IET LABS PRS-300 Programmable Decade Resistance Substituter Operation Manual
Below you will find brief information for Programmable Decade Resistance Substituter PRS-300 PRS-300. The PRS-300 is a high precision manual and programmable decade substituter. It provides direct resistance substitution as well as RTD (Resistance Temperature Detector) simulation. It offers a wide choice of ranges, power, voltage, and accuracies for a wide variety of design and production applications.
Advertisement
Advertisement
♦
PRECISION INSTRUMENTS FOR TEST AND MEASUREMENT
♦
PRS-300 SERIES
Programmable
Decade Resistance Substituters
(IEEE-488.2, Ethernet and USB)
Operation Manual
IET LABS, INC.
www.ietlabs.com
TEL: (516) 334-5959 • (800) 899-8438 • FAX: (516) 334-5988
IET LABS, INC.
www.ietlabs.com
TEL: (516) 334-5959 • (800) 899-8438 • FAX: (516) 334-5988
♦
PRECISION INSTRUMENTS FOR TEST AND MEASUREMENT
♦
PRS-300 SERIES
Programmable
Decade Resistance Substituters
(IEEE-488.2, Ethernet and USB)
Operation Manual
Copyright
©
2016 IET Labs, Inc.
PRS-300 IM / October 2016
IET LABS, INC.
www.ietlabs.com
TEL: (516) 334-5959 • (800) 899-8438 • FAX: (516) 334-5988
♦
PRECISION INSTRUMENTS FOR TEST AND MEASUREMENT
♦
IET LABS, INC.
www.ietlabs.com
Email: [email protected]
TEL: (516) 334-5959 • FAX: (516) 334-5988
PRS-300 Series
WARRANTY
We warrant that this product is free from defects in material and workmanship and, when properly used, will perform in accordance with applicable IET speci fi cations. If within one year after original shipment, it is found not to meet this standard, it will be repaired or, at the option of IET, replaced at no charge when returned to IET. Changes in this product not approved by IET or application of voltages or currents greater than those allowed by the speci fi cations shall void this warranty. IET shall not be liable for any indirect, special, or consequential damages, even if notice has been given to the possibility of such damages.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED,
INCLUDING BUT NOT LIMITED TO, ANY IMPLIED WARRANTY OF MERCHANTABILITY OR
FITNESS FOR ANY PARTICULAR PURPOSE.
i
ii
PRS-300 Series
WARNING
OBSERVE ALL SAFETY RULES
WHEN WORKING WITH HIGH VOLTAGES OR LINE VOLTAGES.
Dangerous voltages may be present inside this instrument. Do not open the case
Refer servicing to quali
fi
ed personnel
HIGH VOLTAGES MAY BE PRESENT AT THE TERMINALS OF THIS INSTRUMENT
WHENEVER HAZARDOUS VOLTAGES (> 45 V) ARE USED, TAKE ALL MEASURES TO
AVOID ACCIDENTAL CONTACT WITH ANY LIVE COMPONENTS.
USE MAXIMUM INSULATION AND MINIMIZE THE USE OF BARE
CONDUCTORS WHEN USING THIS INSTRUMENT.
Use extreme caution when working with bare conductors or bus bars.
WHEN WORKING WITH HIGH VOLTAGES, POST WARNING SIGNS AND
KEEP UNREQUIRED PERSONNEL SAFELY AWAY.
CAUTION
DO NOT APPLY ANY VOLTAGES OR CURRENTS TO THE TERMINALS OF THIS
INSTRUMENT IN EXCESS OF THE MAXIMUM LIMITS INDICATED ON
THE FRONT PANEL OR THE OPERATING GUIDE LABEL.
CONTENTS
WARRANTY ........................................................................................................... i
WARNING ............................................................................................................. ii
CAUTION .............................................................................................................. ii
CONTENTS .......................................................................................................... iii
FIGURES .............................................................................................................. iv
TABLES ............................................................................................................... iv
Chapter 1: INTRODUCTION ................................................................................ 1
Chapter 2: SPECIFICATIONS .............................................................................. 3
2.1 Model Con fi guration and Speci fi cations ........................................................................3
2.2 General Speci fi cations ...................................................................................................4
2.2 Operating Guide .............................................................................................................5
Chapter 3: OPERATION ....................................................................................... 6
3.1 Initial inspection and setup ............................................................................................6
3.2 Connection .....................................................................................................................6
3.2.1 General Considerations ........................................................................................6
3.2.2 Electrical Considerations ......................................................................................6
3.3 Condensed Operating Instructions .................................................................................6
3.4 Menu Structure ...............................................................................................................7
3.4.1 Main Menu ...........................................................................................................8
3.4.2 Setting a Resistance Value ....................................................................................8
3.4.3 Menu Button .........................................................................................................8
3.4.4 Standard Values Screen ........................................................................................8
3.4.6 Increment INCR Key ............................................................................................9
3.4.7 OPEN Key ............................................................................................................9
3.4.8 STO,RCL Keys ...................................................................................................10
3.4.10 2 and 4 Wire Connections ................................................................................10
3.4.11 Thermal emf Considerations ............................................................................10
3.4.12 Environmental Conditions ................................................................................11
3.4.13 Remote Operation .............................................................................................11
Chapter 4: USB Interface .................................................................................. 12
4.1 Introduction ..................................................................................................................12
4.2 Capabilities ..................................................................................................................12
4.3 USB Operation .............................................................................................................12
PRS-300 Series
Table of Contents iii
iv
PRS-300 Series
Chapter 5: ETHERNET INTERFACE OPTION ................................................... 14
5.1 Ethernet Programming .................................................................................................14
5.2 Network Setup .............................................................................................................14
5.3 Ethernet Test Keyboard ................................................................................................14
5.4 Network Con fi guration ................................................................................................15
5.5 Web Browser Con fi guration .........................................................................................15
5.6 VXI Con fi guration Utility ............................................................................................17
5.7 Resetting Default Network Settings .............................................................................19
5.8 PRS Programming .......................................................................................................19
Chapter 6: IEEE INTERFACE OPTION .............................................................. 20
6.1 Introduction ..................................................................................................................20
6.2 Capabilities ..................................................................................................................20
6.3 Address Switch and Communications Settings ............................................................20
6.4 IEEE Option Operation ................................................................................................20
Chapter 7: PROGRAMMING ................................................................... 22
7.1 Introduction
......................................................................................... 22
Chapter 8: MAINTENANCE .................................................................... 23
8.1 Veri fi cation of Performance
.................................................................... 23
8.1.1 Calibration Interval
........................................................................ 23
8.1.2 General Considerations
.................................................................. 23
8.1.3 Calibration Procedure
.................................................................... 23
8.1.4 Adjustment of Internal Resistors
..................................................... 23
APPENDIX A: SCPI COMMAND REFERENCE ...................................... 25
APPENDIX B: IEEE.2 COMMON COMMANDS ...................................... 29
Table of Contents
PRS-300 Series
Chapter 1
INTRODUCTION
The PRS-300 (Figure 1.1) is a high precision manual and programmable decade substituter. The PRS-300 series provide direct resistance substitution as well as RTD (Resistance Temperature Detector) simulation.
