COLPOSCOPE MODEL 985 LED

Add to my manuals
16 Pages

advertisement

COLPOSCOPE MODEL 985 LED | Manualzz

COLPOSCOPE

MODEL 985 LED

OWNER’S MANUAL

PRECISION MICROSCOPES

PRECISION MICROSCOPES

We are here to serve you!

If you have any questions regarding Seiler’s products or services, please feel free to contact us.

Toll-Free: 800.489.2282

Local: 314.968.2282

Fax: 314.968.3601

E-mail: [email protected]

www.seilerinst.com

• Microscope Division Manager

Dane Carlson ext. 365 • [email protected]

• Corporate Marketing/Sales Representative

Nicole Rasch ext. 345 • [email protected]

Tony Leise ext. 363 • [email protected]

Sandra Jeremich ext. 336 • [email protected]

2

Register your product warranty online:

http://www.seilermicro.com/warranty-registration/

TABLE OF CONTENTS

INDICATIONS FOR USE

TECHNICAL DATA

BINOCULARS

OBJECTIVE

COLPOSCOPE ASSEMBLY

COLPOSCOPE HANDLING

HOW TO FOCUS THE COLPOSCOPE

OPTICAL AND FIELD MAGNIFCATION TABLE

CLEANING

TROUBLESHOOTING TIPS

LED LIGHTSOURCE

OWNER’S RECORD

NOTES

7

8

6

7

5

5

4

4

9

15

8

8

15

3

INDICATIONS FOR USE

The Seiler Colposcope is intended to provide magnified visualization of the tissues of the vulva, vagina, cervix, and anogenital area. It is used to evaluate these tissues, select areas for biopsy, as necessary, and to facilitate related procedures, e.g., LEEP, conization, etc.

Caution: Federal law restricts this device to sale by or on the order of a physician or practitioner trained in its use.

This manual is intended to give assistance in the assembly and operation of the Colposcope Model 955.

This handbook includes a chapter to acquaint the user with the assembly process.

Read this handbook carefully prior to operating the colposcope.

Do not allow persons unfamiliar with the equipment to operate the colposcope.

TECHNICAL DATA

A - Colposcope Head

Microscope

Binoculars

Objective (focus)

Eyepieces

Straight, F=170MM (optional: Inclined F=135 mm)

300 mm

Wide angle type 12,5x, adjustable

Magnification

Observation field

Illumination field

Interpupillary distance

See table

See table

Illumination System:

Adjustable between 42 mm and 75 mm

LED Illumination System

Green

Focusing:

Micro

Total weight of the head

B- Colposcope stand

Manual, with rack and pinion

3.0 Kg

Horizontal reach

Angular rotation (arms)

Distance between the floor and the objective

725 mm. - first arm 600 mm, second arm 110 mm.

340º

508 mm to 1143 mm

Base Base with five casters, two with brakes

Total floor stand set weight 37.7 kg

4

BINOCULARS

2

1

3

1. Binoculars body

2. Eyepieces with diopter adjustment

3. Eye Protectors

OBJECTIVE

The Colposcopes are factory shipped with the following objective:

F = 300 mm

The following optional objectives are available:

F = 175 mm

F = 200 mm

F = 250 mm

F = 400 mm

5

COLPOSCOPE ASSEMBLY

Step 1

Place post into base with pin in base, lined up with notch in post.

Step 2

Install bolt and tighten securely with Allen wrench (provided).

Step 3

Place arm into pole.

Step 4

Install extension arm by pulling plunger and inserting into

Pantographic arm.

Plunger

Step 5

Install Optic Pod into the extension arm.

The correct position of the Binoculars to the Colposcope

Align the guide-pin to the groove, adjust the base of the Binoculars to point the flange, and tighten the set screw.

Finally check if the Binoculars are well secured and do not move.

The Binoculars Eyepieces

The eyepieces are mounted to the binocular housings, as shown in the illustration. Make sure that they are well seated.

The Eye Protectors

The Eye Protectors are threaded to the Binoculars as shown in the figure 9. The Eye Protectors protect the eyes, prevent the intake of lateral lights and create a suitable darkroom to the observer.

Precautions

The Binoculars contain optical parts internally that cannot be cleaned without disassembly.

It is therefore advisable to keep the eyepieces inserted in the eye tubes at all times in order to keep impurities from getting inside the housing. It is recommended that if the eyepieces are removed, a clean cloth be placed over the openings.

6

COLPOSCOPE HANDLING

a- MAGNIFICATION SELECTOR

The magnification selector has 6 positions, engraved from 1 to 5. Note that the position 3 repeats. The position being used is the one aligned to the black dot, engraved on the body. The magnification factors are found in the table appended.

b- PANTOGRAPHIC ARM BALANCING

The Colposcope Model 985 is factory adjusted for balancing, using the regular optical head, and may only require adjustment at the tension knob when accessories are added.

