Keysight M9187A PXI Digital IO Module Startup Guide
Below you will find brief information for Digital IO Module PXI M9187A, Connector Block Y1181A, Connector Cable Y1187A, Connector Cable Y1188A, PXIe Chassis M9018A, PCIe Cable Interface Module M9021A, PCIe Desktop Adapter M9047A, PCIe ExpressCard Adapter M9045A, PXI Connector Cable Y1202A, PXI Connector Cable Y1200A. The Keysight PXI M9187A Digital IO module provides 32 high output voltage and current channels along with 32 dual variable threshold input channels. The module also includes a 78 pin D-type connector on the front panel. For detailed usage instructions, refer to the M9187A Digital Soft Front Panel help file on the Digital IO Module Software and Product Information CD.
Advertisement
Advertisement
Startup Guide
Keysight M9187A
PXI Digital IO Module
Notices
© Keysight Technologies, Inc. 2014
No part of this manual may be reproduced in any form or by any means
(including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Keysight Technologies, Inc. as governed by United States and international copyright laws.
Manual Part Number
M9187-90001
Edition
Third Edition, December 2014
Published in U.S.A.
900 S. Taft Avenue,
Loveland, CO. 80537
Trademarks
PICMIG
TCA
®
®
, COMPACT PCI
®
, and Advanced
are registered trademarks of the
PCI Industrial Computer Manufacturers
Group
PCI-SIG
®
, PCI Express
®
, and PCIe registered trademarks of PCI-SIG
®
are
Sales and Technical Support
To contact Keysight for sales and technical support, refer to the support links on the following Keysight websites: www.keysight.com/find/M9187A (product-specific information and support, software and documentation updates) www.keysight.com/find/assist (worldwide contact information for repair and service)
Declaration of Conformity
Declarations of Conformity for this product and for other Keysight products may be downloaded from the Web. Go to http://www.keysight.com/go/conformity and click on “Declarations of Conformity.”
You can then search by product number to find the latest Declaration of Conformity.
T
echnology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
Warranty
THE MATERIAL CONTAINED IN THIS
DOCUMENT IS PROVIDED “AS IS,” AND
IS SUBJECT TO BEING CHANGED,
WITHOUT NOTICE, IN FUTURE EDI-
TIONS. FURTHER, TO THE MAXIMUM
EXTENT PERMITTED BY APPLICABLE
LAW, KEYSIGHT DISCLAIMS ALL WAR-
RANTIES, EITHER EXPRESS OR IMPLIED,
WITH REGARD TO THIS MANUAL AND
ANY INFORMATION CONTAINED
HEREIN, INCLUDING BUT NOT LIMITED
TO THE IMPLIED WARRANTIES OF MER-
CHANTABILITY AND FITNESS FOR A
PARTICULAR PURPOSE. KEYSIGHT
SHALL NOT BE LIABLE FOR ERRORS OR
FOR INCIDENTAL OR CONSEQUENTIAL
DAMAGES IN CONNECTION WITH THE
FURNISHING, USE, OR PERFORMANCE
OF THIS DOCUMENT OR OF ANY INFOR-
MATION CONTAINED HEREIN. SHOULD
KEYSIGHT AND THE USER HAVE A SEP-
ARATE WRITTEN AGREEMENT WITH
WARRANTY TERMS COVERING THE
MATERIAL IN THIS DOCUMENT THAT
CONFLICT WITH THESE TERMS, THE
WARRANTY TERMS IN THE SEPARATE
AGREEMENT SHALL CONTROL.
Keysight Technologies does not warrant third-party system-level (combination of chassis, controllers, modules, etc.) performance, safety, or regulatory compliance unless specifically stated.
DFARS/Restricted Rights
Notices
If software is for use in the performance of a U.S. Government prime contract or subcontract, Software is delivered and licensed as “Commercial computer software” as defined in DFAR 252.227-7014
(June 1995), or as a “commercial item” as defined in FAR 2.101(a) or as “Restricted computer software” as defined in FAR
52.227-19 (June 1987) or any equivalent agency regulation or contract clause.
Use, duplication or disclosure of Software is subject to Keysight Technologies’ standard commercial license terms, and non-
DOD Departments and Agencies of the
U.S. Government will receive no greater than Restricted Rights as defined in FAR
52.227-19(c)(1-2) (June 1987). U.S. Government users will receive no greater than Limited Rights as defined in FAR
52.227-14 (June 1987) or DFAR 252.227-
7015 (b)(2) (November 1995), as applicable in any technical data.
Safety Information
The following general safety precautions must be observed during all phases of operation of this instrument.
Failure to comply with these precautions or with specific warnings or operating instructions in the product manuals violates safety standards of design, manufacture, and intended use of the instrument. Keysight Technologies assumes no liability for the customer's failure to comply with these requirements.
General
Do not use this product in any manner not specified by the manufacturer. The protective features of this product must not be impaired if it is used in a manner specified in the operation instructions.
Before Applying Power
Verify that all safety precautions are taken.
Make all connections to the unit before applying power. Note the external markings described under “Safety Symbols”.
