Thermo Scientific HyClone Single-Use Mixer (SUM) Docking Station

Thermo Scientific HyClone Single-Use Mixer (SUM) Docking Station
Overview:
The S.U.M. is based on the same concept as the
Thermo Scientific HyClone Single-Use Bioreactor
(S.U.B.) with a single-use BioProcess Container
(BPC) with an integrated impeller and a hardware
system providing support, mobility, mixing motor
and controller. In contrast to the other options
in the S.U.M. product family the docking station
system has the mixing motor mounted on a
separate docking station rather than on the tank
itself. This allows the mixing motor to link with a
range of separate tanks. Easy-to-use motor height
and angle adjustment combined with a secure
mechanism to locate and lock the docking station
and tank in place make changeover of tanks simple
and rapid.
The stainless steel tank options include:
• Maximum volumes of 200, 500 and 1000 L
• With or without heating/cooling jackets
• With or without casters
S.U.M. BPC options include:
• Open top liner for media/buffer hydration
• Closed BPC with powder port for contained
media/buffer hydration
• Closed BPC with powder port and
monitoring probe capability
• Closed BPC for liquid-liquid mixing of
critical sterile solutions such as pooling
of fractions or processing of intermediate
solutions.
• Closed BPC for liquid-liquid mixing with
monitoring probe capability.
S.U.M. Recommended Operating Parameters
Temperature
Motor Speed
Volume Range
Maximum Closed Top
Mixing Bag Pressure
Continuous Operating Time
Maximum Floor Scale Height
*Mixing time at nominal volume only.
2-40°C
0-350 rpm
20-100% of Nominal
.2 psig
21 days*
3”
Data Sheet 060 Rev 1
Thermo Scientific HyClone
Single-Use Mixer (S.U.M.)
Docking Station System
The Thermo Scientific HyClone
Single-Use Mixer (S.U.M.) is a
product family of systems providing
powerful and efficient mixing based
on a conventional stirred tank
design. It is available in a wide
range of options to cover mixing
applications from powder hydration
to preparation of critical sterile
solutions and suspensions.
The S.U.M. Docking Station system
is optimized for high throughput
applications for volumes up to 1000 L.
Item
Width
Depth
Height
Weight
Electrical Power Supply Requirement
(voltage, phase, amp)
Top Stair Height
Programmable VFD,
Remote Panel Interface
Mixer Motor Drive
(voltage, phase, amp)
Motor Power Rating
Motor Torque Rating
Gear Reduction
Docking Station Frame
34”
44.8”
61.9”
509 lbs (231 kg)
120 or 240 VAC, single,
15 Amp, Non GFCI circuit
18.2016”
Standard
208 VAC, three, 15 Amp
0.5 hp (372.8 W)
80 lbf-in (9.04 Nm)
5:1
Item
Rated Liquid Working Volume
Minimum Liquid Working Volume
BPC Chamber Diameter
BPC Chamber Shoulder Height
Liquid Height @ Rated Working Volume
Overall Geometry (height/diameter ratio)
Fluid Geometry @ Working Volume
(height/diameter) Ratio
Impeller (quantity x blade count)
Impeller Diameter
Hold-Up Volume (typical)
Skid Width
Skid Depth
Skid Height
Dry Skid Weight (mass) Non-Heated Tank
Dry Skid Weight (mass) Jacketed Tank
Wet Skid Weight—Rated Working Volume (mass)
Non-Heated Tank
Wet Skid Weight—Rated Working Volume (mass)
Jacketed Tank
Minimum Ceiling Height Recommended
Minimum Required Floor Scale Platform Dimensions
200 L Tank
500 L Tank
1000 L Tank
200 L
40 L
20” (50.8 cm)
48.6” (123.4 cm)
39” (99 cm)
1.95:1
500 L
100 L
29.75” (75.6 cm)
57.5” (146 cm)
44” (111.8 cm)
1.5:1
1000 L
200 L
41.5” (105.4 cm)
61.9” (157.2 cm)
45.5” (115.6 cm)
1.1:1
2.2:1
1.7:1
1.2:1
1x3
7.875” (20 cm)
<50 mL
25.5” (64.8 cm)
29” (73.7 cm)
58.2” (147.8 cm)
443 lbs (201 kg)
465 lbs (211 kg)
1x3
7.875” (20 cm)
<100 mL
30.4” (77 cm)
37.4” (95 cm)
65.1” (165.3 cm)
776 lbs (352 kg)
802 lbs (364 kg)
1x3
7.875” (20 cm)
<300 mL
49.6” (126 cm)
49.3” (125.2 cm)
63.4” (161 cm)
880 lbs (399 kg)
913 lbs (414 kg)
884 lbs (401 kg)
1878 lbs (852 kg)
3084 lbs (1399 kg)
906 lbs (411 kg)
1905 lbs (864 kg)
3117 lbs (1414 kg)
10’ (305 cm)
23” x 22”
11’ (335 cm)
27” x 24”
11’ (335 cm)
38” x 38”
S.U.M. Docking Station
