Instruction Manual for MODEL A6xx8 1/8
Instruction Manual for A6000 Series
Universal Digital Panel Meters
! Caution
Temperature Measurement Unit
(1) Applying a voltage or current exceeding its maximum permissible value may cause the unit to be damaged.
(2) Always use the unit within the specified voltage range; otherwise, it may cause a fire, electric shock or personal/equipment damage.
(3)
(4)
For the purpose of functional improvement, the information written herein may be changed without prior notice.
Information contained herein is considered accurate to the best of our knowledge. If you have any question or comment on the information, please contact us or our distributor.
(5) Read this manual carefully and thoroughly before starting to operate the unit, and keep the manual available for future reference.
1 Before Using the Unit
Thank you for purchasing our quality designed and manufactured A6000
Series. Before unpacking the unit, check for damages during transportation.
If you have noticed any damage, directly contact us or our distributor.
1.1 Type Identification
Each model number of the A6000 series has its general specifications, and the following describes each note and the meaning. Before using the unit, check that the model number and specifications of the delivered unit match those of the product you ordered. For optional units, see the separate instruction Manuals.
A 6
P o w e r s u p p l y u n i t
( 2 u n i t t y p e s )
: 1
2
A l t e r n a t i n g c u r r e n t
D i r e c t c u r r e n t
D i s p l a y u n i t
( 2 u n i t t y p e s )
I n p u t u n i t
( 1 1 u n i t t y p e s )
:
:
1
2
1
2
3
4
M u l t i d i s p l a y
S i n g l e d i s p l a y
D C v o l t a g e m e a s u r e m e n t
D C c u r r e n t m e a s u r e m e n t ( s m a l l e l e c t r i c c u r r e n t )
D C c u r r e n t m e a s u r e m e n t ( l a r g e e l e c t r i c c u r r e n t )
A C v o l t a g e m e a s u r e m e n t
C o m p a r a t o r o u t p u t u n i t
( 2 u n i t t y p e s )
O p t i o n u n i t s
( 8 u n i t t y p e s )
:
:
5
6
7
8
9
A
B
0
1
2
0
1
A C c u r r e n t m e a s u r e m e n t ( s m a l l e l e c t r i c c u r r e n t )
A C c u r r e n t m e a s u r e m e n t ( l a r g e e l e c t r i c c u r r e n t )
R e s i s t a n c e m e a s u r e m e n t
T e m p e r a t u r e m e a s u r e m e n t
F r e q u e n c y m e a s u r e m e n t
L o a d c e l l i n p u t
P r o c e s s s i g n a l m e a s u r e m e n t
N o n e
R e l a y c o n t a c t o u t p u t
P h o t o c o u p l e r o u t p u t
N o n e
E x t e r n a l c o n t r o l
2
3
4
5
6
7
B C D o u t p u t ( T T L ) + e x t e r n a l c o n t r o l
A n a l o g o u t p u t ( P W M ) + e x t e r n a l c o n t r o l
R S - 2 3 2 C
R S - 4 8 5
R S - 2 3 2 C + a n a l o g o u t p u t ( P W M )
R S - 4 8 5 + a n a l o g o u t p u t ( P W M ) 8
B C D o u t p u t ( o p e n c o l l e c t o r ) + e x t e r n a l c o n t r o l
1.2 Accessories
Accessory
6-p terminal cover
Mounting band
Unit indication label
Quantity Remarks
2 or 3 each 2 without a comparator output 3 with a comparator output
2 pieces
1 each
2 Mounting Method
2.1 Panel Cut Size
Cut the panel to mount the A6000 series in accordance with the illustration
92
+0.8
-0
120 mm or more
2.2 How to Mount the Unit on the Panel
Mount the A6000 series to the panel in accordance with the illustration below:
(1) Remove the mounting band and insert the case (2) Fix the case using the mounting band from the
from the front of the panel.
rear of the panel.
(1) The recommended panel thickness is 0.8 to 5mm.
(2) Do not install the unit in locations where it is exposed to direct sunlight; where ambient emperature or humidity doesn't meet the requirements below; or where a drastic change in temperature may ause condensation.
Ambient temperature: 0 to 50℃
Ambient humidity: 35 to 85%
(3) Do not install the unit where it is exposed to dust, particles, chemicals harmful to electric components, corrosive gases, etc.
(4)
(5)
When this unit is installed inside other equipment, pay attention to the heat radiation and keep the heat inside the equipment 50℃ or below.
Do not install the unit where it is exposed to excessive vibration or shock.
(6)
Install the unit horizontally; otherwise, ventilation will be adversely affected and may result in deterioration.
3 Terminals and Connections
Illustration of the rear of the A6000
Input unit Comparator output unit
Option unit Power supply unit
Each unit is not replaceable by the customer.
3.1 Power supply connections
Top
Bottom
AC power supply
①NC
②NC
③NC
④NC
⑤AC POWER
⑥AC POWER
DC power supply
①DC POW(+)
②DC POW(-)
③NC
④NC
⑤NC
⑥NC
Suitable solderless terminals
5.8 mm or less
5.8 mm or less
3.2 Input Signal Connections
Top ①TC HI
②TC LO
③RTD A
④RTD B
Suitable solderless terminals
5.8 mm or less
5.8 mm or less
⑤RTD C
Bottom
⑥NC
3.3 Comparator Output Connections
Top
Bottom
Relay output
①HI a
②HI c
③GO a
④GO c
⑤LO a
⑥LO c
Photocoupler output
①HI c
②HI e
③GO c
④GO e
⑤LO c
⑥LO e
Suitable solderless terminals
5.8 mm or less
5.8 mm or less
3.4 Option connections
For connections of the options, see separate optional function instruction
Manuals.
WATANABE ELECTRIC INDUSTRY CO., LTD.