HOW TO FOCUS THE COLPOSCOPE

Preliminarily:

Displace the Articulated Arm of the stand set where the colposcope is fixed, and point its objective to the direction of the field to be observed. For an objective with focal distance of 300 mm, keep a gap of about 300 mm between the lens and the specimen.

NOTE: It is easier to focus when the illumination system is on. The illuminated field coincides with the focused object.

Parfocalizing Procedure:

1. Turn the magnification drum to the highest setting (5) and focus sharply on a target, such as a piece of paper with an “X” drawn on it, by using the focusing knobs.

2. The turn the magnification drum to the lowest setting (1) and focus by turning eyepieces clockwise or counter clockwise until you reach sharp focus. Be sure to depress the orange tab when turning the eyepiece.

3. Now the microscope is in focus for your eyes throughout all the magnification settings.

The Magnifications Selector appears indicated in Figure 10 of the section VI-a) “Magnification Selector”, and below.

Binocular Head

Tension

Adjustment

Magnification

Changer

Objective Lens

Green Light

Filter

Fine Focus

Knob

7

OPTICAL AND FIELD MAGNIFCATION TABLE

The optical magnifications achieved with the Colposcope are determined by 3 variables: the objective focal distance, the position of the Magnification Selector and the by eyepiece used.

The table shows the optical magnification and the observed field diameters, in millimeters, depending on those variables. At the bottom line of the table the illuminated field is shown, corresponding to the light disk incident on the field, with a diameter that depends only on the focus of the work objective.

The magnification selector has 6 positions: 1, 2, 3, 4, and 5. The 3 position is repeated in the selector. The working position is the one aligned to the red dot mark.

 Eyepiece

Posi(on  of  

BuNon

10X/18

12.5X/16

1(0.4)

2(0.6)

3(1.0)

4(1.6)

5(2.5)

1(0.4)

2(0.6)

3(1.0)

4(1.6)

5(2.5)

16X/16

1(0.4)

2(0.6)

3(1.0)

4(1.6)

5(2.5)

Illuminated  Field

Magnifica(on  Table  for  985

Binocular  Head  Focal  Length:  F=170mm

175 200

Magnifica(on Field Magnifica(on

3.89

46.32

5.83

30.88

9.71

18.53

15.54

11.58

24.29

7.41

3.40

5.10

8.50

13.60

21.25

4.86

41.18

7.29

27.45

12.14

16.47

19.43

10.29

30.36

6.59

6.22

41.18

9.33

27.45

15.54

16.47

24.87

10.29

38.86

6.59

65

21.76

34.00

72

4.25

6.38

10.63

17.00

26.56

5.44

8.16

13.60

Field

52.94

35.29

21.18

13.24

8.47

47.06

31.37

18.82

11.76

7.53

47.06

31.37

18.82

11.76

7.53

Objec(ve  Lenses

250

Magnifica(on Field

2.72

4.08

6.80

10.88

17.00

66.18

44.12

26.47

16.54

10.59

3.40

5.10

8.50

13.60

21.25

4.35

6.53

10.88

17.41

27.20

90

58.82

39.22

23.53

14.71

9.41

58.82

39.22

23.53

14.71

9.41

300

Magnifica(on Field

2.27

3.4

5.67

9.07

14.17

2.83

4.25

7.08

11.33

17.71

3.63

5.44

9.07

14.51

22.67

108

79.41

52.94

31.76

19.85

12.71

70.59

47.06

28.24

17.65

11.29

70.59

47.06

28.24

17.65

11.29

Magnifica(on

400

1.70

2.55

4.25

6.80

10.63

2.13

3.19

5.31

8.50

13.28

2.72

4.08

6.80

10.88

17.00

144

Field

105.88

70.59

42.35

26.47

16.94

94.12

62.75

37.65

23.53

15.06

94.12

62.75

37.65

23.53

15.06

CLEANING

The Objective, pointed to the field under clinical or surgical treatment, is exposed to blood and other bodily fluids. Spots on the objective dim the passage of the light. Stains blur the light, reducing the luminosity of the optical observation.

Cotton padding wetted in alcohol rubbed softly, in CIRCULAR MOVEMENTS over the optical surface, is enough to clean the stains, except those of clotted blood. Clotted blood shall be cleaned with cotton and hydrogen peroxide.

If the eyepieces are too dirty, change the cotton at each circular friction motion to avoid spreading out the dirt.

On the other hand, there is an effective protector for the objective, fitted under pressure at the outer edge of the lens which protects the objective against “splatter” and contamination.

To have the eyepieces and objective well cleaned for clinical use, it is sufficient to use cotton.

The metallic parts (chrome-plated or painted) can be cleaned with cotton and alcohol.