Ground the Instrument
Keysight chassis’ are provided with a grounding-type power plug. The instrument chassis and cover must be connected to an electrical ground to minimize shock hazard. The ground pin must be firmly connected to an electrical ground (safety ground) terminal at the power outlet. Any interruption of the protective (grounding) conductor or disconnection of the protective earth terminal will cause a potential shock hazard that could result in personal injury.
Do not operate the module/chassis in the presence of flammable gases or fumes.
Do not operate the module/chassis in the presence of flammable liquids or near containers of such liquids.
Cleaning
Clean the outside of the Keysight module/chassis with a soft, lint-free, slightly dampened cloth. Do not use detergent or chemical solvents.
Do Not Remove Instrument Cover
Only qualified, service-trained personnel who are aware of the hazards involved should remove instrument covers. Always disconnect the power cable and any external circuits before removing the instrument cover.
Keep away from live circuits
Operating personnel must not remove equipment covers or shields. Procedures involving the removal of covers and shields are for use by servicetrained personnel only. Under certain conditions, dangerous voltages may exist even with the equipment switched off. To avoid dangerous electrical shock, DO NOT perform procedures involving cover or shield removal unless you are qualified to do so.
DO NOT operate damaged equipment
Whenever it is possible that the safety protection features built into this product have been impaired, either through physical damage, excessive moisture, or any other reason, REMOVE POWER and do not use the product until safe operation can be verified by servicetrained personnel. If necessary, return the product to an Keysight Technologies Sales and Service Office for service and repair to ensure the safety features are maintained.
DO NOT block the primary disconnect
The primary disconnect device is the appliance connector/power cord when a chassis used by itself, but when installed into a rack or system the disconnect may be impaired and must be considered part of the installation.
Do Not Modify the Instrument
Do not install substitute parts or perform any unauthorized modification to the product. Return the product to an
Keysight Sales and Service Office to ensure that safety features are maintained.
In Case of Damage
Instruments that appear damaged or defective should be made inoperative and secured against unintended operation until they can be repaired by qualified service personnel
Do NOT block vents and fan exhaust:
To ensure adequate cooling and ventilation, leave a gap of at least 50mm
(2") around vent holes on both sides of the chassis.
Do NOT operate with empty slots: To ensure proper cooling and avoid damaging equipment, fill each empty slot with an AXIe filler panel module.
Do NOT stack free-standing chassis:
Stacked chassis should be rackmounted.
All modules are grounded through the chassis: During installation, tighten each module's retaining screws to secure the module to the chassis and to make the ground connection.
Operator is responsible to maintain safe operating conditions. To ensure safe operating conditions, modules should not be operated beyond the full temperature range specified in the
Environmental and physical specification. Exceeding safe operating conditions can result in shorter lifespan, improper module performance and user safety issues. When the modules are in use and operation within the specified full temperature range is not maintained, module surface temperatures may exceed safe handling conditions which can cause discomfort or burns if touched. In the event of a module exceeding the full temperature range, always allow the module to cool before touching or removing modules from the chassis.
iv
Safety Symbols
A CAUTION denotes a hazard. It calls attention to an operating procedure or practice, that, if not correctly performed or adhered to could result in damage to the product or loss of important data.
Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
A WARNING denotes a hazard. It calls attention to an operating procedure or practice, that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a
WARNING notice until the indicated conditions are fully understood and met.
Products display the following symbols:
The instrument has been tested, investigated and found to comply with the requirements of the Standard(s) for
Electrical Measuring & Test Equipment.
ICES/NMB-001
ISM GRP 1-A
Notice for European Community: This product complies with the relevant
European legal Directives: EMC Directive (2004/108/EC) and Low Voltage
Directive (2006/95/EC).
This is the symbol for an Industrial,
Scientific, and Medical Group 1 Class A product.
Waste Electrical and
Electronic
Equipment (WEEE)
Directive
2002/96/EC
This product complies with the WEEE
Directive (2002/96/EC) marking requirement. The affixed product label
(see below) indicates that you must not discard this electrical/electronic product in domestic household waste.
Product Category: With reference to the equipment types in the WEEE directive Annex 1, this product is classified as a “Monitoring and Control instrumentation” product.
Do not dispose in domestic household waste.
To return unwanted products, contact your local Keysight office, or see www.keysight.com/environment/product for more information.
Warning, risk of electric shock
Refer to manual for additional safety information.
Earth Ground.
The Regulatory Compliance Mark
(RCM) mark is a registered trademark.
This signifies compliance with the Australia EMC Framework regulations under the terms of the Radio Communication Act of 1992.
Chassis Ground.
Alternating Current (AC).
Standby Power. Unit is not completely disconnected from AC mains when switch is in standby.
Antistatic precautions should be taken.
ICES/NMB-001 indicates that this ISM device complies with the Canadian
ICES-001.
For localized Safety Warnings, Refer to Keysight Safety document (p/n
9320-6792).
This symbol represents the time period during which no hazardous or toxic substance elements are expected to leak or deteriorate during normal use.
