important safety information warnings caution

important safety information warnings caution
!
PLEASE READ THE ENTIRE MANUAL
BEFORE ATTEMPTING TO INSTALL
OR USE THE LUMEN 200 OR
LUMEN 200PRO
QUICK START INSTRUCTIONS
!
Manual Shutter
Control
Alarm Reset
Button
IMPORTANT SAFETY INFORMATION
Display
• Before using the system, please follow and adhere to all warnings,
safety and operating instructions located either on the product or
in this Quick Start Manual.
• Do not expose the product to water or moisture.
• Connect the AC power cord only to designated power sources as
marked on the product.
• Ensure that the ventilation slots in the controller case have a
minimum of 100mm (4 inches) of clearance from blockages.
Power Switch
Liquid Light Guide
• WARNING: Hg - LAMP CONTAINS MERCURY,
manage in accord with disposal laws.
• WARNING: The method in which the lamps are
disposed of must comply with the local rules & regulations
for disposal of hazardous materials.
WARNINGS
• Eye damage may result from directly viewing the light
produced by the lamp used in this product.
• Always make sure the light guide is properly inserted
into the collimator which is firmly attached to the
microscope using the adapter provided and into the
Lumen 200/PRO, before turning on the power to the unit.
CAUTION
• Never look into the emitting end of a light guide. The
light could severely damage the cornea and retina of the
eye if the light is observed directly. Appropriate eye
shielding must be used at all times, clothing should be
used to protect exposed skin.
• Never place the end of an emitting light guide near skin
as this may result in burning and damage to the skin.
• Never place the end of an emitting light guide near a
flammable substrate as sufficient power is emitted from
the light guide to ignite flammable substances.
Prior Scientific, Inc.
Collimator
Microscope Adapter
CONNECTING TO THE MICROSCOPE
Once the lamp is installed, the Lumen is ready to be connected to the microscope - see opposite side of guide for lamp installation.
1. Place the Lumen in a safe location and ensure a minimum of 100mm (4 inches) of clearance from obstructions for the fan outputs.
2. Unpack the liquid light guide from the foil packaging and remove both red plastic caps from the light guide.
Note: It is important to remove the red caps before connecting to the light guide to avoid damage on power up.
3. Insert the clamping nut end of light guide into the threaded opening on the front of the Lumen and thread the clamping nut
into the Lumen.
4. Tighten the connector firmly.
5. Loosen the thumb screw on the back of the collimator. Firmly push the light guide into the hole ensuring it has reached its
end stop and tighten the thumb screw.
6. Connect collimator to microscope.
STARTING UP THE LUMEN
WARNING: Do not power the Lumen without the light guide attached to both the Lumen and microscope. The Lumen should only
be used on a level surface.
1. Connect the power cable to the Lumen.
2. Ensure the light guide is attached to both the Lumen and microscope.
3. Switch the Lumen power on.
4. Allow 1-5 minutes for light to reach operating output.
WARNING: Do not power down the unit within five minutes of power up. This may reduce the effective lifetime of the lamp.
SHUTTING DOWN THE LUMEN
The following warnings apply as damage to the lamp may result if instructions not followed:
WARNING: Do not shut the unit down within five minutes of powering up the unit.
WARNING: After shutting down the unit allow five minutes before powering up the unit again. Allow 30 minutes before
changing the lamp. Failure to do so is likely to result in damage to the lamp.
80 Reservoir Park Drive, Rockland, MA. 02370
Phone: 781-878-8442
Toll Free: 800-877-2234
Fax: 781-878-8736 Web: www.prior.com
V6-10/10
LUMEN 200 & LUMEN 200PRO LAMP INSTALLATION GUIDE
!
PLEASE READ THE ENTIRE MANUAL
BEFORE ATTEMPTING TO INSTALL
OR USE THE LUMEN 200 OR
LUMEN 200PRO
!
IMPORTANT SAFETY INFORMATION
• Before using the system, please follow and adhere to all warnings,
safety and operating instructions located either on the product or
in the user manual.
• Do not expose the product to water or moisture.
• Connect the AC power cord only to designated power sources as
marked on the product.
• Ensure that the ventilation slots in the controller case have a
minimum of 100mm (4 inches) of clearance from blockages.
• WARNING: Hg LAMP CONTAINS MERCURY,
manage in accord with disposal laws.
• WARNING: The method in which the lamps are disposed
of must comply with the local rules & regulations for
disposal of hazardous materials.
INSTALLING THE LAMP
4
a.
b.
c.
d.
Remove lamp from carton. Take care not to touch the
inside surface of the lamp (see diagram below):
Open top of carton
Remove V-shaped cardboard holder
Push cardboard flaps back
Lift lamp out of carton
Cardboard
flaps
Lamp
spring
CAUTION!
Avoid contact
with sensor
when installing
bulb
V-shape
cardboard
holder
8
With the bulb now firmly in place, plug the data connector into
the PCB.
5
Rotate the bulb so that the cables and connector hang
down into the bulb chamber. Plug connector on lamp
into the socket in lamp chamber, ensure it is pushed
firmly into position.
a. Confirm that the lamp is orientated with the PCB facing
upwards.
PCB
Required Equipment: Lumen 200/Lumen 200PRO unit
• Metric Allen Wrenches • Prior Supplied Lamp
Data
connector
9
Place the Lumen upside down and replace the lamp access panel
using the four screws provided. A complete installation video is
available at www.prior.com.
Disconnect the Lumen power supply, (if the Lumen has been
previously switched on, wait for 30 minutes to allow lamp to cool).
Place the Lumen
upside down on a flat
padded surface. The
packing material that the
Lumen is shipped with
works well for a padded
surface.
3
Lamp
spring
Connector
VISIT WWW.PRIOR.COM TO VIEW THE
LUMEN LAMP INSTALLATION VIDEO
1
2
7
Lift the spring towards the bulb, as you do the bulb will be
pushed into the correct position. Click the springs all the way
into the lamp spring restraints (see images below).
Remove four screws
and the lamp access
panel, as pictured in
the photo on the right.
6
Locate the bottom of the bulb (as indicated in the above
photo). Place the wire at the bottom of the bulb into the
groove in the bottom of the lamp chamber (as indicated
below). Center the bulb in the circular lamp mount.
Lamp spring
restraint
Bottom
groove
Prior Scientific, Inc.
80 Reservoir Park Drive, Rockland, MA. 02370
Phone: 781-878-8442
Toll Free: 800-877-2234 Fax: 781-878-8736