200 L Tank Standard BPC Configurations (CX5-14 Film)
Powder-Liquid Applications—Part number SH30750.01
For use with tubing welder and with CPC® Quick Connects
Back Face
Front Face
2. Addition Line
4. Powder Addition Port
Impeller
1. Bottom Drain
3. Recirculation/Sample Line
Line
Description
Tubing Set (ID X OD X Length)
End Treatment
1
Bottom Drain
1/2” (12.7 mm) ID X 3/4” (19.1 mm) OD C-Flex® X 48” (122 cm)
Capped 1/2” (12.7 mm) MPX insert
2
Addition Line
1/2” (12.7 mm) ID X 3/4” (19.1 mm) OD C-Flex X 48” (122 cm)
Plugged 1/2” (12.7 mm) MPX body
3
Recirculation/Sample line
4
Powder Addition Port
1/2” (12.7 mm) ID X 3/4” (19.1 mm) OD C-Flex X 54” (137 cm) splits to
Capped 1/2” (12.7 mm) MPX insert
1/2” (12.7 mm) ID X 3/4” (19.1 mm) OD C-Flex X 24” (61 cm) and
Plugged 1/2” (12.7 mm) MPX body
1/2” (12.7 mm) ID X 3/4” (19.1 mm) OD C-Flex X 12” (30 cm)
3” (76 mm) Sanitary Fitting, Tri-clamp
Cap with gasket
200 L Tank Standard BPC Configurations (CX5-14 Film)
Liquid-Liquid Applications—Part number SH30753.01
For use with tubing welder and with CPC® Quick Connects
Back Face
Front Face
4. Recirculation/Sample Line
1. Bottom Drain
Impeller
2. Fill Line
3. Vent Filter
5. Addition Line
Line
Description
Tubing Set (ID X OD X Length)
End Treatment
1
Bottom Drain
1/2” (12.7 mm) ID X 3/4” (19.1 mm) OD C-Flex X 48” (122 cm)
Capped 1/2” (12.7 mm) MPX insert
2
Fill Line
3/4” (19.1 mm) ID X 1” (25.4 mm) OD C-Flex X 48” (122 cm)
1 1/2” (38.1 mm)
Tri-clamp SterilEnz™
3
Vent Filter
1/4” (6.4 mm) ID X 1/2” (12.7 mm) OD C-Flex X 4” (10.2 cm)
Sterile Hydrophobic Vent Filter (0.2
micron PVDF - Pall Acro 50)
4
Recirculation/
Sample Line
5
Addition Line
1/2” (12.7 mm) ID X 3/4” (19.1 mm) OD C-Flex X 54” (137 cm) splits to
Capped 1/2” (12.7 mm) MPX insert
1/2” (12.7 mm) ID X 3/4” (19.1 mm) OD C-Flex X 24” (61 cm) and
Plugged 1/2” (12.7 mm) MPX body
1/2” (12.7 mm) ID X 3/4” (19.1 mm) OD C-Flex X 12” (30 cm)
3/8” (9.5 mm) ID X 5/8” (15.9 mm) OD C-Flex X 24” (61 cm)
Plugged 3/8” (9.5 mm) MPX body
S.U.M. Docking Station
200 L Tank Standard BPC Configurations (CX5-14 Film)
Powder-Liquid Applications with Probe Ports and Thermo-well—Part number SH30750.02
For use with tubing welder and with CPC® Quick Connects
Back Face
Front Face
7. Bottom Drain
2. Addition Line
8. Powder Addition Port
3-4. Probe Ports (2)
Impeller
1. Recirculation/Sample Line
5. Unused
6. Thermo-well/Small Volume Sample
Line
Description
Tubing Set (ID X OD X Length)
End Treatment
1
Recirculation/Sample line
1/2” (12.7 mm) ID X 3/4” (19.1 mm) OD C-Flex X 54” (137 cm) splits to
1/2” (12.7 mm) ID X 3/4” (19.1 mm) OD C-Flex X 24” (61 cm) and
1/2” (12.7 mm) ID X 3/4” (19.1 mm) OD C-Flex X 12” (30 cm)
Capped 1/2” (12.7 mm) MPX insert
Plugged 1/2” (12.7 mm) MPX body
2
Addition Line
1/2” (12.7 mm) ID X 5/8” (15.9 mm) OD C-Flex X 48” (122 cm)
Plugged 1/2” (12.7 mm) MPX body
3-4
Probe Ports (2)
None
5
Unused
Thermo-well/
Small Volume Sample
None
Thermo-well Adapter for 1/8” (3.2 mm) Diameter
1/8” (3.2 mm) ID X 1/4” (6.4 mm) OD C-Flex X 12” (30 cm)
7
Bottom Drain
1/2” (12.7 mm) ID X 3/4” (19.1 mm) OD C-Flex X 48” (122 cm)
Plugged 1/2” (12.7 mm) MPX body
8
Powder Addition Port
3” (76 mm) Sanitary Fitting, Tri-clamp
Cap with gasket
6
Pall™ Kleenpak™ Aseptic
Connector KPCHT Series (Female)
Plug
Luer and SmartSite® Valve Port
200 L Tank Standard BPC Configurations (CX5-14 Film)
Liquid-Liquid Applications with Probe Ports and Thermo-well—Part number SH30753.02
For use with tubing welder and with CPC® Quick Connects
Back Face
Front Face
6. Recirculation/Sample Line
1-2. Probe Ports (2)
7. Fill Line
3. Unused
Impeller
8. Vent Filter
9. Addition Line
4. Thermo-well/Small Volume Sample
5. Bottom Drain
Line
Description
Tubing Set (ID X OD X Length)
End Treatment
1-2
Probes Ports (2)
None
Pall Kleenpak Aseptic
Connector KPCHT Series (Female)
3
Unused
None
Plug
5
Thermo-well/
Small Volume Sample
Bottom Drain
Luer and SmartSite
Valve Port
Capped 1/2” (1.27 cm) MPX insert
6
Recirculation/Sample Line
Thermo-well Adapter for 1/8” (3.2 mm) Diameter