Forty years is the expected useful life of this product.
v
vi
Contents
Keysight PXI Digital IO Module Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Step 1: Unpack and Inspect the Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Step 3: Install the Software on the Host Computer . . . . . . . . . . . . . . . . . . . . . 12
Step 4. Connect the Controller to the PXI Chassis. . . . . . . . . . . . . . . . . . . . . . 16
Using an embedded computer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Step 5: Install the Module in the PXI Chassis. . . . . . . . . . . . . . . . . . . . . . . . . . 17
Run Keysight IO Libraries Connection Expert . . . . . . . . . . . . . . . . . . . . . . . 19
Module Front Panel, Connector and Pin Out. . . . . . . . . . . . . . . . . . . . . . . . 23
Y1181A 78 Pin Shielded Connector Block. . . . . . . . . . . . . . . . . . . . . . . . . . 26
Y1187A, Y1188A 78 Pin Male to Female Connector Cable. . . . . . . . . . . . . 27
Disassembling the Connector Block Clam Shells . . . . . . . . . . . . . . . . . . . . 28
Keysight M9187A PXI Digital IO Module Startup Guide vii
viii Keysight M9187A PXI Digital IO Module Startup Guide
M9187A PXI Digital IO Module
Startup Guide
Keysight PXI Digital IO Module Introduction
The Keysight M9187A PXI Digital IO Module provides 32 high output voltage and current channels along with 32 dual variable threshold input channels.
- Each of the output channels can drive the output either high (sourcing up to
0.4A) or low (sinking up to 0.5A), or three-stated (off). All output channels are protected against over-voltage, over-current, and thermal overload. An external power supply is required for high-side driving.
- The 32 input channels are read by comparing the input DC voltage to two threshold voltages that can be set between 0.3 and 50Vdc. The thresholds can programmatically be set with 12.5mVdc resolution, permitting input testing against the system thresholds that define a logic low or logic high.
Note that the two thresholds are the same for all 32 input channels -- the thresholds are not independently set for each input channels.
Inputs to the Keysight M9187A module are DC voltage only.
- The inputs can be connected directly to the outputs without risk of damage, allowing module configuration for operation as 32 channels of independent input and output or as 32 channels of configurable I/O.
- User connection is a 78 pin D-type connector on the module front panel.
Keysight also supplies software drivers allowing you to support the module in all popular PXI chassis’ and programming environments. Soft Front Panel software allows you to exercise the channels for test purposes.
For detailed usage instructions, refer to the M9187A Digital Soft Front Panel help file on the Digital IO Module Software and Product Information CD.
Related documentation
This Startup Guide, and the documentation listed below, are on the Digital IO
Module Software and Product Information CD.
- Help file for the Digital IO Module Soft Front Panel
- Help file for the Digital IO Module IVI-C/IVI-COM device drivers
- Help file for the Digital IO Module LabVIEW G device drivers
- Keysight PXI Digital IO Module Specifications. For the latest specifications, check the Keysight web site at: www.keysight.com/find/pxi-dio .
9
Step 1: Unpack and Inspect the Module
Step 1: Unpack and Inspect the Module
Keysight’s PXI Modules are shipped in materials that prevent static electricity damage. The modules should only be removed from the packaging in an anti-static area ensuring that correct anti-static precautions are taken. Store all modules in anti-static envelopes when not in use.
ESD
Electrostatic discharge (ESD) can damage or destroy electronic components. All work on electronic assemblies should be performed at a static-safe work station.
The following figure shows an example of a static-safe work station using two types of ESD protection.
- Conductive table-mat and wrist-strap combination.
- Conductive floor-mat and heel-strap combination.
Both types, when used together, provide a significant level of ESD protection. Of the two, only the table-mat and wrist-strap combination provides adequate
ESD protection when used alone.
To ensure user safety, the static-safe accessories must provide at least 1 M of isolation from ground. Purchase acceptable ESD accessories from your local supplier.
10 Keysight M9187A PXI Digital IO Startup Guide
Step 1: Unpack and Inspect the Module
Inspect for damage
After unpacking the module, carefully inspect it for any shipping damage. Report any damage to the shipping agent immediately, as such damage is not covered by the warranty.
To avoid damage when handling a module; do not touch exposed connector pins.
Return the module for service
Should it become necessary to return a Keysight module for repair or service, follow the steps below:
1 Review the warranty information shipped with your product.
2 Contact Keysight to obtain a return authorization and return address. If you need assistance finding Keysight contact information go to www.keysight.com/find/assist (worldwide contact information for repair and service) or refer to the Support information on the product web page at: www.keysight.com/find/pxi-dio .
3 Write the following information on a tag and attach it to the malfunctioning equipment.
– Name and address of owner. A Post Office box is not acceptable as a return address.
– Product model number (M9187A).
– Product serial number (for example, MYXXXXXXXX). The serial number label is located on the side of the module.
– A description of failure or service required.
4 Carefully pack the module in its original ESD bag and carton. If the original carton is not available, use bubble wrap or packing peanuts, place the instrument in a sealed container and mark the container “FRAGILE”.
5 On the shipping label, write ATTENTION REPAIR DEPARTMENT and the service order number (if known).
If any correspondence is required, refer to the product by serial number and model number.
Keysight M9187A PXI Digital IO Startup Guide 11
Step 2: Verify Shipment Contents
Step 2: Verify Shipment Contents
Your shipment should have included the following:
- The Keysight PXI M9187A Digital IO module that you ordered.
- This document (Keysight PXI M9187A Module Startup Guide).
- A Digital IO Module Software and Product Information CD. This CD contains software, drivers and all product printed documentation in PDF format for the
PXI Digital IO Module.