UW-33831k
2015.11
Instruction Manual for MODEL A6xx8 2/8
4 Parameter Settings
4.1 Multi Display Unit
Names and major functions
①
④
⑨
H H
H I
G O
L O
L L
Enter
Normal operation
1
1
1
1
1
⑩
Mode
Operation procedure diagram
Enter
1
1
1
1
⑤
⑪
⑥
R E P H D Z T Z
M E P 1 P 2 P 3
⑫
2or3
1
1
1
⑦ ⑧
②
③
+
Location
+ 1 or
1
①
②
③
④
⑤
⑥
⑦
⑧
⑨
⑩
⑪
⑫
2or3
Jog 1
Jog 2
Jog 3
Shift data setting
Name Major function
Menu and value selection at the time of parameter setting; pattern selection when used with
Increment.
Changes judgment values when used with the meter relay; sets a current measurement value to a judgment value; switching maximum values, etc. when used without meter relay.
Changes judgment values when used with the meter relay; sets a current measurement value to a judgment value; switching maximum values, etc. when used without meter relay.
Judgment monitor
Displays results of judgment when used with meter relay.
Main monitor Displays a measured value, menu name or a value at the time of parameter setting.
Sub monitor 1
Function monitor
Sub monitor 2
Enter
Mode
Shift
Increment
Judgment value setting
+ 1
Scaling data pattern selection
Comparator data pattern selection
M E
P 1
P 2
P 3
Displays a judgment value when used with meter relay; displays a maximum value, etc. when used without meter relay.
R E
Illuminates when the unit is set to the remote mode via communication function.
P H
D Z
T Z
Illuminates when peak hold, valley hold, or peak-valley hold is turned on.
Illuminates when Digital Zero is turned on.
Illuminates when Tracking Zero is turned on.
Illuminates when Digital Zero Backup is turned on.
P 1
P 2
P 3
Pattern 1 Pattern 2 Pattern 3 Pattern 4 Pattern 5 Pattern 6 Pattern 7 Pattern 8
O F F
O N
O F F
O F F
O N
O F F
O N
O F F
O F F
O N
O N
O F F
O N
O F F
O N
O N
Displays a judgment value when used with meter relay; displays a maximum value, etc. when used without meter relay.
Switches to the parameter setting mode.
Changes modes at the time of parameter setting; switches to the memory mode at the time of normal measurements (when this button is pushed and held.)
Selects digits at the time of parameter setting; DZ control at the time of normal measurements.
(when this button is pushed and held.)
Changes values at the time of parameter setting; pattern selection at the time of normal measurements (when this button is pushed and held); special operations
Digit selection (from the lowest
2or3 digit to highest digit)
* Returns to measurement after the highest digit.
Set a measured value as a reference.
Pattern selection
2or3
Set low layer values of the condition data.
Increase a value.
2or3
1
1
1
2or3 the highest digit.
Decrease a value.
Returns to measurement after
Select a parameter to change.
Change the parameter selected.
1
1
2
2
Set low layer values of the scaling data.
Set low layer values of the comparator data.
2
Increase or decrease a value.
Digit selection (from the highest digit to the lowest digit)
Mode
1
1
1
1
1
1
Select a parameter to change.
Change the parameter selected.
Select a parameter to change.
Change the parameter selected.
Shift data is reflected upon the measurement data on the Main
Monitor.
Instruction Manual for MODEL A6xx8 3/8
How to set a low layer value (Condition data/scaling data/comparator data)
1
Select the name of a parameter (shown on Sub monitor 2) you want to change by turning Jog 1 up or down.
Remarks:
1 or
Mode
1 1 or 1 1 1
1 or
Mode
1 or
Enter
To change the selected parameter, push the center of Jog 1 or use the Mode key to move from the menu selection mode to value change mode.
*While Sub monitor 2 flashes in the menu selection mode, Main monitor will flash in the value change mode.
If the parameter requires a selection:
Turn Jog 1 up or down to choose a selection.
If the parameter requires a value entry:
Push the center of Jog 1 to choose the digit you want to change, and turn Jog 1 up or down to change the value.
Push the center of Jog 1 or use the Mode key to shift from the value change mode to the menu selection mode.
* While Sub monitor 2 flashes in the menu selection mode, Main monitor will flash in the value change mode.
Turn Jog 1 up or down to select the menu name of the parameter you want to change (shown in Sub monitor 2) or return to measurement by pushing the Enter key.
※The multi display unit can also follow the operation procedure of the single display unit operation system.
※Comparator judgment values can be set not only from Jog 2 and Jog 3 but also from the low layers of the comparator data.
※For operation procedures in the memory mode (maximum value/minimum value/(maximum value-minimum value)),
see the operation procedure diagram of the single display unit.
4.2 Single Display Unit
Names and major functions
①
HH
②
④
HI
GO
LO
LL
Enter
⑤
Mode
⑥ ⑦
③
RE ME
PH P1
DZ
TZ
P2
P3
Location
①
②
③
④
⑤
⑥
⑦
Name
Judgment monitor
Main monitor
Major function
Displays judgment results when used with meter relay.
Displays a measured value as well as menu names and values at the time of parameter setting.
Function monitor
Enter
Mode
Shift
Increment
R E
P H
Illuminates when the unit is set to the remote mode via communication function.
Illuminates when peak hold, valley hold, or peak-valley hold is turned on.
D Z
T Z
M E
P 1
P 2
P 3
Illuminates when Digital Zero is turned on.
Illuminates when Tracking Zero is turned on.
Illuminates when Digital Zero Backup is turned on.
P 1
Pattern 1 Pattern 2 Pattern 3 Pattern 4 Pattern 5 Pattern 6 Pattern 7 Pattern 8
O N O F F
O N O F F
O N O F F
P 2
P 3
O F F
O F F
O N
O F F
Switches to the parameter setting mode.