TROUBLESHOOTING TIPS

ELECTRIC CONNECTIONS

Ensure that none of the following are present:

Broken wires inside the cable.

Loose wires at the connector.

Broken wires at the fixation, inside the connector.

Oxidized connectors terminals.

8

LED LIGHTSOURCE

1. INTRODUCTION

Congratulations on the purchase of your new SSL-2000 Light Source! This user-friendly LED light source is a high efficiency light source utilizing state of-the-art superior illumination technology. It offers a variety of features such as:

• 6500º K daylight brightness for good color definition

• Quiet operation

• Long life, 50,000 hrs.

• Compact and light weight

• Pulse-width Modulation (PWM) Electrical Dimming

This section of the manual will help you to install the device and optimally integrate it with other components of your system. It will also instruct you how to operate the LED Light Source and how to keep it clean. It will give you maintenance and service guidelines as well as recommendations for best performance results.

1.1 INDICATIONS FOR USE: The LED Light Source is used to illuminate the site of surgery during minimally invasive surgical procedures. The light is transmitted from the light source through a fiber optic cable and a scope.

2. WARNINGS

To prevent fire or electric shock, do not open or expose the LED light source unit to rain or moisture. Refer all servicing to qualified personnel only.

Not suitable for use in presence of flammable anesthetic mixture with air or with oxygen or nitrate oxide.

To prevent any potential electro-magnetic interference, do not use any kind of cellular phone near the light source.

This product should be used only with type BF endoscopic instruments which have been certified according to IEC 601-1 for medical equipment and IEC 601-2-18 for endoscopic equipment.

Caution:

This product is not provided as sterile.

Caution:

All devices connecting to the LED Light Source must be classified as medical equipment. When additional information processing equipment is connected to the LED Light Source, the operator must determine that all equipment complies with the appropriate end-product standards (such as IEC 60950 or IEC 60065 and the Standard for Medical System, IEC 60601-1-1).

Caution:

The LED Light source can cause permanent eye damage if viewed directly with unprotected eye.

To reduce the chance of eye damage, set the intensity control always to the minimum level and plug the fiber optic cable into the unit before turning on the power.

EQUIPMENT CONNECTION

Caution:

The fiber optic cable must be a NON-CONDUCTIVE CABLE. It should not have conductive shielding or any other conductive connection between the patient and equipment. Such connection will impair safety of the equipment. It must be rinsed free of soaking/disinfectant solution and dried before plugging into the LED light source receptacle. Ensure the optical surface is clean before engaging into the light source.

9

3. SPECIFICATIONS

Item

Light Source Type

Power

Color Temperature

Led life

Light guide adapter

Brightness control

Input voltage

Rated Power

Regulatory Approvals

Equipment Class

Mode of Operation

Water Resistant

Operating Environment

Temperature

Relative Humidity

Air Pressure

Storage Environment

Temperature

Relative Humidity

Air Pressure

Dimensions

Weight

Specification

LED (Light Emitting Diode)

50 Watt

6500

º K

50,000 hours (typical)

PWM (Pulse-width Modulation) – 0-100% Dimming

100-240V AC, 50/60 Hz

52 watt

UL60601-1,CAN/CSA C22.2 No.601.1 (SUP1+AM2),EN 60601-1-2 and CE marked

Class I, BF-type

Continuous operation

Not Protected Equipment, IPX0

+0° to +40° C (32° to 104° F)

0 to 85%rh, non condensing

700 to 1060 hPa

-20° to +60° C (-4° to 140° F)

30 to 95%rh, non condensing

700 to 1060 hPa

4.65” W x 3.32” H x 7.25” D

2.5 lbs./1.14 kg

4. OPERATING ELEMENTS, SYMBOLS AND FUNCTIONS

4.1 FRONT PANEL

1

2

No.

Name

1

2

3

Power switch

Light Guide Adapter

Intensity Control

10

3

Function

Turns the light source on and off. Fan will operate

Electronically controls the light output

4.2 REAR PANEL

2

3

1

2

3

4

No.

Name

1 AC main inlet

Fuse

Grounding Connector

Venting grids

4.3 Bottom Panel

4

Figure 2. LED Light Source Rear Panel

Function

Accepts AC power cord

2A, 250V (1/4” x 1 ¼”) time delay

For potential equalization

Allows adequate ventilation and cooling of the light source

Figure 3. LED Light Source Product Label

11

5. INSTALLATION

5.1 SETTING UP THE LED LIGHT SOURCE

Place the LED light source on a stable surface (cart, counter, stand, etc.).

NOTE: Avoid places where the light source may be splashed with liquid.

Absolutely DO NOT use in any environment with explosive or flammable gases.

DO NOT block the venting grids of the LED light source.

Make sure the power switch is in OFF position.

