Nova Biomedical Nova StatStrip Glucometer StatStrip Glucometer Instruction Manual

Nova Biomedical Nova StatStrip Glucometer StatStrip Glucometer Instruction Manual

Nova StatStrip Glucometer are glucose meters used for inpatient and ambulatory areas. The meter can measure glucose levels in blood samples and display the results on a color touch screen. The meter can also transmit results automatically when it is placed in the docking station.

advertisement

Assistant Bot

Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.

Nova StatStrip Glucometer Instruction Manual | Manualzz

Nova StatStrip Glucometer

Before Super User hands-on training please do tutorial on KP Learn so your training session will be more meaningful.

1.

2.

3.

4.

5.

6.

Access the KP Learn website at http://kplearn.kp.org

Enter StatStrip Training into Learning Catalog Search window

Click Launch, located on the right side.

Start the tutorial. It will take approximately 25 minutes to complete the entire presentation.

Once completed, access the “Transcript” tab click on the “Print Certificate of Completion”

Submit to the Educator/ADA/Clinical Supervisor

NOVA STAT STRIP GLUCOSE METER

• Results are automatically transmitted when the meter is placed in the docking station.

• The middle green light will flash when data is transmitting.

• An error screen will be displayed if the meter is removed before data transfer is complete.

NOVA STAT STRIP GLUCOSE METER

QC must be done every 24 hrs or as needed (i.e. pt symptoms do not match results, drop meter, etc.)

_________________________________________________________

NOVA STAT STRIP GLUCOSE METER

Touch screen or press OK to “wake up” meter

• Push Login or OK soft key

Operator ID – NUID#

• For INPATIENT areas, enter operator ID by SCANNING your barcode

• For AMBULATORY areas,

MANUALLY enter your operator ID

Press QC button on meter

NOVA STAT STRIP GLUCOSE METER

Scan barcode on Strip vial and press Accept

NOVA STAT STRIP GLUCOSE METER

NOVA STAT STRIP GLUCOSE METER

NOVA STAT STRIP GLUCOSE METER

HIGH value indicated by a single arrow up

LOW value indicated by a single arrow down

CRITICAL VALUES:

>400 HIGH

<40 LOW

To accept a result, press the ACCEPT button

For REJECTED results (i.e. pt symptoms do not match results) and CRITICAL VALUES, you MUST enter a COMMENT

Comments (choose up to three):

• Repeated Test

• Procedure Error

• Treat Patient

• Lab Draw

• Notify MD

• Do not accept this

• Do not accept prior

• Accept this

NOVA STAT STRIP GLUCOSE METER

Meter saves up to

1000 results in its memory

NOVA STAT STRIP GLUCOSE METER

NOVA STAT STRIP GLUCOSE METER

Clean meter after EACH patient use following locally approved infection control policies.

Thoroughly wipes the surface of the meter (top, bottom, left, and right sides) with locally approved germicidal wipe (i.e. PDI Sani-Cloth PLUS, Clorox wipe) a minimum of 3 times horizontally and 3 times vertically, avoiding the meter’s bar code scanner, electrical

connector, and making sure that no fluid enters the test strip port. Allow to air dry. Never reuse a previously soiled wipe. (Clorox wipes used only for disinfecting the meters used on C. difficile isolation patients).

San Diego Service Area

Nova Glucose Meters: Initial Training

Operator Print Last Name: ____________________ First Name: ________________ NUID# __________ EMP#__________

Location: ______________________ Bldg: ______________ Unit/Dept: _______________________ Date: _______________

Training Meter# _______________ Training Patient MRN# ______________________ Training NUID# ________________

Operator Initial each line

Describe Nova Meters

a.

____ Meter stays on the docking station or locks out after 4hours.

b.

___

Carry the meter with carrying case. Leave the case outside the room. c. ____ Color touch screen d. ____ 6-sec analysis time e.

____ 1.2 μL sample volume f. ____ Rechargeable batteries g. ____ Position of strip port

1. Describe touch-screen and buttons

a. ____ On/off/ sleep b. ____ Sleep mode after 90 seconds c. ____ Show when buttons/touch screen are interchangeable

2. Describe Docking Station

a. ____ 3 functions: upload, download, and charge battery b. ____ Describe amber and green lights (RT: battery charge, Ctr: communication, Lt: connected)* c. ____ Show spare battery placement

3. Describe strips

a. ____ Gold layer b. ____ Front fill with one continuous process c. ____ Do NOT under-dose or re-dose; overdosing is not possible. One continuous process d. ____ Date vials with180-day (6Months) expiration which must be before Manufacturer expiration. e. ____ Keep vial closed; affected by heat, humidity.

