements
Non-aqueous – Solvents, alcohols
Electrodes should have a glass body that resists damage
from the sample and a Sure-Flow junction for better
contact with the sample and more stable measurements
8172BNWP
Semi-solids – Fruit, meat, cheese
Electrodes should have a spear tip for piercing samples
and a ROSS or double junction Ag/AgCl reference
8163BNWP
Small Sample Size – Micro-titer plates
Electrodes should have a small diameter that fits
the container
Small Sample Size – NMR tubes
Electrodes should have a small diameter that fits
the container
Small Sample Size –
Test tubes, small flasks and beakers
Electrodes should have a small diameter that fits
the container
8103BNUWP
8115BNUWP
8103BN, 8115BN
8175BNWP
Small Sample Size –
TRIS buffer, proteins, sulfides
Electrodes should have a small diameter that
fits the container and a ROSS or double junction
Ag/AgCl reference
8103BNUWP
8115BNUWP
8103BN, 8115BN
8175BNWP
Titration
Electrodes should have a Sure-Flow or sleeve junction
for better contact with the sample and more stable
measurements
8162SC, 8165BNWP
8172BNWP
Viscous Liquids –
Slurries, suspended solids sludges
Electrodes should have a Sure-Flow or open junction to
prevent the electrode from clogging
8165BNWP
8172BNWP
Waters – Acid rain, boiler feed water,
distilled water, rain water, well water
Electrodes should have a ROSS or double junction
Ag/AgCl reference and be refillable for better contact
with the sample
8102BNUWP
8156BNUWP
8157BNUMD
8102BN, 815600
8165BNWP
8172BNWP
Waters – Drinking water, tap water
Electrodes should have an epoxy body for
added durability
8156BNUWP
8157BNUMD
815600
8165BNWP
Waters – Wastewater, seawater
Electrodes should have a ROSS or double junction
Ag/AgCl reference and have an epoxy body for
added durability
8156BNUWP
8157BNUMD
815600
8165BNWP
GOOD
4
Thermo Scientific Orion Laboratory Products Catalog
8165BNWP
8172BNWP
8102BNUWP
8156BNUWP
8157BNUMD
8102BN, 815600
8165BNWP
8172BNWP
pH electrodes
PerpHecT®
AquaPro
Standard
Ag/AgCl
Green
Micro
Ag/AgCl
Double
Junction
Economy/
Basic
0.01 to 0.05
0.02
0.02
0.02 to 0.05
0.02
0.02
0.05 to 0.1
8202BN, 8207BNU
8256BN, 8257BNU
8272BN
9102AP, 9104APWP
9107APMD
9156APWP
8207BNU, 8256BN
8257BNU, 9206BN
9207BN, 9256BN
9107APMD
9156APWP
9107BNMD
9156BNWP
9157BNMD
8272BN
9272BN
9104APWP
9107APMD
9165BNWP
9172BNWP
8272BN
9272BN
9102AP
9104APWP
9172BNWP
GD9106BNWP
GD9156BNWP
9102DJWP
9156DJWP
GS9106BNWP
GD9106BNWP
9156DJWP
9106BNWP
930101*
8235BN
9135APWP
8235BN
9135APWP
9167SC
8202BN, 8207BNU
8256BN, 8257BNU
9202BN, 9206BN
9207BN, 9256BN
9102AP, 9104APWP
9107APMD
9156APWP
9102BNWP
9107BNMD
9156BNWP
9157BNMD
GS9106BNWP
GD9106BNWP
GS9156BNWP
GD9156BNWP
8207BNU, 9206BN
9207BN
9107APMD
9156APWP
9107BNMD
GS9106BNWP
GD9106BNWP
8272BN
9272BN
9104APWP
9107APMD
9102AP
9165BNWP
9172BNWP
913600
9802BN
9102DJWP
9156DJWP
9106BNWP
9156DJWP
9106BNWP
912600
8202BN, 8256BN
8257BNU, 8272BN
9256BN, 9272BN
9165BNWP
9172BNWP
8272BN
9272BN
9172BNWP
GS9156BNWP
GD9156BNWP
9102DJWP
9120APWP
8220BNWP
9802BN, 9803BN
9810BN
8220BNWP
9826BN
8203BN, 9203BN
8220BNWP
9103APWP
9115APWP
8203BN
8220BNWP
9103APWP
9115APWP
8272BN
9272BN
8272BN, 9272BN
9103BNWP
9167SC
9802BN, 9803BN
9810BN, 9826BN
9110DJWP
911600, 912600
9110DJWP
9166SC
9104APWP
9107APMD
8202BN, 8256BN
8257BNU, 8272BN
9165BNWP
9172BNWP
9165BNWP
9172BNWP
GS9156BNWP
GD9156BNWP
9102DJWP
8256BN, 8257BNU
9256BN
9107APMD
9156APWP
9156BNWP
9165BNWP
GS9106BNWP
GD9106BNWP
9156DJWP
8256BN, 8257BNU
9256BN
9107APMD
9156APWP
9165BNWP
GD9106BNWP
GD9156BNWP
9156DJWP
9106BNWP
2009 Sourcebook
5
pH electrodes
Thermo Scientific Orion pH Electrode Maintenance,
Storage and Cleaning
pH Electrode Maintenance
On a weekly basis, inspect the pH electrode for scratches,
cracks, salt build-up, or membrane/junction deposits.
Rinse off any salt build-up with distilled water. Remove any
membrane/junction deposits by soaking the electrode in a
0.1 M HCl or 0.1 M HNO3 solution for 15 minutes or soaking
the electrode in a 0.1 M KCl solution heated to 55 °C for 15
minutes. If the electrode is dirty, clogged or coated, refer
to the pH Electrode Cleaning section for a more thorough
electrode cleaning procedure.
If a refillable electrode is used, drain the reference chamber,
flush it with distilled water until all of the salt build-up inside of
the electrode is removed, flush it with fresh filling solution and
fill the reference chamber with fresh filling solution.
Soak electrode in ROSS pH electrode storage solution,
Cat. No. 810001, or standard pH electrode storage solution,
Cat. No. 910001, for 1 to 2 hours or overnight. The ROSS pH
electrode storage solution is recommended for enhanced
electrode performance.
To minimize the chance of breakage, the pH electrode should
be attached to an electrode stand and suspended in a beaker
that contains storage solution. The electrode should not touch
the sides or bottom of the beaker.
Long Term Electrode Storage (more than one week)
If the electrode is refillable, fill the reference chamber up to
the fill hole with the appropriate electrode filling solution and
securely cover the fill hole with parafilm. Cover the pH-sensing
bulb and reference junction with a protective cap, sleeve or
storage bottle that contains storage solution. Before returning
the electrode to use, prepare it as a new electrode.
Typical pH Electrode Components
Cap
Fill Hole
pH Electrode Storage
To ensure a quick electrode response and unclogged electrode
junction, the electrode should never be stored dry and the pHsensing bulb and reference junction must not dry out. Always
store the pH electrode in pH electrode storage solution.
Short Term Electrode Storage (up to one week)
Soak the electrode in ROSS pH electrode storage solution,
Cat. No. 810001, or standard pH electrode storage solution,
Cat. No. 910001. For enhanced electrode performance, use
ROSS pH electrode storage solution.
6
Cat. No.
Description
810001
ROSS® pH electrode storage solution, 475 mL bottle
910001
Standard pH electrode storage solution, 475 mL bottle
910003
Storage bottles for electrodes with a 12 mm diameter,
3 pack
910004
Storage bottles for electrodes with an 8 mm diameter, 3 pack
(semi-micro epoxy pH electrodes)
910006
Storage bottles for electrodes with a 6 mm diameter, 3 pack
(semi-micro glass pH electrodes)
090043
Swing arm electrode stand
Thermo Scientific Orion Laboratory Products Catalog
Internal
Reference
pH-sensing
Bulb
Reference
Filling
Solution
Chamber
Reference
Junction
pH electrodes
pH Electrode Cleaning
Good Laboratory Practices (GLP)
One of the most common reasons for a pH electrode to not
work properly is because it is dirty, clogged or coated with
sample. Cleaning a dirty, clogged or coated electrode restores
proper electrode performance and prolongs the useful life of
the electrode.
For the most accurate measurement results, follow good
laboratory practices.
• Proper Maintenance, Storage and Cleaning – Proper
electrode maintenance, storage and cleaning maximizes
the electrode performance and extends the electrode life.
The Thermo Scientific pH electrode cleaning solutions are
designed to simplify pH electrode maintenance. The readyto-use cleaning solutions include a small beaker to hold the
cleaning solution and a plastic pipet for removing the electrode
filling solution.
• Fill Hole Cover and Filling Solution Level – If a refillable
electrode is used, uncover the fill hole during calibration
and measurement and cover the fill hole during storage.
The filling solution level must be at least 1 inch higher
than the sample level.
Optimal cleaning procedures are dependent upon the
sample type, extent of build-up or clogging and the type of
electrode. The following instructions provide a starting point for
developing an effective cleaning procedure.
• Rinsing and Stirring – Rinse the electrode with distilled
water between buffers and samples to prevent
contamination and blot the electrode with a lint-free
tissue (do not wipe or rub the electrode) to remove
excess water. Stir all buffers, standards and samples at a
moderate, uniform rate for accurate measurements and
improved response time.
1. Choose an appropriate cleaning solution. Cleaning solution
D is a mild cleaning solution. If a stronger cleaning solution
is required, use cleaning solution C. Cleaning solution A is
for removing protein deposits and cleaning solution B is for
removing bacterial contaminants.
2. Shake the cleaning solution. Pour enough of the cleaning
solution into the beaker to cover the reference junction.
3. Soak the electrode for five to ten minutes in the cleaning
solution while moderately stirring the solution. Electrodes
with wick and fiber junctions may require more cleaning
time.
4. Remove the electrode from the cleaning solution and rinse
it thoroughly with distilled water to remove all traces of the
cleaning solution.
5. If the electrode is refillable, remove the filling solution from
the electrode using the pipet that is included with the kit
and add fresh filling solution to the electrode. Repeat this
step two or three times for optimal electrode performance.
6. If the electrode is a Sure-Flow® electrode, flush a few drops
of filling solution through the reference junction by pressing
down on the electrode cap. Ensure that the junction resets
properly. Refill the electrode with filling solution.
7. Rinse the electrode thoroughly with distilled water and
measure samples as usual. If the electrode response is
slow or the electrode does not calibrate correctly, repeat
the cleaning procedure. Viscous samples and samples that
contain solid materials often require additional cleaning and
additional filling solution changes.
• Temperature – The pH value of a solution will change if
the temperature of the solution changes. To account for
pH slope and pH buffer value changes due to temperature
variations, use an automatic temperature compensation
(ATC) probe or manual temperature compensation.
• pH Buffers – Keep pH buffer bottles tightly sealed and
always pour fresh aliquots of buffers into clean beakers
for calibration. Verify that buffers are within the expiration
date before using them – unopened buffers have a two
year shelf life and open buffers have a 2 to 3 month shelf
life (1 month for pH 10 buffer).
• Calibration – Electrode calibration verifies the slope and
proper function of the electrode. Calibrate the electrode
daily with at least two buffers that bracket the expected
sample pH range and are one to four pH units apart.
Cat. No.
Description
900020
pH electrode cleaning solution kit, includes 1 x 30 mL bottle
each of cleaning solution A and C, 1 x 60 mL bottle each of
cleaning solution B and D, beaker and pipet
900021
pH electrode cleaning solution A, for removing protein
contaminants, includes 4 x 30 mL bottles, beaker and pipet
900022
pH electrode cleaning solution B, for removing bacterial
contaminants, includes 4 x 60 mL bottles, beaker and pipet
900023
pH electrode cleaning solution C, for general cleaning, includes
4 x 30 mL bottles, beaker and pipet
900024
pH electrode cleaning solution D, for removing oil and grease
contaminants, includes 4 x 60 mL bottles, beaker and pipet
2009 Sourcebook
7
pH electrodes
Thermo Scientific Orion ROSS Ultra® Premium pH Electrodes
Superior stability, measurement speed, accuracy and reproducibility
under temperature variation
Accurate, rapid
results in a halfcell design
The Best of the Best!
The benefits in the ROSS Ultra line cannot
be found in any other pH electrode line.
Consider the advantages before you
purchase your next pH electrode.
BEST
MERCURY
FREE
TRIS
COMPATIBLE
SUPERIOR
TEMPERATURE
CONTROL
Low
maintenance
with built in ATC
ROSS Ultra electrode technology offers
such great stability and drift improvements
that the warranty period is double other
electrodes. The refillable electrodes have
a two year replacement warranty. The gelfilled ROSS Ultra Triode electrodes have an
18 month warranty.
NEW
NEW
Complex samples, such as biological
media, foods and pharmaceuticals, can be
measured easily. ROSS technology can be
used where metallic reference electrodes
would contaminate the sample. All ROSS
Ultra electrodes can be used in samples
that contain TRIS, sulfides or proteins.
ROSS Ultra electrodes offer the
best temperature performance of all
other electrodes in repeated, varying
temperatures.
• Precise, durable and low maintenance
with built in temperature
• Top performance, with a rugged
design and built in temperature
Cat. No.
pH Range
pH Precision
Temp. Range
Temp. Accuracy
Internal Reference
Recommended Accessories
810001
ROSS Storage Solution
810007
ROSS Fill Solution
810199
All-in-One ROSS pH Buffer Kit
Thermo Scientific Orion Laboratory Products Catalog
ROSS Ultra®
pH/ATC Triode
with epoxy body
• Use with
8101BNWP or
8101SC ROSS
pH half-cell or
8411BN ROSS
sodium half-cell
electrodes
• Built-in ATC with
rapid response
800500U
ROSS Ultra®
pH/ATC Triode with
epoxy body, low
maintenance gel
• Low maintenance
gel with built-in ATC
• General purpose,
high performance
8157BNUMD (Star)*
8157BNU (A+ Series)
8257BNU (PerpHecT)
• Rugged epoxy body
8107BNUMD (Star)*
8107BNU (A+ Series)
8207BNU (PerpHecT)
0-14
0-14
0-14
-
0.01
0.01
0-100 °C
0-100 °C
0-100 °C
-
± 1.0 °C
± 1.0 °C
ROSS
ROSS
ROSS
Ceramic
Glass Fiber
Glass Capillary
Fill Solution
3 M KCl
(810007)
3 M KCl
(810007)
-
L- 120 mm
D- 12 mm
L- 120 mm
D- 12 mm
L- 120 mm
D- 12 mm
Pin Tip
BNC - MiniDIN (Star)
BNC - 8 Pin (A+ Series)
BNC - PhonoTip (PerpHecT)
BNC - MiniDIN (Star)
BNC - 8 Pin (A+ Series)
BNC - PhonoTip (PerpHecT)
Connector Type
Cat. No.
ROSS Ultra®
reference half-cell
with glass body
Junction
Dimension
D (Dia) L (Length) T (Tip)
8
Convenience
with built in ATC
pH
pHelectrodes
electrode
Ideal for QC
samples with
size limitations
Top performance
for QC and
research
Precise yet
durable, for QC
and research
For aqueous
samples with
size constraints
Ideal for soft
solid and semisolid samples
Top performance –
precise pH analysis
with durability
BEST
MERCURY
FREE
TRIS
COMPATIBLE
SUPERIOR
TEMPERATURE
CONTROL
ROSS Ultra®
combination pH
electrode with
glass body
• For precise pH
determinations
• General
purpose, high
performance
ROSS Ultra®
combination pH
electrode with
glass body, semimicro tip
ROSS Ultra®
combination
pH electrode
with glass body,
rugged bulb
ROSS Ultra®
combination pH
electrode with
epoxy body,
semi-micro tip
ROSS Ultra®
combination pH
electrode with
epoxy body, flat
surface
• Measures
samples as
small as 0.2 mL
• Toughened bulb
for rugged lab
use
• Measures
samples as
small as 0.2 mL
• Flat surface for
measurements
of moist
surfaces such as
agar gel plates,
meat, and
cheese
• Fits test tubes
and small
containers
• Epoxy body for
ruggedness and
durability
ROSS Ultra®
combination pH
electrode with
epoxy body
• Epoxy body for
ruggedness
• General
purpose, high
performance
8102BNUWP
8103BNUWP
8104BNUWP
8115BNUWP
8135BNUWP
8156BNUWP
Cat. No.
0-14
0-14
0-14
0-14
0-14
0-14
pH Range
0.01
0.01
0.01
0.01
0.01
0.01
pH Precision
0-100 °C
0-100 °C
0-100 °C
0-100 °C
0-100 °C
0-100 °C
Temp. Range
-
-
-
-
-
-
Temp. Accuracy
ROSS
ROSS
ROSS
ROSS
ROSS
ROSS
Internal Reference
Ceramic
Ceramic
Ceramic
Glass Fiber
Glass Fiber
Glass Fiber
Junction
3 M KCl
(810007)
3 M KCl
(810007)
3 M KCl
(810007)
3 M KCl
(810007)
3 M KCl
(810007)
3 M KCl
(810007)
Fill Solution
L- 120 mm
D- 12 mm
L- 165 mm
T- 6 mm x 95 mm L
L- 120 mm
D- 12 mm
L- 165 mm
T- 8 mm x 95 mm L
L- 120 mm
D- 12 mm
L- 120 mm
D- 12 mm
Dimension
D (Dia) L (Length) T (Tip)
BNC Waterproof
BNC Waterproof
BNC Waterproof
BNC Waterproof
BNC Waterproof
BNC Waterproof
Connector Type
Key Information
Environmentally-safe mercury-free option for TRIS, protein and sulfide.
All cap diameters are 16 mm at bottom of cap. All cable lengths are 1 meter unless specified.
We offer more connector options; contact your local sales office for information.
BNC Waterproof
Pin Tip
MiniDIN (Star)
8 Pin (A+ Series) Phono Tip (PerpHecT)
*8157UWMMD and 8107UWMMD electrodes have 3 meter cable lengths and BNC - MiniDIN
connectors for Star meters.
2009 Sourcebook
9
pH electrodes
Thermo Scientific Orion ROSS® pH Electrodes
The standard for accurate, reproducible pH measurements
ROSS Specifications :
• Slope: 92-102% of theoretical
Nernst slope
Accurate, rapid
results in a halfcell design
For precise and
reproducible pH
determinations
ROSS® pH halfcell with glass
body
ROSS®
combination pH
electrode with
glass body
• Isopotential point: pH 7
BEST
MERCURY
FREE
TRIS
COMPATIBLE
• Accuracy of measuring a pH 6.86
buffer after standardization at 25 °C:
Accurate within 0.03 pH for buffer
at any temperature between
0-100 °C using automatic temperature
compensation.
• Speed of response in 6.86 buffer going
from 25 °C to 75 °C: Response stable
to 0.01 pH within 30 seconds
• Speed of response between 6.86
and 4.01 buffers at 25 °C: Response
stable to 0.002 pH within
15 seconds
• Mercury free, TRIS, protein and
sulfide compatible
SUPERIOR
TEMPERATURE
CONTROL
Temperature Response
The graph below shows how ROSS
electrodes respond versus the best
of conventional pH electrodes. The
ROSS electrode continues to show fast
reproducibility and accuracy after many,
dramatic temperature changes.
• Use with
800300 ROSS or
800500U ROSS
Ultra reference
electrodes
pH
4.19
• General
purpose, high
performance
• For QC and
research
applications
Conventional Electrode @ 75˚C
ROSS Electrode
4.13
Cat. No.
Theoretical Response
8101BNWP
8102BN
pH Range
0-14
0-14
pH Precision
0.01
0.01
Temp. Range
0-100 °C
0-100 °C
ROSS
ROSS
Junction
-
Ceramic
Fill Solution
-
3 M KCl
(810007)
L- 120 mm
D- 12 mm
L- 120 mm
D- 12 mm
BNC Waterproof*
BNC*
4.07
25˚C
4.01
Internal Reference
Minutes 1
2
3
4
5
6
Temperature Response of the ROSS Ultra® and
ROSS® Electrodes vs. Conventional Electrodes
7
Dimension
D (Dia) L (Length) T (Tip)
Connector Type
10
Cat. No.
Recommended Accessories
810001
ROSS Storage Solution
810007
ROSS Fill Solution
810199
All-in-One ROSS pH Buffer Kit
Thermo Scientific Orion Laboratory Products Catalog
pH electrodes
Fast, reliable
results in small
samples
Durable, for
precise pH
analysis
Dependable
results in small
samples
For soft solid
and semi-solid
samples
Durable for
precise pH
applications
For piercing
semi-solid
samples
BEST
MERCURY
FREE
TRIS
COMPATIBLE
SUPERIOR
TEMPERATURE
CONTROL
ROSS®
combination
pH electrode
with glass body,
semi-micro tip
ROSS®
combination
pH electrode
with glass body,
rugged bulb
ROSS®
combination pH
electrode with
epoxy body, semimicro tip
ROSS®
combination pH
electrode with
epoxy body, flat
surface
• Measures
samples as
small as 0.2 mL
• Toughened bulb
for rugged lab
use
• Epoxy body for
ruggedness and
durability
• Fits test tubes
and small
containers
• For QC and
research
applications
• Measures
samples as
small as 0.2 mL
• Flat surface for
measurements
of moist
surfaces such
as agar gel
plates, meat,
and cheese
ROSS®
combination pH
electrode with
epoxy body
• Epoxy body for
ruggedness
• General
purpose, high
performance
ROSS®
combination pH
electrode with
glass body,
spear tip
• For cheese,
meat, and fruit
samples
• Measure
samples as
small as 100 μL
8103BN
8104BN
8115BN
8135BN
815600
8163BNWP
Cat. No.
0-14
0-14
0-14
0-14
0-14
0-14
pH Range
0.01
0.01
0.01
0.01
0.01
0.01
pH Precision
0-100 °C
0-100 °C
0-100 °C
0-100 °C
0-100 °C
0-100 °C
Temp. Range
ROSS
ROSS
ROSS
ROSS
ROSS
ROSS
Internal Reference
Ceramic
Ceramic
Glass Fiber
Glass Fiber
Glass Fiber
Ceramic
Junction
3 M KCl
(810007)
3 M KCl
(810007)
3 M KCl
(810007)
3 M KCl
(810007)
3 M KCl
(810007)
3 M KCl
(810007)
Fill Solution
L- 165 mm
D- 6 mm x 95 mm L
L- 120 mm
D- 12 mm
L- 165 mm
T- 8 mm x 95 mm L
L- 120 mm
D- 12 mm
L- 120 mm
D- 12 mm
L- 110 mm
T- 4.5 mm x 20 mm L
Dimension
D (Dia) L (Length) T (Tip)
BNC*
BNC*
BNC*
BNC*
BNC*
BNC Waterproof*
Connector Type
Key Information
Environmentally-safe mercury-free option for TRIS, protein and sulfide.
All cap diameters are 16 mm at bottom of cap. All cable lengths are 1 meter.
* We offer more connector options; contact your local sales office for information.
BNC Waterproof
BNC
2009 Sourcebook
11
pH electrodes
Thermo Scientific Orion ROSS® Sure-Flow® pH Electrodes
Easy to clean junction never clogs
Sure-Flow® Junction
BEST
MERCURY
FREE
The unique, free-flowing liquidto-liquid junction assures you
of the most stable, drift-free
measurements. The easy-to-clean
junction never clogs, simply press
the cap and flush the junction
area. Release the cap and the
junction is reset. Now even the
most problematic, dirty or viscous
samples can be easily measured
without a clogged junction!
Fast response–
even in difficult
samples
High
performance in
all samples
For difficult
samples with
size limitations
Accurate, rapid
results in a halfcell design
ROSS®
Sure-Flow®
combination pH
electrode with
epoxy body
ROSS®
Sure-Flow®
combination pH
electrode with
glass body
ROSS® Sure-Flow®
reference half-cell
with glass body
• Sure-Flow
junction prevents
clogging, ideal for
soil samples
• Sure-Flow
junction
prevents
clogging
ROSS® SureFlow®
combination pH
electrode with
epoxy body,
semi-micro tip
• Durable epoxy
design
• For soil, sludge,
colloids, viscous
material and
organic solvents
TRIS
COMPATIBLE
SUPERIOR
TEMPERATURE
CONTROL
Cat. No.
• Use with
8101BNWP or
8101SC ROSS pH
half-cell or 8411BN
ROSS sodium halfcell electrodes
• Sure-Flow
junction prevents
clogging
• Measure
samples as small
as 0.2 mL
8165BNWP
8172BNWP
8175BNWP
800300
pH Range
0-14
0-14
0-14
0-14
pH Precision
0.01
0.01
0.01
-
Temp. Range
0-100 °C
0-100 °C
0-100 °C
0-100 °C
-
-
-
-
ROSS
ROSS
ROSS
ROSS
Sure-Flow
Sure-Flow
Sure-Flow
Sure-Flow
3 M KCl
(810007)
3 M KCl
(810007)
3 M KCl
(810007)
3 M KCl
(810007)
L- 120 mm
D- 12 mm
L- 120 mm
D- 12 mm
L- 165 mm
T- 8 mm x 95 mm L
L- 120 mm
D- 12 mm
BNC Waterproof*
BNC Waterproof*
BNC Waterproof
Pin Tip
Temp. Accuracy
Internal Reference
Junction
Fill Solution
Dimension
D (Dia) L (Length) T (Tip)
Connector Type
12
• Sure-Flow junction
prevents clogging
Cat. No.
Recommended Accessories
810001
ROSS Storage Solution
810007
ROSS Fill Solution
810199
All-in-One ROSS pH Buffer Kit
Thermo Scientific Orion Laboratory Products Catalog
pH electrodes
Thermo Scientific Orion PerpHecT® ROSS® pH Electrodes
The ideal electrodes for PerpHecT® meters for LogR™ temperature measurements
Ideal for precise
and reliable pH
analysis
Precise
measurements of
small samples
Accurate in
extremely
small samples
For soft solid
and semi-solid
samples
Durability and
reliability for
precise pH
High
performance
in all samples
BEST
MERCURY
FREE
TRIS
COMPATIBLE
SUPERIOR
TEMPERATURE
CONTROL
PerpHecT®
ROSS®
combination pH
electrode with
glass body
PerpHecT®
ROSS®
combination
pH electrode
with glass body,
semi-micro tip
• For QC and
research
applications
• Measure
samples as
small as 0.2
mL
• Fits test tubes
and small
containers
PerpHecT®
ROSS®
combination pH
electrode with
glass body, micro
• Measures
samples as small
as 15 μL in 384
well plates
• Minimum depth
of immersion is
4.5 mm
PerpHecT®
ROSS®
combination pH
electrode with
epoxy body, flat
surface
PerpHecT®
ROSS®
combination
pH electrode
with epoxy
body
PerpHecT®
ROSS®
Sure-Flow®
combination pH
electrode with
glass body
• Flat surface for
measurements
of moist
surfaces such as
agar gel plates,
meat, and
cheese
• Epoxy
body for
ruggedness
and
durability
• Sure-Flow
junction prevents
clogging
View our selection of
refillable and
gel-filled PerpHecT
ROSS Ultra pH/ATC
Triode electrodes;
see page 8.
• For soil, sludge,
colloids, viscous
material, and
organic solvents
8202BN
8203BN
8220BNWP
8235BN
8256BN
8272BN
Cat. No.
0-14
0-14
0-14
0-14
0-14
0-14
pH Range
0.01
0.01
0.01
0.01
0.01
0.01
pH Precision
0-100 °C
0-100 °C
0-100 °C
0-100 °C
0-100 °C
0-100 °C
Temp. Range
± 0.5 °C
± 0.5 °C
± 0.5 °C
± 0.5 °C
± 0.5 °C
± 0.5 °C
Temp. Accuracy
ROSS
ROSS
ROSS
ROSS
ROSS
ROSS
Internal Reference
Ceramic
Ceramic
Sleeve
Glass Fiber
Glass Fiber
Sure-Flow
Junction
3 M KCl
(810007)
3 M KCl
(810007)
3 M KCl
(810007)
3 M KCl
(810007)
3 M KCl
(810007)
3 M KCl
(810007)
Fill Solution
L- 120 mm
D- 12 mm
L- 165 mm
T- 6 mm 95 mm L
L- 155 mm
T- 3 mm 40 mm L
L- 120 mm
D- 12 mm
L- 120 mm
D- 12 mm
L- 120 mm
D- 12 mm
Dimension
D (Dia) L (Length) T (Tip)
BNC
BNC
BNC Waterproof
BNC
BNC
BNC
Connector Type
Key Information
Environmentally-safe mercury-free option for TRIS, protein and sulfide.
All cap diameters are 16 mm at bottom of cap. All cable lengths are 1 meter.
* We offer more connector options; contact your local sales office for information.
BNC Waterproof
BNC
Pin Tip
2009 Sourcebook
13
pH electrodes
Thermo Scientific Orion PerpHecT® pH Electrodes
Measure temperature with LogR™ technology
Ideal for
everyday use
For samples with
size limitations
Easy to use and
low maintenance
Low maintenance
plus built-in ATC
For reliable pH
measurements
For analysis of
difficult samples
PerpHecT®
combination pH
electrode with
glass body
PerpHecT®
combination pH
electrode with
glass body, semimicro tip
PerpHecT®
combination pH
electrode with
epoxy body, low
maintenance gel
PerpHecT® pH/
ATC Triode™ with
epoxy body, low
maintenance gel
PerpHecT®
combination pH
electrode with
epoxy body
PerpHecT® SureFlow® combination pH
electrode with glass
body
• Measure samples
as small as 0.2 mL
• Low maintenance
gel electrode
• Low
maintenance gel
with built-in ATC
• Epoxy body for
ruggedness and
durability
• Sure-Flow junction
prevents clogging
• Fits test tubes and
small containers
• For rugged
field, plant, or
educational use
• pH and ATC
connectors for
PerpHecT meters
• Ideal for routine
or research
applications
BEST
MERCURY
FREE
• General purpose
• For routine
or research
applications
Cat. No.
9202BN
9203BN
9206BN
9207BN
9256BN
9272BN
pH Range
0-14
0-14
0-14
0-14
0-14
0-14
pH Precision
0.01
0.01
0.05
0.02
0.01
0.01
Temp. Range
0-90 °C
0-90 °C
0-80 °C
0-80 °C
0-90 °C
0-100 °C
Temp. Accuracy
± 0.5 °C
± 0.5 °C
± 0.5 °C
± 0.1 °C
± 0.5 °C
± 0.5 °C
Internal Reference
Ag/AgCl
Ag/AgCl
Ag/AgCl
Ag/AgCl
Ag/AgCl
Ag/AgCl
Junction
Ceramic
Ceramic
Wick
Wick
Glass Fiber
Sure-Flow
4 M KCl with Ag/AgCl
(900011)
4 M KCl with Ag/AgCl
(900011)
-
-
4 M KCl with Ag/AgCl
(900011)
4 M KCl with Ag/AgCl
(900011)
L- 120 mm
D- 12 mm
L- 165 mm
T- 6.5 mm x 100 mm L
L- 120 mm
D- 12 mm
L- 120 mm
D- 12 mm
L- 120 mm
D- 12 mm
L- 120 mm
D- 12 mm
BNC
BNC
BNC
BNC - Phono Tip
(PerpHecT)
BNC
BNC
Fill Solution
(Cat. No.)
Dimension
D (Dia) L (Length) T (Tip)
Connector Type
14
• For samples such as
soil, colloids, viscous
material, and organic
solvents
Cat. No.
Recommended Accessories
900011
Ag/AgCl Fill Solution
910199
All-in-One pH Buffer Kit
910001
Electrode Storage Solution
Thermo Scientific Orion Laboratory Products Catalog
pH electrodes
Thermo Scientific Orion AquaPro Professional pH Electrodes
Ideal for environmental and field testing of water and wastewater samples
Reliable and
accurate
Fast response in
small samples
Precise and
durable
Rugged with
built-in ATC
For soft solids
and semi-solids
Rugged and
accurate
BEST
MERCURY
FREE
TRIS
COMPATIBLE
AquaPro
combination pH
electrode with
low maintenance
polymer, glass
body
AquaPro
combination pH
electrode with
low maintenance
polymer, glass body,
semi-micro tip
AquaPro
combination pH
electrode with
low maintenance
polymer, glass
body, rugged bulb
• General
purpose, high
performance
• Measure samples
as small as 0.2 mL
• Toughened bulb
for rugged lab
use
• Fits test tubes and
small containers
AquaPro pH/ATC
Triode™ with low
maintenance
polymer, epoxy
body
• Low maintenance
with built-in ATC
probe for Star
meters
AquaPro
combination pH
electrode with
low maintenance
polymer, epoxy
body, flat surface
AquaPro
combination pH
electrode with
low maintenance
polymer, epoxy
body
• Flat surface for
measurements
of moist surfaces
such as agar gel
plates, meat, and
cheese
• General purpose,
high performance
9102AP
9103APWP
9104APWP
9107APMD
9135APWP
9156APWP
Cat. No.
0-14
0-14
0-14
0-14
0-14
0-14
pH Range
0.02
0.02
0.02
0.02
0.02
0.02
pH Precision
0-60 °C
0-60 °C
0-60 °C
0-60 °C
0-60 °C
0-60 °C
Temp. Range
-
-
-
± 2.0 °C
-
-
Temp. Accuracy
Ag/AgCl
Double Junction
Ag/AgCl
Double Junction
Ag/AgCl
Double Junction
Ag/AgCl
Double Junction
Ag/AgCl
Double Junction
Ag/AgCl
Double Junction
Internal Reference
Open
Open
Open
Open
Open
Open
Junction
-
-
-
-
-
-
Fill Solution
L- 120 mm
D- 12 mm
L- 165 mm
T- 6.5 mm x 100 mm L
L- 120 mm
D- 12 mm
L- 120 mm
D- 12 mm
L- 120 mm
D- 12 mm
L- 120 mm
D- 12 mm
Dimension
D (Dia) L (Length) T (Tip)
BNC
BNC Waterproof
BNC Waterproof
BNC - MiniDIN (Star)
BNC
BNC
Connector Type
Key Information
Environmentally-safe mercury-free option for TRIS, protein and sulfide.
All cap diameters are 16 mm at bottom of cap. All cable lengths are 1 meter.
We offer more connector options; contact your local sales office for information.
BNC Waterproof
BNC
MiniDIN (Star) Phono Tip (PerpHecT)
2009 Sourcebook
15
pH electrodes
Thermo Scientific Orion pH Electrodes
Ideal for all routine pH measurements
Accurate results
in a half-cell
Easy-to-clean
bulb in a halfcell design
General purpose
for everyday use
For samples with
size constraints
Durable and
reliable
Low maintenance
with built-in ATC
pH half-cell with
glass body
Low resistance pH
half-cell with glass
body, rugged bulb
Combination pH
electrode with
glass body
• For routine
or research
applications
Combination pH
electrode with
glass body, rugged
bulb
pH/ATC Triode™ with
epoxy body, low
maintenance gel
• For routine or
clean water
measurements
Combination pH
electrode with
glass body, semimicro tip
• Measure samples
as small as 0.2 mL
• Toughened bulb
for rugged lab use
BETTER
MERCURY
FREE
• Use with 900100
or 900200
reference
electrodes
• Use with 900100
reference
electrode
Cat. No.
• Epoxy body for
ruggedness
• Fits test tubes and
small containers
9107BNMD* l 9107WMMD*
9107WLMD (Star)*
9107BN (A+ Series)
9101BN
9161BN
9102BNWP
9103BNWP
9104BNWP
pH Range
0-14
0-14
0-14
0-14
0-14
0-14
pH Precision
0.02
0.02
0.02
0.02
0.02
0.02
0-80 °C
Temp. Range
0-90 °C
0-90 °C
0-90 °C
0-90 °C
0-90 °C
Internal Reference
Ag/AgCl
Ag/AgCl
Ag/AgCl
Ag/AgCl
Ag/AgCl
Ag/AgCl
Junction
-
-
Ceramic
Ceramic
Ceramic
Wick
Fill Solution
(Cat. No.)
-
-
4 M KCl w/ Ag/AgCl
(900011)
4 M KCl w/ Ag/AgCl
(900011)
4 M KCl w/ Ag/AgCl
(900011)
-
L- 120 mm
D- 12 mm
L- 120 mm
D- 12 mm
L- 120 mm
D- 12 mm
L- 165 mm
T- 6.5 mm x 100 mm L
L- 120 mm
D- 12 mm
L- 120 mm
D- 12 mm
BNC✝
BNC✝
BNC Waterproof✝
BNC Waterproof✝
BNC Waterproof✝
BNC - MiniDIN (Star)
BNC - 8 Pin (A+ Series)✝
Dimension
D (Dia) L (Length) T (Tip)
Connector Type
16
• Low maintenance gel
with built-in ATC
Cat. No.
