TSQ Quantum XLS and TSQ Quantum GC

TSQ Quantum XLS and TSQ Quantum GC
TSQ Series
TSQ Quantum XLS
and TSQ Quantum GC
User Guide
70111-97220 Revision A
March 2010
© 2009 Thermo Fisher Scientific Inc. All rights reserved.
Swagelok™ is a trademark of the Crawford Fitting Company. Dranetz™ is a trademark of Dranetz
Technologies, Inc. Styrofoam™ is a trademark of Dow Chemical Company. Teflon™ is a trademark of E. I. du
Pont de Nemours & Co. Tygon™ is a trademark of Norton Company. Dust-Off™is a trademark of Falcon™
Safety Products. Micro-Blast™ is a trademark of MicroCare™ Corporation. Restek™ is a trademark of Restek
Corporation. Dremel, Vespel
Thermo Fisher Scientific Inc. provides this document to its customers with a product purchase to use in the
product operation. This document is copyright protected and any reproduction of the whole or any part of this
document is strictly prohibited, except with the written authorization of Thermo Fisher Scientific Inc.
The contents of this document are subject to change without notice. All technical information in this
document is for reference purposes only. System configurations and specifications in this document supersede
all previous information received by the purchaser.
Thermo Fisher Scientific Inc. makes no representations that this document is complete, accurate or errorfree and assumes no responsibility and will not be liable for any errors, omissions, damage or loss that might
result from any use of this document, even if the information in the document is followed properly.
This document is not part of any sales contract between Thermo Fisher Scientific Inc. and a purchaser. This
document shall in no way govern or modify any Terms and Conditions of Sale, which Terms and Conditions of
Sale shall govern all conflicting information between the two documents.
Release history: Revision A March 2010
Software version: Xcalibur 2.1, TSQ 2.3
For Research Use Only. Not for use in diagnostic procedures.
Regulatory Compliance
Thermo Fisher Scientific performs complete testing and evaluation of its products to ensure full compliance with
applicable domestic and international regulations. When the system is delivered to you, it meets all pertinent
electromagnetic compatibility (EMC) and safety standards as described in the next section or sections by product name.
Changes that you make to your system may void compliance with one or more of these EMC and safety standards.
Changes to your system include replacing a part or adding components, options, or peripherals not specifically
authorized and qualified by Thermo Fisher Scientific. To ensure continued compliance with EMC and safety standards,
replacement parts and additional components, options, and peripherals must be ordered from Thermo Fisher Scientific
or one of its authorized representatives.
TSQ Quantum XLS (not reviewed)
EMC Directive 89/336/EEC as amended by 92/31/EEC and 93/68/EEC
EMC compliance has been evaluated by TUV Rheinland of North America Inc.
EN 55011
1998, 1999, 2002
EN 61000-4-3
2002
EN 61000-3-2
1995, A1; 1998, A2; 1998, A14; 2000
EN 61000-4-4
1995, A1; 2000, A2; 2001
EN 61000-3-3
1998, 2001
EN 61000-4-5
1995, A1; 2001
EN 61326-1
1998, 2001, 2003
EN 61000-4-6
1996, A1; 2003
EN 61000-4-2
2001
EN 61000-4-11
1994, A1; 2001
CISPR 11
1998
FCC Class A, CFR 47 Part 15
TSQ Quantum GC
EMC Directive 89/336/EEC as amended by 92/31/EEC and 93/68/EEC
EMC compliance has been evaluated by TUV Rheinland of North America Inc.
EN 55011
1998, 1999, 2002
EN 61000-4-3
2002
EN 61000-3-2
1995, A1; 1998, A2; 1998, A14; 2000
EN 61000-4-4
1995, A1; 2000, A2; 2001
EN 61000-3-3
1998, 2001
EN 61000-4-5
1995, A1; 2001
EN 61326-1
1998, 2001, 2003
EN 61000-4-6
1996, A1; 2003
EN 61000-4-2
2001
EN 61000-4-11
1994, A1; 2001
CISPR 11
1998
FCC Class A, CFR 47 Part 15
Low Voltage Safety Compliance
Compliance with safety issues is declared under Thermo Fisher Scientific sole responsibility.
This device complies with Low Voltage Directive 73/23/EEC and harmonized standard EN 61010-1:2001.
Changes that you make to your system may void compliance with one or more of these EMC and safety standards.
Changes to your system include replacing a part or adding components, options, or peripherals not specifically
authorized and qualified by Thermo Fisher Scientific. To ensure continued compliance with EMC and safety standards,
replacement parts and additional components, options, and peripherals must be ordered from Thermo Fisher Scientific
or one of its authorized representatives.
FCC Compliance Statement
THIS DEVICE COMPLIES WITH PART 15 OF THE FCC RULES. OPERATION IS SUBJECT TO
THE FOLLOWING TWO CONDITIONS: (1) THIS DEVICE MAY NOT CAUSE HARMFUL
INTErfERENCE, AND (2) THIS DEVICE MUST ACCEPT ANY INTErfERENCE RECEIVED,
INCLUDING INTErfERENCE THAT MAY CAUSE UNDESIRED OPERATION.
CAUTION Read and understand the various precautionary notes, signs, and symbols contained inside
this manual pertaining to the safe use and operation of this product before using the device.
Notice on Lifting and Handling of
Thermo Fisher Scientific San Jose Instruments
For your safety, and in compliance with international regulations, the physical handling of this Thermo Fisher
Scientific San Jose instrument requires a team effort to lift and/or move the instrument. This instrument is too heavy
and/or bulky for one person alone to handle safely.
Notice on the Proper Use of
Thermo Fisher Scientific San Jose Instruments
In compliance with international regulations: Use of this instrument in a manner not specified by Thermo Fisher
Scientific San Jose could impair any protection provided by the instrument.
Notice on the Susceptibility
to Electromagnetic Transmissions
Your instrument is designed to work in a controlled electromagnetic environment. Do not use radio frequency
transmitters, such as mobile phones, in close proximity to the instrument.
WEEE Compliance
This product is required to comply with the European Union’s Waste Electrical & Electronic Equipment (WEEE)
Directive 2002/96/EC. It is marked with the following symbol:
Thermo Fisher Scientific has contracted with one or more recycling or disposal companies in each European Union
(EU) Member State, and these companies should dispose of or recycle this product. See www.thermo.com/
WEEERoHS for further information on Thermo Fisher Scientific’s compliance with these Directives and the
recyclers in your country.
WEEE Konformität
Dieses Produkt muss die EU Waste Electrical & Electronic Equipment (WEEE) Richtlinie 2002/96/EC erfüllen.
Das Produkt ist durch folgendes Symbol gekennzeichnet:
Thermo Fisher Scientific hat Vereinbarungen mit Verwertungs-/Entsorgungsfirmen in allen EU-Mitgliedsstaaten
getroffen, damit dieses Produkt durch diese Firmen wiederverwertet oder entsorgt werden kann. Mehr Information
über die Einhaltung dieser Anweisungen durch Thermo Fisher Scientific, über die Verwerter, und weitere Hinweise,
die nützlich sind, um die Produkte zu identifizieren, die unter diese RoHS Anweisung fallen, finden sie unter
www.thermo.com/WEEERoHS.
Conformité DEEE
Ce produit doit être conforme à la directive européenne (2002/96/EC) des Déchets d'Equipements Electriques et
Electroniques (DEEE). Il est marqué par le symbole suivant:
Thermo Fisher Scientific s'est associé avec une ou plusieurs compagnies de recyclage dans chaque état membre de
l’union européenne et ce produit devrait être collecté ou recyclé par celles-ci. Davantage d'informations sur la
conformité de Thermo Fisher Scientific à ces directives, les recycleurs dans votre pays et les informations sur les
produits Thermo Fisher Scientific qui peuvent aider la détection des substances sujettes à la directive RoHS sont
disponibles sur www.thermo.com/WEEERoHS.
CAUTION Symbol
CAUTION
VORSICHT
ATTENTION
PRECAUCION
AVVERTENZA
Electric Shock: This instrument uses
high voltages that can cause personal
injury. Before servicing, shut down the
instrument and disconnect the instrument
from line power. Keep the top cover on
while operating the instrument. Do not
remove protective covers from PCBs.
Elektroschock: In diesem Gerät werden
Hochspannungen verwendet, die
Verletzungen verursachen können. Vor
Wartungsarbeiten muß das Gerät
abgeschaltet und vom Netz getrennt
werden. Betreiben Sie Wartungsarbeiten
nicht mit abgenommenem Deckel. Nehmen
Sie die Schutzabdeckung von Leiterplatten
nicht ab.
Choc électrique: L’instrument utilise des
tensions capables d’infliger des blessures
corprelles. L’instrument doit être arrêté et
débranché de la source de courant avant
tout intervention. Ne pas utiliser
l’instrument sans son couvercle. Ne pas
elensver les étuis protecteurs des cartes de
circuits imprimés.
Descarga eléctrica: Este instrumento
utiliza altas tensiones, capaces de
producir lesiones personales. Antes de
dar servicio de mantenimiento al
instrumento, éste debera apagarse y
desconectarse de la línea de alimentacion
eléctrica. No opere el instrumento sin sus
cubiertas exteriores quitadas. No remueva
las cubiertas protectoras de las tarjetas
de circuito impreso.
Shock da folgorazione. L’apparecchio è
alimentato da corrente ad alta tensione
che puo provocare lesioni fisiche. Prima di
effettuare qualsiasi intervento di
manutenzione occorre spegnere ed isolare
l’apparecchio dalla linea elettrica. Non
attivare lo strumento senza lo schermo
superiore. Non togliere i coperchi a
protezione dalle schede di circuito
stampato (PCB).
Chemical: This instrument might contain
hazardous chemicals. Wear gloves when
handling toxic, carcinogenic, mutagenic,
or corrosive or irritant chemicals. Use
approved containers and proper
procedures to dispose waste oil.
Chemikalien: Dieses Gerät kann
gefährliche Chemikalien enthalten. Tragen
Sie Schutzhandschuhe beim Umgang mit
toxischen, karzinogenen, mutagenen oder
ätzenden/reizenden Chemikalien.
Entsorgen Sie verbrauchtes Öl
entsprechend den Vorschriften in den
vorgeschriebenen Behältern.
Chimique: Des produits chemiques
dangereux peuven se trouver dans
l’instrument. Proted dos gants pour
manipuler tous produits chemiques
toxiques, cancérigènes, mutagènes, ou
corrosifs/irritants. Utiliser des récipients
et des procédures homologuées pour se
débarrasser des déchets d’huile.
Química: El instrumento puede contener
productos quimicos peligrosos. Utilice
guantes al manejar productos quimicos
tóxicos, carcinogenos, mutagenos o
corrosivos/irritantes. Utilice recipientes y
procedimientos aprobados para
deshacerse del aceite usado.
Prodotti chimici. Possibile presenza di
sostanze chimiche pericolose
nell’apparecchio. Indossare dei guanti per
maneggiare prodotti chimici tossici,
cancerogeni, mutageni, o
corrosivi/irritanti. Utilizzare contenitori
aprovo e seguire la procedura indicata per
lo smaltimento dei residui di olio.
Heat: Before servicing the instrument,
allow any heated components to cool.
Hitze: Warten Sie erhitzte Komponenten
erst nachdem diese sich abgekühlt haben.
Haute Temperature: Permettre aux
composants chauffés de refroidir avant
tout intervention.
Altas temperaturas: Permita que lop
componentes se enfríen, ante de efectuar
servicio de mantenimiento.
Calore. Attendere che i componenti
riscaldati si raffreddino prima di
effetturare l’intervento di manutenzione.
Fire: Use care when operating the system
in the presence of flammable gases.
Feuer: Beachten Sie die einschlägigen
VorsichtsmaBnahmen, wenn Sie das
System in Gegenwart von entzündbaren
Gasen betreiben.
Incendie: Agir avec précaution lors de
l’utilisation du système en présence de
gaz inflammables.
Fuego: Tenga cuidado al operar el
sistema en presencia de gases
inflamables.
Incendio. Adottare le dovute precauzioni
quando si usa il sistema in presenza di gas
infiammabili.
Eye Hazard: Eye damage could occur
from splattered chemicals or flying
particles. Wear safety glasses when
handling chemicals or servicing the
instrument.
Verletzungsgefahr der Augen:
Verspritzte Chemikalien oder kleine
Partikel können Augenverletzungen
verursachen. Tragen Sie beim Umgang mit
Chemikalien oder bei der Wartung des
Gerätes eine Schutzbrille.
Danger pour les yeux: Dex projections
chimiques, liquides, ou solides peuvent
être dangereuses pour les yeux. Porter des
lunettes de protection lors de toute
manipulationde produit chimique ou pour
toute intervention sur l’instrument.
Peligro par los ojos: Las salicaduras de
productos químicos o particulas que
salten bruscamente pueden causar
lesiones en los ojos. Utilice anteojos
protectores al mnipular productos
químicos o al darle servicio de
mantenimiento al instrumento.
Pericolo per la vista. Gli schizzi di
prodotti chimici o delle particelle presenti
nell’aria potrebbero causare danni alla
vista. Indossare occhiali protettivi quando
si maneggiano prodotti chimici o si
effettuano interventi di manutenzione
sull’apparecchio.
General Hazard: A hazard is present that
is not included in the above categories.
Also, this symbol appears on the
instrument to refer the user to instructions
in this manual.
Allgemeine Gefahr: Es besteht eine
weitere Gefahr, die nicht in den
vorstehenden Kategorien beschrieben ist.
Dieses Symbol wird im Handbuch
auBerdem dazu verwendet, um den
Benutzer auf Anweisungen hinzuweisen.
Danger général: Indique la présence
d;un risque n’appartenant pas aux
catégories citées plus haut. Ce symbole
figure également sur l’instrument pour
renvoyer l’utilisateur aux instructions du
présent manuel.
Peligro general: Significa que existe un
peligro no incluido en las categorias
anteriores. Este simbolo también se utiliza
en el instrumento par referir al usuario a
las instrucciones contenidas en este
manual.
Pericolo generico. Pericolo non
compreso tra le precedenti categorie.
Questo simbolo è utilizzato inoltre
sull’apparecchio per segnalare all’utente
di consultare le istruzioni descritte nel
presente manuale.
When the safety of a procedure is
questionable, contact your local Technical
Support organization for Thermo Fisher
Scientific San Jose Products.
Wenn Sie sich über die Sicherheit eines
Verfahrens im unklaren sind, setzen Sie
sich, bevor Sie fortfahren, mit Ihrer
lokalen technischen
Unterstützungsorganisation für Thermo
Fisher Scientific San Jose Produkte in
Verbindung.
Si la sûreté d’un procédure est incertaine,
avant de continuer, contacter le plus
proche Service Clientèle pour les produits
de Thermo Fisher Scientific San Jose.
Cuando la certidumbre acerca de un
procedimiento sea dudosa, antes de
proseguir, pongase en contacto con la
Oficina de Asistencia Tecnica local para
los productos de Thermo Fisher Scientific
San Jose.
Quando e in dubbio la misura di sicurezza
per una procedura, prima di continuare, si
prega di mettersi in contatto con il
Servizio di Assistenza Tecnica locale per i
prodotti di Thermo Fisher Scientific San
Jose.
CAUTION Symbol
CAUTION
Electric Shock: This instrument uses
high voltages that can cause personal
injury. Before servicing, shut down the
instrument and disconnect the instrument
from line power. Keep the top cover on
while operating the instrument. Do not
remove protective covers from PCBs.
Chemical: This instrument might contain
hazardous chemicals. Wear gloves when
handling toxic, carcinogenic, mutagenic,
or corrosive or irritant chemicals. Use
approved containers and proper
procedures to dispose waste oil.
Heat: Before servicing the instrument,
allow any heated components to cool.
Fire: Use care when operating the system
in the presence of flammable gases.
Eye Hazard: Eye damage could occur
from splattered chemicals or flying
particles. Wear safety glasses when
handling chemicals or servicing the
instrument.
General Hazard: A hazard is present that
is not included in the above categories.
Also, this symbol appears on the
instrument to refer the user to instructions
in this manual.
When the safety of a procedure is
questionable, contact your local Technical
Support organization for Thermo Fisher
Scientific San Jose Products.
C
Contents
Preface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .v
Related Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . v
Safety and Special Notices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . v
Safety Precautions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vii
Solvent and Gas Purity Requirements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .viii
Service Philosophy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .viii
Level of Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .viii
Contacting Us . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ix
Thermo Scientific
Chapter 1
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
Ionization Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Electron Ionization Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Chemical Ionization Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Ion Polarity Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Scan Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Q1MS and Q3MS Scan Modes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Product Scan Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Parent Scan Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Neutral Loss Scan Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Data-Dependent Scan Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Scan Types . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Full Scan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Selected Ion Monitoring (SIM) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Selected Reaction Monitoring (SRM) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Highly-Selective Reaction Monitoring (H-SRM) . . . . . . . . . . . . . . . . . . . . . 13
Intelligent Selected Reaction Monitoring (iSRM) . . . . . . . . . . . . . . . . . . . . . 13
Quantitation-Enhanced, Data-Dependent MS (QED MS) . . . . . . . . . . . . . . 14
AutoSIM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Data Types . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Profile Data Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Centroid Data Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Mass/Charge Range. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Chapter 2
Functional Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
Autosampler . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Gas Chromatograph . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
TSQ Quantum XLS and TSQ Quantum GC User Guide
i
Contents
Direct Sample Probes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Transfer Line. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Mass Spectrometer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Controls and Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
EI CI Ion Source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Inlet Valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Ion Optics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Mass Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Ion Detection System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Vacuum System and Inlet Gasses Hardware . . . . . . . . . . . . . . . . . . . . . . . . . 41
Electronic Assemblies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Data System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Computer Hardware . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Data System / Mass Spectrometer / GC Interface . . . . . . . . . . . . . . . . . . . . . 48
Data System / Local Area Network Interface . . . . . . . . . . . . . . . . . . . . . . . . . 49
ii
Chapter 3
System Shutdown, Startup, and Reset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .51
Shutting Down the System in an Emergency. . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Placing the System in Standby Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Shutting Down the System Completely. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Starting Up the System after a Complete Shutdown . . . . . . . . . . . . . . . . . . . . . 55
Restoring Power to the TSQ Quantum XLS or TSQ Quantum GC
system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Starting Up the GC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Starting Up the Data System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Starting Up the Mass Spectrometer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Starting Up the Autosampler . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Setting Up Conditions for Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Resetting the Mass Spectrometer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Resetting the Data System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Turning Off Selected Mass Spectrometer Components . . . . . . . . . . . . . . . . . . . 60
Chapter 4
Tuning and Calibrating. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .63
Displaying the FC-43 Mass Spectrum . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Running Auto Tune and Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
Saving the Tune and Calibration Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Chapter 5
Changing Ionization Modes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .73
Removing the Ion Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
Installing the Ion Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
Contents
Thermo Scientific
Chapter 6
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .83
Cleaning Ion Source Components. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Cleaning Ion Volumes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
Cleaning Lenses of the TSQ Quantum XLS Ion Source . . . . . . . . . . . . . . . . 88
Cleaning the Lenses of the TSQ Quantum GC Ion Source . . . . . . . . . . . . . . 95
Reinstalling the Ion Source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102
Replacing the Filament of the TSQ Quantum XLS Ion Source . . . . . . . . . . . . 102
Replacing the Filament of the TSQ Quantum GC Ion Source . . . . . . . . . . . . 105
Disassembling and Reassembling the TSQ Quantum XLS Ion Source
Completely . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
Disassembling and Reassembling the TSQ Quantum GC Ion Source
Completely . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 110
Cleaning Stainless Steel Parts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113
Cleaning Non-Stainless Steel or Hybrid Part . . . . . . . . . . . . . . . . . . . . . . . . 115
Maintaining the Forepump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116
Adding Calibration Compound. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117
Replacing the Ball Valve Seal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120
Removing and Installing a GC Capillary Column . . . . . . . . . . . . . . . . . . . . . . 122
Removing a GC Capillary Column . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 122
Installing a GC Capillary Column. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 124
Chapter 7
Diagnostics and Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .129
TSQ Quantum XLS System Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 129
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 131
Communication Issues . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 131
Contamination Issues . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 133
Filament and Lens Control Issues . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 135
Heated Zone Issues . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 136
High Vacuum Issues . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 138
Linearity Issues . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 139
Power Supply Issues. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 140
Sensitivity Issues . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 140
Stability Issues . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143
Tuning Issues . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143
Replacing a Fuse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145
Replacing PCBs and Power Supplies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 146
Appendix 8
Using the Direct Sample Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .147
Creating an Instrument Method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 147
Creating a Sequence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 150
Preparing the Probe and Inlet Valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 151
Preparing the Mass Spectrometer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 151
Running the Sequence. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 153
Examining the Raw Data in Qual Browser . . . . . . . . . . . . . . . . . . . . . . . . . . . 157
Removing the Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 158
TSQ Quantum XLS and TSQ Quantum GC User Guide
iii
Contents
Chapter 9
Replaceable Parts and Consumables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .161
Accessory Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 161
Chemicals Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 162
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .163
iv
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
P
Preface
The Thermo Scientific TSQ Quantum XLS™ and TSQ Quantum GC™ systems are members
of the TSQ™ Quantum family of mass spectrometers.
This TSQ Quantum XLS and TSQ Quantum GC User Guide contains a description of the
modes of operation and principle hardware components of your TSQ Quantum XLS or TSQ
Quantum GC system. In addition, this manual provides step-by-step instructions for cleaning
and maintaining your mass spectrometer.
Related Documentation
In addition to this manual, Thermo Fisher provides the following for the TSQ Quantum XLS
and TSQ Quantum GC:
• Preinstallation Requirements Guide
• Help available from within the software
Safety and Special Notices
Make sure you follow the precautionary statements presented in this guide. The safety and
other special notices appear in boxes.
Safety and special notices include the following:
CAUTION Highlights hazards to humans, property, or the environment. Each CAUTION
notice is accompanied by an appropriate CAUTION symbol.
IMPORTANT Highlights information necessary to prevent damage to software, loss of
data, or invalid test results; or might contain information that is critical for optimal
performance of the system.
Note Highlights information of general interest.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
v
Preface
Tip Highlights helpful information that can make a task easier.
vi
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
Preface
Safety Precautions
Observe the following safety precautions when you operate or perform service on the mass
spectrometer.
CAUTION Do Not Perform Any Servicing Other Than That Contained in the TSQ
Quantum XLS and TSQ Quantum GC User Guide. To avoid personal injury or damage
to the instrument, do not perform any servicing other than that contained in the TSQ
Quantum XLS and TSQ Quantum GC User Guide or related manuals unless you are
authorized to do so.
CAUTION Shut Down the Mass Spectrometer and Disconnect It From Line Power
Before You Service It. High voltages capable of causing personal injury are used in the
instrument. Some maintenance procedures require that the mass spectrometer be shut
down and disconnected from line power before service is performed. Do not operate the
mass spectrometer with the top or side covers off. Do not remove protective covers from
PCBs.
CAUTION Respect Heated Zones. Treat heated zones with respect. The ion source and
transfer line might be very hot and might cause severe burns if they are touched. Allow
heated components to cool before you service them.
CAUTION Provide and Adequate Fume Exhaust System. It is your responsibility to
provide an adequate fume exhaust system. Samples and solvents that are introduced into
the mass spectrometer will eventually be exhausted from the forepump. Therefore, the
forepump should be connected to a fume exhaust system. Consult local regulations for the
proper method of exhausting the fumes from your system.
CAUTION Use Care When Changing Vacuum Pump Oil. Treat drained vacuum pump
oil and pump oil reservoirs with care. Hazardous compounds introduced into the system
might have become dissolved in the pump oil. Always use approved containers and
procedures for disposing of waste oil. Whenever a pump has been operating on a system
used for the analysis of toxic, carcinogenic, mutagenic, or corrosive/irritant chemicals, the
pump must be decontaminated by the user and certified to be free of contamination
before repairs or adjustments are made by a Thermo Fisher Scientific San Jose Customer
Support Engineer or before it is sent back to the factory for service.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
vii
Preface
Solvent and Gas Purity Requirements
Use the highest purity solvents available. The TSQ Quantum XLS and TSQ Quantum GC
mass spectrometers are extremely sensitive to solvent impurities. Liquid chromatography
grade is the minimum acceptable purity. Higher grade solvents are preferred. Distilled water is
recommended. Deionized water contains chemicals and is not recommended. For a wide
variety of solvents and consumables, visit www.FisherLCMS.com.
The TSQ Quantum XLS and TSQ Quantum GC mass spectrometers use argon as a collision
gas. The argon must be high purity (99.995%). The required gas pressure is 135 ± 70 kPa
(20 ± 10 psig). Thermo Fisher has found that particulate filters are often contaminated and
are therefore not recommended.
Service Philosophy
Servicing the TSQ Quantum XLS and TSQ Quantum GC systems consists of performing
procedures required to maintain system performance standards, prevent system failure, restore
the system to an operating condition, or all of the above. Routine and preventive maintenance
procedures are documented in this manual.
The user is responsible for routine and preventive maintenance during and after the warranty
period. Regular maintenance increases the life of the system, maximizes the up-time of your
system, and allows you to achieve optimum system performance.
Only a Thermo Fisher Scientific Customer Support Engineer can perform services not
described in this manual.
Level of Repair
Thermo Fisher Scientific’s service philosophy for the TSQ Quantum XLS and TSQ
Quantum GC systems calls for troubleshooting to the lowest part, assembly, printed circuit
board (PCB), or module listed in the “Replaceable Parts” chapter of this manual.
For mechanical failures: A mechanical assembly typically is to be repaired to the level of the
smallest item listed in the “Replaceable Parts” chapter of this manual.
For electronic failures: PCBs are not repaired to the component level except in certain cases of
fuses, relays, and so on. When these exceptions occur, component information can be found
in the “Replaceable Parts” chapter.
viii
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
Preface
Contacting Us
There are several ways to contact Thermo Fisher Scientific.
Y To contact Technical Support
Phone
Fax
E-mail
Knowledge base
800-685-9535
561-688-8736
[email protected]
www.thermokb.com
Find software updates and utilities to download at www.mssupport.thermo.com.
Y To contact Customer Service for ordering information
Phone
Fax
Web site
800-532-4752
561-688-8731
www.thermo.com/finnigan
Y To suggest changes to documentation or to Help
• Fill out a reader survey online at www.thermo.com/lcms-techpubs.
• Send an e-mail message to the Technical Publications Editor at
[email protected]
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
ix
1
Introduction
The TSQ Quantum XLS™ and TSQ Quantum GC™ are members of the TSQ™ Quantum
family of Thermo Scientific mass spectrometers. The TSQ Quantum XLS and the TSQ
Quantum GC are advanced analytical instrument that includes a mass spectrometer, liquid
chromatograph, and the Xcalibur™ data system. See Figure 1.
Contents
• Ionization Modes
• Ion Polarity Modes
• Scan Modes
• Scan Types
• Data Types
• Mass/Charge Range
In a typical analysis, an autosampler (AS) introduces a sample into the gas chromatograph
(GC). The GC separates the sample into its various components. The components elute from
the GC and pass into the mass spectrometer where they are analyzed.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
1
1
Introduction
Figure 1.
TSQ Quantum GC mass spectrometer, TriPlus autosampler, and TRACE GC Ultra gas
chromatograph
The TSQ Quantum XLS and TSQ Quantum GC mass spectrometers include an electron
ionization/chemical ionization (EI/CI) ion source, ion optics, a triple-stage mass analyzer, and
an ion detection system, all of which are enclosed in a vacuum manifold. Ionization of the
sample takes place in the ion source. The specific process used to ionize the sample is known
as the ionization mode. The ion optics transmit the ions produced in the ion source into the
mass analyzer, where they are filtered according to their mass-to-charge ratio. The polarity of
the potentials applied to the lenses in the ion source and ion optics determines whether
positively charged ions or negatively charged ions are transmitted to the mass analyzer. You
can configure the mass spectrometer to analyze positively or negatively charged ions (called
the positive or negative ion polarity mode).
The mass spectrometer’s triple-stage mass analyzer performs either one or two stages of mass
analysis:
• The mass spectrometer is operated as a conventional mass spectrometer with one stage of
mass analysis. The ion source ionizes the sample and the ion products are subjected to
mass analysis in the first rod assembly. The second and third rod assemblies transmit the
resulting mass-selected ions to the ion detection system.1
2
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
1 Introduction
Ionization Modes
• The mass spectrometer is operated as a tandem mass spectrometer with two stages of mass
analysis. The ion source ionizes the sample and the ion products are mass analyzed by the
first rod assembly. In this case, however, mass-selected ions exiting the first rod assembly
collide with an inert gas in the second rod assembly and fragment to produce a set of ions
known as product ions. (A chamber called the collision cell surrounds the second rod
assembly. The collision cell can be pressurized with an inert gas.) The product ions
undergo further mass analysis in the third rod assembly to detect selected ions. Two stages
of mass analysis yield far greater chemical specificity than a single stage can achieve,
because of the system’s ability to select and determine two discrete but directly related sets
of masses.
In a first stage of mass analysis the mass spectrometer can be used to elucidate the structures of
pure organic compounds and the structures of the components within mixtures. Furthermore,
in a second stage of mass analysis, the mass spectrometer can fragment and separate each ionic
fragment of a molecule formed in the ion source to build up an entire structure for the
molecule, piece by piece. Thus, TSQ Quantum XLS and TSQ Quantum GC systems make
investigating all pathways for the formation and fragmentation of each ion in the mass
spectrum possible.
The two stages of mass analysis, with resultant reduction of chemical noise in the final mass
spectrum, allow for very selective and sensitive analysis.
Each sequence of single or triple-stage mass analysis of the ions is called a scan. The mass
spectrometer uses several different scan modes and different scan types to filter, fragment, or
transmit ions in the mass analyzer. Along with the ionization and ion polarity modes, the
ability to vary the scan mode and scan type affords the user great flexibility in the
instrumentation for solving complex analytical problems.
Ionization Modes
The specific process used to ionize the analyte is referred to as the ionization mode. You can
operate the TSQ Quantum XLS and TSQ Quantum GC mass spectrometers in either of two
ionization modes:
• Electron ionization mode
• Chemical ionization mode
1
Thermo Scientific
The instrument can also be used as a single-stage mass spectrometer by transmitting the ions through the first
and second rod assemblies followed by mass analysis in the third rod assembly.
TSQ Quantum XLS and TSQ Quantum GC User Guide
3
1
Introduction
Ionization Modes
Electron Ionization Mode
In electron ionization (EI) mode, electrons are emitted by a heated, wire filament that has
electric current running through it, by thermionic emission. The filament and its reflector are
typically maintained at a -70 V potential relative to the ion source block. This potential
accelerates the electrons through the ionization space, called the ion volume. These energetic
electrons interact with neutral, gas-phase analyte molecules present in the ion volume and
cause the analyte to lose an electron and produce a radical cation:
M + e - --> M+ + 2e Frequently, numerous cleavage reactions give rise to fragment ions, which provide structural
information about the analyte.