1/8” (3.2 mm) ID X 1/4” (6.4 mm) OD C-Flex X 12” (30 cm)
1/2” (12.7 mm) ID X 3/4” (19.1 mm) OD C-Flex X 48” (122 cm)
1/2” (12.7 mm) ID X 3/4” (19.1 mm) OD C-Flex X 48” (122 cm) splits to
1/2” (12.7 mm) ID X 3/4” (19.1 mm) OD C-Flex X 24” (61 cm) and
1/2” (12.7 mm) ID X 3/4” (19.1 mm) OD C-Flex X 12” (30 cm)
7
Fill Line
3/4” (19.1 mm) ID X 1” (25.4 mm) OD C-Flex X 48” (122 cm)
1 1/2” Tri-clamp SterilEnz
8
Vent Filter
1/4” (6.4 mm) ID X 1/2” (12.7 mm) OD C-Flex X 4” (10.2 cm)
Sterile Hydrophobic Vent Filter
(0.2 micron PVDF - Pall Acro 50)
9
Addition Line
3/8” (9.5 mm) ID X 5/8” (15.9 mm) OD C-Flex X 24” (61 cm)
Plugged 3/8” (9.5 mm) MPX body
4
Capped 1/2” (12.7 mm) MPX insert
Plugged 1/2” (12.7 mm) MPX body
S.U.M. Docking Station
500 L Tank Standard BPC Configurations (CX5-14 Film)
Powder-Liquid Applications—Part number SH30751.01
For use with tubing welder and with CPC® Quick Connects
Back Face
Front Face
2. Addition Line
4. Powder Addition Port
1. Bottom Drain
Impeller
3. Recirculation/Sample Line
Line
Description
Tubing Set (ID X OD X Length)
End Treatment
1
Bottom Drain
1/2” (12.7 mm) ID X 3/4” (19.1 mm) OD C-Flex® X 48” (122 cm)
Capped 1/2” (12.7 mm) MPX insert
2
Addition Line
1/2” (12.7 mm) ID X 5/8” (15.9 mm) OD C-Flex X 60” (152 cm)
Plugged 1/2” (12.7 mm) MPX body
3
4
1/2” (12.7 mm) ID X 3/4” (19.1 mm) OD C-Flex X 66” (168 cm) splits to
Recirculation/Sample Line
1/2” (12.7 mm) ID X 3/4” (19.1 mm) OD C-Flex X 24” (61 cm) and
1/2” (12.7 mm) ID X 3/4” (19.1 mm) OD C-Flex X 12” (30 cm)
Powder Addition Port
Capped 1/2” (12.7 mm) MPX insert
Plugged 1/2” (12.7 mm) MPX body
3” (76 mm) Sanitary Fitting, Tri-clamp
Cap with gasket
500 L Tank Standard BPC Configurations (CX5-14 Film)
Liquid-Liquid Applications—Part number SH30754.01
For use with tubing welder and with CPC® Quick Connects
Back Face
Front Face
4. Recirculation/Sample Line
Impeller
1. Bottom Drain
2. Fill Line
3. Vent Filter
5. Addition Line
Line
Description
Tubing Set (ID X OD X Length)
End Treatment
1
Bottom Drain
1/2” (12.7 mm) ID X 3/4” (19.1 mm) OD C-Flex X 48” (122 cm)
Capped 1/2” (12.7 mm) MPX insert
2
Fill Line
3/4” (19.1 mm) ID X 1” (25.4 mm) OD C-Flex X 60” (152 cm)
3
Vent Filter
1/4” (6.4 mm) ID X 1/2” (12.7 mm) OD C-Flex X 4” (10.2 cm)
4
Recirculation/Sample Line
1/2” (12.7 mm) ID X 3/4” (19.1 mm) OD C-Flex X 66” (168 cm) splits to
1/2” (12.7 mm) ID X 3/4” (19.1 mm) OD C-Flex X 24” (61 cm) and
1/2” (12.7 mm) ID X 3/4” (19.1 mm) OD C-Flex X 12” (30 cm)
Capped 1/2” (12.7 mm) MPX insert
Plugged 1/2” (12.7 mm) MPX body
5
Addition Line
3/8” (9.5 mm) ID X 5/8” (15.9 mm) OD C-Flex X 24” (61 cm)
Plugged 3/8” (9.5 mm) MPX body
1 1/2” (38.1 mm)
Tri-clamp SterilEnz™
Sterile Hydrophobic Vent Filter
(0.2 micron PVDF - Pall Acro 50)
S.U.M. Docking Station
500 L Tank Standard BPC Configurations (CX5-14 Film)
Powder-Liquid Applications with Probe Ports and Thermo-well—Part number SH30751.02
For use with tubing welder and with CPC® Quick Connects
Back Face
Front Face
8. Powder Addition Port
Impeller
5. Unused
3-4. Probe Ports (2)
2. Addition Line
1. Recirculation/Sample Line
6. Thermo-well/Small Volume Sample
7. Bottom Drain
Line
Description
Tubing Set (ID X OD X Length)
End Treatment
1
Recirculation/Sample line
1/2” (12.7 mm) ID X 3/4” (19.1 mm) OD C-Flex X 66” (168 cm) splits to
1/2” (12.7 mm) ID X 3/4” (19.1 mm) OD C-Flex X 24” (61 cm) and
1/2” (12.7 mm) ID X 3/4” (19.1 mm) OD C-Flex X 12” (30 cm)
Capped 1/2” (12.7 mm) MPX insert
Plugged 1/2” (12.7 mm) MPX body
2
Addition Line
1/2” (12.7 mm) ID X 5/8” (15.9 mm) OD C-Flex X 60” (152 cm)
Plugged 1/2” (12.7 mm) MPX body
3-4
Probe Ports (2)
None
Pall™ Kleenpak™ Aseptic
Connector KPCHT Series (Female)
5
Unused
None
Plug
6
Thermo-well/
Small Volume Sample
Thermo-well Adapter for 1/8” (3.2 mm) Diameter
1/8” (3.2 mm) ID X 1/4” (6.4 mm) OD C-Flex X 12” (30 cm)
Luer and SmartSite® Valve Port
7
Bottom Drain
1/2” (12.7 mm) ID X 3/4” (19.1 mm) OD C-Flex X 48” (122 cm)
Plugged 1/2” (12.7 mm) MPX body
8
Powder Addition Port