Recommended Accessories
900011
Ag/AgCl Fill Solution
910199
All-in-One pH Buffer Kit
910001
Electrode Storage Solution
Thermo Scientific Orion Laboratory Products Catalog
pH electrodes
Low maintenance
with built-in ATC
General
purpose and
reliable
Durable with
built-in ATC
Rugged with
an easy-toclean bulb
Fast response
in difficult
samples & soils
For all samples
including
solvents
BETTER
MERCURY
FREE
pH/ATC Triode™
with epoxy body,
low maintenance
gel
• Low
maintenance gel
with built-in ATC
• Epoxy body for
ruggedness
Combination pH
electrode with
epoxy body
pH/ATC Triode™
with epoxy body,
refillable
• Epoxy body for
ruggedness and
durability
• Long-lasting
with built-in ATC
• For routine
or research
applications
• Epoxy body for
ruggedness
Low resistance
combination
pH electrode
with glass body,
rugged bulb
• For routine or
clean water
measurements
Sure-Flow®
combination pH
electrode with
epoxy body
Sure-Flow®
combination pH
electrode with
glass body
• Sure-Flow
junction
prevents
clogging
• Sure-Flow
junction prevents
clogging
• Rugged, durable
epoxy design
• For soil, sludge,
colloids, and
viscous material
9109WP l 9109WL*
9156BNWP
9157BNMD (Star)
9157BN (A+ Series)
9162BNWP
9165BNWP
9172BNWP
Cat. No.
0-14
0-14
0-14
0-14
0-14
0-14
pH Range
0.02
0.02
0.02
0.02
0.02
0.02
pH Precision
0-80 °C
0-90 °C
0-90 °C
0-90 °C
0-100 °C
0-100 °C
Temp. Range
Ag/AgCl
Ag/AgCl
Ag/AgCl
Ag/AgCl
Ag/AgCl
Ag/AgCl
Internal Reference
Wick
Glass Fiber
Glass Fiber
Ceramic
Sure-Flow
Sure-Flow
Junction
-
4 M KCl w/ Ag/AgCl
(900011)
4 M KCl w/ Ag/AgCl
(900011)
4 M KCl w/ Ag/AgCl
(900011)
4 M KCl w/ Ag/AgCl
(900011)
4 M KCl w/ Ag/AgCl
(900011)
Fill Solution
L- 120 mm
D- 12 mm
L- 120 mm
D- 12 mm
L- 120 mm
D- 12 mm
L- 120 mm
D- 12 mm
Dimension
D (Dia) L (Length) T (Tip)
BNC - MiniDIN (Star)
BNC - 8 Pin (A+ Series)
BNC Waterproof✝
BNC Waterproof
BNC Waterproof
Connector Type
L- 120 mm
L- 120 mm
D- 12 mm Cap D- 22 mm D- 12 mm
BNC Waterproof✝
E DIN Waterproof
Banana Plug
Key Information
Environmentally-safe mercury-free.
All cap diameters are 16 mm at bottom of cap. All cable lengths are 1 meter except
where noted.
✝ We offer more connector options; contact your local sales office for information.
BNC Waterproof
BNC
MiniDIN (Star)
8 Pin DIN (A+Series)
* 9107BNMD– 1.5 meter cable l 9107WMMD– 3 meter cable
9107WLMD– 6 meter cable l 9109WL– 6 meter cable
2009 Sourcebook
17
pH electrodes
Thermo Scientific Orion Green Electrodes
No lead or mercury – the world’s most environmentally friendly glass-containing
pH electrode
BETTER
MERCURY
FREE
Pb
The Green electrode line is the
world’s first glass-containing pH
electrode to be free of mercury or
lead. Typically, glass pH electrodes
have up to a 20% lead content in the
glass. Green electrodes use glass
that is lead free, which exceeds the
requirements of RoHS.
Low maintenance,
rugged epoxy
body
Low maintenance,
double junction
NEW
NEW
The rugged epoxy body provides
durability in tough applications.
Green electrodes are easy to use and
are available in low maintenance gel
or refillable designs. Double junction
versions allow measurements
in tough dirty water samples or
samples with TRIS, sulfides and
proteins.
Green electrodes provide fast,
accurate, repeatable results while
eliminating the hassle of hazardous
material disposal.
NEW
TRIS
COMPATIBLE
Green combination
double junction
pH electrode, gel
with epoxy body
Green combination
refillable pH
electrode with
epoxy body
Green combination
double junction
refillable pH electrode
with epoxy body
• Ideal for education,
plant or field use
• Ideal for routine
or research
applications
• Ideal for routine or
research applications
• Use in dirty water
or TRIS, sulfide and
protein samples
GS9106BNWP
GD9106BNWP
GS9156BNWP
GD9156BNWP
0-14
0-14
0-14
0-14
pH Precision
0.05
0.02
0.02
0.02
Temp. Range
0-90 °C
0-90 °C
0-90 °C
0-90 °C
-
-
-
-
Ag/AgCl
Ag/AgCl Dbl. Junction
Ag/AgCl
Ag/AgCl Dbl. Junction
Glass Capillary
Glass Capillary
Glass Fiber
Glass Fiber
-
-
4 M KCl w/ Ag/AgCl
(900011)
3 M KCl
(910008)
L- 120 mm
D- 12 mm
L- 120 mm
D- 12 mm
L- 120 mm
D- 12 mm
L- 120 mm
D- 12 mm
-
-
-
-
BNC Waterproof
BNC Waterproof
BNC Waterproof
BNC Waterproof
Temp. Accuracy
Internal Reference
Junction
Fill Solution
Dimension
D (Dia) L (Length) T (Tip)
Depth of Immersion
Connector Type
18
NEW
• Use in dirty water
or TRIS, sulfide and
protein samples
Cat. No.
Cat. No.
Green combination
pH electrode, gel
with epoxy body
• Ideal for education,
plant or field use
pH Range
Refillable,
double junction
TRIS
COMPATIBLE
LEAD FREE
There is no loss of performance with
a Green electrode. They perform
as well or better than standard pH
electrodes. In fact, low maintenance
green electrodes have a wider
temperature range than standard gel
electrodes.
Refillable,
rugged epoxy
body
Recommended Accessories
910001
Electrode Storage Solution
910199
All-in-One pH Buffer Kit
900011
Ag/AgCl Fill Solution
910008
3 M KCl Fill Solution
Thermo Scientific Orion Laboratory Products Catalog
pH electrodes
Thermo Scientific Orion Micro pH Electrodes
Ideal for small samples; fits 384 microwell plates
Measure
samples as
small as 5 μL
For small test
tubes and
micro-titer plates
Measure
samples as
small as 0.5 μL
For use in
NMR tubes
For piercing
septa
Accurate in
extremely
small samples
BETTER
MERCURY
FREE
TRIS
COMPATIBLE
Micro
combination pH
electrode with
glass body
• Designed for
test tubes and
small containers
• Measure
samples as
small as 5 μL
Micro
combination
pH electrode
with glass body,
small length
Micro
combination
pH electrode
with glass body,
small tip
Micro
combination pH
electrode with
glass body, long
length
• Measure
samples as
small as 5 μL
• Designed
for gels and
small sample
volumes
• Ideal for
measurements
in extremely
small vessels
Micro
combination pH
electrode with
needle tip
PerpHecT® ROSS®
combination pH
electrode with
glass body, micro
• 16 gauge
stainless steel
bevel tip
• Measures samples
as small as 15 μL
in 384 well plates
• Minimum depth
of immersion is
4.5 mm
• Measure
samples as
small as 0.5 μL
9802BN
9803BN
9810BN
9826BN
9863BN
8220BNWP
Cat. No.
0-14
0-14
0-14
0-14
0-14
0-14
pH Range
0.02
0.02
0.02
0.02
0.02
0.01
pH Precision
0-80 °C
0-80 °C
0-80 °C
0-80 °C
0-80 °C
0-100 °C
Temp. Range
-
-
-
-
-
± 0.5 °C
Temp. Accuracy
Ag/AgCl
Ag/AgCl
Ag/AgCl
Ag/AgCl
Ag/AgCl
ROSS
Internal Reference
Ceramic
Ceramic
Ceramic
Ceramic
Ceramic
Sleeve
Junction
4 M KCl w/ Ag/AgCl
(900011)
4 M KCl w/ Ag/AgCl
(900011)
4 M KCl w/ Ag/AgCl
(900011)
4 M KCl w/ Ag/AgCl
(900011)
4 M KCl w/ Ag/AgCl
(900011)
3 M KCl
(810007)
Fill Solution
L- 150 mm
T- 2.5 mm x 18 mm L
L- 83 mm
T- 2.5 mm x 48 mm L
L- 120 mm
T- 1.3 mm x 37 mm L
L- 228 mm
T- 2.5 mm x 2.5 mm L
L- 137 mm
T- 1.7 mm x 40 mm L
L- 155 mm
T- 3 mm x 40 mm L
Dimension
D (Dia) L (Length) T (Tip)
2 mm
2 mm
1 mm
2 mm
3 mm
-
Depth of Immersion
BNC
BNC
BNC
BNC
BNC
BNC Waterproof
Connector Type
Key Information
Environmentally-safe mercury-free.
All cap diameters are 16 mm at bottom of cap. All cable lengths are 1 meter.
We offer more connector options; contact your local sales office for information.
BNC Waterproof
BNC
2009 Sourcebook
19
pH electrodes
Thermo Scientific
Orion No Cal®
Thermo Scientific Orion
Double Junction pH Electrodes
Electrode maintains
calibration for
one year
For meat, cheese
and sludge
samples
Accurate and
reliable
For samples with
size limitations
Durable and low
maintenance
No Cal® combination
pH electrode with epoxy
body and built-in ATC
KNIpHE® Double
Junction combination pH
electrode with stainless
steel cutting blade
Double Junction
combination pH
electrode with
glass body
• For meat, cheese and
sludge where glass
alone may break
• General purpose
Double Junction
combination pH
electrode with
glass body, semimicro tip
Double Junction
combination pH
electrode with
epoxy body, low
maintenance gel
• Measure samples
as small as 0.2 mL
• Low maintenance
gel with durable
epoxy body
BETTER
MERCURY
FREE
TRIS
COMPATIBLE
• Designed for stable,
high performance pH
analysis in the field
• Replacement electrode
filling solution
Cat. No. 510011
• Easy-to-clean
refillable design
• Replacement electrode
Cat. No. 9121APWP
• Easy-to-clean
refillable design
• Replacement blade
Cat. No. 712001
• Replacement sheath
Cat. No. 712002
5107BNMD (Star)
5107NC (A+ Series)
9120APWP
9102DJWP
9110DJWP
9156DJWP
pH Range
0-14
0-14
0-14
0-14
0-14
pH Precision
0.01
0.02
0.02
0.02
0.02
Temp. Range
0-100 °C
0-60 °C
0-60 °C
0-60 °C
0-60 °C
Temp. Accuracy
-
Cat. No.
± 1.0 °C
-
-
-
Internal Reference
Platinum
Ag/AgCl Dbl. Junction
Ag/AgCl Dbl. Junction
Ag/AgCl Dbl. Junction
Ag/AgCl Dbl. Junction
Junction
Ceramic
Ceramic
Ceramic
Ceramic
Wick
No Cal
(510011)
3 M KCl
(910008)
3 M KCl
(910008)
3 M KCl
(910008)
-
L- 120 mm
D- 12 mm
L- 215 mm
D- 16 mm
L- 120 mm
D- 12 mm
L- 150 mm
T- 4.5 mm x 90 mm L
L- 120 mm
D- 12 mm
BNC - MiniDIN (Star)
BNC - 8 Pin (A+ Series)✝
BNC Waterproof
BNC Waterproof
BNC Waterproof
BNC Waterproof
Fill Solution
Dimension
D (Dia) L (Length) T (Tip)
Connector Type
20
Cat. No.
Recommended Accessories
910008
3 M KCl Fill Solution
900011
Ag/AgCl Fill Solution
910001
Electrode Storage Solution
910199
All-in-One pH Buffer Kit
Thermo Scientific Orion Laboratory Products Catalog
pH electrodes
Thermo Scientific
Orion Economy pH Electrodes
For everyday
use in small
samples
For rugged,
everyday use
Thermo Scientific
Specialty Electrodes
For everyday
use in tall
containers
• 9163SC
For solids and
semi-solids
Combination pH electrode
with glass body, piercing
spear tip (screw cap)
• 9166SC
Combination pH electrode
with glass body, sleeve
junction (screw cap)
BETTER
• 9167SC
Combination pH electrode
with glass body, semimicro, flat surface
(screw cap)
• 930101
HF resistant pH module
for samples with
hydrofluoric acid
Thermo Scientific
Basic Electrodes
• 9142BN
Refillable combination pH
electrode with glass body,
900011 fill solution (BNC)
• 9145BN
Low maintenance gel
combination pH electrode
with epoxy body (BNC)
Combination pH
electrode with
epoxy body, low
maintenance gel
• Ideal for field,
plant, or
educational
applications
Combination pH
electrode with
epoxy body, low
maintenance gel,
semi-micro tip
Combination pH
electrode with
epoxy body, low
maintenance gel,
flask length
Combination pH
electrode with
epoxy body, low
maintenance gel,
flat surface
• Measure
samples as small
as 0.2 mL
• Designed for
use in longnecked flasks
• Flat surface
for analysis of
moist surfaces
• 9146BN
Low maintenance gel
combination pH electrode
with epoxy body and
removable bulb guard (BNC)
GOOD
• 9147BN
Low maintenance gel
combination pH electrode
with epoxy body and builtin ATC (BNC and 2.5 mm
phono jack connections,
for use with Russell
RL060P meter)
• 917004
Stainless steel ATC probe
(2.5 mm phono jack, for
use with Russell meters)
Visit www.thermo.com/water for information on Russell meters.
9106BNWP
911600
912600
913600
Cat. No.
0-14
0-12
0-12
0-12
pH Range
0.05
0.1
0.1
0.1
pH Precision
0-80 °C
0-80 °C
0-80 °C
0-80 °C
Temp. Range
-
-
-
-
Temp. Accuracy
Ag/AgCl
Ag/AgCl
Ag/AgCl
Ag/AgCl
Internal Reference
Wick
Wick
Wick
Wick
Junction
-
-
-
-
Fill Solution
L- 120 mm
D- 12 mm
L- 145 mm
D- 6 mm
L- 305 mm
D- 8 mm
L- 110 mm
D- 12 mm
Dimension
D (Dia) L (Length) T (Tip)
BNC Waterproof
BNC
BNC
BNC
Connector Type
Key Information
Environmentally-safe mercury-free; double junction are compatible with TRIS,
protein and sulfide.
BNC Waterproof
BNC
MiniDIN (Star
8 Pin DIN (A Series)
Screw Cap
All cap diameters are 16 mm at bottom of cap. All cable lengths are 1 meter.
✝ We offer more connector options; contact your local sales office for information.
2009 Sourcebook
21
pH electrodes
Thermo Scientific Orion Redox/ORP and ORP Triode™ Electrodes
Durable and
clog-free
Solvent-resistant
glass body
Easy-to-clean
with built-in ATC
Low maintenance
with built-in ATC
Sure-Flow®
combination redox/
ORP electrode with
epoxy body
Combination redox/
ORP electrode with
glass body
ORP/ATC Triode™
with epoxy body,
refillable
• For use in water,
wastewater, metal
plating and organic
solvent samples
• For use in water
and wastewater
ORP/ATC Triode™
with epoxy body,
low maintenance
gel
BEST
BETTER
GOOD
MERCURY
FREE
• For use in water,
wastewater,
metal plating, and
biotech samples
Cat. No.
pH Range
• Epoxy body for
ruggedness and
durability
9778BNWP
9180BNMD (Star)
9180BN (A+ Series)
9179BNMD (Star)
9179BN (A+ Series)
-
-
-
-
Temp. Range
0-90 °C
0-100 °C
0-90 °C
0-80 °C
Internal Reference
Ag/AgCl
Ag/AgCl
Ag/AgCl
Ag/AgCl
Junction
Fill Solution
(Cat. No.)
Dimension
D (Dia) L (Length) T (Tip)
Connector Type
22
9678BNWP
• For use in water
and wastewater
• Epoxy body for
ruggedness and
durability
Cat. No.
Recommended Accessories
900011
Ag/AgCl Fill Solution
967901
ORP Standard 475 mL
967961
ORP Standard 5 x 60 mL
Thermo Scientific Orion Laboratory Products Catalog
Sure-Flow
Ceramic
Glass Fiber
Wick
4 M KCl w/ Ag/AgCl
(900011)
4 M KCl w/ Ag/AgCl
(900011)
4 M KCl w/ Ag/AgCl
(900011)
-
L- 110 mm
D- 13 mm
L- 120 mm
D- 12 mm
L- 120 mm
D- 12 mm
L- 120 mm
D- 12 mm
BNC Waterproof✝
BNC Waterproof✝
BNC - MiniDIN (Star)
BNC - 8 Pin (A+Series)
BNC - MiniDIN (Star)
BNC - 8 Pin (A+Series)
pH electrodes
Thermo Scientific Orion
Sure-Flow® Reference
Electrodes
For routine pH
and many ISE
measurements
Thermo Scientific Orion
Silver Billet and Karl Fischer
Electrodes
For use with
ISE half-cell
electrodes
For Karl Fischer
titrations
For halide
titrations
For halide
titrations
BEST
BETTER
GOOD
MERCURY
FREE
Single Junction
Sure-Flow®
reference halfcell with epoxy
body
• Use with
9101BN, 9101SC,
9161BN or
9161SC pH halfcell electrodes
• Compatible with
several ISEs
Double Junction
Sure-Flow®
reference half-cell
with epoxy body
Karl Fischer
double platinum
electrode with
glass body
• Sure-Flow
junction prevents
clogging
• For accurate Karl
Fischer titrations
Silver Billet
combination
electrode
Silver Billet halfcell electrode
• Use with
900200
reference
electrode
• For accurate and
reliable halide
titrations
• Double junction
allows a variety of
filling solutions to
be used
900100
900200
Cat. No.
977900
9780SC
9781SC
Cat. No.
0-14
0-14
pH Range
Endpoint Indicator
Endpoint Indicator
Endpoint Indicator
pH Range
0-100 °C
0-100 °C
Temp. Range
0-40 °C
0-80 °C
0-80 °C
Temp. Range
Ag/AgCl
Ag/AgCl Dbl. Junction
Internal Reference
Platinum
Ag/AgCl
Ag/AgCl
Internal Reference
Sure-Flow
Sure-Flow
Junction
-
Ceramic
-
Junction
Equitransferent
solution w/ Ag/AgCl
(900001)
Equitransferent
solution (900002)
outer (900003)
Fill Solution
-
4 M KCl w/Ag/AgCl
(900011)
-
Fill Solution
L- 110 mm
D- 13 mm
L- 110 mm
D- 13 mm
Dimension
D (Dia) L (Length) T (Tip)
L- 120 mm
D- 12 mm
L- 150 mm
D- 12 mm
L- 120 mm
D- 12 mm
Dimension
D (Dia) L (Length) T (Tip)
Pin Tip
Pin Tip
Connector Type
KF Input
Screw Cap
Screw Cap
Connector Type
Key Information
Environmentally-safe mercury-free.
All cap diameters are 16 mm at bottom of cap. All cable lengths are 1 meter.
✝ We offer more connector options; contact your local sales office for information.
BNC Waterproof
Screw Cap
Pin Tip
MiniDIN (Star)
8 Pin DIN (A Series)
2009 Sourcebook
23
pH electrodes
Thermo Scientific Orion ATC Probes
Water Quality Specialist Tip
Temperature is the number one
source for pH error! The ideal solution
is an ATC probe – now your pH
measurement is accurate and EPA
compliant.
BEST
Fastest
temperature
response
Rugged
stainless steel
Durable epoxy
Solventresistant glass
Micro ATC Probe
with Stainless
Steel and Epoxy
Body
ATC Probe with
Stainless Steel
Body
ATC Probe with
Epoxy Body
ATC Probe with
Glass Body
• General purpose
• Extremely
durable body
• For aqueous
samples,
intermittent use
in methanol or
ethanol
• For solutions that
contain organic
solvents
Change of pH Slope with
Temperature
Electrode Potential (mV)
500
100˚C (74 mV/pH Unit)
BETTER
50˚C (64 mV/pH Unit)
0˚C (54 mV/pH Unit)
Isopotential Point
0
GOOD
-500
7
14
pH
pH Value versus Temperature
25 °C
0 °C
10 °C
20 °C
40 °C
1.68
1.67
1.67
1.67
1.69
4.01
4.00
4.00
4.00
4.03
6.86
6.98
6.95
6.87
6.84
7.00
7.11
7.08
7.01
6.97
9.18
9.46
9.40
9.23
9.07
10.01
10.32
10.25
10.06
9.89
12.46
13.42
13.21
12.64
11.99
• For critical
temperature
monitoring
• Measures
samples as small
as 10 μL
• For field, plant,
and food
applications
• Minimum depth
of immersion is
3 mm
Cat. No.
Temp. Range
Temp. Accuracy
Dimension
D (Dia) L (Length) T (Tip)
Connector Type
928007MD (Star)
927007MD (Star)
917007 (A+ Series)
927005MD (Star)
917005 (A+ Series)
927006MD (Star)
917006 (A+ Series)
0-100 °C
0-100 °C
0-80 °C
0-100 °C
± 1 °C
± 1 °C
± 1 °C
± 1 °C
L- 117 mm
D- 5 mm
D- 1 mm x 38 mm L
L- 120 mm
D- 6 mm
L- 120 mm
D- 6 mm
L- 120 mm
D- 8 mm
MiniDIN for Star
MiniDIN (Star)
8 Pin DIN (A+ Series)✝
MiniDIN (Star)
8 Pin DIN (A+ Series)✝
MiniDIN (Star)
8 Pin DIN (A+ Series)✝
Key Information
Environmentally-safe mercury-free.
All cap diameters are 16 mm at bottom of cap. All cable lengths are 1 meter.
✝We offer more connector options; contact your local sales office for information.
MiniDIN (Star)
24
8 Pin DIN (A Series)
Thermo Scientific Orion Laboratory Products Catalog
pH electrodes
Thermo Scientific Orion Stirrers
Ideal for laboratory pH and ISE measurements
Automatic Stirrer Probe
NEW
PRICE
Thermo Scientific Orion stirring probes are the answer to
economical and convenient stirring.
• Powered by Orion DUAL STAR and 3-, 4- and 5- Star
benchtop meters
• Eliminates the hassles of magnetic stir plates and
stir bars
• No heat transfer from stir plate
• No need to capture & wash stir bars
• No potential electrode bulb damage from stir bars
ionplus™ Stirrer
• Stirrer easily attaches to electrode body to
mix sample
• Works with any magnetic stirring plate
• Ideal for small volume samples
ionplus Stirrer
Cat. No.
Description
900060
ionplus stirrer
stirrer easily attaches to the
electrode body to mix sample
• New price breakthrough - half the price of similar models
• Conveniently plugs into the back panel of DUAL STAR
and benchtop Star meters for stirrer power, speed
and control
It is so easy-to-use! Plug the stirring probe into the back
panel of an Orion DUAL STAR or benchtop 3-, 4- and 5- Star
meter and place it into an electrode stand with the measuring
electrode. Control the on/off function and stirring speed using
the meter controls. Simply insert the stirring probe into a
sample and rinse it between measurements.
Stirring Probes
with Paddles
Cat. No.
Description
096019
Automatic stirring probe
with paddle
096021
Replacement paddles,
3 pack
Accessories
Orion DUAL STAR™
pages 48-49
For a winning combination, pair the
automatic stirrer probe with an Orion
DUAL STAR meter.
2009 Sourcebook
25
pH electrodes
Thermo Scientific Orion
pH Buffers, Solutions and Test Kits
pH Buffers
Cat. No.
Description
910199
All-in-One pH Buffer Kit–
475 mL each of 4.01, 7.00,
10.01 buffers and pH electrode
storage solution (910001), pH
electrode storage bottle
810199
All-in-One ROSS pH Buffer Kit–
475 mL each of 4.01, 7.00, 10.01
buffers and ROSS pH electrode
storage solution (810001), 30
mL bottle of general cleaning
solution (900023), pH electrode
storage bottle
NEW
911110
911125
pH Rinse Solution–
PerpHecT Individual Rinse
Pouch
10 pack
25 pack
910168
9116860
pH 1.68 Buffer–
475 mL
5 x 60 mL
910104
910460
9104CB
910410
910425
Our pH and ORP buffers are designed to meet all end user
requirements, including laboratory grade NIST traceability and
DIN standards.
To maximize the life of your electrodes, we offer pH electrode
cleaning kits as well as filling, storage and rinse solutions for
proper electrode care.
All bottles are 5 x 60 mL
Cat. No.
Description
For the following pH Electrodes
810007
3 M KCI ROSS
ROSS Ultra, ROSS and PerpHecT ROSS Electrodes
910008
3 M KCl
Double Junction, Double Junction Refillable Green and
Double Junction KNIpHE Electrodes
900011
4 M KCl with
Ag/AgCl
or
2 M KCl with Ag/
AgCl (For low ionic
strength samples)
Standard, Specialty, and PerpHecT Ag/AgCl Electrodes, and
Single Junction Refillable Green Electrodes
pH 5.00 Buffer (orange)– 475 mL
910686
916860
pH 6.86 Buffer–
475 mL
5 x 60 mL
910107
910760
9107CB
910710
4 M KCI with
Ag/AgCl
Micro pH Electrodes and ORP Electrodes
900001
Equitransferent
Solution with
Ag/AgCl
Single Junction Sure-Flow Reference Electrode
900002
Inner Chamber–
Equitransferent
Solution with
Ag/AgCl
Outer Chamber–
10% KNO3
Double Junction Sure-Flow Reference Electrode
510011
No Cal Electrode
Filling Solution
No Cal pH Electrodes
610001
3 M KCl with
Ag/AgCl
pHuTure electrodes
910918
9191860
910110
911060
9110CB
911010
911025
900004
900011
900003
26
Standard, Specialty, and PerpHecT Ag/AgCl Electrodes, and
Single Junction Refillable Green Electrodes used in low
ionic strength samples
910112
911260
pH 7.00 Buffer (yellow)–
475 mL
5 x 60 mL
19 L (5 gallon) cubitainer
PerpHecT Individual Buffer
Pouch,
10 pack
PerpHecT Individual Buffer
Pouch, 25 pack
pH 9.18 Buffer–
475 mL (1 pint)
5 x 60 mL
pH 10.01 Buffer (blue)–
475 mL
5 x 60 mL
19 L (5 gallon) cubitainer
PerpHecT Individual Buffer
Pouch,
10 pack ,PerpHecT Individual
Buffer Pouch, 25 pack
pH 12.46 Buffer–
475 mL
5 x 60 mL
ORP Standards
Cat. No.
Description
967901
475 mL
967961
5 x 60 mL
NIST
Storage Solutions
Cat. No.
Description
910001
910060
9100CB
pH Electrode Storage Solution–
475 mL
5 x 60 mL
19 L (5 gallon) cubitainer
810001
Thermo Scientific Orion Laboratory Products Catalog
pH 4.01 Buffer (red)–
475 mL
5 x 60 mL
19 L (5 gallon) cubitainer
PerpHecT Individual Buffer
Pouch, 10 pack
PerpHecT Individual Buffer
Pouch, 25 pack
910105
910725
pH Electrode Filling Solutions
NIST
ROSS pH Electrode Storage
Solution– Optimize ROSS
and ROSS Ultra pH electrode
performance and life, 475 mL
pH electrodes
pH Testing Kits
Cat. No.
Description
700010
Total Alkalinity Test Kit–
Alkalinity reagent, 475 mL
(700011); alkalinity standard/
control, 475 mL (700012); total
alkalinity conversion chart;
application procedure 517 “Total
Alkalinity in Natural Waters”
700001
Pure Water pH Test Kit–
pH 6.97 buffer, 4 x 475 mL
(700702); pH 4.10 buffer, 2
x 475 mL (700402); pHISA
adjuster, 2 x 50 mL (700003),
application procedure 501 “pH
Measurement in Low Ionic
Strength Solutions”
700702
Pure Water pH 6.97 Buffer A–
4 x 475 mL
700402
Pure Water pH 4.10 Buffer B–
4 x 475 mL
700902
Pure Water pH 9.15 Buffer C–
4 x 475 mL
700003
pHISA Pure Water Ionic Strength
Adjustor– 5 x 60 mL; one
syringe dispenser
Total Alkalinity Test Kit
• Direct Reading 0-200 ppm as
CaCO3
• Requires No Special Equipment
• No Interference from Color or
Turbidity
Total alkalinity is as easy as pH! The total
alkalinity test kit is a two-step direct
measurement for total alkalinity and
pH. Calibrate for pH as usual, add the
alkalinity reagent and read total alkalinity
in ppm CaCO3 from the enclosed
pH conversion chart. Saves time by
eliminating additional equipment set-up,
tedious titrations and calculations.
pH Electrode Cleaning
Pure Water® pH Test Kit
• Increases Measurement Accuracy
Cat. No.
Description
900020
pH Electrode Cleaning Solutions–
pH Cleaning Solution Kit–
Includes 1 x 30 mL each of
cleaning solutions (A and C)
and 1 x 60 mL each of cleaning
solutions (B and D), beaker
and pipet
900021
pH Cleaning Solution A–
For removing protein
contaminants, includes 4 x 30
mL bottles, beaker and pipet
900022
pH Cleaning Solution B–
For removing bacterial
contaminants, includes 4 x 60
mL bottles, beaker and pipet
900023
pH Cleaning Solution C–
For general cleaning,
includes 4 x 30 mL bottles,
beaker and pipet
900024
pH Cleaning Solution D–
For removing oil and grease
contaminants, includes 4 x 60
mL bottles, beaker and pipet
• Reduces Electrode Drift
• Eliminates Contamination of
Sample
Many laboratories take pH
measurements of “pure water” with
low levels of acid, base or dissolved
salts. The low conductivity and limited
buffering capacity of low ionic strength
pure water causes pH electrodes to
drift, producing non-reproducible and
inaccurate results. Using the Pure Water
pH Test Kit provides fast, accurate pH
results in high purity waters. This kit
contains special low ionic strength
buffers for calibration and an additive,
called pHISA Adjustor, to increase
sample ionic strength without changing
its pH. Recommended for samples
such as well water, rain water, distilled/
deionized water, boiler feed water,
and process water. For best results, a
high quality pH electrode such as the
8102BNUWP ROSS Ultra pH electrode
is recommended.
pH Accessories
Cat. No.
Description
910005
Bulb guard for glass pH
electrode, 5 pack
910003
12 mm electrode storage
bottles, 3 pack
910004
8 mm electrode storage bottles,
3 pack
910006
6 mm electrode storage bottles,
3 pack
020017
No Cal electrode storage chamber
510015
Stain lifter solution, 60 mL
Signifies a hazardous solution. See terms and conditions for
important shipping information at www.thermo.com/water.
2009 Sourcebook
27
pH electrodes
Thermo Scientific Orion Electrode Connectors and Cables
Connectors that stay dry and connected
If you have ever lost data – or even
an electrode – because of a failed
connection, you’ll appreciate the Orion
Star Series’ proprietary waterproof BNC
and MiniDIN locking connectors. Adapter
cables can be used to connect Orion Star
Series electrodes to a wide variety of
other meters.
Electrode Adapter Cables
28
Cat. No.
Meter Input
Electrode Connector
Adapter Needed
090033
BNC
U.S. Standard
US Standard to BNC
Adapter
91CBNC
BNC
Screw Cap
Detachable Cable to BNC
Connector
090048
BNC
Karl Fischer Double
Pin Tip
Karl Fischer Adapter
090032
U.S. Standard
BNC
BNC to U.S. Standard
Adapter
91USCB
U.S. Standard
Combination Electrode
with Screw Cap
Detachable Cable to U.S.
Standard Connector
91USHC
U.S. Standard
Half-Cell Electrode with
Screw Cap
Detachable Cable to HalfCell U.S. Std. Connector
090036
F LEMO (Metrohm)
BNC
BNC to F LEMO Adapter
91CLFO
F LEMO (Metrohm)
Screw Cap
Detachable Cable to F
LEMO Connector
090035
LEMO Miniature
(Mettler)
BNC
BNC to LEMO Miniature
Adapter
91CLMO
LEMO Miniature
(Mettler)
Screw Cap
Detachable Cable to
LEMO Connector
91CBBR
British (Kent EIL)
Screw Cap
Detachable Cable to
British Connector
090034
E DIN (Knick, Schott,
WTW)
BNC
BNC to E DIN Adapter
91CDIN
E DIN (Knick, Schott,
WTW)
Screw Cap
Detachable Cable to E Din
Connector
090037
Radiometer No. 7
BNC
BNC to Radiometer
Adapter
91CBRA
Radiometer No. 7
Screw Cap
Detachable Cable to
Radiometer Connector
91USRF
2 mm Pin Tip
Screw Cap
Detachable Cable to
2.5 mm Pin Tip Connector
91CBNT
Other
Screw Cap
Detachable Cable to
Stripped End
Thermo Scientific Orion Laboratory Products Catalog
Electrode Extension
Cables
Cat. No.
Description
910032
15 ft. Extension Cable, 8 PIN
Mini DIN for ATC Probes for
STAR series meters
NEW
910026
15 ft. Extension Cable, Pin Tip
connector
910027
15 ft. Extension Cable, BNC
connector
910028
15 ft. Extension Cable, 8 Pin
DIN connector for 917005,
917006, and 917007 ATC
probes
910029
15 ft. Extension Cable, 3.5
mm Phono Tip connector for
927005, 927006, and 927007
ATC probes
910030
15 ft. Extension Cable, BNC and
3.5 mm Phono Tip connectors
for 9207BN pH/ATC Triode
910031
15 ft. Extension Cable, Banana
Jack connector for 917001 and
917002 ATC probes
Adapter Cables
Meter
Capabilities
Cat. No.
Description
1010050
MiniDIN ATC
Probe to 3.5
mm Phono Tip
Meter
Use on 310
320, 330, 350
370, 555A
550A, 535A
and 162A
meters
1010051
MiniDIN ATC
Probe to 8 Pin
DIN Meter
Use on
210A+, 230A+
250A+, 290A+
410A+, 420A+
520A+, 525A+
710A+, 720A+
and 920A+
meters
1010800
MiniDIN DO
Probe to 8 Pin
Waterproof
DIN DO
Meter
Use on 830A
835A and
862A meters
1010801
MiniDIN DO
Probe to 8
Pin DIN DO
Meter
Use on
805A+, 810A+
and 850A+
meters
1010900
MiniDIN
Conductivity
Probe to 8 Pin
Waterproof
DIN
Conductivity
Meter
Use on, 555A
550A, 162A
135A, and
130A meters
1010901
MiniDIN
Conductivity
Probe to
8 Pin DIN
Conductivity
Meter
Use on
150A+, 145A+
125A+, 115A+
and 105A+
meters
ion selective electrodes
Thermo Scientific Orion Ion Selective Electrode Theory
Measurements using an ion selective electrode (ISE) are
applicable in virtually every laboratory. ISEs can be used to
measure ion concentrations in water, food, pharmaceuticals
and biological samples. Many analytical methods using ISEs
have been developed and published worldwide. The main
advantage of ion selective electrode technology is the variety
of analytical methods that are available. Ion selective electrodes
can be used as endpoint indicators or to perform direct
measurements and incremental techniques.
Electrode measurements are simpler and faster than other
analytical techniques. Time-consuming sample preparation
steps such as filtration and distillations are rarely needed.
Analysis time is typically under one minute. In comparison to
methods such as atomic absorption or ion chromatography, the
setup cost is less and does not require expensive equipment.
Methods are adaptable to the lab or field. Sample color or
turbidity does not affect the measurement. A variety of
analytical techniques are available to the analyst.