EI positive-ion mode is the only commonly used EI mode.
Chemical Ionization Mode
In chemical ionization (CI) mode, ionization of the sample molecules is a multi-step process:
1. Reagent gas is introduced into the CI ion volume at a flow (for methane) of about
2 mL/min, along with sample vapors typically present at partial pressures of less than
one-thousandth that of the reagent gas.
2. The energetic (typically 100 eV) electrons emitted by the heated filament interact to
ionize the reagent gas and form a plasma. This reaction also produces thermal electrons.
3. Reagent gas ions react with reagent gas molecules to form a variety of secondary ions that
are stable with respect to further reaction with reagent gas.
For example, for methane:
CH4 + e - ----> CH4+. + 2e CH4 + e - ----> CH3+ + e - + HCH4+. + CH4 ----> CH5+ +CH3.
CH3+ + CH4 ----> C2H5+ + H2
4. Positive sample ions are formed by one of the following:
• The transfer of a proton from a secondary reagent gas ion to a sample molecule
• The abstraction of an electron by a reagent gas ion
• An ion association reaction in which an adduct ion is formed between a reagent gas
ion and a sample molecule
In methane positive ion mode CI, the relevant peaks observed are MH+, [M+CH5]+, and
[M+C2H5]+; but mainly MH+.
4
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
1 Introduction
Ion Polarity Modes
In isobutane positive ion mode CI, the main peak observed is MH+.
In ammonia positive ion mode CI, the main peaks observed are MH+ and [M+NH4]+.
Negative sample ions are most commonly formed by one of the following:
• Sample molecules capture the secondary thermal electrons present in the ion source
• Electron transfer from ionized reagent gas (e.g. NH2-)
• Proton abstraction
Molecular ions observed in negative ion chemical ionization mass spectra are usually M- or
[M-H]-.
Ion Polarity Modes
You can operate the TSQ Quantum XLS and TSQ Quantum GC mass spectrometers in
either of two ion polarity modes: positive or negative. Both positively charged and negatively
charged ions form in the ion source of the mass spectrometer. The mass spectrometer controls
whether positive ions or negative ions are transmitted to the mass analyzer for mass analysis by
changing the polarity of the potentials applied to the ion source and ion optics. The ion optics
deliver the ions produced in the ion source, in a collimated beam, to the mass analyzer. The
mass spectrometer can switch between positive and negative polarity modes in under 95 ms.
The information obtained from a positive-ion mass spectrum is different from and
complementary to the information from a negative-ion spectrum. Thus, the ability to obtain
both positive-ion and negative-ion mass spectra aids you in the qualitative analysis of your
sample. You can choose the ion polarity mode and ionization mode to obtain maximum
sensitivity for the particular analyte of interest.
Scan Modes
You can operate the TSQ Quantum XLS and TSQ Quantum GC mass spectrometers in a
variety of scan modes. The most commonly used scan modes can be divided into two
categories: single mass spectrometry (MS) scan modes and MS/MS scan modes. The scan
modes in each category are as follows:
• MS scan modes: Q1MS and Q3MS scan modes
• MS/MS scan modes: Product scan mode, Parent scan mode, Neutral Loss scan mode
• Data-dependent scan mode
The scan modes that can be employed depend on the number and type of rod assemblies and
the voltages applied to the rod assemblies.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
5
1
Introduction
Scan Modes
The mass analyzer or the mass spectrometer has three rod assemblies.2 The first and third rod
assemblies, Q1 and Q3, are quadrupoles, and the second rod assembly, Q2, is a square-profile
quadrupole.
Rod assemblies can operate in either of two capacities:
• As ion transmission devices
• As mass analyzers
If only RF voltage is applied, a rod assembly serves as an ion transmission device that passes
all ions within a large range of mass-to-charge ratios (that is, virtually all ions present).
When you apply both RF and dc voltages to a rod assembly, the separation of ions of different
mass-to-charge ratios occurs. This separation allows the rod assembly to serve as a mass
analyzer.
On the TSQ Quantum XLS and TSQ Quantum GC mass spectrometers, the quadrupole rod
assemblies can operate with rf and dc voltages or with only rf voltage. That is, Q1 and Q3 can
act either as mass analyzers or ion transmission devices. The Q2 rod assembly operates
exclusively with rf voltage. Thus, Q2 is always an ion transmission device. Surrounding Q2 is
a collision cell where fragmentation can take place if the argon collision gas is present in the
cell. For a summary of how the rod assemblies function in several of the major scan modes, see
Table 1.
.
Table 1. Summary of scan modes
Scan mode
Q1 quadrupole
Q2 collision cell
Q3 quadrupole
Q1MS
Scana
Pass all ionsb
Pass all ions
Q3MS
Pass all ions
Pass all ions
Scan
Product
Setd
Fragment ionsc, then pass
all fragments
Scan
Parent
Scan
Fragment ions, then pass all
fragments
Set
Neutral Loss
Scan
Fragment ions, then pass all
fragments
Scan
a
Scan = full scan or transmission of selected ions
Pass all ions or fragments = pass ions or fragments within a wide range of mass-to-charge ratios
c
Fragment ions = collisions with argon gas cause ions to fragment
d
Set = set to pass ions of a single mass-to-charge ratio or a set of mass-to-charge ratios
b
2A
rod assembly is a regular array of metal rods. Refer to “Mass Analyzer” on page 32 for a discussion of the rod
assemblies used on the TSQ Quantum XLS instrument.
6
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
1
Introduction
Scan Modes
Q1MS and Q3MS Scan Modes
In the Q1MS and Q3MS scan modes, only one stage of mass analysis is performed. The mass
spectrum obtained is equivalent to the mass spectrum obtained from an instrument with a
single mass analyzer. In the one stage of analysis, ions formed in the ion source enter the
analyzer assembly. One of the mass analyzers (Q1 or Q3) is scanned to obtain a complete mass
spectrum. The other rod assemblies (Q2 and Q3, or Q1 and Q2, respectively) act as ion
transmission devices. In the Q1MS scan mode, Q1 is used as the mass analyzer; in the Q3MS
scan mode, Q3 is used as the mass analyzer.
Product Scan Mode
Product scan mode performs two stages of analysis. In the first stage, ions formed in the ion
source enter Q1, which is set to transmit ions of one mass-to-charge ratio. Ions selected by
this first stage of mass analysis are called parent ions. (As a result, Q1 is referred to as the
parent mass analyzer, and the mass-to-charge ratio of ions transmitted by the parent mass
analyzer is referred to as the parent set mass.) Parent ions selected by Q1 then enter Q2, which
is surrounded by the collision cell.
Note When referring to the first, second, and third rod assemblies as pieces of
hardware, it is convenient to call them Q1, Q2, and Q3, respectively. However, when
discussing their function in MS/MS scan modes, it often adds clarity to refer to them as
the parent mass analyzer, collision cell (ion transmission device surrounded by the
collision cell), and product mass analyzer, respectively.
In the second stage of analysis, ions in the collision cell can fragment further to produce
product ions. Two processes produce product ions: by unimolecular decomposition of
metastable ions or by interaction with argon collision gas present in the collision cell. This
latter step is known as collision-induced dissociation (CID). Ions formed in the collision cell
enter Q3 (the product mass analyzer) for the second stage of mass analysis. Q3 is scanned to
obtain a mass spectrum that shows the product ions produced from the fragmentation of the
selected parent ion.
A mass spectrum obtained in the Product scan mode (product mass spectrum) is the mass
spectrum of a selected parent ion.
The Product scan mode is illustrated in Figure 2.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
7
1
Introduction
Scan Modes
Figure 2.
Illustration of Product scan mode
Q1 Set
Q2
RF Only + Ar
Q3 Scanning
Q3 m/z
Parent Scan Mode
The Parent scan mode also uses two stages of analysis. In the first stage, ions formed in the ion
source are introduced into the parent mass analyzer, which is scanned to transmit parent ions
sequentially into the collision cell.
In the second stage of analysis, in the collision cell, parent ions can fragment to produce
product ions by unimolecular decomposition of metastable ions or by collision-induced
dissociation. Ions formed in the collision cell enter the product mass analyzer, which transmits
a selected product ion. (The product set mass is the mass-to-charge ratio of ions transmitted
by the product mass analyzer.)
The resultant spectrum shows all the parent ions that fragment to produce the selected
product ion. Note that for a mass spectrum obtained in the Parent scan mode (parent mass
spectrum), data for the mass-to-charge ratio axis are obtained from Q1 (the parent ions),
whereas data for the ion intensity axis are obtained from Q3 (from monitoring the product
ion).
The Parent scan mode is illustrated in Figure 3.
8
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
1
Figure 3.
Introduction
Scan Modes
Illustration of the Parent scan mode
Q1 Scanning
Q2
RF Only + Ar
Q3 Set
Q1 m/z
Experiments that employ the parent scan mode (parent experiments) can be used in structure
and fragmentation studies as well as in survey analyses of mixtures. In general, parent
experiments detect all compounds that decompose to a common fragment. The experiments
are useful for the rapid detection of a series of structural homologs that have a common
fragment ion (for example, m/z 149 for the phthalates).
Neutral Loss Scan Modes
In the Neutral Loss scan mode, the two mass analyzers (Q1 and Q3) are linked together so
that they are scanned at the same rate over mass ranges of the same width. The respective mass
ranges, however, are offset by a selected mass, such that the product mass analyzer scans a
selected number of mass units lower than the parent mass analyzer.
Thus, in the Neutral Loss scan mode, there are two stages of mass analysis. In the first stage,
the parent mass analyzer separates ions formed in the ion source by mass-to-charge ratio.
Then the ions are introduced sequentially into the collision cell.
In the second stage of analysis, ions admitted to the collision cell can fragment further by
metastable ion decomposition or by CID to produce product ions. The product mass analyzer
then separates these product ions by mass-to-charge ratio. Neutral Loss scan mode is
illustrated in Figure 4. Examples of compounds with a common neutral loss fragment appear
in Figure 5.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
9
1
Introduction
Scan Modes
To detect an ion, between the time the ion leaves Q1 and enters Q3, it must lose a neutral
moiety whose mass (the neutral loss mass) is equal to the difference in the mass ranges being
scanned by the two mass analyzers. Thus, a neutral loss mass spectrum is a spectrum that
shows all the parent ions that lose a neutral species of a selected mass.
Note that a neutral gain (or association) experiment can also be performed in which the mass
range scanned by Q3 is offset by a selected mass above the mass range scanned by Q1.
For a neutral loss (or neutral gain) mass spectrum, as for a parent mass spectrum, data for the
mass-to-charge ratio axis are obtained from Q1 (the parent ion), whereas data for the ion
intensity axis are obtained from Q3 (the product ion being monitored).
Experiments that use the Neutral Loss scan mode (neutral loss experiments) are useful when a
large number of compounds are being surveyed for common functionality. Neutral moieties
are frequently lost from substituent functional groups (for example, CO2 from carboxylic
acids, CO from aldehydes, HX from halides, and H2O from alcohols).
Figure 4.
Illustration of the Neutral Loss scan mode
Q2
Q1 Scanning
R F O nly + A r
Q3 = Q1 - Δ
Q1 m/z
10
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
1
Figure 5.
Introduction
Scan Modes
Examples of compounds with a common neutral-loss fragment
NH2
N
N
H2 N
N
HO
N
N
N
N
H2 N
N
N
H2 N
N
N
Data-Dependent Scan Mode
The TSQ Quantum XLS and TSQ Quantum GC mass spectrometers use the information in
a data-dependent scan mode experiment to make automatic decisions about the next step of
the experiment without input from a user. In data-dependent scan mode you specify criteria
to select one or more ions of interest on which to perform subsequent scans, such as MS/MS.
You can approach the setup of data-dependent experiments in either of two ways:
• If you have some idea of what the parent ion is, or if you expect a certain kind of parent
ion, you can set up a list of possible parent ions. Then, when one of the parent ions you
specified is detected, you can acquire product spectra and analyze the information.
Conversely, you can also set up a list of ions that you do not want selected for
fragmentation.
• If you have little information about your compound, you can set up the parameters of a
data-dependent experiment so that if the intensity of the ion signal is above a specified
threshold, the mass spectrometer generates product spectra. Later, you can decide if the
information is useful.
Because a data-dependent scan needs to use a target ion from a previous scan, the first scan
event cannot be a data-dependent scan.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
11
1
Introduction
Scan Types
Scan Types
You can operate the TSQ Quantum XLS and TSQ Quantum GC systems with a variety of
scan types. The most common scan types are as follows:
• Full Scan
• Selected Ion Monitoring (SIM)
• Selected Reaction Monitoring (SRM)
• Intelligent Selected Reaction Monitoring (iSRM)
• Quantitation-Enhanced, Data-Dependent MS (QED MS)
• AutoSIM
Full Scan
The full-scan scan type provides a full mass spectrum of each analyte. With full scan, the
scanning mass analyzer is scanned from the first mass to the last mass, without interruption,
in a given scan time.
Full-scan experiments are used to determine or confirm the identity of unknown compounds
or the identity of each component in a mixture of unknown compounds. (Generally, a full
mass spectrum is needed to determine the identity of an unknown compound.)
The full-scan scan type gives you more information about an analyte than does SIM, but a full
scan does not yield the sensitivity that the other two scan types can achieve. With full scan,
you spend less time monitoring the signal for each ion than you do in SIM or SRM. Full scan
provides greater information but lower sensitivity than the other two scan types.
To use the SIM or SRM, you need to know what ions or reactions you are looking for before
you can perform an experiment with these scan types. Thus, you might use a full scan for SIM
to determine the identity of an analyte and to obtain its mass spectrum and a full scan for
SRM to determine the mass spectrum and product mass spectra for parent ions of interest.
Then, you might use SIM or SRM to do routine quantitative analysis of the compound.
Selected Ion Monitoring (SIM)
Selected ion monitoring (SIM) monitors a particular ion or set of ions. SIM experiments are
useful in detecting small quantities of a target compound in a complex mixture when you
know the mass spectrum of the target compound. Thus, SIM is useful in trace analysis and in
the rapid screening of a large number of samples for a target compound.
12
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
1
Introduction
Scan Types
Because SIM monitors only a few ions, it can provide lower detection limits and greater speed
than the full-scan modes. SIM achieves lower detection limits because more time is spent
monitoring significant ions that are known to occur in the mass spectrum of the target
analyte. SIM can achieve greater speed because it monitors only a few ions of interest; SIM
does not monitor regions of the spectrum that are empty or have no ions of interest.
SIM can improve the detection limit and decrease analysis time, but it can also reduce
specificity. Because SIM monitors only specific ions, any compound that fragments to
produce those ions will appear to be the target compound. The result could be a false positive.
Selected Reaction Monitoring (SRM)
Selected reaction monitoring (SRM) monitors a particular reaction or set of reactions, such as
the fragmentation of an ion or the loss of a neutral moiety.
SRM monitors a limited number of parent/product-ion pairs. In product-type experiments, a
parent ion is selected as usual, but generally only one product ion is monitored. SRM
experiments are normally conducted with the product scan mode.
As does SIM, SRM provides for the very rapid analysis of trace components in complex
mixtures. However, because SRM selects two sets of ions, it obtains specificity that is much
greater than what SIM can obtain. Any interfering compound would not only have to form
an ion source product (parent ion) of the same mass-to-charge ratio as the selected parent ion
from the target compound, but that parent ion would also have to fragment to form a product
ion of the same mass-to-charge ratio as the selected product ion from the target compound.
Highly-Selective Reaction Monitoring (H-SRM)
Highly-selective reaction monitoring (H-SRM) is SRM performed at higher resolution (0.4 u
FWHM versus 0.7 u FWHM for SRM).
Intelligent Selected Reaction Monitoring (iSRM)
In intelligent selected reaction monitoring (iSRM) experiments, the TSQ Quantum XLS and
TSQ Quantum GC mass spectrometers perform multiple “primary” selected reaction
monitoring (SRM) scans for each analyte (parent ion) of interest. The mass spectrometer uses
for quantification the sum of the intensities of an analyte’s primary SRM scans.
You specify a trigger threshold for each primary SRM scan. If the intensities of all the listed
primary SRM reactions for an analyte are above their trigger thresholds, and if all these scans
are within a specified retention time window, then the mass spectrometer performs
“secondary” SRM scans to confirm the identity of the analyte. The maximum number of
secondary SRM reactions is 12400.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
13
1
Introduction
Data Types
The mass spectrometer uses reference SRM scans to adjust the acquisition time window of
subsequent SRM scans. It uses the time difference between the expected and observed
retention times of reference SRM reactions to correct the retention times that it uses to make
procedural decisions. The mass spectrometer still records data with the true retention times.
You should specify two to four reference reactions spaced throughout the LC run. The mass
spectrometer uses the preceeding two retention time corrections to calculate a piecewise linear
retention time correction for the primary SRM scans that follow. As with primary SRM scans,
reference SRM scans may trigger secondary SRM scans.
The maximum number of primary plus reference SRM transitions is 3600.
Quantitation-Enhanced, Data-Dependent MS (QED MS)
Quantitation-enhanced, data-dependent MS (QED MS) scan type provides simultaneous
quantitation (SRM) and structural confirmation (full-scan MS/MS product ion spectra). If
the intensity of an SRM reaction that you specify for a parent ion is above its trigger
threshold, the mass spectrometer performs a full-scan MS/MS product ion scan to confirm
the identity of the analyte.
AutoSIM
In the scan type known as AutoSIM, the mass spectrometer automatically selects the most
intense masses (m/z values) in a survey scan, builds a SIM scan list for them, and then acquires
and records ion current at only these selected masses. AutoSIM scans can be performed on any
full scan in any scan mode, but not on data-dependent scans.
Data Types
You can acquire and display mass spectral data (intensity versus mass-to-charge ratio) with the
TSQ Quantum XLS and TSQ Quantum GC mass spectrometers in one of two data types:
• Profile data type
• Centroid data type
Profile Data Type
In the profile data type, you can see the shape of the peaks in the mass spectrum. Each atomic
mass unit is divided into many sampling intervals. The intensity of the ion current is
determined at each of the sampling intervals. The profile data type displays the intensity at
each sampling interval with the intensities connected by a continuous line. In general, use the
profile scan data type when you tune and calibrate the mass spectrometer so that you can
easily see and measure mass resolution.
14
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
1 Introduction
Mass/Charge Range
Centroid Data Type
The centroid data type displays the mass spectrum as a bar graph and sums the intensities of
each set of multiple sampling intervals. This sum is displayed versus the integral center of
mass of the sampling intervals. In general, use the centroid scan data type for data acquisition
for faster scan speed. Data processing is also much faster for centroid data.
Mass/Charge Range
The TSQ Quantum XLS and TSQ Quantum GC mass spectrometers can operate in a
mass/charge range of 10 to 3000 Da.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
15
2
Functional Description
This chapter describes the principal components of the TSQ Quantum XLS and TSQ
Quantum GC systems and their respective functions.
Contents
• Autosampler (optional)
• Gas Chromatograph
• Direct Sample Probes (optional)
• Transfer Line
• Mass Spectrometer
• Data System
A functional block diagram of the TSQ Quantum XLS or TSQ Quantum GC system is
shown in Figure 6. A sample transfer line connects the gas chromatograph (GC) to the mass
spectrometer. The autosampler and GC are installed on the left side of the mass spectrometer.
In analysis by GC/MS, a sample is injected into a GC column. The sample then separates into
its various components. The components elute from the GC column and pass through the
transfer line into the mass spectrometer where they are analyzed. You can also use a direct
sample probe to introduce sample into the mass spectrometer.
Electron ionization (EI) or chemical ionization (CI) ionize sample molecules upon entering
the mass spectrometer. The ion optics focus and accelerate the resulting sample ions into the
mass analyzer where they are filtered according to their mass-to-charge ratios. An ion
detection system then produces a signal proportional to the number of ions detected. The
system electronics receive and amplify the ion current signal from the ion detection system.
That signal is then passed on to the data system for further processing, storage, and display.
The data system provides the primary mass spectrometer user interface.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
17
2
Functional Description
Autosampler
Figure 6.
Functional block diagram of the TSQ Quantum XLS or TSQ Quantum GC system
Sample flow
Electrical connection
Mass spectrometer
Data system
Autosampler
(Optional)
Printer
(optional)
Gas
chromatograph
Ion
source
Ion
optics
Mass
analyzer
Ion detection
system
Transfer
line
Vacuum
system
Instrument
control
electronic
assemblies
Personal
computer
Video
monitor
Autosampler
The (optional) Thermo Scientific TriPlus autosampler is used to inject samples automatically
into the GC inlet. With an autosampler, you can automate your GC/MS/MS analyses. The
TriPlus autosampler is shown in Figure 1 on page 2.
Autosampler Start/Stop signals with the mass spectrometer are provided by contact closure.
You configure the TriPlus autosampler from the data system computer. Select the TriPlus
instrument button in the Instrument Configuration window, which is available by choosing
Start > All Programs > Xcalibur > Instrument Configuration. Refer to Xcalibur Help for a
description of TriPlus configuration options.
You also use the data system to set up the TriPlus autosampler to inject samples. Choose
Start > All Programs > Xcalibur > Xcalibur and click Instrument Setup to open the
Instrument Setup window. Then, click the TriPlus icon to open the TriPlus Autosampler
page. Refer to the Help for instructions on running the TriPlus autosampler.
Refer to the documentation provided with the autosampler for maintenance procedures.
18
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
2
Functional Description
Gas Chromatograph
Gas Chromatograph
The Thermo Scientific TRACE GC Ultra gas chromatograph (GC)) separates a sample
mixture into its chemical components by gas chromatography. In gas chromatography, the
sample mixture is partitioned between a solid stationary phase and a mobile gas. The
stationary phase is adhered to the inside of a small-diameter glass tube: the capillary column.
The molecular structure of each component of the mixture determines in which order each
component elutes from the GC and enters the mass spectrometer. The TRACE GC Ultra gas
chromatograph is shown in Figure 1 on page 2.
Gas chromatography is widely used in analytical chemistry, though the high temperatures
used in GC make it unsuitable for high molecular weight biopolymers, frequently
encountered in biochemistry. It is well suited for use in the petrochemical, environmental
monitoring, and industrial chemical fields. It is also used extensively in chemistry research.
You configure the TRACE GC Ultra gas chromatograph from the data system computer.
Select the TRACE GC Ultra instrument button in the Instrument Configuration window,
which is available by choosing Start > All Programs > Xcalibur > Instrument
Configuration. Refer to Xcalibur Help for a description of TRACE GC Ultra configuration
options.
The TSQ Quantum XLS and TSQ Quantum GC data system computers can directly control
the TRACE GC Ultra. Choose Start > Programs > Xcalibur > Xcalibur and click
Instrument Setup to open the Instrument Setup window. Click the TRACE GC Ultra icon
to open the TRACE GC Ultra page. Refer to the Help for instructions for operating the
TRACE GC Ultra.
Front-panel (keypad) operation of the GC and maintenance procedures for the GC are
described in the documentation provided with the GC. To replace the GC capillary column,
see “Removing and Installing a GC Capillary Column” on page 122.
Direct Sample Probes
With the (optional) direct sample probes you can introduce compounds directly into the ion
source without GC column separation. See Figure 7. The direct sample probes are ideal for
qualitative or semi-quantitative analysis of materials that don't require a GC column
separation, or are difficult, if not impossible, to elute chromatographically, such as solids. The
direct sample probes introduce samples directly into the ion source via a vacuum interlock. A
single controller box (Figure 7) with interchangeable probe tools makes it easy to select the
best method of sample introduction.
The direct sample probe system includes two probe tools:
• Direct exposure probe
• Direct insertion probe
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
19
2
Functional Description
Transfer Line
The direct exposure probe (DEP) has a heated filament that rapidly vaporizes liquids or
solutions. The DEP is ideal for rapid molecular weight confirmation of liquids or solids
dissolved in a suitable solvent. The DEP can vaporize compounds with a high boiling point.
The direct insertion probe (DIP) has a temperature-controlled, heated capillary tube that
slowly vaporizes solid samples. You can use the DIP for rapid analysis of solids or trace
components in solid matrices, such as forensic samples or tissue.
Figure 7. Direct sample probe and controller
Transfer Line
The transfer line is the interface between the GC and mass spectrometer. The transfer line
heats the capillary column as it passes from the GC into the ion source in the mass
spectrometer. This prevents the sample from condensing. The transfer line includes an inlet
for calibration gas and chemical ionization gas.
20
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
2
Figure 8.
Functional Description
Mass Spectrometer
Transfer line
GC end
Inlet for
calibration gas
and CI gas
Mass spectrometer
ion source end
Mass Spectrometer
The TSQ Quantum XLS mass spectrometer provides sample ionization and mass analysis of
samples introduced from a gas chromatograph or direct insertion probe. The mass
spectrometer uses a triple-quadrupole mass analyzer with an ion source external to the mass
analyzer. Several important features of the TSQ Quantum XLS and TSQ Quantum GC mass
spectrometers are as follows:
• High sensitivity and resolution
• m/z 10 to 3000 mass range
• EI and CI ionization modes
• Positive and negative ion polarity modes
• MS and MS/MS scan modes
• Full-scan, SIM, SRM, iSRM, QED MS, AutoSIM, and data-dependent scan types
The mass spectrometer includes the following components:
• Controls and Indicators
• EI CI Ion Source
• Ion Optics
• Mass Analyzer
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
21
2
Functional Description
Mass Spectrometer
• Ion Detection System
• Vacuum System and Inlet Gasses Hardware
• Electronic Assemblies
• Data System
Controls and Indicators
Five light-emitting diodes (LEDs) are located at the upper right side of the front panel of the
mass spectrometer. See Figure 9.
The Power LED illuminates green whenever power is supplied to the vacuum system and
electronic assemblies of the mass spectrometer.
The Vacuum LED illuminates yellow when the turbomolecular pump is nearly at speed (80%
of its operating speed of 750 MHz) and it is safe to turn on the ion gauge. The Vacuum LED
is off if the turbomolecular pump is not at speed. The Vacuum LED illuminates green
whenever the pressure in the analyzer chamber, as measured by the ion gauge, is at or below
the value required to enable high voltages to the mass analyzer. See Table 2.
Figure 9.
Front panel LEDs of the mass spectrometer
Power
Vacuum
System
Communication
Scan
Table 2. Maximum allowed pressure to turn on high voltages
Carrier gas
Ar collision gas
Maximum pressure (Torr)
He
Off
6 × 10-6
He
On
5 × 10-5
H2
Off
5 × 10-5
H2
On
1 × 10-4
The Communication LED illuminates yellow when the mass spectrometer and the data
system are trying to establish a communication link. The Communication LED illuminates
green when the Ethernet communication link between the mass spectrometer and the data
system has been made.
22
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
2
Functional Description
Mass Spectrometer
The System LED illuminates yellow whenever the mass spectrometer is in Standby—that is,
high voltage is not supplied to the ion source, mass analyzer, or ion detection system, but the
mass spectrometer power is on. The System LED illuminates green whenever the high voltage
is enabled and the system is in the On state. High voltage is enabled if the pressure of the
analyzer chamber is below the values listed in Table 2.
The Scan LED flashes blue whenever the mass spectrometer is on and is scanning ions.
The System Power Off button, located on the front of the electronics module, turns off power
to the mass spectrometer, gas chromatograph, and autosampler. See Figure 10. You must use
the main power circuit breaker, located on the back of the electronics module, to restore
power to the mass spectrometer, gas chromatograph, and autosampler.
CAUTION In an emergency, to shut off all power to the mass spectrometer, gas
chromatograph, and autosampler, press the System Power Off button located at the front
of the instrument.
Figure 10. Front panel System Power Off button
System
System Power Off
The main power circuit breaker switch (labeled Main Power) is located on the power panel at
the back of the electronics module. See Figure 11. In the Off (O) position, the circuit breaker
removes all power to the mass spectrometer, gas chromatograph, and autosampler. In the
On (|) position, power is supplied to the mass spectrometer, gas chromatograph, and
autosampler. In the standard operational mode, the circuit breaker is kept in the On (|)
position.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
23
2
Functional Description
Mass Spectrometer
Figure 11. Rear power panel of the electronics module
GC Power Out
Power In
Quantum Power Out
A/S Power Out
!
!
V~230
Hz, 30A MAX
V~230
50/60 Hz, 16A MAX
V~230
50/60 Hz, 10A MAX
V~230
50/60 Hz, 2A MAX
The mass spectrometer main power circuit breaker switch (labeled Main Power) is located on
the mass spectrometer power panel in the lower corner of the right-side panel of the mass
spectrometer. See Figure 12. In the Off position, the circuit breaker removes all power to the
mass spectrometer, including the vacuum pumps. In the On position, power is supplied to the
mass spectrometer. In the standard operational mode, the circuit breaker is kept in the On
position.
The electronics service switch (labeled Electronics) is located next to the main power circuit
breaker on the mass spectrometer power panel (Figure 12). In the Service Mode position the
switch removes power to all components of the mass spectrometer other than the vacuum
system. The Operating Mode position supplies power to all non-vacuum system components
of the mass spectrometer.
The vacuum service switch (labeled Vacuum) is located next to the electronics service switch
on the mass spectrometer power panel (Figure 12). In the Service Mode position the switch
removes power to all components of the vacuum system, including the forepump,
turbomolecular pump, and turbomolecular pump controller. The switch in the Operating
Mode position supplies power to all vacuum system components of the mass spectrometer.
The System Reset button is also located on the mass spectrometer power panel. Pressing the
System Reset button takes the 5 V logic to ground and causes the embedded computer on the
System Control PCB to reboot. The mass spectrometer software is then reloaded from the
data system. See “Resetting the Mass Spectrometer” on page 59 for information on how to
reset the mass spectrometer.
Three LEDs are located on the power panel: The Pump On LED is green when the rough
pump current sensor detects current to the forepump. The LED is off when the rough-pump
current sensor does not detect current to the forepump. If the current sensor detects a loss of
current when the mass spectrometer is on, the vacuum system vents.
The Vent Valve Closed LED is green whenever the vent valve current sensor detects current
through the vent valve and the vent valve is closed. The LED is off when the vent valve is
open.
24
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
2
Functional Description
Mass Spectrometer
The Ethernet Link OK LED is green when the System Control PCB is communicating with
the data system PC. The LED is off when there is no communication between the System
Control PCB and the data system PC.
Figure 12. Right-side power panel of the mass spectrometer
Pum p O n
Electronics
Vacuum
Qualified
Service
Personnel
Only
!
