- Industrial & lab equipment
- Measuring, testing & control
- Multimeters
- Agilent Technologies
- U3402A
- User manual
advertisement
Agilent U3402A
5 1/2 Digit Dual Display
Multimeter
User’s and Service Guide
Agilent Technologies
Notices
© Agilent Technologies, Inc. 2009–2013
No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Agilent
Technologies, Inc. as governed by United
States and international copyright laws.
Manual Part Number
U3402-90001
Edition
Fifth Edition, May 21, 2013
Printed in Malaysia
Agilent Technologies, Inc.
5301 Stevens Creek Blvd.
Santa Clara, CA 95051 USA
Warranty
The material contained in this document is provided “as is,” and is subject to being changed, without notice, in future editions. Further, to the maximum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a particular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or performance of this document or of any information contained herein. Should
Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement shall control.
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
Restricted Rights Legend
U.S. Government Restricted Rights. Software and technical data rights granted to the federal government include only those rights customarily provided to end user customers. Agilent provides this customary commercial license in Software and technical data pursuant to FAR 12.211 (Technical
Data) and 12.212 (Computer Software) and, for the Department of Defense, DFARS
252.227-7015 (Technical Data - Commercial
Items) and DFARS 227.7202-3 (Rights in
Commercial Computer Software or Computer Software Documentation).
Safety Notices
C A U T I O N
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a
CAUTION notice until the indicated conditions are fully understood and met.
WA R N I N G
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
ii U3402A User’s and Service Guide
Safety Symbols
Direct current (DC)
The following symbol on the instrument and in the documentation indicates precautions that must be taken to maintain safe operation of the instrument.
Caution, risk of electric shock
Alternating current (AC)
Both direct and alternating current
Caution, risk of danger (refer to this manual for specific Warning or Caution information)
Out position of a bi-stable push control
Earth (ground) terminal
Protective conductor terminal
In position of a bi-stable push control
Frame or chassis terminal
CAT II
300 V
IEC Measurement Category II. Inputs may be connected to mains (up to 300 VAC) under Category II overvoltage conditions.
U3402A User’s and Service Guide iii
Regulatory Markings
The CE mark is a registered trademark of the European Community. This CE mark shows that the product complies with all the relevant European Legal
Directives.
ICES/NMB-001 indicates that this ISM device complies with the Canadian
ICES-001.
Cet appareil ISM est confomre a la norme NMB-001 du Canada.
The CSA mark is a registered trademark of the Canadian Standards
Association.
The C-tick mark is a registered trademark of the Spectrum
Management Agency of Australia. This signifies compliance with the Australia EMC Framework regulations under the terms of the
Radio Communication Act of 1992.
This instrument complies with the
WEEE Directive (2002/96/EC) marking requirement. This affixed product label indicates that you must not discard this electrical/electronic product in domestic household waste.
This symbol indicates the time period during which no hazardous or toxic substance elements are expected to leak or deteriorate during normal use.
Forty years is the expected useful life of the product.
iv U3402A User’s and Service Guide
General Safety Information
The following general safety precautions must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument. Agilent Technologies assumes no liability for the customer’s failure to comply with these requirements.
WA R N I N G
• Do not defeat power cord safety ground feature. Plug in to a grounded (earthed) outlet.
• Do not use instrument in any manner that is not specified by the manufacturer.
• Double-check the instrument’s operation by measuring a known voltage.
• For current measurement, turn off circuit power before connecting the instrument to the circuit. Always place the instrument in series with the circuit.
• When connecting probes, always connect the common test probe first. When disconnecting probes, always disconnect the live test probe first.
• Do not measure more than the rated voltage (as marked on the multimeter) between terminals, or between terminal and earth ground.
• Do not use repaired fuses or short-circuited fuse-holders. For continued protection against fire, replace the line fuses only with fuses of the same voltage and current rating and recommended type.
• Do not service or perform adjustments alone. Under certain conditions, hazardous voltages may exist, even with the instrument switched off. To avoid dangerous electric shock, service personnel must not attempt internal service or adjustment unless another person, capable of rendering resuscitation or first aid, is present.
• Do not substitute parts or modify instrument to avoid the danger of introducing additional hazards. Return the instrument to Agilent Technologies Sales and
Service Office for service and repair to ensure the safety features are maintained.
• Do not operate damaged instrument as the safety protection features built into this instrument may have been impaired, either through physical damage, excessive moisture, or any other reason. Remove power and do not use the instrument until safe operation can be verified by service-trained personnel. If necessary, return the instrument to Agilent Technologies Sales and Service Office for service and repair to ensure the safety features are maintained.
U3402A User’s and Service Guide v
C A U T I O N
• Turn off circuit power and discharge all high-voltage capacitors in the circuit before you perform resistance, continuity, or diode tests.
• Use the correct terminals, functions, and range for your instrument.
• Do not measure voltage when current measurement is selected.
• Use the instrument with the cables provided.
• Repair or service that is not covered in this manual should only be performed by qualified personnels.
vi U3402A User’s and Service Guide
Environmental Conditions
This instrument is designed for indoor use and in an area with low condensation. The table below shows the general environmental requirements for the instrument.
Environmental Conditions
Operating temperature
Operating humidity
Storage temperature
Altitude
Pollution degree
Requirements
Full accuracy from 0 °C to 50°C (Operating)
Full accuracy up to 80 % R.H. (relative humidity) for temperature up to 28°C
–20 °C to 60 °C (Non-operating)
Operating up to 2,000 metres (6,562 feet)
Pollution Degree 2
N O T E
The Agilent U3402A 5 1/2 digit dual display multimeter complies with the following EMC requirements:
• IEC 61010-1:2001/EN61010-1:2001 (2nd Edition)
• Canada: CAN/CSA-C22.2 No. 61010-1-04
• USA: ANSI/UL 61010-1:2004
• IEC 61326-1:2005/EN 61326-1:2006
• Canada: ICES/NMB-001: Issue 4, June 2006
• Australia/New Zealand: AS/NZS CISPR11:2004
U3402A User’s and Service Guide vii
Waste Electrical and Electronic Equipment (WEEE) Directive
2002/96/EC
This instruction complies with the WEEE Directive (2002/96/EC) marking requirement. This affixed product label indicates that you must not discard this electrical/electronic product in domestic household waste.
Product Category:
With reference to the equipment types in the WEEE directive Annex 1, this instrument is classified as a “Monitoring and Control Instrument” product.
The affixed product label is shown as below:
viii
Do not dispose in domestic household waste
To return this unwanted instrument, contact your nearest Agilent office, or visit www.agilent.com/environment/product for more information.
U3402A User’s and Service Guide
Declaration of Conformity (DoC)
The Declaration of Conformity (DoC) for this instrument is available on the
Web site. You can search the DoC by its product model or description.
http://regulations.corporate.agilent.com/DoC/search.htm
N O T E
If you are unable to search for the respective DoC, please contact your local Agilent representative.
U3402A User’s and Service Guide ix
In This Guide ...
1
This chapter provides an introduction to the U3402A digital multimeter and a tutorial showing how to use the front panel in order to make measurements.
2
Operations and Features
This chapter explains the various functions and features available to the
U3402A digital multimeter.
3
Application Tutorial
This chapter describes the advanced features and the possible applications for effective operation of the multimeter.
4
This chapter contains performance test procedures. The performance test procedures allow you to verify that the multimeter is operating within its published specifications.
5
Disassembly and Repair
This chapter will help you troubleshoot a faulty multimeter. It describes how to disassemble the multimeter, how to obtain repair services, and lists the replaceable parts.
6
Specifications and Characteristics
This chapter describes the multimeter’s specifications and operating specifications
x U3402A User’s and Service Guide
Contents
U1253B User’s and Service Guide
1 Getting Started Tutorial
Introducing the Agilent U3402A Dual Display Multimeter
2
Checking the shipping contents 3
Connecting power to the multimeter
4
5
6
7
7
9
14
16
17
Performing voltage measurements 18
Performing current measurements
20
Performing frequency measurements
22
Performing resistance measurements 23
Performing diode/continuity test
25
30
32
2 Operations and Features
20
22
xi
xii
26
Combination of Math Operations
27
29
31
Changing the Configurable Settings 32
3 Application Tutorial
Applications for Using Dual Display
52
Dual Display Operation Examples
53
Measure DC Voltage and AC Ripple on a Rectification
53
Measure AC and DC Current on a Rectification Circuit 54
Measure AC Voltage and Frequency on an AC Circuit
55
Measure DC Voltage and DC Current on a Transistor Circuit or
56
Measure Resistance Using 2-Wire Mode 58
Measure Resistance Using 4-Wire Mode 59
4 Performance Test
62
Agilent Technologies Calibration Services 62
62
63
64
Performance Verification Test Overview
65
Performance Verification Test 65
5 Disassembly and Repair
76
U1253B User’s and Service Guide
U1253B User’s and Service Guide
77
78
To Replace the Power Line Fuse
79
To Replace a Current Input Fuse 80
Electrostatic Discharge (ESD) Precautions 80
87
87
6 Specifications and Characteristics
90
93
97
Supplemental Measurement Specifications
98
98
98
107
To Calculate Total Measurement Error
109
110
110
110
Configuring for Highest Accuracy Measurements
111
xiii
xiv U1253B User’s and Service Guide
List of Tables
U1253B User’s and Service Guide
Display annunciators 9
Keypad functions 11
Input terminal for different measurement functions 15
Range scale value in slow, medium, and fast reading rate 31
Reading rates for single function measurements 32
Math operations for different measurement functions 20
Descriptions for combined math operations 28
Description for dual display combination 30
Setup menu and communication parameters 31
Typical combinations and applications when using dual display 52
Recommended test equipments 63
DC voltage verification test 66
DC current verification test 67
2-wire W verification test 68
4-wire W verification test 69
Diode verification test 71
Frequency verification test 71
AC volts verification test 72
AC current verification test 73
Type of supplied fuse (according to country of destination) 79
Type of current input fuse 80
DC resolution, full scale reading, and accuracy [±(% of reading + count)] 90
AC voltage resolution, full scale reading, and accuracy [±
(% of reading + count)] 93
AC voltageac+dc resolution, full scale reading, and accuracy [± (% of reading + count)] 94
AC current resolution, full scale reading, burden voltage, and accuracy [± (% of reading + count)] 95
AC current ac+dc resolution, full scale reading, burden voltage, and accuracy [± (% of reading + count)] 96
xv
Frequency resolution and accuracy [± (% of reading + count)] 96
Range and accuracy (±dB) 97
Full scale display update rates 98
Supplemental measurement specifications 98
Reading rates for single display (readings/second
(approx)) 105
Reading rates for the dual display (readings/second
(approx)) 106
General characteristics 107
xvi U1253B User’s and Service Guide
List of Figures
U1253B User’s and Service Guide
Stacking the U3402A 5
Type of handle position 6
Attaching and detaching the handle.
6
U3402A dimensions 7
Front panel 8
VFD full display with all segments illuminated.
9
Keypad 11
Input terminal 14
Rear panel 16
ACV terminal connection and display 18
DCV terminal connection and display 19
ACI RMS or DCI (mA) terminal connection and display 20
ACI RMS or DCI (A) terminal connection and display 21
Frequency terminal connection and display 22
2-wire W terminal connection and display 24
4-wire W terminal connection and display 24
Forward-biased diode/continuity test terminal connection and display 27
Reverse-biased diode/continuity terminal connection and display 27
2-wire/continuity test terminal connection and display 29
Reading rate annunciator 32
Typical dBm operation display 21
Typical Rel operation display 22
Typical Max operation display 23
Typical Min operation display 24
Typical Comp operation display 25
Typical Hold operation display 26
Combined math operations sequence 28
Secondary display 29
Terminal connection when measuring DC voltage and
AC ripple on a rectification circuit 53
xvii
Terminal connection when measuring AC and DC current on a rectification circuit 54
Terminal connection when measuring AC voltage and frequency on an
AC circuit 55
Terminal connection when measuring DC voltage and
DC current on a transistor circuit or load 57
Terminal connection when measuring resistance using
2-wire mode 58
Terminal connection when measuring resistance using
4-wire mode 59
xviii U1253B User’s and Service Guide
U3402A 5 1/2 Digit Dual Display Multimeter
User’s and Service Guide
1
Getting Started Tutorial
Introducing the Agilent U3402A Dual Display Multimeter 2
Checking the shipping contents 3
Connecting power to the multimeter 4
Performing voltage measurements 18
Performing current measurements 20
Performing frequency measurements 22
Performing resistance measurements 23
Performing diode/continuity test 25
This chapter provides an introduction to the U3402A digital multimeter and a tutorial showing how to use the front panel in order to make measurements.
Agilent Technologies
1
1 Getting Started
2
Introducing the Agilent U3402A Dual Display Multimeter
The key features of the U3402A dual display multimeter are:
• 5 ½–digit dual display measurement
• Eleven measurement functions:
•
AC voltage
•
DC voltage
•
AC + DC voltage
•
AC current
•
DC current
•
AC+DC current
•
2- wire resistance
•
4- wire resistance
•
Frequency
•
Continuity test
•
Diode test
• Five math operations:
•
dBm
•
Min/Max
•
Relative (Rel)
•
Compare (Comp)
•
Hold
• True RMS measurement for both AC+DC votage and current.
• Wide AC and DC current measurememt range; from 12 mA to 12 A.
• Resistance measurement up to 120 M
W with 1 mW resolution at slow reading rate or up to 300 M
W with 10 mW and 100 mW resolution at medium and fast reading rate respectively.
• Frequency measurement up to 1 MHz.
• dBm measurement with selectable reference impedance from 2
W to 8000 W and audio power measurement capability.
• Dynamic recording for minimum and/or maximum readings.
U3402A User’s and Service Guide
N O T E
Getting Started 1
Checking the shipping contents
Verify that you have received the following items with your multimeter:
•
Power cord
•
Standard test lead kit
•
Quick start guide
•
Product reference CD
•
Test report
•
Certificate of calibration
Inspect the shipping container for damage. Signs of damage may include a dented or torn shipping container or cushioning material that shows signs of unusual stress or compacting.
Carefully remove the contents from the shipping container and verify that any options ordered are included with the shipment by checking the packing list included with the shipment.
If anything is missing, contact your nearest Agilent Sales Office.
• If the shipping container or packaging material is damaged, it should be kept until the contents have been checked mechanically and electrically. If there is mechanical damage, notify the nearest Agilent Technologies office. Keep the damaged shipping materials (if any) for inspection by the carrier and Agilent representative. If required, you can find a list of Agilent Sales and Service Offices on the last page of this guide.
• Ensure you have read and understand the preceding safety information before you proceed.