Titration
Low Level Measurement
nt
me
su
ea
Ion
ific
th
No
n-
sp
ec
ng
re
St
ic
Ion
ld
Fie
M
.1
>0
nt
me
re
su
M
ea
Ch
ce
du
Re
P
P
P
M
ge
sa
lU
ica
em
ro
be
rg
eN
um
mp
Sa
all
P
P
P
Oc
P
re
s
ple
am
fS
Vo
lu
le
am
lS
na
sio
ca
as
re
Inc
me
ng
pli
on
isi
ec
Pr
ed
No
in
ith
P
P
P
La
Known Addition
Sm
Direct Measurement
W
W
ith
in
Lin
n-
ea
Lin
rR
ea
es
po
rR
ns
es
po
eR
ns
an
ge
eR
an
ge
Choosing the right measurement technique
P
P
P
P
P
Analate Subtraction
Indicator Titration
Direct measurement is a simple procedure for measuring a
large number of samples. Only one meter reading is required
for each sample. Calibration is performed in a series of
standards. The concentration of the samples is determined by
comparison to the standards. Orion ISE meters calculate and
store the calibration curves, saving you time in analysis and
ensuring an accurate result. Ionic strength adjustor is added
to all solutions to ensure that samples and standards have
similar ionic strength, proper pH, and to reduce the effect of
interfering ions.
Low level measurement is a similar method to direct
measurement. This method is recommended when the
expected sample concentration is within the non-liner response
range of the electrode. A minimum 3 point calibration is
recommended to compensate for the electrodes non-linear
response at the concentrations. Calibration is performed in one
beaker, reducing the chance of cross contamination of
the solutions.
Known addition is a useful method for measuring samples,
since calibration is not required. This method is recommended
when measuring only a few samples, or when samples have
a high (>0.1 M) ionic strength, or when there is complicated
background matrix. The electrodes are immersed in the
P
P
P
P
P
sample solution and an aliquot of standard solution, containing
the measured species, is added tot he sample. The original
sample concentration is determined from the changes in
potential before and after the addition. As in direct calibration,
any convenient concentration unit can be used. Many Orion
ISE meters automate this measurement, by performing the
additions and automatically calculating the result.
Analate subtraction is also a useful for method for measuring
samples, since calibration is not required. The electrodes are
immersed in a reagent solution that contains a species that
the electrode senses, and that reacts with the sample. It is
useful when sample size is small, for samples for which a
stable standard is difficult to prepare, and for viscous or very
concentrated samples. The method is not suited for very diluted
samples. It is also necessary to know the stoichiometric ratio
between standard and sample.
Titrations are quantitative analytical techniques for measuring
the concentration of a species by incremental addition of a
reagent (titrant)) that reacts with the sample species. Sensing
electrodes can be used for determination of the titration end point.
Ion selective electrodes are useful as end point detectors,
because they are unaffected by sample color or turbidity.
2009 Sourcebook
29
ion selective electrodes
Thermo Scientific Orion Ion Selective Electrode Selection Guide
Reference
Filling
Solution
Calibration
Standards
Required
ISA
Construction
Gas sensing
combination
5 x 10-7 to 1.0 M
0.01 to 17,000 ppm
0 to 50 °C
Included
951202
0.1 M NH4Cl / 951006
951211
Ammonia
high performance
9512HPBNWP 1 Gas sensing
9512HP01 3
combination
5 x 10-7 to 1.0 M
0.01 to 17,000 ppm
0 to 50 °C
Included
951209
0.1 M NH4Cl / 951006
951011
Ammonium (NH4+)
931801 8
Plastic membrane 5 x 10-7 to 1.0 M
half-cell
0.01 to 17,000 ppm
0 to 50 °C
900200
900002 inner /
900018 outer
1000 ppm as N / 951007
--
Bromide (Br-)
ionplus Design
9635BNWP 1
ionplus SureFlow solid state
combination
5 x 10-6 to 1.0 M
0.40 to 79,900 ppm
0 to 80 °C
Included
900063
0.1 M NaBr / 943506
940011
Bromide (Br-)
9435BN 2
9435SC 4
Solid state
half-cell
5 x 10-6 to 1.0 M
0.40 to 79,900 ppm
0 to 80 °C
900200
900002 inner /
900003 outer
0.1 M NaBr / 943506
940011
Cadmium (Cd2+)
ionplus Design
9648BNWP 1
ionplus SureFlow solid state
combination
1 x 10-7 to 0.1 M
0.01 to 11,200 ppm
0 to 80 °C
Included
900061
Consult user guide
940011
Cadmium (Cd2+)
9448BN 2
9448SC 4
Solid state
half-cell
1 x 10-7 to 0.1 M
0.01 to 11,200 ppm
0 to 80 °C
900200
900002 inner /
900003 outer
Consult user guide
940011
Calcium (Ca2+)
ionplus Design
9720BNWP 1
ionplus Sure-Flow 5 x 10-7 to 1.0 M
plastic membrane 0.02 to 40,100 ppm
combination
0 to 40 °C
Included
900061
0.1 M CaCl2 / 922006
100 ppm CaCO3 / 923206
932011
Calcium (Ca2+)
9320BN 2
Plastic membrane 5 x 10-7 to 1.0 M
half-cell
0.02 to 40,100 ppm
0 to 40 °C
900100
900011
0.1 M CaCl2 / 922006
100 ppm CaCO3 / 923206
932011
Carbon Dioxide
(CO2)
9502BNWP 1
Gas sensing
combination
1 x 10-4 to 1 x 10-2 M
4.4 to 440 ppm
0 to 50 °C
Included
950202
950210
0.1 M NaHCO3 / 950206
1000 ppm as CaCO3 / 950207
Chloride (Cl-)
ionplus Design
9617BNWP 1
ionplus SureFlow solid state
combination
5 x 10-5 to 1.0 M
1.8 to 35,500 ppm
0 to 80 °C
Included
900062
0.1 M NaCl / 941706
100 ppm Cl- / 941707
1000 ppm Cl- / 941708
940011 or
941709 /
CISA
Chloride (Cl-)
9417BN 2
9417SC 4
Solid state
half-cell
5 x 10-5 to 1.0 M
1.8 to 35,500 ppm
0 to 80 °C
900200
900002 inner /
900003 outer
0.1 M NaCl / 941706
100 ppm Cl- / 941707
1000 ppm Cl- / 941708
940011 or
941709 /
CISA
Chloride (Cl-)
931701 8
Plastic membrane 5 x 10-6 to 1.0 M
half-cell
0.18 to 35,500 ppm
0 to 50 °C
900200
900002 inner / 0.1 M NaCl / 941706
0.1 M KCl outer 100 ppm Cl- / 941707
1000 ppm Cl- / 941708
None required
Chlorine (Cl2)
9770BNWP 1
9770SC 4
Solid state
combination
1 x 10-7 to 3 x 10-4 M
0.01 to 20 ppm
0 to 50 °C
Included
None required
100 ppm as Cl2 / 977007
977010 /
iodide reagent
977011 / acid
reagent
Cupric (Cu2+)
ionplus Design
9629BNWP 1
ionplus SureFlow solid state
combination
0 to 80 °C
1 x 10-8 to 0.1 M
6.4 x 10-4 to 6350 ppm
Included
900063
0.1 M Cu(NO3)2 / 942906
940011
Cupric (Cu2+)
9429BN 2
9429SC 4
Solid state
half-cell
0 to 80 °C
1 x 10-8 to 0.1 M
6.4 x 10-4 to 6350 ppm
900200
900002 inner /
900003 outer
0.1 M Cu(NO3)2 / 942906
940011
Cyanide (CN-)
ionplus Design
9606BNWP 1
ionplus SureFlow solid state
combination
8 x 10-6 to 1 x 10-2 M
0.2 to 260 ppm
0 to 80 °C
Included
900062
Consult user guide
951011
Cyanide (CN-)
9406BN 2
9406SC 4
Solid state
half-cell
8 x 10-6 to 1 x 10-2 M
0.2 to 260 ppm
0 to 80 °C
900200
900002 inner /
900003 outer
Consult user guide
951011
Fluoride (F-)
ionplus Design
9609BNWP 1
960900 3
ionplus SureFlow solid state
combination
1 x 10-6 M to saturated 0 to 80 °C
0.02 ppm to saturated
Included
900061
0.1 M NaF / 940906
100 ppm F- / 940907
1 ppm F- w/ TISAB II / 040906
2 ppm F- w/ TISAB II / 040907
10 ppm F- w/ TISAB II / 040908
940909 /
TISAB II
940911 /
TISAB III
Fluoride (F-)
9409BN 2
940900 3
9409SC 4
Solid state
half-cell
1 x 10-6 M to saturated 0 to 80 °C
0.02 ppm to saturated
900100
900001
0.1 M NaF / 940906
100 ppm F- / 940907
1 ppm F- w/ TISAB II / 040906
2 ppm F- w/ TISAB II / 040907
10 ppm F- w/ TISAB II / 040908
940909 /
TISAB II
940911 /
TISAB III
900200
900002 inner / Consult user guide
dilute ISA outer
Species
Cat. No.
Ammonia standard 9512BNWP 1
951201 3
(NH3)
Fluoroborate (BF4-) 9305BN 2
30
Required
Optimum
Temperature Reference
Electrode
Range
Measurement
Range
Plastic membrane 7 x 10-6 to 1.0 M
half-cell
0.6 to 86,800 ppm
Thermo Scientific Orion Laboratory Products Catalog
0 to 40 °C
930711
ion selective electrodes
Measurement
Range
Optimum
Temperature
Range
Required
Reference
Electrode
Reference
Filling
Solution
Calibration
Standards
Required
ISA
Species
Cat. No.
Construction
Iodide (I-)
ionplus Design
9653BNWP 1
ionplus® SureFlow® solid state
combination
5 x 10-8 to 1.0 M
5 x 10-3 to 127,000 ppm
0 to 80 °C
Included
900063
0.1 M NaI / 945306
940011
Iodide (I-)
9453BN 2
9453SC 4
Solid state half-cell
5 x 10-8 to 1.0 M
5 x 10-3 to 127,000 ppm
0 to 80 °C
900200
900002 inner /
900003 outer
0.1 M Na/ 945306
940011
Lead (Pb2+)
ionplus Design
9682BNWP 1
ionplus SureFlow solid state
combination
1 x 10-6 to 0.1 M
0.2 to 20,700 ppm
0 to 80 °C
Included
900062
0.1 M Pb(ClO4)2 / 948206
0.1 M Na2SO4 / 948207
Consult
instruction
manual
Lead (Pb2+)
9482BN 2
9482SC 4
Solid state half-cell
1 x 10-6 to 0.1 M
0.2 to 20,700 ppm
0 to 80 °C
900200
900002 inner /
900003 outer
0.1 M Pb(ClO4)2 / 948206
0.1 M Na2SO4 / 948207
Consult
instruction
manual
Nitrate (NO3-)
ionplus Design
9707BNWP 1
ionplus Sure-Flow
plastic membrane
combination
7 x 10-6 to 1.0 M
0.1 to 14,000 ppm as N
0 to 40 °C
Included
900046
0.1 M NaNO3 / 920706
1000 ppm N / 920707
100 ppm N / 930707
930711 or
930710 /
nitrate ISS
Nitrate (NO3-)
9307BNWP 1
Plastic membrane
half-cell
7 x 10-6 to 1.0 M
0.1 to 14,000 ppm as N
0 to 40 °C
900200
900002 inner /
900046 or ISA
outer
0.1 M NaNO3 / 920706
1000 ppm N / 920707
100 ppm N / 930707
930711 or
930710 /
nitrate ISS
Nitrogen Oxide
(NOx)
9546BN 2
Gas sensing
combination
4 x 10-6 to 5 x 10-3 M
0.18 to 230 ppm
0 to 50 °C
Included
954602
0.1 M NaNO2 / 954606
956410
Perchlorate (CIO4-) 938101 8
Plastic membrane
half-cell
7 x 10-6 to 1.0 M
0.7 to 99,500 ppm
0 to 40 °C
900200
900002 inner /
dilute ISA outer
Consult user guide
930711
Potassium (K+)
ionplus Design
9719BNWP 1
ionplus Sure-Flow
plastic membrane
combination
1 x 10-6 to 1.0 M
0.04 to 39,000 ppm
0 to 40 °C
Included
900065
0.1 M KCl / 921906
931911
Potassium (K+)
9319BN 2
Plastic membrane
half-cell
1 x 10-6 to 1.0 M
0.04 to 39,000 ppm
0 to 40 °C
900200
900002 inner /
dilute ISA outer
0.1 M KCl / 921906
931911
Silver/Sulfide
(Ag+/S2-)
ionplus Design
9616BNWP 1
ionplus SureFlow solid state
combination
0 to 80 °C
1 x 10-7 to 1.0 M
0.01 to 107,900 ppm as Ag+
0.003 to 32,100 ppm as S2-
Included
900062 for Ag+/S2- Consult user guide
900067 for Ag+
900061 for S2-
940011 for Ag+
941609 for S2-
Silver/Sulfide
(Ag+/S2-)
9416BN 2
9416SC 4
Solid state half-cell
1 x 10-7 to 1.0 M
0 to 80 °C
0.01 to 107,900 ppm as Ag+
2
0.003 to 32,100 ppm as S
900200
900002 inner /
900003 outer
Consult user guide
940011 for Ag+
941609 for S2-
Sodium (Na+)
8611BNWP 1
ROSS® Sure-Flow
combination
1 x 10-6 M to saturated
0.02 ppm to saturated
0 to 100 °C
Included
900010 or 900012
for low level Na+
10 ppm Na+ / 941105
100 ppm Na+ / 941107
1000 ppm Na+ / 841108
KA standard kit, 1 M NaCl
with ISA / 650700
0.1 M NaCl / 941706
841111
841113 /
reconditioning
solution
Sodium (Na+)
8411BN 2
ROSS half-cell
1 x 10-6 M to saturated
0.02 ppm to saturated
0 to 100 °C
800300 or
800500U
900010 or 900012
for low level Na+
10 ppm Na+ / 941105
100 ppm Na+ / 941107
1000 ppm Na+ / 841108
KA standard kit, 1 M NaCl
with ISA / 650700
0.1 M NaCl / 941706
841111
841113 /
reconditioning
solution
Sodium (Na+)
9811BN 2
Micro combination
1 x 10-6 M to saturated
0.02 ppm to saturated
0 to 80 °C
Included
900004
0.1 M NaCl / 941706
841111
Surfactant
9342BN 2
Plastic membrane
half-cell
Endpoint indicator
0 to 40 °C
900200
900002 inner /
810007 outer
0.5 M Hyamine titrant /
654201
654203 /
sample additive
Solid state half-cell
5 x 10-6 to 1.0 M
0.29 to 58,100 ppm
0 to 50 °C
900200
900002 inner /
900003 outer
Consult user guide
940011
Plastic membrane
half-cell
6 x 10-6 to 1.0 M
0 to 50 °C
900100
900011
100 ppm CaCO3 / 923206
0.1 M CaCl2 / 922006
None required
Thiocyanate (SCN-) 9458BN 2
9332BNWP 1
Water Hardness
(X2+)
Compliant with EPA testing method
1
2
3
4
Signifies a hazardous solution. See terms and
conditions for important shipping information
at www.thermo.com/water.
Key Information
1). BNC Waterproof connector 2). BNC connector 3). US Standard connector
4). Screw Cap connector, requires separate cable 8). Module only, requires separate
BNC
BNC connector
US Standard
Screw Cap
93 Series electrode handle (9300BNWP, 930000, or 9300SC)
All cap diameters are 16 mm at bottom of cap. All cable lengths are 1 meter.
2009 Sourcebook
31
ion selective electrodes
Thermo Scientific Orion Fluoride Ion Selective Electrodes
The standard in fluoride ion analysis – EPA compliant
Approved ASTM method for fluoride
in drinking water
Analyze free fluoride ions in aqueous
solutions reliably at low limits of detection.
Measurements are quick, simple, accurate
and economical.
BEST
MERCURY
FREE
TRIS
COMPATIBLE
Half cell reference design
provides flexibility
Combination
fluoride ISE
with Sure-Flow
reference
Half cells – Fluoride ISE and
Sure-Flow™ reference
• Fluoride surface
can be easily
cleaned using
toothpaste and a
lint-free wipe
• Fluoride surface can be easily
cleaned using toothpaste and a
lint-free wipe
Thermo Scientific Orion fluoride electrodes
feature high quality lanthanum fluoride
pellet sensors. Choose from combination
electrodes or half cell designs. The fluoride
half cell electrode requires a separate half
cell reference electrode.
Other applications for fluoride
electrodes
• Phosphate: Gran plot titration can
determine phosphate in applications
from animal feed to cleaning solutions
to food and beverage
• Ammonium bifluoride: Multiple
known addition (MKA) titration
method determines levels without
need of removing interfering heavy
metal ions
• Aluminum: Gran plot titration can
determine micro and semi micro
levels of aluminum
Accessories and solutions
A full line of supporting accessories is
offered to meet your measurement needs.
A variety of calibration standards are
available. Low level standards have the
added convenience of being pre-made with
ionic strength adjustor (ISA) and requiring
that TISAB ISA be added only to your
samples. TISAB II requires a 50:50 dilution
with the sample and is available in gallon
bottles. TISAB III is a concentrate and
requires a 1:10 dilution.
• Built-in SureFlow reference
provides fast and
stable readings
Cat. No.
Measurement Range
Temp. Range
Connector Type
Cat. No.
Recommended Accessories
940906
0.1 M NaF standard, 475 mL
940907
040906
32
Convenient
combination
with reference
F–,
100 ppm standard as
475 mL
1 ppm standard as F–,
with TISAB II, 475 mL
Thermo Scientific Orion Laboratory Products Catalog
040907
040908
940909
• Use with the 900100 single
junction or 900200 double
junction reference electrodes
9609BNWP l 960900
9409BN l 940900
9409SC
1 x 10-6 M to saturated
0.02 ppm to saturated
1 x 10-6 M to saturated
0.02 ppm to saturated
0 to 80 °C
0 to 80 °C
0 to 100 °C
BNC l US Standard
BNC l US Standard
Screw Cap
Pin Tip
2 ppm standard as F–,
with TISAB II, 475 mL
10 ppm standard as F–,
with TISAB II, 475 mL
TISAB II, 1 gallon
900100
-
940999
TISAB II, 4 x 1 gallon
940911
TISAB II concentrate, 475 mL
ion selective electrodes
Thermo Scientific Orion Ammonia Ion Selective Electrodes
Compliant with EPA testing methods
High performance
ammonia electrode
Standard ammonia
electrode
NEW
BEST
Cat. No.
951214
951215
951211
951011
Recommended Accessories
951209
Electrode filling solution, 60 mL
951006
0.1 M NH4Cl standard, 475 mL
951007
1000 ppm ammonia as nitrogen
standard, 475 mL
Cat. No.
Temp. Range
Connector Type
951204
951205
Recommended Accessories
Loose membranes for standard
electrodes, box of 20
Bonded membranes for
standard electrodes, pack of 3
951211
pH adjusting ISA, 475 mL
951202
Electrode filling solution, 60 mL
951006
0.1 M NH4Cl standard, 475 mL
951007
1000 ppm ammonia as nitrogen
standard, 475 mL
9512HPBNWP l 9512HP01
5 x 10-7 to 1.0 M
0.01 to 17,000 ppm
0 to 50 °C
Cat. No.
9512BNWP l 951201
Measurement
Range
5 x 10-7 to 1.0 M
0.01 to 17,000 ppm
Temp. Range
BNC l US Standard
BETTER
GOOD
MERCURY
FREE
Provides reliable results at mid
to high ammonia levels
Pre-assembled body and
membrane simplifies use and
achieves optimum performance
Measurement
Range
Cat. No.
Loose membranes for HP
electrodes, box of 20
Pre-assembled outer body and
membrane cap assembly for
HP electrodes, 3 pack
pH adjusting ISA, 475 bottle for
samples greater than 10 ppm
pH adjusting ISA, 475 bottle for
samples less than 10 ppm
Connector Type
Approved ASTM method for ammonia
in wastewater and drinking water
Measurements are quick, simple, accurate
and economical.
Thermo Scientific Orion ammonia
electrodes feature time-tested membrane
technology. Choose from high performance
and standard designs.
The high performance ammonia electrode
offers linear response down to the
lower limits of detection. The electrode
can detect down to 0.01 ppm. The high
performance ammonia electrode can
achieve response times of 1 minute in
TRIS
COMPATIBLE
0 to 50 °C
BNC l US Standard
samples of 1 ppm or higher. It is rugged
and meets the rigorous requirements
of waste water and drinking water
operators. Supplied with pack of 20 loose
membranes, 1 pre-assembled outer body
and 2 bottles of fill solution.
Other applications for ammonia
electrodes
Ammonium or nitrogen: Measure
ammonium after conversion to ammonia or
nitrogen after Kjeldahl digestion of sample.
Accessories and solutions
A full line of supporting accessories is
offered to meet your measurement needs.
2009 Sourcebook
33
ion selective electrodes
Thermo Scientific Orion Nitrate Ion Selective Electrodes
Compliant with EPA testing methods
The easy way to measure nitrate
levels in drinking water, wastewater
and soils
Convenient
combination
reference
Half cell reference design
provides flexibility
Combination
nitrate ISE with
Sure-Flow ™
reference
Half cells – Nitrate ISE
and Sure-Flow™ reference
• Sure-Flow
reference
provides stable
readings and is
easy to clean
• Reference junction is reliable
and easy to maintain
Analyze free nitrate ions in aqueous
solutions reliably at low limits of detection.
Measurements are quick, simple, accurate
and economical.
BEST
Choose from combination electrodes
or half cell designs. The nitrate half cell
electrode requires a separate half cell
reference electrode.
Other applications for nitrate
electrodes
•
Nitric Acid: Multiple known addition
(MKA) titration method determines
levels without need of removing
interfering heavy metal ions
Accessories and solutions
A full line of supporting accessories
is offered to meet your measurement
needs. A variety of calibration standards
are available. Replacement modules are
available individually or in convenient
three packs.
• Replaceable module provides
convenience
• Convenient with
small sample
sizes
Cat. No.
Measurement Range
Temp. Range
Connector Type
Cat. No.
900046
900002
920706
920707
34
9707BNWP
9307BNWP
900200
7 x 10-6 to 1.0 M
7 x 10-6 to 1.0 M
0.1 to 14,000 ppm as N 0.1 to 14,000 ppm as N
0 to 80 °C
0 to 80 °C
0 to 100 °C
BNC
BNC
Pin Tip
Recommended Accessories
Optimum Results F electrode
fill solution, 5 x 60 mL. For
9707BNWP and outer fill
solution for 900200
Inner chamber fill solution for
900200, 5 x 60 mL
0.1 M nitrate standard solution,
475 mL
1000 ppm as nitrogen, 475 mL
Thermo Scientific Orion Laboratory Products Catalog
930707
930711
930710
970701
1000 ppm as nitrogen, 475 mL
Nitrate ionic strength adjustor
(ISA), 475 mL
Nitrate interference suppressor
solution, 475 mL
Replacement module for
9707BNWP (1 each)
930701
Replacement module for
9307BNWP (pack of 3)
930702
Replacement module for
9307BNWP (1 each)
ion selective electrodes
Thermo Scientific Orion ISE Solutions and Accessories
ISE Ionic Strength Adjusters
(ISA) and Special Reagents ISE Filling Solutions
Cat. No.
Description
All are 475 mL bottles unless noted
Cat. No.
951211
Ammonia– pH Adjusting ISA
951202
700005
Nitrate Analysis Test Kit for Ammonia
Electrode–
Alkaline Reagent (951011); Reducing
Reagent (700006); Electrode Filling Solution,
2 x 50 mL (951203); 100 ppm as N Nitrate
Standard (930707); 100 ppm Ammonia
Standard (951207); Application Procedure
No. 115 “Nitrate Measurement in
Environmental Samples”
700006
Nitrate Analysis Test Kit– Reducing Agent
940011
Bromide, Cadmium, Chloride, Cupric, Iodide,
Silver, Thiocyanate– ISA
932011
Calcium– ISA
950210
Carbon Dioxide– ISA
941709
Chloride– Complexation Ionic Strength
Adjuster (CISA) Reagent Pack, 2 x 475 mL
Description
All are 5 x 60 mL bottles unless noted
Ammonia, 1 x 60 mL (for 9512BNWP
and 951201)
951209
Ammonia, 1 x 60 mL (for 9512HPBNWP
and 9512HP01)
900018
Ammonium
950202
Carbon Dioxide, 1 x 60 mL
900017
Chloride
900001
Fluoride
954602
Nitrogen Oxide, 1 x 50 mL
Sodium– Micro Electrode
930710
Nitrate– Nitrate Interference Suppressor
Solution (NISS)
934610
Nitrite– ISA
956410
Nitrogen Oxide– Acid Buffer
931911
Potassium– ISA
941609
Sulfide– Sulfide Anti-Oxidant Buffer (SAOB)
Reagent Pack, 4 x 475 mL
Polishing Strips for Solid State Electrodes
ionplus Stirring Accessory
Chloride–
0.1 M NaCl
100 ppm as Cl1000 ppm as Cl-
977007
Chlorine– 100 ppm as Cl2
942906
Cupric– 0.1 M Cu(NO3)2
Nitrate–
0.1 M NaNO3
100 ppm as N
1000 ppm as N
900004
900060
941706
941707
941708
920706
930707
920707
Fluoroborate, Nitrate, and Perchlorate– ISA
948201
Carbon Dioxide–
0.1 M NaHCO3
1000 ppm as CaCO3
Lead–
0.1 M Pb(CIO4)2 Lead
0.1 M Na2SO4 Sulfate
930711
97 Series ionplus Handle–
9700BNWP BNC Waterproof Connector
950206
950207
948206
948207
Sodium–
2 M NH4Cl
0.1 M NH4Cl for low sodium levels
93 Series Handle–
9300BNWP BNC Waterproof Connector
U.S. Std. Connector
930000
Screw Cap, Requires Separate Cable
9300SC
Calcium– 0.1 M CaCl2
Iodide– 0.1 M Nal
900010
900012
Description
Bromide– 0.1 M NaBr
922006
945306
Fluoride–
TISAB II 1 gallon, 3.8 L
TISAB II 4 gallon, 4 x 3.8 L
TISAB III Concentrate
Cat. No.
943506
Double Junction Reference–
Inner Filling Solution
Outer Filling Solution
940909
940999
940911
ISE Accessories
Ammonia–
0.1 M NH4Cl
100 ppm as N
1000 ppm as N
900002
900003
Cyanide– Alkaline Reagent, 10 M NaOH
Surfactant–
Sample Additive
Non-Ionic Surfactant Reagent
Non-Ionic Surfactant Titrant
951006
951207
951007
940906
940907
040906
040907
040908
940916
951011
654203
654204
654205
Description
All are 475 mL bottles unless noted
900067
900063
900065
900046
900062
Chlorine–
Acid Reagent
Iodide Reagent, 5 x 50 mL
Sodium–
ISA
Reconditioning Solution
Storage Solution
Cat. No.
ionplus Series–
Optimum Results A– Cadmium, Calcium,
Fluoride, Sulfide
Optimum Results B– Chloride, Cyanide, Lead,
Silver/Sulfide
Optimum Results C– Silver
Optimum Results D– Bromide, Cupric, Iodide
Optimum Results E– Potassium
Optimum Results F– Nitrate, Nitrite
900061
977011
977010
841111
841113
841101
ISE Calibration Standards
ISE Membrane
and Modules
Cat. No.
Description
951204
951205
Standard Ammonia–
9512BNWP and 951201
20 Loose Membranes
3 Bonded Membranes
951214
951215
High Performance Ammonia–
9512HPBNWP and 9512HP01
20 Loose Membranes
3 Pre-assembled body and membrane cap
950204
Carbon Dioxide–
4 Membranes with O-rings
954604
Nitrogen Oxide–
20 Loose Membranes
931801
932001
931701
930501
930702
930701
938101
930101
931901
933201
Plastic Membrane Half-Cell Modules–
Ammonium
Calcium
Chloride
Fluoroborate
Nitrate
Nitrate – 3 Pack
Perchlorate
pH – HF Resistant
Potassium
Water Hardness
972001
970701
971901
Plastic Membrane Combination Modules–
ionplus Calcium
ionplus Nitrate
ionplus Potassium
Fluoride–
0.1 M NaF
100 ppm as F1 ppm as F- with TISAB II
2 ppm as F- with TISAB II
10 ppm as F- with TISAB II
Bulk Pack– 4 x 475 mL each of
1 ppm as F- with TISAB II (040906) and 10
ppm as F- with TISAB II (040908)
954606
Nitrite and Nitrogen Oxide–
0.1 M NaNO2
921906
Potassium– 0.1 M KCl
941104
941706
941105
941107
841108
841109
650700
Sodium–
5% NaCl
0.1 M NaCl
10 ppm as Na+
100 ppm as Na+
1000 ppm as Na+
Known Addition Standard,
1000 ppm as Na+ with ISA
Know Addition Kit– 3 x 475 mL of 1 M NaCl
with ISA and 1 x 475 mL ISA (841111)
654201
654202
Surfactant–
0.0 5 M Hyamine
0.01 M Sodium Lauryl Sulfate (SLS), 60 mL
923206
922006
Water Hardness–
100 ppm as CaCO3
0.1 M CaCl2
Signifies a hazardous solution. See terms and
conditions for important shipping information
at www.thermo.com/water.
2009 Sourcebook
35
conductivity cells
Thermo Scientific Orion Conductivity Cell Families
BEST
BETTER
GOOD
A wide range of conductivity cells
are available for every application.
These conductivity cells feature
built-in temperature compensation;
high, standard, and low conductivity
measuring ranges; cell constants
from 0.1 to 10 cm-1 and a variety
of probe materials including
epoxy/graphite, glass/platinum, and
stainless steel.
Rugged
DuraProbe
4-electrode
conductivity cell
Reliable
DuraProbe
4-electrode
conductivity cell
2-electrode
conductivity cell for
ultra-pure water
013005MD A
013010MD A
013605MD A
013610MD A
013610 F
013016MD A
013016A F
013016D B
• Wide
conductivity
range
• Low conductivity
range
DuraProbe™ 4-Electrode
Conductivity Cells
DuraProbe 4-electrode cells provide
the highest accuracy for demanding
laboratory or field applications. The
4-electrode design compensates
for electrode fouling, cable and
connector resistance, polarization
errors, and fringe field interference
errors. The epoxy/graphite cell
material is extremely durable and
chemically resistant. DuraProbe
conductivity cells are ideal for high
and standard conductivity samples
and difficult samples such as
wastewater, runoff water, and mud.
013020MD A
013025MD A
013005A F
013010A F
013005D B
2-Electrode Conductivity Cells
The 2-electrode cells are able to
measure low, standard, and high
conductivity samples, depending
on the cell constant. Cells with a
0.1 cm-1 cell constant are ideal for
low ionic strength solutions, deionized water, and ultra pure water.
Glass/platinum cells are the best
for chemically reactive conductivity
samples, since the cell material is
highly chemical resistant.
• Widest
conductivity
range
• Rugged epoxy/
graphite
• For lab and field
applications
• Rugged epoxy/
graphite body
• Cable lengths
from 1.5 to 10
meters
μS/cm (sample)
0.1
Cell Selection
• For lab and field
applications
1.0
10
mS/cm (sample)
100
10
100
1000
D.I. water
boiler feedwater
K = 0.1
Sample Conductivity Range
drinking water
Conventional 2-Electrode Cells
wastewater
Conventional 4-Electrode Cells
surfacewater
K = 1.0
K = 0.6 and 0.475
K = 1.0
seawater
process water
conc. acids and lyes
K = 10.0
36
Thermo Scientific Orion Laboratory Products Catalog
.0001 M
.001 M .01 M
(14.94 µS)
(147.0 µS)(1.413 mS)(12.9 mS)
.1 M
• For ultra pure
water applications
• Includes
detachable glass
flow cell
conductivity cells
Platinized
2-electrode
conductivity cell
Chemical
resistant
2-electrode
conductivity cell
Precise 2electrode
conductivity cell
Rugged
2-electrode
conductivity cell
High range
2-electrode
conductivity cell
Chemical
resistant
3-electrode
conductivity cell
011020 B
011020A F
011010 B
011010A F
011510MD A
011510 B
018020MD A *
014005 H **
• Low conductivity
range
• Standard
conductivity
range
011050MD A
011050 B
011050A F
• High conductivity
range
• Rugged epoxy/
stainless steel
• For lab
applications
• Removable
guard
• For ultra pure
water applications
• Platinized glass/
platinum
A
B
• Standard
conductivity
range
• For lab
applications
• For lab and field
applications
• Chemical
resistant glass
body
F
• Platinized epoxy/
platinum
• Standard
conductivity
range
• For lab and field
applications
• Rugged epoxy/
graphite
• Chemical
resistant glass
body
• Flow through or
dip cell
H
Key Information
A). MiniDIN connector (Star) B). 8 Pin DIN connector (A+ Series) F). 8 Pin Waterproof
DIN connector (A Series) H). BNC and 2.5 mm Phono Jack connector (Russell)
MiniDIN
8 Pin DIN
WP 8 Pin DIN
2.5 mm phono Jack
* Conductivity cell does not have temperature compensation
** 3-Electrode Conductivity Cell
2009 Sourcebook
37
conductivity cells
Thermo Scientific Orion Conductivity Cell Selection Guide
4-Electrode Conductivity Cells
Cat. No.
Cable Length
Meter Compatibility
Measurement
Range
Application
013005MD A
1.5 m
Star Series
1 μS/cm to 200 mS/cm
Lab and Field
013010MD A
3m
013020MD A
6m
Star Series
10 μS/cm to 200 mS/cm
Lab and Field
1230, 555A, 550A, 550, 162A, 162, 142, 136S, 135A, 131S,
130A, 128
1 μS/cm to 200 mS/cm
Lab and Field
10 μS/cm to 200 mS/cm
Lab and Field
Application
013025MD A
10 m
013605MD A
1.5 m
013610MD A
3m
013005A F
1.5 m
013010A F
3m
013005D B
1.5 m
150 Aplus, 125 Aplus
013610 F
3m
555A, 550A, 162A, 136S, 135A, 131S, 130A
2-Electrode Conductivity Cells
38
Cat. No.
Cable Length
Meter Compatibility
Measurement
Range
013016MD A
1.5 m
Star Series
0.01 μS/cm to 300 μS/cm
013016A F
1m
555A, 550A, 162A, 136S, 135A, 131S, 130A
Boiler feed water,
ultra-pure water
Includes flow cell
013016D B
1m
150Aplus, 145Aplus, 125Aplus, 115Aplus, 105Aplus
011510MD A
3m
Star Series
10 μS/cm to 200 mS/cm
Lab and Field
011510B
3m
150Aplus, 145Aplus, 125Aplus, 115Aplus, 105Aplus
018020MD A
1.5 m
Star Series
10 μS/cm to 2000 mS/cm
High electrolyte
concentrations,
011050MD A
1.5 m
Star Series
1 μS/cm to 20 mS/cm
Lab and Field
011050 B
1m
150Aplus, 145Aplus, 125Aplus, 115Aplus, 105Aplus
011050A F
1m
555A, 550A, 162A, 136S, 135A, 131S, 130A
011020 B
1m
150Aplus, 145Aplus, 125Aplus, 115Aplus, 105Aplus
0.1 μS/cm to 100 μS/cm
Boiler feed water,
ultra-pure water
011020A F
1m
555A, 550A, 162A, 136S, 135A, 131S, 130A
011010 B
1m
150Aplus, 145Aplus, 125Aplus, 115Aplus, 105Aplus
1 μS/cm to 200 mS/cm
Lab
011010A F
1m
555A, 550A, 162A, 136S, 135A, 131S, 130A
014005 H**
1m
RL060C Russell
0 μS/cm to 200 μS/cm
Field
Thermo Scientific Orion Laboratory Products Catalog
conductivity cells
Conductivity Standards
and Accessories
Approximate Cell
Constant
Cell Material
Dimensions
Min/Max Immersion
Cat. No.
