Edge Single Parameter Meters
Single Parameter Meters
Lightweight and versatile pH, EC and DO meters that can be
used in portable, wall-mount and benchtop configurations
Digital electrodes
edge® performs measurements through
its unique digital electrodes. These digital
electrodes are auto-recognized, providing
sensor type, calibration data and a serial
number when connected to edge® by
an easy to plug-in 3.5 mm connector.
edge®pH features Hanna’s exclusive
CAL Check™ to warn you if the electrode
in use is not clean or if your buffers are
contaminated during calibration. We have
added Sensor Check™ for pH sensors
with a matching pin. Our Sensor Check™
feature warns you if the pH bulb is cracked
and/or the junction of the electrode
is compromised. In addition to these
features, edge®pH also measures ORP
with edge® compatible ORP probes.
3
models—pH, EC
and DO
0 0.5 8.8 8 5.5 2
footprint
inch thick
oz. weight
(12.7 mm)
(250 g)
hours battery
life
inch display
(14 cm)
USB ports
dedicated
Innovation dedicated
to a single parameter
edge® dedicated meters are designed to
measure a single parameter. This dedicated
series is thin and lightweight, measuring
just 1⁄2” (12 mm) thick and weighing
less than 9 ounces (250 g). Each edge®
dedicated meter has an incredibly wide
viewing angle, 5.5” (14 cm) LCD and a
sensitive capacitive touch keypad.
edge®
Hybrid meters that can be used in portable,
wall-mount and benchtop configurations
The versatile design of edge® enables it to be used as a portable, wall-mount or benchtop meter. edge® simplifies measurement,
configuration, calibration, diagnostics, logging and transferring data directly to a computer or USB drive.
• Portable field unit
· edge® is ideal for field use due to its light
weight, large screen and thin design. It can be
easily slipped into a backpack or messenger
bag. Up to 8 hours of battery life when used
as a portable device
• Wall mount cradle
· The included wall mount cradle makes it easy
to conserve space on the benchtop and can
charge edge® with the AC adapter. Ideal for
continuous monitoring applications
| edge® | single parameter meters
• Electrode holder with built-in cradle
· The included electrode holder features a
swivel,adjustable arm with a built-in cradle
to hold edge® securely in place at the
optimum viewing angle
Features
edge® technical features
edge® design features
• Two USB ports
edge® includes one standard USB for exporting
data to a flash drive. edge® also includes one
micro USB port for exporting files to your
computer as well as charging edge® when the
cradle is not available.
• Capacitive touch keypad
edge® features a capacitive touch keypad
that gives a distinctive, modern look. Since the
keypad is part of the screen, your buttons can
never get clogged with sample residue.
• Clear, full text readout
edge® features clear, full text guides displayed
on the bottom of the screen. There is no
need to decipher scrambled abbreviations or
symbols; these helpful messages guide you
through every process quickly and easily.
• Easy to read LCD
edge® features a 5.5” (14 cm) LCD display
that you can clearly view from over 5 m (16.4’).
The large display, with its wide 150° viewing
angle, provides one of the easiest to read
LCDs in the industry.
• Data logging
edge® allows you to store up to 1000 log
records of data. Data sets include readings,
GLP data, date and time.
• GLP
Data of the last calibration you perform
is stored in the sensor including the date, time
and buffer/standards. When a compatible
sensor is connected to edge®, GLP data is
automatically transferred.
5.5"
• Zero footprint
Using the wall mount cradle (included),
edge® can be placed on a wall, leaving zero
footprint on the benchtop space. The
cradle has a built-in connector to power
edge® and charge its batteries.
• 3.5 mm probe input
Plugging an electrode in has never
been simpler; no alignments or broken
pins, simply connect the 3.5 mm plug
and begin. Digital electrodes are
automatically recognized.
• Basic mode (HI2002 and HI2003 only)
You can use edge®pH and edge®EC in Basic
Mode—ideal for routine measurements by
displaying a simplified screen and features.
