Quick Guide for Connection of CBM-20A(lite) and SPD-M20A

223-00950
Quick Guide for Connection of CBM-20A(lite) and SPD-M20A
This guide describes how to connect the CBM-20A(lite) and
*Procedure
the SPD-M20A to the network and make the settings to
1)
Setting IP address for PC
enable control from PC.
2)
Changing browser setting for PC
3)
Setting IP address and group name of the
Prepare LAN cable, a hub, and PC in order to make
CBM-20A(lite)
communication setting for CBM-20A(lite) and SPD-M20A
4)
that are at the state of shipment.
Setting IP address and master server of the
SPD-M20A.
5)
Changing IP address of the PC when necessary
Make the settings with the following procedure.
■
■
■
PC settings
IP address setting
.
.
.
Subnet mask setting
.
.
.
Internet Explorer setting
Amount of disk space to use
□done
Adding CBM-20A(lite) to local intranet zone
□done
Adding SPD-M20A to local intranet zone
□done
“Run ActiveX controls and plug-ins” setting
□done
“Active scripting” setting
□done
“Access data sources across domains” setting
□done
CBM-20A(lite)/ SPD-M20A settings
Items
CBM-20A
SPD-M20A
IP address setting
.
.
.
.
.
.
Subnet mask setting
.
.
.
.
.
.
System name setting
Group name setting
Master Name setting
Communication Mode setting
□LCsolution □CLASS-VP
Interface setting
□Ethernet □RS-232C
Set time from PC clock
□done
Help file setting
□done
[With CBM-20Alite only] CBM LINK setting
□done
1
1 Setting IP Address for PC
This section describes how to change the IP address of PC.
CBM-20A(lite) and SPD-M20A have the following IP
address and Subnet mask set at the time of shipment from
the factory.
Table 1 Initial value of the instrument and the IP address
IP address
Subnet mask
CBM-20A(lite)
192.168.200.99
255.255.255.0
SPD-M20A
192.168.200.98
255.255.255.0
■Changing the IP address
1. Open Control Panel from Start.
2. From Control Panel, display the Property of Local
Area Connection Properties window with a series of
operation below:
-
Click on Network and Internet Connections in
Control Panel.
-
In the Network and Internet Connections window,
click on Network Connections.
-
In the Network Connections window, right click
on the Local Area Connection to display the pop
up menu, and then click on Properties button.
3. In the Local Area Connection Properties dialog box,
select Internet Protocol (TCP/IP) and click on the
Properties button.
4. Set the IP address and Subnet mask.
If the factory setting of the CBM-20A(lite) and
[Notice]
CBM-20A(lite)/SPD-M20A shipped from the factory have
IP addresses "192.168.200.99" and "192.168.200.98" in
factory settings. When any PC in the existing LAN has one
of these IP addresses, collision of IP address occurs, which
prevents from configuring any parameters in
CBM-20A(lite)/SPD-M20A. In such a case, connect
CBM-20A(lite)/SPD-M20A with the PC in a closed
configuration (in isolation from the existing LAN), set
their IP addresses, then connect them to the existing LAN.
In addition, collision of IP address occurs when two or
more CBM-20A(lite)/SPD-M20As in factory settings are
connected simultaneously. In this case, configure
CBM-20A(lite)/SPD-M20A one by one without connecting
them to the LAN at once.
SPD-M20A have not been changed, input the IP
address and the subnet mask for the PC as show
below.
Items
Set Values
IP address
192.168.200.90
Subnet mask
255.255.255.0
Click on the OK button after settings are made.
You may be requested to reboot the PC. In that case,
follow the instruction and reboot your PC.
*
The above explanation is given provided that the OS of your PC is Windows XP and Control Panel shows Category View.
*
When you have changed CBM-20A(lite) and SPD-M20A settings, if necessary, change the IP address and Subnet mask of PC to
confirm that connection can be made.
2
2 Setting Internet Explorer for PC
■ Setting Internet Explorer
Set the CBM-20A(lite) and SPD-M20A using the Internet
Explorer. Internet Explorer requires IE6 SP1. Also it is
required to change the IE settings including Amount of
disk space to use, Security, Advanced etc. Change the IE
settings from the Internet Option screen.
[Setting the disk capacity]
Start up IE and click on Tool in the menu followed by
Internet Option to display the Internet Option dialog box.
Setting of the Amount of disk space to use
■
1)
Click on the Settings button of the Temporary
Internet files folder in the General tab to display
the setting dialog.