Original packaging
Containers and materials identical to those used in the factory pakaging are available through Agilent Technologies office. If the multimeter is being returned to Agilent Technologies for servicing, attach a tag indicating the type of service required, return address, model number, and serial number. Also mark the container FRAGILE to assure careful handling. In any correspondence, refer to the multimeter by model number and serial number.
U3402A User’s and Service Guide 3
4
1 Getting Started
N O T E
Connecting power to the multimeter
Connect the power cord and press the power switch to turn on the multimeter.
The front panel display illuminates while the multimeter performs its power- on self- test. (If the multimeter does not power- on, refer to the
“Operating Checklist” on page 76). During the power- on session, press
to hold the full display. Press any key to resume the power- on self- test.
The multimeter powers up in the DC voltage function with autoranging enabled. If self- test is successful, the multimeter goes to normal operation.
If the self- test fails, either a full annunciator or a blank display is displayed without entering the normal operation. In the unlikely event that the self- test repeatedly fails, contact your nearest Agilent Sales and Service Office.
The multimeter will operate at any line voltage between 90 VAC and 264 VAC when the line voltage selector is set properly with frequency range 50 Hz or 60 Hz.
C A U T I O N
• Before turning on the multimeter, make sure the line voltage selector is set to the correct position for the applied line voltage to the power line cord connector.
• Do not apply a line voltage that exceeds the specified range of the line cord connector.
U3402A User’s and Service Guide
Getting Started 1
Stacking the U3402A
The U3402A is shipped with specially designed anti- slip protective bumpers on the front panel and rear panel. The multimeters will not slide off when stacked on top of each other.
To be able to stack the U3402A multimeters, ensure the attached bumpers are
in correct orientation. Refer to Figure 1- 1 .
Figure 1-1
Stacking the U3402A
U3402A User’s and Service Guide 5
1 Getting Started
Adjusting the Handle
To adjust the handle, grasp the handle by the side and pull outward. Then,
rotate the handle to the desired position. Figure 1- 2 below shows the possible
handle positions.
Carrying position
Benchtop position
Handle removal position
Figure 1-2
Type of handle position
To attach or detach the handle, rotate the handle upright and pull it out from the sides of the multimeter. Refer to
.
6
Figure 1-3
Attaching and detaching the handle.
U3402A User’s and Service Guide
Getting Started 1
Product at a Glance
Product dimensions
Front view
Side view
226.0 mm
305.0 mm
Figure 1-4
U3402A dimensions
U3402A User’s and Service Guide
105.0 mm
7
8
1 Getting Started
The front panel at a glance
Primary display
Secondary display
Input terminals and current fuse
Measurement function keypads
Autoranging, manual range and comparator operation
Figure 1-5
Front panel
Math operation keypads
Shift Power on/off switch
U3402A User’s and Service Guide
Getting Started 1
The display at a glance
Figure 1-6
VFD full display with all segments illuminated.
The highly visible vacuum fluorescent display (VFD) annunciator are described in
.
Table 1-1 Display annunciators
Annunciator
Primary display
S
M
F
PEAK
HOLD
MIN
MAX
REL dBm
4 2 W
Description
AUTO
DC
AC
DCAC
Reading rate: Slow
Reading rate: Medium
Reading rate: Fast
Peak measurement. Not applicable for the U3402A.
Data hold
MinMax math operation: Minimum value shown on the primary display
MinMax math operation: Maximum value shown on the primary display
Relative value
Decibel unit relative to 1 mW
4-wire/2-wire resistance
Diode test
Audible continuity test for resistance
Autoranging
Direct current
Alternating current
AC + DC
Polarity, digits, and decimal points for primary display
U3402A User’s and Service Guide 9
1 Getting Started
10
2ND
AUTO
DC
AC
DCAC
COMP dBm
CAL
µnmF
ºC
ºF
% mV
µmA
S
Mk
W
MkHz
EXT
TRIG
REMOTE
LOCAL
STO
RCL
Table 1-1 Display annunciators
Annunciator
ºC
ºF mV
Description
Celcius temperature unit. Not applicable for the U3402A.
Fahrenheit temperature unit. Not applicable for the U3402A.
Voltage unit: mV, V
µmA
µmnF
Mk
W
MkHz
Current unit: µA, mA, A
Capacitance unit: nF, µF, mF. Not applicable for the U3402A.
Resistance unit:
W, kW, MW
Frequency unit: Hz, kHz, MHz
Secondary display
Polarity, digits, and decimal points for secondary display
Capacitance unit: nF, µF, mF. Not applicable for the U3402A.
Celcius temperature unit. Not applicable for the U3402A.
Fahrenheit temperature unit. Not applicable for the U3402A.
Duty cycle measurement. Not applicable for the U3402A.
Voltage unit: mV, V
Current unit: µA, mA, A
Shift mode
Resistance unit:
W, kW, MW
Frequency unit: Hz, kHz, MHz
Secondary display is enabled
Autoranging
Direct current
Alternating current
AC + DC
Compare operation
Decibel unit relative to 1 mW
Calibration mode. Not applicable for the U3402A.
External. Not applicable for the U3402A.
Trigger mode. Not applicable for the U3402A.
Remote interface control. For calibration use only.
Local mode
Store instrument state. Not applicable for the U3402A.
Recall stored instrument state. Not applicable for the U3402A.
U3402A User’s and Service Guide
Getting Started 1
The keypad at a glance
The operation for each key is shown in
. Pressing a key changes the current key operation, illuminates the related symbol on the display and emits a beep.
Figure 1-7
Keypad
Table 1-2 Keypad functions
Key
System related operation
Description
Press to power-on or power-off the U3402A multimeter.
+
+
Press to select Shift.
Press to return the multimeter to front panel operation when it is in remote state.
Press to step through the Setup menu. See
“Using the Setup Menu” on page 31 for more information.
Press to enable the secondary display.
Press to disable the secondary display.
U3402A User’s and Service Guide 11
1 Getting Started
Table 1-2 Keypad functions
Key Description
Measurement related operation
Press to select the DC voltage measurement.
Press to select the AC voltage measurement.
+
+
Press to select the DC current measurement.
Press to select the AC current measurement.
Press to select the AC+DC voltage measurement.
Press to select the AC+DC current measurement.
Press to toggle between the 2-wire resistance or 4-wire resistance measurement.
Press to select the frequency measurement.
+
Press to toggle between the diode and continuity measurement.
Press to select dBm measurement.
Press to toggle between manual ranging and autoranging.
Press to select a higher range and disable autoranging. See “Selecting a
Range” on page 30 for more information.
Press to select a lower range and disable autoranging. See
Range” on page 30 for more information.
12 U3402A User’s and Service Guide
Getting Started 1
Table 1-2 Keypad functions
Key Description
Press to select compare math operation.
+
+
Press to select and set the high limit for compare math operation.
+
Press to select and set the low limit for compare math operation.
Press to enable Hold math operation. See “Hold” on page 26 for more
information.
Press to select the reference impedance for dBm measurement.
+
+
+
Press to enable the MinMax math operation.
Press to select the reading rate. See
“Setting the Reading Rate” on page 32 for more information.
Press to select the relative math operation.
Press to toggle in and out of the relative base (Rel#). See “Rel” on page 22
for more information.
U3402A User’s and Service Guide 13
1 Getting Started
The terminals at a glance
C A U T I O N
To avoid damaging this multimeter, do not exceed the rated input limit.
14
N O T E
Figure 1-8
Input terminal
Voltages above 300 VAC may be measured only in circuits that are isolated from mains.
However, transient overvoltages are also present on circuits that are isolated from mains.
The Agilent U3402A is designed to safely withstand occasional transient overvoltages up to
2500 V PEAK. Do not use this multimeter to measure circuits where transient overvoltages could exceed this level.
U3402A User’s and Service Guide
Getting Started 1
Table 1-3 Input terminal for different measurement functions
Measurement function
DC voltage (VDC)
Input terminal
AC voltage (VAC), frequency (Hz)
Miliampere (mA), frequency (Hz)
12A, frequency (Hz)
V
W Hz (Hi) mA
12A
Lo
Overload protection
1000 VDC
750 VAC RMS, 1100 V PEAK,
2x10
7
V-Hz normal mode, or
1x10
6
V-Hz common mode
1200 mADC or AC RMS
12 ADC or AC RMS for continuous 30 seconds, or
10 ADC or AC RMS
2-wire resistance (
W (2W))
Diode test, continuity test
4-wire resistance (
W (4W))
All functions
V
W Hz
Hi
Any terminal to earth
500 VDC or AC RMS
250 VDC or AC RMS
1000 VDC or AC PEAK
U3402A User’s and Service Guide 15
1 Getting Started
The rear panel at a glance
AC line voltage selector
Line voltage fuse selection table
16
AC power connector
AC line fuse Chassis ground lug
Figure 1-9
Rear panel
RS232 interface connector
(For calibration use only)
U3402A User’s and Service Guide
Getting Started 1
Making Measurements
N O T E
The following pages show you how to make measurement connections and how to select measurement functions from the front panel for each of the measurement functions.
• After measuring a high voltage measurement of up to 1000 VDC, you are recommended to wait for approximate two minutes before measuring a low-level measurement with
1 to 10 µV resolutions.
• After measuring a high current measurement using the A input terminal, you are recommended to wait for approximate ten minutes before measuring a low-level DC measurements of volts, amperes, or ohms; to achieve accurate measurement. This is due to the thermal voltages generated during the high current measurements that may cause errors when measuring the low-level measurements.
U3402A User’s and Service Guide 17
1 Getting Started
Performing voltage measurements
C A U T I O N
Ensure that the terminal connections are connected correctly before making any measurement. To avoid damaging the multimeter, do not exceed the rated input limit.
AC voltage:
• Five ranges:
•
Slow reading rate: 120.000 mV, 1.20000 V, 12.0000 V, 120.000 V, 750.00 V.
•
Medium reading rate: 400.00 mV, 4.0000 V, 40.000 V,400.00 V, 750.0 V.
•
Fast reading rate: 400.0 mV, 4.000 V, 40.00 V, 400.0 V, 750 V.
• Measurement method: AC coupled true RMS—measures the AC component with up to 400 VDC bias on any range
• Crest factor: Maximum 3:0 at full scale
• Input impedance: 1 M
W ± 2% in parallel with < 120 pF on all ranges
• Input protection: 750 V RMS on all ranges
1 Press .
2 Connect the red and black test leads to the respective input terminals as shown in
3 Probe the test points and read the display. In auto- ranging mode, the multimeter automatically selects the appropriate range and the measurement is displayed.
+
ACV voltage source
–
18
Figure 1-10 ACV terminal connection and display
U3402A User’s and Service Guide
Getting Started 1
DC voltage:
• Five ranges:
•
Slow reading rate: 120.000 mV, 1.20000 V, 12.0000 V, 120.000 V, 1000.00 V.
•
Medium reading rate: 400.00 mV, 4.0000 V, 40.000 V, 400.00 V, 1000.0 V.
•
Fast reading rate: 400.0 mV, 4.000 V, 40.00 V, 400.0 V, 1000 V.
• Measurement method: Sigma Delta A-to-D converter
• Input impedance: 10 M
W ± 2% range (typical)
• Input protection: 1000 V on all ranges
1 Press .
2 Connect the red and black test leads to the respective input terminals as shown in
3 Probe the test points and read the display. In autoranging mode, the multimeter automatically selects the appropriate range and the measurement is displayed.
+
DCV voltage source
–
Figure 1-11 DCV terminal connection and display
U3402A User’s and Service Guide 19
1 Getting Started
Performing current measurements
Measuring AC (RMS) or DC Current
in mA
• Three AC current or DC current ranges:
•
Slow reading rate: 12.0000 mA, 120.000 mA, 1.20000 A
•
Medium reading rate: 40.000 mA, 120.00 mA, 1.2000 A
•
Fast reading rate: 40.00 mA, 120.0 mA, 1.200 A
• Shunt resistance: 0.1
W to 10 W for 12 mA to 1.2 A range
• Input protection: Front panel 1.25 A, 500 V FH fuse for one terminal
1 Press
.
2 Power off the measured circuit.
3 Connect the red and black test leads to mA input terminal as shown in
.
4 Probe the test points in series with the circuit
5 Power on the measured circuit and read the display.
+
ACI or DCI current source
–
Figure 1-12 ACI RMS or DCI (mA) terminal connection and display
20 U3402A User’s and Service Guide
Getting Started 1
N O T E
Measuring AC (RMS) or DC Current
up to 12 A
• One range:
•
12.0000A for DC or AC RMS continuous
•
12.0000 ADC or AC RMS for maximum 30 seconds
• Shunt resistance: 0.01
W for 12 A range
• Input protection: Internal 15 A, 600 V fuse for 12A terminal
1 Press
.
2 Power off the measured circuit.
3 Connect the red and black test leads to the A input terminal as shown in
.
4 Probe the test points in series with the circuit.
5 Power on the measured circuit and read the display.
Autoranging is not applicable for current measurement up to 12 A. You are required to select the range manually when a signal is applied to the A terminal.
–
ACI or DCI current source
+
Figure 1-13 ACI RMS or DCI (A) terminal connection and display
U3402A User’s and Service Guide 21
1 Getting Started
Performing frequency measurements
WA R N I N G
Use the frequency counter for low voltage applications. Do not use the frequency counter on AC power line systems.
Measuring frequency
• Five ranges: 120.000 mV, 1.20000 V, 12.0000 V, 120.000 V, 750.00 V — range is based on the voltage level of the signal, not frequency
• Measurement method: Reciprocal counting technique
• Signal level: 10% of range to full scale input on all ranges
• Gate time: 0.1 s or 1 period of the input signal, whichever is longer
• Input protection: 750 V RMS on all ranges
1 Press .
2 Connect the red and black test leads to the input terminal as shown in
.
3 Probe the test points and read the display. In auto- ranging mode, the multimeter automatically selects the appropriate range and the measurement is displayed.
+
Frequency source
–
22
Figure 1-14 Frequency terminal connection and display
U3402A User’s and Service Guide
Getting Started 1
Performing resistance measurements
C A U T I O N
Disconnect circuit power and discharge all high-voltage capacitors before measuring resistance to avoid damaging the multimeter or the device under test.
Measuring resistance
• Seven ranges:
•
Slow reading rate: 120 .000
W, 1.20000 kW, 12.0000 kW, 120.000 kW, 1.20000 MW, 12.0000 MW,
120.000 M
W
•
Medium reading rate: 400 .00
W, 4.0000 kW, 40.000 kW, 400.00 kW, 4.0000 MW, 40.000 MW,
300.00 M
W
•
Fast reading rate: 400 .0
W, 4.000 kW, 40.00 kW, 400.0 kW, 4.000 MW, 40.00 MW, 300.0 MW
• Measurement method: 2-wire ohms or 4-wire ohms, open circuit voltage limited to < 5 V
• Input protection: 500 V on all ranges
1 Press . The default function is 2- wire
W measurement.