EI CI Ion Source
The ion source forms gas phase sample ions from sample molecules that elute from the GC or
are introduced by the direct sample probe. You can operate the ion source in either the
electron ionization (EI) or chemical ionization (CI) mode. See “Ionization Modes” on page 3.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
25
2
Functional Description
Mass Spectrometer
Figure 13. Ion source of the TSQ Quantum XLS
Magnets and
magnet yoke
EI/CI Source
PCB
Filament
Ion source
block
Heater ring
Q00 quadrupole lens
assembly
L1, L2, L3, lens
assembly
26
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
2
Functional Description
Mass Spectrometer
Figure 14. Ion source of the TSQ Quantum GC
Magnets and
magnet yoke
EI/CI Source
PCB
Ion source
block
Filament
Heater ring
L1, L2, L3, and L4 lens
assembly
The ion volume, located in the center of the ion source, is the site where electrons interact
with sample or reagent gas molecules to form ions. Three exchangeable ion volumes and
ion volume holders are available for use in the ion source. See Figure 15. The choice of ion
volume depends on the ionization mode. The open EI ion volume is open on the analyzer
end, with a relatively large electron entrance hole. The closed EI ion volume has a smaller ion
exit hole than the EI ion volume. This results in a higher pressure of analyte, and greater
sensitivity, but less dynamic range. The CI ion volume is closed on the analyzer end, except
for a small ion exit hole and a relatively small electron entrance hole. In the CI mode it is
important to maintain a relatively high reagent gas pressure.
The ion source block holds the ion volume in its center. Samples are introduced from the
GC capillary column through an aperture on the side of the ion source block and then into
the ion volume. Gases for calibration or chemical ionization enter the ion volume through a
gas inlet tube.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
27
2
Functional Description
Mass Spectrometer
Figure 15. EI (left), closed EI (center), and CI (right) ion volumes and holders
Ion volume
holder
Electron
entrance
hole
Ion exit hole
(front)
GC effluent, CI gas, calibration gas
entrance hole
Electron
exit hole
Cartridge heaters heat the ion source block and lens assemblies to minimize the rate at which
deposits form on the ions source block, ion volume, and lens assemblies. The ion source block
is also heated to assist in the rapid temperature equilibration of the ion source block and ion
volume after the filament is turned on. In EI mode, the ion source is typically maintained at
150 oC to 220 oC. However, lower or higher temperatures are sometimes used for certain
applications. In the CI mode, the ion source is typically maintained at 180 oC to 200 oC. The
ion source heater is feedback controlled.
The filament assembly, positioned in a recess at the top of the ions source block, contains the
filament, reflector, and electron lens. The filament is a rhenium wire that is electrically heated
to produce electrons by thermionic emission. The reflector repels electrons emitted by the
filament away from the filament toward the ion volume. The filament and its reflector are
maintained at a negative potential relative to the ion volume. The electron lens prevents
positive ions from traveling up the electron beam. In turn, this prevents positive ions from
leaving the ion volume through the electron entrance hole. The difference in potential
between the filament and the ion volume determines the electron energy.
Two permanent magnets are held in a magnet yoke in the proper position above and below
the ion source block. The permanent magnets collimate the electron beam and cause the
beam to spiral through the ion volume. This ensures optimum ionization of the sample.
Two ion source len assemblies, called the L1, L2, and L3 lens assembly and the Q00
quadrupole lens assembly, extract the ions formed in the ion source and transmit them to the
ion optics. The lens L2 voltage is mass dependent. The rods of the Q00 quadrupole contact
the rods of the Q0 quadrupole and share their rf and dc voltages.
In positive ion mode, the lenses are adjusted to a negative potential, and in negative ion mode,
the lenses are adjusted to a positive potential. The arrangement of the lenses and their spacers
enables quick and efficient pump-out of the ion source.
28
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
2
Functional Description
Mass Spectrometer
Inlet Valve
The inlet valve, which is attached to the front of the vacuum manifold, is a vacuum-sealed
valve that allows you to change ion volumes or insert the direct exposure probe without
venting the mass spectrometer to atmosphere. You use the insertion/removal (I/R) tool to
insert or remove ion volumes. See Figure 16. The forepump evacuates the inlet valve.
The ball valve is a ball with a hole in it, and it is located between the vacuum manifold and
the inlet valve block. The ball valve is open when the hole in the ball is aligned with the
opening between the vacuum manifold and the inlet valve. The ball valve lever opens and
closes the ball valve. When the ball valve is closed, it prevents the vacuum manifold from
venting to atmosphere.
Figure 16. Inlet valve, I/R tool, guide bar, and ball valve lever
Inlet
valve
I/R tool
Ball valve lever
Guide bar
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
29
2
Functional Description
Mass Spectrometer
Ion Optics
The ion optics focus the ions produced in the ion source and transmit them to the mass
analyzer. The ion optics includes the Q0 quadrupole and lenses L11 and L12.
The Q0 quadrupole is a square array of square-profile rods that acts as an ion transmission
device. See Figure 17. An rf voltage applied to the rods gives rise to an electric field that guides
the ions along the axis of the quadrupole. The Q0 offset voltage increases the translational
kinetic energy of ions emerging from the ion source.
Figure 17. Q0 quadrupole
Figure 18 gives a cross-sectional view of the transfer line, ion source, and Q0 quadrupole.
30
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
2
Functional Description
Mass Spectrometer
Figure 18. Cross-sectional view of the transfer line, ion source, and Q0 quadrupole of the TSQ Quantum XLS
Lens L1
Q00 quadrupole
Q0 quadrupole
Ion source
ion volume
Lens L2
Lens L3
Transfer line
Capillary column
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
31
2
Functional Description
Mass Spectrometer
Figure 19. Cross-sectional view of the transfer line, ion source, and Q0 quadrupole of the TSQ Quantum GC
Lens L2
Q0 quadrupole
Lens L1
Lens L3
Ion source
ion volume
Lens L4
Transfer line
Capillary column
The L11 and L12 lenses are metal disks with a circular hole in the center through which the
ion beam can pass. Together they act as a two-element cone lens. An electrical potential can be
applied to the lens to accelerate (or decelerate) ions as they approach the lens and to focus the
ion beam as it passes through the lens. The value ranges between 0 and ±300 V. Lenses L11
and L12 also act as a vacuum baffle between the Q0 quadrupole chamber and the mass
analyzer chamber.
Mass Analyzer
The mass analyzer separates ions according to their mass-to-charge ratio and then passes them
to the ion detection system. The mass analyzer on the TSQ Quantum XLS and TSQ
Quantum GC mass spectrometer consists of three quadrupole rod assemblies (Q1, Q2, and
Q3) and three lens sets.
The mass analyzer is discussed in detail in the following subtopics:
• Quadrupole Rod Assemblies
• rf and DC Fields Applied to the Quadrupoles
32
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
2
Functional Description
Mass Spectrometer
• Mass Analysis
• Collision Cell and CID Efficiency
• Quadrupole Offset Voltage
• Mass Analyzer Lenses
Quadrupole Rod Assemblies
The three rod assemblies used in the mass spectrometer are numbered from the ion source end
of the manifold and are designated Q1, Q2, and Q3. Q1 and Q3 are quadrupoles that enable
high-resolution scans without signal loss.
Q1 and Q3 are square arrays of precision-machined and precision-aligned, hyperbolic-profile
round rods. Q1 or Q3 are shown in Figure 20. Quartz spacers act as electrical insulators
between adjacent rods.
Figure 20. Q1 or Q3 quadrupole
Q2 is a square-profile quadrupole rod assembly. Q2 always acts as an ion transmission device.
The Q2 quadrupole rods are bent through a 90-degree arc. In addition to reducing the
footprint of the instrument, this prevents the transmission of unwanted neutral species to the
detector and dramatically lowers the noise level in the data. Q2 has become synonymous with
the term collision cell. Technically, the collision cell is the chamber that encloses Q2 where
collision-induced dissociation can take place if the argon collision gas is present. See
Figure 21.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
33
2
Functional Description
Mass Spectrometer
Figure 21. Collision cell, Q2 quadrupole, and lenses
Lenses L21, L22,
and L23
Lenses L31, L32, and
L33
rf and DC Fields Applied to the Quadrupoles
In a quadrupole rod assembly, because rods opposite each other in the array connect
electrically, the four rods can be considered as two pairs of two rods each. Ac and dc voltages
are applied to the rods and these voltages are ramped during the scan. Voltages of the same
amplitude and sign are applied to the rods of each pair. However, the voltages applied to the
different rod pairs are equal in amplitude but opposite in sign. See Figure 22.
Figure 22. Polarity of the rf and dc voltages applied to the rods of the Q1 and Q3 mass analyzers
RF voltage
+ dc voltage
RF voltage 180° out of phase
- dc voltage
The ac voltage applied to the quadrupole rods is of constant frequency (1.123 MHz). The
ac voltage applied to the rods varies from 0 to 10000 V P/P, and the dc voltage varies from
0 to ±840 V. Voltages of the same amplitude and sign are applied to each rod pair. However,
the voltages applied to the other rod pair are equal in amplitude but opposite in sign.
34
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
2
Functional Description
Mass Spectrometer
Because the frequency of this ac voltage is in the radio frequency range, it is referred to as rf
voltage. In Figure 23, the solid line represents the combined rf and dc voltage applied to one
rod pair, and the dashed line represents the combined rf and dc voltage applied to the other
rod pair. The ratio of rf voltage to dc voltage determines the ability of the mass spectrometer
to separate ions of different mass-to-charge ratios.
The first and third quadrupole rod assemblies (Q1 and Q3 quadrupoles) can act as mass
analyzers or as ion transmission devices. When both rf and dc voltages are applied, Q1 and
Q3 function as mass analyzers. When only rf voltage is applied, they act as ion transmission
devices. In the ion transmission mode, the quadrupole rod assemblies allow ions in a wide
window of mass-to-charge ratios to pass.
The square quadrupole rod assembly (Q2) operates in the ion transmission mode only.
Surrounding Q2 is a collision cell where collision-induced dissociation (CID) can take place if
the argon collision gas is present in the cell.
Figure 23. Magnitude of the rf and dc voltages applied to the rods of the Q1 and Q3 mass
analyzers
10,000 V P/P
840 V DC VOLTAGE
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
35
2
Functional Description
Mass Spectrometer
Mass Analysis
The Q1 and Q3 rods are charged with a variable ratio of rf voltage and dc voltage (Figure 23).
These potentials give rise to an electrostatic field that gives stable oscillations to ions with a
specific mass-to-charge ratio and unstable oscillations to all others.
At any given instant, one particular set of rf and dc voltage values is being applied to the mass
analyzer rods. Under these conditions, only ions of one mass-to-charge ratio (for example,
m/z 180) are maintained within bounded oscillations as their velocity carries them through
the mass analyzer. During this same time, all other ions undergo unbounded oscillations.
These ions strike one of the rod surfaces, become neutralized, and are pumped away, or they
are ejected from the rod assembly.
Then, at a later time, both rf and dc voltages change, and ions of the next mass-to-charge ratio
(for example, m/z 181) are allowed to pass, while all other ions (including m/z 180) become
unstable and undergo unbounded oscillations. This process continues, with ions of one
mass-to-charge ratio after another being transmitted, as the rf and dc voltages change in value.
At the end of the scan, the rf and dc voltages are discharged to zero, and the process is
repeated.
The potentials on the quadrupole rods can be changed rapidly and precisely. The rf and dc
voltages in the mass spectrometer can be scanned over the full mass range of the system,
m/z 10 to 3000, in 0.6 s.
The precision quadrupole rods of the mass spectrometer provide excellent sensitivity, peak
shape, resolution, and high mass transmission.
Collision Cell and CID Efficiency
In the MS/MS scan modes, the mass spectrometer applies a large voltage of opposite polarity
to the Q2 rod pairs between scans, which empties the collision cell. This process ensures that
no ions remain in the collision cell from scan to scan.
The collision cell quadrupole rod assembly (Q2), which always acts as an ion transmission
device, is a quadrupole array of square-profile rods. A variable rf voltage charges the rods,
which creates an electrostatic field that gives stable oscillations to ions in a wide window of
mass-to-charge ratios.
The collision cell surrounds Q2 and is usually pressurized from about 1 × 10-3 to 4 × 10-3 Torr
with argon collision gas. The collision cell is where collision-induced dissociation (CID) takes
place.
CID is a process in which an ion collides with a neutral atom or molecule and then, because
of the collision, dissociates into smaller fragments. The mechanism of dissociation involves
converting some of the translational kinetic energy (TKE) of the ion into internal energy. This
collision places the ion in an excited state. If the internal energy is sufficient, the ion
fragments.
36
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
2
Functional Description
Mass Spectrometer
Three expression convey the efficiency of the CID process:
• Collection efficiency
• Fragmentation efficiency
• Overall CID efficiency
Collection efficiency
The ion flux ratio measured at the exit of the collision cell and at its entrance. With no
collision gas present, the mass spectrometer obtains virtually 100 percent collection efficiency.
Collection efficiency is a mass-dependent parameter. For example, with mid-range collision
gas pressure, the collection efficiency might vary from about 50 percent for comparatively less
massive ions (which are more prone to scatter) up to 75 percent for comparatively more
massive ions (which are less prone to scatter).
Fragmentation efficiency
The fraction of the ion flux at the exit of the collision cell that results from fragmented ions.
Fragmentation efficiency depends directly on the stability of the ion and indirectly on the
mass of the ion. The more stable the ion, the less likely a given collision will fragment the ion.
Typically, low mass ions require a higher collision energy to dissociate than high mass ions. To
compensate for this, the data system allows you to specify a mass-dependent collision energy
ramp.
With a mid-range collision gas pressure, fragmentation efficiency might vary from 15 percent
to 65 percent for various compounds. As the collision gas pressure increases, the
fragmentation efficiency for all compounds approaches 100 percent due to multiple collisions.
The collection efficiency decreases, however, due to scattering.
Overall CID efficiency
The product of the collection efficiency and the fragmentation efficiency. The overall CID
efficiency exhibits a maximum with intermediate pressure. As the pressure is increased beyond
the optimum value, more and more collisions take place, the probability of scattering
increases, and fewer and fewer ions pass through the collision cell. This results in the
collection efficiency decreasing. The fragmentation efficiency also decreases as the pressure is
decreased from its optimum value, because fewer and fewer collisions take place.
Quadrupole Offset Voltage
The quadrupole offset voltage is a dc potential applied to the quadrupole rods in addition to
the ramping dc voltage. The offset voltage applied to the two rod pairs of the assemblies is
equal in magnitude and equal in sign. The quadrupole offset voltage accelerates or decelerates
ions and, therefore, sets theTKE of the ions as they enter the quadrupole rod assembly.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
37
2
Functional Description
Mass Spectrometer
In general, for a given experiment, the mass spectrometer has fixed offset voltages for Q1 and
Q2. However, in MS/MS experiments, the quadrupole offset voltage applied to Q3 usually
varies as a scan proceeds. The mass spectrometer automatically computes the Q3 quadrupole
offset voltage necessary and then varies the voltage, as appropriate, as each scan proceeds.
The offset voltage applied to Q2 (which contains the collision cell) is responsible for the
collision energy. The collision energy is the difference in potential between the ion source
(where parent ions are formed) and Q2 (where they collide with collision gas). As the offset
voltage on Q2 increases, the TKE of the parent ions also increases. As a result, increases in the
Q2 offset voltage increase the energy of ion/Ar collisions. The collision energy is usually
ramped downward with m/z to promote the fragmentation of low mass ions.
Before obtaining any mass spectra, the mass spectrometer tunes Q1 in the Q1MS scan mode
(Q2 and Q3 rf voltage only), and tunes Q3 in the Q3MS scan mode (Q1 and Q2 rf voltage
only). During tuning, the mass spectrometer determines the optimum quadrupole offset
voltage for Q1 and for Q3.
Mass Analyzer Lenses
The TSQ Quantum XLS and TSQ Quantum GC mass analyzers have three lens sets. See
Figure 21 on page 34. Those between Q1 and Q2 are designated L21, L22, L23; those
between Q2 and Q3 are designated L31, L32, L33; and the lens between Q3 and the ion
detection system is designated as L4 (or exit lens). All of the lenses have circular holes in their
centers through which the ion beam passes.
The lens assemblies also retain the three rod assemblies to ensure accurate and automatic axial
alignment of the rod assemblies.
The L2x lens set (between Q1 and Q2) and the L3x lens set (between Q2 and Q3) serve these
functions:
• To minimize the amount of collision gas that enters the mass analyzers (Q1 and Q3) from
the collision cell (Q2). Lenses L22 and L21 on one side of Q2 and lenses L32 and L33 on
the other side of Q2 act as baffles. (For high-mass transmission, it is important to
maintain a low pressure in the mass analyzers.)
• To retain the collision gas. Lenses L23 and L3 form two of the walls of the collision cell,
so they tend to hold the collision gas in the collision cell. The collision gas escapes,
however, through the same lens holes through which the ion beam passes.
• To shield Q1 from the rf voltage applied to Q2 and vice versa (L2x lens set) and to shield
Q3 from the rf voltage applied to Q2 and vice versa (L3x lens set).
• To focus the ion beam. The three lenses between Q1 and Q2 (and those between Q2 and
Q3) together form a three-element aperture lens. The first and third lenses are generally
set to similar or identical values and the central lens is set to a value different (either
higher or lower) from the other two.
38
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
2
Functional Description
Mass Spectrometer
The voltage applied to each of the lenses can vary from about -300 to +300 V. Typically,
however, the voltage applied to the first and third elements of the L2x lens set is somewhat
greater than the quadrupole offset voltage applied to Q1. Because the Q1 quadrupole offset
voltage is generally set to about ±5 V (depending on the charge of the ions of interest), the
voltage applied to lenses L21 and L23 is typically about -10 V for positive ions and +10 V for
negative ions. The voltage applied to the central lens of the L2x lens set is typically about
±225 V.
In the Q3MS scan mode, the voltage applied to the lenses of the L3x lens set is about the same
as that applied to the corresponding lens in the L2x lens set. Note, however, that in the
MS/MS scan modes, the voltage applied to the L3x lens set automatically varies with the
quadrupole offset voltage applied to Q3. As the Q3 quadrupole offset voltage ramps, the
voltages applied to the lenses ramp correspondingly.
Lens L4 is located between Q3 and the ion detection system. L4 is held at ground potential.
Its purpose is to shield Q3 from the high voltage applied to the ion detection system and to
shield the ion detection system from the high rf voltages applied to Q3.
Ion Detection System
The TSQ Quantum XLS and TSQ Quantum GC mass spectrometers are equipped with a
high-sensitivity, off-axis ion detection system. See Figure 24. This system produces a high
signal-to-noise ratio and allows for voltage polarity switching between positive ion and
negative ion modes of operation. The ion detection system includes a 15 kV conversion
dynode and a channel electron multiplier. The ion detection system is located at the rear of
the vacuum manifold behind the mass analyzer.
The conversion dynode is a concave metal surface located at a right angle to the ion beam.
The mass spectrometer applies to the conversion dynode a potential of +15 kV for negative
ion detection or -15 kV for positive ion detection. When an ion strikes the surface of the
conversion dynode, one or more secondary particles are produced. These secondary particles
can include positive ions, negative ions, electrons, and neutrals. When positive ions strike a
negatively charged conversion dynode, the secondary particles of interest are negative ions and
electrons. When negative ions strike a positively charged conversion dynode, the secondary
particles of interest are positive ions. The curved surface of the conversion dynode focuses
these secondary particles and a voltage gradient accelerates the particles into the electron
multiplier.
The electron multiplier includes a cathode and an anode. The cathode of the electron
multiplier is a lead-oxide, funnel-like resistor. The high voltage ring applies a potential of up
to -2.5 kV to the cathode. The exit end of the cathode (at the anode) is near ground potential.
The anode of the electron multiplier is a small cup located at the exit end of the cathode. The
anode collects the electrons produced by the cathode. The anode screws into the anode
feedthrough in the base plate.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
39
2
Functional Description
Mass Spectrometer
Figure 24. Ion detection system, showing the electron multiplier (top) and conversion dynode
(bottom)
Secondary particles from the conversion dynode strike the inner walls of the electron
multiplier cathode with sufficient energy to eject electrons. The ejected electrons are
accelerated farther into the cathode, drawn by the increasingly positive potential gradient.
The funnel shape of the cathode causes the ejected electrons not to travel far before they again
strike the inner surface of the cathode, which causes the emission of more electrons. A cascade
of electrons is then created that finally results in a measurable current at the end of the
cathode where the anode collects the electrons. The current collected by the anode is
proportional to the number of secondary particles striking the cathode.
Typically, the electron multiplier is set to a gain of about 3 × 105 (that is, for each ion or
electron that enters, 3 × 105 electrons exit) in MS mode and 2 × 106 in MS/MS mode. The
electrometer circuit converts the current that leaves the electron multiplier via the anode to a
voltage and the data system records the voltage.
The ion detection system of the mass spectrometer increases signal and decreases noise. The
high voltage applied to the conversion dynode results in a high conversion efficiency and
increased signal. That is, for each ion striking the conversion dynode, many secondary
particles are produced. The increase in conversion efficiency is more pronounced for more
massive ions than it is for less massive ions.
Because of the off-axis orientation of the ion detection system relative to the mass analyzer,
neutral molecules from the mass analyzer tend not to strike the conversion dynode or electron
multiplier. As a result, noise from neutral molecules is reduced.
40
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
2
Functional Description
Mass Spectrometer
Vacuum System and Inlet Gasses Hardware
The vacuum system evacuates the region around the ion source, ion optics, mass analyzer, and
ion detection system. The principal components of the vacuum system include the following:
• Vacuum Manifold
• Turbomolecular Pump
• Forepump
• Convectron Gauges
• Ion Gauges
The inlet gasses hardware controls the flow of collision gas, CI reagent gas, calibration
compound, and air (during venting) into the mass spectrometer. The inlet gasses hardware
includes the following components:
• Vent Valve
• Collision Gas Flow Control Valves
• Calibration Compound and CI Reagent Gas Flow Control
Figure shows a functional block diagram of the vacuum system and inlet gasses hardware.
Figure 25. Functional block diagram of the vacuum system and inlet gasses hardware
Calibration
compound
vial
Calibration
compound
valve
CI reagent
gas valve
CI reagent
gas port
Air
filter
Ion
gauge
Gas
chromatograph
Transfer line
Inlet
valve
Vent
valve
Ion source Q0 quadrupole
chamber
chamber
(4 x 10 - 5 Torr)
Analyzer
chamber
(3 x 10 - 6 Torr)
Dual-inlet
turbomolecular pump
Convectron
gauge
Exhaust
Collision
cell
Ion
gauge
Convectron
gauge
Collision gas
divert valve
Forepump
Collision
gas valve
Thermo Scientific
Argon
gas port
TSQ Quantum XLS and TSQ Quantum GC User Guide
41
2
Functional Description
Mass Spectrometer
Vacuum Manifold
The vacuum manifold encloses the ion source, ion optics, mass analyzer, and ion detection
system assemblies. The vacuum manifold is a thick-walled, aluminum chamber with two
removable side cover plates, with openings on the front, sides, and top, and various electrical
feedthroughs and gas inlets.
The main vacuum manifold is divided into two chambers by a baffle. See Figure 26. The high
vacuum port of the turbomolecular pump evacuates the region inside the first chamber, called
the analyzer region, to less than 10-5 Torr. The turbomolecular pump then discharges into the
forepump through the foreline.
The region inside the second chamber, called the Q0 quadrupole region, is evacuated to
1 mTorr by the interstage port of the turbomolecular vacuum pump.
A second vacuum manifold houses the ion source, which is open to the Q0 quadrupole
chamber. The interstage port of the turbomolecular pump evacuates the ion source region.
42
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
2
Functional Description
Mass Spectrometer
Figure 26. Vacuum manifold (interior)
O-ring
Ion gauge
Turbomolecular
pump
Baffle
Collision cell
chamber
Analyzer
chamber
Ion gauge
Q0 quadrupole
chamber
Ion source
chamber
When CID is turned on, the collision cell chamber, inside the analyzer chamber, has a user
controlled argon pressure of between 1 and 4 mTorr. The forepump evacuates the argon in the
collision cell when CID is turned off.
Two removable side cover plates on the left side of the vacuum manifold allow access to the
Q0 ion optics, mass analyzer, and ion detection system. Two electrically conductive O-rings
provide a vacuum-tight seal between the side cover plates and the vacuum manifold.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
43
2
Functional Description
Mass Spectrometer
Turbomolecular Pump
A Leybold TW220/150/15S double-inlet turbomolecular pump provides the vacuum for the
ion source, Q0 quadrupole region, and analyzer region of the vacuum manifold. The
turbomolecular pump mounts onto the top of the vacuum manifold (Figure 26).
The turbomolecular pump has two pumping inlets:
• A high-vacuum inlet at the top of the rotor stack, which evacuates the analyzer chamber
• An interstage inlet about halfway down the rotor stack, which evacuates the ion source
and Q0 quadrupole chambers
The turbomolecular pump is controlled by a Leybold TDS controller and powered by a
+24 V dc (250 W) power supply. The mass spectrometer circuit breaker switch and the
vacuum service switch, but not the electronics service switch, turn power for the
turbomolecular pump off and on. A fan draws air in from the front of the instrument cools
the pump.
Power to the turbomolecular pump shuts off if the foreline pressure, as measured by the
Convectron gauge, is too high, or if the turbomolecular temperature is too high.
Forepump
An Edwards E2M30 forepump (or roughing pump) establishes the vacuum necessary for the
proper operation of the turbomolecular pump. The forepump also evacuates the inlet valve
and the collision cell. The pump has a maximum displacement of 650 L/min and maintains a
minimum pressure of approximately 1 Pa (0.01 Torr).
The forepump is connected to the turbomolecular pump by a section of reinforced PVC
tubing. The power cord of the forepump is plugged into the outlet labeled Forepump on the
power panel (see Figure 12 on page 25). This outlet supplies power to the pump and is
controlled by the main power circuit breaker switch and vacuum service switch, not by the
electronics service switch.
CAUTION Always plug the forepump power cord into the outlets labeled Forepump on the
right side power panel of the mass spectrometer. Never plug it directly into a wall outlet.
44
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
2
Functional Description
Mass Spectrometer
Convectron Gauges
A Convectron™ gauge measures the pressure in the inlet valve and the foreline, which
connects the turbomolecular pump and the forepump. A second Convectron gauge measures
the pressure of argon collision gas in the collision cell.
The Convectron gauge uses a Wheatstone bridge with a temperature dependent resistor to
measure pressure down to a fraction of a milliTorr. The voltage present at the top of the bridge
depends on how fast the resistor can radiate heat, which is related to the pressure. The
pressure measured by the Convectron gauge is monitored by vacuum protection circuitry on
the Source PCB, which in turn is monitored by the embedded computer on the System
Control PCB. The vacuum protection circuitry detects whether the foreline pressure is too
high for the proper operation of the turbomolecular pump.
Ion Gauges
A Granville-Phillips® 342™ mini ion gauge measures the pressures in the analyzer region of
the vacuum manifold and the Q0/ion source region. The ion gauge produces energetic
electrons that cause the ionization of molecules in the ion gauge. Positive ions formed in the
ion gauge are attracted to a collector. The collector current is related to the pressure in the
vacuum manifold. The ion gauge is also involved in vacuum protection.
Vent Valve
The vent valve, located in the area of Q2, allows the vacuum manifold to be vented to air that
has been filtered through a sintered nylon filter. The vent valve is a solenoid-operated valve,
which is controlled by the Vent Delay PCB. When the solenoid is energized, the vent valve
closes. If the power fails or the main power circuit breaker is placed in the Off (O) position, a
4-farad capacitor located in the Power Entry Module keeps the solenoid in the closed position
for several minutes. If power is not restored in this time, then the solenoid opens and the
system is vented with filtered air. The vent valve closes after power is restored to the mass
spectrometer.
Collision Gas Flow Control Valves
The collision gas flow control valves control the flow of argon collision gas into and out of the
Q2 collision cell. A solenoid valve acts to shut off argon gas flow to the cell. The collision gas
pressure is regulated by a proportional valve that is controlled by the data system. You can set
the collision gas pressure (0 to 5 milliTorr) in the Define Scan dialog box of the EZ Tune
window.
Ions enter the Q2 collision cell, collide with the argon collision gas, and then, because of the
collision, dissociate into smaller fragments. See “Collision Cell and CID Efficiency” on
page 36.
Argon enters the mass spectrometer through a 1/8 in. port on the left side of the mass
spectrometer.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
45
2
Functional Description
Mass Spectrometer
A second proportional valve allows the forepump to evacuate the Q2 collision cell of argon
and waste gasses when CID is turned off. The mass spectrometer automatically opens and
closes the collision gas evacuation valve, depending on whether Q2 is acting as an ion
transmission device or a collision cell.
Calibration Compound and CI Reagent Gas Flow Control
The calibration compound gas flow control valve controls the flow of calibration compound
gas into the ion source, via the transfer line. The calibration gas flow can be either high or low.
Normally low flow is used, but high flow might be necessary for negative ion CI calibration.
The CI gas flow control valve controls the flow of chemical ionization reagent gas into the ion
source, via the transfer line. The maximum CI gas flow rate is about 4 mL/min, depending on
the gas.
Electronic Assemblies
The electronic assemblies that control the operation of the mass spectrometer are distributed
among various PCBs and other modules located in the tower, in the embedded computer, and
on or around the vacuum manifold of the mass spectrometer.
The Power Entry Module provides mass spectrometer power control, a contact closure
interface, vent valve control, an Ethernet 100 base-T connection from the System Control
PCB to the data system PC, a mechanical pump failure protection circuit (linked to vent valve
control), a system reset button, status LEDs, and service ports. The right-side power panel,
shown in Figure 12 on page 25, is part of the Power Entry Module.
The Power Entry Module accepts line power, filters it, and provides it to various components
of the mass spectrometer. The Power Module includes the following components:
• Main power circuit breaker switch
• Surge suppressor
• Line filter
• Electronics service switch
• Vacuum service switch
The “brains” of the TSQ Quantum XLS and TSQ Quantum GC mass spectrometers are the
System Control PCB. The System Control PCB and embedded computer include the
following:
• PowerPC processor
• Serial Peripheral Interconnect (SPI) bus
• I/O coprocessor
• Super Harvard Architecture Computer (Sharc) bus
46
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
2
Functional Description
Mass Spectrometer
• Scan generator DSP
• Acquisition processor DSP
• Interbus bridge
• 100 base-T Ethernet port
The rf voltage generation electronic assemblies produce the Q0, Q1, Q2, and Q3 rf voltages
that enable ion transmission and mass analysis. All rf voltages are controlled by the Analyzer
Control PCB and the System Control PCB.
The Q1 and Q3 rf voltage amplifier circuits are identical, and the circuits for Q0 and Q2 are
similar.
The rf voltage generation electronic assemblies include the following components:
• rf oscillator
• rf Voltage Amplifier PCB
• Low Pass Filter PCB
• rf voltage coil
• rf voltage detector
• Mass DAC
• Integrating amplifier
The ion detection system electronic assemblies provide high voltage to the electron multiplier
and conversion dynode of the ion detection system. They also receive the electron multiplier
output current signal, convert it to a voltage (by the electrometer circuit), and pass it to the
embedded computer.