3” (76 mm) Sanitary Fitting, Tri-clamp
Cap with gasket
500 L Tank Standard BPC Configurations (CX5-14 Film)
Liquid-Liquid Applications with Probe Ports and Thermo-well—Part number SH30754.02
For use with tubing welder and with CPC® Quick Connects
Back Face
Front Face
6. Recirculation/Sample Line
Impeller
3. Unused
3-4. Probe Ports (2)
7. Fill Line
6. Addition Line
4. Thermo-well/Small Volume Sample
8. Vent Filter
5. Bottom Drain
Line
Description
Tubing Set (ID X OD X Length)
End Treatment
1-2
Probe Ports (2)
None
Pall Kleenpak Aseptic
Connector KPCHT Series (Female)
3
Unused
None
Plug
4
Thermo-well/
Small Volume Sample
Thermo-well Adapter for 1/8” (3.2 mm) Diameter
1/8” (3.2 mm) ID X 1/4” (6.4 mm) OD C-Flex X 12” (30 cm)
Luer and SmartSite
Valve Port
5
Bottom Drain
1/2” (12.7 mm) ID X 3/4” (19.1 mm) OD C-Flex X 48” (122 cm)
Capped 1/2” (1.27 cm) MPX insert
6
Recirculation/Sample line
1/2” (12.7 mm) ID X 3/4” (19.1 mm) OD C-Flex X 66” (168 cm) splits to
1/2” (12.7 mm) ID X 3/4” (19.1 mm) OD C-Flex X 24” (61 cm) and
1/2” (12.7 mm) ID X 3/4” (19.1 mm) OD C-Flex X 12” (30 cm)
Capped 1/2” (12.7 mm) MPX insert
Plugged 1/2” (12.7 mm) MPX body
7
Fill Line
3/4” (19.1 mm) ID X 1” (25.4 mm) OD C-Flex X 60” (152 cm)
1 1/2” Tri-clamp SterilEnz
8
Vent Filter
1/4” (6.4 mm) ID X 1/2” (12.7 mm) OD C-Flex X 4” (10.2 cm)
Sterile Hydrophobic Vent Filter (0.2
micron PVDF - Pall Acro 50)
9
Addition Line
3/8” (9.5 mm) ID X 5/8” (15.9 mm) OD C-Flex X 24” (61 cm)
Plugged 3/8” (9.5 mm) MPX body
S.U.M. Docking Station
1000 L Tank Standard BPC Configurations (CX5-14 Film)
Powder-Liquid Applications—Part number SH30752.01
For use with tubing welder and with CPC® Quick Connects
Back Face
Front Face
Impeller
2. Addition Line
4. Powder Addition Port
1. Bottom Drain
3. Recirculation/Sample Line
Line
Description
Tubing Set (ID X OD X Length)
End Treatment
1
Bottom Drain
1/2” (12.7 mm) ID X 3/4” (19.1 mm) OD C-Flex® X 48” (122 cm)
Capped 1/2” (12.7 mm) MPX insert
2
Addition Line
1/2” (12.7 mm) ID X 3/4” (19.1 mm) OD C-Flex X 72” (183 cm)
Plugged 1/2” (12.7 mm) MPX body
3
Recirculation/Sample line
1/2” (12.7 mm) ID X 3/4” (19.1 mm) OD C-Flex X 78” (198 cm) splits to
1/2” (12.7 mm) ID X 3/4” (19.1 mm) OD C-Flex X 24” (61 cm) and
1/2” (12.7 mm) ID X 3/4” (19.1 mm) OD C-Flex X 12” (30 cm)
Capped 1/2” (12.7 mm) MPX insert
Plugged 1/2” (12.7 mm) MPX body
4
Powder Addition Port
3” (76 mm) Sanitary Fitting, Tri-clamp
Cap with gasket
1000 L Tank Standard BPC Configurations (CX5-14 Film)
Liquid-Liquid Applications—Part number SH30755.01
For use with tubing welder and with CPC® Quick Connects
Back Face
Front Face
4. Recirculation/Sample Line
Impeller
1. Bottom Drain
2. Fill Line
3. Vent Filter
5. Addition Line
Line
Description
Tubing Set (ID X OD X Length)
End Treatment
1
Bottom Drain
1/2” (12.7 mm) ID X 3/4” (19.1 mm) OD C-Flex X 48” (122 cm)
Capped 1/2” (12.7 mm) MPX insert
2
Fill Line
3/4” (19.1 mm) ID X 1” (25.4 mm) OD C-Flex X 72” (183 cm)
3
Vent Filter
1/4” (6.4 mm) ID X 1/2” (12.7 mm) OD C-Flex X 4” (10.2 cm)
4
Recirculation/Sample line
1/2” (12.7 mm) ID X 3/4” (19.1 mm) OD C-Flex X 78” (198 cm) splits to
1/2” (12.7 mm) ID X 3/4” (19.1 mm) OD C-Flex X 24” (61 cm) and
1/2” (12.7 mm) ID X 3/4” (19.1 mm) OD C-Flex X 12” (30 cm)
Capped 1/2” (12.7 mm) MPX insert
Plugged 1/2” (12.7 mm) MPX body
5
Addition Line
3/8” (9.5 mm) ID X 5/8” (15.9 mm) OD C-Flex X 24” (61 cm)
Plugged 3/8” (9.5 mm) MPX body
1 1/2” (38.1 mm) Tri-clamp
SterilEnz™
Sterile Hydrophobic Vent Filter
(0.2 micron PVDF - Pall Acro 50)
S.U.M. Docking Station
1000 L Tank Standard BPC Configurations (CX5-14 Film)
Powder-Liquid Applications with Probe Ports and Thermo-well—Part number SH30752.02
For use with tubing welder and with CPC® Quick Connects
Back Face
Front Face
Impeller
8. Powder Addition Port
3-4. Probe Ports (2)
5. Unused
2. Addition Line
1. Recirculation/Sample Line
6. Thermo-well/Small Volume Sample
Line
7. Bottom Drain
Description
Tubing Set (ID X OD X Length)
End Treatment
1
Recirculation/Sample Line