The PRS-300 substituter is a precision resistance source with excellent characteristics of stability, temperature coef fi cient, and power coef fi cient. These features combined with a low virtually constant “zero resistance” make for very versatile instruments.
Operation is both local using convenient display and keypad and remote standard USB or optional Ethernet, and IEEE-488.2 interfaces. An RS-232 interface is also available upon request.
The PRS-300 employs very low resistance, low thermal emf relays with gold-clad silver-alloy contacts. A special design keeps contact resistance to a minimum. The gold plating keeps the silver contacts from becoming tarnished when unused, or when only low currents are passed through them. This is most often the case when only minute test currents are drawn by digital multimeters and other test instruments. Contact resistance remains low and repeatable.
High-quality gold-plated tellurium-copper fi ve-way binding posts serve to minimize the thermal emf effects, which would produce errors in dc resistance measurements.
With a resolution as low as 1 μ
Ω
and a maximum available resistance of 20 M
Ω
, the
PRS-300 may be used for exacting precision measurement applications requiring high accuracy, good stability, and low zero resistance.
The PRS-300 is suited for automatic and manual calibration and testing, simulation of
RTD’s, programmable loads, and many other laboratory and industrial applications.
The PRS Series may be rack mounted to serve as components in measurement and control systems.
The PRS-300 (Figure 1.1) is part of the
PRS family of high precision manual and programmable decade substituters. This series
Figure 1.1: High Precision Programmable Decade
Resistance Substituter
INTRODUCTION
1
2
PRS-300 Series
offers a wide choice of ranges, power, voltage, and accuracies for a wide variety of design and production applications.
The PRS-300 uses a microcontroller design to enhance the ease of use and calibration. It provides direct resistance substitution as well as RTD (Resistance Temperature Detector) simulation. It allows for higher precision and tighter tolerances.
The PRS-300 is a precision resistance source with excellent characteristics of accuracy, stability, temperature coef fi cient, and power coeffi cient. It is direct reading without the need to subtract “zero resistance.”
The PRS-300 offers the option of true 4-wire
Kelvin measurement for calibration applications and 2-wire mode for use in series as a substitute resistor and for RTD applications.
The front panel display indicates the mode selected: 2-wire or 4-wire.
There are two pre-programmed RTD resistance tables, and others may be added by the user.
The controller allows for other convenient programmed features.
The unique design makes adjustment of the
PRS-300 for calibration semiautomatic. An internal calibration routine guides the user to perform measurements using
only
a precision ohmmeter and enter the measured values into the PRS-300. No other standards are required.
IET Labs offers the PRS-DMM software which works with the PRS-300 and Fluke
8508A or Keysight 3458A to automate this adjustment process and eliminate the need for a technician to manual enter the resistance values manually into the PRS-300
INTRODUCTION
PRS-300 Series
Chapter 2
SPECIFICATIONS
For convenience to the user, the pertinent speci fi cations are given in an OPERATING
GUIDE, shown typically in Figure 2.1, af fi xed to the case of the instrument.
2.1 Specifi cations
User interface:
Numeric keypad to enter resistance value with digital display.
Accuracy:
±(70 ppm + 1 m Ω ) 2 and 4 Terminal
Minimum setting:
0.1 Ω
Resolution:
1 μΩ or 6 digits
Range:
0.1 Ω - 20 M Ω
Stability:
±
50 ppm/year
Thermal emf:
< 15 μ V
Maximum Load:
2 A, 200 V (peak), 0.5 W whichever applies fi rst
Resistors:
Precision wire-wound and metal foil
RTD Tables:
9 RTD tables can be entered into memory to allow user selection of temperature and the correct value of resistance will automatically be programmed.
PT-100 and PT-1000 tables for both Fahrenheit and
Celsius are pre-programmed into memory locations
1 to 4.
Adjustment:
Automatic adjustment procedure utilizing a high precision DMM eliminates the requirement for manual trimming of resistors.
AC Frequency Response:
Residual Capacitance Terminals to GND: < 850 pF
Resistance
0.1 Ω -10 k Ω
10 - 100 k Ω
100 k Ω - 1 M Ω
Typical ac/dc difference @ 1 kHz
<100 ppm
<200 ppm
<1%
1 - 20 M Ω <20%
Terminals
:
Four low-emf, gold-plated, tellurium-cop per 5-way bind ing posts are used for
HI
and
LO
ter mi nal pairs for
CUR RENT
and
SENSE
.
GND
bind ing post is con nect ed to the case, to the chassis ground and to the earth ground.
Switching time:
<1 second per change
Power requirements:
90 - 264 Vac , 47 - 63 Hz. 30 Watts Max.
Dimensions:
Bench model: 43 cm W x 8.9 cm H x 33 cm D (17” x
3.5” x 13”)
in front of panel
: 3.8 cm (1.5”).
Weight:
Bench model: 5.5 kg (12 lb); weight speci fi cations are nominal.
Remote Control:
USB:
USB Type B connector standard on rear panel
Remote Control Options:
IEEE:
GPIB standard 24 pin connector, conforms to IEEE-
488.2; SCPI 1994.0 command set; Hardware or software con fi gurable addressing range of 1 to 30.
Ethernet:
IEEE 802.3 compliant, Speeds 10 BaseT
(10 Mb/s) and 100 BaseT (100 Mb/s), IP Address
Static or DHCP, Factory setting 192.168.0.254 static
SPECIFICATIONS
3
PRS-300 Series
2.2 General Specifi cations
Warmup:
This unit utilizes custom designed low thermal emf relays for high stability and low-level signal applications. This thermal voltage is nominally less than 10
μ
V under laboratory conditions. Please note that a change in this thermal emf will give a false drift in resistance as measured by a dc ohmmeter, since the ohmmeter will interpret this emf as an effective resistance.
The worst case such effective drift after warm-up is less than ±50 ppm, primarily due to stabilization of this emf at the relay contacts as settings are changed.
Unit should be allowed to stabilize for one hour. For maximum stability, allow unit to warm-up over two hours.
Recommended Calibration Interval:
12 months.
4 SPECIFICATIONS
PRS-300 Series
PRS-300 PROGRAMMABLE DECADE RESISTANCE SUBSTITUTER
CONSULT INSTRUCTION MANUAL FOR PROPER INSTRUMENT OPERATION
Accuracy:
±(70 ppm + 1 m Ω ) 4 Terminal
Switching time:
<1 second per change
Minimum setting:
0.1 Ω
Power requirements:
Resolution:
1 μΩ
90 - 264 Vac , 47 - 63 Hz. 30 Watts Max.
Range:
0.1 Ω - 20 M Ω
Stability: ±
50 ppm/year
Thermal emf:
< 15 μ V
Dimensions:
Bench model: 43 cm W x 8.9 cm H x 33 cm D (17" x 3.5" x 13")
in front of panel
:
3.8 cm (1.5").
Weight:
Bench model: 5.5 kg (12 lb) weight speci fi cations are nominal.
Maximum Load:
2 A, 200 V (peak), 0.5 W whichever applies fi rst
Resistors:
Precision wire-wound and metal foil
Remote Control:
RTD Tables:
USB:
Standard USB Type B connector on rear panel, uses FTDI driver
3 RTD tables can be entered into memory to allow user selection of temperature and the correct value of resistance will automatically be programmed. PT-100 and PT-1000 tables are pre-programmed into memory.