The ion detection system electronic assemblies include the following:
• Electron multiplier power supply
• Conversion dynode power supply
• Electrometer PCB
• Acquisition DSP
The Analyzer Control PCB contains circuitry for controlling and monitoring the operation of
the ion source, ion optics, mass analyzer, and ion detection system. These circuits are in turn
monitored by the PowerPC processor of the System Control PCB via the SPI bus.
The Analyzer Control PCB controls and monitors the rf voltages for Q0, Q1, Q2, and Q3
quadrupoles. It also has lens voltage drivers that convert ±330 V dc power from the DC Rod
Driver PCB to dc voltages that are applied to the lenses
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
47
2
Functional Description
Data System
Data System
The data system controls and monitors the TSQ Quantum XLS and TSQ Quantum GC
mass spectrometers. The data system also processes data that the mass spectrometer acquires.
The data system is composed of the following:
• Computer Hardware
• Data System / Mass Spectrometer / GC Interface
• Data System / Local Area Network Interface
Computer Hardware
The data system computer has the following major features:
• Intel® Pentium® IV processor
• High capacity hard disk drive
• Recordable/rewriteable CD drive
• Primary Ethernet port (data system to mass spectrometer)
• Secondary Ethernet port (data system to local area network)
• High performance video graphics card
• CDRW drive
• DVD drive
• 1280×1024 resolution color monitor
• Keyboard and mouse
For more information about the computer, refer to the appropriate manuals.
Data System / Mass Spectrometer / GC Interface
The data system computer contains a 100 base-T Ethernet adapter (called the primary
Ethernet adapter) that is dedicated to data system/mass spectrometer/GC communications.
This primary Ethernet adapter communicates with the mass spectrometer and GC modules
via a 10/100 base-T Ethernet switch. The Ethernet adapter on the mass spectrometer resides
on the System Control PCB. A twisted pair, Ethernet cable connects the primary Ethernet
adapter of the data system to the Ethernet switch, which is connected to the Ethernet
connector on the power panel of the mass spectrometer and to the GC.
48
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
2
Functional Description
Data System
Data System / Local Area Network Interface
The data system computer contains a secondary Ethernet adapter. This secondary Ethernet
adapter is not involved in data system/mass spectrometer or GC communications. You can
use this secondary Ethernet adapter to access your local area network.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
49
3
System Shutdown, Startup, and Reset
Many maintenance procedures for the TSQ Quantum XLS and TSQ Quantum GC systems
require that the mass spectrometer be shut down completely. In addition, you can place the
mass spectrometer in Standby mode if the system is not to be used for 12 hours or more.
Contents
• Shutting Down the System in an Emergency
• Placing the System in Standby Mode
• Shutting Down the System Completely
• Starting Up the System after a Complete Shutdown
• Resetting the Mass Spectrometer
• Resetting the Data System
• Turning Off Selected Mass Spectrometer Components
Shutting Down the System in an Emergency
You can turn off all power to the mass spectrometer, gas chromatograph, and autosampler by
pressing the System Power Off button, located on the front of the electronics module, or the
main power circuit breaker, located on the rear of the electronics module. You can turn on the
system power only with the main power circuit breaker on the rear of the electronics module
and not the System Power Off switch.
Y To turn off the mass spectrometer in an emergency
CAUTION Press the System Power Off button located on the front of the electronics
module. See Figure 27. The System Power Off button turns off all power to the mass
spectrometer (including the vacuum pumps), gas chromatograph, and autosampler.
Although removing power abruptly does no harm to any component within the
system, under normal conditions do not shut down the system with the System Power
Off button. For the recommended procedure, see “Shutting Down the System
Completely” on page 53.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
51
3
System Shutdown, Startup, and Reset
Placing the System in Standby Mode
Figure 27. System Power Off button on the front of the electronics module
System Power
Off button
Figure 28. Power panel on the rear of the electronics module
Power In
GC Power Out
Quantum Power Out
A/S Power Out
!
!
V~230
Hz, 30A MAX
V~230
50/60 Hz, 16A MAX
V~230
50/60 Hz, 10A MAX
V~230
50/60 Hz, 2A MAX
Placing the System in Standby Mode
If you are not going to use the TSQ Quantum XLS or TSQ Quantum GC system for a short
period of time, such as overnight or over weekends, it does not need to be shut down
completely. When you are not going to operate the system for 12 hours or more, you can leave
the system in standby mode.
Y To place the TSQ Quantum XLS or TSQ Quantum GC system in standby mode
1. Wait until data acquisition, if any, is complete.
On
52
Off
Standby
2. From the TSQ EZ Tune window, choose Control > Standby (or click the On/Standby
button) to put the mass spectrometer in standby.
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
3
System Shutdown, Startup, and Reset
Shutting Down the System Completely
When you choose Control > Standby, the TSQ Quantum XLS or TSQ Quantum GC
system turns off the electron multiplier, the conversion dynode, the ion source filament
and lenses (but not the heater), and the mass analyzer and ion optics rf voltages. See
Table 4 on page 60 for the On/Off status of mass spectrometer components when the
mass spectrometer is in standby mode. The System LED on the front panel of the mass
spectrometer illuminates yellow when the system is in standby.
3. Leave the mass spectrometer power on.
4. Leave the GC power on with column flow.
5. Leave the autosampler power on.
6. Leave the data system power on.
Shutting Down the System Completely
Shut down the TSQ Quantum XLS or TSQ Quantum GC system completely only if it is to
be unused for an extended period or if it must be shut down for a maintenance or service
procedure. For a short period of time, such as overnight or over weekends, you can place the
system in standby mode. See “Placing the System in Standby Mode.”
Y To shut down the TSQ Quantum XLS or TSQ Quantum GC system completely
1. Cool the GC, transfer line, and ion source:
• GC oven 30 oC
• Injector off
• Transfer line off
• Ion source 30 oC
Note If you do not plan to change the column or perform maintenance on the gas
chromatograph, you do not have to lower the injector temperature.
2. From the TSQ EZ Tune window, choose Control > Standby (or click the On/Standby
button) to put the mass spectrometer in standby.
3. Place the electronics service switch, located on the right-side power panel, in the Service
Mode.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
53
3
System Shutdown, Startup, and Reset
Shutting Down the System Completely
Figure 29. Right-side power panel of the mass spectrometer
Pum p O n
Electronics
Vacuum
Qualified
Service
Personnel
Only
!
4. Place the vacuum service switch, located on the right-side power panel, in the Service
Mode position.
5. Place the mass spectrometer main power circuit breaker switch, located on the right-side
power panel, in the Off position. When you place the main power circuit breaker switch
in the Off position, the following occurs:
• All power to the mass spectrometer is turned off. (All LEDs on the front panel of the
mass spectrometer are off.)
• A capacitor on the Vent Delay PCB provides power to the vent valve for two to four
minutes (to allow the turbomolecular pump to spin down). After the capacitor
discharges, power to the vent valve solenoid shuts off. When power to the vent valve
solenoid shuts off, the vent valve opens and the vacuum manifold vents to filtered air.
You can hear a hissing sound as the air passes through the air filter.
• After about two minutes, the pressure of the vacuum manifold reaches atmospheric
pressure.
6. Unplug the power cord for the mass spectrometer.
CAUTION Allow heated components to cool before servicing them.
54
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
3 System Shutdown, Startup, and Reset
Starting Up the System after a Complete Shutdown
Note If you plan to perform routine or preventive system maintenance on the mass
spectrometer only, you can leave the argon, data system, GC, and autosampler on. In
this case, the shutdown procedure is complete. However, if you do not plan to operate
your system for an extended period of time, Thermo Fisher Scientific recommends
that you turn off the GC, data system, and autosampler as described in steps 7
through 12 below.
7. Turn off the GC. Follow the procedure described in the manual that came with the GC.
8. Turn off the autosampler by using the main power on/off switch.
9. Press the System Power Off button (Figure 27) to remove all power from the TSQ
Quantum XLS or TSQ Quantum GC system.
10. If the system is not to be used for an extended period, turn off the argon collision gas
supply at the tank.
11. Turn off the data system:
a. Choose Start > Shut Down from the Windows® task bar. The Shut Down Windows
dialog box appears.
b. To start the Windows shutdown procedure, select Shut down and click OK.
12. Turn off the (optional) printer by using the on/off switch.
CAUTION Burn Hazard. The ion source, GC oven, injector, and transfer line are hot.
Allow them to cool to room temperature before touching them.
Starting Up the System after a Complete Shutdown
Starting the TSQ Quantum XLS or TSQ Quantum GC system after it has been shut down
completely requires the following:
• Restoring Power to the TSQ Quantum XLS or TSQ Quantum GC system
• Starting Up the GC
• Starting Up the Data System
• Starting Up the Mass Spectrometer
• Starting Up the Autosampler (autosampler is optional)
• Setting Up Conditions for Operation
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
55
3
System Shutdown, Startup, and Reset
Starting Up the System after a Complete Shutdown
Restoring Power to the TSQ Quantum XLS or TSQ Quantum GC system
Y To restore power to the TSQ Quantum XLS or TSQ Quantum GC system
Place the main power circuit breaker, located on the rear of the electronics module
(Figure 28), in the On position.
Starting Up the GC
Y To start the GC
Follow the startup procedure described in the manual that came with the GC. Verify that the
gas chromatograph is on and there is carrier gas flowing through the column into the mass
spectrometer.
Starting Up the Data System
Y To start the data system
1. Turn on the monitor, computer, and printer.
2. Observe the Windows startup procedure on the monitor and press
CONTROL+ALT+DELETE when you are prompted to do so. To complete the startup
procedure, click OK or enter your password (if you have one) in the Logon Information
dialog box.
Starting Up the Mass Spectrometer
Y To start the mass spectrometer
CAUTION If you turn on the mass spectrometer without column flow, air can damage
the GC column. This large air leak into the TSQ Quantum GC also causes the ion
source to require cleaning.
1. Turn on the flow of argon at the tank if it is off.
IMPORTANT The data system must be running before you start the mass
spectrometer. The mass spectrometer will not operate until it receives software from
the data system.
2. Make sure that the mass spectrometer main power circuit breaker switch, located on the
right side power panel (Figure 29), is in the Off (O) position and the electronics service
switch and the vacuum service switch are both in the Service Mode position.
56
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
3 System Shutdown, Startup, and Reset
Starting Up the System after a Complete Shutdown
3. Plug in the power cord for the mass spectrometer.
4. Place the mass spectrometer main power circuit breaker switch in the On (|) position.
When you place the main power circuit breaker switch in the On (|) position, power is
supplied to those mass spectrometer components that are not affected by the vacuum
service switch and the electronics service switch.
5. Place the vacuum service switch in the Operational position.
6. Place the electronics service switch in the Operational position. When you place the
electronics service switch in the Operational position, the following occurs:
• The Power LED on the mass spectrometer front panel illuminates green to indicate
that power is provided to the mass spectrometer electronics. (The electron multiplier,
conversion dynode, ion source, and mass analyzer and ion optics rf voltages remain
off.)
• The embedded computer reboots. After several seconds the Communication LED on
the front panel illuminates yellow to indicate that the data system and the mass
spectrometer have started to establish a communication link.
• After several more seconds, the Communication LED illuminates green to indicate
that the data system and the mass spectrometer have established a communication
link. Software for the operation of the mass spectrometer is then transferred from the
data system to the mass spectrometer.
• After three minutes, the System LED illuminates yellow to indicate that the software
transfer from the data system to the mass spectrometer is complete and that the
instrument is in standby.
• The Vacuum LED on the front panel of the mass spectrometer remains off until the
turbomolecular pump reaches 80 percent of its operational speed of 750 Hz. At this
time the ion gauge is turned on and the Vacuum LED illuminates yellow. The
Vacuum LED illuminates green, and the high voltage can be turned on, only if the
pressure in the mass analyzer region of the vacuum manifold is less than the
appropriate value listed in Table 3.
Table 3.Maximum allowed pressure to turn on high voltage
Thermo Scientific
Carrier gas
Ar collision gas
Maximum pressure (Torr)
He
Off
6 × 10-6
He
On
5 × 10-5
H2
Off
5 × 10-5
H2
On
1 × 10-4
TSQ Quantum XLS and TSQ Quantum GC User Guide
57
3
System Shutdown, Startup, and Reset
Starting Up the System after a Complete Shutdown
Figure 30. Front panel LEDs of the mass spectrometer
Power
Vacuum
System
Communication
Scan
If you have an autosampler, go to Starting Up the Autosampler. If you do not have an
autosampler, go to Setting Up Conditions for Operation.
Starting Up the Autosampler
Y To start the autosampler
Place the main power switch on the autosampler in the On position. If necessary,
configure the autosampler. For procedures for placing sample vials, preparing solvent and
waste bottles, installing syringes, and so on, refer to the manual that came with the
autosampler.
Setting Up Conditions for Operation
Y To set up your TSQ Quantum XLS or TSQ Quantum GC mass spectrometer for operation
1. Before you begin data acquisition with your TSQ Quantum XLS system, allow the
system to pump down for at least one hour. Operation of the system with excessive air
and water in the vacuum manifold can cause reduced sensitivity, tuning problems, and
reduced lifetime of the electron multiplier.
2. Ensure that the argon pressure is within the operational limits [argon: 135 ±70 kPa
(20 ±10 psig)].
Note Air in the argon line must be purged or given sufficient time to be purged for
normal TSQ Quantum XLS mass spectrometer performance.
3. In the TSQ EZ Tune window, choose View > Instrument Status. Look at the Status
view. Check that the pressure measured by the ion gauge is below about 3 × 10-6 Torr
(2 × 10-5 Torr with H2 carrier gas) with the collision gas turned off.
Note You do not need to calibrate and tune the mass spectrometer each time you
restart the TSQ Quantum XLS or TSQ Quantum GC system. To tune and calibrate
the mass spectrometer, see Chapter 4, “Tuning and Calibrating.”
58
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
3
System Shutdown, Startup, and Reset
Resetting the Mass Spectrometer
Resetting the Mass Spectrometer
If communication between the mass spectrometer and data system computer is lost, it might
be necessary to reset the mass spectrometer using the System Reset button on the right-side
power panel. Pressing the System Reset button creates an interrupt in the embedded
computer. This causes the embedded computer to restart in a known (default) state. See
Figure 29 on page 54 for the location of the System Reset button.
The procedure given here assumes that the mass spectrometer and data system computer are
both powered on and operational. If the mass spectrometer, data system computer, or both are
off, go to “Starting Up the System after a Complete Shutdown” on page 55.
Y To reset the mass spectrometer
Press the System Reset button located on the right-side power panel. Make sure the
Communication LED is extinguished before releasing the System Reset button. When you
press the Reset button, the following occurs:
• An interrupt on the embedded computer causes the CPU to reboot. All LEDs on the
front panel of the mass spectrometer are off except the Power LED.
• After several seconds, the Communication LED illuminates yellow to indicate that the
data system and the mass spectrometer are starting to establish a communication link.
• After several more seconds, the Communication LED illuminates green to indicate that
the data system and the mass spectrometer have established a communication link.
Software for the operation of the mass spectrometer is then transferred from the data
system to the mass spectrometer.
• After three minutes the software transfer is complete. The System LED illuminates either
green to indicate that the instrument is functional and the high voltages are on, or yellow
to indicate that the instrument is functional, and it is in standby.
Resetting the Data System
If possible, use the Windows shutdown and restart procedure to shut down and restart the
data system so that Windows can properly close programs and save changes to files.
Y To reset the data system by using the Windows shutdown and restart procedure
1. Choose Start > Shut Down from the Windows task bar. The Shut Down Windows
dialog box appears.
2. Select Restart and click OK to start the Windows shutdown and restart procedure.
3. Observe the Windows shutdown and restart procedure on the monitor. Press
CTRL+ALT+DELETE when you are prompted to do so. To complete the shutdown and
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
59
3
System Shutdown, Startup, and Reset
Turning Off Selected Mass Spectrometer Components
restart procedure, click OK or enter your password (if you have one) in the Logon
Information dialog box.
Note The communications link between the data system and the mass spectrometer
should automatically reestablish after you reset the data system. When this occurs the
Communication LED on the front panel of the mass spectrometer illuminates yellow and
then green. If the system is unable to reestablish the communications link, press the
System Reset button on the power panel of the mass spectrometer.
Turning Off Selected Mass Spectrometer Components
There are different ways that you can turn off some or all of the mass spectrometer
components:
• Turn off individual mass spectrometer components from the TSQ EZ Tune window.
Turning off individual mass spectrometer components might be necessary when you are
troubleshooting or when you are running certain diagnostic procedures.
• Place the mass spectrometer in Standby mode. Standby is the normal condition to leave
the mass spectrometer in when it is not in use. Choose Control > Standby (or toggle the
On/Standby button) from the TSQ EZ Tune window to place the mass spectrometer in
standby.
• Place the mass spectrometer in the Off condition. The Off condition is similar to
Standby, except all high voltage components of the mass spectrometer are turned off.
Choose Control > Off from the TSQ EZ Tune window to place the mass spectrometer in
the Off condition.
• Place the electronics service switch in the Service Mode position. The electronics service
switch turns off all components in the mass spectrometer other than the +24 V power
supply, forepump, turbomolecular pump, Vent Delay PCB, and fans.
• Place the vacuum service switch in the Service Mode position. The vacuum service switch
turns off all vacuum system components, including the +24 V power supply, forepump,
turbomolecular pump, Vent Delay PCB, and fans.
• Place the right-side power panel circuit breaker switch in the Off position. Placing the
right-side power panel circuit breaker switch in the Off position removes all power to the
mass spectrometer, including the vacuum system.
• Press the System Power Off button. Pressing the System Power Off button removes all
power to the mass spectrometer, gas chromatograph, and autosampler.
Table 4 summarizes the on/off status of mass spectrometer components, voltages, and gas
flows.
60
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
3 System Shutdown, Startup, and Reset
Turning Off Selected Mass Spectrometer Components
Table 4. On/Off status of mass spectrometer components and voltages (Sheet 1 of 2)
Standby
Off
Mass spectrometer component
Electronics
service switch in
Service Mode
position
Vacuum service
switch in
Service Mode
position
MS Main power
circuit breaker
switch in Off
position
Electron multiplier
Off
Off
Off
Off
Off
Conversion dynode
Off
Off
Off
Off
Off
Mass analyzer rf voltage
Off
Off
Off
Off
Off
Mass analyzer dc offset voltage
Off
Off
Off
Off
Off
Q0 ion optics rf voltage
Off
Off
Off
Off
Off
Q0 ion optics dc offset voltage
Off
Off
Off
Off
Off
Ion source filament
Off
Off
Off
Off
Off
Ion source heater
On
On
Off
Off
Off
Ion source lenses
Off
Off
Off
Off
Off
Argon collision gas
Off
Off
Off
Off
Off
Vent valve
Closed
Closed
Closed
Open
(after 2 to 4 min)
Open
(after 2 to 4 min)
Turbomolecular pump
On
On
On
Off
Off
Forepump
On
On
On
Off
Off
Vent Delay PCB
On
On
On
On
Off
Embedded computer
On
On
Off
On
Off
Turbomolecular pump controller On
On
On
Off
Off
Power supply, electron multiplier Off
and conversion dynode
Off
Off
On
Off
Power supply, 8 kV
Off
Off
Off
On
Off
PS1 power supply, +24 V
On
On
On
Off
Off
PS2 power supply, +5, ±15,
±24 V dc
On
On
Off
Off
Off
PS3 power supply, +36,
-28 V dc
On
On
Off
Off
Off
Fan, turbomolecular pump
On
On
On
Off
Off
Fan, above manifold
On
On
On
Off
Off
Fan, center wall
On
On
On
Off
Off
Convectron gauge, foreline
On
On
On
On
Off
Convectron gauge, collision cell
On
On
Off
On
Off
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
61
Table 4. On/Off status of mass spectrometer components and voltages (Sheet 2 of 2)
Standby
Off
Mass spectrometer component
Ion gauge
On
On
Electronics
service switch in
Service Mode
position
Vacuum service
switch in
Service Mode
position
MS Main power
circuit breaker
switch in Off
position
Off
Off
Off
4
Tuning and Calibrating
Tune parameters are instrument parameters that affect the intensity of the ion signal.
Calibration parameters are instrument parameters that affect the mass accuracy and resolution
of the mass spectrum. Tuning and calibrating the TSQ Quantum XLS or TSQ Quantum GC
mass spectrometer automatically with TSQ EZ Tune uses perfluorotributylamine (FC-43) as
the tuning and calibration compound. FC-43 resides in a vial inside the mass spectrometer.
Contents
• Displaying the FC-43 Mass Spectrum
• Running Auto Tune and Calibration
• Saving the Tune and Calibration Report
Record the ion signal intensity of FC-43 (see Figure 32) just after you tune and calibrate the
mass spectrometer. Periodically check the FC-43 ion signal. Re-tune and calibrate the mass
spectrometer if the FC-43 ion signal falls below 50% of the tuned value. Also, re-tune and
calibrate after switching between EI and CI modes or between positive and negative polarity
modes.
Note You may need to calibrate more often if you operate in H-SRM mode (as opposed
to SRM mode only).
Table 5 lists typical peaks that are observed in the FC-43 mass spectrum in EI, positive
polarity mode.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
63
4
Tuning and Calibrating
Displaying the FC-43 Mass Spectrum
Table 5. Typically observed FC-43 peaks in EI, positive polarity mode
m/z 69
m/z 100
m/z 114
m/z 119
m/z 131
m/z 169
m/z 219
m/z 264
m/z 314
m/z 352
m/z 402
m/z 414
m/z 464
m/z 502
m/z 614
Displaying the FC-43 Mass Spectrum
The first step in tuning and calibrating the mass spectrometer is to display the FC-43 mass
spectrum.
Y To display the FC-43 mass spectrum
1. Choose Start > All Programs > Thermo Instruments > TSQ > TSQ Tune to open the
EZ Tune window.
2. Choose a polarity mode. In this example the mass spectrometer is in positive polarity
mode. If necessary, toggle the polarity button.
Positive Negative
Note For EI negative polarity mode, you might have to set the calibration gas flow to
high.
3. Click the Define Scan button to display the Define Scan dialog box.
4. In the Define Scan dialog box, select or enter the following settings:
• Scan type: Full Scan
• Scan mode: Q1MS
• Scan range entry mode: FM/LM
64
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
4 Tuning and Calibrating
Displaying the FC-43 Mass Spectrum
• First mass: 50
• Last mass: 650
• Scan time: 0.500
Figure 31. Define Scan dialog box
5. Click Apply to apply the settings.
6. Click the Calibration Gas button in the EZ Tune toolbar to turn on the flow of
calibration gas into the ion source.
.......... ..Cal Gas Off .........On
7. Click the Filament button in the EZ Tune toolbar to turn on the filament.
.............Filament Off. . ....On
8. Choose Control > On to start the mass spectrometer scanning.
9. Observe the mass spectrum. If your mass spectrum looks very different from the one in
Figure 32, see “Diagnostics and Troubleshooting” on page 129.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
65
4
Tuning and Calibrating
Running Auto Tune and Calibration
Figure 32. EZ Tune - EZ Workspace, showing the FC-43 mass spectrum in EI, positive polarity mode before tuning and
calibrating.
Running Auto Tune and Calibration
Run the tune and calibration procedure after you obtain a good FC-43 ion signal.
Y To tune and calibrate your mass spectrometer automatically in the EI, positive ion
mode
1. Display the FC-43 positive ion mass spectrum as described in the previous section.
2. Click the System Tune and Calibration button (or choose Setup > System Tune and
Calibration) to display the System Tune and Calibration dialog box.
66
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
4 Tuning and Calibrating
Running Auto Tune and Calibration
Figure 33. System Tune and Calibration dialog box
3. In the Compound list, select FC43 pos. ions. This automatically selects the positively
charged FC-43 ions to be used for automatic tuning and calibrating.
4. Select Tune A as the tune algorithm.
Note Tune A is the first choice of tune algorithm if there is a regular EI or CI ion
volume installed in the ion source. Tune B is the first choice of tune algorithm if there
is a closed EI ion volume installed.
5. Click Start to start the automatic tuning and calibration procedure.
The Status box displays real-time messages about the system tune and calibration so that
you can monitor the progress of each sub-procedure. After a sub-procedure is complete,
the result is reported (for example, whether it passed or failed). At the end of the entire
procedure, it displays a summary.
• If errors occur during the automatic tuning and calibration procedure, go to step 6.
• If the automatic tuning and calibration procedure finishes without errors, go
to step 7.
6. If errors occur during the automatic tuning and calibration procedure, restore the
previous mass spectrometer device settings and perform the tuning and calibration
procedure again by completing the following steps:
a. To restore the prior tuning and calibration settings, click Undo.
b. To reload the prior tuning and calibration settings to the mass spectrometer, click
Accept.
c. Troubleshoot and correct the problem that caused the tuning and calibration
procedure to fail. See “Tuning Issues” on page 143.
d. Go to step 5 and restart the tuning and calibration procedure.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
67
4
Tuning and Calibrating
Running Auto Tune and Calibration
Figure 34. Typical parameter optimization plot
7. Click Accept to accept the results of the tuning and calibration procedure.
After you accept the results of the tuning and calibration procedure, a message box asks
whether you want to copy the positive ion tuning and calibration settings to the negative
ion mode.
• If you have already tuned and calibrated the instrument successfully in the negative
ion mode, click No. (Do not copy the positive ion mode parameters to the negative
ion mode.)
• If you have not tuned and calibrated the instrument in the negative ion mode,
click Yes.
Note If you intend to perform high-sensitivity, negative-ion mass spectrum
analysis, Thermo Fisher Scientific recommends that you also perform a full tune
and calibration of the instrument in the negative ion mode.
8. Save the calibration file as follows:
a. Click Save Calib. As to open the Save Calibration File dialog box.
b. In the File Name box, enter a name for your calibration file.
68
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
4 Tuning and Calibrating
Saving the Tune and Calibration Report
c. Click Save to save the calibration file. The Save As dialog box appears.
9. Save the tune method file as follows:
a. In the File Name box, enter a name for your tune method file.
b. Click Save to save the tune method file.
The mass spectrometer is now tuned and calibrated in the positive ion mode.
Saving the Tune and Calibration Report
You can save the information that appears in the Status box and in the parameter optimization
plots as a PDF file. Examples of status information and parameter optimization plots from the
tune and calibration report follow.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
69
4
Tuning and Calibrating
Saving the Tune and Calibration Report
Figure 35. Tune and calibration report, status page
Quantum Tune Master - Tune and Calibration
Results from System Tune and Calibration using FC43 Pos. Ions:
Q1 and Q3 Tuning & Calibrating
10:55:29:
10:55:39:
10:55:41:
10:55:42:
10:56:02:
10:56:02:
10:56:07:
10:56:08:
10:56:12:
10:56:14:
10:56:20:
10:56:20:
10:56:20:
10:56:20:
10:56:26:
10:56:26:
10:56:26:
10:56:26:
10:56:32:
10:56:32:
10:56:32:
10:56:32:
10:56:38:
10:56:38:
10:56:38:
10:56:38:
10:56:45:
10:56:45:
10:56:45:
10:56:45:
10:56:47:
10:56:48:
10:57:04:
10:57:04:
10:57:10:
10:57:11:
10:57:16:
10:57:18:
10:57:24:
10:57:24:
10:57:24:
10:57:24:
10:57:30:
10:57:30:
10:57:30:
10:57:30:
Tuning and Calibrating Q1
Tuning Q1MS at mass 69.00 m/z
Opening resolution
Adjusting resolution
The adjusted resolution is -13.33 at the width 0.78
Coarse resolution done - Quick calibration
The adjusted calibration of mass 69.0 is -0.88
Averaging Stopping Curve...
Q1 Quadrupole Offset set to -0.85 for all reswidths
Optimizing Lens 2 for ion 69.00 m/z
Previous Setting = -7.00, New Setting = -0.80
Maximum Intensity = 1.04e+06
322 % Improvement
Optimizing Lens 4 for ion 69.00 m/z
Previous Setting = -10.00, New Setting = -46.55
Maximum Intensity = 1.61e+06
60 % Improvement
Optimizing Lens 1-1 for ion 69.00 m/z
Previous Setting = -5.00, New Setting = -0.86
Maximum Intensity = 1.93e+06
15 % Improvement
Optimizing Lens 1-2 for ion 69.00 m/z
Previous Setting = -21.90, New Setting = -5.00
Maximum Intensity = 2.43e+06
41 % Improvement
Optimizing Lens 2-1 for ion 69.00 m/z
Previous Setting = -8.45, New Setting = 3.92
Maximum Intensity = 2.56e+06
17 % Improvement
Tuning Q1MS at mass 501.97 m/s
Opening resolution
Adjusting resolution
The adjusted resolution is -14.38 at the width 0.78
Coarse resolution done - Quick calibration
The adjusted calibration of mass 502.0 is -0.16
Averaging Stopping Curve...
Q1 Quadrupole Offset set to -2.00 for all reswidths
Optimizing Lens 2 for ion 501.97 m/z
Previous Setting = -7.00, New Setting = -2.70
Maximum Intensity = 2.97e+05
223 % Improvement
Optimizing Lens 1-2 for ion 501.97 m/z
Previous Setting = -50.14, New Setting = -51.11
Maximum Intensity = 4.35e+05
6 % Improvement
Optimizing Lens 2-1 for ion 501.97 m/z
Signature: __________________________________________
May 09, 2007 11:17:28
70
TSQ Quantum XLS and TSQ Quantum GC User Guide
Page 1 of 10
Thermo Scientific
4 Tuning and Calibrating
Saving the Tune and Calibration Report
Figure 36. Tune and calibration report, parameter optimization plots
Quantum Tune Master - Tune and Calibration
Optimizing Lens 2 for Q1MS
80
60
40
20
Mass 501.97m/z : 223 % Improvement
Mass 218.99m/z : 94 % Improvement
Mass 69.00m/z : 322 % Improvement
0
-50
-40
-30
-20
Lens 2 Voltage (V)
Previous Setting
Optimum Setting
100
R e la t iv e Int e ns it y
R e la t iv e Int e ns it y
Optimizing Lens 2 for Q3MS
Previous Setting
Optimum Setting
100
80
60
40
20
Mass 501.97m/z : 281 % Improvement
Mass 218.99m/z : 57 % Improvement
Mass 69.00m/z : 235 % Improvement
0
-10
0
-50
Optimizing Lens 4 for Q1MS
60
40
20
Mass 69.00m/z : 60 % Improvement
-50
-40
-30
-20
Lens 4 Voltage (V)
60
40
20
Mass 69.00m/z : 15 % Improvement
0
-10
0
-10
40
20
Mass 69.00m/z : 6 % Improvement
-6
-4
Lens 1-1 Voltage (V)
-2
0
May 9, 2007
Previous Setting
Optimum Setting
100
R e la t iv e Int e ns it y
R e la t iv e Int e ns it y
60
-8
-6
-4
Lens 1-1 Voltage (V)
Optimizing Lens 1-2 for Q1MS
80
-10
-8
May 9, 2007
Previous Setting
Optimum Setting
0
0
May 9, 2007
80
Optimizing Lens 1-1 for Q3MS
100
-10
Previous Setting
Optimum Setting
100
R e la t iv e Int e ns it y
R e la t iv e Int e ns it y
80
0
-30
-20
Lens 2 Voltage (V)
Optimizing Lens 1-1 for Q1MS
Previous Setting
Optimum Setting
100
-40
May 9, 2007
80
60
40
20
Mass 501.97m/z : 6 % Improvement
Mass 218.99m/z : No Improvement
Mass 69.00m/z : 41 % Improvement
0
-2
0
-250
-200
May 9, 2007
-150
-100
Lens 1-2 Voltage (V)
-50
0
May 9, 2007
Y To save the tune and calibration report
1. In the System Tune and Calibration dialog box, click Save Report. The Save As dialog
box appears.