- An Automation-Ready CD with Keysight IO Libraries Suite (version 16.0 or later).
- Any other accessories that you ordered (cables, connectors, etc.).
Step 3: Install the Software on the Host Computer
The following illustration shows typical system installations.
Controller
:
Laptop PC
System Interface
:
M9045
PCIe Laptop
Adapter
Y1200A
PCIe Cable
M9021
Cable Interface
Module
PXI System Chassis
M9018A
18 Slot PXIe Chassis
:
Desk mount or
Rack mount PC
M9047
PCIe Desktop
Adapter
Y1202A
PCIe Cable
PXIe Embedded
System Controller
12
Do not install the PXI modules in the PXI chassis yet! You must install the software prior to installing the modules in the chassis so that Keysight IO Libraries Connection Expert finds them.
Keysight M9187A PXI Digital IO Startup Guide
Step 3: Install the Software on the Host Computer
System requirements
Operating
System
Processor
Speed
Available memory
Available hard disk space
*
Video
Browser
The following table lists the minimum system requirements for Keysight IO
Libraries Suite 16. In general, any x86 or x64 (except Itanium) should work but there may be a significant decrease in performance.
Windows XP Service Pack 3 or later
600 MHz or higher required,
800 MHz recommended
256 MB minimum
(1 GB or greater recommended)
Windows Vista SP1 and SP2
(32-bit and 64-bit), Business,
Ultimate, Enterprise, Home
Basic, and Home Premium
1Ghz 32-bit (x86),
1GHz 64-bit (x64), no support for Itanium64
1 GB minimum
Windows 7 (32- and 64-bit)
Starter, Home Basic, Home
Premium, Professional,
Ultimate, Enterprise
1Ghz 32-bit (x86),
1GHz 64-bit (x64), no support for Itanium64
1 GB minimum
1.5 GB available hard disk space, includes:
•
1GB available for Microsoft
.NET Framework 3.5 SP1
†
•
100MB for Keysight IO
Libraries Suite
Super VGA (800x600) 256 colors or more
Microsoft Internet Explorer 6.0 or greater
1.5 GB available hard disk space, includes:
•
1GB available for Microsoft
.NET Framework 3.5 SP1
2
•
100MB for Keysight IO
Libraries Suite
Support for DirectX 9 graphics with 128MB graphics memory recommended (Super VGA graphics is supported)
Microsoft Internet Explorer 7 or greater
1.5 GB available hard disk space, includes:
•
1GB available for Microsoft
.NET Framework 3.5 SP1
2
•
100MB for Keysight IO
Libraries Suite
Support for DirectX 9 graphics with 128MB graphics memory recommended (Super VGA graphics is supported)
Microsoft Internet Explorer 7 or greater
* Because of the installation procedure, less memory may be required for operation than is required for installation.
† .NET Framework Runtime Components are installed by default with Windows Vista. Therefore you may not need this amount of available disk space.
PXI System
PXI system/ host controller
A PXI or PXI Express embedded controller or PC host controller is required.
Keysight M9187A PXI Digital IO Startup Guide 13
14
Step 3: Install the Software on the Host Computer
Power the controller
- If you are using a desktop, laptop, or rack mount PC as remote controller:
1 Install any peripheral devices in the PC (e.g., PCIe Interface adapters, etc.).
Follow the manufacturers instructions.
2 Power up the controller. Do not apply power to the PXI chassis yet.
3 Choose the default option for any “Found Hardware” dialogs.
- If you are using a PXI embedded computer as the host computer:
1 Install the embedded computer module into the PXI chassis following the manufacturers instructions.
2 Connect peripherals such as a mouse, keyboard, monitor, CD drive, etc.
3 Power up the chassis.
4 Choose the default option for any “Found Hardware” dialogs
Install Keysight IO Libraries Suite
Keysight IO Libraries Suite 16.0 (or later) is required for the PXI modules. It includes the Keysight Connection Expert, the IVI Shared Components, and the
VISA Shared Components.
Keysight IO Libraries version 16.0 (or later) must be installed prior to installing and running any other software and prior to powering the chassis. The latest version can be downloaded from: www.keysight.com/find/iosuite .
1 Insert the Automation-Ready CD with Keysight IO Libraries Suite into the
CD-ROM drive of your system controller. Wait a few seconds for the auto-run window to appear. If the auto-run window does not appear automatically:
– Click Start > Run...
– Type: <drive>:Autorun\IOLibraries.hta where <drive> is your CD drive letter.
2 Follow the installation instructions that came with the IO Libraries Suite.
If the IVI Shared Components and VISA Shared Components are not already installed on your PC, Keysight IO Libraries Suite installs them in the default locations. If they are already installed, the installer upgrades them to the latest version, using the same location used by the older version. If this is a first-time installation, you can select installation locations for these components by choosing a Custom Installation.
Keysight M9187A PXI Digital IO Startup Guide
Step 3: Install the Software on the Host Computer
Install instrument drivers
1 Insert the Digital IO Module Software and Product Information CD into the
CD-ROM drive of your PC.
Wait a few seconds for the auto-run window to appear. If the auto-run window does not appear automatically:
– Click Start > Run...
– Type:
<drive>:Autorun.exe
where
<drive> is your CD drive letter.