Remote Control Options:
IEEE:
GPIB standard 24 pin connector, conforms to IEEE-488.2; SCPI 1994.0 command set; Hardware or software con fi gurable addressing range of 1 to 30.
Adjustment:
Automatic adjustment procedure utilizing a high precision DMM eliminates the requirement for manual trimming of resistors.
E t h e r n e t :
I E E E 8 0 2 . 3 c o m p l i a n t , S p e e d s 1 0 B a s e T
(10 Mb/s) and 100 BaseT (100 Mb/s), IP Address Static or DHCP, Factory setting
192.168.0.254 static
Terminals
:
Four low-emf, gold-plated, tellurium-cop per 5-way bind ing posts are used for
HI
and
LO
ter mi nal pairs for
CUR RENT
and
SENSE
.
GND
bind ing post is con-
nect ed to the case, to the chassis ground and to the earth ground.
MODEL: PRS-300 SN: E1-1634626
WARNING
Observe all safety rules when working with high voltages or line voltages. Connect the (GND) terminal to earth ground in order to maintain the case at a safe voltage.
Whenever hazardous voltages (>45 V) are used, take all measures to avoid accidental contact with any live components: a) Use maximum insulation and minimize the use of bare conductors. b) Remove power when adjusting switches. c) Post warning signs and keep personnel safely away.
IET LABS, INC.
CAGE CODE: 62015
• Long Island, New York
• Tel: (516) 334-5959 • Email: [email protected]
www.ietlabs.com
PRS-300 Nov. 2016
FIGURE 2.1 Typical OPERATING GUIDE Af fi xed to Unit
(Please see label af fi xed to your unit)
SPECIFICATIONS
5
PRS-300 Series
6
Chapter 3
OPERATION
3.1 Initial inspection and setup
This instrument was tested and carefully inspected before shipment. It should be in proper electrical and mechanical order upon receipt.
An
OPERATING GUIDE
is attached to the case of the instrument to provide ready reference to speci fi cations.
Mount the unit in a standard 19” rack if the rack mount option is speci fi ed.
3.2 Connection
3.2.1 General Considerations
The PRS-300 Decade Resistor utilizes a
5-terminal connection. The binding posts are standard laboratory type and readily accept banana plugs, spade lugs, alligator clips, and bare wire. Binding posts are located on the front panel of the instrument unless speci fi cally ordered with a Rear Output option.
The Kelvin terminals consisting of a
CURRENT
and a
SENSE
pair, each labeled
HI
and
LO
. These minimize contact resistance.
The
GND
terminal on all models is connected to the case and to earth and chassis grounds. This may be used as a shield terminal.
3.2.2 Electrical Considerations
The performance of the PRS-300 is directly affected by the quality of the connection to the system under test. This is particularly true with the precision series models having higheraccuracy and/or lower-impedance decades.
For optimum performance, contact resistance should be kept to a minimum by using the most substantial mating connection possible, and by assuring that the connection is well secured to the binding posts.
3.3 Condensed Operating Instructions
1. Turn on the PRS-300 ON using POWER
SWITCH
2. Select a desired resistance value in one of the following ways: a) Enter a value using the number keys, b) Press a Quick Value key (
A
– B) to c) Press
3. then press
ENTER
.
display a user-de fi ned value or a factory default value.
Default values are: k Ω
then a number key ( to display a user-de tory default value.
Default values are: k Ω │
2
- 2 k Ω
A
fi
0
: 10 k
- 100
Ω
Ω
│
│
B
0 – 9
1
: 100
- 1
To prevent possible damage to the PRS-300, set the maximum
)
ned value or a fac-
OPERATION
PRS-300 Series
applied voltage by pressing
VOLT
, entering the voltage on the keypad, then pressing
ENTER
. (Default: 0.5 V)
The fi rmware sets the minimum selectable resistance value so that the power does not exceed 1 W. To ensure that the whole range of resistances is selectable,
VOLT
setting must be ≤ 1 V.
Rmin is limited to >V
2
, where V is the
VOLT setting.
4. Select 2-wire or 4-wire mode by pressing the
MENU
button, selecting
CON-
FIGURE
, pressing the “2” will change between 2-wire and 4-wire modes
In 2-wire mode shorting links must be installed across high and low terminal pairs.
In 4-wire remove the shorting links.
5. Connect to the circuit or device under test.
To isolate the PRS-300 from the application, press
OPEN
to open the circuit. To close the circuit, press
OPEN
again.
3.4 Menu Structure
The main table below show menu structure of the PRS-300 and basic function of each button and menu selection.
Key Description Function
Basic operation enter
enter button Sets resistance value
0-9, 000/(-)
Used to enter numerical values. To select desired numerical value, press value as required.
(-)
is used with tables for negative temperature
◄
Deletes the last character from the display.
To delete the last character, press
◄
.
Options
MENU
INCR
Displays the following submenu:
1. STANDARD VALUES
1: 1% STD VALUES
2: 5% STD VALUES
3: 10% STD VALUES
4: RATIOMETRIC
2. CONFIGURE
1: LCD CONTRAST
2: 2W / 4W MODE
3. CALIBRATE
4. VERSION
Press
MENU
to select one of the following:
1. STANDARD VALUES - selects increments for UP and DOWN keys
2. CONFIGURE - allows user to adjust display contrast and change between 2-wire and 4-wire modes
3. CALIBRATE - see Section 8
4. VERSION - Displays current fi rmware version and release date.
Selects the increments for the
UP and DOWN keys
Default value: Ratiometric (0.8% of displayed value)
To adjust increments:
1. Press .
2. Enter desired increment value on the keypad.
3. Press to save.
Storing and recalling resistance values
STO
Stores a resistance value in memory locations A-D and 0-9.
To store a value, press
STO
and then press the desired location button A-D or 0-9.
A, B
RCL
Recalls a stored resistance value.
Default values:
A
: 10 k Ω
B
: 100 k Ω
Recalls a stored resistance value from memory location 0-9.
To change stored values:
1. Enter a resistance value
2. Press and then
A
,
B
To recall a stored value, press
RCL
, then press the desired location button (0-9)
Limiting Current
VOLT
Sets voltage to limit the current in order to protect the PRS-300 from possible damage.
Default value: 0.5 V
To adjust voltage:
1. Press .
2. Enter desired voltage on the keypad.
ENTER
to save.
Electrical Isolation
OPEN
Isolates the connection terminals To open the circuit, press
OPEN
.
by establishing an open circuit.
To close, press
OPEN
again.
Option: Simulating RTD’s and transducers
TABLE
Loads up to 9 stored data tables for simulating RTD’s or transducers.
To select the desired table, press
TABLE
to cycle through the three options.
OPERATION 7
8
PRS-300 Series
3.4.1 Main Display
The main display shows the resistance value in ohms or temperature value if tables are used. The main display also shows if the unit is con fi gured for
4W
(4-wire mode) or
2W
(2-wire mode) in the upper left corner of the display.
3.4.4 Standard Values Screen
There are 4 selections available in the
STNADARD VALUES
menu. The operation of the standard values allows the PRS-300 to be used as an automated resistance carousel with built-in EIA “preferred value” resistance tables of 1% (E96), 5% (E24), 10% (E12) increments or RATIOMETRIC user speci fi ed increment.
