Keysight Technologies M9502A and M9505A

Keysight Technologies
M9502A and M9505A
2 and 5-Slot AXIe Chassis
Data Sheet
02 | Keysight | M9502A and M9505A 2 and 5-Slot AXIe Chassis - Data Sheet
Product description
The Keysight Technologies, Inc. M9502A and M9505A AXIe chassis are fully compatible
with the AXIe 1.0 specification. They provide 2 or 5 slots for AXIe instrument modules,
and have an embedded system module that does not take up an instrument slot. The
embedded system module (ESM) provides Gigabit LAN and Gen 2 x8 PCIe® interfaces for
connecting the chassis to an external controller. To minimize rack space, the instrument
module slots are arranged horizontally. In addition, the chassis are designed to be easily
maintained. Both the power supply and fan tray can be removed with modules installed
while the chassis remains in the rack.
Applications
–– Aerospace and defense
–– Computation
–– Communications
–– Electronics test
–– Semiconductor testing
–– High-energy physics
Features
–– AXIe 1.0 compliant
–– Embedded system module with Gen 2 PCIe x8 and Gigabit LAN interfaces for connecting to host computer
–– Keysight-exclusive, built-in inter-chassis synchronization
for multi-chassis systems
–– 62 differential local bus lines provide very large data
path between adjacent modules
–– High power with 200W cooling per slot
–– Front-removable fan tray
Customer values
–– AXIe chassis provide a high-performance platform to compliment PXI-based
systems
–– The compact, 2-slot chassis is ideal for transportable applications, but is also
rack-mountable
–– The 5-slot chassis provides a cost-effective solution when more AXIe slots are
required
–– The embedded AXIe system module and horizontal module placement save rack
space, requiring only 2U or 4U, depending on model
–– With the innovative cooling design, no additional rack space is required to meet
cooling specifications
03 | Keysight | M9502A and M9505A 2 and 5-Slot AXIe Chassis - Data Sheet
Hardware platform
Compliance
The M9502A and M9505A chassis are fully compliant with the
AXIe 1.0 specification. An additional half-height slot is built into
the chassis and reserved for the embedded system module. This
slot is not compatible with AXIe instrument modules or embedded
computers. Embedded controllers must be installed in one of the
instrument module slots.
Backplane configuration
The AXIe backplane in the Keysight M9502A and M9505A chassis
include all of the following AXIe backplane lines:
–– Timing and triggering signals including: 100 MHz clock, 100
MHz PCIe clock (FCLK), point-to-point star trigger from ESM
(SYNC), bi-directional point-to-point star trigger (STRIG), and
12 signal parallel trigger bus (TRIG)
–– Distributed PCIe Gen 2 data fabric from ESM (x4 to each slot)
plus a secondary PCIe data fatbric
–– Distributed gigabit Ethernet LAN to each slot
–– 62-pair local bus for adjacent module signaling or data
transfer (AXIe only requires 18 pairs)
–– DC power rail (normally 45-53 VDC)
–– Intelligent Platform Management Bus (IPMB) for chassis
control, including module power-up
Figure 1. 5-slot AXIe backplane
04 | Keysight | M9502A and M9505A 2 and 5-Slot AXIe Chassis - Data Sheet
Hardware platform (continued)
Figure 2. M9505A backplane block diagram
Embedded system module (ESM)
Both the M9502A and M9505A chassis contain an embedded
system module that does not take up an instrument slot. This
module performs the chassis management functions contained in
the AXIe specification, including:
–– Managing trigger and clock busses
–– Managing clocks, including internal or external reference
sources
–– Gigabit LAN switching with a front panel RJ45 LAN
connection (see module documentation to determine if it is
supported for LAN communications)
–– AXIe Fabric 1 switching (Gen 2 x4 lanes to each module slot)
–– Monitoring the health of the chassis subsystems and modules
–– Controlling the chassis cooling fans, and managing the
chassis power-up sequence
In addition to the required functions, the system module also
provides:
–– Ability to synchronize multiple chassis with trigger and
clocking signals (MultiFrame In/Out connectors)
–– SMA connectors for external clock in/out and trigger in/out
–– Gen 2 x8 PCIe cable connector for connecting the chassis to
an external computer
–– USB 2.0 option provides USB connectivity for some AXIe
modules. Only supported on a limited set of Keysight
modules. See module documentation to determine if it
supported for USB control.
05 | Keysight | M9502A and M9505A 2 and 5-Slot AXIe Chassis - Data Sheet
Hardware platform (continued)
Innovative cooling
Chassis Web server
The AXIe chassis utilize a cooling design that does not require
additional rack space to cool the modules. This allows the chassis
to fit into 2U or 4U of rack space.
