2 Pt. Acoustical Calibrator 850016

2 Pt. Acoustical Calibrator 850016
2 Pt. Acoustical
Calibrator
850016
Instruction Manual
2 Pt. Acoustical Calibrator 850016
Copyright ©2013 by
Sper Scientific
ALL RIGHTS RESERVED
Printed in the USA
The contents of this manual may not be reproduced
or transmitted in any form or by any means
electronic, mechanical, or other means that do
not yet exist or may be developed, including
photocopying, recording, or any information storage
and retrieval system without the express permission
from Sper Scientific.
2
TABLE OF CONTENTS
1. INTRODUCTION . . . . . . . . . . . . . . . . 2. INSTRUMENT DESCRIPTION . . . . . . . . . 3. OPERATING PROCEDURE . . . . . . . . . . .
4. BATTERY INSTALLATION/REPLACEMENT . . 5. SPECIFICATIONS . . . . . . . . . . . . . . . .
6. WARRANTY . . . . . . . . . . . . . . . . . . .
3
4
5
6
7
8
1. INTRODUCTION
Calibration is suggested by OSHA to ensure good
sampling practices. The 850016 Sper Scientific
2-Point Acoustical Calibrator is designed to calibrate
Sper Scientific digital sound meters, or any other
sound meter that is user-calibrated and has a ½" or
1" microphone. Features 2-point calibration at 94 dB
and 114 dB, and a low battery indicator. Comes with
instructions, 9V battery and a carrying case.
3
2. INSTRUMENT DESCRIPTION OPERATING
1 - 1" Microphone Insertion Socket
2 - 1/2" Microphone Adapter
3 - Power Indicator
4 - 94dB/114dB/Off
Slide Switch
5 - Battery compartment (lift out for access)
1
6 - Battery compartment ring
2
3
4
5
4
6
3. OPERATING PROCEDURE
Note: For accurate calibration, background noise
should be less than 10 dB.
• Determine the sound meter’s microphone size.
• Use the 1/2" Microphone Adapter (2) when
calibrating a 1/2" diameter microphone.
• When calibrating a sound meter with a 1"
microphone, pull off the 1/2" Microphone Adapter
(2) to use the 1" Microphone Insertion Socket (1).
• Fully insert the sound meter’s microphone into
the insertion socket.
• Use the 94dB/114dB/Off Slide Switch (4) to turn
the calibrator on and select either 94 dB or 114 dB.
• Be sure the sound meter is set to an appropriate
scale to measure a matching sound of 94 dB
or 114 db.
• Adjust the sound meter according to it’s
calibration instructions.
5
4. BATTERY INSTALLATION AND REPLACEMENT
• Replace the battery if the Power Indicator (3) does
NOT light when 94dB/114dB/Off Slide Switch (4) is
in the 94 dB or 114 dB position and/or when there is
no audible tone.
• Unscrew the Battery Compartment Ring (5).
• There are wires connected to the Battery
Compartment (6), so SLOWLY lift the center to
access the battery.
• Insert a fresh 9V Alkaline battery.
• Replace the Battery Compartment (6) and tighten
the Battery Compartment Ring (5).
• A fresh Alkaline battery should power your calibrator
for approximately 20 hours of use.
• Remove the battery if the calibrator will not be in use
for a long period of time.
• Tampering with the internal circuitry will void
your warranty.
6
5. SPECIFICATIONS
Standard:
Output Sound Level:
Output Frequency:
Accuracy:
Operating
Environment:
Storage Environment:
Test Conditions
Atmospheric Pressure:
Dimensions:
Weight:
ANSI S1.40-1984 and
IEC942 1988 Class 2
94 dB and 114 dB
re 20 uPA under reference
1kHz ±4%
± 0.5 dB
32~104°F (0~40°C)
Altitude up to 2000 meters
10~70% RH
14~140°F (-10~60°C)
10~90% RH
20°C
65% RH
1013 mbar
1 7/8” x 5” (48 x 125 mm)
10 oz (285g)
7
6. WARRANTY
Sper Scientific warrants this product against defects
in materials and workmanship for five (5) years from
the date of purchase, and agrees to repair or replace
any defective unit without charge. If your model has
since been discontinued, an equivalent Sper Scientific
product will be substituted if available. This warranty
does not cover damage resulting from accident,
misuse, or abuse of the product.
To obtain warranty service, ship
the unit postage prepaid to:
SPER SCIENTIFIC LTD.
8281 East Evans Road, Suite #103
Scottsdale, AZ 85260
The defective unit must be accompanied by a
description of the problem and your return address.