0HI 98280 1GPS Multiparameter Meter with Fast Tracker™ Tag

0HI 98280 1GPS Multiparameter Meter with Fast Tracker™ Tag
0
1
HI 98280
GPS Multiparameter Meter with Fast Tracker™ Tag Identification System
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
Display up to 12 parameters
Track measurement locations with GPS
Waterproof protection for meter (IP67) and probe (IP68)
Fast Tracker™—Tag I.D. System simplifies test logging
Graphic LCD with backlight
Built-in barometer for DO compensation
Quick calibration feature
Measurement check eliminates erroneous readings
Auto recognition of pH and pH/ORP probe
Logger function records the data of all connected sensors
Log on demand and automatic logging (up to 60,000 samples)
Logged data can be displayed as graphs
USB for PC connectivity
Autorange of EC and TDS readings
Good Laboratory Practice feature with last 5 calibrations recorded
Field replaceable sensors
Meter accepts both alkaline and rechargeable batteries
Rugged probe with stainless steel tip has a diameter under 2” for wells and pipes
The HI 98280 multiparameter meter with GPS receiver monitors up to 13 different water quality parameters (6
measured, 7 calculated). Measurements include pH, pH/mV, ORP, % saturation DO, mg/L DO, EC, absolute EC,
resistivity, TDS, salinity, seawater specific gravity, atmospheric pressure and temperature.
Measurements from specific locations are tracked with detailed coordinate information that can be viewed
immediately on the display. GPS information can be transferred to a PC using HANNA’s
HI 929828 software. GPS information can also be viewed by GPS mapping software such as Google™ maps*.
Clicking on visited locations using mapping software such as Google™ Maps displays measurement information.
The built-in 12 channel GPS receiver and antenna guarantees a position accuracy of 30 ft (10 m).
For measuring points within a 30 ft range or where GPS signals are not available, HI 98280’s Fast Tracker™ is
invaluable for associating measurements with their locations. HANNA’s exclusive Fast Tracker™—Tag I.D. System
monitors and records data using iButton®s that can be installed at any number of sampling sites.
The HI 98280 has a graphic, backlit LCD that automatically sizes the digits to fit the screen and allows full
configuration of each parameter measured along with on-screen graphing capability. The meter incorporates
comprehensive GLP features and the downloading of data via USB connection.
Each parameter is supported by on-screen context sensitive help, both in the calibration mode and during
measurement.
Designed for outdoor environments, the meter is impact resistant and waterproof meeting IP67 standards (30
minutes immersion under 1 m of water). The multi-sensor probe can be left underwater in accordance with IP68
standards.
HI 769828 Multiparameter, Intelligent Probe
HANNA’s HI 769828 multiparameter probe incorporates a built-in microprocessor and amplifier that convert the
high impedance signals from the sensors of the probe to eliminate common problems such as cable length
limitations and noise associated with high impedance signals. This allows the probe to have a reliable
communication with the meter and also immediately warns the user of problems such as a broken cable. The
standard cable lengths of the probe are 4, 10 and 20 meters (13, 32 and 64 feet) and custom lengths are also
available.
The probe also features HANNA’s Quick Calibration which allows the user to calibrate pH and conductivity with
one solution in a single, simple step. Dissolved oxygen is also calibrated in one step in saturated air. The probe
houses 5 of the 6 measured parameters: pH, ORP, EC, dissolved oxygen and temperature. The sensors are all
independently replaceable and are easy to maintain and keep clean. The sensors are protected by an outer
PVC/stainless steel sleeve and cap which is suitable for use in 2” wells. The probe housing complies with IP68
standard.
GPS (Global Positioning System) Enabled
The new HI 98280 features an internal 12 channel GPS receiver and antenna that calculates its position to track
locations along with measurement data. The GPS tracks your location using satellites to within 30 ft (10 m).
GPS coordinates are shown on the LCD with up to 10 measurement parameters. The LCD also displays the GPS
signal strength, number of satellites and logged data embedded with GPS coordinates.
Users can associate GPS coordinates with alphanumeric locations and distances between current location and
predefined locations are displayed arranged by distance. In the case of GPS signal loss, the HI 98280 memorizes
the last location and time.
HI 929828 PC Software manages logged data from the HI 98280 and displays GPS coordinates along with logged
data.