2 Connect the red and black test leads to the input terminal as shown in
Figure 1- 15 or Figure 1- 16 on page 24 respectively (according to the
selected measurement method).
3 Probe the test points (by shunting the resistor) and read the display. In auto- ranging mode, the multimeter automatically selects the appropriate range and the measurement is displayed.
U3402A User’s and Service Guide 23
1 Getting Started
Resistance
+
–
Test current
Figure 1-15 2-wire
W
terminal connection and display
Resistance
+
–
Test current
Figure 1-16 4-wire W terminal connection and display
24 U3402A User’s and Service Guide
Getting Started 1
N O T E
Performing diode/continuity test
Testing diodes
The diode test measures the forward voltage of a semiconductor junction of approximately 0.5 mA. The beeper will emit a single beep tone when the input voltage is below +0.7 V (approximately 1.4 k
W) and emits a continuous beep tone when the input voltage is below 50 mV (approximately 100
W).
Measurements are displayed as shown below:
Reading rate
Slow
Medium
Fast
Measurement display
1.2 V range
2.5 V range
4.0 V range
The measurement value will display OL (overload) when the voltage measured is
• > 1.2 V at slow reading rate
• > 2.5 V at medium reading rate
• > 4.0 V at fast reading rate
C A U T I O N
Disconnect circuit power and discharge all high-voltage capacitors before testing diodes to avoid damaging the multimeter.
• Measurement method: 0.83 mA ± 0.2% constant current source, open-circuit voltage limited to < 5 V
• Response time: 70 samples per seconds with audible tone
• Gate time: 0.1 s or 1 period of the input signal, whichever is longer
• Input protection: 500 V RMS on all ranges
U3402A User’s and Service Guide 25
1 Getting Started
N O T E
To test a diode, switch the circuit power off, and remove the diode from the circuit. Then proceed as follows:
1 Press . The default function is diode measurement.
2 Connect the red and black test leads to the input terminal as shown in
.
3 Connect the red test lead to the positive terminal (anode) of the diode and
the black test lead to the negative terminal (cathode). Refer to Figure 1- 17 .
The cathode of a diode is indicated with a band.
4 Read the display.
5 Reverse the probes and measure the voltage across the diode again as shown
. Assess the diode according to the following guidelines:
• A diode is considered good if the multimeter displays
OL in reverse bias mode.
• A diode is considered shorted if the multimeter displays approximately
0 V in both forward and reverse bias modes, and the multimeter beeps continuously.
• A diode is considered open if the multimeter displays
OL in both forward and reverse bias modes.
26 U3402A User’s and Service Guide
Getting Started 1
Forward bias
+
–
Test current
Figure 1-17 Forward-biased diode/continuity test terminal connection and display
Reverse bias
+
Test current
–
Figure 1-18 Reverse-biased diode/continuity terminal connection and display
U3402A User’s and Service Guide 27
1 Getting Started
C A U T I O N
Testing Continuity
The continuity test measures the resistance of a tested circuit with 2- wire method at approximately 0.5 mA and determines whether the circuit is intact.
The beeper emits a continuous beep tone when the input resistance value is less than the approximate 10
W.
Measurement are displayed as shown below:
Reading rate Measurement display
Slow 120.000
W range
Medium
Fast
400.00
W range
400.0
W range
Disconnect circuit power and discharge all high-voltage capacitors before testing the circuit continuity to avoid damaging the multimeter or the device under test.
• Measurement method: 0.83 mA ± 0.2% constant current source, open circuit voltage limited to < 5 V
• Continuity threshold: 10
W fixed
• Input protection: 500 V RMS on all ranges
1 Press
to toggle to the continuity measurement function.
2 Connect the red and black test leads to the input terminal as shown in
.
3 Probe the test points and read the display.
28 U3402A User’s and Service Guide
Resistance
+
–
Test current
Figure 1-19 2-wire/continuity test terminal connection and display
Getting Started 1
U3402A User’s and Service Guide 29
1 Getting Started
Selecting a Range
You can allow the multimeter to select the range automatically by using autoranging, or you can select a fixed range using manual ranging.
Auto- ranging is convenient because the multimeter automatically selects the appropriate range for sensing and displaying each measurement. However, manual ranging results in better performance, since the multimeter does not have to determine which range to use for each measurement.
Selects autoranging and disables manual ranging. Press to toggle between the manual ranging and autoranging.
Selects a higher range and disable autoranging.
Selects a lower range and disable autoranging.
For dual display, the measurement range for primary and secondary display as stated below is similar and is unable to be changed independently.
•
DCV/DCV
•
DCV/ACV
•
ACV/DCV
•
ACV/ACV
•
ACV+DCV/DCV
•
ACV+DCV/ACV
•
•
•
•
•
•
DCI/DCI
DCI/ACI
ACI/DCI
ACI/ACI
ACI+DCI/DCI
ACI+DCI/ACI
• Autoranging is selected at default factory power- on.
• Manual ranging — If the input signal is greater than can be measured on the selected range, the multimeter will display an overload indication, OL on the primary or secondary display front panel.
• The multimeter remembers the selected ranging method (auto or manual) and the selected manual range for each measurement function.
30 U3402A User’s and Service Guide
Getting Started 1
N O T E
• Autorange thresholds — The multimeter shifts ranges as follows:
•
Down range at < 5% of current range
•
Up range at > full scale of current range
•
shows the summary of range values for slow, medium, and fast reading rate respectively.
Autoranging is not applicable for current measurement up to 12 A. You are required to select the range manually when a signal is applied to the A terminal.
Table 1-4 Range scale value in slow, medium, and fast reading rate
Measurement function
Slow reading rate
Range
Medium reading rate Fast reading rate
Autoranging
DCV
ACV, DCV + ACV
DCI, ACI, DCI + ACI
DCI, ACI, DCI + ACI
120.000 mV, 1.20000 V,
12.0000 V, 120.000 V,
1000.00 V
120.000 mV, 1.20000 V,
12.0000 V, 120.000 V,
750.00 V
12.0000 mA, 120.000 mA,
1200.00 mA
12.0000 A
400.00 mV, 4.0000 V,
40.000 V, 400.00 V,
1000.0 V
400.00 mV, 4.0000 V,
40.000 V, 400.00 V,
750.0 V
40.000 mA, 120.00 mA,
1200.0 mA
12.000 A
]
400.0 mV, 4.000 V,
40.00 V, 400.0 V, 1000 V
400.0 mV, 4.000 V,
40.00 V, 400.0 V, 750 V
40.00 mA, 120.0 mA,
1200 mA
12.00 A
]
Manual only
Frequency
Resistance
]
1200.00 Hz, 12.0000 kHz,
120.000 kHz, 1.00000 MHz
120.000
W, 1.20000 kW,
12.0000 k
W, 120.000 kW,
1.20000 M
W, 12.0000 MW,
120.000 M
W
1200.0 Hz, 12.000 kHz,
120.00 kHz, 1.0000 MHz
400.00
W, 4.0000 kW,
40.000 k
W, 400.00 kW,
4.0000 M
W, 40.000 MW,
300.00 M
W,
1200 Hz, 12.00 kHz,
120.0 kHz, 1.000 MHz
400.0
W, 4.000 kW,
40.00 k
W, 400.0 kW,
4.000 M
W, 40.00 MW,
300.0 M
W,
Diode test
1.20000 V 2.5000 V 2.500 V
2-wire
W/120 W
2-wire
W/400 W
2-wire
W/400 W
Fixed range
Continuity
(continuity mode) (continuity mode) (continuity mode)
Notes
:
1 10 A continuous DC or AC RMS; 12 A DC or AC RMS for 30 seconds maximum.
2 A shielded test cable is recommended when measuring resistance more than 120 k
W to eliminate noise interference that might be induced to the test leads.
U3402A User’s and Service Guide 31
1 Getting Started
Setting the Reading Rate
You can select three reading rates for the AC and DC voltage, AC and DC current, and resistance measurement which are Slow (
S), Medium (M), and
Fast (
F)
The selected rate allows you to maximize either the measurement speed or noise rejection, which affects the measurement accuracy. Refer to
.
Table 1-5 Reading rates for single function measurements
Reading rate Resolution Display counts
[1][2]
Slow 5 ½ 119,999
Medium 4 ½ 39,999
Fast
Notes:
3 ½ 3,999
1 In VDC 1000 V range, the display counts is limited up to 1200.00, 1200.0 and 1200 for slow, medium, and fast reading rate respectively.
2 In VAC 750 V range, 1000 V RMS is measureable.
The annunciator S, M, and F (slow, medium and fast respectively) are located at the corner left of the display indicates the selected rate on the primary display. Refer to
.
32
Figure 1-20 Reading rate annunciator
N O T E
Press to cycle through the available reading rates ( S, M, or F).
In the dBm function, the display counts is 0.01 dBm for slow or medium reading rate and
0.1 dBm for fast reading rate.
U3402A User’s and Service Guide
U3402A 5 1/2 Digit Dual Display Multimeter
User’s and Service Guide
2
Operations and Features
Combination of Math Operations 27
Using the Secondary Display 29
Changing the Configurable Settings 32
This chapter explains the various functions and features available to the U3402A digital multimeter
Agilent Technologies
19
2 Operations and Features
Operating Math Operations
presents a summary of the math operations that can be used with each measurement function.
Table 2-1 Math operations for different measurement functions
Measurement functions dBm Rel
Allowed math operations
Min Max Comp Hold
DCV
DCI
Resistance
ACV
ACI
Frequency
Diode/Continuity
—
—
—
—
—
• All math operations can be toggled on and off by reselecting the same math operation.
• Only one math operation can be turned- on at a time. When selecting another math operation when one is already on, you are required to turn- off the first operation and then turn- on the second math operation.
• All math operations are automatically turned- off when changing the measurement functions.
• Range changing is allowed for all math operations.
20 U3402A User’s and Service Guide
Operations and Features 2
dBm
The logarithmic dBm (decibels relative to one milliwatt) scale is often used in
RF signal measurements. The multimeter’s dBm operation takes a measurement and calculates the power delivered to a reference impedance
(typically 50, 75, or 600
W). The formula used for conversion from the voltage reading is:
dBm
= 10 x Log
10
[ 1000 x (Reading
2
/ reference impedance
)
]
Figure 2-1
Typical dBm operation display
The default reference impedance value is 600
W. You can select the following 21 reference impedance values:
2
W, 4 W, 8 W, 16 W, 50 W, 75 W, 93 W, 110 W, 124 W, 125 W, 135 W, 150 W, 250 W,
300
W, 500 W, 600 W, 800 W, 900 W, 1000 W, 1200 W, or 8000 W..
If reference impedance 2
W, 4 W, 8 W, or 16 W is selected, the dBm operation is displayed in watt (power).
Numeric results are in the range of ±120.000 dBm with 0.01 dBm resolution shown, independent of the number of digits setting.
The dBm operation can be applied to DCV and ACV measurement functions only. The multimeter displays the dBm operation on the primary display and displays the reference impedance selection on the secondary display.
Procedure 1
1 Press to enter dBm operation.
2 Use to select the desired reference impedance.
3 Press
to store the selected value and read the display.
U3402A User’s and Service Guide 21
2 Operations and Features
Procedure 2
1 Press . The reference impedance currently used will be displayed on the secondary display.
2 Use to select the desired reference impedance.
3 Press to store the selected value.
4 Press to enable the dBm operation and read the display.
Rel
When making Rel (relative) measurements, each reading is the difference between a stored relative value and the input signal. For example, this feature can be used to make more accurate resistance measurements by nulling the test lead resistance.
After you enable the Rel operation, the multimeter stores the next reading as a
Rel # (relative base) and immediately displays on the primary display:
Primary Display = Reading – Rel #
22
Figure 2-2
Typical Rel operation display
The multimeter allows relative settings for the following measurement functions: DC voltage, AC voltage, DC current, AC current, resistance, and frequency.
U3402A User’s and Service Guide
Operations and Features 2
N O T E
Procedure
1 Press
2 Press
to enable Rel operation.
to toggle in (and out) the Rel #.
3 Use , , , and to modify the Rel #.
4 Press to set the Rel # and read the display.
5 Press
to disable the Rel operation.
• In resistance measurement mode, the multimeter will read a non-zero value even when the two test leads are in direct contact, because of the resistance of these leads. Use the Rel operation to zero-adjust the display.
• In DC voltage measurement mode, the thermal effect will influence the accuracy. Short the test leads and press the display.
once the displayed value is stable to zero-adjust
MinMax
The MinMax (Minimum/Maximum) operation stores the minimum and maximum values of reading during a series of measurements.
When enabled, the MinMax operation turns on the
MINMAX
annunciator and begins accumulating various statistics of the readings being displayed.
Figure 2-3
Typical Max operation display
U3402A User’s and Service Guide 23
2 Operations and Features
24
N O T E
Figure 2-4
Typical Min operation display
Each time a new minimum or maximum value is stored, the multimeter beeps once (if the beeper is enabled) and briefly turns on the appropriate
MAX
or
MIN
annunciator.
Accumulated statistics are:
• MAX—maximum reading since MinMax was enabled
• MIN—minimum reading since MinMax was enabled
• MINMAX—actual readings
When MinMax is enabled, the measurement range changed to manual ranging and the current measurement mode is locked until MinMax is disabled (with other ranges being selected or autoranging is enabled).
Procedure
1 Press
to enable MinMax operation.
2 Press to cycle through the available MINMAX operations ( MIN, MAX, or MINMAX).
3 Press to disable the MinMax operation.
U3402A User’s and Service Guide
Operations and Features 2
Comp
The Comp (compare) operation allows you to perform pass/fail testing against specified upper and lower limits. You can set the upper and lower limits to any value between 0 and
\100% of the highest range for the present function.
Figure 2-5
Typical Comp operation display
When enabled, the actual readings are shown in primary display and the comparison results such as
HI, LO, or PASS is shown in secondary display.
• You should specify the upper limit to always be a more positive number than the lower limit. The initial factory setting for
LO limit is 0.
• The secondary display shows
PASS
when readings are within the specified limits. The secondary display shows
HI
when the reading is outside the high limit and
LO
when the reading is outside the low limit.
• When the beeper is ON (see
“Using the Setup Menu” on page 31), the beeper
beeps on the transition from
PASS
to HI or PASS to LO or when transitioning directly from
HI
to LO or LO to HI (no PASS in between).
• Press to enable Comp operation.
Procedure
1 Press
to enter the upper limit setup mode.
The upper limit is shown on primary display while the HI annunciator is shown on secondary display.