• CAL Check™ (HI2002 only)
edge®pH features Hanna’s exclusive
CAL Check™ technology to warn you
if the electrode bulb is not clean
or if the buffers are contaminated
during calibration.
• Sleek design
Incredibly thin and lightweight,
edge® measures just 1/2”
(12 mm) thick and weighs just
8.8 ounces (250 g).
Model Specific Features
Accepts edge®pH compatible
pH and ORP probes
• Resolution selectable from 0.01 and 0.001 pH
• Range -2.000 to 16.000 pH
• Accuracy ±0.002 pH for 0.001 pH resolution;
±0.01 for 0.01 resolution
• Data logging
· Manual log-on-demand
· Manual log-on-stability
· Interval logging
• Temperature readout (°C or °F)
• Automatic Temperature Compensation (ATC)
• CAL Check ™ Indicators:
· Probe condition
· Response time
· Check buffer
· Clean electrode
• Sensor Check™ Indicators:
· Broken electrode
· Clogged junction
• GLP data
· Records date, time, offset, slope and buffers
used during calibration
• Five-point calibration
· A choice of seven pre-programmed buffers
plus two selectable custom buffers
• Calibration tag on screen
· Identifies buffers used for current calibration
• Calibration expiration warning
Accepts edge®EC compatible
conductivity probe
• Digital four-ring conductivity probe
· Covers all ranges from 0.00 μS/cm to
500 mS/cm (absolute EC)
• Accuracy
· ± 1% of the reading (±0.05 μS/cm
or 1 digit, whichever is greater)
• Calibration
· Offset (0 μS/cm) and cell factor calibration
· Choice of 5 standards (auto-recognition)
• Data logging
· Manual log-on-demand
· Manual log-on-stability
· Interval logging
• Auto-ranging or manual range selection
• EC, TDS and salinity reading modes
• Temperature compensation
· Automatic
· NoTC (absolute)
• GLP data
· Records date, time, offset and cell
factor
· Data of the last performed calibration
is stored in the probe: date, time,
cell constant, temperature coefficient,
reference temperature and battery
status. When the probe is connected
to edge®EC , GLP data is automatically
transferred
• Adjustable EC to TDS conversion factor
• Adjustable temperature correction
coefficient
• Seawater salinity units
· % NaCl
· PSU
· g/L
Accepts edge®DO compatible
dissolved oxygen probe
• Clark type digital polarographic probe
with easy-to-replace membrane cap
· Covers all ranges from
0.00 to 45.00 mg/L (ppm);
0.0 to 300% saturation
• Accuracy ±1.5% full scale
• One or two-point calibration (HI7040),
0% (solution) and 100% (air)
• Data logging
· Manual log-on-demand
· Manual log-on-stability
· Interval logging
• Automatic Temperature Compensation
from 0 to 50 °C
• GLP data
· Records date, time, calibration
standards, altitude value and
salinity value
• Altitude compensation from -500 to
4000 meters (-1640 to 13,123’)
• Salinity compensation from 0 to 40g/L
Specifications
Specifications
Range*
-2.00 to 16.00 pH; -2.000 to 16.000 pH†
Resolution
0.01 pH; 0.001 pH†
Accuracy (@25°C/77°F)
±0.01 pH; ±0.002 pH†
Calibration
automatic, up to three points (five points†) calibration, 5 standard (7 standard†) buffers available
(1.68†, 4.01 or 3.00, 6.86, 7.01, 9.18, 10.01, 12.45†) and two custom buffers†
Temperature Compensation*
automatic, -5.0 to 100.0°C (23.0 to 212.0°F) (using integral temperature sensor)
Electrode Diagnostics
standard mode: probe condition, response time and out of calibration range
pH
edge®pH