Description
0.475 cm-1
Epoxy/Graphite
Dia - 15 mm
Length - 120 mm
35/NA mm
011008
100 μS/cm conductivity/TDS standard,
5 x 60 mL
01100910
147 μS/cm conductivity standard, 10 pouches
011007
1413 μS/cm conductivity/TDS standard,
5 x 60 mL
01100710
1413 μS/cm conductivity/TDS standard,
10 pouches
011006
12.9 mS/cm conductivity/TDS standard,
5 x 60 mL
01100610
12.9 mS/cm conductivity/TDS standard,
10 pouches
011005
111 mS/cm conductivity standard, 5 x 60 mL
01100510
111 mS/cm conductivity standard, 10 pouches
990106
0.1 M KCl conductivity standard, 475 mL
1010001
Conductivity calibration resistor kit for Orion
Star Series conductivity meters, MiniDIN
011220
Conductivity calibration resistor kit for 105A+,
115A+, 125A+, 145A+, and 150A+ meters,
8 Pin DIN
011221
Conductivity calibration resistor kit for 130A,
131S, 135A, 136S, 162A, 550A, and 555A
meters, 8 Pin Waterproof DIN
011001
Conditioning solution for 011050, 011050A, and
011050MD conductivity cells
0105PK
Platinizing kit, includes fixture and solutions
010010
Platinizing solution and vessel
013017
Replacement flow cell for 013016A, 013016D,
and 013016MD
013045
Stainless steel protective probe guard
(for 013005MD, 013010MD, 013025MD,
013005A, 013010A, 013030A, 013060A,
013005D, 013010D)
080045
Stainless steel and plastic protective probe
guard (for 013005MD, 013010MD, 013025MD,
013005A, 013010A, 013030A, 013060A,
013005D, 013010D)
081045
Plastic protective probe guard (for 013005MD,
013010MD, 013025MD, 013005A, 013010A,
013030A, 013060A, 013005D, 013010D)
0.55 cm-1
Epoxy/Graphite
Dia - 12 mm
Length- 120 mm
35/NA mm
0.475 cm-1
Epoxy/Graphite
Dia - 15 mm
Length - 120 mm
35/NA mm
Dia - 12 mm
Length- 120 mm
35/NA mm
0.55 cm-1
Epoxy/Graphite
Approximate Cell
Constant
Cell Material
Dimensions
Min/Max Immersion
0.1 cm-1
Steel, V4A
Dia - 13 mm
Length - 120 mm
Flow Cell Volume
8-12 mL
35/110 mm
1.0 cm-1
Epoxy/Graphite
Dia - 18 mm
Length - 134 mm
35/NA mm
10 cm-1
Glass/Platinum,
Platinized
Dia - 16 mm
Length - 115 mm
55/110 mm
1.0 cm-1
Epoxy/Platinum,
Platinized
Dia - 12 mm
Length - 120 mm
20/90 mm
cm-1
Glass/Platinum,
platinized
Tip Dia - 17 mm
Tip Length - 22 mm
Dia - 13 mm
Length - 120 mm
25/120 mm
1.0 cm-1
Glass/Platinum,
Platinized
Dia - 13 mm
Length - 120 mm
25/120 mm
1.0 cm-1
Epoxy/Steel
Dia - 13 mm
Length - 120 mm
25/120 mm
0.1
A
B
F
H
Key Information
A). MiniDIN connector (Star) B). 8 Pin DIN connector (A+ Series) F). 8 Pin Waterproof
DIN connector (A Series) H). BNC and 2.5 mm Phono Jack connector (Russell)
MiniDIN
8 Pin DIN
WP 8 Pin DIN
2.5 mm phono Jack
* Conductivity cell does not have temperature compensation
** 3-Electrode Conductivity Cell
2009 Sourcebook
39
dissolved oxygen probes
Thermo Scientific Orion RDO® Optical Dissolved Oxygen Probes
RDO technology meets the requirements of ASTM D888-05 Method C
Measuring dissolved oxygen with a luminescence-based sensor
Optical dissolved oxygen is a new
technology with many advantages
that is rapidly gaining acceptance.
BEST
BETTER
GOOD
Rugged, low
maintenance
field probe
The Thermo Scientific Orion RDO probe
provides reliable measurements, no warm
up time, no stirring, no maintenance
(membranes or solution), no thermal
shock and less interferences. The sensor
cap requires easy annual replacement.
The optical sensor cap is durable, able to
withstand harsh environments without
degradation or effect on performance. The
cap has minimal hydration effects and is
not prone to photo bleaching. Calibration
information is stored in the sensor cap
enhancing the ease of use.
NEW
Cat. No.
Recommended Accessories
087001
Replacement RDO sensor cap
087002
Stainless steel protective guard,
for RDO probe
Calibration sleeve for RDO probe
087003
For connector type information, refer to the RDO meter section to
determine the correct meter for your application.
All Thermo Scientific RDO probes come
with a stainless steel guard and calibration
sleeve and are intended for field use.
Optical Sensor/Luminescense-based
Dissolved Oxygen Sensor
Optical Window
Photo Diode
(with Red Filter)
for Returning Light
Sensing Foil
(Lumiphore)
Sensor cap
replaced annually
Sensor Cap
Light Pipe
(for Blue Light)
Blue Filter
Red LED
No membranes
to change
Blue LED
Optical Probe/Luminescense-based Dissolved Oxygen
(LDO) — Newest technology that monitors time
to quench an excited lumiphore, which is inversely
proportional to the concentration of oxygen
Specifications
Probe
Accuracy
+/- 0.1 mg/L from 0 to 8 mg/mL and
+/- 0.2 mg/L from 8 to 20 mg/mL
Response
Time
T90, 30 seconds and T95, 60 seconds
Operating
Range
0 to 20 mg/L and 0 to 200%
Temperature
Range
0 to 50 °C
RDO Probes
40
Thermo Scientific Orion Laboratory Products Catalog
Cat. No.
Description
087010MD
Rugged Dissolved Oxygen probe with
3 meter cable
087020MD
Rugged Dissolved Oxygen probe with
6 meter cable
087030MD
Rugged Dissolved Oxygen probe with
10 meter cable
087050MD
Rugged Dissolved Oxygen probe with
15 meter cable
087100MD
Rugged Dissolved Oxygen probe with
30 meter cable
dissolved oxygen probes
Thermo Scientific Orion Dissolved Oxygen Probe Families
Durable and
accurate DO
probes for the
lab or field
AUTO-STIR™
probe for BOD
applications
Reliable DO
probes for the
lab or field
Rugged DO
probes plus
probe guard
Convert any
pH meter into
a DO meter
A complete line of polarographic
dissolved oxygen probes are
available for BOD (Biochemical
Oxygen Demand) analysis, beer
and wine testing, rugged field
applications, and water and
wastewater analysis. All DO probes
feature automatic temperature
compensation and convenient screw
on membrane caps that reduce
probe servicing time and eliminate
assembly problems.
The AUTO-STIR probe fits directly
into a standard BOD bottle and has
a built-in stirrer. This probe offers
convenience, quick response, and
long life between servicing.
See page 43 for a complete offering of
accessories.
083005MD A
083010MD A
083025MD A
083060MD A
083005A F
083010A F
083005D B
086030MD A, C
086020A G
081010MD A
081010 B
080510MD A
081010F B
970899WP 1
970800 3
• For lab
applications
• For lab and field
applications
• For field
applications
• For lab
applications
• Self-stirring with
One-Touch control
• Rugged epoxy
body
• Includes probe
guard
• Works on pH
meters with BNC
or U.S. Standard
connections
• For lab and field
applications
• Cable lengths
from 1.5 to 20
meters
Key Information
1
3
BNC
Waterproof
U.S. Standard
A
B
MiniDIN
8 Pin DIN
C
F
3.5 mm Phono
Tip Stirrer Jack
8 Pin
Waterproof DIN
G
13 Pin DIN
1). BNC Waterproof connector
3.) US Standard connector
A). MiniDIN connector
B). 8 Pin DIN connector
C). 3.5 mm Phono Tip Stirrer Jack connector
F). 8 Pin Waterproof DIN connector
G). 13 Pin DIN connector
2009 Sourcebook
41
dissolved oxygen probes
Thermo Scientific Orion Dissolved Oxygen Probe Selection Guide
Polarographic Dissolved Oxygen Probes
42
Maximum
Allowed
Overpressure
Sample
Temperature
0.008 μg/h
(mg/L)-1 at
20 °C
6 BAR
0 to 50 °C
10 cm/sec
0.008 μg/h
(mg/L)-1 at
20 °C
6 BAR
0 to 50 °C
90% of final
value in 10
seconds;
95% of final
value in 16
seconds;
99% of final
value in 60
seconds
20 cm/sec
-
-
0 to 50 °C
Field
90% of final
value in 10
seconds;
95% of final
value in 18
seconds;
99% of final
value in 60
seconds
10 cm/sec
0.008 μg/h
(mg/L)-1 at
20 °C
10 BAR
0 to 50 °C
Laboratory
96% response
in less than
30 seconds
between
oxygen-free
and air
saturated
water at 22 °C
-
0.1 mg/hr
-
0 to 45 °C
Meter
Compatibility
Application
Response Time
Laboratory /
Field
Cat. No.
Cable Length
083005A F
1.5 m
830A, 835A,
862A
083010A F
3m
830A, 835A,
862A
083005D B
1.5 m
810Aplus,
850Aplus
083005MD A
1.5 m
Star Series
083010MD A
3m
Star Series
083025MD A
10 m
Star Series
083060MD A
20 m
Star Series
086020A G
2m
862A
086030MD A, C
2m
Star Series,
Benchtop
Models
081010 B
3m
805Aplus,
810Aplus,
850 Aplus
081010MD A
3m
Star Series
081010F B
3m
810Aplus,
850Aplus
080510MD A
3m
Star Series
970800 3*
1m
pH meters
with US
Standard
connector
970899WP 1*
1m
pH meters
with BNC
connector
Thermo Scientific Orion Laboratory Products Catalog
Minimum
Sample Flow
Oxygen
Consumption
90% of final
value in 10
seconds;
95% of final
value in 15
seconds;
99% of final
value in 60
seconds
10 cm/sec
Laboratory
90% of final
value in 10
seconds;
95% of final
value in 15
seconds;
99% of final
value in 60
seconds
Laboratory /
Field
dissolved oxygen probes
Polarographic Dissolved
Oxygen Solutions and
Accessories
Temperature
Sensor
Drift
Dual,
thermally
separated
Approximately
0.1% per day
Electrolyte
lifetime
Storage
Temperature
Membrane Cap
and Solution
Maintenance
Kit
180 days
-5 to 50 °C
Membrane
Cap– 080515
080513
Solution–
080514
Dual,
thermally
separated
Approximately
0.1% per day
180 days
- 5 to 50 °C
Membrane
Cap– 080515
< 1% per day
180 days
- 5 to 50 °C
Membrane
Cap– 081003
Approximately
0.1% per day
180 days
- 5 to 50 °C
Membrane
Cap for
080510MD–
081003
Kit for
080510MD–
080113
Membrane
Cap for
081010F &
083010F–
080515
Kit for
081010F &
083010F–
080513
Solution–
080514
-
Dual,
thermally
separated
-
-
Pre-Filled
Membrane
Cap– 970801
Maintenance Kit, Includes Electrolyte Solution,
Polishing Disk and 2 Membrane Caps for
083005MD, 083010MD, 083025MD, 083060MD,
086030MD, 083005A, 083010A, 083005D,
086020A, 081010F
080113
Maintenance Kit, Includes Electrolyte Solution,
Polishing Disk and 2 Membrane Caps for
080510MD, 081010MD, 081010
080515
1 Membrane Cap for
083005MD, 083010MD, 083025MD,
083060MD, 086030MD, 083005A, 083010A,
083005D, 086020A, 081010F
081003
3 Membrane Caps for
080510MD, 081010MD, 081010
970801
1 Pre-Filled Membrane Cap for
970899WP, 970800
080514
Polarographic Electrolyte Solution for
083005MD, 083010MD, 083025MD,
083060MD, 086030MD, 083005A, 083010A,
083005D, 086020A, 081010F, 080510MD,
081010MD, 081010
080017
Calibration Sleeve for
All DO Probes Excluding AUTO-STIR Probes and
RDO Probe
086021A
Calibration Chamber/Stand for
086020A, 086030MD AUTO-STIR Probes
080360
BOD Adapter for
083005MD, 083010MD, 083025MD,
083060MD, 083005A, 083010A, 083005D
080160
BOD Adapter for
081010MD, 081010
970802
BOD Stir Funnel for
970899WP, 970800
970803
2 Batteries for
970899WP, 970800
013045
Stainless Steel Protective Probe Guard for
080510MD, 083005MD, 083010MD,
083025MD, 083060MD, 083005A, 083005D,
083010A
080045
Stainless Steel and Plastic Protective Probe
Guard for
080510MD, 083005MD, 083010MD,
083025MD, 083060MD, 083005A, 083005D,
083010A
081045
Plastic Protective Probe Guard for
080510MD, 083005MD, 083010MD,
083025MD, 083060MD, 083005A, 083005D,
083010A
080113
Solution–
080514
Dual,
thermally
separated
(080510MD
has a single
temperature
sensor)
Description
080513
080513
Solution–
080514
Single
thermally
separated
Cat. No.
-
* Accuracy– ± 0.05 ppm or 2% of reading, whichever is greater
Battery Life– One year based on 4 hours/day operation
Key Information
1
3
BNC
Waterproof
U.S. Standard
A
B
MiniDIN
8 Pin DIN
C
F
3.5 mm Phono
Tip Stirrer Jack
8 Pin
Waterproof DIN
G
13 Pin DIN
1). BNC Waterproof connector
3.) US Standard connector
A). MiniDIN connector
B). 8 Pin DIN connector
C). 3.5 mm Phono Tip Stirrer Jack connector
F). 8 Pin Waterproof DIN connector
G). 13 Pin DIN connector
2009 Sourcebook
43
star meters
Thermo Scientific Orion Meter Overview
Meters you can count on to deliver accurate, reliable data
Orion DUAL STAR™
3-Star RDO
The products you know:
The best keep getting better:
Our newest additions:
Thermo Scientific Orion Star™ Series
5-Star, 4-Star and 3-Star meters are now
Star Plus meters with the following bonuses:
Thermo Scientific Orion DUAL STAR™
• Bright displays and reference guide
attached to each meter
• SMART STABILITY™ and
SMART AVERAGING™ functions for
optimized results
• Store readings with time and date stamp
• The RS232 connection allows for easy
data downloading and analysis
• Waterproof design for portable meters
• Improved stability algorithm for increased
calibration and measurement accuracy
• Average slope and slope of segment
printed with measurement
• More datalog points with the ability to
turn off datalogging function
• Toggle between simultaneous dual
channel display or single channel display
• Five language choices – English, Spanish,
French, German and Italian
• Informative meter prompts to make this
an easy-to-use, advanced meter
• Increased dissolved oxygen benefits
including:
• EZ Startup™ and shortcut keys for easy
access to meter set-up and features
◗ separate solution and membrane
• CE, CSA, TUV, UL and FCC Class A
limits approvals
◗ printouts containing DO slope and
• 3 year meter replacement warranty
◗ BOD stirrer control on benchtop units
temperature display
separate temperature information
• Enhanced conductivity features such as:
◗ expanded reference temperature
• Obtain temperature-compensated
readings with any pH electrode –
no ATC or 3-in-1 electrode needed,
no temperature connection
compatibility questions
• Works great for ISE readings with
incremental methods and low
concentration range features
• Direct stirrer probe control from
the meter
Thermo Scientific Orion Star Series RDO®
• Optical dissolved oxygen technology
options
◗ cell constant entry up to 3 decimal
points
◗ temperature compensation to 0.01
resolution
• CE, CSA and UL approvals
Thermo Scientific Orion Intrinsically Safe
• Benchtop meter with an adjustable
backlight display
• Temperature calibration capabilities
• Universal power adapter for worldwide
operation included with benchtop meters
Thermo Scientific Orion PerpHecT® LogR™
• Stirrer probe control (AutoStirrer) and
autosampler interface on benchtop meters
• Waterproof and portable
• Approved to meet US intrinsically
safe specifications
44
3-Star pH benchtop meter
Thermo Scientific Orion Laboratory Products Catalog
• No need for membranes or electrolyte
• Waterproof, portable meter design
• Built-in barometric for automatic pressurecorrected readings
star meters
Looking for the Right Thermo Scientific Orion Star™ Meter for Your Job?
Location
Job
Ideal Candidate
Applicant
Resume Highlights
Educational facility
or cost-conscious
laboratory
Part-time position
Easy to work with
Not a lot of training to
get started
Thermo Scientific Orion
2-Star Meter
Includes pull-out reference guide
Measure pH from
clean samples
Work with multiple
users with little lab
experience
QA/QC laboratory
with accuracy needs
Full-time position
0.001 pH resolution
Measure pH from
samples containing
protein/sulfides
Temperature calibration
capabilities
Brings low-maintenance, epoxy-body
pH/ATC electrode with TRIS compatible
experience
Thermo Scientific Orion
3-Star Plus pH Meter
Able to fulfill all requirements of the
ideal candidate
Cat. No. 1112001
Brings a ROSS Ultra® pH/ATC Triode™
electrode, pH buffers and electrode
stand – ready to start on day 1
Pages 62-63
Full-time position
0.001 pH resolution
Need to measure
pH from a variety of
samples
Handles ionic species of
even low concentrations
Thermo Scientific Orion
4-Star Plus pH/ISE
Meter
Work with various
stages of product
samples
Future growth
opportunity to work with
an autosampler
Research laboratory
Full-time position
Good multitasker
Measure pH and
ammonia from samples
throughout the
treatment plant at the
same time
Attention to detail
Full-time position
Measure pH and various
ionic species from
samples at different
testing stages
Easy to work with
and quickly shares
information
May be involved in
complex calculations
involving known
addition/subtraction
Meets all requirements of the ideal
candidate
Cat. No. 1115001
Able to support ISEs and work with
autosampler as need arises
Pages 56-57
Temperature calibration capabilities
Arrives ready to measure pH with a
ROSS Ultra refillable glass pH electrode,
epoxy ATC probe, pH buffers and
electrode stand
Joint project with ISE
periodically
Wastewater laboratory
3 point pH calibration
Page 68
Future growth
opportunity to work with
an autosampler
Involved in product
testing and quality
batch testing
Research laboratory
Cat. No. 1111208
Thermo Scientific Orion
DUAL STAR pH/ISE
Meter
Cat. No. 2115001
Pages 48-49
Displays results from 2 channels
simultaneously
Data storage for up to 1000 points and
20 total methods with calibrations
Screen prompts and hot buttons for
quick access to popular meter functions
Able to handle single and double known
addition/subtraction methods
This candidate has it all: ROSS Ultra
refillable pH glass electrode, stainless
steel ATC probe, electrode stand,
RS232 cable and buffers
Involved in product
developmental, trial
testing and quality
batch testing
Wastewater plant or
watershed, river, lake
or stream
Full-time position
Travel 75%+ of the time
Measure dissolved
oxygen from multiple
sampling locations
Able to handle itself in
water environments
Low-maintenance and
durable
Thermo Scientific Orion
3-Star RDO Meter
DO probe uses no membranes or
electrolyte
Cat. No. 1213302
Stores up to 1000 data sets with time
and date stamp
Pages 50-51
Has its own barometer for automatic
pressure-compensated DO readings
6 meter probe cable length reaches far
and included case keeps it all together
Watershed, river, lake,
stream or estuary
Full-time position
Travel 75%+ of the time
Measure pH, DO,
conductivity and
temperature from a
variety of locations
Multitasker a must
Able to handle itself in
water environments
Low-maintenance and
durable
Thermo Scientific
Orion 5-Star pH/RDO/
Conductivity Meter
Cat. No. 1213322
Pages 50-51
IP67 rated with 6 meter cable
length probes
pH electrode includes temperature
sensor and is low maintenance
DO probe uses no membranes or
electrolyte
Automatic salinity correction of DO
readings with the conductivity probe
Stores up to 500 results
Well organized – comes with a carrying
case and ready to work
2009 Sourcebook
45
star meters
Thermo Scientific Orion Meter Selection Guide
Series
Orion Star™
Model
DUAL STAR
5-Star Plus
5-Star Plus
5-Star Plus
4-Star Plus
4-Star Plus
4-Star Plus
4-Star Plus
3-Star Plus
Parameters
pH/ORP/ISE
pH/ORP/ISE/
Cond/DO
pH/ORP/
Cond/RDO
pH/ORP/
Cond/DO
pH/ORP/ISE
pH/ORP/Cond
pH/ORP/RDO
pH/ORP/DO
pH/ORP
48-49
54-55
50-51
52-53
56-57
58-59
50-51
60-61
62-63
P
P
P
P
P
P
P
P
P
P
P
P
Page(s) in Catalog
Portable Design
P
P
pH
-2.000 to
19.999
-2.000 to
19.999
-2.000 to
19.999
-2.000 to
19.999
-2.000 to
19.999
-2.000 to
19.999
-2.000 to
19.999
-2.000 to
19.999
-2.000 to
19.999
mV/Rel mV/ORP
± 1999.9
± 1999.9
± 1999.9
± 1999.9
± 1999.9
± 1999.9
± 1999.9
± 1999.9
± 1999.9
0 to 19,999
0 to 19,999
Benchtop Design
ISE
0 to 19,999
0.000 to
3000 mS
0.000 to
3000 mS
0.000 to
3000 mS
0.000 to
3000 mS
0.0001 to
100 Mohm
0.0001 to
100 Mohm
0.0001 to
100 Mohm
0.0001 to
100 Mohm
0.1 to
80.0 ppt NaCl
0.1 to
80.0 ppt NaCl
0.1 to
80.0 ppt NaCl
0.1 to
80.0 ppt NaCl
TDS
0 to 19,999
0 to 19,999
0 to 19,999
0 to 19,999
DO
0.00 to 90.0
mg/L,
0.0 to 600%
Conductivity
Resistivity
Salinity
0.00 to 90.0
mg/L,
0.0 to 600%
0.00 to 90.0
mg/L,
0.0 to 600%
RDO (Optical DO)
0.00 to 20.0
mg/L,
0.0 to 200%
0.00 to 20.0
mg/L,
0.0 to 200%
Temperature (RDO only)
0 to 50.0 °C
0 to 50.0 °C
Temperature
-5 to 105.0 °C
-5 to 105.0 °C
-5 to 105.0 °C
-5 to 105.0 °C
-5 to 105.0 °C
-5 to 105.0 °C
-5 to 105.0 °C
-5 to 105.0 °C
-5 to 105.0 °C
1
1
1
1
1
1
1
1
Conductivity/Resistivity/
Salinity/TDS
1*
1*
1*
DO
1
Number of Electrode Probe Connections
pH/ORP and where
available ISE – BNC
2
RDO
Temperature
1
1
1*
Benchtop Versions
with AutoStirrer
P
P
Method Storage
20
10
10
1000
500
GLP
P
Star/Star Plus
Navigator Software
Outputs
46
1
2
Datalog Points
1*
1
1*
P
P
10
10
10
500
500
750
P
P
P
N/A
Included
N/A
RS232, USB
RS232
RS232
Thermo Scientific Orion Laboratory Products Catalog
1*
1
1*
1
1
1
P
P
10
10
10
750
750
750
1000
P
P
P
P
P
Included
Optional
Optional
N/A
Optional
Optional
RS232
RS232
RS232
RS232
RS232
RS232
star meters
PerpHecT® LogR™
Orion Star™
Intrinsically Safe
3-Star Plus
3-Star Plus
3-Star Plus
2-Star Plus
370
350
320
266S
261S
136S
131S
Cond
RDO
DO
pH/ORP
pH/ORP
pH/ORP
pH/mV
pH/mV
pH/mV
Cond
Cond
64-65
50-51
66-67
68
70
70
70
71
71
71
71
P
P
P
P
P
P
P
0.00 to 500 mS
0.00 to 500 mS
0.1 to
80.0 ppt NaCl
0 to 45.0
0 to 45.0
0 to 19,999
0 to 1999 mg/L
0 to 1999 mg/L
20.0 to 120.0 °C
20.0 to 120.0 °C
1
1
1
1
P
P
P
P
P
P
0.000 to
14.999
0.000 to
14.999
0.000 to
14.000
0.000 to
14.000
-2.00 to 16.00
-2.00 to 16.00
± 1999.9
± 1999.9
± 1999.9
± 1999
mV only
± 1300.0
mV only
± 1300.0
mV only
0.000 to 19900
0.000 to
3000 mS
0.0001 to
100 Mohm
0.00 to 90.0
mg/L,
0.0 to 600%
0.00 to 20.0
mg/L,
0.0 to 200%
0 to 50.0 °C
-5 to 105.0 °C
-5 to 105.0 °C
-5 to 105.0 °C
-5 to 105.0 °C
0 to 100.0 °C
0 to 100.0 °C
0 to 100.0 °C
20.0 to 120.0 °C
20.0 to 120.0 °C
1
1
1
1
1
Number of Electrode Probe Connections cont.
1
1*
1
1
1*
1
P
1
1
1
1
1
1
1
P
10
10
10
1000
1000
1000
P
P
P
Optional
N/A
Optional
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
RS232
RS232
RS232
RS232 – printer
format only
RS232,
recorder output
RS232,
recorder output
None
RS232
None
RS232
None
50
100
100
P
P
*The conductivity cell and temperature probe use the same connector.
Only one can be used at a time, and there is a temperature sensor in
the conductivity cell.
2009 Sourcebook
47
star meters
Thermo Scientific Orion DUAL STAR™
Dual Channel pH/ISE Benchtop Meter
Display shows results, electrode
status, ATC temperature and
channel settings. If a method is
selected, the number shows as well
Channel key toggles display to show both
channels simultaneously or just one.
Single channel display also shows electrode
serial number, sample ID, date and time.
Power key turns on/off the
meter and the backlighting
Function keys update according to
the menu for easy selection
Shortcut keys for direct access
to commonly used functions
Numeric keypad for select data
entry in submenus
Features and Benefits
Making it easy to do two
things at once – the Orion
DUAL STAR™ combines
the best of the Orion 720A+/
920A+ and the Orion Star
Plus meters
• Dual channel with
simultaneously displayed
measurements
• Low concentration range
reading stability
• Known addition and subtraction
incremental methods
• 5 language choices: English,
Spanish, French, German
and Italian
• Meter prompts and shortcut
keys
• Thermo Scientific Orion EZ
Startup™ for quick and
easy meter set-up
• Thermo Scientific Orion
SMART STABILITY™ and
SMART AVERAGING™ for
accurate and fast results
48
Thermo Scientific Orion Laboratory Products Catalog
• EH ORP calibration capability
• 10 methods stored per channel
– retains calibrations among
multiple electrodes
• 1000 point datalogging with
time and date stamping
• Calibration, high point and low
point alarms
• Calibration graph with slope
data for on-screen viewing
• Star stirrer probe controlled
directly from the meter
• Autosampler compatibility*
• 3 year meter replacement
warranty
star meters
CE I TUV I EPA
ISE
mV • Rel mV •
ORP
Temperature
pH
Backlight
5 adjustable levels
Range
- 2.000 to 19.999
Method Storage
20 (10 per channel with calibration data)
Resolution
0.1, 0.01, 0.001
Password Protection
Yes
Relative
Accuracy
± 0.002
Datalog Points
1000 with time/date stamp
Calibration
Points
1 to 6
Data Transfer
RS232, USB*
IP Rating
IP54
Auto-Buffer
Recognition
USA/NIST and
EURO/DIN
Power
Universal AC adapter (110V/220V/240V) included with meter
Warranty
36 months replacement
Product Specifications
pH
Features
Range
0 to 19999
Resolution
1,2 or 3 significant
digits
Relative
Accuracy
± 0.2 mV or 0.05%,
whichever is
greater
Units
ppm, ppb, M, % or
unitless
Technique
Options
Single known
addition and
subtraction
Double known
addition and
subtraction
Calibration
Points
2 to 6
Calibration
Options
Linear, point to
point, non-linear
selectable
auto-bank
Range
± 1999.9
ISE
ORP
Cond
DO
Temp
Benchtop Meter Kits
Cat. No.
Description
2115000
• Orion DUAL STAR pH/ISE meter
• stainless steel ATC probe (927007MD)
• swing arm electrode stand (090043)
• RS232 cable (1010053)
• universal power adapter
• user guide
2115101
• Orion DUAL STAR pH/ISE meter
• 8102BNUWP ROSS Ultra® pH electrode
• stainless steel ATC probe (927007MD)
• RS232 cable (1010053)
• swing arm electrode stand (090043)
• universal power adapter
• user guide
2115102
• Orion DUAL STAR pH/ISE meter
Resolution
0.1
• 8157BNUMD ROSS Ultra pH/ATC Triode™ refillable electrode
Relative
Accuracy
± 0.2 or 0.05%,
whichever is
greater
• RS232 cable (1010053)
EH
Calibration
Capability
Yes
Range
- 5 to 105 °C
• stainless steel ATC probe (927007MD)
Resolution
0.1
• RS232 cable (1010053)
Relative
Accuracy
± 0.1 °C
• swing arm electrode stand (090043)
• universal power adapter
• user guide
2115001
• Orion DUAL STAR pH/ISE meter
• 8102BNUWP ROSS Ultra pH electrode
• pint of pH 4.01 buffer (910104)
• pint of pH 7.00 buffer (910107)
• pint of pH 10.01 buffer (910110)
• pint of ROSS® electrode storage solution (810001)
• universal power adapter
• user guide
Accessories
Cat. No.
Description
096019
Thermo Scientific Orion stirrer probe with paddle
910104
pH 4.01 buffer, 475 mL
910107
pH 7.00 buffer, 475 mL
910110
pH 10.01 buffer, 475 mL
*Available as a software update for complimentary download.
Visit www.thermo.com/water in December 2009.
2009 Sourcebook
49
star meters
Thermo Scientific Orion Star RDO®
Optical Dissolved Oxygen Meters
RDO technology meets the requirements of ASTM D888 - 05 Method C –
Measuring dissolved oxygen with a luminescence-based sensor
3-Star RDO Meter
4-Star pH/RDO Meter
5-Star pH/RDO/Conductivity Meter
3-Star RDO Meters Offer:
4-Star pH/RDO Meters Offer:
• Dissolved oxygen and
temperature measurement
capabilities
• Data storage for 1000 data
sets with time and date
stamping
• Dissolved oxygen, pH and
temperature measurement
capabilities
• pH calibration of up to 5 points
with calibration alarm and
automatic buffer recognition of
USA, NIST and DIN buffers
• Data storage for 750 data sets
with time and date stamping
5-Star pH/RDO/Conductivity
Meters Offer:
• Dissolved oxygen, pH,
conductivity, salinity, TDS,
resistivity and temperature
measurement capabilities
• Conductivity / TDS / salinity /
resistivity calibration of up to
5 points and reading reference
temperatures of 5 °C, 10 °C,
15 °C, 20 °C or 25 °C
• Automatic salinity correction
for dissolved oxygen
measurements from the
conductivity probe reading
• Data storage for 500 data sets
with time and date stamping
Features and Benefits for All Versions
Advantages of RDO optical/
luminescence-based dissolved
oxygen (LDO) technology over first
generation LDO technologies on
the market:
• No dehydration concerns,
even if the probe is not
stored in the included
calibration sleeve
• Superior abrasion and
chemical resistance
• Stored data in the RDO
optical/luminescence-based
cap – no codes to enter or
match up
• Dissolved oxygen level
measurements using optical/
luminescence-based dissolved
oxygen (LDO) technology
• Built-in barometer for automatic
pressure correction with option
for manual correction
for dissolved oxygen readings
• No membranes to change –
once-a-year cap replacement
• RS232 port for easy data
downloading and software
updates
• Not affected by color, turbidity
or sulfides
• IP67-rated housing
• RDO probes include calibration
sleeve and stainless steel probe
guard for use anywhere
• Four AA batteries to provide
over 1000 hours of operation in
AUTO-READ™ mode
• Calibrate for dissolved oxygen
using water-saturated air,
air-saturated water, zero point,
or at a custom point with
Winkler titration
• 3 year warranty on meter and
RDO probe body
• Temperature calibration capability
50
Thermo Scientific Orion Laboratory Products Catalog
• 1 year warranty on RDO optical/
luminescence-based cap
star meters
3-Star, 4-Star and
5-Star Meters
CE I CSA I EPA*
Product Specifications
DO
Temperature
Meters Only
Cat. No.
Description
1213300
3-Star RDO meter, batteries and user guide
1213310
4-Star pH/RDO meter, batteries and user guide
1213320
5-Star pH/RDO/Conductivity meter, batteries and user guide
Concentration
Range
0.00 to 20.0 mg/L
Resolution
0.1, 0.01 mg/L
Relative
Accuracy
±0.1 mg/L up to 8 mg/L
±0.2 mg/L from 8 to 20 mg/L
% Saturation
Range
0.0 to 200 %
Resolution
0.1, 1 %
Cat. No.
Description
ORP
Relative
Accuracy
±2%
1213301
3-Star RDO portable meter, RDO probe with 3 meter cable,
stainless steel RDO probe guard, optical/luminescence-based
cap, calibration sleeve, batteries and user guide
Cond
Barometric
Pressure
Correction
450 to 850 mm Hg,
automatic or manual
1213302
3-Star RDO portable meter, RDO probe with 6 meter cable,
stainless steel RDO probe guard, optical/luminescence-based
cap, calibration sleeve, carrying case, batteries and user guide
DO
Calibration
Features
Water saturated air, air saturated water,
manual (Winkler), and zero point
1213311
Probe Type
RDO optical/luminescence-based
Range
0 to 50 °C
4-Star pH/RDO portable meter, RDO probe with 3 meter cable,
stainless steel RDO probe guard, optical/luminescence-based
cap, calibration sleeve, pH/ATC Triode™ with 3 meter cable
(9107WMMD), batteries and user guide
Resolution
0.1 °C
1213312
Relative
Accuracy
± 0.3 °C
4-Star pH/RDO portable meter, RDO probe with 6 meter cable,
stainless steel RDO probe guard, optical/luminescence-based
cap, calibration sleeve, pH/ATC Triode with 6 meter cable
(9107WLMD), carrying case, batteries and user guide
1213321
5-Star pH/RDO/Cond portable meter, RDO probe with 3 meter
cable, stainless steel RDO probe guard, optical/luminescencebased cap, calibration sleeve, pH/ATC Triode with 3 meter cable
(9107WMMD), DuraProbe™ 4-electrode conductivity cell with 3
meter cable (013010MD), batteries and user guide
1213322
5-Star pH/RDO/Cond portable meter, RDO probe with 6 meter
cable, stainless steel RDO probe guard, optical/luminescencebased cap, calibration sleeve, pH/ATC Triode with 6 meter cable
(9107WLMD), DuraProbe 4-electrode conductivity cell with 6
meter cable (013020MD), carrying case, batteries and user guide
4-Star and 5-Star Meters
Product Specifications
pH
mV • Rel mV •
ORP
Temperature
(RDO probe
not used)
Range
-2.000 to 19.999
Resolution
0.1, 0.01, 0.001
Relative
Accuracy
± 0.002
Range
± 1999.9 mV
Resolution
0.1 mV
Cat. No.
Description
Relative
Accuracy
± 0.2 mV or 0.05% of reading,
whichever is greater
087001
Replacement optical/luminescence-based cap
087010MD
Range
-5 to 105 °C
Resolution
0.1 up to 99.9 °C, 1.0 over 99.9 °C
RDO optical/luminescence-based probe with 3 meter
cable, stainless steel guard, calibration sleeve and optical/
luminescence-based cap
087020MD
Relative
Accuracy
± 0.1 °C
RDO optical/luminescence-based probe with 6 meter
cable, stainless steel guard, calibration sleeve and optical/
luminescence-based cap
087030MD
RDO optical/luminescence-based probe with 10 meter
cable, stainless steel guard, calibration sleeve and optical/
luminescence-based cap
087050MD
RDO optical/luminescence-based probe with 15 meter
cable, stainless steel guard, calibration sleeve and optical/
luminescence-based cap
087100MD
RDO optical/luminescence-based probe with 30 meter
cable, stainless steel guard, calibration sleeve and optical/
luminescence-based cap
5-Star Meters
Resistivity
Salinity
TDS
Temp
Accessories
USP
Product Specifications
Conductivity
ISE
Portable Meter Kits
*Check with your regulatory agency for exact acceptance/approval.
pH
Range
0.000 to 3000 mS/cm, auto resolution,
cell constant dependent
Resolution
4 significant digits down to 0.001 μS/cm,
cell constant dependent
Relative
Accuracy
0.5 % ± 1 digit or 0.01 μS/cm,
whichever is greater
087002
Replacement stainless steel RDO optical/luminescencebased probe guard
Cell Constants
0.001 to 199.9 cm-1
087003
Calibration sleeve for RDO optical/luminescence-based
probes
Range
0.0001 to 100 Megohm
9107WMMD
Resolution
Automatic
pH/ATC Triode, 3 meter cable, epoxy body,
low-maintenance gel
Relative
Accuracy
0.5 % ± 1 digit
9107WLMD
pH/ATC Triode, 6 meter cable, epoxy body,
low-maintenance gel
Range
0.1 to 80 ppt NaCl equivalent,
0.1 to 42 ppt practical salinity
013010MD
DuraProbe conductivity cell, 4-electrode, 3 meter cable,
K=0.475 cm-1
Resolution
0.1 ppt
013020MD
Relative
Accuracy
± 0.1 ± 1 digit
DuraProbe conductivity cell, 4-electrode, 6 meter cable,
K=0.475 cm-1
Range
0 to 19999 mg/L
Resolution
1 mg/L
Relative
Accuracy
± 0.5% ± 1 digit
For technical information about RDO optical/luminescencebased dissolved oxygen (LDO) technology; see page 40
2009 Sourcebook
51
star meters
Thermo Scientific Orion 5-Star Plus
pH/ORP/Conductivity/Dissolved Oxygen Portable Meter
One meter to measure a variety of parameters at once –
from a field, stream or any environment
Features and Benefits
• Simultaneous display of pH,
mV or ORP; conductivity;
dissolved oxygen and
temperature measurements on
the backlit LCD
• The Thermo Scientific Orion
SMART STABILITY™ and
SMART AVERAGING™
functions for automatically
optimized accuracy, precision
and response time
• Temperature calibration
capability
• pH calibration of up to 5 points
with calibration alarm and
automatic buffer recognition of
USA, NIST and DIN buffers
• Conductivity / TDS / salinity /
resistivity calibration of up to
5 points and reading reference
temperatures of 5 °C, 10 °C,
15 °C, 20 °C or 25 °C
• Built-in barometer for automatic
pressure correction with
option of manual correction
for dissolved oxygen
measurements
• Automatic salinity correction
for dissolved oxygen
measurements from the
conductivity probe reading
Accessories
Field Cases
page 72
Going out? Keep your meter and
supplies ready to move with our
hard and soft field cases. Perfect
for making accurate measurements
where you need them.