O F F O N
O F F
O N
O N
O N
Changes modes at the time of parameter setting; switches to the memory mode at the time of normal measurements (when this button is pushed and held.)
Selects digits at the time of parameter setting; DZ control at the time of normal measurements.
(when this button is pushed and held.)
Changes values at the time of parameter setting; pattern selection at the time of normal measurements
(when this button is pushed and held) ; special operations
Instruction Manual for MODEL A6xx8 4/8
Operation procedure diagram
Enter
Push and hold.
Mode
Mode
Mode
Mode
Push and hold.
Push and hold.
Mode
Memory mode
Pattern selection
Mode
Mode
When you push the Mode key, modes will change in the order of the maximum value, minimum value, (maximum value-minimum value) and input value. If you push and hold the Mode key when you are in the memory mode, all the values in memory will be cleared.
Keep on pushing the button for automatic increments.
Set low layer values of the condition data.
Set low layer values of the scaling data.
Set low layer values of the comparator data.
Digit selection.
Value change.
Mode
Mode
Mode
Mode
Select a parameter to change.
Change the parameter selected.
Select a parameter to change.
Change the parameter selected.
Select a parameter to change.
Change the parameter selected.
Shift data is reflected upon the measurement data on the Main
Monitor.
How to set a low layer value (Condition data/scaling data/comparator data)
Select the name of a parameter you want to change by pushing the Increment key.
Mode
To change the selected parameter, use the Mode key to move from the menu selection mode to the value change mode.
If the parameter requires a selection:
or &
Use the Increment key to choose a selection.
If the parameter requires a value entry:
Use the Shift key to choose the digit you want to change, and push the Increment key to set the value.
Mode
Use the Mode key to shift from the value change mode to the menu selection mode.
or
Enter
Use the Increment key to select the menu name of the parameter you want to change or return to measurement by pushing the Enter key.
Instruction Manual for MODEL A6xx8 5/8
4.3 Numeric and Character Indications
0
R
1
S
2
A B C
T
3 4
D E
5
F
6
G H
U V W X
7
Y
8
I
Z
9
J
-
K L M N O P Q
4.4 Protection Lev els
Each parameter of the A6000 has an individual protection level, and by setting the protection level of the condition data, you can set an access level. (For the protect level of each parameter, see the P.L. column of the tables in Section 4.5.)
The higher the protection level is, the less the number of settable parameters will be. If you set the protection level to the strictest LV3, you can change the protection level only, and all the other parameters may not be changed. (No comparator judgment value can be changed using the jog switches in this case.)
*The protection level set at the time of shipment is LV1. (Settings of display colors, scaling and judgment-related values only are available.)
4.5 List of the Parameters
4.5.1
Condition Data
Menu indication
A V G
M A V
S . W D
Parameter name
Average times
Average times of movements
Step wide
C L R Display color
C L R . T
H I . C L
G O . C L
L O . C L
Display color type
HI display color
GO display color
LO display color
B L N K Display blank level
J . S W Jog SW
P V H PH Selection
D Z . B U DZ backup
P S
T R . T
T R . W
P . O N
P.SEL
TZ time
TZ correction range
Power on delay time
Initial value
50
1
1
R E D 1
A U T O 1
R E D 1
G R E N 1
R E D
O F F
O N
P H
O F F
1
000
01
0
P.L.
0
0
0
1
0
0
0
0
0
0
0
0
Setting range or alternatives
50/100/200/400/800/1000/200
0/5000
Selects Conversion rate (average times of internal sampling; sampling time: approx. 1 millisecond)
1/2/4/8/16/32
1/2/5/10
RED/GREEN
AUTO/MANU
RED/GREEN
RED/GREEN
RED/GREEN
OFF/LV1/LV2/LV3/ON
ON/OFF
PH/VH/PVH
OFF/ON
1/2/4/8
000 to 999
01 to 99
0 to 9
Major setting purpose and remarks
Selects the average times of movements.Filtering effects: Small<1(OFF)-2-4-8-16-32>Big
Selects the range of display changes to maintain image display consistency.
(If this parameter is set to 5, the lowest digit will display either 0 or 5 only).
Selects display colors.*Only when without meter relay.
Selects automatic (red in the HI or LO mode or green in the GO mode) or manual setting for display color type.
*Only when with meter relay.
Selects red or green as display color at the time of HI judgment.*Only when CLR.T is MANU.
Selects red or green as display color at the time of GO judgment.*Only when CLR.T is MANU.
Selects red or green as display color at the time of LO judgment.*Only when CLR.T is MANU.
Selects the display brightness. <bright OFF-LV1-LV2-LV3-ON >turned off
Selects whether the jog SW is used or not.*With the multi display unit only.
Selects a type that operates when the PH function is turned on (peak hold/valley hold/peak-valley hold).
Selects whether the digital zero value is backed up or not at the time when the unit is turned off.
Selects the number of patterns available for the pattern selection function.
Selects whether the tracking zero function is enabled/disabled and sets the correction time (setup value/conversion rate).
Sets the correction range of the tracking zero function.*Only when TR.T is set to a value other than 000.
Sets the time between the startup and actual start of measurements (setup value x 1 second).
P R O Protection level
U - N O .
Unit number indication
P V H . T PH type
D Z . C DZ control
P S . C P.SEL control
B C D . L
B A U D
BCD logic
Baud rate
D A T A Data length
P . B I T
S T P . B Stop bit
T -
Parity bit
Delimiter
A D R Equipment ID
L V . 1 3
O F F
A
S W
S W
0
0
0
0
N . L O G 0
9600
7
E
2
C R . L F
0 1
1
1
1
1
1
1
Lv.0/LV.1/LV.2/LV.3
OFF/ON
A/B
SW/TERM
SW/TERM
N.LOG/P.LOG
2400/4800/9600/19200/38400
7/8
E/N
1/1.5/2
CR.LF/CR
01 to 99
Selects the protection level to prevent operation mistakes.High< LV3-LV2-LV1-LV0 >Low
Selects whether the code of a unit mounted at the time of startup is displayed or not.