2 Select the Install Software link. Follow the installer prompts. Accept all of the default directories specified during installation if prompted.
3 After the Welcome screen, you will be prompted three times for license agreements. The software installer installs the following drivers:
– Keysight Modular Software License Agreement. The installer program installs the Soft Front Panel (SFP) application and other object code to connect to the module. Accept the license terms and click Next.
– Keysight IVI Driver Source Code License Agreement. The Interchangeable
Virtual Instrument (IVI) driver is available for programming the Keysight modules using Microsoft development environments (e.g., Visual Studio, C,
C++, C#, Visual Basic), Keysight VEE, MATLAB, or National Instruments
Labview. Accept the license terms and click Next.
– Keysight Software License Agreement for drivers for use with LabVIEW
Software. The LAbVIEW driver provides access to the functionality of the module through LabVIEW VIs. This driver works in National Instruments
LabVIEW development environments. Before this driver can be installed, your computer must already have the IVI Shared Components installed.
Accept the license terms and click Next.
Installing Keysight IO Libraries also installs the IVI Shared
Components. The IVI Shared Components are required before IVI drivers (e.g., IVI-COM, IVI-C) can be installed from the product reference CD.
4 Next, the installer indicates the LabVIEW installations found on your host computer. Click Next
5 After accepting the licenses, the driver software is ready to be installed on your host computer. Click Install.
6 When the installation wizard is finished, you will be prompted to reboot your host computer. Do not reboot the host computer at this time! Select “No, I
will restart my computer later
.” Power down the host computer.
Proceed to “Step 4. Connect the Controller to the PXI Chassis” on the next
page.
Keysight M9187A PXI Digital IO Startup Guide 15
Step 4. Connect the Controller to the PXI Chassis
Step 4. Connect the Controller to the PXI Chassis
To ensure proper system operation and the PC’s ability to enumerate all of the PXI modules, you must use an approved embedded PC, desktop PC, or laptop PC along with approved
PCIe adaptor and cable.
Using a remote PC
Refer to the following figure. Make certain that both the PC and the PXI chassis are turned off. Unplug the chassis from the ac power mains. If you are using the
M9018A 18 Slot PXIe chassis, install the Keysight M9021A PCIe Cable Interface module in the chassis.
- If you are using a desktop or rack mount PC, install the M9047A PCIe Desktop
Adapter in the PC. With a Keysight Y1202A cable, connect the adapter to the
System Interface module.
- If you are using a laptop PC, install the M9045A PCIe ExpressCard Adapter in the laptop. With a Keysight Y1200A cable, connect the adapter to the System
.
Interface module on the PXI chassis
Controller
:
System Interface
:
M9045
PCIe Laptop
Adapter
Y1200A
PCIe Cable
M9021
Cable Interface
Module
Laptop PC
Desk mount or
Rack mount PC
PXIe Embedded
System Controller
M9047
PCIe Desktop
Adapter
Y1202A
PCIe Cable
PXI System Chassis:
18 Slot PXIe Chassis
16
Using an embedded computer
Refer to the figure above. If you are using an embedded controller in the PXI chassis, you should have installed it prior to installing the Keysight IO Libraries and instrument drivers. No cables or other adapters are required. After
installation, proceed to “Step 5: Install the Module in the PXI Chassis”
Keysight M9187A PXI Digital IO Startup Guide
Step 5: Install the Module in the PXI Chassis
Step 5: Install the Module in the PXI Chassis
- PXI hard ware does not support “hot-swap” capabilities
(changing modules while power is applied to the chassis).
- Before installing Keysight PXI Modules into the chassis, the chassis must be powered off to prevent damage to the PXI module. Remove all cables/terminal blocks from the module prior to installing or removing the module.
The module can be used in a chassis with a cPCI, PXI-1, or PXIh chassis peripheral slot.
The module can be installed in any standard PXI slot marked with a peripheral slot compatibility image (a circle containing the slot number).
The module can also be installed in any hybrid PXI slot marked with a peripheral slot compatibility image (the letter “H” and a solid circle containing the slot number).
1 Make sure the PXI chassis power is turned off.
2 If the chassis has multiple fan speed settings, ensure that the fans are set to automatic. Do not set the fan speed to low or turn them off.
3 Position the chassis so that there is ample space between the chassis fan intake and exhaust vents. Blockage by walls or obstructions affects the air flow needed for cooling. (Refer to the chassis documentation for more information about cooling).
4 The modules are shipped with thread protectors over the mounting screws.
These must be removed before installing the modules in a chassis.
5 Holding the PXI module by the injector/ejector handle, slide it into an available PXI (or hybrid) slot, as shown in the following figure.
– Install the module in the PXI slot by placing the module card edges into the front module guides (top and bottom).
– Slide the module to the rear of the chassis and assure that the injector/ejector handle is pushed down in the unlatched (downward) position.
– Slide the module completely into the chassis. When you begin to feel resistance, push up on the injector/ejector handle to fully seat the module into the chassis.
Keysight M9187A PXI Digital IO Startup Guide 17
18
Step 5: Install the Module in the PXI Chassis
6 Latch the module by pulling up on the injector/ejector handle and secure the front panel to the chassis using the module mounting screws.