1/2” (12.7 mm) ID X 3/4” (19.1 mm) OD C-Flex X 78” (198 cm) splits to
1/2” (12.7 mm) ID X 3/4” (19.1 mm) OD C-Flex X 24” (61 cm) and
1/2” (12.7 mm) ID X 3/4” (19.1 mm) OD C-Flex X 12” (30 cm)
Capped 1/2” (12.7 mm) MPX insert
Plugged 1/2” (12.7 mm) MPX body
2
Addition Line
1/2” (12.7 mm) ID X 5/8” (15.9 mm) OD C-Flex X 48” (122 cm)
Plugged 1/2” (12.7 mm) MPX body
3-4
Probe Ports (2)
None
Pall™ Kleenpak™ Aseptic
Connector KPCHT Series (Female)
5
Unused
None
Plug
6
Thermo-well/
Small Volume Sample
Thermo-well Adapter for 1/8” (3.2 mm) Diameter
1/8” (3.2 mm) ID X 1/4” (6.4 mm) OD C-Flex X 12” (30 cm)
Luer and SmartSite® Valve Port
7
Bottom Drain
1/2” (12.7 mm) ID X 3/4” (19.1 mm) OD C-Flex X 48” (122 cm)
Plugged 1/2” (12.7 mm) MPX body
8
Powder Addition Port
3” (76 mm) Sanitary Fitting, Tri-clamp
Cap with gasket
1000 L Tank Standard BPC Configurations (CX5-14 Film)
Liquid-Liquid Applications with Probe Ports and Thermo-well—Part number SH30755.02
For use with tubing welder and with CPC® Quick Connects
Back Face
Front Face
6. Recirculation/Sample Line
Impeller
1-2. Probe Ports (2)
3. Unused
7. Fill Line
8. Vent Filter
9. Addition Line
4. Thermo-well/Small Volume Sample
5. Bottom Drain
Line
Description
Tubing Set (ID X OD X Length)
End Treatment
1-2
Probe Ports (2)
None
Pall Kleenpak Aseptic
Connector KPCHT Series (Female)
3
Unused
None
Plug
4
Thermo-well/
Small Volume Sample
Thermo-well Adapter for 1/8” (3.2 mm) Diameter
1/8” (3.2 mm) ID X 1/4” (6.4 mm) OD C-Flex X 12” (30 cm)
Luer and SmartSite
Valve Port
5
Bottom Drain
1/2” (12.7 mm) ID X 3/4” (19.1 mm) OD C-Flex X 48” (122 cm)
Capped 1/2” (1.27 cm) MPX insert
6
Recirculation/Sample line
1/2” (12.7 mm) ID X 3/4” (19.1 mm) OD C-Flex X 78” (198 cm) splits to
1/2” (12.7 mm) ID X 3/4” (19.1 mm) OD C-Flex X 24” (61 cm) and
1/2” (12.7 mm) ID X 3/4” (19.1 mm) OD C-Flex X 12” (30 cm)
Capped 1/2” (12.7 mm) MPX insert
Plugged 1/2” (12.7 mm) MPX body
7
Fill Line
3/4” (19.1 mm) ID X 1” (25.4 mm) OD C-Flex X 72” (183 cm)
1 1/2” Tri-clamp SterilEnz
8
Vent Filter
1/4” (6.4 mm) ID X 1/2” (12.7 mm) OD C-Flex X 4” (10.2 cm)
Sterile Hydrophobic Vent Filter
(0.2 micron PVDF - Pall Acro 50)
9
Addition Line
3/8” (9.5 mm) ID X 5/8” (15.9 mm) OD C-Flex X 24” (61 cm)
Plugged 3/8” (9.5 mm) MPX body
S.U.M. Docking Station
Ordering Information for Standard Products
Autoclave Tray and Probe Kits:
Probe
Handle
KPC
Connector
Probe Assembly
Autoclave Tray
Additional Information on Autoclave Tray:
• Fabricated from stainless steel
• Plastic carry handle for easy transport
right out of the autoclave
• Positions probes on 15% incline for
greater probe/membrane longevity
• Will restrain probe bellows from
collapsing during sterilization
• Probe holder accommodates two probes
Powdertainer Hanger:
Open-Top Mixing (OTM) use:
The Single-Use Mixer (S.U.M.) can be
equipped with a Thermo Scientific HyClone
Powdertainer Hanger if so desired. When
adding solids to liquids, this optional feature
enables utilization of the Powdertainer
powder feed system. The Powdertainer
hanger can be used to support full 1, 5 or 25
Kg Powdertainers and one collar positions
the 3” sanitary connectors for bag-to-bag
connection and solids transfer. The second
collar is provided to support the bearing hub
assembly when the tank is detached from the
docking station motor.
The S.U.M. can be equipped with a reusable
mixing hub accessory to enable use of open
top liners and single-use impeller sleeves.
The OTM option is a non-sterile easy-to-use
mixing option. The reusable hub is simply
placed into the motor’s bearing port receiver
and latched into place. The single-use
impeller sleeve attaches to the hub’s quick
connect end. With an open liner in position
and a drive shaft inserted the system is ready
to use.
Probe Holders (Clips):
Reusable
Mixing Hub
(with quick connect)
Bag Hanger
Standard probes can be used with docking
tanks by employing probe assemblies and
probe holders. The holders are retained in
position by the probe holder band above the
probe window on the tank hatch.