4. Demonstrate how to manipulate the meter

a. ____ Keep horizontal / keep strip well dry/ Make sure patient is not dripping b. ____ Scanner; press and release (don’t hold down) c. ____ Hold meter 4-6” away from barcode d. ____ Apply sample or QC until countdown starts

Using the Meter

5. Direct Observation/Log on/Getting Started/ a. ____ Speaker icon b. ____ Scanning methods; hard key, soft key c. ____ Barcode scan required for Operator NUID# or manual entry for 12 digit patient MRN# allowed d. ____ always check blue dialogue bar to see what meter is doing

Updated: 11/22/2016

Page 1 of 2

San Diego Service Area

Nova Glucose Meters: Initial Training

6. Direct Observation/Quality Control

a. ____ True QC Lockout every 24 hrs/ require once per day of use & New strip lot number b. ____ Date QC vials for 90-day expiration or Manufacturer Expiration date, whichever comes first. c. ____ Scan QC Bottle d. ____ Mix vial technique e. ____ pass/fail result f. ____ Keep meter on counter g. ____ Pass QC (Use Comment: Clean Meter) h. ____ How to follow-up on failed QC (use comments; Mix & Repeat, procedure error & repeat test)

7. Run a sample a. ____ Scan strip lot b. ____ Scan patient barcode or enter 12 digits Medical Number c. ____ Importance of clean and dry patient finger d. ____

Importance of wiping the first drop of blood with dry Gauze

e. ____ Enter comments (must be done before accepting results) f. ____ Explain

REJECT

comment for error or repeated results (Reject, Accept, Reject Previous) g. ____ How to run an unregistered patient/Fill out a form using the number on the Form h. ____ How to test an isolation patient/Nova bag or bio-Hazard Bag*

8. Interpretation of Results / Comment use

a. ____ Linearity range 10-600 mg/dL {Less then 10(<10) Lo and Greater then (>600) Hi} b. ____ Normal Range Adult 70 mg/dL to 140 mg/dL and Neonate 55 mg/dL to 115 mg/dL c. ____ Critical Range Adult <55 mg/dL and >400 mg/dL and Neonate <45mg/dL and >200mg/dL d. ____ Comment Use (Accept, Reject, Reject Previous, Repeat etc.)

9. Review Results / Log Out

a. ____ Sort by date, time / ID / sample type b. ____ Highlight patient to view c. ____ Use of

NEXT / PREV. keys d. ____ Manual logout (2-steps) e. ____ Automatic logout when docking / timed logout after 5 min*

10. Change Battery

a. ____ Rechargeable

11. Cleaning a. ____ Cleaning (after each Use) & Disinfecting (Example: C.Diff patient) b. ____ “Clean Meter” Comment use after each level of QC c. ____ After each patient use

12. Competency a. ____ Initial for New Operator and Annual there after b. ____ Check off list, Direct Observation and Quiz c. ____ Must be done using training barcodes & on training meter

13. General a.

_____ Provided with verbal & written instructions on how to locate POCT Policy & Procedure and Clerical

Error forms

Validator Print Last Name_____________________ First Name: _________________ NUID# _________

Updated: 11/22/2016

Page 2 of 2

San Diego Service Area

NOVA GLUCOMETER COMPETENCY Assessment & Test

Circle One: Initial____ or Annual_____

Operator Print Last Name: ____________________, First Name: _________________ NUID# __________ EMP#_________

Location: _________________ Unit/Dept.: _________________ Date: _____________ Training Meter# _____

Validator Print Last Name_____________________, First Name: ________________ NUID# ____________

All testing personnel using the meter will successfully complete the following tasks:

Tasks (For Initial Use Training NUID# & Patient MR#)

Operator initial

1. Turn Glucose meter on/ Wake meter up.

2. Enter or scan your own NUID# on meter

3. QC Competency for Initial & Annual: Must Perform High and Low QC tests.

4. Select “QC” mode on the meter.

5. Scan test strip lot and Control lot numbers

Keep Stat Strip Glucose Test Strip vial tightly closed when not in use.