Selects an operation type of peak hold (A: real-time display; B: results display).*Only with the external control.
Selects a control method of Digital Zero (SW: front key; TERM: external control terminal).*Only with the external control.
Selects a control method of Pattern Selection (SW: front key; TERM: external control terminal).*Only with the external control.
Selects the BCD output logic (N: negative logic; P: positive logic).*Only when the BCD output is available.
Sets a baud rate for communication.*Only when the communication function is available.
Selects the data length for communication.*Only when the communication function is available.
Selects the parity bit for communication.*Only when the communication function is available.
Selects the stop bit for communication.*Only when the communication function is available.
Selects a delimiter for communication.*Only when the communication function is available.
Selects the equipment ID for the RS-485 function.*Only when the RS-485 function is available.
※Start/hold is applocable to type “A” only in temperature input units.
※Start/hold delay time cannot be set in temperature input units.
Instruction Manual for MODEL A6xx8 6/8
4.5.2
Scaling Data
Menu indication
R A N G
D L H I
D L L O
A . O U T
Parameter name
Input range
Digital limiter HI
Digital limiter LO
Analog output type
A O H I Analog output HI
A O L O Analog output LO
U N I T
Unit of temperature indication
Initial value
P.L.
KA 1
Setting range or alternatives
KA/KB/J/T/S/R/B
PA/JPA/PB/JPB
9999
-9999
0-1
9999
0
C
0
0
1
1
1
1
-9999 to +9999
-9999 to +9999
0-1/0-10/1-5/4-20
-9999 to +9999
-9999 to +9999
C/F
Major setting purpose and remarks
Selects an input range. *Note that input terminals vary depending on the actual input range.
Sets the upper limit of the displayable range. (Any value equal to or exceeding the digital limiter HI setting will not be updated and
kept at the setup value.
Sets the lower limit of the displayable range. (Any value equal to or below the digital limiter LO setting will not be updated and kept at
the setup value.
Selects an analog output range.*Only when the analog output is available.
Sets the relationship between indications and analog outputs.*Only when the analog output is available.
Unit of temperature indications is set.
4.5.3
Comparator Data (for meter relay only)
Menu indication
Parameter name
COM.T Comparator output type
HI-S HI judgment value
LO-S LO judgment value
N.VAL
Nominal value
ERR1 Error 1
HI-H HI hysteresis
LO-H LO hysteresis
ER1.H
HI-L
Error 1 hysteresis
HI logic
GO-L GO logic
Initial value
O/U
1000
500
5000
5.00
0
0
1
N.O
N.O
LO-L LO logic
5 Setting Examples
N.O
1
1
1
0
2
2
0
0
1
2
2
O/U/ERR
-9999 to +9999
-9999 to +9999
-9999 to +9999
0.00 to 99.99
0 to +999
0 to +999
0 to +999
N.O/N.C
N.O/N.C
N.O/N.C
Major setting purpose and remarks
Select either [above or below] or error comparator types.
Sets a HI judgment value.*Only when COM.T is O/U.
Sets a LO judgment value.*Only when COM.T is O/U.
Sets a nominal value.*Only when COM.T is ERR.
Sets an error.*Only when COM.T is ERR.
Sets a HI hysteresis value *Only when COM.T is O/U.
Sets a LO hysteresis value *Only when COM.T is O/U.
Sets an error hysteresis *Only when COM.T is ERR.
Sets a HI output logic (N.O.=normally open or N.C.=normally closed)*Output when the power is OFF is always open (OFF).
Sets a GO output logic (N.O.=normally open or N.C.=normally closed)*Output when the power is OFF is always open (OFF).
Sets a LO output logic (N.O.=normally open or N.C.=normally closed)*Output when the power is OFF is always open (OFF).
5.1 Comparator Data Setting Examples (for meter relay only)
E x a m p l e 1
A b o v e a n d b e l o w s e t u p m o d e
1 0 0 0
9 0 0
I n d i c a t i o n v a l u e s
H y s t e r e s i s r a n g e
E x a m p l e 2
E r r o r j u d g m e n t m o d e
1 1 0 0
7 0 0
H I j u d g m e n t v a l u e
1 0 0 0
5 0 0
4 0 0
9 0 0
L O j u d g m e n t v a l u e
2 0 0
0 0
J u d g m e n t G O H I G O L O G O
※The hysteresis is effective either in the range lower than the HI judgment value or
that higher than the LO judgment value.
J u d g m e n t G O H I
I n d i c a t i o n v a l u e s
G O L O
H y s t e r e s i s r a n g e
G O
U p p e r s i d e e r r o r
N o m i n a l v a l u e
L o w e r s i d e e r r o r
※Error is to be set as a percentage of the nominal value.
※The hysteresis is to be set as xx digits of the indication value.
※The hysteresis is effective either in the range lower than the upper side error or
that higher than the lower side error.
Instruction Manual for MODEL A6xx8 7/8
6 Specifications and External Dimensions
6.1 Input Specifications
●Temperature measurements
Range Sensor Measurement range
KA
KB
K
-50.0 to + 199.9 ℃
-50 to + 1200 ℃
J
T
J
T
-50 to + 1000 ℃
-50 to + 400 ℃
S S 0 to + 1700 ℃
R R -10 to + 1700 ℃
B B 100 to 1800 ℃
PA
JPA
Pt100Ω
JPt100Ω
-100.0 to + 199.9 ℃
PB
JPB
Pt100Ω
JPt100Ω
-100 to + 600 ℃
Sampling rate :
Error
±(0.5% of FS)
±(0.2% of FS)
±(0.6% of FS)
±(0.4% of FS)
±(0.15% of FS)
±(0.3% of FS)
Cold junction compensation error (10 to 40℃)
±2℃
ー
ー
RTD : Approx. 10 times/sec max. (The sampling speed is half of the configured.)