Adjustable
from
20” to 36”
Bottom Drain
Single-use Liner
Bag Support Collars
Single-use Impeller Sleeve
(with quick connect)
Accessories
Part Number
Description
SH30720.01
Probe Assembly (non-sterile)
SV20664.01
Heavy Duty Tubing Clamp
SV20664.02
Heavy Duty Tubing Clamp (4 pack of SV20664.01)
SV20664.03
Heavy Duty Tubing Clamp (10 pack of SV20664.01)
SV20750.01
S.U.M. thermo-well/sample port
Used for RTD calibration/validation.
SV20778.01
RTD
Pt100 Resistance Temperature Detector for use with S.U.M BPC
thermo-well/sampling port.
SV50177.01
Autoclave Tray for Probe Kits (stainless steel)
Support tray which allows the probes to be safely autoclaved.
SH30845.01
Sterile Sampling Manifold with luer lock (individual)
SH30845.02
Sterile Sampling Manifold with luer lock (10 pack)
Available for aseptic luer connection or welding onto the sample line
to take sample sets.
SV50187.05
Powdertainer Hanger with Dual Support Collars
To support Powdertainer and Bearing Port (for bag use in Docking Tanks)
SV50177.77
Reusable Mixing Hub
Required for use with single-use impeller sleeve
SH30762.01
200 L Liner for S.U.M. (gamma-sterilized),
Open Top Mixing
Bottom drain liner with 72” (182.9 cm) of ½” (12.7 mm) C-Flex tubing
and an MPX insert connector, CX3-9 Film
SH30762.02
500 L Liner for S.U.M. (gamma-sterilized),
Open Top Mixing
Bottom drain liner with 72” (182.9 cm) of ½” (12.7 mm) C-Flex tubing
and an MPX insert connector, CX3-9 Film
SH30762.03
1000 L Liner for S.U.M. (gamma-sterilized),
Open Top Mixing
Bottom drain liner with 72” (182.9 cm) of ½” (12.7 mm) C-Flex tubing
and an MPX insert connector, CX3-9 Film
SH30749.08
Single-use Impeller Sleeve for 200 L tanks
For Open-Top Mixing
SH30749.10
Single-use Impeller Sleeve for 500 and 1000 L tanks
For Open-Top Mixing
Used to package probes (not supplied) for sterilization and to
aseptically connect them to the S.U.M. BPC.
Reusable tubing clamp used on probe ports when connecting probe assembly.
S.U.M. Docking Station
Hardware Standard Products
Part Number
Description
SV50222.01
S.U.M. Docking Station System (120 VAC, Single Phase)
SV50222.02
S.U.M. Docking Station System (240 VAC, Single Phase)
SV50224.01
200 L S.U.M. Docking Station Tank, non-jacketed, feet
SV50224.02
200 L S.U.M. Docking Station Tank, non-jacketed, casters
SV50224.03
200 L S.U.M. Docking Station Tank, jacketed, feet
SV50224.04
200 L S.U.M. Docking Station Tank, jacketed, casters
SV50225.01
500 L S.U.M. Docking Station Tank, non-jacketed, feet
SV50225.02
500 L S.U.M. Docking Station Tank, non-jacketed, casters
SV50225.03
500 L S.U.M. Docking Station Tank, jacketed, feet
SV50225.04
500 L S.U.M. Docking Station Tank, jacketed, casters
SV50226.01
1000 L S.U.M. Docking Station Tank, non-jacketed, feet
SV50226.02
1000 L S.U.M. Docking Station Tank, non-jacketed, casters
SV50226.03
1000 L S.U.M. Docking Station Tank, jacketed, feet
SV50226.04
1000 L S.U.M. Docking Station Tank, jacketed, casters
Includes: 304 stainless steel cart with swivel casters, adjustable motor
mounting, motor, variable speed agitation speed controller, drive assembly,
shaft, steps and standard tool set.
Includes: 304 stainless steel cart with swivel casters, adjustable motor
mounting, motor, variable speed agitation speed controller, drive assembly,
shaft, steps and standard tool set.
Includes: 304 stainless steel outer support container with platform on feet.
Includes: 304 stainless steel outer support container with platform on swivel
casters.
Includes: 304 stainless steel outer support container with heating/cooling jacket
and platform on feet.
Includes: 304 stainless steel outer support container with heating/cooling jacket
and platform on swivel casters.
Includes: 304 stainless steel outer support container with platform on feet.
Includes: 304 stainless steel outer support container with platform on swivel
casters.
Includes: 304 stainless steel outer support container with heating/cooling jacket
and platform on feet.
Includes: 304 stainless steel outer support container with heating/cooling jacket
and platform on swivel casters.
Includes: 304 stainless steel outer support container with platform on feet.
Includes: 304 stainless steel outer support container with platform on swivel
casters.
Includes: 304 stainless steel outer support container with heating/cooling jacket
and platform on feet.
Includes: 304 stainless steel outer support container with heating/cooling jacket
and platform on swivel casters.
Custom BPC Options:
Tubing Type
C-Flex (clear, white, and ADCF), Silicone, PVC, PharMed® or PharmaPure®
Tubing Size
Specific Lengths of 1/8 to 1” (3.18 to 25.4 mm) ID — depending on type of tubing chosen
Connectors
Luer — 1/8 to 1/2” (3.18 to 12.7 mm) ID
CPC — 1/4 to 3/4” (6.35 to 19 mm) ID
SIP Connector —1/4 to 3/4” (6.35 to 19 mm) ID
Tri-clamp — 1/8” to 1” (3.18 to 25.4mm) ID
Mini-Tri-clamp — 1/8 to 1/2” (6.35 to 12.7 mm) ID
Aseptic Connection Device — All available sizes of Pall Kleenpak and BioQuate™ DAC
Others
Needle-free sample port (SmartSite or Clave®)
Filter Capsule (Millipore®, Pall, Sartorius™, Domnick Hunter®, Meissner®, Other)
Please Note: Not all options are available for all ports. Customization of port type and location, chamber dimensions or mixing assembly is not available.
