Agilent Infiniium 90000 Q-Series Oscilloscopes User`s Guide

Agilent Infiniium 90000 Q-Series Oscilloscopes User`s Guide
Keysight Infiniium S-Series
Oscilloscopes
User’s Guide
Notices
© Keysight Technologies 2014-2015
Technology Licenses
No part of this manual may be reproduced in
any form or by any means (including electronic storage and retrieval or translation
into a foreign language) without prior agreement and written consent from Keysight
Technologies as governed by United States
and international copyright laws.
The hardware and/or software described in
this document are furnished under a license
and may be used or copied only in accordance with the terms of such license.
Manual Part Number
54911-97004
Ed ition
Second Edition, April 2015
Available in electronic format only
Published by:
Keysight Technologies
1900 Garden of the Gods Rd.
Colorado Springs, CO 80907 USA
Warranty
The material contained in this document
is provided “as is,” and is subject to
being changed, without notice, in future
editions. Further, to the maximum
extent permitted by applicable law,
Keysight disclaims all warranties, either
express or implied, with regard to this
manual and any information contained
herein, including but not limited to the
implied warranties of merchantability
and fitness for a particular purpose.
Keysight shall not be liable for errors or
for incidental or consequential damages
in connection with the furnishing, use,
or performance of this document or of
any information contained herein.
Should Keysight and the user have a
separate written agreement with warranty terms covering the material in this
document that conflict with these
terms, the warranty terms in the separate agreement will control.
U.S. Government Rights
The Software is “commercial computer
software,” as defined by Federal Acquisition
Regulation (“FAR”) 2.101. Pursuant to FAR
12.212 and 27.405-3 and Department of
Defense FAR Supplement (“DFARS”)
227.7202, the U.S. government acquires
commercial computer software under the
same terms by which the software is
customarily provided to the public.
Accordingly, Keysight provides the Software
to U.S. government customers under its
standard commercial license, which is
embodied in its End User License Agreement
(EULA), a copy of which can be found at
www.keysight.com/find/sweula. The
license set forth in the EULA represents the
exclusive authority by which the U.S.
government may use, modify, distribute, or
disclose the Software. The EULA and the
license set forth therein, does not require or
permit, among other things, that Keysight:
(1) Furnish technical information related to
commercial computer software or
commercial computer software
documentation that is not customarily
provided to the public; or (2) Relinquish to,
or otherwise provide, the government rights
in excess of these rights customarily
provided to the public to use, modify,
reproduce, release, perform, display, or
disclose commercial computer software or
commercial computer software
documentation. No additional government
requirements beyond those set forth in the
EULA shall apply, except to the extent that
those terms, rights, or licenses are explicitly
required from all providers of commercial
computer software pursuant to the FAR and
the DFARS and are set forth specifically in
writing elsewhere in the EULA. Keysight
shall be under no obligation to update,
revise or otherwise modify the Software.
With respect to any technical data as
defined by FAR 2.101, pursuant to FAR
12.211 and 27.404.2 and DFARS 227.7102,
the U.S. government acquires no greater
than Limited Rights as defined in FAR 27.401
or DFAR 227.7103-5 (c), as applicable in any
technical data.
Safety Notices
C AU TI O N
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like
that, if not correctly performed or
adhered to, could result in damage
to the product or loss of important
data. Do not proceed beyond a
CAUTION notice until the indicated
conditions are fully understood and
met.
W A RNI NG
A WARNING notice denotes a
hazard. It calls attention to an
operating procedure, practice, or
the like that, if not correctly performed or adhered to, could result
in personal injury or death. Do not
proceed beyond a WARNING
notice until the indicated conditions are fully understood and met.
For more safety information, refer to the For
Your Safety booklet included with your
Infiniium oscilloscope.
Infiniium S-Series Oscilloscopes—At a Glance
Table 1
S-Series Oscilloscope Band wid ths
Model
Analog
Band wid th
Maximum
Sampling Rate
(2-channel)
Standard Memory Depth
(2/4-channel mode)
DSOS054A/MSOS054A
500 MHz
20 GSa/s
100 Mpts/50 Mpts
DSOS104A/MSOS104A
1 GHz
20 GSa/s
100 Mpts/50 Mpts
DSOS204A/MSOS204A
2 GHz
20 GSa/s
100 Mpts/50 Mpts
DSOS254A/MSOS254A
2.5 GHz
20 GSa/s
100 Mpts/50 Mpts
DSOS404A/MSOS404A
4 GHz
20 GSa/s
100 Mpts/50 Mpts
DSOS604A/MSOS604A
6 GHz
20 GSa/s
100 Mpts/50 Mpts
DSOS804A/MSOS804A
8 GHz
20 GSa/s
100 Mpts/50 Mpts
Keysight Infiniium S-Series Oscilloscopes User’s Guide
3
Ease of use with high performance
The Infiniium S-Series oscilloscopes combine unprecedented ease of use with
high-performance digitizing oscilloscope functionality to simplify your design and
analysis measurement tasks.
•
Traditional oscilloscope front-panel interface provides direct access to the
controls needed for most troubleshooting tasks.
•
User interface with menus, windows, dialog boxes, and toolbars provides easy
access to dozens of configuration and analysis tools, ensuring you can set up
and make the most complex measurements.
•
Models with bandwidths from 500 MHz to 8 GHz.
•
16 digital channels on MSO models at 2 GS/s.
Display shows waveforms and user interface
•
User interface allows direct interaction with waveforms, including
drag-and-drop positioning and instant waveform zoom.
•
Large capacitive touch screen display with multi-touch (gestures), handles, and
resizing allows oscilloscope operation without an external pointing device.
•
Waveforms are displayed in color, making correlation easy.
•
Current configuration parameters displayed near the waveform display area are
color-coded to make identification easy.
•
Menus and toolbars simplify complex measurement setups.
Horizontal controls set sweep speed and position
•
Zoom box on main sweep window makes it easy to see what will appear in the
zoom window.
Acquisition and general controls start and stop the oscilloscope and do
basic setup
•
Run, stop, and single controls for continuous or single acquisitions.
•
Clear display before one or more acquisitions.
•
Default setup and Autoscale set initial configuration.
Removable solid-state drive and USB 2.0 and 3.0 ports for saving and
restoring setups and measurement results
•
Store measurement displays for inclusion in reports and test setup guides.
•
Store oscilloscope setups to repeat tests another time.
•
250 Gb removable solid state drive for fast boot-up.
Trigger setup controls set mode and basic parameters
4
•
Select Edge, Glitch, or Advanced Modes.
•
Choose input source and slope.
Keysight Infiniium S-Series Oscilloscopes User’s Guide
•
Use the user interface to simplify configuration of pattern, state, delay, and
violation trigger modes.
•
Use auxiliary trigger to increase triggering flexibility.
Vertical controls set attenuation and position
•
Vertical scaling down to 2 mV/div in hardware.
•
Color-coded knobs make it easy to find the controls that affect each waveform.
Marker and quick measurements help measure waveform parameters
•
Use waveform markers 1 and 2 to check voltage or Δ-time at any point on the
displayed waveform.
Keysight Infiniium S-Series Oscilloscopes User’s Guide
5
In This Guide
This guide provides the information you need to begin using the Infiniium S-Series
oscilloscopes.