Connect AC power cord to the power inlet located on the rear panel of the light source.

CAUTION:

Use only cords provided with the light source.

Plug the AC power cord into a wall outlet using the three-prong plug supplied with the unit.

CAUTION:

To prevent electric shock, connect power cords of peripheral equipment through medical isolation transformers.

NOTE: When using medical isolation transformer, be sure to check the transformer power ratings.

Make sure that the power cord is connected to the main power with three-prong plug (USA use UL2601-1 rated isolation transformers and/or power strips only).

CAUTION:

When used in clinical or residential areas near radio or TV units, this equipment may be subjected to radio interference. To avoid adverse electromagnetic effects, DO NOT operate this equipment near RF energy equipment.

5.2 CONNECTING THE FIBER OPTIC CABLE

Turn intensity control to the minimum setting. Connect the light cable to the instrument and then plug the light cable endfitting into the port on the front panel.

6. OPERATION

6.1 POWERING UP THE LED LIGHT SOURCE

To operate the LED light source: Turn on the power switch. The indicator on power switch will light.

6.2 LIGHT BRIGHTNESS DISPLAY AND CONTROL

The light brightness adjustment enables the user to obtain a brighter or darker illumination of the object of observation.

Adjust the light intensity by turning the intensity control knob.

7. CLEANING

NOTE: Always disconnect the power cord before cleaning the system.

The unit can be cleaned with any cleaning agent, which is used for external cleaning of electric equipment, according to instructions given by the manufacturer of the cleaning solution.

Do not allow excessive moisture or liquids to come in direct contact with the unit.

Do not use cleaning agents that are not permitted for use with plastics, i.e., ammonia, acetone, salty acids (HCl), etc.

Do not allow cleaning agents or liquids to enter the unit outlets.

12

8. DISINFECTION

8.1 DISINFECTING THE UNIT

NOTE: Always disconnect the power cord before cleaning the system.

Use any disinfectant agents which are commonly applied while disinfecting surfaces of electric medical equipment. Such disinfectant agents are usually in the form of sprays or damp cloths.

Follow the instructions given by the manufacturer of the disinfectant solution.

9. SERVICING AND REPAIR

Defective items of equipment are to be serviced and repaired exclusively by persons authorized by the manufacturer. All repair work shall employ original manufacturer’s parts only.

9.1 FUSE REPLACEMENT

CAUTION:

Always disconnect power cord and turn main switch off before fuse replacement.

Turn light source off and unplug power cord. On back of unit remove fuse by turning fuse cover and pulling out fuse.

Replace fuse with 2 AMP (1/4” x 1 ¼”) time delay 250V rated fuse. Insert back into fuse housing.

Re-connect the power cord and turn the LED light source on according to section 6.1.

9.2 LIMITED WARRANTY

Your LED light source carries a 1-year warranty from the date of shipment on workmanship and all defects of material, excluding replaceable sweatbands, headband, fiberoptic cable and lamps. Should your product prove to have such defects within one year of the shipment. Seiler Instrument will repair or replace the product or component part without charge. Should your LED Light Source product(s) need servicing under this warranty, please contact, Seiler Instrument return authorization documentation. You should carefully pack unit in a sturdy carton and ship it to the factory. Please include a note describing the defects, your name, telephone number and a return address. Warranty does not cover equipment subject to misuse, accidental damage, normal wear and tear or if transferred to a new owner without authorization from Seiler Instrument. This warranty gives you specific legal rights and you may also have other rights that vary from state to state.

POST WARRANTY REPAIRS: You may return your product(s) for repair, shipping prepaid to the factory. Your product will be inspected and an estimate of repair charges will be submitted to you for approval. Payment must be received before repairs are completed.

• In the US: 800-489-2282 (toll free)

• FAX number: 314-968-3601

• Customer Service: 314-218-6336

13

10. END OF PRODUCT LIFE

We encourage our customers to recycle this product whenever possible. Disposal of this unit must be performed in accordance with the applicable local environmental regulations.

In the United States a list of recyclers in your area can be found at: http:/www.eiae.org/.

Please contact customer service to issue a return authorization to return product to manufacturer at the end of product life.

11. TROUBLESHOOTING

Problem

The power indicator (refer to 4.1) is not lit.

The power indicator is lit, but the lamp will not ignite.

Solution

A. Check that the AC power cord is properly connected.

B. Check the unit fuses. If necessary, replace.

Turn intensity control know clockwise to increase light output intensity.

14

OWNER’S RECORD

Model Number

Serial Number

Date Purchased

Location Purchased

NOTES

15

Version 09/22/14

PRECISION MICROSCOPES

3433 Tree Court Industrial Blvd.

St. Louis, Missouri 63122

Toll Free: (800) 489-2282

Local: (314) 968-2282

Email: [email protected]

Web: www.seilerinst.com