High Resolution X-ray Imaging System

High Resolution X-ray Imaging System
High Resolution
X-ray Imaging System
▲M11427-62 (Camera: ORCA-Flash4.0 V3)
The high resolution X-ray imaging system is designed
for the application of X-ray beam alignment.
Adopting unique mechanism, it enables to combine
various type of cameras for real time X-ray beam
alignment. Beryllium input window realized wide range
of X-ray energy and coupled with L-shaped quartz
optics makes high X-ray radiation tolerance.
Suitable for high resolution X-ray beam alignment in
large synchrotron radiation facilities.
PRINCIPLE
Beryllium X-rays
window
Scintillator
Objectives
Imaging device
The irradiated X-ray is converted to visible light and conducted
to the lens bended 90 degrees by L-shaped mirror in order to
reduce the X-ray damage onto the digital camera.
FEATURES
X-ray proof design
(adopting a quartz glass plate and L-shaped optics)
Easy to exchange scintillators
Possible to detect low X-ray energy
Small size and light weight
Remote controllable focus adjustment
One touch design for camera attachment
High resolution
APPLICATIONS
X-ray CT
Phase contrast X-ray CT
X-ray optics alignment
X-ray topography
X-ray microscope
Angiography
XAFS
Imaging unit selection
M11427-41, -42
(Standard type)
High resolution X-ray imaging system AA40
Resolution: 10 μm
Scintillator diameter: 17 mm
PRINCIPLE
X-rays
Beryllium
window
Scintillator
P43 (Gd2O2S: Tb)
Visible light
Mirror
M11427-51, -52, -53
Relay lens
Option
Scintillator: CsI
(Small area type)
High resolution X-ray imaging system AA50
Resolution: 2 μm
Scintillator diameter: 10 mm
PRINCIPLE
Beryllium
window
Visible light
Microscope
objective lens
X-rays
Mirror
Amorphous
carbon plate
Single crystal scintillator
(Lu2SiO5: Ce, thickness < 10 μm)
M11427-62
Option
Scintillator: LuAG
(Large area type)
High resolution X-ray imaging system AA60
Resolution: 10 μm or more
Scintillator diameter: 35 mm
PRINCIPLE
X-rays
Scintillator
Glassy carbon
P43 (Gd2O2S: Tb)
window
Visible light
Mirror
Relay lens
Option
Scintillator: CsI
SPECIFICATIONS
M11427-41
Type number
M11427-42
M11427-51 M11427-52 M11427-53
M11427-62
Imaging unit
AA40
AA50
AA60
Scintillator diameter
17 mm
10 mm
35 mm
Beryllium (0.5 mm)
Beryllium (0.5 mm) / Amorphous carbon (2 mm)
Glassy carbon (0.5 mm)
3 keV or more
3 keV or more
6 keV or more
Input window material
Sensitivity range
Scintillator material
P43 (Gd2O2S: Tb)
LSO (Lu2SiO5: Ce)
P43 (Gd2O2S: Tb)
Peak emission wavelength
540 nm
420 nm
540 nm
Minimum thickness of scintillator
10 μm
10 μm
10 μm
Substrate material of scintillator
Quart glass
Amorphous carbon plate
Quart glass
2 μm
1 μm
10 μm or more
50× (NA 0.55)
75 mm (F2.8)
Resolution *
10 μm
10 % Decay time
1 ms
40 ns
10× (NA 0.3)
20× (NA 0.4)
1 ms
First lens
24 mm
Second lens
105 mm, 50 mm, 35 mm
-
105 mm, 50 mm, 35 mm
ND10 / ND1
-
ND10 / ND1
ND filter
50 mm (F1.4)
4 μm
* Depending on the camera type
DIMENSIONAL OUTLINES
(Unit: mm)
High resolution X-ray imaging system AA40
High resolution X-ray imaging system AA50
(Approx. 5.2 kg)
(Approx. 3.7 kg)
95
43
165
115.5
107
23.5
415
High resolution X-ray imaging system AA60
(Approx. 7.5 kg)
180
142
105
114
492
184
Camera selection
R
ORCA-Flash4.0 V3
ImagEM X2-1K
High speed, High resolution,
Low readout noise
High sensitivity
R
Field of view
Camera lens
(f=50 mm)
M11427-42 (f=50 mm)
13.3 mm × 13.3 mm
M11427-52 (20×)
0.665 mm × 0.665 mm
M11427-62 (f=75 mm)
19.95 mm × 19.95 mm
Field of view
Camera lens
(f=50 mm)
M11427-42 (f=50 mm)
13.3 mm × 13.3 mm
M11427-52 (20×)
0.665 mm × 0.665 mm
M11427-62 (f=75 mm)
19.95 mm × 19.95 mm
* Pictures show camera and camera lens. * The field of view is based on calculation. Actual field of view may be different from the calculated values.