Figure 3-1 Main Display showing 4-Wire
See section 3.2.3 for more information on
2-wire and 4-wire connections.
3.4.2 Setting a Resistance Value
To set a resistance value use the numeric keypad to enter the value followed by the
ENTER
button. The
000
button can be used to increase the resistance value by a factor of
1000.
The
UP
and
DOWN
keys can also be used to increment or decrement the resistance value. See 3.4.6
INCR
Key for more information.
3.4.3 Menu Button
Pressing the
MENU
button will show four selections. See Figure 3-2
Figure 3-2 Menu Screen
Press the numerical value 1 to 4 on the keypad for the desired selection.
Press the
MENU
button to cancel and return to previous screen at anytime.
Figure 3-3 Standard Values Menu
Press the numerical value 1 to 4 on the keypad for the desired standard values.
Resistors are available in a number of standard ranges, often called ‘preferred values’ These ranges, or series, are set out by the
Electronic Industries Association (EIA), and common values are E12, E24, E48, and E96.
The number after the ‘E’ denotes the number of values the series contains per decade.
The series are logarithmic and are derived from the resistor tolerance; resistors with a tighter tolerance can have more values in the series that won’t overlap one another. The series are sometimes referred to by the tolerance, the two being related as follows:
E12: 10% tolerance
E24: 5% tolerance
E48: 2% tolerance
E96: 1% tolerance
When a
STANDARD VALUE
is selected on the PRS-300 pressing the
UP
or
DOWN
arrows will increment the resistance value to the
OPERATION
9
PRS-300 Series
next “preferred” resistance value.
RATIOMETRIC allows a custom tolerance value to be entered by the user. The default value is 0.8% but can be set to the required value in %. heit.
TABLE
key allows use of tables. There are 9 tables that can be selected.
The fi
3.4.5 Tables Key
Pressing the
rst 4 tables, 1 to 4, are de
C, PT-100 F, PT-1000 C and PT-1000 F where
C is degrees Celcius and F is degrees Fahren-
Press
TABLE
table as shown in Figure 3-5 fi ned as PT-100
key and then a number key
0 through 9 followed by the select a table. Table 0 is de fi used and resistance is shown.
ENTER
Figure 3-4 No translation Table
key will
ned as no table is
Press the numerical value 1 to 9 on the keypad for the desired table to be used. For example pressing
TABLE
key, then “1” and
ENTER
key will select use of the PT-100 C
Figure 3-5 Table 1 PT-100 C
Once the table is selected the display will change showing temperature on top and resistance on bottom. Any value entered via front panel or remotely will be assumed to be a table value rather than a resistance value. For example if 0.25 is entered then the screen will look like Figure 3-5 showing temperature on top line and resistance on the bottom line.
Figure 3-5 PR-100 C set to 0.25 °C
For negative temperatures press the 000/
(-) key prior to entering the value.
If a value is not listed in the table and is entered using the numeric keypad or selected using the
UP
and
DOWN
keys interpolation is used between the two closest table values.
Custom tables can be entered in locations
5 to 9 by using the USB interface and remote
CONIG:TABLE
commands. See Appendix B for more information.
3.4.6 Increment INCR Key
The
INCR
key allows entry of a numeric value. The numeric value representing the step or increment size the
UP
and
DOWN
keys use.
For example pressing
INCR
key and then
10 would result in the displayed value being incremented or decremented 10 Ω for each time the
UP
or
DOWN
key is pressed.
INCR
can also be used for tables in the same way so that the operator can press the
UP
or
DOWN
key and increment or decrement the displayed value in °C or °F.
3.4.7 OPEN Key
The
OPEN
key opens all relays in the
PRS-300 resulting in high resistance at the resistance terminals. The display will shown
OPEN
Ω
. Pressing the
OPEN
key again will cause the resistance to return to the previous set value.
For short enter lowest value of 0.1 Ω us-
OPERATION
10
PRS-300 Series
ing keypad or remote.
3.4.8 STO, RCL Keys
The
STO
and
RCL
key are used to store user de fi ned values into memory. Press
STO
key followed by 0 to 9 in the numeric keypad stores the currently shown resistance value into that memory location.
Press
RCL
key followed by 0 to 9 in the numeric keypad recalls the resistance value saved into memory to the currently displayed value.
3.4.9 VOLT Key
The
VOLT
key should normally be set to 0.5 V which allows us of the full resistance range of the PRS-300. Pressing the
VOLT
key followed by a numeric value representing the maximum voltage being used in the applicaton will limit the lowest resistance that can be entered to prevent damage to the PRS-300.
The limited is based upon a maximum of
1 W being applied to the resistors. Note that
0.5 W is the maximum recommended and operation at 1 W can cause the internal resistors to drift beyond speci fi cation.
3.4.10 2 and 4 Wire Kelvin Lead Connections
2-wire
or
4-wire
operation can be selected from by pressing the
MENU
button, selecting
CONFIGURE
, pressing the “2” will change between 2-wire and 4-wire modes.
The use of 4W with Kelvin leads minimize the effects of contact resistance and gives best performance. In 4-wire remove the shorting links from the high and low terminal pairs.
Figure 3-6 4-Terminal shorting links not connected
2W mode should be used when connecting to a 2 wire device or if the PRS-300 will be used as a standard 2-terminal resistor such as
RTD simulation. In 2-wire mode shorting links must be installed across high and low terminal pairs.
Figure 3-7 Two terminal with shorting links
3.4.11 Thermal emf Considerations
The PRS Series uses high-quality, lowemf components. Thermal emf is primarily attributable to the temperature difference between the leads of the relay and the contacts when temperature is applied to the coil. This emf is of the order of 5 μ V per relay, but is not usually additive. The typical worst case is <15
μ
V.
If the effect of tens of microvolts is signi fi cant to your application, connect to the instrument with low-thermal-emf materials only. Copper wire and copper alloys are recommended; brass and steel should be avoided.
Tinned copper and solder are acceptable.
OPERATION
11
PRS-300 Series
This emf will
not
be re fl ected if an ac measurement instrument is employed, and can be eliminated by using a meter with “True
Ohm” capability. In other cases, the emf may represent a very small component of a dc resistance measurement.
3.5 Environmental Conditions
The PRS is built, calibrated and intended for use in a laboratory environment with a nominal ambient temperature near 23 o
C. The accuracy of the unit may be affected when operated in non-laboratory environments. Always allow the instrument to stabilize at room temperature after unpacking or relocating the instrument. Humidity should be maintained at laboratory conditions.
3.6 Remote Operation
When the PRS is in
REMOTE
with front panel locked out an REM indication will be shown in the top right of the display.
OPERATION
PRS-300 Series
Chapter 4
USB INTERFACE
4.1 Introduction
The standard USB connection is via a standard USB B female connector located on the rear panel. This can be connected to a
PC via a standard USB 2.0 A Male to B Male
Cable.
Cable
Figure 4-1 USB 2.0 A Male to B Male
The SCPI standard provides a tree like series of standard commands for programmable instruments so that similar instruments by different manufacturers can be controlled by the same program.