TC : Approx. 5 times/sec max. (The sampling speed is quarter of the configured.)
(1) “ERR1” may be indicated immediately after the power is turned on or immediately after the range is switched. This is not an abnormal occurrence, because it is caused by the zero-contact compensation circuit becoming temporarily unstable. The product will automatically recover after several samplings.
(2) If indications fluctuate heavily, decrease the sampling speed.
6.2 General Specifications
D i s p l a y
D i s p l a y r a n g e
O p e r a t i o n a l t e m p e r a t u r e
S t o r a g e t e m p e r a t u r e
P o w e r s u p p l y
P o w e r c o n s u m p t i o n
E x t e r n a l d i m e n s i o n
W e i g h t
W i t h s t a n d v o l t a g e
I n s u l a t i o n r e s i s t a n c e
:
M u l t i d i s p l a y
M a i n d i s p l a y: R e d/ g r e e n 7 - s e g m e n t d i s p l a y
( c h a r a c t e r h e i g h t: a p p r o x. 2 0 m m )
S u b d i s p l a y: R e d 7 - s e g m e n t d i s p l a y
( c h a r a c t e r h e i g h t: a p p r o x. 6 m m )
:
:
S i n g l e d i s p l a y
R e d/ g r e e n 7 - s e g m e n t d i s p l a y ( c h a r a c t e r h e i g h t: a p p r o x.
2 0 m m )
- 9 9 9 9 t o 9 9 9 9
0 t o 5 0 ℃ 3 5 t o 8 5 % R H
: - 1 0 t o 7 0 ℃ ; 6 0 % R H o r l e s s
: A C P S
A C 1 0 0 t o 2 4 0V ± 1 0 %
D C P S
D C 1 2 t o 4 8 V ± 1 0 %
: A C P S
M a x i m u m l o a d: A p p r o x. 8 V A a t 1 0 0 V A C
D C P S
M a x i m u m l o a d: A p p r o x. 7 W a t 2 4 V D C
: 4 8 m m ( H ) x 9 6 m m ( W ) x 9 7 . 5 m m
( w i t h n o D X o p t i o n u n i t m o u n t e d)
: A p p r o x. 4 5 0g
: A C P S
P S - i n p u t, o u t p u t: 1 , 5 0 0 V A C, 1 m i n u t e ( A C P S)
I n p u t- o u t p u t: 5 0 0 V D C, 1 m i n u t e
O u t p u t- o u t p u t: 5 0 0 V D C, 1 m i n u t e
C a s e- P S , i n p u t, o u t p u t: 1 , 5 0 0 V A C, 1 m i n u t e
D C P S
P S - i n p u t, o u t p u t: 5 0 0 V D C, 1 m i n u t e ( D C P S)
I n p u t- o u t p u t: 5 0 0 V D C, 1 m i n u t e
O u t p u t- o u t p u t: 5 0 0 V D C, 1 m i n u t e
C a s e- P S , i n p u t, o u t p u t: 1 , 5 0 0 V A C, 1 m i n u t e
: A C P S
A m o n g t h e a b o v e t e r m i n a l s: 5 0 0 V D C 1 0 0 M Ω o r m o r e
D C P S
A m o n g t h e a b o v e t e r m i n a l s: 5 0 0 V D C 1 0 0 M Ω o r m o r e
The PS voltage must be applied or shut down at once (not gradually).
Take at least a 10-second interval between a shutdown and startup.
6.3 External Dimensions
Enter Mode
Instruction Manual for MODEL A6xx8 8/8
7 Warranty and Service
7.1 W arranty
The manufacturer warrants to the original retail customer its A6000 series universal digital panel meter to be free of defects in material and workmanship for use under normal care and will repair or replace any meter at no charge to the customer during the one (1) year warranty period of the meter.
7.2 After Sales Service
Under strict quality control measures, this product was manufactured, tested, inspected and shipped. Should a defect in manufacture or workmanship be identified, please return the product to our distributor or directly to us. It would be highly appreciated if you could give a detailed account of the fault and enclose it with the product.
WATANABE ELECTRIC INDUSTRY CO., LTD.
WATANABE ELECTRIC INDUSTRY CO., LTD.
ASAHI KEIKI DIVISION
(UV-33624m)
Instruction Manual for MODEL A6xxx-x1 1/2
Instruction Manual for A6000 Series
Universal Digital Panel Meters
External Control Unit
1 Before Using the Unit
This instruction manual is for the external control unit of the A6000 series. Read this manual along with the main unit’s instruction manual before using the unit.
A6000's rear panel
Input unit
Comparator output unit
Option unit
None of the units are replaceable by the customer.
Power supply unit
2 Terminals and Connections
Top
①HOLD
②PH
③DZ
④R.RESET
⑤P.SEL0
⑥P.SEL1
⑦P.SEL2
⑧COM
⑨NC
⑩NC
Bottom
Terminal
S/H
Function
"Start/hold" control terminal. The “hold” state is turned on when this terminal is at the same potential as the
COM terminal or short-circuited.
PH
"Peak hold" control terminal. The peak hold function is turned on when this terminal is at the same potential as the COM terminal or short-circuited.
DZ
"Digital zero" control terminal. The digital zero function is turned on when this terminal is at the same potential as the COM terminal or short-circuited. This setting is valid only when the external control terminal is selected as the control method for digital zero using condition data.