52
Thermo Scientific Orion Laboratory Products Catalog
• Separate dissolved oxygen
membrane and solution
temperature displays and
calibrations
• Full complement of
GLP-related features including
Star Plus Navigator 21 software
• Storage of up to 10 individually
password-protected methods
for easy retrieval of operation
procedures
• Data storage for 500 data sets
with time and date stamping
• RS232 port for easy data
downloading and software
updates
• IP67-rated housing for
waterproof and dustproof
protection
• Four AA batteries deliver over
2,000 hours of continuous
operation
• 3 year meter replacement
warranty
star meters
CE I UL I CSA I TUV I USP I EPA
Product Specifications
pH
mV • Rel mV •
ORP
Conductivity
Range
- 2.000 to 19.999
Resolution
0.1, 0.01, 0.001
Relative
Accuracy
± 0.002
Range
± 1999.9
Resolution
0.1
Relative
Accuracy
± 0.2 or 0.05%,
whichever is
greater
Range
0.000 to 3000 mS/
cm, auto resolution
with cell constant
dependence
Resolution
Relative
Accuracy
Resistivity
Salinity
TDS
DO
Temperature
Meters Only
Package
1218000
Includes portable meter, batteries and user guide
Cell
Constants
0.001 to 199.9 cm-1
Range
0.0001 to 100
Megohm
Resolution
Automatic
Relative
Accuracy
0.5 % ± 1 digit
Range
0.1 to 80.0 ppt
NaCl equivalent,
0.1 to 42 ppt
practical salinity
Resolution
0.1 ppt
Relative
Accuracy
± 0.1 ± 1 digit
Range
0 to 19999 mg/L
Resolution
1 mg/L
Relative
Accuracy
± 0.5 % ± 1 digit
Range
0.00 to 90.0 mg/L
Resolution
0.1, 0.01 mg/L
Relative
Accuracy
± 0.2 mg/L
%
Saturation
Range
0.0 to 600 %
Resolution
0.1, 1 %
Relative
Accuracy
±2%
Barometric
Pressure
Correction
450 to 850 mm
Hg, automatic or
manual
Calibration
Features
Water saturated air,
air saturated water,
manual (Winkler),
and zero point
Probe Type
Polarographic
Range
- 5 to 105 °C
Resolution
0.1 up to 99.9 °C,
1.0 over 99.9 °C
Relative
Accuracy
± 0.1 °C
ISE
Portable Meter Kits
ORP
Cat. No.
Sensor (s)
Accessories and Solutions
1218001
• 9107WMMD low
maintenance epoxy
pH/ATC Triode™ electrode
with 3 meter cable
• Meter impact jacket with electrode
storage chamber
• 083010MD DO probe
with 3 meter cable and
calibration sleeve
• 013010MD DuraProbe™
conductivity cell, K = 0.475,
includes 3 meter cable,
1 μS/cm to 200 mS/cm
4 significant digits
down to 0.001 μS/
cm, cell constant
dependent
0.5% ± 1 digit
or 0.01 μS/cm,
whichever is
greater
pH
Cat. No.
Cond
DO
• Beaker
• Star Plus Navigator 21 software with
RS232 computer cable
Temp
• pH electrode storage solution
(5 x 60 mL bottles)
• pH 4.01, 7.00, 10.01 buffers
(10 x 15 mL pouches)
• DO probe maintenance kit (080513)
• Protective probe guard
• 1413 μS/cm conductivity standard
(10 x 15 mL pouches)
• Hard field case
1218501
• 9107WMMD low
maintenance epoxy
pH/ATC Triode electrode
with 3 meter cable
• 083010MD DO probe
with 3 meter cable and
calibration sleeve
• 013010MD DuraProbe
conductivity cell, K = 0.475,
includes 3 meter cable,
1 μS/cm to 200 mS/cm
• Meter impact jacket with electrode
storage chamber
• Beaker
• Star Plus Navigator 21 software with
RS232 computer cable
• pH electrode storage solution
(5 x 60 mL bottles)
• pH 4.01, 7.00, 10.01 buffers
(10 x 15 mL pouches)
• DO probe maintenance kit (080513)
• Protective probe guard
• 1413 μS/cm conductivity standard
(10 x 15 mL pouches)
• Soft field case
2009 Sourcebook
53
star meters
Thermo Scientific Orion 5-Star Plus
pH/ORP/ISE/Conductivity/Dissolved Oxygen Meters
Features and Benefits
One meter with all the
parameters, including ISE,
to handle every application
in the field, the lab, the
production floor, anywhere
• Simultaneous display of pH,
mV, ORP or ISE; conductivity;
dissolved oxygen and
temperature measurement on
the backlit LCD
• The Thermo Scientific Orion
SMART STABILITY™ and
SMART AVERAGING™
functions for automatically
optimized accuracy, precision
and response time
• Temperature calibration
capability
Accessories
Stirrer Probe
page 25
Mix your sample without a separate
stirrer and don’t worry about a stir bar.
The Thermo Scientific Orion stirrer
probe connects directly to the Star
Plus benchtop meters for sample
mixing the easy way.
54
Thermo Scientific Orion Laboratory Products Catalog
• Separate dissolved oxygen
membrane and solution
temperature displays and
calibrations
• Full complement of GLP-related
features including Star Plus
Navigator 21 software
• Storage of up to 10 individually
password-protected methods
for easy retrieval of operation
procedures
• Data storage for 500 data sets
with time and date stamping
• pH calibration of up to 5 points
with calibration alarm and
automatic buffer recognition of
USA, NIST and DIN buffers
• RS232 port for easy data
downloading and software
updates
• ISE calibration of up to 5 points
with automatic blank correction
feature
• Benchtop units can control an
autosampler and the 096019
stirrer probe (each sold separately)
• Conductivity / TDS / salinity /
resistivity calibration of up to
5 points and reading reference
temperatures of 5 °C, 10 °C,
15 °C, 20 °C or 25 °C
• Benchtop units are splashproof
with an IP54-rated housing
and include a universal power
supply
• Built-in barometer for automatic
pressure correction with
option of manual correction
for dissolved oxygen
measurements
• Automatic salinity correction
for dissolved oxygen
measurements from the
conductivity cell reading
• Portable units are waterproof
with an IP67-rated housing
and run for over 2000 hours on
4 AA batteries
• 3 year meter replacement
warranty
star meters
CE I UL I CSA I TUV I USP I EPA
Cat. No.
Package
Range
- 2.000 to 19.999
1119000
Includes benchtop meter, universal power and user guide
Resolution
0.1, 0.01, 0.001
1219000
Includes portable meter, batteries and user guide
Relative
Accuracy
± 0.002
Product Specifications
pH
ISE
mV • Rel mV •
ORP
Conductivity
Range
0 to 19999
Resolution
1, 2, or 3 significant digits
Relative
Accuracy
± 0.2 mV or 0.05%,
whichever is greater
Units
M, mg/L, %, ppb, or
no units
Calibration
Features
Linear point to point,
non-linear selectable
auto-blank, low
concentration range
stability
Range
± 1999.9
Resolution
0.1
Relative
Accuracy
± 0.2 or 0.05%,
whichever is greater
Range
0.000 to 3000 mS/cm,
auto resolution with cell
constant dependence
Resolution
Resistivity
Salinity
TDS
DO
Temperature
Meters Only
pH
Benchtop Meter Kits
ISE
Cat. No.
Sensor(s)
Accessories and Solutions
1119001
• 8102BNUWP ROSS Ultra®
refillable glass pH electrode
• Star Plus Navigator 21 software with
RS232 computer cable
• 927005MD epoxy ATC probe
• Swing arm electrode stand
• 081010MD DO probe with
3 meter cable
• pH 4.01, 7.00, and 10.01 buffers and pH
electrode storage solution (pint bottles)
DuraProbe™
• 013605MD
conductivity cell, K = 0.55,
10 μS/cm to 200 mS/cm
• 8102BNUWP ROSS Ultra
refillable glass pH electrode
• 013016MD conductivity cell,
K = 0.1, includes flow cell,
0.01 μS/cm to 300 μS/cm
• 096019 stirrer probe
• BOD reservoir, stirrer and adapter
• Star Plus Navigator 21 software with
RS232 computer cable
• Swing arm electrode stand
• pH electrode storage solution
(5 x 60 mL bottles)
• pH 4.01, 7.00, and 10.01 buffers
(pint bottles)
• 100 μS/cm conductivity standard
(5 x 60 mL bottles)
• Method Note - Water Conductivity
USP 645
0.001 to 199.9 cm-1
0.0001 to 100 Megohm
Resolution
Automatic
Relative
Accuracy
1119201
• 8102BNUWP ROSS Ultra
refillable glass pH electrode
• Star Plus Navigator 21 software with
RS232 computer cable
0.5 % ± 1 digit
• 927005MD Epoxy ATC
probe
• Swing arm electrode stand
Range
0.1 to 80.0 ppt NaCl
equivalent, 0.1 to 42 ppt
practical salinity
• 081010MD DO probe with
3 meter cable
• BOD reservoir, stirrer and adapter
Resolution
0.1 ppt
Relative
Accuracy
± 0.1 ± 1 digit
Range
0 to 19999 mg/L
Resolution
1 mg/L
Relative
Accuracy
± 0.5 % ± 1 digit
Concentration
Range
0.00 to 90.0 mg/L
Resolution
0.1, 0.01 mg/L
Relative
Accuracy
± 0.2 mg/L
% Saturation
Range
0.0 to 600 %
Resolution
0.1, 1 %
Relative
Accuracy
±2%
Barometric
Pressure
Correction
450 to 850 mm Hg,
automatic or manual
Calibration
Features
Water saturated air, air
saturated water, manual
(Winkler), and zero point
Probe Type
Polarographic
Range
- 5 to 105 °C
Resolution
0.1 up to 99.9 °C,
1.0 over 99.9 °C
Relative
Accuracy
± 0.1 °C
Temp
• Conductivity calibration resistor kit
0.5% ± 1 digit or 0.01 μS/
cm, whichever is greater
Cell Constants
DO
• DO probe maintenance kit (080113)
• ROSS electrode filling solution
(5 x 60 mL bottles)
4 significant digits down to
0.001 μS/cm, cell constant
dependent
Range
Cond
• DO calibration sleeve
• 1413 μS/cm conductivity standard
(5 x 60 mL bottles)
1010152
ORP
• 013005MD DuraProbe
Conductivity cell, K = 0.475,
1 μS/cm to 200 mS/cm
• DO calibration sleeve
• DO probe maintenance kit (080113)
Portable Meter Kits
Cat. No.
Sensor (s)
Accessories and Solutions
1219001
• 9107WMMD low
maintenance epoxy
pH/ATC Triode electrode with
3 meter cable
• Beaker
• 083010MD DO probe with
3 meter cable and calibration
sleeve
• 013010MD DuraProbe
conductivity cell, K = 0.475,
includes 3 meter cable,
1 μS/cm to 200 mS/cm
• Meter impact jacket with electrode
storage chamber
• Star Plus Navigator 21 software with
RS232 computer cable
• pH electrode storage solution
(5 x 60 mL bottles)
• pH 4.01, 7.00, 10.01 buffers
(10 x 15 mL pouches)
• Protective probe guard
• DO probe maintenance kit (080513)
• 1413 μS/cm conductivity standard
(10 x 15 mL pouches)
• Hard field case
1219101
• 9107WMMD low
maintenance epoxy
pH/ATC Triode electrode with
3 meter cable
• Meter impact jacket with electrode
storage chamber
• 083010MD DO probe with
3 meter cable and calibration
sleeve
• DO probe maintenance kit (080513)
• Star Plus Navigator 21 software with
RS232 computer cable
• 013010MD DuraProbe
conductivity cell, K = 0.475,
includes 3 meter cable,
1 μS/cm to 200 mS/cm
2009 Sourcebook
55
star meters
Thermo Scientific Orion 4-Star Plus
pH/ISE Meters
Features and Benefits
• Simultaneous display of pH,
mV or ISE and temperature
measurements on the backlit
LCD
Ideal for optimum ISE
measurements from
any sample
• The Thermo Scientific Orion
SMART STABILITY™ and
SMART AVERAGING™
functions for automatically
optimized accuracy, precision
and response time
Accessories
Star Plus
Navigator Software
page 69
Thermo Scientific Orion Star Plus
Navigator software makes it easy
to interface your meter with the
computer. Generate real-time data
graphs, download meter data with
one click and perform data analysis
the easy way. The data security are
ideal for meeting FDA 21 CFR Part 11
compliance
56
Thermo Scientific Orion Laboratory Products Catalog
• Data storage for 750 points
with time and date stamp
• Average slope and slope
of segment printed with
measurements
• RS232 port for easy data
downloading and software
communication
• Temperature calibration
capability
• Benchtop units can control an
autosampler and the 096019
stirrer probe (each sold separately)
• pH calibration of up to 5 points
with calibration alarm and
automatic buffer recognition of
USA, NIST and DIN buffers
• Benchtop units are splashproof
with an IP54-rated housing
and include a universal power
supply
• ISE calibration of up to 5 points
with automatic blank correction
feature
• Portable units are waterproof
with an IP67-rated housing and
run for over 2,000 hours on four
AA batteries
• Storage of up to 10 individually
password-protected methods
for easy retrieval of operation
procedures
• 3 year meter replacement
warranty
star meters
CE I UL I CSA I TUV I EPA
Product Specifications
pH
ISE
Range
- 2.000 to 19.999
Resolution
0.1, 0.01, 0.001
Relative
Accuracy
± 0.002
Range
0 to 19999
Resolution
1, 2, or 3 significant
digits
Relative
Accuracy
± 0.2 mV or 0.05%,
whichever is greater
Units
M, mg/L, %, ppb, or
no units
Calibration
Features
mV • Rel mV •
ORP
Temperature
Range
Linear point to point,
non-linear selectable
auto-blank, low
concentration range
stability
Meters Only
pH
Cat. No.
Package
1115000
Includes benchtop meter, universal power and user guide
1215000
Includes portable meter, batteries and user guide
0.1
± 0.2 or 0.05%,
whichever is greater
Range
- 5 to 105 °C
Resolution
0.1 up to 99.9 °C,
1.0 over 99.9 °C
Relative
Accuracy
± 0.1 °C
Cond
Cat. No.
Sensor (s)
Accessories and Solutions
1115001
• 8102BNUWP ROSS Ultra®
refillable glass pH electrode
• Swing arm electrode stand
1115002
1115005
1010109
Relative
Accuracy
ORP
Benchtop Meter Kits
• 927005MD epoxy ATC
probe
• 9172BNWP Sure-Flow®
refillable glass pH electrode
• 927005MD epoxy ATC
probe
• 8611BNWP ROSS® sodium
ion selective electrode
1010103
1010108
1010102
DO
• pH 4.01, 7.00, and 10.01 buffers and pH
electrode storage solution (pint bottles)
Temp
• Swing arm electrode stand
• pH 4.01, 7.00, and 10.01 buffers and pH
electrode storage solution (pint bottles)
• Sodium standards and solutions
(pint bottles)
• 9512HPBNWP ammonia ion
selective electrode
• Swing arm electrode stand
• Ammonia standards and solutions
(pint bottles)
• 096019 stirrer probe
• Swing arm electrode stand
• 9609BNWP fluoride ion
selective electrode
• Method Note - Ammonia in Wastewater
• Fluoride standards and solutions
(pint bottles)
• 096019 stirrer probe
• Swing arm electrode stand
• 9707BNWP nitrate ion
selective electrode
• Method Note - Fluoride in Drinking Water
• Nitrate standards and solutions
(pint bottles)
• 096019 stirrer probe
• Swing arm electrode stand
• 9512HPBNWP ammonia ion
selective electrode
• Method Note - Nitrate in Drinking Water
• Ammonia standards and solutions
(pint bottles)
• 096019 stirrer probe
• Swing arm electrode stand
± 1999.9
Resolution
ISE
• Method Note - Ammonia in Drinking
Water
Portable Meter Kits
Cat. No.
Sensor (s)
Accessories and Solutions
1215001
• 9107BNMD low
maintenance epoxy
pH/ATC Triode™ electrode
• pH electrode storage solution
(5 x 60 mL bottles)
1215501
• 9107BNMD low
maintenance epoxy
pH/ATC Triode electrode
• pH 4.01, 7.00, 10.01 buffers
(10 x 15 mL pouches)
• Hard field case
• pH electrode storage solution
(5 x 60 mL bottles)
• pH 4.01, 7.00, 10.01 buffers
(10 x 15 mL pouches)
• Soft field case
Signifies a hazardous solution. See terms and conditions for
important shipping information at www.thermo.com/water.
2009 Sourcebook
57
star meters
Thermo Scientific Orion 4-Star Plus
pH/Conductivity Meters
Features and Benefits
Measure pH and
conductivity with one
meter, whether it’s for water
quality or pharmacology
• Simultaneous display of pH,
conductivity and temperature
measurements on the backlit
LCD
• Storage of up to 10 individually
password-protected methods
for easy retrieval of operation
procedures
• The Thermo Scientific Orion
SMART STABILITY™ and
SMART AVERAGING™
functions for automatically
optimized accuracy, precision
and response time
• Datalog up to 750 points with
time and date stamp
• Meet all U.S. Pharmacopeia
(USP) requirements
• RS232 port for easy data
downloading and software
updates
• Temperature calibration
capability
• pH calibration of up to 5 points
with calibration alarm and
automatic buffer recognition of
USA, NIST and DIN buffers
Accessories
USB Adapter
page 72
Do you have plenty of USB ports,
but no open COM ports? The RS232
to USB cable adapter 1010005 can
be used to download data from your
meter to your computer through a
USB connection.
58
Thermo Scientific Orion Laboratory Products Catalog
• Conductivity / TDS / salinity /
resistivity calibration of up to 5
points with reading reference
temperatures at 5 °C, 10 °C,
15 °C, 20 °C or 25 °C
• Compatible with 2-electrode
and 4-electrode conductivity
cells
• Average slope and slope of
segment printed with pH
measurements
• Benchtop units can control an
autosampler and the 096019
stirrer probe (each sold separately)
• Benchtop units are splashproof
with an IP54-rated housing
and include a universal power
supply
• Portable units are waterproof
with an IP67-rated housing and
run for over 2,000 hours on four
AA batteries
• 3 year meter replacement
warranty
star meters
CE I UL I CSA I TUV I USP I EPA
Product Specifications
pH
mV • Rel mV •
ORP
Conductivity
Salinity
TDS
Temperature
pH
Cat. No.
Package
Range
- 2.000 to 19.999
1117000
Includes benchtop meter, universal power and user guide
Resolution
0.1, 0.01, 0.001
1217000
Includes portable meter, batteries and user guide
Relative
Accuracy
± 0.002
Range
± 1999.9
Resolution
0.1
Relative
Accuracy
± 0.2 or 0.05%,
whichever is
greater
Range
0.000 to 3000 mS/
cm, auto resolution
with cell constant
dependence
Resolution
Resistivity
Meters Only
4 significant digits
down to 0.001 μS/
cm, cell constant
dependent
Relative
Accuracy
0.5% ± 1 digit or
0.01 μS/cm,
whichever is
greater
Cell
Constants
0.001 to 199.9 cm-1
Range
0.0001 to 100
Megohm
Resolution
Automatic
Relative
Accuracy
0.5 % ± 1 digit
Range
0.1 to 80.0 ppt NaCl
equivalent,
0.1 to 42 ppt
practical salinity
Resolution
0.1
Relative
Accuracy
± 0.1 ± 1 digit
Range
0 to 19999
Resolution
1 mg/L
Relative
Accuracy
± 0.5 % ± 1 digit
Range
- 5 to 105 °C
Resolution
0.1 up to 99.9 °C,
1.0 over 99.9 °C
Relative
Accuracy
± 0.1 °C
ISE
ORP
Benchtop Meter Kits
Cond
Cat. No.
Sensor (s)
Accessories and Solutions
1117001
• 9157BNMD refillable epoxy
pH/ATC Triode™ electrode
• Swing arm electrode stand
• 013605MD DuraProbe™
conductivity cell, K = 0.55,
10 μS/cm to 200 mS/cm
DO
• pH electrode storage solution
(5 x 60 mL bottles)
Temp
• pH 4.01, 7.00, 10.01 buffers
(pint bottles)
• 1413 μS/cm conductivity standard
(5 x 60 mL bottles)
1117201
• 9157BNMD refillable epoxy
pH/ATC Triode electrode
• Swing arm electrode stand
• 013005MD DuraProbe
conductivity cell, K = 0.475,
1 μS/cm to 200 mS/cm
Portable Meter Kits
Cat. No.
Sensor (s)
Accessories and Solutions
1217001
• 9107BNMD low
maintenance epoxy
pH/ATC Triode electrode
• pH electrode storage solution
(5 x 60 mL bottles)
• 013005MD DuraProbe
conductivity cell, K = 0.475,
1 μS/cm to 200 mS/cm
• pH 4.01, 7.00, 10.01 buffers
(10 x 15 mL pouches)
• 1413 μS/cm conductivity standard
(10 x 15 mL pouches)
• Hard field case
1217002
• 9107WMMD low
maintenance epoxy
pH/ATC Triode electrode
with 3 meter cable
• 013010MD DuraProbe
conductivity cell, K = 0.475,
includes 3 meter cable,
1 μS/cm to 200 mS/cm
• Beaker
• pH electrode storage solution
(5 x 60 mL bottles)
• pH 4.01, 7.00, 10.01 buffers
(10 x 15 mL pouches)
• 1413 μS/cm conductivity standard
(10 x 15 mL pouches)
• Hard field case
1217501
• 9107BNMD low
maintenance epoxy
pH/ATC Triode electrode
• Beaker
• 013005MD DuraProbe
conductivity cell, K = 0.475,
1 μS/cm to 200 mS/cm
• pH 4.01, 7.00, 10.01 buffers
(10 x 15 mL pouches)
• pH electrode storage solution
(5 x 60 mL bottles)
• 1413 μS/cm conductivity standard
(10 x 15 mL pouches)
• Soft field case
1217101
• 9107BNMD low
maintenance epoxy
pH/ATC Triode electrode
—
• 013005MD DuraProbe
conductivity cell, K = 0.475,
1 μS/cm to 200 mS/cm
2009 Sourcebook
59
star meters
Thermo Scientific Orion 4-Star Plus
pH/Dissolved Oxygen Meters
Features and Benefits
Check pH and dissolved
oxygen levels at the same
time from a variety of
water sources
Thermo Scientific Orion Laboratory Products Catalog
• Data storage for up to 750
points with time and date
stamp
• Average slope and slope of
segment printed with pH
measurements
• RS232 port for easy data
downloading and software
updates
• pH calibration of up to 5 points
with calibration alarm and
automatic buffer recognition of
USA, NIST and DIN buffers
• Portable units are waterproof
with an IP67-rated housing and
run for over 2,000 hours on four
AA batteries
• Salinity correction and built-in
automatic barometric pressure
compensation for dissolved
oxygen measurements
• Benchtop units can control an
autosampler and the 096019
stirrer probe (each sold separately)
pages 26-27
A great calibration requires great
buffers and solutions. Make sure what
you are using is fresh and the correct
range when you perform calibrations.
Thermo Scientific Orion buffers and
solutions come in a variety of points,
volumes and packaging to meet your
quality needs.
60
• The Thermo Scientific Orion
SMART STABILITY™ and
SMART AVERAGING™
functions for automatically
optimized accuracy, precision
and response time
• Separate dissolved oxygen
membrane and solution
temperature displays and
calibrations
Accessories
Buffers
• Simultaneous display of
pH, dissolved oxygen and
temperature measurements on
the backlit LCD
• Dissolved oxygen calibration to
the Winkler Titration for BOD
analysis
• Storage of up to 10 individually
password-protected methods
for easy retrieval of operation
procedures
• Benchtop units are splashproof
with an IP54-rated housing
and include a universal power
supply
• 3 year meter replacement
warranty
star meters
CE I UL I CSA I TUV I EPA
Product Specifications
pH
mV • Rel mV •
ORP
DO
Temperature
Meters Only
pH
Cat. No.
Package
Range
- 2.000 to 19.999
1116000
Includes benchtop meter, universal power and user guide
Resolution
0.1, 0.01, 0.001
1216000
Includes portable meter, batteries and user guide
Relative
Accuracy
± 0.002
Range
± 1999.9
Resolution
0.1
Relative
Accuracy
± 0.2 or 0.05%,
whichever is
greater
Concentration
Range
0.00 to 90.0 mg/L
Resolution
0.01, 0.1 mg/L
Relative
Accuracy
± 0.2 mg/L
% Saturation
Range
0.0 to 600 %
Resolution
0.1, 1 %
Relative
Accuracy
±2%
Barometric
Pressure
Correction
450 to 850 mm
Hg, automatic or
manual
Calibration
Features
Water saturated air,
air saturated water,
manual (Winkler),
and zero point
ORP
Benchtop Meter Kits
Cond
Cat. No.
Sensor (s)
Accessories and Solutions
1116001
• 8102BNUWP ROSS Ultra®
refillable glass pH electrode
• Swing arm electrode stand
• 927005MD epoxy ATC probe
• 081010MD DO probe with
3 meter cable
DO
• pH 4.01, 7.00, and 10.01 buffers
(pint bottles)
Temp
• DO calibration sleeve
• BOD reservoir, stirrer and adapter
• DO probe maintenance kit (080113)
1116201
• 9157BNMD refillable epoxy
pH/ATC Triode™ electrode
• Swing arm electrode stand
• 081010MD DO probe with
3 meter cable
• BOD reservoir, stirrer and adapter
• DO calibration sleeve
• DO probe maintenance kit (080113)
Portable Meter Kits
Cat. No.
Sensor (s)
Accessories and Solutions
1216001
• 9107BNMD low
maintenance epoxy
pH/ATC Triode electrode
• Beaker
• 083005MD DO probe with
calibration sleeve
• pH 4.01, 7.00, 10.01 buffers
(10 x 15 mL pouches)
• pH electrode storage solution
(5 x 60 mL bottles)
Probe Type
Polarographic
• DO probe maintenance kit (080513)
Range
- 5 to 105 °C
• Protective probe guard
Resolution
0.1 up to 99.9 °C,
1.0 over 99.9 °C
Relative
Accuracy
ISE
• Hard field case
1216002
± 0.1 °C
• 9107WMMD low
maintenance epoxy
pH/ATC Triode electrode with
3 meter cable
• 083010MD DO probe
with 3 meter cable and
calibration sleeve
• Beaker
• pH electrode storage solution
(5 x 60 mL bottles)
• pH 4.01, 7.00, 10.01 buffers
(10 x 15 mL pouches)
• DO probe maintenance kit (080513)
• Protective probe guard
• Hard field case
1216502
• 9107WMMD low
maintenance epoxy
pH/ATC Triode electrode with
3 meter cable
• 083010MD DO probe
with 3 meter cable and
calibration sleeve
• Beaker
• pH electrode storage solution
(5 x 60 mL bottles)
• pH 4.01, 7.00, 10.01 buffers
(10 x 15 mL pouches)
• DO probe maintenance kit (080513)
• Protective probe guard
• Soft field case
2009 Sourcebook
61
star meters
Thermo Scientific Orion 3-Star Plus
pH Meters
Features and Benefits
• Simultaneous display of pH,
mV or ORP, and temperature
measurements on the backlit
LCD
The ideal meter for
measuring pH in water,
process, research and
QA/QC testing
• The Thermo Scientific Orion
SMART STABILITY™ and
SMART AVERAGING™
functions for automatically
optimized accuracy, precision
and response time
• Temperature calibration
capability
• pH calibration of up to 5 points
with calibration alarm
• Automatic buffer recognition of
USA, NIST and DIN buffers
• Storage of up to 10 individually
password protected methods
for easy retrieval of operation
procedures
• Datalog up to 1000 points with
time and date stamp
Accessories
Stirrer Probe
page 25
Mix your sample without a separate
stirrer and don’t worry about a stir bar.
The Thermo Scientific Orion stirrer
probe connects directly to the Star
Plus benchtop meters for sample
mixing the easy way.
62
Thermo Scientific Orion Laboratory Products Catalog
• Average slope and slope of
segment printed with pH
measurements
• RS232 port for data downloads
and software updates
• Benchtop units can control an
autosampler and the 096019
stirrer probe (each sold separately)
• Benchtop units are splashproof
with an IP54-rated housing
and include a universal power
supply
• Portable units are waterproof
with an IP67-rated housing and
run for over 2,000 hours on four
AA batteries
• 3 year meter replacement
warranty
star meters
CE I UL I CSA I TUV I EPA
Product Specifications
pH
mV • Rel mV •
ORP
Temperature
Meters Only
Cat. No.
Package
Range
- 2.000 to 19.999
1112000
Includes benchtop meter, universal power and user guide
Resolution
0.1, 0.01, 0.001
1212000
Includes portable meter, batteries and user guide
Relative
Accuracy
± 0.002
Range
± 1999.9
Resolution
0.1
Cat. No.
Sensor (s)
Accessories and Solutions
Relative
Accuracy
± 0.2 or 0.05%,
whichever is
greater
1112001
• 8157BNUMD ROSS
Ultra® refillable
pH/ATC Triode electrode
• Swing arm electrode stand
Range
- 5 to 105 °C
Resolution
0.1 up to 99.9 °C,
1.0 over 99.9 °C
• Swing arm electrode stand
Relative
Accuracy
± 0.1 °C
• 8102BNUWP ROSS
Ultra refillable glass pH
electrode
1112003
1112004
Cat. No.
Sensor (s)
Accessories and Solutions
1212001
• 9107BNMD low
maintenance epoxy
pH/ATC Triode™
electrode
• Beaker
• Electrode storage chamber
1112006
• pH electrode storage solution
(5 x 60 mL bottles)
• Hard field case
• Beaker
• pH electrode storage solution
(5 x 60 mL bottles)
• pH electrode storage solution
(5 x 60 mL bottles)
• pH 4.01, 7.00, 10.01 buffers
(10 x 15 mL pouches)
• Soft field case
1212503
• 9179BNMD low
maintenance epoxy
ORP/ATC Triode
electrode
1010115
• 9107BNMD low
maintenance epoxy
pH/ATC Triode electrode
—
• 9107BNMD low
maintenance epoxy
pH/ATC Triode electrode
• Meter impact jacket with
electrode storage chamber
• Swing arm electrode stand
• pH 4.01, 7.00, and 10.01 buffers
and pH electrode storage
solution (pint bottles)
• Swing arm electrode stand
• pH 4.10, 6.97, and 9.15 pure
water buffers and pH electrode
storage solution (pint bottles)
• pHISA adjuster solution
(5 x 60 mL bottles)
1112202
• 8172BNWP ROSS
Sure-Flow refillable
glass pH electrode
• Swing arm electrode stand
• 927006MD glass ATC
probe
1112203
• 9156APWP AquaPro
low maintenance epoxy
pH electrode
• Swing arm electrode stand
1112204
• 9107BNMD low
maintenance epoxy
pH/ATC Triode electrode
• Swing arm electrode stand
1112205
• 9156BNWP refillable
epoxy pH electrode
• Swing arm electrode stand
• 927007MD stainless
steel ATC probe
1112102
• Soft field case
1212101
Temp
• pH 4.01, 7.00, and 10.01 buffers
and pH electrode storage
solution (pint bottles)
• Swing arm electrode stand
• Meter impact jacket with
electrode storage chamber
• ORP standard (5 x 60 mL
bottles)
DO
• pH 4.01, 7.00, and 10.01 buffers
and pH electrode storage
solution (pint bottles)
• 9157BNMD refillable
epoxy pH/ATC Triode
electrode
• Beaker
• Electrode storage chamber
ORP
Cond
1112201
• Hard field case
• 9107BNMD low
maintenance epoxy
pH/ATC Triode electrode
• 8102BNUWP ROSS
Ultra refillable glass pH
electrode
• 927005MD epoxy ATC
probe
• pH 4.01, 7.00, 10.01 buffers
(10 x 15 mL pouches)
1212501
• 9172BNWP Sure-Flow®
refillable glass pH
electrode
• 927007MD stainless
steel ATC probe
• pH 4.01, 7.00, 10.01 buffers
(10 x 15 mL pouches)
• 5107BNMD No Cal®
refillable pH/ATC
electrode
ISE
Benchtop Meter Kits
• 927005MD epoxy ATC
probe
Portable Meter Kits
1212002
pH
• 8172BNWP ROSS
Sure-Flow refillable
glass pH electrode
—
• 927006MD glass ATC
probe
1112002
—
• Swing arm electrode stand
• Electrode rinse solution
(25 x 15 mL pouches)
• pH electrode storage solution
(5 x 60 mL bottles)
• pH 4.01, 7.00 and 10.01 buffers
(10 x 15 mL pouches)
• Soft field case
• Method Note - pH of Salt Water
2009 Sourcebook
63
star meters
Thermo Scientific Orion 3-Star Plus
Conductivity Meters
Features and Benefits
Ideal meter to help meet
USP requirements
• Simultaneous display of
conductivity, resistivity, TDS
or salinity and temperature
measurements on the backlit LCD
• Storage of up to 10 individually
password-protected methods
for easy retrieval of operation
procedures
• The Thermo Scientific Orion
SMART STABILITY™ and SMART
AVERAGING™ functions for
automatically optimized accuracy,
precision and response time
• Datalog up to 1000 points with time
and date stamp
• Meet all U.S. Pharmacopeia (USP)
requirements
• Accept cell constants from 0.001
to 199.9 cm-1
• Compatible with 2-electrode and
4-electrode conductivity cells
• Temperature calibration capability
• Conductivity / TDS / salinity /
resistivity calibration of up to
5 points
• Reading reference temperatures of
5 °C, 10 °C, 15 °C, 20 °C or 25 °C
Accessories
Field Cases
page 72
Going out? Keep your meter and
supplies ready to move with our
hard and soft field cases. Perfect
for making accurate measurements
where you need them.
64
Thermo Scientific Orion Laboratory Products Catalog
• RS232 port for easy data
downloading and software updates
• Benchtop units can control an
autosampler and the 096019 stirrer
probe (each sold separately)
• Benchtop units are splashproof with
an IP54-rated housing and include a
universal power supply
• Portable units are waterproof with
an IP67-rated housing and run
for over 2,000 hours on four AA
batteries
• 3 year meter replacement warranty
star meters
CE I UL I CSA I TUV I USP I EPA
Product Specifications
Conductivity
Range
Resolution
Relative
Accuracy
Resistivity
Salinity
TDS
Temperature
0.000 to 3000 mS/
cm, auto resolution
with cell constant
dependence
4 significant digits
down to 0.001 μS/
cm, cell constant
dependent
0.5% ± 1 digit or
0.01 μS/cm,
whichever is
greater
Cell
Constants
0.001 to 199.9
Range
0.0001 to 100
Megohm
Resolution
Automatic
Relative
Accuracy
0.5 % ± 1 digit
Range
0.1 to 80.0 ppt NaCl
equivalent,
0.1 to 42 ppt
practical salinity
Resolution
0.1
Relative
Accuracy
± 0.1 ± 1 digit
Range
0 to 19999
Resolution
1 mg/L
Relative
Accuracy
± 0.5 % ± 1 digit
Range
- 5 to 105 °C
Resolution
0.1 up to 99.9 °C,
1.0 over 99.9 °C
Relative
Accuracy
Meters Only
pH
Cat. No.