Test strips should only be stored in original vial at room temp.

6. Dose strip with QC solution. Warning: The test strip must be completely filled by touching QC

solution droplet to strip. If test strip does not fill completely, do not touch test strip to QC

droplet a second time. Discard test strip and repeat test with a new test strip.

7. Remove and dispose of test strip. Enter “Cleaned meter” under “Comments”.

8. Patient Competency for initial must be performed (Annual may verbalize):

9. Select “Accept” for Patient mode. Enter or Scan provided patient MRN.

10. Obtain Patient blood sample. (capillary, arterial or venous) and perform Glucose test

• May use QC solution for patient blood sample competency.

11. View test data on screen. Accept or reject the result.

12. When applicable, enter a ‘Comment’ (e.g. MD Notified)

13. Recall and review on-board test data.

14. Dock meter in Docking Station to automatically upload stored meter data and download updated setup information.

15. Provided with verbal & written instructions on how to locate POCT Policy & Procedure

All testing personnel will answer the following 5 questions.

Circle Correct Answer (Passing sc ore ≥ 80%

)

1. Once opened, test strips are good for 180 days and control solutions are good for 90 days or until manufacturer’s printed expiration date whichever comes first.

2. Comments are mandatory when documenting: a) Action taken when QC results are out of range. b) Action taken when critical patient results are obtained. c) All of the above

3. Once opened, control solutions and test strips must be dated with open and expiration date.

True False

a b c

True False

4. Fill in Blanks for about cleaning the meter? a) Clean/disinfect the meter surface with locally approved _________________disinfectant wipes between every patient. b) Clean/disinfect Glucose meter surface with _____________________________ for isolation and C. diff. patients.

5. The test strip must fill completely upon touching the blood droplet. If the test strip does not fill completely, do not touch the test strip to the blood droplet a second time. Discard the test strip and repeat the test with a new strip.

True False

Score: /5 ______% Validator Initials: _____Pass/Fail:______

Remediation Score: /5 _____% Remediation Validator Initials: ____

REV: 11/21/2016

San Diego Service Area

Nova Glucose Meter Training Sign-In Log

Circle one

: Initial _______ or Annual ___________

Date:

_______________

Validator Print: Last name: _________________________ First Name: __________________ NUID# ___________

Tie line number: _______________Training Location: ______________ Bldg/Unit /Module/Dept: _______________

*OPERATORS

PLEASE PRINT CLEARLY, ANSWER LAST QUESTION, and Write in Meter #

Operator Print

LAST Name

Operator Print

FIRST Name

HOSPITAL Location: a) Zion____ b) SDMC_____

Unit: ______ Floor: _____

AMBULATORY Location:

MOB Name: _____________

Dept: _______ Mod: _______

OPERATOR MUST

COMPLETE

Direct observation

1. Did you perform

QC? WRITE: YES /

NO

2. Write Meter #

1

2

3

4

5

6

1.

2. #

1.

2. #

1.

2. #

1.

2. #

1.

2. #

1.

7

2. #

1.

2. #

1. 8

9

10

2. #

1.

2. #

1.

2. #

2. #

Validator, please email this completed form to (not scan or fax) POCT Manager:

[email protected]

and

[email protected]

Do NOT Fax or Copy Competency Assessment (CA) check off list or CA Written Test.

For completed original Competency Assessment (CA) forms, follow your department procedure or send

forms to the KP Education Department for filing

Updated: 12/27/2016

advertisement

Key Features

  • Color touch screen
  • 6-sec analysis time
  • 1.2 μL sample volume
  • Rechargeable batteries
  • Automatic result transmission
  • QC lockout
  • Comment entry for rejected results
  • Memory for 1000 results
  • Cleaning after each use

Frequently Answers and Questions

How do I turn on the Nova StatStrip Glucometer?
Touch the screen or press the OK button to wake up the meter.
How do I log in to the Nova StatStrip Glucometer?
Press the Login or OK button, then enter your operator ID. If you are in an inpatient area, you can scan your barcode to enter your operator ID.
How do I perform a quality control test on the Nova StatStrip Glucometer?
Press the QC button on the meter, scan the barcode on the strip vial, and press Accept.
How do I interpret the results on the Nova StatStrip Glucometer?
The meter displays the results on the screen and indicates high or low values with arrows. Critical values are indicated with a symbol. You can accept or reject a result, and you must enter a comment for rejected results.