mV pH
Range
±1000 mV
Resolution
0.1 mV
Accuracy (@25°C/77°F)
±0.2 mV
Range
±2000 mV
Resolution
0.1 mV
ORP
Additional
Specifications
EC
edge®EC
TDS
Salinity†
Additional
Specifications
DO
edge®DO
Additional
Specifications
Temperature
All Models
Additional
Specifications
* temperature limits will be reduced to actual probe limits
** with temperature compensation function disabled
† standard mode only
Accuracy (@25°C/77°F)
±0.2 mV (±999.9 mV); ±1 mV (±2000 mV)
Calibration
one-point calibration
Probe
HI11310 digital glass body pH electrode with 3.5 mm (1/8”) connector and 1 m (3.3’) cable
Logging
up to 1000† (400 for basic mode) records organized in: manual log-on-demand (max. 200 logs),
manual log-on-stability (max. 200 logs), interval logging† (max. 600 samples; 100 lots)
Range
0.00 to 29.99 μS/cm; 30.0 to 299.9 μS/cm; 300 to 2999 μS/cm; 3.00 to 29.99 mS/cm; 30.0 to 200.0 mS/cm;
up to 500.0 mS/cm absolute EC**
Resolution
0.01 μS/cm; 0.1 μS/cm; 1 μS/cm; 0.01 mS/cm; 0.1 mS/cm
Accuracy (@25°C/77°F)
±1% of reading (±0.5 μS or 1 digit, whichever is greater)
Calibration
single cell factor calibration; six standards available: 84 μS/cm, 1413 μS/cm, 5.00 mS/cm, 12.88 mS/cm,
80.0 mS/cm, 118.8 mS/cm, one point offset: 0.00 μS/cm
Temperature Coefficient
0.00 to 6.00%/°C (for EC and TDS only), default value is 1.90%/°C
Range
0.00 to 14.99 mg/L (ppm); 15.0 to 149.9 mg/L (ppm); 150 to 1499 mg/L (ppm); 1.50 to 14.99 g/L; 15.0 to 100.0 g/L;
up to 400.0 g/L absolute TDS using 0.80 conversion factor**
Resolution
0.01 mg/L (ppm); 0.1 mg/L (ppm); 1 (ppm); 0.01 g/L; 0.1 g/L
Accuracy (@25°C/77°F)
±1% of reading (±0.03 ppm or 1 digit, whichever is greater)
Calibration
through EC calibration
TDS Factor
0.40 to 0.80 (default value is 0.50)
Range
0.0 to 400.0 % NaCl; 2.00 to 42.00 PSU; 0.0 to 80.0 g/L
Resolution
0.1 % NaCl; 0.01 PSU; 0.01 g/L
Accuracy (@25°C/77°F)
±1% of reading
Calibration
PSU and g/L through EC calibration; % NaCl — one-point with HI7037 sea water standard
Probe
HI763100 digital four-ring conductivity probe with 3.5 mm (1/8”) connector and 1 m (3.3’) cable
Logging
up to 1000† (400 for basic mode) records organized in: manual log-on-demand (max. 200 logs),
manual log-on-stability (max. 200 logs), interval logging† (max. 600 samples; 100 lots)
Range
0.00 to 45.00 ppm (mg/L); 0.0 to 300.0 % saturation
Resolution
0.01 ppm (mg/L); 0.1 % saturation
Accuracy
± 1.5% of reading ±1 digit
Calibration
one or two-point at 0% (HI7040 solution) and 100% (in air)
Temperature Compensation
ATC (0 to 50°C; 32.0 to 122.0°F)*
Salinity Compensation
0 to 40 g/L (with 1 g/L resolution)
Altitude Compensation
-500 to 4000 m (-1640 to 13120') (with 100 m (328') resolution)
Probe
HI764080 digital dissolved oxygen electrode with 3.5 mm (1/8”) connector and 1 m (3.3’) cable (included)
Logging
up to 1000 records organized in: manual log-on-demand (max. 200 logs), manual log-on-stability (max. 200 logs),
interval logging (max. 600 samples; 100 lots)
Range*
-20.0 to 120.0°C; -4.0 to 248.0°F
Resolution
0.1°C; 0.1°F
Accuracy
±0.5°C; ±0.9°F
Connectivity
1 USB port for storage; 1 micro USB port for charging and PC connectivity
Environment
0 to 50°C (32 to 122°F); RH max 95% non-condensing
Power Supply
5 VDC adapter (included)
Dimensions
202 x 140 x 12 mm (7.9” x 5.5” x 0.5”)
Weight
250 g (8.82 oz.)