2)
Input “100” in the disk capacity to use and click on
the OK button.
Security setting (1)
■
1)
Select [
[Setting Security (1)]
] in the Security tab and click on the
Site button.
2)
Click on the Advanced button.
3)
Add URLs of CBM-20A(lite) and SPD-M20A.
[Setting the URL of CBM-20A]
Input “http://192.168.200.99” and press the Add
button.
[Setting the URL of SPD-M20A]
Input “http://192.168.200.98” and press the Add
button. After input, click on the OK button to
confirm the set contents.
Security setting (2)
■
[Security Setting (2)]
Click on the Custom Level button in the Security tab.
[Advanced]
Set the following items from the setting list.
Items
Setting description
Run ActiveX control and plug-ins
“Enable”
Active scripting
“Enable”
Access data source across domains
“Enable” or “Prompt”
■
You may be warned with a message saying “Are you sure
you want to change the security setting for this zone?”
Select Yes in that case.
Advanced
1)
Display the Advanced tab.
2)
Browse to the checkbox “Display a notification
about every script error” and select this option.
*
Change the URL added in Security setting (1) so that new URL is included in the zone after setting IP of the SPD-M20A.
Delete old URL and add new URL.
3
3 Setting the CBM-20A(lite)
■
Setting IP address of CBM-20A(lite)
Connect to CBM-20A(lite) using Internet Explorer and
change the IP address.
1)
Start IE and input the URL of CBM-20A(lite).
Input “http://192.168.200.99” when newly installed.
2)
Click on System Name on the Status page.
3)
Login the Group_Setting window with the
Administrator authority. When newly installed, input
User ID: Admin and Password: Admin.
Click on Group Settings button. after inputting User
Name and Password.
4)
Set System Name and Group Name.
5)
Set IP address and Subnet Mask.
Clicking on the Advanced Settings button displays
Advance Settings dialog box. The default gateway can
be set.
■ Setting Communication Mode and interface
When shipped out from the factory, the Communication
Mode is set to LCsolution and Interface is set to Ethernet.
If used in the setting other than this, change the settings
with the procedure below.
1)
Connect with Internet Explorer, press System Name
in the Status page of CBM-20A to login.
2)
3)
Click on System Controller.
In the Data Processing Setting System, set
Communication Mode and Interface.
The following items can be selected.
Communication Mode: LCsolution, CLASS-VP
Interface: Ethernet, RS-232C
Setting of the IP address, communication mode, and
interface become effective from the next time when
CMB-20A(lite) starts up.
In addition to the above settings, perform the following
operation.
Checking the case to use CBM-20Alite
Set time from PC clock
In order to have CBM-20Alite link to the instrument
Click on the Set time from PC clock button in the
to which CBM-20Alite is installed, it is necessary to
System Controller tab to set the time.
set CBM LINK at LC to INT setting at system
Setting the Help file
window on LC side. Address (5) will be automatically
Copies the folder below the \ProgramFiles in the
set as the link address.
CD-ROM provided with CBM-20A to \ProgramFiles.
*
If the following units are connected to CBM-20A(lite), it is possible to set IP address and Subnet mask directly from each unit.
For details, see “3.1.2 Setting IP address” in Instruction Manual (Basics) for CBM-20A/20Alite.
Pumps: LC-20A series pumps (LC-20AD, LC-20AT, LC-20AB), Auto sampler: SIL-20A(C), and Detector: SPD-20A/20AV
4
4 Setting the SPD-M20A
■
Setting the IP address of SPD-M20A
Connect to SPD-M20A using Internet Explorer to change
the IP address.
1)
Open the setting screen from IE.
Input “http://192.168.200.98” when newly installed.
2)
Log in as Administrator.
Input user ID and password, and press down the
Login button. When newly installed, input User ID:
administrator and Password: pdaadmin.
3)
Change the IP address and Subnet mask
Click on the Edit button of the Administrator Setting
window.
Input IP address and Subnet mask from the Network
Setup window.
After input, click on the OK button.
4)
Restart SPD-M20A.
The changed IP address becomes valid from the next
time SPD-M20A starts up.
■
Setting Group name and Master name
When starting analysis synchronously with the
CBM-20A(lite), change the Group Name and Master
Name.
Perform setting from the Administrator Setting window.
- Set the same Group Name as CBM.
- Set System name of CBM to the Master Name.
(Setting Example)
Setting of
Setting of
CBM
SPD-M20A
Group Name
HPLC1
HPLC1
System Name
CBM1
M20A1
Master Name
CBM1
5