4.2 Capabilities
The USB interface provides remote control over all functions except
POWER
.
4.3 USB Operation
The USB uses the command set outlined in Appendix B.
The standard USB interface uses a standard FTDI chip set.
Normally when the PRS-300 is connected to a computer using the USB port, Windows
7 and higher will search for the correct driver and as long as the computer is connected to the internet the driver should automatically be installed.
If the driver is not automatically installed the drivers can be installed from http://www.
ftdichip.com/Drivers/VCP.htm
USB INTERFACE 12
PRS-300 Series
13 USB INTERFACE
PRS-300 Series
Chapter 5
ETHERNET INTERFACE OPTION
5.1 Ethernet Programming
PRS units, with Ethernet interface option, come with Ethernet remote programming capability accessed through the RJ45 LAN connector. The connector is a RJ45 industrystandard connector found on the rear panel.
The PRS’s Ethernet connection is based upon the ICS Ethernet/GPIB/USB to Serial Interface
Card. More information can be found at www.
icselect.com on the 9006 Ethernet/GPIB/USB to Serial Interface Card.
Ethernet Protocols
The PRS supports Raw Socket and the
VXI-11 protocol which makes it easy to control from a PC or over the company network.
Raw socket lets you telnet to it and is best used with direct PC to instrument applications.
VXI-11 operates over Sun RPC and is a more secure protocol that mimics GPIB control of an instrument and provides secure communication over a company network or over the Internet.
Raw Socket Operation
The PRS is always ready for a telnet connection. When connected, the PRS outputs its IDN message to confirm the connection.
All raw socket messages are terminated with a linefeed. Carriage returns and ignored. A backspace character causes the prior character to be deleted. Communication timeout is two minutes and it is recommended that the client issue a
Space-BS sequence on an occasional basis, less than the timeout, to reset the timeout counter.
VXI-11 Background
VXI-11 is a communication standard developed in conjunction with the VISA Speci fi cation.
The speci fi cation de fi nes a VXI-11.3 interface like the PRS as an instrument which can be controlled in Windows systems by programs that make VISA or SICL library calls and in
UNIX/LINUX or similar operating system with
RPC calls. The VXI-11 speci fi cation provides an RPCL (Remote Procedure Call Library) that can be used by virtually any operating system to control the PRS. Microsoft’s RPC is not ONC compliant and cannot be used with a VXI-11.3 device like the PRS. An VXI-11.3 interface like the PRS will operate in an LXl system and it supports LXI’s ‘VXI Discovery Method’.
5.2 Network Setup
The PRS includes an internal WebServer with
HTML web pages that can be accessed by a web browser from any computer. The web pages let the user quickly change the PRS network settings.
The default IP address is 192.168.1.253 static.
5.3 Ethernet Test Keyboard
To assist the user in setup and communication the VXI-11kybd program can be used which includes defined RPC calls to change the PRS network settings.
VXI-11 Keyboard Control Program provides interactive control of VXI-11 instruments from the computer keyboard without having to write a program. The VXI-11 Keyboard program is the
ETHERNET INTERFACE OPTION
14
15
PRS-300 Series
ideal utility program for con fi guring and testing the
PRS or any VXI-11 compatible instrument. Use the VXI-11kybd program to exercise the interface or to try out commands before using them in a program. The VXI-11kybd program can be found on the PRS product page at www.ietlabs.com
Versatile Programming
The PRS can be easily controlled by several programming techniques and languages because it is a VXI-11.3 instrument. If you program with LabVIEW, National Instruments’ VISA supports VXI-11.3 instruments like the PRS. NI’s
Measurement and Automation Explorer treats the PRS as a TCP/IP compliant device. Agilent’s
VISA library supports VXI-11.3 instruments and the Agilent Connection Manager sees the PRS as a TCP/IP instrument.
If you are a Visual Basic, VB.Net or C/C++ programmer, you can write your program to call Agilent’s or National Instruments’ VISA or Agilent’s SICL library in the Windows environment.
If you use LINUX or any other fl avor of UNIX like SunOS, IBM-AIX, HP-UX, or Apple’s OS X, you can communicate with the PRS through RPC over TCP/IP. RPC (or Remote Procedure Calls) provides an invisible communication medium for the developer. The VXI-11 speci fi cation provides an RPCL (Remote Procedure Call Library) that can be used by virtually any operating system to control the PRS.
If you program with Java then you can write a
PRS control program that can be easily moved to many different operating systems.
The Java jGpibEnet project on SourceForge was developed using an ICS 8065 Controller similar to the PRS Controller.
5.4 Network Confi guration
This paragraph con fi gures the 9006 card in the
PRS for operation on your network. The board’s digital interface is con fi gured later by sending commands as outlined in Appendix A.
When shipped, the boards are con fi gured with default settings outlined in Table 4-1 Default
Settings.
Command Function Factory Setting
IP Mode Static or DHCP
Mode
Static
IP Address
Net Mask
0.0.0.0 to
255.255.255.255.0
0.0.0.0 to
255.255.255.255.0
Sets socket timeout COMM Timeout
Auto Disconnect Aborts socket if link count goes to 0
Table 4-1 Default Settings
192.168.1.254
255.255.255.0
120 sec.
Off
Review the Table with your network administrator and decide on which settings, if any, that need to be changed. Table 4-1 provides detailed information about each network setting to help you with your decisions. The minimum change is to set a static IP address so your PC can communicate with the board.
The network con fi guration can be changed and the board’s MAC Address can be read with a web browser, by running ICS’s VXI-11 Con fi guration
Utility on a WIN32 or WIN98 PC. ,
5.5 Web Browser Configuration
Method
This method uses a standard browser such as
Firefox, Internet Explorer or Safari to view and change the current network settings.
1. Temporarily disconnect the computer from the company network. Connect the PRS and computer running the browser using a standard
Ethernet Cables to a hub or switch. Temporarily disconnect the local network connection to avoid network con fl icts until the board is con fi gured.
ETHERNET INTERFACE OPTION
PRS-300 Series
Figure 4-1 9006 Connected to the local hub
An alternate connection is to use an Ethernet
Crossover Cable to connect the computer directly to the PRS for initial con fi guration. This will eliminate any potential network con fl icts while con fi guring the PRS.
a different range, record the current settings and temporarily set the following network values:
Check ‘Use the following IP Address’
IP Address 192.168.1.254
Subnet mask 255.255.255.0
4. Open the browser and enter the default IP address of 192.168.1.254 for new units (or your last set address for older units) in the browser address window.
5. A Welcome Page similar to the one shown
Figure 4-3 should appear in your browser.
6. If you want to change any of the settings, press the ‘Update Configuration’ button. A
Con fi guration Page similar to the one shown in
Figure 4-2 9006 Connected to the local hub
2. Apply power to the PRS. Set the Remote local switch to Remote.
3. Check your computer’s network settings to be sure its IP address is in the 192.168.1.xxx range so it can communicate with the PRS
Ethernet board’s default IP address. If it is not, it must be set before proceeding. Use the values shown below. For Windows PCs, right-click on
My Network Places and click on Properties.