2 Use
, , , and to modify the upper limit.
U3402A User’s and Service Guide 25
2 Operations and Features
3 Press
to store the specified value.
4 Press to enter the lower limit setup mode.
The lower limit is shown on primary display while the LO annunciator is shown on secondary display.
5 Use
6 Press
, , , and to modify the lower limit.
to store the specified value.
7 Press
8 Press
to enable the Comp operation.
to disable the Comp operation or exit.
Hold
The reading hold feature allows you to capture and hold a reading on the front panel display. When enabled, the Hold operation turns on the
Hold
annunciator and hold the reading.
Procedure
1 Press
to hold the reading on the display.
Figure 2-6
Typical Hold operation display
26 U3402A User’s and Service Guide
Operations and Features 2
Combination of Math Operations
The math operation can be operated for primary display only.
N O T E
The Agilent U3402A multimeter allows you to use multiple math operation such as dBm, MinMax, Rel, Hold, and Comp simultaneously.
Example:
Set upper and lower limit for Comp operation using multiple math operation as below:
1 Press to enable dBm operation.
2 Press
to take dBm readings as Rel # for new measurement.
3 Press
to record the Max value as the new Rel # (upper limit).
4 Press
to record the Min value as the new Rel # (lower limit).
5 Press to enable the Comp operation.
The step- by- step sequence and readings of the combined math operations are shown in
Table 2- 2 on page 28 when all math operations are used
sequentially. Refer to
.
U3402A User’s and Service Guide 27
2 Operations and Features
Table 2-2 Descriptions for combined math operations
Description
1
2
No.
Math operation sequence
3 dBm
Rel
Min
4
5
Max
Comp
Readings
Reading is calculated to a dBm operation dBm reading, “A” is taken as Rel #
Min reading of relative dB is recorded as a new Rel #, “B”
Max reading of relative dB is recorded as a new Rel #, “B”
Compare operation is performed based on the readings of “C” and “D”
“A”
“B”
“C”
“D”
“E”
1. Press
2. Press
3. Press
4. Press
28
5. Press
Figure 2-7
Combined math operations sequence
U3402A User’s and Service Guide
Operations and Features 2
Using the Secondary Display
To enable the secondary display mode:
1 Press followed by a specified function keys such as ,
, , or to enable the secondary display mode.
The
2ND annunciator is displayed along with the secondary display.
,
Secondary display
2ND annunciator
Figure 2-8
Secondary display
To disable the secondary display mode:
1 Press
.
Table 2- 3 details the available input combinations for both primary and
secondary display when dual display mode is selected.
U3402A User’s and Service Guide 29
2 Operations and Features
Table 2-3 Description for dual display combination
Primary display
DCV ACV
Secondary display
DCI
[
ACI
[
Hz
]
DCV
ACV
DCI
[
ACI
[
[
]
[
]
ACV + DCV
ACI + DCI
]
[
]
Frequency
]
Resistance
]
Diode/Continuity
[
]
[
]
dBm
[
Notes:
1 The range for both primary and secondary display are corresponding to the higher range of two displays (autoranging mode, while the range of secondary display are the same as the primary displays (manual range mode).
2 The frequency measurement corresponding to the current input signals; other measurements are corresponding to the voltage input signals.
3 In dual display mode, users are recommended to measure the resistance up to 1 M
W.
4 At 12 A range, manual range mode is selected by default.
5 Measurable with non-guaranteed accuracy.
6 Autoranging mode by default.
7 The voltage or current range of the frequency function follows the voltage or current range of the other function.
N O T E
In dual display mode, the
The multimeter has an increased key response time (0.6 s to 1 s) when in dual display mode. You may need to press the selected key until the multimeter responses.
30 U3402A User’s and Service Guide
Operations and Features 2
Using the Setup Menu
The Setup menu allows you to customize a number of non- volatile instrument
configurations. The content of the Setup menu are shown in Table 2- 4 .
Table 2-4 Setup menu and communication parameters
First tier menu
Description Second tier menu
Description Default factory setting
Available settings
rS232
]
RS232 interface parameters bAUd
PArtY dAtA
Baud rate for remote communication with a
PC (remote control).
Parity bit for remote communication with a
PC.
Data bit length for remote communication with a PC.
Stop bit.
9600
None
8
300, 600, 1200, 2400,
4800 or 9600
None, odd, or even
7 or 8
StoP 1 1 or 2
Echo
ECHO. Return a character to PC in remote communication.
OFF ON or OFF
Printer-Only. Print measured data to a PC in remote communication.
OFF ON or OFF bEEP
]
Beeper selection ON ON or OFF
Notes:
1 For calibration use only.
2 The beeper is use to simplify the multimeter operation. It is not a communication related parameter.
U3402A User’s and Service Guide 31
2 Operations and Features
Changing the Configurable Settings
The parameters in the Setup menu can be configured by using the following procedures:
1 Press
to access to the Setup menu.
2 Use and tier menu.
If you select bEEP,
to select the desired configurable items in the first
a Use
b Press
and to select the desired parameter.
to confirm the selected parameter.
c Press to quit from the Setup menu.
d The configuration settings is saved and main display is displayed.
If you select rS232,
a Press to enter the second tier menu.
b Use
and to select the desired configurable items.
c Use
and to select the desired parameters.
d Press
to confirm the selected parameter.
e Press
to quit from the tier menu or quit from the Setup menu.
f
The configuration settings is saved and main display is displayed.
32 U3402A User’s and Service Guide
U3402A 5 1/2 Digit Dual Display Multimeter
User’s and Service Guide
3
Application Tutorial
Applications for Using Dual Display 52
Dual Display Operation Examples 53
Measure DC Voltage and AC Ripple on a Rectification Circuit 53
Measure AC and DC Current on a Rectification Circuit 54
Measure AC Voltage and Frequency on an AC Circuit 55
Measure DC Voltage and DC Current on a Transistor Circuit or Load 56
Measure Resistance Using 2-Wire Mode 58
Measure Resistance Using 4-Wire Mode 59
This chapter describes the advanced features and the possible applications for effective operation of the multimeter.
Agilent Technologies
51
3 Application Tutorial
Applications for Using Dual Display
The dual display feature in the multimeter can be used to enhance test and
measurement capabilities. See Table 3- 1 for the available combinations and
application when using dual display.
Table 3-1 Typical combinations and applications when using dual display
No.
Primary display Secondary display
1 DCV ACV
2 ACV + DCV DCV
3
4
5
6
DCV
DCV
ACI + DCI
ACV
7
8
ACI + DCI
ACV
9 ACV
10 ACI
11 DCI
12 ACI + DCI
13 dBm
14 dBm
15 dBm
16 dBm
DCI
ACI
DCV
DCI
ACV
ACI
Frequency
Frequency
ACI
DCI
Reference
W
DCV
ACV
Frequency
Applications
• Test DC to AC or AC to DC converter circuit.
• Measure DC level and AC ripple of power supply.
• Test power supply load regulation.
• Check loop current and voltage drop level.
• Test line and load regulation.
• Test AC to DC or DC to AC converters.
• Measure DC level and AC ripple of power supply.
• Test transformer.
• Measure AC frequency response of amplifier circuit.
• Adjust AC motor control.
• Measure AC ripple and DC current of power supply.
• Measure current dissipation for power supply analysis.
• Set dB reference impedance and show dBm.
• Indicate DC voltage and dBm.
• Indicate AC voltage and dBm.
• Check frequency response.
52 U3402A User’s and Service Guide
Application Tutorial 3
Dual Display Operation Examples
This section describes some practical operations when using dual display feature.
Measure DC Voltage and AC Ripple on a Rectification Circuit
A single measurement for both DC voltage and AC ripple can be displayed through both display while testing a rectifier circuit.
1 Connect the red and black test leads to the input terminal and probe the test
points as shown in Figure 3- 1
.
AC/DC voltage source
+
–
Figure 3-1
Terminal connection when measuring DC voltage and AC ripple on a rectification circuit
2 Press
3 Press
4 Press
to select DC voltage measurement for primary display.
to enable the secondary display. The 2ND is displayed.
to select AC voltage measurement for secondary display.
U3402A User’s and Service Guide 53
3 Application Tutorial
N O T E
to select autoranging or manual ranging for the primary and secondary display. The ranging will be the same for both displays.
• Press to disable the secondary display.
• Press to select the suitable range if DCV+ACV ripple is above the scale of the current range.
Measure AC and DC Current on a Rectification Circuit
A single measurement for both AC current and DC current can be displayed through both display while testing a rectifier circuit.
WA R N I N G
• Make sure you select the correct input terminal according to the input range used.
• Do not apply more than 12 A to the A input terminal to avoid the multimeter from damage.
1 Connect the red and black test leads to the input terminal and probe the test points as shown in
.
AC/DC current source
–
+
54
Figure 3-2
Terminal connection when measuring AC and DC current on a rectification circuit
U3402A User’s and Service Guide
Application Tutorial 3
N O T E
2 Press
to select AC current measurement for primary display.
to select autoranging or manual ranging for secondary display.
4 Press to enable the secondary display. The 2ND is displayed.
5 Press
Press
to select DC current measurement for secondary display.
to disable the secondary display.
Measure AC Voltage and Frequency on an AC Circuit
A single measurement for both AC voltage and frequency can be displayed through both display while testing a rectifier circuit.
1 Connect the red and black test leads to the input terminal and probe the test
points as shown in Figure 3- 3
.
AC voltage source
+
–
Figure 3-3
Terminal connection when measuring AC voltage and frequency on an
AC circuit
2 Press to select AC voltage measurement for primary display.
U3402A User’s and Service Guide 55
3 Application Tutorial
N O T E
to select autoranging or manual ranging for primary display.
4 Press to enable the secondary display. The 2ND is displayed.
5 Press to select frequency measurement for secondary display.
to select autoranging or manual ranging for secondary display.
Press to disable the secondary display.
Measure DC Voltage and DC Current on a Transistor Circuit or
Load
A single measurement for both DC voltage and current can be displayed through both display while testing a transistor amplifier circuit. You may also check the H fe
or calculate the DC load consumption by using the dual display.
WA R N I N G
• Make sure you select the correct input terminal according to the input range used.
• Do not apply more than 12 A to the A input terminal to avoid the multimeter from damage.
56 U3402A User’s and Service Guide
Application Tutorial 3
1 Connect the red and black test leads to the input terminal and probe the test
points as shown in Figure 3- 4
.
DC voltage source
+
Load
–
N O T E
Figure 3-4
Terminal connection when measuring DC voltage and DC current on a transistor circuit or load
2 Press to select DC voltage measurement for primary display.
to select autoranging or manual ranging for primary display.
4 Press to enable the secondary display. The 2ND is displayed.
5 Press
to select DC current measurement for secondary display.
to select autoranging or manual ranging for secondary display.
Press to disable the secondary display.
U3402A User’s and Service Guide 57
3 Application Tutorial
Measure Resistance Using 2-Wire Mode
WA R N I N G
Do not apply voltage exceeding 500 V PEAK between V.
W
.Hz and Lo input terminals.
1 Connect the red and black test leads to the resistor and probe the test points as shown in
.
Resistance
+
–
N O T E
Figure 3-5
Terminal connection when measuring resistance using 2-wire mode
2 Press
to select the 2- wire
W measurement for primary display. The
2W is displayed.
to select autoranging or manual ranging for primary display.
When measuring low resistance, you may use the Rel operation to reduce the measurement error created by the test leads resistance and contact resistance in the test loop (0.1
W~ 0.5 W typical)
58 U3402A User’s and Service Guide
Application Tutorial 3
Measure Resistance Using 4-Wire Mode
WA R N I N G
Do not apply voltage exceeding 250 V PEAK between Sense Hi and Sense Lo terminals, and 500 V PEAK between V.
W
.Hz and Lo input terminals.
1 Connect the red and black test leads to the resistor and probe the test points as shown in
.
Resistance
+
–
Figure 3-6
Terminal connection when measuring resistance using 4-wire mode
2 Press
to select the 4- wire
W measurement for primary display. The
4W is displayed.
to select autoranging or manual ranging for primary display.
U3402A User’s and Service Guide 59
3 Application Tutorial
Measure True RMS AC+DC
The multimeter can measure the true RMS value of the AC voltage and AC current.
1 Press
and , or and simultaneously. The multimeter will measure the DC and AC signals alternatively, calculate and display the AC+DC (RMS) value using the equation below:
AC+DC (RMS) = DC
2
+ AC
2
N O T E
When AC+DC voltage measurement is selected, the DCV input impedance is paralleled with the AC coupled 1.1 M
W AC divider.
60 U3402A User’s and Service Guide
U3402A 5 1/2 Digit Dual Display Multimeter
User’s and Service Guide
4
Performance Test
Agilent Technologies Calibration Services 62
Performance Verification Test Overview 65
Performance Verification Test 65
DC Voltage Verification Test 65
DC Current Verification Test 67
Resistance Verification Test 68
Frequency Verification Test 71
AC Voltage Verification Test 72
AC Current Verification Test 73
This chapter contains performance test procedures. The performance test procedures allow you to verify that the multimeter is operating within its published specifications.
WA R N I N G
Shock hazard. Only service-trained personnel who are aware of the hazards involved should perform the procedures in this chapter. To avoid electrical shock and personal injury, make sure to read and follow all test equipment safety instructions.
Use only completely electrically insulated test lead sets with connectors that prevent contact with test voltages.
Agilent Technologies
61
4 Performance Test
Calibration Overview
N O T E
Make sure you have read “Test Considerations” on page 64 before
calibrating the multimeter.
Agilent Technologies Calibration Services
When your instrument is due for calibration, contact your local Agilent Service
Center. The U3402A is supported on automated calibration systems at Agilent service centres only.
Calibration Interval
A one- year interval is adequate for most applications. Accuracy specifications are warranted only if adjustment is made at regular calibration intervals.
Accuracy specifications are not warranted beyond the one- year calibration interval. Agilent does not recommend extending calibration intervals beyond two years for any application. When an adjustment is required, contact your local Agilent Service Center.
62 U3402A User’s and Service Guide
Performance Test 4
Recommended Test Equipment
The test equipments recommended for the performance verification procedures are listed below. If the exact instrument is not available, substitute calibration standards of equivalent accuracy.
A suggested alternate method would be to use the Agilent 3458A 8½ - digit digital multimeter to measure less accurate yet stable sources. The output value measured from the source can be entered into the instrument as the target calibration value.
