GL800表english_end
Multi-channel measurement
anytime, anywhere
Multifunction input; the number of channels can be expanded from the standard 20 to the maximum 200
Isolated input for all channels; vibration-resistant; outstanding, worry-free measurement performance
Direct data capture to a USB memory
Equipped with a bright, easy-to-read large-format TFT LCD
Long-term data capture for nine hours enabled when the battery is used
Main Unit Specifications
Input Terminal Unit Specifications
Item
Number of analog input terminal units
Item
Number of input channels
Type of input terminal
Method
Description
1unit (20 ch) or Extension unit (max. 200 ch)
Maximum 10 units or 200 ch when using with a PC
Sampling interval *1
100 ms (10 ch) to 1 h
Trigger function
Type
Start (Data capture starts when a trigger is generated)
Stop
(Data capture stops when a trigger is generated)
Condition Start Off, Level, Alarm, Scheduled time, External
Stop
Off, Level, Alarm, Scheduled time, Elasped time, External
Alarm function
Type
Analog, Logic, Pulse (AND and OR operation available)
Condition Analog H, L, Window In, Window Out
Logic H, L at each channel
Pulse H, L, Window In, Window Out
Pulse / Logic input *2 *5
Selectable between Pulse and Logic, Number of channels : 4
Pulse input range Count mode
50 C, 500 C, 5 kC, 50 kC, 500 kC, 5 MC, 50 MC, 500 MC/F.S.
(max. input 50 kHz)
Inst. Mode
50 C, 500 C, 5 kC, 50 kC, 500 kC, 5 MC, 50 MC, 500 MC/F.S.
(max. input 50 kHz)
RPM mode
50 rpm, 500 rpm, 5 krpm, 50 krpm, 500 krpm, 5 Mrpm,
50 Mrpm, 500 Mrpm/F.S. (max. input 50 kHz)
Alarm output *5 No. of channels
4
Output format
Open collector output (5V pull-up resistance 10 k ohm),
5 to 24V (100 mA or less)
Output conditions
Level judgement, Window judgement, Logic pattern judgement,
Pulse judgement
External trigger input *2 *5
1 ch
Interface to PC
Ethernet (10BASE-T / 100BASE-TX),
USB (Compatible with high speed)
Data storage
Measured data
Internal flash memory or USB memory directly
function
Other
Setting conditions and Screen hard copy can be saved into the
internal memory or the USB memory
Internal storage device
Internal flash memory : 12 Mbyte
USB memory slot (Full speed)
Provided as standard
Calculation
Statistics calculation
Average, Peak, maximum, Minimum, RMS
function
Number of calculations 2 calculations can be set simultaneously
Searching function
Searching the necessary point from captured data. Type : Level,
Alarm, Logic, Pulse
Display
Size
5.7 inch TFT color LCD
Displayed items
Waveforms + Digital values, Waveforms only, Digital values only
Operating environment
Temperature : 0 to 45 °C (15 to 40 ∞C when operating by
battery), Humidity : 5 to 85 % R.H.
Power supply
AC adapter : 100 to 240V AC, 50/60 Hz
DC input : 8 to 24 V DC *3
Battery pack : Max. 2 batteries installable, 9 hours operation
(when using under Graphtec specified condition) *3
Power consumption
28 VA or lower (when operating with AC power)
External dimensions (W x D x H) (approx.) 232 mm x 152 mm x 50 mm
Weight (approx.)
990 g including a 20 channel input terminal unit *4
Vibration resistance
Compatible with JIS Vibration testing methods for automobile
Type 1 Class A-equivalent
Certifications
CE, RoHS, China RoHS
Description
20 (maximum 200 channels by the expansion terminal unit)
M3 screw type terminal
Scan
All channels isolated, balanced input
Measurement Voltage
20, 50, 100, 200, 500 mV, 1, 2, 5, 10, 20, 50, V, 1-5 V / F.S.
range
Temperature Thermocouple : K, J, E, T, R, S, B, N, W (WRe5-26)
PT : Pt100, JPt100, Pt1000 (IEC751)
Humidity
0 to 100 % RH (Voltage 0 to 1 V conversion, when using optional
B-530 humidity sensor is used)
Input filter
Off, 2, 5, 10, 20, 40 (Moving average)
Measurement Voltage
+/-0.1 % of F.S.