Right-click on Local Area Connection and click
Properties. Highlight Internet Protocol (TCP/IP) and click Properties. If your PC’s IP address is in
Figure 4-3 9006 Welcome Page
ETHERNET INTERFACE OPTION
16
PRS-300 Series
Figure 4-4 9006 Con fi guration Page
7. Enter the new settings as desired. If you select
DHCP for the TCP/IP Mode, the page blanks out the IP, Net and Gateway addresses as they will be supplied by your DHCP server. Check the entered values carefully as the unit’s webserver does minimal error checking. Press the ‘Update
Flash’ button when done. A Con fi rmation Page similar to the one shown in Figure 4-5 will appear in your browser.
8. Your new settings have been saved in the board’s fl ash memory. You have to reboot the unit or power cycle it for the changes to take affect.
Press the ‘Reboot’ button to reboot the unit now or the ‘Return to the Con fi guration Page’ button to revisit the new settings.
5.6 VXI-11 Configuration Utility
Method
The VXI-11 Configuration Utility program is called ‘VXI11_con fi g.exe’ and runs in any
WIN32 PC with Windows 98, Me, 2K, XP,
Vista, 7 and Server 2003 operating systems. The
VXI11_con fi g.exe program can be found on the
PRS product page at www.ietlabs.com.
The VXI11_con fi g program can be run from the
CD or can be installed onto your hard disk and run from the installation directory.
VXI11_con fi g.exe is a Visual Basic program and requires that either Microsoft’s Visual Studio 6 or VBruntime6 be installed on your PC to run.
1. Connect the board directly to the WIN32 PC that will be running the Con fi guration Utility.
Disconnect the PC from the company network and use the supplied Ethernet Crossover Cable to connect the PC to the board as shown in Figure
4-6. This will eliminate any potential network con fl icts while con fi guring the board.
17
Figure 4-5 9006 Con fi rmation Page
Figure 4-6 9006 Connected to PC with a
Crossover Cable
Alternately, use a standard Ethernet patch cable
ETHERNET INTERFACE OPTION
PRS-300 Series
to connect the 9006 to the same hub or switch that the PC running the Con fi guration Utility is connected to as shown in Figure 4-7. Temporarily disconnect the local network connection to avoid network con fl icts until the board is con fi gured.
scans for all VXI-11 Services connected to the local LAN or to your PC. (The 9006 is an RPC server which provide a VXI-11 Service) The results are displayed in the Results box.
6. When the servers(s) have been found, use the pulldown arrow in the Found Servers box to view the Found Server addresses. The board’s default address is 192.168.1.254. Highlight the board’s IP address and click the Create Link button. If the server is not found, you can enter the default IP address (192.168.1.254) in the
Found Servers box. Click the Create Link button.
Figure 4-7 9006 Connected to the local hub
2. Apply power to the board
3. Check your PC’s network settings to be sure its IP address is in the 192.168.1.xxx range so it can communicate with the board’s default IP address. To check, right-click on My Network
Places and click on Properties. Right-click on
Local Area Connection and click Properties.
Highlight Internet Protocol (TCP/IP) and click
Properties. If your PC’s IP address is in a different range, record the current settings and temporarily set the following network values:
Check ‘Use the following IP Address’
IP Address 192.168.1.254
Subnet mask 255.255.255.0
4. Run the VXI11_config program. The
Con fi guration Utility opens a window as shown in Figure 4-8. Initially only the Find Server, Help and Exit buttons are enabled on the program window. The other buttons will be enabled as you advance through the program.
5. Click on the Find Server button. The program
Figure 4-8 VXI-11 Con fi guration Utility
(Showing all configuration choices with no changes)
7. When the link has been created, device model number(s) will appear in the ‘Select Device to be Con fi gured’ box. Highlight the desired model number and click the Con fi gure button to start the con fi guration process.
8. The Con fi guration Choices box displays only one line with the fi rst parameter to be changed and its current setting. If you like the current setting, click Enter to advance to the next
ETHERNET INTERFACE OPTION
18
19
PRS-300 Series
parameter. If you want to change the setting, type a new value in the New Value box and click
Enter. The program will send your setting to the board and read back the new setting. Repeat as needed to make another change or click Enter again to advance to the next parameter.
9. Repeat step 8 for each con fi guration parameter.
Figure 4-8 shows the VXI-11 Configuration
Utility after all parameters have been entered for a Model 9006. Click the Redo From Start button if you need to start over or if you want to change any of the prior settings
10. When done, the Save Con fi guration button is enabled if you changed any settings. Click the
Save Con fi guration button to save the values in the board’s fl ash memory. If you did not make any changes you can just exit the program.
11. The board has to be power cycled or rebooted before the con fi guration changes take affect.
Click the Reboot button to reboot the board and use the new settings.
12. Press the Exit button to quit the VXI11_con fi g program.
13. If the IP address was changed to an address outside the 192.168.1.xxx range in step 3, your
PC’s network settings will have to be changed to communicate with the board. Exit the VXI11_ con fi g program and restore the PC’s network settings.
5.7 Resetting Default Network
Settings
The board can be reset to the default network settings listed in Table 4-2 at any time by holding the LAN Reset Button in for 5 seconds while applying power to the board. Access to this button is on the rear panel of the PRS-300..
The Digital I/O con fi guration values are not affected by the LAN Reset operation.
Command
IP Mode
IP Address
Net Mask
Function
Static or DHCP
Mode
0.0.0.0 to
255.255.255.255.0
0.0.0.0 to
255.255.255.255.0
Sets socket timeout COMM Timeout
Auto Disconnect Aborts socket if link count goes to 0
Table 4-2 Default Settings
Factory Setting
Static
192.168.0.2
255.255.255.0
120 sec.
Off
5.8 PRS Programming
A PRS SCPI command reference is included in Appendix A. This gives a complete set of commands.
An example of commands to be sent to place the PRS in remote, this must be done fi rst otherwise the will be no remote control of output impedance, and then con fi gure the unit for a speci fi c impedance value.
Where the command string is constructed as described in Appendix A and Appendix B.
A command string might be, for example:
SOURce:DATA 1000000.1
ETHERNET INTERFACE OPTION
PRS-300 Series
Chapter 5
IEEE INTERFACE OPTION
6.1 Introduction
The IEEE interface option makes the
PRS-300 a IEEE-488.2-1987 and SCPI 1994.0 compatible instrument. The PRS-300 IEEE-
488 interface is based upon the ICS Ethernet/
GPIB/USB to Serial Interface Card. More information can be found at www.icselect.
com on the 9006 Ethernet/GPIB/USB to Serial
Interface Card.
The IEEE STD 488.2 covers the electrical and mechanical bus speci fi cations, and state diagrams for each GPIB bus function. It also establishes data formats, common commands for each 488.2 device and controller protocols.
The standard is available on-line at http://www.
ieee.org.
The SCPI standard provides a tree like series of standard commands for programmable instruments so that similar instruments by different manufacturers can be controlled by the same program. SCPI information and a command reference are located in Appendix A.
Other tutorials are available on-line; consult IET for additional information. A software
GPIB “keyboard” can be found at www.ietlabs.
com
6.2 Capabilities
The IEEE option provides remote control over all functions except
POWER
. The IEEE option responds to all Basic Commands in Appendix B and all IEEE.2 Mandated common commands in Appendix A.