2. Enter the name and the path of the report, and click Save.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
71
5
Changing Ionization Modes
You can operate the TSQ Quantum XLS and TSQ Quantum GC mass spectrometers in
either EI or CI ionization mode. Changing ionization modes requires changing ion volumes.
It is not necessary to shut down the mass spectrometer to change ion volumes.
The following sections describe how to change the ionization mode.
Contents
• Removing the Ion Volume
• Installing the Ion Volume
Removing the Ion Volume
Tools Needed
• Gloves, clean, lint- and powder-free
• Insert/removal (I/R) tool and guide bar
Frequency
When you change ionization modes or as needed to clean the ion volume
Y To remove the ion volume
1. Install the guide bar.
a. With the guide ball track facing left, insert the guide bar into the entry housing.
See Figure 37.
b. Push the guide bar in as far as it will go; then rotate it 90° clockwise to lock the guide
bar in the entry housing.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
73
5
Changing Ionization Modes
Removing the Ion Volume
Figure 37. Insertion/removal (I/R) tool and guide bar
I/R tool
Guide ball
Guide ball
hole
First stop
Guide ball
track
Guide bar
Bayonet lock
2. Prepare the inlet valve and I/R tool for insertion.
a. Make sure the inlet valve is closed. Figure 38 shows the inlet valve lever is down for
closed.
b. Loosen the inlet valve knob counter-clockwise and remove the inlet valve plug. The
inlet valve plug prevents air from entering the vacuum manifold in case the inlet valve
is inadvertently opened.
c. Turn the I/R tool to the unlock position
in position to accept the ion volume.
74
TSQ Quantum XLS and TSQ Quantum GC User Guide
, which indicates the I/R tool is
Thermo Scientific
5
Changing Ionization Modes
Removing the Ion Volume
Figure 38. Inlet valve components
Inlet valve
knob
Inlet valve
plug
Entry housing
Inlet valve lever
(down is closed,
up is open)
3. Choose Start > All Programs > Thermo Instruments > TSQ > TSQ Tune to open the
EZ Tune window.
4. Click the Probe
appears (Figure 39).
button in the EZ Tune toolbar. The Insert Probe message
Figure 39. Insert Probe message
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
75
5
Changing Ionization Modes
Removing the Ion Volume
5. Inset the I/R tool into the inlet valve.
a. Insert the guide ball into the guide ball hole.
b. Slide the I/R tool forward into the inlet valve until the guide ball is at the guide bar’s
first stop (see Figure 40).
c. Tighten the inlet valve knob clockwise to ensure a leak-tight seal.
Figure 40. I/R tool at the first stop on guide bar and the inlet valve knob tightened
I/R tool
76
First stop of
guide bar
TSQ Quantum XLS and TSQ Quantum GC User Guide
Inlet valve knob
Thermo Scientific
5
Changing Ionization Modes
Removing the Ion Volume
6. Click OK. The forepump evacuates the inlet valve. Wait until the safe to insert probe
message appears, and then click OK.
Figure 41. Safe to insert the probe message
7. Once evacuation is complete, pull the inlet valve lever up to open the inlet valve.
Figure 42. I/R tool at first stop and inlet valve lever up
I/R tool
Thermo Scientific
First stop of
guide bar
Inlet valve lever
TSQ Quantum XLS and TSQ Quantum GC User Guide
77
5
Changing Ionization Modes
Removing the Ion Volume
8. Remove the ion volume.
a. Slide the I/R tool into the vacuum manifold until the tip of the I/R tool is fully
inserted into the ion volume holder. The head of the arrow on the I/R tool
(Figure 44), when viewed through the window of the ion source manifold, should
not be visible.
Figure 43. I/R tool inserted into the inlet valve
b. Turn the I/R tool handle counterclockwise to put the I/R tool into lock position
. Listen for a click, which indicates that the handle is fully engaged in
the lock position and is holding the ion volume.
c. Withdraw the I/R tool (with the ion volume attached) until the guide ball reaches the
first stop (see Figure 42).
d. Pull the lever down to close the inlet valve.
78
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
5
Changing Ionization Modes
Removing the Ion Volume
CAUTION Do not withdraw the I/R tool beyond the point where the guide ball
reaches the first stop in the guide bar. Close the inlet valve first. Otherwise, the
system vents to the atmosphere.
e. Loosen the inlet valve knob by turning it counterclockwise to release the seal.
f.
Continue withdrawing the I/R tool completely from the inlet valve by sliding the I/R
tool through the guide ball track in the guide bar.
CAUTION Because the ion volume might be too hot to touch, let it cool to room
temperature before handling it.
9. Remove the ion volume from the I/R tool. See Figure 44.
a. Wearing clean gloves, press the ion volume forward into the tip of the I/R tool and
rotate it to disconnect the bayonet pins from the pin guides.
b. Pull the ion volume out of the I/R tool.
Figure 44. Ion volume, ion volume holder, and I/R tool
Arrow
Bayonet pin
Bayonet
lock
Arrow
I/R tool
Bayonet pin
guide
Ion volume holder
Spring washer
Ion volume
10. To clean the ion volume, use the instructions in “Cleaning Stainless Steel Parts” on
page 113. If you are cleaning a CI ion volume, be sure to clean out the small electron
entrance hole. Aluminum oxide can get trapped in this hole, which can adversely affect
sensitivity. Use a dental pick or old syringe needle to clean the hole.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
79
5
Changing Ionization Modes
Installing the Ion Volume
Installing the Ion Volume
You install an ion volume after cleaning or to change ionization modes.
Y To install the ion volume
Note Wear clean, lint- and powder- free gloves when you handle a clean ion volume.
1. Place the clean ion volume on the I/R tool with the arrows aligned. See Figure 44.
CAUTION Make sure the arrows on the I/R tool and ion volume are aligned to avoid
damage to the ion source.
2. Turn the I/R tool handle to the lock position
.
3. Install the guide bar.
a. With the guide ball track facing left, insert the guide bar into the entry housing (see
Figure 37).
b. Push the guide bar in as far as it will go, and then rotate it 90° clockwise to lock the
guide bar in the entry housing.
4. Prepare the inlet valve for insertion.
a. Make sure the inlet valve is closed. Figure 38 shows the inlet valve lever is down for
closed.
b. Loosen the inlet valve knob counter-clockwise and remove the inlet valve plug.
Remove the inlet valve plug. The inlet valve plug prevents air from entering the
vacuum manifold in case the inlet valve is inadvertently opened.
5. Choose Start > All Programs > Thermo Instruments > TSQ > TSQ Tune to open the
EZ Tune window.
6. Click the Probe
appears. See Figure 39.
button in the EZ Tune toolbar. The Insert Probe message
7. Inset the I/R tool into the inlet valve.
a. Insert the guide ball into the guide ball hole.
b. Slide the I/R tool forward into the inlet valve until the guide ball is at the guide bar’s
first stop. See Figure 40.
c. Tighten the inlet valve knob clockwise to ensure a leak-tight seal.
8. Click OK. The forepump evacuates the inlet valve. Wait for the safe to insert probe
message to appear (Figure 41); then click OK.
9. Once evacuation is complete, pull the inlet valve lever up to open the inlet valve.
80
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
5
Changing Ionization Modes
Installing the Ion Volume
10. Install the ion volume.
a. Slide the I/R tool and ion volume assembly into the vacuum manifold until the ion
volume assembly is fully inserted and seated into the ion source block. Listen for a
click, which indicates that the ion volume has connected with the ion source block.
b. Turn the I/R tool handle to the unlock position
volume from the I/R tool.
to release the ion
Figure 45. Ion volume seated in the ion source block
Ion source block
Ion volume
11. Verify that the ion volume is fully seated in the ion source block:
• Withdraw the I/R tool away from the ion volume about 2.5 cm (1 in.), and turn the
I/R tool handle to the lock position.
• Slide the I/R tool back into the vacuum manifold until the end of the I/R tool just
touches the ion volume.
• Test that the I/R tool does not go into the inlet valve completely, which indicates that
the ion volume is seated properly.
12. Remove the I/R tool.
a. Withdraw the I/R tool until the guide ball reaches the first stop (see Figure 42).
b. Close the inlet valve by pulling the inlet valve lever down (see Figure 40).
CAUTION Do not withdraw the I/R tool beyond the point where the guide ball
reaches the first stop in the guide bar. Close the inlet valve first. Otherwise, the
system vents to the atmosphere.
c. Loosen the inlet valve knob by turning it counter-clockwise.
d. Continue withdrawing the I/R tool completely from the inlet valve by sliding the I/R
tool through the guide ball track in the guide bar.
13. Remove the guide bar by rotating it 90° counter-clockwise and sliding it out of the entry
housing.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
81
5
Changing Ionization Modes
Installing the Ion Volume
14. Replace the inlet valve plug. Orient the plug to indicate whether an EI or CI ion volume
is installed.
15. Tighten the inlet valve knob clockwise to create a leak-tight seal.
16. (Optional) To prevent the inlet valve from being opened accidentally, pull free and
remove the inlet valve lever.
82
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
6
Maintenance
Maintenance
This chapter describes routine maintenance procedures that you must perform to ensure
optimum performance of the TSQ Quantum XLS or TSQ Quantum GC mass spectrometer.
Optimum performance of the mass spectrometer depends on the maintenance of all parts of
the instrument. You are responsible for maintaining your instrument properly by
performing the maintenance procedures on a regular basis.
Note When you perform maintenance procedures, be methodical; always wear clean,
lint-free gloves when handling the components of the ion source; always place the
components on a clean, lint-free surface; and never overtighten a screw or use excessive
force.
Table 6 lists routine and infrequent mass spectrometer maintenance procedures. For
instructions on maintaining the GC or autosampler, refer to the manuals that come with the
GC or autosampler.
Table 6. Maintenance procedures (Sheet 1 of 2)
Mass spectrometer
component
Procedure
Frequency
Procedure
location
Ion source
Cleaning the ion
volume
As needed*
Page 87
Ion source
Cleaning the lenses
As needed*
Page 88 (TSQ
Quantum XLS),
page 95 (TSQ
Quantum GC)
Ion source
Cleaning the entire ion As needed*
source
Page 107(TSQ
Quantum XLS),
page 110 (TSQ
Quantum GC)
Ion source
Replacing the filament
Inlet valve
Replacing the ball valve If there is an air
seal
leak in the inlet
valve
Calibration compound Adding calibration
compound
Thermo Scientific
If the filament fails Page 102 (TSQ
Quantum XLS),
page 105 (TSQ
Quantum GC)
Page 120
If EZ Tune error
Page 117
message indicates
low intensity of
calibration gas ions
TSQ Quantum XLS and TSQ Quantum GC User Guide
83
6
Maintenance
Table 6. Maintenance procedures (Sheet 2 of 2)
Mass spectrometer
component
Procedure
location
Procedure
Frequency
Gas chromatograph
Replacing the capillary
column
If chromatographic Page 122
separation is bad or
if a different
stationary phase is
needed for the
application
Q0 ion optics
Cleaning the Q0
quadrupole and L11
and L12 lenses**
As needed*
Mass analyzer
Cleaning the Q1, Q2,
and Q3 quadrupoles
and lenses**
As needed*
Forepump
Purging the forepump
pump oil
If pump oil is
cloudy
Forepump
Adding pump oil
If pump oil level is Manufacturer’s
low
documentation
Forepump
Changing the pump oil Every 4 months or
if pump oil is
cloudy and
discolored
Ion detection system
Replacing the electron
multiplier assembly**
Electronic modules
Replacing an electronic If electronic
module**
module fails
PCBs
Replacing a PCB**
Manufacturer’s
documentation
Manufacturer’s
documentation
If noise in
spectrum is
excessive or proper
electron multiplier
gain cannot be
achieved
If PCB fails
*The frequency of cleaning the components of the mass spectrometer depends on the types and amounts of
samples and solvents that are introduced into the instrument. Cleaning of the Q0, Q1, Q2, and Q3
quadrupoles is rarely (if ever) required.
**A Thermo Fisher Scientific Field Service Engineer must perform this maintenance procedure.
84
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
6 Maintenance
Cleaning Ion Source Components
Contents
• Cleaning Ion Source Components
• Replacing the Filament of the TSQ Quantum XLS Ion Source
• Replacing the Filament of the TSQ Quantum GC Ion Source
• Disassembling and Reassembling the TSQ Quantum XLS Ion Source
Completely
• Disassembling and Reassembling the TSQ Quantum GC Ion Source
Completely
• Maintaining the Forepump
• Adding Calibration Compound
• Replacing the Ball Valve Seal
• Removing and Installing a GC Capillary Column
Cleaning Ion Source Components
An important part of maintaining your TSQ Quantum XLS or TSQ Quantum GC mass
spectrometer is making sure that the ion source components are clean. Follow the cleaning
procedures in this section to clean stainless steel and non-stainless steel parts.
When your mass spectrometer is clean and in good working order, perform benchmark tests
and record the results. When the tested performance of your system decreases significantly
from your benchmark test results, clean the ion volume. If this does not restore performance,
clean the lenses.
How often you clean the ion source depends on the types and amounts of samples and
solvents you introduce into the system. In general, the closer a component is to the sample
introduction point, the more rapidly the component becomes dirty. For example, you clean
the ion volume more often than other parts. If you just want to change or clean the ion
volumes, you do not need to shut down the system.
Most parts can be removed and disassembled by hand. Make sure you have all the tools
needed before starting each procedure.
Figure 46 shows an exploded view of the ion source assembly.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
85
Figure 46. Ion source assembly of the TSQ Quantum XLS, exploded and assembled views
86
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
Figure 47. Ion source assembly of the TSQ Quantum GC, exploded and assembled views
87
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
6
Maintenance
Cleaning Ion Source Components
Cleaning Ion Volumes
You do not need to vent the mass spectrometer to atmosphere to clean the ion volume.
Y To clean the ion volume
1. Remove the ion volume. See “Removing the Ion Volume” on page 73.
CAUTION Burn Hazard. The ion volume might be hot. Allow the ion volume to cool
to room temperature before touching it.
2. Clean the ion volume. See “Cleaning Stainless Steel Parts” on page 113. If you are
cleaning a CI ion volume, be sure to clean out the small electron entrance hole.
Aluminum oxide can get trapped in this hole, which can adversely affect sensitivity. Use a
dental pick or old syringe needle to clean the hole.
3. Reinstall the ion volume. See“Installing the Ion Volume” on page 80.
Cleaning Lenses of the TSQ Quantum XLS Ion Source
If cleaning the ion volume does not restore system performance, clean the ion source lenses.
Y To clean the ion source lenses
1. Shut down and vent the system. See “Shutting Down the System Completely” on
page 53.
CAUTION Shock Hazard. Unplug the TSQ Quantum XLS mass spectrometer before
proceeding.
CAUTION Burn Hazard. The ion source might be hot. Allow the ion source to cool
to room temperature before touching it.
2. Remove the ion source. See “Removing the TSQ Quantum XLS Ion Source” on page 89.
3. Remove the ion source lens assemblies from the ion source and disassemble it. See
“Removing the L1, L2, L3 Lens Assembly and the Q00 Quadrupole Lens Assembly” on
page 91.
4. Clean the lenses and the spacers. See “Cleaning Stainless Steel Parts” on page 113 and
“Cleaning Non-Stainless Steel or Hybrid Part” on page 115, respectively.
5. Reassemble L1, L2, L3 lens assembly and install it and the Q00 quadrupole onto the ion
source. See “Reassembling the L1, L2, L3 Lens Assembly” on page 93.
88
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
6 Maintenance
Cleaning Ion Source Components
6. Reinstall the ion source into the mass spectrometer. See “Reinstalling the TSQ Quantum
XLS Ion Source” on page 94.
7. Restart the system. See “Starting Up the System after a Complete Shutdown” on page 55.
Removing the TSQ Quantum XLS Ion Source
Remove the ion source to clean the lenses or replace the filament.
Tools Needed
• Gloves, cleanroom grade (P/N 23827-0008 and 23827-0009)
• Lint-free cloth
Frequency
As needed to perform ion source maintenance
Y To remove the ion source
CAUTION Do not remove the ion source without first pulling back the GC capillary
column, or else the capillary column might break.
1. Shut down and vent the TSQ Quantum XLS mass spectrometer. See “Shutting Down the
System Completely” on page 53.
CAUTION Shock Hazard. Unplug the TSQ Quantum XLS mass spectrometer before
proceeding.
2. Pull back the capillary column.
a. Lower the oven, injector, and transfer line temperatures to 30 °C and allow them to
cool before continuing.
CAUTION Burn Hazard. The GC oven, injector, and transfer line are hot. Allow
them to cool to room temperature before touching them.
b. Once the oven, injector, and transfer line are cool, turn off the gas chromatograph.
c. Loosen the transfer line nut. See Figure 64 on page 123.
d. Pull back the column. (You can first mark the column’s position with white out or
something similar.)
3. Prepare a clean work area by covering the area with lint-free cloth.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
89
6
Maintenance
Cleaning Ion Source Components
4. Undo the two latches that secure the lid of the ion source vacuum manifold, and open the
lid.
5. Disconnect the connectors from the lead pins at the EI/CI Source PCB, and L1, L2, L3
lens assembly.
Figure 48. TSQ Quantum XLS ion source, installed
Q0 quadrupole mount
Lens L1, L2, L3 connector
EI/CI Source PCB
connector
Thumbscrew
Transfer line
Thumbscrew
CAUTION Burn Hazard. The ion source might be hot. Allow the ion source to cool
to room temperature before touching it.
Note Wear clean, lint- and powder- free gloves when you handle the ion source.
6. While holding the ion source assembly, loosen the two thumbscrews that secure the ion
source assembly to the Q0 quadrupole mount.
7. Pull the ion source assembly away from the Q0 quadrupole mount and 2 mm to the right
to clear the transfer line bellows.
8. Slide the ion source off the magnet yoke.
90
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
6 Maintenance
Cleaning Ion Source Components
9. Place the ion source and magnet yoke on a clean, lint-free cloth.
10. Close the lid of the ion source vacuum manifold.
Removing the L1, L2, L3 Lens Assembly and the Q00 Quadrupole Lens Assembly
The ion source includes the L1, L2, L3 lens assembly and the Q00 quadrupole lens assembly.
You must remove and disassemble the L1, L2, L3 lens assembly to clean the lenses. A retainer
clip fastens the L1, L2, L3 lens assembly and the Q00 quadrupole lens assembly to the ion
source block.
To remove the ion source lens assemblies
1. Wearing clean, lint- and powder- free gloves, unscrew the threaded hexagonal standoff
and remove the retainer clip that secures the lens assembly to the ion source block. See
Figure 49.
2. Remove the Q00 quadrupole lens assembly and the L1, L2, L3 lens assembly from the
heater cartridges of the ion source. Place them on a clean surface.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
91
6
Maintenance
Cleaning Ion Source Components
Figure 49. L1, L2, L3 lens assembly and Q00 quadrupole lens assembly removed from the ion source
Q00 quadrupole
lens assembly
L1, L2, L3 lens
assembly
Ion source
Retainer clip
Threaded standoff
Disassembling the L1, L2, L3 Lens Assembly
Y To disassemble the L1, L2, L3 lens assembly
1. Remove the lens retainer clip by pinching the ends with your fingers or a pair of tweezers.
See Figure 50.
2. Remove the lenses and spacers from the lens holder.
3. Place the components on a clean surface.
92
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
6 Maintenance
Cleaning Ion Source Components
Figure 50. L1, L2, L3 lens assembly, exploded and assembled views
Lens holder
Lens L3
Spacer
Lens L2
Spacer
Lens L1
Lens retainer clip
Note To clean the ion source lenses, follow the procedure in “Cleaning Stainless Steel
Parts” on page 113. To clean the remaining parts, follow the procedure in “Cleaning
Non-Stainless Steel or Hybrid Part” on page 115.
Reassembling the L1, L2, L3 Lens Assembly
Ensure that the lenses are clean and dry before you reassemble the lens assembly.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
93
6
Maintenance
Cleaning Ion Source Components
Y To reassemble the L1, L2, L3 lens assembly
1. Wearing clean, lint- and powder- free gloves, reinstall the lenses and spacers in the lens
holder in the order and orientation shown in Figure 50, starting with lens L3.
2. Reinstall the lens clip by pinching the ends with your fingers or a pair of tweezers and
seating it in the lens holder.
3. Insert the lens alignment tool (in the TSQ Quantum XLS Accessory Kit) into the lens
assembly and align the lenses
Reinstalling the Ion Source Lens Assemblies
Y To reinstall the ion source lens assemblies onto the ion source
1. Wearing clean, lint- and powder- free gloves, insert the three heater cartridges of the ion
source through their corresponding holes in the L1, L2, L3 lens assembly. See Figure 49.
2. Insert the three heater cartridges of the ion source through their corresponding holes in
the Q00 quadrupole lens assembly. Verify the correct orientation of the guide pin hole
and the retainer clip groove on the Q00 quadrupole lens assembly.
3. Use the retainer clip and threaded standoff to secure the lens assemblies to the ion source.
IMPORTANT For accurate temperature readings of the heater block, fully screw the
threaded standoff into the ion source to ensure contact between the temperature
sensor with the heater block.
Reinstalling the TSQ Quantum XLS Ion Source
Y To reinstall the ion source into the vacuum manifold
1. Wearing clean, lint- and powder- free gloves, place the ion source onto the magnet yoke.
2. Position the ion source in the vacuum manifold and ensure the following:
• The transfer line is seated in the sample inlet aperture in the ion source.
• The guide pin on the Q0 quadrupole mount inserts into the guide pin groove on the
Q00 lens assembly.
• The two thumbscrews are aligned with the screw holes in the Q0 quadrupole mount.
See Figure 48.
3. Secure the ion source to the Q0 quadrupole mount by alternately tightening each
thumbscrew one-half turn at a time. Make sure there is no play in the ion source.
4. Align and reconnect the 8-pin connector to the lead pins on the EI/CI Source PCB.
See Figure 48.
5. Align the 3-pin connector such that the side with the beveled edge is closest to the L3 lead
pin and the side with the rounded corners is closest to the L1 lead pin.
94
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
6 Maintenance
Cleaning Ion Source Components
6. Reconnect the 3-pin connector to the lead pins on the L1, L2, and L3 lenses. See
Figure 48.
7. Reinsert the capillary column into the ion source.
a. Using the I/R tool, remove the ion volume. See “Removing the Ion Volume” on
page 73.
b. Push the capillary column in until you can see it in the ion source.
c. Pull the column back just far enough that you cannot see it.
d. Tighten the transfer line nut and transfer line union.
e. Using the I/R tool, install the ion volume. See “Installing the Ion Volume” on
page 80.
8. Close and secure the cover to the ion source vacuum manifold.
Cleaning the Lenses of the TSQ Quantum GC Ion Source
If cleaning the ion volume does not restore system performance, clean the ion source lenses.
Note Lens L4 requires cleaning less often than lens L1, L2, and L3.
Y To clean the ion source lenses
1. Shut down and vent the system. See “Shutting Down the System Completely” on
page 49.
CAUTION Shock Hazard. Unplug the TSQ Quantum GC before proceeding.
CAUTION Burn Hazard. The ion source might be hot. Allow the ion source to cool to
room temperature before touching it.
2. Remove the ion source. See “Removing the TSQ Quantum XLS Ion Source” on page 89.
3. Remove the ion source lens assembly from the ion source and disassemble it. See
“Removing the L1, L2, L3 Lens Assembly and the Q00 Quadrupole Lens Assembly” on
page 91.
4. Clean the lenses and the spacers. See “Cleaning Stainless Steel Parts” on page 113 and
“Cleaning Non-Stainless Steel or Hybrid Part” on page 115, respectively.
5. Reassemble the ion source lens assembly and install it onto the ion source. See
“Reassembling the L1, L2, L3 Lens Assembly” on page 93.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
95
6
Maintenance
Cleaning Ion Source Components
6. Reinstall the ion source into the mass spectrometer. See “Reinstalling the TSQ Quantum
XLS Ion Source” on page 94.
7. Restart the system. See “Starting Up the System after a Complete Shutdown” on page 51.
Removing the TSQ Quantum GC Ion Source
Remove the ion source to clean the lenses or replace the filament.
Tools Needed
• Gloves, cleanroom grade (P/N 23827-0008 and 23827-0009)
• Lint-free cloth
Frequency
As needed to perform ion source maintenance
Y To remove the ion source
CAUTION Do not remove the ion source without first pulling back the capillary
column, or else the capillary column might break.
1. Shut down and vent the TSQ Quantum GC mass spectrometer. See “Shutting Down the
System Completely” on page 49.
CAUTION Shock Hazard. Unplug the TSQ Quantum GC mass spectrometer before
proceeding.
2. Pull back the capillary column.
a. Lower the oven, injector, and transfer line temperatures to 30 °C and allow them to
cool before continuing.
CAUTION Burn Hazard. The GC oven, injector, and transfer line are hot. Allow
them to cool to room temperature before touching them.
b. Once the oven, injector, and transfer line are cool, turn off the gas chromatograph.
c. Loosen the transfer line nut. See Figure 64 on page 123.
d. Pull back the column. (You can first mark the column’s position with white out or
something similar.)
3. Prepare a clean work area by covering the area with lint-free cloth.
4. Undo the two latches that secure the lid of the ion source vacuum manifold, and open the
lid.
96
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
6 Maintenance
Cleaning Ion Source Components
5. Disconnect the connectors from the lead pins at the EI/CI Source PCB, lens L1, L2, L3
assembly, and lens L4 assembly. See Figure 48.
Figure 51. TSQ Quantum GC ion source, installed
Q0 quadrupole mount
Lens L4 connector
Lens L1, L2, L3 connector
EI/CI Source PCB
connector
Thumbscrew
Transfer line
Thumbscrew
CAUTION Burn Hazard. The ion source might be hot. Allow the ion source to cool
to room temperature before touching it.
Note Wear clean, lint- and powder- free gloves when you handle the ion source.
6. While holding the ion source assembly, loosen the two thumbscrews that secure the ion
source assembly to the Q0 quadrupole.
7. Pull the ion source assembly away from the Q0 quadrupole and 2 mm to the right to clear
the transfer line bellows.
8. Slide the ion source off the magnet yoke.
9. Place the ion source and magnet yoke on a clean, lint-free cloth.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
97
6
Maintenance
Cleaning Ion Source Components
10. Close the lid of the ion source vacuum manifold.
Removing the Ion Source Lens Assembly
The ion source lens assembly includes lens L1, L2, L3, and L4. You must remove and
disassemble the ion source lens assembly to clean the lenses. A retainer clip fastens the lens
assembly to the ion source block.
To remove the ion source lens assembly
1. Wearing clean, lint- and powder- free gloves, unscrew the threaded hexagonal standoff
and remove the retainer clip that secures the lens assembly to the ion source block.
See Figure .
2. Pull the lens assembly out of the ion source.
3. Place the ion source and lens assembly on a clean surface.
Figure 52. Lens assembly removed from the ion source
Lens
assembly
Ion source
Retainer clip
Threaded
standoff
Disassembling the Ion Source Lens Assembly
Y To disassemble the lens assembly
1. Pull the L1, L2, L3 lens assembly from the L4 lens assembly.
98
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
6 Maintenance
Cleaning Ion Source Components
Figure 53. Ion source lens assembly separated into L1, L2, L3 lens assembly (left) and L4 lens
assembly (right)
L1, L2, L3 lens
assembly
L4 lens
assembly
2. Disassemble the L1, L2, L3 lens assembly.
a. Remove the lens clip by pinching the ends with your fingers or a pair of tweezers.
b. Remove the lenses and spacers from the lens holder.
c. Place the components on a clean surface.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
99
6
Maintenance
Cleaning Ion Source Components
Figure 54. L1, L2, L3 lens assembly, exploded and assembled views
Lens holder
Lens L3
Spacer
Lens L2
Spacer
Lens L1
Lens retainer clip
3. Disassemble the L4 lens assembly.
a. Remove the lens clip by pinching the ends with your fingers or a pair of tweezers.
b. Remove the lens L4 from the lens holder.
c. Place the components on a clean surface.
100
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
6 Maintenance
Cleaning Ion Source Components
Figure 55. L4 lens assembly, exploded view
Lens clip
Lens L4
Lens holder
Note To clean the ion source lenses, follow the procedure in “Cleaning Stainless Steel
Parts” on page 113.
Reassembling the Ion Source Lens Assembly
Ensure that the lenses are clean and dry before you reassemble the lens assembly.
Y To reassemble the ion source lens assembly
1. Wearing clean, lint- and powder- free gloves, reassemble the L4 lens assembly.
a. Insert lens L4 into the lens holder. See Figure 55.
b. Install the lens clip by pinching the ends with your fingers or a pair of tweezers. Seat
the lens clip in the lens holder.
2. Reassemble the L1, L2, L3 lens assembly. See Figure 54.
a. Reinstall the lenses and spacers in the lens holder in the order and orientation shown
in Figure 54.
b. Reinstall the lens clip by pinching the ends with your fingers or a pair of tweezers and
seating it in the lens holder.
c. Insert the lens alignment tool (in the TSQ Quantum GC Accessory Kit) into the lens
assembly and align the lenses
3. Insert the L4 lens assembly into the L1, L2, L3 lens assembly. Align the lead pins.
See Figure 53.
Reinstalling the Ion Source Lens Assembly
Y To reinstall the ion source lens assembly onto the ion source
1. Wearing clean, lint- and powder- free gloves, insert the lens assembly into the ion source.
See Figure 53.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
101
6
Maintenance
Replacing the Filament of the TSQ Quantum XLS Ion Source
2. Use the retainer clip and threaded standoff to secure the lens assembly to the ion source.
See Figure .
IMPORTANT For accurate temperature readings of the heater block, fully screw the
threaded standoff into the ion source to ensure contact between the temperature
sensor with the heater block.
Reinstalling the Ion Source
Y To reinstall the ion source into the vacuum manifold
1. Wearing clean, lint- and powder- free gloves, place the ion source onto the magnet yoke.
2. Position the ion source in the vacuum manifold and ensure the following:
• The transfer line is seated in the sample inlet aperture in the ion source.