Package
1114000
Includes benchtop meter, universal power and user guide
1214000
Includes portable meter, batteries and user guide
ISE
ORP
Benchtop Meter Kits
Cond
Cat. No.
Sensor (s)
Accessories and Solutions
1114001
• 013005MD DuraProbe™
conductivity cell, K = 0.475,
1 μS/cm to 200 mS/cm
• 1413 μS/cm conductivity standard
(5 x 60 mL bottles)
DO
• Swing arm electrode stand
1114002
• 013016MD conductivity cell,
K = 0.1, includes flow cell,
0.01 μS/cm to 300 μS/cm
• 100 μS/cm conductivity standard
(5 x 60 mL bottles)
Temp
1114003
• 013016MD conductivity cell,
K = 0.1, includes flow cell,
0.01 μS/cm to 300 μS/cm
• 100 μS/cm conductivity standard
(5 x 60 mL bottles)
• Swing arm electrode stand
• Conductivity calibration resistor kit
• Swing arm electrode stand
Portable Meter Kits
Cat. No.
Sensor (s)
Accessories and Solutions
1214001
• 013005MD DuraProbe
conductivity cell, K = 0.475,
1 μS/cm to 200 mS/cm
• 1413 μS/cm conductivity standard
(10 x 15 mL pouches)
1214003
• 013610MD conductivity cell,
K = 0.55, includes 3 meter
cable, 10 μS/cm to
200 mS/cm)
• 1413 μS/cm conductivity standard
(10 x 15 mL pouches)
1214004
• 013016MD conductivity cell,
K = 0.1, includes flow cell,
0.01 μS/cm to 300 μS/cm
• 100 μS/cm conductivity standard
(5 x 60 mL bottles)
1214501
• 013005MD DuraProbe
conductivity cell, K = 0.475,
1 μS/cm to 200 mS/cm
• 1413 μS/cm conductivity standard
(10 x 15 mL pouches)
1214504
• 013016MD conductivity cell,
K = 0.1, includes flow cell,
0.01 μS/cm to 300 μS/cm
• 100 μS/cm conductivity standard
(5 x 60 mL bottles)
1214101
• 013005MD DuraProbe
conductivity cell, K = 0.475,
1 μS/cm to 200 mS/cm
—
± 0.1 °C
• Hard field case
• Hard field case
• Hard field case
• Soft field case
• Soft field case
2009 Sourcebook
65
star meters
Thermo Scientific Orion 3-Star Plus
Dissolved Oxygen Meters
Features and Benefits
Ensure proper oxygen levels
for maximum control
• Simultaneous display of
dissolved oxygen and
temperature measurements on
the backlit LCD
• The Thermo Scientific Orion
SMART STABILITY™ and
SMART AVERAGING™
functions for automatically
optimized accuracy, precision
and response time
• Separate dissolved oxygen
membrane and solution
temperature displays and
calibrations
• Salinity correction and built-in
automatic barometric pressure
compensation for dissolved
oxygen measurements
• Dissolved oxygen calibration to
the Winkler titration for BOD
analysis
Accessories
Star Plus
Navigator Software
page 69
Thermo Scientific Orion Star Plus
Navigator software makes it easy
to interface your meter with the
computer. Generate real-time data
graphs, download meter data with
one click and perform data analysis
the easy way. The data security are
ideal for meeting FDA 21 CFR Part 11
compliance
66
Thermo Scientific Orion Laboratory Products Catalog
• Storage of up to 10 individually
password-protected methods
for easy retrieval of operation
procedures
• Dissolved oxygen slope on
printout
• Datalog up to 1000 points with
time and date stamp
• RS232 port for easy data
downloads and software
update
• Portable units are waterproof
with an IP67-rated housing and
run for over 2,000 hours on four
AA batteries
• Benchtop units can control an
autosampler and the 096019
stirrer probe (each sold separately)
• Benchtop units are splashproof
with an IP54-rated housing
and include a universal power
supply
• 3 year meter replacement
warranty
star meters
CE I UL I CSA I TUV I EPA
Product Specifications
DO
Temperature
Concentration
Range
0.00 to 90.0 mg/L
Resolution
0.01, 0.1 mg/L
Relative
Accuracy
± 0.2 mg/L
% Saturation
Range
0.0 to 600 %
Resolution
0.1, 1 %
Relative
Accuracy
±2%
Barometric
Pressure
Correction
450 to 850 mm
Hg, automatic or
manual
Meters Only
pH
Cat. No.
Package
1113000
Includes benchtop meter, universal power and user guide
1213000
Includes portable meter, batteries and user guide
ISE
ORP
Benchtop Meter Kits
Cond
Cat. No.
Sensor (s)
Accessories and Solutions
1113001
• 081010MD DO probe with
3 meter cable
• DO calibration sleeve
DO
• BOD funnel / stirrer (970802)
• BOD adapter (080160)
Calibration
Features
Water saturated air,
air saturated water,
manual (Winkler),
and zero point
Probe Type
Polarographic
Range
- 5 to 105 °C
Resolution
0.1 up to 99.9 °C,
1.0 over 99.9 °C
Relative
Accuracy
± 0.1 °C
Temp
• DO probe maintenance kit (080113)
• Swing arm electrode stand
1010114
• 086030MD AUTO-STIR™
DO/BOD probe with
calibration sleeve
• Polarographic electrolyte solution
• Method Note - BOD of Wastewater
Portable Meter Kits
Cat. No.
Sensor (s)
Accessories and Solutions
1213101
• 083010MD DO probe
with 3 meter cable and
calibration sleeve
• DO probe maintenance kit (080513)
1213001
• 083010MD DO probe
with 3 meter cable and
calibration sleeve
• DO probe maintenance kit (080513)
• 083010MD DO probe
with 3 meter cable and
calibration sleeve
• DO probe maintenance kit (080513)
• 083010MD DO probe
with 3 meter cable and
calibration sleeve
• Electrode storage chamber
1213501
1010121
• Hard field case
• Soft field case
• Polarographic electrolyte solution
• Hard field case
• Method Note - DO in Freshwater
1010138
• 083010MD DO probe
with 3 meter cable and
calibration sleeve
• Electrode storage chamber
• Polarographic electrolyte solution
• Hard field case
• Method Note - DO in Wine
and Beer
2009 Sourcebook
67
star meters
Thermo Scientific Orion 2-Star
pH Benchtop Meter
The Thermo Scientific Orion 2-Star pH benchtop meter
is easy to use and value priced for your general purpose
pH measurements
pH
ISE
ORP
Features and Benefits
Cond
• The Thermo Scientific Orion
SMART STABILITY™ and
SMART AVERAGING™
functions for automatically
optimized accuracy, precision
and response time
DO
Temp
• Built-in pullout guide for
quick access to calibration
instructions
CE I UL I CSA I TUV I EPA
• Data storage for up to 50 points
with time and date stamp
• RS232 port for printer output
• IP54-rated housing
• Up to 3 point pH calibration
• Includes universal power
supply
• Automatic buffer recognition of
USA or EURO buffers
• 3 year meter replacement
warranty
Meter Only
Product Specifications
Cat. No.
Package
pH
1111000
Includes benchtop meter, universal power and user guide
mV • Rel mV •
ORP
Temperature
68
• EH autocalibration for accurate
ORP calibration and readings
Range
0.000 to 14.999
Resolution
0.1, 0.01, 0.001
Relative
Accuracy
± 0.002
Auto-Buffer
Recognition
4.01, 7.00, 10.01/
1.68, 6.86, 9.18,
12.46
Calibration
Points
1, 2 or 3
Selectable
Calibration
Sets
USA or EURO
Custom
Buffers
Accepted
Yes
Range
Meter Kit
Cat. No.
Sensor (s)
Accessories and Solutions
1111001
• 9157BNMD refillable epoxy
pH/ATC Triode™ electrode
• Swing arm electrode stand
• pH 4.01, 7.00, and 10.01 buffers
and pH electrode storage solution
(pint bottles)
Meter, Electrode and Stand
Cat. No.
Sensor (s)
1111201
9165BNWP Sure-Flow® refillable epoxy pH electrode
± 1999.9
1111202
9157BNMD refillable epoxy pH/ATC Triode electrode
Resolution
0.1
1111203
9156APWP AquaPro low maintenance epoxy pH electrode, TRIS compatible
Relative
Accuracy
± 0.2 or 0.05% of
reading, whichever
is greater
1111204
9142BN refillable glass pH electrode
927007MD stainless steel ATC probe
Range
- 5 to 105 °C
1111206
9145BN low maintenance epoxy pH electrode
Resolution
0.1 up to 99.9 °C,
1.0 over 99.9 °C
1111207
9107BNMD low maintenance epoxy pH/ATC Triode electrode
Relative
Accuracy
± 0.1 °C
1111208
9107APMD AquaPro low maintenance epoxy pH/ATC Triode electrode,
TRIS compatible
Thermo Scientific Orion Laboratory Products Catalog
star meters
Thermo Scientific Orion Star and Star Plus Navigator 21™
Software Packages
Both versions of data-logging
and control software enable
compliance with auditors and
regulatory agencies and set the bar
for industry best-practices in the
pharmaceutical, food and beverage
and environmental industries.
• Provide the tools for FDA 21
CFR Part 11 compliance; on/off
mode flexibility.
Orion Star Navigator 21 Software
• Included with select 5-Star
Plus meters and meter kits,
or available as a stand alone
product to upgrade 3-Star and
4-Star meters.
• Works with portable or
benchtop meters.
• Compatible with the Windows®
XP operating system. The Star
Plus Navigator 21 version also
works with the Windows Vista®
operating system.
Complete control of your meter
from your PC
Orion Star Plus Navigator 21 Software
The Star Plus Navigator 21 software
includes enhancements which take
advantage of the Star Plus meter’s
additional features and allow for:
• Real-time graph of data results
• Automatic date and time
synchronization between the
computer, meter and software
• One button import of stored meter
data into the software
• Autosampler interface for faster
data reading and analysis
• Calibration capabilities for
temperature readings
• Expanded temperature-reference
values for conductivity readings
• Separate temperature viewing
of membrane and solution
temperatures
Now you can efficiently edit,
create and change methods; take
measurements; and perform routine
data browsing, retrieval, printing and
exporting all with a few key strokes.
Microsoft® Windows® operating
system defined password protected
security parameters enable
individual user accounts.
• System Access (Login/Logoff)
– Login is strictly regulated by
entering a valid user name and
password. Security measures
include account lockout after
a predetermined number of
unsuccessful login attempts
and automatic password
protected screen savers.
• Password Rules and
Management – Complex
password capabilities may
be enforced as a means of
extended security.
• Audit Trails – Records important
events and includes user,
date, time, event importance,
application, description of
event, and details of the event.
• Electronic Signatures – All
changes are recorded with a
digital signature and the users
must provide reasons for
changes.
• Data Security – All data is
stored in a database and it is
impossible to remove or alter
data in an uncontrolled manner.
The database can only be
accessed with the proper tools
and user rights. Data integrity
is verified each time data is
accessed.
Cat. No.
Description
1010007
Star Navigator 21 software
and RS232 computer interface
cable; use with 5-Star, 4-Star
and 3-Star meters
1010017
Star Plus Navigator 21 software
and RS232 computer interface
cable; use with 5-Star Plus,
4-Star Plus and 3-Star Plus
meters
Features
Benefits
User friendly Windows® interface
Easy to install and operate
Complete control from your computer
Allows quick operation of custom Orion Star
Series methods
Convenient report templates
Provide framework for compliance reporting,
capability to customize
Electronic records, signatures and audit trails
Provide the tools for FDA 21 CFR Part 11
compliance
Enables paperless record system
Saves time and money in the approval process
Which software is right for my meter?
When you start up your meter, the meter display will show Star or Star Plus corresponding
to the software version compatible with the meter.
Also, if you have any questions, you can contact our technical support experts at
(800) 225-1480 or (978) 232-6000.
2009 Sourcebook
69
star meters
Thermo Scientific Orion PerpHecT® LogR™ Meters
LogR temperature compensation from pH bulb – no ATC probe required!
CE I UL I CSA
pH
Product Specifications
pH
ISE
ORP
Cond
ISE
DO
Temp
Features and Benefits
mV
• All PerpHecT benchtop meters feature
exclusive digital LogR technology
• Allows the pH electrode to read
temperature – no separate ATC
probe and no need to use a 3-in-1
electrode
Temp
• 320 PerpHecT benchtop meter
performs pH or mV and temperature
measurements
• 350 PerpHecT benchtop meter has
more advanced features including a
clock, an RS232 data output, and
2.5 mm recorder output
• 370 PerpHecT benchtop adds an ISE
(concentration) mode to the features
of the 350
• Bright LED display shows
automatically calculated electrode
slope
• Meters accept a separate ATC probe
or manual temperature compensation,
if desired
• Meter only configuration includes
power adapter
• 3 year meter warranty
320 Meters
350 Meters
370 Meters
Range
0.00 to 14.00
0.00 to 14.000
0.00 to 14.000
Resolution
0.1, 0.01
0.1, 0.01, 0.001
0.1, 0.01, 0.001
Relative Accuracy
± 0.02
± 0.005
± 0.005
Auto-Buffer Recognition
4.01, 7.00, 10.01
4.01, 7.00, 10.01
4.01, 7.00, 10.01
Calibration Points
2
3
3
Range
-
-
± 0 to 19999
Resolution
-
-
±3 sig digits
Relative Accuracy
-
-
±0.2 mV or ±0.5%
of reading
Calibration Points
-
-
2
± 1999.9 mV
Range
± 1999 mV
± 1999.9 mV
LogR Range
±1000
±1000.0
±1000.0
Resolution
1.0
0.1, 1.0
0.1, 1.0
Relative Accuracy
±1 or ±0.1% of
reading
±0.2 mV or ±0.05%
of reading
±0.2 mV or ±0.05%
of reading
Range
0 to 100 °C
0 to 100 °C
0 to 100 °C
Resolution
0.1
0.1
0.1
Relative Accuracy
± 0.1 °C
± 0.1 °C
± 0.1 °C
LogR Relative Accuracy
± 0.5 °C
± 0.5 °C
± 0.5 °C
Benchtop Meter Kits
Cat. No.
Meter
Sensors
Accessories and Solutions
032000 (110V)
032001 (220V)
320
• 9202BN PerpHecT glass pH
-
0320A0 (110V)
0320A1 (220V)
320
032004 (110V)
032005 (220V)
320
032006 (110V)
032007 (220V)
320
035000 (110V)
035001 (220V)
350
0350A0 (110V)
0350A1 (220V)
350
035003 (110V)
035004 (220V)
350
electrode
• 8202BN PerpHecT ROSS® glass
-
pH electrode
• 9159APWP AquaPro epoxy pH
• Swing arm electrode stand
electrode
• 8220BNWP PerpHecT ROSS
• pH 4.01, 7.00, and 10.01 buffers
Micro pH electrode
• 9272BN PerpHecT Sure-Flow®
and pH electrode storage
solution (pint bottles)
-
glass pH electrode
• 8272BN PerpHecT ROSS Sure-
-
Flow glass pH electrode
• 8220BNWP PerpHecT ROSS
• pH 4.01, 7.00, and 10.01 buffers
Micro pH electrode
and pH electrode storage
solution (pint bottles)
• Printer and interface cable to
PerpHecT meter
0350G0 (110V)
350
Meters Only
• Printer and interface cable to
Flow glass pH electrode
Cat. No.
70
• 8272BN PerpHecT ROSS Sure-
Meter
110V
220V
240V
320
0320A9
0320A8
0320A7
350
0350A9
0350A8
0350A7
370
0370A9
0370A8
0370A7
Thermo Scientific Orion Laboratory Products Catalog
037000 (110V)
037001 (220V)
370
0370A0 (110V)
0370A1 (220V)
370
0370G0 (110V)
0370G1 (220V)
370
• 9272BN PerpHecT Sure-Flow
PerpHecT meter
-
glass pH electrode
• 8272BN PerpHecT ROSS Sure-
-
Flow glass pH electrode
• 8272BN PerpHecT ROSS Sure-
Flow glass pH electrode
• Printer and interface cable to
PerpHecT meter
star meters
Thermo Scientific Orion Intrinsically Safe* Meters
266S/261S
136S/131S
pH
Range
-2.00 to 16.00
-
Resolution
0.1, 0.01
-
Relative Accuracy
± 0.01
-
Auto-Buffer Recognition
Dependent on buffer set
(7 pre-programmed sets)
-
ORP
Cond
mV
Features and Benefits
Cond
• For use where non-sparking
equipment is desired, these carry
US intrinsically safe approval rating
by Factory Mutual (FM)*
• Waterproof, portable models for
basic field measurements
Salinity
• 266S and 136S meters include
RS232 output, advanced GLPdoc
100 point datalogger with time and
date stamp and includes software
program for simplified data
collection and remote operation on
a computer
TDS
• Accepts Waterproof DIN, dual
banana plug connections
• Units use 3 AA batteries
• 131S and 136S meters offer
linear and non-linear temperature
compensation for natural and
ultrapure water applications
• 3 year meter warranty
* Does not meet European Intrinsically
Safe specification.
Accessories
Cat. No.
Description
013655
Electrode clip
pH
Product Specifications
Temp
ISE
Calibration Points
3
-
Range
± 1300.0
-
Resolution
±0.1
-
Relative Accuracy
±0.1% of value +0.3 mV
-
Range
-
0.00 to 500 mS
Resolution
-
0.01 up to 99.99 μS/cm
0.1 above 99.99 μS/cm
Relative Accuracy
-
±0.5% of measured value
+ 1 count
Calibration Points
-
1
Cell Constants
-
0.010 to 199.0 cm-1
Range
-
0 to 45.0 (10 to 30 °C) g/kg
Resolution
-
0.1
Range
-
0 to 1999 mg/L (10 to
40 °C)
Resolution
-
1
Range
-20 to 120 °C
-20 to 120 °C
Resolution
0.1
0.1
Relative Accuracy
± 0.3 °C
± 0.3 °C
DO
Temp
Portable Meter and Electrode
Cat. No.
Meter
Sensor (s)
0261S0
261S
9109WP low maintenance pH/ATC Triode™
0266S0
266S
9109WP low maintenance pH/ATC Triode
0131S0
131S
013010A DuraProbe™ conductivity cell
0136S0
136S
013010A DuraProbe conductivity cell
Portable Meter, Electrode and Field Kit
Cat. No.
Meter
Sensor (s)
Accessories and Solutions
0261S3
261S
9109WP low maintenance
pH/ATC Triode
• Neck strap • Field case • Stand
• Electrode holder • Beakers
• 4, 7 and 10 buffers
0266S3
266S
9109WP low maintenance
pH/ATC Triode
• Neck strap • Field case • Stand
• Electrode holder • Beakers
• 4, 7 and 10 buffers
• RS232 cable and software
0131S3
131S
013010A DuraProbe
conductivity cell
• Beakers
• Standards
0136S3
136S
013010A DuraProbe
conductivity cell
• Beakers • Standards
• RS232 cable and software
2009 Sourcebook
71
star meters
Thermo Scientific Orion Meter Accessories
To make sure you have everything you need to attain accurate results for your applications,
we offer a multitude of accessories for both field and lab applications. From printers to field
cases, electrode stands and stirrers, we have it all. Below are the top selling accessories
– for complete details, please visit www.thermo.com/water.
Automatic Stirrer Probe with Paddle, Cat. No. 096019
and Swing Arm Stand, Cat. No. 090043
Hard Field Case, Cat. No. 1210004
Aplus™ and PerpHecT® Meter Accessories
Orion Star™ Series Meter Accessories
Cat. No.
Description
Cat. No.
Description
020125
110V Power Adapter
1010003
Universal Power Adapter
020130
220V Power Adapter
1010004
RS232 Computer and Printer Cable Set
020135
240V Power Adapter
1010005
RS232 to USB Cable Adapter
090043
Swing Arm Electrode Stand
1010006
Star Series Inkjet Printer, 110 V / 220V and Cable
090070
Basic Electrode Stand and Holder
1010008
Ink Cartridge for Star Series Printer
910002
Universal Electrode Holder for Basic Electrode Stand
1010009
Paper for Star Series Printer
090045
BNC Shorting Cap
1010053
RS232 Computer Cable
0900A0
900A Thermal Printer for Aplus Meters, 110V
1210001
Portable Star Series Electrode Stand
0900A1
900A Thermal Printer for Aplus Meters, 220V
1210002
Meter Impact Jacket and Electrode Storage Chamber
0900A7
900A Thermal Printer for Aplus Meters, 240V
1210003
Electrode Storage Chamber (For Meter Impact Jacket)
0900A4
Paper for 900A Printer
1210004
Hard Field Case
0900A6
Battery Pack for 900A Printer
1210005
Soft Field Case
PRT300
PRT300 Ink Printer and Cable for Aplus and PerpHecT Meters,
110V
096019
Automatic Stirrer Probe with Paddle
(For Benchtop Star Series Meters)
PRT301
PRT300 Ink Printer and Cable for Aplus and PerpHecT Meters,
220V
096021
Replacement Paddles for 096019 – 3 Pack
090043
Swing Arm Electrode Stand
PRT302
Ink Ribbon for PRT300 Printer
090045
BNC Shorting Cap
0ACBL0
RS232 Computer Cable for Aplus and PerpHecT Meters
1010050
0PBLSK
Aplus Portable Meter Starter Kit – Includes Field Case, Field
Stand, Beaker, 2 x 60mL Bottles Each of pH 4.01, 7.00, and
10.01 Buffers, and 1 x 60mL Bottle of pH Storage Solution
Adapter– MiniDIN ATC Probe to 3.5 mm Phono Tip Meters (310,
320, 330, 350, 370, 555A, 550A, 535A, and 162A)
1010051
Adapter– MiniDIN ATC Probe to 8 Pin DIN Meters (210Aplus,
230Aplus, 250Aplus, 290Aplus, 410Aplus, 420Aplus, 520Aplus,
525Aplus, 710Aplus, 720Aplus, and 920Aplus)
1010800
Adapter– MiniDIN DO Probe to 8 Pin Waterproof DIN DO Meters
(830A, 835A and 862A)
1010801
Adapter– MiniDIN DO Probe to 8 Pin DIN DO Meters (805Aplus,
810Aplus, and 850Aplus)
1010900
Adapter– MiniDIN Conductivity Probe to 8 Pin Waterproof DIN
Conductivity Meters (1230, 555A, 550A, 162A, 135A, and 130A)
1010901
Adapter– MiniDIN Conductivity Probe to 8 Pin DIN Conductivity
Meters (150Aplus, 145Aplus, 125Aplus, 115Aplus, and 105Aplus)
0BCHSK
72
Soft Field Case, Cat. No. 1210005
Aplus Benchtop Meter Starter Kit – Includes Basic Electrode
Stand and Holder, 917005 Epoxy ATC Probe, and 1 x 475 mL
Bottle Each of pH 4.01, 7.00, and 10.01 Buffers and pH Storage
Solution
0607CA
607 Multi-Channel Electrode Switchbox and Cables, 110 V
0607CI
607 Multi-Channel Electrode Switchbox and Cables, 220 V
Thermo Scientific Orion Laboratory Products Catalog
star meters
Thermo Scientific Orion AutoTration™ - 500 Autosampler
High Sample Throughput with Star Plus Benchtop Meters
Autosampler
CE
Cat. No.
Package
AT5050
48-position beaker rack for 50 mL beakers, small electrode holder, drip tray, cable
management system, BNC extension cable, stirrer extension cable, accessories
package, interface cable, line cord, 50 pack of 50 mL plastic beakers, user guide, and
warranty card, 115/220V.
AT5150
28-position beaker rack for 150 mL beakers, small electrode holder, drip tray, cable
management system, BNC extension cable, stirrer extension cable, accessories
package, interface cable, line cord, 50 pack of 150 mL plastic beakers, user guide, and
warranty card, 115/220V.
Autosampler and Star Plus Meter Packages
Features and Benefits
• Performs analyses, calibrations
and standardization procedures
automatically
Cat. No.
Autosampler
Benchtop Meter
Autosampler AT5050
1112000SPA
AT5050 - 48 position
1112000 - 3 Star Plus pH
1114000SPA
AT5050 - 48 position
1114000 - 3 Star Plus conductivity
• Improves data quality by using up
to five rinse stations
1115000SPA
AT5050 - 48 position
1115000 - 4 Star Plus pH/ISE
1117000SPA
AT5050 - 48 position
1117000 - 4 Star Plus pH/conductivity
• Durable and corrosion resistant,
the sealed electronics chassis
and coated metal parts protect
against the harshest chemical
environments
1119000SPA
AT5050 - 48 position
1119000 - 5 Star Plus pH/ISE/Conductivity/
Dissolved Oxygen
• Compatible with Star Plus
meters and 960 Titrator PLUS for
automated techniques
• RS232 interface for easy
communication
• Available as autosampler only
or with meter packages making
accurate data from multiple
samples easier to obtain
• Available with either a 50 mL
or 150 mL beaker rack and
accessories
Autosampler AT5150
1112000SPB
AT5150 - 28 position
1112000 - 3 Star Plus pH
1114000SPB
AT5150 - 28 position
1114000 - 3 Star Plus conductivity
1115000SPB
AT5150 - 28 position
1115000 - 4 Star Plus pH/ISE
1117000SPB
AT5150 - 28 position
1117000 - 4 Star Plus pH/conductivity
1119000SPB
AT5150 - 28 position
1119000 - 5 Star Plus pH/ISE/Conductivity/
Dissolved Oxygen
All packages include beaker rack, electrode holder, drip tray, cable
management system, BNC and 8 pin MiniDIN extension cables, stirrer probe,
interface cable, power cords, plastic beakers and user guides
Product
Specifications
Accuracy
±1.0 mm in all axes
Travel and Resolution Precision
Accessories
Cat. No.
Description
ATA01
Replacement tubing for guide arm
ATA02
RS232 cable for interface between
Star Plus meter and autosampler
ATR050
48-position beaker rack for
50 mL beakers
Maximum X axis
510 mm
Minimum X axis
resolution/
precision
Better than 0.1 mm
Maximum Y axis
270 mm
ATR150
28-position beaker rack for
150 mL beakers
Minimum Y axis
resolution/
precision
Better than 0.1 mm
ATB050
50 mL plastic beakers for
48-position rack, 50 pack
Maximum Z axis
160 mm
ATB150
150 mL plastic beakers for
28-position rack, 50 pack
Minimum Z axis
resolution/
precision
Better than 0.1 mm
096S22
45 mL plastic beakers for
48-position rack, 50 pack
AT0STD
Maximum Z axis
clearance
140 mm
Electrode holder, standard size for
AT5150
AT0050
Electrode holder, small size for AT5050
Environmental
Requirements
5 to 45 °C and
20 to 80% relative
humidity
096019
Replacement automatic stirrer probe
2009 Sourcebook
73
colorimetry
Thermo Scientific Orion Colorimetry Theory Overview
Sample Vial
Incident Light, P0
Transmitted Light, P
Colorimetry is an optical measurement that uses
color to quantitatively determine the concentration
of a component in a liquid sample. A chemical
reagent for the component of interest is added to
the sample being tested. A color will develop that
can be measured by a colorimeter.
Transmission = P/P0 = 10(-abc)
• It is good practice to run a “zero” of deionized
water to correct for stray light and to correct
for internal reflection of light by vial walls.
%T
P is power of transmitted light
a is a constant - the ability of a given
molecule to absorb a particular wavelength
of light
b is the path length - the longer the path,
the less light gets through
Concentration
c is the concentration - the more molecules in
the solution, the more light is absorbed
Absorbance
As expected from Beer’s Law, an exponential
curve results from plotting the percent
transmission versus concentration points, and a
linear curve results from plotting the absorbance
versus concentration points.
Abs
400
450
500
550
nm
DPD Free Chlorine Spectrum
74
600
• Colorimetry is a value oriented test for a
specific concentration range that can be
measured in either the lab or field.
• Both reagent blank and deionized water zero
measurements are important for accuracy
and precision. Be sure to review specific
steps of the method to see if both are used.
P0 is initial power of incident light
1.0
0.9
0.8
0.7
0.6
0.5
0.4
0.3
0.2
0.1
0
• Colorimetry is often a good technique for lowlevel readings in the ppb to ppm range.
As depicted in the figure on the left, colorimetry
is a chemical measurement based on Beer’s Law.
Beer’s Law describes the relationship between the
amount of light transmitted through a sample and
the component’s concentration in the sample.
Where:
Concentration
Colorimetry Tips and Techniques
A typical colorimeter has one or more fixed
wavelength light emitting diode (LED) sources
in the meter. To have the most sensitivity and
selectivity to the compound of interest, the
wavelength of the LED is chosen to match one
of the maxima or peaks, in the visible spectrum
of the compound of interest when reacted with
the reagent. In the free chlorine DPD spectrum,
there is a peak near 520 nm, the wavelength range
used for this test. The colorimeter is designed
to pass light through a sample and collect the
transmitted light through the sample and the vial.
By establishing a calibration curve from readings
of standards with known concentrations for the
component of interest, the concentration of that
component in the sample can be determined. For
many official colorimetry methods, a wavelength
range is specified for the test.
Thermo Scientific Orion Laboratory Products Catalog
• Some methods require or recommend a
reagent blank, which uses deionized water
instead of the sample. The user adds the
test reagent and follows the test method as
if it was a sample. When recommended, a
reagent blank should be performed.
• Reagent blanks can be used to subtract
the amount of color contributed by the
reagents and not due to the sample itself.
This gives greater accuracy in the sample
result. In advanced meters, the subtraction is
performed for you.
• Concentrated samples that are above the
range of the test method may be diluted with
deionized water to meet the concentration
range. These results needs to be corrected
for the dilution amount to obtain the actual
sample value.
• Turbidity and color in the sample itself can
often be “zeroed (subtracted) out” if the level
is not too high.
• Precision depends on concentration and
range and is typically between 5-10% in the
middle of the method range.
• Be sure to insert the vials so the mark on
the vial and the mark on the instrument are
aligned for reproducibility.
• Routinely inspect vials before use and discard
any with cracks or scratches, as they can
affect the results.
colorimetry
Thermo Scientific Orion Colorimetry and Turbidity Markets
You can find colorimetry and turbidity
measurements in a wide variety of industries.
The tests are quick and easy and often cost less
than alternative methods. Thermo Scientific Orion
AQUAfast colorimetry and turbidity products have
approvals and meet the regulatory requirements
in many industries. We also have several types of
tests for common parameters to give you choices
of reagent types, ranges and simplicity to fit your
requirements. The reagent table on pages 84 and
85 highlights our EPA approved, accepted and
equivalent methods.
Commonly Tested Parameters by Market
Agriculture
Beverage
Drinking Water
Environmental
Wastewater
Aquaculture
Boiler/Cooling Water
Chemical
Ammonia
Bromine
Aluminum
Ammonia
Aluminum
Ammonia
Aluminum
Aluminum
Chlorine
Chlorine
Ammonia
Bromine
Ammonia
Bromine
Ammonia
Ammonia
Iron
Chlorine Dioxide
Bromine
Chlorine
Bromine
Chlorine
Bromine
Copper
Nitrate
Copper
Chloride
COD
Chlorine
Copper
Chlorine
Iron
Nitrite
Fluoride
Chlorine
Copper
COD
Nitrate
Chlorine Dioxide
pH
Oxygen
Iron
Chlorine Dioxide
Iron
Copper
Nitrite
Copper
Phosphate
pH
Manganese
Copper
Nitrate
Iron
Oxygen
Hydrazine
Silica
Phosphate
Nitrate
Fluoride
Nitrite
Manganese
Ozone
Iron
Sulfate
Silica
Oxygen
Iron
Ozone
Nitrate
pH
Nitrate
Sulfide
Sulfate
pH
Manganese
pH
Nitrite
Phosphate
Nitrite
Zinc
Turbidity
Phosphate
Nitrate
Silica
Oxygen
Turbidity
Ozone
Silica
Nitrite
Sulfate
Ozone
Turbidity
pH
Sulfide
pH
Phosphate
Phosphate
Zinc
Phosphate
pH
Ozone
Sulfate
Silica
Silica
Sulfide
Zinc
Sulfate
Zinc
Sulfide
Turbidity
Oxygen
Zinc
Turbidity
Metal Plating
Mining
Pulp & Paper
Solid Waste/
Sludge
Petroleum
Pharmaceutical/
Biotech
Semiconductor
Copper
Copper
Aluminum
Aluminum
Ammonia
Chlorine
Aluminum
Iron
Iron
Chlorine
Copper
Chlorine
Cyanide
Copper
pH
pH
Chlorine Dioxide
pH
Iron
Ozone
Fluoride
Phosphate
Phosphate
pH
Phosphate
Nitrate
pH
pH
Zinc
Phosphate
Zinc
pH
Silica
Silica
Silica
Sulfide
Zinc
2009 Sourcebook
75
colorimetry - turbidity
Thermo Scientific Orion AQUAfast® IV Advanced Turbidity Meter
The AQ4500 turbidimeter has dual source LED that
allows readings that comply with both EPA 180.1
and ISO 7027 methodologies.
• Nephelometric and ratiometric measurements
with autoranging
• Data log capacity of up to 100 data points
AQ4500 Kit
The AQ4500 offers ratiometric
range, which will give EPA GLI
method 2 equivalent numbers.
• Readings in the range of 0 to 1000 NTU with
choice of units: NTU, FTU, FNU, ASBC or EBC
• Includes turbidity standards kit, rugged
carrying case, and replacement cuvettes
• The AQ4500 is truly IP67 waterproof to a
depth of 3 meters
Specifications
Description
Light Source
White LED, IR LED
Temperature Effect
LED temperature compensated
Includes Measurement
Modes
EPA 180.1, ISO Nephelometric, ISO Absorbance, IR Ratiometric,
% Transmittance, EPA GLI Method 2, EBC (European Brewing
Chemists), ASBC (American Society of Brewing Chemists)
Measurement Ranges
0 to 4000 NTU, EPA 0 to 2000 NTU, GLI range 0 to 40 NTU,
ISO-7027 0 to 150 FAU, IR Ratio 0 to 4000 NTU, EBC 0 to 24.5,
ASBC 0 to 236
Resolution
0.01 (0 to 9.99 NTU), 0.1 (10 to 99.9), 1 (100 to 1000)
Precision
+/- 2% of reading plus 0.01 NTU from 0 to 500 NTU,
+/- 3% of reading from 500 to 100 NTU,
+/- 5% of reading from 1000 to 2000 NTU
Calibration Standards
Provided primary SDVB standards
Sample Size Minimum
10 mL
User Interface
Automatic sample measurement for each measurement with
simple user interface
RS232
Bidirectional
Datalog
100 points
Real Time Clock
Yes
Power
Long life - low power consumption - about 2500 hours with
4 alkaline AA batteries or about 10000 with lithium batteries
Waterproof Rating
IP67
Warranty
2 years
Regulatory and Safety
CE, FCC part 15
Operating Temperatures
(-40 ˚C to 60 ˚C), 90% relative humidity at 30.0 ˚C max
Ordering Information
76
Cat. No.
Description
AQ4500
AQUAfast IV Turbidimeter and calibration kit
Kit includes meter, 4 primary standards, standard and sample vials, silicone
oil and cloth in carrying case
AQ4CBL
RS232 cable for data download and software updates
AC45FZ
Formazin turbidity standard, 4000 NTU, 475 mL
AC45SI
Silicon oil and cloth kit for AQ4500
AC45ST
Primary turbidity calibration kit
AC2T24
24 mm replacement turbidity vials, 12 pack
Thermo Scientific Orion Laboratory Products Catalog
colorimetry - turbidity
Thermo Scientific Orion AQUAfast® AQ3010 Turbidity Meter
• Ergonomic IP67 turbidity meter for lab
or field use
• Uses Nephelometric principle with
90º detection
• In many locations a separate light cap
or shield is not needed
• Meets ISO 7027 and offers consistent
light source for greater accuracy
• Easy vial indexing with continuous
measurement mode
• No mixing or diluting with polymer bead
primary standards for faster calibration
• Three year warranty
NEW
The new AQUAfast AQ3010 Turbidity Meter makes
it easy to determine the clarity of your sample with
one keystroke and without an extra light shield
cap in many environments. With autoranging from
0 – 1000 NTUs, this small waterproof portable
meter is very easy to use, plus features a large
easy-to-read LCD display. The meter is supplied
with ready-to-use, highly stable EPA-approved
polymer bead primary standards. Calibration is
easy with step-by-step prompts with the option
to choose which points to calibrate. The standards
can be used for a calibration check at any time,
verifying that your readings are accurate.
