Pressure Transmitters for Precision Measurement with Digital Output

Electronic
Pressure Measurement
ressure Transmitters for Precision Measurement
P
with Digital Output RS 232
Model D-10, standard version / Model D-11, flush diaphragm
WIKA Datasheet D-10, D-11
Applications
„„ Automation engineering
„„ Test bench construction
„„ Laboratories
„„ Maintenance shops
Special Features
„„ Accuracy ≤ 0.1 % (optional 0.05 %) of span
„„ Digital output RS 232 with 9-pin Sub-D connector
„„ Communication software EasyCom for Windows ® 95, 2000, NT, XP or Vista
„„ No additional temperature error in the range 0 ... 50 °C
„„ Pressure ranges from 0 ... 250 mbar to 0 ... 1,000 bar
Fig. left Pressure Transmitter D-10
Fig. right Pressure Transmitter D-11
Description
High Precision
The model D-10 pressure transmitter with an accuracy of
0.1 % (or 0.05%) has been designed to enable direct
communication to a PC, which is specifically required in the
field of test, calibration and service technology. The electrical
power of the pressure transmitter is taken directly from the
RS 232-interface of the PC.
Digital signal processing
The digital data processing of the precision pressure
transmitter ensures excellent values regarding linearity
and repeatability. System-related temperature errors occurring
usually in pressure measuring instruments are compensated
by the temperature sensor integrated in the process
connection and the digital data processing via
microprocessor. The result is a total temperature error
below 0.1 % in the range of 0 ... 50 °C (32 ... 122 °F).
WIKA Datasheet WIKA Datasheet D-10, D-11 · 01/2016
Software EasyCom
The communication software Easy Com, which is included
in delivery, allows not only the display of pressure and
temperature but also the storage of the measuring data for
pressure and temperature (data logger function) it also offers
an easy adjustment of zero point and span.
Flush version
The model D-11 with a flush diaphragm is particularly suitable
for the measurement of media that is highly viscous,
crystallizing or contains particulates.
Page 1 of 4
R
Specifications
Model D-10 / D-11
Pressure ranges
bar
0.25
0.4
0.6
1
1.6
2.5
4
6
Pressure ranges
Over pressure safety
bar
bar
25
50
40
80
60
120
100
200
160
320
250
500
400
800
600
1000 1)
1200 1500
2400 2) 3000
Over pressure safety
Burst pressure
Burst pressure
bar
bar
2
2.4
2
2.4
4
4.8
5
6
10
12
10
12
17
20.5
35
42
10
35
42
16
80
96
bar
96
400
550
800
1000 1200 1700 2)
{Vacuum, gauge pressure, compound range, absolute pressure are available}
{compound ranges: minimum span 400 mbar, e.g. -200 mbar ... +200 mbar}
1) Only
2)
Materials
■ Wetted parts
» Model D-10
Model D-10.
For Model D-11: the value specified in the table applies only when sealing is realised with the
sealing ring underneath the hex. Otherwise max. 1500 bar applies.
(other materials see WIKA diaphragm seal program)
Stainless steel (pressure ranges > 25 bar additional Elgiloy®)
» Model D-11
■ Case
Internal transmission fluid 3)
Power supply UB
Stainless steel {Hastelloy}; O-ring: NBR {FPM/FKM or EPDM}
Stainless steel
Synthetic oil {Halocarbon oil for oxygen applications}
{Listed by FDA for Food & Beverage}
3)
Not for D-10 with pressure ranges > 25 bar.
via RS 232-interface
When connecting the D-1X to a notebook/laptop computer, the minimum power
supply of the RS 232 interface must be secured if necessary with an adapter,
which can be ordered as an option (see page 4)
RS 232 (8N1/9600 Baud) {USB via Seriell Converter}
3 adjustable modes of operation
■ Pressure and temperature value on request from host system
■ Cyclic pressure output, time interval adjustable 10 ms 4) ... 10 min
Signal output
Adjustability
■ Zero point
■ Span
Internal measuring rate
Warm-up time
Insulation voltage
Accuracy 6)
■ Cyclic pressure and temperature output, time interval adjustable 10 ms 4) …10 min
Resolution pressure value: 50,000 digits, temperature value 0.5 K
4)
Due to the Windows access times an interval of 10 ms is not feasible via the EasyCom software.