Chapter 1, “Setting Up the Oscilloscope,” starting on page 11 includes power and
air flow requirements, plus other setup information.
Chapter 2, “Using the Oscilloscope,” starting on page 25 gives an overview of the
front and side panel inputs and outputs, front-panel controls, and user interface,
and tells you how to perform basic operations with the oscilloscope.
Chapter 3, “Online Help and Other Information,” starting on page 49 describes the
Infiniium oscilloscope application’s online help contents and online demos. The
online help describes how to use the Infiniium oscilloscope application in detail.
Chapter 4 provides recommendations for working comfortably and safely while
operating the Infiniium oscilloscope.
For More Information
6
•
For detailed information on how the oscilloscope makes measurements and
how to use the oscilloscope, see the Infiniium oscilloscope application’s online
help.
•
For information on controlling the oscilloscope from a remote computer, see
the Oscilloscopes Programmer’s Reference found in the Infiniium oscilloscope
application’s online help.
•
For information on testing and servicing the oscilloscope, see the Service Guide
found in the Infiniium oscilloscope application’s online help.
Keysight Infiniium S-Series Oscilloscopes User’s Guide
Contents
Infiniium S-Series Oscilloscopes—At a Glance 3
In This Guide 6
1 Setting Up the Oscilloscope
Inspecting Package Contents / 12
Environmental Characteristics / 13
Positioning for Proper Airflow / 14
Connecting a Mouse, Keyboard, and LAN Cable / 15
Connecting Power / 16
Connecting Oscilloscope Probes / 17
Tilting the Oscilloscope for Easier Viewing / 18
Turning On the Oscilloscope / 19
Verifying Basic Oscilloscope Operation / 20
Installing Application Programs on Infiniium / 21
Changing Windows Operating System Settings / 22
Turning Off the Oscilloscope / 23
Cleaning the Oscilloscope / 23
2 Using the Oscilloscope
Front Panel Inputs and Outputs / 26
Channel inputs / 26
Digital channels connector / 26
Probe compensation terminal / 26
Ground / 26
Side Panel Inputs and Outputs / 27
Motherboard I/O / 27
Aux Out / 28
10 MHz In, 10 MHz Out / 28
Trig Out / 28
Aux Trig / 28
Front Panel Controls (Keys and Knobs) Overview / 29
Keysight Infiniium S-Series Oscilloscopes User’s Guide
7
User Interface Overview / 30
Menu overview / 32
Using the Setup and Display Controls / 33
Starting and Stopping Waveform Acquisitions / 34
Adjusting the Horizontal Time Scale and Trigger Position / 35
Adjusting the horizontal scale 35
Adjusting the horizontal trigger position (delay) 36
Magnifying a part of the waveform using Zoom 36
Setting the scale, position, and timebase reference 36
Adjusting the Vertical Settings / 38
Turning an analog channel on or off 38
Adjusting the analog channel’s vertical scale and offset 39
Setting Up Triggers / 40
Setting the oscilloscope to trigger on an edge 40
Using Markers and Making a Measurement / 42
Making a measurement on a waveform 43
Using quick measurements 43
Using markers 44
Moving markers using the user interface 44
Controlling digital channels 45
Turning digital channels on or off 45
Decoding serial data 45
Saving and Printing Data / 46
Forcing a Default Setup / 47
Infiniium hard drive recovery 47
3 Online Help and Other Information
Accessing the Online Help / 49
Navigating the Online Help / 51
Using the Demo Wizard / 51
4 Working in Comfort
About Repetitive Strain Injury / 54
What is RSI? / 54
What causes RSI? / 54
What if I experience discomfort? / 54
Mice and Other Input Devices / 55
8
Keysight Infiniium S-Series Oscilloscopes User’s Guide
Index
Keysight Infiniium S-Series Oscilloscopes User’s Guide
9
10
Keysight Infiniium S-Series Oscilloscopes User’s Guide
Keysight Infiniium S-Series Oscilloscopes
User’s Guide
1 Setting Up the Oscilloscope
Inspecting Package Contents 12
Environmental Characteristics 13
Positioning for Proper Airflow 14
Connecting a Mouse, Keyboard, and LAN Cable 15
Connecting Power 16
Connecting Oscilloscope Probes 17
Tilting the Oscilloscope for Easier Viewing 18
Turning On the Oscilloscope 19
Verifying Basic Oscilloscope Operation 20
Installing Application Programs on Infiniium 21
Changing Windows Operating System Settings 22
Turning Off the Oscilloscope 23
Cleaning the Oscilloscope 23
This chapter shows how to set up your Infiniium oscilloscope, connect power and
accessories, and verify general operation.
11
1
Setting Up the Oscilloscope
Inspecting Package Contents
✔ Inspect the shipping container for damage.
•
Keep the shipping container or cushioning material until you have inspected
the contents of the shipment for completeness and have checked the
oscilloscope mechanically and electrically.
•
If the shipping container is damaged, or the cushioning materials show signs
of stress, notify the carrier and your Keysight Technologies Sales Office.
Keep the shipping materials for the carrier’s inspection. The Keysight
Technologies Sales Office will arrange for repair or replacement at
Keysight’s option without waiting for claim settlement.
✔ Inspect the oscilloscope.
If there is mechanical damage or a defect, or if the oscilloscope does not
operate properly or does not pass performance tests, notify your Keysight
Technologies Sales Office.
✔ Verify that you received the following items in the Infiniium oscilloscope
packaging.
•
Infiniium oscilloscope
•
Power cord
•
Keyboard
•
Mouse (USB optical)
•
Accessory pouch (mounts on rear of oscilloscope)
•
Front panel cover
•
Calibration cable
•
Quick Start poster
•
500 MHz passive probes (4)
•
Digital channels cable, BNC probe tip adapter, and 17-channel flying lead
kit (MSO models only)
If anything is missing, contact your nearest Keysight Technologies Sales Office.
✔ Verify that you received the options and accessories you ordered and that none
were damaged.
For a complete list of options and accessories available for the S-Series
oscilloscopes, see the Infiniium S-Series Oscilloscopes Data Sheet.
12
Keysight Infiniium S-Series Oscilloscopes User’s Guide
Setting Up the Oscilloscope
1
Environmental Characteristics
Environment
Indoor use only
Ambient temperature
Operating: 5 °C to +40 °C
Non-operating: –40 °C to +65 °C
Humidity
Operating: up to 90% relative humidity (non-condensing) at +40 °C
Non-operating: up to 90% relative humidity at +65 °C
Altitude
Operating: up to 4,000 meters (13,123 feet)
Non-operating: up to 15,300 meters (50,000 feet)
Weight
26 lbs (11.8 kg)
Dimensions
16.8 in (43 cm) wide, 12.9 in (33 cm) tall, and 9 in (23 cm) deep
Safety
UL61010-1 3rd edition
CAN/CSA-22.2 No. 61010-1-12 3rd edition
Installation/
Measurement
Category
Installation category: II
Measurement category: 1
Voltage Fluctuations
Note that the main supply voltage fluctuations are not to exceed ±10% of the
nominal supply voltage.
Pollution Degree
The Infiniium S-Series oscilloscopes may be operated in environments of
Pollution Degree 2.
Pollution Degree
Definitions
Pollution Degree 1: No pollution or only dry, non-conductive pollution occurs.
The pollution has no influence. Example: A clean room or climate-controlled
office environment.