For additional information, please refer to the Selection Guides in the 2008 BPC Catalog.
Presentation (as dry BPC systems):
Outer Packaging
Supplied ‘flat-packed’
Two polyethylene outer layers
Label
Description
Product code
Lot number
Expiry date on outer packaging and shipping container
Sterilization
Irradiation (25 to 38 kGy) inside outer packaging
Shipping Container
Durable cardboard carton
Documentation
Certificate of Analysis provided with each lot for each delivery
S.U.M. Docking Station
Hardware Features
S.U.M. Docking Station System
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
3
4
Docking Station – including the following features:
Platform with large diameter wheels
Mixer motor
Motor adjustment mechanism
Drive shafts
Control panel
Tool kit
Docking lock lever
Docking lock mechanism (grapple)
Step
ID plate
2
6
5
10
8
9
1
7
S.U.M. Docking Station System
Docking Station Tank (Jacketed version shown)
– including the following features:
1.
2.
3.
4.
5.
6.
7.
8.
7
Platform with large diameter wheels
(available with feet)
One of two hatches for BPC loading and
monitoring probe insertion
Bottom cut outs for BPC alignment
Sight window
Inlet/outlet for temperature control fluid
(jacketed version only)
Grapple receiver
Support arm for Powdertainer (optional
extra)
Handle to assist with tank mobility
8
4
2
5
6
3
1
S.U.M. Docking Station
Operation of S.U.M. Docking Station
The design features which enable the S.U.M. Docking
Station hardware to link up with the 200, 500 and
1000 L tanks are as follows:
1.
The mixer motor mounting is adjustable
vertically and horizontally as shown to
the right. In addition the motor angle is
adjustable. The adjustments are simple to
make with indicators showing the correct
setting for each size of tank and secure
locking mechanism. Adjustments are made
before the S.U.M. BPC is loaded into the tank
2.
The S.U.M. Docking Station and tanks are
mounted on lockable casters to enable
mobility and accurate positioning
3.
Locking jaws (or grapple) which open and
close horizontally are controlled by a lever on
the side of the hardware
4.
Each S.U.M. docking station tank has a
grapple receiver (A) onto which the jaws lock
to secure the tank in the correct orientation
to the docking station.
A
5.
Once the tank is locked in place (as shown
from underneath) the S.U.M. BPC can be
loaded and used in the same way as with
other S.U.M. systems.
S.U.M. Docking Station
1
73 2 "
73.486
23
77 32 "
77.715
Docking Station:
Front View
(Maximum height – adjusted for 1000 L tank)
23
80 32 "
80.715
0"
0"
0"
M
23
80 32 "
80.715
Hardware Dimensions (units in inches [mm])
500L & 1000L
W/ PLATFORM
200L W/O PLATFORM
0"
L & 1000L
PLATFORM
500L & 1000L
W/O PLATFORM
Overhead View
5
46 8 "
46.637
60°
60°
60°
11
2 32 "
2.348
23
40 32 "
40.710
11
2 32 "
2.348
34"
34.000
PROJECTION
BREAK ALL SHARP EDGES
AND REMOVE ALL BURRS
TITLE
25.500
25 1/2"
Hardware Dimensions (units in inches [mm])
200 L Non-Jacketed S.U.M.:
33.375
33 3/8"
28.500
28 1/2"
25.500
25 1/2"
83.231
83 7/32"
3D MODEL: 19343
40.052
40 1/16"
20.250
20 1/4"
58.244
58 1/4"
33.312
33 5/16"
28.500
28 1/2"
13.630
13 5/8"
25.500
25 1/2"
83.231
83 7/32"
200 L Jacketed S.U.M.:
40.052
40 1/16"
33.312
33 5/16"
28.500
28 1/2"
20.250
20 1/4"
58.244
58 1/4"
20.199
20 3/16"
13.630
13 5/8"
25.500
25 1/2"
82.246
82 1/4"
TITLE
PROJECTION
BREAK ALL SHARP EDGES
AND REMOVE ALL BURRS
200L DOCKING SUM TANK W/
CASTERS-NJ
UNLESS OTHERWISE SPECIFIED DIMENSIONS ARE INCHES AND ARE IN
ACCORDANCE WITH ANSI14.5M-1982 SPECIFICATION
NUMBER
TOLERANCES ARE:
DECIMALS: MACHINED 125
SURFACES:
X = .1
XX = .01
XXX = .005 ANGLES: 1
SCALE
1:1
SHEET
1722 S. HyClone Rd.
Logan, UT 84321
THIS DRAWING CONTAINS INFORMATION
PROPRIETARY TO THERMO SCIENTIFIC. ANY
REPRODUCTION, DISCLOSURE OR USE OF THIS
DRAWING IS EXPRESSLY PROHIBITED EXCEPT AS
THERMO SCIENTIFIC MAY AUTHORIZE IN WRITING.
SV50224.02
FINISH/HEAT TREAT
MATERIAL
N/A
ENGINEER
M. Goodwin
BY
2 OF 3 RELEASED
--
N/A
DATE
4/21/2009
EFFECTIVE DATE
--
SIZE
D
58.244
58 1/4"
DRAWING NUMBER
17601
40.067
40 1/16"
20.253
20 1/4"
13.630
13 5/8"
82.246
82 1/4"
58.244
58 1/4"
40.067
40 1/16"
20.253
20 1/4"
13.630
13 5/8"
16.753
16 3/4"
23.108
23 3/32"
16.753
16 3/4"
REV.
