Addendum To Precision Medical Flowmeters User Manual
To Whom It May Concern,
This letter is to supplement cleaning instructions for Precision Medical Air and Oxygen Flowmeters whether
as single or with PTO to include flowmeters installed on wye blocks or tee branches. Precision Medical Air
and Oxygen Flowmeters are not single patient/disposable items. This includes:
1MFA series
3MFA series
4MFA series
6MFA series
7MFA series
8MFA series
19MFA series
Flowmeters can be externally cleaned/disinfected using hospital approved wipes with quaternary cleaning,
sodium hypochlorite, or hydrogen peroxide.
Alcohol based wipes should not be used to clean/disinfect oxygen flow meters and accessories due to the
risk of fire.
We recommend following the directions of the manufacturer of the chosen cleaner.
The flowmeters should be wiped down between patient use.
Precision Medical, Inc.