Pollution Degree 2. Normally only dry non-conductive pollution occurs.
Occasionally a temporary conductivity caused by condensation may occur.
Example: General indoor environment.
Pollution Degree 3: Conductive pollution occurs, or dry, non-conductive
pollution occurs which becomes conductive due to condensation which is
expected. Example: Sheltered outdoor environment.
Keysight Infiniium S-Series Oscilloscopes User’s Guide
13
1
Setting Up the Oscilloscope
Positioning for Proper Airflow
1 Position the oscilloscope where it will have sufficient clearance for airflow
around the back and sides.
Minimum 0 mm
Front panel of oscilloscope
Minimum bottom clearance: No intrusion into the
space under the oscilloscope as defined by the
feet. Feet must rest on hard surface.
Minimum 0 mm
Rear Panel
Minimum 75 mm
Minimum 25.4 mm
Minimum 25.4 mm
Top View
Figure 1
14
Positioning the S-Series Oscilloscope with Sufficient Clearance
Keysight Infiniium S-Series Oscilloscopes User’s Guide
Setting Up the Oscilloscope
1
Connecting a Mouse, Keyboard, and LAN Cable
A mouse and keyboard can be plugged into the USB host ports. Four host ports
are on the side panel, with two more on the front panel. Connect your LAN cable to
the RJ-45 connector on the side panel of the oscilloscope.
NO TE
After you have connected to the LAN card, you must set up the network. Exit the oscilloscope
application before you start setting up your network.
If you do not know how to set up a network in the Windows 7 operating system, see your
network administrator or use the Windows 7 operating system’s online help.
Figure 2
Side Panel
Keysight Infiniium S-Series Oscilloscopes User’s Guide
15
1
Setting Up the Oscilloscope
Connecting Power
Table 2
Power Requirements
Power
100-120 V, 50/60/400 Hz
100-240 V, 50/60 Hz
380 W Max
1 Connect the power cord to the side of the oscilloscope, then to a suitable AC
voltage source.
CAU TI O N
Use only power cords designed for your oscilloscope.
The power cord provided is matched to the country of origin of the order.
WA RN ING
16
To avoid electric shock, be sure the oscilloscope is properly grounded.
Keysight Infiniium S-Series Oscilloscopes User’s Guide
1
Setting Up the Oscilloscope
Connecting Oscilloscope Probes
1 Attach the probe connector to the desired oscilloscope channel or trigger input
using the probe instructions.
2 Connect the probe to the circuit of interest using the browser or other probing
accessories.
3 Disconnect the probe.
.
Figure 3
S-Series Oscilloscope Probe Connectors
CAU TI O N
Do not exceed the maximum input voltage rating. The maximum input voltage for the
50 Ω inputs is ± 5 Vpeak, and for the 1 M Ω it is 300 Vrms.
Keysight Infiniium S-Series Oscilloscopes User’s Guide
17
1
Setting Up the Oscilloscope
Tilting the Oscilloscope for Easier Viewing
Tabs under the front feet of the oscilloscope can be flipped out to tilt the
oscilloscope.
Figure 4
18
Latching the Front Feet
Keysight Infiniium S-Series Oscilloscopes User’s Guide
1
Setting Up the Oscilloscope
Turning On the Oscilloscope
•
Figure 5
Press the power switch in the lower left corner of the oscilloscope front panel.
Turning On the Oscilloscope
After a short initialization period, the oscilloscope display appears. The
oscilloscope is ready to use.
•
You can connect and disconnect probes and cables while the oscilloscope is
turned on.
Keysight Infiniium S-Series Oscilloscopes User’s Guide
19
1
Setting Up the Oscilloscope
Verifying Basic Oscilloscope Operation
1 Connect one end of the passive probe cable to oscilloscope input channel 1.
2 Connect the other end of the passive probe cable to the front panel probe comp
with the square wave label.
3
Push
Push
for Vernier
to Zero
4
Probe Comp
DIGITAL D15-D0
Front panel connector
with square wave label
Figure 6
1
All Inputs
2
~ 14pF
1M
~
300 V RMS
+ 5V MAX
50
3
4
Channel 1 input
Verifying Basic Oscilloscope Operation
3 Press [Defaul t Setup] on the front panel.
The display will pause momentarily while the oscilloscope is configured to its
default settings.
4 Press [Auto Scale] on the front panel.
The display will pause momentarily while the oscilloscope adjusts the time/div
setting and vertical scale. You should then see a square wave with about four
cycles on screen and a peak-to-peak amplitude of approximately five divisions.
If you do not see the waveform, make sure your power source is adequate, the
oscilloscope is properly powered on, and the probe is connected securely to the
front panel connector output.
5 Move the mouse around the mouse surface and verify that the on-screen
pointer follows the mouse movement.
6 Press the [Touch] key on the front panel to turn on the touch screen. Press and
hold your finger to the screen. A right-click menu appears, which verifies that
the touch screen is working properly.
20
Keysight Infiniium S-Series Oscilloscopes User’s Guide
Setting Up the Oscilloscope
1
Installing Application Programs on Infiniium
Infiniium has an open Windows operating system, which lets you install your own
application software. Any application that runs on Microsoft Windows 7 Embedded
and uses 8 GB of RAM or less may be installed on your Infiniium oscilloscope.
NO TE
CAU TI O N
Exit the oscilloscope application before installing any software.
Installing an application that does not meet these requirements may break the
oscilloscope application and require a hard drive recovery.
Keysight Infiniium S-Series Oscilloscopes User’s Guide
21
1
Setting Up the Oscilloscope
Changing Windows Operating System Settings
NO TE
Exit the oscilloscope application before changing any Windows operating system settings
outside of the oscilloscope application.
Many Windows operating system settings can be changed to suit your own
personal preferences. However, some operating system settings should not be
changed because doing so would interfere with the proper operation of the
oscilloscope.
22
•
Do not change the Power Options.
•
Do not change the Language settings.
•
Do not remove Fonts.
•
Do not change the screen resolution from 1024 by 768 pixels.
•
Do not use the Administrative Tools to enable or disable Internet Information
Services. Use the Infiniium Remote Setup dialog box to enable or disable the
Web Server.
•
Do not delete or modify the Infiniium Administrator user account.
Keysight Infiniium S-Series Oscilloscopes User’s Guide
1
Setting Up the Oscilloscope
Turning Off the Oscilloscope
•
Press the power switch at the lower left corner of the oscilloscope front panel.
The oscilloscope will go through a normal Windows operating system
shutdown process.
Cleaning the Oscilloscope
Clean the oscilloscope with a soft cloth dampened with a mild soap and water
solution.
CAU TI O N
Do not use too much liquid in cleaning the oscilloscope. Water can enter the Infiniium
panels, damaging sensitive electronic components.
Keysight Infiniium S-Series Oscilloscopes User’s Guide
23
1
24
Setting Up the Oscilloscope
Keysight Infiniium S-Series Oscilloscopes User’s Guide
Keysight Infiniium S-Series Oscilloscopes
User’s Guide
2 Using the Oscilloscope
Front Panel Inputs and Outputs 26
Side Panel Inputs and Outputs 27
Front Panel Controls (Keys and Knobs) Overview 29
User Interface Overview 30
Menu overview 32
Using the Setup and Display Controls 33
Starting and Stopping Waveform Acquisitions 34
Adjusting the Horizontal Time Scale and Trigger Position 35
Adjusting the Vertical Settings 38
Setting Up Triggers 40
Using Markers and Making a Measurement 42
Saving and Printing Data 46
Forcing a Default Setup 47
This chapter describes how to use the Infiniium S-Series oscilloscope’s inputs and
outputs, front panel controls, and user interface.