Both AXIe chassis include a built-in Web server for configuration,
control, and monitoring of the chassis.
This Web server can be accessed by either LAN or PCIe.
The cooling design uses auto-speed fans to pull in cool air. Air
flows from right-to-left within the chassis. The ESM provides
intelligent control of the power supplies and fans and ensures
there is sufficient power/cooling for each module.
The following functions are provided:
Lower maintenance costs
The chassis were designed to reduce maintenance costs. The
innovative air-flow design does not require air filters to replace.
In addition, the power supply and fans can be removed while the
chassis is mounted in a rack, allowing the chassis to be serviced
while keeping DUT cabling in place.
Software platform
–– Chassis information
–– Chassis LAN configuration
–– Instrument module inventory
–– Trigger routing setup
–– Chassis health (temperature, fan, power supply)
Soft front panel
A soft front panel interface is also provided to monitor and
control the AXIe chassis. It has some of the basic functionality of
the chassis Web page and communicates to the chassis using the
IVI drivers. The soft front panel has two screens: trigger routing
and chassis monitor.
Drivers
The M9502A and M9505A chassis come complete with IVICOM, IVI-C, and LabVIEW drivers. Windows XP, Windows 7, and
Windows 8.1 operating systems are supported and applications
can be completed using many different software tools including
LabVEW, LabWindows/CVI, MATLAB, VEE, VisualStudio.NET
(C/C++, C#, VB.NET).
Figure 3. Chassis Web page
Figure 4. Chassis soft front panel interface
06 | Keysight | M9502A and M9505A 2 and 5-Slot AXIe Chassis - Data Sheet
Technical Specifications and Characteristics
Chassis characteristics
Standards compliance
AXIe 1.0 Base Architecture specification
Backplane
M9502A
M9505A
Total slots
2
5
Instrument module slots
2
5
System module
Embedded
Embedded
M9502A
M9505A
Operating voltage range
100-240 VAC
100-240 VAC
Input frequency range
50-60 Hz
50-60 Hz
Input power consumption
800 VA
1280 VA
Overcurrent protection
Auto-recovery
Auto-recovery
Efficiency (typical)
85-90%
85-93%
Power
supply characteristics
AC input
DC
supply
Model
DC output
Total DC
module power
Total max module
current
Load
regulation
Maximum ripple and noise (20 MHz BW)
M9502A
50 V
400 W
8A
2%
1% pk-pk
M9505A
52 V
1000 W
19.2 A
2%
500 mV pk-pk
Chassis cooling and power dissipation
Slot airflow direction
Right to left
Chassis cooling intake
Right side of chassis
Chassis cooling exhaust
Left side of chassis
Chassis cooling fans
M9502A
Three 90.0 cfm fans on fan tray with HIGH/AUTO speed selector
M9505A
Six 90.0 cfm fans on fan tray with HIGH/AUTO speed selector
Power dissipation, instrument slot
200 W max
General characteristics
Mechanical
Weight (nom)1
Size with bumpers
Size without bumpers
M9502A
7.7 kg
(17 lbs)
462mm W x 102mm H x 414mm D
432mm W x 87mm H x 414mm D
M9505A
13.3 kg
(29.3 lbs)
462mm W x 193mm H x 436mm D
432mm W x 177mm H x 436mm D
1. Without modules
07 | Keysight | M9502A and M9505A 2 and 5-Slot AXIe Chassis - Data Sheet
Technical Specifications and Characteristics (continued)
Clocks and triggers
100 MHz system clock (CLK100)
Accuracy
± 20 ppm
Duty cycle
45/55%
Maximum slot-to-slot skew
100 ps
External clock in (SMA)
Input frequency
10 MHz ± 100 ppm
Input level
± 5 V, AC coupled
Minimum swing
250 mV
External clock out (SMA)
Output frequency
10 MHz
Output level
3.3 V CMOS
Output load
50 ohm
AXIe differential star trigger (STRIG)
Maximum slot-to-slot skew
100 ps
External trigger out (SMA)
Output level
3.3 V CMOS
Output load
50 ohm
External trigger in (SMA)
Input level
± 5 V, adjustable threshold
Input impedance
4 kohm (pulled up to 2.5 V)
Minimum swing
Environmental characteristics 1, 2
250 mV
Operating and storage conditions
Operating
Storage
Temperature
0˚C to 50˚C
-25˚C to 60˚C
Humidity
Type tested at 95%, +40˚C
(non-condensing)
Type tested at 95%, +40˚C
(non-condensing)
Altitude
up to 3000 meters
up to 4600 meters
Shock and vibration
Operating random vibration: type tested at 5 to 500 Hz, 0.21 g rms
Survival random vibration: type tested at 5 to 500 Hz, 2.09 g rms
Acoustical sound power (LWA dB, ref 1pW)
Worst case
Normal operation (25˚C air intake)
M9502A
74.3
55.2
M9505A
77.6
62.8
1. Samples of this product have been type tested in accordance with the Keysight Environmental Test Manual and verified to be robust against the
environmental stresses of storage, transportation, and end-use; those stresses include but are not limited to temperature, humidity, shock, vibration,
altitude, and power line conditions.
