Manual Supplement
Manual Supplement
Manual Title: T5-600/T5-1000 Service Information Supplement Issue: 2
Part Number: 686953
Issue Date:
11/12
Print Date:
March 1998
Page Count:
3
Revision/Date:1, 9/01
This supplement contains information necessary to ensure the
accuracy of the above manual.
© 2011-2012 Fluke Corporation. All rights reserved.


T5-600/T5-1000 Service Information
Manual Supplement
Change #1, 63860
On page 1, replace the Safety Information with the following:
 Warning
To avoid possible electric shock or personal injury, follow these guidelines:
 Read all safety Information before you use the Product.
 Carefully read all instructions.
 Do not use the Product if it is damaged or operating abnormally.
Protection may be impaired.
 Make sure the battery door is closed and latched.
 Inspect the Product and test leads. Look for cracks, missing plastic,
exposed metal, or damaged insulation. Replace damaged test leads
before using the tester.
 Verify Product operation by measuring a known voltage.
 Replace the batteries as soon as the low battery indicator () appears.
 Do not use the Product around explosive gas, vapor or dust.
 Do not apply more than the rated voltage, as marked on the tester,
between terminals or between any terminal and earth ground.
 Use caution when working above 30 V ac rms, 42 V ac peak, or 60 V dc.
 When using the probes, keep your fingers behind the finger guards on the
probes.
 Remove the batteries if the Product is not used for an extended period
of time, or if stored in temperatures above 50 °C. If the batteries are not
removed, battery leakage can damage the Product.
 Limit operation to the specified measurement category, voltage, or
amperage ratings.
 Use the correct terminals, function, and range for measurements.
 Do not measure current while the test leads are in the input jacks.
 Hold the Product behind the tactile barrier.
 Do not use and disable the Product if it is damaged.
 Do not use the Product if it operates incorrectly.
 Do not use test leads if they are damaged. Examine the test leads for
damaged insulation, exposed metal, or if the wear indicator shows.
Check test lead continuity.
 Use only current probes, test leads, and adapters supplied with the
Product.
 Remove all probes, test leads, and accessories before the battery door
is opened.
 Remove all probes, test leads, and accessories that are not necessary
for the measurement.
 Do not exceed the Measurement Category (CAT) rating of the lowest
rated individual component of a Product, probe, or accessory.
 Repair the Product before use if the battery leaks.
11/12
1
Manual Supplement
T5-600/T5-1000 Service Information
 Be sure that the battery polarity is correct to prevent battery leakage.
 Do not operate the Product with covers removed or the case open.
Hazardous voltage exposure is possible.
 Use only specified replacement parts.
 Have an approved technician repair the Product.
 Disconnect power and discharge all high-voltage capacitors before you
measure resistance, continuity, capacitance, or a diode junction.
On page 2 and 3, replace Serial Number 79560000 and Above General Specifciations
table, with:
Serial Numbers 79560000 and Above
General Specifications
Calibration
One-year calibration cycle.
Maximum Voltage
Between any Terminal
and Earth Ground
T5-600: 600 V rms, Measurement Category III, Pollution Degree 2
T5-1000: 1000 V rms, Measurement Category III, 600 V RMS, Category IV, Pollution Degree 2
Maximum Voltage
Between Current Fork
and Earth Ground
T5-600 and T5-1000: 1000 V rms, Catergory III, also 600 V rms, Category IV, Pollution Degree 2,
(This dual rating applies to the current fork only.)
Temperature
Operating: -10 C to +50 C (14 F to 122 F); Storage: -30 C to +60 C (-22 F to +140 F)
Altitude
Operating: 2000 m (6562 ft); Storage: 10,000 m (32808 ft)
Relative Humidity
0 % to 95 %, 5 C to 30 C (41 F to 86F); 0 % to 75 %, 30 C to 40 C (86 F to 104 F);
0 % to 45 %, 40 C to 50 C (104 F to 122 F)
Battery Type and Life
Two (2) AA; 360 hours continuous with alkaline; 125 hours continuous with zinc chloride
On page 3 and 4, replace the Serial Number 79559999 and Below General
Specifciations table, with:
Serial Numbers 79559999 and Below
General Specifications
Calibration
One-year calibration cycle.
Maximum Voltage
Between any Terminal
and Earth Ground
T5-600: 600 V rms, Measurement Category III
T5-1000: 1000 V rms, Measurement Category III
Temperature
Operating: -10 C to +50 C (14 F to 122 F); Storage: -30 C to +60 C (-22 F to +140 F)
Altitude
Operating: 2000 m (6562 ft); Storage: 10,000 m (32808 ft)
Relative Humidity
0 % to 95 %, 5 C to 30 C (41 F to 86F); 0 % to 75 %, 30 C to 40 C (86 F to 104 F);
0 % to 45 %, 40 C to 50 C (104 F to 122 F)
Battery Type and Life
Two (2) AA; 400 hours continuous with alkaline; 200 hours continuous with zinc chloride
Enclosure Rating
IP 52 per IEC 60529, no vacuum applied
RF Field Specification
0.5 % full scale + (specified accuracy) at 3 V/m
2
11/12
T5-600/T5-1000 Service Information
Manual Supplement
On page 4, following Input Characteristics, add:
CAT II Measurement Category II is applicable to test and measuring circuits
connected directly to utilization points (socket outlets and similar points) of the
low-voltage MAINS installation