•
The familiar front-panel oscilloscope interface with knobs and keys is optimized
for common tasks and basic measurements.
•
The Infiniium oscilloscope application’s user interface with menus, windows,
dialog boxes, and toolbars provides easy logical access to dozens of
configuration and analysis tools, making it easy for you to set up and make
complex measurements.
•
You have the option of using either the front panel controls or the user interface
for many common tasks.
25
2
Using the Oscilloscope
Front Panel Inputs and Outputs
On the Infiniium S-Series oscilloscopes, the channel inputs, digital channels
connector, probe compensation terminal, and ground plug appear on the lower
part of the front panel. Two USB 2.0 host ports are also located here.
Figure 7
Infiniium S-Series Oscilloscope Front Panel I/O
Channel inputs
Your Infiniium oscilloscope comes with four 10:1 500 MHz passive probes.
The AutoProbe interface works with the InfiniiMax III probing system. See
“Connecting Oscilloscope Probes" on page 17.
For the latest information about probes for your Infiniium oscilloscope, install the
Probe Resource Center. Instructions for downloading are in the online Help
described in Chapter 3, “Online Help and Other Information”.
Digital channels connector
MSO models include a 17-channel flying lead set logic probe, an MSO cable, and a
calibration fixture.
Probe compensation terminal
This terminal has a square wave signal that is used to adjust compensated passive
probes.
You can also output a DC level on this terminal using the Infiniium oscilloscope
application's Calibration Output dialog box (Utilities > Calibration Output...).
Ground
The ground plug is convenient for ESD wrist straps.
26
Keysight Infiniium S-Series Oscilloscopes User’s Guide
Using the Oscilloscope
2
Side Panel Inputs and Outputs
The Infiniium S-Series oscilloscope’s right side panel has the motherboard I/O
connectors, reference clock synchronization connectors, and BNC connectors.
Figure 8
Infiniium S-Series Oscilloscope Side Panel I/O
Motherboard I/O
The motherboard provides these inputs/outputs/ports in the oscilloscope: four
USB ports for peripherals, an external monitor connector, a USB III device port (for
remote control of the oscilloscope from a PC), a LAN port, and speaker and
microphone connectors.
Keysight Infiniium S-Series Oscilloscopes User’s Guide
27
2
Using the Oscilloscope
Aux Out
This output signal is selected by the Infiniium oscilloscope application's
Calibration Output dialog box. It can be a DC level, the probe compensation signal
(a square wave used to adjust compensated passive probes), the trigger out
signal, or a demo signal.
10 MHz In, 10 MHz Out
The 10 MHz In BNC connector is used to synchronize the oscilloscope's horizontal
timebase system to a reference clock that you provide. The clock that you provide
must meet the following specifications:
•
Amplitude: 800 mV peak to 1.26 V peak
•
Frequency: 10 MHz ±5 ppm high-quality sine wave or square wave
To use an external reference clock, connect the external clock to the 10 MHz In
BNC connector; then, in the Infiniium oscilloscope application's Horizontal dialog
box (Setup > Horizontal...), enable the External 10 MHz Reference Clock.
You can use the 10 MHz Out BNC connector to send the oscilloscope’s 10 MHz
reference clock output signal to another instrument’s reference clock input.
Trig Out
Pulses corresponding to oscilloscope triggers can be sent to this BNC output.
Aux Trig
You can set up the oscilloscope to trigger on the auxiliary trigger signal connected
to this BNC input.
28
Keysight Infiniium S-Series Oscilloscopes User’s Guide
Using the Oscilloscope
2
Front Panel Controls (Keys and Knobs) Overview
The Infiniium S-Series oscilloscope front panel gives you direct access to the
functions needed to perform the most common measurements, using a traditional
oscilloscope interface. Knobs and keys let you directly set vertical and horizontal
parameters. The front panel also has a set of LED indicators; by using these and
the display, you see the oscilloscope’s configuration at a glance.
The oscilloscope uses color consistently throughout the front panel and user
interface. For example, the color of the knob for channel 1 is the same color as the
waveform for channel 1. All the configuration items and values related to channel
1 are displayed in the same color.
Figure 9
Infiniium S-Series Oscilloscope Front Panel
Keysight Infiniium S-Series Oscilloscopes User’s Guide
29
2
Using the Oscilloscope
User Interface Overview
With the user interface for the Infiniium oscilloscope, you can access all the
configuration and measurement features of the oscilloscope through an
easy-to-use system of windows, menus, toolbars, dialog boxes, icons, wizards,
and buttons.
The user interface is arranged so the most common functions affecting the
waveform display are located around the edge of the waveform display area.
Context-sensitive menus are available when you right-click something in the
waveform display area, such as the grid, a signal, a bookmark, or a measurement.
You can mouse over or touch other areas, such as the drag & drop measurements
area and horizontal and acquisition control regions, to find more information about
those areas or to enter data.
The following figures call out the areas and controls of the user interface.
Clear display
Open the
Digital dialog box
Memory bar
Turn waveforms on/off
Grid selection:
Drag waveform
Pin/unpin controls Draw rectangle
mode
mode
Menu bar
ertical scale
ertical offset
Figure 10
Access the Channel
dialog box
Access the
Trigger
dialog box
Set
trigger level
Infiniium Oscilloscope Top of Display
The selected grid mode in the upper right corner determines whether you draw a
selection box or manipulate waveforms when you touch the screen.
30
Keysight Infiniium S-Series Oscilloscopes User’s Guide
2
Using the Oscilloscope
Scale settings
Drag & Drop measurements
Ground reference indicator
Figure 11
Infiniium Oscilloscope Waveform Display Area
The waveform display area shows up to eight waveforms areas. Several display
options are available, such as a grid, or horizontal and vertical scales.
NOT E
Avoid Overdriving Vertical Input Amplifiers
When zooming on a waveform with the oscilloscope running, be careful to keep the signal
within the screen vertically to avoid overdriving the vertical input amplifiers. Overdriving
causes waveform distortion and erroneous measurement results.
Ground reference indicators appear for each displayed channel, waveform
memory, or math function. The symbol represents the ground reference point for
each waveform; it moves when you change the vertical offset. You can also drag
this symbol up and down to change the vertical offset for that waveform.
Keysight Infiniium S-Series Oscilloscopes User’s Guide
31
2
Using the Oscilloscope
Expand drag & drop
measurements
Zoom on/off
Set horizontal
position
Pin/unpin
controls
Markers turn on/
set up
Set horizontal scale
Access the Horizontal
dialog box
Figure 12
Infiniium Oscilloscope Bottom of Display
A Results pane is visible at the very bottom of the display when you do anything
that requires it, such as taking a measurement. When it is not needed, the Results
pane is not visible.
Figure 13
Results Pane
Menu overview
You can use menus to perform defined operations, set up measurement
parameters, and access every function the oscilloscope provides.
Take a few minutes to look through the menus to get an overview of the many
features and capabilities of your Infiniium oscilloscope.