Ordering Information
Digital Electrodes for edge®
All edge® single parameter meters
are supplied with:
pH electrodes
(for HI2002 only)
Sensor Check™
HI11310
HI11311
Single ceramic, double junction,
glass body, refillable pH electrode with
temperature sensor
Single ceramic, double junction,
glass body, refillable pH electrode with
temperature sensor and matching pin
Recommended for laboratory and
general purpose
Recommended for laboratory and
general purpose
Sensor Check™
HI12300
HI12301
Single ceramic, double junction, gel filled, PEI
body, pH electrode with temperature sensor
Single ceramic, double junction, gel filled, PEI
body, pH electrode with temperature sensor
and matching pin
Recommended for field applications
Recommended for field applications
benchtop docking
station with
electrode holder
wall-mount cradle
USB cable
5 VDC power
adapter
edge® and
electrode quality
certificates
instruction
manual
In addition to these components, the following meter-specific
items are also included:
edge®pH: HI2002-01 (115V) and HI2002-02 (230V) also includes:
HI11310
glass body,
refillable pH
electrode
4 sachets of
pH 7 buffer
solutions
2 sachets of
pH 4 buffer
solutions
2 sachets of
pH 10 buffer
solutions
2 sachets
of electrode
cleaning
solutions
HI10530
HI10430
Triple ceramic, double junction, glass body,
refillable pH electrode with conical tip and
temperature sensor
Triple ceramic, double junction, glass body,
refillable pH electrode with temperature
sensor
Recommended for fats, creams, soil and low
conductivity samples
Recommended for low conductivity samples
HI10480
FC2320
PTFE reference, double junction, Clogging
Prevention System (CPS), glass body pH
electrode with temperature sensor
Open viscolene reference electrolyte, double
junction, PVDF body pH electrode with conical
tip and temperature sensor
Recommended for wine analysis and
solutions with a high concentration of
suspended solids
Recommended for meat applications with
use of optional FC098 20 mm (0.8’’) or FC099
35 mm (1.4’’) stainless steel blade
FC2100
FC2020
Open viscolene reference electrolyte, double
junction, glass body pH electrode with conical
tip and temperature sensor
Open viscolene reference electrolyte, double
junction, PVDF body pH electrode with conical
tip and temperature sensor
Recommended for dairy analysis including milk
Recommended for dairy analysis including
cheese, yogurt, and other semi-solids
edge®EC: HI2003-01 (115V) and HI2003-02 (230V) also includes:
ORP probes
(for HI2002 only)
HI763100
conductivity
probe
4 sachets of
1413 μS/cm
conductivity
standard
2 sachets of
12880 μS/cm
conductivity
standard
2 sachets of
5000 μS/cm
conductivity
standard
2 sachets of
electrode rinse
solution
edge®DO: HI2004-01 (115V) and HI2004-02 (230V) also includes:
HI36180
HI36200
Single ceramic, double junction, glass body,
refillable ORP probe with temperature sensor
Single ceramic, single junction, gel filed, PEI
body, ORP probe with temperature sensor
Recommended for laboratory and
general purpose
Recommended for field applications
Conductivity probe
Dissolved oxygen
electrode
(for HI2003 only)
edgededicated 1.0 02/15 PRINTED IN USA
(for HI2004 only)
HI764080
dissolved oxygen
electrode
HI7041S
refill electrolyte
solution
2 DO membrane
caps
HannaNorden AB
Energigatan 15B
434 37 KUNGSBACKA
SWEDEN
2 DO membrane
cap o-rings
HI763100
HI764080
Conductivity probe with temperature sensor
Dissolved oxygen electrode with
temperature sensor
Recommended for general purpose
Tel:
+46 (0)300 404018
e-mail: info@hannanorden.com
www.hannanorden.com
Recommended for general purpose
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project

Download PDF

advertising