-
Hardware Dimensions (units in inches [mm])
35.219
35 7/32"
500 L Non-Jacketed S.U.M.:
35.790
35 25/32"
35.219
35 7/32"
90.523
90 17/32"
38.891
38 29/32"
65.101
65 3/32"
39.910
39 29/32"
35.110
35 1/8"
20.538
20 17/32"
13.630
13 5/8"
90.523
90 17/32"
500 L Jacketed S.U.M.:
65.101
65 3/32"
39.910
39 29/32"
38.891
38 29/32"
20.538
20 17/32"
20.538
20 17/32"
13.630
13 5/8"
35.110
35 1/8"
500L DOCKING SUM TANK W/
CASTERS-NJ
BREAK ALL SHARP EDGES
AND REMOVE ALL BURRS
UNLESS OTHERWISE SPECIFIED DIMENSIONS ARE INCHES AND ARE IN
ACCORDANCE WITH ANSI14.5M-1982 SPECIFICATION
NUMBER
TOLERANCES ARE:
DECIMALS: MACHINED 125
SURFACES:
X = .1
XX = .01
XXX = .005 ANGLES: 1
SCALE
1:1
89.662
89 21/32"
TITLE
PROJECTION
SHEET
2 OF 3
1722 S. HyClone Rd.
Logan, UT 84321
FINISH/HEAT TREAT
N/A
N/A
ENGINEER
DATE
RELEASED BY
EFFECTIVE DATE
M. Goodwin
--
THIS DRAWING CONTAINS INFORMATION
PROPRIETARY TO THERMO SCIENTIFIC. ANY
REPRODUCTION, DISCLOSURE OR USE OF THIS
DRAWING IS EXPRESSLY PROHIBITED EXCEPT AS
THERMO SCIENTIFIC MAY AUTHORIZE IN WRITING.
SV50225.02
MATERIAL
4/21/2009
--
SIZE
D
40.070
40 1/16"
23.393
23 13/32"
13.630
13 5/8"
17.038
17 1/32"
89.662
89 21/32"
65.101
17603
65 3/32"
DRAWING NUMBER
REV.
-
47 7/32"
Data Sheet
Hardware Dimensions (units in inches [mm])
45.026
45 1/32"
1000 L Non-Jacketed S.U.M.:
47.234
47 7/32"
50.395
50 13/32"
45.026
45 1/32"
45.130
45 1/8"
89.125
89 1/8"
63.457
63 15/32"
40.062
40 1/16"
20.538
20 17/32"
13.630
13 5/8"
89.125
89 1/8"
PROJECT
BREAK
AND R
63.457
63 15/32"
1000 L Jacketed S.U.M.:
UNLESS OT
SIONS A
ACCORDA
TOLERAN
DECIMAL
X = .1
XX = .01
XXX = .
SCALE
40.062
40 1/16"
20.538
20 17/32"
50.395
50 13/32"
20.538
20 17/32"
13.630
13 5/8"
87.712
87 23/32"
45.130
45 1/8"
TITLE
PROJECTION
BREAK ALL SHARP EDGES
AND REMOVE ALL BURRS
1000L DOCKING SUM TANK W/
CASTERS-NJ
UNLESS OTHERWISE SPECIFIED DIMENSIONS ARE INCHES AND ARE IN
ACCORDANCE WITH ANSI14.5M-1982 SPECIFICATION
NUMBER
63.457
63 15/32"
TOLERANCES ARE:
DECIMALS: MACHINED 125
SURFACES:
X = .1
XX = .01
XXX = .005 ANGLES: 1
SCALE
1:1
SHEET
1722 S. HyClone Rd.
Logan, UT 84321
THIS DRAWING CONTAINS INFORMATION
PROPRIETARY TO THERMO SCIENTIFIC. ANY
REPRODUCTION, DISCLOSURE OR USE OF THIS
DRAWING IS EXPRESSLY PROHIBITED EXCEPT AS
THERMO SCIENTIFIC MAY AUTHORIZE IN WRITING.
SV50226.02
FINISH/HEAT TREAT
MATERIAL
N/A
ENGINEER
M. Goodwin
BY
2 OF 3 RELEASED
--
N/A
DATE
4/21/2009
EFFECTIVE DATE
--
SIZE
D
DRAWING NUMBER
17605
REV.
-
40.061
40 1/16"
3.393
13/32"
13.630
13 5/8"
©2009 Thermo Fisher Scientific Inc. All rights reserved. CPC® is a registered trademark of Colder Product Company. SterilEnz™ is a registered trademark of PAW
BioScience Products. SmartSite® is a trademark of ALARIS Medical Systems, Inc. C-Flex® and PharMed® are registered trademarks of Saint-Gobain. PharmaPure®
is a registered trademark of ParmaPure Tubing. Clave® is a registered trademark of ICU Medical. Pall™ and Kleenpak™ are registered trademark of Pall Corporation.
BioQuate™ is a registered trademark of BioQuate Incorporated. Millipore® is a registered trademark of Millipore Corporation. Meissner® is a registered trademark of
Meissner Filtration Products. Domnick Hunter® is a registered trademark of Parker Hannifin Ltd. Sartorius™ is a registered trademark of Sartorius Stedim Biotech. All
other trademarks, registered trademarks, or logos are properties of their respective owners. Specifications, terms and pricing are subject to change. Not all products
are available in all countries. Please consult your local sales representative for details.
HyClone - Media•BPC•Sera
87.712925 West 1800 South
Logan, UT
87 23/32"
84321
63.457
63 15/32"
In Americas/Asia
435-792-8000
435-792-8001 fax
In Europe
+32 53 85 75 59
+32 53 85 74 31 fax
www.thermo.com/hyclone
17.038
17 1/32"
HyClone Laboratories, Inc. Logan, UT USA is ISO Certified.
1
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project

Download PDF

advertising