2. Test methods are aligned with IEC 60068-2 and levels are similar to MIL-PRF-28800F Class 3
08 | Keysight | M9502A and M9505A 2 and 5-Slot AXIe Chassis - Data Sheet
Technical Specifications and Characteristics (continued)
Regulatory charateristics
Safety
Complies with the essential requirements of the European LVD Directive of the following standards (dates and editions are
cited in the Declaration of Conformity):
– IEC/EN 61010-1
– Canada: CSA C22.2 No. 61010-1
– USA: UL std no. 61010-1
Acoustic statement
(European Machinery
Directive)
Acoustic noise emission
LpA < 70 dB
Operator position
Normal operation mode per ISO 7779
EMC
Complies with European EMC Directive of the following standards (dates and editions are cited in the Declaration of Conformity):
- IEC/EN 61326-1
- CISPR pub 11 group 1, class A
- AS/NZS CISPR 11
- ICES/NMB-001
This ISM device complies with Canadian ICES-001.
Cet appareil ISM est conforme a la norme NMB-001 du Canada
09 | Keysight | M9502A and M9505A 2 and 5-Slot AXIe Chassis - Data Sheet
Technical Specifications and Characteristics (continued)
ESM front panel connectors
10 | Keysight | M9502A and M9505A 2 and 5-Slot AXIe Chassis - Data Sheet
Configuration
Hardware
Model
Description
M9502A
2-slot AXIe chassis with embedded system module
M9505A
5-slot AXIe chassis with embedded system module
Includes:
AXIe filler modules, getting started guide, drivers, and
Keysight I/O libraries
Accessories
Figure 5. Configuration
Model
Description
Y1225A
Rack mount kit for M9502A
Y1226A
Rack mount kit for M9505A
Y1223A
AXIe MultiFrame cable: 0.5m*
Y1224A
AXIe MultiFrame cable: 3m*
Y1232A
Soft carry bag for M9502A
N5650-00080
Single slot AXIe filler module
*Check module documentation to see if MultiFrame cables are supported
Recommended configuration
Configure the Keysight M9502A and M9505A AXIe chassis, as
follows:
1. 1. Configure the chassis connection to the host:
–– If you are using PCIe to connect an external computer,
select an appropriate PC interface card; the Keysight
M9048A is recommended
–– For a list of tested computers, see 5990-7632EN.
–– If the AXIe modules support USB (see
www.keysight.com/find/axie-chassisusb) then you can
use a USB cable to connect the AXIe chassis to the
host. A PCIe interface card/cable is not required
2. Select an appropriate cable to connect the computer
interface board to the system module; the Y1202A is
recommended to connect the M9048A to the M9502A or M9505A
3. Select rack mount and MultiFrame cables as required
Related products
Model
Description
M9537A
Embedded AXIe PC controller
M9048A
PCIe desktop PC adapter: Gen 2, x8
Y1202A
PCIe cable: x8, 2.0 m (used with M9048A)
Software
Model
Description
Supported operating
systems
Microsoft Windows XP (32-bit),
Microsoft Windows 7 (32/64-bit),
Microsoft Windows 8.1 (32/64-bit)
Standard compliant drivers
IVI-COM, IVI-C, MATLAB
Supported application
development environments
(ADE)
Visual Studio (VB.NET, C#, C/C++),
LabVIEW, LabWindows/CVI, VEE
Keysight IO Libraries
Supported versions: 17.1 Update 1 (or
later). Use of latest version recommended.
Includes: VISA Libraries, Keysight
Connection Expert, IO Monitor
11 | Keysight | M9502A and M9505A 2 and 5-Slot AXIe Chassis - Data Sheet
Warranty and Calibration
Ordering
Model
Description
Advantage services: Calibration and warranty
M9502A
AXIe chassis: 2-slot with embedded system module
M9505A
AXIe chassis: 5-slot with embedded system module
Keysight Advantage Services is committed to your success throughout
your equipment’s lifetime.