R.RESET
"Relay reset control terminal. The relay reset function is turned on (only for the meter relay) when this terminal is at the same potential as the COM terminal or short-circuited.
P.SEL0
P.SEL1
P.SEL2
"Pattern selection" control terminals. Patterns can be selected by setting these terminals at the same potential as the COM terminal or short-circuiting them as shown in the table below. These settings are valid only when the external control terminal is selected as the control method for pattern selection using condition data.
P.SEL0
P.SEL1
P.SEL2
Pattern 1
Open
Pattern 2
Short-circuited
Open
Pattern 3
Open
Short-circuited
Open
Pattern 4
Short-circuited
Open
Pattern 5
Open
Short-circuited
Pattern 6
Short-circuited
Open
Short-circuited
Pattern 7
Open
Pattern 8
Short-circuited
Short-circuited
COM
External control’s common terminals, which are at the same potential as the LO or AG terminal for the input circuit.
3 Timing Charts
Start/hold type A
Internal sampling
Specified sampling
S/H terminal
Comparator output update
*1
*4
*3
*3 + time delay +(*1×2)+500μs+*4
Start/hold type B
Internal sampling
Specified sampling
*1
S/H terminal
Comparator output update
Time delay +(*1×2)+500μs+*4
*4
*1 Specified sampling
The sampling speed is determined by the AVG parameter setting of condition data as shown in the table below:
AVG setting
1
2
4
8
10
20
50
Sampling speed
1041.65 times/sec
520.825 times/sec
260.4125 times/sec
130.20625 times/sec
104.165 times/sec
52.0825 times/sec
20.833 times/sec
Sampling cycle
Approx. 960
s
Approx. 1.92 ms
Approx. 3.84 ms
Approx. 7.68 ms
Approx. 9.6 ms
Approx. 19.2 ms
Approx. 48 ms
AVG setting
100
200
400
800
1000
2000
5000
Sampling speed Sampling cycle
10.4165 times/sec
5.20825 times/sec
2.604125 times/sec
1.3020625 times/sec
1.04165 times/sec
0.520825 times/sec
0.20833 times/sec
Approx. 96 ms
Approx. 192 ms
Approx. 384 ms
Approx. 768 ms
Approx. 960 ms
Approx. 1.92 sec
Approx. 4.8 sec
※Using a temperature measuring unit, the sampling speed is half of the value in the table above when the RTD is selected and quarter of it when the TC is selected.
※It is different, and a frequency measurement unit is indication update time with a range.
Please refer to input specifications (Clause 6.1) of an instruction manual for frequency measurement.
*3 External start signal delay can be set for the external start using the S/H delay time parameter of the condition data.
*4 Time delay for comparator output
4 Specifications and External Dimensions
External control terminal
Internal circuit: Pulled up to approximately +5 V (through a 10 k
resistor).
Control signal’s HI level: 4.2 to 5 V relative to the COM terminal
Control signal’s LO level: 0 to 0.4 V relative to the COM terminal
The external control circuit’s COM terminals and the input circuit’s LO or AG terminal are set at the same DC potential.
Enter Mode
WATANABE ELECTRIC INDUSTRY CO., LTD.
UW-48222h
2015.11
Instruction Manual for MODEL A6xxx-x1 2/2
5 Warranty and Service
5.1 Warranty
The manufacturer grants a warranty to the original retail customer stating that its A6000 series universal digital panel meter is free of defects in material and workmanship for use under normal care, and will repair or replace the meter at no charge to the customer during the one
(1) year warranty period.
5.2 After Sales Service
This product was manufactured, tested, inspected and shipped under strict quality control measures. Should a defect in manufacture or workmanship be identified, please return the product to our distributor or directly to us. It would be highly appreciated if you could give a detailed account of the problem and enclose it with the product.
WATANABE ELECTRIC INDUSTRY CO., LTD.
WATANABE ELECTRIC INDUSTRY CO., LTD.
ASAHI KEIKI DIVISION
(UV-47964j)
Instruction Manual for MODEL A6xxx-x2 or A6xxx-x3 1/2
Instruction Manual for A6000 Series
Universal Digital Panel Meters
BCD Output and External Control Unit
1 Before Using the Unit
This instruction manual is for the BCD output and external control unit of the A6000 series. Read this manual along with the main unit’s instruction manual before using the unit.
A6000's rear panel
Input unit
Comparator output unit
Option unit
Power supply unit
None of the units are replaceable by the customer.
2 Terminals and Connections
Top
NC
NC
10 3
8
10
3
10
2
10
2
10
1
2
8
2
8
10
10
10
1
0
0
2
8
2
POL
P.C
D.COM
NC
PH
R.RESET
P.SEL1
COM
Bottom
NC
NC
10 3
4
10
3
10
2
10
2
10
1
1
4
1
4
10
10
10
1
0
0
1
4
1
OVER
ENABLE
D.COM
S/H
DZ
P.SEL0
P.SEL2
COM
Suitable connector (receptacle):
KEL Corporation’s 8822E-035-171-F
Terminal
10 0 1
10 0 2
10 0 4
10
0
8
Function
Least significant bit output of least significant BCD digit. The output logic can be switched using condition data.
Second least significant bit output of least significant BCD digit. The output logic can be switched using condition data.
Second most significant bit output of least significant BCD digit. The output logic can be switched using condition data.
Most significant bit output of least significant BCD digit. The output logic can be switched using condition data.
10
3
1
10
3
2
10 3 4
10 3 8
POL
OVER
P.C
Least significant bit output of most significant BCD digit. The output logic can be switched using condition data.
Second least significant bit output of most significant BCD digit. The output logic can be switched using condition data.
