insulation resistance meter
nr 214551 QM/UM
ISO 9001
ISO 14001
nr 214551 OH PL
PN-N-18001
Index: WMGBMIC2
MIC-2
INSULATION
RESISTANCE
METER
Digital insulation meter MIC-2 automatically select insulation measurements range at 250 and 500V test voltages. This
insulation meter is suited for insulation resistance measurements of electric installations, motors and other electric
devices. MIC-2 can also be used to measure voltage up to 600V AC and resistance up to 1999Ω. Instrument is easy to use,
professional tool in small and incredibly ergonomic case.
Standard accessories of the meter MIC-2:
- „crocodile” clip K01; black
- pin probe with banana connector - black
- batteries 1,5V (AAA) (2 pcs.)
WAKROBL20K01
WASONBLOGB1
- calibration certificate issued by SONEL S.A.
- operating manual
Optional accessories of the meter MIC-2:
Sonel S.A.
ul. Wokulskiego 11
58-100 Świdnica, PL
tel. +48 74 85 83 860
fax +48 74 85 83 809
export@sonel.pl
www.sonel.pl
- AGT-16P (triple phase socket adapter)
WAADAAGT16P
- carrying case S2
- AGT-32P (triple phase socket adapter)
WAADAAGT32P
- calibration certificate issued
- AGT-63P (triple phase socket adapter)
WAADAAGT63P
by calibration laboratory
WAFUTS2
LSWPLMIC2
nr 214551 QM/UM
ISO 9001
ISO 14001
nr 214551 OH PL
PN-N-18001
MIC-2
 Insulation resistance measurement:
- the measurement voltage is selected between the two values 250V or 500V,
- on-line insulation resistance measurement,
- automatic discharge of tested object’s capacitance after measurement
of insulation resistance,
- acoustic determination of 5 seconds time intervals that facilitates to take time
characteristics during insulation resistance measurement,
- protection against live objects.
 Low voltage resistance measurement in range 0...2000Ω:
Insulation resistance measurement:
UISO=250V
Measurement range: IEC 61557-2: 250kΩ...1000MΩ
Range
Resolution
Accuracy
not specified
1...249kΩ
1kΩ
250...1999kΩ
1kΩ
2,00...19.99MΩ
0.01MΩ
20,0...199.9MΩ
0.1MΩ
200...1000MΩ
1MΩ
±(3% m.v. + 8 digits)
- current measurement <10mA with resolution 0.1Ω,
- quick acoustic signalling (beeper) for circuit resistance below 10Ω.
 AC/DC voltage measurement in range 0...600V:
- automatic detection AC/DC voltage.
 Autoranging.
 Suitable for 50Hz and 60Hz network.
UISO=500V
Measurement range: IEC 61557-2: 500kΩ...1999MΩ
Range
Resolution
Accuracy
1...499kΩ
1kΩ
not specified
 Backlit display.
500...1999kΩ
1kΩ
2,00...19.99MΩ
0.01MΩ
 „AUTO-OFF” function.
20,0...199.9MΩ
0.1MΩ
 Meter meets the requirements of the standard EN 61557.
200...1999MΩ
1MΩ
Electric security:
- type of insulation
- measurement category
- protection class acc. to EN 60529
±(3% m.v. + 8 digits)
Resistance measurement:
double, according to EN 61010-1 and IEC 61557
CAT IV 600V acc. to EN 61010-1
IP65
Other technical data:
- power supply
batteries LR03 (AAA) (2 pcs.) or rechargeable battery
- measurement current RISO
1,2mA ± 0,2mA
- max interference voltage at which a measurement is realised RISO
20V
- number of RISO measurements
> 500
- dimensions
approx. 240 x 60 x 30 mm
- weight (including batteries)
approx. 0,3kg
- auto power OFF
5 min.
- electromagnetic compatibility
according to EN 61000-6-3 and EN 61000-6-2
- quality standard
ISO 9001
The group of Insulation Resistance Meters conforms the following directives or
standards according to CE requirements.
Directive:
Low Voltage Directive (LVD) 2005/95/WE.
Standards:
EN 61010-1:2001 Safety requirements for electrical equipment for
measurement, control and laboratory use. General requirements.
EN 61010-031:2002/A1:2008: Safety requirements for electrical
equipment for measurement, control and laboratory use. Safety
requirements for hand-held probe assemblies for electrical
measurement and test.
EN 61557-2:2007 Electrical safety in low-voltage distribution systems
up to 1000 V AC and 1500 V DC - Equipment for testing, measuring,
or monitoring of protective measures Part 2: Insulation resistance.
This part of IEC 61557 specifies the requirements applicable to
equipment for measuring the insulation resistance of equipment and
installations in the de-energized state.
Directive:
Electromagnetic compatibility (EMC): 2004/108/WE.
Standards:
EN 61325-1:2006 Electrical equipment for measurement, control and
laboratory use - EMC requirements.
Range
Resolution
0,0...199.9Ω
0.1Ω
200...1999Ω
1Ω
Accuracy
±(4% m.v. + 3 digits)
continuous audible signal for R < 10Ω
measurement current (with open terminals for Ubat > 3.0V): < 10mA
max voltage on open terminals: 4...24V
max interference voltage for which measurement is still possible: +7V/-1V DC, 5V AC
AC/DC voltage measurement
Range
Resolution
Accuracy
0...600V
1V
±(3% m.v. + 2 digits)
„m.v.” - measured value.
Rated operational conditions:
- operating temperature
- reference temperature
- storage temperature
0...+40°C
23 ± 2°C
-20...+70°C
v.3