Table 4-1 Recommended test equipments
Application
Zero calibration
DC voltage
DC current
Resistance
AC voltage
AC current
Frequency
Diode
Recommended equipment Recommended accuracy requirements
Shorting plug — Dual banana plug with copper wire short between the two terminals
Fluke 5520A
Fluke 5520A
Fluke 5520A
Fluke 5520A
Fluke 5520A
Fluke 5520A
Fluke 5520A
—
<1/5 instrument 1 year spec
<1/5 instrument 1 year spec
<1/5 instrument 1 year spec
<1/5 instrument 1 year spec
<1/5 instrument 1 year spec
<1/5 instrument 1 year spec
<1/5 instrument 1 year spec
U3402A User’s and Service Guide 63
4 Performance Test
Test Considerations
For optimum performance, all procedures should comply with the following recommendations:
• Ensure that the calibration ambient temperature is stable and between
18 °C and 28 °C. Ideally the calibration should be performed at 23 °C ±1 °C.
• Ensure ambient relative humidity is less than 80%.
• Allow a one- hour warm- up period with a shorting plug connected to the Hi and Lo input terminals.
• Use shielded twisted pair PTFE- insulated cables to reduce settling and noise errors. Keep the input cables as short as possible.
N O T E
Please ensure that the calibration standards and test procedures used do not introduce additional errors. Ideally, the standards used to verify and adjust the instrument should be an order of magnitude more accurate than each instrument range full-scale error specification.
64 U3402A User’s and Service Guide
Performance Test 4
Performance Verification Test Overview
Performance verification test is an extensive set of tests that are recommended as an acceptance test when you first received the instrument .
Use the performance verification test to verify the measurement performance of the instrument. The performance verification test uses the instrument's specifications listed in
Chapter 6 , “Specifications and Characteristics”.
N O T E
Performance Verification Test
The performance verification test is recommended as acceptance tests when you first received the instrument. The acceptance test results should be compared against the one- year test limits. After acceptance, you should repeat the performance verification tests at every calibration interval.
If the instrument fails performance verification, adjustment or repair is required. Contact your local Agilent Service Center for details.
Make sure you have read
“Test Considerations” on page 64 before doing the performance
verification test.
This test checks the full- scale reading accuracy of the instrument.
DC Voltage Verification Test
1 Connect the calibrator to the front panel Hi and Lo input terminals.
2 Select each function and range in the order shown in
Table 4- 2 . Provide the input shown in Table 4- 2 .
3 Make a measurement and observe the result. Compare measurement results
to the appropriate test limits shown in the Table 4- 2
. (Be certain to allow for appropriate source settling when using the Fluke 5520A.)
U3402A User’s and Service Guide 65
4 Performance Test
C A U T I O N
Table 4-2 DC voltage verification test
Function Reading rate Input
DC voltage Slow
Medium
Slow
Medium
0.000 V
0.00000 V
0.0000 V
0.000 V
0.00 V
0.00 V
0.0000 V
0.000 V
0.00 V
0.0 V
100.000 mV
1.00000 V
10.0000 V
100.000 V
1000.00 V
360.00 mV
3.6000 V
36.000 V
360.00 V
1000.0 V
Range
120 mV
1.2 V
12 V
120 V
1000 V
400 mV
4 V
40 V
400 V
1000 V
120 mV
1.2 V
12 V
120 V
1000 V
400 mV
4 V
40 V
400 V
1000 V
±0.5V
±0.02 mV
±0.17 mV
±1.7 mV
±17 mV
±170 mV
±90 µV
±900 µV
±9 mV
±90 mV
±0.6 V
Error from nominal one year
±8 µV
±50 µV
±0.5 mV
±5 mV
±50 mV
±50 µV
±500 µV
±5 mV
±50 mV
Set the calibrator output to 0 V before disconnecting it from the multimeter input terminals.
66 U3402A User’s and Service Guide
Performance Test 4
DC Current Verification Test
1 Connect the calibrator to the front panel Hi and Lo input connectors.
2 Select each function and range in the order shown in
Table 4- 3 . Provide the input shown in Table 4- 3 .
3 Make a measurement and observe the result. Compare measurement results to the appropriate test limits shown in
. (Be certain to allow for appropriate source settling when using the Fluke 5520A.)
Table 4-3 DC current verification test
Function Reading rate Input
DC current
Slow
Medium
Slow
Medium
0.0000 mA
0.000 mA
0.00 mA
0.0000 A
0.000 mA
0.00 mA
0.0 mA
0.000 A
10.0000 mA
100.000 mA
1000.00 mA
10.0000 A
36.000 mA
100.00 mA
1000.0 mA
10.000 A
Range
12 mA
120 mA
1200 mA
12 A
40 mA
120 mA
1200 mA
12 A
12 mA
120 mA
1200 mA
12 A
40 mA
120 mA
1200 mA
12 A ±3 mA
±6.5 µA
±55 µA
±2.05 mA
±20.5 mA
±42 µA
±0.13 mA
±2.3 mA
±23 mA
Error from nominal one year
±1.5 µA
±5 µA
±50 µA
±0.5 mA
±6 µA
±30 µA
±0.3 mA
C A U T I O N
Connect calibrator to multimeter’s A and Lo terminals before applying 10 A.
U3402A User’s and Service Guide 67
4 Performance Test
68
Resistance Verification Test
Configuration: 2- wire
W
1 Select the resistance function.
2 Select each range in the order shown in
Table 4- 4 . Provide the resistance
value indicated. Compare measurement results to the appropriate test limits shown in
. (Be certain to allow for appropriate source settling.)
Table 4-4 2-wire
W verification test
Function Reading rate Input
2-wire
W
Slow
Medium
Slow
0.000
W
0.00000
W
0.0000
W
0.000
W
0.00000
W
0.0000
W
0.000
W
0.00
W
0.0000
W
0.000
W
0.00
W
0.0000
W
0.000
W
0.00
W
100.000
W
1.00000 k
W
10.0000 k
W
100.000 k
W
1.00000 M
W
10.0000 M
W
100.000 M
W
Range
120
W
1.2 k
W
12 k
W
120 k
W
1.2 M
W
12 M
W
120 M
W
400
W
4 k
W
40 k
W
400 k
W
4 M
W
40 M
W
300 M
W
120
W
1.2 k
W
12 k
W
120 k
W
1.2 M
W
12 M
W
120 M
W
±0.3
W
[1]
±3
W
[1]
±30
W
±0.3 k
W
±3 k
W
±50 k
W
±108 m
W
[1]
±0.85
W
[1]
±6.5
W
[1]
±65
W
±0.65 k
W
±30.5 k
W
±3.008 M
W
Error from nominal one year
±8 m
W
[1]
±50 m
W
[1]
±0.5
W
[1]
±5
W
±50
W
±0.5 k
W
±8 k
W
±50 m
W
[1]
U3402A User’s and Service Guide
Performance Test 4
Table 4-4 2-wire
W verification test
Function Reading rate Input Range Error from nominal one year
2-wire
W
Medium 360.00
W
3.6000 k
36.000 k
W
W
400
4 k
W
40 k
W
W
±0.41
W
[1]
±3.2
W
[1]
±25
W
[1]
360.00 k
W
3.6000 M
W
20.000 M
W
400 k
4 M
W
40 M
W
W
±250
±5.7 k
W
±303 k
W
W
200.00 M
W
300 M
W
±10.05 M
W
Notes:
1 Specifications are for 2-wire ohms function using the Rel operation enabled to eliminate lead resistance.
Configuration: 4- wire
W
1 Select the resistance function.
2 Select each range in the order shown in
Table 4- 5 . Provide the resistance
value indicated. Compare measurement results to the appropriate test limits shown in
. (Be certain to allow for appropriate source settling.)
Table 4-5 4-wire
W verification test
Function Reading rate Input
4-wire
W
Slow 0.000
W
0.00000
W
0.0000
W
0.000
W
0.00000
W
0.0000
W
0.000
W
Range
120
W
1.2 k
W
12 k
W
120 k
W
1.2 M
W
12 M
W
120 M
W
Error from nominal one year
±8 m
W
[1]
±50 m
W
[1]
±0.5
W
±5
W
±50
W
±0.5 k
W
±8 k
W
U3402A User’s and Service Guide 69
4 Performance Test
Table 4-5 4-wire
W verification test
Function Reading rate Input Range Error from nominal one year
4-wire
W
Medium
Slow
Medium
0.00
0.00
W
0.0000
0.000
0.00
W
0.000
W
W
W
1.00000 M
360.00
W
0.0000
W
100.000
W
W
1.00000 k
10.0000 k
100.000 k
W
W
W
3.6000 k
W
36.000 k
W
360.00 k
W
3.6000 M
W
20.000 M
W
200.00 M
W
W
10.0000 M
W
100.000 M
W
400
4 k
W
40 k
400 k
W
4 M
W
40 M
300 M
120
1.2 k
12 k
4 k
W
W
W
W
W
W
120 k
1.2 M
4 M
W
W
W
12 M
W
120 M
W
400
W
W
40 k
W
400 k
W
40 M
W
W
300 M
W
±50 m
W
[1]
±0.3
W
±3
W
±30
W
±0.3 k
W
±3 k
W
±50 k
W
±58 m
W
[1]
±0.55
W
[1]
±5.5
W
±55
W
±0.55 k
W
±30.5 k
W
±3.008 M
W
±0.23
W
[1]
±2.1
W
±21
W
±0.21 k
±5.7 k
W
W
±0.303 k
W
±10.05 M
W
[1]
Specifications are for 4-wire ohms function using the Rel operation enabled to eliminate lead resistance.
70 U3402A User’s and Service Guide
Performance Test 4
Diode Verification Test
Configuration: Diode
1 Connect the calibrator to the front panel Hi and Lo input terminals.
2 Select each function and range in the order shown in
Table 4- 6 . Provide the input shown in Table 4- 6 .
3 Make a measurement and observe the result. Compare measurement results to the appropriate test limits shown in
. (Be certain to allow for appropriate source settling when using the Fluke 5520A.)
Table 4-6 Diode verification test
Function Reading rate Voltage
Diode
Slow
Medium
0.50000 V
1.00000 V
0.5000 V
2.0000 V
Range
1.2 V
1.2 V
2.5 V
2.5 V
Error from nominal one year
±0.11 mV
±0.17 mV
±0.6 mV
±0.7 mV
Frequency Verification Test
Configuration: Frequency
1 Select the frequency function.
2 Select each range in the order shown in Table 4- 7
. Provide the input voltage and frequency indicated. Compare measurement results to the appropriate test limits shown in
. (Be certain to allow for appropriate source settling.)
Table 4-7 Frequency verification test
Function Voltage
Frequency
Reading rate
Slow 1 V
Input frequency
1000.00 Hz
Range
1200 Hz
Error from nominal one year
±0.08 Hz
U3402A User’s and Service Guide 71
4 Performance Test
72
AC Voltage Verification Test
Configuration: AC volts
1 Select the AC voltage function.
2 Select each range in the order shown in
Table 4- 8 . Provide the indicated
input voltage and frequency. Compare measurement results to the
appropriate test limits shown in Table 4- 8 . (Be certain to allow for
appropriate source settling.)
Table 4-8 AC volts verification test
V RMS Function Reading rate
AC voltage
Slow
Medium
12.000 mV
100.000 mV
0.12000 V
1.00000 V
1.2000 V
10.0000 V
12.000 V
100.000 V
120.00 V
750.00 V
40.00 mV
360.00 mV
0.4000 V
3.6000 V
4.000 V
36.000 V
40.00 V
360.00 V
120.0 V
750.0 V
1 kHz
1 kHz
1 kHz
1 kHz
1 kHz
1 kHz
1 kHz
1 kHz
1 kHz
1 kHz
1 kHz
1 kHz
1 kHz
1 kHz
Input frequency
1 kHz
1 kHz
1 kHz
1 kHz
1 kHz
1 kHz
Range
750 V
750 V
400 mV
400 mV
4 V
4 V
40 V
40 V
120 mV
120 mV
1.2 V
1.2 V
12 V
12 V
120 V
120 V
400 V
400 V
750 V
750 V
Error from nominal one year
±0.124 mV
±0.3 mV
±1.24 mV
±3 mV
±12.4 mV
±30 mV
±124 mV
±0.3 V
±1.24 V
±2.5 V
±0.48 mV
±1.12 mV
±4.8 mV
±11.2 mV
±48 mV
±112 mV
±480 mV
±1.12 V
±4.2 V
±5.5 V
C A U T I O N
Set the calibrator output to 0 V before disconnecting it from the multimeter input terminals.
U3402A User’s and Service Guide
Performance Test 4
AC Current Verification Test
Configuration: AC current
1 Select the AC current function.
2 Select each range in the order shown in
Table 4- 9 . Provide the input
current and frequency indicated. Compare measurement results to the
appropriate test limits shown in Table 4- 9 . (Be certain to allow for
appropriate source settling.)
Table 4-9 AC current verification test
Current Function Reading rate
AC current
Slow
Medium
1.2000 mA
10.0000 mA
12.000 mA
100.000 mA
120.00 mA
1000.00 mA
1.2000 A
10.000 A
4.000 mA
36.000 mA
12.00 mA
100.00 mA
120.0 mA
1000.0 mA
1.200 A
10.000 A
1 kHz
1 kHz
1 kHz
1 kHz
1 kHz
1 kHz
1 kHz
1 kHz
1 kHz
1 kHz
Input frequency
1 kHz
1 kHz
1 kHz
1 kHz
1 kHz
1 kHz
Range
40 mA
40 mA
120 mA
120 mA
1200 mA
1200 mA
12 A
12 A
12 mA
12 mA
120 mA
120 mA
1200 mA
1200 mA
12 A
12 A ±110 mA
±60 µA
±0.22 mA
±0.18 mA
±0.62 mA
±1.8 mA
±6.2 mA
±24 mA
±112 mA
Error from nominal one year
±16 µA
±60 µA
±0.16 mA
±0.6 mA
±1.6 mA
±6 mA
±22 mA
C A U T I O N
Connect calibrator to multimeter’s A and Lo terminals before applying 10 A.
U3402A User’s and Service Guide 73
4 Performance Test
74 U3402A User’s and Service Guide
U3402A 5 1/2 Digit Dual Display Multimeter
User’s and Service Guide
5
Disassembly and Repair
To Replace the Power Line Fuse 79
To Replace a Current Input Fuse 80
Electrostatic Discharge (ESD) Precautions 80
This chapter will help you troubleshoot a faulty multimeter.
It describes how to disassemble the multimeter, how to obtain repair services, and lists the replaceable parts.
Agilent Technologies
75
5 Disassembly and Repair
Operating Checklist
Before returning your multimeter to Agilent for service or repair check the following items:
Is the multimeter inoperative?
q Verify the power line voltage setting.
q Verify the power line fuse is installed.
q Verify that the power cord is connected to the multimeter and to AC line power.
q Verify the front panel power switch is depressed.
See
.
Is the multimeter’s current input inoperative?
q Verify the current input fuse.