%
-5 ... +20 (adjustment via software EasyCom)
%
-5 ... +5 (adjustment via software EasyCom)
Hz
100 5)
5)
50 Hz with pressure ranges ≤ 1 bar or compound pressure ranges ≤ 3 bar span
min
< 10
VDC
500
% of span
≤ 0.1 in the range 0 ... +50 °C / +32 ... 122 °F {< 0.05 at 20 °C / 68 °F} 7)
Including non-linearity, hysteresis, zero point and full scale error
(corresponds to error of measurement per IEC 61298-2).
Adjusted in vertical mounting position with lower pressure connection.
6)
Cannot be manufactured for: compound ranges and pressure ranges ≤ 0.4 bar.
% of span
≤ 0.04 (BFSL) according to IEC 61298-2
% of span
≤ 0.1
(at reference conditions)
7)
Non-linearity
1-year stability
Permissible temperature of
■ Medium *)
■ Ambience
■ Storage
Compensated temp. range
Temperature coefficients within
■ Mean TC of zero
■ Mean TC of span
■ CE-conformity
■ Pressure equipment directive
■ EMC directive
Shock resistance
Vibration resistance
Page 2 of 4
% of span
% of span
g
g
-20 ... +80 °C {other on request}
-4 ... +176 °F {other on request}
-20 ... +80 °C
-4 ... +176 °F
-40 ... +85 °C
-40 ... +185 °F
-20 ... +80 °C
-4 ... +176 °F
(the temperature related deviations in the range 0 ... +50 °C (+32 ... +122 °F)
are already covered by the accuracy above)
≤ 0.1 / 10 K
≤ 0.1 / 10 K
97/23/EC
89/336/EEC emission (class B) and immunity according to EN 61 326
< 100 according to IEC 60068-2-27
< 5 according to IEC 60068-2-6
(mechanical shock)
(vibration under resonance)
WIKA Datasheet D-10, D-11 · 01/2016
Specifications
Model D-10 / D-11
Wiring protection
■ Reverse polarity protection
Software
Weight
kg
UB+ towards UBCommunication software EasyCom (for Windows® 95, 98, 2000, NT, XP or Vista)
Approx. 0,3
*) In an oxygen version model D-11 is not available. In an oxygen version model D-10 is only available with media temperatures between
-20 … +60 °C/ -4 … +140 °F.
{ } Items in curved brackets are optional extras for additional price.
Dimensions in mm
Ingress Protection IP per IEC 60529.
Electrical connection
9-pin Sub-D connector,
cable length 1.5 m
IP 67 (on the instrument side)
Case
Pressure connections D-10
G 1/2
Order code: GD
G 1/4
Order code: GB
1/ 2 NPT
per „Nominal size for
US-standard tapered
pipe thread NPT“
Order code: ND
1/ 4 NPT
per „Nominal size for
US-standard tapered
pipe thread NPT“
Order code: NB
Others on request
Pressure connections D-11, flush diaphragm
G1
0 ... 0.25 up to 0 ... 1.6 bar
Order code: 85
Sealing ring
29.7 x 35.7 x 2.0
O-ring 26 x 2
G 1/2
> 1.6 bar
Order code: 86
Sealing ring
18.5 x 23.9 x 1.5
O-ring 15 x 2
Others on request
For installation and safety instructions see the operating instructions for this product.
For tapped holes and welding sockets please see Technical Information IN 00.14 for download at www.wika.com
{ } Items in curved brackets are optional extras for additional price
WIKA Datasheet D-10, D-11 · 01/2016
Page 3 of 4
Wiring detail
Sub-D connector
Sub-D connector at PC
Sub-D connector Pressure Transmitter
white
brown
yellow
green
Pressure
Transmitter
The pressure transmitter is
supplied via the pins RTS and
DTR of the PC interface.
The maximum permissible
cable length between pressure transmitter and PC is 3 m.
Accessories
Adapter for stabilising the RS 232 interface for the operation of a D-1X on a laptop/notebook
USB Seriell Converter, converts USB-signal into RS-232 signal
Order No.
7429407
2470327
Further information
You can obtain further information (datasheets, instructions, etc.) via Internet address www.wika.com
Page 4 of 4
WIKA Datasheet D-10, D-11 01/2016
R
WIKA Instrument Corporation
1000 Wiegand Boulevard
Lawrenceville, GA 30043-5868
Tel: 888-WIKA-USA • 770-513-8200
Fax:770-338-5118
E-Mail:info@wika.com
www.wika.com