The sub-menu selections will vary slightly, depending on which licenses are
installed.
Figure 14
32
Menu bar
Keysight Infiniium S-Series Oscilloscopes User’s Guide
2
Using the Oscilloscope
Using the Setup and Display Controls
You can set the oscilloscope to a known starting condition and set the display to
suit your preferences.
Auto
Scale
Figure 15
Touch
Clear
Display
Default
Setup
Setup and Display Control Keys
•
To automatically configure the oscilloscope for the current input signal(s), press
the [Auto Scale] key or click Control > Autoscale from the menu bar.
•
To enable or disable the touch screen, press the [Touch] key. Or, click Utilities >
User Preferences... to open the User Preferences dialog box, and click the Enable
Touch Screen check box.
•
To clear the waveform display, press the [Clear Display] key or click the Clear
Display button
.
The oscilloscope clears acquired waveform data from the display in preparation
for another acquisition. If the oscilloscope is in Run mode and is receiving
triggers, it will update the display as it collects new waveform data.
Clearing the waveform display also resets measurements, averaging, infinite
persistence, color grade persistence, histograms, and the mask testing
database.
•
To reset the oscilloscope to its default setup, press the [Defaul t Setup] key or
click Control > Defaul t Setup.
You can click Control > Undo Defaul t Setup to return the oscilloscope to its original
configuration.
NOT E
Save the Current Oscilloscope Configuration
Before using the default setup, you may want to save the current oscilloscope configuration
for later use. See the online help (described in chapter 3) for instructions on saving and
recalling setups, and for information on the exact configuration that is set when you use the
default setup.
Keysight Infiniium S-Series Oscilloscopes User’s Guide
33
2
Using the Oscilloscope
Starting and Stopping Waveform Acquisitions
Use the acquisition run controls to run and stop acquisitions or make a single
acquisition. The boxed area of the memory bar above the waveform display area
shows how much of acquisition memory is displayed on the screen.
Figure 16
Front Panel Acquisition Run Control Keys
Figure 17
User Interface Acquisition Run Control Buttons
The [Run/Stop] key is lit green and the green Run button is highlighted when the
oscilloscope is running (acquiring data). The [Run/Stop] key is red and the Stop
button is highlighted when the acquisition is stopped.
•
To start waveform acquisition, press the [Run/Stop] key or click the Run button.
The oscilloscope begins acquiring data. When it receives a trigger signal, it
finishes acquiring data, updates the display, and then starts another acquisition
cycle if it is in Trig’d or Auto trigger mode.
•
To stop waveform acquisition, press the [Run/Stop] key or click the Stop button.
The oscilloscope stops acquiring data. Whatever data was last acquired
remains on the screen.
34
•
To make a single acquisition, press the [Single] key or click the Single button.
•
You can also click the Run, Run Single, and Stop commands from the Control
menu.
•
To set up how you want the signals to be sampled, such as sampling rate and
mode, click Setup > Acquisition....
Keysight Infiniium S-Series Oscilloscopes User’s Guide
Using the Oscilloscope
2
Adjusting the Horizontal Time Scale and Trigger Position
Use the horizontal controls to configure the oscilloscope’s horizontal scale (time
per division) and horizontal position of the waveform. You can view a magnified
section of the waveform using the zoom window.
Use the horizontal scale and position knobs, the horizontal controls, or the
Horizontal dialog box to adjust the horizontal scale and position.
Figure 18
Horizontal Scale and Position Knobs, [Zoom] key
Access the
Horizontal dialog box
Set horizontal scale
Figure 19
Turn Zoom mode on/off
Set horizontal position
(delay)
User Interface Horizontal Controls
Adjusting the horizontal scale
The horizontal scale knob is the larger of the two horizontal control knobs.
•
To stretch the waveform horizontally, turn the horizontal scale knob clockwise.
To shrink it horizontally, turn the knob counter-clockwise.
Stretching the waveform means fewer seconds are displayed per division.
Shrinking it means more seconds are displayed per division.
•
Push and turn the horizontal scale knob to change the scaling in finer (Vernier)
increments.
•
Use the multi-touch gesture capability to stretch or shrink the waveform.
Keysight Infiniium S-Series Oscilloscopes User’s Guide
35
2
Using the Oscilloscope
•
You can also use the controls in the horizontal toolbar to adjust the horizontal
scale. Mouse over or touch the horizontal scale field and use the resulting
controls to set a particular horizontal scale. You can click the Scale field to enter
an exact value, or click the “narrower” or “wider” buttons.
Adjusting the horizontal trigger position (delay)
The horizontal position knob is the smaller of the two horizontal control knobs.
•
To move the waveform to the right, turn the horizontal position knob clockwise.
To move it to the left, turn the knob counter-clockwise.
Moving the waveform to the right shows more of the pre-trigger data (data
acquired before the trigger event). Moving the waveform to the left shows more
of the post-trigger data.
When you click and drag the desired waveform, the horizontal position will
change for all channels and functions on the display. Waveform memories will
also move if you check the Tie to Timebase box in the Waveform Memories
dialog box.
•
You can also use the controls in the horizontal toolbar to adjust the horizontal
position. Mouse over the horizontal position field and use the resulting controls
to set a particular horizontal position (time relative to the trigger at the
highlighted horizontal reference point).
Magnifying a part of the waveform using Zoom
•
To turn on zoom, press the
key or click the Zoom button.
The waveform display area splits into two regions. The top one is the main
timebase. The bottom is the zoomed timebase, which represents an expansion
of the acquired waveform data. A section of the waveform in the main timebase
window is highlighted to indicate the part shown in the zoomed timebase
window.
The horizontal scale and horizontal position controls now change how the
waveform is shown in the zoomed timebase window. The horizontal scale will
change the amount of magnification, while the position will change the part of
the waveform in the main window that is shown in the zoomed window.
•
Press the
key or click the Zoom button again to turn off Zoom.
Setting the scale, position, and timebase reference
You can use the Horizontal dialog box to set scale, position, and timebase
reference. You can also set up the zoomed timebase window.
•
36
To access the Horizontal dialog box, click the
click Setup > Horizontal... from the menu bar.
in the horizontal toolbar, or
Keysight Infiniium S-Series Oscilloscopes User’s Guide
Using the Oscilloscope
•
2
The Timebase Reference control is a slider from 0 to 100% of the screen. The
center of the screen is at 50%. The slider correlates to the solid orange triangle
at the bottom of the display area, showing you where the horizontal offset is on
the screen.
Keysight Infiniium S-Series Oscilloscopes User’s Guide
37
2
Using the Oscilloscope
Adjusting the Vertical Settings
Use the vertical controls to set the vertical scaling (volts per division) and vertical
offset for each analog channel. You can also turn the display on or off for a
particular channel.
Figure 20
Channel Keys and Vertical Scale and Offset Knobs
Open the Channel
dialog box
Turn signals on/off
Set vertical scale
Figure 21
Set vertical offset
User Interface Vertical Controls
Turning an analog channel on or off
•
38
To turn an analog channel on or off, press the channel number key on the front
panel or click the Add Waveforms button
. When you turn off a channel, the
current vertical scale and offset fields for that channel disappear.
Keysight Infiniium S-Series Oscilloscopes User’s Guide
2
Using the Oscilloscope
If you are not using a particular analog channel, you can turn it off to simplify
the waveform display and increase the display update rate. Functions continue
to run on a channel source that is turned off. Data acquisition continues for a
channel if a function requires it.