The AQUAfast AQ3010 meter comes with
everything you need in a field case for easy storage
and transport. This meter is covered by a 3 year
warranty!
AQ3010 Meter Kit
Ordering Information
Specifications
Description
Cat. No.
Description
Contents
Display
LCD
AQ3010
AQUAfast turbidity meter kit
Keys
Power, Cal, Scroll Up,
Scroll Down, Read/Enter
Automatic Range
Selections and Resolution
0.01 to 19.99 NTU with
0.01 NTU resolution
Meter, set of EPA-approved
polymer-based primary
turbidity standards, batteries,
light shield cover, seven vials,
lint-free cloth, silicone oil,
MSDS and user guide
AC301S
Replacement set of EPAapproved polymer-based
primary turbidity standards
60 mL bottle of 0.02, 20.0,
100, and 800 NTU
100-1000 NTU with
1 NTU resolution
AC3SIL
Replacement silicone oil
and cloth
1 bottle of silicone oil and
one cloth
+/- 2% of reading
+/- 1 LSD for 0-500 NTU
AC3V25
Replacement 25 mm vials,
pack of 3
3 empty 25 mm vials
with caps
+/- 3% of reading
+/- 1 LSD for 501-1000
NTU
AC3CBR
Replacement vial cleaning
brush
Brush, package of 10
20.0 to 99.9 NTU with
0.1 NTU resolution
Precision
Response Time
< 6 seconds
Sample Requirement
10 mL
Make you job easier!
Included Calibration
Standards
0.02, 20.0, 100, and
800 NTU
Sample Vials
25 mm diameter with
indexing mark
Light Source
IR LED for ISO 7027
Turbidity measurement
Visit our new applications webpage at
http://www.thermo.com/waterapps for
regulatory links, methods, applications and
technical information.
Power
4 x AAA > 3000 tests
(4 x AAA Lithium > 12000
tests)
Battery conservation
feature
Auto shutoff after 20
minutes of inactivity
Ambient Operating
Temperature
0 to 50 C (0 to 90%
Non-Condensing Relative
Humidity at 30 C)
Self Diagnostics
Yes, with troubleshooting
error codes
Waterproof rating
IP67
Regulatory Approvals
CE, TUV 3-in-1 for
CSA-US, FCC part 15
Warranty
3 year
2009 Sourcebook
77
colorimetry
Thermo Scientific Orion AQUAfast® AQ3070 Chlorine Meter
NEW
Measures free and total chlorine for EPA reporting – and more!
• EPA-approved AC4P71 free chlorine and
AC4P72 total chlorine methods for drinking
water and wastewater
• Single meter programmed for free chlorine,
total chlorine, pH, chlorine dioxide and
cyanuric acid measurements
• Additional capability for liquid pH test, chlorine
dioxide and cyanuric acid – no need to
purchase separate meters for additional tests
• Chlorine dioxide and cyanuric acid tablets
feature long shelf lives, so they are ready
when you need them
Designed with the user in mind, these small
rugged IP67 portable meters have a large easyto-read display with a very simple user interface.
The expanded range for free and total chlorine
measurements gives you a real number if you
are out of the EPA measurement range to quickly
recover your process. The results are accurate
and repeatable. You save time on measuring by
not having to cover and uncover the vial for each
measurement. The field case keeps the meter and
test supplies always ready for testing.
AQ3070 Meter
Specifications
Description
Display
LCD
Keys
Power, mode, zero, read/enter
Sample Requirement
10 mL
Light Shield Cover over Vial for Measuring
Not needed
Power with Auto Shutoff
4 x AAA, > 3000 tests
Regulatory Approvals
CE, TUV for US and Canada,
FCC part 15
Warranty
2 years
Operating Range
0 to 50 ˚C, 0 to 90% relative humidity,
non-condensing at 30 ˚C
Test Parameter Ranges
AQ3070 Meter Kit
Free Chlorine
0.02 to 4.0 mg/L;
US EPA range 0.03 to 2.0 mg/L
Total Chlorine
0.02 to 4.0 mg/L;
US EPA range 0.03 to 2.0 mg/L
Chlorine Dioxide
0 to 11.4 mg/L
pH
5.9 to 8.2 pH
Cyanuric Acid
5 to 90 mg/L
Ordering Information
Visit www.thermo.com/waterapps
for our EPA letters and methods.
78
Cat. No.
Description
Contents
AQ3070
AQUAfast III chlorine meter kit
Meter, 6 sample vials, AC4P71 free
chlorine, AC4P72 total chlorine,
batteries, vial cleaning brush,
tablet tamping stir rod, user guide,
MSDSs, and carrying case.
AC4P71
AQUAfast AC4P71 free chlorine
powder packets
100 tests
AC4P72
AQUAfast AC4P72 total chlorine
powder packets
100 tests
AC2099
Chlorine dioxide tablets
100 tests
AC3001
pH, 2 x 15 mL dropper bottles
60 tests
AC2098
Cyanuric acid, tablets
100 tests
AC3V25
Replacement 25 mm vials,
pack of 3
3 vials
AC3ROD
Replacement tablet tamping
stir rod
1 package of 10
AC3CBR
Replacement vial cleaning brush
1 package of 10
Thermo Scientific Orion Laboratory Products Catalog
colorimetry
Thermo Scientific Orion AQUAfast® AQ3700 Colorimeter
NEW
Featuring digestion tests for nitrogen, phosphate and COD
Reagents for AQ3700
The AQUAfast 3700 is the newest addition
to the colorimetry line. With over 70 preprogrammed method options, this IP67
waterproof portable meter offers liquid,
powder and tablet tests for expanded ranges
and test options. There are six wavelengths
generated by the long life LEDs in the meter.
This instrument has a graphical display that is
easy to understand and can be used in English,
Spanish, German, French, or Italian.
Cat. No.
New Test
Range, mg/L
Reagent Type
AC3002AD
Acid demand pH 4.3
0.1 to 4
Tablet
AC3002M
Alkalinity-m
5 to 200
Tablet
AC3002P
Alkalinity-p
5 to 500
Tablet
AC4P27
Aluminum
0.01 to 0.25
Tablet & liquid
ACD012
Ammonia, LR
0.02 to 2.5
Thermal digestion
tube
ACD011
Ammonia, HR
1 to 50
Thermal digestion
tube
AC3089
Boron
0.1 to 2
Tablet
AC2070LQ
Chlorine
0.02 to 4
Liquid
AC3072
Chlorine total HR
5 to 200
Tablet
AC4P06
Cyanide
0.01 to 0.5
Liquid & powder
Specifications
Description
AC2098
Cyanuric acid
2 to 160
Tablet
Incident Light Source
6 Light Emitting Diodes (LED)
AC3002AD
DEHA
0.02 to 0.5
Liquid & powder
Wavelengths
430, 530, 560, 580, 610, 660 nm
AC4P88
DEHA
20 to 500
Liquid & powder
Photometric Precision
2 % full scale as measured with standard solutions
AC3032C
Hardness, calcium
50 to 900
Tablet
Photometric Resolution
+/- 0.005 Abs
AC3032T
Hardness, total
2 to 50
Tablet
Software
AC3032HR
Hardness, total HR
20 to 500
Tablet
Test Methods
AC4P30
Hydrazine
0.05 to 0.5
Powder & spoon
AC3030LQ
Hydrazine
0.01 to 0.6
Liquid
AC3069
Hydrogen peroxide
0.03 to 3
Tablet
AC3053
Iodine
0.05 to 3.6
Tablet
AC3078
Iron (ll, lll) soluble
0.02 to 1
Tablet
AC4P54
Manganese, LR
0.01 to 0.7
Powder & liquid
AC3042
Molybdate
1.0 to 50.0
Tablet
ACR007
Nitrate
1.0 to 30.0
Powder pack &
reaction tube
AC4P46
Nitrite
0.01 to 0.3
Powder pack
ACD004
Nitrogen, total, LR
0.5 to 25
Powder pack &
digestion tube
See pages 80-81 and 83-85 for additional
AQ3700 reagents.
Preprogrammed with user prompts and feedback, user’s
favorite list
Autoranging
Yes
Blank Capability
Per method
Test Reagent Formats
Powder, tablet, liquid, reagent tubes, acid digestion tubes
Datalog Points
Up to 1000, time and date stamp
New Method Upload
Capability
Yes, via internal RJ45 port
Meter Hardware
Display
Backlit, graphic display
Inputs
12 button numeric keypad, plus 3 function keys
Sample Required
10 mL
Sample Cells Accepted
24 mm sample vials and 13 mm ampoules and 16 mm tubes
with adapters
ACD007
Nitrogen, total, HR
5 to 150
Powder pack &
digestion tube
Outputs
IRDA (adapter required)
AC4008LR
Oxygen, dissolved
0.01 to 0.8
Ampoule
Auto Shut Down
20 min after last key-press, wakes to last location
AC3093
PHMB (biguanide)
2.0 to 60.0
Tablet
Power
4 AA batteries (>3000 tests)
ACR095
Phosphate, ortho
0.06 to 5
Powder pack &
reaction tube
Regulatory and Safety
CE
ACD095AH
Phosphate, acid
hydrolyzable
0.02 to 1.6
Powder pack &
digestion tube
Environmental Conditions
5 to 40 °C (30 to 90% non-condensing at 30 °C)
Meter Waterproof Rating
IP67
ACD095
Phosphate, total
0.02 to 1.1
Powder pack &
digestion tube
Warranty
2 years
AC3001BP
pH
5.2 to 6.8
Tablet
AC3001
pH
6.5 to 8.4
Liquid
AC3001TB
pH
8.0 to 9.6
Tablet
Cat. No.
Description
AC3019
Potassium
0 to 12
Tablet
AQ3700
AC4P60LR
Silica LR
0.1 to 1.6
Powder & liquid
AC3073
Sodium hypochlorite
0.2 to 16
Tablet
Meter kit includes AQ3700 meter, 4 batteries, 4 vials (24 mm), 3 vials
(16 mm), 1 adapter each for 16 mm and 13 mm vials, carrying case, user
guide with test methods, tablet tampering stir rod and brush.
Order Starter Kits and/or Reagents and IR Adapter Separately.
AC4P82
Sulfate
5 to 100
Powder pack
AC3013
Urea
0.1 to 3
Tablet & liquid
Ordering Information
See page 80 for AQ3700 accessories.
2009 Sourcebook
79
colorimetry
Thermo Scientific Orion Digestion Test Reagents
Tube digestions for nitrogen and phosphate
nutrient removal process monitoring
AQUAfast AQ3700 Meter
The benefit of efficient operation is important to
all production facilities, no matter what they
make. By optimizing the process in wastewater
treatment, the dirty water can move to the next
stage of the process at the perfect time – not
prior to the being fully treated and not overtreated.
Overtreatment wastes chemicals and slows
down the processing. Undertreatment can
cause problems if the discharge does not meet
the permit levels. Typically total nitrogen and
total phosphate and other digestions were very
long reflux digestion processes with expensive
glassware. Now the results can be ready in an hour
or so using persulfate digestion tube methods.
This greatly improves the turnaround time for the
results. These in process checks of the nutrient
removal status can ensure you are on track for the
final levels you need to attain or give you timely
information to correct the process.
Total nitrogen LR test
Nitrogen
Cat. No.
Description
Quantity
ACD004
Total nitrogen LR reagents
(0.5 to 25 mg/L)
50 tests
ACD007
Total nitrogen HR reagents
(5 to 150 mg/L)
50 tests
AC2V16
16 mm vials
Total phosphate test
Phosphate
Cat. No.
Description
Quantity
ACD095
Total phosphorous reagents
(0.02 to 1.1 mg/L)
50 tests
AC2V16
16 mm vials
Accessories for AQ3700 Meter
80
Cat. No.
Description
AC2V16
Replacement 16 mm vials, package of 10
AC2V24
Replacement 24 mm vials, package of 12
AC37SR24
Sealing ring for 24 mm vial, package of 12
AC37A16
Replacement 16 mm adapter for AQ3700
AC37A13
Replacement 13 mm adapter for AQ3700
AC3ROD
Stir rod, package of 10
AC3CBR
Cleaning brush, package of 10
AC3VSK
Meter verification check standards kit
AQ37IRIM
IR data transfer interface module
Thermo Scientific Orion Laboratory Products Catalog
NEW
With the AQ3700 meter, total nitrogen and total
phosphate tube digestion tests are now available.
The kit includes the reagents and accessories for
the test.
If you are already testing COD levels, you can
quickly add these important tests and start reaping
the benefits!
Overview of nitrogen and phosphorus methods
• Persulfate digestion for nitrogen; persulfate/
ascorbic acid digestion for phosphorus
• Nitrogen tests require two tubes – one for
a blank and one for the test; phosphate test
needs just one tube for the blank and test
• 100 °C digestion temperature for 30 minutes
and let cool about 10 more minutes for
additional reagent reaction times
• AQUAfast COD digestors have the
temperatures needed for nitrogen and
phosphate tests
colorimetry
Thermo Scientific Orion COD Test Systems
COD testing correlated to BOD readings can save time and money
• Ideal for environmental and wastewater
monitoring with EPA accepted methods
• Datalogging and GLP features
• Bidirectional RS232 interface on AQ4000
• Three ranges of COD reagents
• IR Interface on AQ3700 (IRDA required)
• New bulk COD reagent packs for added
convenience
• COD reactor can digest up to 25 samples
simultaneously in 2 hours or less and meet
fast COD temperature requirements
AQ4000 Meter
• Works with the new AQ3700 multiparameter
colorimeter, the AQ4000 advanced
colorimeter or the economical AQ2040
colorimeter kit
Just add the sample to the COD vial, digest
sample in the reactor and read results with your
AQUAfast colorimeter.
The AQ4000 and AQ3700 colorimeters offer:
• Pre-programmed methods for water
quality testing
• Compatibility with 13, 16 and 24 mm
diameter vials
• IP67 waterproof housing
• 2 year warranty
The AQ2040 colorimeter offers:
• Low-cost, dedicated colorimeter design
for COD testing
• Measures all three COD ranges
• 1 year warranty
Please visit www.thermo.com/water for more
product, methods information and method
approval letters from EPA.
• Multiple wavelength and more test options
COD Reactor, Test Reagents and
Standards with AQ4000 Meter
Select One Meter
The AQ4000 and AQ3700
colorimeters support
additional tests, including
chlorine, fluoride and
ammonia (see pages 79-85).
Cat. No.
Description
AQ4000
AQUAfast AQ4000 advanced colorimeter with COD
AQ2040
AQUAfast II dedicated COD meter
AQ3700
AQUAfast AQ3700 colorimeter with COD, total nitrogen and
total phosphate
Order Digestor
Cat. No.
Description
COD125
AQUAfast thermoreactor for digestion methods to 150 ˚C
COD165
AQUAfast thermoreactor for digestion methods to 165 ˚C
Select Your Test Reagents by
Range and Quantity
Cat. No.
Range
No. of Vials
EPA Approved
CODL00
0 to 150 ppm
25
Yes
CODL00B1
0 to 150 ppm
100
Yes
CODL00B3
0 to 150 ppm
300
Yes
CODH00
0 to 1500 ppm
25
Yes
CODH00B1
0 to 1500 ppm
100
Yes
CODH00B3
0 to 1500 ppm
300
Yes
CODHP0
0 to 15000 ppm
25
No requirement
Select a Standard
Cat. No.
Description
CODS01
AQUAfast 1000 ppm COD standard, 475 mL
CODS10
AQUAfast 10000 ppm COD standard, 475 mL
2009 Sourcebook
81
colorimetry
Thermo Scientific Orion AQUAfast® AQ4000 Colorimeter
The most advanced microprocessor-based LED colorimeter on the market today!
The AQ4000 colorimeter stores up to 189 preprogrammed methods and can record 100 data
points in the field with a time and date stamp.
Download later to a printer or computer in the lab.
The AQ4000 colorimeter takes all the guesswork
out of sample preparation with the Auto-Test™
ampoules and the exclusive Auto-ID™ feature.
Versatile AQ4000 Meter
Quality of the measurement is assured since the
ampoules are only opened when a measurement
is about to be taken.
• Auto-ID: Automatically identifies the species
to be measured, selects the method and the
wavelength plus initiates the method – no
buttons to push!
• Auto-Test Ampoules: Pre-measured liquid
reagent ampoules are specially designed to
work with the AQ4000
Selection of display units is push-button easy
Use the AQ4000 with our tablets or powders
• Tablets: Quick dissolve tablets that are easy
to add to the sample. Tablets can be used
with the AQUAfast II and IV colorimeters.
• Powders: Pre-measured and ready-to-use
packets. Just add the packet to the sample,
allow it to dissolve and then take your
measurement. Powders can be used with any
colorimeter on the market.
• COD Reagents: Available in three
measurement ranges – low, mid and
high range.
Updating the AQ4000 with new methods is simple
The latest software updates are available at
www.thermo.com/water. A simple download
procedure updates the colorimeter with the most
current programs in a few quick, easy steps.
See pages 84 and 85 for reagent chemistries that you
can use with the AQ4000 colorimeter.
Choose from concentration, absorbance or percent
transmittance. The AQ4000 accepts round vials
with diameters of 13, 16 or 24 mm. Best of all, the
AQ4000 carries a full two year warranty.
AQ4000 Colorimeter and Accessories
Cat. No.
Description
AQ4000
AQUAfast IV colorimeter only (order reagents separately)
AQ40FK
Field case to carry AQ4000 meter, accessories and reagents
AQ4EK1
Choice clean water kit featuring the AQ4000
AQ4CBL
RS232 cable for data download and software updates
AQ4ZER
AQUAfast zero kit
AC2V16
16 mm vials, 10 pack
AC2V24
24 mm vials, 12 pack
AQII Series Accessories
Cat. No.
Description
AC2V16
16 mm replacement vials, 10 pack
AC2V24
24 mm replacement vials, 12 pack
AC2T24
24 mm replacement turbidity vials, 12 pack
AC2073
Chlorine secondary reference standards kit for AQ2070
AC201S
Turbidity standards for AQ2010
Signifies a hazardous solution. See terms and conditions for
important shipping information at www.thermo.com/water.
82
Thermo Scientific Orion Laboratory Products Catalog
colorimetry
Thermo Scientific Orion AQUAfast® Chemistries Overview
For many of the parameters, there are several
different choices of reagents in the AQUAfast
line. Each offers a variety of beneficial features
beyond test range.
Auto-ID™ Ampoules
Auto-ID ampoules for our AQ4000 colorimeter
offer minimal user intervention – the vial
contains the reagents, the vial pulls the proper
amount of the sample into the vial, and the first
in a portable colorimeter Auto-ID label selects
and starts the test in the AQ4000. When the
test is completed, the results are available for
viewing and download. No chemical handling
or sample pipeting, plus no pushing buttons,
which minimizes errors! The liquid remains in
the vial for easy transport back to the lab for
disposal. What could be simpler! With proper
storage, most ampoules have a shelf life of a
year or more.
available. Each powder chemistry comes with
enough packs to do 100 tests. Powder packs
are pre-measured reagents that are ready to
use. The steps of the test method are already
pre-programmed into the more advanced
AQUAfast colorimeters such as AQ4000 and
AQ3700. For added convenience, powder
packs can be used on other manufacturers’
colorimeters. With proper storage, powders
have a five year shelf life.
Tablets
Convenient tablets are individually packaged
to minimize handling. They can be pushed out
of the foil directly into the sample vial so you
do not touch the reagent chemical. A tablet
crushing stir rod is used to mix and quickly
dissolve the tablet for the measurement. The
tablets have a long shelf life – many maintain
their activity for 10 years.
Powder Packets
Liquids
AQUAfast powder chemistry packs allow for
easy and cleaner measurements. We have
a wide range of EPA equivalent chemistries
Some reagent methods such as the SPADNS
fluoride test kit and pH indicator reagents
are liquids that are added to the sample. The
SPADNS test follows the EPA method and
our pH reagents are a phenol red indicator to
determine the pH quickly. The sample’s pH is
very important for the formation of the colored
species being tested.
Digestion Test Reagents
COD, total nitrogen and total phosphate
digestion methods also are liquids that are sold
in 16 mm tubes. Just add the sample, digestion
in a thermoreactor and after the reaction with
the reagent, the measurements are taken in
the 16 mm tubes used in the preparation steps.
Similar to digestion tubes, reaction tubes do
not require a digestor to heat the samples for
the reaction to occur. We also offer COD and
other standards to be used for spiking the
samples.
The AQUAfast line of products offers choices
for sample preparation to maximize your
testing efficiency.
Convenient Reagent Starter Kits for AQUAfast AQ3700
Drinking Water – Chloramines: Cat. No. AC3K01
Nutrients Reagents: Cat. No. AC3K04
Cat. No.
Parameter
Tests
Format
Ranges mg/L
Cat. No.
Parameter
Tests
Format
Ranges mg/L
AC3072
Total chlorine HR
100
Tablet
5 to 200
ACD004
Total nitrogen LR
50
0.2 to 25
AC4P72
Total chlorine LR
100
Powder pack
0.02 to 2
Thermal
digestion
tube
AC4P29
Copper
100
Powder pack
0.05 to 5
ACD095
Total phosphorous LR
50
Nitrate
100
Reaction tube 1 to 30
Thermal
digestion
tube
0.02 to 1.6
ACR007
AC2046
Nitrite
100
Tablet
0.01 to 0.5
AC4P95
Reactive phosphorous
100
Powder
pillow
0.06 to 2.5
ACD012
Free ammonia LR
100
Thermal
digestion
tube
0.02 to 2.5
ACD011
Free ammonia HR
50
Thermal
digestion
tube
1 to 50
AC2070
Monochloramine
50
Tablet
0.02 to 2
ACR007
Nitrate
50
Reaction tube 1 to 30
AC2009
Fluoride
50
Reaction
tube
0.2 to 1.4
AC2046
Nitrite
100
Tablet
Wastewater Systems: Cat. No. AC3K03
0.01 to 0.5
Drinking Water - Free Chlorines: Cat. No. AC3K02
Cat. No.
Parameter
Tests
Format
Ranges mg/L
Cat. No.
Parameter
Tests
Format
Ranges mg/L
AC3072
Total chlorine LR
100
Tablet
5 to 200
AC4P71
Free Chlorine HR
100
Powder pack
0.02 to 2
AC2095
Phosphate
100
Tablet
0.05 to 4
AC4P72
Total Chlorine LR
100
Powder pack
0.02 to 2
AC4P82
Sulfate
100
Powder pack
5 to 100
AC4078
Iron
100
Ampoule
0.2 to 6
ACR007
Nitrate
100
Reaction tube 1 to 30
ACR007
Nitrate
100
Reaction tube 1 to 30
AC2046
Nitrite
100
Tablet
0.01 to 0.5
AC4055
Manganese
100
Ampoule
1.0 to 30
AC4012
Free ammonia LR
50
Ampoule
0.5 to 7
AC2009
Fluoride
50
Reaction
tube
0.2 to 1.4
2009 Sourcebook
83
colorimetry
Thermo Scientific Orion AQUAfast® Colorimeter
Reagent Chemistries
Parameter
Reagent
Type
Catalog
Number
Description
Range
Method
Alkalinity-m
Tablet
AC2002
AQUAfast II Alkalinity-m
5.0 - 200 mg/L
Acid/Indicator
Aluminum
Auto-Test
Ampoule
AC4027
AQUAfast IV Aluminum
Tablet
AC2027
Meter Models
EPA Approved
Methods
100
AQ4000, AQ3700,
AQ2001
-
0.04 - 0.25 mg/L Eriochrome Cyanine R
30
AQ4000
WW Equivalent
AQUAfast II Aluminum
0.05 - 0.30 mg/L Eriochrome Cyanine R
50
AQ4000, AQ3700,
AQ2005, AQ2020
WW Equivalent
Powder
AC4P12
AQUAfast Ammonia-Nitrogen, Powder
0.1 - 0.50 mg/L
Salicylate
100
AQ4000, AQ3700
-
Auto-Test
Ampoule
AC4010
AQUAfast IV Ammonia (Ultra Low Range)
0.10 - 3.0 mg/L
Salicylate
30
AQ4000
-
Auto-Test
Ampoule
AC4012
AQUAfast IV Ammonia (Low Range)
0.5 - 7.0 mg/L
Nesslerization
30
AQ4000
WW Equivalent
Auto-Test
Ampoule
AC4011
AQUAfast IV Ammonia (High Range)
1.0 - 14 mg/L
Nesslerization
30
AQ4000
WW Equivalent
Tablet
AC2012
AQUAfast II Ammonia, as Nitrogen
0.05 - 1.0 mg/L;
0.5 - 10.0 mg/L
Indophenole/phenate
50
AQ4000, AQ3700,
AQ2004
WW Equivalent
Auto-Test
Ampoule
AC4035
AQUAfast IV Bromine
0.75 - 9.0 mg/L
DPD
30
AQ4000
No requirement
Tablet
AC2035
AQUAfast II Bromine
0.1 - 13.5 mg/L
DDPD
100
AQ3700, AQ2005
No requirement
Auto-Test
Ampoule
AC4017
AQUAfast IV Chloride
2.0 - 40 mg/L
Mercuric Nitrate
30
AQ4000
-
Tablet
AC2017
AQUAfast II Chloride
0.5 - 25 mg/L
Silver nitrate/turbidity
50
AQ4000, AQ3700,
AQ2005
-
Auto-Test
Ampoule
AC4099
AQUAfast IV Chlorine Dioxide
0.80 - 12 mg/L
DPD
30
AQ4000
No requirement
Tablet
AC2099
AQUAfast II Chlorine Dioxide
0.04 - 2.8 mg/L
DPD
100
AQ4000, AQ3070,
AQ3700, AQ2020
No requirement
Powder
AC4P71
AQUAfast Chlorine, Free, DPD, Powder Packets, 0.03 - 2.00 mg/L DPD
EPA-approved Method
100
AQ4000, AQ3070
DW
Powder
AC4P71B
AQUAfast Chlorine, Free, Powder Packets
Bulk Packaging, EPA-approved Method
0.03 - 2.00 mg/L DPD
1000
AQ4000, AQ3070
DW
Tablet
AC2071
AQUAfast II Chlorine, DPD, Free Only
0.05 - 6.0 mg/L
DPD
100
AQ4000, AQ2070
-
Auto-Test
Ampoule
AC4070
AQUAfast IV Chlorine, Free and Total,
EPA-approved Method
0.05 - 6.0 mg/L
DPD
30
AQ4000
DW
Tablet
AC2070
AQUAfast II Chlorine, Free and Total
0.05 - 6.0 mg/L
DPD
50 each
AQ4000, AQ3700,
AQ2070, AQ2020
-
Powder
AC4P72
AQUAfast Chlorine, Total, DPD, Powder Packets, 0.03 - 2.00 mg/L DPD
EPA-approved Method
100
AQ4000, AQ3070
DW & WW
Powder
AC4P72B
AQUAfast Chlorine, Total, Powder Packets
Bulk Packaging, EPA-approved Method
0.03 - 2.00 mg/L DPD
1000
AQ4000, AQ3070
DW & WW
Tablet
AC2072
AQUAfast II Chlorine, Total,
EPA-approved Method
0.05 - 6.0 mg/L
50
AQ4000, AQ3700,
AQ2070
DW & WW
Powder
AC4P29
AQUAfast Copper, Bicinchoninate, Powder
0.05 - 5.00 mg/L Bicinchoninate
100
AQ4000, AQ3700
WW Equivalent
Auto-Test
Ampoule
AC4029
AQUAfast IV Copper (Soluble)
0.5 - 12 mg/L
Bathocuproine
30
AQ4000
WW Equivalent
Tablet
AC2065
AQUAfast II Zinc/Copper
0.02 - 1.0 mg/L
Zincon
50
AQ4000, AQ3700,
AQ2002
-
Tablet
AC2029
AQUAfast II Copper
0.05 - 5.0 mg/L
Biquinoline
50
AQ4000, AQ3700,
AQ2005
-
Cyanide
Auto-Test
Ampoule
AC4006
AQUAfast IV Cyanide (Free)
0.20 - 0.50 mg/L Pyridine-barbituric acid
30
AQ4000, AQ3700
ETV Verified, DW
& WW Equivalent
Cyanuric Acid
Tablet
AC2098
AQUAfast II Cyanuric Acid
5 - 90 mg/L
Melamine
100
AQ3070, AQ3700,
AQ2003
No requirement
Fluoride
Auto-Test
Ampoule
AC4009
AQUAfast IV Fluoride SPADNS
0.20 - 2.0 mg/L
SPADNS
30
AQ4000
DW & WW
Liquid
AC2009
AQUAfast II Fluoride SPADNS Kit,
EPA-approved Method
0.05 - 20 mg/L
SPADNS
50
AQ4000, AQ3700,
AQ2009, AQ2020
DW & WW
Auto-Test
Ampoule
AC4030
AQUAfast IV Hydrazine
0.10 - 1.2 mg/L
PDMAB
30
AQ4000, AQ3700
No requirement
Tablet
AC2030
AQUAfast II Hydrazine Kit
0.05 - 0.50 mg/L Dimethylaminobenzaldehyde
100
AQ4000, AQ2003
No requirement
Ammonia
(Nitrogen)
Bromine
Chloride
Chlorine
Dioxide
Chlorine
Copper
Hydrazine
84
Thermo Scientific Orion Laboratory Products Catalog
DPD
Number
of Tests
colorimetry
Parameter
Reagent
Type
Catalog
Number
Description
Range
Method
Number
of Tests
Meter Models
EPA Approved
Methods
Iron
Powder
AC4P79
AQUAfast Iron Total TPTZ, Powder
0 - 1.80 mg/L
TPTZ
100
AQ4000, AQ3700
-
Powder
AC4P78
AQUAfast Iron Total Ferro, Powder
Auto-Test
Ampoule
0 - 3.00 mg/L
1,10 Phenanthroline
100
AQ3700
WW Equivalent
AC4078
AQUAfast IV Iron 1 (Total & Soluble)
0.45 - 7.0 mg/L
1,10 Phenanthroline
30
AQ4000
WW Equivalent
TPTZ
100
AQ4000, AQ3700,
AQ2005
-
Tablet
AC2078
AQUAfast II Iron
0.02 - 1.0 mg/L;
0.2 - 10 mg/L
Powder
AC4P55
AQUAfast Manganese HR, Powder
0.2 - 20.0 mg/L
Periodate Oxidation
100
AQ4000
WW Equivalent
Auto-Test
Ampoule
AC4055
AQUAfast IV Manganese
1.0 - 30 mg/L
Periodate Oxidation
30
AQ4000
WW Equivalent
Tablet
AC2055
AQUAfast II Manganese
0.05 - 4 mg/L
Formaldoxime
50
AQ4000, AQ3700,
AQ2003
-
Powder
AC4P42
AQUAfast Molybenum, Molybdate, HR, Powder 0 - 40.0 mg/L
Mercaptoacetic acid
100
AQ4000, AQ3700
-
Auto-Test
Ampoule
AC4042
AQUAfast IV Molybdate
1.0 - 25 mg/L
Catechol
30
AQ4000
-
Auto-Test
Ampoule
AC4007
AQUAfast IV Nitrate 3 (as NO3)
5.0 - 50 mg/L
Cadmium reduction
30
AQ4000
DW & WW
Equivalent
Auto-Test
Ampoule
AC4005
AQUAfast IV Nitrate 2 (as N)
0.40 - 3.0 mg/L
Cadmium reduction
30
AQ4000
DW & WW
Equivalent
Auto-Test
Ampoule
AC4004
AQUAfast IV Nitrate (as N) LR
0.20 - 1.5 mg/L
Cadmium reduction
30
AQ4000
DW & WW
Equivalent
Tablet
AC2007
AQUAfast II Nitrate
1 - 40 mg/L
Zinc reduction
50
AQ4000, AQ2007
-
Powder
AC4P46
AQUAfast Nitrite LR, Powder
0 - 0.350 mg/L
Diazotization (Azo)
100
AQ4000, AQ3700
DW & WW
Equivalent
Auto-Test
Ampoule
AC4046
AQUAfast IV Nitrite
0.080 - 0.80
mg/L
Diazotization (Azo)
30
AQ4000
DW & WW
Equivalent
Tablet
AC2046
AQUAfast II Nitrite, EPA-approved Method
0.05 - 0.50 mg/L Diazotization (Azo)
100
AQ4000, AQ3700,
AQ2046
DW & WW
Oxygen
Auto-Test
Ampoule
AC4008
AQUAfast IV Oxygen
2.0 - 15 mg/L
Azo dye
30
AQ4000
-
Ozone
Auto-Test
Ampoule
AC4048
AQUAfast IV Ozone
0.20 - 2.5 mg/L
DDPD
30
AQ4000
-
Tablet
AC2048
AQUAfast II Ozone
0.05 - 0.5 mg/L
DPD/Glycine
100
AQ3700, AQ2001
-
Liquid
AC3001
AQUAfast II pH, Dropper bottle
5.9 - 8.2
Phenol red
30
AQ3070, AQ3700
-
Tablet
AC2001
AQUAfast II pH
6.5 - 8.4
Phenol red
100
AQ4000, AQ3700,
AQ2001
-
Powder
AC4P95
AQUAfast Phosphorus 3, Powder
0.07 - 2.5 mg/L
Phosphormolybdic
blue/Ascorbic acid
100
AQ4000, AQ3700
DW & WW
Equivalent
Auto-Test
Ampoule
AC4096
AQUAfast IV Phosphate 1-HR
5.0 - 80 mg/L
Vanado molybdate
30
AQ4000, AQ3700
-
Auto-Test
Ampoule
AC4095
AQUAfast IV Phosphate 2-LR
0.75 - 8.0 mg/L
Stannous chloride
30
AQ4000, AQ3700
-
Tablet
AC2096
AQUAfast II Phosphorus HR
10 - 100 mg/L
Phosphormolybdic
blue/Ascorbic acid
50
AQ4000, AQ3700,
AQ2003
DW & WW
Equivalent
Tablet
AC2095
AQUAfast II Phosphorus LR,
EPA Approval Method
0.05 - 4.0 mg/L
Vanado Molybdate
50
AQ4000, AQ3700,
AQ2004
DW & WW
Powder
AC4P60
AQUAfast Silica HR, Powder Packets
1.0 - 75.0 mg/L
Silicomolybdate
100
AQ4000, AQ3700
DW & WW
Equivalent
Auto-Test
Ampoule
AC4060
AQUAfast IV Silica
1.0 - 14 mg/L
Heteropoly Blue
30
AQ4000
DW Equivalent
Powder
AC2061
AQUAfast II Silica Phosphate Removal
NA
NA
100
AQ4000, AQ3700,
AQ2002
-
Tablet
AC2060
AQUAfast II Silica
0.05 - 4.0 mg/L
Silicomolybdate
50
AQ4000, AQ3700,
AQ2002
DW & WW
Equivalent
Powder
AC4P82
AQUAfast Sulfate 4, Powder Packets
2 - 70 mg/L
Barium sulfate/
Turbidity
100
AQ4000, AQ3700
DW & WW
Equivalent
Auto-Test
Ampoule
AC4082
AQUAfast IV Sulfate
8.0 - 100 mg/L
Barium sulfate/
Turbidity
30
AQ4000
DW & WW
Equivalent
Tablet
AC2082
AQUAfast II Sulfate
5 - 200 mg/L
Barium sulfate/
Turbidity
100
AQ4000, AQ3700,
AQ2005
DW & WW
Equivalent
Auto-Test
Ampoule
AC4016
AQUAfast IV Sulfide (Soluble)
0.3 - 3.0 mg/L
Methylene Blue
30
AQ4000
WW Equivalent
Tablet
AC2016
AQUAfast II Sulfide
0.05 - 0.50 mg/L Methylene Blue
50
AQ4000, AQ3700,
AQ2004
WW Equivalent
Auto-Test
Ampoule
AC4065
AQUAfast IV Zinc
0.30 - 3.0 mg/L
Zincon
30
AQ4000
WW Equivalent
Tablet
AC2065
AQUAfast IV Zinc/Copper
0.02 - 1.0 mg/L
Zincon
50
AQ4000, AQ3700,
AQ2002
WW Equivalent
Manganese
Molybdate
Nitrate
Nitrite
pH
Phosphate
Silica
Sulfate
Sulfide
Zinc
2009 Sourcebook
85
cross reference
Orion Part Numbers
Cross Reference Guide
Orion
Cat. No.