• The two thumbscrews are aligned with the screw holes in the Q0 quadrupole mount.
See Figure 48.
3. Secure the ion source to the Q0 quadrupole mount by alternately tightening each
thumbscrew one-half turn at a time. Make sure there is no play in the ion source.
4. Reconnect the connectors to the lead pins on the EI/CI Source PCB, lens L1, L2, L3
assembly (the orientation is not important), and lens L4 assembly. See Figure 48.
5. Reinsert the capillary column into the ion source.
a. Using the I/R tool, remove the ion volume. See “Removing the Ion Volume” on
page 69.
b. Push the column in until you can see it through the inlet valve.
c. Pull the column back just far enough that you cannot see it.
d. Tighten the transfer line nut and transfer line union.
e. Using the I/R tool, install the ion volume. See “Installing the Ion Volume” on
page 76.
6. Close and secure the cover to the ion source vacuum manifold.
Replacing the Filament of the TSQ Quantum XLS Ion Source
The number of ions produced in the ion source is approximately proportional to the filament
emission current. If ion production is lacking, you might have to replace the filament. If the
measured emission current is substantially less than the value that the emission current is set
to, or if the measured emission current decreases over time, then the filament has failed or is
failing and requires replacement.
102
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
6 Maintenance
Replacing the Filament of the TSQ Quantum XLS Ion Source
See Figure 46 on page 86 and Figure 56 on page page 104 for the location of the ion source
components.
Tools Needed
• Filament (P/N 120320-0030)
• Gloves, cleanroom grade (P/N 23827-0008 and 23827-0009)
• Lint-free cloth
Frequency
If filament fails
Y To replace the ion source filament
1. Prepare a clean work area by covering the area with lint-free cloth.
2. Shut down and vent the TSQ Quantum XLS mass spectrometer. See “Shutting Down the
System Completely” on page 53.
CAUTION Shock Hazard. Unplug the TSQ Quantum XLS mass spectrometer before
proceeding.
3. Remove the ion source. See “Removing the TSQ Quantum XLS Ion Source” on page 89.
4. Wearing clean, lint- and powder- free gloves, remove the ion source from the magnet
yoke.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
103
6
Maintenance
Replacing the Filament of the TSQ Quantum XLS Ion Source
Figure 56. Disassembling the TSQ Quantum XLS ion source to replace the filament
Filament retainer clip
Filament
Centering ring
Heater ring and lens
assemblies
Heater ring and lens
assemblies retainer clip
Ion source block and
EI/CI Source PCB
Threaded standoff
5. Unscrew the threaded standoff and remove the retainer clip that secures the heater ring
and lens assemblies to the ion source block. See Figure 56.
6. Pull the heater ring and lens assemblies off of the cartridge heaters on the ion source
block. Place them on a clean surface.
7. Remove the retainer clip that secures the filament and centering ring to the ion source
block.
8. Remove the filament and centering ring from the ion source block.
Note Now is a good time to clean the ion source block and lenses. Use the procedure
described in “Cleaning Stainless Steel Parts” on page 113 to clean the ion source block
and lenses.
9. Inspect and install a new filament (P/N 120320-0030, in the TSQ Quantum XLS
Accessory Kit).
a. Verify that the filament wire is centered in the electron lens hole.
b. Insert the filament into the centering ring.
104
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
6 Maintenance
Replacing the Filament of the TSQ Quantum GC Ion Source
c. Align the filament leads with the EI/CI Source PCB connectors and gently press the
leads into the connectors. Normally, there is a small gap (about 0.016 in.) between
the filament and the connectors. The gap allows the ceramic centering ring to
properly position and align the electron lens hole with the ion volume.
d. Secure the filament and centering ring with the retainer clip.
10. Press the heater ring and lens assemblies onto the cartridge heaters on the ion source
block.
11. Use the retainer clip and threaded standoff to secure the lens assemblies to the ion source.
IMPORTANT For accurate temperature readings of the heater block, fully screw the
threaded standoff into the ion source to ensure contact between the temperature
sensor with the heater block.
12. Reinstall the ion source onto the magnet yoke.
13. Reinstall the ion source into the vacuum manifold. See “Reinstalling the TSQ Quantum
XLS Ion Source” on page 94.
Restart the system. See “Starting Up the System after a Complete Shutdown” on page 55.
Replacing the Filament of the TSQ Quantum GC Ion Source
The number of ions produced in the ion source is approximately proportional to the filament
emission current. If ion production is lacking, you might have to replace the filament. If the
measured emission current is substantially less than the value that the emission current is set
to, or if the measured emission current decreases over time, then the filament has failed or is
failing and requires replacement.
See Figure 46 on page 86 and Figure 56 on page page 104 for the location of the ion source
components.To replace the ion source filament
Tools Needed
• Filament (P/N 120320-0030)
• Gloves, cleanroom grade (P/N 23827-0008 and 23827-0009)
• Lint-free cloth
Frequency
If filament fails
1. Prepare a clean work area by covering the area with lint-free cloth.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
105
6
Maintenance
Replacing the Filament of the TSQ Quantum GC Ion Source
2. Shut down and vent the TSQ Quantum GC. See “Shutting Down the System
Completely” on page 49.
CAUTION Shock Hazard. Unplug the TSQ Quantum GC before proceeding.
3. Remove the ion source. See “Removing the TSQ Quantum GC Ion Source” on page 96.
4. Wearing clean, lint- and powder- free gloves, remove the ion source from the magnet
yoke.
Figure 57. Disassembling the ion source to replace the filament
Filament retainer clip
Filament retainer clip
Filament
Centering ring
Filament
Centering ring
Heater ring and lens
assembly
Heater ring and lens
assembly retainer clip
Ion source block and
EI/CI Source PCB
Threaded standoff
5. Unscrew the threaded standoff and remove the retainer clip that secures the heater ring
and lens assembly to the ion source block. See Figure 56.
6. Pull the heater ring and lens assembly out of the ion source block.
7. Remove the retainer clip that secures the filament and centering ring to the ion source
block.
106
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
6 Maintenance
Disassembling and Reassembling the TSQ Quantum XLS Ion Source Completely
8. Remove the filament and centering ring from the ion source block.
Note Now is a good time to clean the ion source block. Use the procedure described
in “Cleaning Stainless Steel Parts” on page 113 to clean the ion source block.
9. Inspect and install a new filament (P/N 120320-0030, in the TSQ Quantum GC Accessory
Kit).
a. Verify that the filament wire is centered in the electron lens hole.
b. Insert the filament into the centering ring.
c. Seat the filament on the EI/CI Source PCB.
d. Secure the filament with the retainer clip.
10. Press the heater ring and lens assembly onto the ion source block.
11. Use the retainer clip and threaded standoff to secure the lens assembly to the ion source.
See Figure 56.
IMPORTANT For accurate temperature readings of the heater block, fully screw the
threaded standoff into the ion source to ensure contact between the temperature
sensor with the heater block.
12. Reinstall the ion source onto the magnet yoke.
13. Reinstall the ion source into the vacuum manifold. See “Reinstalling the TSQ Quantum
XLS Ion Source” on page 94.
14. Restart the system. See “Starting Up the System after a Complete Shutdown” on page 51.
Disassembling and Reassembling the TSQ Quantum XLS Ion Source
Completely
Disassemble the ion source completely to clean the ion source block or to replace the EI/CI
Source PCB.
See Figure 58 for the location of the ion source components.
Tools Needed
• Gloves, cleanroom grade (P/N 23827-0008 and 23827-0009)
• Lint-free cloth
Frequency
As needed
Y To disassemble the ion source completely
1. Prepare a clean work area by covering the area with lint-free cloth.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
107
6
Maintenance
Disassembling and Reassembling the TSQ Quantum XLS Ion Source Completely
2. Shut down and vent the TSQ Quantum XLS. See“Shutting Down the System
Completely” on page 49.
CAUTION Shock Hazard. Unplug the TSQ Quantum XLS mass spectrometer before
proceeding.
3. Wearing clean, lint- and powder- free gloves, remove the ion source. See “Removing the
TSQ Quantum XLS Ion Source” on page 89.
4. Remove the ion source from the magnet yoke.
Figure 58. TSQ Quantum XLS ion source, disassembled
Filament retainer clip
Q00 quadrupole
lens assembly
Filament
Heater ring
Centering ring
Ion source block
L1, L2, L3 lens
assembly
Threaded standoff
Heater ring and lens
assemblies retainer clip
EI/CI Source PCB
Base studs
5. Pull the Q00 quadrupole lens assembly and the L1, L2, L3 lens assembly off the ion
source.
6. Pull the heater ring off the ion source.
7. Unscrew and remove the three base studs.
108
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
6 Maintenance
Disassembling and Reassembling the TSQ Quantum XLS Ion Source Completely
CAUTION Do not pull the EI/CI Source PCB out of the ion source block. This
might damage the EI/CI Source PCB.
8. Remove the EI/CI Source PCB.
a. Place the ion source on a clean surface with the cartridge heaters down and the EI/CI
Source PCB up.
b. Press down on the ion source block to separate it from the EI/CI Source PCB.
9. Remove the retainer clip that secures the filament and centering ring to the ion source
block.
10. Remove the filament and centering ring from the ion source block.
11. Clean the ion source block and lenses. See “Cleaning Stainless Steel Parts” on page 113.
Clean non-stainless steel parts as described in “Cleaning Non-Stainless Steel or Hybrid
Part” on page 115.
Y To reassemble the ion source
1. Reinstall the EI/CI Source PCB and filament.
a. Align and insert the cartridge heaters and temperature sensor on the EI/CI Source
PCB with the corresponding holes in the ion source block.
IMPORTANT Ensure that the temperature sensor is seated snugly.
b. Reinstall the three base studs.
c. Align the filament leads with the EI/CI Source PCB connectors and gently press the
leads into the connectors. Normally, there is a small gap (about 0.016 in.) between
the filament and the connectors. The gap allows the ceramic centering ring to
properly position and align the electron lens hole with the ion volume.
2. Press the heater ring and lens assemblies onto the ion source block.
3. Use the retainer clip and threaded standoff to secure the lens assembly to the ion source.
See Figure 58.
IMPORTANT For accurate temperature readings of the heater block, fully screw the
threaded standoff into the ion source to ensure contact between the temperature
sensor with the heater block.
4. Reinstall the ion source into the vacuum manifold. See “Reinstalling the TSQ Quantum
XLS Ion Source” on page 94.
5. Restart the system. See “Starting Up the System after a Complete Shutdown” on page 51.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
109
6
Maintenance
Disassembling and Reassembling the TSQ Quantum GC Ion Source Completely
Disassembling and Reassembling the TSQ Quantum GC Ion Source
Completely
Disassemble the ion source completely to clean the ion source block or replace the EI/CI
Source PCB.
See Figure 58 for the location of the ion source components.
Tools Needed
• Gloves, cleanroom grade (P/N 23827-0008 and 23827-0009)
• Lint-free cloth
Frequency
As needed
Y To disassemble the ion source completely
1. Prepare a clean work area by covering the area with lint-free cloth.
2. Shut down and vent the TSQ Quantum XLS. See“Shutting Down the System
Completely” on page 49.
CAUTION Shock Hazard. Unplug the TSQ Quantum XLS before proceeding.
3. Wearing clean, lint- and powder- free gloves, remove the ion source. See “Removing the
TSQ Quantum XLS Ion Source” on page 89.
4. Remove the ion source from the magnet yoke.
110
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
6 Maintenance
Disassembling and Reassembling the TSQ Quantum GC Ion Source Completely
Figure 59. Disassembling the ion source
Filament retainer clip
Filament
Heater ring
Centering ring
Ion source block
Lens assembly
Threaded standoff
Heater ring and
lens assembly
retainer clip
EI/CI Source PCB
Base studs
5. Pull the lens assembly off the ion source.
6. Pull the heater ring off the ion source.
7. Unscrew and remove the three base studs.
CAUTION Shock Hazard. Do not pull the EI/CI Source PCB out of the ion source
block. This might damage the EI/CI Source PCB.
8. Remove the EI/CI Source PCB.
a. Place the ion source on a clean surface with the cartridge heaters down and the EI/CI
Source PCB up.
b. Press down on the ion source block to separate it from the EI/CI Source PCB.
9. Remove the retainer clip that secures the filament and centering ring to the ion source
block.
10. Remove the filament and centering ring from the ion source block.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
111
6
Maintenance
Disassembling and Reassembling the TSQ Quantum GC Ion Source Completely
11. Clean the ion source block and lenses. See “Cleaning Stainless Steel Parts” on page 113.
Clean non-stainless steel parts as described in “Cleaning Non-Stainless Steel or Hybrid
Part” on page 115.
Y To reassemble the ion source
1. Reinstall the EI/CI Source PCB and filament.
a. Align and insert the cartridge heaters and temperature sensor on the EI/CI Source
PCB with the corresponding holes in the ion source block.
IMPORTANT Ensure that the temperature sensor is seated snugly.
b. Reinstall the three base studs.
c. Align the filament leads with the EI/CI Source PCB connectors and gently press the
leads into the connectors. Normally, there is a small gap (about 0.016 in.) between
the filament and the connectors. The gap allows the ceramic centering ring to
properly position and align the electron lens hole with the ion volume.
2. Press the heater ring and lens assembly onto the ion source block.
3. Use the retainer clip and threaded standoff to secure the lens assembly to the ion source.
See Figure 58.
IMPORTANT For accurate temperature readings of the heater block, fully screw the
threaded standoff into the ion source to ensure contact between the temperature
sensor with the heater block.
4. Reinstall the ion source into the vacuum manifold. See “Reinstalling the TSQ Quantum
XLS Ion Source” on page 94.
5. Restart the system. See “Starting Up the System after a Complete Shutdown” on page 51.
112
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
6 Maintenance
Disassembling and Reassembling the TSQ Quantum GC Ion Source Completely
Cleaning Stainless Steel Parts
Tools Needed
• Acetone, reagent grade (or other suitable polar solvent)
• Aluminum oxide abrasive, number 600 (P/N 32000-60340)
• Applicators, cotton-tipped (P/N 00301-01-00015)
• Beaker, 450 mL
• Clean, dry gas
• De-ionized water
• Detergent (Alconox, Micro, or equivalent)
• Dremel rotary tool or equivalent (recommended)
• Foil, aluminum
• Forceps (P/N 76360-0400)
• Gloves, cleanroom grade (P/N 23827-0008 and 23827-0009)
• Glycerol, reagent grade
• Lint-free cloth
• Protective eyewear
• Tap water
• Toothbrush, soft
• Ultrasonic cleaner
Frequency
As needed to clean these stainless steel parts:
• Ion volumes
• Ion source block
• Ion source lenses
CAUTION Material and Eye Hazard. Wear impermeable laboratory gloves and eye
protection when performing cleaning procedures.
Y To clean stainless-steel parts
1. Remove contamination from all the surfaces you are cleaning.
a. Use a slurry of number 600 aluminum oxide in glycerol and a cleaning brush or
cotton-tipped applicator. Contamination appears as dark or discolored areas, but
often is not visible. The heaviest contamination is usually found around the
apertures, such as the electron entrance hole on an ion volume.
b. Clean each part thoroughly, even if no contamination is visible.
c. To clean the inside corners, use the wooden end of an applicator cut at an angle.
d. Use a Dremel® tool with the polishing swab at its lowest speed to increase cleaning
efficiency, as well as decrease the time required to clean the items. To prevent personal
injury, be sure to keep the Dremel tool away from possible hazards, such as standing
water or flammable solvents.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
113
6
Maintenance
Disassembling and Reassembling the TSQ Quantum GC Ion Source Completely
2. Rinse the parts with clean water. Use a clean applicator or toothbrush to remove the
aluminum oxide slurry. Do not let the slurry dry on the metal; dried aluminum oxide is
difficult to remove.
3. Sonicate the parts in a warm detergent solution.
a. Using forceps, place the parts in a beaker containing warm detergent solution.
b. Place the beaker and contents in an ultrasonic bath for five minutes.
c. Rinse the parts with tap water to remove the detergent.
4. Sonicate the parts in deionized water.
a. Using forceps, place the parts in a beaker containing deionized water.
b. Place the beaker and contents in an ultrasonic bath for five minutes.
c. If the water is cloudy after sonicating, pour off the water, add fresh water, and place
the beaker and its contents in an ultrasonic bath again for five minutes. Repeat until
the water is clear.
5. Sonicate the parts in acetone.
a. Using forceps, place the parts in a beaker containing acetone.
b. Using forceps, transfer the parts to a beaker containing fresh acetone.
c. Place the beaker and contents in an ultrasonic bath again for five minutes.
6. Blow-dry the parts immediately. Use clean, dry gas to blow the acetone off the parts.
7. Using forceps, place the parts in a beaker, cover the beaker with aluminum foil, and put
the beaker in the oven.
8. Dry the parts in an oven set at 100 oC for 30 minutes.
9. Allow the parts to cool before putting them back together.
114
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
6 Maintenance
Disassembling and Reassembling the TSQ Quantum GC Ion Source Completely
Cleaning Non-Stainless Steel or Hybrid Part
Tools Needed
• Acetone, reagent grade (or other suitable polar solvent)
• Aluminum oxide abrasive, number 600 (P/N 32000-60340)
• Applicators, cotton-tipped (P/N 00301-01-00015)
• Beaker, 450 mL
• Clean, dry gas
• De-ionized water
• Detergent (Alconox, Micro, or equivalent)
• Dremel rotary tool or equivalent (recommended)
• Forceps (P/N 76360-0400)
• Gloves, cleanroom grade (P/N 23827-0008 and 23827-0009)
• Glycerol, reagent grade
• Lint-free cloth
• Protective eyewear
• Tap water
• Toothbrush, soft
Frequency
As needed to clean non-stainless steel parts (such as aluminum, ceramic, or
gold-plated), or to clean hybrid parts that are partially made of stainless steel:
• Lens holder and spacers
• Filament spacer
• Heater ring
CAUTION Material and Eye Hazard. Wear impermeable laboratory gloves and eye
protection when performing cleaning procedures.
Y To clean non-stainless steel or hybrid parts
1. Remove contamination from stainless steel surfaces. It is only necessary to clean a surface
that comes in contact with the ion beam.
a. Use a slurry of number 600 aluminum oxide in glycerol and a cleaning brush or
cotton-tipped applicator. Contamination appears as dark or discolored areas, but
often is not visible.
b. Clean each part thoroughly, even if no contamination is visible.
c. To clean the inside corners, use the wooden end of an applicator cut at an angle.
d. Use a Dremel tool with the polishing swab at its lowest speed to increase cleaning
efficiency, as well as decrease the time required to clean the items. To prevent personal
injury, be sure to keep the tool away from possible hazards, such as standing water or
flammable solvents.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
115
6
Maintenance
Maintaining the Forepump
2. Rinse the parts with clean water. Use a clean applicator or toothbrush to remove the
aluminum oxide slurry. Do not let the slurry dry on the metal; dried aluminum oxide is
difficult to remove.
3. Scrub all of the parts with a warm detergent solution.
a. Scrub the parts with a toothbrush or clean applicator. Do not soak or sonicate the
parts in detergent.
b. Using forceps, rinse the parts thoroughly with tap water to remove the detergent.
CAUTION Do not leave aluminum parts, such as the heater ring, in the detergent.
Basic solutions, like detergent, discolor aluminum.
4. Rinse the parts in deionized water. Using forceps, dip the parts in a beaker of deionized
water. Change the water if it becomes cloudy. Do not soak or sonicate the parts.
5. Rinse the parts with acetone. Using forceps, dip the parts in a beaker of acetone. Change
the acetone if it becomes cloudy. Do not soak or sonicate the parts.
6. Blow-dry the parts immediately. Use clean, dry gas to blow the acetone off the parts.
Maintaining the Forepump
Inspecting, adding, purging, and changing the forepump oil are all that is required to
maintain the forepump.
Forepump oil is a translucent, light amber color, which you must check often. During normal
operation, oil must always be visible in the oil level sight glass between the MIN and MAX
marks. If the oil level is below the MIN mark, add oil. If the oil is cloudy or discolored, purge
the oil to decontaminate dissolved solvents. If the pump oil is still discolored, change it. You
should change the pump oil every 3000 hours (about four months) of operation. Refer to the
manufacturer’s documentation for procedures for purging, adding, and changing the
forepump oil.
CAUTION If you use ammonia as a chemical ionization reagent gas, change the oil every
month. Ammonia is highly basic and quickly damages seals in the forepump. Purging the
oil with the Gas Ballast Control helps remove dissolved ammonia from the oil.
116
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
6 Maintenance
Adding Calibration Compound
Adding Calibration Compound
Tools Needed
• Calibration compound (P/N 50010-02500)
• Syringe (P/N 36502019)
• 9/16 in., open-ended wrench
Frequency
Yearly or as needed
The calibration compound is a liquid whose mass spectrum of ions is used to tune and
calibrate the TSQ Quantum XLS and TSQ Quantum GC mass spectrometers. The mass
spectrometer uses FC-43 as its calibration compound. Although you cannot visually
determine when a calibration compound needs to be added, Xcalibur Automatic Tune
produces an error message indicating that “the intensity of calibration gas ions is too low.”
This message indicates that calibration compound must be added. However, there are many
factors other than a decrease in the calibration compound that can cause a shortage of
calibration gas ions. As a rule, add calibration compound once a year.
Y To add calibration compound
1. Shut down and vent the mass spectrometer. See “Shutting Down the System Completely”
on page 53.
2. Lower the GC oven and transfer line temperatures to 30 °C and allow them to cool before
continuing.
CAUTION The transfer line is hot. Allow it to cool to room temperature before
moving the GC.
3. Move the GC far enough from the mass spectrometer to access the calibration compound
vial compartment. See the location of the calibration compound vial compartment in
Figure 60.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
117
6
Maintenance
Adding Calibration Compound
Figure 60. Location of the calibration compound vial compartment
Calibration compound vial
compartment
4. Remove the cover to the calibration compound vial compartment.
5. Remove the calibration compound vial from the calibration gas flow module. See
Figure 61.
a. Use a 9/16 in. wrench to remove the nut from the calibration compound vial.
b. Slide the vial away from the calibration valve fitting with the ferrule and nut attached.
c. Remove the ferrule and nut.
CAUTION Having more than 0.1 mL of the calibration compound can damage
the calibration gas flow module. Be sure that there is less than 0.1 mL of the
calibration compound in the vial.
6. Add the calibration compound.
a. Draw up no more than 0.1 mL of calibration compound into a clean syringe.
b. Insert the syringe in the calibration compound vial until it comes in contact with the
white frit at the bottom of the vial.
c. Inject no more than 0.1 mL of calibration compound into the white frit. The frit
absorbs the calibration compound, so if the calibration compound is pooled
(calibrant covers the top of the frit), then you have too much in the vial. Immediately
pour out the excess calibrant according to local environmental regulations.
118
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
6 Maintenance
Adding Calibration Compound
7. Remove the syringe from the calibration compound vial.
Figure 61. Calibration compound vial
Ferrule
Nut
Calibration
compound vial
8. Inspect the 1/4 in. ferrule for damage. Replace it if necessary (P/N 95001-20310).
9. Reinstall the calibration compound vial (P/N 96000-40013) into the module.
a. Hold the vial vertically and place the nut on the neck of the vial.
b. Place the 1/4 in. ferrule over the neck of the vial and into the nut.
c. Connect the vial to the calibration valve fitting with the nut and ferrule.
10. Reinstall the cover to the calibration compound vial compartment.
11. Position the GC next to the mass spectrometer.
12. Restart your system. See “Starting Up the System after a Complete Shutdown” on
page 55.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
119
6
Maintenance
Replacing the Ball Valve Seal
Replacing the Ball Valve Seal
Tools Needed
• Ball valve seal kit (P/N 119265-0003)
Frequency
If the ball valve seal leaks air after you insert the direct insertion probe or
I/R tool into the inlet valve and tighten the inlet valve knob
The ball valve seal in the inlet valve can become worn over time. Replace the ball valve seal if
it leaks air after you insert the direct insertion probe or I/R tool and tighten the inlet valve
knob. The ball valve seal kit (P/N 119265-0003) is included in the TSQ Quantum XLS
Accessory Kit.
The ball valve seal consists of a Teflon® spool and two O-rings. See Figure 62.
Figure 62. Components of the ball valve seal and inlet valve
Inlet valve
housing
Ball valve seal
components
Inlet valve
insert
Inlet valve knob
You use the ball valve seal extraction tool (part of the ball valve seal kit, P/N 119265-0003) to
remove the ball valve seal. See Figure 63.
120
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
6 Maintenance
Replacing the Ball Valve Seal
Y To remove the ball valve seal
1. Loosen the inlet valve knob counter-clockwise until you can remove it.
2. Remove the inlet valve knob, plug, and insert.
3. Insert the not-engaged ball valve seal extraction tool into the inlet valve housing.
4. Engage the ball valve seal by pushing the plunger of the ball valve seal extraction tool
forward.
5. Remove the seal and tool.
6. Pull the plunger back to disengage the seal.
Figure 63. Ball valve seal extraction tool and ball valve seal, not engaged (left) and engaged (right)
Y To install the ball valve seal
1. Obtain the ball valve seal kit (P/N 119265-0003) from the Accessory Kit.
2. Assemble the ball valve seal by installing the two O-rings onto the Teflon™ spool.
3. Insert the ball valve seal into the ball valve manifold with your finger.
4. Install the inlet valve knob and insert.
5. Install the inlet valve plug and tighten the inlet valve knob clockwise.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
121
6
Maintenance
Removing and Installing a GC Capillary Column
Removing and Installing a GC Capillary Column
You replace the GC capillary column if the chromatographic separation is bad or if it does not
have the appropriate stationary phase for your application.
Tools Needed
• Gloves, cleanroom grade (P/N 23827-0008 and 23827-0009)
• Wrench, open-ended, 5/16 in.
• Wrench, open-ended, 7/16 in.
• Wrench, open-ended, 6 mm
Frequency
As needed for maintenance or column replacement
Removing a GC Capillary Column
Y To remove a GC capillary column
1. Shut down the TSQ Quantum XLS or TSQ Quantum GC and the gas chromatograph.
a. Shut down and vent the mass spectrometer. See “Shutting Down the System
Completely” on page 53.
b. Lower the oven, injector, and transfer line temperatures to 30 °C and allow them to
cool before continuing.
CAUTION The oven, injector, and transfer line are hot. Allow them to cool to
room temperature before touching them.
c. Once the oven, injector, and transfer line are cool, turn off the gas chromatograph.
2. Remove the column from the transfer line. See Figure 64.
a. Unscrew the transfer line nut.
b. Remove the column from the transfer line.
122
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
6 Maintenance
Removing and Installing a GC Capillary Column
Figure 64. GC injector and transfer line unions
3. Remove the column from the injector. Refer to the Column chapter of the TRACE GC
Ultra Operating Manual for instructions.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
123
6
Maintenance
Removing and Installing a GC Capillary Column
Installing a GC Capillary Column
Tools Needed
• Capillary column
• Gloves, cleanroom grade (P/N 23827-0008 and 23827-0009)
• Injector ferrule, for 0.25 mm column (P/N 290 134 88)
• Leak detector, hand-held electronic (GL Sciences, Inc., model LD-228, or
equivalent)
• Lint-free tissue
• Magnifying glass
• Methanol or other suitable solvent
• Scoring wafer (or sapphire scribe) to cut capillary column
• Transfer line ferrule, 0.4 mm ID (P/N A0101-18100)
• Wrench, open-ended, 5/16 in.
• Wrench, two, open-ended, 7/16 in.
• Wrench, open-ended, 6 mm
Frequency
As needed
Refer to the TRACE GC Ultra Operating Manual and the TRACE GC Ultra Maintenance and
Troubleshooting Manual for additional GC column installation information.
Y To install a GC capillary column
CAUTION Burn Hazard. The injector, oven, and transfer line might be hot. Allow them
to cool to room temperature before touching them.
Note Wear clean, lint- and powder-free gloves when you handle the column and injector
ferrule.
1. Connect the column to the injector. Refer to the Column chapter of the TRACE GC
Ultra Operating Manual for instructions. Refer to the TRACE GC Ultra Quick Reference
card for the column depth length for your injector.
2. Set up the gas chromatograph.
a. Turn on the gas chromatograph.
b. Set the oven and injector temperatures to 30 °C.
c. Set the injector flow to 1.0 mL/min.
d. Turn off vacuum compensation (under the Right or Left Carrier menu).
124
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
6 Maintenance
Removing and Installing a GC Capillary Column
e. Dip the column outlet in a small vial of methanol. Bubbles indicate there is flow
through the column.
f.
Allow the column to purge for at least 10 minutes.
3. Perform a manual column leak check. Refer to the Ensuring Tightness chapter of the
TRACE GC Ultra Maintenance and Troubleshooting Manual for instructions.
4. Perform a column evaluation. Refer to the Column chapter of the TRACE GC Ultra
Operating Manual for instructions. Refer to the K-Factor Quick Reference card for expected
K factors for various column diameters, column lengths, and carrier gases.
Note The K factor is a measure of the resistance of the column. A K factor that is too
low might indicate a leak in the system, while a K factor that is too high might
indicate a blockage.
5. Perform an automatic column leak check.
a. Run an automatic leak check. Refer to the Column chapter of the TRACE GC Ultra
Operating Manual for instructions.
b. If the report indicates a leak, look for leaks and use the leak detector to fix leaks at all
the fittings in the GC.
c. Repeat column evaluation and leak check procedures until no leaks can be found.
CAUTION Do not raise the oven temperature until you are sure the system is
leak-free. At temperatures above 100 °C, oxygen exposure will destroy the
column.
6. Condition the capillary column according to the manufacturer’s instructions. New
columns must be conditioned before inserting them in the mass spectrometer.
CAUTION Do not insert the column into the transfer line during the co. The material
released from the column during the conditioning (column bleed) will contaminate
the ion source. You must then clean the ion source.
CAUTION Never exceed the column manufacturer’s maximum operating
temperature.
7. Connect the column to the transfer line.
a. Shut down and vent the mass spectrometer. See“Shutting Down the System
Completely” on page 53.
b. Lower the GC oven temperature to 30 °C and allow it to cool before continuing.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
125
6
Maintenance
Removing and Installing a GC Capillary Column
CAUTION Burn Hazard. The GC oven and transfer line might be hot. Allow
them to cool to room temperature before touching them. Do not touch the
injector when it is hot.
c. Wearing clean, lint- and powder-free gloves, unwind about one turn of the capillary
column (shown in Figure 64) from the column outlet end.
d. Wipe about 450 mm (18 in.) of the column with a tissue soaked in methanol.
Note Sliding a septum on the column before the transfer line nut helps you
measure the proper distance between the nut and the end of the column. The
column should extend approximately 1 to 2 mm past the end of the transfer line.
e. Insert the column through the transfer line nut and ferrule. Wipe the column again
with a tissue soaked in methanol.
f.