NOT E
Using an Analog Channel as Trigger
Any analog channel can be used as a trigger source. If you need a trigger but do not need all
analog channels, you can use an analog channel as a trigger without displaying it by turning
the analog channel display off.
Adjusting the analog channel’s vertical scale and offset
You can use the vertical scale and offset knobs, the vertical user interface controls,
or the Channel dialog box to adjust the vertical scale and offset.
The vertical scale knob is the larger of the two knobs for a channel. It is marked
with a set of sine wave symbols.
•
To make the waveform bigger (fewer volts per division), turn the vertical scale
knob clockwise. To make it smaller (more volts per division), turn the knob
counter-clockwise.
Decreasing the vertical scale makes the waveform bigger, and increasing it
makes the waveform smaller.
•
You can also mouse over or touch the vertical scaling field and use the resulting
controls to set an exact value for the scaling.
The vertical offset knob is the smaller of the two knobs for a channel. It is marked
with a set of arrows.
•
To move the waveform toward the top of the display, turn the vertical offset
knob clockwise. To move it toward the bottom of the display, turn the knob
counter-clockwise.
•
You can click and drag the waveform or its ground reference indicator to the
desired vertical offset if the grid is in drag mode
.
•
Click Setup > Channel N... or click a channel number to open the Channel dialog
box, in which you can set the vertical scale, offset, skew, and labels. You can
also specify the characteristics of a probe, or perform a probe calibration.
For Keysight Technologies probes that are compatible with AutoProbe II
interfaces, the oscilloscope will automatically set these characteristics (except
for skew) after identifying the probe when it is connected to the channel input.
Keysight Infiniium S-Series Oscilloscopes User’s Guide
39
2
Using the Oscilloscope
Setting Up Triggers
Use the trigger controls to set the conditions on which the oscilloscope will trigger
and acquire an input signal. You can set up a variety of trigger conditions. Edge
triggers and the parameters for edge triggering can be set up from the front panel.
Trigger configuration settings you make using the user interface are reflected in
the front panel status indicators, and will remain set unless you change them or
press the [Defaul t Setup] key.
Figure 22
Front Panel Trigger Controls and Indicators
Figure 23
User Interface Trigger Controls
Setting the oscilloscope to trigger on an edge
1 Press the [Source] key until the desired source LED is lit.
You can choose any of the channels or the Aux Trig or Line input as the source
for an edge trigger.
2 Press the [Slope] key until the desired slope LED is lit.
You can have an edge trigger on a rising or falling edge, or both.
3 Press the [Sweep] key until the desired LED is lit (Trig’d or Auto).
40
Keysight Infiniium S-Series Oscilloscopes User’s Guide
Using the Oscilloscope
2
When Trig’d is selected, the oscilloscope must find the trigger before capturing
and displaying data.
When Auto is selected, if a trigger does not occur within a certain amount of
time, an acquisition is automatically saved and displayed. In Auto trigger mode,
you are able to see your signals while setting up the desired trigger.
4 Turn the Level knob to adjust the voltage level at which the oscilloscope will
trigger.
Use the Trigger dialog box to select any of the different modes of triggering, the
parameters and conditions for each trigger mode, and advanced configuration
items.
You can also mouse over the Trigger Level field and use the resulting controls to
set a particular trigger level when the scope is set for edge trigger on a particular
channel. You can also drag the trigger reference indicator at the left side of the
display.
Keysight Infiniium S-Series Oscilloscopes User’s Guide
41
2
Using the Oscilloscope
Using Markers and Making a Measurement
With the measurement controls you can display and adjust markers, define a key
to perform automatic measurements or other quick actions, and make
measurements.
Figure 24
Front Panel Measure Controls
Drag & drop measurement
icons showing most
commonly used measurements
Expand and collapse the
measurement icons
Figure 25
42
Drag & Drop Measurement Toolbar
Keysight Infiniium S-Series Oscilloscopes User’s Guide
2
Using the Oscilloscope
Making a measurement on a waveform
Click and drag a measurement icon to the waveform event you want to measure,
or click a measurement icon and specify which source you want to measure in the
dialog box that appears.
For measurements on waveform features, such as those that involve waveform
edges, if you click the measurement icon and specify a source, the measurement
defaults to using the feature closest to the horizontal reference point. When you
make the measurement using drag-and-drop, the measurement uses the
waveform feature closest to the point where you drop the icon.
The most commonly used measurements are available in the drag and drop area.
Others are available from the Add Measurement dialog box.
When you drag and drop a measurement icon on a waveform, the icon outline
changes color to match the color of each waveform it touches so you can easily
see which waveform will be measured.
For edge-sensitive measurements, when you drop the measurement icon on a
waveform, a circled number appears in the waveform marker color. This number
shows exactly where the measurement is being made. It is displayed next to the
measurement readout in the Results area.
This feature helps you distinguish measurement results from each other when you
make multiple measurements on the same waveform, but at different waveform
features.
Using quick measurements
•
You can define the [Mul ti Purpose] key to perform one of these quick actions:
•
QuickMeas—performs automatic measurements
•
QuickPrint—prints the screen image to the default printer
•
QuickScreen—saves the screen image to a file
•
QuickSetup—loads a setup file
•
QuickWaveform—saves waveform data to a file
•
QuickEmail—sends an email of the current screen image
•
QuickExecute—runs an executable file
•
QuickControl—cycles through the possible front panel control sources
The action taken when the [Mul ti Purpose] key is pressed depends on the feature
selected in the Customize Multipurpose dialog box (Utilities > Customize
Mul tipurpose...). The default feature is QuickMeas.
•
To turn on the quick measurement display, press the [Mul ti Purpose] key.
The 10 preset measurements defined in the Quick Measurement configuration
are enabled and results are displayed on the screen for the first waveform
source.
Keysight Infiniium S-Series Oscilloscopes User’s Guide
43
2
Using the Oscilloscope
•
To measure parameters for another waveform, press the [Mul ti Purpose] key until
that waveform is the one shown in the measurement readout.
Continuing to press the [Mul ti Purpose] key cycles through each of the
waveforms available.
•
To turn off the quick measurement display, press and release the [Mul ti Purpose]
key until the measurements are turned off.
See the Infiniium oscilloscope application’s online help for information on how to
configure the quick measurement capability.
Using markers
Markers make it easier to make precise measurements because the marker
measurement readouts show exact voltage and time positions for the markers. The
measurements are based on actual waveform data from the acquisition system,
not on approximations based on the display position, so you can be sure the values
are highly accurate.
Using the marker and measurement controls, you control two sets of markers
within the oscilloscope grid.
Both time and voltage differences between the markers are updated continuously
on the screen. By default, the markers track the source waveform. Voltage
measurements from the markers are the value of the waveform at the time set with
the marker arrow keys.
•
To select the type of marker mode you want to use, press the [Markers] key.
•
To turn on Marker 1 (X1), push the Position knob. Turn the knob to move the
marker. Push the knob again to select Marker 1 (Y1).
Marker 1 (X1 and Y1) has a solid line pattern on the waveform display. It is
associated with the first available source on the display.
•
To turn on Marker 2 (X2), push the Position knob again.
