YSC Technologies
BRIEF
Thank you for purchasing 3D Microscope System.
To use the product correctly, please read this “Operation Instruction” carefully.
Please understand the symbol first.
Graph
Meaning
Danger
Caution
Common
Forbidden
Common
Behavior
Cautions:
1. As the 3D microscope is precise equipment, always handle with care; avoid abrupt
movement and collision in usage and transportation.
2. Avoid exposure to direct sunlight, high temperature, humidity, dust and vibration.
3. Avoid leaving dirt or fingerprint on the lens surfaces, as the dirty mirror or lenses
may reduce image clarity.
Maintenance and Storage:
1. Place the 3D microscope in a cool, dry place where is free from dust, acid, alkali,
steam, etc.
2. Do not disassemble lens without authorization as all lens have been pre-adjusted
and tested. Should dirt is on lens; slightly wipe it with gauze with liquid mixture of
ether (70%) and alcohol (30%). Dust on lens might be blown away with a blower or
gently wiped out by clean brush or cleaning tissue. Be careful not to touch any optical
elements while in cleaning mechanical parts and smearing non-corrosive lubricant.
3. Do not use organic solvent to clean other component surfaces, especially for the
plastic components.
4. Do properly store reflector module not in use.
5. Do switch off power supply after usage. Unplug the power cord if the instrument is
out of use for long time.
6. Handle the instrument carefully and gently during usage and transportation. It is
strictly prohibited to put the instrument upside down.
3D Microscope Operation Procedures:
1. Adjust Working Distance and Focus: Rotate the focus adjustment knob counter
clockwise or clockwise to instrument main body up (down). After seeing the
image, rotate the knob slightly until the image is sharp.
2. Adjust Focus Knob Tension: Hold 1 side of the focus knob and rotate another side
clockwise (counter clockwise) to tighten (loosen) the knob tension. User can
adjust the tension freely up to its feeling.
Adjust Zoom: Rotate Zoom Adjustment Knob to change the zoom ratio.
3
510.226.0889
www.ysctech.com
info@ysctech.com
YSC Technologies
Description
Drawing
3
510.226.0889
www.ysctech.com
info@ysctech.com
YSC Technologies
No.
Name
No.
Name
1
Platform
7
3D Adapter
2
LCD
8
12V Power Socket
3
LED Light Adjusting
9
Power On/Off
6
Lifting Knob
10
Microscope Lens
11
Small Keyboard Socket
12
VGA Output
THE MAIN PARAMETERS:
:
Name
Parameters
Name
Power In
Parameters
Zoom Range
0.75~5
CCD Camera
1.3 or 0.3MP CCD
Min Fov
2mm×1.5mm
Working Humidity
Below 85%RH
Max Fov
11.5mm×9.2mm
Working Temperature
-20℃~50℃
Output Signal
12V/3A
VGA
Microscope operation:
1, the working distance and focal length adjustment: clockwise (counterclockwise) rotate the
handwheel, can make the lens upper (lower) move. When objects appears, slightly rotate the
handwheel until see the clear pictures.
2, rate adjust: adjust the zoom knob, it will continuously change the magnification.
3, 3D Angle adjustment: by rotating the lens adapter can adjust view angle, clockwise
(counterclockwise) adjust to the angle users needed.
4, LED lighting adjust: by turning the knob at the right side, clockwise and counterclockwise
rotate the lights to be bright or dark; Users can according to image needed to adjust.
3-d seismic steps
1, Adjust the zoom knob to lowest ratio
2, Manual adjust the lens adapter, pictures begin to spin
3, Adjust the working distance and focal length, clockwise (counterclockwise) rotate the
handwheel to make the lens upper (lower) move until the pictures around a fixed point without
moving.
4, Adjust the zoom knob to highest ratio, stop rotate the atapter, adjust the handwheel to the
picture center till it is clear. (right now, because magnification is larger, previewer is smaller, some
points may be not clear, please be patient to adjust)
Power adapter output must be 12V!
Before use, please make sure the power adapter cable connect with Lens.
3
510.226.0889
www.ysctech.com
info@ysctech.com
YSC Technologies
How to Measure
1. Setting
It will show below items after “Cross Cursor”is checked under system sub menu.
Cursor Color: 8 cursor color to be chosen
Measure Line Color: Set the measure line color
Calibration: Calibrate the real size. Within the FOV (Field of View), the length is longer, the
measurement accuracy is higher.
2. How to Calibrate
Put standard calibrator on the table.
Calibrate X: Long press CAL for 3 secs to enter X Calibration Mode.”Calibration Mode dx,
Calibration length= ** mm”will be shown on upper left of display. Using SET, ←,→ to
adjust parallel lines distance as calibration distance.Then long press CAL until “Calibration
dx OK”shown on upper left of display.
Calibrate Y: Long press CAL for 3 secs to enter Y Calibration Mode.”Calibration Mode dy,
Calibration length= ** mm”will be shown on upper left of display. Using SET, ↑,↓ to adjust
parallel lines distance as calibration distance.Then long press CAL until “Calibration dy
OK”shown on upper left of display.
3. Mesurement
Put object to be measure on the table. Maintain the same manification, adjust focus for
sharp image. Using SEL、↑、↓、←、→ to measure the object real size.The meausrment
accuracy depends on manification. The manification is higher, accuracy is higher.
Note: Maintain the same manification in measurement. Otherwise it will need
recalibration.
4. Image Freeze
Hold POWER for 1 sec to freeze the image.To measure on freeze image will be more
stable. Hold POWER for 1 sec to release.
5. Power Off
Hold POWER for 3 sec to power off.
Camera Specification:
1. 1.3 MP pixel CMOS, 1280(H)*1240(V)
,15 frames/sec,600TV-Lines。
2. VGA output, 1024X768 60HZ、1024X768 75HZ、1280X1024 60HZ
3. Chinese/English menu, Brightness, Contrast, Saturation, Acutance,Blue Offset, Red
Offset,Exposure Control,Negative,Cross, Line Color. All specifications is memory
power off proof, except vertical mirror image, horizontal mirror image, Negative.
Note: Using auto exposure control will achieve better image quality with brightness is
adjustable.
4. Accuracy= pixels for standard length/ calibrated length. For example, when it is
displayed in 1024*768 pixels, the cross will move in 1 pixel unit. When you magnify
1mm to full screen, the measurement accuracy is 1/1024mm, nearly 1 micron. If
0.1mm is magnified to full screen, accuracy will be 0.1um
3
510.226.0889
www.ysctech.com
info@ysctech.com