7 Tighten the screws on the module (or remote controller) front panel.
Performance may suffer if the screws are not securely tightened.
8 Install all chassis covers, filler panels, and air inlet modules after installing the module. Missing filler panels may disrupt necessary air circulation in the chassis.
9 If you are using a remote controller, connect the System Interface Card in the chassis to the host computer.
10 Plug in and power up the PXI chassis. Verify that the chassis fans are operating and free of obstructions that may restrict airflow.
If you are using a remote controller linked to the M9021A Cable
Interface, you must power up the chassis BEFORE you power up the PC. When you power down your system, you must Shut Down the PC BEFORE you power down the chassis.
If you are using MXI-3 to connect a desktop PC to a PXI chassis or link to multiple chassis, power up the system as follows:
- For a system with a PC and one chassis, power up the chassis before powering the PC.
- For a system with more than one chassis, power on the last chassis in the system followed by the penultimate, etc. Finally, turn on the PC or chassis containing the system controller.
11 If you are using a remote host computer (rack mount, desktop, or laptop PC), power-on the computer. Choose the default option for any “Found New
Hardware” dialogs.
After all of the “Found New Hardware” dialogs are complete, you must reboot the host computer.
Chassis Power Down Process
If you are using a remote controller linked to the M9021A Cable
Interface, you must Shut Down the PC BEFORE you power down the chassis. When you restore power, you must power up the chassis BEFORE you power up the PC.
Keysight M9187A PXI Digital IO Startup Guide
Step 6. Verify Operation of the Module
Step 6. Verify Operation of the Module
Run Keysight IO Libraries Connection Expert
If Keysight Connection Expert is already running on the system controller, click the Refresh All button to identify any hardware you have just installed or re-connected.
If Connection Expert is not already running, run it now to verify your I/O configuration. In the Windows Notification Area, click the IO icon ( ), then click
Keysight Connection Expert.
Locate your interfaces and instruments in the Keysight Connection Expert
Explorer Pane. The following graphic shows the Connection Expert screen.
Select the module in the center pane (Instrument I/O on this PC). The right-hand
Pane shows the instrument properties. Select the Installed Drivers tab then click the Start SFP button. You can use the M9187A Soft Front Panel
(SFP) software to set the outputs read the inputs, and verify operation of the module. The Soft Front Panel software was installed as part of the Software installation process.
For detailed instructions for using the M9187A Digital IO module, refer to the
Soft Front Panel help file on the Digital IO Module Software and Product
Information CD.
Keysight M9187A PXI Digital IO Startup Guide 19
Step 6. Verify Operation of the Module
Refer to the SFP help file on the Digital IO Module Software and Product
Information CD for specific detailed information on the SFP. The following graphic shows an example of the SFP for the M9187A PXI Digital IO module.
20
If you have another application, either your own program or another instance of the SFP interface, that has initialized the M9187A module, then the SFP enters it’s
“monitor” mode. In this mode, you must check the “Allow Control” check box (upper left corner) before you can actually set the digital output channels. If Allow Control is not checked, the Output Channels section of the SFP display is grayed out.
Selecting the Continuous button allows the SFP monitor mode to continuously read and update the state of the digital inputs regardless of whether the Allow Control check box is checked or not. The “light” next to the Continuous button flashes to indicate the SFP is reading and updating the display.
Verify operation
There are no specific operational verification or self test procedures. However, you can use the Soft Front Panel software to set the outputs and read the inputs.
Module specifications are guaranteed by design.
Keysight M9187A PXI Digital IO Startup Guide
Module Functional Description
Characteristics
For detailed specifications, refer to the module Data Sheet on the Digital IO
Module Software and Product Information CD or the module data sheet online at www.keysight.com/find/pxi-dio .
Module Functional Description
The Keysight PXI M9187A Digital IO module conforms to the single slot, 3U form factor (100mm by 160mm / 3.94 in. by 6.3 in.) Eurocard standard. These modules meet PXI Specification 2.2. Local bus, trigger bus, and star trigger are not included. The modules also include:
- CPCI Ejector Handle
- Front panel 78 pin D type connector
- The front panel secures to the chassis by two M2.5 x 6mm pan-head
Posi-drive screws.
Figure 1 shows a functional block diagram for the module. The PCI connector provides +5Vdc, backplane.
12Vdc, ground, as well as the control signals from the chassis
Figure 1
M9187A Digital IO Functional Block Diagram
Keysight M9187A PXI Digital IO Startup Guide 21
Module Functional Description
The module’s 32 channel output section comprises High Side and Low Side drivers and Shift Registers. The shift registers control the output drivers and are supplied with serial data from the PCI Bridge via the control logic. Each output can be switched to either sink current to the GND terminal via a Low Side driver device, or source current from the V ext outputs can also be tri-stated (off).
terminal via a High Side driver device. The
For serial input, the module’s 32 input channels are attenuated, then multiplexed through an analog multiplexer. A programmable threshold detector circuit converts the analog signal into a digital data stream determined by the programmable high and low thresholds. The input data is then read by the PCI
Bridge via the control logic.