Score and break the end of the column with a scoring wafer. With the magnifying
glass, check for an even, flat cut. Repeat if necessary.
g. Insert the column into the transfer line.
i.
Open the lid of the ion source vacuum chamber so that you can get a better view
of the column.
ii. Using the I/R tool, remove the ion volume. See “Removing the Ion Volume” on
page 73.
iii. Insert the column into the transfer line and tighten the transfer line nut by hand.
iv. Push the column in until you can see it through the inlet valve.
v.
Pull the column back just far enough that you cannot see it.
vi. Tighten the transfer line nut and transfer line union.
vii. Using the I/R tool, replace the ion volume. See “Installing the Ion Volume” on
page 80.
8. Condition the transfer line ferrule. Graphite/Vespel™ ferrules like the transfer line ferrule
require conditioning to ensure a leak-tight seal.
a. Raise the oven temperature to the maximum temperature at which you will operate
the column (normally 300 °C).
b. Wait 10 minutes.
c. Lower the oven temperature to 30 °C and allow it to cool before continuing.
CAUTION The oven might be hot. Allow it to cool to room temperature before
opening it. Do not touch the injector, which will still be hot.
d. Re-tighten the transfer line nut and the transfer line union.
126
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
6 Maintenance
Removing and Installing a GC Capillary Column
9. Set up the gas chromatograph.
a. Make sure the column does not have any sharp bends and that it does not touch any
metal objects or walls inside the oven.
b. Raise the oven temperature to the initial temperature that you will use (normally
40 °C).
c. Turn on the vacuum compensation (under the Right or Left Carrier menu).
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
127
7
Diagnostics and Troubleshooting
The TSQ Quantum XLS and TSQ Quantum GC system diagnostics can test many
components of the TSQ Quantum XLS and TSQ Quantum GC mass spectrometers. If there
is a problem with the instrument electronics, the diagnostics can often locate the problem.
Replacing a faulty PCB or assembly can usually correct the problem. After the PCB or
assembly is replaced, the diagnostic tests are rerun to verify the instrument is functioning
properly.
Contents
• TSQ Quantum XLS and TSQ Quantum GC System Diagnostics
• Troubleshooting
• Replacing a Fuse
• Replacing PCBs and Power Supplies
TSQ Quantum XLS and TSQ Quantum GC System Diagnostics
The TSQ Quantum XLS and TSQ Quantum GC system diagnostics are used to test the
major electronic circuits within the instrument and indicate whether the circuits pass or fail
the tests. If there is a problem with the instrument electronics, the system diagnostics can
often locate the problem.
The system diagnostics do not diagnose problems that are not electrical in nature. For
example, they do not diagnose poor sensitivity due to misaligned or dirty components or to
improper tuning. Therefore, it is important that the person running the diagnostics be
familiar with system operation and basic hardware theory as well as the details of the
diagnostics.
IMPORTANT Typically, only a Thermo Fisher Scientific Field Service Engineer runs
diagnostic tests because certain tests can overwrite system parameters.
Before calling a Thermo Fisher Scientific Field Service Engineer to run diagnostics, consider
the following:
• Did the system fail when you were running samples?
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
129
7
Diagnostics and Troubleshooting
TSQ Quantum XLS and TSQ Quantum GC System Diagnostics
• Did problems occur after you performed maintenance on the instrument, data system, or
peripherals?
• Did you change the configuration of the system, cables, or peripherals just before the
problem occurred?
If the answer is yes to the first item above, there is the possibility of a hardware failure, and
running the diagnostics is appropriate.
If the answer is yes to either of the last two questions above, the problem is probably
mechanical, not electrical. Reverify that alignment, configurations, and cable connections are
correct before you call the Thermo Fisher Scientific Field Service Engineer. Keep careful notes
documenting the nature of the problem and the corrective steps you have taken. If you are not
successful in correcting the problem, you can e-mail this information to Field Service
engineer. Field Service can then do a preliminary evaluation of the problem before the Field
Service engineer arrives at your site.
130
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
7
Diagnostics and Troubleshooting
Troubleshooting
Troubleshooting
The following topics discuss possible TSQ Quantum XLS and TSQ Quantum GC system
problems and solutions.
Communication Issues
Communication issues likely involve links between the data system and the mass
spectrometer, gas chromatograph, or autosampler. This section does not address
communication issues with other devices.
Communication issues can occur during the following:
• Data transfer between the mass spectrometer and the data system
• Mass spectrometer, gas chromatograph, and autosampler current status readbacks
• Instrument control, method downloading, and uploading
• Start, stop, pause, and initialize functions
• Error messages
How does the MS communication work?
The flow of digital information in the TSQ Quantum XLS and TSQ Quantum GC systems
is bi-directional; the data system downloads analytical methods to the instrument and
activates functions to start, stop, shut down, start up, and initialize. The TSQ Quantum XLS
and TSQ Quantum GC systems report their readiness state, current tasks, various voltages,
heated zones, and pressure readings. It also delivers a steady stream of mass spectral data
during acquisition.
Why does the MS have communication issues?
Some communication issues are due to mechanical faults—for instance, a cable might be
unplugged, or a device might be turned off. In other cases, the instrument method could be
incorrect for TSQ Quantum XLS or TSQ Quantum GC system operations. Less common
communication issues are due to defective electronic hardware components.
How do I detect communication issues?
You might detect communication issues from an error message delivered by the data system,
or you might notice the failure of the data system to perform some expected task.
Issue: Data system is
unable to initiate
communication with the
mass spectrometer.
Thermo Scientific
Possible causes/solutions
• The software is not configured correctly. Select and configure the mass spectrometer from
the Instrument Configuration window.
TSQ Quantum XLS and TSQ Quantum GC User Guide
131
7
Diagnostics and Troubleshooting
Troubleshooting
• The Ethernet cable is unplugged. Verify that the Ethernet cable is connected to the
Ethernet port on the mass spectrometer. See Figure 12 on page page 25.
• The Ethernet card in the PC is not configured or is faulty. Check the configuration or
replace the Ethernet card if necessary.
• The mass spectrometer is in Service Mode. Return the electronics service switch to
operating position.
• An incorrect Ethernet cable is used for the mass spectrometer. Use the supplied 10 Base-T
Category 5 crossover cable (P/N 76396-0052, in TSQ Quantum XLS Accessory Kit).
Note You can extend the length of the Ethernet cable by plugging a standard
Category 5 cable in series with the supplied crossover cable.
Issue: Computer
intermittently loses
communication with the
mass spectrometer.
Possible cause/solution
Issue: Unable to
download methods to
the mass spectrometer.
Possible cause/solution
Issue: Acquisition does
not begin as expected.
• The Ethernet cable is loose or damaged. Inspect the cable and replace it if necessary.
• The software is not properly configured. Verify the correct settings in Instrument
Configuration.
Possible causes/solutions
• The TSQ Quantum XLS or TSQ Quantum GC start mode is not properly configured.
Verify that the instrument is configured properly. Refer to Xcalibur Help for more
information.
• The TSQ Quantum XLS or TSQ Quantum GC Instrument Setup file has incorrect
settings. Check the start time in the Instrument Setup file.
• The forepressure is too high due to solvent peak. Reduce injection volume or extend
filament/multiplier delay time until after solvent peak.
• The remote start cable to the GC is disconnected. Connect the remote start cable.
• The GC did not start. Verify GC instrument method and configuration. Verify the
connection between the autosampler and GC.
• The autosampler did not start. Verify autosampler instrument method and configuration.
Verify that the sample is present.
• The heated zone stipend is not attained. See “Heated Zone Issues” on page 136.
132
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
7
Issue: Acquisition
terminates
unexpectedly.
Diagnostics and Troubleshooting
Troubleshooting
Possible causes/solutions
• The end run time in the TSQ Quantum XLS or TSQ Quantum GC instrument method
is incorrect. Check the instrument method for the GC and the MS.
• The data system is out of disk space. Check disk space; back up and remove files.
Issue: Unable to
initialize the gas
chromatograph.
Possible causes/solutions
• The GC is not turned on. Turn on the GC.
• The GC is not configured properly in Xcalibur. Check Instrument Configuration.
• The cable between COM1 and GC is disconnected. Verify connection.
• COM1 is not configured properly. Verify COM port configuration.
Issue: Unable to
download methods to
the gas chromatograph.
Possible causes/solutions
• There is a discrepancy between the instrument method and configuration. Verify
consistency between method and instrument configuration.
• COM1 is not configured properly. Verify COM port configuration.
Issue: Unable to
initialize the
autosampler.
Possible causes/solutions
• The autosampler is not turned on. Turn on autosampler.
• The autosampler is not configured properly in Xcalibur. Check Instrument
Configuration.
• The cable between the GC and autosampler is disconnected. Verify connection.
• The autosampler instrument configuration is not set to the correct port. Verify
connection and configuration.
Issue: Unable to
download methods to
the autosampler.
Possible cause/solution
• There is a discrepancy between instrument method and configuration. Verify consistency
between instrument method and instrument configuration.
Contamination Issues
Chemical noise is always present in any mass spectrometer. As a result, the high sensitivity of
the mass spectrometer can cause new users to confuse background with a contamination
problem. Additionally, the spectra shown in Xcalibur Tune and Real-Time Display are
auto-normalized, which can make the background appear high-level.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
133
7
Diagnostics and Troubleshooting
Troubleshooting
Some chemical noise does occur, such as septum bleed after a series of injections, vial sample
bleed (which occurs if more than one injection is made from a sample vial), and siloxane peaks
that appear in the chromatogram at regular intervals from focusing at the head of the column
or in the injector.
Other possible contamination sources include hydrocarbon contamination of the carrier gas,
pump oil, or instrument cleaning solvents.
Always wear clean, lint- and powder- free gloves when handling ion source and mass analyzer
components, and ensure that the carrier gas filter, carrier gas lines, and gas regulators are free
of contamination and leaks.
Issue: Excessive
chemical background
due to a column bleed
(m/z 429, 355, 281).
Possible causes/solutions
• The capillary column has not been properly conditioned. Condition the capillary
column.
• The capillary column is damaged as a result of exposure to oxygen. Find the source of the
oxygen in the carrier gas or air leak. Recondition or replace the capillary column.
Issue: Excessive
injection port septum
bleed (Typical m/z 207,
429, 355, 281).
Possible causes/solutions
Issue: Phthalate
background (m/z 149,
167, 279).
Possible causes/solutions
• The septum is worn out or damaged. Replace the septum.
• Small pieces of septum are in the injection port liner. Replace the injection port liner;
condition the capillary column.
• Phthalate contamination has occurred due to sample handling or solvent contamination.
• Packaging materials could be the source of phthalates.
• Isolate source of phthalates such as vial lids or plastic solvent containers and remedy.
Issue: Excessive
hydrocarbon
contamination (Typical
ions are m/z 43, 57, 71,
85, 99).
Possible causes/solutions
• Carrier gas tubing is contaminated.
• Isolate source of hydrocarbon contamination and remedy.
• Replace carrier gas tubing.
• Change carrier gas filters.
134
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
7
Issue: Chemical
background due to
rhenium oxide (m/z
185/187, 201/203,
217/219, 233/235,
250/252).
Issue: Spectra are
observed due to the
following solvents:
Diagnostics and Troubleshooting
Troubleshooting
Possible causes/solutions
• These series of rhenium oxide ions come from oxidation of the rhenium filament wire
due to the introduction of air into the ion source while the filament is on.
• Check for air leaks and remedy. See “High Vacuum Issues” on page 138 for more
information.
Solvent Spectra
Acetone (m/z 43, 58, 59)
Hexane (m/z 41, 43, 56, 57, 58, 85, 86)
Methanol (m/z 31)
Methylene chloride (m/z 84, 83)
Toluene (m/z 91, 92)
Trichloroethane (m/z 151, 153)
Xylene (m/z 105, 106)
Possible causes/solutions
• There is residual solvent from a cleaning procedure or laboratory background
contamination. When you finish performing a cleaning procedure, allow cleaned
components to dry thoroughly. Warm parts in the GC oven to drive off residual solvent.
• The observed compounds have been introduced through sample injection. The ultimate
source is either a sample solvent or the autosampler rinsing solvent. Optimize GC
method to separate solvent peak from the area of interest in the chromatogram.
Filament and Lens Control Issues
The lifetime of a filament depends on its exposure to oxygen and solvent vapors. The filament
assembly protects the filament and increases its lifetime for many months.
Xcalibur diagnostics test the filament for continuity and current regulation. Testing the
filament for continuity before each acquisition ensures that an open filament condition will
stop an autosampler sequence and generate an error message.
Diagnostics test the lenses of the TSQ Quantum XLS and TSQ Quantum GC mass
spectrometers. A flat line, which represents voltage readback versus the predicted voltage
ramp, indicates a lens or other control fault. Contamination causes lens performance to
deteriorate over time; the amount of time depends on what type of sample and ionization
mode you use.
Be careful to handle the lenses with care, and do not use harsh cleaning techniques. Damaged
lenses cause short circuits, which in turn can cause damage to the lens drivers.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
135
7
Diagnostics and Troubleshooting
Troubleshooting
Issue: Diagnostics
indicate that the
filament is open.
Possible cause/solution
Issue: Unstable
emission current.
Possible cause/solution
Issue: Short
filament lifetime.
• The filament is open. Vent the system and remove the filament. Normal resistance is
1.0 Ω. Replace if open.
• The filament is near the end of its life span. Replace the filament.
Possible causes/solutions
• There is an air leak contributing to short filament lifetime. Check for leaks; repair if any
are found. See “High Vacuum Issues” on page 138 for more information.
• The filament is on during the solvent peak. Increase acquisition delay time until the
solvent peak has passed.
• High emission current is being used. Use lower emission current to extend lifetime.
Issue: Diagnostics
indicate a flat lens
response.
Possible causes/solutions
• The power supply to the lens has a fault. Contact Thermo Fisher Scientific Technical
Support.
• There is a fault with the lens drivers. Contact Thermo Fisher Scientific Technical
Support.
Heated Zone Issues
The ion source and transfer line are heated zones. The ion source heater is controlled by the
mass spectrometer and the transfer line heater is controlled by the Aux1 heated zone of the
TRACE GC.
A heated zone problem is often the result of downloading an instrument method to the
instrument that has a different setpoint than the current setting, causing a delay while the
heated zone heats or cools.
Component failures are less common but can occur. These are usually related to open circuits
in heater cartridges or faulty temperature sensors.
This manual does not discuss heated zones in the gas chromatograph.
Issue: Excessive
chromatographic peak
tailing.
136
Possible causes/solutions
• The ion source, transfer line, or both are not hot enough. Increase the transfer line
temperature. It should be at least as hot as the highest GC oven temperature. Then, try
increasing the ion source temperature.
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
7
Diagnostics and Troubleshooting
Troubleshooting
• The sample analyte is adsorbing in the GC injector. Clean and deactivate the injection
liner. You can also try liners made of different materials.
• The GC oven is not ramped to a high enough temperature. Extend upper oven
temperature.
• The GC column must be replaced or does not have the appropriate stationary phase for
your application. Change the GC column. See “Removing and Installing a GC Capillary
Column” on page 122.
• The GC column does not extend far enough past the end of the transfer line tip. If the
end of the column is inside the tip, an excessive amount of GC effluent will contact the
inside wall of the tip. Follow the procedure described in “Installing a GC Capillary
Column” on page 124.
Issue: Source heater
will not heat.
Possible causes/solutions
• No contact between the ion source temperature sensor and the heater block. For accurate
temperature readings of the heater block, fully screw the threaded standoff into the ion
source to ensure contact between the temperature sensor and the heater block. See
“Reassembling the L1, L2, L3 Lens Assembly” on page 93.
• One or more heater cartridges on the EI/CI Source PCB is defective. Replace the EI/CI
Source PCB. See “Disassembling and Reassembling the TSQ Quantum XLS Ion Source
Completely” on page 107 (TSQ Quantum XLS), or “Disassembling and Reassembling
the TSQ Quantum GC Ion Source Completely” on page 110 (TSQ Quantum GC).
• The ion source temperature sensor (RTD) is defective. Replace the EI/CI Source PCB.
See “Disassembling and Reassembling the TSQ Quantum XLS Ion Source Completely”
on page 107 (TSQ Quantum XLS), or “Disassembling and Reassembling the TSQ
Quantum GC Ion Source Completely” on page 110 (TSQ Quantum GC).
Issue: Source
heater overheats.
Possible cause/solution
Issue: Transfer
line will not heat.
Possible causes/solutions
• The ion source temperature sensor (RTD) is defective. Replace the EI/CI Source PCB if
necessary. See “Disassembling and Reassembling the TSQ Quantum XLS Ion Source
Completely” on page 107 (TSQ Quantum XLS), or “Disassembling and Reassembling
the TSQ Quantum GC Ion Source Completely” on page 110 (TSQ Quantum GC).
• Aux1 zone of the GC is not configured for MS Transfer Line. Configure Aux1 zone of the
GC and set transfer line temperature in the GC Method Editor.
• Transfer line heater elements are defective. Replace the transfer line.
• The transfer line temperature sensor (RTD) is defective. Replace the transfer line.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
137
7
Diagnostics and Troubleshooting
Troubleshooting
Issue: Transfer
line overheats.
Possible cause/solution
• The transfer line temperature sensor is defective. Replace the transfer line.
High Vacuum Issues
High vacuum problems can manifest themselves in two ways:
• An intermittent vacuum condition (the vacuum pressure in the vacuum manifold
fluctuates intermittently) can cause chromatographic signals to drop out, or, if the
pressures exceed the maximum allowed pressures set in Xcalibur (See Table 3 on page 57),
then the TSQ Quantum XLS or TSQ Quantum GC turns off.
• If the vacuum is consistent enough that it does not exceed the maximum allowed pressure
and avoids the Xcalibur automatic shutdown, non-reproducible false chromatographic
peaks can be generated in the chromatogram.
Typical forepressure readbacks are 30 to 40 mTorr (in EI mode), and typical manifold
pressure (ion gauge readback) is 2 × 10-5 Torr with argon collision gas on, and 2 × 10-6 Torr
with Ar collision gas off.
The most reliable way to find vacuum leaks is to spray a gas around the vacuum manifold and
look for the characteristics ions in full-scan EI mode. Argon produces m/z 40. Alternatively,
you can use compressed electronic dusting spray containing an HFC. For example, Falcon®
Dust-Off®and MicroCare® Micro-Blast™ contain tetrafluoroethane, which produces ions at
m/z 69 and 83.
Issue: Forepump will
not turn on.
Possible causes/solutions
• The forepump is off. Check the forepump switch.
• The vacuum service switch is in the Off position. Place the vacuum service switch in the
operating position.
• The forepump power cable from the TSQ Quantum XLS or TSQ Quantum GC is not
connected. Connect the power cable.
• The forepump is faulty. Replace the forepump.
Issue: Forepump
powers on, but will not
pump down.
Possible causes/solutions
• The oil level in the forepump is insufficient. Check oil level; add oil if necessary.
• The foreline is leaking. Check the clamps and connectors. Replace the foreline hose if a
hole is found.
• The vacuum manifold cover is leaking.
• The forepump is faulty. Replace the forepump.
138
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
7
Issue: Turbo pump
shuts off during
operation.
Issue: Unexpected full
ventilation occurs.
Diagnostics and Troubleshooting
Troubleshooting
Possible cause/solution
• The foreline pressure is too high. Check for leaks in the foreline.
Possible causes/solutions
• A gross leak is present. Check for leaks.
• The system was vented through the inlet valve. Close and plug the inlet valve.
• The GC column broke at the transfer line. Replace the GC column. See “Removing and
Installing a GC Capillary Column” on page 122.
• The foreline was cut. Replace the foreline hose.
Issue: Vacuum is faulty.
Possible Solutions
• A gross leak is present. Check for leaks.
• The pressure in the analyzer region of the vacuum manifold, as measured by the ion
gauge, must be below the pressures listed in Table 3 on page 57.
Linearity Issues
Linearity issues occur when a plot of intensity versus concentration of a known compound is
not linear. Poor instrument operating conditions can cause linearity problems. Additionally,
certain compounds do not give a desired linear response due to chromatographic activity.
A well-maintained instrument provides good linear response over a wide range of
concentrations for most compounds. Like any instrument, however, the TSQ Quantum XLS
and TSQ Quantum GC has a saturation point.
Perform routine injector and column maintenance to minimize linearity problems. Usually, a
hardware fault that affects linearity shows different issues than those that might be attributed
to linearity.
Issue: Calibration plots
not sufficiently linear.
Possible causes/solutions
• High-end standards are too concentrated for the MS. Use the split injection technique to
decrease the amount of sample or lower emission current to reduce MS sensitivity.
• The ion volume and lenses are dirty. Clean ion volume and lenses as described in
“Cleaning Ion Source Components” on page 85.
• The electron multiplier setting is incorrect. Run Auto Tune - Calibration as described in
“Tuning and Calibrating” on page 63.
• The injection port liner or capillary column is dirty. Change the injection port liner and
trim the capillary column.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
139
7
Diagnostics and Troubleshooting
Troubleshooting
• The capillary column stationary phase is too thin for high concentration samples. Use a
higher capacity capillary column with a thicker stationary phase, or use a split injection
technique.
• The capillary column is bad. Replace the capillary column. See “Removing and Installing
a GC Capillary Column” on page 122.
Power Supply Issues
Xcalibur diagnostics detect most power supply issues. Power supply problems often involve a
blown fuse, faulty electronic components, or something as simple as a disconnected cable.
Fuses should be replaced by a Thermo Fisher Scientific Field Service Engineer.
Issue: TSQ Quantum
XLS will not power on.
Possible causes/solutions
• The power cord is disconnected. Verify that the power cord is plugged in.
• Voltage is not coming from the electrical outlet. Verify that the electrical outlet is
operational.
• The Power Module is faulty. Contact Thermo Fisher Scientific Technical Support.
Issue: The TSQ
Quantum XLS powers
on, but trips the circuit
breaker.
Possible causes/solutions
• The Power Module is faulty. Contact Thermo Fisher Scientific Technical Support.
• The forepump causes the circuit breaker to trip. Check forepump; replace it if necessary.
Sensitivity Issues
If you observe a drop in instrument sensitivity, you should determine if the sensitivity drop
was sudden or if it occurred gradually. A sudden loss of sensitivity can be the result of sudden
component failure or an unnoticed change in the analytical method. Simple errors such as a
plugged autosampler syringe or too low sample level in the sample vial can give the
appearance of instrument failure.
Gradual drops in sensitivity are usually the result of ion volume or lens contamination, and
are easily remedied by cleaning the ion volume or lenses.
The electron multiplier influences sensitivity and has a limited lifetime. Eventually the
electron multiplier must be replaced.
Improper GC maintenance is another cause of diminished sensitivity. It is important to
establish a routine maintenance program for the GC. See the TRACE GC Maintenance
Manual for more information on establishing a maintenance program.
140
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
7
Issue: The
chromatogram has a
low total ion current
signal.
Diagnostics and Troubleshooting
Troubleshooting
Possible causes/solutions
• The GC temperature ramp does not continue to a high enough temperature to elute high
boiling point compounds. Multiple injections cause these compounds to accumulate in a
column, reducing sensitivity. Extend upper temperature or the time at upper temperature
in the GC oven ramp.
• The instrument is out of tune, or the tune file is incorrect. Select correct tune file for the
method or run automatic tune as described in “Tuning and Calibrating” on page 63.
• The ion volume is contaminated. Clean the ion volume.
• The ion volume is incorrectly positioned. Position the ion volume properly.
• The magnets above and below the ion source are installed incorrectly. Position both
magnets so the south pole is on top. Electrons will not focus into the ion volume if one
magnet is upside down.
• Dust has collected in the electron multiplier or on the conversion dynode. Contact
Thermo Fisher Scientific Technical Support.
• The emission current is set too low. Check the setting listed for the emission current.
Choose Tune > Manual and select the Controls tab.
• The ground connection between the electrometer PCB and anode feedthrough is faulty.
Contact Thermo Fisher Scientific Technical Support.
• There is a problem with the filament or lens control. See “Filament and Lens Control
Issues” on page 135 for more information.
• The EI/CI Source PCB is faulty, allowing emission current to leak to ground. Contact
Thermo Fisher Scientific Technical Support.
• Reagent gas is leaking into the analyzer, suppressing EI signal. Check for presence of CI
reagent ions in spectrum. Replace CI Gas Flow Module if necessary.
Issue: Poor
compound sensitivity.
Possible causes/solutions
• The syringe, injection port liner, and column depth in injector are incorrectly matched.
See TRACE GC Ultra Operators Manual for the correct combination.
• Sample delivery is insufficient due to a plugged syringe needle. Clean or replace the
syringe.
• The injection port liner is contaminated. Clean or replace the injection port liner.
• Graphite or septa particles contaminate the injection port. Clean the injection port.
• The injector or septum is leaking. Replace septum and perform leak check on the GC.
• The capillary column is at the end of its life span. Replace the capillary column.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
141
7
Diagnostics and Troubleshooting
Troubleshooting
• Method development problems are present. Contact Thermo Fisher Scientific Technical
Support.
Issue: Sensitivity is
unstable or shows
decrease with repeated
injections.
Possible causes/solutions
• The GC temperature ramp does not continue to a high enough temperature to elute high
boiling point compounds. Multiple injections cause these compounds to accumulate in a
column, reducing sensitivity. Extend the upper temperature or the time at upper
temperature in the GC oven ramp.
• The ion volume or lenses are contaminated. Clean the ion volume and lenses as described
in “Cleaning Ion Source Components” on page 85.
• The ion source temperature is too low and causes the ion source to contaminate too
quickly. Clean ion volume and lenses as described in “Cleaning Ion Source Components”
on page 85 and then raise ion source temperature.
• There is a problem with the filament emission current control. See “Filament and Lens
Control Issues” on page 135 for more information.
• The electron multiplier is faulty. Contact Thermo Fisher Scientific Technical Support.
• The injection port liner or capillary column is contaminated. Replace the injection port
liner and trim the capillary column.
Issue: Poor
sensitivity in
CI mode.
Possible causes/solutions
• An EI ion volume is installed. Install a CI ion volume as described in “Changing
Ionization Modes” on page 73.
• The small hole in the CI ion volume is plugged. Use a dental pick or old syringe needle to
clear it.
• The filament is not aligned. Remove the ion volume and check if the ion burn is centered
around the small electron entrance hole. Be sure the filament is properly inserted into the
connector. Carefully bend the filament wire to better align it with the ion volume.
• The ion volume is not inserted properly. While running the instrument, you can push on
the ion volume with the I/R tool. Be sure to monitor the pressure to ensure that you get a
good seal around the I/R tool. Otherwise, you may damage the filament. An increase in
signal calibration gas as you push on the ion volume usually indicates that the filament is
not aligned properly.
Stability Issues
Stability helps provide consistent instrument precision and the reproducibility of accurate
results. Good operating conditions for the mass spectrometer, gas chromatograph, and
autosampler contribute to instrument stability.
142
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
7
Diagnostics and Troubleshooting
Troubleshooting
Sample preparation, spiking errors, sample injection errors, and lack of routine maintenance
on the instruments can cause false stability issues.
When hardware faults affect instrument stability, investigate simple solutions first, such as
cleaning the ion volume and lenses, or checking for air leaks. Usually, a hardware fault that
affects stability shows different issues than those which might be attributed to stability.
Issue: The signal
response is unstable or
drops out unexpectedly.
Possible causes/solutions
• There is a problem with the filament or lens control. See “Filament and Lens Control
Issues” on page 135 for more information.
• There is an air leak. Check for leaks. See “High Vacuum Issues” on page 138 for more
information.
• There is a high vacuum problem. See “High Vacuum Issues” on page 138 for more
information.
• There is a contamination problem. See “Contamination Issues” on page 133 for more
information.
• Dust has collected in the electron multiplier or on the conversion dynode. Contact
Thermo Fisher Scientific Technical Support.
Tuning Issues
You can suspect a tuning problem when Xcalibur Auto Tune - Calibration fails. Auto Tune Calibration performs several functions, and issues or error messages indicate different
problems. Diagnostics can usually uncover a tuning problem.
Issue: “Cannot find
Calibration Gas” error
message received.
Possible causes/solutions
• There is a mechanical problem with the ion source or lenses. Verify that an EI ion volume
is installed. Verify correct orientation and cleanliness of ion volume. Verify cleanliness and
correct operation of lenses.
• Tune file settings are out of usable range. Restore default tune settings and calibration
settings. Run Auto Tune - Calibration. See “Tuning and Calibrating” on page 63.
• The calibration gas vial is empty. Add 100 μL (max) of calibration compound to
calibration gas vial. See “Adding Calibration Compound” on page 117.
Issue: The electron
multiplier gain
calibration fails.
Possible causes/solutions
• The electron multiplier has not been given sufficient time to outgas since venting. Allow
more time to pump out.
• The GC column flow is too high. Lower it to1 mL/min.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
143
7
Diagnostics and Troubleshooting
Troubleshooting
• The multiplier is near the end of its lifetime. Multipliers typically last about 2 to 3 years
before they are too noisy for the gain to be set accurately. Contact Thermo Fisher
Scientific Technical Support.
• The multiplier power supply is faulty. Contact Thermo Fisher Scientific Technical
Support.
• The filament is the source of too much background noise. See “Filament and Lens
Control Issues” on page 135 for more information.
• Chemical background in the manifold is elevated. Remedy leaks or sources of water
contamination in carrier gas. See “High Vacuum Issues” on page 138 for more
information.
• The difference between the electron energy and the setpoint is greater than 5 V. Contact
Thermo Fisher Scientific Technical Support for assistance.
Issue: Poor
high mass
response.
Possible causes/solutions
• Poor high m/z ion intensity because the ion source temperature is too high. Reduce the
ion source temperature to reduce the amount of thermal decomposition and
fragmentation of your analyte.
• Poor high m/z ion intensity because of bad ion source lens settings. Run Auto Tune Calibration. See “Tuning and Calibrating” on page 63.
• Helium pressure in the ion source is too high or low. Set flow to 1.0 mL per minute for
most applications.
• There is a vacuum leak. Find and repair any leaks. Be sure to check the transfer line fitting
in the GC oven. See “High Vacuum Issues” on page 138 for more information.
• An excess amount of low m/z ions, such as hydrocarbons or column bleed. Remedy the
source of these low m/z ions.
• The ion volume or lenses are contaminated. Clean the ion volume or lenses. See
“Cleaning Ion Source Components” on page 85.
• m/z 131 is not base peak because the multiplier gain is too low. The multiplier gain
calibration may set the multiplier voltage too low for a noisy multiplier. Electron
multipliers older than 2 to 3 years are often noisy. Contact Thermo Fisher Scientific
Technical Support.
• There is poor high m/z intensity and poor resolution of low m/z ions because the
conversion dynode is faulty. Contact Thermo Fisher Scientific Technical Support.
