Flexible 3000 Amp current probe

Flexible 3000 Amp current probe
Battery power, AC to DC adapter power
Flexible 3000 Amp current probe
Model : CP-3000-B
LUTRON ELECTRONIC
ISO-9001, CE, IEC1010
30 A/300 A/3000 A, 100 mV/A, 10 mV/A, 1 mV/A
For 3 phase power analyzer DW-6093, DW-6095
3000A FLEXIBLE CURRENT PROBE
Model : CP-3000
LUTRON ELECTRONIC
ISO-9001, CE, IEC1010
Flexible 3000 Amp
CURRENT PROBE
Model : CP-3000
FEATURES
* Range : 30 A/300 A/3000 A.
* Output sensitivity : 100 mV/A, 10 mV/A, 1 mV/A.
* Output voltage : 3 ACV ( 30 A, 300 A, 3000 A ).
* Flexible probe length : 600 mm ( 23.6 inch ).
* Optional current probe for DW-6093, DW-6095.
* When change the new probe, the calibration
procedures are not necessary to be executed.
* Signal output use the banana plugs, durable and safety.
* Power supply is from the meter (DW-6093, DW-6095 ),
battery is not necessary.
Operating
Temperature
Operating
Humidity
Dimension
0℃ to 50℃ ( 32℉ to 122℉ ).
Less than 80% RH.
Main circuit unit :
105 x 56.5 x 33 mm.
( 4.1 x 2.2 x 1.3 inch ).
Flexible probe :
600 mm x 14.5 mm DIA.
23.6 inch x 0.57 inch DIA.
Cable length :
SPECIFICATIONS
Measurement 30 A/300 A/3000 A, 3 ranges,
ranges
switch selectable
1st range value : 30 A
2nd range value : 300 A
3rd range value : 3000 A
Output
sensitivity
Accuracy
100 mV/A, 10 mV/A, 1 mV/A,
AC coupled.
± 1% of range value
3.4 meters ( 11.2 feet ).
DC plug size :
Weight
Accessory
Included
5.5 mm x 2.5 mm DIA.
480 g ( 1.06 LB ).
Operating manual .....................1 PC.
FRONT PANEL DESCRIPTIONS
* Accuracy is specified that the
measured conductor's position is
on the center of " Flexible current
probe ".
* 50/60 Hz..
* 23 ± 5 ℃
Frequency
range
Linearity
Phase error
3-1 Flexible current probe
3-2 Probe locker
45 to 400 Hz.
* Accuracy tested under 50/60 Hz.
± 0.2 % of reading value from 10 %
to 100% of range value.
< ± 1.5°
3-3 Main circuit unit
3-4 3000A/300A/30A
range switch
* 50/60 Hz..
Conductor
position
sensitivity
± 2 % of range value.
External
field
influence
increase ± 1.5 % of range value max..
Minimum
load
Enclosure
material
Safety
Power Supply
* Recommend the measured conductor
position should > 25 mm ( 1" )
away from the " Probe locker ".
* Recommend the distance between
different " Flexibel current
probe " should > 200 mm ( 8" ).
3-5 Cable
3-6 DCV power plug
3-7 Signal output plug
( +, red )
100 K ohm for specified accuracy.
3-8 Signal output plug
( -, black )
UL 94-V0 rated.
3-9 Power LED lamp
IEC 1010, Categories III 600V.
Power supply is from the meter
(DW-6093, DW-6095 ), battery is not
necessary.
* Power supply voltage from DW-6093,
DW-6095 is DC 8V± 1V..
* Appearance and specifications listed in this brochure are subject to change without notice.
1201-CP3000