76 U3402A User’s and Service Guide
Disassembly and Repair 5
Types of Service Available
If your instrument fails during the warranty period, Agilent Technologies will repair or replace it under the terms of your warranty. After your warranty expires, Agilent offers repair services at competitive prices.
Extended Service Contracts
Many Agilent products are available with optional service contracts that extend the covered period after the standard warranty expires. If you have such a service contract and your instrument fails during the covered period,
Agilent Technologies will repair or replace it in accordance with the contract.
Obtaining Repair Service (Worldwide)
To obtain service for your instrument (in- warranty, under service contract, or post- warranty), contact your nearest Agilent Technologies Service Center.
They will arrange to have your instrument repaired or replaced, and can provide warranty or repair- cost information where applicable.
To obtain warranty, service, or technical support information you can contact
Agilent Technologies at one of the following telephone numbers:
In the United States: (800) 829–4444
In Europe: 31 20 547 2111
In Japan: 0120–421–345
Or use our Web link for information on contacting Agilent worldwide:
www.agilent.com/find/assist
Or contact your Agilent Technologies representative.
Before shipping your instrument, ask the Agilent Technologies Service Center to provide shipping instructions, including what components to ship. Agilent recommends that you retain the original shipping carton for use in such shipments.
U3402A User’s and Service Guide 77
5 Disassembly and Repair
Repackaging for Shipment
If the instrument is to be shipped to Agilent for service or repair, be sure to:
• Attach a tag to the unit identifying the owner and indicating the required service or repair. Include the model number and full serial number.
• Place the unit in its original container with appropriate packaging material for shipping.
• Secure the container with strong tape or metal bands.
• If the original shipping container is not available, place your instrument in a container which will ensure at least 4 inches of compressible packaging material around all sides for the instrument. Use static- free packaging materials to avoid additional damage to your instrument.
Agilent suggests that you always insure shipments.
Cleaning
Clean the outside of the multimeter with a soft, lint- free, slightly dampened cloth. Do not use detergent. Disassembly is not required or recommended for cleaning.
78 U3402A User’s and Service Guide
Disassembly and Repair 5
To Replace the Power Line Fuse
The power line fuse is located within the multimeter’s fuse- holder assembly on the rear panel. The multimeter is shipped from the factory with a power- line fuse installed (according to country of destination). Refer to
Table 5- 1 . If you determine that the fuse is faulty, replace it with one of the
same size and rating.
Table 5-1 Type of supplied fuse (according to country of destination)
Type of fuse (time-lag, low breaking fuse)
0.25A, 250V, 5x20 mm
0.125A, 250V, 5x20 mm
Input line voltage
100 V to 120 V
220 V to 240 V
Part number
A02-62-25592-3U
A02-62-25648-1U
1 Disconnect the power cord. Depress tab 1 and
2 and pull the fuse holder from rear panel.
2 Remove the line voltage selector from fuse holder assembly.
2
1
3 Rotate the line voltage selector and reinstall it back, so that the correct voltage appears in the fuse holder window.
4 Replace fuse holder assembly in rear panel.
100, 120, 220 (230), or 240 Vac
U3402A User’s and Service Guide 79
5 Disassembly and Repair
To Replace a Current Input Fuse
Both the mA and the A current input terminals are fuse protected. The fuse
for the mA input terminal is located on the front panel (see page 14 ). The fuse
is a 1.25 A, 500 V fuse (refer to Table 5- 2
). If you determine that the fuse is faulty, replace it with one of the same size and rating.
The fuse for the A current input terminal is located inside the multimeter (see
) and requires partial disassembly of the multimeter. The fuse is a
15 A, 600 V fast- acting fuse (refer to Table 5- 2
). If you determine that the fuse is faulty, replace it with one of the same size and rating.
Table 5-2 Type of current input fuse
Type of fuse
1.25 A, 500 V fuse
15 A, 600 V fast-acting fuse
Part number
2110-1394
2110-1396
Electrostatic Discharge (ESD) Precautions
Almost all electrical components can be damaged by electrostatic discharge
(ESD) during handling. Component damage can occur at electrostatic discharge voltages as low as 50 volts.
The following guidelines will help prevent ESD damage when servicing the instrument or any electronic device.
• Disassemble instruments only in a static- free work area.
• Use a conductive work area to reduce static charges.
• Use a conductive wrist strap to reduce static charge accumulation.
• Minimize handling.
• Keep replacement parts in original static- free packaging.
• Remove all plastic, foam, vinyl, paper, and other static- generating materials from the immediate work area.
• Use only anti–static solder suckers.
80 U3402A User’s and Service Guide
Disassembly and Repair 5
Mechanical Disassembly
For procedures in this manual, the following tools are required for disassembly:
• T15 Torx driver
• T20 Torx driver
• #2 Pozi–drive screw driver
WA R N I N G
Shock hazard. Only service–trained personnel who are aware of the hazards involved should remove the instrument covers. To avoid electrical shock and personal injury, make sure to disconnect the power cord from the multimeter before removing the covers. Some circuits are active and have power applied even when the power switch is turned off.
General Disassembly
1 Remove power and all cables from the multimeter.
2 Remove the carrying handle by rotating the handle upright and pulling it out from the sides of the multimeter.
U3402A User’s and Service Guide 81
5 Disassembly and Repair
3 Remove the multimeter’s bumpers. Pull from a corner and stretch the bumpers off the multimeter.
4 Remove the rear bezel. Loosen the two captive screws in the rear bezel and remove the rear bezel.
82 U3402A User’s and Service Guide
Disassembly and Repair 5
5 Remove the cover. Remove the screw at the bottom of the cover and slide the cover off the multimeter.
Front Panel Removal
1 Remove on/off switch push rod. Gently move the power switch push rod toward the front of the multimeter to disengage it from the switch. Be careful not to twist or bend the push rod.
U3402A User’s and Service Guide 83
5 Disassembly and Repair
2 Remove the screw holding the front panel.
3 Disconnect the two ribbon cable connectors from the front panel.
84 U3402A User’s and Service Guide
Disassembly and Repair 5
4 Disconnect the individual front panel wires shown below.
15 A current fuse
5 There is now enough play to allow the side of the front panel to be pried from the chassis and removed as an assembly.
U3402A User’s and Service Guide 85
5 Disassembly and Repair
Front Panel Disassembly
1 Remove the keyboard and display assembly. Remove the two screws holding the circuit board. Lift the keyboard and display assembly from the plastic housing.
a The rubber keypad can now be pulled from the plastic housing.
86 U3402A User’s and Service Guide
Disassembly and Repair 5
Replaceable Parts
This section contains information for ordering replacement parts for your instrument. You can find the instrument support part list at Agilent’s Test &
Measurement Parts Catalog at http://www.agilent.com/find/parts
This parts list includes a brief description of each part with applicable Agilent part number.
To order replaceable parts
You can order replaceable parts from Agilent using the Agilent part number.
Note that not all parts listed are available as field- replaceable parts.
To order replaceable parts from Agilent, do the following:
1 Contact your nearest Agilent Sales Office or Service Center.
2 Identify the parts by the Agilent part number shown in the support parts list.
3 Provide the instrument model number and serial number.
U3402A User’s and Service Guide 87
5 Disassembly and Repair
88 U3402A User’s and Service Guide
U3402A 5 1/2 Digit Dual Display Multimeter
User’s and Service Guide
6
Specifications and Characteristics
True RMS AC Voltage (AC Coupling Mode) 93
True RMS AC Voltage (AC+DC Coupling Mode) 94
True RMS AC Current (AC Coupling Mode) 95
True RMS AC Current (AC+DC Coupling Mode) 96
Supplemental Measurement Specifications 98
To Calculate Total Measurement Error 109
Configuring for Highest Accuracy Measurements 111
This chapter describes the multimeter’s specifications and operating specifications
Agilent Technologies
89
6 Specifications and Characteristics
DC Specifications
Table 6-1 DC resolution, full scale reading, and accuracy [±(% of reading + count)]
Function
Voltage
Rate
Slow
Med
Fast
Range
120.000 mV
1.20000 V
12.0000 V
120.000 V
1000.00 V
400.00 mV
4.0000 V
40.000 V
400.00 V
1000.0 V
400.0 mV
4.000 V
40.00 V
400.0 V
1000 V
Resolution
0.001 mV
0.00001 V
0.0001 V
0.001 V
0.01 V
0.01 mV
0.0001 V
0.001 V
0.01 V
0.1 V
0.1 mV
0.001 V
0.01 V
0.1 V
1 V
Maximum reading
119.999
1.19999
11.9999
119.999
1000.00
[
399.99
3.9999
39.999
399.99
1000.0
[
399.9
3.999
39.99
399.9
1000
[
Accuracy
(One year; 23ºC
± 5ºC)
±0.012% + 8
]
±0.012% + 5
±0.012% + 5
±0.012% + 5
±0.012% + 5
±0.012% + 5
±0.012% + 5
±0.012% + 5
±0.012% + 5
±0.012% + 5
±0.012% + 2
±0.012% + 2
±0.012% + 2
±0.012% + 2
±0.012% + 2
Typical input impedance
10.0 M
W
10.0 M
W
11.1 M
W
10.1 M
W
10.0 M
W
10.0 M
W
11.1 M
W
10.1 M
W
10.0 M
W
10.0 M
W
10.0 M
W
11.1 M
W
10.1 M
W
10.0 M
W
10.0 M
W
Burden voltage
[
—
—
—
90 U3402A User’s and Service Guide
Specifications and Characteristics 6
Table 6-1 DC resolution, full scale reading, and accuracy [±(% of reading + count)] (continued)
Function
Current
Diode/
Continuity
Resistance
/
Continuity
(2-wire)
Rate Range
Slow
Med
Fast
12.0000 mA
120.000 mA
1200.00 mA
12.0000 A
40.000 mA
120.00 mA
1200.0 mA
12.000 A
40.00 mA
120.0 mA
1200 mA
12.00 A
Slow
Med
—
—
Fast —
Slow 120.000
W
Med 400.00
W
Fast 400.0
W
Resolution
0.0001 mA
0.001 mA
0.01 mA
0.0001A
0.001 mA
0.01 mA
0.1 mA
0.001 A
0.01 mA
0.1 mA
1 mA
0.01 A
0.00001 V
0.0001 V
0.001 V
0.001
W
0.01
W
0.1
W
Maximum reading
11.9999
119.999
1199.99
11.9999
39.999
119.99
1199.9
11.999
39.99
119.9
1199
11.99
1.19999 V 0.012% + 5
2.4999 V 0.012% + 5
2.499 V
119.999
399.99
399.9
0.012% + 2
0.1% + 8
0.1% + 5
0.1% + 2
Accuracy
(One year; 23ºC
± 5ºC)
0.05% + 15
[
]
0.05% + 5
0.2% + 5
0.2% + 5
Typical input impedance
—
0.1% + 6
0.1% + 3
0.2% + 3
0.2% + 3
0.1% + 2
0.1% + 2
0.2% + 2
0.2% + 2
—
—
—
—
—
—
—
—
Burden voltage
[
< 0.15 V
< 1.5 V
< 0.3 V
< 0.6 V
< 0.5 V
< 1.5 V
< 0.3 V
< 0.6 V
< 0.5 V
< 1.5 V
< 0.3 V
< 0.6 V
—
—
—
—
—
—
U3402A User’s and Service Guide 91
6 Specifications and Characteristics
Table 6-1 DC resolution, full scale reading, and accuracy [±(% of reading + count)] (continued)
Function
Resistance
Rate
Slow
Med
Fast
Range
[
120.000
W
1.20000 k
W
12.0000 k
W
120.000 k
W
1.20000 M
W
12.0000 M
W
120.000 M
W
400.00
W
4.0000 k
W
40.000 k
W
400.00 k
W
4.0000 M
W
40.000 M
W
300.00 M
W
400.0
W
4.000 k
W
40.00 k
W
400.0 k
W
4.000 M
W
40.00 M
W
300.0 M
W
Resolution
0.001
0.00001 k
0.0001 k
0.001 k
W
W
0.00001 M
0.0001 M
0.001 M
0.01
W
W
0.0001 k
0.001 k
0.01 k
0.001 M
0.01 M
0.1
W
W
W
W
W
0.001 k
W
0.01 k
W
0.1 k
W
W
W
0.001 M
W
0.01 M
W
0.1 M
W
W
0.0001 M
W
W
W
Maximum reading
119.999
1.19999
11.9999
119.999
1.19999
11.9999
119.999
399.99
3.9999
39.999
399.99
3.9999
39.999
299.99
399.9
3.999
39.99
399.9
3.999
39.99
299.9
Current source
0.5 mA
0.5 mA
100 µA
10 µA
1 µA
100 nA
10 nA
0.5 mA
100 µA
50 µA
5 µA
500 nA
50 nA
10 nA
0.5 mA
100 µA
50 µA
5 µA
500 nA
50 nA
10 nA
Accuracy (One year; 23ºC ±
2-wire
0.1% + 8
]
0.08% + 5
]
0.06% + 5
]
0.06% + 5
0.06% + 5
0.3% + 5
3.0% + 8
0.1% + 5
]
0.08% + 3
]
0.06% + 3
0.06% + 3
0.15% + 3
1.5% + 3
5.0% + 5
0.1% + 2
0.08% + 2
0.06% + 2
0.06% + 2
0.15% + 2
1.5% + 2
5.0% + 2
]
5ºC)
4-wire
0.05% + 8
0.05% + 5
0.05% + 5
0.05% + 5
0.05% + 5
0.3% + 5
3.0% + 8
0.05% + 5
0.05% + 3
0.05% + 3
0.05% + 3
0.15% + 3
1.5% + 3
5.0% + 5
0.05% + 2
0.05% + 2
0.05% + 2
0.05% + 2
0.15% + 2
1.5% + 2
5.0% + 2
]
]
]
Notes:
1 Input impedance is in parallel with capacitance < 120 pF.
2 Use Rel operation.
3 In VDC 1000 V range, 5% over-range (1050 VDC) is readable.
4 Typical at full scale reading and voltage across the input terminals.
5 Use Rel operation. If Rel operation is not used, add 0.2
W additional error.