SPECIFICATIONS
ORCA-Flash4.0 V3 (C13440-20CU)
ImagEM X2-1K (C9100-24B)
Scientific CMOS image sensor
2048 (H) × 2048 (V)
6.5 μm (H) × 6.5 μm (V)
13.3 mm (H) × 13.3 mm (V)
30 000 electrons
100 frames/s (Camera Link), 40 frames/s (USB 3.0)
1.4 electrons (slowscan)
0.06 electrons/pixel/s (-10 ˚C)
0.006 electrons/pixel/s (-30 ˚C)
Forced air
Water cooled
-10 ˚C (Ambient temperature: +20 ˚C)
-30 ˚C (Water temperature: +15 ˚C)
16 bit
Camera Link full configuration Deca mode / USB 3.0
Electron multiplying back-thinned frame transfer
1024 (H) × 1024 (V)
13 μm (H) × 13 μm (V)
13.3 mm (H) × 13.3 mm (V)
400 000 electrons (EM-CCD mode)
18.5 to 314 frames/s
3 electrons (0.6875 MHz)
1 electron max. (0.6875 MHz)
10 electrons (0.6875 MHz)
0.01 electrons/pixel/s (-50 ˚C)
0.001 electrons/pixel/s (-70 ˚C)
Forced air
Water cooled
-50 ˚C (Ambient temperature: 0 ˚C to +30 ˚C) -70 ˚C (Water temperature: +20 ˚C)
16 bit
IEEE1394b
Camera type number
Imaging device
Effective number of pixels
Cell size
Effective area
Full well capacity
Readout speed
Readout noise EM 10×
(rms)
EM 1200×
Normal
Air cooled
Dark current
Water cooled
Cooling method
Cooling temperature
Digital output
Interface
*Refer to the camera catalog for detail information.
ORCA is registered trademark of Hamamatsu Photonics K.K. (France, Germany, Japan, U.K., U.S.A.)
ImagEM is registered trademark of Hamamatsu Photonics K.K. (EU, Japan, U.K., U.S.A.)
Product and software package names noted in this documentation are trademarks or registered trademarks of their respective manufacturers.
Subject to local technical requirements and regulations, availability of products included in this promotional material may vary. Please consult your local sales representative.
Information furnished by HAMAMATSU is believed to be reliable. However, no responsibility is assumed for possible inaccuracies or omissions.
Specifications and external appearance are subject to change without notice.
© 2017 Hamamatsu Photonics K.K.
HAMAMATSU PHOTONICS K.K.
www.hamamatsu.com
HAMAMATSU PHOTONICS K.K., Systems Division
812 Joko-cho, Higashi-ku, Hamamatsu City, 431-3196, Japan, Telephone: (81)53-431-0124, Fax: (81)53-435-1574, E-mail: export@sys.hpk.co.jp
U.S.A.: Hamamatsu Corporation: 360 Foothill Road, Bridgewater, NJ 08807, U.S.A., Telephone: (1)908-231-0960, Fax: (1)908-231-1218 E-mail: usa@hamamatsu.com
Germany: Hamamatsu Photonics Deutschland GmbH.: Arzbergerstr. 10, D-82211 Herrsching am Ammersee, Germany, Telephone: (49)8152-375-0, Fax: (49)8152-265-8 E-mail: info@hamamatsu.de
France: Hamamatsu Photonics France S.A.R.L.: 19, Rue du Saule Trapu, Parc du Moulin de Massy, 91882 Massy Cedex, France, Telephone: (33)1 69 53 71 00, Fax: (33)1 69 53 71 10 E-mail: infos@hamamatsu.fr
United Kingdom: Hamamatsu Photonics UK Limited: 2 Howard Court,10 Tewin Road, Welwyn Garden City, Hertfordshire AL7 1BW, UK, Telephone: (44)1707-294888, Fax: (44)1707-325777 E-mail: info@hamamatsu.co.uk
North Europe: Hamamatsu Photonics Norden AB: Torshamnsgatan 35 16440 Kista, Sweden, Telephone: (46)8-509-031-00, Fax: (46)8-509-031-01 E-mail: info@hamamatsu.se
Italy: Hamamatsu Photonics Italia S.r.l.: Strada della Moia, 1 int. 6, 20020 Arese (Milano), Italy, Telephone: (39)02-935-81-733, Fax: (39)02-935-81-741 E-mail: info@hamamatsu.it
China: Hamamatsu Photonics (China) Co., Ltd.: 1201 Tower B, Jiaming Center, 27 Dongsanhuan Beilu, Chaoyang District, 100020 Beijing, China, Telephone: (86)10-6586-6006, Fax: (86)10-6586-2866 E-mail: hpc@hamamatsu.com.cn
Taiwan: Hamamatsu Photonics Taiwan Co., Ltd.: 8F-3, No.158, Section2, Gongdao 5th Road, East District, Hsinchu, 300, Taiwan R.O.C. Telephone: (886)03-659-0080, Fax: (886)07-811-7238 E-mail: info@tw.hpk.co.jp
Cat. No. SCAS0011E18
MAR/2017 HPK
Created in Japan
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project