Second most significant bit output of most significant BCD digit. The output logic can be switched using condition data.
Most significant bit output of most significant BCD digit. The output logic can be switched using condition data.
BCD polarity output. The output logic can be switched using condition data.
BCD overflow output. The output logic can be switched using condition data.
BCD print instruction output.
ENABLE
"Enable BCD" control terminal. The BCD output is set to the high-impedance state or the transistor is turned off when this terminal is short-circuited or at the same potential as the D.COM terminal.
D.COM
BCD output•fs common terminals
S/H
"Start/hold" control terminal. The “hold” state is turned on when this terminal is at the same potential as the COM terminal or short-circuited.
PH
"Peak hold" control terminal. The peak hold function is turned on when this terminal is at the same potential as the COM terminal or short-circuited.
DZ
"Digital zero" control terminal. The digital zero function is turned on when this terminal is at the same potential as the
COM terminal or short-circuited. This setting is valid only when the external control terminal is selected as the control method for digital zero using condition data.
R.RESET
P.SEL0
P.SEL1
P.SEL2
COM
"Relay reset control terminal. The relay reset function is turned on (only for the meter relay) when this terminal is at the same potential as the COM terminal or short-circuited.
"Pattern selection" control terminals. Patterns can be selected by setting these terminals at the same potential as the COM terminal or short-circuiting them as shown in the table below. These settings are valid only when the external control terminal is selected as the control method for pattern selection using condition data.
P.SEL0
P.SEL1
P.SEL2
Pattern 1
Open
Pattern 2
Short-circuited
Open
Pattern 3
Open
Short-circuited
Open
Pattern 4
Shortcircuited
Open
Pattern 5
Open
Short-circuited
Pattern 6
Short-circuited
Open
Short-circuited
Pattern 7
Open
Shortcircuited
Pattern 8
Shortcircuited
External control’s common terminals, which are at the same potential as the LO or AG terminal for the input circuit.
External control terminal
Internal circuit: Pulled up to approximately +5 V (through a 10 k
resistor).
Control signal’s HI level: 4.2 to 5 V relative to the COM terminal
Control signal’s LO level: 0 to 0.4 V relative to the COM terminal
WATANABE ELECTRIC INDUSTRY CO., LTD.
3 Timing Charts
Start/hold type A
Internal sampling
Specified sampling
S/H terminal
*1
P.C
(BCD determined)
Comparator output
*2 update
*3
*3 + time delay + (*1x 2) + 500
s
*3 + time delay + (*1x 2) + 500
s + *4
*4
Start/hold type B
Internal sampling
Specified sampling
*1
S/H terminal
*2
Time delay + (*1x 2) + 500
s
P.C
(BCD determined)
Comparator output update
Time delay + (*1 x 2) + 500
s + *4
*4
*1 Specified sampling
The sampling speed is determined by the AVG parameter setting of condition data as shown in the table below:
AVG setting
1
2
4
8
10
20
50
1041.65 times/sec
520.825 times/sec
260.4125 times/sec
130.20625 times/sec Approx. 7.68 ms
20.833 times/sec
P.C
output width
Approx. 220
s
Approx. 0.25 ms
Approx. 5.8 ms
Approx. 14.4 ms
Approx. 960
s
Approx. 1.92 ms
Approx. 3.84 ms
Approx. 9.6 ms
Approx. 19.2 ms
Approx. 48 ms
100
800
5000
10.4165 times/sec
1.3020625 times/sec
0.20833 times/sec
Approx. 192 ms
Approx. 384 ms
Approx. 960 ms
Approx. 1.92 sec
※Using a temperature measuring unit, the sampling speed is half of the value in the table above when the RTD is selected and quarter of it when the TC is selected.
※It is different, and a frequency measurement unit is indication update time with a range.
Please refer to input specifications (Clause 6.1) of an instruction manual for frequency measurement.
*2 P.C output signal width
The output signal width is determined by the specified sampling setting as shown in the table below:
AVG setting
1
2
4
8
10
20
50
Sampling speed
104.165 times/sec
52.0825 times/sec
Approx. 1.15 ms
Approx. 2.3 ms
Approx. 2.9 ms
※The output width is always 30 ms for the frequency measurement unit.
*3 External start signal
Sampling cycle
AVG setting
100
200
400
800
1000
2000
5000
AVG setting
P.C
output width
200
400
1000
2000
Approx. 30 ms
Sampling speed
5.20825 times/sec
2.604125 times/sec
1.04165 times/sec
0.520825 times/sec
Sampling cycle
Approx. 96 ms
Approx. 768 ms
Approx. 4.8 sec delay can be set for the external start using the S/H delay time parameter of the condition data.
*4 Time delay for comparator output
4 Specifications and External Dimensions
BCD output
Output type : open collector or TTL-compatible
Polarity : The applicable transistor (open collector) or polarity bit (TTL) turns on when negative values are indicated.
Overflow : The applicable transistor (open collector) or bit (TTL) turns on when the overflow state is indicated.
Output logic : Can be switched alternatively (the logic for the P.C bit cannot be changed).
Output rating : 30 V DC and 10 mA (open collector), or fan-out 2 (TTL)
The BCD output logic can be switched using the BCD logic parameter (BCD.L) of the condition data, which is positive when BCD.L is set to P.LOG and negative when BCD.L is set to N.LOG for an output type of Open collector. This logic is reversed if the output type is "TTL."
UW-48220L
2015.11
Instruction Manual for MODEL A6xxx-x2 or A6xxx-x3 2/2
1) The BCD output circuit and the external control circuit are electrically isolated from each other (withstand voltage: 500 V DC for one minute).
2) The external control circuit's COM terminals and the input circuit's LO or AG terminal are set at the same DC potential.