6 In order to eliminate the noise interference, which might be induced to the test leads, it is recommended to use a shielded test cable for measuring resistance above 100 k
W
92 U3402A User’s and Service Guide
Specifications and Characteristics 6
AC Specifications
True RMS AC Voltage (AC Coupling Mode)
Table 6-2 AC voltage resolution, full scale reading, and accuracy [± (% of reading + count)]
Rate Range Resolution
Maximum reading
Slow
Med
120.000 mV
1.20000 V
12.0000 V
120.000 V
750.00 V
400.00 mV
4.0000 V
40.000 V
400.00 V
750.0 V
0.001 mV
0.00001 V
0.0001 V
0.001 V
0.01 V
0.01 mV
0.0001 V
0.001 V
0.01 V
0.1 V
Fast
400.0 mV
4.000 V
40.00 V
400.0 V
0.1 mV
0.001 V
0.01 V
0.1 V
399.9
3.999
39.99
399.9
750 V 1 V 750
Notes:
1 Specified accuracy at input > 5% of full scale.
2 For input < 200 V RMS.
3 For input < 500 V RMS.
4 In VAC 750 V range, 787.5 V RMS is readable.
119.999
1.19999
11.9999
119.999
750.00
[
399.99
3.9999
39.999
399.99
750.0
20 Hz to 45 Hz
Accuracy (One year; 23ºC ± 5ºC)
]
45 Hz to 10kHz
10 kHz to
30 kHz
1% + 100
1% + 100
1% + 100
1% + 100
1% + 100
]
1% + 40
1% + 40
1% + 40
1% + 40
]
1% + 40
1% + 5
1% + 5
1% + 5
]
1% + 5
[
1% + 5
[
0.2% + 100
0.2% + 100
0.2% + 100
0.2% + 100
0.2% + 100
0.2% + 40
0.2% + 40
0.2% + 40
0.2% + 40
0.2% + 40
0.2% + 5
0.2% + 5
0.2% + 5
0.2% + 5
0.2% + 5
1.5% + 300
1% + 100
1% + 100
1% + 100
1% + 100
1.5% + 80
1% + 40
1% + 40
1% + 40
1% + 40
1.5% + 10
1% + 5
1% + 5
1% + 5
1% + 5
30 kHz to
100 kHz
]
5% + 300
3% + 200
3% + 200
3% + 200
3% + 200
]
5% +120
3% + 80
3% + 80
3% + 80
3% + 80
]
5% + 15
3% + 10
3% + 10
3% + 10
3% + 10
]
U3402A User’s and Service Guide 93
6 Specifications and Characteristics
True RMS AC Voltage (AC+DC Coupling Mode)
Table 6-3 AC voltage ac+dc
resolution, full scale reading, and accuracy [± (% of reading + count)]
Rate Range
[
Resolution
Maximum reading
Slow
Med
120.000 mV
1.20000 V
12.0000 V
120.000 V
750.00 V
400.00 mV
4.0000 V
40.000 V
400.00 V
750.0 V
0.001 mV
0.00001 V
0.0001 V
0.001 V
0.01 V
0.01 mV
0.0001 V
0.001 V
0.01 V
0.1 V
119.999
1.19999
11.9999
119.999
750.00
]
399.99
3.9999
39.999
399.99
750.0
Fast
400.0 mV
4.000 V
40.00 V
400.0 V
0.1 mV
0.001 V
0.01 V
0.1 V
399.9
3.999
39.99
399.9
750 V 1 V 750
Notes:
1 VDC and VAC are automatically set at the same range.
2 Specified accuracy at input > 5% of full scale.
3 In VAC 750 V range, 787.5 V RMS is readable.
4 For input < 500 V RMS.
Accuracy (One year; 23ºC ± 5ºC)
[
45 Hz to 10kHz 10 kHz to 30 kHz 30 kHz to 100 kHz
0.2% + 100
0.2% + 100
0.2% + 100
0.2% + 100
0.2% + 100
0.2% + 45
0.2% + 43
0.2% + 43
0.2% + 43
0.2% + 43
0.2% + 7
0.2% + 7
0.2% + 7
0.2% + 7
0.2% + 7
1.5% + 300
1% + 100
1% + 100
1% + 100
1% + 100
1.5% + 83
1% + 43
1% + 43
1% + 43
1% + 43
1.5% + 12
1% + 7
1% + 7
1% + 7
1% + 7
5% + 300
3% + 200
3% + 200
3% + 200
3% + 200
]
5% + 125
3% + 83
3% + 83
3% + 83
3% + 83
]
5% + 18
3% + 12
3% + 12
3% + 12
3% + 12
]
94 U3402A User’s and Service Guide
Specifications and Characteristics 6
True RMS AC Current (AC Coupling Mode)
Table 6-4 AC current resolution, full scale reading, burden voltage, and accuracy [± (% of reading + count)]
Rate Range Resolution
Maximum reading
Burden voltage
[
Slow
12.0000 mA
120.000 mA
1200.00 mA
12.0000 A
0.0001 mA
0.001 mA
0.01 mA
0.0001A
11.9999
119.999
1199.99
11.9999
< 0.15 V
< 1.5 V
< 0.3 V
< 0.6 V
Med
40.000 mA
120.00 mA
1200.0 mA
12.000 A
0.001 mA
0.01 mA
0.1 mA
0.001 A
39.999
119.99
1199.9
11.999
< 0.5 V
< 1.5 V
< 0.3 V
< 0.6 V
Fast
40.00 mA
120.0 mA
1200 mA
12.00 A
0.01 mA
0.1 mA
1 mA
0.01 A
39.99
119.9
1199
11.99
< 0.5 V
< 1.5 V
< 0.3 V
< 0.6 V
Notes:
1 Typical at full scale reading and voltage across the input terminals.
2 Specified accuracy at input > 5% of full scale.
Accuracy (One year; 23ºC ± 5ºC)
[
20 Hz to 45 Hz 45 Hz to 2 kHz
2 kHz to 10 kHz
1.5% + 100
1.5% + 100
1.5% + 100
2% + 100 (<1.2 A)
0.5% + 100
0.5% + 100
0.5% + 100
1% + 100
2% + 200
2% + 200
2% + 200
–
1.5% + 40
1.5% + 12
1.5% + 12
1.5% + 12 (<1.2 A)
1.5% + 5
1.5% + 2
1.5% + 2
2% + 2 (<1.2 A)
0.5% + 40
0.5% + 12
0.5% + 12
1% + 12
0.5% + 5
0.5% + 2
0.5% + 2
1% + 2
2% + 80
2% + 30
2% + 30
–
2% + 10
2% + 5
2% + 5
–
U3402A User’s and Service Guide 95
6 Specifications and Characteristics
True RMS AC Current (AC+DC Coupling Mode)
Table 6-5 AC current ac+dc
resolution, full scale reading, burden voltage, and accuracy [± (% of reading + count)]
Rate Range Resolution
Maximum reading
Burden voltage
Slow
Med
12.0000 mA
120.000 mA
1200.00 mA
12.0000 A
40.000 mA
120.00 mA
1200.0 mA
12.000 A
0.0001 mA
0.001 mA
0.01 mA
0.0001 A
0.001 mA
0.01 mA
0.1 mA
0.001 A
11.9999
119.999
1199.99
11.9999
39.999
119.99
1199.9
11.999
Fast
40.00 mA
120.0 mA
1200 mA
12.00 A
0.01 mA
0.1 mA
1 mA
0.01 A
39.99
119.9
1199
11.99
< 0.5 V
< 1.5 V
< 0.3 V
< 0.6 V
Notes:
1 Typical at full scale reading and voltage across the input terminals.
2 Specified accuracy at input > 5% of full scale.
< 0.15 V
< 1.5 V
< 0.3 V
< 0.6 V
< 0.5 V
< 1.5 V
< 0.3 V
< 0.6 V
[
Accuracy (one year; 23ºC ± 5ºC)
[
45 Hz to 2 kHz 2 kHz to 10 kHz
0.5% + 100
0.5% + 100
0.5% + 100
1% + 100
0.5% + 42
0.5% + 15
0.5% + 15
1% + 15
0.5% + 7
0.5% + 4
0.5% + 4
1% + 4
2% + 200
2% + 200
2% + 200
–
2% + 80
2% + 30
2% + 30
–
2% + 12
2% + 7
2% + 7
–
Frequency
Table 6-6 Frequency resolution and accuracy [± (% of reading + count)]
1200 Hz
12 kHz
120 kHz
5 Hz to 1200 Hz
10 Hz to 12 kHz
100 Hz to 120 kHz
0.01 Hz
0.0001 kHz
0.001 kHz
1 MHz 1 kHz to 1 MHz 0.0001 MHz
Notes:
1 Specified accuracy at input > 5% of full scale.
Maximum reading
1199.99
11.9999
119.999
1.1999
Accuracy (One year;
23ºC ± 5ºC)
[
0.005% + 3
0.005% + 2
0.005% + 2
0.005% + 2
Input sensitivity
(Sine wave)
40 mV RMS
40 mV RMS
40 mV RMS
0.5 V RMS
96 U3402A User’s and Service Guide
Specifications and Characteristics 6
Decibel (dB) Calculation
Table 6-7 Range and accuracy (±dB)
Rate
Voltage range
[
Input voltage dBm
[
range at
600
W ref
Slow
Medium
120.000 mV
1.20000 V
12.0000 V
120.000 V
1000.00 VDC
750.00 VAC
400.00 mV
4.0000 V
40.000 V
400.00 V
1000.0 VDC
750.0 VAC
6 mV to 120 mV
120 mV to 1.2 V
1.2 V to 12 V
12 V to 120 V
120 V to 1000 V
120 V to 750 V
20 mV to 400 mV
400 mV to 4 V
4 V to 40 V
40 V to 400 V
400 V to 1000 V
400 V to 750 V
– 42.20 to – 16.20
– 16.20 to 3.80
3.80 to 23.80
23.80 to 43.80
43.80 to 62.22
43.80 to 59.72
– 31.76 to – 5.74
– 5.74 to 14.26
14.26 to 34.26
34.26 to 54.26
54.26 to 62.22
54.26 to 59.72
Fast
400.0 mV
4.000 V
40.00 V
400.0 V
20 mV to 400 mV
400 mV to 4 V
4 V to 40 V
40 V to 400 V
– 31.76 to – 5.74
– 5.74 to 14.26
14.26 to 34.26
34.26 to 54.26
1000 VDC
750 VAC
400 V to 1000 V
400 V to 750 V
54.26 to 62.22
54.26 to 59.72
Notes:
1 Autoranging is used when dBm operation is enabled.
2 In VAC 750 V range, 5% over-range is readable.
3 Reading displayed in dB when Rel operation is used.
4 For input voltage at frequency between 45 Hz to 1 kHz.
–
–
1.0
0.8
0.8
0.8
–
–
1.0
0.8
0.8
0.8
Accuracy (One year; 23ºC ± 5ºC)
]
20 Hz to 45 Hz 45 Hz to 10 kHz 10 kHz to 100 kHz
–
–
1.0
0.8
0.8
0.8
0.2
0.1
0.1
0.1
1.0
]
1.0
]
1.0
0.8
0.8
0.8
–
–
0.2
0.1
0.1
0.1
1.0
]
1.0
]
0.2
0.1
0.1
0.1
1.0
]
1.0
]
1.0
0.8
0.8
0.8
–
–
1.0
0.8
0.8
0.8
–
–
U3402A User’s and Service Guide 97
6 Specifications and Characteristics
Supplemental Measurement Specifications
Display update rate
Table 6-8 Full scale display update rates
Display update rates Display counts
Slow
Medium
Fast
119,999
39,999
3,999
Measurement Specifications
Table 6-9 Supplemental measurement specifications
DC voltage
• Measurement method
•
Sigma Delta A-to-D converter
• Input resistance
•
10 M
W ± 2% range (typical)
• Maximum input voltage
•
1000 VDC or PEAK AC on all ranges
• Input protection
•
1000 V on all ranges
• Response time
•
Approximately 1.0 second when the displayed reading reaches 99.9% DC value of the tested input signal at the same range.
98 U3402A User’s and Service Guide
Specifications and Characteristics 6
Table 6-9 Supplemental measurement specifications (continued)
DC current
• Shunt resistance
•
0.1
W to 10 W for 12 mA to 1.2 A ranges
•
0.01
W for 12 A range
• Maximum input and overload protection
•
mA input terminal: 1200 mADC or AC RMS. Protected with 1.25 A/500 V, IEC-127 sheet, FB fuse
•
12 A input terminal: 10 ADC or AC RMS continuous, or 12 ADC or AC RMS for 30 seconds maximum. Protected with 15 A/600 V, breaking capacity 10,000 A FB fuse
• Response time
•
Approximately 1.0 seconds when the displayed reading reaches 99.9% DC value of the tested input signal at the same range
AC voltage (True RMS AC Coupling Mode)
• Measurement method
•
AC coupled true RMS—measure the AC component with up to 400 VDC bias on any range
• Crest factor
•
Maximum 3:0 at full scale
• Input impedance
•
1 M
W ± 2% in parallel with <120 pF on all ranges
• Maximum input voltage
•
750 V RMS /1100 V PEAK
•
2x10
7
V-Hz product on any range, normal mode input
•
1x10
6
V-Hz product on any range, common mode input
• Overload ranging
•
Will select higher range if peak input overload is detected during autorange. Overload is reported in manual ranging
• Input protection
•
750 V RMS on all ranges
U3402A User’s and Service Guide 99
6 Specifications and Characteristics
Table 6-9 Supplemental measurement specifications (continued)
• Response time
•
Approximately 1.5 seconds when the displayed reading reaches 99.9% AC RMS value of the tested input signal at the same range.
AC voltage (True RMS AC + DC Coupling Mode)
• Measurement method
•
AC+DC coupled true RMS—measure the AC component with up to 400 VDC bias on any range
• Crest factor
•
Maximum 3:0 at full scale
• Input impedance
•
1 M
W ± 2% in parallel with <120 pF of all ranges
• Maximum input voltage
•
750 V RMS /1100 V PEAK
•
2x10
7
V-Hz product on any range, normal mode input
•
1x10
6
V-Hz product on any range, common mode input
• Overload ranging
•
Will select higher range if peak input overload is detected during autorange. Overload is reported in manual ranging
• Input protection
•
750 V RMS on all ranges
• Response time
•
Approximately 2.5 seconds when the displayed reading reaches 99.9% AC RMS value of the tested input signal at the same range
100 U3402A User’s and Service Guide
Specifications and Characteristics 6
Table 6-9 Supplemental measurement specifications (continued)
AC current (True RMS, AC Coupling Mode)
• Measurement method
•
DC coupled to the fuse and current shunt, AC coupled true RMS measurement (measures the
AC component only)
• Crest factor
•
Maximum 3:0 at full scale
• Shunt resistance
•
0.1
W to 10 W for 10 mA to 1.2 A ranges
•
0.01
W for 12 A range
• Input protection
•
mA input terminal: 1200 mADC or AC RMS. Protected with 1.25 A/500 V, IEC-127 sheet, FH fuse
•
12 A input terminal: 10 ADC or AC RMS continuous, or 12 ADC or AC RMS for 30 seconds maximum. Protected with 15 A/600 V, breaking capacity 10,000 A FH fuse.