6.3 Address Switch and
Communications Settings
The default GPIB address is 4. This address can be changed remotely by sending the command SYSTem:ADDR <n> where n is 0
- 30.
6.4 GPIB Test Keyboard
To aid the user in operating the PRS, a
GPIB “Keyboard” Controller program - the easiest way to control GPIB devices without writing a program - is available from IET. This
GPIB Keyboard program automatically fi nds your device at start-up and it lets you enter just the data that you want to send to the device.
This program works with ICS, Measurement
Computing and National Instruments controllers.
To implement, request a download of
ICS_GPIBkybd_Install.zip
from IET Labs
Tech Support.
IEEE INTERFACE OPTION 20
PRS-300 Series
Unzip the fi le and follow instructions to install.
Open the application. You may use the
Find Listeners
button to con fi rm that the PRS unit is recognized. Other instruments may also be recognized at this time.
21
IEEE INTERFACE OPTION
Chapter 7
PROGRAMMING
7.1.1 Calibration Interval
The standard USB interface can be programmed via any of the commands listed in
Appendix B.
For USB and RS-232 interfaces, <lf>,
<cr> or <crlf>, can be used as terminator. The device returns in this case <crlf>. The device performs all commands previously sent after it receives a terminator.
The optional Ethernet/IEEE/USB card can be programmed via any of the commands listed in Appendix A and Appendix B.
PRS-300 Series
PROGRAMMING
22
PRS-300 Series
Chapter 8
MAINTENANCE
8.1 Verifi cation of Performance
8.1.1 Calibration Interval
The PRS Series instruments should be veri fi ed for performance at a calibration interval of twelve (12) months. This procedure may be carried out by the user, if a calibration capability is available, by IET Labs, or by a certi fi ed calibration laboratory. If the user should choose to perform this procedure, then the considerations below should be observed.
8.1.2 General Considerations
It is important, whenever testing the PRS
Series Decade units, to be very aware of the capabilities and limitations of the test instruments used. Such instruments have to be signi fi cantly more accurate than the speci fi ed accuracy for all applicable ranges in order to perform this task, allowing for a band of uncertainty of the instrument itself, the test setup and the environment; consult IET Labs for information.
It is important to allow both the testing instrument and the PRS-300 to stabilize for a number of hours at the nominal operating temperature of 23
0
C, and at nominal laboratory conditions of humidity. There should be no temperature gradients across the unit under test.
In the case of the PRS-300 a Kelvin type
4-wire test terminals should be used to obtain accurate low-resistance readings.
Steps should be taken to minimize thermal-emf effects, by maintaining an even temperature and by using only low-emf connectors. Use of meters with a “True Ohm” function is recommended.
Proper metrology practices should be followed in performing this veri fi cation.
8.1.3 Calibration Procedure
1. Determine the allowable upper and lower limits for each resistance setting as would be done for a decade box from 0.1 Ω to 20 M Ω s based on the speci fi ed accuracy.
2. Con fi rm that the resistances fall within these limits
3. If any resistances fall outside these limits, See section 8.1.3 Adjustment of internal resistors.
8.1.4 Adjustment of Internal Resistors
During the adjustment process a DMM such as the Fluke 8508A or Keysight 3458A is connected to the binding posts of the PRS-300 using a 4 terminal Kelvin connection.
The PRS-300 has 43 internal resistors which are combined in either series or parallel to achieve the programmed resistance.
These resistors require periodic measurement to keep the PRS-300 within published speci fi cations. The resistors are measured using a DMM. The measured value is then
MAINTENANCE
23
PRS-300 Series
stored in a table within the memory of the
PRS-300. The measured values of the resistors are used in an algorithm to determine which resistors are used to create the programmed resistance value.
In addition to the 43 resistors the residual internal lead resistance is also measured at time of manufacture and entered into memory. The internal lead resistance does not need to be measured again unless repair has been performed to the PRS-300.
Below is a list of the 43 resistors and their nominal resistance value.
Res. Number Nominal Resistance Value
1 0.17 Ω
2
3
4
0.35
0.7 Ω
1.37
Ω
Ω
5
6
7
8
9
2.66
5.15
10.2
20.0
38.8
Ω
Ω
Ω
Ω
Ω
11 148.0 Ω
During the adjustment process the fi rmware cycles through each resistor and the technician can enter the measured resistance value in ohms. This should be done with a minimum of 6.5 digits of resolution.
IET Labs offers the PRS-DMM software which works with the PRS-300 and Fluke
8508A or Keysight 3458A to automate this adjustment process.
MAINTENANCE
24
25
PRS-300 Series
Appendix A
SCPI COMMAND REFERENCE
SCPI is an acronym for “Standard Commands for Programmable Instruments”. For additional information or an on-line copy of this standard, see:
http://www.scpiconsortium.org.
The IEEE 488.2 Standard was established in 1987 to standardize message protocols and status reporting and to de fi ne a set of common commands for use on the IEEE 488 bus. IEEE
488.2 devices are supposed to receive messages in a more fl exible manner than they send.
A message sent from GPIB controller to GPIB device is called: PROGRAM MESSAGE.
A message sent from device to controller is called: RESPONSE MESSAGE. As part of the protocol standardization the following rules were generated:
(;)
(:) words.
Semicolons separate messages.
Colons separate command
(,) Commas separate data fi elds.
<nl> Line feed and/or EOI as last character
terminates a ‘program message’.
Line feed (ASCII 10) and EOI
terminates a RESPONSE MESSAGE.
(*) Asterisk de fi nes a IEEE 488.2 common command.
(?) Ends a query where a reply is expected.
SCPI builds on the programming syntax of IEEE-488.2 to give the programmer the capability of handling a wide variety of instrument functions in a common manner. This gives all instruments a common “look and feel”. SCPI commands are not case-sensitive.
The portion of commands shown in capitals denotes the abbreviated form of the keyword. Either the abbreviated or whole keyword may be used when entering a complete command. There must be a space between the command and a parameter or channel list. Multiple
SCPI commands may be concatenated together as a compound command by using semicolons as command separators.
APPENDIX A
PRS-300 Series
Keywords shown inside braces [ ] are defaults, and are optional when constructing a
PROGRAM MESSAGE.
Commands not recognized have no effect on the unit’s operation and will set the corresponding bits in the Standard Event Status
Register. SCPI commands that end with a question mark ‘?’ are queries. All queries should be followed by reading their response to avoid data loss.
command line must end with line feed <lf>.
Response from the device also returns line feed
<lf>.
Terminators: For GPIB interface each
Semicolon ‘;’ is used to separate more commands written on one line.
IEEE-488.2 Common Commands
*IDN?
Returns instrument identi fi cation “IET Labs,PRS-300,<Serial Number>,<software version>”
*ESR?
Returns the read of the event status register.
*STB?
Returns the read of the status byte register.
*ESE?
Returns the read of the event status enable register.
*SRE?
Returns the read of the service request enable register.
*ESE
Set the event status enable register value
*SRE
Set the service request enable register value
*RST
Reset the buffer
*TST?
Self test query
*CLS
Clear standard event status register
*OPC
Operation complete
*OPC?
Operation complete query
*WAIT
Wait until operation is complete before executing next command(No Function in PRS)
APPENDIX A
26
PRS-300 Series
*PCS
Power-on Status Clear
Note: Remote command can start with or without * symbol for compatibility.