Opt 900-932
Optional power cords
Warranty1
Opt U20
ESM USB 2.01
Includes USB cable
Included
3-year warranty (return to Keysight), standard
R-51B-001-5Z
5-year return to Keysight warranty assurance plan
1. USB port is only supported for a very limited set of Keysight AXIe
modules. Check module documentation or go to:
www.keysight.com/find/axie-chassisusb for compatibility information
1. Options not available in all countries.
Definitions
Specification
Warranted performance. Specifications include guardbands to account for the expected statistical performance distribution, measurement uncertainties,
and changes in performance due to environmental conditions. All specifications and characteristics apply over the operating environment outlined in the
“Environmental and Regulatory” section of this data sheet. In addition, the following conditions must be met:
–– Instrument is within its calibration cycle if calibration is required.
–– Instrument has been stored for a minimum of 1 hour within the operating temperature range prior to turn-on and after a 30 minute warm-up period.
Characteristics
Characteristics describe product performance that is useful in the application of the product, but that is not covered by the product warranty. Characteristics are often referred to as Typical or Nominal values.
Typical
Expected performance of an average unit when operated over a 20 to 30 °C temperature range. Typical performance is not warranted. The instrument
must be within its calibration cycle if calibration is required.
Nominal
Nominal describes representative performance that is useful in the application of the product when operated over a 20 to 30 °C temperature range.
Nominal performance is not warranted.
Additional information
All data are measured from multiple units at room temperature and are representative of product performance within the operating temperature range
unless otherwise noted.
The specifications contained in this document are subject to change.
12 | Keysight | M9502A and M9505A 2 and 5-Slot AXIe Chassis - Data Sheet
Evolving
Our unique combination of hardware, software, support, and people can help
you reach your next breakthrough. We are unlocking the future of technology.
From Hewlett-Packard to Agilent to Keysight
myKeysight
www.keysight.com/find/mykeysight
A personalized view into the information most relevant to you.
www.axiestandard.org
AdvancedTCA® Extensions for Instrumentation and Test (AXIe) is an
open standard that extends the AdvancedTCA for general purpose and
semiconductor test. Keysight is a founding member of the AXIe consortium.
ATCA®, AdvancedTCA®, and the ATCA logo are registered US trademarks of
the PCI Industrial Computer Manufacturers Group.
PCI eXtensions for Instrumentation (PXI) modular instrumentation delivers a
rugged, PC-based high-performance measurement and automation system.
Three-Year Warranty
www.keysight.com/find/ThreeYearWarranty
Keysight’s committed to superior product quality and lower total cost
of ownership. Keysight is the only test and measurement company with
three-year warranty standard on all instruments, worldwide. And, we provide
a one-year warranty on many accessories, calibration devices, systems and
custom products.
Keysight Assurance Plans
www.keysight.com/find/AssurancePlans
Up to ten years of protection and no budgetary surprises to ensure your
instruments are operating to specification, so you can rely on accurate
measurements.
Keysight Channel Partners
www.keysight.com/find/channelpartners
Get the best of both worlds: Keysight’s measurement expertise and product
breadth, combined with channel partner convenience.
For more information on Keysight
Technologies’ products, applications or
services, please contact your local Keysight
office. The complete list is available at:
www.keysight.com/find/contactus
Americas
Canada
Brazil
Mexico
United States
(877) 894 4414
55 11 3351 7010
001 800 254 2440
(800) 829 4444
Asia Pacific
Australia
China
Hong Kong
India
Japan
Korea
Malaysia
Singapore
Taiwan
Other AP Countries
1 800 629 485
800 810 0189
800 938 693
1 800 11 2626
0120 (421) 345
080 769 0800
1 800 888 848
1 800 375 8100
0800 047 866
(65) 6375 8100
Europe & Middle East
Austria
Belgium
Finland
France
Germany
Ireland
Israel
Italy
Luxembourg
Netherlands
Russia
Spain
Sweden
Switzerland
United Kingdom
0800 001122
0800 58580
0800 523252
0805 980333
0800 6270999
1800 832700
1 809 343051
800 599100
+32 800 58580
0800 0233200
8800 5009286
800 000154
0200 882255
0800 805353
Opt. 1 (DE)
Opt. 2 (FR)
Opt. 3 (IT)
0800 0260637
For other unlisted countries:
www.keysight.com/find/contactus
(BP-06-08-16)
DEKRA Certified
ISO9001 Quality Management System
www.keysight.com/find/axie-chassis
www.keysight.com/find/modular
www.keysight.com/go/quality
Keysight Technologies, Inc.
DEKRA Certified ISO 9001:2015
Quality Management System
This information is subject to change without notice.
© Keysight Technologies, 2013 - 2016
Published in USA, December 19, 2016
5990-6584EN
www.keysight.com
Download PDF