5.1 Warranty
Enter Mode
5 Warranty and Service
91.5
102.5
96
The manufacturer grants a warranty to the original retail customer stating that its A6000 series universal digital panel meter is free of defects in material and workmanship for use under normal care, and will repair or replace the meter at no charge to the customer during the one
(1) year warranty period.
5.2 After Sales Service
This product was manufactured, tested, inspected and shipped under strict quality control measures. Should a defect in manufacture or workmanship be identified, please return the product to our distributor or directly to us. It would be highly appreciated if you could give a detailed account of the problem and enclose it with the product.
WATANABE ELECTRIC INDUSTRY CO., LTD.
WATANABE ELECTRIC INDUSTRY CO., LTD.
ASAHI KEIKI DIVISION
(UV-47983L)
Instruction Manual for MODEL A6xxx-x4 1/2
Instruction Manual for A6000 Series
Universal Digital Panel Meters
Analog Output and External Control Unit
1 Before Using the Unit
This instruction manual is for the analog output and external control unit of the A6000 series. Read this manual along with the main unit’s instruction manual before using the unit.
2 Terminals and Connections
A6000's rear panel
Input unit Comparator output unit Option unit Power supply unit
Terminal
S/H
PH
DZ
R.RESET
COM
A.OUT
V(+)
A.OUT
I(+)
A.OUT
COM(-)
None of the units are replaceable by the customer.
Top
①HOLD
②PH
③DZ
④R.RESET
⑤NC
⑥NC
⑦COM
⑧A.OUT V(+)
⑨A.OUT I(+)
⑩A.OUT COM(-)
Bottom
Function
"Start/hold" control terminal. The “hold” state is turned on when this terminal is at the same potential as the
COM terminal or short-circuited.
"Peak hold" control terminal. The peak hold function is turned on when this terminal is at the same potential as the COM terminal or short-circuited.
"Digital zero" control terminal. The digital zero function is turned on when this terminal is at the same potential as the COM terminal or short-circuited. This setting is valid only when the external control terminal is selected as the control method for digital zero using condition data.
"Relay reset control terminal. The relay reset function is turned on (only for the meter relay) when this terminal is at the same potential as the COM terminal or short-circuited.
External control’s common terminals, which are at the same potential as the LO or AG terminal for the input circuit.
Voltage output terminal of analog output.
Current output terminal of analog output.
Common terminal of analog output.
Insulated from input LO or AG terminal/external control COM terminal.
3 Timing Charts
Start/hold type A
Internal sampling
Specified sampling
S/H terminal
Comparator output update
*1
Start/hold type B
*4
*3
*3 + time delay +(*1×2)+500μs+*4
Internal sampling
Specified sampling
*1
S/H terminal
Comparator output update
Time delay +(*1×2)+500μs+*4
*4
*1 Specified sampling
The sampling speed is determined by the AVG parameter setting of condition data as shown in the table below:
AVG setting
Sampling speed Sampling cycle
AVG setting
Sampling speed Sampling cycle
1
2
4
8
10
20
1041.65 times/sec
520.825 times/sec
260.4125 times/sec Approx. 3.84 ms
130.20625 times/sec Approx. 7.68 ms
104.165 times/sec
52.0825 times/sec
Approx. 960
s
Approx. 1.92 ms
Approx. 9.6 ms
Approx. 19.2 ms
100
200
400
800
1000
2000
10.4165 times/sec
5.20825 times/sec
2.604125 times/sec
1.3020625 times/sec
1.04165 times/sec
0.520825 times/sec
Approx. 96 ms
Approx. 192 ms
Approx. 384 ms
Approx. 768 ms
Approx. 960 ms
Approx. 1.92 sec
50 20.833 times/sec the RTD is selected and quarter of it when the TC is selected.
※It is different, and a frequency measurement unit is indication update time with a range.
*3 External start signal
Approx. 48 ms 5000 0.20833 times/sec Approx. 4.8 sec
※Using a temperature measuring unit, the sampling speed is half of the value in the table above when
Please refer to input specifications (Clause 6.1) of an instruction manual for frequency measurement.
delay can be set for the external start using the S/H delay time parameter of the condition data.
*4 Time delay for alarm output
4 Specifications and External Dimensions
●Analog output (PWM)
Output type
0 to 1V
0 to 10V
1 to 5V
4 to 20mA
Conversion system
Resolution
Scaling
Internal circuit
Control signal’s HI level
Load resistance
10kΩ or more
●External control terminal
Control signal’s LO level
550Ω or less
: PWM conversion
: Digital scaling
:
Accuracy
±(0.5% of FS)
: Equivalent to 14 bits
Response speed
Approx. 0.5 second
: Pulled up to approximately +5V (through a 10kΩ resistor).
: 4.2 to 5V relative to the COM terminal
0 to 0.4V relative to the COM terminal
(1) Analog output circuits and external control circuits are insulated (withstand voltage 500V DC, one minute) electric potential at direct current.
Enter Mode
Ripple
50mVp-p
25mVp-p
(2) COM terminals of external control circuits and LO terminals of input circuits have the same
WATANABE ELECTRIC INDUSTRY CO., LTD.
UW-48252i
2015.11
problem and enclose it with the product.
Instruction Manual for MODEL A6xxx-x4 2/2
5 Warranty and Service
5.1 Warranty
The manufacturer grants a warranty to the original retail customer stating that its A6000 series universal digital panel meter is free of defects in material and workmanship for use under normal care, and will repair or replace the meter at no charge to the customer during the one
(1) year warranty period.
5.2 After Sales Service
This product was manufactured, tested, inspected and shipped under strict quality control measures. Should a defect in manufacture or workmanship be identified, please return the product to our distributor or directly to us. It would be highly appreciated if you could give a detailed account of the
WATANABE ELECTRIC INDUSTRY CO., LTD.