Marker 2 has a dashed line pattern on the waveform display. It is associated
with the first available source on the display.
•
In Track Measurements mode, the marker position cannot be changed.
Moving markers using the user interface
1 Turn on the markers by clicking Measure > Markers... and selecting a mode. You
can also click the Markers button
Markers dialog box.
to either turn on markers or bring up the
2 Drag one of the markers to the position you want on the waveform.
Dragging a marker makes it easy to quickly move the marker to the desired
waveform event. You can use the front-panel Position knob for fine adjustment,
or click Measure > Markers... and set the marker position precisely.
44
Keysight Infiniium S-Series Oscilloscopes User’s Guide
Using the Oscilloscope
2
Controlling digital channels
If your oscilloscope is an MSO model, click Setup > Digital Channels... to open the
Digital dialog box so you can set up controls for the digital channels.
Turning digital channels on or off
To turn the digital channels on, click the Add Waveforms button
check box next to the
, or press the [Digital] key.
and select the
Decoding serial data
•
To open the Protocol Decode dialog box so you can define parameters for
selected decodes, click Setup > Protocol Decode... or press [Serial Decode].
You can perform up to four decodes at the same time using p1-p4.
•
After selecting the protocol decode parameters, click Auto Setup to
automatically configure the oscilloscope for the selected decode type.
•
You can view decoded acquisition data in the Digital Listing Window that
appears.
Keysight Infiniium S-Series Oscilloscopes User’s Guide
45
2
Using the Oscilloscope
Saving and Printing Data
46
•
Click File > Save > to save your composite, setup, waveform, screen image, or
measurement data.
•
Click File > Copy Screen Image to easily copy and paste a screen image into a
document.
•
Click File > Print... to print waveform and setup data to a specified file.
•
You can customize the [Mul ti Purpose] key to perform a QuickPrint.
Keysight Infiniium S-Series Oscilloscopes User’s Guide
2
Using the Oscilloscope
Forcing a Default Setup
If your Infiniium oscilloscope is not working properly when you start it up, follow
these steps to perform a default setup and return the Infiniium to normal
operation.
1 Click Control > Defaul t Setup or press the [Defaul t Setup] key.
2 If the oscilloscope is still not working properly, click Control > Factory Defaul t to
return the oscilloscope to the default settings it had when it left the factory.
3 If the oscilloscope is still not working properly, turn it off.
4 Turn the oscilloscope back on. If it does not successfully restart, try recycling
the power again.
5 As soon as the Windows 7 load screen disappears, press [Defaul t Setup]. If the
oscilloscope still does not successfully restart, follow the instructions for
recovering the hard drive.
Infiniium hard drive recovery
Follow these steps to recover your Infiniium hard drive.
1 Turn off the oscilloscope.
2 Connect the keyboard to the keyboard connector on the rear panel of the
oscilloscope.
3 Connect the mouse to the mouse connector on the rear panel of the
oscilloscope.
4 Turn on the oscilloscope.
5 As soon as you see the message press F2 to boot the Infiniium Recovery
System, press [F2] and follow the on-screen instructions.
6 Once the recovery process is finished and the oscilloscope is running, check in
the About Infiniium dialog box under installed options to see if all of the options
you ordered are installed. If the options are not installed, please install them
using the license keys provided on the oscilloscope option license certificates
you received, or refer to the rear panel of the oscilloscope.
Keysight Infiniium S-Series Oscilloscopes User’s Guide
47
2
48
Using the Oscilloscope
Keysight Infiniium S-Series Oscilloscopes User’s Guide
Keysight Infiniium S-Series Oscilloscopes
User’s Guide
3 Online Help and Other Information
Accessing the Online Help 49
Navigating the Online Help 51
Using the Demo Wizard 51
Most of the information about using the Infiniium oscilloscope effectively is
included in the online help.
The online help is accessible from the menu bar and by clicking the question mark
in the top right corner of dialog boxes.
Accessing the Online Help
•
Click Help > Contents....
The Using the Oscilloscope topic appears on the home page. It looks like the
following figure.
49
3
Online Help and Other Information
Hides the left pane
Shows the current topic
in the Contents
Figure 26
Prints current topic
on the default printer
Online Help Home Page
•
For information about items in a dialog box, click the Help button in that dialog
box.
Displays the Help topic
for this dialog box
Figure 27
50
Help Button for Dialog Box Help
Keysight Infiniium S-Series Oscilloscopes User’s Guide
Online Help and Other Information
3
Navigating the Online Help
The online help provides several ways to find the information you need.
Use the Contents tab to browse topics in the help system by clicking topics in the
left pane.
Use the Index tab to type in a keyword and search the index for that keyword or
scroll through the list to find a topic.
Use the Search tab to type in a keyword and if that word exists in the online help, a
list of topics containing the keyword appears.
Use the Favorites tab to add preferred help topics to a list for easy reference.
Using the Demo Wizard
Your S-Series oscilloscope comes with a built-in demo wizard that showcases
many of the oscilloscope’s capabilities. To see the demos, click Help > Show Demo
Wizard.... You can then select a specific demo, such as a particular protocol or a UI
demo showing bookmarks.
For example, the following screen display shows the initial demo page for the FFT
function. To experiment with the demo, click Load Demo....
Figure 28
FFT Demo
Keysight Infiniium S-Series Oscilloscopes User’s Guide
51
3
52
Online Help and Other Information
Keysight Infiniium S-Series Oscilloscopes User’s Guide
Keysight Infiniium S-Series Oscilloscopes
User’s Guide
4 Working in Comfort
About Repetitive Strain Injury 54
Mice and Other Input Devices 55
To optimize your comfort and productivity, it is important that you set up your
work area correctly and use your Infiniium oscilloscope properly. With that in mind,
we have developed some setup and use recommendations for you to follow based
on established ergonomic principles.
Improper and prolonged use of keyboards and input devices are among those
tasks that have been associated with repetitive strain injury (RSI) to soft tissues in
the hands and arms. If you experience discomfort or pain while using the
oscilloscope, discontinue use immediately and consult your physician as soon as
possible. For more information on RSI, read the “About Repetitive Strain Injury”
section, next.
Please study the recommendations offered here in this chapter. Included are
references to relevant parts of international standards, regulations, and guidelines,
such as ISO 9241 and the European Community Display Screen Equipment
directive. You may also consult your employer’s human resources department or
other relevant departments for guidance specific to your company.
53
4
Working in Comfort
About Repetitive Strain Injury
Because your comfort and safety are our primary concern, we strongly recommend
that you use the Infiniium oscilloscope in accordance with established ergonomic
principles and recommendations. Scientific literature suggests that there may be a
relationship between injury to soft tissues—especially in the hands and arms—and
prolonged improper use of keyboards or other equipment requiring repeated
motions of the hands and forearms. This literature also suggests that many other
risk factors may increase the chance of such injury, commonly called Repetitive
Strain Injury.
What is RSI?
Repetitive Strain Injury (RSI—also known as cumulative trauma disorder or
repetitive motion injury) is a type of injury where soft tissues in the body, such as
muscles, nerves, or tendons, become irritated or inflamed. RSI has been a reported
problem for those who perform repetitive tasks such as assembly line work,
meatpacking, sewing, playing musical instruments, and computer work. RSI also
has been observed in those who frequently engage in activities such as carpentry,
knitting, housework, gardening, tennis, windsurfing, and lifting children.