Module Programming
Refer to the Soft Front Panel Help file for detailed operation of the module. For programming information, refer to the IVI C and LabVIEW driver help files. These help files are located on the Digital IO Module Software and Product Information
CD
22 Keysight M9187A PXI Digital IO Startup Guide
Module Functional Description
Module Front Panel, Connector and Pin Out
Figure 2 below shows the M9187A module front panel.
shows the M9187A front panel connector (as viewed from the module front
panel) and connector pinout. Table 1 on page 25 lists the front panel pin out
connections by signal (channel number).
Figure 2
M9187A Front Panel
Figure 3 shows the M9187A front panel connector (as viewed from the module
front panel) and the connector pinout. Table 1 on page 25 lists the connector by
signal (channel) number.
Keysight M9187A PXI Digital IO Startup Guide 23
Module Functional Description
Pin
No.
Signal
60 V ext
61 Output 4
62 Output 8
63 Output 12
64 Output 16
65 Output 20
66 Output 24
67 Output 28
68 Output 32
69 GND
70 Input 4
71 Input 8
72 Input 12
73 Input 16
74 Input 20
75 Input 24
76 Input 28
77 Input 32
78 GND
51
52
53
54
47
48
49
50
55
56
57
58
59
43
44
45
46
Pin
No.
40
M9187A 78-Pin D-Type Connector
Signal Pin
No.
21
Signal
41
42
V ext
Output 3
Output 7
22
23
V ext
Output 2
Output 6
Output 11
Output 15
Output 19
Output 23
24
25
26
27
Output 10
Output 14
Output 18
Output 22
Output 27
Output 31
GND
Input 3
Input 7
Input 11
Input 15
Input 19
Input 23
Input 27
Input 31
GND
GND
28
29
30
31
32
33
34
35
36
37
38
39
Input 22
Input 26
Input 30
GND
Output 26
Output 30
GND
Input 2
Input 6
Input 10
Input 14
Input 18
Figure 3
M9187A Front Panel Connector and Pinout
10
11
12
13
8
9
6
7
4
5
2
3
Pin
No.
1
18
19
20
14
15
16
17
Signal
V ext
Output 1
Output 5
Output 9
Output 13
Output 17
Output 21
Output 25
Output 29
GND
Input 1
Input 5
Input 9
Input 13
Input 17
Input 21
Input 25
Input 29
GND
GND
24 Keysight M9187A PXI Digital IO Startup Guide
M9187A Module Accessories
Signal
V ext
Gnd
Pin No.
1, 21, 40, 60
10, 30, 49, 69, 19, 20, 39, 58, 59, 78
Signal
Input 1
Input 5
Input 9
Input 13 14
Input 17 15
Input 21 16
Input 25 17
Input 29 18
Pin No.
11
12
13
Signal
Input 2
Input 6
Input 10
Input 14
Input 18
Input 22
Input 26
Input 30
34
35
36
37
38
Pin No.
31
32
33
Signal.
Pin No.
Input 3
Input 7
50
51
Input 11 52
Input 15 53
Input 19 54
Input 23 55
Input 27 56
Input 31 57
Signal.
Pin No.
Input 4
Input 8
70
71
Input 12 72
Input 16 73
Input 20 74
Input 24 75
Input 28 76
Input 32 77
Output 1 2
Output 5 3
Output 9 4
Output 13 5
Output 17 6
Output 21 7
Output 25 8
Output 29 9
Output 2 22
Output 6 23
Output 10 24
Output 14 25
Output 18 26
Output 22 27
Output 26 28
Output 30 29
Output 3 41
Output 7 42
Output 11 43
Output 15 44
Output 19 45
Output 23 46
Output 27 47
Output 31 48
Table 1 M9187A Signal (Channel) Number to Connector Pin Out
Output 4 61
Output 8 62
Output 12 63
Output 16 64
Output 20 65
Output 24 66
Output 28 67
Output 32 68
M9187A Module Accessories
Model Description
Y1181A PXI Connector Block: 78 pin, shielded, female D Sub
Y1187A PXI Connector Cable: 78 pin, male to female, 1Meter
Y1188A PXI Connector Cable: 78 Pin, male to female, 2Meter
Keysight M9187A PXI Digital IO Startup Guide 25
M9187A Module Accessories
Y1181A 78 Pin Shielded Connector Block
This shielded connector block provides a simple method of connecting wires to
Keysight M9187A PXI Digital IO module with a 78 pin D-Type front panel connector. The connector block has a simple bar clamp for strain relief. The screw terminals accept wires up to 20AWG; the recommended torque for the screw terminals is 0.12 - 0.15 Nm. The following diagram shows the pinout of the connector block. The pin numbers match the pin numbers on the module’s front panel connector. Refer to
“Disassembling the Connector Block Clam Shells” on page 28 for disassembly instructions.
26
Figure 4
Y1181A 78 Pin Terminal Block Wiring Diagram
Voltages greater than 60 Vdc present an electric shock hazard. Disconnect all source voltages before connecting or removing the source-to-modules I/O connector or wiring the connector block. All field wiring must be rated for the highest voltage applied to any single channel.
Keysight M9187A PXI Digital IO Startup Guide
M9187A Module Accessories
Y1187A, Y1188A 78 Pin Male to Female Connector Cable
These cable assemblies are used to extend the front panel connections of the
M9187A PXI Digital IO module. The cable is PFA copper/tin, 26AWG with a 3A current rating. Nominal resistance is 0.2/M. The Y1187A is 1m long and the
Y1188A is 2m long.