• Response time
•
Approximately 1.5 seconds when the displayed reading reaches 99.9% AC RMS value of the tested input signal at the same range
AC current (True RMS, AC Coupling Mode)
• Measurement method
•
AC+DC coupled to the fuse and current shunt, AC+DC coupled true RMS measurement
(measures the AC component only)
• Crest factor
•
Maximum 3:0 at full scale
• Measurement range
•
VDC and VAC are automatically set at the same range.
• Response time
•
Approximately 1.5 seconds when the displayed reading reaches 99.9% AC RMS value of the tested input signal at the same range
U3402A User’s and Service Guide 101
6 Specifications and Characteristics
Table 6-9 Supplemental measurement specifications (continued)
Resistance (2-wire
W and 4-wire W)
• Measurement method
•
2-wire Ohms or 4-wire Ohms
• Open-circuit voltage
•
Limited to < +5 VDC
• Zeroing error
•
0.05
W or less (excluding test lead resistance) in each range when Rel operation is used
• Input protection
•
500 V on all ranges
• Response time
•
Approximately 1.5 seconds for 12 M
W and ranges below 12 MW
•
Approximately 5 seconds for 40 M
W
•
Approximately 10 seconds for 120 M
W
•
Approximately 23 seconds for 300 M
W
Diode/continuity
• Measurement method
•
0.83 mA ±0.2% constant current source, open-circuit voltage limited to < 5 V
• Test current
•
Approximately 0.5 mADC
• Open-circuit voltage
•
Limited to < +5 VDC
• Continuity threshold
•
10
W fixed
• Continuity level
•
Approximately < +50 mVDC
102 U3402A User’s and Service Guide
Specifications and Characteristics 6
Table 6-9 Supplemental measurement specifications (continued)
• Audible tone
•
Continuous beep for continuity and single tone for normal forward-biased diode or semiconductor junction
• Input protection
•
500 V RMS on all ranges
Resistance/continuity (2-wire
W)
• Measurement method
•
2-wire Ohms
• Test current
•
Approximately 0.5 mADC
• Open-circuit voltage
•
Limited to < +5 VDC
• Audible tone
•
Continuous beep for continuity and single tone for normal forward-biased diode or semiconductor junction
• Zeroing error
•
0.05
W or less (excluding test lead resistances) in each range when Rel operation is used
• Input protection
•
500 V RMS on all ranges
Frequency
• Measurement method
•
Reciprocal counting technique. AC coupled input using AC voltage function
• Crest factor
•
Maximum 3:0 at full scale
• Signal level
•
10% of range to full scale input on all ranges
•
Auto or manual range selection
U3402A User’s and Service Guide 103
6 Specifications and Characteristics
Table 6-9 Supplemental measurement specifications (continued)
• Gate time
•
0.1 second or 1 period of the input signal, whichever is longer
• Input impedance
•
1 M
W ± 2% in parallel with <120 pF of all ranges
• Maximum input voltage
•
750 V RMS /1100 V PEAK
•
2x10
7
V-Hz product on any range, normal mode input
•
1x10
6
V-Hz product on any range, common mode input
• Input protection
•
750 V RMS on all ranges
• Response time
•
Approximately 1.5 seconds when the displayed reading reaches 99.9% of frequency value
Measurement Noise Rejection
• Common mode reject ratio (CMRR) for 1 k
W unbalanced LO lead
•
50/60 Hz ± 0.1%: DC > 90 dB
• Normal mode rejection ratio (NMRR)
•
50/60 Hz ± 0.1%: > 50 dB
dBm Operation
• 0 dBm
•
1 mW at 600
W reference impedance
• Resolution
•
Slow: 0.01 dB for all ranges
•
Medium: 0.01 dB for all ranges
•
Fast: 0.1 dB for all ranges
• Reference impedance
[1]
•
2
W
[2]
, 4
W
[2]
, 8
W
[2]
, 16
W
[2]
, 50
W, 75 W, 93 W, 110 W, 124 W, 125 W, 135 W, 150 W, 250 W, 300
W, 500 W, 600 W, 800 W, 900 W, 1000 W, 1200 W, or 8000 W
104 U3402A User’s and Service Guide
Specifications and Characteristics 6
Table 6-9 Supplemental measurement specifications (continued)
Math Operation
• dBm, Rel, MinMax, Comp, Hold
I/O Interface
• RS232
[3]
Notes:
1 Reference impedance is displayed on the secondary display.
2 Reading displayed in watts (Audio power).
3 For calibration use only.
N O T E
When V ac+dc
measurement function is selected, the VDC input impedance is parallel with an AC-couples 1.1 M
W divider.
Reading rates
Table 6-10 Reading rates for single display (readings/second (approx))
Measurement functions
DCV
DCA
Diode
ACV
ACA
2-wire
W
4-wire
W, 4MW/1.2MW range and below
4-wire
W, 12MW range and above
Frequency (5Hz)
ACV + DCV
ACA + DCA
Slow
2.2
2.2
2.2
2.2
2.2
2.2
0.7
1.5
1.0
0.8
0.5
Medium
5.5
5.5
5.5
5.5
5.5
5.5
0.9
1.8
1.7
1.1
0.7
Fast
22.0
22.0
22.0
22.0
22.0
22.0
1.0
2.0
2.3
1.6
0.9
U3402A User’s and Service Guide 105
6 Specifications and Characteristics
Table 6-11 Reading rates for the dual display (readings/second (approx))
Measurement functions Slow Medium Fast
DCV / ACV
0.8
1.1
1.6
DCA / ACA
0.5
0.7
0.9
DCV / DCA
0.7
1.1
1.6
DCV / ACA
0.1
0.4
0.5
ACV / ACA
0.1
0.4
0.5
ACV / DCA
0.8
1.1
1.6
ACV / Frequency
[
1.0
1.7
2.3
ACA / Frequency
[
]
1.2
1.7
2.5
ACV + DCV / DCV
0.7
1.1
1.6
ACA + DCA / DCV
0.1
0.3
0.4
ACA + DCA / ACV
0.1
0.3
0.4
ACA + DCA / DCA
0.5
0.7
0.9
dBm (ACV) / Reference
2.2
5.5
22.0
dBm (ACV) / ACV
2.2
4.3
11.0
dBm (ACV) / DCV
0.8
1.1
1.8
dBm (ACV) / Frequency
[
1.0
1.7
2.4
Notes:
1 This reading rate depends on the frequency of the signal, the table itself shows the worst possible case.
2 This table shows the typical combinations of measurement functions for the dual display.
106 U3402A User’s and Service Guide
Specifications and Characteristics 6
General Characteristics
Table 6-12 General characteristics
Power Supply
•
100 V/120 V/220 V/240 V ± 10%
•
AC line frequency 50 Hz to 60Hz
Power Consumption
•
16 VA maximum
Input Power Option
•
Manual ranging (100 VAC to 240 VAC ±10%)
Fuse
• Terminal:
•
1.25 A, 500 V FB fuse
•
15 A, 600 V FB fuse (internal)
• Power line (according to country of destination):
•
0.25 A, 250 V SB fuse OR
•
0125 A, 250 V SB fuse
Display
•
Highly visible vacuum fluorescent display (VFD)
Operating Environment
•
Operating temperature from 0 °C to + 50 °C
•
Relative humidity up to 80% at 28 °C RH (non-condensing)
•
Altitude up to 2000 meters
•
Pollution degree 2
•
For indoor use only
U3402A User’s and Service Guide 107
6 Specifications and Characteristics
Table 6-12 General characteristics
Storage Compliance
•
– 20 °C to 60 °C
•
Relative humidity at 5% to 90% RH (non-condensing)
Safety Compliance
•
IEC 61010-1:2001/EN61010-1:2001 (2nd Edition)
•
Canada: CAN/CSA-C22.2 No. 61010-1-04
•
USA: ANSI/UL 61010-1:2004
EMC Compliance
•
IEC 61326-1:2005/EN61326-1:2006
•
Canada: ICES/NMB-001:2004
•
Australia/New Zealand: AS/NZS CISPR11:2004
Shock and Vibration
•
Tested to IEC/EN 60068-2
I/O Connector
•
Output connectors
I/O Interface
•
RS232 (For calibration use only)
Dimension (HxWxD)
•
226.00 × 105.00 × 305.00 mm (with bumpers)
•
215.00 × 87.00 × 282.00 mm (without bumpers)
Weight
•
3.44 kg (with bumpers)
Warm Up Time
•
30 minutes
Calibration Cycle
•
1 year
Warranty
•
3 years
•
3 months for standard shipped accessories
108 U3402A User’s and Service Guide
Specifications and Characteristics 6
To Calculate Total Measurement Error
The multimeter's accuracy specifications are expressed in the form of:
(% of reading + count)
In addition to the reading error and count error, you may need to add additional errors for certain operating conditions. Check the list below to make sure you include all measurement errors for a given function. Also, make sure you apply the conditions as described in the footnotes on the specification pages.
• If you are operating the multimeter outside the temperature range specified, apply an additional temperature coefficient error.
• For AC voltage and AC current measurements, you may need to apply an additional low frequency error or crest factor error.
U3402A User’s and Service Guide 109
6 Specifications and Characteristics
Accuracy Specifications
Transfer Accuracy
Transfer accuracy refers to the error introduced by the multimeter due to noise and short – term drift. This error becomes apparent when comparing two nearly equal signals for the purpose of "transferring" the known accuracy of one device to the other.
One-Year Accuracy
These long
– term accuracy specifications are valid at the calibration temperature (T cal
) ± 5 °C temperature range. These specifications include the initial calibration errors plus the multimeter's long
– term drift errors.
Temperature Coefficients
Accuracy is usually specified at the calibration temperature (T cal
) ± 5 °C temperature range. This is a common temperature range for many operating environments. You must add additional temperature coefficient errors to the accuracy specification if you are operating the multimeter at 0 °C to 18 °C and
28 °C to 50 °C temperature range (the specification is per °C).
Temperature Coefficient = add ± 0.15 x [the applicable accuracy)/°C]
110 U3402A User’s and Service Guide
Specifications and Characteristics 6
Configuring for Highest Accuracy Measurements
The measurement configurations shown below assume that the multimeter is in its power – on or reset state. It is also assumed that auto–ranging is enabled to ensure proper full scale range selection.
• Select 5½ digits.
• Null the test lead resistance for 2–wire ohms measurements, and to remove any interconnection offset for DC voltage measurements.
U3402A User’s and Service Guide 111
6 Specifications and Characteristics
112 U3402A User’s and Service Guide
www.agilent.com
Contact us
T o obtain service, warranty or technical support assistance, contact us at the following phone numbers:
United States:
(tel) 800 829 4444 (fax) 800 829 4433
Canada:
(tel) 877 894 4414 (fax) 800 746 4866
China:
(tel) 800 810 0189 (fax) 800 820 2816
Europe:
(tel) 31 20 547 2111
Japan:
(tel) (81) 426 56 7832 (fax) (81) 426 56 7840
Korea:
(tel) (080) 769 0800 (fax) (080) 769 0900
Latin America:
(tel) (305) 269 7500
Taiwan:
(tel) 0800 047 866 (fax) 0800 286 331
Other Asia Pacific Countries:
(tel) (65) 6375 8100 (fax) (65) 6755 0042
Or visit Agilent worlwide Web at: www.agilent.com/find/assist
Product specifications and descriptions in this document are subject to change without notice.
Always refer to the Agilent Web site for the latest revision.
©
Agilent Technologies, Inc. 2009–2013
Printed in Malaysia
Fifth Edition, May 21, 2013
U3402-90001
Agilent Technologies
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
advertisement
Table of contents
- 3 Safety Symbols
- 4 Regulatory Markings
- 5 General Safety Information
- 7 Environmental Conditions
- 8 Waste Electrical and Electronic Equipment (WEEE) Directive 2002/96/EC
- 9 Declaration of Conformity (DoC)
- 2 Notices
- 10 In This Guide ...
- 11 Contents
- 15 List of Tables
- 17 List of Figures
- 19 Getting Started Tutorial
- 20 Introducing the Agilent U3402A Dual Display Multimeter
- 21 Checking the shipping contents
- 22 Connecting power to the multimeter
- 23 Stacking the U3402A
- 24 Adjusting the Handle
- 25 Product at a Glance
- 25 Product dimensions
- 26 The front panel at a glance
- 27 The display at a glance
- 29 The keypad at a glance
- 32 The terminals at a glance
- 34 The rear panel at a glance
- 35 Making Measurements
- 36 Performing voltage measurements
- 38 Performing current measurements
- 40 Performing frequency measurements
- 41 Performing resistance measurements
- 43 Performing diode/continuity test
- 48 Selecting a Range
- 50 Setting the Reading Rate
- 51 Operations and Features
- 52 Operating Math Operations
- 53 dBm
- 54 Rel
- 55 MinMax
- 57 Comp
- 58 Hold
- 59 Combination of Math Operations
- 61 Using the Secondary Display
- 63 Using the Setup Menu
- 64 Changing the Configurable Settings
- 65 Application Tutorial
- 66 Applications for Using Dual Display
- 67 Dual Display Operation Examples
- 67 Measure DC Voltage and AC Ripple on a Rectification Circuit
- 68 Measure AC and DC Current on a Rectification Circuit
- 69 Measure AC Voltage and Frequency on an AC Circuit
- 70 Measure DC Voltage and DC Current on a Transistor Circuit or Load
- 72 Measure Resistance Using 2-Wire Mode
- 73 Measure Resistance Using 4-Wire Mode
- 74 Measure True RMS AC+DC
- 75 Performance Test
- 76 Calibration Overview
- 76 Agilent Technologies Calibration Services
- 76 Calibration Interval
- 77 Recommended Test Equipment
- 78 Test Considerations
- 79 Performance Verification Test Overview
- 79 Performance Verification Test
- 89 Disassembly and Repair
- 90 Operating Checklist
- 91 Types of Service Available
- 92 Repackaging for Shipment
- 92 Cleaning
- 93 To Replace the Power Line Fuse
- 94 To Replace a Current Input Fuse
- 94 Electrostatic Discharge (ESD) Precautions
- 95 Mechanical Disassembly
- 101 Replaceable Parts
- 101 To order replaceable parts
- 103 Specifications and Characteristics
- 104 DC Specifications
- 107 AC Specifications
- 111 Decibel (dB) Calculation
- 112 Supplemental Measurement Specifications
- 112 Display update rate
- 112 Measurement Specifications
- 119 Reading rates
- 121 General Characteristics
- 123 To Calculate Total Measurement Error
- 124 Accuracy Specifications
- 124 Transfer Accuracy
- 124 One-Year Accuracy
- 124 Temperature Coefficients
- 125 Configuring for Highest Accuracy Measurements