accuracy
Thermocouple Measurement range
Accuracy
± 5.2 °C
R/S
0 ≤ TS ≤ 100 °C
100 < TS ≤ 300 °C
± 3.0 °C
R : 300 < TS ≤ 1600 °C
± (0.05 % of rdg + 2.0 °C)
± (0.05 % of rdg + 2.0 °C)
S : 300 < TS ≤ 1760 °C
B
400 ≤ TS ≤ 600 °C
± 3.5 °C
600 < TS ≤ 1820 °C
± (0.05 % of rdg + 2.0 °C)
K
-200 ≤ TS ≤ -100 °C
± (0.05 % of rdg + 2.0 °C)
-100 < TS ≤ 1370 °C
± (0.05 % of rdg + 1.0 °C)
± (0.05 % of rdg + 2.0 °C)
E
-200 ≤ TS ≤ -100 °C
-100 < TS ≤ 800 °C
± (0.05 % of rdg + 1.0 °C)
T
-200 ≤ TS ≤ -100 °C
± (0.1 % of rdg + 1.5 °C)
± (0.1 % of rdg + 0.5 °C)
-100 < TS ≤ 400 °C
J
-200 ≤ TS ≤ -100 °C
± 2.7 °C
± 1.7 °C
-100 < TS ≤ 100 °C
100 < TS ≤ 1100 °C
± (0.05 % of rdg + 1.0 °C)
N
0 ≤ TS ≤ 1300 °C
± (0.1 % of rdg + 1.0 °C)
≤
≤
°C
± (0.1 % of rdg + 1.5 °C)
W
0 TS 2000
Pt100 -200 to 850 °C (FS = 1050 °C) ± 1.0 °C
PT
JPt100 -200 to 500 °C (FS = 700 °C) ± 0.8 °C
Pt1000 -200 to 500 °C (FS = 700 °C) ± 0.8 °C
* If the reference junction compensation is internal, add ± 0.5 °C to each of the above values.
A/D converter
16 bit (out of which 14 are internally acknowledged)
Maximum input voltage
60 Vp-p (Between ±)
60 Vp-p (Between input terminals)
60 Vp-p (Between input terminal and chassis)
Withstand voltage
350 Vp-p (between input terminal and GND) 1 minute
*1
*2
*3
*4
*5
The available number of channels varies according to the sampling interval.
Max. input voltage : 24 V, Input threshold voltage : approx. 2.5 V, Hysteresis approx. 0.5 V (+2.5 to 3 V)
Option
Excluding the battery and AC adapter
Logic/Alarm cable (B-513) is required
GL800 PC software specifications
Item
Compatible OS
Function
Main unit settings
Maximum number of controlable units
Captured data
Display
Display mode
Monitor functions
File conversion
Direct Excel function
Statistcs calculation, report
Reprot function
Description
Windows 2000/XP/Vista
Control of main unit, Realtime data capture, File conversion
Input settings, memory settings, alarm settings, trigger
settings, report settings, others
Either 10 units or 500 channles
Realtime data, Internal memory data, USB menory data
Analog waveforms, logic waveforms, pulse waveforms,
digital values
Y-T waveform view, X-Y waveform view, Digital view, Meter
view, Report view
Alarm monitor enables sending of email to the specified
address
Data conversion between cursors/All data to CSV file
format, batch conversion function
Max. 1 sec sampling data can be saved in to Excel file.
Maximum, minimum, avarage data during measurement
Automatic creation of daily or monthly files
Standard Accessories
Item
Quick Start Guide
CD-ROM
AC Adapter
Description
User’s manual, Application software
100 to 240 V AC 50 / 60 Hz
with specified type of power cord
Options
Item
Extension terminal base kit
20 ch extension terminal set
Battery pack
Humidity sensor
Logic / Alarm cable
DC power cable
Model
B-537
B-538
B-517
B-530
B-513
B-514
Q’Ty
1
1
1
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project

Download PDF