What causes RSI?
The specific causes of RSI have not been established. Nevertheless, the incidence
of RSI has been associated with a variety of risk factors, including:
•
Too many uninterrupted repetitions of an activity or motion
•
Performing an activity in an awkward or unnatural posture
•
Maintaining static posture for prolonged periods
•
Failing to take frequent short breaks
•
Other environmental and psychosocial factors
In addition, there have been reports associating the occurrence of RSI with the use
of keyboards, mice, and other input devices. Also, certain medical conditions, such
as rheumatoid arthritis, obesity, and diabetes, may predispose some people to this
type of injury.
What if I experience discomfort?
If you are experiencing any discomfort, seek professional medical advice
immediately. Typically, the earlier a problem is diagnosed and treated, the easier it
is to resolve.
54
Keysight Infiniium S-Series Oscilloscopes User’s Guide
Working in Comfort
4
Mice and Other Input Devices
Various aspects of using mice and other input devices may increase your risk of
discomfort or injury. Observing the following recommendations may reduce that
risk.
•
Try to keep your hand, wrist, and forearm in a neutral position while using your
mouse or other input device.
•
If you use your thumb to rotate the ball on a trackball or spaceball, keep it in a
relaxed, natural shape, and maintain a neutral posture in your hand, wrist, and
forearm.
•
Hold the mouse gently by draping your fingers over it. Keep your hand relaxed
and fingers loose. Do not grip the mouse tightly.
•
It takes very little pressure or force from your fingers to activate the buttons or
scroll wheel on your mouse, scrolling mouse, trackball, or other input device.
Using too much force can place unnecessary stress on the tendons and
muscles in your hands, wrists, and forearms.
•
If you are using a scrolling mouse, be sure to keep your fingers and hand in a
relaxed, neutral position when activating the scroll wheel. Also, this type of
mouse features software that can minimize the number of mouse movements or
button clicks.
•
When using a mouse, trackball, or other input device, position it as close to the
keyboard as possible, and keep it at the same level so you do not have to
stretch while using it.
•
Be sure to keep your mouse and trackball clean. Regular removal of
accumulated dust and dirt helps ensure proper tracking and reduces
unnecessary hand and wrist motions.
Keysight Infiniium S-Series Oscilloscopes User’s Guide
55
4
56
Working in Comfort
Keysight Infiniium S-Series Oscilloscopes User’s Guide
Index
A
accessing
index, 50, 51
online help, 49
accessories
and options, 12
acquisition
starting and stopping, 34
adjusting
analog channel’s vertical offset, 39
analog channel’s vertical scaling, 39
airflow requirements, 14
application software, 21
associating measurement results with
waveforms, 43
Auto Scale key, 33
cycling
markers through channels, 44
measurements through channels, 43
D
B
default setup, 20
Default Setup key, 33
demo
online, 51
digital channels, 45
Digital key, 45
disk
hard disk recovery, 47
display layout
customizing, 43
display update rate
increasing, 38
drag-and-drop measurements, 43
buttons, 29
E
C
edge trigger
setting, 40
environmental characteristics, 13
calibration
output, 20
cautions
cleaning, 23
channel keys, 38
channels
turning on or off analog channels, 38
checking the oscilloscope, 12
cleaning the oscilloscope, 23
Clear Display key, 33
color
use of, 29, 43
configuration
default, 20
edge trigger, 40
configuring
quick measurements, 43
connecting
power, 16
probes, 17
contents
online help, 49
contents topic, 49
customizing display layout, 43
F
finding information, 49, 51
front panel
connections, 17
interface, 25, 29
H
hard disk recovery, 47
horizontal
zoomed timebase, 36
horizontal position
in zoom mode, 36
I
icons
measurement, 43
increasing display update rate, 38
Keysight Infiniium S-Series Oscilloscopes User’s Guide
index
accessing, 50, 51
information
methods for accessing, 49
inspecting the oscilloscope, 12
installing
application software, 21
K
keys, 29
analog channel, 38
Auto Scale, 33
Clear Display, 33
Default Setup, 33
Digital, 45
Markers, 44
Menu, 40
Multi Purpose, 43
Run, 34
Serial Decode, 45
Single, 34
Slope, 40
Source, 40
Stop, 34
Sweep, 40
Touch, 33
Zoom, 36
knobs, 29
analog channel’s vertical offset, 39
analog channel’s vertical scaling, 39
trigger level, 40
L
LED indicators, 29
level
trigger, 40
line voltage, 16
M
magnifying
a section of the waveform, 36
main timebase, 36
making
measurements, 43
quick measurements, 43
waveform bigger or smaller, 39
57
Index
markers
and measurements, 43
moving, 44
positioning, 44
turning on or off, 44
measurement results
associating with waveform, 43
measurements, 43
and markers, 43
quick, 43
using markers, 44
Menu key, 40
methods for accessing information, 49
mouse
connecting, 15
verifying, 20
moving
markers, 44
waveform vertically, 39
Multi Purpose key, 43
N
navigating the online help, 51
notes
online help access, 49
zooming, 31
notices, 2
O
offset
adjusting analog channels, 39
online help
accessing, 49
index, 51
navigating, 51
overview, 49
opening the online help, 49
options
and accessories, 12
oscilloscope
cleaning, 23
inspecting, 12
using, 25
oscilloscope front panel, 29
oscilloscope operation
verifying, 20
oscilloscope probes, see probes
overdriving vertical amplifiers, 31
overvoltage category, 13
P
pointing device
connecting, 15
verifying operation, 20
pollution degree, 13
pollution degree, definitions, 13
58
position
adjusting analog channel’s vertical, 39
positioning markers, 44
power
connecting, 16
requirements, 16
turning off, 23
preset measurements, 43
probes
connecting, 17
probing a circuit, 17
U
Q
use of color in interface, 29, 43
user interface, 25
front panel, 29
use of color, 29
using
markers, 44
oscilloscope, 25
quick measurements, 43
R
Repetitive Strain Injury, 53, 54
using the mouse, 55
RSI, 53, 54
using the mouse, 55
Run key, 34
V
S
scaling
adjusting analog channel’s vertical, 39
seconds per division
in zoom mode, 36
Serial Decode key, 45
setting
edge trigger, 40
setup
default, 20
Single key, 34
slope
trigger, 40
Slope key, 40
source
trigger, 40
Source key, 40
starting and stopping acquisition, 34
status indicators
trigger, 40
Stop key, 34
sweep
key, 40
T
topics
contents, 49
navigating, 51
Touch key, 33
tracking waveforms
using markers, 44
trigger
input coupling, 40
level, 40
setting edge, 40
slope, 40
source, 40
status indicators, 40
triggered sweep, 40
turning analog channels on or off, 38
turning markers on and off, 44
verifying basic operation, 20
vertical inputs
overdriving, 31
vertical offset
adjusting analog channels, 39
vertical scaling
adjusting analog channels, 39
volts per division
adjusting, 39
W
waveform
magnifying, 36
waveform features
measurements on, 43
moving markers to, 44
waveforms
acquiring, 34
associating with measurement
results, 43
drag-and-drop measurements on, 43
making quick measurements, 43
measuring using markers, 44
turning off, 38
window
zoom, 36
Windows operating system settings, 22
Z
zoom display, 36
Zoom key, 36