2.13 Formats
IEEE 488.2 enables remote programming of all instrument functions, measurement conditions and comparator settings etc. Outputs include measurement conditions, open/short corrections, and measured values.
Data Formats
Data will be transmitted in ASCII NR3 format per IEEE488.2 sec. 8.7.4 and reproduced below.
Note that there is always precisely one digit before the decimal point, and precisely three digits in the exponent.
Multiple results
For the case where a measurement produces multiple results (e.g. MEASure Cs, and DF), the individual numbers will be separated by commas per IEEE488.2 para. 8.4.2.2.
Sequences of Test (Sequence Mode) will be treated as a single Message Unit, with results separated by commas. If a particular test has “None” selected as a secondary parameter, no place will be reserved for the null result. As an example, a sequence of three tests asking for C/D, ESR, and Z/ would appear as follows:
<data>,<data>,<data>,<data>,<data><NL>
All response messages will be terminated by the NL character together with the EOI line asserted.
Status Byte Register
Decimal
6
5
64
32
Decimal
Bit Value
7 128
SRQ, SPOL Resets
Summary of Standard Event
4 16
3 8 None
2 4 None
1 2 None
0 1 None
*The Status Byte Regester is readable via the standard STB? as de fi ned in para. 11.2.2.2 of the
IEEE spec. The 7600 Plus will also implement an SRE register to enable each bit of the Status Byte
Register per para 11.3.2 of the IEEE spec. This register shall be readable by a SRE? command and writeable by a SRE <#> command.
Standard Event Status Register
Use
Power Up Since Last Query
APPENDIX A
27
PRS-300 Series
6 64
5
4
3
32
16
8
None
Command Error (Syntax)
Execution Error (Over Range, etc.)
No Contact
2 4
1 2
Query Error
None
0 1 Operation Complete
This register is read by executing an “ESR?” command, page 76 (except no *). Note that this is a destructive read. Reading the register clears it. Each bit of the Event register must be enabled in order to cause the ESB bit of the Status Register to be set. This enabling is done in the Standard
Event Status Enable Register by issuing an ESE command.
APPENDIX A
28
29
PRS-300 Series
Appendix B
PRS-300 USB and COMMON COMMANDS
COMMON Commands:
*IDN? Returns instrument identi fi cation “IET Labs Inc.,PRS-300,<Serial number>,<software version>”
Serial number in the form “Ax-xxxxxxxx” x = 0 to 9. A = Alpha can be A to Z.
*TST? Self-Test Returns ASCII 0 or 1, 0 = bad or 1 = good
*RST Returns PRS-300 to default “power on” condition
*SAV <number> Saves current resistance value into memory location 0 to 9
Number is 0 to 9
*RCL< number> Recalls resistance value from memory location 0 to 9
Number is 0 to 9
SYSTEM Commands
ber>
SYSTem:SERIal <Ax-xxxxxxxx> Sets the serial number in non-volatile memory < serial num-
Serial number in the form “Ax-xxxxxxxx” x = 0 to 9. A = Alpha can be A to Z.
SYSTem:SERIal? Returns Serial number
SYSTem:CONTrast <number> Sets system contrast speci fi c number from 0 to 15
SYSTem:KEY <number>
<Number> is 0 to 24 keys 0-9 0 – 9
A,
,
10 – 13
Same as pressing button on front panel of PRS-300
TABLE
RCL, SAV
16
17 – 18
000 / (-)
ENTER
21
22
24
APPENDIX B
30
PRS-300 Series
SYSTem:LOCal: This command places device in the LOCAL mode and unlocks all keys on front panel of the device. The Command is valid only for RS232, and USB interfaces.
SYSTem:REMote This command places device in the REMOTE mode and locks all keys. The Command is valid only for RS232, and USB interfaces.
CALIBRATION Commands
CALIbrate:RESistance Goes into calibration mode starting at resistor 1 and then progressing to resistor 43. Each time the command CALI:RES:SET is used the resistor number is automatically incremented by 1.
CALIbrate:RESistance? Returns resistance number to be calibrated.
Numbers 1 to 43
CALIbrate:RESistance:SET <number> Sets calibrated value of resistance in ohms into memory. <number> is a fl oating point number representing the resistance value in ohms for the internal resistor being measured. Each time the command CALI:RES:SET is used the resistor number is automatically incremented by 1.
CALIbrate:RESistance:SET? Returns the calibrated value of resistance in ohms
CALIbrate:DATE <mmddyyyy> Sets calibration date into memory
CALIbrate:DATE?
SOURCE Commands
Returns calibration date in format mmddyyyy
SOURce:DATA <display value> Sets output resistance of the PRS-300 if table is set to 0 or the resistance value corresponding to the display value for the table selected.
Display value in form of number and decimal point representing desired display value.
example SOURce:DATA 1.000002 would set resistance value to 1.000002 Ω if Table is set to
0.
SOURce:DATA? Returns display value currently shown in the PRS display
APPENDIX B
31
PRS-300 Series
SOURce:RESistance <display value> Sets output resistance of the PRS-300 if table is set to 0 or the resistance value corresponding to the display value for the table selected.
Display value is in the form of number and decimal point representing desired display value.
This is the same as SOURce:DATA command.
example SOURce:RES 0 would set display value to 0 and if Table is set to 1 to 4 this would represent the resistance value depending upon how the table is con fi gured. If in Celsius and PT-100 then resistance of 100 Ω would be set at output of PRS-300.
CONFIGURE Commands
CONFigure:TABLe:SELect <number> Selects Table to be used, <number> is 0 to 4. 0 is no table with display shown in ohms and 1 , 2 and 3 are user de fi ned tables
CONFigure:TABLe:SELect? Returns table number that is selected
CONFigure:TABLe:NAME <name> Sets Table name, <name> is ASCII up to 20 characters such as “PT100”
CONFigure:TABLe:NAME? Returns the number, 1 to 4 ,of currently selected table
CONFigure:TABLe:ERASE Erases all rows in the table
CONFigure:TABLe:ADD <user_value,resistance_value> Adds a row to the end of table showing user value and resistance value in ohms. example CONF:TABL:ADD 20, 95 means that if the table is selected and the operator enters 20 from the front panel the output will be set to 90 Ω .
CONFigure:TABLe:ADD? Returns the last entry in the table
CONFigure:TABLe:DISPlay? Returns all data in the table
Example of information returned
PT100
20, 95
CONFigure:TABLe:UNIT <Unit> Set units for tables up to 8 ASCII characters
As an example “C” could be send for Celsius or “F” for Fahrenheit
CONFigure:TABLe:UNIT? Returns ASCII string representing the units
CONFigure:SELECT <”2” or “4”> Sets “2” Wire or “4” Wire mode
APPENDIX B
Advertisement
Key Features
- High precision decade resistance substitution
- RTD (Resistance Temperature Detector) simulation
- Remote control via USB, Ethernet, and IEEE-488.2
- Low thermal emf relays
- Automatic adjustment procedure
- Resolution of 1 μΩ
- Range of 0.1 Ω to 20 MΩ
- Maximum load of 2 A, 200 V, 0.5 W