Figure 5
Y1187A, Y1188A 78 Pin Male to Female Connector Cable
Voltages greater than 60 Vdc present an electric shock hazard. Disconnect all source voltages before connecting or removing the source-to-modules I/O connector or wiring the connector block. All field wiring must be rated for the highest voltage applied to any single channel.
Keysight M9187A PXI Digital IO Startup Guide 27
M9187A Module Accessories
Disassembling the Connector Block Clam Shells
1 Loosen the two screws indicated in Figure 6. These screws are captive to the
bottom half of the clam shell. Note: the other two screws are for the strain relief.
Screws for
Strain Relief
Loosen these two captive screws to separate clam shell halves
Figure 6
Disassembling the Terminal Block Clam Shell
2 Separate the two clam shell halves. Be careful not to lose the long screw locks used to secure the terminal block to the module.
3 To use the strain relief, loosen the two strain relief screws and lift up on the
strain relief bar. Refer to Figure 7.
Wires go under strain relief bar.
Do not over tighten strain relief clamp.
28
Figure 7
Using the Strain Relief Clamp
4 Insert your cable under the strain relief bar and retighten the two strain relief screws.
Do not over tighten the strain relief screws. Possible damage to the cable(s) may occur if overtightened. Use only sufficient tension to secure the cable in place.
5 To reassemble the clamshell, make certain the screw locks are in place. Place the calm shell halves together. Tighten the two clam shell screws.
Keysight M9187A PXI Digital IO Startup Guide
Electrical Operating Conditions
Electrical Operating Conditions
To avoid electric shock, turn off the chassis and disconnect or de-energize all field wiring to the modules before installing or removing any module or chassis slot cover.
Transients
The Digital IO module described in this manual is intended for low voltage applications only and should not be connected to circuits that may generate or conduct large transient voltages.
Not for connection to mains. Do not connect any of the module’s I/O pins directly to a mains power outlet.
High Energy Sources
These modules are designed to handle inputs up to their rated currents or their rated powers, whichever is less. Under certain fault conditions, high energy sources could provide substantially more current or power than a module can handle. It is important to provide external current limiting, such as fuses, if the module inputs are connected to high energy sources.
Install current limiting devices between high energy sources and the module inputs.
Electrical Operating Conditions
This module is designed to operate in a temperature range of 0 non-condensing humidity. The maximum humidity is 95% at 40
°
°
C to +55
°
C with
C. Do not use in locations where conductive dust or electrolytic salt dust may be present.
These modules should be operated in an indoor environment where temperature and humidity are controlled. Condensation can pose a potential shock hazard.
Condensation can occur when the modules are moved from a cold to a warm environment, or if the temperature and/or humidity of the environment changes quickly.
Refer to the data sheet for maximum voltage, current, and power dissipation ratings for each module. For the latest specifications, check the Keysight web site at: www.keysight.com/find/pxi-dio .
Keysight M9187A PXI Digital IO Startup Guide 29
Electrical Operating Conditions
30 Keysight M9187A PXI Digital IO Startup Guide
MyKeysight www.keysight.com/find/mykeysight
A personalized view into the information most relevant to you.
www.axiestandard.org
Advanced TCA
®
Extensions for Instrumentation and Test (AXIe) is an open standard that extends the Advanced TCA for general purpose and semiconductor test. Keysight is a founding member of the AXIe consortium.
www.lxistandard.org
LAN eXtensions for instruments puts the power of Ethernet and the Web inside your test systems. Keysight is a founding member of the LXI consortium.
www.pxisa.org
PCI eXtensions for Instrumentation (PXI) modular instrumentation delivers a rugged PC-based high-performance measurement and automation platform.
www.keysight.com/find/ThreeYearWarranty
Keysight’s combination of product reliability and three-year warranty coverage is another way we help you achieve your business goals: increased confidence in uptime, reduced coat of ownership and greater convenience.
Keysight Assurance Plans
Keysight Advantage Services www.keysight.com/find/AssurancePlans
Five years of protection and no budgetary surprises to ensure your instruments are operating to specifications and you can continually rely on accurate measurements.
www.keysight.com/quality
Keysight Electronic Measurement Group DEKRA Certified ISO 90001:2008
Quality Management System
Keysight Channel Partners www.keysight.com/find/channelpartners
Get the best of both worlds: Keysight’s measurement expertise and product breadth, combined with channel partner convenience.
www.keysight.com
www.keysight.com/find/modular www.keysight.com/find/pxi-switch
For more information on Keysight Technologies’ products, applications or services, please contact your local Keysight office. The complete list is available at: www.keysight.com/find/assist
For other unlisted Countries: www.keysight.com/find/contactus
Keysight M9195A PXIe Digital Stimulus/Response Startup Guide 1
This information is subject to change without notice.
© Keysight Technologies, 2014
Edition 3 December 2014
*M9187-90001*
M9187-90001 www.keysight.com

Public link updated
The public link to your chat has been updated.
Advertisement
Key features
- 32 high output voltage and current channels
- 32 dual variable threshold input channels
- 78 pin D-type connector
- High side and Low side drivers
- Programmable threshold detector