Page
No.
Orion
Cat. No.
Page
No.
Orion
Cat. No.
Orion
Cat. No.
Page
No.
010010
39
080513
43
615701
*
900017
35
011001
39
080514
43
616500
*
900018
35
011005
39
080515
43
616501
*
900019
*
011006
39
081003
43
650700
35
900020
27
011007
39
081010
43
652012
*
900021
27
011008
39
081015
*
654201
35
900022
27
011010
37
081045
43
654202
35
900023
27
011020
37
087001
40
654203
35
900024
27
011050
37
087002
40
654204
35
900046
35
011220
39
087003
40
654205
35
900060
35
011221
39
090032
28
700001
27
900061
35
011510
37
090033
28
700003
27
900062
35
013017
39
090034
28
700005
35
900063
35
35
900065
35
013045
39
090035
28
700006
013610
36
090036
28
700010
27
900067
35
013655
71
090037
28
700011
27
900100
23
014005
*
090043
72
700012
27
900200
23
020017
27
090044
*
700402
27
910001
26
020125
72
090045
72
700702
27
910002
72
020130
72
090048
28
700902
27
910003
27
020135
72
090070
72
711002
*
910004
27
032000
70
095000
*
712001
20
910005
27
032001
70
095001
*
712002
20
910006
27
032002
*
095002
*
800300
10
910008
27
032004
70
095003
*
0807001
*
910026
28
032005
70
095004
*
810001
26
910027
28
032006
70
096000
*
810007
26
910028
28
032007
70
096002
*
810200
*
910029
28
035000
70
096010
*
810300
*
910030
28
035001
70
096011
*
810400
*
910031
28
*
910060
26
035002
*
096014
*
811500
035003
70
096016
*
815500
*
910104
26
035004
70
096017
*
815600
11
910105
26
037000
70
096018
*
841101
35
910107
26
037001
70
096019
*
841108
35
910110
26
037002
*
096021
*
841109
35
910112
26
040030
*
096025
*
841111
35
910168
26
040032
*
096035
*
841113
35
910199
26
040906
65
096038
*
900001
35
910410
26
040907
65
096039
*
900002
35
910425
26
040908
65
096045
*
900003
35
910460
26
080017
43
096050
*
900004
35
910500
*
080045
43
510011
19
900010
35
910686
26
080113
43
510015
27
900011
26
910710
26
080160
43
610011
*
900012
35
910725
26
080360
43
615700
*
900014
*
910760
26
* For the latest information on these products, visit www.thermo.com/water
86
Page
No.
Thermo Scientific Orion Laboratory Products Catalog
cross reference
Orion
Cat. No.
Page
No.
Orion
Cat. No.
Page
No.
Orion
Cat. No.
Page
No.
Orion
Cat. No.
Page
No.
910918
26
940906
35
967901
26
1010901
72
911010
26
940907
35
967961
26
01100510
39
911025
26
940909
35
970701
35
01100610
39
911060
26
940911
35
970800
41
01100710
39
911110
26
940916
35
970801
43
01100910
39
911125
26
940999
35
970802
43
1110001
*
911260
26
941104
35
970803
43
1111000
68
911600
21
941105
35
971901
35
1111001
68
912600
21
941107
35
972001
35
1111101
*
913600
21
941609
35
974601
*
1111102
*
915500
*
941706
35
977001
*
1111103
*
916860
26
941707
35
977007
35
1111104
*
917001
28
941708
35
977010
35
1111105
*
917002
28
941709
35
977011
35
1111106
*
917004
21
942906
35
977900
23
1111107
*
917005
24
943506
35
990101
*
1111108
*
917006
24
945306
35
990106
39
1111201
68
917007
24
948201
35
1010001
39
1111202
68
917008
*
948206
35
1010003
72
1111203
68
920706
35
948207
35
1010004
72
1111204
68
920707
35
950202
35
1010005
72
1111206
68
921906
35
950204
35
1010006
72
1111207
68
922006
35
950206
35
1010007
69
1111208
68
923206
35
950207
35
1010008
72
1112000
63
927005
24
950210
35
1010009
72
1112001
63
927006
24
951004
*
1010010
*
1112002
63
927007
24
951006
*
1010017
69
1112003
63
930000
35
951007
*
1010020
*
1112004
63
930101
35
951011
35
1010021
*
1112006
63
930501
35
951201
33
1010050
72
1112102
63
930701
35
951202
33
1010051
72
1112104
*
930702
35
951203
35
1010053
72
1112201
63
930707
35
951204
33
1010102
57
1112202
63
930710
35
951205
33
1010103
57
1112203
63
930711
35
951207
35
1010108
57
1112204
63
931701
35
951209
33
1010109
57
1112205
63
931801
35
951211
33
1010111
*
1112906
*
931901
35
951214
33
1010114
67
1113000
67
931911
35
951215
33
1010115
63
1113001
67
932001
35
954602
35
1010121
67
1114000
65
932011
35
954604
35
1010138
67
1114001
65
933201
35
954606
35
1010139
*
1114002
65
934610
35
956410
35
1010152
55
1114003
65
938101
35
960900
32
1010800
72
1115000
57
940011
35
961700
*
1010801
72
1115001
57
940900
32
967800
32
1010900
72
1115002
57
* For the latest information on these products, visit www.thermo.com/water
2009 Sourcebook
87
cross reference
Orion Part Numbers
Cross Reference Guide
Orion
Cat. No.
Page
No.
Orion
Cat. No.
Page
No.
Orion
Cat. No.
Orion
Cat. No.
Page
No.
1115005
57
1216002
61
013020MD
38
083025MD
41
1116000
61
1216502
61
013025MD
38
083060MD
41
1116001
61
1217000
59
0131S0
71
086020A
41
1116201
61
1217001
59
0131S2
*
086021A
43
1117000
59
1217002
59
0131S3
71
086030MD
41
1117001
59
1217101
59
013605MD
36
087010MD
40
1117201
59
1217501
59
013610MD
36
087020MD
40
1119000
55
1218000
53
0136S0
71
087030MD
40
1119001
55
1218001
53
0136S2
*
087050MD
40
1119201
55
1218101
*
0136S3
71
087100MD
40
1210001
72
1218501
53
018020MD
37
0900A0
72
1210002
72
1219000
55
0261S0
71
0900A1
72
1210003
72
1219001
55
0261S2
*
0900A2
*
1210004
72
1219101
55
0261S3
71
0900A4
72
1210005
72
2115000
49
0266S0
71
0900A6
72
1212000
63
2115001
49
0266S2
*
0900A7
72
1212001
63
2115101
49
0266S3
71
096025L
*
1212002
63
2115102
49
0320A0
*
0960KF
*
1212101
63
5122001
*
0320A1
*
0960PC
*
1212501
63
5122003
*
0320A7
*
096S22
73
1212503
63
5124002
*
0320A8
*
096SRT
*
1213000
67
5222004
*
0320A9
*
0ACBL0
72
1213001
67
5222006
*
0350A0
*
0BCHSK
72
1213101
67
5224003
*
0350A1
*
0GLP05
*
1213300
51
9116860
26
0350A7
*
0GLP30
*
1213301
51
9191860
26
0350A8
*
0GLP31
*
1213302
51
10009999
*
0350A9
*
0GLP35
*
1213310
51
10060003
*
0350G0
*
0PBLSK
72
1213311
51
10060020
*
0370A0
*
1112000SPA
73
1213312
51
10060040
*
0370A1
*
1112000SPB
73
1213320
51
10060080
*
0370A7
*
1114000SPA
73
1213321
51
0105PK
39
0370A8
*
1114000SPB
73
1213322
51
011010A
37
0370A9
*
1115000SPA
73
1213501
*
011020A
37
0370G0
*
1115000SPB
73
1214000
65
011050A
37
0370G1
*
1117000SPA
73
1214001
65
011050MD
37
0607CA
72
1117000SPB
73
1214003
65
011510A
37
0607CI
72
1119000SPA
73
1214004
65
011510MD
37
080510MD
41
1119000SPB
73
1214101
65
013005A
38
081010F
41
254210-001
?
1214501
65
013005D
38
081010MD
41
5107BNMD
20
1214504
65
013005MD
38
083005A
41
5107NC
20
1215000
57
013010A
38
083005D
41
5109NC
*
1215001
57
013010MD
38
083005MD
41
55PR40
*
1215501
57
013016A
38
083010A
41
55PR45
*
1216000
61
013016D
38
083010MD
41
55PR49
*
1216001
61
013016MD
38
083025A
*
7102BN
*
* For the latest information on these products, visit www.thermo.com/water
88
Page
No.
Thermo Scientific Orion Laboratory Products Catalog
cross reference
Orion
Cat. No.
Page
No.
Orion
Cat. No.
Page
No.
Orion
Cat. No.
Page
No.
Orion
Cat. No.
Page
No.
7120BN
*
8411BN
31
9157BN
17
9300SC
35
7121BN
*
8611BNWP
31
9157BNMD
17
9305BN
30
800500U
8
9100CB
26
9161BN
16
9307BNWP
34
8101BNWP
10
9101BN
16
9161SC
*
9319BN
31
8101SC
8
9101SC
23
9161SCE
*
9320BN
30
8102BN
10
9101SCE
*
9162BNWP
17
9332BNWP
31
8102BNUWP
9
9102AP
15
9162SC
*
9342BN
31
8102SC
*
9102BNWP
16
9162SCE
*
9406BN
30
8103BN
11
9102DJWP
20
9163SC
21
9406SC
30
8103BNUWP
9
9102SC
*
9163SCE
*
9409BN
30
8103SC
*
9102SCE
*
9165BNWP
17
9409SC
30
8104BN
11
9103APWP
15
9166SC
21
9416BN
31
8104BNUWP
9
9103BNE
*
9166SCE
*
9416SC
31
8104SC
*
9103BNWP
16
9167SC
21
9417BN
30
8107BNU
8
9103SC
*
9167SCE
*
9417SC
30
8107BNUMD
8
9103SCE
*
9172BNWP
17
9429BN
30
8107UWMMD
9
9104APWP
15
9179BN
22
9429SC
30
8115BN
11
9104BNWP
16
9179BNMD
22
9435BN
30
8115BNUWP
9
9104CB
26
9180BN
22
9435SC
30
8115SC
*
9104SC
*
9180BNMD
22
9448BN
30
8135BN
11
9104SCE
*
91CBBR
28
9448SC
30
8135BNUWP
9
9106BNWP
21
91CBBRE
*
9453BN
31
8135SC
*
9107APMD
*
91CBNC
28
9453SC
31
8155SC
*
9107BN
16
91CBNCE
*
9458BN
31
8156BNUWP
9
9107BNMD
16
91CBNT
28
9482BN
31
8157BNU
8
9107CB
26
91CBRA
28
9482SC
31
8157BNUMD
8
9107WLMD
51
91CBRAE
*
9502BNWP
30
8157UWMMD
9
9107WMMD
51
91CDIN
28
950DPU
*
8162SC
*
9107WP
*
91CDINE
*
950EPN
*
8162SCE
*
9109WL
17
91CLM0
*
9512BNWP
30
8163BNWP
11
9109WP
17
91USCB
28
9512DN
*
8163SC
*
9110CB
26
91USCBE
*
9512HP01
30
8163SCE
*
9110DJWP
20
91USHC
28
9512HPBNWP
30
8165BNWP
12
9115APWP
5
91USHCE
*
9546BN
31
8165DN
*
9120APWP
20
91USRF
28
9606BNWP
30
8172BNWP
12
9121APWP
20
9202BN
14
9609BNWP
30
8172DN
*
9135APWP
15
9203BN
14
960DPU
*
8175BNWP
12
9142BN
21
9206BN
14
9616BNWP
31
8202BN
13
9145BN
21
9207BN
14
9617BNWP
30
8203BN
13
9146BN
21
9256BN
14
9629BNWP
30
8207BNU
8
9147BN
21
927005MD
24
9635BNWP
30
8220BNWP
13
9156APWP
15
927006MD
24
9648BNWP
30
8235BN
13
9156BNWP
17
927007MD
24
9653BNWP
31
8256BN
13
9156DJWP
20
9272BN
14
9678BNWP
22
8257BNU
8
9156SC
*
928007MD
24
9682BNWP
31
8272BN
13
9156SCE
*
9300BNWP
35
9700BNWP
35
* For the latest information on these products, visit www.thermo.com/water
2009 Sourcebook
89
cross reference
Orion Part Numbers
Cross Reference Guide
Orion
Cat. No.
Page
No.
Orion
Cat. No.
Page
No.
Orion
Cat. No.
Orion
Cat. No.
Page
No.
9707BNWP
31
AC2095
85
AC4P78
85
CODL00
81
970899WP
41
AC2096
85
AC4P79
85
CODS01
81
9719BNWP
31
AC2099
78
AC4P82
85
CODS10
81
9720BNWP
30
AC2T24
82
AC4P95
85
DTCL20
*
9746BNWP
*
AC2V16
82
AQ2001
85
DTCL21
*
9770BNWP
30
AC2V24
82
AQ2002
84
DTPL20
*
9770SC
30
AC4004
85
AQ2003
84
DTPL21
*
9778BNE
*
AC4005
85
AQ2004
84
DTPL22
*
9778BNWP
22
AC4006
84
AQ2005
84
GD9106BNWP
18
9778SC
*
AC4007
85
AQ2007
84
GD9156BNWP
18
9778SCE
*
AC4008
85
AQ2009
84
GS9106BNWP
18
977900E
*
AC4009
84
AQ2020
84
GS9156BNWP
18
9780SC
23
AC4010
84
AQ2040
81
IQOQT2
*
9780SCE
*
AC4011
84
AQ2046
85
M361
*
9781SC
23
AC4012
84
AQ3010
77
M361F
*
9781SCE
*
AC4016
85
AQ4000
81
M362
*
9802BN
19
AC4017
84
AQ40FK
82
M362F
*
9803BN
19
AC4027
84
AQ4500
76
M365
*
9810BN
19
AC4029
84
AQ4CBL
76
M365F
*
9811BN
31
AC4030
84
AQ4EFC
*
PRT300
72
9826BN
19
AC4035
84
AQ4EK1
82
PRT301
72
9863BN
19
AC4042
85
AQ4ZER
82
PRT302
72
AC2001
85
AC4046
85
AS0050
*
RS0SML
*
AC2002
84
AC4048
85
AS0ACC
*
RS0STD
*
AC2007
85
AC4055
85
AS0CON
*
AC2009
83
AC4060
85
AS0STD
*
AC2012
84
AC4065
85
AS100K
*
AC2016
85
AC4070
84
ASB050
*
AC2017
84
AC4078
85
ASB100
*
AC201S
82
AC4082
85
ASB150
*
AC2027
83
AC4095
85
ASB250
*
AC2029
84
AC4096
85
ASCBL1
*
AC2030
84
AC4099
84
ASEADP
*
AC2035
84
AC45FZ
76
ASEXT1
*
AC2046
85
AC45SI
76
ASEXT2
*
AC2048
85
AC45ST
76
AT0050
73
AC2055
85
AC4A13
*
AT0STD
73
AC2060
85
AC4A16
*
AT5050
73
AC2061
85
AC4P12
84
AT5150
73
AC2065
85
AC4P29
84
ATCBL1
*
AC2070
83
AC4P42
85
ATR050
73
AC2071
84
AC4P46
85
ATR150
73
AC2072
84
AC4P55
85
CBL300
*
AC2073
82
AC4P60
85
COD125
81
AC2078
85
AC4P71
78
CODH00
81
AC2082
85
AC4P72
78
CODHP0
81
* For the latest information on these products, visit www.thermo.com/water
90
Page
No.
Thermo Scientific Orion Laboratory Products Catalog
notes
Catalog Notes
2009 Sourcebook
91
index
Index
A
Accessories
Conductivity Cell 39
Dissolved Oxygen Probe 43
ISE 35
Meter 72
pH Electrode 26-27
RDO Optical DO 40, 51
BOD (see also Dissolved Oxygen)
Adapters 28
AUTO-STIR Probe 41, 42
Meter Kits 54, 60, 66
Adapters, Electrode 28
Bromide ISE
Electrode 30
Solutions 30, 35
Adapters, Power 72
Buffers, pH 26-27
Ag/AgCl Electrodes
pH 14-21
ORP 22
Reference 23
Bulb Guard, Electrode 27
Alkalinity Test Kit, Total 27
C
Ammonia ISE
Electrode 30, 33
Membranes 33, 35
Solutions 33, 35
Cables 28, 72
Ammonium ISE
Handle 35
Module 30, 35
Solutions 30, 35
Calcium ISE
Electrode 30
Modules 35
Solutions 30, 35
AQUAfast
COD 81
Colorimeter 78, 79, 82
Turbidity Meter 76, 77
Calibration Standards
Conductivity 39
ISE 35
ORP 26
AquaPro Professional pH Electrodes 15
Calomel Replacement Electrodes
AquaPro Professional 15
Double Junction 20
KNIpHE 20
ROSS 10-13
ROSS Ultra 8-9
ATC (Automatic Temp. Compensation) Probes 24
AutoTration-500 Autosampler 73
B
Basic pH Electrodes 21
Beaker Racks 73
Beakers 73
Benchtop Meters
Accessories 72
Concentration 48, 54, 56, 70
Conductivity 54, 58, 64
Dissolved Oxygen 54, 60, 66
ISE 48, 54, 56, 70
Millivolt (mV) 48, 54, 56, 58, 60, 62, 68, 70
Multi-parameter 48, 54, 56, 58, 60, 70
Orion DUAL STAR 48-49
PerpHecT 70
pH 48, 54, 56, 58, 60, 62, 68, 70
Selection Guide 46-47
Star Series 48-68
92
BNC Shorting Cap 72
Thermo Scientific Orion Laboratory Products Catalog
Cadmium ISE
Electrode 30
Solutions 30, 35
Carbon Dioxide ISE
Electrode 30
Membranes 35
Solutions 30, 35
Chloride ISE
Electrode 30
Handle 35
Module 30, 35
Solutions 30, 35
Chlorine ISE
Electrode 30
Solutions 30, 35
Cleaning Solutions, pH Electrode 7, 27
COD (Chemical Oxygen Demand) 81
index
E
Colorimeters 74-85
Colorimetry Chemistries 83-85
Computer
Cables 72
Software 69
Concentration (see ISE)
Conductivity
2-Electrode 36-39
4-Electrode 36, 38-39
Accessories 39
Cells 36-39
DuraProbe 36, 38-39
Meter Calibration Kit 39
Meters 50, 52, 54, 58, 64, 71
Selection Guide 38-39
Standards 39
USP Kit 55
Connectors 28
Cupric ISE
Electrode 30
Solutions 30, 35
Cyanide ISE
Electrode 30
Solutions 30, 35
D
Data Collection
Star Navigator 21 Software 69
Dissolved Oxygen (see also BOD)
Accessories 43
Calibration Sleeve 40, 43
Maintenance Kits 43
Membrane Caps 43
Meters 50, 52, 54, 60, 66
Probes 40-43
RDO Optical DO 40, 50
Selection Guide 42-43
Double Junction pH Electrodes 20
Economy pH Electrodes 21
Electrode (see also ISE, pH)
Adapters 28
Cables 28
Cleaning Kits 7, 27
Filling Solutions 26
ORP/Redox 22
Reference 8, 12, 23
Selection Guide 4-5
Stands 72
Storage Bottles 27
Storage Solutions 26
Sure-Flow 12, 13, 14, 17, 22, 23
Triode 8, 14, 15, 16, 17, 20, 22
Extension Cables 28
F
Filling Solutions
DO 43
ISE 35
pH 26
Fluoride ISE
Electrode 30, 32
Solutions 30, 32, 35
Fluoride Standards Bulk Pack 35
Fluoroborate ISE
Electrode 30
Module 35
Solutions 30, 35
G
Gas Sensing Ion Selective Electrodes 30-31, 33
GLP Printer 72
Green Electrodes 18
Guards, Electrode 27, 39, 40, 43
Dual Channel Meters 48-49
DUAL STAR 48-49
DuraProbe Conductivity Cells 36, 38-39
H
Half-Cells
ISE 30-31, 32, 34
pH 10, 16
Reference 8, 12, 23
2009 Sourcebook
93
index
I
M
Interface Cables 28, 72
Maintenance Kits for DO Probes 43
Intrinsically Safe Meters 71
Membrane Caps for DO Probes 43
Iodide ISE
Electrode 31
Solutions 31, 35
Membranes for Gas Sensing ISE 35
Ionic Strength Adjusters (ISA) 30-31, 35
Meters
Concentration 48, 54, 56, 70
Conductivity 50, 52, 54, 58, 64, 71
Dissolved Oxygen 50, 52, 54, 60, 66
Intrinsically Safe 71
ISE 48, 54, 56, 70
Millivolt 48, 50, 52, 54, 56, 58, 60, 62, 68, 70, 71
Multi-parameter 48, 50, 52, 54, 56, 58, 60, 70
Orion DUAL STAR 48-49
ORP 48, 50, 52, 54, 56, 58, 60, 62, 68, 70
PerpHecT 70
pH 48, 50, 52, 54, 56, 58, 60, 62, 68, 70, 71
Selection Guide 46-47
Star Series 48-68
ionplus Ion Selective Electrodes 30-31
ISE (Ion Selective Electrode)
Accessories 35
Calibration Standards 35
Electrodes 30-31, 32, 33, 34
Filling Solutions 35
Handles 35
Ionic Strength Adjusters (ISA) 35
Membranes 35
Meters 48, 54, 56, 70
Modules 35
Reference Electrodes 8, 12, 23
Selection Guide 30, 31
Special Reagents 35
K
KNIpHE
pH Electrode 20
Replacement Blade 20
Replacement Electrode 20
L
Lead ISE
Electrode 31
Solutions 31, 35
LogR Temperature Compensation 70
Meter Calibration Kit for Conductivity 39
Micro Electrodes
ATC 24
PerpHecT ROSS 13
pH 19
Sodium 31
Multi-parameter Meters 48, 50, 52, 54, 56, 58, 60, 70
N
Nitrate ISE
Electrode 31, 34
Modules 34, 35
Solutions 31, 34, 35
Nitrite ISE
Electrode 31
Modules 35
Solutions 31, 35
Nitrogen Oxide ISE
Electrode 31
Membranes 35
Solutions 31, 35
No Cal pH Electrodes 20
94
Thermo Scientific Orion Laboratory Products Catalog
index
O
Optical DO 40, 50
Orion DUAL STAR 48-49
Oxidation Reduction Potential (ORP)
Electrodes 22
Meters 48, 50, 52, 54, 56, 58, 60, 62, 68, 70
Standards 26
Oxygen Electrodes (see Dissolved Oxygen, BOD)
Potassium ISE
Electrode 31
Modules 35
Solutions 31, 35
Power Adapters 72
Printers 72
Probes (see ATC, BOD, Dissolved Oxygen,
Conductivity)
Protective Guards 27, 39, 40, 43
P
Perchlorate ISE
Handle 31
Module 35
Solutions 31, 35
PerpHecT
Digital LogR Meters 70
pH Electrodes 13, 14
ROSS Micro pH Electrode 13
PerpHecT Pouches
Conductivity Standards 39
pH Buffers 26
Rinse Solution 26
pH Electrode
Accessories 26-28
Buffers 26-27
Cleaning Solution Kits 7, 27
Filling Solutions 26
Rinse Solution 26
Selection Guide 4-5
Storage Solution 26
pH Meter
Kits 48, 50, 52, 54, 56, 58, 60, 62, 68, 70, 71
Selection Guide 46-47
pH/ISE Meter
Kits 48, 54, 56, 70
Selection Guide 46-47
Pure Water pH Test Kit 27
R
RDO
Meters 50
Probes 40
Reagents
Colorimetry 78-85
ISE 35
Redox/ORP Electrodes 22
Reference Electrodes
ROSS Sure-Flow 12
ROSS Ultra 8
Sure-Flow 23
Rinse Solution 26
ROSS Electrodes
Micro pH 13
pH 8-13
Reference 8, 12
Sodium 31
ROSS Ultra Electrodes
pH 8- 9
Reference 8
Plastic Membrane Ion Selective Electrodes 30-31, 34
S
Platinizing Kit, for Conductivity Cells 39
Salinity 50, 52, 54, 58, 64, 71
Portable Meters
Accessories 72
Concentration 54, 56
Conductivity 50, 52, 54, 58, 64, 71
Dissolved Oxygen 50, 52, 54, 60, 66
Intrinsically Safe 71
ISE 54, 56
Millivolt 50, 52, 54, 56, 58, 60, 62, 71
Multi-parameter 50, 52, 54, 56, 58, 60
pH 50, 52, 54, 56, 58, 60, 62, 71
Selection Guide 46-47
Star Series 48-68
Selection Guide
Conductivity Cell 38-39
DO Probe 42-43
ISE 30-31
Meter 46-47
pH Electrode 4-5
Shorting Cap, BNC 72
2009 Sourcebook
95
index
Silver Billet Electrode 23
Silver/Sulfide ISE
Electrode 31
Solutions 31, 35
Sodium ISE
Electrode 31
Micro Electrode 31
Solutions 31, 35
TRIS Compatible pH Electrodes
AquaPro Professional 15
Double Junction 20
KNIpHE 20
PerpHecT ROSS 13
ROSS 10-12
ROSS Ultra 8-9
Turbidity Meter 76-77
Solid State Ion Selective Electrodes 30- 31, 32
Stands, for Electrodes 72
Star Navigator 21 Software 69
Star Series
2-Star 68
3-Star 50-51, 62-67
4-Star 50-51, 56-61
5-Star 50-51, 52-55
Accessories 72
Computer Cable 72
Meters 48-68
Orion DUAL STAR 48-49
Printer 72
Selection Guide 46-47
Software 69
Stirrer Probe 25, 72
Storage Solutions 26, 35
Sure-Flow Electrodes 12, 13, 14, 17, 22, 23
Surfactant ISE
Electrode 31
Module 35
Solutions 31, 35
T
TDS (Total Dissolved Solids) 50, 52, 54, 58, 64, 71
Temperature Probes 24
Test Kits
Alkalinity 27
Pure Water 27
Thiocyanate ISE
Electrode 31
Solutions 31, 35
Total Alkalinity Test Kit 27
Total Ionic Strength Adjustment Buffer
(TISAB) 30, 32, 35
Triode 8, 14, 15, 16, 17, 20, 22
96
Thermo Scientific Orion Laboratory Products Catalog
U
Ultra Pure Water Kits 27
USP Kit 55
W
Water Hardness ISE
Electrode 31
Module 35
Solutions 31, 35
Thermo Scientific Orion Service Program Options
Our leadership shows up in everything we do including:
A letter of introduction from
Pekka Ronka, Vice President and General Manager
to our valued customers.
• Minimized ramp up time
• Ongoing support
• Flexible data reporting
and analysis
• Low cost of ownership
• Comprehensive support solutions for
all phases of the product life cycle
• Immediate access to experts worldwide
Dear Valued Customer,
I would like to take this opportunity to welcome you to the NEW 2009 Thermo Scientific
Orion Laboratory Products Catalog, and to introduce myself as the new Vice-President &
General Manager of Water Analysis Instruments.
For over 40 years Thermo Scientific Orion products have set the standard for innovation
and quality in electrochemistry. Today, Water Analysis Instruments continue to meet all
your electrochemistry needs with the highest quality products, service, and support.
Service Programs
Flexible service programs provide
fixed-price solutions, offering benefits
over the traditional time and materials,
per-quote based model. So you get
improved ROI, minimized downtime,
unmatched flexibility, better staff
utilization and increased productivity.
Titration Product Service Options
In this catalog you will find a comprehensive selection of meters, electrodes, solutions
and accessories for pH, ion selective electrodes (ISE), conductivity and dissolved
oxygen. We also feature innovative new products for water analysis, including:
•
Orion Star™ RDO® Meters, pages 50-51
Introducing rugged optical dissolved oxygen technology for maintenance-free,
better accuracy measurements.
•
Orion DUAL STAR™ Meters, pages 48-49
The next generation of Orion 720A/920A meters make it easier than ever to
simultaneously measure pH and ISE.
•
Colorimetry and Turbidity Meters, pages 74-85
Ideal for EPA monitoring and reporting of critical water analysis parameters.
•
Green Electrodes, page 18
Go green with environmentally-friendly electrodes that meet RoHS compliance
and contain no lead, mercury or hazardous materials.
•
ROSS® pH/ATC Triode™ Electrodes, page 8
Providing the convenience of built-in temperature with the performance and
accuracy of ROSS technology.
We are excited about the new products and welcome you to contact our technical
experts at (800) 225-1480 or fax (987) 232-6015 with your product questions.
Thank you for your continued support of Thermo Scientific Orion products. I am excited
to be a part of this world–class organization, and look forward to leading the company
into a future of innovation and excellence.
Sincerely,
Pekka Ronka
Vice President & General Manager
Water Analysis Instruments
A variety of service options are now
available for all titration products to
get you up and running quickly and
to make sure you are operating at
the highest performance possible.
Set-up services and training options
are available to minimize ramp up
time so results are expedited rapidly
and accurately. Avoid downtime by
choosing a preventative maintenance
program that meets your high
demands for quality and accurate
results, or select a two or three year
extended warranty to ensure you are
minimizing the possibility of downtime
due to an unforeseen failure.
Orion Star™ Series Service Options,
Calibration and Validation
All Thermo Scientific Orion Products In
Warranty Support
You’ll feel confident your results
are accurate and exact when taking
advantage of our calibration and
validation options. We have a range
of testing, validation and calibration
programs to satisfy even the most
exacting standard. Our dedicated
validation and calibration consultants
offer in-depth expertise and a
full understanding of compliance
requirements. It is recommended that
these options be performed annually,
based on your application demands.
Orion Star series and PerpHecT® meters
carry a three-year warranty that includes
return-to-factory, repair or replace
on a reasonable efforts basis, phone
support, web support, and any software
maintenance releases. Russell meters
include a 2 year warranty.
Stay in specification compliance by
simply sending your Orion Star series
meter to us, where calibration and
validation will be performed to your
choice of standards. We will return
your meter along with a certificate
of validation that includes calibration
details and test results.
IQ/OQ Package
Save time and money when you
purchase the IQ/OQ package to
ensure that your meter performs
to the manufacturer’s published
specifications. Full IQ/OQ
documentation is available for the
Orion Star series meters to perform
a comprehensive installation and
operation on your meter and satisfy
the requirements of regulated
industries such as pharmaceutical.
For product failures in the first thirty
days after shipment, we issue a product
replacement, no questions asked.
Go to thermo.com/water for
comprehensive warranty and service
option information.
Spare Parts
To help you keep your instruments
operating at the highest levels of
accuracy and efficiency, we offer factory
authorized parts at prices comparable to
those of generics. All parts, accessories,
and consumables are performancecertified and fully warranted to perform
to our products’ rigorous design
specifications.
Product Service Options
Calibration Options
Description
Cat. No.
2 Year Extended
Warranty
3 Year Extended
Warranty
Start-up
Preventative
Maintenance
Cat. No.
Star Series Calibration
and Validation
960 Titrator
096000
096000EXT2
096000EXT3
—
—
111200CAL
3-Star pH Benchtop
—
121200CAL
3-Star pH Portable
CBODEZ00PM
111300CAL
3-Star DO Benchtop
121300CAL
3-Star DO Portable
111400CAL
3-Star Cond Benchtop
121400CAL
3-Star Cond Portable
111500CAL
4-Star pH/ISE Benchtop
121500CAL
4-Star pH/ISE Portable
111600CAL
4-Star pH/DO Benchtop
121600CAL
4-Star pH/DO Portable
111700CAL
4-Star pH/Cond Benchtop
121700CAL
4-Star pH/Cond Portable
121800CAL
5-Star pH/DO/Cond Portable
111900CAL
5-Star pH/DO/Cond/ISE Benchtop
121900CAL
5-Star pH/DO/Cond/ISE Portable
STAR-IQOQ
Star Series IQ/OQ Package
960 Titrator
BOD AutoEZ
096002
10060020
096002EXT2
—
096002EXT3
—
—
CBODEZ00SU
Preventative Maintenance (PM) – contracts and Start-up/Training (SU) are available in N. America only and are represented by
Catalog Numbers ending in PM and SU.
Extended Warranties (EXT) – are available world wide and are represented by Catalog Numbers ending in EXT2 (2 year) and EXT3 (3 year).
ROSS and the COIL trade dress are trademarks of
Thermo Fisher Scientific Inc.
PerpHecT ROSS electrodes are protected by U.S. patent
6,168,707.
AQUAfast, Cahn, ionplus, KNIpHE, No Cal, ORION,
ORION Series A meters and 900A printer are protected
perpHect, PerpHecT, PerpHecTion, pHISA, pHuture, Pure
by U.S. patents 5,198,093, D334,208 and D346,753.
ROSS Ultra, Sure-Flow, Titrator PLUS and TURBO2 are
registered trademarks of Thermo Fisher.
1-888-pHAX-ION, A+, All-in-One, Aplus, AQUAsnap,
AssuredAccuracy, AUTO-BAR, AUTO-CAL, AUTO
DISPENSER, Auto-ID, AUTO-LOG, AUTO-READ, AUTOSTIR, Auto-Test, BOD AutoEZ, Cable-Free, CERTI-CAL,
CISA, DataCOLLECT, DataPLUS, digital LogR, DirectCal,
DuraProbe, Environmental Product Authority, Extra Easy/
Extra Value, FAST QC, GAP, GLPcal, GLPcheck, GLPdoc,
ISEasy, KAP, LabConnect, LogR, Low Maintenance
Triode, Minimum Stir Requirement, MSR, NISS, OneTouch, One-Touch Calibration, One-Touch Measurement,
Optimum Results, Orion Star, Pentrode, pHuture MMS,
pHuture Pentrode, pHuture Quatrode, pHuture Triode,
Quatrode, QuiKcheK, rf link, ROSS Resolution, SAOB,
SMART AVERAGING, Smart CheK, SMART STABILITY,
Stacked, Star Navigator 21, Stat Face, The Enhanced
Lab, ThermaSense, Triode, TRIUMpH, Unbreakable pH,
Universal Access are trademarks of Thermo Fisher.
Guaranteed Success and The Technical Edge are service
marks of Thermo Fisher.
PerpHecT meters are protected by U.S. patent
6,168,707.
ionplus electrodes and Optimum Results solutions are
protected by U.S. patent 5,830,338.
ROSS Ultra electrodes are protected by U.S. patent
6,793,787.
ORP standard is protected by U.S. patent 6,350,367.
No Cal electrodes with stabilized potential are protected
by U.S. patent 7,276,142.
© 2009 Thermo Fisher Scientific Inc. RDO is a
registered trademark of In-Situ Inc., Fort Collins, CO
USA. Microsoft and Windows are registered trademarks
of Microsoft Corporation in the United States and/or
other countries. All other trademarks are the property of
Thermo Fisher Scientific Inc. and its subsidiaries.
The specifications, descriptions, drawings, ordering
information and part numbers within this document are
subject to change without notice.
This publication supersedes all previous publications on
this subject.
2009 Thermo Scientific Orion Laboratory Products Catalog
Water, Sage, Sensing the Future, SensorLink, ROSS,
C-09ORION-E 0209 Rev. A
Thermo Scientific Orion
Laboratory Products Catalog
pH l ISE l Dissolved Oxygen l Conductivity l Colorimetry
www.thermo.com/water
Environmental Instruments
Water Analysis Instruments
9001:2008
North America
166 Cummings Center
Beverly, MA 01915 USA
Toll Free: 1-800-225-1480
Tel: 1-978-232-6000
Dom. Fax: 1-978-232-6015
Int’l Fax: 978-232-6031
Europe
P.O. Box 254, 3860 AG Nijkerk
Wallerstraat 125K, 3862 BN
Nijkerk, Netherlands
Tel: (31) 033-2463887
Fax: (31) 033-2460832
Asia Pacific
Blk 55, Ayer Rajah Crescent
#04-16/24, Singapore 139949
Tel: 65-6778-6876
Fax: 65-6773-0836
www.thermo.com/water
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project

Download PDF

advertising