• The injection rf is not calibrated. Run Auto Tune - Calibration as described in “Tuning
and Calibrating” on page 63.
• There is a problem with the filament or lens control. See “Filament and Lens Control
Issues” on page 135 for more information.
144
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
7
Diagnostics and Troubleshooting
Replacing a Fuse
• Reagent gas is leaking into the analyzer, suppressing EI signal. Check for the presence of
CI Reagent ions in spectrum. Replace the CI Reagent Gas Flow Module if necessary.
Issue: Weak signal.
Possible causes/solutions
• The ion volume or ion source lenses are dirty. Clean the contaminated components as
described in “Cleaning Ion Source Components” on page 85.
• The multiplier is set too low. Run multiplier gain calibration in Auto Tune - Calibration
as described in “Tuning and Calibrating” on page 63. An electron multiplier older than
2–3 years may be too noisy for the multiplier gain calibration to accurately set the voltage.
The electron multiplier might need to be replaced. Contact Thermo Fisher Scientific
Technical Support.
• The wrong type of ion volume is installed. EI and CI require different ion volumes. They
may not be used interchangeably. Change the ion volume as described in “Changing
Ionization Modes” on page 73.
• The ion volume is absent or incorrectly positioned. Verify that the ion volume is
positioned correctly.
• There is a filament or lens control problem. See “Filament and Lens Control Issues” on
page 135 for more information.
• The electron multiplier is faulty. Contact Thermo Fisher Scientific Technical Support.
Replacing a Fuse
CAUTION Fuses protect the various circuits by opening the circuits whenever overcurrent
occurs. On the TSQ Quantum XLS and TSQ Quantum GC mass spectrometers, a failed
fuse indicates a failed board or electronic module that must be replaced by a Thermo
Fisher Scientific Field Service Engineer.
Replacing PCBs and Power Supplies
CAUTION The TSQ Quantum XLS and TSQ Quantum GC mass spectrometer
electronic assemblies are close-packed to minimize the size of the system. Due to the
complexity of removing and reinstalling the mass spectrometer electronic assemblies, only
a Thermo Fisher Scientific Field Service Engineer can replace electronic assemblies.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
145
7
Diagnostics and Troubleshooting
Replacing PCBs and Power Supplies
146
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
8
Using the Direct Sample Probe
With the direct sample probes (direct insertion probe and direct exposure probe) you can
introduce compounds directly into the ion source without GC column separation. This
chapter covers, step-by-step, a direct insertion probe (DIP) experiment using
perfluorotetracosane (CF3(CF2)22CF3, molecular weight 1238.18, melting point 190 ºC).
You can perform a similar experiment using your compound.
P
Contents
• Creating an Instrument Method
• Creating a Sequence
• Preparing the Probe and Inlet Valve
• Preparing the Mass Spectrometer
• Running the Sequence
• Examining the Raw Data in Qual Browser
• Removing the Probe
Creating an Instrument Method
An instrument method contains the settings that the TSQ Quantum XLS or TSQ Quantum
GC mass spectrometer uses during an experiment. Use the Instrument Setup view to create
the instrument method.
Y To create an instrument method
1. On the Xcalibur Home Page, click the Instrument Setup icon to open the Instrument
Setup view.
2. Click the TSQ icon in the viewbar.
3. Choose TSQ > Regular Method to open the regular method Scan Editor page.
4. On the Scan Editor page, create a one segment, one scan event instrument method. In
this example, the segment is 8 minutes long (see Figure 65).
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
147
8
Using the Direct Sample Probe
Creating an Instrument Method
5. Specify the mass spectrometer settings, for this example, full-scan scan type, Q1MS scan
mode, and positive polarity.
Figure 65. Scan Editor page of Instrument Setup
6. Click the EI/CI tab to display the EI/CI page (see Figure 66).
148
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
8
Using the Direct Sample Probe
Creating an Instrument Method
Figure 66. EI/CI page of Instrument Setup
7. Enter or select the following settings:
• Number of States: 1
• State at Start of Run: On
• Calibration Gas Setting: Off
• Use DIP/DEP Probe:
This example is an EI experiment with a 200 μA emission current.
8. Save the instrument method.
a. Choose File > Save As to open the Save As dialog box.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
149
8
Using the Direct Sample Probe
Creating a Sequence
b. Save the instrument method as a .meth file. In this example we name the instrument
method Dip test.meth.
Creating a Sequence
A sequence contains sample information—one sample per row. In this example the sequence
has one sample.
Note If the sequence has more than one sample, Xcalibur prompts you to reload the
probe after each sample.
Y To create a sequence
1. On the Xcalibur Home Page, click the Sequence Setup icon to open the Sequence Setup
view (see Figure 67).
2. Enter the name and path of the instrument method. In this example, it is
C:\Xcalibur\methods\DIP test.
3. Enter a file name and path for the raw file that contains the acquired data. In this
example, it is C:\Xcalibur\Data\data01.
Figure 67. Sequence Setup view, showing one sample in the sequence
150
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
8
Using the Direct Sample Probe
Preparing the Probe and Inlet Valve
Preparing the Probe and Inlet Valve
Y To prepare the probe and inlet valve
1. Connect the direct sample probe controller to the mass spectrometer by contact closure
with the probe start cable (P/N 70111-63627, in TSQ Quantum XLS or TSQ Quantum
GC Accessory Kit).
2. Create a probe method on the controller. Refer to the probe manual. This example uses
the following DIP settings:
• Initial temperature: 50 ºC
• Initial time: 5 s
• Ramp rate: 100 ºC/min
• Final temperature: 200 ºC
• Ramp hold time: 200 s
3. Load sample on the probe.
4. Install the guide bar.
a. With the guide ball track facing left, insert the guide bar into the entry housing. See
Figure 16 on page 29.
b. Push the guide bar in as far as it goes. Rotate it 90° clockwise to lock the guide bar in
the entry housing.
5. Prepare the inlet valve.
a. Make sure the inlet valve is closed with the inlet valve lever down, as shown in
Figure 16 on page 29.
b. Loosen the inlet valve knob counter-clockwise, and remove the inlet valve plug.
Preparing the Mass Spectrometer
Y To prepare the TSQ Quantum XLS for the experiment
1. Choose Start > All Programs > Thermo Instruments > TSQ > TSQ Tune to open the
EZ Tune window if it is not already open.
IMPORTANT The TSQ EZ Tune window must be open for the mass spectrometer to
initiate the probe insertion sequence.
2. Ensure that the TSQ Quantum XLS or TSQ Quantum GC mass spectrometer is the only
instrument in use and that there are no start instruments.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
151
8
Using the Direct Sample Probe
Preparing the Mass Spectrometer
a. On the Xcalibur Home Page, click the Sequence Setup icon to open Sequence Setup,
if it is not already open.
b. In Sequence Setup, choose Actions > Run Sequence. The Run Sequence dialog box
opens (see Figure 68).
Figure 68. Run Sequence dialog box, showing instruments in use
c. Click Change Instruments to open the Change Instruments In Use dialog box.
152
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
8
Using the Direct Sample Probe
Running the Sequence
Figure 69. Change Instruments In Use dialog box, showing the TSQ Quantum mass
spectrometer in use and no start instruments
d. Click any Yes in the In Use column to clear all instruments other than the TSQ
Quantum.
e. Ensure that the TSQ Quantum has a Yes in the In Use column. Click the area if
necessary.
f.
Click any Yes in the Start Instrument column to clear all start instruments.
g. Click OK to save your settings and close the Change Instruments In Use dialog box.
Running the Sequence
You acquire data by running the sequence. Xcalibur saves the acquired data in a raw file.
Note You must have the TSQ EZ Tune or QuickQuan application running so that
Xcalibur can display instrument methods.
Y To run the sequence
1. In the Run Sequence dialog box (see Figure 68 on page 152), click OK to start the
acquisition. Xcalibur displays the Insert Probe message.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
153
8
Using the Direct Sample Probe
Running the Sequence
Figure 70. Insert Probe message
2. Inset the probe into the inlet valve.
a. Insert the guide ball on the DIP probe into the guide ball hole on the guide bar.
b. Slide the probe forward into the inlet valve until the guide ball is at the guide bar’s
first stop. See Figure 71.
c. Tighten the inlet valve knob clockwise to ensure a leak-tight seal.
Figure 71. Probe at the first stop on guide bar
3. Click OK. The forepump evacuates the inlet valve.
4. When the Safe to Open Valve message to appears, click OK.
154
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
8
Using the Direct Sample Probe
Running the Sequence
Figure 72. Safe to Open Valve message
5. Pull the inlet valve lever up to open the inlet valve.
6. Slide the probe into the vacuum manifold until the tip of the probe is fully inserted into
the ion volume holder. See Figure 73.
Figure 73. Probe inserted into the inlet valve
7. When you receive the Acquisition Ready message, click OK. The TSQ Quantum XLS is
now waiting for contact closure.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
155
8
Using the Direct Sample Probe
Running the Sequence
Figure 74. Acquisition Ready message
.
8. In Sequence Setup, choose View > Real Time Plot View to display the mass spectrum
and chromatogram.
9. Start the probe method from the probe controller to initiate data acquisition. See
Figure 75 and Figure 76.
Figure 75. Perfluorotetracosane mass spectrum at 0.56 minute retention time
156
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
8 Using the Direct Sample Probe
Examining the Raw Data in Qual Browser
Figure 76. Perfluorotetracosane chromatogram
Examining the Raw Data in Qual Browser
You can use Qual Browser to display the mass spectrum at various times in the chromatogram.
Y To examine the raw data
1. On the Xcalibur Home Page, click the Qual Browser icon to open the Qual Browser
window.
2. Click on the chromatogram to display the mass spectrum at that retention time. See
Figure 77.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
157
8
Using the Direct Sample Probe
Removing the Probe
Figure 77. Raw file displayed in Qual Browser
Removing the Probe
Y To remove the probe
1. Withdraw the probe until the guide ball on the probe reaches the first stop on the guide
bar (Figure 71).
2. Close the inlet valve by pushing the lever down.
CAUTION Close the inlet valve first. Otherwise, the system vents to the atmosphere.
Do not withdraw the probe beyond the point where the guide ball reaches the first
stop in the guide bar.
3. Loosen the inlet valve knob by turning it counter-clockwise.
4. Continue withdrawing the probe completely from the inlet valve by sliding it through the
guide ball track in the guide bar.
158
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
8
Using the Direct Sample Probe
Removing the Probe
5. Store the probe in its case.
6. Remove the guide bar by rotating it 90° counter-clockwise and sliding it out of the entry
housing.
7. Replace the inlet valve plug and tighten the inlet valve knob clockwise to form a seal. The
inlet valve plug prevents air from entering the vacuum manifold in case the inlet valve is
accidentally opened. You can also remove the inlet valve lever by pulling it free.
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
159
9
Replaceable Parts and Consumables
This chapter contains TSQ Quantum XLS and TSQ Quantum GC part numbers for
replaceable parts and consumables. To ensure proper results in servicing the TSQ Quantum
XLS or TSQ Quantum GC mass spectrometer, order only the parts listed or their equivalent.
Contact Thermo Fisher Scientific San Jose and have your TSQ Quantum XLS or TSQ
Quantum GC serial number ready.
These kits with replaceable parts and consumables are available for the TSQ Quantum XLS
and TSQ Quantum GC:
Contents
• Accessory Kit
• Chemicals Kit
Accessory Kit
Table 7. Accessory Kit (P/N 70111-62077) (Sheet 1 of 2)
Description
Thermo Scientific
Function
Part number
Fitting, ferrule, SWG, 1/8 in., brass, set
00101-08-00009
Fitting, connection, 1/4 to 1/8 in.
00101-01709
Fitting, connection, 1/4 to 1/8 in.
00101-01712
Fitting, ferrule, SWG, front, 3/8 in.,
brass
00101-11500
Ferrule, 0.4mm ID × 1/16
00101-18100
Restek® capillary-grade hydrocarbon
trap
For removing hydrocarbons from the GC
carrier gas
00106-99-00001
Restek high-capacity oxygen trap
For removing oxygen from the GC carrier
gas
00106-99-00002
Cotton swabs
For applying cleaning paste to stainless
steel parts
00301-01-00015
Tubing, 1/8 in., copper, pre-cleaned
Gas lines
00301-22701
Ball valve seal kit (replacement)
For replacing the ball valve seal in the inlet
valve
119265-0003
TSQ Quantum XLS and TSQ Quantum GC User Guide
161
9
Replaceable Parts and Consumables
Chemicals Kit
Table 7. Accessory Kit (P/N 70111-62077) (Sheet 2 of 2)
Description
Function
Part number
Lens alignment tool
For aligning ion source lenses 1, 2, and 3
120271-0001
Filament assembly
For replacing a failed filament in the ion
source
120320-0030
Nitrile gloves medium
For handling clean parts that are under
vacuum
23827-0008
Nitrile gloves large
For handling clean parts that are under
vacuum
23827-0009
Syringe, 10 uL, 80mm
For adding calibration compound
36502019
Tee union, 1/8 in.
00101-01-00012
Cable, start, probe
Contact closure cable between the direct
sample probe and the mass spectrometer
70111-63627
Cable, start, GC
Contact closure cable between the GC and
the mass spectrometer
70111-63626
Forceps, 10 in., stainless steel
For handling ion volumes and clean
components
76360-0400
Cable, 10 base-T, crossover, 15-ft,
shielded
76396-0052
Chemicals Kit
Table 8. Chemicals Kit (P/N 70111-62078)
162
Description
Function
Part number
Perfluorotributylamine (FC-43)
Tuning and calibration compound
50010-02500
Aluminum Oxide
Cleaning stainless steel parts
32000-60340
Benzohexane
Performance testing
120150-TEST
Octafluoronaphthalene
Performance testing
120150-TEST
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
I
Index
Symbols
133
A
Accessory Kit 161
aluminum oxide 162
analyzer assembly, lenses, voltages applied to 39
analyzer chamber,location (figure) 43
Analyzer Control PCB, discussion 47
analyzer region, vacuum manifold, description 42
analyzer region, vacuum manifold, location 42
and 41
anode, electron multiplier, description 39
Auto Tune - Calibration 66
automatically 66
autosampler
cofiguration 18
photo 2
power outlet 24
startup 55, 58
TriPlus 18
troubleshooting communication problems 133
AutoSIM scan type 14
B
ball valve
extraction tool (figure) 121
extraction tool, using 121
seal, replacing 121
ball valve seal kit 161
benzohexane 162
C
cables 162
calibrating
and H-SRM (note) 63
discussed 63
frequency (note) 63
Thermo Scientific
running 66
tune and calibration report 69
calibration compound
adding 117
flow control, description 46
mass spectrum, displaying 64
observed peaks 64
on, off 64
spectrum, positive EI mode 66
calibration compound vial
location (figure) 118
removing and reinstalling 118
calibration gas valve, description 46
capillary column
conditioning 125
installing 124
removing 122
removing ion source (Caution) 89, 96
cathode, electron multiplier, description 39
Cautions
emergency shutdown 23, 51
forepump and line power 44
I/R tool removal, withdrawing 79
Ion Source PCB removal 109, 111
removing ion source and capillary column 89, 96
replacing fuses 145
replacing PCBs and power supplies 145
centroid scan, defined 15
chemical ionization (CI)
discussed 4
gas valve, description 46
methane 4
removing ion volume 73
Chemicals Kit 162
circuit breaker 23–24
cleaning
aluminum parts 115
ceramic parts 115
frequency 83
ion volumes 88
lenses L1, L2, L3, L4 88, 95
TSQ Quantum XLS User Guide
163
Index: D
list (table) 83
stainless steel parts 113
collision cell
figure 34
location (figure) 43
collision energy (Q2 offset voltage) 38
collision gas
description 36
collision gas valve, description 45
collision-induced dissociation
discussion 36
column bleed contamination 133
Communication LED
description 22
figure 22
mass spectrometer reset 59
mass spectrometer startup 57
compliance
regulatory iii
WEEE vii
computer, features 48
consumables 161
contamination
column bleed 133
septum bleed 133
solvent ions 135
troubleshooting 133
Convectron gauge, description 45
conversion dynode
description 39
figure 40
D
data dependent scan mode, discussed 11
data system
description 48
GC interface 48
instrument interface 48
LAN interface 49
personal computer 48
primary Ethernet adapter 48
shutdown 55
data types, discussed 14
detector system 39
diagnostics 129
running 129
DIP, example procedure
1 creating an instrument method 147
2 creating a sequence 150
3 preparing the probe and inlet valve 151
4 preparing the mass spectrometer 151
164
TSQ Quantum XLS User Guide
5 running the sequence 153
6 examining the data 157
perfluorotetracosane 147
direct insertion probe (DIP)
perfluorotetracosane example procedure 147
direct sample probe
and inlet valve 151
direct exposure probe (DEP) 20
direct insertion probe (DIP) 20
inserting into inlet valve (figure) 154
mass spectrometer, preparing 151
photo 20
using 147
dynode 39
E
electron ionization (EI)
discussed 4
removing ion volume 73
electron multiplier
anode, description 39
description 39
figure 40
electronic assemblies
description 46
ion detection system 47
Power Entry Module, description 46
RF voltage generation, description 47
electronics service switch
description 24
location (figure) 24–25
mass spectrometer components On/Off status 60
embedded computer
description 46
resetting 58
emergency shutdown
Caution 23
front panel system power off button 23
procedure 51
system power off button 51
Ethernet 48
Ethernet Link OK LED 25
F
FC-43
adding 117
mass spectrum, displaying 64
observed peaks 64
on, off 64
part number 162
spectrum, positive EI mode 66
Thermo Scientific
Index: G
ferules 161
filament
description 28
location (figure) 26–27, 104, 106
on, off 64
part number 102, 105
problems, troubleshooting 135
replaceable parts 162
replacing 102, 105
filters 46
fittings 161
forepump
description 44
maintainence 116
power cord (Caution) 44
forpump
power outlet (figure) 25
front panel LEDs
Communication 22
description 22
figure 22
Power 22
Scan 23
System 23
Vacuum 22
front panel system power off button
figure 23
full scan type 12
functional description
autosampler 18
data system 48
direct sample probes 20
gas chromatograph (GC) 19
inlet valve 29
ion optics 30
mass analyzer 32
mass spectrometer 21
Q0 quadrupole 30
transfer line 20
TSQ Quantum GC 17
fuses, mass spectrometer, replacing 145
G
gas chromatograph (GC)
capillary column, installing 124
capillary column, removing 122
column installation 124
communication with data system 48
configuring 19
functional description 19
leak check 125
photo 2
Thermo Scientific
power outlet 24
startup 55–56
troubleshooting communication problems 133
guide bar
figure 29, 74
intalling 73
H
hydrocarbon trap 161
I
inlet gasses hardware
calibration compound 46
CI gas valve 46
collision gas valve 45
description 41
functional block diagram 41
inlet valve
and direct sample probe 151
Caution 79
components (figure) 75
description 29
figure 29
guide bar 29
Insert Probe message 75
insertion/removal (I/R) tool, inserting 76
ion volumes, installing 80
lever, removing 82
plug orientation 82
procedure 74
Safe to Insert Probe message 77
Insert Probe message (figure) 75
insertion/removal (I/R) tool
figure 29, 74
inserting 76
ion volumes, removing and installing 79
lock position 78
unlock position 74
withdrawal (Caution) 79
Instrument Configuration window, opening 18–19
instrument method, creating 147
intelligent selected reaction monitoring (iSRM) 13
ion detection system
conversion dynode 39
description 39
electron multiplier 39
electron multiplier gain 40
electronic assemblies 47
ion gauge
description 45
location (figure) 43
TSQ Quantum XLS User Guide
165
Index: L
ion optics
description 30
ion polarity modes,discussed 5
ion source
cleaning 102, 105
cleaning components 85
cross sectional view (figure) 31
description 25
dissassembling completely 107, 110
electrical connections 89, 96
exploded view (figure) 86–87
figure 26–27
filament, replacing 102, 105
filament,replacing 102, 105
ion source lenses, reinstalling 94, 101
Ion Source PCB, location (figure) 108, 111
lens assembly, disassembling 92, 98
lens assembly, removing 91, 98
lenses, cleaning 88, 95
lenses, reassembling 93, 101
less assembly, exploded view 93, 100
location (figure) 89, 96
reinstalling 94, 102
removing 89, 96
ion source block
location (figure) 26–27
ion source chamber
location (figure) 43
Ion Source PCB
location (figure) 26–27, 89, 96, 107, 110
reinstalling 109, 112
removal 109, 111
removal (Caution) 109, 111
ion transmission device
defined 6
rod assembly 6
ion volumes
attaching and removing from I/R tool 79
cleaning 88
figure 28
installing 80
location (figure) 81
removing 73
ionization modes
changing 73
chemical ionization (CI) 3
discussed 3
electron ionization (EI) 3
iSRM scan type 13
L
leak check, GC 125
166
TSQ Quantum XLS User Guide
leak checking 138
LEDs
Communication LED 22
Communication LED, mass spectrometer reset 59
Communication LED, mass spectrometer startup 57
Ethernet Link OK 24–25
Forepump On 24
Power LED 22
Power LED, mass spectrometer reset 59
Power LED, mass spectrometer startup 57
Pump On, location (figure) 25
Scan LED 23
System LED 23
System LED, mass spectrometer reset 59
System LED, mass spectrometer startup 57
Vacuum LED 22
Vent Valve Closed 24
Vent Valve Closed (figure) 25
lens alignment tool 161
lens L1, L2, L3, and L4
location (figure) 26–27
lenses
analyzer assembly, voltages applied to 39
ion source lenses exploded view 93, 100
ion source lenses, reassembling 93, 101
ion source lenses, reinstalling 94, 101
L1, L2, L3 exploded view 93, 100
L1, L2, L3, L4 figure (TSQ Quantum XLS) 26
L1, L2, L3, L4 location (figure) 27
L1, L2, L3, L4, cleaning 88, 95
L21, L22, L23, figure 34
L31, L32, L33, figure 34
linearity problems, troubleshooting 139
M
main power circuit breaker
description 23–24
location (figure) 24
Maintaining 116
maintenance
ball valve seal, replacing 120
forepump 116
frequency 83
GC column installation 124
ion source filament, replacing 105
ion source lenses, cleaning 88
ion volumes, cleaning 88
procedures (table) 83
mass analysis
collision-induced dissociation 36
discussion 34
mass analyzer
Thermo Scientific
Index: N
defined 6
description 32
quadrupole rod assembly, functional description 33
mass anaysis
RF and dc fields (figure) 34
mass range 15
mass spectrometer
and direct sample probe 151
CI gas valve 46
circuit breaker, location (figure) 25
collision gas valve 45
diagnostics 129
electronic assemblies 46
electronics service switch, location (figure) 25
emergency shutdown 51
front panel LEDs 22
functional description 21
fuses, replacing 145
inlet gasses hardware 41
inlet valve (figure) 29
ion detection system 39
ion detection system electronic assemblies 47
ion gauge 45
ion optics 30
ion source 26–27
LEDs and system startup 59
maintenance procedures (table) 83
mass analyzer 32
Off condition 60
On/Off status of components 60
Power Entry Module, description 46
power outlet 24
Q0 quadrupole 30
resetting 58
RF voltage generation electronic assemblies 47
right-side power panel 54
shutdown 53
Standby condition 52, 60
startup 55
turbomolecular pump 44
vacuum manifold 42
vacuum service switch, location (figure) 25
vacuum system 41
vent valve 45
mass spectrometer circuit breaker
location (figure) 25
mechanical pump 44
MS main power circuit breaker
mass spectrometer components On/Off status 60
MS/MS scan modes
Neutral Loss 9
Parent 8
Product 7
Thermo Scientific
N
Neutral Loss scan mode
discussed 9
example (figure) 11
illustration (figure) 10
O
octafluoronaphthalene 162
Off condition, mass spectrometer components On/Off status
60
offset voltage, quadrupole 37
oxygen trap 161
P
Parent 9
Parent scan mode
discussed 8
illustration (figure) 8
Perfluorotetracosane 157
personal computer, features 48
phthalate contamination 133
power cord, forepump (Caution) 44
Power Entry Module, description 46
Power LED
description 22
figure 22
mass spectrometer reset 59
mass spectrometer startup 57
power supply problems, troubleshooting 140
pressure, maximim allowed (table) 57
probe button 75
procedures
adding calibration compound 117
changing ionization modes 73
cleaning ion source components 85
cleaning ion source lenses 88, 95
cleaning ion volumes 88
cleaning stainless steel parts 113
complete system shutdown 53
conditioning capillary columns 125
disassembling ion source lens assembly 92, 98
displaying FC-43 spectrum 64
dissassembling ion source completely 107, 110
emergency shutdown 51
GC leak checking 125
installing capillary column 124
installing the ion volume 80
maintaining forepump 116
operating the inlet valve 73
placing system in standby 52
reassembly ion source lenses 93, 101
TSQ Quantum XLS User Guide
167
Index: Q
reinstalling ion source 94, 102
reinstalling ion source lenses 94, 101
reinstalling Ion Source PCB 109, 112
removing capillary column 122
removing ion source 89, 96
removing ion source lens assembly 91, 98
removing the ion volume 73
replacing filament 102, 105
replacing ion source filament 102, 105
resetting mass spectrometer 58
saving tune and calibration report 71
starting mass spectrometer 56
system startup 55
tuning and calibrating 66
Product scan mode
discussed 7
illustration (figure) 8
profile scan, defined 14
pumps
forepump 44
turbomolecular pump 44
Q
Q0 30
Q0 quadrupole
cross sectional view (figure) 31
description 30
figure 30
location (figure) 43
Q1 33
Q1 quadrupole
figure 33
scan modes 6
Q1MS and Q3MS scan modes 7
Q2 rod assembly
figure 34
scan modes 6
Q3 quadrupole
figure 33
scan modes 6
QED MS scan type 14
quadrupole 30
quadrupole mass analyzer
functional description 33
quadrupole offset voltage 37
quadrupoles
mass analysis 33
Q0, description 30
Q1 and Q3 33
RF and dc fields 34
RF and dc fields (figure) 34–35
Qual Browser
168
TSQ Quantum XLS User Guide
direct sample probe data 157
reviewing data 157
quantitation-enhanced data-dependent MS (QED MS) scan
type 14
R
rear power panel
location (figure) 52
reference SRM reactions 14
regulatory compliance iii
removed 98
Replaceable 161
replaceable parts 161
Replacing 120
RF / dc voltages applied to mass analyzer, discussion 33
RF voltage generation, discussion 47
right side power panel
figure 25
right-side power panel
description 54
figure 54
rod assemblies
description 34
ion transmission 6
mass analysis 6
mass analyzer(s), quadrupole 33
Q1, Q2, and Q3 (note) 7
roughing pump 44
S
Safe to Open Valve message
direct sample probe 154
safety precautions vii
safety standards iii
Save 68
scan data types
centroid scan 15
profile scan 14
Scan LED
description 23
figure 22
scan modes
data dependent 11
discussed 5
mass spectrometer 5
Neutral Loss 9
Neutral Loss, example (figure) 11
Neutral Loss, illustrated (figure) 10
Parent 8
Parent, illustrated (figure) 9
Product 7
Thermo Scientific
Index: T
Product, illustrated (figure) 8
Q1MS and Q3MS 7
summary (table) 6
scan types
AutoSIM 14
discussed 12
full scan 12
intelligent selected reaction monitoring (iSRM) 13
quantitation-enhanced data-dependent MS (QED MS) 14
selected ion monitoring (SIM) 12
selected reaction monitoring (SRM) 13
sectional 31–32
selected ion monitoring (SIM) scan type 12
selected reaction monitoring (SRM) scan type 13
sensitivity problems, troubleshooting 140
septum bleedcontamination 133
sequence 150
creating 150
direct sample probe, running 153
Sequence Setup view 150
shutdown
data system 55
emergency procedure 51
emergency, front panel system power off button 23
mass spectrometer 53
non-emergency 53
side cover plate (vacuum manifold), description 43
solvent ions masses 135
solvents
purity requirements viii
spectrometer 14
square 33
stability problems, troubleshooting 143
stainless steel, cleaning 113
Standby condition
mass spectrometer components On/Off status 60
placing system in 52
startup 55
autosampler 58
gas chromatograph (GC) 56
operating conditions, setting 58
switch 24
System Control PCB, description 46
System LED
figure 22
mass spectrometer reset 59
mass spectrometer startup 57
system power off button
emergency shutdown 23
location (figure) 52
system reset button
description 24
Thermo Scientific
location (figure) 24
mass spectrometer reset 59
system shutdown
emergency procedure 51
System Tune and Calibration workspace
figure 66
T
tables
Accessory Kit 161
Chemicals Kit 162
mass spectrometer components on/off status 60
mass spectrometer maintenance procedures 83
maximum allowed pressures 57
summary of scan modes 6
the 46
TRACE GC Ultra gas chromatograph
configuration 19
photo 2
transfer line
capillary column, removing 123
cross sectional view (figure) 31
figure 21
functional description 20
location (figure) 89, 96
union (figure) 123
TriPlus autosampler
cofiguration 18
photo 2
troubleshooting
autosampler communication 133
communication 131
contamination 133
filament and lens control 135
gas chromatograph communication 133
heated zones 136
high vacuum 138
leak checking 138
linearity 139
power supplies 140
sensitivity 140
stability 143
tuning 143
TSQ Quantum GC
autosampler 18
calibrating 63
data system 48
data types 14
diagnostics 129
front panel LEDs 22
functional block diagram (figure) 17
functional description 17
TSQ Quantum XLS User Guide
169
Index: V
gas chromatograph 19
ion polarity modes 5
LEDs and system startup 59
mass range 15
mass spectrometer 21
operating conditions, setting 58
overview 1
photo 2
scan modes 5
scan types 12
shutdown 55
Standby condition 52–53
startup 55
tuning 63
TSQ Tune Master Application
diagnostics 129
tune and calibration report
figure 70
saving 69
tuning
discussed 63
frequency (note) 63
running 66
tune and calibration report 69
tuning problems, troubleshooting 143
turbomolecular pump, description 44
powering off 24
problems, troubleshooting 138
turbomolecular pump 44
Vacuum LED 22
vacuum service switch 24
vent valve 45
vent valve
description 45
LED 24
Vent Valve Closed LED 24
voltages
conversion dynode 39
Q0 offset 30
quadrupole RF and DC 34
W
WEEE compliance vii
V
Vacuum LED
and analyzer region pressure 22
description 22
figure 22
vacuum manifold
description 42
location (figure) 43
side cover plate, description 43
vacuum pumps
forepump 44
turbomolecular pump 44
vacuum service switch
description 24
location (figure) 25
On/Off status of mass spectrometer components 60
vacuum system
collision gas valve 45
Convectron gauge 45
description 41
forepump 44
functional block diagram (figure) 42
ion gauge 45
leak checking 138
maximim allowed pressures 57
170
TSQ Quantum XLS User Guide