Track Measurement Locations
Users can connect to GPS tracking software such as Google™ Maps* to view locations where samples have been
taken. Measurement information is shown right on the map.
Fast Tracker™—Tag Identification System
HANNA’s Fast Tracker™—Tag Identification System simplifies test logging. iButton®s with a unique ID can be
installed at various sampling sites. When the matching connector on the meter contacts the location button,
measurements are logged and labeled with the alphanumeric user entered location ID. Location, date, time and
measurements are logged into the meter which can be transferred to a PC. The Fast Tracker™ system
complements the GPS for ultimate tracking.
Choose your configuration HI 982804-01 HI 9828 with GPS and 4 m (13.1’) probe cable, 115VAC
HI 982804-02 HI 9828 with GPS and 4 m (13.1’) probe cable, 230VAC
HI 9828010-01 HI 9828 with GPS and 10 m (33’) probe cable, 115VAC
HI 9828010-02 HI 9828 with GPS and 10 m (33’) probe cable, 230VAC
HI 9828020-01 HI 9828 with GPS and 20 m (66’) probe cable, 115VAC
HI 9828020-02 HI 9828 with GPS and 20 m (66’) probe cable, 230VAC
HI 9828030-01 HI 9828 with GPS and 30 m (98’) probe cable, 115VAC
HI 9828030-02 HI 9828 with GPS and 30 m (98’) probe cable, 230VAC
HI 9828040-01 HI 9828 with GPS and 40 m (131’) probe cable, 115VAC
HI 9828040-02 HI 9828 with GPS and 40 m (131’) probe cable, 230VAC
HI 9828050-01 HI 9828 with GPS and 50 m (164’) probe cable, 115VAC
HI 9828050-02 HI 9828 with GPS and 50 m (164’) probe cable, 230VAC
HI 9828060-01 HI 9828 with GPS and 60 m (200’) probe cable, 115VAC
HI 9828060-02 HI 9828 with GPS and 60 m (200’) probe cable, 230VAC
HI 9828070-01 HI 9828 with GPS and 70 m (230’) probe cable, 115VAC
HI 9828070-02 HI 9828 with GPS and 70 m (230’) probe cable, 230VAC
HI 9828080-01
HI 9828080-02
HI 9828090-01
HI 9828090-02
HI
HI
HI
HI
9828
9828
9828
9828
with
with
with
with
GPS
GPS
GPS
GPS
and
and
and
and
80
80
90
90
m
m
m
m
(262’)
(262’)
(295’)
(295’)
probe
probe
probe
probe
cable,
cable,
cable,
cable,
115VAC
230VAC
115VAC
230VAC
Order Information:
HI 9828 configurations are supplied with HI 769828 multisensor probe (pH/ORP, EC, DO, temperature), HI 982825 quick calibration standard solution (500 mL), probe maintenance kit, rechargeable C size Ni-MH batteries (4),
power adapter & cable, car 12V accessory outlet adapter, HI 7698281 USB interface cable, HI 92828 Windows®
compatible software and instruction manual in a rugged carrying case.
