QA-30 External Pacemaker Analyzer

QA-30 External Pacemaker Analyzer
QA-30
External Pacemaker Analyzer
PRODUCT HIGHLIGHTS
• Tests Invasive and Non-invasive External
Pacemakers (Transthoracic and A-V Sequential).
• Measurements of Amplitude, Rate, Pulse Width.
• Tests Demand Sensitivity, Refractory Period,
and Succeptability to 50/60 Hz Interference.
• Small and portable.
• Very simple operation.
• 2 Line, 16 Character LCD Display.
• Battery Operation or AC Adapter.
OVERVIEW
The QA-30 External Pacemaker Analyzer is an
advanced microprocessor-based, compact, high
performance external pacemaker tester designed for accurate testing of all types of external pacemakers, including invasive and transthoracic. The device is menu-driven and simple
to operate.
All functions are set from a 2-line, 16-character
LCD display and a keypad with 8 tactile keys.
The QA-30 provides superior accuracy and
performance.
QA-30 Specifications
PULSE ACCURACY TESTS
DEMAND SENSITIVITY TEST
Ventricular Pacing Signal:
Amplitude:
0.1 to 25 mA
(500 ohm load)
Accuracy:
±10% 0.1-1 mA,
±5% 1-25 mA
Waveforms:
Atrial Pacing Signal:
Amplitude:
0.1 to 25 mA
(500 ohm load)
Accuracy:
±10% 0.1-1 mA,
±5% 1-25 mA
Pacer rate:
Amplitude:
Waveform width:
Square(SQR),
Triangle(TRI),
and Haversine (SSQ)
0 - 100 mV(500 ohm)
10, 25, 40, 100
and 200 ms
30 to 120 ppm
A-V Time Interval: 0 to 999 ms
Accuracy:
±1% or 1ms
IMMUNITY TEST
Transthoracic Pacing Signal:
Amplitude:
1 to 200 mA
(50 ohm load)
Accuracy:
±10 % 1-10 mA,
±5 % 10-200 mA
Rate:
Accuracy:
30 to 800 ppm
±1 % or 2 ppm
Pulse Width:
Accuracy:
0.5 to 80 ms
±10 % 0.5-5 ms,
±5 % 5-25 ms,
±1 % 25-80 ms
50/60 Hz Interference Signal:
0 to 102 mV in 6.8 mV steps in Ventricular/Atrial
measurements.
0 to 10.6 mV in 0.7 mV steps in Transthoracic
measurements.
Refractory Period Measurement:
20 to 500 ms (both Pacing and Sensing)
Accuracy:
±10 ms
Housing:
High impact ABS plastic case.
Dimensions:
D x W x H:
145 mm x 90 mm x 54 mm.
5.7 in x 3.6 in x 2.1 in
Weight:
0.8 kg/0.4 lbs with battery.
QA-30
Ordering Information
Order no:
16500:
16505:
Standard Accessories:
16510:
User and Service Manual QA-30
Optional Accessories:
GENERAL INFORMATION
16503:
16504:
17021:
Power:
17027:
9 Volt Alkaline Battery or 240
VAC (Optional Battery Eliminator).
QA-30 External Pacemaker Analyzer
QA-30 External Pacemaker Analyzer
with RS-232
RS-232 Interface
Carrying Case
230 V AC to 9 V DC Battery Eliminator
-or115 V AC to 9 V DC Battery Eliminator
Note: Specifications may be subject
to change without notice.
www.metron-biomed.com
USA
1345 Monroe NW, Suite 255A
Grand Rapids, MI 49505
Phone: (+1) 888 863-8766
Fax:
(+1) 616 454-3350
E-mail: support.us@metron-biomed.com
FRANCE
30, rue Paul Claudel
91000 Evry, France
Phone: (+33) 1 6078 8899
Fax:
(+33) 1 6078 6839
E-mail: info@metron.fr
NORWAY
Travbaneveien 1
N-7044 Trondheim, Norway
Phone: (+47) 7382 8500
Fax:
(+47) 7391 7009
E-mail: support@metron.no
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project

Download PDF