x
Specifications
x
x
Accessories
Downloads
pH
mV of pH
Input
ORP
Dissolved
Oxygen
Range
0.00 to 14.00 pH
Resolution
0.01 pH
Accuracy
±0.02 pH
Range
±600.0 mV
Resolution
0.1 mV
Accuracy
±0.5 mV
Range
±2000.0 mV
Resolution
0.1 mV
Accuracy
±1.0 mV
Range
0.0 to 500.0% / 0.00 to 50.00 mg/L
Resolution
0.1% / 0.01 mg/L
0.0 to 300.0%: ±1.5% of reading or ±1.0% whichever is greater; 300.0 to 500.0%: ±3% of
Accuracy
reading; 0.00 to 30.00 mg/L: ±1.5% of reading or 0.10 mg/L whichever is greater; 30.00 mg/L to
50.00 mg/L: ±3% of reading
Conductivity
Range
0.000 to 200.000 mS/cm (absolute EC up to 400 mS/cm)
manual: 1 μS/cm; 0.001 mS/cm; 0.01 mS/cm; 0.1 mS/cm; 1 mS/cm; automatic: 1 μS/cm from 0
Resolution
to 9999 μS/cm; 0.01 mS/cm from 10.00 to 99.99 mS/cm; 0.1 mS/cm from 100.0 to 400.0 mS/cm;
automatic mS/cm: 0.001 mS/cm from 0.000 to 9.999 mS/cm; 0.01 mS/cm from 10.00 to 99.99
mS/cm; 0.1 mS/cm from 100.0 to 400.0 mS/cm
Resistivity
Accuracy
±1% of reading or ±1 μS/cm whichever is greater
Range
0 to 999999 ƻ•cm; 0 to 1000.0 kƻ•cm; 0 to 1.0000 Mƻ•cm
TDS
Resolution
dependent on resistivity reading
Range
0 to 400000 mg/L or ppm (the maximum value depends on the TDS factor)
manual: 1 mg/L (ppm); 0.001 g/L (ppt); 0.01g/L (ppt); 0.1 g/L (ppt); 1 g/L (ppt); autorange
scales: 1 mg/L (ppm) from 0 to 9999 mg/L (ppm); 0.01 g/L (ppt) from 10.00 to 99.99 g/L (ppt);
Resolution
0.1 g/L (ppt) from 100.0 to 400.0 g/L (ppt); autorange g/L (ppt) scales: 0.001 g/L (ppt) from
0.000 to 9.999 g/L (ppt); 0.01 g/L (ppt) from 10.00 to 99.99 g/L (ppt); 0.1 g/L (ppt) from 100.0
to 400.0 g/L (ppt)
Salinity
Seawater
Specific
Accuracy
±1% of reading or ±1 mg/L (ppm) whichever is greater
Range
0.00 to 70.00 PSU (Extended Practical Salinity Scale)
Resolution
0.01 PSU
Accuracy
±2% of reading or 0.01 PSU whichever is greater
Range
0.0 to 50.0 ıt, ı0, ı15
Resolution
0.1 ıt, ı0, ı15
Accuracy
±1 ıt, ı0, ı15
Gravity
Atm. Pressure Range
Temperature
Calibration
450 to 850 mmHg; 17.72 to 33.46 inHg; 600.0 to 1133.2 mbar; 8.702 to 16.436 psi; 0.5921 to
1.1184 atm; 60.00 to 113.32 kPa
Resolution
0.1 mmHg; 0.01 inHg; 0.1 mbar; 0.001 psi; 0.0001 atm; 0.01 kPa
Accuracy
±3 mmHg within ±15°C from the calibration temperature
Range
-5.00 to 55.00°C; 23.00 to 131.00°F; 268.15 to 328.15K
Resolution
0.01°C; 0.01°F; 0.01K
Accuracy
±0.15°C; ±0.27°F; ±0.15K
pH
ORP
Conductivity
DO
Resistivity,
TDS, ı
Salinity
automatic one, two, or three points with five memorized standard buffers (pH 4.01, 6.86, 7.01,
9.18, 10.01) or one custom buffer
automatic at one custom point
automatic one point with six memorized standards (84 μS/cm, 1413 μS/cm, 5.00 mS/cm, 12.88
mS/cm, 80.0 mS/cm, 111.8 mS/cm) or custom point
automatic one or two points at 0, 100% or one custom point
based on conductivity or salinity calibration
one custom point
Atm.
Pressure,
automatic at one custom point
Temperature
Temperature
Compensation
Logging
Memory
automatic from -5 to 55°C (23 to 131°F)
up to 60,000 samples with 13 measurements each; Up to 45,000 samples with 15 measurements
each (GPS)
Logging
Interval
1 second to 3 hours
Computer
Interface
USB (with HI 929829 software)
Waterproof
Protection
meter IP67, probe IP68
Environment
0 to 50°C (32 to 122°F); RH 100%
Power Supply
1.5V alkaline C cells (approximately 150 hours of continuous use without backlight) (4) / 1.2V
rechargeable C cells (approximately 70 hours of continuous use without backlight) (4)
Dimensions
Meter/Probe
221 x 115 x 55 mm (8.7 x 4.5 x 2.2”) / 270 x 46 mm DIA (10.6 x 1.8” DIA)
Weight
meter: 750 g (26.5 oz.); probe: 750 g (26.5 oz.)